EmbryoCellect. RHS Scanning and Analysis Instructions. for. Genepix Pro Software

Size: px
Start display at page:

Download "EmbryoCellect. RHS Scanning and Analysis Instructions. for. Genepix Pro Software"

Transcription

1 EmbryoCellect RHS Scanning and Analysis Instructions for Genepix Pro Software EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October

2 Copyright Reproductive Health Science Limited West Thebarton Road Thebarton, South Australia, 5031 EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October 2015 Reproductive Health Science Limited (RHS) makes no representations and gives no warranties of any kind in relation to the contents of this document and all warranties, conditions and other terms implied by statute or common law are, to the fullest extent permitted by law, hereby excluded. In particular, RHS assumes no responsibility for any errors or omissions that may appear in this document and makes no commitment to update or keep current the information contained in this document. RHS retains the right to make changes to this document (including any specifications contained herein) at any time without notice. This document is confidential. No part of this document may be modified, copied, reproduced, republished, published, transmitted or distributed in any form or by any means without the prior written consent of RHS. The contents of this document are to be used solely for the purpose for which they are provided by RHS and for no other purpose. All content, text, graphics and all other materials contained in this document are owned by RHS, and all proprietary and intellectual property (names, logos and trademarks) wherever arising in relation to this document, rest in RHS (or its licensors) and all such rights are reserved. EmbryoCellect is a Research Use Only product and is not to be used for diagnostic procedures EmbryoCellect is for Research Use Only and should not be used in diagnostic procedures. You are responsible for ensuring that you accurately follow the protocols provided in this Genepix Pro Scanning and Analysis Technical Data Sheet (TDS) and analysing and interpreting the results you obtain. RHS does not guarantee any results obtained. EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October

3 Contents 1. Genepix Pro Software Compatibility General Instructions Slide format Scanner requirements Prescan Scanning Scan parameters Saving TIFF images Scanning Instructions Check the settings Perform and save a Preview Scan of the slide Scan each array individually Array quality control Quantifying the Data Analysing the Data using the RHS Macro Final Result Quality Control Macro Troubleshooting EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October

4 1. Genepix Pro Software Compatibility The instructions described in this technical data sheet for the analysis of the EmbryoCellect TM array using the Genepix Pro software (Molecular Devices, USA) are compatible with version 6.1 and should be compatible with other versions. These instructions require the use of the Genepix Pro.gal files (Genepix_A, Genepix_B, Genepix_C, Genepix_D) and Macro 6.2, which are available from the RHS dropbox or the RHS website. The RHS analysis macro only works when the Excel decimal separator is a period (.) not a comma (,). While every effort has been made to ensure the protocol steps are compatible across all versions of Genepix Pro software, there may be a version where the user may experience a run error during the analyses of array data. If an error is encountered, please contact RHS immediately by ing support@rhsc.com.au and we will provide you with assistance. 2. General Instructions 2.1 Slide format Each EmbryoCellect TM slide contains four microarrays with identical probes in each (Figure 1). A B C D Figure 1: Microarray slide template for sample loading displaying the locations of individual arrays Four sample batching is possible, making EmbryoCellect TM ideal for processing small numbers of samples. Slides cannot be reused so the recommended minimum number of samples to be hybridised is four. If all four microarrays are not hybridised, the unused microarrays will not be able to be utilised in subsequent hybridisations. 2.2 Scanner requirements A dual channel scanner equipped with a green and red laser (532nm and 635nm wavelength, respectively) is required to scan EmbryoCellect TM slides. 2.3 Prescan A prescan allows low-resolution visualisation of the entire slide surface, showing the array locations (see Figure 2). Most scanners will also read the slide barcode. When the prescan has completed, the scan area can be adjusted independently for each microarray for the higher resolution analysis scan. Figure 2: Example of a prescan image illustrating individual array locations EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October

5 2.4 Scanning Some microarray scanners have automatic scanning settings which adjust the power of the laser and/or gain of the photomultiplier tube (PMT). The effects of these settings can be viewed on a histogram (see Figure 3). If required, these settings can usually be changed manually and should be adjusted until the red and green lines on the histogram overlap as much as possible (see Figure 4). Figure 3: Example of Histogram before scanning parameter optimisation. The high peak close to the Y axis represents the background signal. The secondary smaller peaks represent the microarray spot signal. The green line represents results from the green laser (532 wavelength test) and the red line represents results from the red laser (635 wavelength - reference). This image shows poor overlap between the green and red signals. Figure 4: Example of Histogram with ideal scan parameters. The red and green lines for the test and reference signal are overlapping. 2.5 Scan parameters In addition to the laser and PMT settings, other scanner software parameters should be adjusted for ideal EmbryoCellect TM slide scanning. Spot signals depend on several parameters, from sample DNA quality to hybridization and washing process success. Scan parameters should be adapted to each slide or array in order to fit within the following image quality control parameters; Pixel Saturation: EmbryoCellect TM slides must be scanned with minimal saturation because saturated pixels are excluded from the image. GenePix Pro s Auto-PMT option allows you to set the percentage of saturated pixels allowed. Set the option Saturation tolerance (%) to 1e-4 Pixel size: Set the Pixel Size (µm) to 10µm EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October

6 2.6 Saving TIFF images Images should be saved with the following settings as a TIFF file containing both colour images (this is usually the default scanner setting). RHS recommends that files are saved using the slide barcode plus the array suffix as a minimum. E.g. RHS001234_A. Figure 5: Save settings 3. Scanning Instructions Switch on Scanner Start GenePix Pro Software Allow scanner to warm up for ~15 minutes before use Insert Slide Ensure that the slide will be scanned from the top to the bottom array by placing the slide in the correct orientation (the barcode is at the bottom of the slide) EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October

7 3.1 Check the settings Open the options box Open the Workflow tab: Check the box: Search for Barcode after each Data Scan or Image load. Check the box: Register Images after a Data Scan. Open the Analysis tab: In the Ratio Formulation select 532/635. Open the Alignment tab: Change value for Minimum Diameter % to 50. Change value for Maximum Diameter % to 200. Change value for Composite Pixel Intensity (CPI) threshold to include a pixel in a feature during alignment to 500. Open the File save tab: In the Image and Results File Naming section, check the Prefix-Barcode box. Select OK EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October

8 3.2 Perform and save a Preview Scan of the slide Open the Hardware Settings dialog box: Set PMT levels at 600 for both 635 and 532. Leave the Power % setting at 100% Set the desired Pixel size to 10µm Leave the Lines to Average setting at 1. Leave the desired Focus Position setting at 0. Press Preview Scan Save the Preview Scan Select File/save Select Export Images Save the preview scan in a desired location as a JPEG 3.3 Scan each array individually Define the region of interest: Select the Scan Area tool Press F10 to bring up the Scan Area options (alternatively right click with the mouse and select Scan Area Options). Array Left Top Width Height A 7000 B C D Enter the following coordinates to select the scan areas for each of the RHS arrays, starting with array A. Press OK EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October

9 Load the Array Layout (.gal file) Click the File button Select the Load Array List command. (Shortcut: Alt + Y) Select the corresponding.gal file for the array you are analysing. E.g. Use the.gal file Genepix_A when scanning and analysing the A array. Use AUTO PMT to set the PMT Gain automatically Click the Auto-PMT button in the Hardware Settings box. Set the Saturation tolerance (%) to 1e-4. Click the Start button. Once finished, press Apply button. Perform a Data Scan Set the resolution for the data scan to 10µm in the Pixel Size field in the Hardware Settings box Press Data Scan EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October

10 Check for pixel saturation If there are any saturated spots; decrease PMT by 20 for both 635 and 532 then repeat the data scan. This image shows saturation in the 532 laser for some spots. The microarray should be rescanned with the PMT lowered by 20 for both lasers until no saturation is present. Align Template Line up the grid to overlap with the array by pressing F8 or click on the Gridding button and select the Find Array, Find Blocks, Align Features option. Ensure that GenePix Pro has correctly identified the position of the array and the blocks. If necessary due to incorrect spot positioning, incorrect flag, presence of debris or a scratch; If the array or the blocks are not correctly identified, move the grid manually into place with the mouse while in Block Mode. Move, resize, include or exclude spots by selecting Feature Mode Click once on a spot to select the spot + Use Ctrl + arrow keys to resize a grid spot Use the arrow keys to move a grid spot EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October

11 Grid spots should be adjusted to avoid dust or debris If necessary; right click on a selected spot and select Flag as Not Found or Flag as Bad to flag a spot and exclude it from the dataset Right click on selected spot and select Clear flags except Absent or Clear flags to include a spot that was previously flagged as not found EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October

12 If a halo effect occurs you may need to resize the spots to the inner ring of strong hybridisation otherwise it may be excluded from the dataset by the RHS macro. Save Image Select file, save images (Shortcut: Ctrl + S) In the Filename field, the slide name should already be filled in. Add the position of the array behind the barcode number (e.g. _A; _B; _C or _D Choose the correct folder for saving your image Click the Save button Proceed to steps 3.4 and 4 below before scanning other arrays on the same slide or another slide. EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October

13 3.4 Array quality control Each array has four replicate blocks. Spots are printed in duplicate in each of the four blocks to make a total of 8 spots for each chromosome (for example in the image to the right each spot marked with a * is the same chromosome). Verify that each of the replicates are consistent. * * * * * * * * In this example there was a bubble which caused uneven hybridisation of the spots in the middle of the array. You can see two of the replicates encompassed in this bubble hybridised differently. These spots should be flagged as bad and excluded from the data analysis. In this example some spots in the bottom left quadrant have been scratched and they are no longer consistent between the 8 spot replicates. These spots should be flagged as bad and excluded from the data analysis. EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October

14 Verify that there is consistent hybridisation across the whole array An array which appears like the above is considered less reliable and the sample should be re-hybridised Verify that there is consistent hybridisation across each spot 4. Quantifying the Data Press the Analyze button Arrays that have spots that have dark areas of hybridisation within them, as in the image above, are considered less reliable and the sample should be rehybridised EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October

15 When the analysis is finished, it will tab to the Results tab located at the top. The raw data is presented. Click the Save Results as... button (Shortcut: Alt + U) to save the results as a.gpr file. Save this in the same location as the scanned images Use the following save parameters: Repeat from 3.3 Scan each array individually with the remaining arrays Repeat from 3.2 Perform and save a Preview Scan of the slide to scan additional slides EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October

16 5. Analysing the Data using the RHS Macro Ensure macros are enabled Select enable macros when prompted or change settings in Excel Options Trust Centre Trust Centre Settings Macro Settings Open the RHS Mapix Analysis Macro The macro will open in the background of your raw data or will appear as an empty excel file Open the.gpr file with Microsoft Office Excel With the raw data file and the macro open Press Ctrl + Shift + E + + EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October

17 6. Final Result Quality Control The Analysis tab displays the resultant Karyotype and Array SD Check the Array SD - Results where SD > 0.10 are less reliable Check the number of spots analysed for each chromosome - Results where less than 4 spots are analysed per chromosome are less reliable If a small number of spots were analysed (eg 4 or less), check the reason for exclusion. Common reasons for exclusion include: - Flagged by Mapix as bad - Flagged by user as bad - Flagged by Macro as bad This can be due to saturation, the spot being too small or too big, the spot not being uniform (eg having a halo (which is an area of strong hybridisation in the middle of the spot surrounded by an area of weaker hybridisation), donuts (which is an area of weak hybridisation in the middle of a spot surrounded by strong hybridisation) or poor hybridisation. It is recommended that the sample is rehybridised. Contact support@rhsc.com.au for additional information. EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October

18 The ArrayData Graph tab displays the raw signal overlap between test and reference for each chromosome. The 2 coloured lines corresponding to the raw signal of both lasers should be as close to overlapping as possible. Where the test and reference lines do not overlap, this suggests: - A difference in the labelling efficiency between test and reference samples - Poor scanning parameters Results where the lines do not overlap are less reliable Good overlap X (Inv) Poor overlap X (Inv) Ensure there is sufficient fluorescent signal by checking the signal intensity on the Y axis. RHS generally sees intensities in the range of 30,000-50,000 units Weak signals may be due to: - Poor scanning parameters i.e. not enough laser power or the PMT detection gain is too low, or; - Samples being poorly labelled. This can be seen in the spectrophotometer readings and on the agarose gel of the labelled products - Loss of labelled sample during purification and/or precipitation Results with low intensities are less reliable Normal signal intensities X (Inv) Weak signal intensities X (Inv) EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October

19 X (Inv) Yp Negative Mean of normalized ratios X (Inv) Y Negative Mean of normalized ratios The Array Graph tab displays the mean normalized ratio for each chromosome; along with the standard deviation across each of the 8 spots analysed. Check the error bars on the Array Graph. Large error bars indicate: - inconsistent hybridisation across the array replicates This may be caused by debris, smearing of the microarray or bubbles, all of which are evident on the scanned image (jpeg file). Results with large error bars are less reliable than array results with smaller standard deviations per chromosome (see example) Large error bars Chromosome Small error bars Chromosome EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October

20 7. Macro Troubleshooting The RHS analysis macro only works when the Excel decimal separator is a period (.) not a comma (,). You can change these settings in Excel Options under the advanced tab. Please contact support@rhsc.com.au for technical assistance. EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October

21 Reproductive Health Science Ltd (ABN West Thebarton Road THEBARTON SA AUSTRALIA 5031 E: Ph: ASX:RHS EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October

In our previous lecture, we understood the vital parameters to be taken into consideration before data acquisition and scanning.

In our previous lecture, we understood the vital parameters to be taken into consideration before data acquisition and scanning. Interactomics: Protein Arrays & Label Free Biosensors Professor Sanjeeva Srivastava MOOC NPTEL Course Indian Institute of Technology Bombay Module 7 Lecture No 34 Software for Image scanning and data processing

More information

Instructions for Mapping * µarray Images using GenePix 5.0

Instructions for Mapping * µarray Images using GenePix 5.0 Instructions for Mapping * µarray Images using GenePix 5.0 Preliminary Information Make sure that the GenePix 5.0 software has been installed on your computer and you have the USB hardware dongle that

More information

NPTEL VIDEO COURSE PROTEOMICS PROF. SANJEEVA SRIVASTAVA

NPTEL VIDEO COURSE PROTEOMICS PROF. SANJEEVA SRIVASTAVA HANDOUT LECTURE-31 MICROARRAY WORK-FLOW: IMAGE SCANNING AND DATA PROCESSING Slide 1: This module contains the summary of the discussion with Mr. Pankaj Khanna, an application specialist from Spinco Biotech,

More information

Scan slides (Axon Genepix 4200AL)

Scan slides (Axon Genepix 4200AL) Page 1 Scan slides (Axon Genepix 4200AL) We need to scan the slides on both channels (Cy3 and Cy5) to obtain a 16-bit grayscale TIFF file for each. Typically these files are about 20-26Mb per channel,

More information

ScanArray Overview. Principle of Operation. Instrument Components

ScanArray Overview. Principle of Operation. Instrument Components ScanArray Overview The GSI Lumonics ScanArrayÒ Microarray Analysis System is a scanning laser confocal fluorescence microscope that is used to determine the fluorescence intensity of a two-dimensional

More information

GenePix Application Note

GenePix Application Note GenePix Application Note Biological Relevance of GenePix Results Shawn Handran, Ph.D. and Jack Y. Zhai, Ph.D. Axon Instruments, Inc. 3280 Whipple Road, Union City, CA 94587 Last Updated: Aug 22, 2003.

More information

Instructions for Howto Scan µarrays

Instructions for Howto Scan µarrays Instructions for Howto Scan µarrays Introduction After probing the µarray slides with samples, one is now ready to scan them. To scan a µarrays slide is too convert the biological information trapped on

More information

NPTEL VIDEO COURSE PROTEOMICS PROF. SANJEEVA SRIVASTAVA

NPTEL VIDEO COURSE PROTEOMICS PROF. SANJEEVA SRIVASTAVA LECTURE-31 MICROARRAY WORK-FLOW: IMAGE SACNNING AND DATA PROCESSING TRANSCRIPT Welcome to the proteomics course. In today s lecture we will talk about microarray work-flow the image scanning and processing.

More information

Products - Microarray Scanners - Laser Scanners - InnoScan 900 Series and MAPIX Software

Products - Microarray Scanners - Laser Scanners - InnoScan 900 Series and MAPIX Software Products - Microarray Scanners - Laser Scanners - InnoScan 900 Series and MAPIX Software Arrayit offers the world s only next generation microarray scanning technology, with proprietary rotary motion control,

More information

Quick Guide. NucleoCounter NC-3000

Quick Guide. NucleoCounter NC-3000 Quick Guide NucleoCounter NC-3000 Table of contents Setting up the FlexiCyte Protocol 2 Editing Image Capture and Analysis Parameters 3 Optimizing Exposure Time 4 Compensation for Spectral Overlap 6 Creating

More information

GenePix Application Note

GenePix Application Note GenePix Application Note Determining the Signal-to-Noise Ratio and Optimal Photomultiplier gain setting in the GenePix 4000B Siobhan Pickett, M.S., Sean Carriedo, Ph.D. and Chang Wang, Ph.D. Axon Instruments,

More information

Product Information. Introduction

Product Information. Introduction Microarray Scanner Calibration Slide To quantitatively analyze scanners performance and output, adjust and fine-tune scanners, and perform comparative analysis for multiple scanner units. To verify scanners

More information

Arcturus XT Laser Capture Microdissection System AutoScanXT Software Module. User Manual

Arcturus XT Laser Capture Microdissection System AutoScanXT Software Module. User Manual Arcturus XT Laser Capture Microdissection System AutoScanXT Software Module User Manual For Research Use Only. Not intended for any animal or human therapeutic or diagnostic use. Information in this document

More information

Volocity Tutorial Creating a Measurement Protocol in Volocity Software

Volocity Tutorial Creating a Measurement Protocol in Volocity Software TUTORIAL NOTE Cellular Imaging and Analysis Volocity Tutorial Creating a Measurement Protocol in Volocity Software This tutorial will illustrate the process of creating a Measurement Protocol in Volocity

More information

Microarray Data Pre-processing. Ana H. Barragan Lid

Microarray Data Pre-processing. Ana H. Barragan Lid Microarray Data Pre-processing Ana H. Barragan Lid Hybridized Microarray Imaged in a microarray scanner Scanner produces fluorescence intensity measurements Intensities correspond to levels of hybridization

More information

Scanning Procedure Using SilverFast Ai Software (6.6.1r2b) 1

Scanning Procedure Using SilverFast Ai Software (6.6.1r2b) 1 last modified 6.16.2015 Scanning Procedure Using SilverFast Ai Software (6.6.1r2b) 1 PREP In scanner notebook, fill out Incoming Checklist column on a Scanner Equipment Checklist sheet. Use flash drive

More information

Guidance on Using Scanning Software: Part 5. Epson Scan

Guidance on Using Scanning Software: Part 5. Epson Scan Guidance on Using Scanning Software: Part 5. Epson Scan Version of 4/29/2012 Epson Scan comes with Epson scanners and has simple manual adjustments, but requires vigilance to control the default settings

More information

Scanning and Image Processing -by Steve Clough

Scanning and Image Processing -by Steve Clough Scanning and Image Processing -by Steve Clough cdna microarrays use two dyes with well separated emission spectra such as Cy3 and Cy5 to allow direct comparisons on single slide GSI Lumonics Ratio of Expression

More information

6.12. SilverFastJobManager. SilverFast JobManager for Film Scanner. Overview

6.12. SilverFastJobManager. SilverFast JobManager for Film Scanner. Overview SilverFast JobManager for Film Scanner Overview To activate the JobManager, click on JobManager -button in the vertical list of buttons to the left of the large SilverFastAi preview window SilverFastAi

More information

Developed by BioDiscovery, Inc. DualChip evaluation software User Manual Version 1.1

Developed by BioDiscovery, Inc. DualChip evaluation software User Manual Version 1.1 Developed by BioDiscovery, Inc. DualChip evaluation software User Manual Version 1.1 1 Table of contents 1. INTRODUCTION...3 2. SCOPE OF DELIVERY...4 3. INSTALLATION PROCEDURES...5 3.1. SYSTEM REQUIREMENTS...

More information

MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION

MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION Olympus VS120-L100 Slide Scanner Standard Operating Procedure Startup 1) Red power bar switch (behind monitor) 2) Computer 3) Login: UserVS120 account (no password) 4) Double click: WAIT FOR INITIALIZATION

More information

Brightness and Contrast Control Reference Guide

Brightness and Contrast Control Reference Guide innovation Series Scanners Brightness and Contrast Control Reference Guide A-61506 Part No. 9E3722 CAT No. 137 0337 Using the Brightness and Contrast Control This Reference Guide provides information and

More information

Before you start, make sure that you have a properly calibrated system to obtain high-quality images.

Before you start, make sure that you have a properly calibrated system to obtain high-quality images. CONTENT Step 1: Optimizing your Workspace for Acquisition... 1 Step 2: Tracing the Region of Interest... 2 Step 3: Camera (& Multichannel) Settings... 3 Step 4: Acquiring a Background Image (Brightfield)...

More information

Scanning Setup Guide for TWAIN Datasource

Scanning Setup Guide for TWAIN Datasource Scanning Setup Guide for TWAIN Datasource Starting the Scan Validation Tool... 2 The Scan Validation Tool dialog box... 3 Using the TWAIN Datasource... 4 How do I begin?... 5 Selecting Image settings...

More information

Using the TWAIN Datasource

Using the TWAIN Datasource Using the TWAIN Datasource Starting the Scan Validation Tool... 2 The Scan Validation Tool dialog box... 2 Using the TWAIN Datasource... 4 How do I begin?... 4 Creating a new Setting Shortcut... 5 Changing

More information

3. When you import the scanner for the first time make sure you change it from Full Auto Mode to that of Professional Mode.

3. When you import the scanner for the first time make sure you change it from Full Auto Mode to that of Professional Mode. PhotoShop Tutorials Scanning Photographic Film WorkFlow 1. Open PhotoShop 2. File > Import > choose scanner 3. When you import the scanner for the first time make sure you change it from Full Auto Mode

More information

dlsoft Barcode Analyser By dlsoft

dlsoft Barcode Analyser By dlsoft dlsoft Barcode Analyser By dlsoft This manual was produced using ComponentOne Doc-To-Help. Contents BarAnalyser 1 Introduction... 1 Barcode symbologies... 5 How to use BarAnalyser... 5 Walk through...

More information

Spotxel 1.7 Microarray Image and Data Analysis Software User s Guide

Spotxel 1.7 Microarray Image and Data Analysis Software User s Guide Spotxel 1.7 Microarray Image and Data Analysis Software User s Guide 27 April 2017 - Rev 7 Spotxel is only intended for research and not intended or approved for diagnosis of disease in humans or animals.

More information

Using the TWAIN Datasource

Using the TWAIN Datasource Using the TWAIN Datasource Starting the Scan Validation Tool... 2 The Scan Validation Tool dialog box... 3 Using the TWAIN Datasource... 5 How do I begin?... 5 Changing or creating a Setting Shortcut...

More information

Color and More. Color basics

Color and More. Color basics Color and More In this lesson, you'll evaluate an image in terms of its overall tonal range (lightness, darkness, and contrast), its overall balance of color, and its overall appearance for areas that

More information

Managing images with NewZapp

Managing images with NewZapp Managing images with NewZapp This guide is for anyone using the NewZapp Fixed editor as opposed to the Drag and Drop editor. The Image Manager is where images are uploaded and stored in your NewZapp account

More information

Learning Photo Retouching techniques the simple way

Learning Photo Retouching techniques the simple way Learning Photo Retouching techniques the simple way Table of Contents About the Workshop... i Workshop Objectives... i Getting Started... 1 Photoshop Workspace... 1 Setting up the Preferences... 2 Retouching

More information

User Guide V10 SP1 Addendum

User Guide V10 SP1 Addendum Alibre Design User Guide V10 SP1 Addendum Copyrights Information in this document is subject to change without notice. The software described in this document is furnished under a license agreement or

More information

Scanning Setup Guide for the TWAIN Datasource

Scanning Setup Guide for the TWAIN Datasource Scanning Setup Guide for the TWAIN Datasource Product overview... 2 Capture system... 2 Starting the Scan Validation Tool... 3 The Scan Validation Tool dialog box... 3 How do I begin?... 5 Selecting Image

More information

Monaco ColorWorks User Guide

Monaco ColorWorks User Guide Monaco ColorWorks User Guide Monaco ColorWorks User Guide Printed in the U.S.A. 2003 Monaco Systems, Inc. All rights reserved. This document contains proprietary information of Monaco Systems, Inc. No

More information

i1800 Series Scanners

i1800 Series Scanners i1800 Series Scanners Scanning Setup Guide A-61580 Contents 1 Introduction................................................ 1-1 About this manual........................................... 1-1 Image outputs...............................................

More information

COPYRIGHT. Limited warranty. Limitation of liability. Note. Customer remedies. Introduction. Artwork 23-Aug-16 ii

COPYRIGHT. Limited warranty. Limitation of liability. Note. Customer remedies. Introduction. Artwork 23-Aug-16 ii ARTWORK Introduction COPYRIGHT Copyright 1998-2016. Wilcom Pty Ltd, Wilcom International Pty Ltd. All Rights reserved. All title and copyrights in and to Digitizer Embroidery Software (including but not

More information

Extreme Makeovers: Photoshop Retouching Techniques

Extreme Makeovers: Photoshop Retouching Techniques Extreme Makeovers: Table of Contents About the Workshop... 1 Workshop Objectives... 1 Getting Started... 1 Photoshop Workspace... 1 Retouching Tools... 2 General Steps... 2 Resolution and image size...

More information

Using Adobe Photoshop to enhance the image quality. Assistant course web site:

Using Adobe Photoshop to enhance the image quality. Assistant course web site: Using Adobe Photoshop to enhance the image quality Assistant course web site: http://www.arches.uga.edu/~skwang/edit6170/course.htm Content Introduction 2 Unit1: Scan images 3 Lesson 1-1: Preparations

More information

DataCapture Transcript Module Getting Started Guide

DataCapture Transcript Module Getting Started Guide DataCapture Transcript Module Getting Started Guide Version: 6.6 Written by: Product Documentation, R&D Date: February 2011 ImageNow and CaptureNow are registered trademarks of Perceptive Software, Inc.

More information

Using the TWAIN Datasource

Using the TWAIN Datasource Using the TWAIN Datasource Starting the Scan Validation Tool... 2 The Scan Validation Tool dialog box... 2 Using the TWAIN Datasource... 4 How do I begin?... 4 Creating a new Setting Shortcut... 5 Changing

More information

Chapter 6: TVA MR and Cardiac Function

Chapter 6: TVA MR and Cardiac Function Chapter 6 Cardiac MR Introduction Chapter 6: TVA MR and Cardiac Function The Time-Volume Analysis (TVA) optional module calculates time-dependent behavior of volumes in multi-phase studies from MR. An

More information

Chlorophyll Fluorescence Imaging System

Chlorophyll Fluorescence Imaging System Quick Start Guide Chlorophyll Fluorescence Imaging System Quick Start Guide for Technologica FluorImager software for use with Technlogica CFImager hardware Copyright 2006 2015 TECHNOLOGICA LIMITED. All

More information

Steps involved in microarray analysis after the experiments

Steps involved in microarray analysis after the experiments Steps involved in microarray analysis after the experiments Scanning slides to create images Conversion of images to numerical data Processing of raw numerical data Further analysis Clustering Integration

More information

Technical Bulletin. Guide for Using BD Cytometric Bead Array (CBA) Flex Sets with the BD Accuri C6 Flow Cytometer. Introduction

Technical Bulletin. Guide for Using BD Cytometric Bead Array (CBA) Flex Sets with the BD Accuri C6 Flow Cytometer. Introduction Guide for Using BD Cytometric Bead Array (CBA) Flex Sets with the BD Accuri C6 Contents 1 Introduction 2 Materials and Methods 3 Setup Procedure Introduction BD Cytometric Bead Array (CBA) flex sets provide

More information

User Manual Veterinary

User Manual Veterinary Veterinary Acquisition and diagnostic software Doc No.: Rev 1.0.1 Aug 2013 Part No.: CR-FPM-04-022-EN-S 3DISC, FireCR, Quantor and the 3D Cube are trademarks of 3D Imaging & Simulations Corp, South Korea,

More information

Importing and processing gel images

Importing and processing gel images BioNumerics Tutorial: Importing and processing gel images 1 Aim Comprehensive tools for the processing of electrophoresis fingerprints, both from slab gels and capillary sequencers are incorporated into

More information

TOON BOOM HARMONY Advanced Edition - Compositing and Effects Guide (Server)

TOON BOOM HARMONY Advanced Edition - Compositing and Effects Guide (Server) TOON BOOM HARMONY 12.1 - Advanced Edition - Compositing and Effects Guide (Server) Legal Notices Toon Boom Animation Inc. 4200 Saint-Laurent, Suite 1020 Montreal, Quebec, Canada H2W 2R2 Tel: +1 514 278

More information

FLIR Tools for PC 7/21/2016

FLIR Tools for PC 7/21/2016 FLIR Tools for PC 7/21/2016 1 2 Tools+ is an upgrade that adds the ability to create Microsoft Word templates and reports, create radiometric panorama images, and record sequences from compatible USB and

More information

Stitching MetroPro Application

Stitching MetroPro Application OMP-0375F Stitching MetroPro Application Stitch.app This booklet is a quick reference; it assumes that you are familiar with MetroPro and the instrument. Information on MetroPro is provided in Getting

More information

CALIBRATION MANUAL. Version Author: Robbie Dowling Lloyd Laney

CALIBRATION MANUAL. Version Author: Robbie Dowling Lloyd Laney Version 1.0-1012 Author: Robbie Dowling Lloyd Laney 2012 by VirTra Inc. All Rights Reserved. VirTra, the VirTra logo are either registered trademarks or trademarks of VirTra in the United States and/or

More information

Contents Foreword 1 Feedback 2 Legal information 3 Getting started 4 Installing the correct Capture One version 4 Changing the version type 5 Getting

Contents Foreword 1 Feedback 2 Legal information 3 Getting started 4 Installing the correct Capture One version 4 Changing the version type 5 Getting Contents Foreword 1 Feedback 2 Legal information 3 Getting started 4 Installing the correct Capture One version 4 Changing the version type 5 Getting to know Capture One Pro 6 The Grand Overview 6 The

More information

AUTODESK INVENTOR Trial Projects

AUTODESK INVENTOR Trial Projects AUTODESK INVENTOR Trial Projects Drawing Creation Create detailed drawings of a collar flange PART 1: CREATING DRAWING VIEWS page: 2 1. 2. 3. Start by clicking the Projects icon in the ribbon. Navigate

More information

Océ Colour Copy 7.0. User Manual

Océ Colour Copy 7.0. User Manual Océ Colour Copy 7.0 User Manual Océ-Technologies B.V. Trademarks Products in this manual are referred to by their trade names. In most, if not all cases, these designations are claimed as trademarks or

More information

Operating Rausch ScanCam within POSM.

Operating Rausch ScanCam within POSM. Operating Rausch ScanCam within POSM. POSM (Pipeline Observation System Management) // posmsoftware.com // info@posmsoftware.com // 859-274-0041 RAUSCH USA // www.rauschusa.com // reusa@rauschusa.com //

More information

UN-SCAN-IT gel TM Version 6.3. User's Manual for. Macintosh (OS X)

UN-SCAN-IT gel TM Version 6.3. User's Manual for. Macintosh (OS X) UN-SCAN-IT gel TM Version 6.3 User's Manual for Macintosh (OS X) Silk Scientific Inc. software and manuals are copyrighted, which means that your right to copy any part of them is limited by copyright

More information

HP Designjet HD Scanner and T1200 HD Multifunction Printer

HP Designjet HD Scanner and T1200 HD Multifunction Printer HP Designjet HD Scanner and T1200 HD Multifunction Printer Introductory information USB 2.0 high-speed certified Introductory Information Other sources of information The User s Guide for your scanner

More information

SCANNING METHODS. Edited 7/2014 JKH REPRESENTING ORIGINAL IMAGE CONDITION - 2 SCANS. 1) Open Adobe Photoshop.

SCANNING METHODS. Edited 7/2014 JKH REPRESENTING ORIGINAL IMAGE CONDITION - 2 SCANS. 1) Open Adobe Photoshop. SCANNING METHODS REPRESENTING ORIGINAL IMAGE CONDITION - 2 SCANS 1) Open Adobe Photoshop. 2) Go to File > Import > Epson Perfection V500. 3) Place item on scanner. Align the item in the upper left corner

More information

Machinery HDR Effects 3

Machinery HDR Effects 3 1 Machinery HDR Effects 3 MACHINERY HDR is a photo editor that utilizes HDR technology. You do not need to be an expert to achieve dazzling effects even from a single image saved in JPG format! MACHINERY

More information

DOCUMENT SCANNER INSTRUCTIONS. Space. Backup. Count Only. New File. Scanner. Feeding Option Manual Auto Semi-Auto

DOCUMENT SCANNER INSTRUCTIONS. Space. Backup. Count Only. New File. Scanner. Feeding Option Manual Auto Semi-Auto E FILM F Scanner A Space Count Only New File Feeding Option Manual Auto Semi-Auto Backup DOCUMENT SCANNER INSTRUCTIONS NOTICE q Copyright 2001 by CANON ELECTRONICS INC. All rights reserved. No part of

More information

EVOS M5000 Imaging System

EVOS M5000 Imaging System EVOS M5000 Imaging System Pub. No. MAN0017765 Doc. Part No. 710209 Rev. A.0 This document is intended as a benchtop reference for the users of the EVOS M5000 Imaging System (Cat. No. AMF5000). For detailed

More information

GALILEO TMA CK 4500 HTS Tissue Microarray Platform

GALILEO TMA CK 4500 HTS Tissue Microarray Platform GALILEO TMA CK 4500 HTS Tissue Microarray Platform Tissue Microarray (TMA) A Block Of Samples From Hundreds Of Blocks (S. M. Hewitt, M.D., Ph.D., Tissue Array Research Program, LP, CCR, NCI, NIH) TMA technology

More information

Drawing output in imos ix 2017 SR2. Date of creation: March 2018; last changes Version used: ix 2017 SR2

Drawing output in imos ix 2017 SR2. Date of creation: March 2018; last changes Version used: ix 2017 SR2 Drawing output in imos ix 2017 SR2 Date of creation: March 2018; last changes 13.04.2018 Version used: ix 2017 SR2 We make every effort to ensure the content of this document is complete, accurate and

More information

PASS4TEST. IT Certification Guaranteed, The Easy Way! We offer free update service for one year

PASS4TEST. IT Certification Guaranteed, The Easy Way!  We offer free update service for one year PASS4TEST IT Certification Guaranteed, The Easy Way! \ We offer free update service for one year Exam : 9A0-125 Title : Adobe Photoshop Lightroom 2 ACE Exam Vendors : Adobe Version : DEMO Get Latest &

More information

We recommend downloading the latest core installer for our software from our website. This can be found at:

We recommend downloading the latest core installer for our software from our website. This can be found at: Dusk Getting Started Installing the Software We recommend downloading the latest core installer for our software from our website. This can be found at: https://www.atik-cameras.com/downloads/ Locate and

More information

W i n d o w s. ScanGear CS-S 4.3 for CanoScan FB1200S Color Image Scanner. User's Guide

W i n d o w s. ScanGear CS-S 4.3 for CanoScan FB1200S Color Image Scanner. User's Guide f o r W i n d o w s ScanGear CS-S 4.3 for CanoScan FB1200S Color Image Scanner User's Guide How to Make Best Use of the Manuals 2 When you open the box Printed Manual CanoScan FB1200S Quick Start Guide

More information

ScanGear CS-U 5.3 for CanoScan FB630U/FB636U Color Image Scanner User s Guide

ScanGear CS-U 5.3 for CanoScan FB630U/FB636U Color Image Scanner User s Guide ScanGear CS-U 5.3 for CanoScan FB630U/FB636U Color Image Scanner User s Guide Copyright Notice 1999 Canon Inc. This manual is copyrighted with all rights reserved. Under the copyright laws, this manual

More information

RAGE TOOL KIT FAQ. Terms and Conditions What legal terms and conditions apply to the RAGE Tool Kit?

RAGE TOOL KIT FAQ. Terms and Conditions What legal terms and conditions apply to the RAGE Tool Kit? RAGE TOOL KIT FAQ Terms and Conditions What legal terms and conditions apply to the RAGE Tool Kit? Editing and Building Maps What are the recommended system specifications for running the RAGE Tool Kit?

More information

Quality control of microarrays

Quality control of microarrays Quality control of microarrays Solveig Mjelstad Angelskår Intoduction to Microarray technology September 2009 Overview of the presentation 1. Image analysis 2. Quality Control (QC) general concepts 3.

More information

ColorPony User Manual

ColorPony User Manual ColorPony 2013-07 User Manual ColorPony User Manual The major sections of the ColorPony user manual 1. Prerequisites.. p3 2. Terminology p4 3. Capturing & Yoking Artwork... p5 3.1. Measure colors p8 3.2.

More information

Image and Data Acquisition

Image and Data Acquisition Image and Data Acquisition LCP Image Acquisition Procedures This section provides guidelines for scanning images that will be added to the LCP image archive. By scanning the image, we obtain a digital

More information

An Introduction to Lasercut 5.3 Preparing the Artwork

An Introduction to Lasercut 5.3 Preparing the Artwork An Introduction to Lasercut 5.3 Preparing the Artwork Version 0.1, December 8th 2015 Precautions Introduction Importing from.dxf Setting up the layers Checking the Operations Tips, Mistakes and Problems

More information

Definiens. Tissue Studio 4.2. Tutorial 1: Composer and Nuclear Markers

Definiens. Tissue Studio 4.2. Tutorial 1: Composer and Nuclear Markers Definiens Tissue Studio 4.2 Tutorial 1: Composer and Nuclear Markers Tutorial 1: Composer and Nuclear Markers Imprint and Version Copyright 2015 Definiens AG. All rights reserved. This document may be

More information

Configuring User Settings at the Scan Station

Configuring User Settings at the Scan Station Configuring User Settings at the Scan Station Contents Touchscreen configurations... 2 Getting familiar with the touch screen... 4 Destination screen... 4 Icon descriptions: Destination screen... 5 Settings

More information

Locating Molecules Using GSD Technology Project Folders: Organization of Experiment Files...1

Locating Molecules Using GSD Technology Project Folders: Organization of Experiment Files...1 .....................................1 1 Project Folders: Organization of Experiment Files.................................1 2 Steps........................................................................2

More information

Image Capture TOTALLAB

Image Capture TOTALLAB 1 Introduction In order for image analysis to be performed on a gel or Western blot, it must first be converted into digital data. Good image capture is critical to guarantee optimal performance of automated

More information

Center for Microscopy and Image Analysis Axio Scan.Z1 Operating Manual

Center for Microscopy and Image Analysis Axio Scan.Z1 Operating Manual No index entries found. Center for Microscopy and Image Analysis Axio Scan.Z1 Operating Manual Table of contents 1. Starting procedure... 3 1.1. Turn on hardware... 3 1.2. Starting ZEN blue... 4 2. Load

More information

Label Studio Quick Start Guide

Label Studio Quick Start Guide Label Studio Quick Start Guide Overview The goal of the LabelStudio program is to help you layout and manage bulk jobs containing a mix of different logos, that all require a specified quanity. LabelStudio

More information

UNIVERSITY OF WATERLOO Physics 360/460 Experiment #2 ATOMIC FORCE MICROSCOPY

UNIVERSITY OF WATERLOO Physics 360/460 Experiment #2 ATOMIC FORCE MICROSCOPY UNIVERSITY OF WATERLOO Physics 360/460 Experiment #2 ATOMIC FORCE MICROSCOPY References: http://virlab.virginia.edu/vl/home.htm (University of Virginia virtual lab. Click on the AFM link) An atomic force

More information

Printer Software Guide

Printer Software Guide Printer Software Guide (For Canon CP Printer Solution Disk Version 4) Macintosh 1 Contents Safety Precautions...3 Read This First...4 About the Manuals...4 Printing Flow Diagram...5 Printing...7 Starting

More information

The ideal K-12 science microscope solution. User Guide. for use with the Nova5000

The ideal K-12 science microscope solution. User Guide. for use with the Nova5000 The ideal K-12 science microscope solution User Guide for use with the Nova5000 NovaScope User Guide Information in this document is subject to change without notice. 2009 Fourier Systems Ltd. All rights

More information

ScanMate. i920 Scanner. Scanning Setup Guide for TWAIN Applications A-61733

ScanMate. i920 Scanner. Scanning Setup Guide for TWAIN Applications A-61733 ScanMate i920 Scanner Scanning Setup Guide for TWAIN Applications A-61733 Scanning Setup Guide for the TWAIN Datasource Starting the Scan Validation Tool... 2 The Scan Validation Tool dialog box... 3 Using

More information

Version 4.2 For WINDOWS

Version 4.2 For WINDOWS SIGMA Photo Pro User Guide Companion Processing Software for SIGMA Digital Cameras Version 4.2 For WINDOWS I M P O R T A N T : Please Read Before You Start Thank you very much for purchasing this SIGMA

More information

Making an Architectural Drawing Template

Making an Architectural Drawing Template C h a p t e r 8 Addendum: Architectural Making an Architectural Drawing Template In this chapter, you will learn the following to World Class standards:! Starting from Scratch for the Last time! Creating

More information

Mullingar Camera Club Basic introduction to Digital Printing using Photoshop CC.

Mullingar Camera Club Basic introduction to Digital Printing using Photoshop CC. Mullingar Camera Club Basic introduction to Digital Printing using Photoshop CC. Table of Contents Course aims: 1 Course presentation notes: 1 Introducing Photoshop: 1 Adjusting the Brightness or Contrast

More information

Collage Images. By Rob MacIlreith For Use with GIMP 2.8 Last Updated July 2013

Collage Images. By Rob MacIlreith For Use with GIMP 2.8 Last Updated July 2013 Collage Images By Rob MacIlreith For Use with GIMP 2.8 Last Updated July 2013 Before Beginning: Have your collage images saved in your H:\Drive. You will have to copy and paste them from your flash drive

More information

Manual. Cell Border Tracker. Jochen Seebach Institut für Anatomie und Vaskuläre Biologie, WWU Münster

Manual. Cell Border Tracker. Jochen Seebach Institut für Anatomie und Vaskuläre Biologie, WWU Münster Manual Cell Border Tracker Jochen Seebach Institut für Anatomie und Vaskuläre Biologie, WWU Münster 1 Cell Border Tracker 1. System Requirements The software requires Windows XP operating system or higher

More information

Adobe Photoshop PS2, Part 3

Adobe Photoshop PS2, Part 3 Adobe Photoshop PS2, Part 3 Basic Photo Corrections This guide steps you through the process of acquiring, resizing, and retouching a photo intended for posting on the Web as well as for a print layout.

More information

SKF TKTI. Thermal Camera Software. Instructions for use

SKF TKTI. Thermal Camera Software. Instructions for use SKF TKTI Thermal Camera Software Instructions for use Table of contents 1. Introduction...4 1.1 Installing and starting the Software... 5 2. Usage Notes...6 3. Image Properties...7 3.1 Loading images

More information

Training Guide for Carl Zeiss LSM 880 with AiryScan FAST

Training Guide for Carl Zeiss LSM 880 with AiryScan FAST Training Guide for Carl Zeiss LSM 880 with AiryScan FAST ZEN 2.3 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2018) Power ON Routine 1 2 Turn ON Main Switch from the remote control

More information

Drawing Layouts Paper space & Model Space

Drawing Layouts Paper space & Model Space Drawing Layouts Paper space & Model Space Users of Bricscad will have seen the tabs at the bottom left of the drawings area labelled: Model, Layout1, Layout2 but may not know how to use them or what they

More information

Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope

Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope ZEN 2012 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 If you require fluorescence imaging,

More information

Positive Pixel Count Algorithm. User s Guide

Positive Pixel Count Algorithm. User s Guide Positive Pixel Count Algorithm User s Guide Copyright 2004, 2006 2008 Aperio Technologies, Inc. Part Number/Revision: MAN 0024, Revision B Date: December 9, 2008 This document applies to software versions

More information

IncuCyte ZOOM Fluorescent Processing Overview

IncuCyte ZOOM Fluorescent Processing Overview IncuCyte ZOOM Fluorescent Processing Overview The IncuCyte ZOOM offers users the ability to acquire HD phase as well as dual wavelength fluorescent images of living cells producing multiplexed data that

More information

Océ Colour Copy 5.0. User Manual

Océ Colour Copy 5.0. User Manual Océ Colour Copy 5.0 User Manual Océ-Technologies B.V. Trademarks Products in this manual are referred to by their trade names. In most, if not all cases, these designations are claimed as trademarks or

More information

ab Firefly Cytometer Setup Particles

ab Firefly Cytometer Setup Particles Version 1 Last updated 12 October 2016 ab211043 Firefly Cytometer Setup Particles For cytometer performance optimization for use with Firefly Multiplex particles. This product is for research use only

More information

Signal Integrity Analyzer

Signal Integrity Analyzer 1 Norlinvest Ltd, BVI. is a trade name of Norlinvest Ltd. All Rights Reserved. No part of the Signal Integrity Analyzer document can be reproduced in any form or by any means without the prior written

More information

SilverFast. Colour Management Tutorial. LaserSoft Imaging

SilverFast. Colour Management Tutorial. LaserSoft Imaging SilverFast Colour Management Tutorial LaserSoft Imaging SilverFast Copyright Copyright 1994-2006 SilverFast, LaserSoft Imaging AG, Germany No part of this publication may be reproduced, stored in a retrieval

More information

Toon Boom Harmony Reference Guide 3. Chapter 1: Dialog Boxes 7. Add Column Dialog Box 7. Add Drawing Layer Dialog Box 9

Toon Boom Harmony Reference Guide 3. Chapter 1: Dialog Boxes 7. Add Column Dialog Box 7. Add Drawing Layer Dialog Box 9 TOC Toon Boom Harmony 12.2.1 Reference Guide 3 Chapter 1: Dialog Boxes 7 7 Add Column Dialog Box 7 Add Drawing Layer Dialog Box 9 Add Frames Dialog Box 10 Advanced Save Dialog Box 11 Auto-Matte Dialog

More information

ISO Verification Fact Sheet

ISO Verification Fact Sheet Introduction To maintain scanning efficiency, barcodes need to be printed correctly and be of a high quality. Verification equipment can assist in providing an indication of the print quality of barcodes.

More information