Instructions for Howto Scan µarrays

Size: px
Start display at page:

Download "Instructions for Howto Scan µarrays"

Transcription

1 Instructions for Howto Scan µarrays Introduction After probing the µarray slides with samples, one is now ready to scan them. To scan a µarrays slide is too convert the biological information trapped on the slide into digital information for data analysis. We use a GenePix 4000B scanner ( or The act of scanning involves using a laser to excite the fluorescent markers on either the antigens or secondary antibodies and detecting the intensity of the fluorescence given off by these markers after being excited. The fluorescent intensity given off by the markers is dependent on three things, 1) the number of markers present, 2) the power of the excitation laser, and 3) the setting on the detector (PMT * ). For any given experiment, the entire set of µarrays slides probed in the experiment is scanned at the same laser power and PMT settings. Therefore, for any given scan session, the fluorescent intensity depends only on the number of markers present on the µarrays slides. Thus, in theory, we could quantify the number of antibodies from a sample that bind to a particular antigen. In practice, we use the fluorescent intensity for relative measurements, meaning, we report the results of our data analysis as sample A has autoantibodies with more/less/equal reactivity against these antigens when compared with sample B. Nonetheless, the data is quantitative and the fact that we could report absolute values makes the technique a very powerful tool. It just doesn t seem necessary at this point because our view of biology is not so precise as to require absolute values. Scanning a µarray Slide Launch GenePix 5.0 A program will search for the scanner and a screen might appear like the one pictured below. Uncheck the update reminder and click Continue. * PMT is an acronym for photo multiplier tube. It is a device that converts how much light striking the detector into a numeric value.

2 I ve added some tags (shown in red with their keyboard shortcut in parentheses) for the more common buttons or features used in the scan. Manual Brightness/Contrast Data Scan (Alt+D) Auto Scale Brightness/Contrast (Alt+T) Zoom Mode (Z) Hand Mode (H) View Scan Area (Ctrl+V) File Hardware Settings (Alt+Q)

3 Setting up the Scan Click Hardware Settings (Alt+Q) Adjust the Power for both channels (635 & 532) to 100%. Adjust the PMT settings for each wavelength to a value that maximizes the brightness of the signal intensity without saturating the detector. The spots will appear white when the detector is saturated. An optimal setting is where the entire µarray shows only a couple of saturated spots. The PMT settings represent the trade-off between signal and noise. One wants to have the Gain as high as possible where the spots are bright and not saturated. One also wants to keep the background level low. Unfortunately, the Gain setting enhances both the signal and the noise, thus, the trade-off. Set the Pixel Size to 10 µm Lines to average 1 (we only need to scan the slide once) Focus Position 0 µm The Scan area: (XXXX x XXXX pixels) and Scan duration: (XX seconds) are set by the scan area and the number of lines to average. Select Hand Mode (H). Select View Scan Area (Ctrl+V) and adjust the rectangle to the best guess for including all the spots on the µarray. It s best to make the border slightly larger so there is some error room because each slide will have the grid of spots in a slightly different location. One saves time by not having to adjust the frame for each slide. The savings may seem like a small amount, say 10 seconds, but over the course of an entire scan of 60 slides it adds an additional 10 minutes of time sitting in front of the scanner. The power setting adjusts how much power the laser will use to excite the sample. There are two collection wavelengths (635 nm and 532 nm) corresponding to red and green colors respectively.

4 Select File, Save Settings As (Alt+H). Name the settings for this scan session. It s a good idea to include the experiment name/number/date and the print date for the µarray slides. For example, here s the format I use: ScanSettings_expXX_YYYY_MM_DD_print_YYYY_MM_DD Note, since we re using the same settings for all the slides from a given experiment, it s important to save the settings in case one needs to come back and scan the same slides at a later date. Scanning the Slide Open up the scanner, lift up the slide clamp, and place the slide, label side up, into the scanning slot. Close the slide clamp. Close the scanner. The scanner will make some noise as it loads the slide into position to scan. Select Scan (Alt+D). The scanner will make some noise as the laser scans the slide. Select Zoom Mode (Z). One can zoom into the square region for a closer view of the µarray spots. After the scanner is finished, determine whether the settings are optimal for the experiment. Non-optimal settings would be incorrect PMT settings (spots are too dim or too many saturated spots) or the square is in the wrong place. Adjust the settings so they are optimal and save the settings. If the settings were not optimal on a scan, then one needs to scan the slide again. In this case, it s better not to save the image. If the scan settings are optimal, then save the settings (see above) one more time. Again, remember to keep the same scan settings for the entire set of slides.

5 Saving the Image Select File, Save Images (Ctrl+S). Choose single image tiff and check both channels (532 & 635) to have an image of the red and green intensities for each slide. We use the information from both channels for some analyses so it s best to collect it now if you re not sure. That way, you don t need to come back and scan all the slides again. Name the slide. It s a good idea to include the slide number and print date for the µarray slide. For example, here s the format I use: slxx_print_yyyy_mm_dd Continue Scanning the rest of the slides in the experiment.

Instructions for Mapping * µarray Images using GenePix 5.0

Instructions for Mapping * µarray Images using GenePix 5.0 Instructions for Mapping * µarray Images using GenePix 5.0 Preliminary Information Make sure that the GenePix 5.0 software has been installed on your computer and you have the USB hardware dongle that

More information

EmbryoCellect. RHS Scanning and Analysis Instructions. for. Genepix Pro Software

EmbryoCellect. RHS Scanning and Analysis Instructions. for. Genepix Pro Software EmbryoCellect RHS Scanning and Analysis Instructions for Genepix Pro Software EmbryoCellect Genepix Pro Scanning and Analysis Technical Data Sheet Version 1.0 October 2015 1 Copyright Reproductive Health

More information

MIF ZEISS LSM510 CONFOCAL USER PROTOCOL

MIF ZEISS LSM510 CONFOCAL USER PROTOCOL MIF ZEISS LSM510 CONFOCAL USER PROTOCOL START-UP Turn on the Mercury Bulb Power Supply (if needed). Power-on the Control Box. Turn on the computer. Open the LSM 510 software. Choose Scan New Images and

More information

GenePix Application Note

GenePix Application Note GenePix Application Note Determining the Signal-to-Noise Ratio and Optimal Photomultiplier gain setting in the GenePix 4000B Siobhan Pickett, M.S., Sean Carriedo, Ph.D. and Chang Wang, Ph.D. Axon Instruments,

More information

In our previous lecture, we understood the vital parameters to be taken into consideration before data acquisition and scanning.

In our previous lecture, we understood the vital parameters to be taken into consideration before data acquisition and scanning. Interactomics: Protein Arrays & Label Free Biosensors Professor Sanjeeva Srivastava MOOC NPTEL Course Indian Institute of Technology Bombay Module 7 Lecture No 34 Software for Image scanning and data processing

More information

ScanArray Overview. Principle of Operation. Instrument Components

ScanArray Overview. Principle of Operation. Instrument Components ScanArray Overview The GSI Lumonics ScanArrayÒ Microarray Analysis System is a scanning laser confocal fluorescence microscope that is used to determine the fluorescence intensity of a two-dimensional

More information

NPTEL VIDEO COURSE PROTEOMICS PROF. SANJEEVA SRIVASTAVA

NPTEL VIDEO COURSE PROTEOMICS PROF. SANJEEVA SRIVASTAVA HANDOUT LECTURE-31 MICROARRAY WORK-FLOW: IMAGE SCANNING AND DATA PROCESSING Slide 1: This module contains the summary of the discussion with Mr. Pankaj Khanna, an application specialist from Spinco Biotech,

More information

Scan slides (Axon Genepix 4200AL)

Scan slides (Axon Genepix 4200AL) Page 1 Scan slides (Axon Genepix 4200AL) We need to scan the slides on both channels (Cy3 and Cy5) to obtain a 16-bit grayscale TIFF file for each. Typically these files are about 20-26Mb per channel,

More information

Point Calibration. July 3, 2012

Point Calibration. July 3, 2012 Point Calibration July 3, 2012 The purpose of the Point Calibration process is to generate a map of voltages (for galvos) or motor positions of the pointing device to the voltages or pixels of the reference

More information

Leica SP8 Resonant Confocal. Quick-Start Guide

Leica SP8 Resonant Confocal. Quick-Start Guide Leica SP8 Resonant Confocal Quick-Start Guide Contents Start-up Preparing for Imaging Part 1 On the scope Part 2 Software interface Part 3 Heat & CO2 incubation Part 4 Other hardware options Shut-down

More information

Confocal imaging on the Leica TCS SP8. 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software:

Confocal imaging on the Leica TCS SP8. 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software: Confocal imaging on the Leica TCS SP8 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software: 4) Do not touch the microscope while the software is initializing. Choose your options: Turn

More information

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017.

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017. Nikon AZ100 Laser Scanning Macro Confocal Microscope Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin May 2017 Contents 1 Introduction 2 2 Hardware - Startup 2 3 Software/Operation 4 3.1 Multidimensional

More information

Leica Sp5 II Confocal User Guide

Leica Sp5 II Confocal User Guide Leica Sp5 II Confocal User Guide Turning on the Confocal System (instructions are posted in the room) 1. Turn on Laser Power Button 2. Turn Key to On position 3. Turn on Scanner Power Button 4. Turn on

More information

Imaging Beyond the Basics: Optimizing Settings on the Leica SP8 Confocal

Imaging Beyond the Basics: Optimizing Settings on the Leica SP8 Confocal Imaging Beyond the Basics: Optimizing Settings on the Leica SP8 Confocal Todays Goal: Introduce some additional functionalities of the Leica SP8 confocal HyD vs. PMT detectors Dye Assistant Scanning By

More information

(Quantitative Imaging for) Colocalisation Analysis

(Quantitative Imaging for) Colocalisation Analysis (Quantitative Imaging for) Colocalisation Analysis or Why Colour Merge / Overlay Images are EVIL! Special course for DIGS-BB PhD program What is an Image anyway..? An image is a representation of reality

More information

3 Choose the Channels button and set the Channel Settings. Set the Pinhole to 1 Airy unit.

3 Choose the Channels button and set the Channel Settings. Set the Pinhole to 1 Airy unit. 1 Set up the confocal light path for imaging a green dye (e.g. Alexa488-EGFP). For example, under the Configuration Control window the light path could be set up as shown here using the 488 nm LASER (found

More information

MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL

MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL START-UP On the Switchbox, turn both black switches to the ON position. Wait for the microscope to boot up completely (watch the screen on the side of the microscope).

More information

LSM 710 Confocal Microscope Standard Operation Protocol

LSM 710 Confocal Microscope Standard Operation Protocol LSM 710 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Switch on Main power switch 2. Switch on System / PC power button 3. Switch on Components power button 4.

More information

Scanning and Image Processing -by Steve Clough

Scanning and Image Processing -by Steve Clough Scanning and Image Processing -by Steve Clough cdna microarrays use two dyes with well separated emission spectra such as Cy3 and Cy5 to allow direct comparisons on single slide GSI Lumonics Ratio of Expression

More information

NPTEL VIDEO COURSE PROTEOMICS PROF. SANJEEVA SRIVASTAVA

NPTEL VIDEO COURSE PROTEOMICS PROF. SANJEEVA SRIVASTAVA LECTURE-31 MICROARRAY WORK-FLOW: IMAGE SACNNING AND DATA PROCESSING TRANSCRIPT Welcome to the proteomics course. In today s lecture we will talk about microarray work-flow the image scanning and processing.

More information

Supplemental Method Information Zeiss LSM710

Supplemental Method Information Zeiss LSM710 Supplemental Method Information Zeiss LSM710 1 Under the Light Path window set up the confocal for imaging a green dye (Alexa488-EGFP). For example, set up the light path as shown here using the 488 nm

More information

Topics. - How to calibrate the LSM scanner. - How to clean the microscope. - How to adjust the pinhole alignment. - How to adjust the Collimator

Topics. - How to calibrate the LSM scanner. - How to clean the microscope. - How to adjust the pinhole alignment. - How to adjust the Collimator Topics - How to calibrate the LSM scanner - How to measure the PSF - How to clean the microscope - How to adjust the pinhole alignment - How to adjust the Collimator How to calibrate the LSM scanner The

More information

LSM 780 Confocal Microscope Standard Operation Protocol

LSM 780 Confocal Microscope Standard Operation Protocol LSM 780 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Sign on log sheet according to Actual start time 2. Check Compressed Air supply for the air table 3. Switch

More information

LSM 510 Meta Training Notes

LSM 510 Meta Training Notes LSM 510 Meta Training Notes Turning on the system Turn on X-Cite power supply. This supplies light for epifluorescence for viewing your samples through the microscope. Turn on the remote control switch.

More information

Zeiss LSM 880 Protocol

Zeiss LSM 880 Protocol Zeiss LSM 880 Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.

More information

LSM 510 Training Notes

LSM 510 Training Notes LSM 510 Training Notes Turning on the system Turn on the arc lamp, found on the bench top left of the microscope. This supplies light for epifluorescence for viewing your samples through the microscope.

More information

TRAINING MANUAL. Olympus FV1000

TRAINING MANUAL. Olympus FV1000 TRAINING MANUAL Olympus FV1000 September 2014 TABLE OF CONTENTS A. Start-Up Procedure... 1 B. Visual Observation under the Microscope... 1 C. Image Acquisition... 4 A brief Overview of the Settings...

More information

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling Zeiss LSM 510 Confocor III Training Notes Center for Cell Analysis & Modeling Confocor 3 Start Up Go to System Module Turn on Main Switch, System/ PC, and Components Switches Do you need the arc lamp?

More information

Zeiss 780 Training Notes

Zeiss 780 Training Notes Zeiss 780 Training Notes Turn on Main Switch, System PC and Components Switches 780 Start up sequence Do you need the argon laser (458, 488, 514 nm lines)? Yes Turn on the laser s main power switch and

More information

Bruker Dimension Icon AFM Quick User s Guide

Bruker Dimension Icon AFM Quick User s Guide Bruker Dimension Icon AFM Quick User s Guide March 3, 2015 GLA Contacts Jingjing Jiang (jjiang2@caltech.edu 626-616-6357) Xinghao Zhou (xzzhou@caltech.edu 626-375-0855) Bruker Tech Support (AFMSupport@bruker-nano.com

More information

Cell Biology and Bioimaging Core

Cell Biology and Bioimaging Core Cell Biology and Bioimaging Core Leica TCS SP5 Operating Instructions Starting up the instrument 1. First, log in the log book located on the confocal desk. Include your name, your lab s PI, an account

More information

The Zeiss AiryScan System, Confocal Four.

The Zeiss AiryScan System, Confocal Four. The Zeiss AiryScan System, Confocal Four. Overview. The Zeiss AiryScan module is a segmented, radially stacked GaASP detector and collector system designed to subsample the airy disk of a point emission

More information

Zeiss 880 Training Notes Zen 2.3

Zeiss 880 Training Notes Zen 2.3 Zeiss 880 Training Notes Zen 2.3 1 Turn on the HXP 120V Lamp 2 Turn on Main Power Switch Turn on the Systems PC Switch Turn on the Components Switch. 3 4 5 Turn on the PC and log into your account. Start

More information

Things to check before start-up.

Things to check before start-up. Byeong Cha Page 1 11/24/2009 Manual for Leica SP2 Confocal Microscope Enter you name, the date, the time, and the account number in the user log book. Things to check before start-up. Make sure that your

More information

Axioscan - Startup. 1. Turn on the Axioscan (button to the left) and turn on the computer. 2. Log on and start the ZEN Blue software from the desktop

Axioscan - Startup. 1. Turn on the Axioscan (button to the left) and turn on the computer. 2. Log on and start the ZEN Blue software from the desktop Axioscan - Startup 1. Turn on the Axioscan (button to the left) and turn on the computer 2. Log on and start the ZEN Blue software from the desktop 3. Press ZEN slidescan and Start System 4. Start by changing

More information

Practical work no. 3: Confocal Live Cell Microscopy

Practical work no. 3: Confocal Live Cell Microscopy Practical work no. 3: Confocal Live Cell Microscopy Course Instructor: Mikko Liljeström (MIU) 1 Background Confocal microscopy: The main idea behind confocality is that it suppresses the signal outside

More information

Bruker Dimension Icon AFM Quick User s Guide

Bruker Dimension Icon AFM Quick User s Guide Bruker Dimension Icon AFM Quick User s Guide August 8 2014 GLA Contacts Jingjing Jiang (jjiang2@caltech.edu 626-616-6357) Xinghao Zhou (xzzhou@caltech.edu 626-375-0855) Bruker Tech Support (AFMSupport@bruker-nano.com

More information

1 Co Localization and Working flow with the lsm700

1 Co Localization and Working flow with the lsm700 1 Co Localization and Working flow with the lsm700 Samples -1 slide = mousse intestine, Dapi / Ki 67 with Cy3/ BrDU with alexa 488. -1 slide = mousse intestine, Dapi / Ki 67 with Cy3/ no BrDU (but with

More information

Mid_Term_Review_PhotoShop_Design Test B Name

Mid_Term_Review_PhotoShop_Design Test B Name Mid_Term_Review_PhotoShop_Design Test B Name Multiple Choice Identify the choice that best completes the statement or answers the question. 1. Photoshop uses a mathematical process called when it changes

More information

Bi/BE 227 Winter Assignment #3. Adding the third dimension: 3D Confocal Imaging

Bi/BE 227 Winter Assignment #3. Adding the third dimension: 3D Confocal Imaging Bi/BE 227 Winter 2016 Assignment #3 Adding the third dimension: 3D Confocal Imaging Schedule: Jan 20: Assignment Jan 20-Feb 8: Work on assignment Feb 10: Student PowerPoint presentations. Goals for this

More information

Spatial intensity distribution analysis Matlab user guide

Spatial intensity distribution analysis Matlab user guide Spatial intensity distribution analysis Matlab user guide August 2011 Guide on how to use the SpIDA graphical user interface. This little tutorial provides a step by step tutorial explaining how to get

More information

Product Information. Introduction

Product Information. Introduction Microarray Scanner Calibration Slide To quantitatively analyze scanners performance and output, adjust and fine-tune scanners, and perform comparative analysis for multiple scanner units. To verify scanners

More information

Nikon A1Rsi Confocal Start-Up Sequence

Nikon A1Rsi Confocal Start-Up Sequence 1. Turn the key on the Nikon LUN-V Laser Launch. Nikon A1Rsi Confocal Start-Up Sequence 2. Press the button the left side of the A1Rsi Controller unit. 3. Turn on the power strip underneath the microscope.

More information

Image Viewing. with ImageScope

Image Viewing. with ImageScope Image Viewing with ImageScope ImageScope Components Use ImageScope to View These File Types: ScanScope Virtual Slides.SVS files created when the ScanScope scanner scans glass microscope slides. JPEG files

More information

Supplementary Figure S1: Schematic view of the confocal laser scanning STED microscope used for STED-RICS. For a detailed description of our

Supplementary Figure S1: Schematic view of the confocal laser scanning STED microscope used for STED-RICS. For a detailed description of our Supplementary Figure S1: Schematic view of the confocal laser scanning STED microscope used for STED-RICS. For a detailed description of our home-built STED microscope used for the STED-RICS experiments,

More information

Leica SPEII confocal microscope. Short Manual

Leica SPEII confocal microscope. Short Manual Leica SPEII confocal microscope Short Manual Switching ON sequence: 1. Turn on the Workstation under the bench (top, far right). 2. Turn on the Supply Unit - Laser box (big green switch first and then

More information

1 Set up the confocal light path for imaging a green dye (Alexa488-EGFP). For example, the

1 Set up the confocal light path for imaging a green dye (Alexa488-EGFP). For example, the 1 Set up the confocal light path for imaging a green dye (Alexa488-EGFP). For example, the light path as shown here using the 488 nm LASER (Laser Unit 1) reflecting off of the 405/488 nm Dichroic mirror

More information

Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope

Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope ZEN 2009 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn Chameleon TiS laser key from Standby

More information

Products - Microarray Scanners - Laser Scanners - InnoScan 900 Series and MAPIX Software

Products - Microarray Scanners - Laser Scanners - InnoScan 900 Series and MAPIX Software Products - Microarray Scanners - Laser Scanners - InnoScan 900 Series and MAPIX Software Arrayit offers the world s only next generation microarray scanning technology, with proprietary rotary motion control,

More information

User manual for Olympus SD-OSR spinning disk confocal microscope

User manual for Olympus SD-OSR spinning disk confocal microscope User manual for Olympus SD-OSR spinning disk confocal microscope Ved Prakash, PhD. Research imaging specialist Imaging & histology core University of Texas, Dallas ved.prakash@utdallas.edu Once you open

More information

GenePix Application Note

GenePix Application Note GenePix Application Note Biological Relevance of GenePix Results Shawn Handran, Ph.D. and Jack Y. Zhai, Ph.D. Axon Instruments, Inc. 3280 Whipple Road, Union City, CA 94587 Last Updated: Aug 22, 2003.

More information

Commercial Art 1 Photoshop Study Guide. 8) How is on-screen image resolution measured? PPI - Pixels Per Inch

Commercial Art 1 Photoshop Study Guide. 8) How is on-screen image resolution measured? PPI - Pixels Per Inch Commercial Art 1 Photoshop Study Guide To help prepare you for the Photoshop test, be sure you can answer the following questions: 1) What are the three things should you do when you first open a Photoshop

More information

Zeiss LSM 780 Protocol

Zeiss LSM 780 Protocol Zeiss LSM 780 Protocol 1) System Startup F Please note the sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.

More information

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope AIM 4.2 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Verify that main power switches on the

More information

Quick Start Guide. Leica SP5 X

Quick Start Guide. Leica SP5 X Quick Start Guide Leica SP5 X Please note: Some of the information in this guide was taken from Leica Microsystems Leica TCS SP5 LAS AF Guide for New Users. This work is licensed under the Creative Commons

More information

Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each

Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each lens with a 1 Airy unit pinhole setting. Many water lenses

More information

ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL

ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL START THE SYSTEM... 2 START ZEN SOFTWARE... 3 SET THE TEMPERATURE AND THE CO2 CONTROLLERS... OBSERVATION AT OCULARS... 5 STATIF PRESENTATION... 6 ACQUIRE ONE

More information

Quick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s

Quick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s LSM 5 MP, LSM 510 and LSM 510 META M i c r o s c o p y f r o m C a r l Z e i s s Quick Guide Laser Scanning Microscopes LSM Software ZEN 2007 August 2007 We make it visible. Contents Page Contents... 1

More information

LEICA TCS SP5 AOBS TANDEM USER MANUAL

LEICA TCS SP5 AOBS TANDEM USER MANUAL LEICA TCS SP5 AOBS TANDEM USER MANUAL STARTING THE SYSTEM...2 THE LAS AF SOFTWARE...3 THE «ACQUIRE» MENU...5 CHOOSE AND CREATE A SETTING...6 THE CONTROL PANEL...8 THE DMI6000B MICROSCOPE...10 ACQUIRE ONE

More information

BIM Toolbox. User Guide. Version: Copyright 2017 Computer and Design Services Ltd GLOBAL CONSTRUCTION SOFTWARE AND SERVICES

BIM Toolbox. User Guide. Version: Copyright 2017 Computer and Design Services Ltd GLOBAL CONSTRUCTION SOFTWARE AND SERVICES BIM Toolbox User Guide Version: 2018.0 Copyright 2017 Computer and Design Services Ltd GLOBAL CONSTRUCTION SOFTWARE AND SERVICES Contents Introduction... 1 Create a new project... 2 Trace around a site

More information

Nikon Eclipse Ti A1-A Confocal Operating Manual. Start-up. Microscope

Nikon Eclipse Ti A1-A Confocal Operating Manual. Start-up. Microscope Nikon Eclipse Ti A1-A Confocal Operating Manual Start-up 1. Turn on Excite Fluorescent light power supply- metal halide. a. Cool down as for mercury bulb b. Wheel closed liquid light guide 2. Turn on power

More information

Usermanual for Leica SP8 confocal

Usermanual for Leica SP8 confocal Usermanual for Leica SP8 confocal Contact information: hege.dale@uib.no & endy.spriet@uib.no 1 Table of content Important information 3 Start up procedure 4 Shut down procedure 5 Operating the DMI 8 microscope

More information

Title: Leica SP5 Confocal User Manual

Title: Leica SP5 Confocal User Manual Title: Leica SP5 Confocal User Manual Date of first issue: 23/10/2015 Date of review: Version: Admin For assistance or to report an issue Office: CG07 or 05 Email: Igmm-imaginghelpdesk@igmm.ed.ac.uk Website:

More information

Jeol JEM Responsible personell: Endy ( ) Online booking is compulsory!

Jeol JEM Responsible personell: Endy ( ) Online booking is compulsory! Jeol JEM 1230 Responsible personell: Endy (45279377) Online booking is compulsory! After training you will have access to working alone on the instrument. All insertion of samples is done by responsible

More information

Single Slit Diffraction

Single Slit Diffraction PC1142 Physics II Single Slit Diffraction 1 Objectives Investigate the single-slit diffraction pattern produced by monochromatic laser light. Determine the wavelength of the laser light from measurements

More information

Leica SP8 TCS Users Manual

Leica SP8 TCS Users Manual Leica SP8 TCS Users Manual Follow the procedure for start up and log on as posted in the lab. Please log on with your account only and do not share your password with anyone. We track and confirm usage

More information

Horiba Jobin-Yvon LabRam Raman Confocal Microscope (GERB 120)

Horiba Jobin-Yvon LabRam Raman Confocal Microscope (GERB 120) Horiba Jobin-Yvon LabRam Raman Confocal Microscope (GERB 120) Please contact Dr. Amanda Henkes for training requests and assistance: 979-862-5959, amandahenkes@tamu.edu Hardware LN 2 FTIR FTIR camera 1

More information

Nikon C1si Spectral Laser Scanning Confocal Microscope. User Guide

Nikon C1si Spectral Laser Scanning Confocal Microscope. User Guide Nikon C1si Spectral Laser Scanning Confocal Microscope User Guide Contents: C1Si Turn-On/ShutDown Procedures... 2 Overview... 4 Setup for epi-illumination to view through the eyepieces:... 5 Setup for

More information

University Libraries ScanPro 3000 Microfilm Scanner

University Libraries ScanPro 3000 Microfilm Scanner University Libraries ScanPro 3000 Microfilm Scanner Help Guide Table of Contents Getting Started 3 Loading the Film 4-5 Viewing Your Film 6-7 Motorized Roll Film Control 6 Crop Box 7 Using the Toolbar

More information

Internal Medicine Imaging Core Emory University Department of Medicine

Internal Medicine Imaging Core Emory University Department of Medicine Internal Medicine Imaging Core Emory University Department of Medicine 1 OPERATION OF THE ZEISS LSM 510 META YOU MUST SIGN UP TO USE THE MICROSCOPE OR COMPUTER EVERY TIME NO EXCEPTIONS Before attempting

More information

Leica SP8 TCS Users Manual

Leica SP8 TCS Users Manual Version : 07/08/0 Leica SP8 TCS Users Manual Start up:. Turn the PC Microscope, Scanner Power, Laser Power, and the Laser Emission key to on (bottom right of desk).. Turn on the fluorescent lamp (top left

More information

Comparing FCS and FRAP as methodologies for calculating diffusion

Comparing FCS and FRAP as methodologies for calculating diffusion Bi/BE 227 Winter 2018 Assignment #4 Comparing FCS and FRAP as methodologies for calculating diffusion Schedule: Jan 29: Assignment Jan 29-Feb 14: Work on assignment Feb 14: Student PowerPoint presentations.

More information

Digital Portable Overhead Document Camera LV-1010

Digital Portable Overhead Document Camera LV-1010 Digital Portable Overhead Document Camera LV-1010 Instruction Manual 1 Content I Product Introduction 1.1 Product appearance..3 1.2 Main functions and features of the product.3 1.3 Production specifications.4

More information

Micro-Image Capture 8 Installation Instructions & User Guide

Micro-Image Capture 8 Installation Instructions & User Guide Micro-Image Capture 8 Installation Instructions & User Guide Software installation: Micro-Image Capture Software 1. Load Micro-Image Capture software CD onto host PC. Auto Run should start driver/software

More information

Standard Operating Procedure of Atomic Force Microscope (Anasys afm+)

Standard Operating Procedure of Atomic Force Microscope (Anasys afm+) Standard Operating Procedure of Atomic Force Microscope (Anasys afm+) The Anasys Instruments afm+ system incorporates an Atomic Force Microscope which can scan the sample in the contact mode and generate

More information

Nasmyth Ultraview Vox User Protocol

Nasmyth Ultraview Vox User Protocol Nasmyth Ultraview Vox User Protocol Switch on all wall sockets labelled Nasmyth, switch camera on (power supply located on table behind monitor), switch on laser switch in laser rack, switch computer on

More information

XXXX - ILLUSTRATING FROM SKETCHES IN PHOTOSHOP 1 N/08/08

XXXX - ILLUSTRATING FROM SKETCHES IN PHOTOSHOP 1 N/08/08 INTRODUCTION TO GRAPHICS Illustrating from sketches in Photoshop Information Sheet No. XXXX Creating illustrations from existing photography is an excellent method to create bold and sharp works of art

More information

Overview. About other software. Administrator password. 58. UltraVIEW VoX Getting Started Guide

Overview. About other software. Administrator password. 58. UltraVIEW VoX Getting Started Guide Operation 58. UltraVIEW VoX Getting Started Guide Overview This chapter outlines the basic methods used to operate the UltraVIEW VoX system. About other software Volocity places great demands on the computer

More information

Developed by BioDiscovery, Inc. DualChip evaluation software User Manual Version 1.1

Developed by BioDiscovery, Inc. DualChip evaluation software User Manual Version 1.1 Developed by BioDiscovery, Inc. DualChip evaluation software User Manual Version 1.1 1 Table of contents 1. INTRODUCTION...3 2. SCOPE OF DELIVERY...4 3. INSTALLATION PROCEDURES...5 3.1. SYSTEM REQUIREMENTS...

More information

Manual. Cell Border Tracker. Jochen Seebach Institut für Anatomie und Vaskuläre Biologie, WWU Münster

Manual. Cell Border Tracker. Jochen Seebach Institut für Anatomie und Vaskuläre Biologie, WWU Münster Manual Cell Border Tracker Jochen Seebach Institut für Anatomie und Vaskuläre Biologie, WWU Münster 1 Cell Border Tracker 1. System Requirements The software requires Windows XP operating system or higher

More information

Microscopy from Carl Zeiss

Microscopy from Carl Zeiss Microscopy from Carl Zeiss Contents Page Contents... 1 Introduction... 1 Starting the System... 2 Introduction to ZEN Efficient Navigation... 5 Setting up the microscope... 10 Configuring the beam path

More information

Registering and Distorting Images

Registering and Distorting Images Written by Jonathan Sachs Copyright 1999-2000 Digital Light & Color Registering and Distorting Images 1 Introduction to Image Registration The process of getting two different photographs of the same subject

More information

IncuCyte ZOOM Fluorescent Processing Overview

IncuCyte ZOOM Fluorescent Processing Overview IncuCyte ZOOM Fluorescent Processing Overview The IncuCyte ZOOM offers users the ability to acquire HD phase as well as dual wavelength fluorescent images of living cells producing multiplexed data that

More information

A guide to SalsaJ. This guide gives step-by-step instructions on how to use SalsaJ to carry out basic data analysis on astronomical data files.

A guide to SalsaJ. This guide gives step-by-step instructions on how to use SalsaJ to carry out basic data analysis on astronomical data files. A guide to SalsaJ SalsaJ is free, student-friendly software developed originally for the European Hands- On Universe (EU-HOU) project. It is designed to be easy to install and use. It allows students to

More information

SOP: EDAX Eagle III Microspot XRF

SOP: EDAX Eagle III Microspot XRF SOP: EDAX Eagle III Microspot XRF Page 1 of 6 SOP: EDAX Eagle III Microspot XRF 1. Scope 1.1 This document describes the standard operating procedure (SOP) for the EDAX Eagle III Microspot XRF. This X-ray

More information

PHOTOSHOP PUZZLE EFFECT

PHOTOSHOP PUZZLE EFFECT PHOTOSHOP PUZZLE EFFECT In this Photoshop tutorial, we re going to look at how to easily create a puzzle effect, allowing us to turn any photo into a jigsaw puzzle! Or at least, we ll be creating the illusion

More information

Automatic gene expression estimation from microarray images. Daniel O. Dantas Adviser: : Junior Barrera

Automatic gene expression estimation from microarray images. Daniel O. Dantas Adviser: : Junior Barrera Automatic gene expression estimation from microarray images Daniel O. Dantas Adviser: : Junior Barrera IME-USP Summary Introduction Problem definition Solution strategy Image segmentation Signal estimation

More information

Operating Instructions for Zeiss LSM 510

Operating Instructions for Zeiss LSM 510 Operating Instructions for Zeiss LSM 510 Location: GNL 6.312q (BSL3) Questions? Contact: Maxim Ivannikov, maivanni@utmb.edu 1 Attend A Complementary Training Before Using The Microscope All future users

More information

For customers in USA This device complies with Part 15 of the FCC rules. Operation is subject to the following two conditions:

For customers in USA This device complies with Part 15 of the FCC rules. Operation is subject to the following two conditions: User manual For customers in North and South America For customers in USA This device complies with Part 15 of the FCC rules. Operation is subject to the following two conditions: (1) This device may not

More information

EVOS M5000 Imaging System

EVOS M5000 Imaging System EVOS M5000 Imaging System Pub. No. MAN0017765 Doc. Part No. 710209 Rev. A.0 This document is intended as a benchtop reference for the users of the EVOS M5000 Imaging System (Cat. No. AMF5000). For detailed

More information

MEDIA CONVERSION PROJECT: DIGITIZING HISTORIC PHOTOS

MEDIA CONVERSION PROJECT: DIGITIZING HISTORIC PHOTOS MEDIA CONVERSION PROJECT: DIGITIZING HISTORIC PHOTOS What you need to know: Basic understanding of a computer What you need: Computer Epson Perfection V550 Scanner (available in the Southglenn and Smoky

More information

INGENIUS 3 LOW COST, HIGH PERFORMANCE GEL DOCUMENTATION AND ANALYSIS

INGENIUS 3 LOW COST, HIGH PERFORMANCE GEL DOCUMENTATION AND ANALYSIS INGENIUS 3 LOW COST, HIGH PERFORMANCE GEL DOCUMENTATION AND ANALYSIS The InGenius 3 uses a high performance 3m pixel camera. The darkroom assembly is easily connected to a PC. GeneSys image acquisition

More information

Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev. Microscopy course, Michmoret Dec 2005

Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev. Microscopy course, Michmoret Dec 2005 Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev Why use confocal microscopy? Principles of the laser scanning confocal microscope. Image resolution. Manipulating the

More information

Customized Foam for Tools

Customized Foam for Tools Table of contents Make sure that you have the latest version before using this document. o o o o o o o Overview of services offered and steps to follow (p.3) 1. Service : Cutting of foam for tools 2. Service

More information

Experiment: P34 Resonance Modes 1 Resonance Modes of a Stretched String (Power Amplifier, Voltage Sensor)

Experiment: P34 Resonance Modes 1 Resonance Modes of a Stretched String (Power Amplifier, Voltage Sensor) PASCO scientific Vol. 2 Physics Lab Manual: P34-1 Experiment: P34 Resonance Modes 1 Resonance Modes of a Stretched String (Power Amplifier, Voltage Sensor) Concept Time SW Interface Macintosh file Windows

More information

Introduction to Photoshop Elements

Introduction to Photoshop Elements John W. Jacobs Technology Center 450 Exton Square Parkway Exton, PA 19341 610.280.2666 ccljtc@ccls.org www.ccls.org Facebook.com/ChesterCountyLibrary Introduction to Photoshop Elements Chester County Library

More information

Guide to Confocal 5. Starting session

Guide to Confocal 5. Starting session Guide to Confocal 5 Remember that when booking and before starting session you can check for any problems at https://www.bris.ac.uk/biochemistry/uobonly/cif/index.html Starting session Switch on microscope

More information

Opterra II Multipoint Scanning Confocal Microscope. Innovation with Integrity

Opterra II Multipoint Scanning Confocal Microscope. Innovation with Integrity Opterra II Multipoint Scanning Confocal Microscope Enabling 4D Live-Cell Fluorescence Imaging through Speed, Sensitivity, Viability and Simplicity Innovation with Integrity Fluorescence Microscopy The

More information

LAB 11 Color and Light

LAB 11 Color and Light Cabrillo College Name LAB 11 Color and Light Bring colored pencils or crayons to lab if you already have some. What to learn and explore In the previous lab, we discovered that some sounds are simple,

More information

INGENIUS 3. Low cost, high performance gel documentation and analysis

INGENIUS 3. Low cost, high performance gel documentation and analysis INGENIUS 3 Low cost, high performance gel documentation and analysis INGENIUS 3 When simplicity and budget matter. The InGenius 3 gel documentation and analysis system is compact, easy to use and offers

More information