Essentials of Digital Imaging

Size: px
Start display at page:

Download "Essentials of Digital Imaging"

Transcription

1 Essentials of Digital Imaging Module 4 Transcript 2016 ASRT. All rights reserved.

2 Essentials of Digital Imaging Module 4 Image Analysis 1. ASRT Animation 2. Welcome Welcome to Essentials of Digital Imaging: Module 4 Image Analysis. 3. License Agreement 4. Objectives After completing this module, you will be able to: Evaluate the effects of exposure changes on the appearance of a digital image. Evaluate the appearance of a digitally acquired image to assess quantum mottle and saturation. Discuss the relationship between exposure field recognition and the appearance of a digital image. Recognize motion on a digital image. Locate different types of artifacts on a digital image and state their probable cause. 5. Evaluating Digital Images The changes in exposure that affect density and contrast for film do not have as much of an effect on the appearance of digital images. This is due to digital processing that creates the image for display. However there are a number of factors that can affect the brightness of the image and also the contrast. The overall appearance of a digital image can be affected by a number of factors such as under exposure, or quantum mottle, over exposure, or saturation, scatter radiation, grid cut, and motion from the patient. Additional factors that can impact image quality are exposure field recognition errors, an unexpected attenuator in the field of view, or the menu selection for the procedure. 6. Quantum Mottle Quantum mottle, more often referred to as noise, in a digital image can be caused by multiple factors that affect the appearance of the image. Noise results in a mottled, or grainy, appearance caused by random variations of signal intensity on the image. Noise caused by low exposure to the image receptor means a low number of x-ray photons strike the receptor resulting in the phenomena known as quantum mottle. Basically, quantum mottle in digital images is caused by insufficient signal strength. Remember that the distribution of photons in the x-ray beam is not uniform. On occasion, when low levels of radiation are used to produce an image, the captured data are the random pattern of x-ray photons within the beam along with patient anatomical information. In other words, there are not enough x-ray photons striking the receptor. Oftentimes the only way to correct for the quantum mottle is to produce a new image that has an increased number of x-ray photons striking the image receptor. When the repeated image is produced with a greater number of photons striking the receptor it results in an image of better diagnostic quality. 7. Evaluating Noise In addition to quantum mottle the exposure index for the image indicates that the image was obtained with a low level exposure. The only way to reduce the amount of noise present in the image would be to

3 increase the amount of x-ray photons striking the image receptor which increases the signal strength available to create the image. The impact of quantum mottle is most noticeable when assessing the low contrast resolution of a radiograph. The impact on visibility from quantum mottle of low contrast structures is easily demonstrated using a low contrast detail phantom. As quantum mottle increases the ability to visualize subtle differences in differential attenuation is decreased. The circles in the phantom represent variations in tissue attenuation. 8. Over Exposure A misconception about digital imaging systems is that the system has the capability of correcting for errors involving receptor exposure. Digital systems are designed to analyze the signal extracted from the image receptor and to display that image with a specific image appearance. The image will fail to display properly due to the over exposure in instances where the signal strength approaches or exceeds the system s dynamic range. Saturation is the result of over exposure to the point where the dynamic range of the digital system is no longer capable of accurately reproducing the image data for display. The point at which saturation occurs is dependent upon the digital imaging system being used. An example of this is a thoracic spine image that demonstrates saturation in the lung field. The presence of saturation in the image results in an inability to visualize the lung field within the over exposed area regardless of manipulating image brightness. 9. Saturation The elbow radiograph on the left represents an acceptable level of exposure in the image. The image on the right represents a level of exposure that has resulted in saturation. The level of receptor exposure beneath the soft tissue portions of the saturated image exceeded the dynamic range of the pixel values. The outcome is that the soft tissues surrounding the bony anatomy are not able to be visualized. To correct for saturation the technologists must reduce the level of exposure striking the image receptor. The lateral cervical spine image on the left shows a level of exposure with a signal strength that exceeded the dynamic range of the pixel values in soft tissue areas and some bony structures. Attempts to adjust the brightness of the image in order to demonstrate soft tissue and bony detail were unsuccessful. The image must be repeated at a lower level of exposure in order to produce a satisfactory image. The repeat image was performed at a lower level of exposure which eliminated saturation and allows for the visualization of soft tissue and bony detail. 10. Scatter Radiation Scatter radiation does not contribute useful information to the radiographic image. In situations where scatter makes up a significant portion of the radiation striking the image receptor, there is a decrease in the image contrast. An example of scatter affecting image quality is observable on an abdominal radiograph where a significant portion of radiation striking the image receptor seemingly is composed of scattered photons. The image was produced without using a grid which clearly demonstrates the loss of visibility because of scatter radiation hitting the receptor. When a grid was used during the repeat examination, much of the scatter was attenuated. The visibility of structures is significantly improved by reducing the amount of scatter radiation striking the image receptor.

4 11. Grid Use Using a grid for certain radiographic examinations can aid the technologist in improving image quality. An abdominal radiograph produced without the use of a grid demonstrates decreased image contrast resulting in poor visibility. After attempts to manipulate the brightness and contrast of the image it still was not possible to achieve an optimal level of visibility of structures that is achieved by producing a radiograph on the same patient with a grid. Regardless of the imaging system used the presence of significant scatter will result in worsened image quality. 12. Grid Cutoff Proper alignment with the central ray can be difficult when performing a radiographic examination using a grid. The grid is designed to attenuate scatter radiation to improve image quality. Failure to properly align the central ray with the grid also will result in the absorption of the portion of the primary beam that represents anatomical information. The effect on the finished radiograph is an image that is displayed with a significant reduction in visibility referred to as grid cutoff. More specifically, grid cutoff occurs when the beam is not centered transversely to the grid. In addition to poor image display, in some circumstances the exposure to the image receptor is low enough to produce visible quantum mottle because of the insufficient signal striking the receptor. The presence of grid cutoff often is accompanied by an exposure indicator that suggests a low level of radiation exposure was used to produce the image. Additionally, grid cutoff might occur from an off level grid error or when using an improper SID or an upside down focused grid. 13. Off Center Grid Another complication in medical imaging associated with improper transverse alignment of the central ray and the grid is the impact of grid cutoff on image contrast. It might appear that an image is degraded by significant scatter radiation striking the image receptor. In reality, the exposure indicator proves that a low level of exposure was used to produce the image and the presence of quantum mottle tells the radiographer that poor image contrast is a result of grid cutoff. Image quality is easily improved by accurately aligning the central ray to the grid. 14. Inverted Focused Grid When a focused grid is inverted, grid cutoff is visible along the edges of anatomy on the finished radiograph. The diminished image quality results because the lead strips of the grid are no longer parallel to the divergent primary x-ray beam. 15. Knowledge Check 16. Knowledge Check 17. Improper SID Grid cutoff also can occur as a result of a source-to-image receptor distance, or SID, that is too close to the grid. The cutoff is demonstrated along the lateral margins of the radiograph. In addition to the presence of grid cutoff, the exposure index likely will indicate a low level of exposure was used for the examination even if the technologist has selected proper exposure factors. Sometimes two different types of grid cutoff can be visualized on a single radiograph. An example is an elbow radiograph demonstrating misalignment between the x-ray tube and the grid that is performed using a SID that is too short. The system was a cassette-less DR unit with a built in grid. The technologist

5 performing the examination did not ensure that the beam and grid were properly aligned before making the exposure and the incorrect SID was displayed on the DICOM header of the image. 18. Exposure Field Recognition Recognition of the edges of the exposure field play an important role in ensuring an image will display correctly. An exposure field recognition error occurs when the computer fails to identify collimated borders of a radiograph. When the computer can t find the borders, it causes a rescaling error, which often affects the brightness and contrast of an image. An exposure field recognition error during image processing will often result in an image displayed with inadequate brightness. An example of when this error can occur is when 2 exposures can be made on 1 image receptor. The individual exposure fields should be recognized during image processing although; sometimes the analysis of the image data fails to recognize the exposure field borders. Details about the examination suggest that the radiographic exposure was accurate for the examination evident by the absence of quantum mottle. Mottle, or noise, would normally be present if a low level of exposure produced the image. This validates that the diminished image quality is the result of an exposure field recognition error. In this instance, the area within the white box determined the level of image brightness as well as the exposure index. The lack of exposure within the region of interest, or ROI, led to a rescaling error that produced a low level of brightness and the exposure indicator reflected a low level of exposure even though the level of exposure was adequate. 19. Recognition Errors Some radiographic exams can present a unique challenge in attempting to produce diagnostic quality images. An example is a cross table lateral lumbar spine. Locating the spine and properly centering the anatomy of interest to the image receptor is not difficult to achieve but is a skill that is acquired after much practice. Improper alignment of the body part with the central ray and image receptor can result in an exposure field recognition error. An exposure field that is not accurately localized during image processing produces a final image that fails to display the proper brightness even though the appropriate level of radiation exposure was used. The system has the ability to reprocess the image data by realigning the region of interest to match the collimation margins. Upon making that adjustment, the image can be significantly improved. 20. Unexpected Attenuator Joint replacement surgery is common and is something that the technologist needs to be aware of before obtaining a radiographic image. Knowing about the surgery beforehand can aid the technologist in selecting the most appropriate technique for the examination. When an unexpected, highly attenuating material is present, it might negatively impact the brightness of the image when it is displayed. Metal rods contribute to an image displayed with inadequate brightness. Rescaling, following the computer s analysis of the image data, can result in a radiograph that is too dark. The rods represent a significant portion of low exposure within the data and therefore the rescaling process adjusts image brightness to take into account the low level of radiation exposure to the area of interest. 21. Menu Selection

6 The menu selected for image processing is critical to displaying the image properly. The menu selected for each unique examination contains specific levels of brightness, contrast, edge enhancement, and equalization. Selecting a menu that is not assigned to the examination being performed because it displays the image better can lead to potential medicolegal concerns. In some circumstances, selecting the incorrect processing menu can provide a more pleasing appearance to the viewer even though it holds changes in image processing that might be unacceptable for the exam. Keep in mind, the change in processing also can produce an exposure indicator that suggests there was an increase in radiation exposure to the patient, when in fact; there is no difference in exposure between the images being displayed. Image processing menus are carefully developed and should only be used for the examinations that they have been assigned to. 22. Image Blur Blurred images are caused by motion or changes in a geometric factor during image production. Motion is the predominant cause of blur in radiographic images. Technologists can minimize geometric blur when selecting technical factors, such as SID, OID, and focal spot size. Blur caused by geometric factors is best assessed using specific test phantoms. Image blur often is caused by patient movement. It is sometimes difficult to immediately recognize motion on radiograph but when using the zoom feature blur from motion becomes apparent when the edges of bony structures of interest are fuzzy or unclear. When motion produces blurring in an image it is often difficult to observe on the technologist workstation without using image zoom. The image zoom feature is essential for assessing the presence of motion in an image. Image zoom enlarges the radiograph making some of the finer details easier to see when motion is otherwise undetected. This function makes any blurred anatomy clearly visible and indicates the need to repeat the examination. 23. Size Distortion Radiographic distortion occurs as a result of changes in the relationship between the primary beam and image receptor. Size distortion is a misrepresentation of the true size of the object due to changes in distance between the tube, body part, and image receptor. An example of size distortion is magnification. The size of the image receptor affects the size of the image display. This change in display size should not be confused with magnification from changes in SID and OID. The two most common factors that are responsible for size distortion because of their effect on the distance between the object of interest and the image receptor are source-to-image receptor distance and object-to-image receptor distance. The Waters method is used to demonstrate the orbits. By tilting the head and placing the patient's chin on the image receptor, a 25- to 37-degree angle is formed between the orbitomeatal line and image receptor. This angulation not only projects the petrous portion of the temporal bones below the inferior border of the orbits, allowing the orbits to be evaluated free of superimposition by the petrous ridges but also brings the orbital margins further from the image receptor resulting in slight magnification of the structures. The combination of these effects makes it possible to better evaluate the orbits for injury such as blow out fractures. 24. Shape Distortion

7 Shape distortion is a misrepresentation of the true shape of an object due to a change in the angular relationship of the tube, body part, and image receptor. Shape distortion causes objects of interest to be elongated or foreshortened. When a patient is positioned for an oblique projection of the hip as opposed to using a standard AP projection the site of injury and the orthopedic hardware can be better evaluated. The wing, or ala, of the ilium appears elongated, the pubis is foreshortened, and the ischium is in profile achieving the goal of demonstrating placement of the surgical pin outside of the joint. Another example of shape distortion is a PA axial projection of the scaphoid using ulnar deviation. The scaphoid is elongated because of the 20o angle of the tube to the part and image receptor. Another option is to instead angle the image receptor 20o using a radiolucent sponge. Size distortion also can be applied to the PA axial scaphoid projection by placing a radiolucent positioning device underneath the wrist to bring the wrist further from the image receptor, increasing the OID. 25. Exposure Index It s important to understand what the exposure index values represent, along with factors that affect its accuracy. If exposure index values are outside the optimal range, however, noise or other problems appear. Factors that can influence the accuracy of the exposure indicator, regardless of the system used, include exposure field recognition, collimation, and unexpected structures in the exposure field. The implementation of the international standardized exposure indicator known as the Exposure Index started in The purpose was to establish a common Exposure Index across manufacturers and vendors of digital x-ray equipment. The end result is that the manufacturers utilize an exposure index that is linearly proportional to the amount of radiation received by the detector. Each department then develops a Target Exposure Index that results in the production of quality images at acceptable dose levels. In addition to the Target Exposure Index, the user of the equipment will be able to determine compliance with the desired value by assessing the Deviation Index for the exposure. The Deviation Index is displayed using a scale of minus 3 to plus 3. A Deviation Index of 0 indicates that the exposure matches the Target Exposure Index. A Deviation Index of +1 represents 25% more exposure than the Target Exposure Index where a Deviation Index of -1 represents 20% less exposure than the Target Exposure Index. Furthermore, a Deviation Index +3 represents 2X more exposure than the Target Exposure Index and a Deviation Index of -3 represents that the exposure was about ½ the Target Exposure Index. 26. Evaluating Exposure Indicator When all the exposure received by the receptor, not just the areas within the collimated borders, is included in calculating the exposure indicator, the exposure index likely will be miscalculated. Often this occurs when 2 radiographs are exposed onto a single image receptor. The area between the 2 images represents a very low exposure level. This area was incorporated in the histogram values and subsequently altered the calculated exposure indicator. The result was a value that indicated an insufficient amount of radiation. This 2-on-1 radiograph of the forearm also demonstrates a faulty exposure indicator calculation. The technique used was acceptable, but the exposure field borders weren't recognized, which decreased contrast and brightness. The histogram is very wide with an increase at the lower end of exposure

8 intensities, where it includes the area between collimated fields. If that area had not been included in producing this image, the display would have been appropriate and the exposure indicator would have been properly calculated. 27. Histogram Analysis Error It is possible to produce a chest radiograph using satisfactory kvp and mas values; however, if a lead apron is included in the exposure field the apron can cause an error in the histogram analysis. The apron is an additional attenuator that produces an extra peak at the low-intensity area of the a priori histogram. As a result, the values result in an exposure indicator miscalculation. The miscalculation suggests that the plate is grossly underexposed, when in fact it isn't. Additionally, a rescaling error occurs that leads to an inappropriate display of image contrast and brightness. Another example of a histogram analysis error occurs when including too much of the abdomen in the exposure field for a portable chest examination. The radiograph is processed using a chest menu that expects a chest-shaped histogram therefore; the calculation of the exposure indicator is inaccurate. The low-level exposure area below the diaphragm is included in the histogram and produces an incorrect exposure indicator, leading the technologist to believe that the image receptor was underexposed. Using the zoom feature on the image allows the technologist to evaluate for noise on the radiograph. If no significant noise is found it suggests that the image was not underexposed. Technologists must make sure that collimation is appropriate when looking at exposure indicators. Checking for additional attenuators within the exposure field and being certain the expected anatomy is shown ensures the histogram values are calculated correctly. 28. Recording Multiple Images Recall that recording multiple images on a single receptor can cause overexposed and underexposed areas. Exposure field margins are altered by overlapping exposure fields in the center of the receptor and nonparallel alignment of exposure field edges. The lookup table uses all the data on the plate, rather than just the image data, to produce the image display. The current recommendation is to record only 1 image per receptor to avoid potential data analysis errors. 29. Knowledge Check 30. Knowledge Check 31. Digital Image Artifacts Artifacts in digital imaging can be grouped by image receptor type. The image receptor is a photostimulable phosphor plate, called a PSP, or flat panel detector with a thin-film transistor, or TFT. Artifacts also might be categorized by their causes or where they occur and can be produced by Image receptors, processing equipment or electronic component issues. In addition, artifacts occur because of mistakes made in patient positioning or preparation. The latter errors are called exposure artifacts. 32. PSP Artifacts Various types of artifacts occur with photostimulable phosphor plate systems. These systems use a phosphor plate, transport mechanism, optical components, and assorted electronic components. The potential for unique artifacts that can be replicated are easily identifiable by disruptions in plate transport, dirty or damaged phosphor plates, dirty or damaged optical components, and malfunctioning

9 electronic devices. However, it is possible that the cause of an intermittently appearing artifact might not be identified until it is replicated. Artifacts associated with the plate and optical components will reduce the intensity of light collected by the optical system. The reduced intensity causes an artifact to appear bright on the image. Scratches, dents, and stains are artifacts on a PSP plate that can severely damage the phosphors. When the plate is scratched or dented, the affected area can t absorb x-rays. If the plate is stained, light emission is reduced or cannot happen. The artifact along the edge of the image is where the phosphor has chipped away from the support base. The white line in the soft tissue is a scratch and the white specs are dust spots on the phosphor. 33. Plate Based Artifacts At first an artifact might be missed. When looking at an image using magnification the artifact should become more apparent. The bright line artifact in the medial aspect of the humeral head was caused by a dirty plate. Many digital artifacts appear bright, or white, on the finished image. Once identified, the cause can be determined. Primary causes of bright artifacts likely are a damaged or dirty plate or damaged or dirty optical components. Plate based artifacts remain localized to the same area of an individual plate. Sometimes artifacts begin to look familiar. This occurs because the same plate is used for multiple radiographs. If technologists don t notice the artifact on the image, and the plate isn t taken out of service to be cleaned, the artifact will continue to appear degrading numerous patient examinations throughout the day. For this reason, it is important that technologists check images for artifacts continuously. 34. Routine PSP Cleaning PSP plates require routine cleaning and inspection by technologists to make sure the plates produce artifact-free images. One reason it s so important to regularly clean PSP plates is that many artifacts that can t be seen on a technologist s workstation are easily seen by the radiologist on his or her workstation. The quality, resolution, brightness, and contrast resolution capabilities of monitors used by radiologists are far superior to the monitors used by technologists. Sometimes the artifact can't be seen on a technologist s workstation unless the image is significantly magnified and altered. However, on a radiologist s workstation, the artifact might be immediately apparent and look like a foreign body. It also is important that a manufacturer approved cleaning solution is used during the periodic cleaning of PSP plates. Cleaning plates with the wrong solution can cause an irreparable artifact created as the solution soaks into the phosphor layer and leaving a stain. The stain disrupts the normal emission of light from the plate during the plate reading process. The PSP must be removed from service and disposed of according to the facility s waste removal policy. 35. Processor Artifacts Differentiating between the artifacts located on a plate and one that occurs during plate processing can be challenging when the questionable plate is not immediately available. The artifact s appearance generally provides some clues. Plate artifacts generally don t cross the entire axis of a plate; they only appear on images from the defective plate. On the other hand, processing artifacts run along an entire plate axis and appear on nearly all images regardless of the plate used.

10 36. Optical Component Artifacts Based on the construction of a plate reader, it might seem improbable for the inside to become dirty. Certain artifacts represent that they assuredly do. The actions of inserting the cassette into and removing the cassette from the reader allow for artifact generating contaminants to be introduced into the equipment. Artifacts such as a piece of hair located on an optical component transmitting light from the plate are a perfect example. A hair stuck to one of the optics can block the signal in that area of the image causing a "wig-wag" pattern as the hair moves back and forth during processing. Some plate readers have a built in cleaning feature that technologists can employ to decontaminate a portion of the optical system, removing the debris from the device. 37. Dirty Light Guide Some artifacts are more subtle than others. Arrows identify a line that runs along the second phalanx. The line represents a loss of image data caused by a dirty light guide in the plate reader. Dirt blocked the plate s signal and therefore the signal couldn't travel up the light guide through the photomultiplier tube to the analog-to-digital converter. The line runs the entire length of the plate and is parallel to the direction the plate traveled through the reader. 38. Flat-panel DEL Artifacts A source of artifact with flat panel units occurs within the detector elements, or DELs. In some cases the artifact is difficult to distinguish on a full-sized image. The arrows point to a small bright line in the chest, just below the humeral head. The partial loss of a row of DELs on the flat-panel detector caused this artifact. To fix this problem, the detector is recalibrated. If the artifact is still present following calibration then a field service engineer should be contacted to further investigate the issue. 39. Image Evaluation Factors There are several factors that are used to evaluate digital images. In the past, factors that were used for analog imaging were simply visual characteristics, such as; density, contrast, image blur, and distortion that couldn t have been manipulated. With digital evaluation, however, factors are a mix of visual cues and numerical values which include brightness, image blur, distortion, noise, and the exposure indicator. To better evaluate the digital image, it can be magnified and the brightness, and contrast levels can be adjusted, to further investigate artifacts and image blur. 40. Knowledge Check 41. Knowledge Check 42. Image Analysis What is causing the degraded image quality? a. exposure field recognition error. b. quantum mottle. c. saturation d. scatter. 43. Image Analysis A 3-on-1 projection of a finger demonstrates an exposure field recognition error. A rescaling error occurred during the computer s analysis of the image data. Because the field borders weren't recognized, the exposure data were included in the image data, including the intrafield area between

11 collimation borders on the receptor. When the values were incorporated into the analysis process there was a rescaling error and the lookup table application didn't display the appropriate brightness. Such 3- on-1 exposures are not uniformly accepted as a method to acquire images of extremities. Using 1-on-1 exposures basically ensures that this type of error won't occur, provided that collimation borders adhere to the rules for exposure field alignment. This error is evident because the image displays too dark and the technical factor selection used were appropriate for the exam. 44. Image Analysis What is causing the degraded image quality in the cervical region? a. exposure field recognition error. b. quantum mottle. c. saturation. d. scatter. 45. Image Analysis The cervical region is saturated. Imaging structures with large variations in part thickness that are adjacent to each other is challenging. The use of a compensating filter would be appropriate in order to create a more uniform attenuation of the beam between the upper thoracic and cervical vertebrae. 46. Image Analysis What is causing the degraded image quality on these radiographs? a. scatter. b. fog. c. grid cutoff. d. saturation. 47. Image Analysis The technologist confirmed that the correct technique was used; however, grid cutoff occurred. Improper alignment of the grid reduces the demonstrated contrast regardless of the receptor used. If this image had been repeated with the grid appropriately aligned with the x-ray beam, the image would have been acceptable. 48. Image Analysis What is causing the degraded image quality? a. exposure field recognition error. b. quantum mottle. c. saturation d. scatter. 49. Image Analysis The grainy appearance of the image represents under exposure. Quantum mottle occurs when there is insufficient exposure to the image receptor. When the strength of the signal is weak from insufficient x- ray photon interaction with the image receptor the result often is a mottled effect, or image noise.

12 50. Image Analysis Both images were both produced with the same technique factors. What is causing the difference in appearance of the two images? a. menu selection. b. grid cutoff. c. saturation d. quantum mottle. 51. Image Analysis The selection of the correct menu for an exam is important to ensure an image displays correctly. In some circumstances the exposure indicator is accurate although here the image appearance is different and the exposure indicator values are different by a factor of Image Analysis The image on the left is a PA projection of the wrist and the image on the right is a lateral projection. What is causing the degraded image quality? a. scatter radiation. b. exposure field recognition error. c. underexposure of lateral wrist. d. overexposure of the PA wrist. 53. Image Analysis The thickness of the body part changed the attenuation of the x-ray beam. One fundamental principle of imaging is when the thickness of the object being imaged changes, the level of exposure also must change. Working with a digital image receptor still requires technologists to adjust the exposure level to achieve an appropriate image display. The exposure level for a lateral projection should increase so the same amount of radiation strikes the receptor as in the PA image. In doing so, both images would display with equal brightness and contrast. 54. Image Analysis What is causing the artifact on this radiograph? a. stain. b. scratch. c. dent. d. dirt. 55. Image Analysis The artifact of the scratch on the full image is more apparent using magnification. The bright line in the tibia is a scratch on the photostimulable phosphor plate. 56. Image Analysis What is causing the artifact on this radiograph? a. scratched PSP. b. stained PSP.

13 c. dented PSP. d. dirty PSP. 57. Image Analysis The dirty PSP artifact is difficult to see on a full-sized image but the magnified image makes the artifact more apparent. The bright specks in the upper abdomen are there because the PSP needs to be cleaned. 58. Image Analysis What is causing the artifact on these radiographs? a. scratched PSP. b. stained PSP. c. dented PSP. d. dirty PSP. 59. Image Analysis On the first set of radiographs the arrow points to a small bright dot in the acromial process. On the radiographs of an AP shoulder the arrow points to a similar small bright dot in the soft tissue above the clavicle. The artifact is similar because it is actually the same artifact in both sets of images. Until the plate is cleaned every radiograph produced using this digital unit will demonstrate the same small bright dot. 60. Image Analysis What is causing the artifact? a. dirty light guide. b. corroded optical component. c. hair stuck to an optical component. d. scratch on the PSP plate. 61. Image Analysis The line identified by the white arrows runs through the metacarpals and represents the image data loss caused by a dirty light guide. Dirt blocked the signal from the plate so that the signal couldn t travel to the analog-to-digital converter. The line runs the entire width of the plate and is parallel to the direction the plate passed through the reader. 62. Image Analysis What 2 artifacts are visible on this radiograph? 1. dirty light guide. 2. corroded optical component. 3. hair stuck to an optical component. 4. scratch on the PSP plate. a. 1 and 2 b. 2 and 3 c. 1 and 4 d. 3 and Image Analysis A dirty light guide in the plate reader produced the line identified by the white arrow labeled "A." Dirt blocked the signal from the plate stopping the signal from travelling up the light guide, through the

14 photomultiplier tube, and to the analog-to-digital converter. The horizontal line identified by the white arrow labeled "B" was caused by a scratch on the PSP plate. 64. Conclusion This concludes Essentials of Digital Imaging: Module 4 Image Analysis. You should now be able to: Evaluate the effects of exposure changes on the appearance of a digital image. Evaluate the appearance of a digitally acquired image to assess quantum mottle and saturation. Discuss the relationship between exposure field recognition and the appearance of a digital image. Recognize motion on a digital image. Locate different types of artifacts on a digital image and state their probable cause. 65. References American Association of Physicists in Medicine. AAPM Report No An exposure indicator for digital radiography. Published July Accessed February 14, Bushong SC. Radiologic Science for Technologists: Physics, Biology, and Protection. 10th ed. St Louis, MO: Mosby Elsevier; Carlton RR, Adler AM. Principles of Radiographic Imaging: An Art and A Science. 5th ed. Clifton Park, NY: Thomson Delmar Learning; Carroll QB. Radiography in the Digital Age: Physics, Exposure, Radiation Biology. Springfield, IL: Charles C Thomas; Carter CE, Vealé BL. Digital Radiography and PACS. St Louis, MO: Mosby Elsevier; Practice Standards for Medical Imaging and Radiation Therapy. Radiography practice standards. American Society of Radiologic Technologists website. June 19, Accessed February 14, Seeram E. Digital Radiography: An Introduction for Technologists. Clifton Park, NY: Delmar Cengage Learning; Strategic Document. Version , April 11, Digital Imaging and Communications in Medicine website. Accessed February 14, 2013.

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura New Technology https://www.youtube.com/watch?v=ptkzznazb 7U COMPUTED

More information

Essentials of Digital Imaging

Essentials of Digital Imaging Essentials of Digital Imaging Module 2 Transcript 2016 ASRT. All rights reserved. Essentials of Digital Imaging Module 2 Processing 1. ASRT Animation 2. Welcome Welcome to Essentials of Digital Imaging

More information

Essentials of Digital Imaging

Essentials of Digital Imaging Essentials of Digital Imaging Module 7 Transcript 2016 ASRT. All rights reserved. Essentials of Digital Imaging Module 7 Quality 1. ASRT Animation 2. Welcome Welcome to the Essentials of Digital Imaging:

More information

SYLLABUS. TITLE: Equipment Operation I. DEPARTMENT: Radiologic Technology

SYLLABUS. TITLE: Equipment Operation I. DEPARTMENT: Radiologic Technology CODE: RADT 156 INSTITUTE: Health Science TITLE: Equipment Operation I DEPARTMENT: Radiologic Technology COURSE DESCRIPTION: This course covers the principles of equipment operation and maintenance of radiographic

More information

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question.

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question. RA110 test 3 Multiple Choice Identify the letter of the choice that best completes the statement or answers the question. 1. An object 35 cm in width is radiographed at 100 cm SID and at a 50 cm SOD. What

More information

3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development

3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development Teaching Radiographic Technique in a Digital Imaging Paradigm Objectives 1. Discuss the historical development of digital imaging. Dawn Couch Moore, M.M.Sc., RT(R) Assistant Professor and Director Emory

More information

LECTURE 1 The Radiographic Image

LECTURE 1 The Radiographic Image LECTURE 1 The Radiographic Image Prepared by:- KAMARUL AMIN ABDULLAH @ ABU BAKAR UiTM Faculty of Health Sciences Medical Imaging Department 11/23/2011 KAMARUL AMIN (C) 1 Lesson Objectives At the end of

More information

RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II

RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II APPROVED 12/O2/2011 EFFECTIVE SPRING 2013-14 Prefix & Number RAD 150 Course Title: Radiologic Exposure Technique II & Lab Purpose of this submission: New Change/Updated

More information

CR Basics and FAQ. Overview. Historical Perspective

CR Basics and FAQ. Overview. Historical Perspective Page: 1 of 6 CR Basics and FAQ Overview Computed Radiography is a term used to describe a system that electronically records a radiographic image. Computed Radiographic systems use unique image receptors

More information

Acquisition, Processing and Display

Acquisition, Processing and Display Acquisition, Processing and Display Terri L. Fauber, R.T. (R)(M) Department of Radiation Sciences School of Allied Health Professions Virginia Commonwealth University Topics Image Characteristics Image

More information

Digital radiography: Practical advantages of Digital Radiography. Practical Advantages in image quality

Digital radiography: Practical advantages of Digital Radiography. Practical Advantages in image quality Digital radiography: Digital radiography is set to become the most common form of processing radiographic images in the next 10 years. This is due to a number of practical and image quality issues. Practical

More information

Digital radiography (DR) post processing techniques for pediatric radiology

Digital radiography (DR) post processing techniques for pediatric radiology Digital radiography (DR) post processing techniques for pediatric radiology St Jude Children s Research Hospital Samuel Brady, MS PhD DABR samuel.brady@stjude.org Purpose Review common issues and solutions

More information

2217 US Highway 70 East Garner, NC Main: Fax:

2217 US Highway 70 East Garner, NC Main: Fax: Viztek is committed to providing the highest image quality possible in our CR & DR product lines. There are several factors that directly affect the overall quality of CR & DR based images. The eposure

More information

RADIOGRAPHIC EXPOSURE

RADIOGRAPHIC EXPOSURE RADIOGRAPHIC EXPOSURE Receptor Exposure Receptor Exposure the that interacts with the receptor. Computed Radiography ( ) requires a. Direct Digital Radiography (DR) requires a. Exposure Indicators Exposure

More information

Essentials of Digital Imaging

Essentials of Digital Imaging Essentials of Digital Imaging Module 6 Transcript 2016 ASRT. All rights reserved. Essentials of Digital Imaging Module 6 Dose Reduction and Patient Safety 1. ASRT Animation 2. Welcome Welcome to Essentials

More information

CONTRASTING VIEWS DIGITAL VS CONVENTIONAL RADIOGRAPHY

CONTRASTING VIEWS DIGITAL VS CONVENTIONAL RADIOGRAPHY Vet Times The website for the veterinary profession https://www.vettimes.co.uk CONTRASTING VIEWS DIGITAL VS CONVENTIONAL RADIOGRAPHY Author : PETRA AGTHE Categories : Vets Date : April 7, 2008 PETRA AGTHE

More information

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of radiation to the population due to Medical Imaging

More information

Essentials of Digital Imaging

Essentials of Digital Imaging Essentials of Digital Imaging Module 1 Transcript 2016 ASRT. All rights reserved. Essentials of Digital Imaging Module 1 Fundamentals 1. ASRT Animation 2. Welcome Welcome to Essentials of Digital Imaging

More information

Do you have any other questions? Please call us at (Toll Free) or , or

Do you have any other questions? Please call us at (Toll Free) or , or INSTRUCTIONS Read the appropriate course/ textbook. This is an open book test. A score of 75% or higher is needed to receive CE credit. You will have a maximum of three attempts to pass this course. Please

More information

STUDENT REVIEW QUESTION SET K CR/DR CONTENT AREA

STUDENT REVIEW QUESTION SET K CR/DR CONTENT AREA STUDENT REVIEW QUESTION SET K CR/DR CONTENT AREA RADT 2913 COMPREHENSIVE REVIEW 1 The CR cassette is backed by aluminum that: A. reflects x-rays B. absorbs x-rays C. captures the image D. transmits x-rays

More information

Digital Imaging Considerations Computed Radiography

Digital Imaging Considerations Computed Radiography Digital Imaging Considerations Digital Radiography Computed Radiography o Cassette based Direct or Indirect Digital Radiography o Cassetteless Computed Radiography 1 CR Image Acquisition Most like conventional

More information

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto X-ray Imaging PHYS Lecture Carlos Vinhais Departamento de Física Instituto Superior de Engenharia do Porto cav@isep.ipp.pt Overview Projection Radiography Anode Angle Focal Spot Magnification Blurring

More information

10/3/2012. Study Harder

10/3/2012. Study Harder This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura Study Harder CR detection is inefficient, inferior to film screen

More information

Outline ASRT Changes Impact on current curriculum Potential new courses WECM Changes Last update Resources and needs

Outline ASRT Changes Impact on current curriculum Potential new courses WECM Changes Last update Resources and needs Change nd Annual Blinn College 2 nd Educator s Workshop For Radiologic Sciences July 28, 2007 Christi Carter, MSRS, RT(R) Outline ASRT Changes Impact on current curriculum Potential new courses WECM Changes

More information

Teaching Digital Radiography and Fluoroscopic Radiation Protection

Teaching Digital Radiography and Fluoroscopic Radiation Protection Teaching Digital Radiography and Fluoroscopic Radiation Protection WCEC 20 th Student Educator Radiographer Conference Dennis Bowman, RT(R), CRT (R)(F) Community Hospital of the Monterey Peninsula (CHOMP)

More information

Beam-Restricting Devices

Beam-Restricting Devices Beam-Restricting Devices Three factors contribute to an increase in scatter radiation: Increased kvp Increased Field Size Increased Patient or Body Part Size. X-ray Interactions a some interact with the

More information

10/26/2015. Study Harder

10/26/2015. Study Harder This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura Study Harder CR detection is inefficient, inferior to film screen

More information

THE ART OF THE IMAGE: IDENTIFICATION AND REMEDIATION OF IMAGE ARTIFACTS IN MAMMOGRAPHY

THE ART OF THE IMAGE: IDENTIFICATION AND REMEDIATION OF IMAGE ARTIFACTS IN MAMMOGRAPHY THE ART OF THE IMAGE: IDENTIFICATION AND REMEDIATION OF IMAGE ARTIFACTS IN MAMMOGRAPHY William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org INTRODUCTION

More information

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Acceptance Testing of a Digital Breast Tomosynthesis Unit Acceptance Testing of a Digital Breast Tomosynthesis Unit 2012 AAPM Spring Clinical Meeting Jessica Clements, M.S., DABR Objectives Review of technology and clinical advantages Acceptance Testing Procedures

More information

Visibility of Detail

Visibility of Detail Visibility of Detail Radiographic Quality Quality radiographic images represents the, and information is for diagnosis. The of the anatomic structures and the accuracy of their ( ) determine the overall

More information

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Definition and Validation of Exposure Indices Ingrid Reiser, PhD DABR Department of Radiology University of Chicago

More information

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over a lifetime Breast cancer screening programs rely on

More information

While digital techniques have the potential to reduce patient doses, they also have the potential to significantly increase them.

While digital techniques have the potential to reduce patient doses, they also have the potential to significantly increase them. In press 2004 1 2 Guest Editorial (F. Mettler, H. Ringertz and E. Vano) Guest Editorial (F. Mettler, H. Ringertz and E. Vano) Digital radiology An appropriate analogy that is easy for most people to understand

More information

4. Contrast is the. There must The function of contrast is to:. The types of contrast are.

4. Contrast is the. There must The function of contrast is to:. The types of contrast are. RADIOGRAPHIC VISIBILITY OF DETAIL STUDY QUESTIONS 1. What is visibility of detail? It is controlled by properties. What are the two factors that affect it? 2. What is sharpness of detail? It is controlled

More information

ADC COMPACT FULL-LEG/FULL-SPINE APPLICATION SOFTWARE USER MANUAL

ADC COMPACT FULL-LEG/FULL-SPINE APPLICATION SOFTWARE USER MANUAL ADC COMPACT FULL-LEG/FULL-SPINE APPLICATION SOFTWARE USER MANUAL TABLE OF CONTENTS Full-leg/full-spine image stitching: a new and accurate CR-based imaging software 1 The conventional approach 1 CR and

More information

The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System

The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System Manus

More information

A Practical Overview of the Clinical and Operational Impact of Computed Radiography(CR) Implementations. Shirley Weddle, RT(R)(M), CIIP, BBA

A Practical Overview of the Clinical and Operational Impact of Computed Radiography(CR) Implementations. Shirley Weddle, RT(R)(M), CIIP, BBA A Practical Overview of the Clinical and Operational Impact of Computed Radiography(CR) Implementations Shirley Weddle, RT(R)(M), CIIP, BBA OBJECTIVES Define Computed Radiography (CR) Discuss CR vendor

More information

Artefacts found in computed radiography

Artefacts found in computed radiography The British Journal of Radiology, 74 (2001), 195 202 E 2001 The British Institute of Radiology Pictorial review Artefacts found in computed radiography L J CESAR, RT(R)(QM), B A SCHUELER, PhD, F E ZINK,

More information

Appropriate Inspection Distance of Digital X-Ray Imaging Equipment for Diagnosis

Appropriate Inspection Distance of Digital X-Ray Imaging Equipment for Diagnosis Indian Journal of Science and Technology Vol 8(S8), 380-386, April 2015 ISSN (Print) : 0974-6846 ISSN (Online) : 0974-5645 DOI: 10.17485/ijst/2015/v8iS8/70528 Appropriate Inspection Distance of Digital

More information

COMPUTED RADIOGRAPHY (CR)

COMPUTED RADIOGRAPHY (CR) COMPUTED RADIOGRAPHY (CR) Moving with the time Avi Avner BVSc BSc CVR DVDI MRCVS CR-Basics A five step process: 1. X-ray image received on phosphor plate 2. Image extracted from phosphor plate by Laser

More information

DIGITAL RADIOGRAPHY ARTIFACTS

DIGITAL RADIOGRAPHY ARTIFACTS IMAGING LAB MPHY 487 DIGITAL RADIOGRAPHY ARTIFACTS Mohammad Esmael Alsulimane B.Sc, M.Sc Medical Physics Lecturer - Physics Department All Rights Reserved: Some information and figures in this presentation

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 07-644 Grid Alignment Test Tool Users Manual March 2005 Manual No. 07-644-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks

More information

1. Carlton, Richard R., and Arlene M. Adler. Principles of Radiographic Imaging: An Art and a Science, 5th edition (2013).

1. Carlton, Richard R., and Arlene M. Adler. Principles of Radiographic Imaging: An Art and a Science, 5th edition (2013). CODE: RADT 151 INSTITUTE: Health Science TITLE: Radiographic Exposure DEPARTMENT: Radiologic Technology COURSE DESCRIPTION: This course covers the principles of radiographic exposure selection and manipulation

More information

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah X-RAY IMAGING EE 472 F2017 Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook Stewart C. Bushong, Radiologic Science for Technologists: Physics, Biology, and Protection, 10 th ed., Mosby,

More information

- KiloVoltage. Technique 101: Getting Back to Basics

- KiloVoltage. Technique 101: Getting Back to Basics Why do I need to know technique? Technique 101: Getting Back to Basics Presented by: Thomas G. Sandridge, M.S., M.Ed., R.T.(R) Program Director Northwestern Memorial Hospital School of Radiography Chicago,

More information

A Comprehensive Review of Image Production

A Comprehensive Review of Image Production A Comprehensive Review of Image Production Presented by: John Fleming, M.Ed., RT(R)(MR)(CT) St. Petersburg College Office: (727) 341-3758 E-mail: flemingj@spcollege.edu Lesson Objectives: ARRT Content

More information

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 2 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 COMPUTERIZED IMAGING Section I: Chapter 2 RADT

More information

2017 West Coast Educators Conference Orlando. Projection Geometry. 1. Review hierarchy of image qualities (amplified version):

2017 West Coast Educators Conference Orlando. Projection Geometry. 1. Review hierarchy of image qualities (amplified version): Spatial Resolution in the Digital Age: NOTES Quinn B. Carroll, MEd, RT 2017 West Coast Educators Conference Orlando Projection Geometry 1. Review hierarchy of image qualities (amplified version): a. Maximum

More information

Image Quality Artifacts in Digital Imaging

Image Quality Artifacts in Digital Imaging MAHIDOL UNIVERSITY Wisdom of the Land Image Quality Artifacts in Digital Imaging Napapong Pongnapang, Ph.D. Department of Radiological Technology Faculty of Medical Technology Mahidol University, Bangkok,

More information

Y11-DR Digital Radiography (DR) Image Quality

Y11-DR Digital Radiography (DR) Image Quality Y11-DR Digital Radiography (DR) Image Quality Image quality is stressed for all systems in Safety Code 35. In the relevant sections Health Canada s advice is the manufacturer s recommended test procedures

More information

Veterinary Science Preparatory Training for the Veterinary Assistant. Floron C. Faries, Jr., DVM, MS

Veterinary Science Preparatory Training for the Veterinary Assistant. Floron C. Faries, Jr., DVM, MS Veterinary Science Preparatory Training for the Veterinary Assistant Floron C. Faries, Jr., DVM, MS Radiology Floron C. Faries, Jr., DVM, MS Objectives Determine the appropriate machine settings for making

More information

Contrast. Contrast: the difference in density on adjacent areas of a radiograph or other image receptor. Subjective. Long Scale (Low Contrast)

Contrast. Contrast: the difference in density on adjacent areas of a radiograph or other image receptor. Subjective. Long Scale (Low Contrast) Contrast Contrast: the difference in density on adjacent areas of a radiograph or other image receptor. Subject Subjective Radiographic Long Scale (Low Contrast) Short Scale (High Contrast) Factors affecting

More information

Introduction. Sam R. Kottamasu Lawrence R. Kuhns

Introduction. Sam R. Kottamasu Lawrence R. Kuhns Pediatr Radiol (1997) 27: 119 123 Springer-Verlag 1997 Sam R. Kottamasu Lawrence R. Kuhns Musculoskeletal computed radiography in children: scatter reduction and improvement in bony trabecular sharpness

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity)

Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity) Vascular Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO Medical Systems Division, Shimadzu Corporation Yoshiaki Miura 1. Introduction In recent years, digital cardiovascular

More information

1.1.Clinical Artefacts

1.1.Clinical Artefacts 1.1.Clinical Artefacts While the incidence of artefact on digital mammographic images 1 is typically less than with film based mammography, artefacts can be produced on digital systems. This section provides

More information

CT Basics: Image Quality Module 6

CT Basics: Image Quality Module 6 Module 6 For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions. Any broadcast,

More information

SPRINGFIELD TECHNICAL COMMUNITY COLLEGE ACADEMIC AFFAIRS

SPRINGFIELD TECHNICAL COMMUNITY COLLEGE ACADEMIC AFFAIRS SPRINGFIELD TECHNICAL COMMUNITY COLLEGE ACADEMIC AFFAIRS Course Number: RADG 112 Department: Radiography Course Title: Image Production & Eval. Semester: Spring Year: 1997 Objectives/ Unit One: Introduction

More information

Outline. Digital Radiography. Understanding Digital Modalities: Image Quality and Dose. Image Quality. Dose Control

Outline. Digital Radiography. Understanding Digital Modalities: Image Quality and Dose. Image Quality. Dose Control Understanding Digital Modalities: Image Quality and Dose S. Jeff Shepard, M.S. University of Texas M. D. Anderson Cancer Center Houston, Texas Special Acknowledgement: Stephen K. Thompson, M.S. William

More information

Digital Imaging CT & MR

Digital Imaging CT & MR Digital Imaging CT & MR January 22, 2008 Digital Radiography, CT and MRI generate images in a digital format What is a Digital Image? A digital image is made up of picture elements, pixels row by column

More information

Nuclear Associates , , , , , ,

Nuclear Associates , , , , , , Nuclear Associates 57-411, 57-412, 57-413 57-426, 57-431, 57-432 57-433, 57-435, 57-436 CLEAR-Pb Transparent X-Ray Compensation Filters Users Manual March 2005 Manual No. 57-XXX-1 Rev. 2 2003, 2005 Fluke

More information

Open. the Digitized world. Fuji Computed Radiography

Open. the Digitized world. Fuji Computed Radiography Open the Digitized world Fuji Computed Radiography If just one of these applies to you... Managing developing fluid is hard and darkroom work is a hassle... Images are not stable... Isn t digitalization

More information

Small Animal Radiographic Techniques and Positioning COPYRIGHTED MATERIAL

Small Animal Radiographic Techniques and Positioning COPYRIGHTED MATERIAL Small Animal Radiographic Techniques and Positioning COPYRIGHTED MATERIAL Section 1 Theory and Equipment 1 Introduction to Digital Imaging Small animal radiography has changed dramatically in the past

More information

Digital Radiography of the Musculoskeletal System: The Optimal Image

Digital Radiography of the Musculoskeletal System: The Optimal Image Digital Radiography of the Musculoskeletal System: The Optimal Image Matthew Freedman and Dot Steller Digital radiography (DR) is replacing screen-film (SF} radiography for musculoskeletal examinations

More information

SECTION I - CHAPTER 1 DIGITAL RADIOGRAPHY: AN OVERVIEW OF THE TEXT. Exam Content Specifications 8/22/2012 RADT 3463 COMPUTERIZED IMAGING

SECTION I - CHAPTER 1 DIGITAL RADIOGRAPHY: AN OVERVIEW OF THE TEXT. Exam Content Specifications 8/22/2012 RADT 3463 COMPUTERIZED IMAGING RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 1 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 1 DIGITAL RADIOGRAPHY: AN OVERVIEW OF THE TEXT RADT 3463 COMPUTERIZED IMAGING Section I: Chapter

More information

Overview of Safety Code 35

Overview of Safety Code 35 Common Quality Control Procedures for All s Quality Control Procedures Film All s Daily Quality Control Tests Equipment Warm-up (D1) According to manufacturers instructions Can include auto calibration(d1)

More information

Photostimulable phosphor plates (PSPs)

Photostimulable phosphor plates (PSPs) DIGITAL IMAGING Digital imaging Photostimulable phosphor plates (PSPs) Indirect digital PSPs are composed of a polyester base with a phosphor layer (europium activated barium fluorohalide) on one side.

More information

Features and Weaknesses of Phantoms for CR/DR System Testing

Features and Weaknesses of Phantoms for CR/DR System Testing Physics testing of image detectors Parameters to test Features and Weaknesses of Phantoms for CR/DR System Testing Spatial resolution Contrast resolution Uniformity/geometric distortion Dose response/signal

More information

DIGITAL RADIOGRAPHY. Digital radiography is a film-less technology used to record radiographic images.

DIGITAL RADIOGRAPHY. Digital radiography is a film-less technology used to record radiographic images. DIGITAL RADIOGRAPHY Digital radiography is a film-less technology used to record radiographic images. 1 The purpose of digital imaging is to generate images that can be used in the diagnosis and assessment

More information

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893 Digital Radiography D. J. Hall, Ph.D. x20893 djhall@ucsd.edu Background Common Digital Modalities Digital Chest Radiograph - 4096 x 4096 x 12 bit CT - 512 x 512 x 12 bit SPECT - 128 x 128 x 8 bit MRI -

More information

Maximizing clinical outcomes

Maximizing clinical outcomes Maximizing clinical outcomes Digital Tomosynthesis Dual Energy Subtraction Automated Long Length Imaging Improved image quality at a low dose Xray Xray Patented ISS capture technology promotes high sensitivity

More information

Evaluation of a quality control phantom for digital chest radiography

Evaluation of a quality control phantom for digital chest radiography JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 2, NUMBER 2, SPRING 2001 Evaluation of a quality control phantom for digital chest radiography Eugene Mah* Department of Radiology, Medical University

More information

IBEX TECHNOLOGY APPLIED TO DIGITAL RADIOGRAPHY

IBEX TECHNOLOGY APPLIED TO DIGITAL RADIOGRAPHY WHITE PAPER: IBEX TECHNOLOGY APPLIED TO DIGITAL RADIOGRAPHY IBEX Innovations Ltd. Registered in England and Wales: 07208355 Address: Discovery 2, NETPark, William Armstrong Way, Sedgefield, UK Patents:

More information

Radiology. Radiograph: Is the image of an object made with use of X- ray instead of light.

Radiology. Radiograph: Is the image of an object made with use of X- ray instead of light. Radiology د. اريج Lec. 3 X Ray Films Radiograph: Is the image of an object made with use of X- ray instead of light. Dental x- ray film: Is a recording media on which image of the object was made by exposing

More information

WHITE PAPER. Moving from CR to DR. Optimizing Image Quality and Dose

WHITE PAPER. Moving from CR to DR. Optimizing Image Quality and Dose Moving from CR to DR Optimizing Image Quality and Dose Table of contents 1. Introduction...3 2. CR and DR: similarities and differences...4 3. Image capture technologies...5 3.1 DR panel technology...5

More information

NJDEP Medical Physicist s Radiographic QC Survey Registration Number:

NJDEP Medical Physicist s Radiographic QC Survey Registration Number: Facility Name NJDEP ID # NJDEP Medical Physicist s Radiographic QC Survey PLEASE PRINT Facility Information Unit Information Manufacturer Model Console Model # Console serial # Tube serial # Location (room)

More information

Mammography: Physics of Imaging

Mammography: Physics of Imaging Mammography: Physics of Imaging Robert G. Gould, Sc.D. Professor and Vice Chair Department of Radiology and Biomedical Imaging University of California San Francisco, California Mammographic Imaging: Uniqueness

More information

Breast Imaging Basics: Module 10 Digital Mammography

Breast Imaging Basics: Module 10 Digital Mammography Module 10 Transcript For educational and institutional use. This test bank is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions.

More information

R/F. Comparison of Long View Radiography Systems. 1. Introduction. 2. Methods of Long View Radiography

R/F. Comparison of Long View Radiography Systems. 1. Introduction. 2. Methods of Long View Radiography R/F Comparison of Long View Radiography Systems Department of Radiology, Tokyo Women's Medical University Medical Center East 1 Department of Central Radiology, Tokyo Women's Medical University Hospital

More information

Digital Radiography : Flat Panel

Digital Radiography : Flat Panel Digital Radiography : Flat Panel Flat panels performances & operation How does it work? - what is a sensor? - ideal sensor Flat panels limits and solutions - offset calibration - gain calibration - non

More information

Moving from film to digital: A study of digital x-ray benefits, challenges and best practices

Moving from film to digital: A study of digital x-ray benefits, challenges and best practices Moving from film to digital: A study of digital x-ray benefits, challenges and best practices H.U. Pöhler 1 and N. D Ademo 2 DÜRR NDT GmbH & Co. KG, Höpfigheimer Straße 22, Bietigheim-Bissingen, 74321,

More information

RADIOGRAPHY TERMS TO KNOW SELF STUDY DENTALELLE TUTORING

RADIOGRAPHY TERMS TO KNOW SELF STUDY DENTALELLE TUTORING RADIOGRAPHY TERMS TO KNOW SELF STUDY DENTALELLE TUTORING PLEASE NOTE You DO NOT need to study these for the board exam if this is why you bought our Radiography course, however if you come across any terms

More information

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Studies on reduction of exposure dose using digital scattered X-ray removal processing Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.

More information

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Studies on reduction of exposure dose using digital scattered X-ray removal processing Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 07-647 R/F QC Phantom Operators Manual March 2005 Manual No. 07-647-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. All product names are trademarks of their respective companies

More information

The effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography

The effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography Journal of Physics: Conference Series OPEN ACCESS The effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography To cite this article: N A A Daud et al 2014 J. Phys.:

More information

Image Display and Perception

Image Display and Perception Image Display and Perception J. Anthony Seibert, Ph.D. Department of Radiology UC Davis Medical Center Sacramento, California, USA Image acquisition, display, & interpretation X-rays kvp mas Tube filtration

More information

NOTES Teaching Digital Processing and Display Quinn B Carroll, MEd, RT WCEC Conference, 2015

NOTES Teaching Digital Processing and Display Quinn B Carroll, MEd, RT WCEC Conference, 2015 NOTES Teaching Digital Processing and Display Quinn B Carroll, MEd, RT WCEC Conference, 2015 1. Teaching Pitfalls: Accuracy Vs. Simplicity 2. Teaching Decisions: 1. Is it relevant? (Not: Is it in my comfort

More information

Fabrício Sampaio Péres Kury Federal University of Rio de Janeiro Medical School

Fabrício Sampaio Péres Kury Federal University of Rio de Janeiro Medical School Fabrício Sampaio Péres Kury Federal University of Rio de Janeiro Medical School Harvard Medical School Exchange Clerkship Program Primary Care Radiology Clerkship Gillian Lieberman, M. D. Monday, September

More information

Collimation Assessment Using GAFCHROMIC XR-M2

Collimation Assessment Using GAFCHROMIC XR-M2 Collimation Assessment Using GAFCHROMIC XR-M2 I. Introduction A method of collimation assessment for GE Senographe full-field digital mammography (FFDM) systems is described that uses a self-developing

More information

X o- System. Chiro- Systems. Chiropractic X-Ray Systems and Digital Solutions. Digital Radiography Computerized Radiography Chiropractic Systems

X o- System. Chiro- Systems. Chiropractic X-Ray Systems and Digital Solutions. Digital Radiography Computerized Radiography Chiropractic Systems X o- System Chiropractic X-Ray Systems and Digital Solutions Chiro- Systems Digital Radiography Computerized Radiography Chiropractic Systems Quantum s Chiropractic Imaging Solutions Digital Work Flow

More information

Photomultiplier Tube

Photomultiplier Tube Nuclear Medicine Uses a device known as a Gamma Camera. Also known as a Scintillation or Anger Camera. Detects the release of gamma rays from Radionuclide. The radionuclide can be injected, inhaled or

More information

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE

More information

Comparison of computed radiography and filmõscreen combination using a contrast-detail phantom

Comparison of computed radiography and filmõscreen combination using a contrast-detail phantom JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 1, WINTER 2003 Comparison of computed radiography and filmõscreen combination using a contrast-detail phantom Z. F. Lu,* E. L. Nickoloff, J.

More information

X-RAYS - NO UNAUTHORISED ENTRY

X-RAYS - NO UNAUTHORISED ENTRY Licencing of premises Premises Refer Guidelines A radiation warning sign and warning notice, X-RAYS - NO UNAUTHORISED ENTRY must be displayed at all entrances leading to the rooms where x-ray units are

More information

radiography detector

radiography detector Clinical evaluation of a full field digital projection radiography detector Gary S. Shaber'1, Denny L. Leeb, Jeffrey Belib, Gregory Poweii1', Andrew D.A. Maidment'1 a Thomas Jefferson University Hospital,

More information

Get more from your images with Symphony Image Processing

Get more from your images with Symphony Image Processing DIRECT RADIOGRAPHY The user-friendly DelWorks image acquisition and processing software possesses a wide range of tools for a variety of image manipulations. Its user interface simplifies every step of

More information

DIGITAL IMAGING Recognise the importance of quality assurance

DIGITAL IMAGING Recognise the importance of quality assurance DIGITAL IMAGING Recognise the importance of quality assurance There are two types of digital image receptor both of which capture a two dimensional image of the three dimensional patient. These are Computed

More information

Amorphous Selenium Direct Radiography for Industrial Imaging

Amorphous Selenium Direct Radiography for Industrial Imaging DGZfP Proceedings BB 67-CD Paper 22 Computerized Tomography for Industrial Applications and Image Processing in Radiology March 15-17, 1999, Berlin, Germany Amorphous Selenium Direct Radiography for Industrial

More information

BENNETT RENAISSANCE SERIES. Dedicated to Quality, Innovation and the Chiropractic Profession

BENNETT RENAISSANCE SERIES. Dedicated to Quality, Innovation and the Chiropractic Profession BENNETT RENAISSANCE SERIES TM Dedicated to Quality, Innovation and the Chiropractic Profession About HCMI, Inc. For over 20 years, HCMI, Inc. has been providing quality equipment for the chiropractic profession.

More information