Biomedical. Measurement and Design ELEC4623/ELEC9734. Electrical Safety and Performance Standards

Size: px
Start display at page:

Download "Biomedical. Measurement and Design ELEC4623/ELEC9734. Electrical Safety and Performance Standards"

Transcription

1 Biomedical Instrumentation, Measurement and Design ELEC4623/ELEC9734 Electrical Safety and Performance Standards

2 Contents Physiological Effects of Electrical Currents Safety Standards for Medical Instrumentation ECG Performance standards

3 Physiological effects of electricity E is proportional to m Ionic Concentration (outside) Ionic Concentration (inside) Effects of electrical currents on the body manifest themselves in three different ways: Injury to tissue Uncontrollable muscle contraction Ventricular fibrillation All living mammalian cells exhibit a potential difference across the cell membrane (Nernst)

4 Nerve excitation Excitation of a nerve occurs at a threshold level of ~ - 60 mv, and the action potential is propagated along the whole nerve membrane. At the end of the nerve, the ACTION POTENTIAL causes a release of the chemical ACETYCHOLINE A WAVE of DEPOLARISATION propagates down the entire muscle membrane. This wave causes movement of specific ions (CALCIUM) into the muscle cell to activate the contractile machine

5 Muscle contraction The muscle can be stimulated by contract by: activating the nerve connected to the muscle direct stimulation of the muscle with ELECTRIC current. The temporal dependence of the various phases of action potential (i.e., it takes a finite time for ions to pass across the membrane) the properties which define the excitation of nerve and muscle cells are described by the STRENGTH- DURATION CURVE, i. e. the relationship between the strength (magnitude) of current and the duration of times required to produce excitation and therefore contraction.

6 Strength/Duration curves Current 2I 0 I 0 τ Excites T (secs) There is a minimum value of current (I O = constant rheobase current) to produce excitation τ = Chronaxie time constant It τ = I0 1+ t Hence the frequency of stimulation is important in terms of the effectiveness of a current in producing excitation The optimal frequency range is Hz (!)

7 Stimulation frequency Tetanic or subtetanic contractions can cause respiration to stop or uncontrolled movement of the limbs. High frequency and short duration stimulation does not produce muscle contraction but is used to produce local heating as in the process of DIATHERMY ( Electric Scalpel ) Tetanic contraction = contraction of maximum amplitude

8 Ventricular Ectopic Beats Ectopic = occurring in an abnormal position or in an unusual manner or form Ventricular ectopic beats (premature ventricular contractions) are common in healthy people Under normal conditions the heart is refractory for about 300 msec. A VEB will normally annihilate itself and will not be self-sustaining.

9 Ventricular Fibrillation If the heart tissue becomes damaged (e.g. from ischaemia or infarction) it can affect the normal conduction pathways By the time the ectopic activity travels around, the transient refractory area is sensitive again, causing a CIRCUS MOTION leading to FIBRILLATION VF is almost invariably fatal unless defibrillated: the uncoordinated ventricular contractions result in little or no blood flow to the body. In the ECG VF appears as an erratic rhythm where the waves cannot be identified.

10 Sensitivity to electric current

11 Perception current The minimum value of current perceived by a subject. Mucous membranes are the most sensitive, but current is usually measured in the grasped hand 45 μa for the tongue 1 ma for the hand The frequency dependence of perception current is a direct result of the strength-duration curve for membrane excitation. Slide 6 The perception current, in itself, is not hazardous, but accidents may occur indirectly. It is important in determining levels of leakage current for household and industrial appliances

12 Let-go current The current at which a subject is unable to release an energised conductor grasped in the hand. The current stimulates all the forearm muscles, causing tetanic contractions. Since the flexor muscles are stronger than the extensor muscles, the subject is unable to release the grip. The following have little effect on let-go current The location of the return electrode (hand or foot) The size of the stimulating conductor Moisture conditions of the site of the electrode contact t

13 Let-go current Let-go current is also frequency dependent Relative minimum in the range Hz mean ~ 16 ma for men 2/3*16 for women. The let-go current increases at both LOW and HIGH frequencies e.g. 24 ma at 5 Hz and 1000 Hz

14 Physiological sensations at DC Physiological sensations Percentage of test subjects 5% 50% 95% Slight prickle in palms and Finger-tips at ma Feeling of warmth and increased prickle in at ma palms, slight pressure in wrists Pressure increasing to shooting pain, developing in wrists and palms Prickle in forearm, pressure in wrists, shooting pain in hands, increased feeling of warmth at ma at ma Increased pain caused by pressure in at ma wrists, prickle reaching up to elbow Acute pain caused by pressure in wrists, shooting pain in hands Current path: hand-body-hand; current in ma at ma

15 Fibrillatory Current K IRMS =, ma (0.083 < t < 5 secs) t K (ma.s -0.5 ) 50 Kg 18 Kg Adult child Experiments may not be conducted directly in humans Non-Fibrillating 116 ma* 52 ma Extrapolation ti from animals or analyses of accidental Fibrillating 185 ma 69 ma electrocutions For a 95% probability of fibrillation or a 95% probability of non-fibrillation with shock Hz ma duration of 1 sec Fibrillatory current is also frequency dependent < ~

16 Factors affecting fibrillatory current 1. Electrode Position Experiments are usually performed in dogs with saline filled catheters (see next slide) Most sensitive areas are: apex of right ventricle (for internal placement) midpoint of right ventricle (for external placement on the pericardium). Five fold variation of current with position and orientation of contact with electrode. A current of 20 microamps caused ventricular fibrillation in dogs. (Hence LEAKAGE current set to 10 microamps by many electrical standards).

17 Canine ventricular arrhythmia, pump failure and fibrillation threshold currents of 60 Hz and 3 seconds duration Catheter position Random Most Sensitive: (n = 54) (n = 50) Ventricular arrhythmia 65 ± ± 2.5 Pump failure 125 ± ± 4.4 Fibrillation 335 ± ± 36 Source: Ledsome et al. Cardiovascular Research 8: , * Values in microamperes, mean ± SD.

18 Factors affecting fibrillatory current 2. Electrode size and current density. Threshold for fibrillation is inversely proportional to current density Fibrillation threshold for application of current to exposed heart of >3 sec: ~ 80 microamps / mm 2 for area of 2 mm 2 ~ 5 microamps / mm 2 for area > 200 mm 2 3. Time factors Current must flow through the heart during the RELATIVE REFRACTORY PERIOD (T wave) for fibrillation to occur. Threshold decreases with increasing duration of current over range 83msec sec (AC) and msec - 21 sec (DC)

19 Fibrillatory window There is a variable upper limit of current above which fibrillation will not occur i.e. fibrillation occurs only in a current window. The longer the duration of applied current, the wider the window. For short pulses, the window is 3-4 times the threshold limit. Upper limit is ~2-3A through the intact chest, above this limit the heart will remain contracted during time of current application. Current greater than 2 Amps through the intact chest may not result in fibrillation but can cause temporary cardiac arrest and paralysis. Current greater than 6 Amps can cause cardiac arrest, paralysis and deep burns On removal of the large super-fibrillation threshold applied current, heart usually resumes its normal rhythm DEFIBRILLATION apply a large high-current pulse simultaneously to ALL heart muscle cells All depolarise at once and relax together Hopefully sinus rhythm returns

20 Impedance of the human body TOTAL BODY IMPEDANCE, essentially consists of the sum of THREE impedances; SKIN impedance at the TWO contact points: extremely variable with moisture, contact area, pressure, temperature, current duration, touch voltage, time. p,, g, INTERNAL impedance: essentially resistive (approx. 500 Ohm) and not too variable.

21 Total body impedance at 50Hz V1= R V(Volts) Z t (Ohms) I (ma) Estimated total body impedances at 50Hz

22 Internal body resistances Friedberger (1934): simplified representation for a reference body (Height 165 cm; Weight 56 kg)

23 Protection against electric shock with floating circuits A floating or isolated circuit itis a one that thas no connection to any other circuit or to ground. The secondary circuit it is isolated from all other circuits by the transformer functional insulation, and the other functional insulations. However, there is a very small leakage current in the stray impedances of the functional insulations. A floating or isolated circuit

24 Protection Circuits Pole-to-opposite pole. Figure 2A depicts the circuit path when a man simultaneously touches both poles of the floating or isolated circuit. In this situation, there is no insulation. The current is limited it only by the impedance of the body.

25 Protection Circuits If we interpose an insulating barrier between one pole of the floating circuit and the man, then we can define that barrier as Basic Insulation. In the event of failure of that t Basic Insulation, there is an electric shock current in the man. If we extend that same insulation such that it is interposed between the OPPOSITE pole and the man, then we can define the OPPOSITE pole portion of the insulation as Supplementary Insulation (because it provides insulation against the SECOND body connection). Failure of ONLY one insulation leads to no shock current.

26 Biomedical Instrumentation, Measurement and Design ELEC4483 Lectures 21 and 22 Diagnostic Ultrasound

27 Physical properties of sound Ultrasound = 20KHz 1 GHz Diagnostic Ultrasound is usually in the range of 1 10 MHz Remember sound is a longitudinal pressure wave v = f*λ, f = frequency, λ =wavelength Velocity of ultrasound in soft tissue is relatively constant ~1540m/sec Variations in v are functions of the density and elasticity

28 Sound waves

29 Acoustic (characteristic) impedance Acoustic (characteristic) impedance Z = ρ *v (kg/m 2 /sec, v = velocity, ρ = density) Think of this as the substance s resistance to penetration of sound waves When a sound wave reaches the boundary between two media of different characteristic impedance, it may be transmitted, reflected, refracted or scattered. What actually happens depends on: The difference between the characteristic ti impedance of the two media. The orientation of the sound beam relative to the boundary plane. The smoothness or roughness of the boundary. The distance between sound source and boundary The frequency of the ultrasound wave.

30 Reflection and Transmission When a sound wave strikes a boundary at right angles, part of the energy is transmitted and the rest is reflected back The ratio of the amplitude of the reflected wave (echo) to that of the incident wave is called the amplitude reflection coefficient R a, and is given by Z Z R a = Z + Z Z 1 = Z 2 means no echo Liver to kidney R a ~ 6%, Soft tissue to lung R a ~ 50% Soft tissue to bone R a ~ 40%. Soft tissue to air R a ~ 99% the use of a gel as a coupling medium is essential.

31 Schematic illustration of the reflection of an ultrasound pulse

32 Refraction When the sound sits the interface at an angle, reflected and transmitted beams travel in different directions to the incident beam incident angle = reflection angle transmitted wave is refracted angle of refraction depends d on the difference of the characteristic impedances of the two media, and hence to the difference in wave velocities. The angle of refraction is smaller than the incident angle when the wave velocity in the second medium is slower than in the first medium When the refraction angle becomes 90, no energy is transmitted. Critical angle cannot see any deeper structures

33 Types of echoes (reflections) Specular - originate from smooth, large, regularly shaped objects Such echoes are relatively strong and depend on incident angle Scattered - originate from relatively small, weakly reflective, irregularly shaped objects are less angle dependant and less intense. (ie. blood cells)

34 Attenuation All media attenuate ultrasound Attenuation of ultrasound increases with frequency (thus penetration drops) Ultrasound amplitude attenuation coefficient α Ultrasound intensity attenuation coefficient μ NOTE: Intensity is proportional to (amplitude) 2 So μ = 2 α Ultrasound rule of thumb for soft tissues α = A db cm -1 MHz -1 For many tissues A ~ 1 I( x) = I e, Q( x) = Q e μ x 0 0 α x

35 Absorption and power Absorption results in the conversion of the wave energy to heat This causes a temperature rise Ultimately the best images come from highest S/N ratio So just use more power! Regulations? IEC and FDA Power restricted to about 200mW max for pulsed, 50% that for CW Can cause up to a C temperature rise

36 Transducers Piezoelectric effect transducers always used Quartz and some ceramic materials are piezoelectric Apply voltage get a deformation or vice versa AC voltage applied will produce vibrations at the same f This produces sound waves

37 Continuous wave (CW) vs. pulsed wave (PW) ultrasound Previously we looked at CW ultrasound for measuring blood flow. CW Doppler separate transmitter and receiver, continuous signal transmission PW Doppler same transducer for transmit and receive Pulses sent in bursts with time delay in between

38 CW vs PW Continuous Wave Advantages Accurately measures high velocity flows Disadvantages Lacks range resolution o Pulsed wave Ability to measure velocities at a specific location (range resolution) Aliasing of velocities above the Nyquist limit (inability to measure high velocities accurately)

39 Pulsed wave signal Typically, the pulse duration is 1 micro second and the pulse repetition period is 1 msec. Since the duty factor is of the order of to 0.005, the same element can transmit the ultrasound and also receive the echo. This technique is known as the Pulse-Echo mode.

40 Time gain control Voltage controlled It accounts for the attenuation of sound in the tissue Time it takes for pulse to return is proportional to distance travelled Simplest form is a logarithmic voltage ramp

41 Near and far field zones Far field zone at distance 2 D > 4 4λ λ

42 Mechanical focusing Internal Focusing - the piezoelectric element is formed in a curved element that focus the beam. External Focusing - a lens is placed on the transducer face

43 Ultrasound scan heads (a) Rotating mechanical device 5 to 10 rotations per second (b) linear sequenced array (c) curved linear array (d) phased array (beam steering)

44 Linear sequenced arrays Pulse is applied simultaneously to all elements in a small group (aperture) E.g. 1 to 4 then 2 5 etc. Each element is about a wavelength wide Rectangular image formed, with width equal to scanning head width

45 Curved linear arrays Curved linear arrays send pulses out in different directions Cover wider area than linear arrays for size of head Operation identical to linear sequenced array

46 Linear phased arrays Pulses are applied to ALL elements (not a small pp ( group) Each element has a programmable phase delay End result is to steer or focus beam in different directions

47 Phased array focussing and steering Focussing and steering an acoustic beam using a phased array. (a) transmit mode (b) Receive mode. Dynamic focussing allows the focus to change the focal point.

48 Lateral resolution The ability to resolve objects side by side. Minimum reflector (object) separation perpendicular to beam direction that produces two separate echoes Equals the beam width in the scan plane

49 Axial resolution The ability to resolve objects that lie behind one another in the direction of the beam. Equals half the spatial pulse length (Length of a pulse from front to back = length of each cycle times the number of cycles in the pulse) Can improve by Reducing number of cycles per pulse (can t reduce below 1 cycle!) increasing frequency (but decreases depth of penetration)

50 Axial resolution

The Physics of Echo. The Physics of Echo. The Physics of Echo Is there pericardial calcification? 9/30/13

The Physics of Echo. The Physics of Echo. The Physics of Echo Is there pericardial calcification? 9/30/13 Basic Ultrasound Physics Kirk Spencer MD Speaker has no disclosures to make Sound Audible range 20Khz Medical ultrasound Megahertz range Advantages of imaging with ultrasound Directed as a beam Tomographic

More information

12/26/2017. Alberto Ardon M.D.

12/26/2017. Alberto Ardon M.D. Alberto Ardon M.D. 1 Preparatory Work Ultrasound Physics http://www.nysora.com/mobile/regionalanesthesia/foundations-of-us-guided-nerve-blockstechniques/index.1.html Basic Ultrasound Handling https://www.youtube.com/watch?v=q2otukhrruc

More information

Physics of ultrasound

Physics of ultrasound 1 Physics of ultrasound Basic principles Nature of ultrasound Sound = longitudinal, mechanical wave particles move parallel to direction of travel Audible sound < 20 khz Ultrasound > 20 khz Sound cannot

More information

Ultrasound Physics. History: Ultrasound 2/13/2019. Ultrasound

Ultrasound Physics. History: Ultrasound 2/13/2019. Ultrasound Ultrasound Physics History: Ultrasound Ultrasound 1942: Dr. Karl Theodore Dussik transmission ultrasound investigation of the brain 1949-51: Holmes and Howry subject submerged in water tank to achieve

More information

Lesson 02: Sound Wave Production. This lesson contains 24 slides plus 11 multiple-choice questions.

Lesson 02: Sound Wave Production. This lesson contains 24 slides plus 11 multiple-choice questions. Lesson 02: Sound Wave Production This lesson contains 24 slides plus 11 multiple-choice questions. Accompanying text for the slides in this lesson can be found on pages 2 through 7 in the textbook: ULTRASOUND

More information

The physics of ultrasound. Dr Graeme Taylor Guy s & St Thomas NHS Trust

The physics of ultrasound. Dr Graeme Taylor Guy s & St Thomas NHS Trust The physics of ultrasound Dr Graeme Taylor Guy s & St Thomas NHS Trust Physics & Instrumentation Modern ultrasound equipment is continually evolving This talk will cover the basics What will be covered?

More information

Physics of Ultrasound Ultrasound Imaging and Artifacts รศ.นพ.เดโช จ กราพาน ชก ล สาขาหท ยว ทยา, ภาคว ชาอาย รศาสตร คณะแพทยศาสตร ศ ร ราชพยาบาล

Physics of Ultrasound Ultrasound Imaging and Artifacts รศ.นพ.เดโช จ กราพาน ชก ล สาขาหท ยว ทยา, ภาคว ชาอาย รศาสตร คณะแพทยศาสตร ศ ร ราชพยาบาล Physics of Ultrasound Ultrasound Imaging and Artifacts รศ.นพ.เดโช จ กราพาน ชก ล สาขาหท ยว ทยา, ภาคว ชาอาย รศาสตร คณะแพทยศาสตร ศ ร ราชพยาบาล Diagnosis TTE TEE ICE 3D 4D Evaluation of Cardiac Anatomy Hemodynamic

More information

Interaction of Sound and. logarithms. Logarithms continued. Decibels (db) Decibels (db) continued. Interaction of Sound and Media continued

Interaction of Sound and. logarithms. Logarithms continued. Decibels (db) Decibels (db) continued. Interaction of Sound and Media continued Interaction of Sound and Media continued Interaction of Sound and Media Chapter 6 As sound travels through a media and interacts with normal anatomical structures its intensity weakens through what is

More information

Chapter 4. Pulse Echo Imaging. where: d = distance v = velocity t = time

Chapter 4. Pulse Echo Imaging. where: d = distance v = velocity t = time Chapter 4 Pulse Echo Imaging Ultrasound imaging systems are based on the principle of pulse echo imaging. These systems require the use of short pulses of ultrasound to create two-dimensional, sectional

More information

The question pool for the current General Class license took effect on July 1, 2007.

The question pool for the current General Class license took effect on July 1, 2007. General Class License Manual Errata and Corrections 22 Dec 2010 The following text is intended to support or correct the 6 th edition of the General Class License Manual. Determine the version of the manual

More information

Fig. 1

Fig. 1 PhysicsAndMathsTutor.com 1 1. Fig. 1 shows data for the intensity of a parallel beam of X-rays after penetration through varying thicknesses of a material. intensity / MW m 2 thickness / mm 0.91 0.40 0.69

More information

Dr.Rajaa CH.11. أ.م.د. رجاء سهيل جنم جامعة تكريت كلية طب االسنان

Dr.Rajaa CH.11. أ.م.د. رجاء سهيل جنم جامعة تكريت كلية طب االسنان Dr.Rajaa CH.11. أ.م.د. رجاء سهيل جنم طب االسنان Applications of Electricity and Magnetism In Medicine Electrical Shock: When an electrode is connected to each hand & 60 Hz currets of different levels are

More information

Ques on (2): [18 Marks] a) Draw the atrial synchronous Pacemaker block diagram and explain its operation. Benha University June 2013

Ques on (2): [18 Marks] a) Draw the atrial synchronous Pacemaker block diagram and explain its operation. Benha University June 2013 Benha University June 2013 Benha Faculty of Engineering Electrical Department Hospital Instrumentations (E472) 4 Th year (control) Dr.Waleed Abdel Aziz Salem Time: 3 Hrs Answer the following questions.

More information

Electrical Safety. Principles and examples of safety. History of (electrical) safety. Rules and regulations. Safety in the Hospital

Electrical Safety. Principles and examples of safety. History of (electrical) safety. Rules and regulations. Safety in the Hospital Electrical Safety History of (electrical) safety Safety in the Hospital Role of Clinical Engineer Rules and regulations Food and Drug Administration (FDA), Underwriter s Laboratory (UL), Federal Communication

More information

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging Medical Imaging X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging From: Physics for the IB Diploma Coursebook 6th Edition by Tsokos, Hoeben and Headlee And Higher Level Physics 2 nd Edition

More information

Medical Imaging (EL582/BE620/GA4426)

Medical Imaging (EL582/BE620/GA4426) Medical Imaging (EL582/BE620/GA4426) Jonathan Mamou, PhD Riverside Research Lizzi Center for Biomedical Engineering New York, NY jmamou@riversideresearch.org On behalf of Prof. Daniel Turnbull Outline

More information

Ultrasound & Artifacts

Ultrasound & Artifacts ISSN 2005-7881 Journal of Neurosonology 3(Suppl. 2):1-17, 2011 Ultrasound & Artifacts Siryung Han The Catholic University of Korea Artifacts False image- echoes without anatomic correlate US image dose

More information

Ultrasonic Linear Array Medical Imaging System

Ultrasonic Linear Array Medical Imaging System Ultrasonic Linear Array Medical Imaging System R. K. Saha, S. Karmakar, S. Saha, M. Roy, S. Sarkar and S.K. Sen Microelectronics Division, Saha Institute of Nuclear Physics, 1/AF Bidhannagar, Kolkata-700064.

More information

Attenuation and velocity of ultrasound in solid state materials (transmission)

Attenuation and velocity of ultrasound in solid state materials (transmission) Attenuation and velocity of ultrasound in solid 5.1.6.08 Related Topics Propagation of ultrasonic waves, time of flight, sound velocity, damping of ultrasonic waves (scattering, reflection, absorption),

More information

Sonic Distance Sensors

Sonic Distance Sensors Sonic Distance Sensors Introduction - Sound is transmitted through the propagation of pressure in the air. - The speed of sound in the air is normally 331m/sec at 0 o C. - Two of the important characteristics

More information

Electrical Safety Medical Instrumentation Application and Design, 4th Edition, Chapter 14

Electrical Safety Medical Instrumentation Application and Design, 4th Edition, Chapter 14 Electrical Safety Medical Instrumentation Application and Design, 4th Edition, Chapter 14 John G. Webster, Univ. of Wisconsin, Madison ISBN: 978-0-471-67600-3 Taught Matter of Lectures Introduction Basic

More information

I. Introduction to Animal Sensitivity and Response

I. Introduction to Animal Sensitivity and Response I. Introduction to Animal Sensitivity and Response The term stray voltage has been used to describe a special case of voltage developed on the grounded neutral system of a farm. If this voltage reaches

More information

Equipment for Attenuation and velocity of ultrasound in solid state materials (transmission), experimental set-up

Equipment for Attenuation and velocity of ultrasound in solid state materials (transmission), experimental set-up Attenuation and velocity of ultrasound in solid TEAS Related Topics Propagation of ultrasonic waves, time of flight, sound velocity, damping of ultrasonic waves (scattering, reflection, absorption), transmission

More information

INTRODUCTION. Have applications for imaging, detection and navigation.

INTRODUCTION. Have applications for imaging, detection and navigation. ULTRASONICS INTRODUCTION The word ultrasonic combines the Latin roots ultra - beyond sonic - sound. Having frequencies above the audible range i.e. above 20000Hz Have applications for imaging, detection

More information

Introduction to Ultrasound Physics

Introduction to Ultrasound Physics Introduction to Ultrasound Physics Vassilis Sboros Medical Physics and Cardiovascular Sciences University of Edinburgh Transverse waves Water remains in position Disturbance traverse producing more wave

More information

3. Ultrasound Imaging(2)

3. Ultrasound Imaging(2) 3. Ultrasound Imaging(2) Lecture 13, 14 Medical Imaging Systems Jae Gwan Kim, Ph.D. jaekim@gist.ac.kr, X 2220 Department of BioMedical Science and Engineering Gwangju Institute of Sciences and Technology

More information

Ultrasound Beamforming and Image Formation. Jeremy J. Dahl

Ultrasound Beamforming and Image Formation. Jeremy J. Dahl Ultrasound Beamforming and Image Formation Jeremy J. Dahl Overview Ultrasound Concepts Beamforming Image Formation Absorption and TGC Advanced Beamforming Techniques Synthetic Receive Aperture Parallel

More information

BME 701 Lecture 1. Measurement and Instrumentation

BME 701 Lecture 1. Measurement and Instrumentation BME 701 Lecture 1 Measurement and Instrumentation 1 Cochlear Implant 2 Advances in Vision (Retinal Stimulation) 3 Mini Gastric Imaging 4 5 Aspects of Measurement General Instrumentation Transducers (Electrodes)

More information

Ultrasound Bioinstrumentation. Topic 2 (lecture 3) Beamforming

Ultrasound Bioinstrumentation. Topic 2 (lecture 3) Beamforming Ultrasound Bioinstrumentation Topic 2 (lecture 3) Beamforming Angular Spectrum 2D Fourier transform of aperture Angular spectrum Propagation of Angular Spectrum Propagation as a Linear Spatial Filter Free

More information

BIO-ELECTRIC MEASUREMENTS

BIO-ELECTRIC MEASUREMENTS BIO-ELECTRIC MEASUREMENTS OBJECTIVES: 1) Determine the amplitude of the electrical "noise" in the body. 2) Observe and measure the characteristics and amplitudes of muscle potentials due to the biceps.

More information

Beamforming in ultrasound

Beamforming in ultrasound Peter Pazmany Catholic University Faculty of Information Technology www.itk.ppke.hu Medical diagnostic systems (Orvosbiológiai képalkotó rendszerek) Beamforming in ultrasound ( Nyalábalkotás az ultrahangban)

More information

Used to overcome ventricular fibrillation may be due to coronary occlusion, shock, or abnormalities in blood chemistry

Used to overcome ventricular fibrillation may be due to coronary occlusion, shock, or abnormalities in blood chemistry Used to overcome ventricular fibrillation may be due to coronary occlusion, shock, or abnormalities in blood chemistry Main problem: heart muscle fibers are continuously stimulated by adjacent muscles

More information

Optimisation of Image Acquisition Bordeaux 16th November J.S. McGhie W.B. Vletter R. Frowijn No disclosures

Optimisation of Image Acquisition Bordeaux 16th November J.S. McGhie W.B. Vletter R. Frowijn No disclosures Optimisation of Image Acquisition Bordeaux 16th November 2016 J.S. McGhie W.B. Vletter R. Frowijn No disclosures Image optimisation: The Echo machine It looks difficult to drive an echo machine!! Some

More information

Ultrasound Imaging Ultr Michael Dadd 2007

Ultrasound Imaging Ultr Michael Dadd 2007 Ultrasound Imaging Ultrasound Physics & Instrumentation - Recommended Reading 1. Diagnostic Ultrasound: Principles and Instruments (7th Ed) Frederick W Kremkau W B Saunders Company 2. Applied Physics &

More information

APPLICATION NOTE. Overview

APPLICATION NOTE. Overview Application Note 111 APPLICATION NOTE Nerve Conduction Velocity 42 Aero Camino, Goleta, CA 93117 Tel (805) 685-0066 Fax (805) 685-0067 info@biopac.com www.biopac.com 06.05.2018 This application note details

More information

MICROWAVE DIATHERMY AND SURGICAL DIATHERMY DIATHERMICS

MICROWAVE DIATHERMY AND SURGICAL DIATHERMY DIATHERMICS MICROWAVE DIATHERMY AND SURGICAL DIATHERMY 1 Microwave diathermy Microwave diathermy uses microwaves to generate heat in the body. It can be used to evenly warm deep tissues without heating the skin. Microwave

More information

EENG473 Mobile Communications Module 3 : Week # (12) Mobile Radio Propagation: Small-Scale Path Loss

EENG473 Mobile Communications Module 3 : Week # (12) Mobile Radio Propagation: Small-Scale Path Loss EENG473 Mobile Communications Module 3 : Week # (12) Mobile Radio Propagation: Small-Scale Path Loss Introduction Small-scale fading is used to describe the rapid fluctuation of the amplitude of a radio

More information

Explain what is meant by a photon and state one of its main properties [2]

Explain what is meant by a photon and state one of its main properties [2] 1 (a) A patient has an X-ray scan taken in hospital. The high-energy X-ray photons interact with the atoms inside the body of the patient. Explain what is meant by a photon and state one of its main properties....

More information

Lesson 06: Pulse-echo Imaging and Display Modes. These lessons contain 26 slides plus 15 multiple-choice questions.

Lesson 06: Pulse-echo Imaging and Display Modes. These lessons contain 26 slides plus 15 multiple-choice questions. Lesson 06: Pulse-echo Imaging and Display Modes These lessons contain 26 slides plus 15 multiple-choice questions. These lesson were derived from pages 26 through 32 in the textbook: ULTRASOUND IMAGING

More information

I. Introduction to Animal Sensitivity and Response

I. Introduction to Animal Sensitivity and Response Stray Voltage Field Guide Douglas J. Reinemann, Ph.D. Professor of Biological Systems Engineering University of Wisconsin Madison September 2007 Update I. Introduction to Animal Sensitivity and Response

More information

Biomedical Engineering Evoked Responses

Biomedical Engineering Evoked Responses Biomedical Engineering Evoked Responses Dr. rer. nat. Andreas Neubauer andreas.neubauer@medma.uni-heidelberg.de Tel.: 0621 383 5126 Stimulation of biological systems and data acquisition 1. How can biological

More information

NCERT solution for Sound

NCERT solution for Sound NCERT solution for Sound 1 Question 1 How does the sound produce by a vibrating object in a medium reach your ear? When an object vibrates, it vibrates the neighboring particles of the medium. These vibrating

More information

Physics of Ultrasound & Doppler. Sang Jae Rhee. MD., PhD. Division of Cardiovascular Medicine Wonkwang University Hospital

Physics of Ultrasound & Doppler. Sang Jae Rhee. MD., PhD. Division of Cardiovascular Medicine Wonkwang University Hospital Physics of Ultrasound & Doppler Sang Jae Rhee. MD., PhD. Division of Cardiovascular Medicine Wonkwang University Hospital Classification of Sound Infrasound Audible sound Ultrasound < 20 Hz 20-20,000 Hz

More information

Sonicator Plus 994 Specifications

Sonicator Plus 994 Specifications Sonicator Plus 994 Specifications General Specifications: Input: Certification: Domestic model ETL and C-ETL Listed: Domestic model Classification: CE model Year 2000 Compliant Weight: Dimensions: Operating

More information

Properties and Applications

Properties and Applications Properties and Applications What is a Wave? How is it Created? Waves are created by vibrations! Atoms vibrate, strings vibrate, water vibrates A wave is the moving oscillation Waves are the propagation

More information

Chapter 7. Waves and Sound

Chapter 7. Waves and Sound Chapter 7 Waves and Sound What is wave? A wave is a disturbance that propagates from one place to another. Or simply, it carries energy from place to place. The easiest type of wave to visualize is a transverse

More information

Introduction to Medical Engineering (Medical Imaging) Ultrasound Imaging. Ho Kyung Kim Pusan National University

Introduction to Medical Engineering (Medical Imaging) Ultrasound Imaging. Ho Kyung Kim Pusan National University Introduction to Medical Engineering (Medical Imaging) Suetens 6 Ultrasound Imaging Ho Kyung Kim Pusan National University Sound Sonic: 20 Hz 20 khz (audible frequency) Subsonic () Ultrasound

More information

PhysicsAndMathsTutor.com 1

PhysicsAndMathsTutor.com 1 PhysicsAndMathsTutor.com. A ear drum [or tympanic membrane] () transfers sound waves from the outer ear to the ossicles of the middle ear () B ossicles [or bones of the middle ear] () system of levers

More information

*Notebook is excluded

*Notebook is excluded Biomedical Measurement Training System This equipment is designed for students to learn how to design specific measuring circuits and detect the basic physiological signals with practical operation. Moreover,

More information

Wave Review Questions Updated

Wave Review Questions Updated Name: Date: 1. Which type of wave requires a material medium through which to travel? 5. Which characteristic is the same for every color of light in a vacuum? A. radio wave B. microwave C. light wave

More information

Frequently Asked Question on Isolated Power Supply(IPS)

Frequently Asked Question on Isolated Power Supply(IPS) Frequently Asked Question on Isolated Power Supply(IPS) Q: What is an IPS? A: IPS is called as Isolated Power Supply or System and is used in a Hospital to ensure safe power supply to OT Operation Theaters

More information

used to diagnose and treat medical conditions. State the precautions necessary when X ray machines and CT scanners are used.

used to diagnose and treat medical conditions. State the precautions necessary when X ray machines and CT scanners are used. Page 1 State the properties of X rays. Describe how X rays can be used to diagnose and treat medical conditions. State the precautions necessary when X ray machines and CT scanners are used. What is meant

More information

COMPUTER PHANTOMS FOR SIMULATING ULTRASOUND B-MODE AND CFM IMAGES

COMPUTER PHANTOMS FOR SIMULATING ULTRASOUND B-MODE AND CFM IMAGES Paper presented at the 23rd Acoustical Imaging Symposium, Boston, Massachusetts, USA, April 13-16, 1997: COMPUTER PHANTOMS FOR SIMULATING ULTRASOUND B-MODE AND CFM IMAGES Jørgen Arendt Jensen and Peter

More information

OPERATING MANUAL MINIDOP ES-100VX POCKET DOPPLER

OPERATING MANUAL MINIDOP ES-100VX POCKET DOPPLER OPERATING MANUAL MINIDOP ES-100VX POCKET DOPPLER CONTENTS * Features.......................... 1 * Cautions.......................... 2 * Clinical applications................. 4 * Operating controls..................

More information

4.6.1 Waves in air, fluids and solids Transverse and longitudinal waves Properties of waves

4.6.1 Waves in air, fluids and solids Transverse and longitudinal waves Properties of waves 4.6 Waves Wave behaviour is common in both natural and man-made systems. Waves carry energy from one place to another and can also carry information. Designing comfortable and safe structures such as bridges,

More information

Ultrasonic Level Detection Technology. ultra-wave

Ultrasonic Level Detection Technology. ultra-wave Ultrasonic Level Detection Technology ultra-wave 1 Definitions Sound - The propagation of pressure waves through air or other media Medium - A material through which sound can travel Vacuum - The absence

More information

A Study on analysis of intracranial acoustic wave propagation by the finite difference time domain method

A Study on analysis of intracranial acoustic wave propagation by the finite difference time domain method A Stud on analsis of intracranial acoustic wave propagation b the finite difference time domain method 4.5 Wa Biological effects of ultrasound, ultrasonic tomograph Yoko Tanikaga, Toshikazu Takizawa, Takefumi

More information

Intext Exercise 1 Question 1: How does the sound produced by a vibrating object in a medium reach your ear?

Intext Exercise 1 Question 1: How does the sound produced by a vibrating object in a medium reach your ear? Intext Exercise 1 How does the sound produced by a vibrating object in a medium reach your ear? When an vibrating object vibrates, it forces the neighbouring particles of the medium to vibrate. These vibrating

More information

Sensors. CSE 666 Lecture Slides SUNY at Buffalo

Sensors. CSE 666 Lecture Slides SUNY at Buffalo Sensors CSE 666 Lecture Slides SUNY at Buffalo Overview Optical Fingerprint Imaging Ultrasound Fingerprint Imaging Multispectral Fingerprint Imaging Palm Vein Sensors References Fingerprint Sensors Various

More information

EXPERIMENT 8 Bio-Electric Measurements

EXPERIMENT 8 Bio-Electric Measurements EXPERIMENT 8 Bio-Electric Measurements Objectives 1) Determine the amplitude of some electrical signals in the body. 2) Observe and measure the characteristics and amplitudes of muscle potentials due to

More information

Physics I Notes: Chapter 13 Sound

Physics I Notes: Chapter 13 Sound Physics I Notes: Chapter 13 Sound I. Properties of Sound A. Sound is the only thing that one can hear! Where do sounds come from?? Sounds are produced by VIBRATING or OSCILLATING OBJECTS! Sound is a longitudinal

More information

Anaesthesia. Update in. Electricity and Magnetism. David Sanders, Bruce McCormick* *Correspondence

Anaesthesia. Update in. Electricity and Magnetism. David Sanders, Bruce McCormick* *Correspondence Update in Anaesthesia Electricity and Magnetism David Sanders, Bruce McCormick* *Correspondence Email: Bruce.McCormick@rdeft.nhs.uk ELECTRICAL CHARGE Charge is a fundamental property of matter. Atoms,

More information

Fundamentals of Pacing Therapy

Fundamentals of Pacing Therapy 1 4 7 2 5 8 3 6 9 Fundamentals of Pacing Therapy This presentation is provided with the understanding that the slide content must not be altered in any manner as the content is subject to FDA regulations.

More information

Application Note 175 Using the STMISOC Stimulus Isolator

Application Note 175 Using the STMISOC Stimulus Isolator APPLICATION NOTE 42 Aero Camino, Goleta, CA 93117 Tel (805) 685-0066 Fax (805) 685-0067 info@biopac.com www.biopac.com 02.14.2018 Application Note 175 Using the The MP160/150 System stimulation features

More information

Acoustic resolution. photoacoustic Doppler velocimetry. in blood-mimicking fluids. Supplementary Information

Acoustic resolution. photoacoustic Doppler velocimetry. in blood-mimicking fluids. Supplementary Information Acoustic resolution photoacoustic Doppler velocimetry in blood-mimicking fluids Joanna Brunker 1, *, Paul Beard 1 Supplementary Information 1 Department of Medical Physics and Biomedical Engineering, University

More information

describe sound as the transmission of energy via longitudinal pressure waves;

describe sound as the transmission of energy via longitudinal pressure waves; 1 Sound-Detailed Study Study Design 2009 2012 Unit 4 Detailed Study: Sound describe sound as the transmission of energy via longitudinal pressure waves; analyse sound using wavelength, frequency and speed

More information

ELECTRICAL CURRENT FOR PAIN CONTROL

ELECTRICAL CURRENT FOR PAIN CONTROL ELECTRICAL CURRENT FOR PAIN CONTROL 2 Pain modulation ES for pain modulation (TENS) Parameter for ES for pain modulation Pain modulation 3 Gate control theory of pain modulation Pain perception nociceptor

More information

DIATHERMY UNITS HIGH FREQUENCY HEAT THERAPY:

DIATHERMY UNITS HIGH FREQUENCY HEAT THERAPY: DIATHERMY UNITS HIGH FREQUENCY HEAT THERAPY: The dipole molecules of the body are normally placed randomly. Under electric field, they rotate in the direction of the field lines so that the positively

More information

Nuove tecnologie per ecografia ad ultrasuoni: da 2D a 4D

Nuove tecnologie per ecografia ad ultrasuoni: da 2D a 4D DINFO Dipartimento di Ingegneria dell Informazione Department of Information Engineering Nuove tecnologie per ecografia ad ultrasuoni: da 2D a 4D Piero Tortoli Microelectronics Systems Design Lab 1 Introduction

More information

How are X-ray slides formed?

How are X-ray slides formed? P3 Revision. How are X-ray slides formed? X-rays can penetrate soft tissue but not bone. X-rays are absorbed more by some materials than others. Photographic film can be used to detect X-rays, but these

More information

ACOUSTIC MICRO IMAGING ANALYSIS METHODS FOR 3D PACKAGES

ACOUSTIC MICRO IMAGING ANALYSIS METHODS FOR 3D PACKAGES ACOUSTIC MICRO IMAGING ANALYSIS METHODS FOR 3D PACKAGES Janet E. Semmens Sonoscan, Inc. Elk Grove Village, IL, USA Jsemmens@sonoscan.com ABSTRACT Earlier studies concerning evaluation of stacked die packages

More information

4.6 Waves Waves in air, fluids and solids Transverse and longitudinal waves

4.6 Waves Waves in air, fluids and solids Transverse and longitudinal waves 4.6 Waves Wave behaviour is common in both natural and man-made systems. Waves carry energy from one place to another and can also carry information. Designing comfortable and safe structures such as bridges,

More information

Floating Circuits - Protection against electric shock

Floating Circuits - Protection against electric shock Floating Circuits - Protection against electric shock Introduction TECHNICALLY SPEAKING by Richard Nute 1998 Richard Nute FLOATING CIRCUITS PROTECTION AGAINST ELECTRIC SHOCK Probably the least-understood

More information

Introductory Physics, High School Learning Standards for a Full First-Year Course

Introductory Physics, High School Learning Standards for a Full First-Year Course Introductory Physics, High School Learning Standards for a Full First-Year Course I. C ONTENT S TANDARDS 4.1 Describe the measurable properties of waves (velocity, frequency, wavelength, amplitude, period)

More information

Real Time Deconvolution of In-Vivo Ultrasound Images

Real Time Deconvolution of In-Vivo Ultrasound Images Paper presented at the IEEE International Ultrasonics Symposium, Prague, Czech Republic, 3: Real Time Deconvolution of In-Vivo Ultrasound Images Jørgen Arendt Jensen Center for Fast Ultrasound Imaging,

More information

Design, Development and Characterization of. Wideband Polymer Ultrasonic Probes. for Medical Ultrasound Applications. A Thesis

Design, Development and Characterization of. Wideband Polymer Ultrasonic Probes. for Medical Ultrasound Applications. A Thesis Design, Development and Characterization of Wideband Polymer Ultrasonic Probes for Medical Ultrasound Applications A Thesis Submitted to the Faculty of Drexel University by Vadivel Devaraju in partial

More information

Section 1: Sound. Sound and Light Section 1

Section 1: Sound. Sound and Light Section 1 Sound and Light Section 1 Section 1: Sound Preview Key Ideas Bellringer Properties of Sound Sound Intensity and Decibel Level Musical Instruments Hearing and the Ear The Ear Ultrasound and Sonar Sound

More information

SODAR- sonic detecting and ranging

SODAR- sonic detecting and ranging Active Remote Sensing of the PBL Immersed vs. remote sensors Active vs. passive sensors RADAR- radio detection and ranging WSR-88D TDWR wind profiler SODAR- sonic detecting and ranging minisodar RASS RADAR

More information

The Middle East Distributor for AMBISEA Technology Corp. Electro-Medical Product Line

The Middle East Distributor for AMBISEA Technology Corp. Electro-Medical Product Line The Middle East Distributor for AMBISEA Technology Corp. Electro-Medical Product Line AV-9100 Single Channel ECG 1 2 AV-9300 3-Channels ECG 3 4 5 AV-9000B Multi-Parameter Patient Monitor 6 7 8 AV-9000C

More information

Chapter 12. Preview. Objectives The Production of Sound Waves Frequency of Sound Waves The Doppler Effect. Section 1 Sound Waves

Chapter 12. Preview. Objectives The Production of Sound Waves Frequency of Sound Waves The Doppler Effect. Section 1 Sound Waves Section 1 Sound Waves Preview Objectives The Production of Sound Waves Frequency of Sound Waves The Doppler Effect Section 1 Sound Waves Objectives Explain how sound waves are produced. Relate frequency

More information

A VIEW OF ELECTROMAGNETIC LIFE ABOVE 100 MHz

A VIEW OF ELECTROMAGNETIC LIFE ABOVE 100 MHz A VIEW OF ELECTROMAGNETIC LIFE ABOVE 100 MHz An Experimentalist's Intuitive Approach Lothar O. (Bud) Hoeft, PhD Consultant, Electromagnetic Effects 5012 San Pedro Ct., NE Albuquerque, NM 87109-2515 (505)

More information

A SHEAR WAVE TRANSDUCER ARRAY FOR REAL-TIME IMAGING. R.L. Baer and G.S. Kino. Edward L. Ginzton Laboratory Stanford University Stanford, CA 94305

A SHEAR WAVE TRANSDUCER ARRAY FOR REAL-TIME IMAGING. R.L. Baer and G.S. Kino. Edward L. Ginzton Laboratory Stanford University Stanford, CA 94305 A SHEAR WAVE TRANSDUCER ARRAY FOR REAL-TIME IMAGING R.L. Baer and G.S. Kino Edward L. Ginzton Laboratory Stanford University Stanford, CA 94305 INTRODUCTION In this paper we describe a contacting shear

More information

INSPECTION OF THERMAL BARRIERS OF PRIMARY PUMPS WITH PHASED ARRAY PROBE AND PIEZOCOMPOSITE TECHNOLOGY

INSPECTION OF THERMAL BARRIERS OF PRIMARY PUMPS WITH PHASED ARRAY PROBE AND PIEZOCOMPOSITE TECHNOLOGY INSPECTION OF THERMAL BARRIERS OF PRIMARY PUMPS WITH PHASED ARRAY PROBE AND PIEZOCOMPOSITE TECHNOLOGY J. Poguet Imasonic S.A. France E. Abittan EDF-GDL France Abstract In order to meet the requirements

More information

ISSUED BY KENDRIYA VIDYALAYA - DOWNLOADED FROM CHAPTER 12 Sound

ISSUED BY KENDRIYA VIDYALAYA - DOWNLOADED FROM  CHAPTER 12 Sound 1. Production of Sound CHAPTER 12 Sound KEY CONCEPTS [ *rating as per the significance of concept] 1 Production of Sound **** 2 Propagation of Sound ***** 3 Reflection of Sound ***** 4 Echo **** 5 Uses

More information

Artifacts. Artifacts. Causes. Imaging assumptions. Common terms used to describe US images. Common terms used to describe US images

Artifacts. Artifacts. Causes. Imaging assumptions. Common terms used to describe US images. Common terms used to describe US images Artifacts Artifacts Chapter 20 What are they? Simply put they are an error in imaging These artifacts include reflections that are: not real incorrect shape, size or position incorrect brightness displayed

More information

INSTITUTE OF AERONAUTICAL ENGINEERING Dundigal, Hyderabad ELECTRONICS AND COMMUNIACTION ENGINEERING QUESTION BANK

INSTITUTE OF AERONAUTICAL ENGINEERING Dundigal, Hyderabad ELECTRONICS AND COMMUNIACTION ENGINEERING QUESTION BANK INSTITUTE OF AERONAUTICAL ENGINEERING Dundigal, Hyderabad - 500 04 ELECTRONICS AND COMMUNIACTION ENGINEERING QUESTION BANK Course Name : Antennas and Wave Propagation (AWP) Course Code : A50418 Class :

More information

Module 1: Introduction to Experimental Techniques Lecture 2: Sources of error. The Lecture Contains: Sources of Error in Measurement

Module 1: Introduction to Experimental Techniques Lecture 2: Sources of error. The Lecture Contains: Sources of Error in Measurement The Lecture Contains: Sources of Error in Measurement Signal-To-Noise Ratio Analog-to-Digital Conversion of Measurement Data A/D Conversion Digitalization Errors due to A/D Conversion file:///g /optical_measurement/lecture2/2_1.htm[5/7/2012

More information

Electrocardiogram (ECG)

Electrocardiogram (ECG) Vectors and ECG s Vectors and ECG s 2 Electrocardiogram (ECG) Depolarization wave passes through the heart and the electrical currents pass into surrounding tissues. Small part of the extracellular current

More information

IMAGING OF DEFECTS IN CONCRETE COMPONENTS WITH NON-CONTACT ULTRASONIC TESTING W. Hillger, DLR and Ing. Büro Dr. Hillger, Braunschweig, Germany

IMAGING OF DEFECTS IN CONCRETE COMPONENTS WITH NON-CONTACT ULTRASONIC TESTING W. Hillger, DLR and Ing. Büro Dr. Hillger, Braunschweig, Germany IMAGING OF DEFECTS IN CONCRETE COMPONENTS WITH NON-CONTACT ULTRASONIC TESTING W. Hillger, DLR and Ing. Büro Dr. Hillger, Braunschweig, Germany Abstract: The building industries require NDT- methods for

More information

1. Introduction. 2. Concept. reflector. transduce r. node. Kraftmessung an verschiedenen Fluiden in akustischen Feldern

1. Introduction. 2. Concept. reflector. transduce r. node. Kraftmessung an verschiedenen Fluiden in akustischen Feldern 1. Introduction The aim of this Praktikum is to familiarize with the concept and the equipment of acoustic levitation and to measure the forces exerted by an acoustic field on small spherical objects.

More information

Electrical noise in the OR

Electrical noise in the OR Electrical noise in the OR Chris Thompson Senior Staff Specialist Royal Prince Alfred Hospital SYDNEY SOUTH WEST AREA HEALTH SERVICE NSW HEALTH Electrical noise in the OR Root causes Tiny little signals

More information

BENG 186B Principles of Bioinstrumentation. Week 7 Review. Solutions

BENG 186B Principles of Bioinstrumentation. Week 7 Review. Solutions BENG 186B Principles of Bioinstrumentation Week 7 Review Solutions Selections from: 2015 Homework 5 2015 Homework 6 C d = 0.001 1 2 1.5 Normalized Voltage 1 0.5 0-0.5-1 -1.5-2 Time A B C b C b BENG 186B

More information

Chapter 05: Wave Motions and Sound

Chapter 05: Wave Motions and Sound Chapter 05: Wave Motions and Sound Section 5.1: Forces and Elastic Materials Elasticity It's not just the stretch, it's the snap back An elastic material will return to its original shape when stretched

More information

Ihor TROTS, Andrzej NOWICKI, Marcin LEWANDOWSKI

Ihor TROTS, Andrzej NOWICKI, Marcin LEWANDOWSKI ARCHIVES OF ACOUSTICS 33, 4, 573 580 (2008) LABORATORY SETUP FOR SYNTHETIC APERTURE ULTRASOUND IMAGING Ihor TROTS, Andrzej NOWICKI, Marcin LEWANDOWSKI Institute of Fundamental Technological Research Polish

More information

Waves, Sound and Light. Grade 10 physics Robyn Basson

Waves, Sound and Light. Grade 10 physics Robyn Basson Waves, Sound and Light Grade 10 physics Robyn Basson Heartbeat Flick in hose pipe What is a pulse? A single disturbance that moves through a medium. Stone in water Other? moving Transverse pulse: A pulse

More information

P6 Quick Revision Questions

P6 Quick Revision Questions P6 Quick Revision Questions H = Higher tier only SS = Separate science only Question 1... of 50 Define wavelength Answer 1... of 50 The distance from a point on one wave to the equivalent point on the

More information

BME 3113, Dept. of BME Lecture on Introduction to Biosignal Processing

BME 3113, Dept. of BME Lecture on Introduction to Biosignal Processing What is a signal? A signal is a varying quantity whose value can be measured and which conveys information. A signal can be simply defined as a function that conveys information. Signals are represented

More information

CHAPTER 7 INTERFERENCE CANCELLATION IN EMG SIGNAL

CHAPTER 7 INTERFERENCE CANCELLATION IN EMG SIGNAL 131 CHAPTER 7 INTERFERENCE CANCELLATION IN EMG SIGNAL 7.1 INTRODUCTION Electromyogram (EMG) is the electrical activity of the activated motor units in muscle. The EMG signal resembles a zero mean random

More information

DOPPLER RADAR. Doppler Velocities - The Doppler shift. if φ 0 = 0, then φ = 4π. where

DOPPLER RADAR. Doppler Velocities - The Doppler shift. if φ 0 = 0, then φ = 4π. where Q: How does the radar get velocity information on the particles? DOPPLER RADAR Doppler Velocities - The Doppler shift Simple Example: Measures a Doppler shift - change in frequency of radiation due to

More information

Answer:- School bell starts vibrating when heated which creates compression and rarefaction in air and sound is produced.

Answer:- School bell starts vibrating when heated which creates compression and rarefaction in air and sound is produced. Sound How does the sound produced by a vibrating object in a medium reach your ear? - Vibrations in an object create disturbance in the medium and consequently compressions and rarefactions. Because of

More information