Electrical Safety. Principles and examples of safety. History of (electrical) safety. Rules and regulations. Safety in the Hospital

Size: px
Start display at page:

Download "Electrical Safety. Principles and examples of safety. History of (electrical) safety. Rules and regulations. Safety in the Hospital"

Transcription

1 Electrical Safety History of (electrical) safety Safety in the Hospital Role of Clinical Engineer Rules and regulations Food and Drug Administration (FDA), Underwriter s Laboratory (UL), Federal Communication (FCC) Principles and examples of safety

2 Physiological Response to Electricity Electrical shock from devices in the home to lightening Macroshock (lightning, instruments) Microshock (implantable catheters, leakage) Susceptibility Parameters Frequency, Duration, Body weight, Point-of-entry

3 Frequency Susceptibility Note: the thresholds for preception, let-go currents.

4 Duration/Body Weight Susceptibility Strengthduration curve is a very important plot, applicable to shocks, pacemaker/defib rillator type devices, etc. Note the chronaxie point (at the lower left corner of each curve)

5 Current vs. Frequency Difference in DC vs AC shocks Note that the threshold is lowest at the powerline frequencies! Very high threshold/low sensitivity at high frequencies used in many medical applications impediance check, respiration, communications.

6 Isolated-Power Systems Transformer isolation Transformer isolation used in hospital based clinical instruments: Isolation of power supply and also of the instrument/circuit

7 Isolated-Power Systems Optical Isolation Simple LED, photodiode circuit Advanced circuit for linearization from the book Optical isolation is used in simple instrument/circuit designs; e.g. interfacing ECG amplifier to the outside world; optical path provides break of electrical continuity usually interrupting path to ground for the leakage current

8 Micro- vs. Macroshock Hazards Points of Entry Microshock occurs because there is a low resistance invasive path to the heart Note the path to ground in any microshock circuit Microshock due to ventricular fibrillation Lowered safety in hospital, bathroom

9 Sources of Microshock Failure of power outlet,cord Failure of transformer Failure of catheter/lead Ground path and ground loop leakage

10 Sources of Macroshock Direct contact to Live power supply Live instrument chassis Failed transformer Electrical spark/discharge Lightning

11 Conductive Paths to the Heart Pacemaker Leads Epi- or endocardial electrodes Intracardiac electrogram (EGM) electrodes Liquid-filled Catheters for: Blood pressure monitoring Blood sampling Drug/dye injection Ground loop and inadvertent leakage path

12 Safety Codes & Standards NFPA 99 Standard for Health Care Facilities National Electrical Code Article 517: Health Care Facilities Association for the Advancement of Medical Instrumentation (AAMI) Developed American National Standard on Safe Current Limits for Electromedical Apparatus

13 Leakage Current Limits Electric Appliance Not intended for patient contact Not intended for patient contact and single fault With nonisolated patient leads With nonisolated leads and single fault With isolated patient leads Chassis Leakage, µa Patient-lead Leakage, µa N/A N/A With isolated leads and single fault

14 Protection Circuits Pole-to-opposite pole. Figure 2A depicts the circuit path when a man simultaneously touches both poles of the floating or isolated circuit. In this situation, there is no insulation. The current is limited only by the impedance of the body.

15 Protection Circuits If we interpose an insulating barrier between one pole of the floating circuit and the man, then we can define that barrier as Basic Insulation. In the event of failure of that Basic Insulation, there is no electric shock current in the man. If we extend that same insulation such that it is interposed between the OPPOSITE pole and the man, then we can define the OPPOSITE pole portion of the insulation as Supplementary Insulation (because it provides insulation against the SECOND body connection).

16 References Webster, JG (1998). Medical Instrumentation. John Wiley & Sons, Inc., New York, NY. Chapters 1 & 14. Nute, R (1998). Floating circuits: protection against electric shock. In Technically Speaking section of Product Safety Technical Committee Newsletter (online).

17 Problems Describe a sensor or a measurement system in which accuracy is important. In contrast, describe a sensor or a measurement in which precision is important. A temperature sensor, such as a thermistor can be described by a first order system. Write down the general equation for a first order system (you can write a differential equation or a transfer function). Plot the output of the first order system in response to a step change in temperature. A blood pressure sensor is described by a second order system. Write down the general equation for a second order system (you can write a differential equation or a transfer function). Plot the output of the second order underdamped pressure system in response to a blood pressure signal. Systolic pressure Distolic pressure Dichrotic notch

18 PROBLEMS Draw the circuit of a differentiator OR integrator. Now, through circuit analysis show why that circuit works like a differentiator/integrator (i.e. derive the relationship between output and input of the amplifier to show using your equation that the input signal is differentiated/integrated. Next, obtain the frequency response of this circuit. E.g. you can derive a transfer function (output over input) in the Laplace form, substitute s=jω and then show the frequency response. Point out one major advantage and one major disadvantage of an analog differentiator/integrator over a digital/software version. Describe (very briefly) a biomedical instrumentation application of an integrator or differentiator.

Electrical Safety Medical Instrumentation Application and Design, 4th Edition, Chapter 14

Electrical Safety Medical Instrumentation Application and Design, 4th Edition, Chapter 14 Electrical Safety Medical Instrumentation Application and Design, 4th Edition, Chapter 14 John G. Webster, Univ. of Wisconsin, Madison ISBN: 978-0-471-67600-3 Taught Matter of Lectures Introduction Basic

More information

Floating Circuits - Protection against electric shock

Floating Circuits - Protection against electric shock Floating Circuits - Protection against electric shock Introduction TECHNICALLY SPEAKING by Richard Nute 1998 Richard Nute FLOATING CIRCUITS PROTECTION AGAINST ELECTRIC SHOCK Probably the least-understood

More information

Frequently Asked Question on Isolated Power Supply(IPS)

Frequently Asked Question on Isolated Power Supply(IPS) Frequently Asked Question on Isolated Power Supply(IPS) Q: What is an IPS? A: IPS is called as Isolated Power Supply or System and is used in a Hospital to ensure safe power supply to OT Operation Theaters

More information

Introduction to Biomedical Engineering

Introduction to Biomedical Engineering Introduction to Biomedical Engineering Biomedical Instrumentation Kung-Bin Sung 5/8/007 Outline Chapter 8 and chapter 5 of st edition: Bioinstrumentation Bridge circuit Operational amplifiers, instrumentation

More information

Biomedical. Measurement and Design ELEC4623/ELEC9734. Electrical Safety and Performance Standards

Biomedical. Measurement and Design ELEC4623/ELEC9734. Electrical Safety and Performance Standards Biomedical Instrumentation, Measurement and Design ELEC4623/ELEC9734 Electrical Safety and Performance Standards Contents Physiological Effects of Electrical Currents Safety Standards for Medical Instrumentation

More information

To Float or Not to Float? Analysis of a floating vs. grounded output Associated Power Technologies

To Float or Not to Float? Analysis of a floating vs. grounded output Associated Power Technologies To Float or Not to Float? Analysis of a floating vs. grounded output Associated Power Technologies Introduction In electrical circuits, voltage is always measured between two points: a point of high potential

More information

BME 701 Lecture 1. Measurement and Instrumentation

BME 701 Lecture 1. Measurement and Instrumentation BME 701 Lecture 1 Measurement and Instrumentation 1 Cochlear Implant 2 Advances in Vision (Retinal Stimulation) 3 Mini Gastric Imaging 4 5 Aspects of Measurement General Instrumentation Transducers (Electrodes)

More information

Biomedical Sensor Systems Laboratory. Institute for Neural Engineering Graz University of Technology

Biomedical Sensor Systems Laboratory. Institute for Neural Engineering Graz University of Technology Biomedical Sensor Systems Laboratory Institute for Neural Engineering Graz University of Technology 2017 Bioinstrumentation Measurement of physiological variables Invasive or non-invasive Minimize disturbance

More information

Fundamentals of Pacing Therapy

Fundamentals of Pacing Therapy 1 4 7 2 5 8 3 6 9 Fundamentals of Pacing Therapy This presentation is provided with the understanding that the slide content must not be altered in any manner as the content is subject to FDA regulations.

More information

BENG 186B Winter 2013 Final

BENG 186B Winter 2013 Final Name (Last, First): BENG 186B Winter 2013 Final This exam is closed book, closed note, calculators are OK. Circle and put your final answers in the space provided; show your work only on the pages provided.

More information

A Comprehensive Model for Power Line Interference in Biopotential Measurements

A Comprehensive Model for Power Line Interference in Biopotential Measurements IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, VOL. 49, NO. 3, JUNE 2000 535 A Comprehensive Model for Power Line Interference in Biopotential Measurements Mireya Fernandez Chimeno, Member, IEEE,

More information

Used to overcome ventricular fibrillation may be due to coronary occlusion, shock, or abnormalities in blood chemistry

Used to overcome ventricular fibrillation may be due to coronary occlusion, shock, or abnormalities in blood chemistry Used to overcome ventricular fibrillation may be due to coronary occlusion, shock, or abnormalities in blood chemistry Main problem: heart muscle fibers are continuously stimulated by adjacent muscles

More information

Technical data ESA614 Electrical Safety Analyzer

Technical data ESA614 Electrical Safety Analyzer Technical data ESA614 Electrical Safety Analyzer The ESA614 Electrical Safety Analyzer brings fast and simple automated testing in the form of a portable analyzer to healthcare technology professionals

More information

BENG 186B Principles of Bioinstrumentation. Week 7 Review. Solutions

BENG 186B Principles of Bioinstrumentation. Week 7 Review. Solutions BENG 186B Principles of Bioinstrumentation Week 7 Review Solutions Selections from: 2015 Homework 5 2015 Homework 6 C d = 0.001 1 2 1.5 Normalized Voltage 1 0.5 0-0.5-1 -1.5-2 Time A B C b C b BENG 186B

More information

EE 230 Experiment 10 ECG Measurements Spring 2010

EE 230 Experiment 10 ECG Measurements Spring 2010 EE 230 Experiment 10 ECG Measurements Spring 2010 Note: If for any reason the students are uncomfortable with doing this experiment, please talk to the instructor for the course and an alternative experiment

More information

Using Optical Isolation Amplifiers in Power Inverters for Voltage, Current and Temperature Sensing

Using Optical Isolation Amplifiers in Power Inverters for Voltage, Current and Temperature Sensing Using Optical Isolation Amplifiers in Power Inverters for Voltage, Current and Temperature Sensing by Hong Lei Chen, Product Manager, Avago Technologies Abstract Many industrial equipments and home appliances

More information

Dr.Rajaa CH.11. أ.م.د. رجاء سهيل جنم جامعة تكريت كلية طب االسنان

Dr.Rajaa CH.11. أ.م.د. رجاء سهيل جنم جامعة تكريت كلية طب االسنان Dr.Rajaa CH.11. أ.م.د. رجاء سهيل جنم طب االسنان Applications of Electricity and Magnetism In Medicine Electrical Shock: When an electrode is connected to each hand & 60 Hz currets of different levels are

More information

GSM based Patient monitoring system

GSM based Patient monitoring system For more Project details visit: http://www.projectsof8051.com/patient-monitoring-through-gsm-modem/ Code Project Title 1615 GSM based Patient monitoring system Synopsis for GSM based Patient monitoring

More information

EG medlab. Three Lead ECG OEM board. Version Technical Manual. Medlab GmbH Three Lead ECG OEM Module EG01010 User Manual

EG medlab. Three Lead ECG OEM board. Version Technical Manual. Medlab GmbH Three Lead ECG OEM Module EG01010 User Manual Medlab GmbH Three Lead ECG OEM Module EG01010 User Manual medlab Three Lead ECG OEM board EG01010 Technical Manual Copyright Medlab 2008-2016 Version 1.03 1 Version 1.03 28.04.2016 Medlab GmbH Three Lead

More information

Anaesthesia. Update in. Electricity and Magnetism. David Sanders, Bruce McCormick* *Correspondence

Anaesthesia. Update in. Electricity and Magnetism. David Sanders, Bruce McCormick* *Correspondence Update in Anaesthesia Electricity and Magnetism David Sanders, Bruce McCormick* *Correspondence Email: Bruce.McCormick@rdeft.nhs.uk ELECTRICAL CHARGE Charge is a fundamental property of matter. Atoms,

More information

Application Notes. Electrosurgical Analyzer Primer

Application Notes. Electrosurgical Analyzer Primer Electrosurgical Analyzer Primer Purpose Scope Overview The purpose is to outline the characteristics of electrosurgical analyzers, describe the challenges of measuring RF power and review the inspection

More information

BENG 186B Winter 2012 Quiz 3. March 7, NAME (Last, First): This quiz is closed book and closed note. You may use a calculator for algebra.

BENG 186B Winter 2012 Quiz 3. March 7, NAME (Last, First): This quiz is closed book and closed note. You may use a calculator for algebra. BENG 186B Winter 2012 Quiz 3 March 7, 2012 NAME (Last, First): This quiz is closed book and closed note. You may use a calculator for algebra. Circle your final answers in the space provided; show your

More information

Testing the Ground Circuit

Testing the Ground Circuit Ground of electrical products Class I vs. Class II products Ground Continuity Test Ground Bond Test What is tested during each test Testing the Ground Circuit Meet Our Team Webinar Notes Please use the

More information

ECE 528 Understanding Power Quality

ECE 528 Understanding Power Quality ECE 528 Understanding Power Quality http://www.ece.uidaho.edu/ee/power/ece528/ Paul Ortmann portmann@uidaho.edu 208-316-1520 (voice) 1 Today Wiring and grounding Why it s important References Terms and

More information

Special-Purpose Operational Amplifier Circuits

Special-Purpose Operational Amplifier Circuits Special-Purpose Operational Amplifier Circuits Instrumentation Amplifier An instrumentation amplifier (IA) is a differential voltagegain device that amplifies the difference between the voltages existing

More information

IFB270 Advanced Electronic Circuits

IFB270 Advanced Electronic Circuits IFB270 Advanced Electronic Circuits Chapter 14: Special-purpose op-amp circuits Prof. Manar Mohaisen Department of EEC Engineering eview of the Precedent Lecture Introduce the level detection op-amp circuits

More information

Overview of Grounding for Industrial and Commercial Power Systems Presented By Robert Schuerger, P.E.

Overview of Grounding for Industrial and Commercial Power Systems Presented By Robert Schuerger, P.E. Overview of Grounding for Industrial and Commercial Power Systems Presented By Robert Schuerger, P.E. HP Critical Facility Services delivered by EYP MCF What is VOLTAGE? Difference of Electric Potential

More information

Biomedical Engineering Evoked Responses

Biomedical Engineering Evoked Responses Biomedical Engineering Evoked Responses Dr. rer. nat. Andreas Neubauer andreas.neubauer@medma.uni-heidelberg.de Tel.: 0621 383 5126 Stimulation of biological systems and data acquisition 1. How can biological

More information

Biopotential Amplifiers. Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan

Biopotential Amplifiers. Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan Biopotential Ampliiers Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan chanhl@mail.cgu.edu.tw Operational ampliier Practical Ideal Biopotential ampliiers Ideal vs. practical

More information

E-health Project Examination: Introduction of an Applicable Pulse Oximeter

E-health Project Examination: Introduction of an Applicable Pulse Oximeter E-health Project Examination: Introduction of an Applicable Pulse Oximeter Mona asseri & Seyedeh Fatemeh Khatami Firoozabadi Electrical Department, Central Tehran Branch, Islamic Azad University, Tehran,

More information

GSM BASED PATIENT MONITORING SYSTEM

GSM BASED PATIENT MONITORING SYSTEM GSM BASED PATIENT MONITORING SYSTEM ABSTRACT This project deals with the monitoring of the patient parameters such as humidity, temperature and heartbeat. Here we have designed a microcontroller based

More information

DESIGN OF A PHOTOPLETHYSMOGRAPHY BASED PULSE RATE DETECTOR

DESIGN OF A PHOTOPLETHYSMOGRAPHY BASED PULSE RATE DETECTOR DESIGN OF A PHOTOPLETHYSMOGRAPHY BASED PULSE RATE DETECTOR Srijan Banerjee 1, Subhajit Roy 2 1 Department of Electrical Engineering, Siliguri Institute of Technology, 2 Department of Electrical Engineering,

More information

Biomedical Electrodes, Sensors, and Transducers. Definition of Biomedical Electrodes, Sensors, and Transducers. Electrode: Sensor: Transducer:

Biomedical Electrodes, Sensors, and Transducers. Definition of Biomedical Electrodes, Sensors, and Transducers. Electrode: Sensor: Transducer: Biomedical Electrodes, Sensors, and Transducers from: Chaterjee, Biomedical Instrumentation, chapter 6 Key Points Electrodes, Sensors, and Transducers: - types of electrodes - voltaic - electrolytic -

More information

NeuVision 500. Abundant and friendly display interface, multifold ECG display screen:

NeuVision 500. Abundant and friendly display interface, multifold ECG display screen: NeuVision 500 Features This monitoring system may be used to monitor patient s 6 physiological parameters: ECG, respiratory rate, body temperature, non-invasive blood pressure (NIBP), pulse oxygen saturation

More information

Grounding for Power Quality

Grounding for Power Quality Presents Grounding for Power Quality Grounding for Power Quality NEC 250.53 states that ground resistance should be less than 25 ohms. Is this true? Grounding for Power Quality No! NEC 250.53 states

More information

Chapter F Protection against electric shocks

Chapter F Protection against electric shocks Chapter F Protection against electric shocks 1 2 3 4 5 6 7 8 Contents General 1.1 Electric shock F2 1.2 Protection against electric shock F3 1.3 Direct and indirect contact F3 Protection against direct

More information

Chapter 4 4. Optoelectronic Acquisition System Design

Chapter 4 4. Optoelectronic Acquisition System Design 4. Optoelectronic Acquisition System Design The present chapter deals with the design of the optoelectronic (OE) system required to translate the obtained optical modulated signal with the photonic acquisition

More information

Amplificador de Biopotencial

Amplificador de Biopotencial Amplificador de Biopotencial Prof. Sérgio F. Pichorim Baseado no cap 6 do Webster e cap 17 do Kutz & Towe From J. G. Webster (ed.), Medical instrumentation: application and design. 3 rd ed. New York: John

More information

GROUNDED ELECTRICAL POWER DISTRIBUTION. Excerpt from Inverter Charger Series Manual BY: VIJAY SHARMA ENGINEER

GROUNDED ELECTRICAL POWER DISTRIBUTION. Excerpt from Inverter Charger Series Manual BY: VIJAY SHARMA ENGINEER GROUNDED ELECTRICAL POWER DISTRIBUTION Excerpt from Inverter Charger Series Manual BY: VIJAY SHARMA ENGINEER .0 Conductors for Electrical Power Distribution For single-phase transmission of AC power or

More information

Ques on (2): [18 Marks] a) Draw the atrial synchronous Pacemaker block diagram and explain its operation. Benha University June 2013

Ques on (2): [18 Marks] a) Draw the atrial synchronous Pacemaker block diagram and explain its operation. Benha University June 2013 Benha University June 2013 Benha Faculty of Engineering Electrical Department Hospital Instrumentations (E472) 4 Th year (control) Dr.Waleed Abdel Aziz Salem Time: 3 Hrs Answer the following questions.

More information

Grounding Recommendations for On Site Power Systems

Grounding Recommendations for On Site Power Systems Grounding Recommendations for On Site Power Systems Revised: February 23, 2017 2017 Cummins All Rights Reserved Course Objectives Participants will be able to: Explain grounding best practices and code

More information

This document is a preview generated by EVS

This document is a preview generated by EVS TECHNICAL SPECIFICATION ISO/TS 22077-2 First edition 2015-08-01 Health informatics Medical waveform format Part 2: Electrocardiography Informatique de santé Forme d onde médicale Partie 2: Electrocardiographie

More information

Chapter F Protection against electric shocks

Chapter F Protection against electric shocks Chapter F Protection against electric shocks 1 2 3 4 5 6 7 8 Contents General 1.1 Electric shock F2 1.2 Protection against electric shock F3 1.3 Direct and indirect contact F3 Protection against direct

More information

IEC ED.1 Safety secondary circuits in series circuits General safety requirements. a new draft

IEC ED.1 Safety secondary circuits in series circuits General safety requirements. a new draft Helmut Schmidt Project leader of IEC /TC97 project team 62870 IEC 62870 ED.1 Safety secondary circuits in series circuits General safety requirements a new draft Schmidt Personal Safety IES ALC Conference

More information

Lab E5: Filters and Complex Impedance

Lab E5: Filters and Complex Impedance E5.1 Lab E5: Filters and Complex Impedance Note: It is strongly recommended that you complete lab E4: Capacitors and the RC Circuit before performing this experiment. Introduction Ohm s law, a well known

More information

Balanced Constant Current Excitation for RTD Sensor Measurements

Balanced Constant Current Excitation for RTD Sensor Measurements Balanced Constant Current Excitation for RTD Sensor Measurements Douglas R. Firth Alan R. Szary Precision Filters, Inc. Ithaca, New York (607) 277-3550 1 Balanced Constant Current Excitation for RTD Sensor

More information

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE TEC-B-01M VOLTAGE CLAMP MODULE FOR EPMS SYSTEMS. VERSION 1.2 npi 2014

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE TEC-B-01M VOLTAGE CLAMP MODULE FOR EPMS SYSTEMS. VERSION 1.2 npi 2014 OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE TEC-B-01M VOLTAGE CLAMP MODULE FOR EPMS SYSTEMS VERSION 1.2 npi 2014 npi electronic GmbH, Bauhofring 16, D-71732 Tamm, Germany Phone +49 (0)7141-9730230;

More information

Design Guidelines for Optocoupler Safety Agency Compliance

Design Guidelines for Optocoupler Safety Agency Compliance ISHAY SEMICONDUCTORS www.vishay.com Optocouplers and Solid-State Relays INTRODUCTION TO ELECTRICAL SAFETY Traditionally, electrical isolation from hazardous voltages has been the most common application

More information

AD8232 EVALUATION BOARD DOCUMENTATION

AD8232 EVALUATION BOARD DOCUMENTATION One Technology Way P.O. Box 9106 Norwood, MA 02062-9106 Tel: 781.329.4700 Fax: 781.461.3113 www.analog.com AD8232 EVALUATION BOARD DOCUMENTATION FEATURES Ready to use Heart Rate Monitor (HRM) Front end

More information

Medlab GmbH EG04000 User Manual. medlab. Four Lead ECG OEM board EG Technical Manual. Copyright Medlab Version Version 1.

Medlab GmbH EG04000 User Manual. medlab. Four Lead ECG OEM board EG Technical Manual. Copyright Medlab Version Version 1. Medlab GmbH EG04000 User Manual medlab Four Lead ECG OEM board EG04000 Technical Manual Copyright Medlab 2014 1 Medlab GmbH EG04000 User Manual Medlab medizinische Diagnosegeräte GmbH Helmholtzstrasse

More information

Arterial pulse waves measured with EMFi and PPG sensors and comparison of the pulse waveform spectral and decomposition analysis in healthy subjects

Arterial pulse waves measured with EMFi and PPG sensors and comparison of the pulse waveform spectral and decomposition analysis in healthy subjects Arterial pulse waves measured with EMFi and PPG sensors and comparison of the pulse waveform spectral and decomposition analysis in healthy subjects Matti Huotari 1, Antti Vehkaoja 2, Kari Määttä 1, Juha

More information

NL800A - Stimulus Isolator

NL800A - Stimulus Isolator NL800A - Stimulus Isolator Introduction The NL800A STIMULUS ISOLATOR is a battery powered, opto-coupled isolator. It has a constant current output, with very high output impedance, making it suitable for

More information

Review Article Electrical Safety in Operating Room. Abstract

Review Article Electrical Safety in Operating Room. Abstract Review Article Electrical Safety in Operating Room Dipasri Bhattacharya 1 M.D., MNAMS, Professor and H.O.D., Dept of Anaesthesiology, R.G.Kar Medical College, Kolkata, West Bengal, India dipasribhattacharya123@gmail.com

More information

Health and Fitness Analog solution. Wenbin Zhu Medical BDM June, 2015

Health and Fitness Analog solution. Wenbin Zhu Medical BDM June, 2015 Health and Fitness Analog solution Wenbin Zhu Medical BDM June, 2015 1 A Broad Market TI in Medical Devices Today TI HealthTech Engineering components for life. TI Solutions for Wearable Optical Bio-Sensing

More information

FEATURES OF VOLTAGE PULSE PLETHYSMOGRAPHY

FEATURES OF VOLTAGE PULSE PLETHYSMOGRAPHY FEATURES OF VOLTAGE PULSE PLETHYSMOGRAPHY Martina Melinščak, B.Sc., Polytechnic of Karlovac, 7 Karlovac, I. Meštrovića, Croatia, martina.melinscak@vuka.hr Ante Šantić, Prof. D.Sc., Faculty of Electrical

More information

Webinar: An Effective Arc Flash Safety Program

Webinar: An Effective Arc Flash Safety Program Webinar: An Effective Arc Flash Safety Program Daleep Mohla September 10 th, 2015: 2pm ET Agenda Arc Flash Defined and Quantified NFPA 70E / CSA Z 462 - Recent Updates What is the ANSI Z10 Hierarchy of

More information

Data Acquisition Basics Lab

Data Acquisition Basics Lab Data Acquisition Basics Lab Introduction Many systems in the body can be modeled as electrical systems that interact with various organs, such as the heart, the brain, and body muscle. These systems communicate

More information

Biomedical Instrumentation (BME420 ) Chapter 6: Biopotential Amplifiers John G. Webster 4 th Edition

Biomedical Instrumentation (BME420 ) Chapter 6: Biopotential Amplifiers John G. Webster 4 th Edition Biomedical Instrumentation (BME420 ) Chapter 6: Biopotential Amplifiers John G. Webster 4 th Edition Dr. Qasem Qananwah BME 420 Department of Biomedical Systems and Informatics Engineering 1 Biopotential

More information

Chapter 25 Alternating Currents

Chapter 25 Alternating Currents Chapter 25 Alternating Currents GOALS When you have mastered the contents of this chapter, you will be able to achieve the following goals: Definitions Define each of the following terms and use it in

More information

Electrical Severity Measurement Tool Revision 4

Electrical Severity Measurement Tool Revision 4 Electrical Severity Measurement Tool Revision 4 November 2017 Electrical Severity Measurement Tool 1.0 Purpose: This tool is intended to measure the severity of exposure to an electrical safety event based

More information

International Journal of Electronics and Communication Engineering & Technology (IJECET), ISSN COMMUNICATION ENGINEERING & TECHNOLOGY (IJECET)

International Journal of Electronics and Communication Engineering & Technology (IJECET), ISSN COMMUNICATION ENGINEERING & TECHNOLOGY (IJECET) 0976 INTERNATIONAL 6464(Print), ISSN 0976 6472(Online) JOURNAL Volume OF 4, Issue ELECTRONICS 1, January- February (2013), AND IAEME COMMUNICATION ENGINEERING & TECHNOLOGY (IJECET) ISSN 0976 6464(Print)

More information

Company Directive STANDARD TECHNIQUE : TP21A/2. Relating to Safety Limits for Touch and Step Voltages - Earthing System Design/Assessment

Company Directive STANDARD TECHNIQUE : TP21A/2. Relating to Safety Limits for Touch and Step Voltages - Earthing System Design/Assessment Company Directive STANDARD TECHNIQUE : TP21A/2 Relating to Safety Limits for Touch and Step Voltages - Earthing System Design/Assessment Policy Summary This document defines safety limits for touch and

More information

4.2 SHORT QUESTIONS AND ANSWERS 1. What is meant by cell? The basic living unit of the body is cell. The function of organs and other structure of the body is understood by cell organization. 2. Give the

More information

AP7301 ELECTROMAGNETIC INTERFERENCE AND COMPATIBILITY L T P C COURSE OBJECTIVES:

AP7301 ELECTROMAGNETIC INTERFERENCE AND COMPATIBILITY L T P C COURSE OBJECTIVES: AP7301 ELECTROMAGNETIC INTERFERENCE AND COMPATIBILITY L T P C 3 0 0 3 COURSE OBJECTIVES: To understand the basics of EMI To study EMI Sources To understand EMI problems To understand Solution methods in

More information

Healthcare Isolation Power Systems. Presented by: David Knecht

Healthcare Isolation Power Systems. Presented by: David Knecht Healthcare Isolation Power Systems Presented by: David Knecht E-Mail: David.Knecht@Bender-us.com Healthcare Isolation Power Systems Theory Offer Codes & Standards FAQs Isolated Power System Purpose To

More information

UNIVERSITY OF CALIFORNIA College of Engineering Department of Electrical Engineering and Computer Sciences

UNIVERSITY OF CALIFORNIA College of Engineering Department of Electrical Engineering and Computer Sciences UNIVERSITY OF CALIFORNIA College of Engineering Department of Electrical Engineering and Computer Sciences EECS 145L: Electronic Transducer Laboratory FINAL EXAMINATION Fall 2013 You have three hours to

More information

Signal Conditioning Systems

Signal Conditioning Systems Note-13 1 Signal Conditioning Systems 2 Generalized Measurement System: The output signal from a sensor has generally to be processed or conditioned to make it suitable for the next stage Signal conditioning

More information

Physics 303 Fall Module 4: The Operational Amplifier

Physics 303 Fall Module 4: The Operational Amplifier Module 4: The Operational Amplifier Operational Amplifiers: General Introduction In the laboratory, analog signals (that is to say continuously variable, not discrete signals) often require amplification.

More information

P2110B 915 MHz RF Powerharvester Receiver

P2110B 915 MHz RF Powerharvester Receiver DESCRIPTION The Powercast Powerharvester is an RF energy harvesting device that converts RF to DC. Housed in a compact SMD package, the receiver provides RF energy harvesting and power management for battery-free,

More information

The question pool for the current General Class license took effect on July 1, 2007.

The question pool for the current General Class license took effect on July 1, 2007. General Class License Manual Errata and Corrections 22 Dec 2010 The following text is intended to support or correct the 6 th edition of the General Class License Manual. Determine the version of the manual

More information

Instrumentation amplifier

Instrumentation amplifier Instrumentationamplifieris a closed-loop gainblock that has a differential input and an output that is single-ended with respect to a reference terminal. Application: are intended to be used whenever acquisition

More information

IBES - Introduction to Biomedical Electronic Systems

IBES - Introduction to Biomedical Electronic Systems Coordinating unit: 230 - ETSETB - Barcelona School of Telecommunications Engineering Teaching unit: 710 - EEL - Department of Electronic Engineering Academic year: Degree: 2018 MASTER'S DEGREE IN ELECTRONIC

More information

BME 405 BIOMEDICAL ENGINEERING SENIOR DESIGN 1 Fall 2005 BME Design Mini-Project Project Title

BME 405 BIOMEDICAL ENGINEERING SENIOR DESIGN 1 Fall 2005 BME Design Mini-Project Project Title BME 405 BIOMEDICAL ENGINEERING SENIOR DESIGN 1 Fall 2005 BME Design Mini-Project Project Title Basic system for Electrocardiography Customer/Clinical need A recent health care analysis have demonstrated

More information

MCU-Based Multi Parameter Patient Monitor

MCU-Based Multi Parameter Patient Monitor International Journal of Sciences: Basic and Applied Research (IJSBAR) ISSN 2307-4531 (Print & Online) http://gssrr.org/index.php?journal=journalofbasicandapplied ---------------------------------------------------------------------------------------------------------------------------

More information

CPC1706Y. Single-Pole, Normally Open 4-Pin OptoMOS DC Power SIP Relay INTEGRATED CIRCUITS DIVISION. Description. Features. Applications.

CPC1706Y. Single-Pole, Normally Open 4-Pin OptoMOS DC Power SIP Relay INTEGRATED CIRCUITS DIVISION. Description. Features. Applications. CPC176 Single-Pole, Normally Open 4-Pin OptoMOS DC Power SIP Relay Parameter Rating Units Blocking Voltage 6 V P Load Current 4 A DC On-Resistance (max).9 Features Handle Load Currents Up to 4A DC 2V rms

More information

Wireless Sensor Networks. EP2980

Wireless Sensor Networks. EP2980 Wireless Sensor Networks EP2980 Jonas.Wahslen@sth.kth.se Sensors What to sense? How to sense/measure? Available sensors Technology Medical ECG Pulsoximeter Applications Smart Grid Industrial Automation

More information

Concepts to be Reviewed

Concepts to be Reviewed Introductory Medical Device Prototyping Analog Circuits Part 3 Operational Amplifiers, http://saliterman.umn.edu/ Department of Biomedical Engineering, University of Minnesota Concepts to be Reviewed Operational

More information

Facility Grounding & Bonding Based on the EMC/PI/SI Model for a High Speed PCB/Cabinet

Facility Grounding & Bonding Based on the EMC/PI/SI Model for a High Speed PCB/Cabinet Facility Grounding & Bonding Based on the EMC/PI/SI Model for a High Speed PCB/Cabinet and: SILICON LABS AN203 PRINTED CIRCUIT BOARD DESIGN NOTES www.silabs.com William Bush (wbush@ieee.org) Industry Consultant

More information

Laboratory Kit for Oscillometry Measurement of Blood Pressure

Laboratory Kit for Oscillometry Measurement of Blood Pressure Laboratory Kit for Oscillometry Measurement of Blood Pressure Jan Dvořák, Jan Havlík Department of Circuit Theory Faculty of Electrical Engineering, Czech Technical University in Prague Technická 2, CZ-16627

More information

Biopotential Electrodes

Biopotential Electrodes Biomedical Instrumentation Prof. Dr. Nizamettin AYDIN naydin@yildiz.edu.tr naydin@ieee.org http://www.yildiz.edu.tr/~naydin Biopotential Electrodes 1 2 Electrode electrolyte interface The current crosses

More information

GE Healthcare. Dash 2500 The standard of excellence for sub-acuity monitoring

GE Healthcare. Dash 2500 The standard of excellence for sub-acuity monitoring GE Healthcare Dash 2500 The standard of excellence for sub-acuity monitoring The clinical With monitored patients, even small measurement variations can indicate significant changes in your patient s condition.

More information

Best Practices for Power and Transient Protection on Rosemount Radar Transmitters

Best Practices for Power and Transient Protection on Rosemount Radar Transmitters Technical Note Rosemount Radar Transmitters Best Practices for Power and Transient Protection on Rosemount Radar Transmitters BACKGROUND INTRODUCTION This document describes best practices for power and

More information

Residual Current Operated Circuit-Breakers (RCCBs)

Residual Current Operated Circuit-Breakers (RCCBs) Product Overview Residual Current Operated Circuit-Breakers (RCCBs) Residual current operated circuit-breakers Number of poles Rated current A Rated residual current ma MW Auxiliary contacts can be mounted

More information

Options to Improve the MEN System into the 21 st Century

Options to Improve the MEN System into the 21 st Century Options to Improve the MEN System into the 21 st Century Chris Halliday Electrical Consulting and Training Pty Ltd, Gladstone NSW, Australia. Email: chris@elect.com.au Web: www.elect.com.au Abstract Network

More information

GROUND TESTERS For all of your Ground Testing needs...

GROUND TESTERS For all of your Ground Testing needs... GROUND TESTERS For all of your Ground Testing needs... An array of Ground Testers to choose from Whether you are doing a simplified 2-Point, a more complete 3- or 4-Point Fall-of-Potential test, a soil

More information

CHAPTER 3. Instrumentation Amplifier (IA) Background. 3.1 Introduction. 3.2 Instrumentation Amplifier Architecture and Configurations

CHAPTER 3. Instrumentation Amplifier (IA) Background. 3.1 Introduction. 3.2 Instrumentation Amplifier Architecture and Configurations CHAPTER 3 Instrumentation Amplifier (IA) Background 3.1 Introduction The IAs are key circuits in many sensor readout systems where, there is a need to amplify small differential signals in the presence

More information

Removal of Power-Line Interference from Biomedical Signal using Notch Filter

Removal of Power-Line Interference from Biomedical Signal using Notch Filter ISSN:1991-8178 Australian Journal of Basic and Applied Sciences Journal home page: www.ajbasweb.com Removal of Power-Line Interference from Biomedical Signal using Notch Filter 1 L. Thulasimani and 2 M.

More information

Biomedical and Wireless Technologies for Pervasive Healthcare

Biomedical and Wireless Technologies for Pervasive Healthcare Miodrag Bolic Associate Professor School of Electrical Engineering and Computer Science (EECS) Faculty of Engineering Biomedical and Wireless Technologies for Pervasive Healthcare Active Research Areas

More information

Physics 481 Experiment 3

Physics 481 Experiment 3 Physics 481 Experiment 3 LAST Name (print) FIRST Name (print) TRANSISTORS (BJT & FET) npn BJT n-channel MOSFET 1 Experiment 3 Transistors: BJT & FET In this experiment transistor properties and transistor

More information

ARC FLASH PPE GUIDELINES FOR INDUSTRIAL POWER SYSTEMS

ARC FLASH PPE GUIDELINES FOR INDUSTRIAL POWER SYSTEMS The Electrical Power Engineers Qual-Tech Engineers, Inc. 201 Johnson Road Building #1 Suite 203 Houston, PA 15342-1300 Phone 724-873-9275 Fax 724-873-8910 www.qualtecheng.com ARC FLASH PPE GUIDELINES FOR

More information

Supplementary Information

Supplementary Information Supplementary Information Fiber-based Generator for Wearable Electronics and Mobile Medication Junwen Zhong 1,, Yan Zhang 2, 3,, Qize Zhong 1,, Qiyi Hu 1, Bin Hu 1, Zhong Lin Wang 2,4 and Jun Zhou 1,*

More information

Some Power Quality Issues in Hospital Facilities

Some Power Quality Issues in Hospital Facilities European Association for the Development of Renewable Energies, Environment and Power Quality (EA4EPQ) International Conference on Renewable Energies and Power Quality (ICREPQ 12) Santiago de Compostela

More information

Operations Manual. Model NT11 & NT11-E. June 18 th 2002

Operations Manual. Model NT11 & NT11-E. June 18 th 2002 Operations Manual Model NT11 & NT11-E June 18 th 2002 3609 North 44 th Street Phoenix, AZ 85018-6023 Internet: www.northstarcontrols.com Fax: (941) 426-0807 Tel: (941) 426-6396 Manual Contents Glossary

More information

Intelligent Pillow for Heart Rate Monitor

Intelligent Pillow for Heart Rate Monitor International Journal of Engineering Research and Development e-issn: 2278-067X, p-issn: 2278-800X, www.ijerd.com Volume 8, Issue 4 (August 2013), PP.47-52 Intelligent Pillow for Heart Rate Monitor Everlyn

More information

EECS 145L Final Examination Solutions (Fall 2013)

EECS 145L Final Examination Solutions (Fall 2013) UNIVERSITY OF CALIFORNIA, BERKELEY College of Engineering, Electrical Engineering and Computer Sciences Department 1.1 Instrumentation amplifier (1) differential amplification (2) very high input impedance

More information

Los A LA-UR Los Alamos National Laboratory Los Alamos, New Mexico 87545

Los A LA-UR Los Alamos National Laboratory Los Alamos, New Mexico 87545 LA-UR-98-1 Los Alamos NationalLaboratory is operated by the University of California for the United States Department of Energy under contract W-7405-ENG-36 TITLE: SUBMITTED TO: Electrical Potential Transfer

More information

* Notebook is excluded. Features KL-720 contains nine modules, including Electrocardiogram Measurement, E lectromyogram Measurement,

* Notebook is excluded. Features KL-720 contains nine modules, including Electrocardiogram Measurement, E lectromyogram Measurement, KL-720 Biomedical Measurement System Supplied by: 011 683 4365 This equipment is intended for students to learn how to design specific measuring circuits and detect the basic physiological signals with

More information

APPLICATION OF HEART PHOTOPLETHYSMOGRAPHY

APPLICATION OF HEART PHOTOPLETHYSMOGRAPHY APPLICATION OF HEART PHOTOPLETHYSMOGRAPHY 1 VICKY KUMAR SINGH, 2 SUMIT KUMAR THAKUR, 3 VINOD KUMAR 1,2,3 Department of Electronics Engineering, Bharati Vidyapeeth College of Engineering Pune E-mail: vickysingh229@gmail.com,

More information

Webinar Organizers. Ryan Shea. Don Miller. Joe Ryan. Support Specialist. Applications Specialist. Product Manager. Precision Digital Corporation

Webinar Organizers. Ryan Shea. Don Miller. Joe Ryan. Support Specialist. Applications Specialist. Product Manager. Precision Digital Corporation Webinar Organizers Joe Ryan Product Manager Precision Digital Corporation Ryan Shea Applications Specialist Precision Digital Corporation Don Miller Support Specialist Precision Digital Corporation Agenda,

More information

AMENDMENT NO. 1 SEPTEMBER IS (Part 1) : 2001/IEC (1991) SURGE ARRESTORS

AMENDMENT NO. 1 SEPTEMBER IS (Part 1) : 2001/IEC (1991) SURGE ARRESTORS AMENDMENT NO. 1 SEPTEMBER 2011 TO IS 15086 (Part 1) : 2001/IEC 60099-1 (1991) SURGE ARRESTORS PART 1 NON-LINEAR RESISTOR TYPE GAPPED SURGE ARRESTORS FOR a.c. SYSTEMS (The Amendment was originally published

More information