Biomedical Engineering Evoked Responses
|
|
- Clement Watson
- 6 years ago
- Views:
Transcription
1 Biomedical Engineering Evoked Responses Dr. rer. nat. Andreas Neubauer Tel.: Stimulation of biological systems and data acquisition 1. How can biological systems be stimulated? Visually Acoustically Via drugs 2. How to record the response of stimuli? Distinguishing signal from noise Artifacts and responses 3. Computer aided data analysis AI vs. PI Andreas Neubauer I Slide 2 I
2 Linear Systems Andreas Neubauer I Slide 3 I How to describe linear systems Components behave in a proportional manner e. g. electrical systems mechanical systems Consider a system with Input and output Goal: Find a function to describe the system Andreas Neubauer I Slide 4 I
3 Frequency responses Easiest test for frequency response: function infinitely high and infinitely short Fourier integral of : 1 1 Testing the response of a low pass filter Fourier transform of the input: Andreas Neubauer I Slide 5 I Stimuli Andreas Neubauer I Slide 6 I
4 Neural stimulation I The body is a good conductor Nerves an muscles work electrically neural stimulation with electricity is possible muscle contraction induced by a current between two surface electrodes Andreas Neubauer I Slide 7 I Neural stimulation II Apart from an alternating current, one can apply a short pulse between two surface electrodes Pulse duration is between 50 s "# 2 %& to avoid high amplitudes The stimulation occurs under the electrode which sees the pulse as negative pulse the transmembrane potential will be reduced where the current leaves the nerve R: Rheobase C: Chronaxie Andreas Neubauer I Slide 8 I
5 More about currents and voltages Stimulation of a nerve is achieved if the transmembrane potential is reduced by about 30% Stimulation current must flow for a minimum time generation of an action potential Position of the electrodes Approx. charge required for stimulation Inside a nerve axon 10 ) On the surface of a nerve trunk 10 * ) On the skin several mm from the nerve 10 + ) Andreas Neubauer I Slide 9 I Depolarization with an electrode Charge has to be removed from one node of Ranvier depolarization generation of an action potential Membrane capacitance is approx. 10 Applied polarization potential: Approx /" % ) Andreas Neubauer I Slide 10 I
6 Electrical stimulation a practical example Energy required to stimulate a nerve with electrons placed directly over the nerve: Approx corresponds to 10 %6 flowing for 100 s through an electrical resistance of 10 7Ω Energy stored in a battery with an output voltage of 9 - and a capacity of 500 %6/: : ; 500 % & a small battery is enough to stimulate a nerve more than times if the nerve is not superficial a much larger amount of energy may be required several joules for a human heart Andreas Neubauer I Slide 11 I Strength duration curves The stimulation energy is not the same for different nerves Large fibers need less energy than small fibers Cutting the innervation of a muscle leads to a slow death of the muscle Denervation Electrical stimuli can be used to stimulate the muscle directly threshold will be increased record the threshold current in dependence of the applied pulse length Rheobase R: Minimum current to stimulate the muscle (independent of pulse length) Chronaxie C: Pulse width with threshold current 2 Rheobase Andreas Neubauer I Slide 12 I
7 Which other stimuli can be used? Stimuli can be applied to each of our senses mechanically auditory taste/smell (difficult in practice) Andreas Neubauer I Slide 13 I visually Andreas Neubauer I Slide 14 I
8 Auditory stimuli Auditory equivalent of a function? Very short click containing a wide range of frequencies Mostly used in objective hearing tests What if the audiologist wants to know which frequencies have been heard? Hypothesis: Give an auditory stimulus, record the response, apply Fourier transform to the response obtain transfer function of auditory system Unfortunately the brain and the hearing system are nonlinear systems Andreas Neubauer I Slide 15 I Auditory stimulus in practice Short burst of single frequency Obtain the response of the ear by testing different frequencies Signal equation: = cos 2@f B Fourier transform leads to: = = B C sin 2@ BF C G sin 2@ BG C 2@ B F C 2@ B G C Andreas Neubauer I Slide 16 I
9 Visual stimuli Visual equivalent of a function? Flash of light The eye cannot deal with frequencies above 20 H Pattern recognition is much easier with visual stimuli Fourier transformation may be more useful in the spatial than in the temporal domain Neubauer, AADA 2014 Andreas Neubauer I Slide 17 I Detour to spatial frequencies Courtesy of M. Ruttorf Neubauer, AADA 2014 Andreas Neubauer I Slide 18 I
10 Detection of small signals Andreas Neubauer I Slide 19 I Bandwidth and signal-to-nose ratios Bandwidth: Frequency range over which the gain remains constant F3 I bandwidth: Frequency range over which the gain is not less than 3 I below the maximum gain Signal Bandwidth [1/&] ECG 0.5F100 EEG Arterial pressure wave 0.5 F75 L) F40 Body temperature L) 1 Respiration L) 10 Electromyograph Nerve action potentials O O Smooth muscle potentials 0.05 F 10 Andreas Neubauer I Slide 20 I
11 Choice of the amplifier Which signal should be amplified? Current? Voltage? Noise will also be amplified! The amplifier will contribute some noise to the signal usually quoted in terms of the equivalent noise at the input, Noise Referred to the Input (RTI) noise (RTI) = (noise at output)/(amplifier gain) RTI can be directly compared with the size of the input signal Example: Measurement of an ECG: 1 %- Required signal to noise ratio: 40 I 100:1 RTI < QR BB 10 - Andreas Neubauer I Slide 21 I Input resistance Amplification may be frequency dependent The choice of the input resistance must be related to the recorded signal Source and input resistance form a potential divider If the input resistance is equal to the source resistance the signal will be reduced to half it s value Reducing this error to 1% requires an input resistance 100 times higher than the source resistance => Source resistance must be known! Andreas Neubauer I Slide 22 I
12 Differential amplifier II In practice almost all bioelectric amplifiers are differential amplifiers ability to reject external signals Use operational amplifiers for the design instrumentation amplifier configuration input voltages Rules for analyzing operational amplifier circuits 1. There will be no voltage between the inputs of an operational amplifier 2. No current will flow into the inputs of an operational amplifier Andreas Neubauer I Slide 23 I Differential amplifier II In practice almost all bioelectric amplifiers are differential amplifiers ability to reject external signals Use operational amplifiers for the design instrumentation amplifier configuration input voltages Rules for analyzing operational amplifier circuits 1. There will be no voltage between the inputs of an operational amplifier 2. No current will flow into the inputs of an operational amplifier Andreas Neubauer I Slide 24 I
Biomedical Engineering Electrophysiology
Biomedical Engineering Electrophysiology Dr. rer. nat. Andreas Neubauer Sources of biological potentials and how to record them 1. How are signals transmitted along nerves? Transmit velocity Direction
More informationBiomedical Instrumentation B2. Dealing with noise
Biomedical Instrumentation B2. Dealing with noise B18/BME2 Dr Gari Clifford Noise & artifact in biomedical signals Ambient / power line interference: 50 ±0.2 Hz mains noise (or 60 Hz in many data sets)
More informationFundamentals of Pacing Therapy
1 4 7 2 5 8 3 6 9 Fundamentals of Pacing Therapy This presentation is provided with the understanding that the slide content must not be altered in any manner as the content is subject to FDA regulations.
More informationEDL Group #3 Final Report - Surface Electromyograph System
EDL Group #3 Final Report - Surface Electromyograph System Group Members: Aakash Patil (07D07021), Jay Parikh (07D07019) INTRODUCTION The EMG signal measures electrical currents generated in muscles during
More informationBME 405 BIOMEDICAL ENGINEERING SENIOR DESIGN 1 Fall 2005 BME Design Mini-Project Project Title
BME 405 BIOMEDICAL ENGINEERING SENIOR DESIGN 1 Fall 2005 BME Design Mini-Project Project Title Basic system for Electrocardiography Customer/Clinical need A recent health care analysis have demonstrated
More information780. Biomedical signal identification and analysis
780. Biomedical signal identification and analysis Agata Nawrocka 1, Andrzej Kot 2, Marcin Nawrocki 3 1, 2 Department of Process Control, AGH University of Science and Technology, Poland 3 Department of
More informationiworx Sample Lab Experiment AN-2: Compound Action Potentials
Experiment AN-2: Compound Action Potentials Exercise 1: The Compound Action Potential Aim: To apply a brief stimulus at the proximal end of the nerve and record a compound action potential from the distal
More informationCHAPTER 7 INTERFERENCE CANCELLATION IN EMG SIGNAL
131 CHAPTER 7 INTERFERENCE CANCELLATION IN EMG SIGNAL 7.1 INTRODUCTION Electromyogram (EMG) is the electrical activity of the activated motor units in muscle. The EMG signal resembles a zero mean random
More informationPrinciples of nerve stimulation
Principles of nerve stimulation An introduction to nerve stimulation, and the use of PowerLab stimulators in physiology. Robert Purves, ADnstruments ntroduction Excitable tissue may be activated by a wide
More informationBME 3113, Dept. of BME Lecture on Introduction to Biosignal Processing
What is a signal? A signal is a varying quantity whose value can be measured and which conveys information. A signal can be simply defined as a function that conveys information. Signals are represented
More informationElectrical noise in the OR
Electrical noise in the OR Chris Thompson Senior Staff Specialist Royal Prince Alfred Hospital SYDNEY SOUTH WEST AREA HEALTH SERVICE NSW HEALTH Electrical noise in the OR Root causes Tiny little signals
More informationCompound Action Potentials
Compound Action Potentials Background As last week s lab showed, even a piece of thread can conduct electricity from stimulating electrodes to recording electrodes in the nerve chamber. You know from lecture
More informationQues on (2): [18 Marks] a) Draw the atrial synchronous Pacemaker block diagram and explain its operation. Benha University June 2013
Benha University June 2013 Benha Faculty of Engineering Electrical Department Hospital Instrumentations (E472) 4 Th year (control) Dr.Waleed Abdel Aziz Salem Time: 3 Hrs Answer the following questions.
More informationLab #9: Compound Action Potentials in the Toad Sciatic Nerve
Lab #9: Compound Action Potentials in the Toad Sciatic Nerve In this experiment, you will measure compound action potentials (CAPs) from an isolated toad sciatic nerve to illustrate the basic physiological
More informationApplication Note 175 Using the STMISOC Stimulus Isolator
APPLICATION NOTE 42 Aero Camino, Goleta, CA 93117 Tel (805) 685-0066 Fax (805) 685-0067 info@biopac.com www.biopac.com 02.14.2018 Application Note 175 Using the The MP160/150 System stimulation features
More informationEE 230 Experiment 10 ECG Measurements Spring 2010
EE 230 Experiment 10 ECG Measurements Spring 2010 Note: If for any reason the students are uncomfortable with doing this experiment, please talk to the instructor for the course and an alternative experiment
More informationBiomedical Signals. Signals and Images in Medicine Lecture 1 Dr Nabeel Anwar
Biomedical Signals Signals and Images in Medicine Lecture 1 Dr Nabeel Anwar Books 1. The Scientist and Engineer's Guide to Digital Signal Processing By Steven W. Smith, Ph.D. This book is written in simple
More informationHearing and Deafness 2. Ear as a frequency analyzer. Chris Darwin
Hearing and Deafness 2. Ear as a analyzer Chris Darwin Frequency: -Hz Sine Wave. Spectrum Amplitude against -..5 Time (s) Waveform Amplitude against time amp Hz Frequency: 5-Hz Sine Wave. Spectrum Amplitude
More informationPerception of pitch. Definitions. Why is pitch important? BSc Audiology/MSc SHS Psychoacoustics wk 4: 7 Feb A. Faulkner.
Perception of pitch BSc Audiology/MSc SHS Psychoacoustics wk 4: 7 Feb 2008. A. Faulkner. See Moore, BCJ Introduction to the Psychology of Hearing, Chapter 5. Or Plack CJ The Sense of Hearing Lawrence Erlbaum,
More informationImagine the cochlea unrolled
2 2 1 1 1 1 1 Cochlea & Auditory Nerve: obligatory stages of auditory processing Think of the auditory periphery as a processor of signals 2 2 1 1 1 1 1 Imagine the cochlea unrolled Basilar membrane motion
More informationBiomedical Sensor Systems Laboratory. Institute for Neural Engineering Graz University of Technology
Biomedical Sensor Systems Laboratory Institute for Neural Engineering Graz University of Technology 2017 Bioinstrumentation Measurement of physiological variables Invasive or non-invasive Minimize disturbance
More informationBiomedical Instrumentation (BME420 ) Chapter 6: Biopotential Amplifiers John G. Webster 4 th Edition
Biomedical Instrumentation (BME420 ) Chapter 6: Biopotential Amplifiers John G. Webster 4 th Edition Dr. Qasem Qananwah BME 420 Department of Biomedical Systems and Informatics Engineering 1 Biopotential
More informationAUDL GS08/GAV1 Signals, systems, acoustics and the ear. Loudness & Temporal resolution
AUDL GS08/GAV1 Signals, systems, acoustics and the ear Loudness & Temporal resolution Absolute thresholds & Loudness Name some ways these concepts are crucial to audiologists Sivian & White (1933) JASA
More informationFlorida Atlantic University Biomedical Signal Processing Lab Experiment 2 Signal Transduction: Building an analog Electrocardiogram (ECG)
Florida Atlantic University Biomedical Signal Processing Lab Experiment 2 Signal Transduction: Building an analog Electrocardiogram (ECG) 1. Introduction: The Electrocardiogram (ECG) is a technique of
More informationComplex Sounds. Reading: Yost Ch. 4
Complex Sounds Reading: Yost Ch. 4 Natural Sounds Most sounds in our everyday lives are not simple sinusoidal sounds, but are complex sounds, consisting of a sum of many sinusoids. The amplitude and frequency
More informationBiosignal filtering and artifact rejection. Biosignal processing I, S Autumn 2017
Biosignal filtering and artifact rejection Biosignal processing I, 52273S Autumn 207 Motivation ) Artifact removal power line non-stationarity due to baseline variation muscle or eye movement artifacts
More informationPerception of pitch. Definitions. Why is pitch important? BSc Audiology/MSc SHS Psychoacoustics wk 5: 12 Feb A. Faulkner.
Perception of pitch BSc Audiology/MSc SHS Psychoacoustics wk 5: 12 Feb 2009. A. Faulkner. See Moore, BCJ Introduction to the Psychology of Hearing, Chapter 5. Or Plack CJ The Sense of Hearing Lawrence
More informationInstrumentation amplifier
Instrumentationamplifieris a closed-loop gainblock that has a differential input and an output that is single-ended with respect to a reference terminal. Application: are intended to be used whenever acquisition
More informationEMG. The study of muscle function through the investigation of the electrical signal the muscles produce
EMG The study of muscle function through the investigation of the electrical signal the muscles produce Niek van Ulzen, 23-11-2010 niekroland.vanulzen@univr.it Program A. Theory (today) 1. Background Electricity
More informationAUDL 4007 Auditory Perception. Week 1. The cochlea & auditory nerve: Obligatory stages of auditory processing
AUDL 4007 Auditory Perception Week 1 The cochlea & auditory nerve: Obligatory stages of auditory processing 1 Think of the ear as a collection of systems, transforming sounds to be sent to the brain 25
More informationEC209 - Improving Signal-To-Noise Ratio (SNR) for Optimizing Repeatable Auditory Brainstem Responses
EC209 - Improving Signal-To-Noise Ratio (SNR) for Optimizing Repeatable Auditory Brainstem Responses Aaron Steinman, Ph.D. Director of Research, Vivosonic Inc. aaron.steinman@vivosonic.com 1 Outline Why
More informationElectrocardiogram (ECG)
Vectors and ECG s Vectors and ECG s 2 Electrocardiogram (ECG) Depolarization wave passes through the heart and the electrical currents pass into surrounding tissues. Small part of the extracellular current
More informationChanging the sampling rate
Noise Lecture 3 Finally you should be aware of the Nyquist rate when you re designing systems. First of all you must know your system and the limitations, e.g. decreasing sampling rate in the speech transfer
More informationSpecial-Purpose Operational Amplifier Circuits
Special-Purpose Operational Amplifier Circuits Instrumentation Amplifier An instrumentation amplifier (IA) is a differential voltagegain device that amplifies the difference between the voltages existing
More informationPerception of pitch. Importance of pitch: 2. mother hemp horse. scold. Definitions. Why is pitch important? AUDL4007: 11 Feb A. Faulkner.
Perception of pitch AUDL4007: 11 Feb 2010. A. Faulkner. See Moore, BCJ Introduction to the Psychology of Hearing, Chapter 5. Or Plack CJ The Sense of Hearing Lawrence Erlbaum, 2005 Chapter 7 1 Definitions
More informationUsed to overcome ventricular fibrillation may be due to coronary occlusion, shock, or abnormalities in blood chemistry
Used to overcome ventricular fibrillation may be due to coronary occlusion, shock, or abnormalities in blood chemistry Main problem: heart muscle fibers are continuously stimulated by adjacent muscles
More informationPhysical Therapy. Methods in Physical Therapy. Non-electric heat therapy Therapeutique application of electric current
Physical Therapy Methods in Physical Therapy Hubbard Hydrotherapy Tank, Carlos Andreson, Watercolour, 943 Non-Electric heat therapy (heating or cooling) Electrotherapy Ultrasound therapy Magneto-therapy
More informationSignals & Systems for Speech & Hearing. Week 6. Practical spectral analysis. Bandpass filters & filterbanks. Try this out on an old friend
Signals & Systems for Speech & Hearing Week 6 Bandpass filters & filterbanks Practical spectral analysis Most analogue signals of interest are not easily mathematically specified so applying a Fourier
More informationThe EarSpring Model for the Loudness Response in Unimpaired Human Hearing
The EarSpring Model for the Loudness Response in Unimpaired Human Hearing David McClain, Refined Audiometrics Laboratory, LLC December 2006 Abstract We describe a simple nonlinear differential equation
More informationPhase and Feedback in the Nonlinear Brain. Malcolm Slaney (IBM and Stanford) Hiroko Shiraiwa-Terasawa (Stanford) Regaip Sen (Stanford)
Phase and Feedback in the Nonlinear Brain Malcolm Slaney (IBM and Stanford) Hiroko Shiraiwa-Terasawa (Stanford) Regaip Sen (Stanford) Auditory processing pre-cosyne workshop March 23, 2004 Simplistic Models
More informationdescribe sound as the transmission of energy via longitudinal pressure waves;
1 Sound-Detailed Study Study Design 2009 2012 Unit 4 Detailed Study: Sound describe sound as the transmission of energy via longitudinal pressure waves; analyse sound using wavelength, frequency and speed
More informationBiomechatronic Systems
Biomechatronic Systems Unit 4: Control Mehdi Delrobaei Spring 2018 Open-Loop, Closed-Loop, Feed-Forward Control Open-Loop - Walking with closed eyes - Changing sitting position Feed-Forward - Visual balance
More informationBiomechatronic Systems
Biomechatronic Systems Unit 4: Control Mehdi Delrobaei Spring 2018 Open-Loop, Closed-Loop, Feed-Forward Control Open-Loop - Walking with closed eyes - Changing sitting position Feed-Forward - Visual balance
More informationEXPERIMENT 5 Bioelectric Measurements
Objectives EXPERIMENT 5 Bioelectric Measurements 1) Generate periodic signals with a Signal Generator and display on an Oscilloscope. 2) Investigate a Differential Amplifier to see small signals in a noisy
More informationAcoustics, signals & systems for audiology. Week 9. Basic Psychoacoustic Phenomena: Temporal resolution
Acoustics, signals & systems for audiology Week 9 Basic Psychoacoustic Phenomena: Temporal resolution Modulating a sinusoid carrier at 1 khz (fine structure) x modulator at 100 Hz (envelope) = amplitudemodulated
More informationECG Project. Raphal Blanchet, Axel Boland, Thomas Donnay, Mario Jose Teles Varandas, University of Liege
ECG Project Raphal Blanchet, Axel Boland, Thomas Donnay, Mario Jose Teles Varandas, University of Liege Abstract We were asked to design our own Electrocardiogram. Obviously, recording heart beats without
More informationPsychology in Your Life
Sarah Grison Todd Heatherton Michael Gazzaniga Psychology in Your Life FIRST EDITION Chapter 5 Sensation and Perception 2014 W. W. Norton & Company, Inc. Section 5.1 How Do Sensation and Perception Affect
More informationUsing Rank Order Filters to Decompose the Electromyogram
Using Rank Order Filters to Decompose the Electromyogram D.J. Roberson C.B. Schrader droberson@utsa.edu schrader@utsa.edu Postdoctoral Fellow Professor The University of Texas at San Antonio, San Antonio,
More informationAcoustics, signals & systems for audiology. Week 4. Signals through Systems
Acoustics, signals & systems for audiology Week 4 Signals through Systems Crucial ideas Any signal can be constructed as a sum of sine waves In a linear time-invariant (LTI) system, the response to a sinusoid
More informationAPPLICATION NOTE. Overview
Application Note 111 APPLICATION NOTE Nerve Conduction Velocity 42 Aero Camino, Goleta, CA 93117 Tel (805) 685-0066 Fax (805) 685-0067 info@biopac.com www.biopac.com 06.05.2018 This application note details
More informationInterface Electronic Circuits
Lecture (5) Interface Electronic Circuits Part: 1 Prof. Kasim M. Al-Aubidy Philadelphia University-Jordan AMSS-MSc Prof. Kasim Al-Aubidy 1 Interface Circuits: An interface circuit is a signal conditioning
More informationCOMMUNICATIONS BIOPHYSICS
XVI. COMMUNICATIONS BIOPHYSICS Prof. W. A. Rosenblith Dr. D. H. Raab L. S. Frishkopf Dr. J. S. Barlow* R. M. Brown A. K. Hooks Dr. M. A. B. Brazier* J. Macy, Jr. A. ELECTRICAL RESPONSES TO CLICKS AND TONE
More informationRFID Systems: Radio Architecture
RFID Systems: Radio Architecture 1 A discussion of radio architecture and RFID. What are the critical pieces? Familiarity with how radio and especially RFID radios are designed will allow you to make correct
More informationBENG 186B Winter 2013 Final
Name (Last, First): BENG 186B Winter 2013 Final This exam is closed book, closed note, calculators are OK. Circle and put your final answers in the space provided; show your work only on the pages provided.
More informationSyllabus Recording Devices
Syllabus Recording Devices Introduction, Strip chart recorders, Galvanometer recorders, Null balance recorders, Potentiometer type recorders, Bridge type recorders, LVDT type recorders, Circular chart
More informationNeurophysiology. The action potential. Why should we care? AP is the elemental until of nervous system communication
Neurophysiology Why should we care? AP is the elemental until of nervous system communication The action potential Time course, propagation velocity, and patterns all constrain hypotheses on how the brain
More informationBrain-computer Interface Based on Steady-state Visual Evoked Potentials
Brain-computer Interface Based on Steady-state Visual Evoked Potentials K. Friganović*, M. Medved* and M. Cifrek* * University of Zagreb, Faculty of Electrical Engineering and Computing, Zagreb, Croatia
More informationKanchan S. Shrikhande. Department of Instrumentation Engineering, Vivekanand Education Society s Institute of.
ISOLATED ECG AMPLIFIER WITH RIGHT LEG DRIVE Kanchan S. Shrikhande Department of Instrumentation Engineering, Vivekanand Education Society s Institute of Technology(VESIT),kanchans90@gmail.com Abstract
More informationLimulus eye: a filter cascade. Limulus 9/23/2011. Dynamic Response to Step Increase in Light Intensity
Crab cam (Barlow et al., 2001) self inhibition recurrent inhibition lateral inhibition - L17. Neural processing in Linear Systems 2: Spatial Filtering C. D. Hopkins Sept. 23, 2011 Limulus Limulus eye:
More informationDetection of external stimuli Response to the stimuli Transmission of the response to the brain
Sensation Detection of external stimuli Response to the stimuli Transmission of the response to the brain Perception Processing, organizing and interpreting sensory signals Internal representation of the
More informationF produce muscle activation resembling that which occurs
212 IEEE TRANSACTIONS ON REHABILITATION ENGINEERING, VOL. 3, NO. 3, SEPTEMBER 1995 Modeling Study of Activation and Propagation Delays During Stimulation of Peripheral Nerve Fibers with a Tripolar Cuff
More informationSOCRATES. Auditory Evoked Potentials
SOCRATES Auditory Evoked Potentials SOCRATES A complete clinical system to record auditory evoked potentials SOCRATES is a PC-based professional medical device which can detect auditory evoked potentials
More informationDesign on Electrocardiosignal Detection Sensor
Sensors & Transducers 203 by IFSA http://www.sensorsportal.com Design on Electrocardiosignal Detection Sensor Hao ZHANG School of Mathematics and Computer Science, Tongling University, 24406, China E-mail:
More informationPhysiological Signal Processing Primer
Physiological Signal Processing Primer This document is intended to provide the user with some background information on the methods employed in representing bio-potential signals, such as EMG and EEG.
More informationBiomedical Electrodes, Sensors, and Transducers. Definition of Biomedical Electrodes, Sensors, and Transducers. Electrode: Sensor: Transducer:
Biomedical Electrodes, Sensors, and Transducers from: Chaterjee, Biomedical Instrumentation, chapter 6 Key Points Electrodes, Sensors, and Transducers: - types of electrodes - voltaic - electrolytic -
More informationANC: Section 2. Unidirectional Propagation - 1 J Thomas Mortimer & Narendra Bhadra
ANC: Section 2. Unidirectional Propagation - 1 J Thomas Mortimer & Narendra Bhadra Under physiological conditions, a nerve action potential (AP) is generated at one end of an axon and proceeds towards
More informationV Technical Textiles Interconnect PRESENTED BY V TECHNICAL TEXTILES, INC.
1 V Technical Textiles Interconnect PRESENTED BY V TECHNICAL TEXTILES, INC. Introduction 2 Flexible conductive fabrics are taking the place of thin film technologies for sensing, heating, and flexible
More information*Notebook is excluded
Biomedical Measurement Training System This equipment is designed for students to learn how to design specific measuring circuits and detect the basic physiological signals with practical operation. Moreover,
More informationChapter 2 A Silicon Model of Auditory-Nerve Response
5 Chapter 2 A Silicon Model of Auditory-Nerve Response Nonlinear signal processing is an integral part of sensory transduction in the nervous system. Sensory inputs are analog, continuous-time signals
More information698 IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, VOL. 51, NO. 5, MAY 2004
698 IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, VOL. 51, NO. 5, MAY 2004 Different Pulse Shapes to Obtain Small Fiber Selective Activation by Anodal Blocking A Simulation Study Aleksandra Vučković*, Nico
More informationBIO-ELECTRIC MEASUREMENTS
BIO-ELECTRIC MEASUREMENTS OBJECTIVES: 1) Determine the amplitude of the electrical "noise" in the body. 2) Observe and measure the characteristics and amplitudes of muscle potentials due to the biceps.
More informationPsycho-acoustics (Sound characteristics, Masking, and Loudness)
Psycho-acoustics (Sound characteristics, Masking, and Loudness) Tai-Shih Chi ( 冀泰石 ) Department of Communication Engineering National Chiao Tung University Mar. 20, 2008 Pure tones Mathematics of the pure
More informationSpectro-Temporal Methods in Primary Auditory Cortex David Klein Didier Depireux Jonathan Simon Shihab Shamma
Spectro-Temporal Methods in Primary Auditory Cortex David Klein Didier Depireux Jonathan Simon Shihab Shamma & Department of Electrical Engineering Supported in part by a MURI grant from the Office of
More informationResults of Egan and Hake using a single sinusoidal masker [reprinted with permission from J. Acoust. Soc. Am. 22, 622 (1950)].
XVI. SIGNAL DETECTION BY HUMAN OBSERVERS Prof. J. A. Swets Prof. D. M. Green Linda E. Branneman P. D. Donahue Susan T. Sewall A. MASKING WITH TWO CONTINUOUS TONES One of the earliest studies in the modern
More informationPortable EEG Signal Acquisition System
Noor Ashraaf Noorazman, Nor Hidayati Aziz Faculty of Engineering and Technology, Multimedia University, Jalan Ayer Keroh Lama, 75450 Melaka, Malaysia Email: noor.ashraaf@gmail.com, hidayati.aziz@mmu.edu.my
More informationBiosignal filtering and artifact rejection. Biosignal processing, S Autumn 2012
Biosignal filtering and artifact rejection Biosignal processing, 521273S Autumn 2012 Motivation 1) Artifact removal: for example power line non-stationarity due to baseline variation muscle or eye movement
More informationAnalog Circuits and Systems
Analog Circuits and Systems Prof. K Radhakrishna Rao Lecture 3 Role of Analog Signal Processing in Electronic Products Part 11 1 Cell Phone o The most dominant product of present day world o Its basic
More informationProduct Overview. npi 07/18
Product Overview For meanwhile over 33 years npi electronic GmbH develops and produces equipment for in vitro and in vivo research in the life sciences (especially physiological and pharmacological basic
More informationLaboratory Activities Handbook
Laboratory Activities Handbook Answer Key 0 P a g e Contents Introduction... 2 Optical Heart Rate Monitor Overview... 2 Bare Board Preparation... 3 Light Indicator... 5 Low Pass Filter... 7 Amplifier...
More informationIntroduction to Electronic Circuit for Instrumentation
Introduction to Electronic Circuit for Instrumentation Fundamental quantities Length Mass Time Charge and electric current Heat and temperature Light and luminous intensity Matter (atom, ion and molecule)
More information40 Hz Event Related Auditory Potential
40 Hz Event Related Auditory Potential Ivana Andjelkovic Advanced Biophysics Lab Class, 2012 Abstract Main focus of this paper is an EEG experiment on observing frequency of event related auditory potential
More informationAmplitude Modulation Effects in Cardiac Signals
Abstract Amplitude Modulation Effects in Cardiac Signals Randall Peters 1, Erskine James 2 & Michael Russell 3 1 Physics Department and 2 Medical School, Department of Internal Medicine Mercer University,
More informationDESIGN OF OTA-C FILTER FOR BIOMEDICAL APPLICATIONS
DESIGN OF OTA-C FILTER FOR BIOMEDICAL APPLICATIONS Sreedhar Bongani 1, Dvija Mounika Chirumamilla 2 1 (ECE, MCIS, MANIPAL UNIVERSITY, INDIA) 2 (ECE, K L University, INDIA) ABSTRACT-This paper presents
More information* Notebook is excluded. Features KL-720 contains nine modules, including Electrocardiogram Measurement, E lectromyogram Measurement,
KL-720 Biomedical Measurement System Supplied by: 011 683 4365 This equipment is intended for students to learn how to design specific measuring circuits and detect the basic physiological signals with
More informationApplied Electronics II
Applied Electronics II Chapter 3: Operational Amplifier Part 1- Op Amp Basics School of Electrical and Computer Engineering Addis Ababa Institute of Technology Addis Ababa University Daniel D./Getachew
More informationTNS Journal Club: Efficient coding of natural sounds, Lewicki, Nature Neurosceince, 2002
TNS Journal Club: Efficient coding of natural sounds, Lewicki, Nature Neurosceince, 2002 Rich Turner (turner@gatsby.ucl.ac.uk) Gatsby Unit, 18/02/2005 Introduction The filters of the auditory system have
More informationAnalog Circuits and Systems
Analog Circuits and Systems Prof. K Radhakrishna Rao Lecture 21: Filters 1 Review Integrators as building blocks of filters Frequency compensation in negative feedback systems Opamp and LDO frequency compensation
More informationBasic Electronics Prof. Dr. Chitralekha Mahanta Department of Electronics and Communication Engineering Indian Institute of Technology, Guwahati
Basic Electronics Prof. Dr. Chitralekha Mahanta Department of Electronics and Communication Engineering Indian Institute of Technology, Guwahati Module: 3 Field Effect Transistors Lecture-7 High Frequency
More informationLecture Fundamentals of Data and signals
IT-5301-3 Data Communications and Computer Networks Lecture 05-07 Fundamentals of Data and signals Lecture 05 - Roadmap Analog and Digital Data Analog Signals, Digital Signals Periodic and Aperiodic Signals
More informationInstrumentationTools.com
Author: Instrumentation Tools Categories: Multiple Choice Questions Measurement and Instrumentation Objective Questions Part 4 Measurement and Instrumentation Objective Questions 1. The decibel is a measure
More informationBME 701 Lecture 1. Measurement and Instrumentation
BME 701 Lecture 1 Measurement and Instrumentation 1 Cochlear Implant 2 Advances in Vision (Retinal Stimulation) 3 Mini Gastric Imaging 4 5 Aspects of Measurement General Instrumentation Transducers (Electrodes)
More informationMagnetoencephalography and Auditory Neural Representations
Magnetoencephalography and Auditory Neural Representations Jonathan Z. Simon Nai Ding Electrical & Computer Engineering, University of Maryland, College Park SBEC 2010 Non-invasive, Passive, Silent Neural
More informationA Comprehensive Model for Power Line Interference in Biopotential Measurements
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, VOL. 49, NO. 3, JUNE 2000 535 A Comprehensive Model for Power Line Interference in Biopotential Measurements Mireya Fernandez Chimeno, Member, IEEE,
More informationCHAPTER 3. Instrumentation Amplifier (IA) Background. 3.1 Introduction. 3.2 Instrumentation Amplifier Architecture and Configurations
CHAPTER 3 Instrumentation Amplifier (IA) Background 3.1 Introduction The IAs are key circuits in many sensor readout systems where, there is a need to amplify small differential signals in the presence
More informationAn optimized layout for multipolar neural recording electrode
An optimized layout for multipolar neural recording electrode Fabien Soulier, Olivier Rossel, Serge Bernard, Guy Cathébras, David Guiraud To cite this version: Fabien Soulier, Olivier Rossel, Serge Bernard,
More informationSection 6 Chapter 2: Operational Amplifiers
03 Section 6 Chapter : Operational Amplifiers eference : Microelectronic circuits Sedra sixth edition 4//03 4//03 Contents: - DC imperfections A. Offset voltage B. Solution of offset voltage C. Input bias
More informationELECTRICAL CURRENT FOR PAIN CONTROL
ELECTRICAL CURRENT FOR PAIN CONTROL 2 Pain modulation ES for pain modulation (TENS) Parameter for ES for pain modulation Pain modulation 3 Gate control theory of pain modulation Pain perception nociceptor
More informationLab E5: Filters and Complex Impedance
E5.1 Lab E5: Filters and Complex Impedance Note: It is strongly recommended that you complete lab E4: Capacitors and the RC Circuit before performing this experiment. Introduction Ohm s law, a well known
More informationChapter 12. Preview. Objectives The Production of Sound Waves Frequency of Sound Waves The Doppler Effect. Section 1 Sound Waves
Section 1 Sound Waves Preview Objectives The Production of Sound Waves Frequency of Sound Waves The Doppler Effect Section 1 Sound Waves Objectives Explain how sound waves are produced. Relate frequency
More informationDEVELOPMENT OF AN AUDIO EVOKED RESPONSE SYSTEM TO FACILITATE ANAESTHESIA MONITORING
DEVELOPMENT OF AN AUDIO EVOKED RESPONSE SYSTEM TO FACILITATE ANAESTHESIA MONITORING A Thesis Submitted by Shabnam Noor Student ID: 06210016 and Saida Hossain Student ID: 07310092 Department of Electronics
More information