NMR Hardware 06/06/2017. Outline. Instrumentation: Magnet. Increasing magnetic field increases Sensitivity, by power of 3/2 Dispersion, linearly

Size: px
Start display at page:

Download "NMR Hardware 06/06/2017. Outline. Instrumentation: Magnet. Increasing magnetic field increases Sensitivity, by power of 3/2 Dispersion, linearly"

Transcription

1 NMR Hardware Outline Magnet Lock Shims Gradient Probe Signal generation and transmitters Receiver and digitizer Variable temperature system Solids hardware Instrumentation: Magnet Often the most impressive (and expensive) part of the NMR system Must be large, stable and homogeneous Superconducting wire typical Made more homogenous with Cryoshims Room temperature shims Increasing magnetic field increases Sensitivity, by power of 3/2 Dispersion, linearly 1

2 T, G, MHz, Hz and ppm Magnetic field is measured in Tesla (T) or Gauss (G) (1 T = 10 4 G), but in NMR usually just refer to 1 H frequency at the given field magnetic field chemical shift T G MHz Hz ppm , , , Independent of the field d ppm = (ν ν 0 ) 10 6 ν 0 The advantages of greater magnetic field strength Signal dispersion Sensitivity ol-acetate Oestradiol-acetate 900 MHz 900 MHz 900 MHz Hz Current Data P Hz NAME txioe EXPNO Hz Estradiol Current Data Parameters 600 NAME MHz acetate PROCNO txioestradiol Current Data Parameters EXPNO 3 F2 - Acquisition NAME txioestradiol PROCNO 1 Date_ 200 EXPNO MHz Time 1 PROCNO MHz F2 - Acquisition Parameters INSTRUM Date_ PROBHD 5 m F2 - Acquisition Parameters Time PULPROG Date_ INSTRUM spect TD 65 Time PROBHD 5 mm TXI 13C Z SOLVENT INSTRUM spect PULPROG zg NS PROBHD 5 mm TXI 13C ZTD DS PULPROG zg SOLVENT CDCl3 SWH TD NS 4 FIDRES 0 SOLVENT CDCl3 HzDS 0 AQ 4.40 NS 4 SWH Hz RG 1 DS FIDRES Hz DW 67 Hz SWH Hz AQ sec DE FIDRES Hz Hz RG 180 TE AQ MHz sec DW usec D RG DE 6.00 usec DW usec TE K ======== CHA 400 MHz DE 6.00 usec D sec NUC1 TE K P MHz D sec ======== CHANNEL f1 ======== PL1-4 NUC1 1H SFO ======== CHANNEL f1 ======== P usec NUC1 1H PL db F2 - Processing P usec SFO MHz SI 655 PL db SF SFO MHzF2 - Processing parameters WDW SI SSB Hz F2 - Processing parameters SF MHz LB -0. SI ppm WDW GM GB 0 SF MHz SSB 0 PC Hz WDW GM LB Hz SSB GB 0.3 Hz LB Hz PC 4.00 GB 0.3 PC 4.00 Lock The lock is a system to keep the magnetic field stable It works by monitoring the deuterium NMR signal arising from the solvent on a nearly continuous basis If the deuterium signal shifts, implying that the magnetic field had shifted, then a compensatory field would be adjusted The compensatory field is called z0 on Varian instruments The lock field on instruments 2

3 What does the lock do? Stabilises the field: Corrects for drift (decay of superconducting magnet) Corrects for transient disturbances Makes the field correct Means unreferenced spectra have meaningful chemical shifts TMS should appear at ~ 0 ppm How does the lock work? The lock monitors the deuterium spectrum of the solvent signal, and keeps the frequency constant by adjusting the magnetic field. If the signal is on resonance, the dispersion mode spectrum has zero intensity at the O1 frequency: If the field changes, the intensity of the spectrum becomes non-zero, and the field is adjusted accordingly. Shimming Purpose: create a homogeneous magnetic field across sample Means: adjust current in shim coils putting current in a shim coil establishes a magnetic field each shim coil has its own geometry (Z, X, Z2, etc.) sum of main magnetic field and effect of shim coils should be a homogeneous magnetic field shim sample c b a bad field good field lineshape 3

4 Good shimming 1.3 Hz (0.3 Hz line broadening) 29 Si satellites 1 H 1.0 Hz (0.3 Hz line broadening) 1 H 1 H Sample preparation Sample should be Clear (filter if necessary, for example through a ball of cotton at the neck of a glass pipette) Homogeneous 5 cm long Labeled on top with marker only NMR tube should be Clean Dry (use oven up to 50 C, keep tubes standing as upright as possible) Free of chips Wiped with a Kimwipe immediately before insertion in magnet Why? Shimming Better observation of small J-couplings Better S/N Avoiding finger grease in probe The grease contaminates your sample and other people s, too Positioning sample in depth gauge It is better to dilute sample to correct depth than to run with a short sample (because concentrated, short sample will have a hump at the bottom of the peak; the added concentration won t increase peak intensity) But, if you have a short sample anyway 4

5 Pulsed magnetic field gradients A pulsed magnetic field gradient is a temporary distortion of the magnetic field. The gradient can be turned on and off quickly is linear and reproducible can be scaled in amplitude (+100% to 100%) normal field + - Gradient field -+ sample Normal evolution In a normal NMR magnet all spins experience the same field and evolve at the same rate. normal field signal The sum of all spins is coherent, giving a strong, sharp signal Gradient pulse Applying a gradient pulse means each section of the sample sees a different field; nuclei in each section therefore evolve at a different frequency gradient field no net signal At the end of a typical gradient pulse the spins are completely distributed in phase, giving zero net magnetisation for the whole sample 5

6 Second gradient pulse Local coherence has not been lost! Apply a gradient of the same strength but opposite polarity - the spins rewind and recreate the observable signal negative gradient signal refocussed Dephasing unwanted signals and rephasing wanted signals is the key to most gradient experiments Gradient shimming Automated shimming routine using gradients to identify regions of inhomogeneity within the magnetic field On instruments, also touch up off-axis shims using the lock level (as you would do manually) Gradients in 2D experiments Apply gradients to both wanted and unwanted magnetisation Allow wanted and unwanted magnetisation to evolve differently Apply gradients to refocus desired magnetisation only 1 H pulses, delays pulses, delays 13 C pulses, delays pulses, delays Grad ratio of gradients = ratio of H / X frequency 6

7 Inverse 2D with Gradients no gradients 22 mins, t1-noise Inverse (proton-detected) experiments are intrinsically more sensitive than carbon-detected versions But they require good spectrometer stability to suppress the 12 C- 1 H signal Using gradients, unwanted signals can be eliminated at source, removing the suppression problem. This gives much quicker and cleaner experiments gradients 10 mins, clean Instrumentation: Probe The heart of the NMR system Transmits and receives rf between spectrometer console and sample Optimised for a particular nucleus or set of nuclei Determines the maximum diameter NMR tube that can be used Can be used for special applications Flow Solids Extra sensitivity with a cryoprobe Probes and channel configurations The transmit / receive coils of an NMR probe are each part of an rf circuit that can be tuned to transmit or receive to a particular range of frequencies Tuning changes as a function of solvent, especially for the most sensitive nuclei Most recent probes have gradient capabilities Most recent probes have VT capabilities Nucleus Frequency (MHz) in 7.0 T magnet Frequency (MHz) in 9.4 T magnet Frequency (MHz) in 11.7 T magnet 1 H F P C H N

8 Ten minute 13 C comparison: 3 mg strychnine Varian Mercury 300 Avance 500 Ten-minute 13 C solution: HSQC??? Varian Mercury 300 Avance 500 Indirect probes Indirect probe: 1 H observe, X decouple Most sensitive for 1 H (and sometimes for 19 F) Good for water suppression (residual water signal is very narrow) Can detect X ( 13 C, 15 N, 31 P, etc.) directly, but insensitively Can detect X indirectly, or decouple X so run 2D experiments (HSQC, HMQC, HMBC) In biological work, especially, three channel inverse probes are common Tuned to 1 H, 13 C, and 15 N (proteins) Or 1 H, 13 C, and 31 P (RNA) Four channel and even five channel equivalents are also possible: 1 H, 13 C, 15 N, and 19 F, or 1 H, 13 C, 15 N, 31 P, etc. 2 H decoupling sometimes included 8

9 Observe probes Standard observe probes Optimized for X ( 13 C, 15 N, 31 P) observe Can also observe 1 H (and, often, 19 F), but water suppression may not be ideal Varian 500 MHz (McGill) Has tuning rods to change the X frequency range (good for 15 N, usable but not ideal for 31 P) High frequency ( 1 H) channel can also be used for 19 F Varian 300 MHz (McGill) X channel tuned to 13 C as standard High-frequency channel always tuned to both 1 H and 19 F SmartProbes Broadband probes (X optimized, but can also observe 1 H) X range stretches up to 19 F Reasonable water suppression Automated tuning on both channels Cryoprobes Transmit/receive coil is cooled to reduce thermal noise Helium-cooled Transmit/receive coil cooled to ca. 20 K Obtain an increase in S/N by a factor of 4 Savings of factor of 16 in terms of time Nitrogen-cooled Transmit/receive coil cooled with liquid nitrogen Obtain an increase in S/N by a factor of 2-3 Savings of factor of 4-9 in terms of time Some sensitivity figures McGill Chemistry QANUC Test Varian 300 Varian Varian Cryoprobe H C without decoupling P Concordia Sherbrooke UQAM U de Montréal Test Fourier (BBFO) 600 (TXI) 400 (AV I) 400 (Prodigy) 700 DCH 1 H C without decoupling P

10 What spectrometer should I use? SNR for peak at 8.1 ppm of same sealed strychnine sample (approximately 13 mm or 3 mg), using default acquisition parameters Varian Mercury 300: scans, ca. 5 min. Varian Mercury 400: 308 (435) 8 (16) scans, ca. 3 min. Avance 400: scans, ca. 7 min. Varian VNMRS 500: 260 (368) 8 (16) scans, ca. 5 min. Avance 500: scans, ca. 7 min. Why tune probes? Varian 500, detuned: 104 (147) 8 (16) scans, ca. 5 min. Varian 500, tuned: 260 (368) 8 (16) scans, ca. 5 min. High-resolution magic-angle spinning appropriate for samples in which magnetic susceptibility anisotropy limits resolution interaction removed by spinning at the magic angle liquid B 0 Motional averaging of anisotropic interactions semi-solid Solid Liquid Solid Solid solid 10

11 Solid phase synthesis: Tetrapeptide Ala-Ile-Gly- Met bound to a Merrifield resin swollen in DMF Spin-half nuclei in presence of MAS static + H 3 N O - O C O MAS, 5 khz θ m ppm 13 C / ppm over all crystallite orientations and spinning at the magic angle yields Generating a pulse Signal Generating Unit (SGU) creates shape of pulse Amplifier amplifies it, maintaining rf profile and phase 11

12 Hard pulses Shaped pulses Selective Broadband (usually adiabatic) Preamplifiers Amplify the NMR signal shortly after it is detected by the probe, so that the signal will be proportionately larger than thermal noise picked up in the rest of the rf circuitry Contain the tuning circuitry 12

13 Receiver Detects signal excited from the z direction into the xy plane Needs to work in concert with a signal shifted by 90 in order to know which side of the centreband the signal is from But, if the two channels are uneven, will get mirror image peaks about the centre So, use phase cycling Phase cycling The appearance of the line depends on the phase of the initial pulse (which determines where the signal ends up) and the phase of the receiver By making the receiver follow the phase and use all four directions in four sequential co-added experiments, any imbalance between the two channels will be eliminated, removing so-called quadrature artifacts Modern receivers Another solution to the problem of knowing where the signal is supposed to be is accomplished by: Using a very wide receiver bandwidth Centering that bandwidth well to the side of the expected signals Using a digital filter to reduce the bandwidth to the range needed This requires a receiver with a very high sampling rate (6 MHz) 13

14 Receiving signal: Digitisation Data is stored digitally but generated as analog data Analog-to-Digital Converter (ADC) converts data Small and large measurements must fit within ADC bandwidth receiver gain setting determines whether it fits too large a signal: ADC overflow (artefacts in spectrum) too small a signal: poor digitisation of noise (poor signal-to-noise ratio) Too high receiver gain / too much signal Digitisation in action SW Hz FT DW =1 / (2 SW) AQ = TD DW 14

15 Modern receiver and digital filter 15

If the magnetic field is larger, more energy is required to excite a given nucleus.

If the magnetic field is larger, more energy is required to excite a given nucleus. 1 2 If an NMR-active nucleus such as 1 H or 13 C is put into a magnet field, then it will come into resonance if it is irradiated with rf at the correct frequency. The correct frequency depends mainly

More information

NMR Basics. Lecture 2

NMR Basics. Lecture 2 NMR Basics Lecture 2 Continuous wave (CW) vs. FT NMR There are two ways of tuning a piano: - key by key and recording each sound (or frequency). - or, kind of brutal, is to hit with a sledgehammer and

More information

MT-540. D-Ribavirin, [ 3 H]- Lot A DG

MT-540. D-Ribavirin, [ 3 H]- Lot A DG MT-54 D-Ribavirin, [ 3 H]- Lot 165-145-225-A-2835-DG A) All chromatograms were run using the HPLC method described on the Product Data Sheet. Concentrations and volumes: Standard solution concentration

More information

MT Gemcitabine, [5-3 H(N)]- Lot A NK

MT Gemcitabine, [5-3 H(N)]- Lot A NK MT-1572 Gemcitabine, [5-3 H(N)]- Lot 194-14-234-A-2514-NK A) All chromatograms were run using the HPLC method described on the Product Data Sheet. Concentrations and volumes: Standard solution concentration

More information

Chem 203. Organic Spectroscopy. Midterm Examination, Part II (60 points total) Problem 1 of 4 (three out of four required, 20 points)

Chem 203. Organic Spectroscopy. Midterm Examination, Part II (60 points total) Problem 1 of 4 (three out of four required, 20 points) NAME Chem 203 Organic Spectroscopy Midterm Examination, Part II (60 points total) Problem 1 of 4 (three out of four required, 20 points) Saturday, November 15, 2014, 9 am -??? SUBMIT THREE OF THE FOUR

More information

Chem 203 December 15, Final Exam Part II Problem 3 of 3 (30 points)

Chem 203 December 15, Final Exam Part II Problem 3 of 3 (30 points) Name: Chem 203 December 15, 2012 Final Exam Part II Problem 3 of 3 (30 points) Select and submit TWO OUT OF THE THREE PROBLEMS FROM PART II for grading. Do not submit three problems. If you wish to unstaple

More information

HMBC 17. Goto. Introduction AVANCE User s Guide Bruker 185

HMBC 17. Goto. Introduction AVANCE User s Guide Bruker 185 Chapter HMBC 17 Introduction 17.1 Goto Heteronuclear Multiple Bond Correlation spectroscopy is a modified version of HMQC suitable for determining long-range 1 H- 13 C connectivity. This is useful in determining

More information

Chem 203. Organic Spectroscopy. Midterm Examination, Part II (60 points total) Problem 4 of 4 (three out of four required, 20 points)

Chem 203. Organic Spectroscopy. Midterm Examination, Part II (60 points total) Problem 4 of 4 (three out of four required, 20 points) NAME Chem 203 Organic Spectroscopy Midterm Examination, Part II (60 points total) Problem 4 of 4 (three out of four required, 20 points) Saturday, November 15, 2014, 9 am -??? SUBMIT THREE OF THE FOUR

More information

Two Dimensional Homonuclear Correlation Spectroscopy

Two Dimensional Homonuclear Correlation Spectroscopy Two Dimensional Homonuclear Correlation Spectroscopy Gradient COSY William D. Wheeler, Ph.D. Department of Chemistry University of Wyoming April 16, 1999 Revised September 22, 1999 2 INTRODUCTION Correlation

More information

Gradients. Effects of B0 gradients on transverse magnetisation Similar to figure 10 of Sattler review Progr. NMR 34 (1999), 93

Gradients. Effects of B0 gradients on transverse magnetisation Similar to figure 10 of Sattler review Progr. NMR 34 (1999), 93 Gradients 1. What are gradients? Modern high-resolution NMR probes contain -besides the RF coils - additional coils that can be fed a DC current. The coils are built so that a pulse (~1 ms long) of DC

More information

Chem 203. Organic Spectroscopy. Midterm Examination, Part II (60 points total) Problem 4 of 4 (three out of four required, 20 points)

Chem 203. Organic Spectroscopy. Midterm Examination, Part II (60 points total) Problem 4 of 4 (three out of four required, 20 points) NAME Chem 203 Organic Spectroscopy Midterm Examination, Part II (60 points total) Problem 4 of 4 (three out of four required, 20 points) Saturday, November 9, 2013, 9 am -??? SUBMIT THREE OF THE FOUR PROBLEMS

More information

Two Dimensional Homonuclear Correlation Spectroscopy

Two Dimensional Homonuclear Correlation Spectroscopy Two Dimensional Homonuclear Correlation Spectroscopy DQF-COSY William D. Wheeler, Ph.D. Department of Chemistry University of Wyoming September 23, 1999 2 INTRODUCTION Correlation Spectroscopy Correlation

More information

Chem 203 December 15, Final Exam Part II Problem 2 of 3 (30 points)

Chem 203 December 15, Final Exam Part II Problem 2 of 3 (30 points) Name: Chem 203 December 15, 2012 Final Exam Part II Problem 2 of 3 (30 points) Select and submit TWO OUT OF THE THREE PROBLEMS FROM PART II for grading. Do not submit three problems. If you wish to unstaple

More information

User manual Bruker DPX200 NMR spectrometer

User manual Bruker DPX200 NMR spectrometer User manual Bruker DPX200 NMR spectrometer Insert the NMR tube in the spinner in such a way that the bottom of the tube reaches the grey disc at the bottom of the spinnerholder. Make sure that the NMR

More information

Development of an Enantioselective Route towards the. Lycopodium Alkaloids: Total Synthesis of Lycopodine.

Development of an Enantioselective Route towards the. Lycopodium Alkaloids: Total Synthesis of Lycopodine. Development of an Enantioselective Route towards the Lycopodium Alkaloids: Total Synthesis of Lycopodine. Hua Yang and Rich G. Carter* Department of Chemistry, Oregon State University, Corvallis, OR 97331.

More information

Principios Básicos de RMN en sólidos destinado a usuarios. Gustavo Monti. Fa.M.A.F. Universidad Nacional de Córdoba Argentina

Principios Básicos de RMN en sólidos destinado a usuarios. Gustavo Monti. Fa.M.A.F. Universidad Nacional de Córdoba Argentina Principios Básicos de RMN en sólidos destinado a usuarios Gustavo Monti Fa.M.A.F. Universidad Nacional de Córdoba Argentina magnet 1 2 4 5 6 computer 3 Block diagrama of a traditional NMR spectrometer.

More information

The Agilent OneNMR Probe

The Agilent OneNMR Probe The Agilent OneNMR Probe Technical Overview Introduction The Agilent OneNMR probe represents a new class of NMR probes. This technology is the most signifi cant advance in solution-state probes in over

More information

400 MHz spectrometer user manual

400 MHz spectrometer user manual 400 MHz spectrometer user manual january 2017 Sandrine Denis-Quanquin 1. THE NMR SPECTROMETER... 3 2. MANUAL MODE / AUTOMATION... 4 2.1 SAMPLE CHANGER... 4 2.2 MANUAL MODE... 4 2.3 AUTOMATION... 4 3. PRELIMINARY

More information

Arenium Acid-Catalyzed Deuteration of Aromatic Hydrocarbons

Arenium Acid-Catalyzed Deuteration of Aromatic Hydrocarbons Supporting Information for Arenium Acid-Catalyzed euteration of Aromatic Hydrocarbons Simon uttwyler 1, Anna Butterfield 2, and Jay S. Siegel 2 * 1 epartment of Chemistry, Yale University, 350 Edwards

More information

NMR spectrometer usage at the BioNMR facility ETH Zürich

NMR spectrometer usage at the BioNMR facility ETH Zürich NMR spectrometer usage at the BioNMR facility ETH Zürich Accounts 2 Safety precautions: strong magnetic fields 2 Parts of an NMR spectrometer 3 NMR data storage 3 Start topspin software 3 Initial steps

More information

Chapter 1. 1 The NMR Spectrometer. 1.1 Components of an NMR Spectrometer The Magnet

Chapter 1. 1 The NMR Spectrometer. 1.1 Components of an NMR Spectrometer The Magnet Chapter 1 1 The NMR Spectrometer 1.1 Components of an NMR Spectrometer 1.1.1 The Magnet In most current NMR spectrometers the magnetic field is generated by a superconducting magnet (Fig. 1.1). The first

More information

Implementing ultrafast 2D NMR experiments on a Bruker Avance Spectrometer

Implementing ultrafast 2D NMR experiments on a Bruker Avance Spectrometer Implementing ultrafast 2D NMR experiments on a Bruker Avance Spectrometer Laetitia Rouger, Benoît Charrier, Serge Akoka, Patrick Giraudeau EBSI group CEISAM laboratory http://www.sciences.univ-nantes.fr/ceisam/en_ebsi1.php

More information

γ-trifluoromethyl proline: Evaluation as a structural substitute of proline for solid state 19 F-NMR peptide studies

γ-trifluoromethyl proline: Evaluation as a structural substitute of proline for solid state 19 F-NMR peptide studies Electronic Supplementary Material (ESI) for Organic & Biomolecular Chemistry. This journal is The Royal Society of Chemistry 2015 γ-trifluoromethyl proline: Evaluation as a structural substitute of proline

More information

2D heteronuclear correlation experiments

2D heteronuclear correlation experiments 2D heteronuclear correlation experiments Assistant Professor Kenneth Kongstad Bioanalytical Chemistry and Metabolomics Research Group Section for Natural Products and Peptides Department of Drug Design

More information

SUPPORTING INFORMATION

SUPPORTING INFORMATION Eur. J. Org. Chem. 2008 WILEY-VCH Verlag GmbH & Co. KGaA, 69451 Weinheim, 2008 ISSN 1434 193X SUPPORTING INFORMATION Title: Structural Elucidation with NMR Spectroscopy: Practical Strategies for Organic

More information

(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010

(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010 (N)MR Imaging Lab Course Script FMP PhD Autumn School Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder Date: November 3rd, 2010 1 Purpose: Understanding the basic principles of MR imaging

More information

10. Phase Cycling and Pulsed Field Gradients Introduction to Phase Cycling - Quadrature images

10. Phase Cycling and Pulsed Field Gradients Introduction to Phase Cycling - Quadrature images 10. Phase Cycling and Pulsed Field Gradients 10.1 Introduction to Phase Cycling - Quadrature images The selection of coherence transfer pathways (CTP) by phase cycling or PFGs is the tool that allows the

More information

Step by step procedure for NMR data acquisition

Step by step procedure for NMR data acquisition Step by step procedure for NMR data acquisition Spectrometers The UTHSCSA 500, 600, and 700 MHz spectrometers are each equipped with 4 independent RF channels and are each operated by a Red Hat Linux workstation

More information

NMR FACILITY NEWSLETTER

NMR FACILITY NEWSLETTER NMR FACILITY NEWSLETTER Department of Chemistry and Biochemistry Matt Revington-Facility Coordinator mrevingt@uwindsor.ca Ext 3997 Workshop Announcement : Advanced Topics in NMR There will be an Advanced

More information

7 Equipments. Spectrometers

7 Equipments. Spectrometers 7 Equipments Spectrometers There are three spectrometers located in the NMR laboratory: Varian UNITYplus 500 MHz (NMR500), Varian UNITYplus 500 MHz (Nightmare) and Varian INOVA 600 MHz. All spectrometers

More information

Instruction for Operating the Bruker Avance III 800 MHz NMR Spectrometers in UTMB

Instruction for Operating the Bruker Avance III 800 MHz NMR Spectrometers in UTMB Instruction for Operating the Bruker Avance III 800 MHz NMR Spectrometers in UTMB Written by Tianzhi Wang, date: February 8, 2013. No food, no drink in NMR room and no internet in NMR host computer except

More information

H Micro-Imaging. Tuning and Matching. i. Open any 1H data set and type wobb.

H Micro-Imaging. Tuning and Matching. i. Open any 1H data set and type wobb. - 1-1 H Micro-Imaging The NMR-specific properties of the objects are visualized as multidimensional images. Translational motion can be observed and spectroscopic information can be spatially resolved.

More information

Program. short intro the hardware locking and shimming 1D proton setup presaturation spectra

Program. short intro the hardware locking and shimming 1D proton setup presaturation spectra Program short intro the hardware locking and shimming 1D proton setup presaturation spectra 13 C spectra, DEPT processing of 1D file transfer and backup 2D homonuclear + 2D processing 2D heteronuclear

More information

1D Transient NOE on the Bruker DRX-500 and DRX-600

1D Transient NOE on the Bruker DRX-500 and DRX-600 1D Transient NOE on the Bruker DRX-500 and DRX-600 Reference: Stott, K., Stonehouse, J., Keeler, T.L. and Shaka, A.J., J. Amer. Chem. Soc. 1995, 117 (14), pp. 4199-4200. At thermal equilibrium in a strong

More information

In a typical biological sample the concentration of the solute is 1 mm or less. In many situations,

In a typical biological sample the concentration of the solute is 1 mm or less. In many situations, Water suppression n a typical biological sample the concentration of the solute is 1 mm or less. n many situations, the signals of interest are those of amide protons that exchange with the solvent water.

More information

Two Dimensional Heteronuclear Correlation Spectroscopy

Two Dimensional Heteronuclear Correlation Spectroscopy Two Dimensional Heteronuclear Correlation Spectroscopy Gradient HMQC William D. Wheeler, Ph.D. Department of Chemistry University of Wyoming Revised September 7, 2006 2 INTRODUCTION Correlation Spectroscopy

More information

2015 Spin echoes and projection imaging

2015 Spin echoes and projection imaging 1. Spin Echoes 1.1 Find f0, transmit amplitudes, and shim settings In order to acquire spin echoes, we first need to find the appropriate scanner settings using the FID GUI. This was all done last week,

More information

NMR Spectrometer Hardware

NMR Spectrometer Hardware NMR Spectrometer Hardware Manoj Naik NMR Facility, TIFR Workshop on NMR & its applications TIFR, Mumbai November 23, 2009 Principle of NMR ω = γ B Nuclear Magnetic Resonance A Simplified 6 0 MHz NMR Spectrometer

More information

MRI SYSTEM COMPONENTS Module One

MRI SYSTEM COMPONENTS Module One MRI SYSTEM COMPONENTS Module One 1 MAIN COMPONENTS Magnet Gradient Coils RF Coils Host Computer / Electronic Support System Operator Console and Display Systems 2 3 4 5 Magnet Components 6 The magnet The

More information

Open acqi window if the button has been lost. autolocking routine, alock= y for autolocking, alock= n for typical manual locking

Open acqi window if the button has been lost. autolocking routine, alock= y for autolocking, alock= n for typical manual locking Glossary of Common NMR Commands and Terms aa acqi ai alock aph array at points (np) axis='p' axis= pd BPsvf bc bs cd directory abort acquisition, hard stop Open acqi window if the button has been lost

More information

Workshop on Rapid Scan EPR. University of Denver EPR Center and Bruker BioSpin July 28, 2013

Workshop on Rapid Scan EPR. University of Denver EPR Center and Bruker BioSpin July 28, 2013 Workshop on Rapid Scan EPR University of Denver EPR Center and Bruker BioSpin July 28, 2013 Direct detection Direct detected magnetic resonance that is, without modulation and phase-sensitive detection

More information

8 COSY. 8.1 Introduction. 8.2 Magnitude COSY

8 COSY. 8.1 Introduction. 8.2 Magnitude COSY 8 COSY 8.1 Introduction COSY (COrrelation SpectroscopY) is a homonuclear 2D technique that is used to correlate the chemical shifts of 1 H nuclei which are J-coupled to one another. In this chapter, two

More information

Ultrahigh-resolution Total Correlation NMR Spectroscopy

Ultrahigh-resolution Total Correlation NMR Spectroscopy Ultrahigh-resolution Total Correlation NMR Spectroscopy Supporting Information Mohammadali Foroozandeh, Ralph W. Adams, Mathias Nilsson and Gareth A. Morris* All experimental spectra were recorded at a

More information

Background (~EE369B)

Background (~EE369B) Background (~EE369B) Magnetic Resonance Imaging D. Nishimura Overview of NMR Hardware Image formation and k-space Excitation k-space Signals and contrast Signal-to-Noise Ratio (SNR) Pulse Sequences 13

More information

1 Introduction. 2 The basic principles of NMR

1 Introduction. 2 The basic principles of NMR 1 Introduction Since 1977 when the first clinical MRI scanner was patented nuclear magnetic resonance imaging is increasingly being used for medical diagnosis and in scientific research and application

More information

Fast Methods for Small Molecules

Fast Methods for Small Molecules Fast Methods for Small Molecules Technical Overview Throughput is a key concern in many NMR laboratories, and using faster methods is one way to increase it. Traditionally, multidimensional NMR requires

More information

Fundamentals of NMR MRI Workshop Dayananda Sagar Institutes 29 May 2014 K.V.RAMANATHAN NMR Research Centre Indian Institute of Science Bangalore

Fundamentals of NMR MRI Workshop Dayananda Sagar Institutes 29 May 2014 K.V.RAMANATHAN NMR Research Centre Indian Institute of Science Bangalore Fundamentals of NMR MRI Workshop Dayananda Sagar Institutes 29 May 2014 K.V.RAMANATHAN NMR Research Centre Indian Institute of Science Bangalore NMR It is an Ubiquitous Technique Physics Chemistry Structural

More information

MRI RF-Coils. Innovation with Integrity. Highest sensitivity for your preclinical MRI and MRS applications. Preclinical Imaging

MRI RF-Coils. Innovation with Integrity. Highest sensitivity for your preclinical MRI and MRS applications. Preclinical Imaging MRI RF-Coils Highest sensitivity for your preclinical MRI and MRS applications Innovation with Integrity Preclinical Imaging Molecular and Preclinical Imaging Preclinical magnetic resonance imaging of

More information

NMR Spectrometer Operation: xwinnmr

NMR Spectrometer Operation: xwinnmr NMR Spectrometer Operation: xwinnmr Dr. Robert Peterson Facility Manager NMR Technology Center UCLA-DOE Institute for Genomics and Proteomics UCLA Dept. of Chemistry and Biochemistry Overview This is a

More information

PINMRF. Varian 300 MHz NMR Spectrometers User Guide for Advanced 1D and Basic 2D NMR Experiments

PINMRF. Varian 300 MHz NMR Spectrometers User Guide for Advanced 1D and Basic 2D NMR Experiments PINMRF Varian 300 MHz NMR Spectrometers User Guide for Advanced 1D and Basic 2D NMR Experiments INCLUDING: Inova-300-1 w/ 5mm 4-nucleus probe 365 WTHR Inova-300-2 w/ 5mm 4-nucleus probe 4100 BRWN Table

More information

PULSED/CW NUCLEAR MAGNETIC RESONANCE

PULSED/CW NUCLEAR MAGNETIC RESONANCE PULSED/CW NUCLEAR MAGNETIC RESONANCE The Second Generation of TeachSpin s Classic Explore NMR for both Hydrogen (at 21 MHz) and Fluorine Nuclei Magnetic Field Stabilized to 1 part in 2 million Homogenize

More information

HETERONUCLEAR IMAGING. Topics to be Discussed:

HETERONUCLEAR IMAGING. Topics to be Discussed: HETERONUCLEAR IMAGING BioE-594 Advanced MRI By:- Rajitha Mullapudi 04/06/2006 Topics to be Discussed: What is heteronuclear imaging. Comparing the hardware of MRI and heteronuclear imaging. Clinical applications

More information

KJM Version 1.0. Topspin 3.5 Windows 7 Topspin 1.3 Windows XP

KJM Version 1.0. Topspin 3.5 Windows 7 Topspin 1.3 Windows XP KJM 9250 1 H NMR spectra on the AVI-600 and AVII-600 Version 1.0 Topspin 3.5 Windows 7 Topspin 1.3 Windows XP Professor Emeritus Alistair Lawrence Wilkins, University of Waikato, New Zealand. January 2018

More information

STELAR s.r.l. via E. Fermi, Mede (PV) - Italy tel Uni-Pavia 24/11/06

STELAR s.r.l. via E. Fermi, Mede (PV) - Italy tel Uni-Pavia 24/11/06 STELAR s.r.l. via E. Fermi,4 27035 Mede (PV) - Italy tel. +39 0384 820096 www.stelar.it info@stelar.it Uni-Pavia 24/11/06 From theory to practice... Gianni Ferrante Stelar School on Field Cycling NMR relaxometry

More information

Laboratory Experiments for Nuclear Magnetic Resonance Spectroscopy May 6, 2004

Laboratory Experiments for Nuclear Magnetic Resonance Spectroscopy May 6, 2004 CONTENTS 1 Contents Laboratory Experiments for Nuclear Magnetic Resonance Spectroscopy May 6, 2004 1 Introduction 3 2 Safety 3 2.1 High Magnetic Fields......................................... 3 2.1.1

More information

Efficient Palladium-catalyzed Coupling Reactions of Aryl Bromides and Chlorides with Phenols

Efficient Palladium-catalyzed Coupling Reactions of Aryl Bromides and Chlorides with Phenols Efficient Palladium-catalyzed Coupling Reactions of Aryl Bromides and Chlorides with Phenols Tongjie Hu, a Thomas Schulz, b Christian Torborg, b Xiaorong Chen, a Jun Wang, a Matthias Beller b* and Jun

More information

NMR Spectroscopy with Radio Frequency Gradients.

NMR Spectroscopy with Radio Frequency Gradients. RF GRASP TM NMR Spectroscopy with Radio Frequency Gradients. BRUKER Werner E. Maas Bruker Instruments, Inc. 19 Fortune Drive Billerica, MA 01821 USA version 1.2 February, 1996 Copyright 1996 Bruker Instruments,

More information

KJM D-SELECTIVE NMR Experiments on the AVIIIHD-800. Version 1.0. Topspin 3.5 Windows 7

KJM D-SELECTIVE NMR Experiments on the AVIIIHD-800. Version 1.0. Topspin 3.5 Windows 7 KJM 9250 1D-SELECTIVE NMR Experiments on the AVIIIHD-800 Version 1.0 Topspin 3.5 Windows 7 Professor Emeritus Alistair Lawrence Wilkins, University of Waikato, New Zealand. January 2018 1D-SELECTIVE NMR

More information

Module 2. Artefacts and Imaging Optimisation for single shot methods. Content: Introduction. Phase error. Phase bandwidth. Chemical shift review

Module 2. Artefacts and Imaging Optimisation for single shot methods. Content: Introduction. Phase error. Phase bandwidth. Chemical shift review MRES 7005 - Fast Imaging Techniques Module 2 Artefacts and Imaging Optimisation for single shot methods Content: Introduction Phase error Phase bandwidth Chemical shift review Chemical shift in pixels

More information

Your first NMR measurement

Your first NMR measurement Your first NMR measurement Introduction Select 10mM water in D2O as NMR sample. The NMR spectrum of such sample consists of only two signals: the water signal and the peak of the reference (TSP). Follow

More information

H 2 O and fat imaging

H 2 O and fat imaging H 2 O and fat imaging Xu Feng Outline Introduction benefit from the separation of water and fat imaging Chemical Shift definition of chemical shift origin of chemical shift equations of chemical shift

More information

S600X SQUID M AGNETOMETER. S600X - For better magnetic measurements. The Better Choice. AC and DC measurements.

S600X SQUID M AGNETOMETER. S600X - For better magnetic measurements. The Better Choice. AC and DC measurements. S600X SQUID M AGNETOMETER S600X - For better magnetic measurements AC and DC measurements. lo -8 EMU sensitivity for total moment. Oscillator and extraction mode. MilliTesla field resolution and setting.

More information

Student Name: Date Completed: Supervisor:

Student Name: Date Completed: Supervisor: 2 NMR Training for the 600 MHz NMR with Chempack INOVA 600 Tests and Assignment Certification Student Name: 600-Test #1: The student will be given a written test administered by Dr. Lee. This test will

More information

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia M R I Physics Course Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia M R I Physics Course Magnetic Resonance Imaging Spatial

More information

KJM D-SELECTIVE NMR Experiments on the AVI-600 and AVII-600. Version 1.0. Topspin 3.5 Windows 7 Topspin 1.3 Windows XP

KJM D-SELECTIVE NMR Experiments on the AVI-600 and AVII-600. Version 1.0. Topspin 3.5 Windows 7 Topspin 1.3 Windows XP KJM 9250 1D-SELECTIVE NMR Experiments on the AVI-600 and AVII-600 Version 1.0 Topspin 3.5 Windows 7 Topspin 1.3 Windows XP Professor Emeritus Alistair Lawrence Wilkins, University of Waikato, New Zealand.

More information

Heterodyne Sweeping Radiometer

Heterodyne Sweeping Radiometer 46 Robezu str. LV-1004 Riga, Latvia Fax: +371-7-065102 Mm-wave Division in St. Petersburg, Russia Fax: +7-812- 326-10-60 Tel: +7-812-326-59-24 E-mail: ivanovph@nnz.ru Heterodyne Sweeping Radiometer Operation

More information

The Pulsed Resistive Low-Field MR Scanner

The Pulsed Resistive Low-Field MR Scanner 39 Chapter 3 The Pulsed Resistive Low-Field MR Scanner 3.1 Background In the remaining part of this work we are going to describe hyperpolarized gas relaxation, diffusion and MR imaging experiments. These

More information

Supplementary Information

Supplementary Information Supplementary Information CP HISQC: a better version of HSQC experiment for intrinsically disordered proteins under physiological conditions. Tairan Yuwen a & Nikolai R. Skrynnikov a,b * (a) Department

More information

Magnet technology Insights & recent UHF results

Magnet technology Insights & recent UHF results Magnet technology Insights & recent UHF results Daniel Baumann, Rainer Kümmerle Bruker Biospin AG Switzerland 25. November 2016 Brussels Innovation with Integrity Innovation with Integrity 1 GHz Aeon at

More information

Nuclear Magnetic Resonance Spectrometer (600MHz)

Nuclear Magnetic Resonance Spectrometer (600MHz) Nuclear Magnetic Resonance Spectrometer (600MHz) Specifications: State of the art analytical 600 MHz Nuclear Magnetic Resonance Spectrometer with Multi Nuclear Liquid and Solid CP-MAS Probes. I. Spectrometer

More information

Lab 8 6.S02 Spring 2013 MRI Projection Imaging

Lab 8 6.S02 Spring 2013 MRI Projection Imaging 1. Spin Echos 1.1 Find f0, TX amplitudes, and shim settings In order to acquire spin echos, we first need to find the appropriate scanner settings using the FID GUI. This was all done last week, but these

More information

Fourier Transform. louder softer. louder. softer. amplitude. time. amplitude. time. frequency. frequency. P. J. Grandinetti

Fourier Transform. louder softer. louder. softer. amplitude. time. amplitude. time. frequency. frequency. P. J. Grandinetti Fourier Transform * * amplitude louder softer amplitude louder softer frequency frequency Fourier Transform amplitude What is the mathematical relationship between two signal domains frequency Fourier

More information

Eclipse+ NMR Training Guide

Eclipse+ NMR Training Guide Eclipse+ NMR Training Guide Version 4.3.4 ECLIPSE+ NMR Training Guide Revision 20050617 Copyright 2005 by JEOL USA, Inc. Analytical Instruments Division 11 Dearborn Road Peabody, MA 01960 (978) 535-5900

More information

7 Telsa SQUID Magnetometer

7 Telsa SQUID Magnetometer 7 Telsa SQUID Magnetometer Cryogen Free / Liquid Helium Cooled www.cryogenic.co.uk Introduction S700X - For better magnetic measurements Cryogen free or Liquid Helium based system High homogeneity 7 Tesla

More information

System/Imaging Imperfections

System/Imaging Imperfections System/Imaging Imperfections B0 variations: Shim, Susceptibility B1 variations: Transmit, Receive Gradient Imperfections: Non-linearities Delays and Eddy currents Concomitant terms 1 B0 Variations - Off-Resonance

More information

An FT-ICR detection system for KATRIN

An FT-ICR detection system for KATRIN An FT-ICR detection system for KATRIN Marta Ubieto Díaz 1, Klaus Blaum 1, R. Burcu Cakirli 1, Michael Heck 1, Strahinja Lukić, Daniel Rodríguez 3, Lutz Schweikhard 4, Stefan Stahl 5 1 Max-Planck-Institut

More information

SPECIFICATIONS FOR A 4.7 TESLA/310MM BORE ACTIVELY SHIELDED MAGNET SYSTEM

SPECIFICATIONS FOR A 4.7 TESLA/310MM BORE ACTIVELY SHIELDED MAGNET SYSTEM SPECIFICATIONS FOR A 4.7 TESLA/310MM BORE ACTIVELY SHIELDED MAGNET SYSTEM Prepared by:- Magnex Scientific Limited The Magnet Technology Centre 6 Mead Road Oxford Industrial Park Yarnton, Oxford OX5 1QU,

More information

BOLD fmri: signal source, data acquisition, and interpretation

BOLD fmri: signal source, data acquisition, and interpretation BOLD fmri: signal source, data acquisition, and interpretation Cheryl Olman 4 th year student, Department of Neuroscience and Center for Magnetic Resonance Research Discussion series Week 1: Biological

More information

Using Frequency Diversity to Improve Measurement Speed Roger Dygert MI Technologies, 1125 Satellite Blvd., Suite 100 Suwanee, GA 30024

Using Frequency Diversity to Improve Measurement Speed Roger Dygert MI Technologies, 1125 Satellite Blvd., Suite 100 Suwanee, GA 30024 Using Frequency Diversity to Improve Measurement Speed Roger Dygert MI Technologies, 1125 Satellite Blvd., Suite 1 Suwanee, GA 324 ABSTRACT Conventional antenna measurement systems use a multiplexer or

More information

2 Hardware for Magnetic Resonance Imaging

2 Hardware for Magnetic Resonance Imaging Hardware for Magnetic Resonance Imaging 13 2 Hardware for Magnetic Resonance Imaging Kenneth W. Fishbein, Joseph C. McGowan, and Richard G. Spencer CONTENTS 2.1 Introduction 13 2.2 Magnets 13 2.2.1 Permanent

More information

MAKING TRANSIENT ANTENNA MEASUREMENTS

MAKING TRANSIENT ANTENNA MEASUREMENTS MAKING TRANSIENT ANTENNA MEASUREMENTS Roger Dygert, Steven R. Nichols MI Technologies, 1125 Satellite Boulevard, Suite 100 Suwanee, GA 30024-4629 ABSTRACT In addition to steady state performance, antennas

More information

Radio Receivers. Al Penney VO1NO

Radio Receivers. Al Penney VO1NO Radio Receivers Al Penney VO1NO Role of the Receiver The Antenna must capture the radio wave. The desired frequency must be selected from all the EM waves captured by the antenna. The selected signal is

More information

Relaxation-encoded NMR experiments for mixture analysis: REST and beer. Electronic Supporting Information

Relaxation-encoded NMR experiments for mixture analysis: REST and beer. Electronic Supporting Information Electronic Supplementary Material (ESI) for ChemComm. This journal is The Royal Society of Chemistry 2017 Relaxation-encoded NMR experiments for mixture analysis: REST and beer Electronic Supporting Information

More information

Pulse Sequence Design and Image Procedures

Pulse Sequence Design and Image Procedures Pulse Sequence Design and Image Procedures 1 Gregory L. Wheeler, BSRT(R)(MR) MRI Consultant 2 A pulse sequence is a timing diagram designed with a series of RF pulses, gradients switching, and signal readout

More information

Chapter 11 Coherence Editing: Pulse-field Gradients and Phase Cycling

Chapter 11 Coherence Editing: Pulse-field Gradients and Phase Cycling Chapter 11 Coherence Editing: Pulse-field Gradients and Phase Cycling Coherence editing is used to remove unwanted signals from NMR spectra. For example, in the double quantum filtered COSY experiment,

More information

1.5T HIGH FIELD SMALL ANIMAL MRI

1.5T HIGH FIELD SMALL ANIMAL MRI 1.5T HIGH FIELD SMALL ANIMAL MRI Designed Specifically for Veterinarians TECHNICAL GUIDE ADVANCING THE ART AND SCIENCE OF VETERINARY MRI The PetVet is the only high-field MRI system designed specifically

More information

Temps can range -130 to +120 C. See instructions below (part I). Such data can also be acquired on Athena and the AVANCE-360.

Temps can range -130 to +120 C. See instructions below (part I). Such data can also be acquired on Athena and the AVANCE-360. Flourine-19 Experiments on Varian Spectrometers updated: 2010 May 12 The difficulty with 19 F experiments involving decoupling (or 2D heterocorrelation) is the close proximity of the 19 F resonant frequency

More information

Using Spectral Analysis to Determine the Resonant Frequency of Vibrating Wire Gages HE Hu

Using Spectral Analysis to Determine the Resonant Frequency of Vibrating Wire Gages HE Hu 4th International Conference on Machinery, Materials and Computing Technology (ICMMCT 2016) Using Spectral Analysis to Determine the Resonant Frequency of Vibrating Wire Gages HE Hu China Institute of

More information

SPECIFICATION FOR A 7.0 TESLA/400MM ROOM TEMPERATURE BORE MAGNET SYSTEM

SPECIFICATION FOR A 7.0 TESLA/400MM ROOM TEMPERATURE BORE MAGNET SYSTEM SPECIFICATION FOR A 7.0 TESLA/400MM ROOM TEMPERATURE BORE MAGNET SYSTEM Prepared by:- Magnex Scientific Limited The Magnet Technology Centre 6 Mead Road Oxford Industrial Park Yarnton, Oxford OX5 1QU,

More information

Signs of Frequencies and Phases in NMR: The Role of Radiofrequency Mixing

Signs of Frequencies and Phases in NMR: The Role of Radiofrequency Mixing Journal of Magnetic Resonance 142, 190 194 (2000) Article ID jmre.1999.1929, available online at http://www.idealibrary.com on Signs of Frequencies and Phases in NMR: The Role of Radiofrequency Mixing

More information

PRELIMINARY SPECIFICATIONS MRBR 7.0 TESLA / 210MM ACTIVELY SHIELDED CRYO-COOLED MAGNET SYSTEM

PRELIMINARY SPECIFICATIONS MRBR 7.0 TESLA / 210MM ACTIVELY SHIELDED CRYO-COOLED MAGNET SYSTEM PRELIMINARY SPECIFICATIONS MRBR 7.0 TESLA / 210MM ACTIVELY SHIELDED CRYO-COOLED MAGNET SYSTEM Prepared by:- Magnex Scientific Limited The Magnet Technology Centre 6 Mead Road Oxford Industrial Park Yarnton,

More information

A Complete Digital Magnetic Resonance Imaging (MRI) System at Low Magnetic Field (0.1 Tesla)

A Complete Digital Magnetic Resonance Imaging (MRI) System at Low Magnetic Field (0.1 Tesla) IEEE Instrumentation and Measurement Technology Conference Anchorage, AK, USA, 21-23 May 2002 A Complete igital Magnetic Resonance Imaging (MRI) System at Low Magnetic Field (0.1 Tesla) Kosai RAOOF*, IEEE

More information

Optical Isolator Tutorial (Page 1 of 2) νlh, where ν, L, and H are as defined below. ν: the Verdet Constant, a property of the

Optical Isolator Tutorial (Page 1 of 2) νlh, where ν, L, and H are as defined below. ν: the Verdet Constant, a property of the Aspheric Optical Isolator Tutorial (Page 1 of 2) Function An optical isolator is a passive magneto-optic device that only allows light to travel in one direction. Isolators are used to protect a source

More information

Hardware. MRI System. MRI system Multicoil Microstrip. Part1

Hardware. MRI System. MRI system Multicoil Microstrip. Part1 Hardware MRI system Multicoil Microstrip MRI System Part1 1 The MRI system is made up of a variety of subsystems. the Operator Workspace Gradient Driver subsystem The Physiological Acquisition Controller

More information

SPECIFICATIONS FOR AN MRBR 7.0 TESLA / 210MM ACTIVELY SHIELDED MAGNET SYSTEM

SPECIFICATIONS FOR AN MRBR 7.0 TESLA / 210MM ACTIVELY SHIELDED MAGNET SYSTEM SPECIFICATIONS FOR AN MRBR 7.0 TESLA / 210MM ACTIVELY SHIELDED MAGNET SYSTEM Prepared by:- Magnex Scientific Limited The Magnet Technology Centre 6 Mead Road Oxford Industrial Park Yarnton, Oxford OX5

More information

System Options. Magnetic Property Measurement System. AC Susceptibility. AC Susceptibility Specifications

System Options. Magnetic Property Measurement System. AC Susceptibility. AC Susceptibility Specifications System Options AC Susceptibility Magnetic Property Measurement System Many materials display dissipative mechanisms when exposed to an oscillating magnetic field, and their susceptibility is described

More information

B12. SIMPSON Exercises

B12. SIMPSON Exercises B12 SIMPSON Exercises Thomas Vosegaard SIMPSON may be downloaded from http://www.bionmr.chem.au.dk/download/b12. SIMPSON Exercises Everything typed in Courier refers to commands to be typed on the keyboard

More information

PCS-150 / PCI-200 High Speed Boxcar Modules

PCS-150 / PCI-200 High Speed Boxcar Modules Becker & Hickl GmbH Kolonnenstr. 29 10829 Berlin Tel. 030 / 787 56 32 Fax. 030 / 787 57 34 email: info@becker-hickl.de http://www.becker-hickl.de PCSAPP.DOC PCS-150 / PCI-200 High Speed Boxcar Modules

More information

Data and Computer Communications Chapter 3 Data Transmission

Data and Computer Communications Chapter 3 Data Transmission Data and Computer Communications Chapter 3 Data Transmission Eighth Edition by William Stallings Transmission Terminology data transmission occurs between a transmitter & receiver via some medium guided

More information

Radio Receivers. Al Penney VO1NO

Radio Receivers. Al Penney VO1NO Radio Receivers Role of the Receiver The Antenna must capture the radio wave. The desired frequency must be selected from all the EM waves captured by the antenna. The selected signal is usually very weak

More information