NMR Spectroscopy with Radio Frequency Gradients.

Size: px
Start display at page:

Download "NMR Spectroscopy with Radio Frequency Gradients."

Transcription

1 RF GRASP TM NMR Spectroscopy with Radio Frequency Gradients. BRUKER Werner E. Maas Bruker Instruments, Inc. 19 Fortune Drive Billerica, MA USA version 1.2 February, 1996 Copyright 1996 Bruker Instruments, Inc. P/N B1161 wem 96 Page 1

2 Warning THE RF GRASP TM ACCESSORY OPERATES WITH HIGH VOLTAGES. ALWAYS DISCONNECT THE POWER CORD FROM THE RF CHANNEL SELECTOR UNIT BEFORE OPENING THE PROBE, THE RF CHANNEL SELECTOR OR THE RF DUPLEXER Copyright 1996 Bruker Instruments, Inc. The information contained in this manual may be altered without notice. Bruker accepts no responsibility for actions taken as a result of the use of this manual. Bruker accepts no liability for any mistakes contained in the manual, leading to coincidental damage, whether during installation or operation of the instrument. Unauthorized reproduction of manual contents, without written permission from the publishers, or translation into another language, either in full or in part, is forbidden. wem 96 Page 2

3 Table of Contents 1 Introduction Geometry of the RF Gradient Field Design of the RF GRASP TM probe Spin Dynamics Hamiltonians Gradient Conversion Cycles Coherence Orders Radiation Damping Experiments with the RF GRASP TM probe Experimental Set-Up Calibration of the RF Gradient P/N type Cosy selection Double Quantum filtered Cosy Solvent suppression with RF gradients Heteronuclear Single Quantum Correlation References Appendix Definitions for gradient switching Gradient conversion cycle rfg.nutation rfg.cosy rfg.dqfcosy rfg.watersup rfg.hsqc wem 96 Page 3

4 1 Introduction Gradients have been shown to be extremely useful for coherence selection in high resolution NMR. The use of gradients to dephase unwanted coherences prior to acquisition leads to an increase in dynamic range, as compared to the traditional phase cycling methods. Whereas B 0 or DC gradients are the most commonly employed gradient fields, RF or B 1 gradients are not new, but their application has been limited to the residual inhomogeneity of nominally homogeneous RF coils. These gradients are very useful, but their strength is limited since the coils are optimized for a homogeneous RF field. Stronger gradients can be obtained using specially designed RF coils and recently the use of a separate RF gradient coil, in addition to a homogeneous RF coil has been introduced 1-3. A reason for investigating B 1 gradients is that RF gradients offer some advantages over DC gradients, such as: short switching times no eddy currents are induced frequency selective, either for a given nucleus or for a specific resonance no need for preemphasis Both types of gradient fields, however, differ markedly in their interaction with the spin system. B 0 gradients are secular and therefore: can not induce coherence transformations couple into the spin system via I z are always orthogonal to transverse magnetization the spin dynamics are independent of the gradient coil geometry. On the other hand RF gradients are non-secular and therefore can induce coherence transformations couple into the spin system from anywhere in the transverse plane the spin dynamics are very dependent of the gradient coil geometry the orthogonality condition is only met if the gradient coil and the homogeneous coil are the same. wem 96 Page 4

5 2 Geometry of the RF Gradient Field Figure 1 Geometry of the RF fields generated by a homogeneous coil and a quadrupolar coil, related to the laboratory axes u and v. The spatial variation of the RF gradient field can, by design, be contained in the amplitude of the field or in the phase of the field or in both simultaneously. In the case of a planar gradient field the amplitude of the RF field is spatially dependent, while the phase is constant. The residual inhomogeneity of a homogeneous RF coil is a perfectly planar gradient, since the gradient and homogeneous RF are generated by one and the same coil. An RF gradient in which only the phase varies over space is used in radial spectroscopy 4. In our approach to RF gradient spectroscopy we have used a quadrupolar RF coil. The field generated by a quadrupolar coil has a spatially dependent phase and amplitude which vary in the transverse plane and are constant along the z-direction. The RF amplitude is zero in the center of the transverse plane and increases with increasing radius, while the phase at constant radius varies over 2π (see Figure 1). The quadrupolar coil design is chosen because it optimizes RF efficiency and filling factor and it generates the maximum gradient strength. 3 Design of the RF GRASP TM probe The Bruker RF GRASP TM is equipped with both a homogeneous 1 H RF coil and a quadrupolar gradient 1 H RF coil. The probe is similar to an inverse detection probe, but the 2 D lock will appear weaker than one is accustomed to using conventional probes. As a result one may have to use higher values for the lock power than usual. Problems with locking are anticipated when using small amounts of lock substance or when the magnetic field is inhomogeneous. It is therefore wem 96 Page 5

6 suggested to shim the probe while locking on a strong deuterium signal (such as at least 50% D 2 O or 50% D-acetone) before attempting to lock on less sensitive samples. The homogeneous and gradient RF coils are electrically isolated via an active switching mechanism. The coil switching is driven by the RF-Channel Selector TM, which is controlled from the pulse program via a control signal. By convention NMRword8 bit 7 is used to switch the probe between the gradient mode and the homogeneous mode. In the normal state (CH1 high, light on) the homogeneous coil is active and the gradient coil is turned off. By changing the status of the nmrword from the pulse programmer (see appendix and the experimental section) the gradient coil is activated (CH2 high) and the homogeneous coil is switched of. RF is delivered to both coils via separate inputs on the probe. The RF power can be generated from two different amplifiers or it can be derived from a single amplifier by using the RF GRASP TM -Duplexer. The Duplexer has a single RF input port and two RF output ports, one for each proton channel on the probe. In normal operation (CH1 and CH3 high) the RF power is routed to RF output 1 (RF OUT1) on the duplexer, which should be connected to the homogeneous RF coil via the preamp. Upon switching the probe to the RF gradient channel (CH2 and CH4 high) the duplexer switches the RF power to RF OUT2, which is connected to the gradient channel on the probe. 4 Spin Dynamics 4.1 Hamiltonians The interaction of an RF pulse with phase θ on a homogeneous coil with the sample can be described by a Hamiltonian H h (1) H h = γ B 1h ( I x cosθ + I y sinθ) In the case of an x-pulse (θ=0), magnetization perpendicular to the rotating frame x-axis will rotate in the yz-plane and magnetization along the x-axis will remain invariant. The Hamiltonian H g for an RF field with phase φ and generated by a quadrupolar gradient coil is given by: (2) H g = γb 1g G x ( I x cosθ+ I y sinθ) γb 1g G y ( I y cosθ sinθ) I x in which the gradient terms of the x and y components are written as G x and G y respectively and B i G i = u + u B i v v and i=x,y, with u and v the directions of the laboratory frame. A quadrupolar RF gradient pulse spreads the magnetization over the surface of a sphere rather than dispersing the magnetization in a plane in the case of B 0 or planar B 1 gradient fields. Magnetization starting along the z-axis will thus be completely dephased, whereas magnetization components along the rotating frame x and y-axes will remain partially invariant. wem 96 Page 6

7 4.2 Gradient Conversion Cycles Figure 2 Radial to planar gradient conversion cycle, based on π pulses. In experiments one sometimes wishes to use a quadrupolar RF gradient while other experiments demand a planar gradient evolution. A quadrupolar RF gradient pulse will by itself dephase all z- components of the magnetization but spin-lock part of the x and y components. Some experiments, such as the solvent suppression sequence 5, the double quantum filtered COSY experiments 6 and a version of the heteronuclear correlation experiment 7 described below, are designed to use the quadrupolar gradient evolutions as is, by a combination of RF gradient pulses and homogeneous pulses. Other experiments, however, need planar gradient evolutions to eliminate unwanted coherence pathways while retaining a specific coherence. A planar gradient evolution can be achieved with a quadrupolar gradient coil via so-called radialto-planar conversion cycles 2. These conversion cycles consist of a string of homogeneous RF pulses, interleaved with periods of gradient evolution. if the gradient evolution is small compared to the nutation by the homogeneous pulses, then the average Hamiltonian adopts the symmetry dictated by the homogeneous pulses. An example of a conversion cycle which results in a planar gradient evolution is based on homogeneous π pulses: (3) G φ π x G ψ G ψ π x G φ where G φ and G ψ are the gradient RF pulses with phases φ and ψ, respectively and π x and π x are the homogeneous RF pulses, with a rotation angle of π radians and phases x and -x, respectively. The cycle is diagrammed in Figure 2. The average Hamiltonian for this sequence is: (4) H = φ + ψ γb 1g G x φ + ψ cos G 2 y φ ψ I x φ ψ cos + I 2 y wem 96 Page 7

8 φ ψ 2 This averaged Hamiltonian is equal to the Hamiltonian for an RF field with phase and 2 φ + ψ 2 2 B 1 g G x Gy φ + ψ 2 magnitude cos The phase of the resultant gradient field is no longer 2 sin dependent on the location in the gradient coil, while the spatial dependence is entirely contained in the magnitude of the resultant field. This sequence therefore equals a planar gradient which rotating frame phase is determined by the difference of the phases of the quadrupolar gradient pulses and which gradient direction in the lab frame is determined by the sum of the phases of the quadrupolar gradient pulses. 4.3 Coherence Orders z- quantized x- quantized y- quantized I z + =I x +ii y I x + =I y +ii z I y + =I z +ii x I z - =I x -ii y I x - =I y -ii z I y - =I z -ii x Table 1: Raising and lowering operators quantized along x, y and z. In the analysis of B 0 gradient experiments it is common to use the concept of coherence orders. These coherence orders are defined with respect to an I z operator and since B 0 gradients are secular and do not induce coherence transformations this concept conveniently deals with this type of gradient fields. In RF gradient spectroscopy the gradient pulse is responsible for both the dephasing/rephasing and for introducing coherence transformations. This leads to a complex mixture of spinstates due to simultaneous gradient evolution and coherence transformations, but to have a convenient picture of the dynamics it is useful to separate the two operations. One approach to this separation is to quantize the states along the rotating frame axis of the gradient field, so that there are no coherence transformations during the gradient pulse. The re-quantization involves an expansion of the Cartesian single spin operators into raising and lowering operators that are quantized along I x and I y. A straight forward expansion of the single spin Cartesian operators leads to the set of raising and lowering operators listed in Table 1. Notice that we have introduced a slight modification to the raising and lowering operator nomenclature to explicitly show the axis of quantization. From Table 1, the effective coherence numbers for the full range of two-spin Cartesian product operators can be calculated for the three axes of quantization, and are shown in Table 2. The details of these redefined coherence orders are described in a publication by Zhang et al. 8 wem 96 Page 8

9 5 Radiation Damping k x k y k z I x 0 +/-1 +/-1 I y +/-1 0 +/-1 I z +/-1 +/-1 0 I x I x 0 0,+/-2 0,+/-2 I x I y +/-1 +/-1 0,+/-2 I x I z +/-1 0,+/-2 +/-1 I y I x +/-1 +/-1 0,+/-2 I y I y 0,+/-2 0 0,+/-2 I y I z 0,+/-2 +/-1 +/-1 I z I x +/-1 0,+/-2 +/-1 I z I y 0,+/-2 +/-1 +/-1 I z I z 0,+/-2 0,+/-2 0 Table 2: Effective Coherence Numbers. Radiation damping is caused by the mutual coupling between a strong NMR signal and the receiver coil. The rotating magnetization induces a current in the receiver coil and in the case of a strong signal this current creates a rotating magnetic field which tilts the magnetization back to its equilibrium position. The occurrence of radiation damping during non-observation windows in the pulse sequence can be prevented by eliminating the mutual coupling between the sample and the receiver coil. In the RF GRASP TM probe this is simply done by switching the probe to the gradient channel during free precession delays, which prevents a current to be generated in the homogeneous coil. Although the magnetization can now in principle couple to the RF gradient coil, the magnetization has the wrong symmetry to induce a current in this coil, so that radiation damping is prevented. Two recent publications 9,10 describe experiments involving Q-switching to avoid radiation damping, using Bruker's new line of Q-Switch TM probes. Similar results can be obtained with the RF GRASP TM probe. wem 96 Page 9

10 6 Experiments with the RF GRASP TM probe In the following sections some examples are given of RF gradient experiments 2,5-8. The pulse programs for these experiments are given in the appendix. Further experiments involving RF gradients are described in publications by Canet et al , Keeler et al. 1,15,16 and Eads Experimental Set-Up The set-up of the RF GRASP TM accessory is shown schematically in Figure 3. Please read the instruction manual for the rf channel selector unit. The RF GRASP TM accessory consist of an RF gradient probe, an RF duplexer and an RF channel selector unit. The RF duplexer is used to direct proton RF power to either the gradient or the homogeneous coil in the probe, when using a single RF amplifier. Under normal operation, i.e. observing the NMR signal on the homogeneous RF coil, RF OUT1 is connected to the preamp and RF OUT2 is connected directly to the RF gradient connection of the probe. (If a second proton amplifier is used, its output is connected directly to the RF gradient connection of the probe and the RF duplexer is not needed.) A burndy cable connects the DUPLEXER burndy output on the RF channel selector with the duplexer and another burndy cable connects the PROBE burndy output on the RF channel selector with the probe. For probe tuning purposes the RF channel selector can be manually switched between the homogeneous channel (CH1/CH3 HIGH) and the gradient channel (CH2/CH4 HIGH). Switching of the probe between the homogeneous and gradient RF channels is performed from the pulse program via NMRword8 bit 7. The following definitions are used: #define HOM setnmr8^7 #define RFG setnmr8 7 These definitions are included in each pulse program or alternatively can be added to the Avance.incl file (see appendix). wem 96 Page 10

11 Figure 3 Set-up of the RF GRASP TM accessory. wem 96 Page 11

12 6.2 Calibration of the RF Gradient RF GRASP TM Figure 4 Pulse sequence of the 1D nutation experiment. RF field strengths can be calibrated by measuring the RF nutation frequency. Whereas in the case of a homogeneous coil the nutation frequency is simply determined from the length of a π/2 pulse (υ rf =1/(4*P 90 ), this is not possible for an RF gradient coil due to the distribution in field strength across the sample. In the latter case the gradient strength is obtained from a so-called nutation experiment, which registers the evolution of the magnetization under an RF field. A nutation experiment can be performed in a 2D fashion by acquiring Fid's as a function of increasing RF gradient pulse length, or in a 1D manner by acquiring datapoints during intervals in which the RF pulse is interrupted. The 1D nutation sequence is easily used to calibrate the RF gradient strength. In this experiment the pulse length p2 takes the place of the dwell time in a regular acquisition. An axis labeling in khz is obtained by setting DW =p2*1000 (in µs), FW=3e6, AQMOD=QF. (Note that both excitation and receiving occur on the RF gradient channel, so in the set-up connect the RF for the gradient via the preamp to the probe!). wem 96 Page 12

13 Figure 5 Proton nutation spectra of a homogeneous RF coil (top) and a quadrupolar RF gradient coil (bottom). wem 96 Page 13

14 6.3 P/N type Cosy selection Figure 6 Pulse sequence for the N-type selected COSY experiment. The N-type selected COSY experiment (Figure 6) is performed in a single scan with RF gradients in composite z-rotations 2. An effective z-rotation is achieved by sandwiching a planar gradient π between two π/2 pulses: -- π. A planar gradient evolution is created with quadrupolar 2 G y x -- 2 y RF gradients, combined with π pulses in a conversion cycle. In the conversion cycle in the pulse sequence of Figure 6, φ and ψ are chosen equal, resulting in an effective RF gradient G x. The strength of the gradient is adjusted either by changing the length of the individual quadrupolar RF gradient pulses or by changing the number of repetitions of the conversion cycle. wem 96 Page 14

15 Figure 7 N-type 400 MHz proton COSY spectrum of 1-chloro-3-nitrobenzene in deuterated benzene, represented in magnitude mode and obtained with a single scan per t 1 increment. The gradient strength is 60 khz/cm and the total duration of the gradient pulse is 4 m wem 96 Page 15

16 6.4 Double Quantum filtered Cosy RF GRASP TM Figure 8 Pulse program of the RFG DQF COSY experiment with quadrupolar RF gradients. A version of a double quantum filtered Cosy experiment with RF gradients is depicted in Figure 8. This particular phase sensitive RFG DQF COSY experiment 6 is robust because the use of two 90 degree phase shifted gradients (ratio 1:2) renders it independent of the phase difference between the gradient and homogeneous RF. In samples in which a strong solvent signal may cause radiation damping, the probe is switched to the gradient mode during the evolution period t 1. This eliminates line broadening due to radiation damping in the first frequency domain and improves the suppression of the solvent signal by the double quantum filter. wem 96 Page 16

17 Figure MHz proton phase sensitive RFG DQF COSY spectrum of sucrose in 90% H 2 O. The quadrupolar gradient pulse duration is 700 µs. wem 96 Page 17

18 6.5 Solvent suppression with RF gradients Figure 10 Pulse program of the RFG solvent suppression experiment with quadrupolar RF gradients. The dephasing of the magnetization due to an RF gradient pulse can be refocused by applying a π pulse to the spin system followed by a second gradient pulse of equal strength. An RF gradient spin echo sequence can be used for solvent suppression by making the inversion π pulse frequency selective, such that the solvent magnetization is not inverted and therefore is not refocused by the second gradient 5. An example is shown in Figure 10, consisting of two quadrupolar RF gradient pulses and a frequency selective inversion sequence consisting of a hard π pulse sandwiched between two soft π 2 pulses, selective for the solvent frequency. The inversion sequence is repeated with a 90 degree phase shift in order to recover the full signal intensity. The magnetization of spins other than the solvent spins, is refocused along the +z-axis and is observed following a π 2 pulse. wem 96 Page 18

19 Figure 11 Single scan 400 MHz proton spectra of sucrose in water. The top spectrum is a single pulse spectrum and the bottom spectrum is obtained by suppressing the water resonance with quadrupolar RF gradients. The RF gradient pulse duration is 500 µs and the selective pulse length is 2.78 ms. wem 96 Page 19

20 6.6 Heteronuclear Single Quantum Correlation Figure 12 Pulse program of an HSQC experiment with quadrupolar RF gradients. RF gradients are frequency selective and in the RF GRASP TM setup act only on proton spins. Heteronuclear coherences are, however, affected but only through the coupling to proton spins 7,18. This situation therefore demands an experimental design which differs from a certain class of heteronuclear experiments in which the evolution under a B 0 gradient of proton magnetization can be refocussed by a matched gradient when the coherence has a transverse heteronuclear component 19. An example of an RF gradient HSQC experiment involves longitudinal storage of the X-nucleus magnetization 7. After the creation of antiphase S-magnetization via an INEPT sequence, the antiphase coherence is allowed to evolve into in-phase S-magnetization, which is subsequently stored along the main magnetic field. The magnetization of protons not coupled to a heteronucleus is pulsed back to the z-axis and dephased by a quadrupolar RF gradient pulse. The S-magnetization is returned to the transverse plane and evolves during t 1 after which it evolves into an antiphase coherence which is returned to proton magnetization via the reverse INEPT step. wem 96 Page 20

21 Figure MHz 1 H- 13 C HSQC experiment of sucrose in D 2 O, employing longitudinal storage of the carbon magnetization while the uncoupled proton magnetization is dephased with a quadrupolar RF gradient pulse. wem 96 Page 21

22 7 References [1] G. Estcourt, A.L. Davis, and J. Keeler, J.Magn.Reson. 96, 191, (1992). [2] W.E. Maas, F.H. Laukien, and D.G. Cory. J.Magn.Reson. A103, 115, (1993). [3] D. Canet, D. Boudot, A. Belmajdoub, A. Retournard, and J. Brondeau. J.Magn.Reson. 79, 168, (1988). [4] D.G. Cory, F.H. Laukien, and W.E. Maas. Chem.Phys.Lett. 212, 487, (1993). [5] W.E. Maas and D.G. Cory. J.Magn.Reson. A106, 256, (1994). [6] D.G. Cory, F.H. Laukien, and W.E. Maas. J.Magn.Reson. A105, 223, (1993). [7] W.E. Maas and D.G. Cory. J.Magn.Reson. A112, 229, (1995). [8] Y. Zhang, W.E. Maas, and D.G. Cory. Molec. Phys. 86, 347, (1995). [9] C.G. Anklin, M. Rindlisbacher, G. Otting, and F.H. Laukien. J.Magn.Reson. B106, 199, (1995). [10] W.E. Maas, F.H. Laukien, and D.G. Cory. J.Magn.Reson. A113, 274, (1995). [11] D. Canet, P. Tekely, N. Mahieu, and D. Boudot. Chem.Phys.Lett. 182, 541, (1991). [12] J. Brondeau, D. Boudot, P. Mutzenhardt, and D. Canet. J.Magn.Reson. 100, 611, (1992). [13] D. Canet, J. Brondeau, E. Mischler, and F. Humbert. J.Mag.Reson. A105, 239, (1993). [14] P. Mutzenhardt, J. Brondeau, and D. Canet. J.Magn.Reson. A108, 110, (1994). [15] J.J. Titman, A.L. Davis, E.D. Laue, and J. Keeler. J.Magn.Reson. 89, 176, (1990). [16] A.L. Davis, G. Estcourt, J. Keeler, E.D. Laue, and J.J. Titman. J.Magn.Reson. A105, 167, (1993). [17] C.D. Eads. J.Magn.Reson. A107, 109, (1994). [18] G. Otting and K. Wuthrich. J.Magn.Reson. 76, 569, (1988). [19] R.E. Hurd and B.K. John. J.Magn.Reson. 91, 648, (1991). wem 96 Page 22

23 8 Appendix 8.1 Definitions for gradient switching Add the following definitions to the preamble of the pulse programs for RF gradient spectroscopy or add these definitions to the Avance.incl file. #define HOM setnmr8^7 #define RFG setnmr8 7 ;switch to homogeneous channel ;switch to RF gradient channel 8.2 Gradient conversion cycle ;rfg.conv ;quadrupolar to planar gradient conversion cycle ;insert in pulse programs for planar gradient evolution 10 5u RFG p20:f1 ph20 ;RF gradient pulse phase phi 5u HOM p3:f1 ph11 ;180 degree pulse phase x 5u RFG p20:f1 ph21 ;RF gradient pulse phase psi 5u p20:f1 ph21 ;RF gradient pulse phase psi 5u HOM p3:f1 ph12 ;180 degree pulse phase -x 5u RFG p20:f1:c4 ph20 ;RF gradient pulse phase phi lo to 10 times l1 8.3 rfg.nutation ;rfg.nutation ;1D nutation sequence ;excitation and acquisition on RF gradient channel #include <Avance.incl> #define HOM setnmr8^7 ;switch to homogeneous channel #define RFG setnmr8 7 ;switch to RF gradient channel 1 ze 2 d1 RFG ;switch to gradient mode wem 96 Page 23

24 3u ph0 adc 3 3u p2:f1 ph1 d2 5u:x lo to 3 times l1 rcyc=2 ph31 wr #0 5u HOM exit ;gradient pulse ;acquire 1 datapoint ;repeat l1 times; l1=td ;switch back to homogeneous mode ph0=0 ph1=0 ph31=0 ;p2 RF gradient pulse ;AQMOD=QF ;DW=p2*1000 (for axis labeling in khz) ;FW=3e6 8.4 rfg.cosy ;rfg.cosy ;N-type COSY with RF gradients in composite z-rotation ;reference: W.E. Maas et al. ; J. Magn.Reson. A103, 115, 1993 #include <Avance.incl> #define HOM setnmr8^7 #define RFG setnmr8 7 ;switch to homogeneous channel ;switch to RF gradient channel 1 ze 2 d1 HOM 3 p1:f1 ph1 ;90 degree pulse d0 p1:f1 ph10 ;start composite z-rotation 10 5u RFG ;start gradient conversion cycle p20:f1 ph20 ;gradient pulse 5u HOM p3:f1 ph11 ;180 degree pulse 5u RFG wem 96 Page 24

25 p20:f1 ph21 5u p20:f1 ph21 5u HOM p3:f1 ph12 5u RFG p20:f1 ph20 lo to 10 times l1 5u HOM p1:f1 ph13 p1:f1 ph2 p1:f1 ph10 ;end gradient conversion cycle ;end composite z-rotation ;cosy mixing pulse ;start composite z-rotation 11 5u RFG ;start gradient conversion cycle p20:f1 ph20 ;gradient pulse 5u HOM p3:f1 ph11 ;180 degree pulse 5u RFG p20:f1 ph21 5u p20:f1 ph21 5u HOM p3:f1 ph12 5u RFG p20:f1 ph20 ;end gradient conversion cycle lo to 11 times l1 5u HOM p1:f1 ph13 ;end composite z-rotation go=2 ph31 d1 wr #0 if #0 id0 zd lo to 3 times td1 exit ph1=0 ph2=0 ph10=0 ph11=2 ph12=1 ph13=3 ph20=0 ph21=0 ph31=0 ;p1 : 90 degree pulse wem 96 Page 25

26 ;p3 : 180 degree pulse ;p20: RF gradient pulse ;d0 : incremented delay (2D) [3 usec] ;d1 : relaxation delay; 1-5 * T1 ;in0: 1/(1 * SW) = 2 * DW ;nd0: 1 ;td1: number of experiments ;MC2: QF 8.5 rfg.dqfcosy ;rfg.dqfcosy ;DQF COSY with quadrupolar RF gradients ;phase sensitive using TPPI ;reference: D.G.Cory et al. ; J. Magn.Reson. A105, 223, 1993 #include <Avance.incl> #define HOM setnmr8^7 #define RFG setnmr8 7 ;switch to homogeneous channel ;switch to RF gradient channel 1 ze 2 d1 3 p1:f1 ph1 ; 5u RFG ; optional switch to gradient channel d0 ; to avoid radiation damping during t1 ; 5u HOM p1:f1 ph2 5u RFG p20:f1 ph20 5u p21:f1 ph21 5u HOM go=2 ph31 d1 wr #0 if #0 id0 ip1 zd lo to 3 times td1 exit ph1=0 ph2=0 ph20=0 ph21=1 ph31=0 wem 96 Page 26

27 ;p1 : 90 degree pulse ;p20: RF gradient pulse ;p21: RF gradient pulse=2*p21 ;d0 : incremented delay (2D) [3 usec] ;d1 : relaxation delay; 1-5 * T1 ;in0: 1/(2 * SW) = DW ;nd0: 1 ;td1: number of experiments ;MC2: TPPI 8.6 rfg.watersup ;rfg.watersup ;Water suppression with quadrupolar RF gradients ;reference: W.E. Maas et al. ; J. Magn.Reson. A106, 256, 1994 #include <Avance.incl> #define HOM setnmr8^7 #define RFG setnmr8 7 1 ze 2 d1 5u RFG p20:f1 ph20 ;switch to homogeneous channel ;switch to RF gradient channel ;first RF gradient pulse 5u HOM ;selective 180)x pulse p10 pl2:f1 ph10:r ; selective pulse power pl2 p3:f1 ph11 ; 180 degree hard pulse p10 pl2:f1 ph10:r ; selective pulse power pl2 ;selective 180)y pulse p10 pl2:f1 ph12:r ; p3:f1 ph13 ; (see above) p10 pl2:f1 ph12:r ; 5u RFG p20:f1 ph20 5u HOM p1:f1 ph1 go=2 ph31 d1 wr #0 if #0 id0 ip1 zd lo to 3 times td1 exit ;second RF gradient pulse ;readout pulse wem 96 Page 27

28 ph1=0 ph10=0 ph11=2 ph12=1 ph13=3 ph20=0 ph31=0 ;p1 : 90 degree pulse ;p3 : 180 degree pulse ;p10: selective 90 degree pulse ;p20: RF gradient pulse ; adjust phases ph10 and ph12 with phcor10 and phcor12 to be ; exactly opposite to ph11 and ph rfg.hsqc ;rfg.hsqc ;1H-X HSQC with quadrupolar RF gradients ;phase sensitive using TPPI ;reference: W.E. Maas et al. ; J. Magn.Reson. A112, 229, 1995 #include <Avance.incl> #define HOM setnmr8^7 #define RFG setnmr8 7 ;switch to homogeneous channel ;switch to RF gradient channel 1 ze 2 d1 3 (p1 ph1):f1 d4 (p3 ph2):f1 (p13 ph10):f2 d4 (p1 ph3):f1 (p11 ph11):f2 d4 (p3 ph2):f1 (p13 ph10):f2 d4 (p1 ph4):f1 (p11 ph12):f2 5u RFG (p20 ph20):f1 5u HOM (p11 ph13):f2 d0 (p3 ph2):f1 wem 96 Page 28

29 d0 d4 (p3 ph2):f1 (p13 ph11):f2 d4 (p1 ph4):f1 (p11 ph11):f2 go=2 ph31 d1 wr #0 if #0 id0 ip13 zd lo to 3 times td1 exit ph1=0 ph2=0 ph3=1 ph4=0 ph10=0 ph11=0 ph12=3 ph13=1 3 ph20=0 ph31=0 2 ;p1 : proton 90 degree pulse ;p3 : proton 180 degree pulse ;p11: X 90 degree pulse ;p13: X 180 degree pulse ;p20: RF gradient pulse ;d0 : incremented delay (2D) [3 usec] ;d1 : relaxation delay; 1-5 * T1 ;d4 : 1/(4J)XH ;in0: 1/(4 * SW_X) = 1/2* DW_X ;nd0: 4 ;td1: number of experiments ;MC2: TPPI wem 96 Page 29

10. Phase Cycling and Pulsed Field Gradients Introduction to Phase Cycling - Quadrature images

10. Phase Cycling and Pulsed Field Gradients Introduction to Phase Cycling - Quadrature images 10. Phase Cycling and Pulsed Field Gradients 10.1 Introduction to Phase Cycling - Quadrature images The selection of coherence transfer pathways (CTP) by phase cycling or PFGs is the tool that allows the

More information

Ultrahigh-resolution Total Correlation NMR Spectroscopy

Ultrahigh-resolution Total Correlation NMR Spectroscopy Ultrahigh-resolution Total Correlation NMR Spectroscopy Supporting Information Mohammadali Foroozandeh, Ralph W. Adams, Mathias Nilsson and Gareth A. Morris* All experimental spectra were recorded at a

More information

Gradients. Effects of B0 gradients on transverse magnetisation Similar to figure 10 of Sattler review Progr. NMR 34 (1999), 93

Gradients. Effects of B0 gradients on transverse magnetisation Similar to figure 10 of Sattler review Progr. NMR 34 (1999), 93 Gradients 1. What are gradients? Modern high-resolution NMR probes contain -besides the RF coils - additional coils that can be fed a DC current. The coils are built so that a pulse (~1 ms long) of DC

More information

Chapter 11 Coherence Editing: Pulse-field Gradients and Phase Cycling

Chapter 11 Coherence Editing: Pulse-field Gradients and Phase Cycling Chapter 11 Coherence Editing: Pulse-field Gradients and Phase Cycling Coherence editing is used to remove unwanted signals from NMR spectra. For example, in the double quantum filtered COSY experiment,

More information

HMBC 17. Goto. Introduction AVANCE User s Guide Bruker 185

HMBC 17. Goto. Introduction AVANCE User s Guide Bruker 185 Chapter HMBC 17 Introduction 17.1 Goto Heteronuclear Multiple Bond Correlation spectroscopy is a modified version of HMQC suitable for determining long-range 1 H- 13 C connectivity. This is useful in determining

More information

1 Introduction. 2 The basic principles of NMR

1 Introduction. 2 The basic principles of NMR 1 Introduction Since 1977 when the first clinical MRI scanner was patented nuclear magnetic resonance imaging is increasingly being used for medical diagnosis and in scientific research and application

More information

NMR Basics. Lecture 2

NMR Basics. Lecture 2 NMR Basics Lecture 2 Continuous wave (CW) vs. FT NMR There are two ways of tuning a piano: - key by key and recording each sound (or frequency). - or, kind of brutal, is to hit with a sledgehammer and

More information

2015 Spin echoes and projection imaging

2015 Spin echoes and projection imaging 1. Spin Echoes 1.1 Find f0, transmit amplitudes, and shim settings In order to acquire spin echoes, we first need to find the appropriate scanner settings using the FID GUI. This was all done last week,

More information

Principios Básicos de RMN en sólidos destinado a usuarios. Gustavo Monti. Fa.M.A.F. Universidad Nacional de Córdoba Argentina

Principios Básicos de RMN en sólidos destinado a usuarios. Gustavo Monti. Fa.M.A.F. Universidad Nacional de Córdoba Argentina Principios Básicos de RMN en sólidos destinado a usuarios Gustavo Monti Fa.M.A.F. Universidad Nacional de Córdoba Argentina magnet 1 2 4 5 6 computer 3 Block diagrama of a traditional NMR spectrometer.

More information

In a typical biological sample the concentration of the solute is 1 mm or less. In many situations,

In a typical biological sample the concentration of the solute is 1 mm or less. In many situations, Water suppression n a typical biological sample the concentration of the solute is 1 mm or less. n many situations, the signals of interest are those of amide protons that exchange with the solvent water.

More information

H Micro-Imaging. Tuning and Matching. i. Open any 1H data set and type wobb.

H Micro-Imaging. Tuning and Matching. i. Open any 1H data set and type wobb. - 1-1 H Micro-Imaging The NMR-specific properties of the objects are visualized as multidimensional images. Translational motion can be observed and spectroscopic information can be spatially resolved.

More information

Background (~EE369B)

Background (~EE369B) Background (~EE369B) Magnetic Resonance Imaging D. Nishimura Overview of NMR Hardware Image formation and k-space Excitation k-space Signals and contrast Signal-to-Noise Ratio (SNR) Pulse Sequences 13

More information

If the magnetic field is larger, more energy is required to excite a given nucleus.

If the magnetic field is larger, more energy is required to excite a given nucleus. 1 2 If an NMR-active nucleus such as 1 H or 13 C is put into a magnet field, then it will come into resonance if it is irradiated with rf at the correct frequency. The correct frequency depends mainly

More information

(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010

(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010 (N)MR Imaging Lab Course Script FMP PhD Autumn School Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder Date: November 3rd, 2010 1 Purpose: Understanding the basic principles of MR imaging

More information

Lab 8 6.S02 Spring 2013 MRI Projection Imaging

Lab 8 6.S02 Spring 2013 MRI Projection Imaging 1. Spin Echos 1.1 Find f0, TX amplitudes, and shim settings In order to acquire spin echos, we first need to find the appropriate scanner settings using the FID GUI. This was all done last week, but these

More information

Implementing ultrafast 2D NMR experiments on a Bruker Avance Spectrometer

Implementing ultrafast 2D NMR experiments on a Bruker Avance Spectrometer Implementing ultrafast 2D NMR experiments on a Bruker Avance Spectrometer Laetitia Rouger, Benoît Charrier, Serge Akoka, Patrick Giraudeau EBSI group CEISAM laboratory http://www.sciences.univ-nantes.fr/ceisam/en_ebsi1.php

More information

The Agilent OneNMR Probe

The Agilent OneNMR Probe The Agilent OneNMR Probe Technical Overview Introduction The Agilent OneNMR probe represents a new class of NMR probes. This technology is the most signifi cant advance in solution-state probes in over

More information

Fast Methods for Small Molecules

Fast Methods for Small Molecules Fast Methods for Small Molecules Technical Overview Throughput is a key concern in many NMR laboratories, and using faster methods is one way to increase it. Traditionally, multidimensional NMR requires

More information

CLIP-HSQMBC: Easy measurement of small proton-carbon coupling constants in organic molecules

CLIP-HSQMBC: Easy measurement of small proton-carbon coupling constants in organic molecules CLIP-HSQMBC: Easy measurement of small proton-carbon coupling constants in organic molecules Josep Saurí, a Teodor Parella a and Juan F. Espinosa* b Supporting information 1. Pulse sequence code (Bruker)

More information

Pulsed NMR Experiment Guide Kenneth Jackson Physics 173, Spring 2014 Professor Tsai

Pulsed NMR Experiment Guide Kenneth Jackson Physics 173, Spring 2014 Professor Tsai Pulsed NMR Experiment Guide Kenneth Jackson Physics 173, Spring 2014 Professor Tsai 1. Introduction NMR or nuclear magnetic resonance occurs when nuclei are placed in a magnetic field. It is a physical

More information

8 COSY. 8.1 Introduction. 8.2 Magnitude COSY

8 COSY. 8.1 Introduction. 8.2 Magnitude COSY 8 COSY 8.1 Introduction COSY (COrrelation SpectroscopY) is a homonuclear 2D technique that is used to correlate the chemical shifts of 1 H nuclei which are J-coupled to one another. In this chapter, two

More information

Supplementary Information

Supplementary Information Supplementary Information CP HISQC: a better version of HSQC experiment for intrinsically disordered proteins under physiological conditions. Tairan Yuwen a & Nikolai R. Skrynnikov a,b * (a) Department

More information

H 2 O and fat imaging

H 2 O and fat imaging H 2 O and fat imaging Xu Feng Outline Introduction benefit from the separation of water and fat imaging Chemical Shift definition of chemical shift origin of chemical shift equations of chemical shift

More information

MEASUREMENT AND MINIMIZATION OF FIELD INHOMOGENEITIES IN HIGH RESOLUTION NMR

MEASUREMENT AND MINIMIZATION OF FIELD INHOMOGENEITIES IN HIGH RESOLUTION NMR MEASUREMENT AND MINIMIZATION OF FIELD INHOMOGENEITIES IN HIGH RESOLUTION NMR SAMPO MATTILA Department of Chemistry, University of Oulu OULU 2001 SAMPO MATTILA MEASUREMENT AND MINIMIZATION OF FIELD INHOMOGENEITIES

More information

PHY3902 PHY3904. Nuclear magnetic resonance Laboratory Protocol

PHY3902 PHY3904. Nuclear magnetic resonance Laboratory Protocol PHY3902 PHY3904 Nuclear magnetic resonance Laboratory Protocol PHY3902 PHY3904 Nuclear magnetic resonance Laboratory Protocol GETTING STARTED You might be tempted now to put a sample in the probe and try

More information

COMMUNICATIONS Volume-Selective Multipulse Spin-Echo Spectroscopy

COMMUNICATIONS Volume-Selective Multipulse Spin-Echo Spectroscopy JOURNAL OF MAGNETC RESONANCE 72,379-384 (1987) COMMUNCATONS Volume-Selective Multipulse Spin-Echo Spectroscopy R. KMMCH* AND D. HOEPFEL? *Universitri t Urn, Sektion Kernresonanzspektroskopie, D-7900 Urn,

More information

Nuclear Magnetic Resonance (NMR)

Nuclear Magnetic Resonance (NMR) California Institute of Technology Physics 77 Nuclear Magnetic Resonance (NMR) Eric D. Black October 3, 2008 1 Theory Read Section 14.4 of Shankar, Spin Dynamics, including the optional digression on negative

More information

NMR Hardware 06/06/2017. Outline. Instrumentation: Magnet. Increasing magnetic field increases Sensitivity, by power of 3/2 Dispersion, linearly

NMR Hardware 06/06/2017. Outline. Instrumentation: Magnet. Increasing magnetic field increases Sensitivity, by power of 3/2 Dispersion, linearly NMR Hardware Outline Magnet Lock Shims Gradient Probe Signal generation and transmitters Receiver and digitizer Variable temperature system Solids hardware Instrumentation: Magnet Often the most impressive

More information

Nutation Spectra of Nuclear Quadrupole Resonance in Off-Resonance Conditions

Nutation Spectra of Nuclear Quadrupole Resonance in Off-Resonance Conditions Nutation Spectra of Nuclear Quadrupole Resonance in Off-Resonance Conditions Nicolay Sinyavsky and Mariusz Mackowiak a Baltic State Academy, Molodiozhnaya str. 6, 236029 Kaliningrad, Russia a Institute

More information

Pulse Sequence Design and Image Procedures

Pulse Sequence Design and Image Procedures Pulse Sequence Design and Image Procedures 1 Gregory L. Wheeler, BSRT(R)(MR) MRI Consultant 2 A pulse sequence is a timing diagram designed with a series of RF pulses, gradients switching, and signal readout

More information

HETERONUCLEAR IMAGING. Topics to be Discussed:

HETERONUCLEAR IMAGING. Topics to be Discussed: HETERONUCLEAR IMAGING BioE-594 Advanced MRI By:- Rajitha Mullapudi 04/06/2006 Topics to be Discussed: What is heteronuclear imaging. Comparing the hardware of MRI and heteronuclear imaging. Clinical applications

More information

SUPPORTING INFORMATION

SUPPORTING INFORMATION Eur. J. Org. Chem. 2008 WILEY-VCH Verlag GmbH & Co. KGaA, 69451 Weinheim, 2008 ISSN 1434 193X SUPPORTING INFORMATION Title: Structural Elucidation with NMR Spectroscopy: Practical Strategies for Organic

More information

Two Dimensional Heteronuclear Correlation Spectroscopy

Two Dimensional Heteronuclear Correlation Spectroscopy Two Dimensional Heteronuclear Correlation Spectroscopy Gradient HMQC William D. Wheeler, Ph.D. Department of Chemistry University of Wyoming Revised September 7, 2006 2 INTRODUCTION Correlation Spectroscopy

More information

1D Transient NOE on the Bruker DRX-500 and DRX-600

1D Transient NOE on the Bruker DRX-500 and DRX-600 1D Transient NOE on the Bruker DRX-500 and DRX-600 Reference: Stott, K., Stonehouse, J., Keeler, T.L. and Shaka, A.J., J. Amer. Chem. Soc. 1995, 117 (14), pp. 4199-4200. At thermal equilibrium in a strong

More information

A Conceptual Tour of Pulsed NMR*

A Conceptual Tour of Pulsed NMR* A Conceptual Tour of Pulsed NMR* Many nuclei, but not all, possess both a magnetic moment, µ, and an angular momentum, L. Such particles are said to have spin. When the angular momentum and magnetic moment

More information

PULSED NUCLEAR MAGNETIC RESONANCE. Advanced Laboratory, Physics 407 University of Wisconsin Madison, Wisconsin 53706

PULSED NUCLEAR MAGNETIC RESONANCE. Advanced Laboratory, Physics 407 University of Wisconsin Madison, Wisconsin 53706 (revised, 2/12/07) PULSED NUCLEAR MAGNETIC RESONANCE Advanced Laboratory, Physics 407 University of Wisconsin Madison, Wisconsin 53706 Abstract A pulsed nuclear magnetic resonance technique (spin-echo)

More information

Workshop on Rapid Scan EPR. University of Denver EPR Center and Bruker BioSpin July 28, 2013

Workshop on Rapid Scan EPR. University of Denver EPR Center and Bruker BioSpin July 28, 2013 Workshop on Rapid Scan EPR University of Denver EPR Center and Bruker BioSpin July 28, 2013 Direct detection Direct detected magnetic resonance that is, without modulation and phase-sensitive detection

More information

Student Name: Date Completed: Supervisor:

Student Name: Date Completed: Supervisor: 2 NMR Training for the 600 MHz NMR with Chempack INOVA 600 Tests and Assignment Certification Student Name: 600-Test #1: The student will be given a written test administered by Dr. Lee. This test will

More information

Relaxation-encoded NMR experiments for mixture analysis: REST and beer. Electronic Supporting Information

Relaxation-encoded NMR experiments for mixture analysis: REST and beer. Electronic Supporting Information Electronic Supplementary Material (ESI) for ChemComm. This journal is The Royal Society of Chemistry 2017 Relaxation-encoded NMR experiments for mixture analysis: REST and beer Electronic Supporting Information

More information

Chapter 1. 1 The NMR Spectrometer. 1.1 Components of an NMR Spectrometer The Magnet

Chapter 1. 1 The NMR Spectrometer. 1.1 Components of an NMR Spectrometer The Magnet Chapter 1 1 The NMR Spectrometer 1.1 Components of an NMR Spectrometer 1.1.1 The Magnet In most current NMR spectrometers the magnetic field is generated by a superconducting magnet (Fig. 1.1). The first

More information

B12. SIMPSON Exercises

B12. SIMPSON Exercises B12 SIMPSON Exercises Thomas Vosegaard SIMPSON may be downloaded from http://www.bionmr.chem.au.dk/download/b12. SIMPSON Exercises Everything typed in Courier refers to commands to be typed on the keyboard

More information

NMR FACILITY NEWSLETTER

NMR FACILITY NEWSLETTER NMR FACILITY NEWSLETTER Department of Chemistry and Biochemistry Matt Revington-Facility Coordinator mrevingt@uwindsor.ca Ext 3997 Workshop Announcement : Advanced Topics in NMR There will be an Advanced

More information

2D heteronuclear correlation experiments

2D heteronuclear correlation experiments 2D heteronuclear correlation experiments Assistant Professor Kenneth Kongstad Bioanalytical Chemistry and Metabolomics Research Group Section for Natural Products and Peptides Department of Drug Design

More information

SUPPLEMENTARY INFORMATION

SUPPLEMENTARY INFORMATION Figure S. Experimental set-up www.nature.com/nature Figure S2. Dependence of ESR frequencies (GHz) on a magnetic field (G) applied in different directions with respect to NV axis ( θ 2π). The angle with

More information

Nuclear Magnetic Resonance (NMR)

Nuclear Magnetic Resonance (NMR) California Institute of Technology Physics 77 Nuclear Magnetic Resonance (NMR) Eric D. Black September 27, 2005 1 Theory Read Section 14.4 of Shankar, Spin Dynamics, including the optional digression on

More information

EE225E/BIOE265 Spring 2011 Principles of MRI. Assignment 6 Solutions. (y 0 + vt) dt. 2 y 0T + 3 )

EE225E/BIOE265 Spring 2011 Principles of MRI. Assignment 6 Solutions. (y 0 + vt) dt. 2 y 0T + 3 ) EE225E/BIOE265 Spring 211 Principles of MRI Miki Lustig Handout Assignment 6 Solutions 1. Nishimura 6.7 (Thanks Galen!) a) After the 9 y pulse, the spin is in the ˆx direction (using left-handed rotations).

More information

400 MHz spectrometer user manual

400 MHz spectrometer user manual 400 MHz spectrometer user manual january 2017 Sandrine Denis-Quanquin 1. THE NMR SPECTROMETER... 3 2. MANUAL MODE / AUTOMATION... 4 2.1 SAMPLE CHANGER... 4 2.2 MANUAL MODE... 4 2.3 AUTOMATION... 4 3. PRELIMINARY

More information

Module 2. Artefacts and Imaging Optimisation for single shot methods. Content: Introduction. Phase error. Phase bandwidth. Chemical shift review

Module 2. Artefacts and Imaging Optimisation for single shot methods. Content: Introduction. Phase error. Phase bandwidth. Chemical shift review MRES 7005 - Fast Imaging Techniques Module 2 Artefacts and Imaging Optimisation for single shot methods Content: Introduction Phase error Phase bandwidth Chemical shift review Chemical shift in pixels

More information

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia M R I Physics Course Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia M R I Physics Course Magnetic Resonance Imaging Spatial

More information

Steady-state sequences: Spoiled and balanced methods

Steady-state sequences: Spoiled and balanced methods Steady-state sequences: Spoiled and balanced methods Karla L Miller, FMRIB Centre, University of Oxford What is steady-state imaging? In the context of MRI pulse sequences, the term steady state typically

More information

Pulse Sequence Design Made Easier

Pulse Sequence Design Made Easier Pulse Sequence Design Made Easier Gregory L. Wheeler, BSRT(R)(MR) MRI Consultant gurumri@gmail.com 1 2 Pulse Sequences generally have the following characteristics: An RF line characterizing RF Pulse applications

More information

Avance GRASP. Installation/User Manual BRUKER. Version

Avance GRASP. Installation/User Manual BRUKER. Version Avance GRASP Installation/User Manual Version 002 BRUKER The information in this manual may be altered without notice. BRUKER accepts no responsibility for actions taken as a result of use of this manual.

More information

User manual Bruker DPX200 NMR spectrometer

User manual Bruker DPX200 NMR spectrometer User manual Bruker DPX200 NMR spectrometer Insert the NMR tube in the spinner in such a way that the bottom of the tube reaches the grey disc at the bottom of the spinnerholder. Make sure that the NMR

More information

Signs of Frequencies and Phases in NMR: The Role of Radiofrequency Mixing

Signs of Frequencies and Phases in NMR: The Role of Radiofrequency Mixing Journal of Magnetic Resonance 142, 190 194 (2000) Article ID jmre.1999.1929, available online at http://www.idealibrary.com on Signs of Frequencies and Phases in NMR: The Role of Radiofrequency Mixing

More information

PINMRF. Varian 300 MHz NMR Spectrometers User Guide for Advanced 1D and Basic 2D NMR Experiments

PINMRF. Varian 300 MHz NMR Spectrometers User Guide for Advanced 1D and Basic 2D NMR Experiments PINMRF Varian 300 MHz NMR Spectrometers User Guide for Advanced 1D and Basic 2D NMR Experiments INCLUDING: Inova-300-1 w/ 5mm 4-nucleus probe 365 WTHR Inova-300-2 w/ 5mm 4-nucleus probe 4100 BRWN Table

More information

JEDI. Status of the commissioning of the waveguide RF Wien Filter

JEDI. Status of the commissioning of the waveguide RF Wien Filter COSY Beam Time Request For Lab. use Exp. No.: Session No. E 005.4 7 Collaboration: JEDI Status of the commissioning of the waveguide RF Wien Filter Spokespersons for the beam time: Ralf Gebel (Jülich)

More information

EE469B: Assignment 1 Solutions

EE469B: Assignment 1 Solutions EE469B Fall 26-7 RF Pulse Design for MRI EE469B: Assignment Solutions Due Thursday Oct 6 Introduction This assignment concerns typical Fourier transform designs of excitation pulses. This includes designing

More information

RAD 229: MRI Signals and Sequences

RAD 229: MRI Signals and Sequences RAD 229: MRI Signals and Sequences Brian Hargreaves All notes are on the course website web.stanford.edu/class/rad229 Course Goals Develop Intuition Understand MRI signals Exposure to numerous MRI sequences

More information

Two Dimensional Homonuclear Correlation Spectroscopy

Two Dimensional Homonuclear Correlation Spectroscopy Two Dimensional Homonuclear Correlation Spectroscopy DQF-COSY William D. Wheeler, Ph.D. Department of Chemistry University of Wyoming September 23, 1999 2 INTRODUCTION Correlation Spectroscopy Correlation

More information

MRI: From Signal to Image

MRI: From Signal to Image MRI: From Signal to Image Johannes Koch physics654 2013-05-06 1 / 27 Tomography Magnetic Resonance Tomography Tomography: tomos: section graphein: to write Signal measured as function of space 2 / 27 Tomography

More information

Two Dimensional Homonuclear Correlation Spectroscopy

Two Dimensional Homonuclear Correlation Spectroscopy Two Dimensional Homonuclear Correlation Spectroscopy Gradient COSY William D. Wheeler, Ph.D. Department of Chemistry University of Wyoming April 16, 1999 Revised September 22, 1999 2 INTRODUCTION Correlation

More information

PULSED/CW NUCLEAR MAGNETIC RESONANCE

PULSED/CW NUCLEAR MAGNETIC RESONANCE PULSED/CW NUCLEAR MAGNETIC RESONANCE The Second Generation of TeachSpin s Classic Explore NMR for both Hydrogen (at 21 MHz) and Fluorine Nuclei Magnetic Field Stabilized to 1 part in 2 million Homogenize

More information

Noninvasive Blood Flow Mapping with Arterial Spin Labeling (ASL) Paul Kyu Han and Sung-Hong Park

Noninvasive Blood Flow Mapping with Arterial Spin Labeling (ASL) Paul Kyu Han and Sung-Hong Park Noninvasive Blood Flow Mapping with Arterial Spin Labeling (ASL) Paul Kyu Han and Sung-Hong Park Department of Bio and Brain Engineering, Korea Advanced Institute of Science and Technology (KAIST), Daejeon,

More information

Open acqi window if the button has been lost. autolocking routine, alock= y for autolocking, alock= n for typical manual locking

Open acqi window if the button has been lost. autolocking routine, alock= y for autolocking, alock= n for typical manual locking Glossary of Common NMR Commands and Terms aa acqi ai alock aph array at points (np) axis='p' axis= pd BPsvf bc bs cd directory abort acquisition, hard stop Open acqi window if the button has been lost

More information

MAGNETIC RESONANCE IMAGING

MAGNETIC RESONANCE IMAGING CSEE 4620 Homework 3 Fall 2018 MAGNETIC RESONANCE IMAGING 1. THE PRIMARY MAGNET Magnetic resonance imaging requires a very strong static magnetic field to align the nuclei. Modern MRI scanners require

More information

ECE 185 ELECTRO-OPTIC MODULATION OF LIGHT

ECE 185 ELECTRO-OPTIC MODULATION OF LIGHT ECE 185 ELECTRO-OPTIC MODULATION OF LIGHT I. Objective: To study the Pockels electro-optic (E-O) effect, and the property of light propagation in anisotropic medium, especially polarization-rotation effects.

More information

CHAPTER - 6 PIN DIODE CONTROL CIRCUITS FOR WIRELESS COMMUNICATIONS SYSTEMS

CHAPTER - 6 PIN DIODE CONTROL CIRCUITS FOR WIRELESS COMMUNICATIONS SYSTEMS CHAPTER - 6 PIN DIODE CONTROL CIRCUITS FOR WIRELESS COMMUNICATIONS SYSTEMS 2 NOTES 3 INTRODUCTION PIN DIODE CONTROL CIRCUITS FOR WIRELESS COMMUNICATIONS SYSTEMS Chapter 6 discusses PIN Control Circuits

More information

Step by step procedure for NMR data acquisition

Step by step procedure for NMR data acquisition Step by step procedure for NMR data acquisition Spectrometers The UTHSCSA 500, 600, and 700 MHz spectrometers are each equipped with 4 independent RF channels and are each operated by a Red Hat Linux workstation

More information

Your first NMR measurement

Your first NMR measurement Your first NMR measurement Introduction Select 10mM water in D2O as NMR sample. The NMR spectrum of such sample consists of only two signals: the water signal and the peak of the reference (TSP). Follow

More information

Gradient Spoiling. Average balanced SSFP magnetization Reduce sensitivity to off-resonance. FFE, FISP, GRASS, GRE, FAST, Field Echo

Gradient Spoiling. Average balanced SSFP magnetization Reduce sensitivity to off-resonance. FFE, FISP, GRASS, GRE, FAST, Field Echo Gradient Spoiling Average balanced SSFP magnetization Reduce sensitivity to off-resonance FFE, FISP, GRASS, GRE, FAST, Field Echo 1 Gradient-Spoiled Sequence (GRE, FFE, FISP, GRASS) RF TR G z G y G x Signal

More information

7 Equipments. Spectrometers

7 Equipments. Spectrometers 7 Equipments Spectrometers There are three spectrometers located in the NMR laboratory: Varian UNITYplus 500 MHz (NMR500), Varian UNITYplus 500 MHz (Nightmare) and Varian INOVA 600 MHz. All spectrometers

More information

Single-photon excitation of morphology dependent resonance

Single-photon excitation of morphology dependent resonance Single-photon excitation of morphology dependent resonance 3.1 Introduction The examination of morphology dependent resonance (MDR) has been of considerable importance to many fields in optical science.

More information

THE INSTRUMENT. I. Introduction

THE INSTRUMENT. I. Introduction THE INSTRUMENT I. Introduction Teach Spin's PS1-A is the first pulsed nuclear magnetic resonance spectrometer signed specifically for teaching. It provides physics, chemistry, biology, geology, and other

More information

Excitation and detection schemes in pulsed EPR

Excitation and detection schemes in pulsed EPR Pure&Appl. Chem., Vol. 64, No. 6, pp. 809-814, 1992. Printed in Great Britain. @ 1992 IUPAC Excitation and detection schemes in pulsed EPR Arthur Schweiger Laboratorium fur Physikalische Chemie, Eidgenossische

More information

Evaluation of Adiabatic Frequency-Modulated Schemes for Broadband Decoupling in Isotropic Liquids

Evaluation of Adiabatic Frequency-Modulated Schemes for Broadband Decoupling in Isotropic Liquids JOURNAL OF MAGNETIC RESONANCE, Series A 119, 129 133 (1996) ARTICLE NO. 0062 Evaluation of Adiabatic Frequency-Modulated Schemes for Broadband Decoupling in Isotropic Liquids RIQIANG FU AND GEOFFREY BODENHAUSEN*

More information

Pulse Sequences: Rapid Gradient Echo

Pulse Sequences: Rapid Gradient Echo Pulse Sequences: Rapid Gradient Echo M229 Advanced Topics in MRI Holden H. Wu, Ph.D. 2018.04.17 Department of Radiological Sciences David Geffen School of Medicine at UCLA Class Business Office hours -

More information

EE225E/BIOE265 Spring 2012 Principles of MRI. Assignment 7. Due March 16, 2012

EE225E/BIOE265 Spring 2012 Principles of MRI. Assignment 7. Due March 16, 2012 EE225E/BIOE265 Spring 2012 Principles of MRI Miki Lustig Assignment 7 Due March 16, 2012 1. From Midterm I 2010: You ve just programmed up your first 2DFT pulse sequence, and are trying it out on the scanner.

More information

Program. short intro the hardware locking and shimming 1D proton setup presaturation spectra

Program. short intro the hardware locking and shimming 1D proton setup presaturation spectra Program short intro the hardware locking and shimming 1D proton setup presaturation spectra 13 C spectra, DEPT processing of 1D file transfer and backup 2D homonuclear + 2D processing 2D heteronuclear

More information

Eclipse+ NMR Training Guide

Eclipse+ NMR Training Guide Eclipse+ NMR Training Guide Version 4.3.4 ECLIPSE+ NMR Training Guide Revision 20050617 Copyright 2005 by JEOL USA, Inc. Analytical Instruments Division 11 Dearborn Road Peabody, MA 01960 (978) 535-5900

More information

EE225E/BIOE265 Spring 2014 Principles of MRI. Assignment 6. Due Friday March 7th, 2014, Self Grading Due Monday March 10th, 2014

EE225E/BIOE265 Spring 2014 Principles of MRI. Assignment 6. Due Friday March 7th, 2014, Self Grading Due Monday March 10th, 2014 EE225E/BIOE265 Spring 2014 Principles of MRI Miki Lustig 1. Read Nishimura Ch. 6 Assignment 6 Due Friday March 7th, 2014, Self Grading Due Monday March 10th, 2014 2. Nishimura assignment 6.5 3. Mimimum-Phase

More information

Data Processing at the Flaw Detector with Combined Multisector Eddy-Current Transducer

Data Processing at the Flaw Detector with Combined Multisector Eddy-Current Transducer Proc. of the nd International Conference on Applied Innovations in IT, (ICAIIT), March 04 Data Processing at the Flaw Detector with Combined Multisector Eddy-Current Transducer Evgeny Yakimov, Alexander

More information

APPLICATIONS OF PORTABLE NEAR-FIELD ANTENNA MEASUREMENT SYSTEMS

APPLICATIONS OF PORTABLE NEAR-FIELD ANTENNA MEASUREMENT SYSTEMS APPLICATIONS OF PORTABLE NEAR-FIELD ANTENNA MEASUREMENT SYSTEMS Greg Hindman Nearfield Systems Inc. 1330 E. 223rd Street Bldg. 524 Carson, CA 90745 (213) 518-4277 ABSTRACT Portable near-field measurement

More information

RF Amplifier. User Manual EDACB BRUKER. Version

RF Amplifier. User Manual EDACB BRUKER. Version RF Amplifier User Manual EDACB Version 001 BRUKER The information in this manual may be altered without notice. BRUKER accepts no responsibility for actions taken as a result of use of this manual. BRUKER

More information

Half-Pulse Excitation Pulse Design and the Artifact Evaluation

Half-Pulse Excitation Pulse Design and the Artifact Evaluation Half-Pulse Excitation Pulse Design and the Artifact Evaluation Phillip Cho. INRODUCION A conventional excitation scheme consists of a slice-selective RF excitation followed by a gradient-refocusing interval

More information

QPR No SPONTANEOUS RADIOFREQUENCY EMISSION FROM HOT-ELECTRON PLASMAS XIII. Academic and Research Staff. Prof. A. Bers.

QPR No SPONTANEOUS RADIOFREQUENCY EMISSION FROM HOT-ELECTRON PLASMAS XIII. Academic and Research Staff. Prof. A. Bers. XIII. SPONTANEOUS RADIOFREQUENCY EMISSION FROM HOT-ELECTRON PLASMAS Academic and Research Staff Prof. A. Bers Graduate Students C. E. Speck A. EXPERIMENTAL STUDY OF ENHANCED CYCLOTRON RADIATION FROM AN

More information

Extending Vector Signal Analysis to 26.5 GHz with 20 MHz Information Bandwidth Product Note

Extending Vector Signal Analysis to 26.5 GHz with 20 MHz Information Bandwidth Product Note H Extending Vector Signal Analysis to 26.5 GHz with 20 MHz Information Bandwidth Product Note 89400-13 The HP 89400 series vector signal analyzers provide unmatched signal analysis capabilities from traditional

More information

Laboratory Experiments for Nuclear Magnetic Resonance Spectroscopy May 6, 2004

Laboratory Experiments for Nuclear Magnetic Resonance Spectroscopy May 6, 2004 CONTENTS 1 Contents Laboratory Experiments for Nuclear Magnetic Resonance Spectroscopy May 6, 2004 1 Introduction 3 2 Safety 3 2.1 High Magnetic Fields......................................... 3 2.1.1

More information

Instruction for Operating the Bruker Avance III 800 MHz NMR Spectrometers in UTMB

Instruction for Operating the Bruker Avance III 800 MHz NMR Spectrometers in UTMB Instruction for Operating the Bruker Avance III 800 MHz NMR Spectrometers in UTMB Written by Tianzhi Wang, date: February 8, 2013. No food, no drink in NMR room and no internet in NMR host computer except

More information

Filters And Waveform Shaping

Filters And Waveform Shaping Physics 3330 Experiment #3 Fall 2001 Purpose Filters And Waveform Shaping The aim of this experiment is to study the frequency filtering properties of passive (R, C, and L) circuits for sine waves, and

More information

QNP Accessory. For BLAXH300/ MHz Operating & Service Manual. Version

QNP Accessory. For BLAXH300/ MHz Operating & Service Manual. Version QNP Accessory For BLAXH300/100 200-600MHz Operating & Service Manual Version 001 The information in this manual may be altered without notice. BRUKER accepts no responsibility for actions taken as a result

More information

- 1 - Rap. UIT-R BS Rep. ITU-R BS.2004 DIGITAL BROADCASTING SYSTEMS INTENDED FOR AM BANDS

- 1 - Rap. UIT-R BS Rep. ITU-R BS.2004 DIGITAL BROADCASTING SYSTEMS INTENDED FOR AM BANDS - 1 - Rep. ITU-R BS.2004 DIGITAL BROADCASTING SYSTEMS INTENDED FOR AM BANDS (1995) 1 Introduction In the last decades, very few innovations have been brought to radiobroadcasting techniques in AM bands

More information

EIA STANDARD TP-27B. Mechanical Shock (Specified Pulse) Test Procedure for Electrical Connectors EIA B ELECTRONIC INDUSTRIES ASSOCIATION

EIA STANDARD TP-27B. Mechanical Shock (Specified Pulse) Test Procedure for Electrical Connectors EIA B ELECTRONIC INDUSTRIES ASSOCIATION ANSI/-1996 Approved: April 17, 1996 EIA STANDARD TP-27B Mechanical Shock (Specified Pulse) Test Procedure for Electrical Connectors (Revision of EIA-364-27A) MAY 1996 ELECTRONIC INDUSTRIES ASSOCIATION

More information

A TECHNIQUE TO EVALUATE THE IMPACT OF FLEX CABLE PHASE INSTABILITY ON mm-wave PLANAR NEAR-FIELD MEASUREMENT ACCURACIES

A TECHNIQUE TO EVALUATE THE IMPACT OF FLEX CABLE PHASE INSTABILITY ON mm-wave PLANAR NEAR-FIELD MEASUREMENT ACCURACIES A TECHNIQUE TO EVALUATE THE IMPACT OF FLEX CABLE PHASE INSTABILITY ON mm-wave PLANAR NEAR-FIELD MEASUREMENT ACCURACIES Daniël Janse van Rensburg Nearfield Systems Inc., 133 E, 223rd Street, Bldg. 524,

More information

Fundamentals of NMR MRI Workshop Dayananda Sagar Institutes 29 May 2014 K.V.RAMANATHAN NMR Research Centre Indian Institute of Science Bangalore

Fundamentals of NMR MRI Workshop Dayananda Sagar Institutes 29 May 2014 K.V.RAMANATHAN NMR Research Centre Indian Institute of Science Bangalore Fundamentals of NMR MRI Workshop Dayananda Sagar Institutes 29 May 2014 K.V.RAMANATHAN NMR Research Centre Indian Institute of Science Bangalore NMR It is an Ubiquitous Technique Physics Chemistry Structural

More information

Understanding the Magnetic Resonance Spectrum of Nitrogen Vacancy Centers in an Ensemble of Randomly-Oriented Nanodiamonds, Supporting Information

Understanding the Magnetic Resonance Spectrum of Nitrogen Vacancy Centers in an Ensemble of Randomly-Oriented Nanodiamonds, Supporting Information Understanding the Magnetic Resonance Spectrum of Nitrogen Vacancy Centers in an Ensemble of Randomly-Oriented Nanodiamonds, Supporting Information Keunhong Jeong *1,2, Anna J. Parker *1,2, Ralph H. Page

More information

P. Robert, K. Kodera, S. Perraut, R. Gendrin, and C. de Villedary

P. Robert, K. Kodera, S. Perraut, R. Gendrin, and C. de Villedary P. Robert, K. Kodera, S. Perraut, R. Gendrin, and C. de Villedary Polarization characteristics of ULF waves detected onboard GEOS-1. Problems encountered and practical solutions XIXth U.R.S.I. General

More information

RF Pulse Toolkit: Application Specific Design

RF Pulse Toolkit: Application Specific Design RF Pulse Toolkit: Application Specific Design William A Grissom Department of Biomedical Engineering, Vanderbilt University, Nashville, TN, USA will.grissom@vanderbilt.edu Introduction RF excitation is

More information

KJM D-SELECTIVE NMR Experiments on the AVIIIHD-800. Version 1.0. Topspin 3.5 Windows 7

KJM D-SELECTIVE NMR Experiments on the AVIIIHD-800. Version 1.0. Topspin 3.5 Windows 7 KJM 9250 1D-SELECTIVE NMR Experiments on the AVIIIHD-800 Version 1.0 Topspin 3.5 Windows 7 Professor Emeritus Alistair Lawrence Wilkins, University of Waikato, New Zealand. January 2018 1D-SELECTIVE NMR

More information

Echo-Planar Imaging for a 9.4 Tesla Vertical-Bore Superconducting Magnet Using an Unshielded Gradient Coil

Echo-Planar Imaging for a 9.4 Tesla Vertical-Bore Superconducting Magnet Using an Unshielded Gradient Coil Magn Reson Med Sci, Vol. XX, No. X, pp. XXX XXX, 2015 2016 Japanese Society for Magnetic Resonance in Medicine TECHNICAL NOTE by J-STAGE doi:10.2463/mrms.tn.2015-0123 Echo-Planar Imaging for a 9.4 Tesla

More information

CH 1. Large coil. Small coil. red. Function generator GND CH 2. black GND

CH 1. Large coil. Small coil. red. Function generator GND CH 2. black GND Experiment 6 Electromagnetic Induction "Concepts without factual content are empty; sense data without concepts are blind... The understanding cannot see. The senses cannot think. By their union only can

More information