Gradients. Effects of B0 gradients on transverse magnetisation Similar to figure 10 of Sattler review Progr. NMR 34 (1999), 93

Size: px
Start display at page:

Download "Gradients. Effects of B0 gradients on transverse magnetisation Similar to figure 10 of Sattler review Progr. NMR 34 (1999), 93"

Transcription

1 Gradients 1. What are gradients? Modern high-resolution NMR probes contain -besides the RF coils - additional coils that can be fed a DC current. The coils are built so that a pulse (~1 ms long) of DC current creates a gradient of the B 0 field along z (x or y, these refer to the lab coordinate system), as if the z (x/y) shims were mis-set so that the linewidths would be ~100 khz. A signal of 100 khz line width will be completely dephased after 1 ms. Effects of B0 gradients on transverse magnetisation Similar to figure 10 of Sattler review Progr. NMR 34 (1999), µs after the gradient pulse the B0 field is homogenous again, but the magnetisation is still dephased. But this holds only for the integrated signal over the whole active volume of the sample, the magnetisation is in a welldefined state in each volume element of the active volume. Therefore, if one applies another gradient after some time, the magnetisation can be refocussed if the gradients have the right ratio.. What gradients are good for. Experiment time, resolution and sensitivity: The time one needs to spend recording a multi-d experiment can depend on the desired sensitivity or on resolution. If the sensitivity is limiting, there is no way around long experiment time. But imagine that the first H N, CA/CB plane of a CBCA(CO)NH contains all expected signals. Then the S/N collected during this time is sufficient, and the 3D doesn t need to run longer than the first plane. But if one wants certain resolutions in the CA/CB and N dimensions of certain sweep widths, there is no way around collecting a number of points given by the Nyquist condition. How long the experiment takes depends then only on the number of scans per increment. Then one should keep this number to a minimum, which means that one can only afford a short phase cycle. Gradients allow to shorten phase cycles.

2 Reducing Phase Cycle by Gradients a) Phase-cycled HSQC 1 H x y t φ R X φ 1 t 1 φ φ 3 φ 1 = x -x A φ = x x -x -x φ 3 = x x x x -x -x -x -x φ R = x -x -x x 8-step phase cycle needed, subtraction artifacts. b) Gradient enhanced HSQC: reduced Phase cycle 1 H X x y φ 1 t 1 ε ε x x 1 1 y y x ε' ε' t φ R G z G 1 G φ 1 = x -x φ R = x -x -step phase cycle fully sufficient.

3 Phase cycles in standard HSQC: At the time A of the first carbon 90 pulse with phase cycle φ 1, 13 C-bound protons have formed the operators -H z C z (amplitude sin(πj)) and H y (amplitude cos(πj)). 1 C-bound protons form only H y. Of these, only the desired -H z C z is sensitive to φ 1 and is transformed into +- H z C y. The receiver phase follows, so that the signals originating from both scans add up. The signals of the H y operators from 1 C- or 13 C-bound protons are subtracted by the inversion of the receiver phase. The φ 1 phase cycle also achieves axial peak suppression: Signals originating from relaxation of the selected operator +- H z C y will not have opposite sign at t, and will therefore also get subtracted. The phase cycles on φ and φ 3 suppress signals that could create artefacts if the 180 pulses are imperfect. Phase cycle in the gradient-enhanced HSQC: Only the φ 1 phase cycle is kept, the gradients take care of pulse imperfections (details below). gradients have more advantages: phase cycling and gradients are not equivalent: With phase cycling, the suppression is achieved by subtraction after several scans. With gradients, the suppression is achieved in each scan. Two consequences: First, Suppression by phase cycling is imperfect, which can lead to t 1 noise. Second, unwanted signals (H O) which will get subtracted away can limit the RG, which can in the extreme reduce the sensitivity. 3. Understanding gradients coherence order: Are the operators I x, I y good to understand gradients? In NOESY one often uses a gradient during the mixing time to purge unwanted signals. It turns out that this purge gradient eliminates I x,y, but in coupled spin systems antiphase signals can survive that during the mixing time were of the form I 1 xi y. How these can survive is hard to see in the I x description. I x evolves as I x cos(ωt), this one can write as I x cos(ωt) = ½( I + exp(-iωt) + I - exp(iωt); ii y sin(ωt) = ½( I + exp(-iωt) - I - exp(iωt); (That cos and sin can be expressed as complex exponentials is just the math of trigonometric functions, another way to look at it is that the cosmodulated signal is the sum of two signals I +, I -, rotating clockwise and counterclockwise) I +, I - are single element operators of coherence order p = ±1. Then the operator I 1 xi y equals ½(I I 1 -) ½(I + - I -) = ¼ (I 1 + I + - I 1 + I - + I 1 - I + - I 1 - I -) I 1 + I +, I 1 - I - are ± quanta, I 1 + I - and I 1 - I + are 0 quanta. The single element operators I +,- rotate with frequency -,+ Ω. They are dephased by gradients proportional to the gyromagnetic ratio γ and the coherence order. 0 quanta

4 rotate with the difference frequency of I 1 and I and are not dephased by gradient pulses. So I x gets dephased, because it consists of +- 1 quanta, while the 0 quantum contribution of I 1 xi y survives a purge gradient. The single element operators are more useful to understand what gradients do. How to use gradients: In the above example, the second gradient simply undid the dephasing of the first. More useful are gradients surrounding a 180 pulse. Purge Gradient: HSQC 1 Purge Gradient suppresses C-bound protons. 1 H X x y t 1 t φ 1 φ φ 3 φ R G G 1 G G p H p X H: H z > -H y > H y > H: y suppressed C-H: H z > -H y> HC x z> HC: z z unaffected

5 The purge gradient G 1 and the pair of gradient pulses are two different categories of gradients: 1. Gradients that do not dephase the desired signal, such as the purge gradient G 1 : Purge gradient in HSQC. The gradient will only suppress the water well if the pulses are perfect. The exact gradient strength is not critical.. Gradients that dephase an operator that shall in the end contribute to the signal. The pair of gradients G around the 180 pulse belongs to this category. The 180 pulse inverts the 1 H coherence order. Therefore a gradient of the same sign will refocus the signal. One can write G p before + G p after = 0. The ratio of gradients must be exactly right, otherwise the desired signal also gets destroyed. In general, one can add up the effect of all gradients on operators in a pulse sequence. A coherence transfer pathway will give a signal if the condition Σγ i p i G i =0 is fulfilled. The magnetogyric ratio γ of the nuclei had to be introduced, because gradients dephase coherences proportional to the γ (or frequency) of the transverse nucleus (nuclei) of a product operator. 4. gradient-enhanced HSQC (sometimes called gradient coherence selection). To encode chemical shifts in t 1 (any incremented dimension), the signal that is detected must be modulated with t 1. How the signal is modulated is a characteristic of the pulse sequence. Figure 1 of review Progr. NMR 34 (1999), 93 amplitude-modulating pulse sequences: Traditional pulse sequences (such as the HSQC with purge gradient) record two transients for each t 1 increment, one cos(ωt 1 ) and one sin(ωt 1 ) modulated (Ruben States Haberkorn); that is the amplitude of the signal varies with t 1 (Fig. 1a). To cos and sin modulation is linked that these sequences select C x H z or C y H z for transfer in the two transients. Fourier transformation of the individual cos or sin modulated partial FIDs along ω 1 gives the signals displayed in figure 1a, because cos and sin modulation cannot distinguish ±Ω. In the Fourier transformation that one normally carries out, the four signals of both transients are merged, so that positive and negative frequencies can be distinguished and an absorptive line is obtained. 4 signals of intensity½ get added, but also noise of 4 signals contributes:

6 S/N ratio: 4 signals of intensity½ / 4 (noise of 4 signals) = 1 phase-modulating pulse sequences: These pulse sequences also record two transients per t 1 increment, but the signal is modulated with exp(iωt 1 ) or exp(-iωt 1 ) (echo-antiecho). The amplitude of the signal is not modulated with t 1, but the phase. Product operators during double INEPT of gradient enhanced HSQC The two INEPT steps of the (gradient) enhanced HSQC transfer C x H z and C y H z. As noted above are I x,y and I +,- related. So transferring (C x + ic y )H z or (C x - ic y )H z is equivalent to transferring I + H z or I - H z (figure 1c), that is to transfer selected coherence orders: Coherence Order Selective Coherence Transfer (COS-CT). Fourier transformation of these individual signals gives signals at frequencies - Ω or +Ω, because we are selecting pure (counter)clockwise rotation. In the full Fourier transformation of the HSQC, both signals are combined into a single absorptive line. signals of intensity 1 get added, noise of signals contributes: S/N ratio: signals of intensity 1 / (noise of signals) = Another way to look at this is as follows: The first increments of a standard HSQC and a (gradient) enhanced HSQC have the same intensities. But when t 1 is incremented, the later transients contribute less signal in the HSQC, but contribute equally in the gradient enhanced HSQC. Advantages of gradient enhancement: The (gradient) enhanced HSQC is for X-H fragments more sensitive than a standard HSQC. For X-H groups, the delay 1 can be reduced from 1/(J) to 1(4J), then the gradient enhanced HSQC has equal sensitivity for X-H, and still 0% more sensitive for X-H (Schleucher et al., JBN 4 (1994), 301). In multi-d experiments, such gains can in principle be achieved for every incremented dimension (Sattler et al., JBN 6 (1995), 11). This is because each incremented dimension with amplitude modulation reduces the sensitivity of the experiment - even in the absence of relaxation. In contrast, incremented dimensions with phase modulation do - in the absence of relaxation - not reduce sensitivity. Each transfer between amplitude-modulating incremented dimensions takes 1/(J), but 1/J for a transfer between amplitude-modulating dimensions, because two components are transferred each during a duration 1/(J). So because of relaxation, double enhancement is only useful in a few experiments, such as HCCH-TOCSY (figure 3b of review) and TOCSY- HSQC. The discussion of the sensitivity is a matter of the pulse sequence, whether or not one has gradients in it. The big advantage of gradient enhancement is that the pair of gradients during t 1 and before t gives extremely good water- and artefact suppression:

7 In alternate scans, the first gradient acts on coherence order +1 or -1 of the X nucleus, and the gradient before acquisition acts on coherence order -1 of 1 H (by definition the detected signal). To select these transfers the equation Σγ i p i G i = γ H (-1) G ± γ X G 1 = 0 must hold. So the gradient ratio G 1 : G must be ± γ H : γ X. All protons that are not bound to X cannot form coherences that get refocussed after two gradients of this ratio. They get suppressed in each scan, very little help from subtraction is needed to get pure spectra. Finally, it turns out that a gradient-enhanced HSQC without decoupling in t 1 or t automatically is a TROSY pulse sequence (Yang and Kay JBN 13 (1999), 3); in principle both the doubly-narrow and doubly-broad signals get selected, but the doubly-broad signal can be suppressed or simply relaxes away during longer pulse sequences.

8 Water suppression Problems with digitisation: The digitizer can handle 16-bit numbers, therefore the water ( mm) and protein resonances (1 mm) cannot be digitized at the same time. Exchangeable protons Exchange of NH and OH ( saturation transfer ): If the water polarization is saturated before the pulse sequence, chemical exchange of H O and sample reduces the polarization of the biomolecule. Saturating water attenuates iminos at 83 K. a) Even 400 ms short presaturation destroys iminos b) Watergate does not saturate water before excitation c) Watergate and flip-back give increased signal, even for s relaxation delay

9 Problem with T 1 relaxation - long T 1 of water (T 1 3s): water saturated during pulse sequence takes longer than normal recycle delays to recover and saturates biomolecule resonances via exchange and NOE during the recycle delay. This can be avoided by "water flip-back", which restores the water along z before acquisition. Then exchange and NOE help to restore biomolecule polarization more quickly (Bax, Pervushin). Gradient HSQC with flip-back, fig. 14C of review The water-selective pulse with phase -x brings the water to -z, the gradient purges any transverse components. The 180 pulse in t 1 returns it to z, and the remaining pulses together act like a 0 pulse on the water. So the water comes out of the pulse sequence along +z, and doesn t need time to relax. This also improves water suppression. Radiation damping: H O doesn t behave like weak signals, but restores itself along the z axis much faster than by T 1. The voltages that the water (weak signals) can induce in the receiver coil are comparable to (small compared to) a weak pulse. If the full water signal is excited, the voltage induced by in the receiver coil acts as a selective pulse at the water frequency, which brings the water back to z. The water signal appears broad. If the water is inverted, there typically will be a small transverse component, which again induces a voltages, which tips the magnetisation away from -z, which again increases the transverse component. This is visible as dispersive phase of the water signal in the inversion recovery. Finally, the water will return to +z much faster than by T 1. T 1 relaxation and radiation damping of HO (next page) Radiation damping restored the water in the watergate NOESY to z at the end of the mixing time. Radiation damping can be suppressed if the water signal dephased and rephased in a controlled way during t 1 (Otting, JMR B103 (1994), 88; Sklenar JMR A114 (1995), 13).

10 T and Radiation Damping of HO 1 a) After saturation, water relaxes with T > 1 s relaxation delay, s b) After inversion, water recovers after < 300 ms relaxation delay, s

11 Criteria for water suppression: - excitation profile: is even excitation possible / needed? - What is the duration of the water suppression element, compared to the T of the signals to be observed? - Does the water suppression element dephase or saturate the water signal? Must exchange of water and biomolecule protons be expected? Can radiation damping occur? Can "water flip-back" be incorporated? - Can resonances under the water signal (H α ) be observed at least in principle? The following table is anything but an exhaustive overview of water suppression techniques, but contains only the most common techniques. Water suppression - characteristics of techniques Technique Excitation profile duration a) Dephased / flip-back Presaturation Notch 0.1 ppm µs yes / no Jump-return Very uneven 100 µs no / yes Watergate Notch 1 ppm 3 ms Avoidable / yes Gradient echo Even 5 ms Avoidable /yes spin lock Even ms yes / no Purge gradients Even ms e) Avoidable / yes a) time added to a pulse sequence without water suppression element. Presaturation is still useful during shimming, and very weak for DO samples Jump-return can be useful to observe very broad iminos. Spin lock and purge gradients are not used much any more.

12 Cryo Probe - How the sensitivity is achieved; The signal is the same, the noise can be reduced. This one can achieve if one reduces the thermal noise, by cooling the receiver coil and the 1 H preamp to 50 K. You must make sure that you are really limited by thermal noise, so they built dedicated preamps with special bandpass filters for 13C and 15N. sketch of cryo probe hardware This is basically a sphisticated fridge, that gets He out of a bottle, then has a He circulation with a compressor and an expansion unit, that provides 0 K He. This He is circulated through the probe to cool the receiver coil and the 1 H preamp. Now the noise is smaller, but to get an advantage, you still have to digitise the noise, so you must operate with higher receiver gain. Presaturation on H O doesn t work well, because it is hard achieve high enough receiver gain. - Basically used like any other probe, but some special things: - The resonance circuits have a higher Q. Much more narrow tuning curve. Tuning has to be done more carefully, Less power needed for any pulse length. Jump-return doesn t work well, because the adjustment of the two pulses becomes too sensitive. - Examples A nice advantage of the lower practicable concentrations is that linewidths decrease, so that one actually gains resolution as well. HMBC on 0.3 mm RNA

13 Adiabatic pulses Shaped pulses, often, but not necessarily with a frequency sweep. Square and adiabatic inversion pulses square inversion pulse adiabatic inversion pulse z z M y M B eff y B eff x x square pulse, non-adiabatic shapes: magnetization rotates on a cone around effective field adiabatic pulse: a spin lock that starts far off-resonance and carries spinlocked magnetization into destination state. Magnetization is aligned along the time-dependent effective field. What is the problem? A 90 pulse of some field strength γb 1 excites a band width ± γb 1. A 1.5 µs C pulse has γb 1 = 0 khz and 40 khz bandwidth, which covers 00 ppm 13 C at 800 MHz. In contrast, a 1.5 µs 180 pulse inverts only 1 khz properly, corresponding to only 100 ppm at 500 MHz. So the problem is to have inversion pulses with large bandwidth.

14 Square versus adiabatic inversion pulse test in 1D HSQC: fid dec. 5 µ s, 0 khz square pulse 400 µ s, 10 khz chirp pulse offset, khz ppm -3

15 Adiabatic inversion pulses don t need careful calibration (but just a minimum amplitude, and are therefore also less sensitive to inhomogeneity of pulse over sample volume ( B1-inhomogeneity ). Adiabatic inversion pulses can be used to create decoupling sequences for broader decoupling with less power, but more side band problems.

16 Water suppression: criteria - excitation profile: is even excitation possible / needed? - What is the duration of the water suppression element, compared to the T of the signals to be observed? - Does the water suppression element dephase or saturate the water signal? Must exchange of water and biomolecule protons be expected? Can radiation damping occur? Can "water flip-back" be incorporated? - Can resonances under the water signal (Hα) be observed at least in principle?

In a typical biological sample the concentration of the solute is 1 mm or less. In many situations,

In a typical biological sample the concentration of the solute is 1 mm or less. In many situations, Water suppression n a typical biological sample the concentration of the solute is 1 mm or less. n many situations, the signals of interest are those of amide protons that exchange with the solvent water.

More information

NMR Basics. Lecture 2

NMR Basics. Lecture 2 NMR Basics Lecture 2 Continuous wave (CW) vs. FT NMR There are two ways of tuning a piano: - key by key and recording each sound (or frequency). - or, kind of brutal, is to hit with a sledgehammer and

More information

10. Phase Cycling and Pulsed Field Gradients Introduction to Phase Cycling - Quadrature images

10. Phase Cycling and Pulsed Field Gradients Introduction to Phase Cycling - Quadrature images 10. Phase Cycling and Pulsed Field Gradients 10.1 Introduction to Phase Cycling - Quadrature images The selection of coherence transfer pathways (CTP) by phase cycling or PFGs is the tool that allows the

More information

Background (~EE369B)

Background (~EE369B) Background (~EE369B) Magnetic Resonance Imaging D. Nishimura Overview of NMR Hardware Image formation and k-space Excitation k-space Signals and contrast Signal-to-Noise Ratio (SNR) Pulse Sequences 13

More information

Principios Básicos de RMN en sólidos destinado a usuarios. Gustavo Monti. Fa.M.A.F. Universidad Nacional de Córdoba Argentina

Principios Básicos de RMN en sólidos destinado a usuarios. Gustavo Monti. Fa.M.A.F. Universidad Nacional de Córdoba Argentina Principios Básicos de RMN en sólidos destinado a usuarios Gustavo Monti Fa.M.A.F. Universidad Nacional de Córdoba Argentina magnet 1 2 4 5 6 computer 3 Block diagrama of a traditional NMR spectrometer.

More information

NMR Hardware 06/06/2017. Outline. Instrumentation: Magnet. Increasing magnetic field increases Sensitivity, by power of 3/2 Dispersion, linearly

NMR Hardware 06/06/2017. Outline. Instrumentation: Magnet. Increasing magnetic field increases Sensitivity, by power of 3/2 Dispersion, linearly NMR Hardware Outline Magnet Lock Shims Gradient Probe Signal generation and transmitters Receiver and digitizer Variable temperature system Solids hardware Instrumentation: Magnet Often the most impressive

More information

The Agilent OneNMR Probe

The Agilent OneNMR Probe The Agilent OneNMR Probe Technical Overview Introduction The Agilent OneNMR probe represents a new class of NMR probes. This technology is the most signifi cant advance in solution-state probes in over

More information

Ultrahigh-resolution Total Correlation NMR Spectroscopy

Ultrahigh-resolution Total Correlation NMR Spectroscopy Ultrahigh-resolution Total Correlation NMR Spectroscopy Supporting Information Mohammadali Foroozandeh, Ralph W. Adams, Mathias Nilsson and Gareth A. Morris* All experimental spectra were recorded at a

More information

1 Introduction. 2 The basic principles of NMR

1 Introduction. 2 The basic principles of NMR 1 Introduction Since 1977 when the first clinical MRI scanner was patented nuclear magnetic resonance imaging is increasingly being used for medical diagnosis and in scientific research and application

More information

If the magnetic field is larger, more energy is required to excite a given nucleus.

If the magnetic field is larger, more energy is required to excite a given nucleus. 1 2 If an NMR-active nucleus such as 1 H or 13 C is put into a magnet field, then it will come into resonance if it is irradiated with rf at the correct frequency. The correct frequency depends mainly

More information

2D heteronuclear correlation experiments

2D heteronuclear correlation experiments 2D heteronuclear correlation experiments Assistant Professor Kenneth Kongstad Bioanalytical Chemistry and Metabolomics Research Group Section for Natural Products and Peptides Department of Drug Design

More information

1D Transient NOE on the Bruker DRX-500 and DRX-600

1D Transient NOE on the Bruker DRX-500 and DRX-600 1D Transient NOE on the Bruker DRX-500 and DRX-600 Reference: Stott, K., Stonehouse, J., Keeler, T.L. and Shaka, A.J., J. Amer. Chem. Soc. 1995, 117 (14), pp. 4199-4200. At thermal equilibrium in a strong

More information

(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010

(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010 (N)MR Imaging Lab Course Script FMP PhD Autumn School Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder Date: November 3rd, 2010 1 Purpose: Understanding the basic principles of MR imaging

More information

Supplementary Information

Supplementary Information Supplementary Information CP HISQC: a better version of HSQC experiment for intrinsically disordered proteins under physiological conditions. Tairan Yuwen a & Nikolai R. Skrynnikov a,b * (a) Department

More information

Pulse Sequence Design and Image Procedures

Pulse Sequence Design and Image Procedures Pulse Sequence Design and Image Procedures 1 Gregory L. Wheeler, BSRT(R)(MR) MRI Consultant 2 A pulse sequence is a timing diagram designed with a series of RF pulses, gradients switching, and signal readout

More information

NMR Spectroscopy with Radio Frequency Gradients.

NMR Spectroscopy with Radio Frequency Gradients. RF GRASP TM NMR Spectroscopy with Radio Frequency Gradients. BRUKER Werner E. Maas Bruker Instruments, Inc. 19 Fortune Drive Billerica, MA 01821 USA version 1.2 February, 1996 Copyright 1996 Bruker Instruments,

More information

H 2 O and fat imaging

H 2 O and fat imaging H 2 O and fat imaging Xu Feng Outline Introduction benefit from the separation of water and fat imaging Chemical Shift definition of chemical shift origin of chemical shift equations of chemical shift

More information

Fast Methods for Small Molecules

Fast Methods for Small Molecules Fast Methods for Small Molecules Technical Overview Throughput is a key concern in many NMR laboratories, and using faster methods is one way to increase it. Traditionally, multidimensional NMR requires

More information

Module 2. Artefacts and Imaging Optimisation for single shot methods. Content: Introduction. Phase error. Phase bandwidth. Chemical shift review

Module 2. Artefacts and Imaging Optimisation for single shot methods. Content: Introduction. Phase error. Phase bandwidth. Chemical shift review MRES 7005 - Fast Imaging Techniques Module 2 Artefacts and Imaging Optimisation for single shot methods Content: Introduction Phase error Phase bandwidth Chemical shift review Chemical shift in pixels

More information

Pulse Sequence Design Made Easier

Pulse Sequence Design Made Easier Pulse Sequence Design Made Easier Gregory L. Wheeler, BSRT(R)(MR) MRI Consultant gurumri@gmail.com 1 2 Pulse Sequences generally have the following characteristics: An RF line characterizing RF Pulse applications

More information

H Micro-Imaging. Tuning and Matching. i. Open any 1H data set and type wobb.

H Micro-Imaging. Tuning and Matching. i. Open any 1H data set and type wobb. - 1-1 H Micro-Imaging The NMR-specific properties of the objects are visualized as multidimensional images. Translational motion can be observed and spectroscopic information can be spatially resolved.

More information

SUPPORTING INFORMATION

SUPPORTING INFORMATION Eur. J. Org. Chem. 2008 WILEY-VCH Verlag GmbH & Co. KGaA, 69451 Weinheim, 2008 ISSN 1434 193X SUPPORTING INFORMATION Title: Structural Elucidation with NMR Spectroscopy: Practical Strategies for Organic

More information

Implementing ultrafast 2D NMR experiments on a Bruker Avance Spectrometer

Implementing ultrafast 2D NMR experiments on a Bruker Avance Spectrometer Implementing ultrafast 2D NMR experiments on a Bruker Avance Spectrometer Laetitia Rouger, Benoît Charrier, Serge Akoka, Patrick Giraudeau EBSI group CEISAM laboratory http://www.sciences.univ-nantes.fr/ceisam/en_ebsi1.php

More information

Student Name: Date Completed: Supervisor:

Student Name: Date Completed: Supervisor: 2 NMR Training for the 600 MHz NMR with Chempack INOVA 600 Tests and Assignment Certification Student Name: 600-Test #1: The student will be given a written test administered by Dr. Lee. This test will

More information

A Conceptual Tour of Pulsed NMR*

A Conceptual Tour of Pulsed NMR* A Conceptual Tour of Pulsed NMR* Many nuclei, but not all, possess both a magnetic moment, µ, and an angular momentum, L. Such particles are said to have spin. When the angular momentum and magnetic moment

More information

Pulsed NMR Experiment Guide Kenneth Jackson Physics 173, Spring 2014 Professor Tsai

Pulsed NMR Experiment Guide Kenneth Jackson Physics 173, Spring 2014 Professor Tsai Pulsed NMR Experiment Guide Kenneth Jackson Physics 173, Spring 2014 Professor Tsai 1. Introduction NMR or nuclear magnetic resonance occurs when nuclei are placed in a magnetic field. It is a physical

More information

NMR FACILITY NEWSLETTER

NMR FACILITY NEWSLETTER NMR FACILITY NEWSLETTER Department of Chemistry and Biochemistry Matt Revington-Facility Coordinator mrevingt@uwindsor.ca Ext 3997 Workshop Announcement : Advanced Topics in NMR There will be an Advanced

More information

Workshop on Rapid Scan EPR. University of Denver EPR Center and Bruker BioSpin July 28, 2013

Workshop on Rapid Scan EPR. University of Denver EPR Center and Bruker BioSpin July 28, 2013 Workshop on Rapid Scan EPR University of Denver EPR Center and Bruker BioSpin July 28, 2013 Direct detection Direct detected magnetic resonance that is, without modulation and phase-sensitive detection

More information

PULSED NUCLEAR MAGNETIC RESONANCE. Advanced Laboratory, Physics 407 University of Wisconsin Madison, Wisconsin 53706

PULSED NUCLEAR MAGNETIC RESONANCE. Advanced Laboratory, Physics 407 University of Wisconsin Madison, Wisconsin 53706 (revised, 2/12/07) PULSED NUCLEAR MAGNETIC RESONANCE Advanced Laboratory, Physics 407 University of Wisconsin Madison, Wisconsin 53706 Abstract A pulsed nuclear magnetic resonance technique (spin-echo)

More information

2015 Spin echoes and projection imaging

2015 Spin echoes and projection imaging 1. Spin Echoes 1.1 Find f0, transmit amplitudes, and shim settings In order to acquire spin echoes, we first need to find the appropriate scanner settings using the FID GUI. This was all done last week,

More information

Your first NMR measurement

Your first NMR measurement Your first NMR measurement Introduction Select 10mM water in D2O as NMR sample. The NMR spectrum of such sample consists of only two signals: the water signal and the peak of the reference (TSP). Follow

More information

RAD 229: MRI Signals and Sequences

RAD 229: MRI Signals and Sequences RAD 229: MRI Signals and Sequences Brian Hargreaves All notes are on the course website web.stanford.edu/class/rad229 Course Goals Develop Intuition Understand MRI signals Exposure to numerous MRI sequences

More information

Lecture 7: Basics of magnetic resonance imaging (MRI): one dimensional Fourier imaging

Lecture 7: Basics of magnetic resonance imaging (MRI): one dimensional Fourier imaging Lecture 7: Basics of magnetic resonance imaging (MRI): one dimensional Fourier imaging Lecture aims to explain: 1. Basic aims of magnetic resonance imaging 2. Signal demodulation in magnetic resonance

More information

HMBC 17. Goto. Introduction AVANCE User s Guide Bruker 185

HMBC 17. Goto. Introduction AVANCE User s Guide Bruker 185 Chapter HMBC 17 Introduction 17.1 Goto Heteronuclear Multiple Bond Correlation spectroscopy is a modified version of HMQC suitable for determining long-range 1 H- 13 C connectivity. This is useful in determining

More information

HETERONUCLEAR IMAGING. Topics to be Discussed:

HETERONUCLEAR IMAGING. Topics to be Discussed: HETERONUCLEAR IMAGING BioE-594 Advanced MRI By:- Rajitha Mullapudi 04/06/2006 Topics to be Discussed: What is heteronuclear imaging. Comparing the hardware of MRI and heteronuclear imaging. Clinical applications

More information

PULSED/CW NUCLEAR MAGNETIC RESONANCE

PULSED/CW NUCLEAR MAGNETIC RESONANCE PULSED/CW NUCLEAR MAGNETIC RESONANCE The Second Generation of TeachSpin s Classic Explore NMR for both Hydrogen (at 21 MHz) and Fluorine Nuclei Magnetic Field Stabilized to 1 part in 2 million Homogenize

More information

MEASUREMENT AND MINIMIZATION OF FIELD INHOMOGENEITIES IN HIGH RESOLUTION NMR

MEASUREMENT AND MINIMIZATION OF FIELD INHOMOGENEITIES IN HIGH RESOLUTION NMR MEASUREMENT AND MINIMIZATION OF FIELD INHOMOGENEITIES IN HIGH RESOLUTION NMR SAMPO MATTILA Department of Chemistry, University of Oulu OULU 2001 SAMPO MATTILA MEASUREMENT AND MINIMIZATION OF FIELD INHOMOGENEITIES

More information

NMR spectrometer usage at the BioNMR facility ETH Zürich

NMR spectrometer usage at the BioNMR facility ETH Zürich NMR spectrometer usage at the BioNMR facility ETH Zürich Accounts 2 Safety precautions: strong magnetic fields 2 Parts of an NMR spectrometer 3 NMR data storage 3 Start topspin software 3 Initial steps

More information

Lab 8 6.S02 Spring 2013 MRI Projection Imaging

Lab 8 6.S02 Spring 2013 MRI Projection Imaging 1. Spin Echos 1.1 Find f0, TX amplitudes, and shim settings In order to acquire spin echos, we first need to find the appropriate scanner settings using the FID GUI. This was all done last week, but these

More information

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia M R I Physics Course Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia M R I Physics Course Magnetic Resonance Imaging Spatial

More information

Fundamentals of NMR MRI Workshop Dayananda Sagar Institutes 29 May 2014 K.V.RAMANATHAN NMR Research Centre Indian Institute of Science Bangalore

Fundamentals of NMR MRI Workshop Dayananda Sagar Institutes 29 May 2014 K.V.RAMANATHAN NMR Research Centre Indian Institute of Science Bangalore Fundamentals of NMR MRI Workshop Dayananda Sagar Institutes 29 May 2014 K.V.RAMANATHAN NMR Research Centre Indian Institute of Science Bangalore NMR It is an Ubiquitous Technique Physics Chemistry Structural

More information

Mapping the Flip Angle in Magnetic Resonance Imaging Using the Accelerated 3D Look-Locker Sequence

Mapping the Flip Angle in Magnetic Resonance Imaging Using the Accelerated 3D Look-Locker Sequence Western University Scholarship@Western Electronic Thesis and Dissertation Repository January 2011 Mapping the Flip Angle in Magnetic Resonance Imaging Using the Accelerated 3D Look-Locker Sequence Trevor

More information

B12. SIMPSON Exercises

B12. SIMPSON Exercises B12 SIMPSON Exercises Thomas Vosegaard SIMPSON may be downloaded from http://www.bionmr.chem.au.dk/download/b12. SIMPSON Exercises Everything typed in Courier refers to commands to be typed on the keyboard

More information

Relaxation-encoded NMR experiments for mixture analysis: REST and beer. Electronic Supporting Information

Relaxation-encoded NMR experiments for mixture analysis: REST and beer. Electronic Supporting Information Electronic Supplementary Material (ESI) for ChemComm. This journal is The Royal Society of Chemistry 2017 Relaxation-encoded NMR experiments for mixture analysis: REST and beer Electronic Supporting Information

More information

A DNP-NMR setup for sub-nanoliter samples

A DNP-NMR setup for sub-nanoliter samples A DNP-NMR setup for sub-nanoliter samples Pedro Freire da Silva, Supervisor(s): Prof. Giovanni Boero, Prof. Pedro Sebastião The electrical and mechanical design for a non-magnetic, 2.5x2.5 cm 2 setup with

More information

Fourier Transform. louder softer. louder. softer. amplitude. time. amplitude. time. frequency. frequency. P. J. Grandinetti

Fourier Transform. louder softer. louder. softer. amplitude. time. amplitude. time. frequency. frequency. P. J. Grandinetti Fourier Transform * * amplitude louder softer amplitude louder softer frequency frequency Fourier Transform amplitude What is the mathematical relationship between two signal domains frequency Fourier

More information

400 MHz spectrometer user manual

400 MHz spectrometer user manual 400 MHz spectrometer user manual january 2017 Sandrine Denis-Quanquin 1. THE NMR SPECTROMETER... 3 2. MANUAL MODE / AUTOMATION... 4 2.1 SAMPLE CHANGER... 4 2.2 MANUAL MODE... 4 2.3 AUTOMATION... 4 3. PRELIMINARY

More information

EE225E/BIOE265 Spring 2011 Principles of MRI. Assignment 6 Solutions. (y 0 + vt) dt. 2 y 0T + 3 )

EE225E/BIOE265 Spring 2011 Principles of MRI. Assignment 6 Solutions. (y 0 + vt) dt. 2 y 0T + 3 ) EE225E/BIOE265 Spring 211 Principles of MRI Miki Lustig Handout Assignment 6 Solutions 1. Nishimura 6.7 (Thanks Galen!) a) After the 9 y pulse, the spin is in the ˆx direction (using left-handed rotations).

More information

PHY3902 PHY3904. Nuclear magnetic resonance Laboratory Protocol

PHY3902 PHY3904. Nuclear magnetic resonance Laboratory Protocol PHY3902 PHY3904 Nuclear magnetic resonance Laboratory Protocol PHY3902 PHY3904 Nuclear magnetic resonance Laboratory Protocol GETTING STARTED You might be tempted now to put a sample in the probe and try

More information

Step by step procedure for NMR data acquisition

Step by step procedure for NMR data acquisition Step by step procedure for NMR data acquisition Spectrometers The UTHSCSA 500, 600, and 700 MHz spectrometers are each equipped with 4 independent RF channels and are each operated by a Red Hat Linux workstation

More information

Signs of Frequencies and Phases in NMR: The Role of Radiofrequency Mixing

Signs of Frequencies and Phases in NMR: The Role of Radiofrequency Mixing Journal of Magnetic Resonance 142, 190 194 (2000) Article ID jmre.1999.1929, available online at http://www.idealibrary.com on Signs of Frequencies and Phases in NMR: The Role of Radiofrequency Mixing

More information

Instruction for Operating the Bruker Avance III 800 MHz NMR Spectrometers in UTMB

Instruction for Operating the Bruker Avance III 800 MHz NMR Spectrometers in UTMB Instruction for Operating the Bruker Avance III 800 MHz NMR Spectrometers in UTMB Written by Tianzhi Wang, date: February 8, 2013. No food, no drink in NMR room and no internet in NMR host computer except

More information

Open acqi window if the button has been lost. autolocking routine, alock= y for autolocking, alock= n for typical manual locking

Open acqi window if the button has been lost. autolocking routine, alock= y for autolocking, alock= n for typical manual locking Glossary of Common NMR Commands and Terms aa acqi ai alock aph array at points (np) axis='p' axis= pd BPsvf bc bs cd directory abort acquisition, hard stop Open acqi window if the button has been lost

More information

Half-Pulse Excitation Pulse Design and the Artifact Evaluation

Half-Pulse Excitation Pulse Design and the Artifact Evaluation Half-Pulse Excitation Pulse Design and the Artifact Evaluation Phillip Cho. INRODUCION A conventional excitation scheme consists of a slice-selective RF excitation followed by a gradient-refocusing interval

More information

RF Pulse Toolkit: Application Specific Design

RF Pulse Toolkit: Application Specific Design RF Pulse Toolkit: Application Specific Design William A Grissom Department of Biomedical Engineering, Vanderbilt University, Nashville, TN, USA will.grissom@vanderbilt.edu Introduction RF excitation is

More information

System/Imaging Imperfections

System/Imaging Imperfections System/Imaging Imperfections B0 variations: Shim, Susceptibility B1 variations: Transmit, Receive Gradient Imperfections: Non-linearities Delays and Eddy currents Concomitant terms 1 B0 Variations - Off-Resonance

More information

IR/SR TrueFISP. Works-in-Progress package Version 1.2. For the SIEMENS Magnetom. Installation and User s Guide NUMARIS/4VA21B.

IR/SR TrueFISP. Works-in-Progress package Version 1.2. For the SIEMENS Magnetom. Installation and User s Guide NUMARIS/4VA21B. Works-in-Progress package Version 1.2 For the Installation and User s Guide NUMARIS/4VA21B January 22, 2003 Section of Medical Physics, University Hospital Freiburg, Germany Contact: Klaus Scheffler PhD

More information

Chapter 1. 1 The NMR Spectrometer. 1.1 Components of an NMR Spectrometer The Magnet

Chapter 1. 1 The NMR Spectrometer. 1.1 Components of an NMR Spectrometer The Magnet Chapter 1 1 The NMR Spectrometer 1.1 Components of an NMR Spectrometer 1.1.1 The Magnet In most current NMR spectrometers the magnetic field is generated by a superconducting magnet (Fig. 1.1). The first

More information

Supplementary Figure 1. Scanning Electron Microscopy images of the pristine electrodes. (a) negative electrode and (b) positive electrode.

Supplementary Figure 1. Scanning Electron Microscopy images of the pristine electrodes. (a) negative electrode and (b) positive electrode. a b Supplementary Figure 1. Scanning Electron Microscopy images of the pristine electrodes. (a) negative electrode and (b) positive electrode. Images were performed using a FEI/Philips XL4 microscope with

More information

MAGNETIC RESONANCE IMAGING

MAGNETIC RESONANCE IMAGING CSEE 4620 Homework 3 Fall 2018 MAGNETIC RESONANCE IMAGING 1. THE PRIMARY MAGNET Magnetic resonance imaging requires a very strong static magnetic field to align the nuclei. Modern MRI scanners require

More information

Development of a new Q-meter module

Development of a new Q-meter module A. Berlin,, W. Meyer, G. Reicherz Experimentalphysik I, Ruhr-Universität Bochum E-mail: jonas.herick@rub.de In the research field of polarized target physics the Q-meter is a well established technique

More information

MR Advance Techniques. Flow Phenomena. Class II

MR Advance Techniques. Flow Phenomena. Class II MR Advance Techniques Flow Phenomena Class II Flow Phenomena In this class we will explore different phenomenona produced from nuclei that move during the acquisition of data. Flowing nuclei exhibit different

More information

6.S02 MRI Lab Acquire MR signals. 2.1 Free Induction decay (FID)

6.S02 MRI Lab Acquire MR signals. 2.1 Free Induction decay (FID) 6.S02 MRI Lab 1 2. Acquire MR signals Connecting to the scanner Connect to VMware on the Lab Macs. Download and extract the following zip file in the MRI Lab dropbox folder: https://www.dropbox.com/s/ga8ga4a0sxwe62e/mit_download.zip

More information

Efficacy of Wavelet Transform Techniques for. Denoising Polarized Target NMR Signals

Efficacy of Wavelet Transform Techniques for. Denoising Polarized Target NMR Signals Efficacy of Wavelet Transform Techniques for Denoising Polarized Target NMR Signals James Maxwell May 2, 24 Abstract Under the guidance of Dr. Donal Day, mathematical techniques known as Wavelet Transforms

More information

Nutation Spectra of Nuclear Quadrupole Resonance in Off-Resonance Conditions

Nutation Spectra of Nuclear Quadrupole Resonance in Off-Resonance Conditions Nutation Spectra of Nuclear Quadrupole Resonance in Off-Resonance Conditions Nicolay Sinyavsky and Mariusz Mackowiak a Baltic State Academy, Molodiozhnaya str. 6, 236029 Kaliningrad, Russia a Institute

More information

BOLD fmri: signal source, data acquisition, and interpretation

BOLD fmri: signal source, data acquisition, and interpretation BOLD fmri: signal source, data acquisition, and interpretation Cheryl Olman 4 th year student, Department of Neuroscience and Center for Magnetic Resonance Research Discussion series Week 1: Biological

More information

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging Medical Imaging X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging From: Physics for the IB Diploma Coursebook 6th Edition by Tsokos, Hoeben and Headlee And Higher Level Physics 2 nd Edition

More information

RLC-circuits with Cobra4 Xpert-Link TEP. 1 2 π L C. f res=

RLC-circuits with Cobra4 Xpert-Link TEP. 1 2 π L C. f res= Related topics Damped and forced oscillations, Kirchhoff s laws, series and parallel tuned circuit, resistance, capacitance, inductance, reactance, impedance, phase displacement, Q-factor, band-width Principle

More information

MRI: From Signal to Image

MRI: From Signal to Image MRI: From Signal to Image Johannes Koch physics654 2013-05-06 1 / 27 Tomography Magnetic Resonance Tomography Tomography: tomos: section graphein: to write Signal measured as function of space 2 / 27 Tomography

More information

The Pulsed Resistive Low-Field MR Scanner

The Pulsed Resistive Low-Field MR Scanner 39 Chapter 3 The Pulsed Resistive Low-Field MR Scanner 3.1 Background In the remaining part of this work we are going to describe hyperpolarized gas relaxation, diffusion and MR imaging experiments. These

More information

KJM D-SELECTIVE NMR Experiments on the AVI-600 and AVII-600. Version 1.0. Topspin 3.5 Windows 7 Topspin 1.3 Windows XP

KJM D-SELECTIVE NMR Experiments on the AVI-600 and AVII-600. Version 1.0. Topspin 3.5 Windows 7 Topspin 1.3 Windows XP KJM 9250 1D-SELECTIVE NMR Experiments on the AVI-600 and AVII-600 Version 1.0 Topspin 3.5 Windows 7 Topspin 1.3 Windows XP Professor Emeritus Alistair Lawrence Wilkins, University of Waikato, New Zealand.

More information

Program. short intro the hardware locking and shimming 1D proton setup presaturation spectra

Program. short intro the hardware locking and shimming 1D proton setup presaturation spectra Program short intro the hardware locking and shimming 1D proton setup presaturation spectra 13 C spectra, DEPT processing of 1D file transfer and backup 2D homonuclear + 2D processing 2D heteronuclear

More information

Qualion NMR Process\Lab NMR (Nuclei Magnetic Resonance) Analyzer Model MASH, Style D

Qualion NMR Process\Lab NMR (Nuclei Magnetic Resonance) Analyzer Model MASH, Style D Qualion NMR Process\Lab NMR (Nuclei Magnetic Resonance) Analyzer Model MASH, Style D User Guide UM SW50407.05 Table of Contents Table of Contents Table of Contents...1 Preface...6 1. SYSTEM OVERVIEW...

More information

Nuclear Magnetic Resonance (NMR)

Nuclear Magnetic Resonance (NMR) California Institute of Technology Physics 77 Nuclear Magnetic Resonance (NMR) Eric D. Black September 27, 2005 1 Theory Read Section 14.4 of Shankar, Spin Dynamics, including the optional digression on

More information

Evaluation of Adiabatic Frequency-Modulated Schemes for Broadband Decoupling in Isotropic Liquids

Evaluation of Adiabatic Frequency-Modulated Schemes for Broadband Decoupling in Isotropic Liquids JOURNAL OF MAGNETIC RESONANCE, Series A 119, 129 133 (1996) ARTICLE NO. 0062 Evaluation of Adiabatic Frequency-Modulated Schemes for Broadband Decoupling in Isotropic Liquids RIQIANG FU AND GEOFFREY BODENHAUSEN*

More information

EE225E/BIOE265 Spring 2012 Principles of MRI. Assignment 7. Due March 16, 2012

EE225E/BIOE265 Spring 2012 Principles of MRI. Assignment 7. Due March 16, 2012 EE225E/BIOE265 Spring 2012 Principles of MRI Miki Lustig Assignment 7 Due March 16, 2012 1. From Midterm I 2010: You ve just programmed up your first 2DFT pulse sequence, and are trying it out on the scanner.

More information

EE225E/BIOE265 Spring 2014 Principles of MRI. Assignment 6. Due Friday March 7th, 2014, Self Grading Due Monday March 10th, 2014

EE225E/BIOE265 Spring 2014 Principles of MRI. Assignment 6. Due Friday March 7th, 2014, Self Grading Due Monday March 10th, 2014 EE225E/BIOE265 Spring 2014 Principles of MRI Miki Lustig 1. Read Nishimura Ch. 6 Assignment 6 Due Friday March 7th, 2014, Self Grading Due Monday March 10th, 2014 2. Nishimura assignment 6.5 3. Mimimum-Phase

More information

NUMERICAL DESIGN OF RESONATOR COILS FOR HIGH FIELD MAGNETIC RESONANCE IMAGING. A Thesis

NUMERICAL DESIGN OF RESONATOR COILS FOR HIGH FIELD MAGNETIC RESONANCE IMAGING. A Thesis NUMERICAL DESIGN OF RESONATOR COILS FOR HIGH FIELD MAGNETIC RESONANCE IMAGING A Thesis Presented in Partial Fulfillment of the Requirements for the Degree Bachelor of Science in the Graduate School of

More information

Arrayed Acquisition in VNMR and on the Gemini 1

Arrayed Acquisition in VNMR and on the Gemini 1 Arrayed Acquisition in VNMR and on the Gemini 1 I. Arrays in VNMR Vnmr allows you to quickly define experiments in which a series of spectra can be obtained as a function of any NMR parameter. For example,

More information

3D-Printed Microstrip Resonators for 4.7T MRI. Saeed Javidmehr. A thesis submitted in partial fulfillment of the requirements for the degree of

3D-Printed Microstrip Resonators for 4.7T MRI. Saeed Javidmehr. A thesis submitted in partial fulfillment of the requirements for the degree of 3D-Printed Microstrip Resonators for 4.7T MRI by Saeed Javidmehr A thesis submitted in partial fulfillment of the requirements for the degree of Master of Science in Electromagnetics and Microwaves Department

More information

RF pulse design and the Small Tip Angle Approximation

RF pulse design and the Small Tip Angle Approximation RF pulse design and the Small Tip Angle Approximation Dr Shaihan J Malik Lecturer in Imaging Sciences Division of Imaging Sciences & Biomedical Engineering King s College London shaihan.malik@kcl.ac.uk

More information

Nuclear Magnetic Resonance (NMR)

Nuclear Magnetic Resonance (NMR) California Institute of Technology Physics 77 Nuclear Magnetic Resonance (NMR) Eric D. Black October 3, 2008 1 Theory Read Section 14.4 of Shankar, Spin Dynamics, including the optional digression on negative

More information

Chapter 11 Coherence Editing: Pulse-field Gradients and Phase Cycling

Chapter 11 Coherence Editing: Pulse-field Gradients and Phase Cycling Chapter 11 Coherence Editing: Pulse-field Gradients and Phase Cycling Coherence editing is used to remove unwanted signals from NMR spectra. For example, in the double quantum filtered COSY experiment,

More information

SC5307A/SC5308A 100 khz to 6 GHz RF Downconverter. Datasheet SignalCore, Inc.

SC5307A/SC5308A 100 khz to 6 GHz RF Downconverter. Datasheet SignalCore, Inc. SC5307A/SC5308A 100 khz to 6 GHz RF Downconverter Datasheet 2017 SignalCore, Inc. support@signalcore.com P RODUCT S PECIFICATIONS Definition of Terms The following terms are used throughout this datasheet

More information

MRI Metal Artifact Reduction

MRI Metal Artifact Reduction MRI Metal Artifact Reduction PD Dr. med. Reto Sutter University Hospital Balgrist Zurich University of Zurich OUTLINE Is this Patient suitable for MR Imaging? Metal artifact reduction Is this Patient suitable

More information

NUCLEAR MAGNETIC RESONANCE 1

NUCLEAR MAGNETIC RESONANCE 1 NUCLEAR MAGNETIC RESONANCE 1 v3.0 Last Revision: R. A. Schumacher, May 2018 I. INTRODUCTION In 1946 nuclear magnetic resonance (NMR) in condensed matter was discovered simultaneously by Edward Purcell

More information

Problems from the 3 rd edition

Problems from the 3 rd edition (2.1-1) Find the energies of the signals: a) sin t, 0 t π b) sin t, 0 t π c) 2 sin t, 0 t π d) sin (t-2π), 2π t 4π Problems from the 3 rd edition Comment on the effect on energy of sign change, time shifting

More information

8 COSY. 8.1 Introduction. 8.2 Magnitude COSY

8 COSY. 8.1 Introduction. 8.2 Magnitude COSY 8 COSY 8.1 Introduction COSY (COrrelation SpectroscopY) is a homonuclear 2D technique that is used to correlate the chemical shifts of 1 H nuclei which are J-coupled to one another. In this chapter, two

More information

An EPR Primer 2. Basic EPR Theory 2.1. Introduction to Spectroscopy 2.1.1

An EPR Primer 2. Basic EPR Theory 2.1. Introduction to Spectroscopy 2.1.1 An EPR Primer 2 This chapter is an introduction to the basic theory and practice of EPR spectroscopy. It gives you sufficient background to understand the following chapters. In addition, we strongly encourage

More information

KJM D-SELECTIVE NMR Experiments on the AVIIIHD-800. Version 1.0. Topspin 3.5 Windows 7

KJM D-SELECTIVE NMR Experiments on the AVIIIHD-800. Version 1.0. Topspin 3.5 Windows 7 KJM 9250 1D-SELECTIVE NMR Experiments on the AVIIIHD-800 Version 1.0 Topspin 3.5 Windows 7 Professor Emeritus Alistair Lawrence Wilkins, University of Waikato, New Zealand. January 2018 1D-SELECTIVE NMR

More information

Magnet technology Insights & recent UHF results

Magnet technology Insights & recent UHF results Magnet technology Insights & recent UHF results Daniel Baumann, Rainer Kümmerle Bruker Biospin AG Switzerland 25. November 2016 Brussels Innovation with Integrity Innovation with Integrity 1 GHz Aeon at

More information

Multimode Optical Fiber

Multimode Optical Fiber Multimode Optical Fiber 1 OBJECTIVE Determine the optical modes that exist for multimode step index fibers and investigate their performance on optical systems. 2 PRE-LAB The backbone of optical systems

More information

Localization of microscale devices in vivo using addressable transmitters operated as magnetic spins

Localization of microscale devices in vivo using addressable transmitters operated as magnetic spins SUPPLEMENTARY INFORMATION Articles DOI: 10.1038/s41551-017-0129-2 In the format provided by the authors and unedited. Localization of microscale devices in vivo using addressable transmitters operated

More information

Rotating Coil Measurement Errors*

Rotating Coil Measurement Errors* Rotating Coil Measurement Errors* Animesh Jain Superconducting Magnet Division Brookhaven National Laboratory, Upton, NY 11973, USA 2 nd Workshop on Beam Dynamics Meets Magnets (BeMa2014) December 1-4,

More information

Optical Pumping Control Unit

Optical Pumping Control Unit (Advanced) Experimental Physics V85.0112/G85.2075 Optical Pumping Control Unit Fall, 2012 10/16/2012 Introduction This document is gives an overview of the optical pumping control unit. Magnetic Fields

More information

Image Quality/Artifacts Frequency (MHz)

Image Quality/Artifacts Frequency (MHz) The Larmor Relation 84 Image Quality/Artifacts (MHz) 42 ω = γ X B = 2πf 84 0.0 1.0 2.0 Magnetic Field (Tesla) 1 A 1D Image Magnetic Field Gradients Magnet Field Strength Field Strength / Gradient Coil

More information

RLC-circuits with Cobra4 Xpert-Link

RLC-circuits with Cobra4 Xpert-Link Student's Sheet RLC-circuits with Cobra4 Xpert-Link (Item No.: P2440664) Curricular Relevance Area of Expertise: Physics Subtopic: Inductance, Electromagnetic Oscillations, AC Circuits Topic: Electricity

More information

Understanding the Magnetic Resonance Spectrum of Nitrogen Vacancy Centers in an Ensemble of Randomly-Oriented Nanodiamonds, Supporting Information

Understanding the Magnetic Resonance Spectrum of Nitrogen Vacancy Centers in an Ensemble of Randomly-Oriented Nanodiamonds, Supporting Information Understanding the Magnetic Resonance Spectrum of Nitrogen Vacancy Centers in an Ensemble of Randomly-Oriented Nanodiamonds, Supporting Information Keunhong Jeong *1,2, Anna J. Parker *1,2, Ralph H. Page

More information

Eddy Current Testing (ET) Technique

Eddy Current Testing (ET) Technique Research Group Eddy Current Testing (ET) Technique Professor Pedro Vilaça * * Contacts: Address: Puumiehenkuja 3 (room 202), 02150 Espoo, Finland pedro.vilaca@aalto.fi October 2017 Contents Historical

More information

PINMRF. Varian 300 MHz NMR Spectrometers User Guide for Advanced 1D and Basic 2D NMR Experiments

PINMRF. Varian 300 MHz NMR Spectrometers User Guide for Advanced 1D and Basic 2D NMR Experiments PINMRF Varian 300 MHz NMR Spectrometers User Guide for Advanced 1D and Basic 2D NMR Experiments INCLUDING: Inova-300-1 w/ 5mm 4-nucleus probe 365 WTHR Inova-300-2 w/ 5mm 4-nucleus probe 4100 BRWN Table

More information