1 Large area position-sensitive CVD diamond detectors for X-ray beam monitoring with extreme position resolution M. Pomorski, P. Bergonzo, Ch. Mer, M. Rebisz-Pomorska D. Tromson, N. Tranchant Diamond Sensors Laboratory, CEA LIST, France S.Hustache, K. Desjardins Synchrotron Soleil, France J. Morse ESRF, France J. Gruenert, Ch.Bressler, W. Gawelda European XFEL, Germany
2 Talk Outline Motivation/Goal Why Diamond? Detector Fabrication and Read-out Measurements and Results in-beam duo-lateral sccvd-psd pulse-mode MHz regime four-corner PSD - lab tests Summary and Outlook 1/78
3 X-ray Synchrotron Beam Monitoring Motivation/Goal Development of novel solid-state PSDs for x-ray beam monitoring including: a hodoscope for EU XFEL commissioning (and normal operation?) beam tracking within ~700m (pulse-mode, 4mm 2, σ 7µm) EU XFEL (Jan Gruenert talk 1st CARAT workshop) 2/15
4 Why Diamond? Physical Properties transmission transparency 1 10 diamond 40 microns Si 40 microns Si 5 microns diamond 3 microns X-ray energy [kev] induced signal [a.u.] µm Diam ~ 5 µm Si fast response sccvd-psd, 3x3x pF, 1kΩ Si-PSD, 3x3x0.005 mm 166pF,1kΩ time [ns] equivalent (absorption) diamond and Si PSDs heat conductivity, radiation hardness, solar-blind, no dark-current, compact device 3/15
5 CVD Diamond Radiation (Particles) Detectors Basics diamond position sensitive detectors for X-ray beam monitoring pixel, strip large sensitive area beam position and profile quadrant fast only 4 channels resistive electrode large sensitive area only 4 channels sophisticated electronic pixel size limits position resolution small active area beam size dependence speed limited by RC 4/15
6 Detector Fabrication duo-lateral PSD four-corner PSD metal collecting electrodes metal collecting electrodes DLC resistive layer DLC resistive layer intr. diamond slab intr. diamond slab DLC resistive layer metal collecting electrode metal collecting electrodes diamond material: electronic grade sccvd DLC resistive layer: PVD, R from 1kOhm to few MOHms metal collecting electrodes: PVD, Al or CrAu, photolithogr. 5/15
7 Duo-lateral PSD: Read-out Electronics X-ray beam DAQ DAQ DAQ DAQ position coordinates from signal division (reconstucted position) σ pos ~ L*σ ele /I0 ( slope.. similar to timing) DC mode (10Hz) Keithley electrometers ; pulse mode FEMTO ampl. (200MHz) 6/15
8 Duo-lateral PSD: Experimental Environment masurement ID06 ESRF sccvd Si y x measurement setup@µxas SLS sccvd X-ray beam y x X-ray beam ESRF, Grenoble FR, ID06 Beam energy: 10.5 kev Beam size: 100 x 100 µm (slits limited) Beam flux: 1.6 x ph/sec Flux absorbed = 1.56 x 10 8 ph/sec/µm sccvd XBIC ~ 0.5 µa DC DC Soleil, Gif-sur-Yvette FR, Proxima1 & SIXS Beam energy: 12.6 kev & 5.6 kev Beam size: 300 x 300 µm (slits limited) Beam flux: 2.3 x ph/sec Flux absorbed = 4.6 x 10 8 ph/sec/µm sccvd XBIC ~ 10 µa DC DC Swiss Light Source, Villigen CH, µxas Beam energy: 7.05 kev Beam size: 50 µm FWHM Beam flux: 4.8 x ph/sec Flux absorbed = 10 3 ph/pulse sccvd XBIC ~ 19 µa DC pulse 7/15
9 XBIC (normalized to beam decay) [a.u.] In-beam Performance: Beam Intensity Monitoring (DC mode) Beam intensity ID06 ESRF (X1+X2 and Y1+Y2 ) sccvd-psd 40 µm bias voltage [V] Beam intensity Proxima 1 Soleil (X1+X2) sccvd-psd 300 µm current [A] beam intensity [normalised] /15 stable non-injecting 100% CCE with E avg ~13eV/e-h XTIC dark current bias voltage [V] X1+X2 - X plane Y1+Y2 - Y plane Si photodiode XBIC X1+X2 [µa] 21,340 21,335 21,330 21,325 21,320 21,315 21,310 21,305 time [min] refill beam intensity variation ~0.6 % measurement point 0,738 0,737 0,736 0,735 0,734 0,733 0,732 beam intensity [normalized] 0.5% 1,00 0,99 0,98 0,97 0,003 0,002 0,001 0,000
10 In-beam Performance: Reconstructed Position Pattern and Linearity 40 µm sccvd-psd cartography (ESRF ID06) 1,0 fine raster scan with 25 µm step (80x80 points) 450µm local linearity FWHM=1,15 µm FWHM=0,51 µm Calculated vertical distance from centre (mm) 0,5 0,0-0,5-1,0-1,0-0,5 0,0 0,5 1,0 Calculated horizontal distance from centre (mm) 30 (µm) Counts 2D beam tracking with beam stabilization 0-2,0-1,5-1,0-0,5 0,0 0,5 1,0 1,5 2,0 no beam-stabilization residuals [µm] deviation from linear fits in the probed region plane X σ L 0.1% plane Y σ L 0.25% step motors crash (during the second scan): 0 integral absorbed dose ~0.3 Giga Gy no radiation damage signs (µm) 9/15
11 In-beam Performance: Position resolution (DC mode) Counts X beam position [µm] 20 kωsq. sccvd-psd 40 µm, S/N ~10 4, active area 2 x 2 mm 2, ESRF ID time [min] residuals [µm] beam position Y [µm] σ y 258 nm Counts 1 MΩsq. sccvd-psd 300 µm, S/N ~10 5, active area 2.5 x 2.5 mm 2, Soleil Proxima1 15 min. beam position X [µm] ,50 0,25 0,00-0,25-0, measurement point residuals [µm] σ 27 nm 10/15
12 In-beam Performance: Long Term ESRF ID06 40 µm sccvd-psd permanently mounted at the ID06, vacuum 10-7 mbar beam drift measurement with sccvd-psd and IC-PSD after few months in the X-ray beam I0 X IC-PSD 30µm Y sccvd-psd Y IC+scCVD PSDs 11/15 no feed-back data here courtesy: Thomas Roth and Carsten Detlefs ID06, ESRF
13 In-beam Performance: Pulse-mode MHz Regime PRELIMINARY 1kΩsq. lateral 1D sccvd-psd 400µm, active area 1.4 x 2 mm 2 SLS to understand applicability range of diamond detectors for EU XFEL irradiation geometry L electrode collimator induced signal [V] beam time structure with sccvd-psd pulse train cam shaft 20 ns pulse train signal amplitude [a.u]] detector's active area ~1.4 mm refill right electrode left electrode I0 collimator cut 0 3,0 3,5 4,0 4,5 5,0 X motor displacement [mm] 12/15 beam scan R electrode reconstucted position [a.u.] σ L ~0.2% central area detector's active area ~1.4 mm center of the device time [ns] collimator cut σ~2µm X motor displacement [mm]
14 Four-corner PSD: preliminary lab tests irradiation throught a metal mesh -1,0 1,0-0,5 0,5 0,0 0,0 Y[L] Y [L] full area irradiation 0,5 1,0 1,0-0,5 0,5 0,0-0,5-1,0 X [L] -1,0-1,0-0,5 0,0 X[L] 0,5 1,0 0,3 241Am α-particles injection ~ 4MeV (50 fc) (measurement in air) 0,0 Pulse mode with, fast CSA (Mircea Ciobanu), 200ns integration time -0,3-0,3 13/15 0,0 0,3 Michal Pomorski, Diamond Sensors Laboratory, 2nd CARAT workshop@gsi, Darmstadt, 14/12/2010
15 Summary A novel type of PSD based on sccvd diamond and DLC resistive electrodes has been built and tested for XBPM at synchrotrons: semitransparent in-beam detector, stable and reliable operation absolute beam intensity monitoring with a precision of ~0.1% sub-micron position resolution (reaching 27 nm σ) very good (<<1%) linearity up to 4x4 mm 2 active area pulse-mode operation within 20ns integration time demonstrated (can be used also as a single particle PSD) 14/15
16 Future Work to build and test more prototypes long term stability tests detector performance evaluation in function of the bandwidth tests in extreme conditions: white beams, XFEL ultra-thin sccvd-psd for low energy (<3keV) lines tetra-lateral approach + resistive lateral lines boron doped CVD diamond layer (PIM structure) dedicated electronics missing Thank you for your attention! 15/15
Results from Diamond Detector tests at ELETTRA
Results from Diamond Detector tests at ELETTRA Wolfgang Freund, WP74 wolfgang.freund@xfel.eu European XFEL User s Meeting 2013 Satellite Workshop on Photon Beam Diagnostics, 24 Jan 2013 Acknowledgements
More informationCharacterization of SC CVD diamond detectors for heavy ions spectroscopy
Characterization of SC CVD diamond detectors for heavy ions spectroscopy Characterization of SC CVD diamond detectors for heavy and ions MIPsspectroscopy timing and MIPs timing Michal Pomorski and GSI
More informationESRF. Tests of a diamond quadrant detector at Hasylab (DESY) using the Libera Brilliance. European Synchrotron Radiation Facility, France
Tests of a diamond quadrant detector at Hasylab (DESY) using the Libera Brilliance J Morse European Synchrotron Radiation Facility, France H Graafsma Hasylab, DESY, Germany ESRF B Solar Instrumentation
More informationcividec DIAMOND DETECTORS & SYSTEMS FOR X-RAYS Instrumentation CIVIDEC Instrumentation GmbH Vienna The Netherlands +31 (0)
cividec Instrumentation DIAMOND DETECTORS & SYSTEMS FOR X-RAYS CIVIDEC Instrumentation GmbH Vienna +32 (0)3 309 32 09 info@gotopeo.com www.gotopeo.com CONTENTS Introduction... 3 Monitors Diamond XBPM System...
More informationX-Ray Beam Position Monitor Based on a Single Crystal Diamond Performing Bunch by Bunch Detection
X-Ray Beam Position Monitor Based on a Single Crystal Diamond Performing Bunch by Bunch Detection M. Di Fraia, M. Antonelli, A. Tallaire, J. Achard, S. Carrato, R. H. Menk, G. Cautero, D. Giuressi, W.
More informationIntroduction to X-ray Detectors for Synchrotron Radiation Applications
Introduction to X-ray Detectors for Synchrotron Radiation Applications Pablo Fajardo Instrumentation Services and Development Division ESRF, Grenoble EIROforum School on Instrumentation (ESI 2011) Outline
More informationSimulation of High Resistivity (CMOS) Pixels
Simulation of High Resistivity (CMOS) Pixels Stefan Lauxtermann, Kadri Vural Sensor Creations Inc. AIDA-2020 CMOS Simulation Workshop May 13 th 2016 OUTLINE 1. Definition of High Resistivity Pixel Also
More informationUV/EUV CONTINUOUS POSITION SENSOR
UV/EUV CONTINUOUS POSITION SENSOR ODD-SXUV-DLPSD FEATURES Submicron position resolution Stable response after exposure to UV/EUV 5 mm x 5 mm active area TO-8 windowless package RoHS ELECTRO-OPTICAL CHARACTERISTICS
More informationR3B Heavy Ion Tracking
R3B Heavy Ion Tracking Roman Gernhäuser, TU-München High Rate Diamond Detectors for Heavy Ion Tracking and TOF material investigations detector concept (a reminder) electronics development prototype production
More informationDiamond Sensors. John Smedley DOE BES Neutron & Photon Detector Workshop
Diamond Sensors John Smedley DOE BES Neutron & Photon Detector Workshop Diamond for Light Sources In May 2011, the APS held the 4 th Diamond for Modern Light Sources Monochromators Kineform lenses Windows
More informationFirst test experiments with FMB- Oxford direct drive DCM at the Sirius beamline of Synchrotron SOLEIL
First test experiments with FMB- Oxford direct drive DCM at the Sirius beamline of Synchrotron SOLEIL Ciatto G., Moreno T., Aubert N., Feret P., Fontaine P. Synchrotron SOLEIL, L'Orme des Merisiers, Saint-Aubin,
More informationx-ray Beam Size Monitor
x-ray Beam Size Monitor J. Alexander, N. Eggert, J. Flanagan, W. Hopkins, B. Kreis, M. McDonald, D. Peterson, N. Rider Goals: 2 products: tuning tool with rapid feedback of beam height during LET measurements
More informationpcvd diamond beam position monitors for PETRA III
pcvd diamond beam position monitors for PETRA III Eckhard Wörner Diamond Materials GmbH Tullastraße 72, 79108 Freiburg, Germany CARAT workshop 13-15.12.09 1/39 Outline Some news about Diamond Materials
More informationcividec PORTFOLIO Instrumentation CIVIDEC Instrumentation GmbH Vienna België / Belgique Nederland
cividec Instrumentation PORTFOLIO Nederland België / Belgique T +31 (0)24 648 86 88 T +32 (0)3 309 32 09 info@gotopeo.com www.gotopeo.com CIVIDEC Instrumentation GmbH Vienna CONTENTS Preface...3 A Monitors
More informationLarge Area Continuous Position Sensitive Diamond Detector: First tests
Large Area Continuous Position Sensitive Diamond Detector: First tests Mircea Ciobanu, ISS-Bucharest 1 st ADAMAS Collaboration Meeting 16 18 December 2012 GSI-Darmstadt OOutline Preparatory work for understanding
More informationVIBRATING WIRE SENSORS FOR BEAM INSTRUMENTATION Suren Arutunian
VIBRATING WIRE SENSORS FOR BEAM INSTRUMENTATION Suren Arutunian Yerevan Physics Institute Yerevan Physics Institute S.Arutunian, VIBRATING WIRE SENSORS FOR BEAM INSTRUMENTATION BIW 2008, Lake Tahoe, USA
More informationStrip Detectors. Principal: Silicon strip detector. Ingrid--MariaGregor,SemiconductorsasParticleDetectors. metallization (Al) p +--strips
Strip Detectors First detector devices using the lithographic capabilities of microelectronics First Silicon detectors -- > strip detectors Can be found in all high energy physics experiments of the last
More informationDiamond sensors as beam conditions monitors in CMS and LHC
Diamond sensors as beam conditions monitors in CMS and LHC Maria Hempel DESY Zeuthen & BTU Cottbus on behalf of the BRM-CMS and CMS-DESY groups GSI Darmstadt, 11th - 13th December 2011 Outline 1. Description
More informationCase Study: Custom CCD for X-ray Free Electron Laser Experiment
Introduction The first XFEL (X-ray Free Electron Laser) experiments are being constructed around the world. These facilities produce femto-second long bursts of the most intense coherent X-rays ever to
More informationOrbit Stability Challenges for Storage Rings. Glenn Decker Advanced Photon Source Beam Diagnostics March 8, 2012
Orbit Stability Challenges for Storage Rings Glenn Decker Advanced Photon Source Beam Diagnostics March 8, 2012 Outline Beam stability requirements RF beam position monitor technology NSLS II developments
More informationDevelopment of scalable laser technology for EUVL applications
Development of scalable laser technology for EUVL applications Tomáš Mocek, Ph.D. Chief Scientist & Project Leader HiLASE Centre CZ.1.05/2.1.00/01.0027 Lasers for real-world applications Laser induced
More informationRANDY W. ALKIRE, GEROLD ROSENBAUM AND GWYNDAF EVANS
S-94,316 PATENTS-US-A96698 BEAM POSITION MONITOR RANDY W. ALKIRE, GEROLD ROSENBAUM AND GWYNDAF EVANS CONTRACTUAL ORIGIN OF THE INVENTION The United States Government has rights in this invention pursuant
More informationImprovement in High-Frequency Properties of Beam Halo Monitor using Diamond Detectors for SPring-8 XFEL
32 nd International Free Electron Laser Conference FEL 2010 Improvement in High-Frequency Properties of Beam Halo Monitor using Diamond Detectors for SPring-8 XFEL August 26, 2010 Thursday, THOC4 1 Hideki
More informationCharacterisation of a novel super-polished bimorph mirror
Characterisation of a novel super-polished bimorph mirror Kawal Sawhney 1, Simon Alcock 1, Hongchang Wang 1, John Sutter 1 and Riccardo Signorato 2 1 Diamond Light Source Ltd. UK 2 BASC, D-51429 Bergisch
More informationX-ray Detectors: What are the Needs?
X-ray Detectors: What are the Needs? Sol M. Gruner Physics Dept. & Cornell High Energy Synchrotron Source (CHESS) Ithaca, NY 14853 smg26@cornell.edu 1 simplified view of the Evolution of Imaging Synchrotron
More informationPSD Characteristics. Position Sensing Detectors
PSD Characteristics Position Sensing Detectors Silicon photodetectors are commonly used for light power measurements in a wide range of applications such as bar-code readers, laser printers, medical imaging,
More informationX-Ray Detection Using SOI Monolithic Sensors at a Compact High-Brightness X-Ray Source Based on Inverse Compton Scattering
Abstract #: 1054 Conference: NSS (Oral) Accelerator Technologies and Beam Line Instrumentation X-Ray Detection Using SOI Monolithic Sensors at a Compact High-Brightness X-Ray Source Based on Inverse Compton
More informationCESRTA Low Emittance Tuning Instrumentation: x-ray Beam Size Monitor
CESRTA Low Emittance Tuning Instrumentation: x-ray Beam Size Monitor xbsm group: (those who sit in the tunnel) J. Alexander, N. Eggert, J. Flanagan, W. Hopkins, B. Kreis, M. McDonald, D. Peterson, N. Rider
More informationWorkshop IGLEX Andromède & ThomX 23 June 2016, LAL Orsay. The X-line of ThomX.
Workshop IGLEX Andromède & ThomX 23 June 2016, LAL Orsay The X-line of ThomX jerome.lacipiere@neel.cnrs.fr mjacquet@lal.in2p3.fr Brightness panorama of X-ray (10-100 kev) sources Synchrotron : not very
More informationThe TSIS Spectral Irradiance Monitor: Prism Optical Degradation Studies
The TSIS Spectral Irradiance Monitor: Prism Optical Degradation Studies Lo Erik Richard, Dave Harber, Joel Rutkowski, Matt Triplett, Kasandra O Malia Laboratory for Atmospheric and Space Physics (LASP)
More informationPolarimetry Concept Based on Heavy Crystal Hadron Calorimeter
Polarimetry Concept Based on Heavy Crystal Hadron Calorimeter for the JEDI Collaboration CALOR 216 May 17, 216 Irakli Keshelashvili Introduction JEDI Polarimetry Concept MC Simulations Laboratory and Beam
More informationPixel hybrid photon detectors
Pixel hybrid photon detectors for the LHCb-RICH system Ken Wyllie On behalf of the LHCb-RICH group CERN, Geneva, Switzerland 1 Outline of the talk Introduction The LHCb detector The RICH 2 counter Overall
More informationPin photodiode Quality Assurance Procedure
GENEVE, SUISSE GENEVA, SWITZERLAND ORGANISATION EUROPEENE POUR LA RECHERCHE NUCLEAIRE EUROPEAN ORGANIZATION FOR NUCLEAR RESEARCH Laboratoire Européen pour la Physique des Particules European Laboratory
More informationShenglan Xu. GM/CA CAT Argonne National Laboratory
MECHANICAL DESIGN OF NEW DUAL PINHOLE MINI- BEAM COLLIMATOR WITH MOTORIZED PITCH AND YAW ADJUSTER PROVIDES LOWER BACKGROUND FOR X-RAY CRYSTALLOGRAPHY AT GMCA@APS Shenglan Xu GM/CA CAT Argonne National
More informationThe SIRAD irradiation facility at the INFN - Legnaro National Laboratory
The SIRAD irradiation facility at the INFN - Legnaro National Laboratory I. Introduction 2 The INFN - Legnaro National Laboratory (LNL) SIRAD beamline http://www.lnl.infn.it 3 What is SIRAD? SIRAD is the
More informationSimulation and test of 3D silicon radiation detectors
Simulation and test of 3D silicon radiation detectors C.Fleta 1, D. Pennicard 1, R. Bates 1, C. Parkes 1, G. Pellegrini 2, M. Lozano 2, V. Wright 3, M. Boscardin 4, G.-F. Dalla Betta 4, C. Piemonte 4,
More informationPhoton Diagnostics. FLASH User Workshop 08.
Photon Diagnostics FLASH User Workshop 08 Kai.Tiedtke@desy.de Outline What kind of diagnostic tools do user need to make efficient use of FLASH? intensity (New GMD) beam position intensity profile on the
More informationOPERA RPC: installation and underground test results
VII Workshop on Resistive Plate Chambers and Related Detectors Korea University, Seoul October 10-12, 2005 The OPERA RPC system: installation and underground test results A. Longhin (INFN & Padova University)
More informationCHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES
CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES The current multiplication mechanism offered by dynodes makes photomultiplier tubes ideal for low-light-level measurement. As explained earlier, there
More informationNano-structured superconducting single-photon detector
Nano-structured superconducting single-photon detector G. Gol'tsman *a, A. Korneev a,v. Izbenko a, K. Smirnov a, P. Kouminov a, B. Voronov a, A. Verevkin b, J. Zhang b, A. Pearlman b, W. Slysz b, and R.
More informationAdvances in laboratory modeling of wave propagation
Advances in laboratory modeling of wave propagation Physical Acoustics Lab Department of Geosciences Boise State University October 19, 2010 Outline Ultrasonic laboratory modeling Bridge between full-size
More informationPoS(VERTEX2015)008. The LHCb VELO upgrade. Sophie Elizabeth Richards. University of Bristol
University of Bristol E-mail: sophie.richards@bristol.ac.uk The upgrade of the LHCb experiment is planned for beginning of 2019 unitl the end of 2020. It will transform the experiment to a trigger-less
More informationA 100MHz voltage to frequency converter
A 100MHz voltage to frequency converter R. Hino, J. M. Clement, P. Fajardo To cite this version: R. Hino, J. M. Clement, P. Fajardo. A 100MHz voltage to frequency converter. 11th International Conference
More informationXH Germanium Microstrip Detector for EDAS.
XH Germanium Microstrip Detector for EDAS. Janet Groves /Jon Headspith STFC Daresbury Laboratory STFC Technology Slide title Outline Brief History of EDXAS detectors at STFC Photodiode array (PDA) Prototype
More informationNano Beam Position Monitor
Introduction Transparent X-ray beam monitoring and imaging is a new enabling technology that will become the gold standard tool for beam characterisation at synchrotron radiation facilities. It allows
More informationThe High-Voltage Monolithic Active Pixel Sensor for the Mu3e Experiment
The High-Voltage Monolithic Active Pixel Sensor for the Mu3e Experiment Shruti Shrestha On Behalf of the Mu3e Collaboration International Conference on Technology and Instrumentation in Particle Physics
More informationGas scintillation Glass GEM detector for high-resolution X-ray imaging and CT
Gas scintillation Glass GEM detector for high-resolution X-ray imaging and CT Takeshi Fujiwara 1, Yuki Mitsuya 2, Hiroyuki Takahashi 2, and Hiroyuki Toyokawa 2 1 National Institute of Advanced Industrial
More information9 th International Workshop on Personal Computers and Particle Accelerator Controls, VECC, Kolkata, India
Master slave topology based, remotely operated, precision X-ray beam profiler and placement system for high pressure physics experiment at Indus-2 beam line 9 th International Workshop on Personal Computers
More informationLarge dynamic range 64 channels ASIC for CZT or CdTe detectors
Large dynamic range 64 channels ASIC for CZT or CdTe detectors F. Glasser, P. Villard, J.P. Rostaing, M. Accensi, N. Baffert, J.L. Girard. CEA-LETI, Grenoble, France This work was supported by CEA-DAM
More informationNIST EUVL Metrology Programs
NIST EUVL Metrology Programs S.Grantham, C. Tarrio, R.E. Vest, Y. Barad, S. Kulin, K. Liu and T.B. Lucatorto National Institute of Standards and Technology (NIST) Gaithersburg, MD USA L. Klebanoff and
More informationTHE Belle II [1] detector is currently under construction
The s-cvd Radiation Monitoring and Beam Abort System of the Belle-II Vertex Detector L. Bosisio, C. La Licata, L. Lanceri, L. Vitale arxiv:1711.06823v1 [physics.ins-det] 18 Nov 2017 Abstract The Belle-II
More informationSDD from device modeling to mass production - practical experience
SDD from device modeling to mass production - practical experience Outline Motivations ALICE at LHC ITS&SDD system From specs to detector HV divider & stability Injectors & speed variations NTD fluctuations
More informationMuon detection in security applications and monolithic active pixel sensors
Muon detection in security applications and monolithic active pixel sensors Tracking in particle physics Gaseous detectors Silicon strips Silicon pixels Monolithic active pixel sensors Cosmic Muon tomography
More informationSupplementary Information
Supplementary Information Supplementary Figure 1. Modal simulation and frequency response of a high- frequency (75- khz) MEMS. a, Modal frequency of the device was simulated using Coventorware and shows
More informationHigh collection efficiency MCPs for photon counting detectors
High collection efficiency MCPs for photon counting detectors D. A. Orlov, * T. Ruardij, S. Duarte Pinto, R. Glazenborg and E. Kernen PHOTONIS Netherlands BV, Dwazziewegen 2, 9301 ZR Roden, The Netherlands
More informationFast CMOS Transimpedance Amplifier and Comparator circuit for readout of silicon strip detectors at LHC experiments
Fast CMOS Transimpedance Amplifier and Comparator circuit for readout of silicon strip detectors at LHC experiments Jan Kaplon - CERN Wladek Dabrowski - FPN/UMM Cracow Pepe Bernabeu IFIC Valencia Carlos
More informationOn-line spectrometer for FEL radiation at
On-line spectrometer for FEL radiation at FERMI@ELETTRA Fabio Frassetto 1, Luca Poletto 1, Daniele Cocco 2, Marco Zangrando 3 1 CNR/INFM Laboratory for Ultraviolet and X-Ray Optical Research & Department
More informationOverview of performance and improvements to fixed exit double crystal monochromators at Diamond. Andrew Dent, Physical Science Coordinator, DLS
Overview of performance and improvements to fixed exit double crystal monochromators at Diamond Andrew Dent, Physical Science Coordinator, DLS Overview Diffraction limit Geometric magnification Source
More informationRecent Development and Study of Silicon Solid State Photomultiplier (MRS Avalanche Photodetector)
Recent Development and Study of Silicon Solid State Photomultiplier (MRS Avalanche Photodetector) Valeri Saveliev University of Obninsk, Russia Vienna Conference on Instrumentation Vienna, 20 February
More informationProduction of HPDs for the LHCb RICH Detectors
Production of HPDs for the LHCb RICH Detectors LHCb RICH Detectors Hybrid Photon Detector Production Photo Detector Test Facilities Test Results Conclusions IEEE Nuclear Science Symposium Wyndham, 24 th
More informationCTPPS Detector Performance
CTPPS Detector Performance Run 2016 Data summary SiStrips Performance Data Quality Radiation Damage Alignment Optics Validation Acceptance Diamond Performance Data Quality Data consistency checks Run 2017
More informationDevelopment of LYSO detector modules for a charge-particle EDM polarimeter
Mitglied der Helmholtz-Gemeinschaft Development of LYSO detector modules for a charge-particle EDM polarimeter on behalf of the JEDI collaboration Dito Shergelashvili, PhD student @ SMART EDM_Lab, TSU,
More informationRadiation hardness and precision timing study of Silicon Detectors for the CMS High Granularity Calorimeter (HGC)
Radiation hardness and precision timing study of Silicon Detectors for the CMS High Granularity Calorimeter (HGC) Esteban Currás1,2, Marcos Fernández2, Christian Gallrapp1, Marcello Mannelli1, Michael
More informationImplementation of A Nanosecond Time-resolved APD Detector System for NRS Experiment in HEPS-TF
Implementation of A Nanosecond Time-resolved APD Detector System for NRS Experiment in HEPS-TF LI Zhen-jie a ; MA Yi-chao c ; LI Qiu-ju a ; LIU Peng a ; CHANG Jin-fan b ; ZHOU Yang-fan a * a Beijing Synchrotron
More informationThe trigger system of the muon spectrometer of the ALICE experiment at the LHC
The trigger system of the muon spectrometer of the ALICE experiment at the LHC Francesco Bossù for the ALICE collaboration University and INFN of Turin Siena, 09 June 2010 Outline 1 Introduction 2 Muon
More informationMPI Halbleiterlabor. MPI Semiconductor Laboratory. MPI mf
MPI Halbleiterlabor MPI Semiconductor Laboratory MPI mf LCLS User Workshop, SLAC, Menlo Park, 18. 10. 2008 Lothar Strüder, MPI Halbleiterlabor and Universität Siegen 1 Prepared by 1. MPI-HLL (MPE and MPP)
More informationVErtex LOcator (VELO)
Commissioning the LHCb VErtex LOcator (VELO) Mark Tobin University of Liverpool On behalf of the LHCb VELO group 1 Overview Introduction LHCb experiment. The Vertex Locator (VELO). Description of System.
More informationInitial Results from a Cryogenic Proton Irradiation of a p-channel CCD
Centre for Electronic Imaging Initial Results from a Cryogenic Proton Irradiation of a p-channel CCD Jason Gow Daniel Wood, David Hall, Ben Dryer, Simeon Barber, Andrew Holland and Neil Murray Jason P.
More informationFull characterization tests of Micromegas with elongated pillars
University of Würzburg Full characterization tests of Micromegas with elongated pillars B. Alvarez1 Gonzalez, L. Barak1, J. Bortfeldt1, F. Dubinin3, G. Glonti1, F. Kuger1,2, P. Iengo1, E. Oliveri1, J.
More informationInvestigate the characteristics of PIN Photodiodes and understand the usage of the Lightwave Analyzer component.
PIN Photodiode 1 OBJECTIVE Investigate the characteristics of PIN Photodiodes and understand the usage of the Lightwave Analyzer component. 2 PRE-LAB In a similar way photons can be generated in a semiconductor,
More informationWhy p-type is better than n-type? or Electric field in heavily irradiated silicon detectors
Why p-type is better than n-type? or Electric field in heavily irradiated silicon detectors G.Kramberger, V. Cindro, I. Mandić, M. Mikuž, M. Milovanović, M. Zavrtanik Jožef Stefan Institute Ljubljana,
More informationThe CoSAXS Beamline at MAX IV: A Small Angle X-Ray Scattering Beamline to Study Structure and Dynamics
The CoSAXS Beamline at MAX IV: A Small Angle X-Ray Scattering Beamline to Study Structure and Dynamics SAS Sample Environment workshop, September 10-11 th, 2015- Lund tomas.plivelic@maxlab.lu.se Aims Take
More informationSensor production readiness
Sensor production readiness G. Bolla, Purdue University for the USCMS FPIX group PMG review 02/25/2005 2/23/2005 1 Outline Sensor requirements Geometry Radiation hardness Development Guard Rings P stops
More informationSmall-pad Resistive Micromegas for Operation at Very High Rates. M. Alviggi, M.T. Camerlingo, V. Canale, M. Della Pietra, C. Di Donato, C.
Small-pad Resistive Micromegas for Operation at Very High Rates CERN; E-mail: paolo.iengo@cern.ch M. Alviggi, M.T. Camerlingo, V. Canale, M. Della Pietra, C. Di Donato, C. Grieco University of Naples and
More informationUpgrade of the ultra-small-angle scattering (USAXS) beamline BW4
Upgrade of the ultra-small-angle scattering (USAXS) beamline BW4 S.V. Roth, R. Döhrmann, M. Dommach, I. Kröger, T. Schubert, R. Gehrke Definition of the upgrade The wiggler beamline BW4 is dedicated to
More informationLaBr 3 :Ce, the latest crystal for nuclear medicine
10th Topical Seminar on Innovative Particle and Radiation Detectors 1-5 October 2006 Siena, Italy LaBr 3 :Ce, the latest crystal for nuclear medicine Roberto Pani On behalf of SCINTIRAD Collaboration INFN
More informationLarge-Area CdTe Photon-Counting Pixel Detectors
Large-Area CdTe Photon-Counting Pixel Detectors Tilman Donath, Application Scientist 22.6.2015, DIR2015, Ghent DECTRIS Ltd. 5400 Baden Switzerland www.dectris.com Agenda 1. Introduction Hybrid Photon Counting
More informationRed, Green, Blue (RGB) SiPMs
Silicon photomultipliers (SiPMs) from First Sensor are innovative solid-state silicon detectors with single photon sensitivity. SiPMs are a valid alternative to photomultiplier tubes. The main benefits
More information200 MHz 350 MHz 750 MHz Linac2 RFQ2 202 MHz 0.5 MeV /m Weight : 1000 kg/m Ext. diameter : 45 cm
M. Vretenar, CERN for the HF-RFQ Working Group (V.A. Dimov, M. Garlasché, A. Grudiev, B. Koubek, A.M. Lombardi, S. Mathot, D. Mazur, E. Montesinos, M. Timmins, M. Vretenar) 1 1988-92 Linac2 RFQ2 202 MHz
More informationRadiation Test Report Paul Scherer Institute Proton Irradiation Facility
the Large Hadron Collider project CERN CH-2 Geneva 23 Switzerland CERN Div./Group RadWG EDMS Document No. xxxxx Radiation Test Report Paul Scherer Institute Proton Irradiation Facility Responsibility Tested
More informationAGIPD calibration status report
AGIPD calibration status report A. Allahgholi 2, R. Dinapoli 1, P. Goettlicher 2, M. Gronewald 4, H. Graafsma 2,5, D. Greiffenberg 1, B.H. Henrich 1, H. Hirsemann 2, S. Jack 2, R. Klanner 3, A. Klyuev
More information3D Optical Motion Analysis of Micro Systems. Heinrich Steger, Polytec GmbH, Waldbronn
3D Optical Motion Analysis of Micro Systems Heinrich Steger, Polytec GmbH, Waldbronn SEMICON Europe 2012 Outline Needs and Challenges of measuring Micro Structure and MEMS Tools and Applications for optical
More informationLCLS project update. John Arthur. LCLS Photon Systems Manager
LCLS project update LCLS Photon Systems Manager LCLS major construction nearly finished Technical systems turning on with good performance Experimental instruments Expectations for early operation First
More informationUsing Higher Order Modes in the Superconducting TESLA Cavities for Diagnostics at DESY
Using Higher Order Modes in the Superconducting TESLA Cavities for Diagnostics at FLASH @ DESY N. Baboi, DESY, Hamburg for the HOM team : S. Molloy 1, N. Baboi 2, N. Eddy 3, J. Frisch 1, L. Hendrickson
More informationThe on-line detectors of the beam delivery system for the Centro Nazionale di Adroterapia Oncologica(CNAO)
The on-line detectors of the beam delivery system for the Centro Nazionale di Adroterapia Oncologica(CNAO) A. Ansarinejad1,2, A. Attili1, F. Bourhaleb2,R. Cirio1,2,M. Donetti1,3, M. A. Garella1, S. Giordanengo1,
More informationGas Monitor Detector Electronics
Gas Monitor Detector Electronics from charged particles to digital value Fini Jastrow, TTF/VUV-FEL TTF/VUV-FEL meeting, meeting, 21. Mar 21. 06Mar 06 The Detector Ion Current Measurement Measured Data
More informationNear Ultraviolet (NUV) SiPMs
Silicon photomultipliers (SiPMs) from First Sensor are innovative solid-state silicon detectors with single photon sensitivity. SiPMs are a valid alternative to photomultiplier tubes. The main benefits
More informationINTEGRATED CIRCUITS DATA SHEET. TDA1596 IF amplifier/demodulator for FM radio receivers. Product specification File under Integrated Circuits, IC01
INTEGRATED CIRCUITS DATA SHEET File under Integrated Circuits, IC01 April 1991 GENERAL DESCRIPTION The provides IF amplification, symmetrical quadrature demodulation and level detection for quality home
More informationAn Overview of MAX IV Insertion Devices & Magnetic Measurement System. Hamed Tarawneh On behalf of Insertion Devices Team
An Overview of MAX IV Insertion Devices & Magnetic Measurement System Hamed Tarawneh On behalf of Insertion Devices Team MAX IV IDs & MagLab 1 Outlook: MAX IV Facility. ID Magnet Lab @ MAX IV. IDs @ 3
More informationRedefining Measurement ID101 OEM Visible Photon Counter
Redefining Measurement ID OEM Visible Photon Counter Miniature Photon Counter for OEM Applications Intended for large-volume OEM applications, the ID is the smallest, most reliable and most efficient single-photon
More informationPADI, a new ASIC for RPC's RPC's and other timing detectors Mircea Ciobanu
PADI, a new ASIC for RPC's RPC's and other timing detectors Mircea Ciobanu NoRDHia Meeting August 30-1 September 2006 GSI-Darmstadt Outline Introduction to timing measurements Simulations Integrated circuits(fee3)
More informationLaBr 3 :Ce scintillation gamma camera prototype for X and gamma ray imaging
8th International Workshop on Radiation Imaging Detectors Pisa 2-6 July 2006 LaBr 3 :Ce scintillation gamma camera prototype for X and gamma ray imaging Roberto Pani On behalf of SCINTIRAD Collaboration
More informationWater-Window Microscope Based on Nitrogen Plasma Capillary Discharge Source
2015 International Workshop on EUV and Soft X-Ray Sources Water-Window Microscope Based on Nitrogen Plasma Capillary Discharge Source T. Parkman 1, M. F. Nawaz 2, M. Nevrkla 2, M. Vrbova 1, A. Jancarek
More information190 nm (6.5 ev) Laser-ARPES. Laser ARPES using a Tunable ps UV Source
190 nm (6. ev) Laser-ARPES Laser ARPES using a Tunable ps UV Source 1 190 nm (6. ev) UV Source Laser ARPES/ARUPS with a High-Power UV Source Tunable deep-uv pulses from 190... 21 nm (6.....77 ev) are available
More informationMeasurement results of DIPIX pixel sensor developed in SOI technology
Measurement results of DIPIX pixel sensor developed in SOI technology Mohammed Imran Ahmed a,b, Yasuo Arai c, Marek Idzik a, Piotr Kapusta b, Toshinobu Miyoshi c, Micha l Turala b a AGH University of Science
More informationSub-ps (and sub-micrometer) developments at ELETTRA
Sub-ps (and sub-micrometer) developments at ELETTRA Mario Ferianis SINCROTRONE TRIESTE, Italy The ELETTRA laboratory ELETTRA is a 3 rd generation synchrotron light source in Trieste (I) since 1993 up to
More informationScanning X-ray microscopy with a single photon counting 2D detector
Scanning X-ray microscopy with a single photon counting 2D detector Karolina Stachnik Faculty of Physics and Applied Computer Science, AGH University of Science and Technology, Kraków XLVIII Zakopane School
More informationTPC Readout with GEMs & Pixels
TPC Readout with GEMs & Pixels + Linear Collider Tracking Directional Dark Matter Detection Directional Neutron Spectroscopy? Sven Vahsen Lawrence Berkeley Lab Cygnus 2009, Cambridge Massachusetts 2 Our
More informationCavity BPMs for the NLC
SLAC-PUB-9211 May 2002 Cavity BPMs for the NLC Ronald Johnson, Zenghai Li, Takashi Naito, Jeffrey Rifkin, Stephen Smith, and Vernon Smith Stanford Linear Accelerator Center, 2575 Sand Hill Road, Menlo
More informationReadout electronics for LGAD sensors
Readout electronics for LGAD sensors O. Alonso, N. Franch, J. Canals, F. Palacio, M. López, A. Vilà and A. Diéguez SIC, Departament d Enginyeries: Electrònica, Universitat de Barcelona, Spain M. Carulla,
More information