1 See discussions, stats, and author profiles for this publication at: Measurement of normalized spectral responsivity of digital imaging devices by using a LED-based tunable uniform source Article in Applied Optics February 2013 Impact Factor: 1.78 DOI: /AO Source: PubMed CITATIONS 5 READS 86 4 authors, including: Seung-Nam Park Korea Research Institute of Standards and 74 PUBLICATIONS 241 CITATIONS Dong-Hoon Lee Korea Research Institute of Standards and 71 PUBLICATIONS 255 CITATIONS SEE PROFILE SEE PROFILE All in-text references underlined in blue are linked to publications on ResearchGate, letting you access and read them immediately. Available from: Seung-Nam Park Retrieved on: 09 May 2016

2 Measurement of normalized spectral responsivity of digital imaging devices by using a LED-based tunable uniform source Khaled Mahmoud, 1,2 Seongchong Park, 2 Seung-Nam Park, 1,2 and Dong-Hoon Lee 1,2, * 1 Science of Measurement Department, University of Science and Technology (UST), Daejeon , South Korea 2 Division of Physical Metrology, Korea Research Institute of Standards and Science (KRISS), 267 Gajeong-Ro, Yuseong-Gu, Daejeon , South Korea *Corresponding author: dh.lee@kriss.re.kr Received 31 October 2012; revised 10 January 2013; accepted 11 January 2013; posted 14 January 2013 (Doc. ID ); published 15 February 2013 We present an instrumentation solution for measurement of normalized spectral responsivity of digital imaging sensors and cameras. The instrument consists of multiple light-emitting diodes (LEDs), a single-grating monochromator, and a small-size integrating sphere. Wavelength tuning is achieved by a proper selection of LED in accordance with the monochromator setting in a range from 380 to 900 nm. High spectral purity with a bandwidth of 5 nm is realized without using double gratings and order-sorting filters. Experimental characteristics and calibration of the instrument are described with the related error and uncertainty sources. The performance is demonstrated by measuring a monochrome charge-coupled device and a trichromatic complementary metal oxide semiconductor device. The measurement uncertainty is evaluated to be less than 1% (k 2) except several wavelengths with low LED power Optical Society of America OCIS codes: , Introduction Digital imaging devices based on charge-coupled device (CCD) or complementary metal oxide semiconductor (CMOS) are widely used in a variety of scientific applications [1]. In radiometry, imaging devices are of great importance for carrying out certain types of measurements that should otherwise be performed by means of scanning devices. They are also the essential elements for multispectral imaging, including color imaging, that becomes a powerful tool in many application fields like biomedical science, remote sensing, arts, and color science. For scientific applications of digital imaging devices, the technical specifications, such as spatial resolution, quantum efficiency, spectral responsivity, linearity, geometric fidelity, and response time are X/13/ $15.00/ Optical Society of America required to be carefully characterized. In the radiometric and colorimetric applications, measurement of spectral responsivity is of particular importance, which is defined as the ratio of output digital reading to input radiometric quantity as a function of wavelength [2]. Since the gain of digital sensors can be easily adjusted and manipulated, the normalized values of spectral responsivity provide the essential information for further characterization and calibration of the radiometric and colorimetric imaging devices [3]. Once the normalized spectral responsivity is measured, absolute calibration of the instrument, i.e., adjustment of the electronic gain and data acquisition algorithm, can be performed at a spectrally integrated calibration source suitable for the target application. A straightforward way to measure the spectral responsivity of an imaging device is to irradiate it uniformly so that each individual detector receives the same and known amount of optical radiation 20 February 2013 / Vol. 52, No. 6 / APPLIED OPTICS 1263

3 from a wavelength-tunable monochromatic source. A uniform source can be effectively realized by using an integrating sphere [4]. However, the choice of a wavelength-tunable monochromatic source requires a careful consideration. The conventional choice for this is the combination of a white-spectrum (incandescent or discharge) lamp and a monochromator. Such a monochromator-based tunable source can scan a wide wavelength range with adjustable interval and bandwidth, but the spectral power and spectral purity are limited by the performance of the used monochromator [2]. In order to achieve a high spectral purity, a double-grating configuration for a low stray level is required which, on the other hand, reduces the spectral output power passing through the monochromator. Recently, tunable laser-based integrating sphere sources are demonstrated at several national metrology institutes as the highaccuracy solution for spectral responsivity measurement of imaging devices [5 8]. However, the coverage of a large wavelength range requires a large and expensive laser facility, which is far from a practical instrumentation for industrial or field applications. In this paper, we present a new instrumentation scheme for measuring the normalized spectral responsivity of digital imaging devices, which consists of light-emitting diodes (LEDs), a single-grating monochromator, and an integrating sphere [9,10]. LEDs are compact, cost-effective, efficient, and commercially available at different wavelengths covering from UV through IR. Various tunable quasi-monochromatic sources based on LEDs for radiometric applications are already reported [11 13]. However, the relatively large spectral bandwidth ranging from 20 to 50 nm limits the use of LEDs as a spectral source. To overcome this shortcoming, we use a compact singlegrating monochromator as a tunable spectral filter between LEDs and an integrating sphere to develop a uniform source with a spectral bandwidth smaller than 5 nm. By selecting a proper LED in accordance to the wavelength setting of the monochromator, the spectral power could be optimized without reducing the spectral purity. We show measurement results for two digital imaging device samples, one monochrome (B/W) CCD and one trichromic CMOS, by using the LED-based tunable uniform source. The measurements are performed both in irradiance mode for bare sensors and in radiance mode for cameras. Measurement accuracy is verified by evaluating the measurement uncertainty and systematic error sources for the case of normalized spectral responsivity. The achieved performance shows a high potential of the instrument as a practical solution for industrial applications with a compact and cost-effective setup. 2. Instrumentation A. Experimental Setup Figure 1 schematically shows the experimental setup for measuring the normalized spectral response of Fig. 1. (Color online) Schematic diagram of the normalized spectral responsivity measurement setup (top view). digital imaging sensors and cameras. The LED-based tunable uniform source consists of LEDs mounted on a rotational disk, a compact single-grating monochromator, and a barium sulfate (BaSO 4 ) coated integrating sphere with an inner diameter of 50 mm. The LED disk operates the selected one from multiple mounted LEDs at a predefined position [12]. We used 25 LEDs of different wavelengths to cover the wavelength range from 380 to 900 nm. The radiation emitted from the selected LED is coupled into the monochromator (Spectral Products, model CM110, focal length 110 mm, f 3.9, slit width 0.6 mm) by using two planoconvex lenses with a diameter of 25.4 mm and a focal length of 20 mm. The spectrally filtered radiation from the monochromator is then coupled into the integrating sphere by using a focusing lens through a sphere port with a diameter of 12.7 mm. A focusing lens is used to form an image of the monochromator output slit with a size of 3 mm 10 mm on the sphere inner wall opposite to the input port so that the light is distributed inside the sphere only after the first diffuse reflection at the sphere wall. No baffle is used inside the sphere as the direct irradiation of the ports by the incident light is effectively avoided though the beam spot control. The choice of the sphere size was the result of a compromise between the irradiance level and the spatial uniformity of the irradiance distribution inside the sphere. The irradiance level inside the integrating sphere is monitored by a Si photodiode attached at another sphere port with a diameter of 12.7 mm. As we calibrate the signal of this photodiode with respect to spectral irradiance at the position of the device under test (DUT), we refer it as the reference detector (REF). The third port of the integrating sphere has a diameter of 25.4 mm and is used as the output port of the tunable uniform source. All three ports are placed so that the normal axes of all the ports are mutually perpendicular (see Fig. 1). Two different types of DUT can be measured with the LED-based tunable uniform source. A bare imaging sensor, which needs to be calibrated with respect to spectral irradiance at a specific position, 1264 APPLIED OPTICS / Vol. 52, No. 6 / 20 February 2013

4 is placed in front of the output port at a distance of approximately 10 mm (irradiance mode measurement). Alternatively, an imaging camera including a lens system for calibration with respect to spectral radiance is placed at a distance corresponding to the focal length of the lens system (radiance mode measurement). For the irradiance mode measurement, the DUT sensor is mounted on a computer-controlled three-axis translation stage for a precise and reproducible positioning. The complete setup is mounted on a ground plate with a size of less than 1 m 1 m and controlled with computer software for an automated measurement. B. Source Characteristics and Measurement Procedure The basic characteristics of the LED-based tunable uniform source, which can influence the accuracy of the normalized spectral responsivity measurement, are experimentally tested. The measurement procedure is then optimized to avoid or minimize the related error or uncertainty sources. 1. Output Radiant Power Figure 2(a) shows spectral power distribution of 25 LEDs used for the tunable uniform source, which is measured at the entrance slit of the monochromator Fig. 2. (Color online) (a) Spectral power distribution of the individual LEDs used for the LED-based tunable uniform source. Different colors of the curves correspond to different 25 LEDs, which had been lit one by one during the wavelength scan. (b) Plot of the relative irradiance available at the output port of the integrating sphere source as a function of wavelength measured with an interval of 5 nm (dot symbols). The error bars correspond to repeatability uncertainties of each measurement points. by using a spectroradiometer (Instrument System, model CAS140) with a resolution bandwidth of approximately 3 nm. Depending on materials and constructions of LEDs, each LED provides different power level and different spectral bandwidth. Note that we used only the LEDs, which are commercially available (supplied by Roithner Laser Technik, Austria; super bright lamp type, 5 mm diameter round package with clear dome lens, 8 viewing angle). Selection of LED is correlated with the wavelength setting of the monochromator: as the measurement wavelength is set by an operator, the computer software automatically controls the monochromator, the rotational disk position, and the LED driving circuit to turn on one LED that can provide the maximum radiant power at that wavelength. The output radiant power from the monochromator, therefore, depends not only on the input LED power but also on the coupling and transmission through the grating monochromator at each wavelength. Figure 2(b) shows the relative change of irradiance as the wavelength scans from 380 to 900 nm with an interval of 5 nm. The irradiance at each wavelength is measured by using a calibrated photodiode with an active area of 1 cm 2 at a distance of 10 mm in front of the sphere output port. The uncertainties shown in Fig. 2(b) as error bars are calculated from repeated scans of the normalized irradiance measurement. We can confirm that the repeatability uncertainty is large at the wavelengths with low irradiance values, which will propagate to the uncertainty of spectral responsivity measurement. The absolute values of irradiance are not accurately known because we could not accurately determine the size of the active area as well as the uniformity of its responsivity. Nevertheless, based on the specification of the photodiode, we could estimate that all the values of the irradiance in Fig. 2(b) lie within a range from 10 to 100 nw cm 2. Considering the geometric condition, this range of irradiance corresponds to a range of radiance from 5 to 50 nw cm 2 sr. We note that we have intentionally reduced the coupling efficiency of LED at several set wavelengths, at which the intrinsic output LED power is extremely high compared to other LEDs, in order to minimize error due to nonlinearity of signal readings. 2. Spectral Purity The spectral characteristics of output radiant power at each wavelength are checked by using the scanning-type spectroradiometer (Instrument System, model Spectro 320) with a resolution bandwidth of 0.25 nm. The spectral bandwidth is measured to be less than 5 nm as a full width at half-maximum for the whole tuning range from 380 to 900 nm. The measured spectral bandwidth was in accordance with the specification of the used single-grating monochromator with a f -number of 3.9 and a slit width of 0.6 mm. In addition, the spectral stray light is measured by scanning the spectroradiometer with a large 20 February 2013 / Vol. 52, No. 6 / APPLIED OPTICS 1265

5 resolution bandwidth of 5 nm in the range far from the set wavelength of the monochromator. Up to the dynamic range limit of 10 4, no out-of-band stray light could be identified. This high spectral purity with low stray light is one of the important advantage of the colored LED-monochromator system in comparison with the conventional white lampmochromator system [2]. Since the LEDs are intrinsically quasi-monochromatic with emission bandwidth of less than 50 nm, no output of higherorder diffraction is expected. Therefore, no ordersorting filter is required, which is indispensable for the lamp system to cover a wide wavelength tuning range. Moreover, the ratio between the in-band power and the out-of-band power for the LED system is much less than the case for the lamp system, which results in a reduced stray light level. As a consequence, the LED-monochromator system could be realized based on a simple high-throughput, single-grating monochromator without concern about the spectral purity. 3. Temporal Power Stability The measurement of normalized spectral responsivity is based on a comparison between the REF signal and the DUT signal at the same radiation provided by the tunable uniform source. Temporal instability of radiant power during the both signal readings, therefore, causes an error in their ratio measurement. In order to achieve a stable radiant power, each LED selected for a specific wavelength is operated with a constant forward current by using a precision source meter. However, we noticed that each LED shows a characteristic power drift within a duration of several minutes after turning on, which is probably due to a thermalization in the LED [9,10]. Since the emission characteristics of LEDs is sensitive to temperature at the p n junction, such a power drift is not to be avoided for the setup where the junction temperature of the LEDs is not actively stabilized [14]. In order to minimize the error due to the temporal power instability, we built a drift check algorithm in the measurement software. At each set value of wavelength, the readings of the REF and DUT signals is repeated twice in the sequence REF t 1 DUT t 2 REF t 3, where REF t and DUT t denote the recordings of the REF and DUT signals at a time of t, respectively. The recording of the REF corresponds to a reading of the averaged REF photocurrent by a calibrated ammeter, while the recording of the DUT corresponds to capturing of its digital image data as a single shot. The software compares then the relative difference of the REF readings at t 1 and t 3, which corresponds to the relative power drift during the signal recordings for the ratio measurement. The software repeats the measurement sequence at this wavelength until the relative power drift between REF t 1 and REF t 3 is less than a predefined criterion of 0.1%. This has a consequence that the measurement software waits until the LED lit for that wavelength reaches a sufficient power stability. The measurement error due to a power drift is limited to be smaller than 0.1% for the whole wavelength range. We note that the short-term power instability is additionally considered as a repeatability uncertainty of the readings of the averaged REF photocurrent both at t 1 and t 3 with a number of sampling of more than 10 for each reading. 4. Spatial Uniformity For an integrating sphere with a finite port size, the spectral irradiance at the output port is not perfectly uniform, but shows a distribution which varies smoothly across the port [4]. If two detectors are to be compared at such a nonuniform irradiance field, the difference in detector size or position causes a systematic error [12]. Therefore, characterization of spatial uniformity of the irradiance distribution at the position of DUT provides useful information for error and uncertainty analysis. Figure 3(a) shows the spatial irradiance distribution of the LED-based tunable uniform source, which is measured at a distance of 10 mm from the output port through a horizontal scan at different wavelengths from 400 to 800 nm. For this measurement, a monochrome CCD sensor is used as a DUT mounted on the Fig. 3. (Color online) (a) Spatial irradiance distribution of the LED-based tunable uniform source measured at a distance of 10 mm from the output port through a horizontal scan in 1 mm steps at different wavelengths (dot symbols). The connecting curves are only for better visibility. (b) Plot of the relative difference of the spatial irradiance distribution with respect to the average values of the different wavelengths within a range of 15 mm at the center of the output port APPLIED OPTICS / Vol. 52, No. 6 / 20 February 2013

6 computer-controlled translation stage. While the CCD sensor is moved on the horizontal axis crossing the output port through the center (x-axis in Fig. 1), the digital images are captured at different positions. The raw data at each position was the averaged value of the pixels in a fixed area of interest (AOI) with a size of pixels corresponding to approximately 0.1 mm 0.1 mm. The measured raw data are then normalized to the value at the center position of the output port. Such a measurement of the normalized irradiance distribution is repeated at different wavelengths. We verify in Fig. 3(a) that the spatial uniformity of irradiance is limited only in a small area at the center of the output port. Such a nonuniform distribution can cause a significant error when the absolute irradiance responsivity should be measured by comparing two detectors with different sizes [12]. However, concerning only measurement of the normalized values of the spectral irradiance responsivity, the nonuniformity of irradiance distribution itself does not affect the accuracy, as long as the positioning of the DUT is reproducible. Instead, the major concern is the spectral dependence of the spatial irradiance distribution. From Fig. 3(a), we confirm that the spatial irradiance distribution measured at different wavelengths show a good overlap. In order to verify it more precisely, we plotted in Fig. 3(b) the relative difference of the spatial irradiance distribution with respect to the average values of the different wavelengths, which are calculated from the data of Fig. 3(a) within a range of 15 mm at the center of the output port. We confirm that the wavelength dependence of irradiance distribution is smaller than 0.5% in the range of 5 mm near the center, which is comparable to the uncertainty of the signal readings (see Table 2 and Fig. 5). Outside the central range, the random scattering of the data increases, as expected, from the decreased level of the DUT signal. No significant wavelength dependence of spatial irradiance distribution could be identified within the limit of the measurement uncertainty. From the spatial uniformity measurements in Fig. 3, we conclude that the error due to spatial nonuniformity of irradiance can be minimized for the normalized spectral responsivity measurement by positioning a calibration detector or a test detector with high precision and reproducibility. In our setup of Fig. 1, we used the translation stage which can control the absolute position within 0.01 mm (Sigma Koki, SGSP model series). 5. Stray Light The REF detector attached to the integrating sphere monitors the radiant power level delivered from the LED-based tunable source into the sphere, and will be calibrated against the spectral irradiance at a fixed point in front of the output port (see Fig. 1). Any ambient stray light which enters through the output port will change the REF reading and cause a measurement error. In general, the ambient stray can be easily corrected by a dark signal reading with a shutter inside the monochromator. We performed the dark signal readings of both REF and DUT prior to the main measurement procedure and subtracted them from the final signal readings. The complete setup is placed inside a light-tight box to keep the ambient stray as low as possible. The specific problem related to our instrumentation is the stray due to reflection from the DUT when it is placed close to the output port for the irradiance mode measurement. In this condition, the light from the output port is reflected from the detector surface and chassis of the test sensor and comes back into the integrating sphere as an additional stray. Since this stray exists only in the presence of the radiant power input, it cannot be corrected by the dark readings. We have experimentally tested the amount of the reflection stray by using different chassis materials at the position of DUT. We compared a chassis of bare black-anodized Aluminum and a chassis coated with a low-reflectance paint (3M/Mankiewicz, NEXTEL Velvet ) at a distance of below 20 mm from the output port. As a result, we confirmed a significant change of the REF readings depending on the chassis materials. Especially for wavelengths above 700 nm, where the reflectance of black-anodized aluminum increases, the difference between two chassis materials were as large as 10% in the REF signal. In order to eliminate such an error due to the stray by reflection, we designed the measurement procedure so that the DUT in the irradiance mode is moved to a predefined zero position outside the output port, whenever the REF reading is performed. For the DUT reading, the DUT is moved back to the test position in front of the output port. This procedure is easily realized with the computer-controlled translation stage mounting the DUT in the irradiance measurement mode. We note that such a DUT movement is not required for the radiance mode measurement with a DUT at a large distance (>100 mm) from the output port, because the change of the REF reading due to reflection from the DUT was negligible. C. Instrument Calibration Calibration of the instrument is performed in two steps. First, the wavelength scale of the monochromator is calibrated. Two pencil-type calibration lamps of Hg(Ar) and Ar are placed at the input slit of the monochromator and the transmitted light is recorded at the output slit by using a photodiode while scanning the wavelength with a small step of 0.2 nm. The instrument values of the peak wavelengths of selected isolated emission lines are compared with the literature values. The maximum difference of the instrument wavelength from the literature value was 0.73 nm in a range from 380 to 900 nm. From this measurement, we declared that the wavelength scale of the tunable source is calibrated with a standard uncertainty of wavelength u λ 0.42 nm, which is estimated from 0.73 nm p 3 by assuming a rectangular probability distribution [15]. 20 February 2013 / Vol. 52, No. 6 / APPLIED OPTICS 1267

7 Second, the responsivity of the REF photocurrent reading (in A) is calibrated against the irradiance (in W m 2 ) at the position of the DUT close to the output port as a function of wavelength. However, as our aim is to measure the normalized spectral responsivity of the DUT, the absolute values of the REF responsivity (in AW 1 m 2 ) is not of interest. Instead, we use only the normalized values of the REF responsivity. For calibration of the REF responsivity, a singleelement Si photodiode (Hamamatsu S BQ, active area 10 mm 10 mm) with the known spectral responsivity is mounted as a standard detector (STD) at the place of DUT in the irradiance mode. The spectral power responsivity of the STD, which we denote as S STD λ, is calibrated against the KRISS spectral responsivity scale from 300 to 1000 nm with relative standard uncertainties from 0.2% to 0.4% (k 1) depending on wavelength [16]. The calibration equation for the REF responsivity S REF λ is then given by S REF λ y REF y STD S STD λ ; (1) where y REF and y STD denote the readings of the REF and of the STD, respectively. Note that the reading of detectors used in this paper always corresponds to the dark-subtracted signal. For the measurement of the normalized spectral responsivity of an imaging sensor or camera, we use only the relative values of the REF spectral responsivity S REF λ which are normalized, e.g., to its maximum value: S REF λ S REF λ S max : (2) REF Figure 4 shows the experimental calibration result of the instrument which is a plot of the normalized REF responsivity S REF λ as a function of wavelength measured from 380 to 900 nm with a step of 5 nm. The calibration measurements are repeated and averaged at different distances of the STD to the Fig. 4. (Color online) Spectral responsivity of the REF as measured against the standard photodiode mounted at a close distance to the output port. The error bars represent the associated uncertainties of this calibration. output port from 10 to 25 mm with a step of 5 mm. The distance dependence of the normalized REF responsivity is included as a random uncertainty component into the uncertainty budget. The error bars in Fig. 4 indicate the combined expanded uncertainty with a coverage factor k 2 for a confidence level of approximately 95% [15]. The uncertainty depends on wavelength mainly because the radiant power of the LED-based tunable source strongly varies with wavelength as shown in Fig. 2(b). In addition, the uncertainty of the STD responsivity is also wavelength dependent. The uncertainty budget for the REF spectral responsivity S REF λ according to Eq. (1) is shown in Table 1, as an example, at 435 nm, where the combined uncertainty was the lowest. Expressed as relative uncertainties, the budget is valid also for the normalized values in Eq. (2). The uncertainty component of type A and B correspond to the component evaluated with a statistical method and with a nonstatistical method, respectively [15]. For the type A components, the number of repetition is also indicated in the parentheses. All the uncertainty components in Table 1 can be evaluated directly from the data of each quantity in Eq. (1) except the uncertainty of the signal ratio with respect to the wavelength calibration uncertainty. The propagation of the wavelength uncertainty u λ to the uncertainty of a ratio measurement is described in [17]. From Eq. (1), we derive the relation between the relative uncertainty of the ratio y REF y STD and the absolute uncertainty of wavelength u λ as yref 1 u rel y REF 1 y STD u λ : (3) y STD y REF λ y STD λ Note that the uncertainty of a ratio of two quantities due to an inaccurate setting of wavelength depends on the difference of the spectral slopes of both the quantities. D. Data Acquisition of Imaging Device Once the instrument is calibrated, the standard photodiode is removed and an imaging device as a DUT is mounted for measurement. If the DUT is a bare digital image sensor without a lens system, it is measured in the irradiance mode at a distance of approximately 10 mm to the output port. If the DUT is a digital camera with an imaging optics, the measurement is performed in the radiance mode at a distance of the focal length of the imaging lens system (see Fig. 1). The imaging device is connected with the computer via an USB interface for control and data communication. In the measurement procedure of normalized spectral responsivity of the DUT, one DUT reading at each wavelength corresponds to a capture of its digital image data as a single shot. All the DUT readings during the wavelength scan of the whole measurement range should be performed at a fixed gain 1268 APPLIED OPTICS / Vol. 52, No. 6 / 20 February 2013

8 Table 1. Uncertainty Budget of the REF Responsivity Calibration at 435 nm Related Quantity in Eq. (1) Uncertainty Component Description Evaluation Type Probability Distribution Relative Standard Uncertainty y REF Repeatability of the REF reading A (50) t 0.03% Nonlinearity of current measurement device B Rectangular 0.05% y STD Repeatability of the STD reading A (50) t 0.02% Nonlinearity of current measurement device B Rectangular 0.3% y REF y STD Wavelength calibration of monochromator B Rectangular 0.01% Distance dependence of calibration A (5) t 0.22% Radiant power drift B Rectangular <0.1% S STD λ Calibration uncertainty of the STD responsivity B Normal 0.25% S REF λ Combined standard uncertainty of the REF responsivity Normal 0.46% and exposure of the DUT, which are to be set for an optimal signal-to-noise ratio prior to the scan measurement. The collected batch of digital images for one wavelength scan is then treated by a separate data acquisition program, which calculates the average and the standard deviation of pixel values within a predefined AOI for each image. Normally, we set pixels at the center of the image as AOI. The average value of the pixel data in the AOI is denoted as the reading of the DUT at a wavelength of λ, y DUT λ, and the measurement equation of the normalized spectral responsivity of the DUT, S DUT λ, is now given by S DUT λ y DUT S REF λ : (4) y REF If necessary, the final result can be renormalized to its maximum value or to any value at a specific normalization wavelength λ norm : 3. Measurement Results S DUT λ S DUT λ S DUT λ norm : (5) A. Monochrome CCD Sensor and Camera The first DUT we have measured was a monochrome (B/W) progressive scan CCD sensor (Sony, model ICX205AL, pixels, pixel size 4.65 μm 4.65 μm), which is integrated into a digital camera device with an 8 bit analog-to-digital (A/D) converter (ThorLabs, model DCU 224M). Figure 5 shows the measured normalized spectral responsivity of the CCD sensor with expanded uncertainty as error bars (k 2). We see the increased uncertainties at several wavelengths due to weak radiant power of the LED-based tunable source. As the average digital counts at these wavelengths are low, the uncertainty of resolution becomes significant, which also causes abrupt changes of the spectral data. Nevertheless, most of the measurement data in the range from 380 to 900 nm have the relative expanded uncertainties of less than 1% (k 2). The lowest uncertainty is achieved again at the points, where the average count of the DUT was the highest. The uncertainty budget of the measurement result of the monochrome CCD sensor in Fig. 5 is representatively presented in Table 2 for the wavelength of 435 nm. The most significant component beside the calibration uncertainty is the resolution uncertainty of the DUT reading. As the resolution of the digital data is fixed to be unity, the resolution uncertainty depends on the average count of the pixel data in the AOI. For y DUT 244 at 435 nm, we calculate the relative uncertainty due to resolution by p % by assuming a rectangular probability distribution. Compared to Table 1, several systematic components are absent in Table 2 because these components are once considered in the uncertainty of the REF responsivity. For Table 2, the equation for the uncertainty due to wavelength inaccuracy in Eq. (3) should be modified in accordance with Eq. (4) to ydut 1 u rel y DUT 1 y REF u λ : y REF y DUT λ y REF λ (6) The radiance mode measurement is demonstrated for an imaging camera, in which the same monochrome CCD sensor is used. A camera lens system (AVENIR zoom lens, back focal length 50 mm, f 1.3) is attached to the CCD sensor, and the camera Fig. 5. (Color online) Normalized spectral responsivity of a monochrome CCD sensor measured in the irradiance mode. The error bars represent the associated uncertainties of this measurement. 20 February 2013 / Vol. 52, No. 6 / APPLIED OPTICS 1269

9 Table 2. Uncertainty Budget of the Normalized Spectral Responsivity Measurement for a Monochrome CCD Sensor at 435 nm Related Quantity in Eq. (4) Uncertainty Component Description Evaluation Type Probability Distribution Relative Standard Uncertainty y DUT Repeatability of the DUT reading A (400) t 0.10% Resolution B Rectangular 0.12% y REF Repeatability of the REF reading A (50) t 0.03% Nonlinearity of current measurement device B Rectangular 0.05% y DUT y REF Wavelength calibration of monochromator B Rectangular 0.01% Radiant power drift B Rectangular <0.1% S REF λ Calibration uncertainty of the REF responsivity (see Table 1) B Normal 0.46% S DUT λ Combined standard uncertainty of the REF responsivity Normal 0.49% is placed at a distance of approximately 500 mm. The focal length of the lens system is adjusted so that we see a well-focused image of the output port. The camera is aligned so that the AOI is placed at the center of the output port while the size of the AOI is kept much smaller than that of the output port image. Figure 6 shows the measured normalized spectral responsivity of the CCD camera as dot symbols. The error bars indicate the expanded uncertainty of the measurement (k 2). The dashed curve corresponds to the CCD sensor result of Fig. 5, which is plotted together for comparison. The measurement uncertainty of the camera is not very different from that of the sensor. However, we can clearly recognize the effect of the lens system to the spectral responsivity from comparison of the two results. Especially in the short wavelength range, the transmission of the used lens system seems to be low so that the responsivity of the camera tends to nearly zero. These measurement results demonstrate the practicability of our instrument, which can measure both imaging sensors and cameras without extra calibration. B. Trichromatic CMOS Sensor and Camera As the second DUT, we measured a trichromatic CMOS-based device with an 8 bit A/D converter (HVS, model HVR-2300CA, pixels, pixel size 3.2 μm 3.2 μm). The measurement procedure was the same as the case with the CCD device. However, the data acquisition is modified because three different values of red (R), green (G), and blue (B) channels are assigned to each pixel. We treated three channels as independent devices. However, the normalization of three channels is done to the same wavelength of 550 nm in order to compare their responsivities to each other. Figure 7 shows the measured normalized spectral responsivity of the trichromatic CMOS sensor in the irradiance mode with expanded uncertainty as error bars (k 2). The spectral shapes and overlaps of the three channels can be clearly recognized. The responsivity above 680 nm is close to zero because the CMOS sensor contains an infrared-cutting filter in front of the sensor. Figure 8 shows the result of the CMOS camera. For this measurement, the camera lens system used for the monochromatic CCD sensor is attached to the CMOS sensor. For comparison, the results of the CMOS sensor of Fig. 7 are plotted together as dashed curves. We see a significant and complex change of the spectral responsivity for the case of the CMOS camera through attaching the lens system. We presume that a multireflection interference between the lens system and the infrared-cutting filter can cause such a change. Figure 8 shows that such influences from lens systems or other optical components to the spectral responsivity of the device can be effectively identified and investigated by using our instrument. Fig. 6. (Color online) Normalized spectral responsivity of a monochrome CCD camera measured in the radiance mode (dot symbols). The error bars represent the associated uncertainties of this measurement. The dashed curve is the result of the CCD sensor in the camera, which is measured in the irradiance mode and shown in Fig. 5. Fig. 7. (Color online) Normalized spectral responsivity of a trichromatic CMOS sensor measured in the irradiance mode. The error bars represent the associated uncertainties of this measurement APPLIED OPTICS / Vol. 52, No. 6 / 20 February 2013

10 bridge these points. Further improvements of the LED-based source are expected by using, e.g., its excellent modulation property. This work was supported by the Korea Research Institute of Standards and Science (KRISS) under the project Establishment of National Physical Measurement Standards and Improvements of Calibration/ Measurement Capability, grant Fig. 8. (Color online) Normalized spectral responsivity of a trichromatic CMOS camera measured in the radiance mode (dot symbols). The error bars represent the associated uncertainties of this measurement. The dashed curve is the result of the CMOS sensor in the camera, which is measured in the irradiance mode and shown in Fig Summary We realized a new instrumentation scheme for measurement of normalized spectral responsivity of imaging sensors and cameras, which is based on a LED-based tunable spectral source with an integrating sphere. A combination of 25 LEDs with different colors and a single-grating monochromator provided tunable radiation with a bandwidth of less than 5 nm in a range from 380 to 900 nm. A high spectral purity could be achieved without using order-sorting filters and double-grating configuration. We experimentally tested the characteristics of the LED-based tunable uniform source, which can affect the accuracy of the measurement. The measurement procedure is then designed and modified to minimize the error and uncertainty sources. The calibration of the instrument is performed by using spectral lamps and a standard photodiode for spectral responsivity. The instrument is applied to a monochrome CCD device and a trichrome CMOS device. The normalized spectral responsivity is measured both in the irradiance mode for the bare sensors and in the radiance mode for the cameras. The measurement uncertainty is evaluated to be less than 1% except several wavelengths where the radiant power was too low. In conclusion, we demonstrated a high potential of the LED-based tunable uniform source for practical metrology applications. Replacement of conventional light source by LEDs provides a possibility to realize a cost-effective and compact setup of a monochromatic source. For instance, the intrinsically low spectral stray of the LED-based source makes the doublemonochromator configuration and the order-sorting filters unnecessary. Although LEDs cannot provide sufficient radiant power at several wavelengths, careful uncertainty evaluation of the spectral measurement provides sufficient information to identify and References 1. G. C. Holst and T. S. Lomheim, CMOS/CCD Sensors and Camera Systems (SPIE, 2007). 2. Spectral responsivity measurements of detectors, radiometers, and photometers, International Commission on Illumination (CIE) Publication No. 202 (2011). 3. H.-C. Lee, Introduction to Color Imaging Science (Cambridge University, 2005). 4. W. A. Hovis and J. S. Knoll, Characteristics of an internally illuminated calibration sphere, Appl. Opt. 22, (1983). 5. A. Ferrero, J. Campos, and A. Pons, Low-uncertainty absolute radiometric calibration of a CCD, Metrologia 43, S17 S21 (2006). 6. V. E. Anderson, N. P. Fox, and D. H. Nettleton, Highly stable, monochromatic and tunable optical radiation source and its application to high accuracy spectrophotometry, Appl. Opt. 31, (1992). 7. S. W. Brown, G. P. Eppeldauer, and K. R. Lykke, Facility for spectral irradiance and radiance responsivity calibrations using uniform sources, Appl. Opt. 45, (2006). 8. M. Schuster, S. Nevas, A. Sperling, and S. Voelker, Spectral calibration of radiometric detectors using tunable laser sources, Appl. Opt. 51, (2012). 9. K. Mahmoud, S. Park, S.-N. Park, and D.-H. Lee, LED-based tunable monochromatic uniform source for calibration of imaging sensors and cameras, in Proceedings of the 11th International Conference on New Developments and Applications in Optical Radiometry (NEWRAD, 2011), pp K. Mahmoud, S. Park, S.-N. Park, and D.-H. Lee, Measurement of normalized spectral response of digital imaging sensors using a LED-based tunable uniform source, in Proceedings of the XX IMEKO World Congress (IMEKO, 2012), paper TC7-O I. Fryc, S. W. Brown, G. P. Eppeldauer, and Y. Ohno, LEDbased spectrally tunable source for radiometric, photometric, and colorimetric applications, Opt. Eng. 44, (2005). 12. G. Zaid, S.-N. Park, S. Park, and D.-H. Lee, Differential spectra responsivity measurement of photovoltaic detectors with a LED-based integrating sphere source, Appl. Opt. 49, (2010). 13. B. H. Hamadani, J. Roller, B. Dougherty, and H. W. Yoon, Versatile light-emitting-diode-based spectral response measurement system for photovoltaic device characterization, Appl. Opt. 51, (2012). 14. S. Park, Y.-W. Kim, D.-H. Lee, and S.-N. Park, Preparation of a standard light-emitting diode (LED) for photometric measurements by functional seasoning, Metrologia 43, (2006). 15. Evaluation of measurement data guide to the expression of uncertainty in measurement (JCGM, 2008). 16. R. Goebel and M. Stock, Report on the comparison CCPR- K2.b of spectral responsivity measurements in the range 300 nm to 1000 nm, Metrologia 41, (2004). 17. S. Park, D.-H. Lee, Y.-W. Kim, and S.-N. Park, Uncertainty evaluation for the spectroradiometric measurement of the averaged light-emitting diode intensity, Appl. Opt. 46, (2007). 20 February 2013 / Vol. 52, No. 6 / APPLIED OPTICS 1271

New automated laser facility for detector calibrations

New automated laser facility for detector calibrations CORM annual conference, NRC, Ottawa, CANADA June 1, 2012 New automated laser facility for detector calibrations Yuqin Zong National Institute of Standards and Technology Gaithersburg, Maryland USA Overview

More information

High Illuminance Calibration Facility and Procedures

High Illuminance Calibration Facility and Procedures Final manuscript for J. IES, 27-2, 132-140 (1998) High Illuminance Calibration Facility and Procedures Yoshi Ohno Optical Technology Division National Institute of Standards and Technology Metrology A320,

More information

ISS-30-VA. Product tags: Integrating Sphere Source. https://www.gigahertz-optik.de/en-us/product/iss-30-va. Gigahertz-Optik GmbH 1/5

ISS-30-VA. Product tags: Integrating Sphere Source. https://www.gigahertz-optik.de/en-us/product/iss-30-va. Gigahertz-Optik GmbH 1/5 ISS-30-VA https://www.gigahertz-optik.de/en-us/product/iss-30-va Product tags: Integrating Sphere Source Gigahertz-Optik GmbH 1/5 Description standards for spectral radiance Spectroradiometers and other

More information

Integrating Spheres. Why an Integrating Sphere? High Reflectance. How Do Integrating Spheres Work? High Damage Threshold

Integrating Spheres. Why an Integrating Sphere? High Reflectance. How Do Integrating Spheres Work? High Damage Threshold 1354 MINIS Oriel Integrating Spheres Integrating spheres are ideal optical diffusers; they are used for radiometric measurements where uniform illumination or angular collection is essential, for reflectance

More information

Developement of the method for realization of spectral irradiance scale featuring system of

Developement of the method for realization of spectral irradiance scale featuring system of Home Search Collections Journals About Contact us My IOPscience Developement of the method for realization of spectral irradiance scale featuring system of spectral comparisons This article has been downloaded

More information

Effect of Beam Size on Photodiode Saturation

Effect of Beam Size on Photodiode Saturation Effect of Beam Size on Photodiode Saturation Experiments were conducted to demonstrate a change in the saturation point for a FDS1010 silicon photodiode as a function of beam diameter. The saturation point

More information

Experiment 1: Fraunhofer Diffraction of Light by a Single Slit

Experiment 1: Fraunhofer Diffraction of Light by a Single Slit Experiment 1: Fraunhofer Diffraction of Light by a Single Slit Purpose 1. To understand the theory of Fraunhofer diffraction of light at a single slit and at a circular aperture; 2. To learn how to measure

More information

Technical Notes. Integrating Sphere Measurement Part II: Calibration. Introduction. Calibration

Technical Notes. Integrating Sphere Measurement Part II: Calibration. Introduction. Calibration Technical Notes Integrating Sphere Measurement Part II: Calibration This Technical Note is Part II in a three part series examining the proper maintenance and use of integrating sphere light measurement

More information

ECEN. Spectroscopy. Lab 8. copy. constituents HOMEWORK PR. Figure. 1. Layout of. of the

ECEN. Spectroscopy. Lab 8. copy. constituents HOMEWORK PR. Figure. 1. Layout of. of the ECEN 4606 Lab 8 Spectroscopy SUMMARY: ROBLEM 1: Pedrotti 3 12-10. In this lab, you will design, build and test an optical spectrum analyzer and use it for both absorption and emission spectroscopy. The

More information

\Ç à{x ÇtÅx Éy ALLAH à{x `xüv yâä

\Ç à{x ÇtÅx Éy ALLAH à{x `xüv yâä \Ç à{x ÇtÅx Éy ALLAH à{x `xüv yâä Ultraviolet Radiation from Some Types of Outdoor Lighting Lamps Dr.Essam El-Moghazy Photometry and Radiometry division, National Institute for Standards (NIS), Egypt.

More information

Spectral responsivity uncertainty of silicon photodiodes due to calibration spectral bandwidth

Spectral responsivity uncertainty of silicon photodiodes due to calibration spectral bandwidth INSTITUTE OF PHYSICS PUBLISHING Meas. Sci. Technol. 12 (2001) 1926 1931 MEASUREMENT SCIENCE AND TECHNOLOGY Spectral responsivity uncertainty of silicon photodiodes due to calibration spectral bandwidth

More information

Capabilities of NIST SIRCUS for Calibrations of SSI Vis-IR Instruments

Capabilities of NIST SIRCUS for Calibrations of SSI Vis-IR Instruments Capabilities of NIST SIRCUS for Calibrations of SSI Vis-IR Instruments Steve Brown National Institute of Standards & Technology Gaithersburg, MD steve.brown@nist.gov; 301.975.5167 Answer: Ask LASP folks

More information

Instruction Manual of Luminance and Spectral Radiance Calibrations

Instruction Manual of Luminance and Spectral Radiance Calibrations Aalto University School of Electrical Engineering Metrology Research Institute Jari Hovila Pasi Manninen Tuomas Poikonen Petri Kärhä Instruction Manual of Luminance and Version 2.6 07/12/2015 Page 2 (19)

More information

UV-VIS-IR Spectral Responsivity Measurement System for Solar Cells

UV-VIS-IR Spectral Responsivity Measurement System for Solar Cells November 1998 NREL/CP-52-25654 UV-VIS-IR Spectral Responsivity Measurement System for Solar Cells H. Field Presented at the National Center for Photovoltaics Program Review Meeting, September 8 11, 1998,

More information

NFMS THEORY LIGHT AND COLOR MEASUREMENTS AND THE CCD-BASED GONIOPHOTOMETER. Presented by: January, 2015 S E E T H E D I F F E R E N C E

NFMS THEORY LIGHT AND COLOR MEASUREMENTS AND THE CCD-BASED GONIOPHOTOMETER. Presented by: January, 2015 S E E T H E D I F F E R E N C E NFMS THEORY LIGHT AND COLOR MEASUREMENTS AND THE CCD-BASED GONIOPHOTOMETER Presented by: January, 2015 1 NFMS THEORY AND OVERVIEW Contents Light and Color Theory Light, Spectral Power Distributions, and

More information

Solid State Luminance Standards

Solid State Luminance Standards Solid State Luminance Standards Color and luminance correction of: - Imaging colorimeters - Luminance meters - Imaging spectrometers Compact and Robust for Production Environments Correct for instrument

More information

Department of Electrical Engineering and Computer Science, Massachusetts Institute of Technology, 77. Table of Contents 1

Department of Electrical Engineering and Computer Science, Massachusetts Institute of Technology, 77. Table of Contents 1 Efficient single photon detection from 500 nm to 5 μm wavelength: Supporting Information F. Marsili 1, F. Bellei 1, F. Najafi 1, A. E. Dane 1, E. A. Dauler 2, R. J. Molnar 2, K. K. Berggren 1* 1 Department

More information

An imaging device for multispectral analysis in the visible range. P. Fiorentin, E. Pedrotti, A. Scroccaro

An imaging device for multispectral analysis in the visible range. P. Fiorentin, E. Pedrotti, A. Scroccaro An imaging device for multispectral analysis in the visible range P. Fiorentin, E. Pedrotti, A. Scroccaro University of Padova, Department of Electrical Engineering, via Gradenigo 6/a, Padov, Italy, +39498277914,

More information

Improving the Collection Efficiency of Raman Scattering

Improving the Collection Efficiency of Raman Scattering PERFORMANCE Unparalleled signal-to-noise ratio with diffraction-limited spectral and imaging resolution Deep-cooled CCD with excelon sensor technology Aberration-free optical design for uniform high resolution

More information

Laser Beam Analysis Using Image Processing

Laser Beam Analysis Using Image Processing Journal of Computer Science 2 (): 09-3, 2006 ISSN 549-3636 Science Publications, 2006 Laser Beam Analysis Using Image Processing Yas A. Alsultanny Computer Science Department, Amman Arab University for

More information

Novel Approach for LED Luminous Intensity Measurement

Novel Approach for LED Luminous Intensity Measurement Novel Approach for LED Luminous Intensity Measurement Ron Rykowski Hubert Kostal, Ph.D. * Radiant Imaging, Inc., 15321 Main Street NE, Duvall, WA, 98019 ABSTRACT Light emitting diodes (LEDs) are being

More information

Fully Portable Spectrophotometer for Transmission Measurement

Fully Portable Spectrophotometer for Transmission Measurement November 2008 LCRT-2005-S Portable Spectrophotometer for Transmission Measurement Spectral Transmission in d/0 and 0/0 Measurement Geometries Transparency Measurement in real in-line Set-up Photometric

More information

Development of 2 Total Spectral Radiant Flux Standards at NIST

Development of 2 Total Spectral Radiant Flux Standards at NIST CIE/USA Annual Conference, October 7, 2014, Seattle, WA Development of 2 Total Spectral Radiant Flux Standards at NIST Yuqin Zong National Institute of Standards and Technology Gaithersburg, Maryland Outline

More information

Multispectral. imaging device. ADVANCED LIGHT ANALYSIS by. Most accurate homogeneity MeasureMent of spectral radiance. UMasterMS1 & UMasterMS2

Multispectral. imaging device. ADVANCED LIGHT ANALYSIS by. Most accurate homogeneity MeasureMent of spectral radiance. UMasterMS1 & UMasterMS2 Multispectral imaging device Most accurate homogeneity MeasureMent of spectral radiance UMasterMS1 & UMasterMS2 ADVANCED LIGHT ANALYSIS by UMaster Ms Multispectral Imaging Device UMaster MS Description

More information

Part 1: New spectral stuff going on at NIST. Part 2: TSI Traceability of TRF to NIST

Part 1: New spectral stuff going on at NIST. Part 2: TSI Traceability of TRF to NIST Part 1: New spectral stuff going on at NIST SIRCUS-type stuff (tunable lasers) now migrating to LASP Absolute Spectrally-Tunable Detector-Based Source Spectrally-programmable source calibrated via NIST

More information

Instructions for the Experiment

Instructions for the Experiment Instructions for the Experiment Excitonic States in Atomically Thin Semiconductors 1. Introduction Alongside with electrical measurements, optical measurements are an indispensable tool for the study of

More information

Calibration of a High Dynamic Range, Low Light Level Visible Source

Calibration of a High Dynamic Range, Low Light Level Visible Source Calibration of a High Dynamic Range, Low Light Level Visible Source Joe LaVeigne a, Todd Szarlan a, Nate Radtke a a Santa Barbara Infrared, Inc., 30 S. Calle Cesar Chavez, #D, Santa Barbara, CA 93103 ABSTRACT

More information

Influence of the light spectral distribution used in the radiometers calibration

Influence of the light spectral distribution used in the radiometers calibration Influence of the light spectral distribution used in the radiometers calibration J C Moraes and L O Guerra Allergisa Pesquisa Dermato-Cosmética Ltda GRUPO INVESTIGA, 452. Dr. Romeu Tórtima Avenue, Postcode:

More information

Development of a new multi-wavelength confocal surface profilometer for in-situ automatic optical inspection (AOI)

Development of a new multi-wavelength confocal surface profilometer for in-situ automatic optical inspection (AOI) Development of a new multi-wavelength confocal surface profilometer for in-situ automatic optical inspection (AOI) Liang-Chia Chen 1#, Chao-Nan Chen 1 and Yi-Wei Chang 1 1. Institute of Automation Technology,

More information

OL Series 426 Low Light Level lntegrating Sphere Calibration Standards

OL Series 426 Low Light Level lntegrating Sphere Calibration Standards OL Series 426 Low Light Level lntegrating Sphere Calibration Standards GENERAL The OL Series 426 Low-Light-Level Calibration Standard is designed for accurately calibrating very sensitive microphotometers,

More information

RADIOMETRIC AND PHOTOMETRIC MEASUREMENTS AT THE LNE-INM/CNAM

RADIOMETRIC AND PHOTOMETRIC MEASUREMENTS AT THE LNE-INM/CNAM RADIOMETRIC AND PHOTOMETRIC MEASUREMENTS AT THE LNE-INM/CNAM Jean Bastie, Laura Patricia Gonzalez Galvan. Institut National de Métrologie Conservatoire National des Arts et Métiers 61 rue du Landy 93210

More information

Current and Future Realizations NRC Photometric and Spectroradiometric Calibration Chains

Current and Future Realizations NRC Photometric and Spectroradiometric Calibration Chains Measurement Science and Standards (MSS) Current and Future Realizations NRC Photometric and Spectroradiometric Calibration Chains Arnold A. Gaertner Photometry, Radiometry and Thermometry (PRT) CORM 30

More information

Instruction manual for Ocean Optics USB4000 and QE65 Pro spectroradiometers

Instruction manual for Ocean Optics USB4000 and QE65 Pro spectroradiometers Aalto University School of Electrical Engineering Metrology Research Institute Hans Baumgartner Instruction manual for Version 1.0 24/11/2016 Page 2 (9) 1. Table of contents 1. Table of contents... 2 2.

More information

Keysight Technologies Optical Power Meter Head Special Calibrations. Brochure

Keysight Technologies Optical Power Meter Head Special Calibrations. Brochure Keysight Technologies Optical Power Meter Head Special Calibrations Brochure Introduction The test and measurement equipment you select and maintain in your production and qualification setups is one of

More information

LINEARPYROMETER LP4. Technical Documentation KE November TN

LINEARPYROMETER LP4. Technical Documentation KE November TN 1 LINEARPYROMETER LP4 Technical Documentation KE 256-6.2007 November 2010 5-TN-1622-100 2 1. General Description With the Linearpyrometer Type LP4 a measuring instrument has been made available for pyrometric

More information

Chameleon Spectroradiometer

Chameleon Spectroradiometer Chameleon Spectroradiometer SPEED / STABILITY / ACCURACY The Chameleon Series is designed to measure the absolute spectral of power, of the light under test. Adding optical probes and absolute calibration

More information

Optical Coherence: Recreation of the Experiment of Thompson and Wolf

Optical Coherence: Recreation of the Experiment of Thompson and Wolf Optical Coherence: Recreation of the Experiment of Thompson and Wolf David Collins Senior project Department of Physics, California Polytechnic State University San Luis Obispo June 2010 Abstract The purpose

More information

HR2000+ Spectrometer. User-Configured for Flexibility. now with. Spectrometers

HR2000+ Spectrometer. User-Configured for Flexibility. now with. Spectrometers Spectrometers HR2000+ Spectrometer User-Configured for Flexibility HR2000+ One of our most popular items, the HR2000+ Spectrometer features a high-resolution optical bench, a powerful 2-MHz analog-to-digital

More information

Photometry for Traffic Engineers...

Photometry for Traffic Engineers... Photometry for Traffic Engineers... Workshop presented at the annual meeting of the Transportation Research Board in January 2000 by Frank Schieber Heimstra Human Factors Laboratories University of South

More information

LSST All-Sky IR Camera Cloud Monitoring Test Results

LSST All-Sky IR Camera Cloud Monitoring Test Results LSST All-Sky IR Camera Cloud Monitoring Test Results Jacques Sebag a, John Andrew a, Dimitri Klebe b, Ronald D. Blatherwick c a National Optical Astronomical Observatory, 950 N Cherry, Tucson AZ 85719

More information

Estimation of spectral response of a consumer grade digital still camera and its application for temperature measurement

Estimation of spectral response of a consumer grade digital still camera and its application for temperature measurement Indian Journal of Pure & Applied Physics Vol. 47, October 2009, pp. 703-707 Estimation of spectral response of a consumer grade digital still camera and its application for temperature measurement Anagha

More information

Spectral Analysis of the LUND/DMI Earthshine Telescope and Filters

Spectral Analysis of the LUND/DMI Earthshine Telescope and Filters Spectral Analysis of the LUND/DMI Earthshine Telescope and Filters 12 August 2011-08-12 Ahmad Darudi & Rodrigo Badínez A1 1. Spectral Analysis of the telescope and Filters This section reports the characterization

More information

Exposure schedule for multiplexing holograms in photopolymer films

Exposure schedule for multiplexing holograms in photopolymer films Exposure schedule for multiplexing holograms in photopolymer films Allen Pu, MEMBER SPIE Kevin Curtis,* MEMBER SPIE Demetri Psaltis, MEMBER SPIE California Institute of Technology 136-93 Caltech Pasadena,

More information

Introduction to the operating principles of the HyperFine spectrometer

Introduction to the operating principles of the HyperFine spectrometer Introduction to the operating principles of the HyperFine spectrometer LightMachinery Inc., 80 Colonnade Road North, Ottawa ON Canada A spectrometer is an optical instrument designed to split light into

More information

Photometry for Traffic Engineers...

Photometry for Traffic Engineers... Photometry for Traffic Engineers... Workshop presented at the annual meeting of the Transportation Research Board in January 2000 by Frank Schieber Heimstra Human Factors Laboratories University of South

More information

CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES

CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES The current multiplication mechanism offered by dynodes makes photomultiplier tubes ideal for low-light-level measurement. As explained earlier, there

More information

PHYS General Physics II Lab Diffraction Grating

PHYS General Physics II Lab Diffraction Grating 1 PHYS 1040 - General Physics II Lab Diffraction Grating In this lab you will perform an experiment to understand the interference of light waves when they pass through a diffraction grating and to determine

More information

BTS2048-UV. Product tags: UV, Spectral Data, LED Binning, Industrial Applications, LED. https://www.gigahertz-optik.de/en-us/product/bts2048-uv

BTS2048-UV. Product tags: UV, Spectral Data, LED Binning, Industrial Applications, LED. https://www.gigahertz-optik.de/en-us/product/bts2048-uv BTS2048-UV https://www.gigahertz-optik.de/en-us/product/bts2048-uv Product tags: UV, Spectral Data, LED Binning, Industrial Applications, LED Gigahertz-Optik GmbH 1/8 Description UV CCD spectroradiometer

More information

Radiometric and Photometric Measurements with TAOS PhotoSensors

Radiometric and Photometric Measurements with TAOS PhotoSensors INTELLIGENT OPTO SENSOR DESIGNER S NUMBER 21 NOTEBOOK Radiometric and Photometric Measurements with TAOS PhotoSensors contributed by Todd Bishop March 12, 2007 ABSTRACT Light Sensing applications use two

More information

Miniaturized Spectroradiometer

Miniaturized Spectroradiometer Miniaturized Spectroradiometer Thomas Morgenstern, Gudrun Bornhoeft, Steffen Goerlich JETI Technische Instrumente GmbH, Jena, Germany Abstract This paper describes the basics of spectroradiometric instruments

More information

We bring quality to light. CAS 120 CCD Array Spectrometer

We bring quality to light. CAS 120 CCD Array Spectrometer We bring quality to light. CAS 120 CCD Array Spectrometer The features at a glance Precision spectrograph with integrated density filter wheel (OD 1 4) Shutter for automatic dark current correction 2048

More information

Chemistry Instrumental Analysis Lecture 7. Chem 4631

Chemistry Instrumental Analysis Lecture 7. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 7 UV to IR Components of Optical Basic components of spectroscopic instruments: stable source of radiant energy transparent container to hold sample device

More information

BTS2048-BS. Product tags: VIS, Spectral Data, Industrial Applications, Laser.

BTS2048-BS. Product tags: VIS, Spectral Data, Industrial Applications, Laser. BTS2048-BS http://www.gigahertz-optik.de/en-us/product/bts2048-bs Product tags: VIS, Spectral Data, Industrial Applications, Laser Gigahertz-Optik GmbH 1/6 Description BTS2048-BS CCD spectroradiometer

More information

Characteristics of point-focus Simultaneous Spatial and temporal Focusing (SSTF) as a two-photon excited fluorescence microscopy

Characteristics of point-focus Simultaneous Spatial and temporal Focusing (SSTF) as a two-photon excited fluorescence microscopy Characteristics of point-focus Simultaneous Spatial and temporal Focusing (SSTF) as a two-photon excited fluorescence microscopy Qiyuan Song (M2) and Aoi Nakamura (B4) Abstracts: We theoretically and experimentally

More information

Chemistry Instrumental Analysis Lecture 10. Chem 4631

Chemistry Instrumental Analysis Lecture 10. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 10 Types of Instrumentation Single beam Double beam in space Double beam in time Multichannel Speciality Types of Instrumentation Single beam Requires stable

More information

LBIR Fluid Bath Blackbody for Cryogenic Vacuum Calibrations

LBIR Fluid Bath Blackbody for Cryogenic Vacuum Calibrations LBIR Fluid Bath Blackbody for Cryogenic Vacuum Calibrations Timothy M. Jung*, Adriaan C. Carter*, Dale R. Sears*, Solomon I. Woods #, Dana R. Defibaugh #, Simon G. Kaplan #, Jinan Zeng * Jung Research

More information

LED Tester BTS256-LED

LED Tester BTS256-LED 1 LED Tester BTS256-LED The BTS256-LED tester is one of the most compact light measurement devices with integrated integrating sphere for high accuracy measurement of luminous flux, spectral and color

More information

Be aware that there is no universal notation for the various quantities.

Be aware that there is no universal notation for the various quantities. Fourier Optics v2.4 Ray tracing is limited in its ability to describe optics because it ignores the wave properties of light. Diffraction is needed to explain image spatial resolution and contrast and

More information

QE65000 Spectrometer. Scientific-Grade Spectroscopy in a Small Footprint. now with. Spectrometers

QE65000 Spectrometer. Scientific-Grade Spectroscopy in a Small Footprint. now with. Spectrometers QE65000 Spectrometer Scientific-Grade Spectroscopy in a Small Footprint QE65000 The QE65000 Spectrometer is the most sensitive spectrometer we ve developed. Its Hamamatsu FFT-CCD detector provides 90%

More information

POTENTIAL OF MULTISPECTRAL TECHNIQUES FOR MEASURING COLOR IN THE AUTOMOTIVE SECTOR

POTENTIAL OF MULTISPECTRAL TECHNIQUES FOR MEASURING COLOR IN THE AUTOMOTIVE SECTOR POTENTIAL OF MULTISPECTRAL TECHNIQUES FOR MEASURING COLOR IN THE AUTOMOTIVE SECTOR Meritxell Vilaseca, Francisco J. Burgos, Jaume Pujol 1 Technological innovation center established in 1997 with the aim

More information

Intercomparison of radiation temperature measurements over the temperature range from 1600 K to 3300 K

Intercomparison of radiation temperature measurements over the temperature range from 1600 K to 3300 K INSTITUTE OF PHYSICS PUBLISHING Metrologia 40 (2003) S39 S44 METROLOGIA PII: S0026-1394(03)56056-X Intercomparison of radiation temperature measurements over the temperature range from 1600 K to 3300 K

More information

Application Note (A11)

Application Note (A11) Application Note (A11) Slit and Aperture Selection in Spectroradiometry REVISION: C August 2013 Gooch & Housego 4632 36 th Street, Orlando, FL 32811 Tel: 1 407 422 3171 Fax: 1 407 648 5412 Email: sales@goochandhousego.com

More information

A novel tunable diode laser using volume holographic gratings

A novel tunable diode laser using volume holographic gratings A novel tunable diode laser using volume holographic gratings Christophe Moser *, Lawrence Ho and Frank Havermeyer Ondax, Inc. 85 E. Duarte Road, Monrovia, CA 9116, USA ABSTRACT We have developed a self-aligned

More information

LED Lighting Flux and Color Measurement System (LFC)

LED Lighting Flux and Color Measurement System (LFC) LED Lighting Flux and Color Measurement System (LFC) Fast, accurate and complete System for research and production LFC System is a high cost efficiency light measurement system. The system is designed

More information

RGB Laser Meter TM6102, RGB Laser Luminance Meter TM6103, Optical Power Meter TM6104

RGB Laser Meter TM6102, RGB Laser Luminance Meter TM6103, Optical Power Meter TM6104 1 RGB Laser Meter TM6102, RGB Laser Luminance Meter TM6103, Optical Power Meter TM6104 Abstract The TM6102, TM6103, and TM6104 accurately measure the optical characteristics of laser displays (characteristics

More information

Single Slit Diffraction

Single Slit Diffraction PC1142 Physics II Single Slit Diffraction 1 Objectives Investigate the single-slit diffraction pattern produced by monochromatic laser light. Determine the wavelength of the laser light from measurements

More information

Observational Astronomy

Observational Astronomy Observational Astronomy Instruments The telescope- instruments combination forms a tightly coupled system: Telescope = collecting photons and forming an image Instruments = registering and analyzing the

More information

Aqualog. CDOM Measurements Made Easy PARTICLE CHARACTERIZATION ELEMENTAL ANALYSIS FLUORESCENCE GRATINGS & OEM SPECTROMETERS OPTICAL COMPONENTS RAMAN

Aqualog. CDOM Measurements Made Easy PARTICLE CHARACTERIZATION ELEMENTAL ANALYSIS FLUORESCENCE GRATINGS & OEM SPECTROMETERS OPTICAL COMPONENTS RAMAN Aqualog CDOM Measurements Made Easy ELEMENTAL ANALYSIS FLUORESCENCE GRATINGS & OEM SPECTROMETERS OPTICAL COMPONENTS PARTICLE CHARACTERIZATION RAMAN SPECTROSCOPIC ELLIPSOMETRY SPR IMAGING CDOM measurements

More information

Imaging Photometer and Colorimeter

Imaging Photometer and Colorimeter W E B R I N G Q U A L I T Y T O L I G H T. /XPL&DP Imaging Photometer and Colorimeter Two models available (photometer and colorimetry camera) 1280 x 1000 pixels resolution Measuring range 0.02 to 200,000

More information

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer Page 1 of 11 Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer The Aramis Raman system is a software selectable multi-wavelength Raman system with mapping capabilities with a 400mm monochromator and

More information

SPECTRAL SCANNER. Recycling

SPECTRAL SCANNER. Recycling SPECTRAL SCANNER The Spectral Scanner, produced on an original project of DV s.r.l., is an instrument to acquire with extreme simplicity the spectral distribution of the different wavelengths (spectral

More information

CCT/10-13 MeP-K direct methods. Section 3.2: Absolute (spectral-band) radiometry (radiation thermometry)

CCT/10-13 MeP-K direct methods. Section 3.2: Absolute (spectral-band) radiometry (radiation thermometry) CCT/10-13 MeP-K direct methods Section 3.2: Absolute (spectral-band) radiometry (radiation thermometry) Authors: Graham Machin (NPL, chair), Klaus Anhalt (PTB), Pieter Bloembergen (NIM, formerly NMIJ &

More information

Spectroscopy in the UV and Visible: Instrumentation. Spectroscopy in the UV and Visible: Instrumentation

Spectroscopy in the UV and Visible: Instrumentation. Spectroscopy in the UV and Visible: Instrumentation Spectroscopy in the UV and Visible: Instrumentation Typical UV-VIS instrument 1 Source - Disperser Sample (Blank) Detector Readout Monitor the relative response of the sample signal to the blank Transmittance

More information

Camera Calibration Certificate No: DMC III 27542

Camera Calibration Certificate No: DMC III 27542 Calibration DMC III Camera Calibration Certificate No: DMC III 27542 For Peregrine Aerial Surveys, Inc. #201 1255 Townline Road Abbotsford, B.C. V2T 6E1 Canada Calib_DMCIII_27542.docx Document Version

More information

BTS256-E WiFi - mobile light meter for photopic and scotopic illuminance, EVE factor, luminous color, color rendering index and luminous spectrum.

BTS256-E WiFi - mobile light meter for photopic and scotopic illuminance, EVE factor, luminous color, color rendering index and luminous spectrum. Page 1 BTS256-E WiFi - mobile light meter for photopic and scotopic illuminance, EVE factor, luminous color, color rendering index and luminous spectrum. The BTS256-E WiFi is a high-quality light meter

More information

Optical transfer function shaping and depth of focus by using a phase only filter

Optical transfer function shaping and depth of focus by using a phase only filter Optical transfer function shaping and depth of focus by using a phase only filter Dina Elkind, Zeev Zalevsky, Uriel Levy, and David Mendlovic The design of a desired optical transfer function OTF is a

More information

A stray light corrected array spectroradiometer for complex high dynamic range measurements in the UV spectral range.

A stray light corrected array spectroradiometer for complex high dynamic range measurements in the UV spectral range. A stray light corrected array spectroradiometer for complex high dynamic range measurements in the UV spectral range Mike Clark Gigahertz-Optik GmbH m.clark@gigahertz-optik.de Array spectroradiometers

More information

A Quantix monochrome camera with a Kodak KAF6303E CCD 2-D array was. characterized so that it could be used as a component of a multi-channel visible

A Quantix monochrome camera with a Kodak KAF6303E CCD 2-D array was. characterized so that it could be used as a component of a multi-channel visible A Joint Research Program of The National Gallery of Art, Washington The Museum of Modern Art, New York Rochester Institute of Technology Technical Report March, 2002 Characterization of a Roper Scientific

More information

NIST Agency Report May 2012 OUTLINE. The case for traceability NMI capabilities A view to the future the HIP Current/recent NIST activities

NIST Agency Report May 2012 OUTLINE. The case for traceability NMI capabilities A view to the future the HIP Current/recent NIST activities NIST Agency Report May 2012 OUTLINE The case for traceability NMI capabilities A view to the future the HIP Current/recent NIST activities The case for traceability Earth Radiation Budget: Solar irradiance

More information

We bring quality to light. Spectro 320 Scanning Spectrometer

We bring quality to light. Spectro 320 Scanning Spectrometer We bring quality to light. Spectro 320 Scanning Spectrometer light measurement Instrument description Applications Spectroradiometry NVIS Display measurements Specifications Key features at a glance Fast-Scan

More information

Measurements of Infrared Sources with the Missile Defense Transfer Radiometer

Measurements of Infrared Sources with the Missile Defense Transfer Radiometer Measurements of Infrared Sources with the Missile Defense Transfer Radiometer Simon G. Kaplan #, Solomon I. Woods #, Adriaan C. Carter, and Timothy M. Jung * # National Institute of Standards and Technology

More information

Coherent Laser Measurement and Control Beam Diagnostics

Coherent Laser Measurement and Control Beam Diagnostics Coherent Laser Measurement and Control M 2 Propagation Analyzer Measurement and display of CW laser divergence, M 2 (or k) and astigmatism sizes 0.2 mm to 25 mm Wavelengths from 220 nm to 15 µm Determination

More information

Experimental Analysis of Luminescence in Printed Materials

Experimental Analysis of Luminescence in Printed Materials Experimental Analysis of Luminescence in Printed Materials A. D. McGrath, S. M. Vaezi-Nejad Abstract - This paper is based on a printing industry research project nearing completion [1]. While luminescent

More information

Quantum Efficiency Measurement System with Internal Quantum Efficiency Upgrade

Quantum Efficiency Measurement System with Internal Quantum Efficiency Upgrade Quantum Efficiency Measurement System with Internal Quantum Efficiency Upgrade QE / IPCE SYSTEM Upgraded with Advanced Features Includes IV Testing, Spectral Response, Quantum Efficiency System/ IPCE System

More information

PoS(PhotoDet 2012)058

PoS(PhotoDet 2012)058 Absolute Photo Detection Efficiency measurement of Silicon PhotoMultipliers Vincent CHAUMAT 1, Cyril Bazin, Nicoleta Dinu, Véronique PUILL 1, Jean-François Vagnucci Laboratoire de l accélérateur Linéaire,

More information

Spectroradiometer specbos 1201

Spectroradiometer specbos 1201 Spectroradiometer specbos 1201 specbos 1201 is a precise and compact VIS spectroradiometer. It can be used in laboratory as well as production environment to measure the following quantities: Luminance,

More information

Application Note (A16)

Application Note (A16) Application Note (A16) Eliminating LED Measurement Errors Revision: A December 2001 Gooch & Housego 4632 36 th Street, Orlando, FL 32811 Tel: 1 407 422 3171 Fax: 1 407 648 5412 Email: sales@goochandhousego.com

More information

PROCEEDINGS OF SPIE. Measuring the image quality of digital-camera sensors by a pingpong

PROCEEDINGS OF SPIE. Measuring the image quality of digital-camera sensors by a pingpong PROCEEDINGS OF SPIE SPIEDigitalLibrary.org/conference-proceedings-of-spie Measuring the image quality of digital-camera sensors by a pingpong ball Antonio M. Pozo, Manuel Rubiño, José J. Castro, Carlos

More information

The Issues of Measurement of Optical Hazard Using Photometers EMRP JRP ENG05 Metrology for Solid State Lighting

The Issues of Measurement of Optical Hazard Using Photometers EMRP JRP ENG05 Metrology for Solid State Lighting The Issues of Measurement of Optical Hazard Using Photometers EMRP JRP ENG05 Metrology for Solid State Lighting Simon Hall,Paul Miller, Neil Haigh, Ben Thornton, Neil Haigh (Lux TSI) 25 th April 2013 Background

More information

Preliminary Characterization Results: Fiber-Coupled, Multi-channel, Hyperspectral Spectrographs

Preliminary Characterization Results: Fiber-Coupled, Multi-channel, Hyperspectral Spectrographs Preliminary Characterization Results: Fiber-Coupled, Multi-channel, Hyperspectral Spectrographs Carol Johnson, NIST MODIS-VIIRS Team Meeting January 26-28, 2010 Washington, DC Marine Optical System & Data

More information

1/8 m GRATING MONOCHROMATOR

1/8 m GRATING MONOCHROMATOR 1/8 m GRATING GRATING OUTPUT PORT INPUT PORT 77250 1/8 m Monochromator with 6025 Hg(Ar) Spectral Calibration Lamp. Low cost, compact size and high performance, ideal for OEM applications Very efficient

More information

Image acquisition. In both cases, the digital sensing element is one of the following: Line array Area array. Single sensor

Image acquisition. In both cases, the digital sensing element is one of the following: Line array Area array. Single sensor Image acquisition Digital images are acquired by direct digital acquisition (digital still/video cameras), or scanning material acquired as analog signals (slides, photographs, etc.). In both cases, the

More information

14.2 Photodiodes 411

14.2 Photodiodes 411 14.2 Photodiodes 411 Maximum reverse voltage is specified for Ge and Si photodiodes and photoconductive cells. Exceeding this voltage can cause the breakdown and severe deterioration of the sensor s performance.

More information

An Optical Characteristic Testing System for the Infrared Fiber in a Transmission Bandwidth 9-11μm

An Optical Characteristic Testing System for the Infrared Fiber in a Transmission Bandwidth 9-11μm An Optical Characteristic Testing System for the Infrared Fiber in a Transmission Bandwidth 9-11μm Ma Yangwu *, Liang Di ** Center for Optical and Electromagnetic Research, State Key Lab of Modern Optical

More information

Luminous Flux (lumens) System LM-ISP-XXXX

Luminous Flux (lumens) System LM-ISP-XXXX Company Profile Taiwan Optimum Optoelectrics Co., Ltd. is Taiwan's the first of photometric and colorimetric measurement equipments and professional supplier of calibration services in Taiwan. In particular,

More information

Design Description Document

Design Description Document UNIVERSITY OF ROCHESTER Design Description Document Flat Output Backlit Strobe Dare Bodington, Changchen Chen, Nick Cirucci Customer: Engineers: Advisor committee: Sydor Instruments Dare Bodington, Changchen

More information

metcon meteorologieconsultgmbh, Instruments for Atmospheric Research W1aa_Feb_2017_1.doc 1 -

metcon meteorologieconsultgmbh, Instruments for Atmospheric Research W1aa_Feb_2017_1.doc 1 - metcon meteorologieconsultgmbh, Instruments for Atmospheric Research W1aa_Feb_2017_1.doc 1 - ACTINIC FLUX SPECTRAL RADIOMETERS Ultra-fast CCD-Detector Spectrometer, UVB enhanced Cooled CCD, 512 pixel *

More information

SR-5000N design: spectroradiometer's new performance improvements in FOV response uniformity (flatness) scan speed and other important features

SR-5000N design: spectroradiometer's new performance improvements in FOV response uniformity (flatness) scan speed and other important features SR-5000N design: spectroradiometer's new performance improvements in FOV response uniformity (flatness) scan speed and other important features Dario Cabib *, Shmuel Shapira, Moshe Lavi, Amir Gil and Uri

More information

Ultra-stable flashlamp-pumped laser *

Ultra-stable flashlamp-pumped laser * SLAC-PUB-10290 September 2002 Ultra-stable flashlamp-pumped laser * A. Brachmann, J. Clendenin, T.Galetto, T. Maruyama, J.Sodja, J. Turner, M. Woods Stanford Linear Accelerator Center, 2575 Sand Hill Rd.,

More information

Laser Speckle Reducer LSR-3000 Series

Laser Speckle Reducer LSR-3000 Series Datasheet: LSR-3000 Series Update: 06.08.2012 Copyright 2012 Optotune Laser Speckle Reducer LSR-3000 Series Speckle noise from a laser-based system is reduced by dynamically diffusing the laser beam. A

More information