UV-VIS-IR Spectral Responsivity Measurement System for Solar Cells

Size: px
Start display at page:

Download "UV-VIS-IR Spectral Responsivity Measurement System for Solar Cells"

Transcription

1 November 1998 NREL/CP UV-VIS-IR Spectral Responsivity Measurement System for Solar Cells H. Field Presented at the National Center for Photovoltaics Program Review Meeting, September 8 11, 1998, Denver, Colorado National Renewable Energy Laboratory 1617 Cole Boulevard Golden, Colorado A national laboratory of the U.S. Department of Energy Managed by Midwest Research Institute for the U.S. Department of Energy under Contract No. DE-AC36-83CH193 Prepared under Task No. PV9341 November 1998

2 UV-VIS-IR Spectral Responsivity Measurement System for Solar Cells Halden Field National Renewable Energy Laboratory (NREL) 1617 Cole Blvd., Golden, Colorado, USA 841 Abstract. NREL s PV Cell and Module Performance Characterization group has built a new spectral responsivity measurement system for solar cells. It uses a xenon arc lamp source, a single, grating monochrometer, and a fiber-optic bundle to couple the monochromatic light to the test device. The system has a spectral bandwidth of 2 nm, minimum spot diameter of 1.6 mm, a spectral range of nm, and uncertainty better than ±3% over most of this range. It is capable of incorporating light bias with intensities exceeding one sun. This paper discusses the system s features, capabilities, calibration, and measurement uncertainties. BACKGROUND A photovoltaic (PV) device s spectral responsivity describes its ability to convert light of various wavelengths to electricity. It is often reported as the ratio of device current divided by incident-beam power (e.g., A/W) or device current divided by incident photon flux (i.e., quantum efficiency). Researchers can use spectral responsivity measurements to help understand device operation. Data from such measurements are also used in spectral mismatch parameter (1) calculations used to set solar simulator intensity for solar cell and module performance measurements. SYSTEM DESCRIPTION Figure 1 illustrates the system s major components and their configuration. Table 1 lists major system specifications. Calibration devices are NIST-calibrated silicon photodiodes and a Laser Probe RS-59 electrically-calibrated pyroelectric radiometer calibrated by the manufacturer with NIST traceability. Other component details are available from the author. PROCEDURES To determine a device s spectral responsivity, one must know the power or irradiance reaching the test device at each wavelength and the current produced by the device at each of those wavelengths. In this system, the power is measured with a calibrated photodiode or a pyroelectric radiometer. At the same time, the current produced by the monitor photodiode is measured. The computer controlling the system records the ratio of these two quantities for later use when the test device s response is measured.

3 Xe arc lamp ellipsoidal reflector 75 W Chopper Stray light and order-sorting filters.19 m Monochrometer 3-6 g/mm gratings Lock-in amplifier Monitor detector 1 fiber 18 fibers Solar cell Bias light amplifier Lock-in Fiber bundle Computer control FIGURE 1. Equipment configuration for spectral responsivity measurement system. TABLE 1. System Specifications Item Specification Notes Spectral range no light bias Spectral resolution 2 nm monochromatic beam spectral bandwidth Spectral step size.14 nm minimum Uncertainty 3% nm, no light bias 1% < 31 nm, >16 nm, no light bias Wavelength uncertainty ±2 nm Beam size 1.6-mm diameter minimum Beam power ~8 µw maximum Beam power density ~4 mw/cm 2.4 "suns" Light-bias capability 1.5 "suns" small or apertured devices During a test, the computer records the currents produced by the test device and the monitor photodiode at the same time for each wavelength in the test. Using the powerto-current ratio previously recorded, it converts the monitor current to a beam power quantity. The ratio of test device current I TD (λ) to beam power is the device responsivity, which is converted to units of quantum efficiency by using the monochrometer s wavelength setting λ: QE( λ) = ( ) ( ), 1% h c I TD λ e λ I MON ( λ) CV MON λ where h is Planck s constant, c is the speed of light, e is the electron charge, I MON (λ) is the monitor-cell current, and CV MON (λ) is the monitor cell s calibration value in W/A units. (1)

4 UNCERTAINTIES This system estimates the uncertainties in its measurements by quantitatively combining uncertainty estimates from various sources during the measurement procedure. It specifies systematic and random components explicitly in the measurement report. Table 2 lists the uncertainties considered for this estimation process. The uncertainty introduced during system calibration depends on the reference device used. The pyroelectric radiometer has uncertainty in the factor applied to correct its readings because the chopped waveform is not square, in its electrical optical equivalence, in its amplifier gain, and in its analog-to-digital converter. The estimate treats as random uncertainties the drift in instrumentation gain during the measurement and the potential gain or detector nonlinearities because they change during the calibration as signal levels vary. TABLE 2. Uncertainty Estimates Source Bias Random Calibration pyroelectric radiometer 2% 2% + as measured Calibration photodiode.2-4.4% + 1% 1% + as measured Measurement 1% 1% + as measured The photodiode calibration report from NIST (2) provides uncertainty estimates for each wavelength with the spectral responsivity data. The software combines these with an additional estimate to account for other uncertainties, including the effect of multiple light reflections between the photodiode and the beam-delivery optics. Uncertainties are combined with the root-sum-square method for the 95% confidence estimate. Fixed instrumentation-gain errors do not contribute to the total uncertainty, because the same instruments are used to amplify the photodiode and test-device signals. The computer collects multiple readings from the calibration device and monitor detector at each wavelength and computes the ratio for each reading. It combines the standard deviation of these ratios, multiplied by the appropriate student s t factor, with the random-error estimate associated with the calibration device used, to estimate the as measured part of the random uncertainty for the calibration shown in Table 2. The computer combines the bias and random uncertainties in the calibration with an additional uncertainty to include the effects of wavelength uncertainty on the quantumefficiency calculation, multiple reflections between the test device and the beamdelivery optics, and others. This result is the bias uncertainty estimate for the data report. The program estimates the measurement random uncertainty by the method described above for the calibration s random uncertainty. It combines this with an additional uncertainty to account for potential instrumentation gain-drift and nonlinearity during the test. Figure 2 illustrates the uncertainty of a measurement made using this system.

5 Relative Uncertainty (%) Spectral Responsivity Measurement Result Total Uncertainty Estimate Bias Uncertainty Estimate Random Uncertainty Estimate 6 8 Wavelength (nm) 1 FIGURE 2. Measurement example and uncertainty estimates (24 ma/cm 2 bias light) Quantum Efficiency (%) WAVELENGTH CALIBRATION AND SPECTRAL BANDWIDTH Adjusting the monochrometer offsets to minimize the difference between its wavelength settings and the results of wavelength calibration checks provides the system s wavelength calibration. Calibration points are provided by line filters calibrated by NREL s CARY 5G spectrophotometer, xenon arc lamp emission lines (3), and a helium-neon laser. Narrow-bandpass (~1-nm) filters commonly used to isolate laser wavelengths were placed over a photodiode, and the responsivity of the combination was measured. Filters with center wavelengths of 324.7, 44.7, 514.5, and nm produced responsivity peaks within.8 nm of the expected wavelength. Emission lines in the system s source at 823.2, 98., and nm appeared 1.2 to 1.8 nm higher than expected in the current vs. wavelength profiles of a bare photodiode s current. Light from a helium-neon laser aligned with the center of the monochrometer appeared within.7 nm of the expected wavelength using first-, second-, and third-order diffractions for all three gratings. The physical positions of the monochrometer s diffraction grating, its other optical components, and the line of optical fibers simulating its exit slit determine the wavelength of the light that reaches the fiber bundle. The relative positions of these components change with temperature. In addition, position repeatability limits for the grating (the one moving part) affect the wavelength calibration. Optical properties of the gratings and fiber optics can contribute to wavelength errors. Finally, errors in the wavelength calibration sources themselves contribute to wavelength uncertainty. Rather than analytically characterize all of the known, potential error sources, an uncertainty estimate of ±2 nm is assigned to encompass these observations. Wavelength calibration errors can also affect the calculation of device quantum efficiency, as the wavelength enters the conversion from power to photon flux (see Equation 1).

6 Small wavelength changes can cause large changes in monochromatic beam intensity when the light source has strong emission lines and the monochrometer wavelength is set near one of those lines. The use of a monitor cell in this system enables the measurement to be relatively insensitive to such changes, as the ratio of test-cell and monitor-cell currents is used to determine the test-device responsivity. With the helium-neon laser used in place of the xenon arc lamp source, the current vs. wavelength profile of a photodiode signal indicates that the system bandwidth is 1.4 nm, consistent with the monochrometer s design specifications. However, similar measurements across xenon emission lines and line filters indicate that the bandwidth may be slightly higher. Though this may be due to finite bandwidth of the line filters and broadening of the emission lines, the system s spectral bandwidth is specified to be 2 nm. CONFIGURATION ISSUES Light Modulation and Filters Use of a light chopper with the monochrometer enables the lock-in amplifier to discriminate between the test-device current resulting from monochromatic light and that from stray and bias light. Order-sorting filters attenuate light that would appear at the monochrometer exit because of higher-order diffractions than the intended one. Stray-light filters attenuate light that could reach the monochrometer exit resulting from reflections from the various surfaces inside the monochrometer, including the mirrors and grating themselves. The light chopper and filter wheel are outside the monochrometer entrance because the presence of the fiber-optic bundle prevents them from being placed at the exit. In this position, the filters reduce the total light reaching the monochrometer, thus reducing heating of the instrument. A disadvantage is that the filters get hot, changing their spectral transmittance. Delays to stabilize filter temperature are incorporated in the software to avoid errors from this problem. Heat can also damage the filters, but a broken filter does not appear to affect this system's performance. Fiber-Optic Bundle Common light fibers used for communication have poor transmission in the ultraviolet (UV). This system uses a fiber doped with OH - ions to boost its UV transmission. A drawback is that the OH - causes substantial absorption between 134 and 141 nm, limiting the system's continuous spectral range. The fiber's numerical aperture of.22 enables it to accept most of the light from the f/3.9 monochrometer. The light fibers are linearly arranged at the bundle entrance to optically resemble a common monochrometer's output slit. Eighteen fibers convey light to the test or calibration device, where the fibers are arranged in a circle. One fiber conveys a sample of the light to the monitor cell, which functions as a calibration transfer standard. One additional fiber can convey light from a bias light source to the test

7 device (see Figure 1), but this method is not yet sufficiently developed. At present, a projector lamp provides bias light. Multiple Reflections Light reflected from the test or calibration device can, in turn, be reflected back to the device by the end of the fiber bundle. Additionally, a small amount of light exits the fiber outside the expected cone. The effect appears as a dependence of signal magnitude on fiber-to-device distance. To minimize this effect, a painted cap covers the ferrule holding the fibers. The paint's reflectivity is about 4%, with little spectral dependence. Even with these precautions, the signal produced by a relatively highreflectivity photodiode varies about 1% between fiber-to-device distances of 3 and 12 mm. BEAM POWER AND STRAY LIGHT Figure 3 presents the spectral beam power for this system. Signal-to-noise performance and the leakage of low-level, out-of-band light through the filters and monochrometer limit measurement capability where the monochromatic beam power is lowest. Figure 4 shows that stray light is well under control in the UV and infrared regions. The left graph shows the device quantum efficiency reported, with the UV component of the monochromatic beam blocked by a 3-75 colorglass filter. The low signal levels represent the extent that stray light influences the measurement results. The right graph shows the device quantum efficiency reported in wavelength ranges for which the test device (a photodiode) should not respond. Beam Power (µw) ~.4 "suns" UV-VIS-IR Solar Cell Spectral Responsivity Measurement System (2-nm bandwidth) Semiconductor-Calibrated Quantum Efficiency Measurement System (8-nm bandwidth) Wavelength (nm) FIGURE 3. This system s beam power is compared to that of another spectral responsivity measurement system at NREL.

8 Quantum Efficiency (%) QE Measurement Results for U124 (UV photodiode) 26 Normal Measurement Measurement with UV light blocked by 3-75 colorglass filter 28 3 Wavelength (nm) 32 Quantum Efficiency (%) Signal from a silicon photodiode in the wavelength region where it should not respond. Background signal level has not been subtracted Wavelength (nm) FIGURE 4. These graphs illustrate that stray light is minimal in wavelength regions where it is most likely to be a problem. FURTHER IMPROVEMENTS Measurements on this system, performed on request to the author by participants in DOE s PV Program, provide a valuable diversity of device characteristics and configurations, revealing opportunities for the system s continuing development. The author would appreciate feedback from potential measurement requesters on additional ideas for this list and suggestions for how to prioritize these items: Substitute rigorous, quantitative analysis for uncertainty judgements. Reduce measurement uncertainty. Improve wavelength calibration. Extend spectral range below 28 nm or above 133 nm. Include uncertainty estimates in measurement reports graphically. Add option to measure responsivity in equal ev or wavenumber increments. Increase bias-light capability. Add temperature control for test devices. ACKNOWLEDGEMENTS The author thanks Ramesh Dhere of NREL for spectral transmissivity and spectral reflection measurements used to explore wavelength-calibration and multiplereflection issues. DOE supported this work under Contract Number DE-AC36-83CH193. REFERENCES 1. Standard Test Method for Determination of the Spectral Mismatch Parameter Between a Photovoltaic Device and a Photovoltaic Reference Cell [Metric], ASTM Standard E 973M 96, West Conshohocken, PA: American Society for Testing and Materials, T.C. Larason, S.S. Bruce, A.C. Parr, NIST Special Publication Spectroradiometric Detector Measurements, Washington, D.C.: U.S. Government Printing Office, 1998, pp. A Also available at 3. Weast, R., Astle, M., and Beyer, W., CRC Handbook of Chemistry and Physics, Boca Raton, Florida: CRC Press, Inc., 1983, pp. E

Quantum Efficiency Measurement System with Internal Quantum Efficiency Upgrade

Quantum Efficiency Measurement System with Internal Quantum Efficiency Upgrade Quantum Efficiency Measurement System with Internal Quantum Efficiency Upgrade QE / IPCE SYSTEM Upgraded with Advanced Features Includes IV Testing, Spectral Response, Quantum Efficiency System/ IPCE System

More information

Measuring optical filters

Measuring optical filters Measuring optical filters Application Note Author Don Anderson and Michelle Archard Agilent Technologies, Inc. Mulgrave, Victoria 3170, Australia Introduction Bandpass filters are used to isolate a narrow

More information

BLACKBODY RADIATION PHYSICS 359E

BLACKBODY RADIATION PHYSICS 359E BLACKBODY RADIATION PHYSICS 359E INTRODUCTION In this laboratory, you will make measurements intended to illustrate the Stefan-Boltzmann Law for the total radiated power per unit area I tot (in W m 2 )

More information

Keysight Technologies Optical Power Meter Head Special Calibrations. Brochure

Keysight Technologies Optical Power Meter Head Special Calibrations. Brochure Keysight Technologies Optical Power Meter Head Special Calibrations Brochure Introduction The test and measurement equipment you select and maintain in your production and qualification setups is one of

More information

Influence of the light spectral distribution used in the radiometers calibration

Influence of the light spectral distribution used in the radiometers calibration Influence of the light spectral distribution used in the radiometers calibration J C Moraes and L O Guerra Allergisa Pesquisa Dermato-Cosmética Ltda GRUPO INVESTIGA, 452. Dr. Romeu Tórtima Avenue, Postcode:

More information

1) Solar simulator with I-V measurement setup and software

1) Solar simulator with I-V measurement setup and software Department of Optoelectronics, University of Kerala, Kariavattom, Thiruvananthapuram, Kerala, India 695581, Ph: 91 471 2308167 OPTO/Nanophotonics-Phase II/P-1/2014-15 Quotation Notice Quotations are invited

More information

Spectrophotometer. An instrument used to make absorbance, transmittance or emission measurements is known as a spectrophotometer :

Spectrophotometer. An instrument used to make absorbance, transmittance or emission measurements is known as a spectrophotometer : Spectrophotometer An instrument used to make absorbance, transmittance or emission measurements is known as a spectrophotometer : Spectrophotometer components Excitation sources Deuterium Lamp Tungsten

More information

Observational Astronomy

Observational Astronomy Observational Astronomy Instruments The telescope- instruments combination forms a tightly coupled system: Telescope = collecting photons and forming an image Instruments = registering and analyzing the

More information

Application Note (A11)

Application Note (A11) Application Note (A11) Slit and Aperture Selection in Spectroradiometry REVISION: C August 2013 Gooch & Housego 4632 36 th Street, Orlando, FL 32811 Tel: 1 407 422 3171 Fax: 1 407 648 5412 Email: sales@goochandhousego.com

More information

THE spectral response (SR) measurement of a solar cell is

THE spectral response (SR) measurement of a solar cell is 944 IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, VOL. 48, NO. 5, OCTOBER 1999 A Fast Low-Cost Solar Cell Spectral Response Measurement System with Accuracy Indicator S. Silvestre, L. Sentís, and

More information

Fundamentals of Infrared Detector Operation and Testing

Fundamentals of Infrared Detector Operation and Testing Fundamentals of Infrared Detector Operation and Testing JOHN DAVID VINCENT Santa Barbara Research Center Goleta, California WILEY A Wiley-Interscience Publication John Wiley & Sons New York I Chichester

More information

1/8 m GRATING MONOCHROMATOR

1/8 m GRATING MONOCHROMATOR 1/8 m GRATING GRATING OUTPUT PORT INPUT PORT 77250 1/8 m Monochromator with 6025 Hg(Ar) Spectral Calibration Lamp. Low cost, compact size and high performance, ideal for OEM applications Very efficient

More information

Chemistry 524--"Hour Exam"--Keiderling Mar. 19, pm SES

Chemistry 524--Hour Exam--Keiderling Mar. 19, pm SES Chemistry 524--"Hour Exam"--Keiderling Mar. 19, 2013 -- 2-4 pm -- 170 SES Please answer all questions in the answer book provided. Calculators, rulers, pens and pencils permitted. No open books allowed.

More information

Chemistry Instrumental Analysis Lecture 10. Chem 4631

Chemistry Instrumental Analysis Lecture 10. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 10 Types of Instrumentation Single beam Double beam in space Double beam in time Multichannel Speciality Types of Instrumentation Single beam Requires stable

More information

Supporting Information for: Getting to the (Square) Root of the Problem: How to Make Non-Coherent Pumped Upconversion Linear

Supporting Information for: Getting to the (Square) Root of the Problem: How to Make Non-Coherent Pumped Upconversion Linear Supporting Information for: Getting to the (Square) Root of the Problem: How to Make Non-Coherent Pumped Upconversion Linear Alexandre Haefele, Jörg Blumhoff, Rony S. Khnayzer, and Felix N. Castellano*

More information

ECEN 4606, UNDERGRADUATE OPTICS LAB

ECEN 4606, UNDERGRADUATE OPTICS LAB ECEN 4606, UNDERGRADUATE OPTICS LAB Lab 10: Photodetectors Original: Professor McLeod SUMMARY: In this lab, you will characterize the fundamental low-frequency characteristics of photodiodes and the circuits

More information

Radiometric Measurement Traceability Paths for Photovoltaic Calibrations. Howard W. Yoon Physical Measurement Laboratory NIST

Radiometric Measurement Traceability Paths for Photovoltaic Calibrations. Howard W. Yoon Physical Measurement Laboratory NIST Radiometric Measurement Traceability Paths for Photovoltaic Calibrations Howard W. Yoon Physical Measurement Laboratory NIST Solar energy and PV Solar radiation: free and abundant! Photovoltaics (PV):

More information

LASP / University of Colorado

LASP / University of Colorado The SORCE SIM Instrument: Progress Toward Spectral Irradiance Time Series Throughout the 300-3000 nm Region. Jerald Harder, Juan Fontenla, Byron Smiley, Sean Davis, George Lawrence and Gary Rottman LASP

More information

Basic Components of Spectroscopic. Instrumentation

Basic Components of Spectroscopic. Instrumentation Basic Components of Spectroscopic Ahmad Aqel Ifseisi Assistant Professor of Analytical Chemistry College of Science, Department of Chemistry King Saud University P.O. Box 2455 Riyadh 11451 Saudi Arabia

More information

CONFIGURING. Your Spectroscopy System For PEAK PERFORMANCE. A guide to selecting the best Spectrometers, Sources, and Detectors for your application

CONFIGURING. Your Spectroscopy System For PEAK PERFORMANCE. A guide to selecting the best Spectrometers, Sources, and Detectors for your application CONFIGURING Your Spectroscopy System For PEAK PERFORMANCE A guide to selecting the best Spectrometers, s, and s for your application Spectral Measurement System Spectral Measurement System Spectrograph

More information

Voltage-dependent quantum efficiency measurements of amorphous silicon multijunction mini-modules

Voltage-dependent quantum efficiency measurements of amorphous silicon multijunction mini-modules Loughborough University Institutional Repository Voltage-dependent quantum efficiency measurements of amorphous silicon multijunction mini-modules This item was submitted to Loughborough University's Institutional

More information

High Illuminance Calibration Facility and Procedures

High Illuminance Calibration Facility and Procedures Final manuscript for J. IES, 27-2, 132-140 (1998) High Illuminance Calibration Facility and Procedures Yoshi Ohno Optical Technology Division National Institute of Standards and Technology Metrology A320,

More information

The University of Toledo R. Ellingson and M. Heben

The University of Toledo R. Ellingson and M. Heben focal length, f Spectral Measurement Using a Monochromator, Thermopile Detector, and Lock-In Amplifier September 18, 2012 The University of Toledo R. Ellingson and M. Heben Where are We, Where we are Going?

More information

Capabilities of NIST SIRCUS for Calibrations of SSI Vis-IR Instruments

Capabilities of NIST SIRCUS for Calibrations of SSI Vis-IR Instruments Capabilities of NIST SIRCUS for Calibrations of SSI Vis-IR Instruments Steve Brown National Institute of Standards & Technology Gaithersburg, MD steve.brown@nist.gov; 301.975.5167 Answer: Ask LASP folks

More information

Spectral responsivity uncertainty of silicon photodiodes due to calibration spectral bandwidth

Spectral responsivity uncertainty of silicon photodiodes due to calibration spectral bandwidth INSTITUTE OF PHYSICS PUBLISHING Meas. Sci. Technol. 12 (2001) 1926 1931 MEASUREMENT SCIENCE AND TECHNOLOGY Spectral responsivity uncertainty of silicon photodiodes due to calibration spectral bandwidth

More information

Project full title: "Nanowire based Tandem Solar Cells" Project acronym: Nano-Tandem Grant agreement no: Deliverable D6.1:

Project full title: Nanowire based Tandem Solar Cells Project acronym: Nano-Tandem Grant agreement no: Deliverable D6.1: Ref. Ares(2016)1038382-01/03/2016 Project full title: "Nanowire based Tandem Solar Cells" Project acronym: Nano-Tandem Grant agreement no: 641023 Deliverable D6.1: Report on adaption of EQE and IV measurement

More information

Spectroscopy in the UV and Visible: Instrumentation. Spectroscopy in the UV and Visible: Instrumentation

Spectroscopy in the UV and Visible: Instrumentation. Spectroscopy in the UV and Visible: Instrumentation Spectroscopy in the UV and Visible: Instrumentation Typical UV-VIS instrument 1 Source - Disperser Sample (Blank) Detector Readout Monitor the relative response of the sample signal to the blank Transmittance

More information

SPECTRORADIOMETRY METHODS: A GUIDE TO PHOTOMETRY AND VISIBLE SPECTRORADIOMETRY

SPECTRORADIOMETRY METHODS: A GUIDE TO PHOTOMETRY AND VISIBLE SPECTRORADIOMETRY SPECTRORADIOMETRY METHODS: A GUIDE TO PHOTOMETRY AND VISIBLE SPECTRORADIOMETRY Written By William E. Schneider, Richard Young, Ph.D 1. SPECTRORADIOMETRY METHODS 1.1. Spectroradiometrics vs Photometric

More information

TracQ. Basic Data Acquisition and Spectroscopy Software

TracQ. Basic Data Acquisition and Spectroscopy Software Basic Data Acquisition and Spectroscopy Software TracQ Basic main application window. Many common spectroscopic measurements require coordinated operation of a detection instrument and light source, as

More information

Powerful DMD-based light sources with a high throughput virtual slit Arsen R. Hajian* a, Ed Gooding a, Thomas Gunn a, Steven Bradbury a

Powerful DMD-based light sources with a high throughput virtual slit Arsen R. Hajian* a, Ed Gooding a, Thomas Gunn a, Steven Bradbury a Powerful DMD-based light sources with a high throughput virtual slit Arsen R. Hajian* a, Ed Gooding a, Thomas Gunn a, Steven Bradbury a a Hindsight Imaging Inc., 233 Harvard St. #316, Brookline MA 02446

More information

Working in Visible NHMFL

Working in Visible NHMFL Working in Visible Optics @ NHMFL NHMFL Summer School 05-19-2016 Stephen McGill Optical Energy Range Energy of Optical Spectroscopy Range SCM3 Optics Facility Energy Range of Optical Spectroscopy SCM3

More information

SCCH 4: 211: 2015 SCCH

SCCH 4: 211: 2015 SCCH SCCH 211: Analytical Chemistry I Analytical Techniques Based on Optical Spectroscopy Atitaya Siripinyanond Office Room: C218B Email: atitaya.sir@mahidol.ac.th Course Details October 19 November 30 Topic

More information

NIST EUVL Metrology Programs

NIST EUVL Metrology Programs NIST EUVL Metrology Programs S.Grantham, C. Tarrio, R.E. Vest, Y. Barad, S. Kulin, K. Liu and T.B. Lucatorto National Institute of Standards and Technology (NIST) Gaithersburg, MD USA L. Klebanoff and

More information

A High-Concentration Programmable Solar Simulator for Testing Multi-Junction Concentrator Photovoltaics

A High-Concentration Programmable Solar Simulator for Testing Multi-Junction Concentrator Photovoltaics A High-Concentration Programmable Solar Simulator for Testing ulti-junction Concentrator Photovoltaics Tasshi Dennis 1, Brent Fisher 2, att eitl 2, and John Wilson 2 1 National Institute of Standards and

More information

It s Our Business to be EXACT

It s Our Business to be EXACT 671 LASER WAVELENGTH METER It s Our Business to be EXACT For laser applications such as high-resolution laser spectroscopy, photo-chemistry, cooling/trapping, and optical remote sensing, wavelength information

More information

LINEARPYROMETER LP4. Technical Documentation KE November TN

LINEARPYROMETER LP4. Technical Documentation KE November TN 1 LINEARPYROMETER LP4 Technical Documentation KE 256-6.2007 November 2010 5-TN-1622-100 2 1. General Description With the Linearpyrometer Type LP4 a measuring instrument has been made available for pyrometric

More information

Developement of the method for realization of spectral irradiance scale featuring system of

Developement of the method for realization of spectral irradiance scale featuring system of Home Search Collections Journals About Contact us My IOPscience Developement of the method for realization of spectral irradiance scale featuring system of spectral comparisons This article has been downloaded

More information

Measuring photometric accuracy using the double aperture method

Measuring photometric accuracy using the double aperture method Measuring photometric accuracy using the double aperture method Application Note Author Robert Francis Agilent Technologies, Inc. Mulgrave, Victoria 3170, Australia. Introduction Photometric accuracy is

More information

Components of Optical Instruments

Components of Optical Instruments Components of Optical Instruments General Design of Optical Instruments Sources of Radiation Wavelength Selectors (Filters, Monochromators, Interferometers) Sample Containers Radiation Transducers (Detectors)

More information

GLOSSARY OF TERMS. Terminology Used for Ultraviolet (UV) Curing Process Design and Measurement

GLOSSARY OF TERMS. Terminology Used for Ultraviolet (UV) Curing Process Design and Measurement GLOSSARY OF TERMS Terminology Used for Ultraviolet (UV) Curing Process Design and Measurement This glossary of terms has been assembled in order to provide users, formulators, suppliers and researchers

More information

771 Series LASER SPECTRUM ANALYZER. The Power of Precision in Spectral Analysis. It's Our Business to be Exact! bristol-inst.com

771 Series LASER SPECTRUM ANALYZER. The Power of Precision in Spectral Analysis. It's Our Business to be Exact! bristol-inst.com 771 Series LASER SPECTRUM ANALYZER The Power of Precision in Spectral Analysis It's Our Business to be Exact! bristol-inst.com The 771 Series Laser Spectrum Analyzer combines proven Michelson interferometer

More information

Chemistry Instrumental Analysis Lecture 7. Chem 4631

Chemistry Instrumental Analysis Lecture 7. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 7 UV to IR Components of Optical Basic components of spectroscopic instruments: stable source of radiant energy transparent container to hold sample device

More information

Solar Cell Parameters and Equivalent Circuit

Solar Cell Parameters and Equivalent Circuit 9 Solar Cell Parameters and Equivalent Circuit 9.1 External solar cell parameters The main parameters that are used to characterise the performance of solar cells are the peak power P max, the short-circuit

More information

Disclosure to Promote the Right To Information

Disclosure to Promote the Right To Information इ टरन ट म नक Disclosure to Promote the Right To Information Whereas the Parliament of India has set out to provide a practical regime of right to information for citizens to secure access to information

More information

Systems & Accessories

Systems & Accessories Light Source Sample Chambers Stepping Motor Controller Detector Variable Wavelength Fiber Optics Modules Tunable Light Sources For Applications In: Analytical Chemistry Physics Life Sciences Engineering

More information

CHAPTER 7. Components of Optical Instruments

CHAPTER 7. Components of Optical Instruments CHAPTER 7 Components of Optical Instruments From: Principles of Instrumental Analysis, 6 th Edition, Holler, Skoog and Crouch. CMY 383 Dr Tim Laurens NB Optical in this case refers not only to the visible

More information

Integrating Spheres. Why an Integrating Sphere? High Reflectance. How Do Integrating Spheres Work? High Damage Threshold

Integrating Spheres. Why an Integrating Sphere? High Reflectance. How Do Integrating Spheres Work? High Damage Threshold 1354 MINIS Oriel Integrating Spheres Integrating spheres are ideal optical diffusers; they are used for radiometric measurements where uniform illumination or angular collection is essential, for reflectance

More information

DETECTING THE RATIO OF I AC

DETECTING THE RATIO OF I AC T E C H N O L O G Y F O R P O L A R I Z A T I O N M E A S U R E M E N T DETECTING THE RATIO OF I AC MEASUREMENT OF THE RAGE INTENSITY OF A MODULATED LIGHT BEAM In any experiment using photoelastic modulators

More information

Department of Electrical Engineering and Computer Science, Massachusetts Institute of Technology, 77. Table of Contents 1

Department of Electrical Engineering and Computer Science, Massachusetts Institute of Technology, 77. Table of Contents 1 Efficient single photon detection from 500 nm to 5 μm wavelength: Supporting Information F. Marsili 1, F. Bellei 1, F. Najafi 1, A. E. Dane 1, E. A. Dauler 2, R. J. Molnar 2, K. K. Berggren 1* 1 Department

More information

Infrared Detectors an overview

Infrared Detectors an overview Infrared Detectors an overview Mariangela Cestelli Guidi Sinbad IR beamline @ DaFne EDIT 2015, October 22 Frederick William Herschel (1738 1822) was born in Hanover, Germany but emigrated to Britain at

More information

Detecting the Ratio of I ac. /I ave. photoelastic modulators

Detecting the Ratio of I ac. /I ave. photoelastic modulators Measurement of the Average Intensity of a Modulated Light Beam In any experiment using (PEMs it is necessary to compare the time average intensity of the light at the detector with the amplitude of a single

More information

Optoelectronic Components Testing with a VNA(Vector Network Analyzer) VNA Roadshow Budapest 17/05/2016

Optoelectronic Components Testing with a VNA(Vector Network Analyzer) VNA Roadshow Budapest 17/05/2016 Optoelectronic Components Testing with a VNA(Vector Network Analyzer) VNA Roadshow Budapest 17/05/2016 Content Introduction Photonics & Optoelectronics components Optical Measurements VNA (Vector Network

More information

ECEN. Spectroscopy. Lab 8. copy. constituents HOMEWORK PR. Figure. 1. Layout of. of the

ECEN. Spectroscopy. Lab 8. copy. constituents HOMEWORK PR. Figure. 1. Layout of. of the ECEN 4606 Lab 8 Spectroscopy SUMMARY: ROBLEM 1: Pedrotti 3 12-10. In this lab, you will design, build and test an optical spectrum analyzer and use it for both absorption and emission spectroscopy. The

More information

Characterisation of Photovoltaic Materials and Cells

Characterisation of Photovoltaic Materials and Cells Standard Measurement Services and Prices Reference 1 Large area, 0.3-sun bias spectral response Wavelength measurement range: 300 1200 nm; Beam power monitoring and compensation; Measurement cell size:

More information

Components of Optical Instruments 1

Components of Optical Instruments 1 Components of Optical Instruments 1 Optical phenomena used for spectroscopic methods: (1) absorption (2) fluorescence (3) phosphorescence (4) scattering (5) emission (6) chemiluminescence Spectroscopic

More information

Modern Physics Laboratory MP4 Photoelectric Effect

Modern Physics Laboratory MP4 Photoelectric Effect Purpose MP4 Photoelectric Effect In this experiment, you will investigate the photoelectric effect and determine Planck s constant and the work function. Equipment and components Photoelectric Effect Apparatus

More information

DETECTORS Important characteristics: 1) Wavelength response 2) Quantum response how light is detected 3) Sensitivity 4) Frequency of response

DETECTORS Important characteristics: 1) Wavelength response 2) Quantum response how light is detected 3) Sensitivity 4) Frequency of response DETECTORS Important characteristics: 1) Wavelength response 2) Quantum response how light is detected 3) Sensitivity 4) Frequency of response (response time) 5) Stability 6) Cost 7) convenience Photoelectric

More information

Characterisation of Photovoltaic Materials and Cells

Characterisation of Photovoltaic Materials and Cells Standard Measurement Services and Prices No. Measurement Description Reference 1 Large area, 0.35-sun biased spectral response (SR) 2 Determination of linearity of spectral response with respect to irradiance

More information

New automated laser facility for detector calibrations

New automated laser facility for detector calibrations CORM annual conference, NRC, Ottawa, CANADA June 1, 2012 New automated laser facility for detector calibrations Yuqin Zong National Institute of Standards and Technology Gaithersburg, Maryland USA Overview

More information

Lasers PH 645/ OSE 645/ EE 613 Summer 2010 Section 1: T/Th 2:45-4:45 PM Engineering Building 240

Lasers PH 645/ OSE 645/ EE 613 Summer 2010 Section 1: T/Th 2:45-4:45 PM Engineering Building 240 Lasers PH 645/ OSE 645/ EE 613 Summer 2010 Section 1: T/Th 2:45-4:45 PM Engineering Building 240 John D. Williams, Ph.D. Department of Electrical and Computer Engineering 406 Optics Building - UAHuntsville,

More information

Properties of a Detector

Properties of a Detector Properties of a Detector Quantum Efficiency fraction of photons detected wavelength and spatially dependent Dynamic Range difference between lowest and highest measurable flux Linearity detection rate

More information

We bring quality to light. Spectro 320 Scanning Spectrometer

We bring quality to light. Spectro 320 Scanning Spectrometer We bring quality to light. Spectro 320 Scanning Spectrometer light measurement Instrument description Applications Spectroradiometry NVIS Display measurements Specifications Key features at a glance Fast-Scan

More information

Sensor Fusion Enables Comprehensive Analysis of Laser Processing in Additive Manufacturing

Sensor Fusion Enables Comprehensive Analysis of Laser Processing in Additive Manufacturing MKS Instruments 1 of 6 Sensor Fusion Enables Comprehensive Analysis of Laser Processing in Additive Manufacturing By Kevin Kirkham, Senior Manager, Product Development, Ophir Sensor: "A device that detects

More information

Irradiance Calibration Using a Cryogenic Radiometer and a Broadband Light Source

Irradiance Calibration Using a Cryogenic Radiometer and a Broadband Light Source Irradiance Calibration Using a Cryogenic Radiometer and a Broadband Light Source Jeff Morrill (1), Donald McMullin (2), Linton Floyd (3), Steven Lorentz (4), and Clarence Korendyke (1) (1) Naval Research

More information

Measurements of Infrared Sources with the Missile Defense Transfer Radiometer

Measurements of Infrared Sources with the Missile Defense Transfer Radiometer Measurements of Infrared Sources with the Missile Defense Transfer Radiometer Simon G. Kaplan #, Solomon I. Woods #, Adriaan C. Carter, and Timothy M. Jung * # National Institute of Standards and Technology

More information

SPECTRAL IRRADIANCE DATA

SPECTRAL IRRADIANCE DATA The radiometric data on the following pages was measured in our Standards Laboratory. The wavelength calibrations are based on our spectral calibration lamps. Irradiance data from 250 to 2500 nm is based

More information

Development of 2 Total Spectral Radiant Flux Standards at NIST

Development of 2 Total Spectral Radiant Flux Standards at NIST CIE/USA Annual Conference, October 7, 2014, Seattle, WA Development of 2 Total Spectral Radiant Flux Standards at NIST Yuqin Zong National Institute of Standards and Technology Gaithersburg, Maryland Outline

More information

Commissioning of Thomson Scattering on the Pegasus Toroidal Experiment

Commissioning of Thomson Scattering on the Pegasus Toroidal Experiment Commissioning of Thomson Scattering on the Pegasus Toroidal Experiment D.J. Schlossberg, R.J. Fonck, L.M. Peguero, G.R. Winz University of Wisconsin-Madison 55 th Annual Meeting of the APS Division of

More information

Optical Power Meter Basics

Optical Power Meter Basics Optical Power Meter Basics Introduction An optical power meter measures the photon energy in the form of current or voltage from an optical detector such as a semiconductor, a thermopile, or a pyroelectric

More information

Applications of Steady-state Multichannel Spectroscopy in the Visible and NIR Spectral Region

Applications of Steady-state Multichannel Spectroscopy in the Visible and NIR Spectral Region Feature Article JY Division I nformation Optical Spectroscopy Applications of Steady-state Multichannel Spectroscopy in the Visible and NIR Spectral Region Raymond Pini, Salvatore Atzeni Abstract Multichannel

More information

Advanced Test Equipment Rentals ATEC (2832) EDFA Testing with the Interpolation Technique Product Note

Advanced Test Equipment Rentals ATEC (2832) EDFA Testing with the Interpolation Technique Product Note Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) EDFA Testing with the Interpolation Technique Product Note 71452-1 Agilent 71452B Optical Spectrum Analyzer Table of

More information

Radiometric Solar Telescope (RaST) The case for a Radiometric Solar Imager,

Radiometric Solar Telescope (RaST) The case for a Radiometric Solar Imager, SORCE Science Meeting 29 January 2014 Mark Rast Laboratory for Atmospheric and Space Physics University of Colorado, Boulder Radiometric Solar Telescope (RaST) The case for a Radiometric Solar Imager,

More information

CHAPTER 5 FINE-TUNING OF AN ECDL WITH AN INTRACAVITY LIQUID CRYSTAL ELEMENT

CHAPTER 5 FINE-TUNING OF AN ECDL WITH AN INTRACAVITY LIQUID CRYSTAL ELEMENT CHAPTER 5 FINE-TUNING OF AN ECDL WITH AN INTRACAVITY LIQUID CRYSTAL ELEMENT In this chapter, the experimental results for fine-tuning of the laser wavelength with an intracavity liquid crystal element

More information

Coherent Laser Measurement and Control Beam Diagnostics

Coherent Laser Measurement and Control Beam Diagnostics Coherent Laser Measurement and Control M 2 Propagation Analyzer Measurement and display of CW laser divergence, M 2 (or k) and astigmatism sizes 0.2 mm to 25 mm Wavelengths from 220 nm to 15 µm Determination

More information

Experimental Analysis of Luminescence in Printed Materials

Experimental Analysis of Luminescence in Printed Materials Experimental Analysis of Luminescence in Printed Materials A. D. McGrath, S. M. Vaezi-Nejad Abstract - This paper is based on a printing industry research project nearing completion [1]. While luminescent

More information

Maya2000 Pro Spectrometer

Maya2000 Pro Spectrometer now with triggering! Maya2000 Pro Our Maya2000 Pro Spectrometer offers you the perfect solution for applications that demand low light-level, UV-sensitive operation. This back-thinned, 2D FFT-CCD, uncooled

More information

Components of Optical Instruments. Chapter 7_III UV, Visible and IR Instruments

Components of Optical Instruments. Chapter 7_III UV, Visible and IR Instruments Components of Optical Instruments Chapter 7_III UV, Visible and IR Instruments 1 Grating Monochromators Principle of operation: Diffraction Diffraction sources: grooves on a reflecting surface Fabrication:

More information

LBIR Fluid Bath Blackbody for Cryogenic Vacuum Calibrations

LBIR Fluid Bath Blackbody for Cryogenic Vacuum Calibrations LBIR Fluid Bath Blackbody for Cryogenic Vacuum Calibrations Timothy M. Jung*, Adriaan C. Carter*, Dale R. Sears*, Solomon I. Woods #, Dana R. Defibaugh #, Simon G. Kaplan #, Jinan Zeng * Jung Research

More information

3550 Aberdeen Ave SE, Kirtland AFB, NM 87117, USA ABSTRACT 1. INTRODUCTION

3550 Aberdeen Ave SE, Kirtland AFB, NM 87117, USA ABSTRACT 1. INTRODUCTION Beam Combination of Multiple Vertical External Cavity Surface Emitting Lasers via Volume Bragg Gratings Chunte A. Lu* a, William P. Roach a, Genesh Balakrishnan b, Alexander R. Albrecht b, Jerome V. Moloney

More information

Doppler-Free Spetroscopy of Rubidium

Doppler-Free Spetroscopy of Rubidium Doppler-Free Spetroscopy of Rubidium Pranjal Vachaspati, Sabrina Pasterski MIT Department of Physics (Dated: April 17, 2013) We present a technique for spectroscopy of rubidium that eliminates doppler

More information

Aqualog. CDOM Measurements Made Easy PARTICLE CHARACTERIZATION ELEMENTAL ANALYSIS FLUORESCENCE GRATINGS & OEM SPECTROMETERS OPTICAL COMPONENTS RAMAN

Aqualog. CDOM Measurements Made Easy PARTICLE CHARACTERIZATION ELEMENTAL ANALYSIS FLUORESCENCE GRATINGS & OEM SPECTROMETERS OPTICAL COMPONENTS RAMAN Aqualog CDOM Measurements Made Easy ELEMENTAL ANALYSIS FLUORESCENCE GRATINGS & OEM SPECTROMETERS OPTICAL COMPONENTS PARTICLE CHARACTERIZATION RAMAN SPECTROSCOPIC ELLIPSOMETRY SPR IMAGING CDOM measurements

More information

Improved Radiometry for LED Arrays

Improved Radiometry for LED Arrays RadTech Europe 2017 Prague, Czech Republic Oct. 18, 2017 Improved Radiometry for LED Arrays Dr. Robin E. Wright 3M Corporate Research Process Laboratory, retired 3M 2017 All Rights Reserved. 1 Personal

More information

NON-AMPLIFIED HIGH SPEED PHOTODETECTOR USER S GUIDE

NON-AMPLIFIED HIGH SPEED PHOTODETECTOR USER S GUIDE NON-AMPLIFIED HIGH SPEED PHOTODETECTOR USER S GUIDE Thank you for purchasing your Non-amplified High Speed Photodetector. This user s guide will help answer any questions you may have regarding the safe

More information

Silicon Photodiodes - SXUV Series with Platinum Silicide Front Entrance Windows

Silicon Photodiodes - SXUV Series with Platinum Silicide Front Entrance Windows Silicon Photodiodes - SXUV Series with Platinum Silicide Front Entrance Windows SXUV Responsivity Stability It is known that the UV photon exposure induced instability of common silicon photodiodes is

More information

TIME-PRESERVING MONOCHROMATORS FOR ULTRASHORT EXTREME-ULTRAVIOLET PULSES

TIME-PRESERVING MONOCHROMATORS FOR ULTRASHORT EXTREME-ULTRAVIOLET PULSES TIME-PRESERVING MONOCHROMATORS FOR ULTRASHORT EXTREME-ULTRAVIOLET PULSES Luca Poletto CNR - Institute of Photonics and Nanotechnologies Laboratory for UV and X-Ray Optical Research Padova, Italy e-mail:

More information

RADIOMETRIC AND PHOTOMETRIC MEASUREMENTS AT THE LNE-INM/CNAM

RADIOMETRIC AND PHOTOMETRIC MEASUREMENTS AT THE LNE-INM/CNAM RADIOMETRIC AND PHOTOMETRIC MEASUREMENTS AT THE LNE-INM/CNAM Jean Bastie, Laura Patricia Gonzalez Galvan. Institut National de Métrologie Conservatoire National des Arts et Métiers 61 rue du Landy 93210

More information

Coherent Receivers Principles Downconversion

Coherent Receivers Principles Downconversion Coherent Receivers Principles Downconversion Heterodyne receivers mix signals of different frequency; if two such signals are added together, they beat against each other. The resulting signal contains

More information

Spectrophotometric calibration system for DECam

Spectrophotometric calibration system for DECam Spectrophotometric calibration system for DECam J.-P. Rheault*, D. L. DePoy, T. W. Behm, E. W. Kylberg, Kris Cabral, Rick Allen, J. L. Marshall, Department of Physics and Astronomy, Texas A&M University,

More information

ISS-30-VA. Product tags: Integrating Sphere Source. https://www.gigahertz-optik.de/en-us/product/iss-30-va. Gigahertz-Optik GmbH 1/5

ISS-30-VA. Product tags: Integrating Sphere Source. https://www.gigahertz-optik.de/en-us/product/iss-30-va. Gigahertz-Optik GmbH 1/5 ISS-30-VA https://www.gigahertz-optik.de/en-us/product/iss-30-va Product tags: Integrating Sphere Source Gigahertz-Optik GmbH 1/5 Description standards for spectral radiance Spectroradiometers and other

More information

OL Series 426 Low Light Level lntegrating Sphere Calibration Standards

OL Series 426 Low Light Level lntegrating Sphere Calibration Standards OL Series 426 Low Light Level lntegrating Sphere Calibration Standards GENERAL The OL Series 426 Low-Light-Level Calibration Standard is designed for accurately calibrating very sensitive microphotometers,

More information

A Measurement of the Photon Detection Efficiency of Silicon Photomultipliers

A Measurement of the Photon Detection Efficiency of Silicon Photomultipliers A Measurement of the Photon Detection Efficiency of Silicon Photomultipliers A. N. Otte a,, J. Hose a,r.mirzoyan a, A. Romaszkiewicz a, M. Teshima a, A. Thea a,b a Max Planck Institute for Physics, Föhringer

More information

Part 1: New spectral stuff going on at NIST. Part 2: TSI Traceability of TRF to NIST

Part 1: New spectral stuff going on at NIST. Part 2: TSI Traceability of TRF to NIST Part 1: New spectral stuff going on at NIST SIRCUS-type stuff (tunable lasers) now migrating to LASP Absolute Spectrally-Tunable Detector-Based Source Spectrally-programmable source calibrated via NIST

More information

Optodevice Data Book ODE I. Rev.9 Mar Opnext Japan, Inc.

Optodevice Data Book ODE I. Rev.9 Mar Opnext Japan, Inc. Optodevice Data Book ODE-408-001I Rev.9 Mar. 2003 Opnext Japan, Inc. Section 1 Operating Principles 1.1 Operating Principles of Laser Diodes (LDs) and Infrared Emitting Diodes (IREDs) 1.1.1 Emitting Principles

More information

NON-AMPLIFIED PHOTODETECTOR USER S GUIDE

NON-AMPLIFIED PHOTODETECTOR USER S GUIDE NON-AMPLIFIED PHOTODETECTOR USER S GUIDE Thank you for purchasing your Non-amplified Photodetector. This user s guide will help answer any questions you may have regarding the safe use and optimal operation

More information

The TSIS Spectral Irradiance Monitor: Prism Optical Degradation Studies

The TSIS Spectral Irradiance Monitor: Prism Optical Degradation Studies The TSIS Spectral Irradiance Monitor: Prism Optical Degradation Studies Lo Erik Richard, Dave Harber, Joel Rutkowski, Matt Triplett, Kasandra O Malia Laboratory for Atmospheric and Space Physics (LASP)

More information

Photodiode: LECTURE-5

Photodiode: LECTURE-5 LECTURE-5 Photodiode: Photodiode consists of an intrinsic semiconductor sandwiched between two heavily doped p-type and n-type semiconductors as shown in Fig. 3.2.2. Sufficient reverse voltage is applied

More information

Single Photon Interference Laboratory

Single Photon Interference Laboratory Single Photon Interference Laboratory Renald Dore Institute of Optics University of Rochester, Rochester, NY 14627, U.S.A Abstract The purpose of our laboratories was to observe the wave-particle duality

More information

Performance Comparisons of InGaAs, extended InGaAs, and Short-wave HgCdTe Detectors between 1 µm and 2.5 µm

Performance Comparisons of InGaAs, extended InGaAs, and Short-wave HgCdTe Detectors between 1 µm and 2.5 µm Performance Comparisons of, extended, and Short-wave HgCdTe Detectors between 1 µm and 2.5 µm Howard W. Yoon, Matt C. Dopkiss, and George P. Eppeldauer Optical Technology Division National Institute of

More information

Oriel Cornerstone 130 1/8 m Monochromator

Oriel Cornerstone 130 1/8 m Monochromator 1 Oriel Cornerstone 130 1/8 m Monochromator Cornerstone 130 1/8 m Monochromator The Cornerstone 130 family of Oriel Monochromators supports two gratings simultaneously, which can be easily interchanged,

More information

Summary of Laser Stepping Apparatus

Summary of Laser Stepping Apparatus Overview Summary of Laser Stepping Apparatus Jason Hiltner, CSU Photovoltaic Lab August 6, 1999 The laser stepping apparatus was constructed to measure the spatial uniformity of solar cell photo-response

More information