1 CORM annual conference, NRC, Ottawa, CANADA June 1, 2012 New automated laser facility for detector calibrations Yuqin Zong National Institute of Standards and Technology Gaithersburg, Maryland USA

2 Overview Introduction Existing spectral calibration facilities New 1 khz optical parametric oscillator (OPO) based calibration system Conclusion

3 Spectral calibration of optical sensors Light source (Watt) Detector i o (λ) (Ampere) Spectral POWER responsivity (Underfilled geometry) (Ampere/Watt) Light source (Watt.m -2 ) i o (λ) (Ampere) SI base unit -luminous intensity: candela SI base unit radiance temperature: Kelvin Colorimetry and radiometry Spectral irradiance scale (FEL lamps) Spectral IRRADIANCE responsivity (Overfilled geometry) (Ampere/(Watt.m -2 ))

4 Existing NIST spectral calibration facilities NIST POWR Cryogenic Radiometer (Power resp.) (0.01 %) NIST SIRCUS facility Trap detectors (Irrad resp.) (0.025 %)* NIST SCF Trap detectors (Irrad resp.) (0.025 %)* Traps to photodiodes Irradiance responsivity (Ampere/(Watt.m -2 )) ( 0.05 %)* NIST SCF Working std. photodiodes (Power resp.) * BEST expanded uncertainty (k=2) POWR: primary optical watt radiometer SIRCUS: Spectral irradiance and radiance responsivity calibrations using uniform sources SCF: Spectral comparator facility Trap detector: specially configured, multi-element silicon photodiodes detector with high performance. Power resp. (Ampere/Watt) ( 0.20 %)*

5 The NIST SIRCUS facility Standard Trap Detector integrating sphere Test Detector Continuous spectral coverage from UV to NIR Continuous wave (CW) or quasi CW tunable lasers based research facility high power (e.g., 100 mw), narrow bandwidth (<0.01 nm) Used for realization of SI base units: Kelvin and candela Provide calibrations for primary radiometric standards and for remote sensing instruments Difficult to automate & high-cost

6 NIST SCF Double monochromator Working standard detector Fully automated Lamp-monochromator based calibration facility, no fringe problem Major facility to disseminate NIST Scale to industry Low radiant power (µw level), broad bandwidth (4 nm) Designed for power responsivity Large uncertainties to acquire irradiance responsivities (mapping method does not work well!) Test Detector

7 Spatial uniformity of a Si photodiode 205 nm 230 nm 250 nm Acknowledgement to Ping-shine Shaw, NIST Y (mm) 6 4 Y (mm) 6 4 Y (mm) X (mm) X (mm) X (mm) nm nm nm Y (mm) 6 4 Y (mm) 6 4 Y (mm) X (mm) X (mm) X (mm)

8 Fully automated tunable OPO-based laser sources OPO: optical parametric oscillators Fully automated Large tunable range Portable Much lower cost pump mirror Low repetition rate (10 Hz to 1000 Hz) Narrow pulse width, extremely low duty cycle (e.g., 10-6 ) Pulse to pulse variation and difficult to stabilize Trans-impedence amplifiers don t work well Have not been used as calibration source yet nonlinear crystal mirror

9 Key questions Can pulse lasers be used for calibration of detectors with small uncertainties? How to overcome fluctuation of a pulsed laser and obtain repeatable results? Will detectors be saturated? Is a pulse laser equivalent to a CW laser for detector calibrations?

10 Schematic of the new automated calibration system integrating sphere baffle Monitor Detector Standard Trap Detector Electrometers (2) Optical fiber Ultrasound bath shutter Test Detector 1000 Hz OPO Computer Shutter controller R test ( λ) = R ( λ) Q standard M test ( λ) / Q M standard ( λ)

11 The automated OPO system Average power (mw) Wavelength, λ (nm) 210 nm to 2400 nm tunable range 1 khz repetition rate 5 ns pulse width 5 8 cm -1 bandwidth ( 0.2 nm in visible range)

12 OPO Pulse waveforms Rel. laser power OPO laser Rel. laser power After 5 m MM fiber Time, t (ns) Time, t (ns) Rel. laser power After 50 mm sphere Time, t (ns)

13 OPO spectra nm 0.8 Relative SPD Wavelength, λ (nm) nm nm Relative SPD Relative SPD Wavelength, λ (nm) Wavelength, λ (nm)

14 The electrometers Pulse train i(t) C T Q = i( t) dt = C V 0 Detector - + V DMM Charge measurement function from 2 nc to 2 μc using a charge amplifier < 3 fa bias current < 20 μv burden voltage High performance multichannel switching card

15 Measurement Timing Electrometer s synchronized charge measurement start end Laser shutter open close

16 Measurement repeatability 1 s integration time for each point Relative charge ratio Relative charge ratio Measurement No., i Measurement No., i two Hamamatus S2281 silicon photodiodes (PD) standard deviation = 7 ppm! one 3 silicon PD trap and one S2281 Si PD standard deviation = 12 ppm!

17 Detector non-linearity test Relative responsivity Obtained by normalizing the charge ratio r(p i ) of the test detector (S2281 PD) to reference detector (S2281 PD with 2 orders of magnitude lower signal) E E E E E-05 Averaged photocurrent, I (A) OPO at 450 nm. Saturation starts at peak=100 ma, averaged=1 µa. 1) Nonlinearity depends on the detector and laser wavelength. 2) The instantaneous photocurrent without causing nonlinearity is several orders of magnitude higher than the threshold nonlinear DC photocurrent (0.1 1 ma typically). 3) The level of allowed averaged photocurrent is several orders of magnitude lower than the threshold nonlinear DC photocurrent.

18 Validation: charge amp vs trans-impedance amp Relative Difference ΔR(λ) (%) (a) charge amplifiers vs. trans-impedance amplifiers Wavelength, λ /nm CW laser + charge amplifiers vs. CW laser + trans-impedance amplifiers Difference in measured responsivity is 0.02 %.

19 Validation: 1 khz pulsed OPO vs CW lasers Relative Difference ΔR(λ) (%) (b) pulsed OPO vs. CW laser Wavelength, λ /nm Pulsed OPO + charge amplifiers vs. CW laser + trans-impedance amplifiers Difference in measured responsivity is also 0.02 %

20 Comparison of Results Relative Difference ΔR(λ) (%) (a) charge amplifiers vs. trans-impedance amplifiers (b) pulsed OPO vs. CW laser Wavelength, λ (nm) Replacing CW laser with pulsed OPO for charge amplifiers does not make difference in measured responsivity. Pulsed OPO CW laser.

21 Uncertainty budget Relative standard unc. (%) Uncertainty component Type A Type B Reference trap detector OPO wavelength (0.02 nm) Sphere source irradiance non-uniformity Detector reference plane Detector non-linearity Transfer to test detector Electrometer (range to range gain error) Combined uncertainty (%) Expanded uncertainty (k=2) (%)

22 Conclusions A fully automated laser facility using a khz pulsed OPO for calibration of detectors and instruments has been developed and validated. The estimated uncertainty is 0.05 % (k=2). The new facility is to be used for calibration of photometers, colorimeters, and spectroradiometers with significantly reduced uncertainty. The developed method may be used in other applications: - transmittance and reflectance - surface color and appearance - optical medical imaging -

23 Acknowledgements Keith Lykke Steve Brown George Eppeldauer Yoshi Ohno and many other colleagues

24 THANK YOU

Development of 2 Total Spectral Radiant Flux Standards at NIST

Development of 2 Total Spectral Radiant Flux Standards at NIST CIE/USA Annual Conference, October 7, 2014, Seattle, WA Development of 2 Total Spectral Radiant Flux Standards at NIST Yuqin Zong National Institute of Standards and Technology Gaithersburg, Maryland Outline

More information

Capabilities of NIST SIRCUS for Calibrations of SSI Vis-IR Instruments

Capabilities of NIST SIRCUS for Calibrations of SSI Vis-IR Instruments Capabilities of NIST SIRCUS for Calibrations of SSI Vis-IR Instruments Steve Brown National Institute of Standards & Technology Gaithersburg, MD steve.brown@nist.gov; 301.975.5167 Answer: Ask LASP folks

More information

Part 1: New spectral stuff going on at NIST. Part 2: TSI Traceability of TRF to NIST

Part 1: New spectral stuff going on at NIST. Part 2: TSI Traceability of TRF to NIST Part 1: New spectral stuff going on at NIST SIRCUS-type stuff (tunable lasers) now migrating to LASP Absolute Spectrally-Tunable Detector-Based Source Spectrally-programmable source calibrated via NIST

More information

Current and Future Realizations NRC Photometric and Spectroradiometric Calibration Chains

Current and Future Realizations NRC Photometric and Spectroradiometric Calibration Chains Measurement Science and Standards (MSS) Current and Future Realizations NRC Photometric and Spectroradiometric Calibration Chains Arnold A. Gaertner Photometry, Radiometry and Thermometry (PRT) CORM 30

More information

NIST Agency Report May 2012 OUTLINE. The case for traceability NMI capabilities A view to the future the HIP Current/recent NIST activities

NIST Agency Report May 2012 OUTLINE. The case for traceability NMI capabilities A view to the future the HIP Current/recent NIST activities NIST Agency Report May 2012 OUTLINE The case for traceability NMI capabilities A view to the future the HIP Current/recent NIST activities The case for traceability Earth Radiation Budget: Solar irradiance

More information

Results of J1 VIIRS testing using NIST s Traveling SIRCUS

Results of J1 VIIRS testing using NIST s Traveling SIRCUS Results of J1 VIIRS testing using NIST s Traveling SIRCUS Steven W. Brown and Keith R. Lykke, NIST, Gaithersburg, MD Joel McCorkel, Brendan McAndrew, Gene Waluschka, and Jeff McIntire (Sigma Space), NASA

More information

JPSS1 VIIRS RSB Sensor calibration using monochromator-based and laser-based methods

JPSS1 VIIRS RSB Sensor calibration using monochromator-based and laser-based methods JPSS1 VIIRS RSB Sensor calibration using monochromator-based and laser-based methods Jinan Zeng 1, Tom Schwarting 2, Jeff McIntire 2, Jack Ji 2, Hassan Oudrari 2, Jack Xiong 3, and Jim Butler 3 1 Fibertek

More information

TRACEABLE SOLAR RADIATION MEASUREMENTS

TRACEABLE SOLAR RADIATION MEASUREMENTS TRACEABLE SOLAR RADIATION MEASUREMENTS Julian Gröbner, Natalia Kouremeti, Gregor Hülsen, Stelios Kazadzis, Luca Egli Physikalisch-Meteorologisches Observatorium Davos, World Radiation Center (PMOD/WRC)

More information

High Illuminance Calibration Facility and Procedures

High Illuminance Calibration Facility and Procedures Final manuscript for J. IES, 27-2, 132-140 (1998) High Illuminance Calibration Facility and Procedures Yoshi Ohno Optical Technology Division National Institute of Standards and Technology Metrology A320,

More information

NIST MEASUREMENT SERVICES:

NIST MEASUREMENT SERVICES: NIST Special Publication 250-37 NIST MEASUREMENT SERVICES: PHOTOMETRIC CALIBRATIONS Yoshihiro Ohno Optical Technology Division Physics Laboratory National Institute of Standards and Technology Gaithersburg,

More information

P-CUBE-Series High Sensitivity PIN Detector Modules

P-CUBE-Series High Sensitivity PIN Detector Modules High Sensitivity PIN Detector Modules Description The P-CUBE-series manufactured by LASER COMPONENTS has been designed for customers interested in experimenting with low noise silicon or InGaAs pin detectors.

More information

CCT/10-13 MeP-K direct methods. Section 3.2: Absolute (spectral-band) radiometry (radiation thermometry)

CCT/10-13 MeP-K direct methods. Section 3.2: Absolute (spectral-band) radiometry (radiation thermometry) CCT/10-13 MeP-K direct methods Section 3.2: Absolute (spectral-band) radiometry (radiation thermometry) Authors: Graham Machin (NPL, chair), Klaus Anhalt (PTB), Pieter Bloembergen (NIM, formerly NMIJ &

More information

Integrating Spheres. Why an Integrating Sphere? High Reflectance. How Do Integrating Spheres Work? High Damage Threshold

Integrating Spheres. Why an Integrating Sphere? High Reflectance. How Do Integrating Spheres Work? High Damage Threshold 1354 MINIS Oriel Integrating Spheres Integrating spheres are ideal optical diffusers; they are used for radiometric measurements where uniform illumination or angular collection is essential, for reflectance

More information

The Art of Light Measurement. Avantes BV Apeldoorn, The Netherlands

The Art of Light Measurement. Avantes BV Apeldoorn, The Netherlands The Art of Light Measurement Avantes BV Apeldoorn, The Netherlands Who am I? Ger Loop Product Manager Avantes BV Oude Apeldoornseweg 28 7333 NS APELDOORN The Netherlands Phone: (+31) 313 670170 Fax: (+31)

More information

Solid State Luminance Standards

Solid State Luminance Standards Solid State Luminance Standards Color and luminance correction of: - Imaging colorimeters - Luminance meters - Imaging spectrometers Compact and Robust for Production Environments Correct for instrument

More information

OL Series 426 Low Light Level lntegrating Sphere Calibration Standards

OL Series 426 Low Light Level lntegrating Sphere Calibration Standards OL Series 426 Low Light Level lntegrating Sphere Calibration Standards GENERAL The OL Series 426 Low-Light-Level Calibration Standard is designed for accurately calibrating very sensitive microphotometers,

More information

Radiometric Measurement Traceability Paths for Photovoltaic Calibrations. Howard W. Yoon Physical Measurement Laboratory NIST

Radiometric Measurement Traceability Paths for Photovoltaic Calibrations. Howard W. Yoon Physical Measurement Laboratory NIST Radiometric Measurement Traceability Paths for Photovoltaic Calibrations Howard W. Yoon Physical Measurement Laboratory NIST Solar energy and PV Solar radiation: free and abundant! Photovoltaics (PV):

More information

A stray light corrected array spectroradiometer for complex high dynamic range measurements in the UV spectral range.

A stray light corrected array spectroradiometer for complex high dynamic range measurements in the UV spectral range. A stray light corrected array spectroradiometer for complex high dynamic range measurements in the UV spectral range Mike Clark Gigahertz-Optik GmbH m.clark@gigahertz-optik.de Array spectroradiometers

More information

Pulsed Laser Power Measurement Systems

Pulsed Laser Power Measurement Systems Pulsed Laser Power Measurement Systems Accurate, reproducible method of determining total laser and laser diode power Ideal for Beam Power Measurement Labsphere s Pulsed Laser Power Measurement Systems

More information

Silicon Photodiodes - SXUV Series with Platinum Silicide Front Entrance Windows

Silicon Photodiodes - SXUV Series with Platinum Silicide Front Entrance Windows Silicon Photodiodes - SXUV Series with Platinum Silicide Front Entrance Windows SXUV Responsivity Stability It is known that the UV photon exposure induced instability of common silicon photodiodes is

More information

\Ç à{x ÇtÅx Éy ALLAH à{x `xüv yâä

\Ç à{x ÇtÅx Éy ALLAH à{x `xüv yâä \Ç à{x ÇtÅx Éy ALLAH à{x `xüv yâä Ultraviolet Radiation from Some Types of Outdoor Lighting Lamps Dr.Essam El-Moghazy Photometry and Radiometry division, National Institute for Standards (NIS), Egypt.

More information

BTS2048-UV. Product tags: UV, Spectral Data, LED Binning, Industrial Applications, LED. https://www.gigahertz-optik.de/en-us/product/bts2048-uv

BTS2048-UV. Product tags: UV, Spectral Data, LED Binning, Industrial Applications, LED. https://www.gigahertz-optik.de/en-us/product/bts2048-uv BTS2048-UV https://www.gigahertz-optik.de/en-us/product/bts2048-uv Product tags: UV, Spectral Data, LED Binning, Industrial Applications, LED Gigahertz-Optik GmbH 1/8 Description UV CCD spectroradiometer

More information

NON-AMPLIFIED PHOTODETECTOR USER S GUIDE

NON-AMPLIFIED PHOTODETECTOR USER S GUIDE NON-AMPLIFIED PHOTODETECTOR USER S GUIDE Thank you for purchasing your Non-amplified Photodetector. This user s guide will help answer any questions you may have regarding the safe use and optimal operation

More information

RADIOMETRIC AND PHOTOMETRIC MEASUREMENTS AT THE LNE-INM/CNAM

RADIOMETRIC AND PHOTOMETRIC MEASUREMENTS AT THE LNE-INM/CNAM RADIOMETRIC AND PHOTOMETRIC MEASUREMENTS AT THE LNE-INM/CNAM Jean Bastie, Laura Patricia Gonzalez Galvan. Institut National de Métrologie Conservatoire National des Arts et Métiers 61 rue du Landy 93210

More information

A CW seeded femtosecond optical parametric amplifier

A CW seeded femtosecond optical parametric amplifier Science in China Ser. G Physics, Mechanics & Astronomy 2004 Vol.47 No.6 767 772 767 A CW seeded femtosecond optical parametric amplifier ZHU Heyuan, XU Guang, WANG Tao, QIAN Liejia & FAN Dianyuan State

More information

2008 Stray Light Correction Work

2008 Stray Light Correction Work 2008 Stray Light Correction Work MLML Presenter: Stephanie Flora MLML: Michael Feinholz, Mark Yarbrough NIST: Carol Johnson, Steve Brown, Keith Lykke, Al Parr, Dennis Clark, Eric Shirley, Bob Saunders

More information

NON-AMPLIFIED HIGH SPEED PHOTODETECTOR USER S GUIDE

NON-AMPLIFIED HIGH SPEED PHOTODETECTOR USER S GUIDE NON-AMPLIFIED HIGH SPEED PHOTODETECTOR USER S GUIDE Thank you for purchasing your Non-amplified High Speed Photodetector. This user s guide will help answer any questions you may have regarding the safe

More information

Influence of the light spectral distribution used in the radiometers calibration

Influence of the light spectral distribution used in the radiometers calibration Influence of the light spectral distribution used in the radiometers calibration J C Moraes and L O Guerra Allergisa Pesquisa Dermato-Cosmética Ltda GRUPO INVESTIGA, 452. Dr. Romeu Tórtima Avenue, Postcode:

More information

Optoelectronic Components Testing with a VNA(Vector Network Analyzer) VNA Roadshow Budapest 17/05/2016

Optoelectronic Components Testing with a VNA(Vector Network Analyzer) VNA Roadshow Budapest 17/05/2016 Optoelectronic Components Testing with a VNA(Vector Network Analyzer) VNA Roadshow Budapest 17/05/2016 Content Introduction Photonics & Optoelectronics components Optical Measurements VNA (Vector Network

More information

Testing with Femtosecond Pulses

Testing with Femtosecond Pulses Testing with Femtosecond Pulses White Paper PN 200-0200-00 Revision 1.3 January 2009 Calmar Laser, Inc www.calmarlaser.com Overview Calmar s femtosecond laser sources are passively mode-locked fiber lasers.

More information

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK 21-23 Campus Road Contact: Mr K G Dove Listerhills Science Park Tel: +44 (0)1274 393857 Bradford Fax: +44 (0)1274 393336 BD7 1HR E-Mail: sales@otc.co.uk

More information

Red Laser for Monitoring Light Source

Red Laser for Monitoring Light Source Red Laser for Monitoring Light Source Liyuan Zhang, Kejun Zhu and Ren-yuan Zhu Caltech Duncan Liu JPL CMS ECAL Week, CERN April 16, 22 A Brief History. Red Laser Specification. Result of Market Survey.

More information

Optical Communications

Optical Communications Optical Communications Telecommunication Engineering School of Engineering University of Rome La Sapienza Rome, Italy 2005-2006 Lecture #4, May 9 2006 Receivers OVERVIEW Photodetector types: Photodiodes

More information

Certificate of Accreditation

Certificate of Accreditation PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of:, Inc. (Hereinafter called the Organization) and hereby

More information

Measurements of Infrared Sources with the Missile Defense Transfer Radiometer

Measurements of Infrared Sources with the Missile Defense Transfer Radiometer Measurements of Infrared Sources with the Missile Defense Transfer Radiometer Simon G. Kaplan #, Solomon I. Woods #, Adriaan C. Carter, and Timothy M. Jung * # National Institute of Standards and Technology

More information

Model S471-LED Optometer

Model S471-LED Optometer Model S471-LED Optometer For over 40 years UDT Instruments, a Gamma Scientific company, has been trusted by the world s leading organizations to provide accurate light measurement systems. UDT Instruments

More information

Status of MOLI development MOLI (Multi-footprint Observation Lidar and Imager)

Status of MOLI development MOLI (Multi-footprint Observation Lidar and Imager) Status of MOLI development MOLI (Multi-footprint Observation Lidar and Imager) Tadashi IMAI, Daisuke SAKAIZAWA, Jumpei MUROOKA and Toshiyoshi KIMURA JAXA 1 Outline of This Presentation 1. Overview of MOLI

More information

MDS-3 EVALUATION SYSTEM FOR METHANE DETECTION INSTRUCTION MANUAL

MDS-3 EVALUATION SYSTEM FOR METHANE DETECTION INSTRUCTION MANUAL MDS-3 EVALUATION SYSTEM FOR METHANE DETECTION INSTRUCTION MANUAL rev. 281014 TABLE OF CONTENTS General Information 3 Application 3 Packaging arrangement 3 Operation conditions 3 Brief Overview of the Components

More information

UNMATCHED OUTPUT POWER AND TUNING RANGE

UNMATCHED OUTPUT POWER AND TUNING RANGE ARGOS MODEL 2400 SF SERIES TUNABLE SINGLE-FREQUENCY MID-INFRARED SPECTROSCOPIC SOURCE UNMATCHED OUTPUT POWER AND TUNING RANGE One of Lockheed Martin s innovative laser solutions, Argos TM Model 2400 is

More information

NFMS THEORY LIGHT AND COLOR MEASUREMENTS AND THE CCD-BASED GONIOPHOTOMETER. Presented by: January, 2015 S E E T H E D I F F E R E N C E

NFMS THEORY LIGHT AND COLOR MEASUREMENTS AND THE CCD-BASED GONIOPHOTOMETER. Presented by: January, 2015 S E E T H E D I F F E R E N C E NFMS THEORY LIGHT AND COLOR MEASUREMENTS AND THE CCD-BASED GONIOPHOTOMETER Presented by: January, 2015 1 NFMS THEORY AND OVERVIEW Contents Light and Color Theory Light, Spectral Power Distributions, and

More information

Improved Radiometry for LED Arrays

Improved Radiometry for LED Arrays RadTech Europe 2017 Prague, Czech Republic Oct. 18, 2017 Improved Radiometry for LED Arrays Dr. Robin E. Wright 3M Corporate Research Process Laboratory, retired 3M 2017 All Rights Reserved. 1 Personal

More information

Measurement of normalized spectral responsivity of digital imaging devices by using a LED-based tunable uniform source

Measurement of normalized spectral responsivity of digital imaging devices by using a LED-based tunable uniform source See discussions, stats, and author profiles for this publication at: https://www.researchgate.net/publication/235716439 Measurement of normalized spectral responsivity of digital imaging devices by using

More information

Unprecedented Resolution and Accuracy For Camera & Image Sensor Calibration

Unprecedented Resolution and Accuracy For Camera & Image Sensor Calibration SpectralLED RS-7-1-VIS Benchtop Uniform Light Source For the ultimate in resolution and accuracy, the SpectralLED Tunable VIS source incorporates 35 discrete wavelengths for synthesis of commercially available

More information

Program for UV Intercomparison 2014 in Davos:

Program for UV Intercomparison 2014 in Davos: Program for UV Intercomparison 2014 in Davos: June 2014 Date: 7 16 July 2014 Location: PMOD/WRC Davos Switzerland. Information Update: http://projects.pmodwrc.ch/env03/index.php/8-emrp-uv/project/24- intercomparison-2014

More information

Technical Notes. Integrating Sphere Measurement Part II: Calibration. Introduction. Calibration

Technical Notes. Integrating Sphere Measurement Part II: Calibration. Introduction. Calibration Technical Notes Integrating Sphere Measurement Part II: Calibration This Technical Note is Part II in a three part series examining the proper maintenance and use of integrating sphere light measurement

More information

Mira OPO-X. Fully Automated IR/Visible OPO for femtosecond and picosecond Ti:Sapphire Lasers. Superior Reliability & Performance. Mira OPO-X Features:

Mira OPO-X. Fully Automated IR/Visible OPO for femtosecond and picosecond Ti:Sapphire Lasers. Superior Reliability & Performance. Mira OPO-X Features: Fully Automated IR/Visible OPO for femtosecond and picosecond Ti:Sapphire Lasers Mira OPO-X is a synchronously pumped, widely tunable, optical parametric oscillator (OPO) accessory that dramatically extends

More information

Spectroradiometer characterisations for traceable solar radiation measurements

Spectroradiometer characterisations for traceable solar radiation measurements Spectroradiometer characterisations for traceable solar radiation measurements Julian Gröbner, Natalia Kouremeti, Stelios Kazadzis, Gregor Hülsen, Luca Egli Physikalisch-Meteorologisches Observatorium

More information

Photometry and Light Measurement

Photometry and Light Measurement Photometry and Light Measurement Adrian Waltho, Analytik Ltd adrian.waltho@analytik.co.uk What is Light? What is Light? What is Light? Ultraviolet Light UV-C 180-280 nm UV-B 280-315 nm UV-A 315-400 nm

More information

Measurement Services Optics Laboratory

Measurement Services Optics Laboratory Federal Institute of Metrology METAS CH-3003 Bern-Wabern, 1. February 2018 Measurement Services Optics Laboratory Valid from: 01.02.2018 In our laboratory we perform high accuracy calibrations of your

More information

DIFFERENTIAL ABSORPTION LIDAR FOR GREENHOUSE GAS MEASUREMENTS

DIFFERENTIAL ABSORPTION LIDAR FOR GREENHOUSE GAS MEASUREMENTS DIFFERENTIAL ABSORPTION LIDAR FOR GREENHOUSE GAS MEASUREMENTS Stephen E. Maxwell, Sensor Science Division, PML Kevin O. Douglass, David F. Plusquellic, Radiation and Biomolecular Physics Division, PML

More information

NIST EUVL Metrology Programs

NIST EUVL Metrology Programs NIST EUVL Metrology Programs S.Grantham, C. Tarrio, R.E. Vest, Y. Barad, S. Kulin, K. Liu and T.B. Lucatorto National Institute of Standards and Technology (NIST) Gaithersburg, MD USA L. Klebanoff and

More information

Chemistry Instrumental Analysis Lecture 10. Chem 4631

Chemistry Instrumental Analysis Lecture 10. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 10 Types of Instrumentation Single beam Double beam in space Double beam in time Multichannel Speciality Types of Instrumentation Single beam Requires stable

More information

dnx/dt = -9.3x10-6 / C dny/dt = -13.6x10-6 / C dnz/dt = ( λ)x10-6 / C

dnx/dt = -9.3x10-6 / C dny/dt = -13.6x10-6 / C dnz/dt = ( λ)x10-6 / C Lithium Triborate Crystal LBO Lithium triborate (LiB3O5 or LBO) is an excellent nonlinear optical crystal for many applications. It is grown by an improved flux method. AOTK s LBO is Featured by High damage

More information

LBIR Fluid Bath Blackbody for Cryogenic Vacuum Calibrations

LBIR Fluid Bath Blackbody for Cryogenic Vacuum Calibrations LBIR Fluid Bath Blackbody for Cryogenic Vacuum Calibrations Timothy M. Jung*, Adriaan C. Carter*, Dale R. Sears*, Solomon I. Woods #, Dana R. Defibaugh #, Simon G. Kaplan #, Jinan Zeng * Jung Research

More information

LINEARPYROMETER LP4. Technical Documentation KE November TN

LINEARPYROMETER LP4. Technical Documentation KE November TN 1 LINEARPYROMETER LP4 Technical Documentation KE 256-6.2007 November 2010 5-TN-1622-100 2 1. General Description With the Linearpyrometer Type LP4 a measuring instrument has been made available for pyrometric

More information

Carbon Nanotube Radiometer for Cryogenic Calibrations

Carbon Nanotube Radiometer for Cryogenic Calibrations Carbon Nanotube Radiometer for Cryogenic Calibrations Solomon I. Woods a, Julia K. Scherschligt a, Nathan A. Tomlin b, John H. Lehman b a National Institute of Standards and Technology, 100 Bureau Drive,

More information

The Importance of Total Measured Optic Response in UV LED Measurement. Joe May, Jim Raymont, Mark Lawrence EIT Instrument Markets

The Importance of Total Measured Optic Response in UV LED Measurement. Joe May, Jim Raymont, Mark Lawrence EIT Instrument Markets The Importance of Total Measured Optic Response in UV LED Measurement Joe May, Jim Raymont, Mark Lawrence EIT Instrument Markets October 18, 2017 Measurement Expectations Temperature Industrial thermometry:

More information

SPECTRAL IRRADIANCE DATA

SPECTRAL IRRADIANCE DATA The radiometric data on the following pages was measured in our Standards Laboratory. The wavelength calibrations are based on our spectral calibration lamps. Irradiance data from 250 to 2500 nm is based

More information

GLOSSARY OF TERMS. Terminology Used for Ultraviolet (UV) Curing Process Design and Measurement

GLOSSARY OF TERMS. Terminology Used for Ultraviolet (UV) Curing Process Design and Measurement GLOSSARY OF TERMS Terminology Used for Ultraviolet (UV) Curing Process Design and Measurement This glossary of terms has been assembled in order to provide users, formulators, suppliers and researchers

More information

BTS2048-BS. Product tags: VIS, Spectral Data, Industrial Applications, Laser.

BTS2048-BS. Product tags: VIS, Spectral Data, Industrial Applications, Laser. BTS2048-BS http://www.gigahertz-optik.de/en-us/product/bts2048-bs Product tags: VIS, Spectral Data, Industrial Applications, Laser Gigahertz-Optik GmbH 1/6 Description BTS2048-BS CCD spectroradiometer

More information

HIGH SPEED FIBER PHOTODETECTOR USER S GUIDE

HIGH SPEED FIBER PHOTODETECTOR USER S GUIDE HIGH SPEED FIBER PHOTODETECTOR USER S GUIDE Thank you for purchasing your High Speed Fiber Photodetector. This user s guide will help answer any questions you may have regarding the safe use and optimal

More information

TracQ. Basic Data Acquisition and Spectroscopy Software

TracQ. Basic Data Acquisition and Spectroscopy Software Basic Data Acquisition and Spectroscopy Software TracQ Basic main application window. Many common spectroscopic measurements require coordinated operation of a detection instrument and light source, as

More information

CU-LASP Test Facilities! and Instrument Calibration Capabilities"

CU-LASP Test Facilities! and Instrument Calibration Capabilities CU-LASP Test Facilities! and Instrument Calibration Capabilities" Ginger Drake Calibration Group Manager 303-492-5899 Ginger.Drake@lasp.colorado.edu Thermal Vacuum Test Facilities" 2 Multiple Optical Beam

More information

Calibration of a High Dynamic Range, Low Light Level Visible Source

Calibration of a High Dynamic Range, Low Light Level Visible Source Calibration of a High Dynamic Range, Low Light Level Visible Source Joe LaVeigne a, Todd Szarlan a, Nate Radtke a a Santa Barbara Infrared, Inc., 30 S. Calle Cesar Chavez, #D, Santa Barbara, CA 93103 ABSTRACT

More information

Product tags: VIS, Spectral Data, Color Temperature, CRI, Waterproof, WiFi, Luminous Color, LED, Photometry, General lighting

Product tags: VIS, Spectral Data, Color Temperature, CRI, Waterproof, WiFi, Luminous Color, LED, Photometry, General lighting BTS256-EF https://www.gigahertz-optik.de/en-us/product/bts256-ef Product tags: VIS, Spectral Data, Color Temperature, CRI, Waterproof, WiFi, Luminous Color, LED, Photometry, General lighting Gigahertz-Optik

More information

Photometry for Traffic Engineers...

Photometry for Traffic Engineers... Photometry for Traffic Engineers... Workshop presented at the annual meeting of the Transportation Research Board in January 2000 by Frank Schieber Heimstra Human Factors Laboratories University of South

More information

NIRCam optical calibration sources

NIRCam optical calibration sources NIRCam optical calibration sources Stephen F. Somerstein, Glen D. Truong Lockheed Martin Advanced Technology Center, D/ABDS, B/201 3251 Hanover St., Palo Alto, CA 94304-1187 ABSTRACT The Near Infrared

More information

INTEGRATED PHOTODIODE AND AMPLIFIER

INTEGRATED PHOTODIODE AND AMPLIFIER FPO 7% ABRIDGED DATA SHEET For Complete Data Sheet Call FaxLine -8-8-633 Request Document Number 8 INTEGRATED PHOTODIODE AND AMPLIFIER FEATURES PHOTODIODE SIZE:.9 x.9 inch (.9 x.9mm) FEEDBACK RESISTOR

More information

picoemerald Tunable Two-Color ps Light Source Microscopy & Spectroscopy CARS SRS

picoemerald Tunable Two-Color ps Light Source Microscopy & Spectroscopy CARS SRS picoemerald Tunable Two-Color ps Light Source Microscopy & Spectroscopy CARS SRS 1 picoemerald Two Colors in One Box Microscopy and Spectroscopy with a Tunable Two-Color Source CARS and SRS microscopy

More information

Balanced Photoreceivers Models 1607-AC & 1617-AC

Balanced Photoreceivers Models 1607-AC & 1617-AC USER S GUIDE Balanced Photoreceivers Models 1607-AC & 1617-AC NEW FOCUS, Inc. 2630 Walsh Ave. Santa Clara, CA 95051-0905 USA phone: (408) 980 8088 Fax: (408) 980 8883 e-mail: contact@newfocus.com www.newfocus.com

More information

Meeting Measurement Challenges For Low-Power, Pulsed, Or Modulated Light Sources

Meeting Measurement Challenges For Low-Power, Pulsed, Or Modulated Light Sources Meeting Measurement Challenges For Low-Power, Pulsed, Or Modulated Light Sources By Denise Ullery, Sylvia Tan, and Jay Jeong, Newport Corporation (www.newport.com) Traditionally, power meters have been

More information

Continuum White Light Generation. WhiteLase: High Power Ultrabroadband

Continuum White Light Generation. WhiteLase: High Power Ultrabroadband Continuum White Light Generation WhiteLase: High Power Ultrabroadband Light Sources Technology Ultrafast Pulses + Fiber Laser + Non-linear PCF = Spectral broadening from 400nm to 2500nm Ultrafast Fiber

More information

Advances in Measuring UV LED Arrays

Advances in Measuring UV LED Arrays Advances in Measuring UV LED Arrays Joe May, Jim Raymont, Mark Lawrence EIT Instrument Markets May 8, 2018 Measurement Expectations Temperature Industrial thermometry: 1% accuracy Laboratory thermometry:

More information

pulsecheck The Modular Autocorrelator

pulsecheck The Modular Autocorrelator pulsecheck The Modular Autocorrelator Pulse Measurement Perfection with the Multitalent from APE It is good to have plenty of options at hand. Suitable for the characterization of virtually any ultrafast

More information

Spectroradiometer specbos 1201

Spectroradiometer specbos 1201 Spectroradiometer specbos 1201 specbos 1201 is a precise and compact VIS spectroradiometer. It can be used in laboratory as well as production environment to measure the following quantities: Luminance,

More information

Study on a spectrophotometer system for measuring LED s photometric parameters

Study on a spectrophotometer system for measuring LED s photometric parameters Study on a spectrophotometer system for measuring LED s photometric parameters XIAO-LI ZhOU 1, MU-QING LIU 1, YONG QIAN 2, HUI WANG 2, SHAO-LONG ZHU 1 1 Institute for Electric Light Sources, 2 Department

More information

BTS256-EF. Product tags: VIS, Spectral Measurement, Waterproof, WiFi. Gigahertz-Optik GmbH 1/7

BTS256-EF. Product tags: VIS, Spectral Measurement, Waterproof, WiFi.   Gigahertz-Optik GmbH 1/7 BTS256-EF http://www.gigahertz-optik.de/en-us/product/bts256-ef Product tags: VIS, Spectral Measurement, Waterproof, WiFi Gigahertz-Optik GmbH 1/7 Description Traditional lux meters are increasingly being

More information

Keysight Technologies Optical Power Meter Head Special Calibrations. Brochure

Keysight Technologies Optical Power Meter Head Special Calibrations. Brochure Keysight Technologies Optical Power Meter Head Special Calibrations Brochure Introduction The test and measurement equipment you select and maintain in your production and qualification setups is one of

More information

GHOLOGRAPHIC COMPANY LIGHT MEASUREMENT SYSTEM

GHOLOGRAPHIC COMPANY LIGHT MEASUREMENT SYSTEM GHOLOGRAPHIC COMPANY LIGHT MEASUREMENT SYSTEM Goniophotometer / Integrating Sphere/ Spectroradiometer/ Illuminance Meter Luminance Meter / UV Radiometer Benefits Easy to Use High End Product Precise Measurement

More information

CO 2 Remote Detection Using a 2-µm DIAL Instrument

CO 2 Remote Detection Using a 2-µm DIAL Instrument CO 2 Remote Detection Using a 2-µm DIAL Instrument Erwan Cadiou 1,2, Dominique Mammez 1,2, Jean-Baptiste Dherbecourt 1,, Guillaume Gorju 1, Myriam Raybaut 1, Jean-Michel Melkonian 1, Antoine Godard 1,

More information

It s Our Business to be EXACT

It s Our Business to be EXACT 671 LASER WAVELENGTH METER It s Our Business to be EXACT For laser applications such as high-resolution laser spectroscopy, photo-chemistry, cooling/trapping, and optical remote sensing, wavelength information

More information

JETI Specbos Instruments

JETI Specbos Instruments Spectral measuring instruments for various applications JETI Specbos Instruments The new Specbos family offers compact, spectrometric instruments, designed to measure the color coordinates, spectral characteristics

More information

UV-VIS-IR Spectral Responsivity Measurement System for Solar Cells

UV-VIS-IR Spectral Responsivity Measurement System for Solar Cells November 1998 NREL/CP-52-25654 UV-VIS-IR Spectral Responsivity Measurement System for Solar Cells H. Field Presented at the National Center for Photovoltaics Program Review Meeting, September 8 11, 1998,

More information

DESIGN OF AN ANALOG FIBER OPTIC TRANSMISSION SYSTEM

DESIGN OF AN ANALOG FIBER OPTIC TRANSMISSION SYSTEM DESIGN OF AN ANALOG FIBER OPTIC TRANSMISSION SYSTEM OBJECTIVE To design and build a complete analog fiber optic transmission system, using light emitting diodes and photodiodes. INTRODUCTION A fiber optic

More information

BTS2048-VL-CP-ILED-B-IS-1.0-HL

BTS2048-VL-CP-ILED-B-IS-1.0-HL BTS2048-VL-CP-ILED-B-IS-1.0-HL https://www.gigahertz-optik.de/en-us/product/bts2048-vl-cp-iled-b-is-1.0-hl Product tags: VIS, NIR, Spectral Data, LED Binning, CIE 127, ILED-B, LED Gigahertz-Optik GmbH

More information

Working in Visible NHMFL

Working in Visible NHMFL Working in Visible Optics @ NHMFL NHMFL Summer School 05-19-2016 Stephen McGill Optical Energy Range Energy of Optical Spectroscopy Range SCM3 Optics Facility Energy Range of Optical Spectroscopy SCM3

More information

Ultra-sensitive, room-temperature THz detector using nonlinear parametric upconversion

Ultra-sensitive, room-temperature THz detector using nonlinear parametric upconversion 15 th Coherent Laser Radar Conference Ultra-sensitive, room-temperature THz detector using nonlinear parametric upconversion M. Jalal Khan Jerry C. Chen Z-L Liau Sumanth Kaushik Ph: 781-981-4169 Ph: 781-981-3728

More information

Terahertz Wave Spectroscopy and Analysis Platform. Full Coverage of Applications From R&D to Industrial Testing

Terahertz Wave Spectroscopy and Analysis Platform. Full Coverage of Applications From R&D to Industrial Testing Terahertz Wave Spectroscopy and Analysis Platform Full Coverage of Applications From R&D to Industrial Testing Terahertz Wave Spectroscopy and Analysis Platform Optimal for a wide range of terahertz research

More information

Department of Electrical Engineering and Computer Science, Massachusetts Institute of Technology, 77. Table of Contents 1

Department of Electrical Engineering and Computer Science, Massachusetts Institute of Technology, 77. Table of Contents 1 Efficient single photon detection from 500 nm to 5 μm wavelength: Supporting Information F. Marsili 1, F. Bellei 1, F. Najafi 1, A. E. Dane 1, E. A. Dauler 2, R. J. Molnar 2, K. K. Berggren 1* 1 Department

More information

SIGNAL RECOVERY: Sensors, Signals, Noise and Information Recovery

SIGNAL RECOVERY: Sensors, Signals, Noise and Information Recovery SIGNAL RECOVERY: Sensors, Signals, Noise and Information Recovery http://home.deib.polimi.it/cova/ 1 Signal Recovery COURSE OUTLINE Scenery preview: typical examples and problems of Sensors and Signal

More information

Standards for PV metrology IEC and IEC Stefan Winter, PTB German member of IEC TC82-WG2

Standards for PV metrology IEC and IEC Stefan Winter, PTB German member of IEC TC82-WG2 Standards for PV metrology IEC 60904 and IEC 61853 Stefan Winter, PTB German member of IEC TC82-WG2 12.10.2016 Overview o Demand for high accuracy solar cell calibrations o Standard Test Conditions o DSR

More information

Sometimes the axis of the I-U-dependence are shown in reverse order. In this case the graph shows the stabilized current and measured voltage.

Sometimes the axis of the I-U-dependence are shown in reverse order. In this case the graph shows the stabilized current and measured voltage. 2. Electrical and other parameters 2.1. absolute maximum ratings are a listing of the environmental and electrical stresses that may be applied to a device without resulting in short term or catastrophic

More information

Photometry for Traffic Engineers...

Photometry for Traffic Engineers... Photometry for Traffic Engineers... Workshop presented at the annual meeting of the Transportation Research Board in January 2000 by Frank Schieber Heimstra Human Factors Laboratories University of South

More information

High-Power Femtosecond Lasers

High-Power Femtosecond Lasers High-Power Femtosecond Lasers PHAROS is a single-unit integrated femtosecond laser system combining millijoule pulse energies and high average power. PHAROS features a mechanical and optical design optimized

More information

771 Series LASER SPECTRUM ANALYZER. The Power of Precision in Spectral Analysis. It's Our Business to be Exact! bristol-inst.com

771 Series LASER SPECTRUM ANALYZER. The Power of Precision in Spectral Analysis. It's Our Business to be Exact! bristol-inst.com 771 Series LASER SPECTRUM ANALYZER The Power of Precision in Spectral Analysis It's Our Business to be Exact! bristol-inst.com The 771 Series Laser Spectrum Analyzer combines proven Michelson interferometer

More information

Variable Gain Photoreceiver - Fast Optical Power Meter

Variable Gain Photoreceiver - Fast Optical Power Meter The picture shows model -FC with fiber optic input. Features Conversion gain switchable from 1 x 10 3 to 1 x 10 11 V/W InGaAs-PIN detector Spectral range 900-1700 nm Calibrated at 1550 nm (fiber optic

More information

LED Sources in Photometry at NIST

LED Sources in Photometry at NIST 23 rd Meeting of CCPR, Sep. 22-23, 2016, BIPM, Sevres, France LED Sources in Photometry at NIST Yoshi Ohno (NIST Fellow, IESNA Fellow, CIE President) Collaborators: Cameron Miller Yuqin Zong Maria Nadal

More information

200 MHz Photoreceiver with Si PIN Photodiode

200 MHz Photoreceiver with Si PIN Photodiode The picture shows the -FS with free space input. The photoreceiver will be delivered without post holder and post. Features Si PIN Detector, 0.8 mm Active Diameter Spectral Range 320... 1000 nm Bandwidth

More information

Measuring Kinetics of Luminescence with TDS 744 oscilloscope

Measuring Kinetics of Luminescence with TDS 744 oscilloscope Measuring Kinetics of Luminescence with TDS 744 oscilloscope Eex Nex Luminescence Photon E 0 Disclaimer Safety the first!!! This presentation is not manual. It is just brief set of rule to remind procedure

More information

THE TUNABLE LASER LIGHT SOURCE C-WAVE. HÜBNER Photonics Coherence Matters.

THE TUNABLE LASER LIGHT SOURCE C-WAVE. HÜBNER Photonics Coherence Matters. THE TUNABLE LASER LIGHT SOURCE HÜBNER Photonics Coherence Matters. FLEXIBILITY WITH PRECISION is the tunable laser light source for continuous-wave (cw) emission in the visible and near-infrared wavelength

More information