Current and Future Realizations NRC Photometric and Spectroradiometric Calibration Chains
|
|
- Aleesha Floyd
- 6 years ago
- Views:
Transcription
1 Measurement Science and Standards (MSS) Current and Future Realizations NRC Photometric and Spectroradiometric Calibration Chains Arnold A. Gaertner Photometry, Radiometry and Thermometry (PRT) CORM 30 May 2012 NRC, 100 Sussex Drive, Ottawa
2 Photometry and Spectroradiometry Outline Calibration and measurement services Present calibration chain Future calibration chain SSL facility
3 Photometry and Spectroradiometry Measurement Geometries - 4 basic optical radiation measurement configurations total flux illuminance / irradiance luminance / radiance intensity total flux intensity luminance radiance illuminance irradiance
4 Calibration services Photometry of Incandescent Lamps - total luminous flux (lumen standards) - luminous intensity standards - illuminance standards Required Working Standards Total luminous flux lamp Luminous intensity lamp Calibration of Photometric Instruments and Sources - illuminance meters - luminance meters - luminance sources Spectroradiometry - spectral irradiance lamp standards - spectral radiance source standards - spectroradiometers - irradiance - radiance + Diffuse reflectance standard Spectral irradiance lamp + Diffuse reflectance standard Spectral irradiance lamp + Diffuse reflectance standard
5 Working standard lamps Total luminous flux Luminous intensity Spectral irradiance Auxiliary standards Diffuse reflectance standard
6 Primary calibration Measurement of absolute amount of optical radiation power - SI units (watt) - absolute radiometer Room Temperature ESR Cryogenic Radiometer
7 Photometric calibration chain (present) Luminous Intensity: Absolute Radiometer (room temperature) SI CIE Secondary Radiometers (Si photodiodes) V(l)-correcting filter LIS lamp Luminous Intensity Standard NRC candela maintained on lamps 0.5%
8 Photometric calibration chain (present) Total Luminous Flux: Absolute Radiometer (room temperature) Secondary Radiometers (Si photodiodes) V(l)-correcting filter NRC lumen maintained on lamps 1.0% LIS lamp Goniophotometer 0.5%
9 Spectroradiometric calibration chain Spectral Irradiance (present): Absolute Radiometer (room temperature) narrow-band filter detector based FEL lamp NRC spectral irradiance maintained on lamps (FEL)
10 Spectroradiometric calibration chain Spectral Irradiance (present): NIST NRC Absolute Radiometer NIST FEL 1000W incandescent quartz-halogen NRC CIE NRC spectral irradiance maintained on lamps (FEL)
11 International Comparisons (CCPR) k=2 CCPR = Consultative Committee for Photometry and Radiometry
12 Optical radiation calibration chain Present: detector - based Absolute Radiometer (room temperature) Secondary Radiometers (Si photodiodes) V(l)-correcting filter narrow-band filter LFS lamp gonio photometer LIS lamp FEL lamp NIST
13 Optical radiation calibration chain Under construction: detector and source - based Absolute Radiometer (cryogenic) Transfer Radiometers Filter Radiometers temperature measurement high temperature blackbody (HTBB) Primary Standard Source High Temperature Fixed Points LIS lamp FEL lamp Primary Standard Pyrometer Pyrometers Photometry Spectroradiometry Radiation Thermometry
14 Optical radiation calibration chain Filter radiometers (5) Thermometric Photometric Spectroradiometric
15 Optical radiation calibration chain Photometer filter radiometer V(l)-correcting filter aperture photodiode aluminum cell thermo-electric coolers
16 Optical radiation calibration chain Photometer filter radiometer Source of Uncertainty Type of Uncertainty Standard Uncertainty Spectral responsivity B 0.14% Spatial uniformity B 0.20% Aperture area B 0.07% Spectral mismatch B 0.06% Temperature control B 0.04% Aging B 0.03% Lamp to photometer distance d A 0.20% Photocurrent A 0.02% Combined standard uncertainty 0.33%
17 Optical radiation calibration chain Under construction: detector and source - based Absolute Radiometer (cryogenic) Transfer Radiometers Filter Radiometers temperature measurement high temperature blackbody (HTBB) Primary Standard Source High Temperature Fixed Points LIS lamp FEL lamp Primary Standard Pyrometer Pyrometers Photometry Spectroradiometry Radiation Thermometry
18 Optical radiation calibration chain Spectroradiometry: spectral irradiance Absolute Radiometer (room temperature) narrow-band filter detector based interpolation fit extrapolation is not valid high temperature blackbody (HTBB) Primary Standard Source thermodynamic T=? L l,t c 1 l 5 exp c 2 1 lt Planck equation
19 Optical radiation calibration chain Filter radiometers for spectroradiometry T L m (l m,t) i T HTBB filter radiometer L m L m l m,t l m 5 c 1 exp c 2 1 l m T m measured value l m Measure the BB output at one wavelength
20 Optical radiation calibration chain Filter radiometers for spectroradiometry HTBB FR0 FR4 FR2 FR1 FR3
21 HTBB Temperature (K) Optical radiation calibration chain Filter radiometers for spectroradiometry k=2 uncertainties
22 Optical radiation calibration chain Spectral Irradiance: NIST NRC Absolute Radiometer NIST NRC CIE FEL Lamps HTBB: K
23 Optical radiation calibration chain Spectral Irradiance: transfer facility (under construction) Filter Radiometer transfer standard lamps HTBB High-Temperature Black-Body Transfer standard lamps (incandescent) spectroradiometer filter radiometers
24 Calibration services Incandescent Sources Solid State Sources peak intensity axis #2 mechanical axis optical axis peak intensity axis
25 Calibration services Solid State Sources (facility under development) LED front tip Averaged LED Intensity (CIE 127:2007) detector aperture circular acceptance area of 100 mm 2 mechanical axis Condition B d B = 100 mm Condition A d A = 316 mm
26 Calibration services Solid State Sources (facility under development) Partial LED Flux (CIE 127:2007) d 50 mm circular aperture
27 Calibration services Photometry of Incandescent Lamps - total luminous flux (lumen standards) - luminous intensity standards - illuminance standards Calibration of Photometric Instruments and Sources - illuminance meters - luminance meters - luminance sources Spectroradiometry - spectral irradiance lamp standards - spectral radiance source standards - spectroradiometers - irradiance - radiance SSL facility
28 Discussion Thank you
New automated laser facility for detector calibrations
CORM annual conference, NRC, Ottawa, CANADA June 1, 2012 New automated laser facility for detector calibrations Yuqin Zong National Institute of Standards and Technology Gaithersburg, Maryland USA Overview
More informationIntercomparison of radiation temperature measurements over the temperature range from 1600 K to 3300 K
INSTITUTE OF PHYSICS PUBLISHING Metrologia 40 (2003) S39 S44 METROLOGIA PII: S0026-1394(03)56056-X Intercomparison of radiation temperature measurements over the temperature range from 1600 K to 3300 K
More informationRADIOMETRIC AND PHOTOMETRIC MEASUREMENTS AT THE LNE-INM/CNAM
RADIOMETRIC AND PHOTOMETRIC MEASUREMENTS AT THE LNE-INM/CNAM Jean Bastie, Laura Patricia Gonzalez Galvan. Institut National de Métrologie Conservatoire National des Arts et Métiers 61 rue du Landy 93210
More informationNIST MEASUREMENT SERVICES:
NIST Special Publication 250-37 NIST MEASUREMENT SERVICES: PHOTOMETRIC CALIBRATIONS Yoshihiro Ohno Optical Technology Division Physics Laboratory National Institute of Standards and Technology Gaithersburg,
More informationCapabilities of NIST SIRCUS for Calibrations of SSI Vis-IR Instruments
Capabilities of NIST SIRCUS for Calibrations of SSI Vis-IR Instruments Steve Brown National Institute of Standards & Technology Gaithersburg, MD steve.brown@nist.gov; 301.975.5167 Answer: Ask LASP folks
More informationPhotometry and Light Measurement
Photometry and Light Measurement Adrian Waltho, Analytik Ltd adrian.waltho@analytik.co.uk What is Light? What is Light? What is Light? Ultraviolet Light UV-C 180-280 nm UV-B 280-315 nm UV-A 315-400 nm
More informationSPECTRORADIOMETRY METHODS: A GUIDE TO PHOTOMETRY AND VISIBLE SPECTRORADIOMETRY
SPECTRORADIOMETRY METHODS: A GUIDE TO PHOTOMETRY AND VISIBLE SPECTRORADIOMETRY Written By William E. Schneider, Richard Young, Ph.D 1. SPECTRORADIOMETRY METHODS 1.1. Spectroradiometrics vs Photometric
More informationInstruction Manual of Luminance and Spectral Radiance Calibrations
Aalto University School of Electrical Engineering Metrology Research Institute Jari Hovila Pasi Manninen Tuomas Poikonen Petri Kärhä Instruction Manual of Luminance and Version 2.6 07/12/2015 Page 2 (19)
More informationDevelopment of 2 Total Spectral Radiant Flux Standards at NIST
CIE/USA Annual Conference, October 7, 2014, Seattle, WA Development of 2 Total Spectral Radiant Flux Standards at NIST Yuqin Zong National Institute of Standards and Technology Gaithersburg, Maryland Outline
More informationReport No.: HZ b. Total Luminous Flux (Lumens) Stabilization Time (Light & Power) Table 1: Executive Data Summary CRI
Test Summary Sample Tested: LDP2450L35U1 2X4 3500K Luminous Efficacy (Lumens /Watt) Total Luminous Flux (Lumens) Power (Watts) Power Factor 102.3 5129.5 50.12 0.9883 CCT (K) CRI Stabilization Time (Light
More informationReport No.: HZ h/R1. Total Luminous Flux (Lumens) Stabilization Time (Light & Power) Table 1: Executive Data Summary CRI
Test Summary Sample Tested: KT-PLED50-24-850-VDIM /G2 Luminous Efficacy (Lumens /Watt) Total Luminous Flux (Lumens) Power (Watts) Power Factor 117.9 6083.5 51.62 0.9926 CCT (K) CRI Stabilization Time (Light
More informationApplication Note (A16)
Application Note (A16) Eliminating LED Measurement Errors Revision: A December 2001 Gooch & Housego 4632 36 th Street, Orlando, FL 32811 Tel: 1 407 422 3171 Fax: 1 407 648 5412 Email: sales@goochandhousego.com
More informationReport No.: HZ b. Total Luminous Flux (Lumens) Stabilization Time (Light & Power) Table 1: Executive Data Summary CRI
Test Summary Sample Tested: MLFP24DS4241/SD Luminous Efficacy (Lumens /Watt) Total Luminous Flux (Lumens) Power (Watts) Power Factor 108.6 4723.5 43.49 0.9902 CCT (K) CRI Stabilization Time (Light & Power)
More informationReport No.: HZ a/R2. Total Luminous Flux (Lumens) Stabilization Time (Light & Power) Table 1: Executive Data Summary CRI
Test Summary Sample Tested: FLS50U50B Luminous Efficacy (Lumens /Watt) Total Luminous Flux (Lumens) Power (Watts) Power Factor 108.1 5291.8 48.96 0.9859 CCT (K) CRI Stabilization Time (Light & Power) 5119
More informationReport No.: HZ c/R1. Total Luminous Flux (Lumens) Stabilization Time (Light & Power) Table 1: Executive Data Summary CRI
Test Summary Sample Tested: MLLWP40LED50DS Luminous Efficacy (Lumens /Watt) Total Luminous Flux (Lumens) Power (Watts) Power Factor 96.2 3353.0 34.85 0.9900 CCT (K) CRI Stabilization Time (Light & Power)
More informationReport No.: HZ a/R2. Total Luminous Flux (Lumens) Stabilization Time (Light & Power) Table 1: Executive Data Summary CRI
Test Summary Sample Tested: WPL40AU50B Luminous Efficacy (Lumens /Watt) Total Luminous Flux (Lumens) Power (Watts) Power Factor 91.6 3480.4 37.99 0.9907 CCT (K) CRI Stabilization Time (Light & Power) 5257
More information(Lumens) Stabilization Time CRI (K) (Light & Power) Table 1 Executive Data Summary
Test Summary Sample Tested: LFC41327W/V2 Luminous Efficacy Total Luminous Flux Power (Lumens /Watt) (Lumens) (Watts) Power Factor 45.3 544.6 12.02 0.9892 CCT Stabilization Time CRI (K) (Light & Power)
More informationReport No.: EASZE Page 2 of 14
Report No.: EASZE03220005 Page 2 of 14 1 SUMMARY Parameter Result Total Luminous Flux 1680.9 lm Luminous Efficacy 83.34 lm/w Power Factor 0.9829 Color Rendering Index (Ra) 83.7 Correlated Color Temperature
More informationIESNA standards on LED and SSL: LM-79, LM-80, and future standards
CORM 2009 Conference, Gaithersburg, MD, May 6-8, 2009 IESNA standards on LED and SSL: LM-79, LM-80, and future standards Yoshi Ohno Director of CIE Division 2 Member of ANSI C78-09, IESNA TPC National
More informationNFMS THEORY LIGHT AND COLOR MEASUREMENTS AND THE CCD-BASED GONIOPHOTOMETER. Presented by: January, 2015 S E E T H E D I F F E R E N C E
NFMS THEORY LIGHT AND COLOR MEASUREMENTS AND THE CCD-BASED GONIOPHOTOMETER Presented by: January, 2015 1 NFMS THEORY AND OVERVIEW Contents Light and Color Theory Light, Spectral Power Distributions, and
More informationLM-79 Test Report. Electrical and Photometric tests as required by IESNA LM-79 test standard
Lamp specifics: Luminaire Used: none none Rated Watts Voltage: 6.75 Watts Lamp Type: LPL-W18-120C-930 / 78875 / Rated Orientation CCT: VBD Test Requester E.Z. Cola Rated Voltage 12 V DC New Product: 1
More informationPhotometry for Traffic Engineers...
Photometry for Traffic Engineers... Workshop presented at the annual meeting of the Transportation Research Board in January 2000 by Frank Schieber Heimstra Human Factors Laboratories University of South
More information(Lumens) Stabilization Time CRI (K) (Light & Power) Table 1 Executive Data Summary
Test Summary Sample Tested: LFC61541W/V2 Luminous Efficacy Total Luminous Flux Power (Lumens /Watt) (Lumens) (Watts) Power Factor 63.3 988.8 15.61 0.9810 CCT Stabilization Time CRI (K) (Light & Power)
More informationLM Test Report. for Maxlite Inc. LED Canopy Model: MLCAN20LED50
LM-79-08 Test Report for Maxlite Inc. 12 York Ave West Caldwell NJ 07006 LED Canopy Model: MLCAN20LED50 Laboratory: Leading Testing Laboratories NVLAP CODE: 200960-0 Tel: +86-571-56680806 www.ledtestlab.com
More informationReport No.: HZ f. Total Luminous Flux (Lumens) Stabilization Time (Light & Power) Table 1: Executive Data Summary CRI
Test Summary Sample Tested: L15T8SE450-G Luminous Efficacy (Lumens /Watt) Total Luminous Flux (Lumens) Power (Watts) Power Factor 125.1 1904.0 15.22 0.9679 CCT (K) CRI Stabilization Time (Light & Power)
More informationLM Test Report. for EiKO Global, LLC. 25W Floodlight Model: FLM-2C-N-U
Quality Assured NVLAP LAB CODE 200960-0 LM-79-08 Test Report for EiKO Global, LLC 23220 W. 84th St. Shawnee, KS 66227 25W Floodlight Model: FLM-2C-N-U Laboratory: Leading Testing Laboratories NVLAP CODE:
More informationLM Test Report. for Maxlite SK America Inc. 2*4 Retrofit Kits Model: RKT4514U5550DV
LM-79-08 Test Report for Maxlite SK America Inc. 12 York Ave West Caldwell NJ 07006 2*4 Retrofit Kits Model: RKT4514U5550DV Laboratory: Leading Testing Laboratories NVLAP CODE: 200960-0 Tel: +86-571-56680806
More informationLighting Terminologies Introduction
Lighting Terminologies Introduction A basic understanding of lighting fundamentals is essential for specifiers and decision makers who make decisions about lighting design, installation and upgrades. Radiometry
More informationRadiometry vs. Photometry. Radiometric and photometric units
Radiometry vs. Photometry Radiometry -- the measurement and specification of the power (energy) of a source of electromagnetic radiation.! total energy or numbers of quanta Photometry -- the measurement
More informationLM Test Report. for Maxlite SK America Inc. WALLPACK Model: MLSWP30LED50DS
LM-79-08 Test Report for Maxlite SK America Inc. 12 York Ave West Caldwell NJ 07006 WALLPACK Model: MLSWP30LED50DS Laboratory: Leading Testing Laboratories NVLAP CODE: 200960-0 Tel: +86-571-56680806 www.ledtestlab.com
More informationReport No.: HZ j. Total Luminous Flux (Lumens) Stabilization Time (Light & Power) Table 1: Executive Data Summary CRI
Test Summary Sample Tested: T8FR17/835/DIR/LED Luminous Efficacy (Lumens /Watt) Total Luminous Flux (Lumens) Power (Watts)/2 Power Factor 105.4 1761.0 16.70 0.9970 CCT (K) CRI Stabilization Time (Light
More informationPhotometry for Traffic Engineers...
Photometry for Traffic Engineers... Workshop presented at the annual meeting of the Transportation Research Board in January 2000 by Frank Schieber Heimstra Human Factors Laboratories University of South
More informationLED Sources in Photometry at NIST
23 rd Meeting of CCPR, Sep. 22-23, 2016, BIPM, Sevres, France LED Sources in Photometry at NIST Yoshi Ohno (NIST Fellow, IESNA Fellow, CIE President) Collaborators: Cameron Miller Yuqin Zong Maria Nadal
More informationHigh Illuminance Calibration Facility and Procedures
Final manuscript for J. IES, 27-2, 132-140 (1998) High Illuminance Calibration Facility and Procedures Yoshi Ohno Optical Technology Division National Institute of Standards and Technology Metrology A320,
More informationReport No.: HZ w. Stabilization Time (Light & Power) CRI. Table 1: Executive Data Summary. Figure 1- Overview of the sample
Test Summary Sample Tested: AR-MAL100UT3-40X Luminous Efficacy (Lumens /Watt) Total Luminous Flux (Lumens) Power (Watts) Power Factor 131.6 12593.0 95.69 0.9967 CCT (K) CRI Stabilization Time (Light &
More informationReport No.: HZ h. Total Luminous Flux (Lumens) Stabilization Time (Light & Power) Table 1: Executive Data Summary CRI
Test Summary Sample Tested: 8PLE26LED35 Luminous Efficacy (Lumens /Watt) Total Luminous Flux (Lumens) Power (Watts) Power Factor 101.5 707.7 6.97 0.9815 CCT (K) CRI Stabilization Time (Light & Power) 3498
More informationPhotometric Calibrations
I NlST Measurement Services: Photometric Calibrations NlST Special Publication 250-37 U.S. Department of Commerce Technology Administration National Institute of Standards and Technology NlST Special Publication
More informationInfluence of the light spectral distribution used in the radiometers calibration
Influence of the light spectral distribution used in the radiometers calibration J C Moraes and L O Guerra Allergisa Pesquisa Dermato-Cosmética Ltda GRUPO INVESTIGA, 452. Dr. Romeu Tórtima Avenue, Postcode:
More informationMeasurements of Infrared Sources with the Missile Defense Transfer Radiometer
Measurements of Infrared Sources with the Missile Defense Transfer Radiometer Simon G. Kaplan #, Solomon I. Woods #, Adriaan C. Carter, and Timothy M. Jung * # National Institute of Standards and Technology
More informationCCT/10-13 MeP-K direct methods. Section 3.2: Absolute (spectral-band) radiometry (radiation thermometry)
CCT/10-13 MeP-K direct methods Section 3.2: Absolute (spectral-band) radiometry (radiation thermometry) Authors: Graham Machin (NPL, chair), Klaus Anhalt (PTB), Pieter Bloembergen (NIM, formerly NMIJ &
More informationLM-79 Test Report. GE Infusion M1000/830/W/N. Electrical and Photometric tests as required by IESNA LM-79 test standard
specifics: Luminaire Used: SRTGLEDGI SRTGLEDGI 15W 30K WD E1 GES66 15W 30KRated Voltage: Watts 15 watt Driver Type: Roal Strato GE RSLD035-7A Infusion M1000/83 Rated CCT: Orientation VBU GE Infusion M1000/830/W/N
More informationAdvantages of white LED lamps and new detector technology in photometry
OPEN (2015) 4, e332; ß 2015 CIOMP. All rights reserved 2047-7538/15 www.nature.com/lsa ORIGINAL ARTICLE Advantages of white LED lamps and new detector technology in photometry Tomi Pulli 1, Timo Dönsberg
More informationSample Report. IES LM Test Report. LED Lamp 001. Test results reported for: Orb Optronix report: LEDC WG. Original issue date:
IES LM 79 08 Test Report Test results reported for: report: Original issue date: LED Lamp 001 LEDC001 010 WG 1 January 2012 Prepared for: Testing performed by: LED ABC, Inc. Memorial Highway 1003 7th Ave.
More informationLM-79 Test Report. GE Infusion M2000/830/W/N. Electrical and Photometric tests as required by IESNA LM-79 test standard
specifics: Luminaire Used: SGE4SQLEDGI SGE4SQLEDGI 30W 30K ND E1 AR0204LGI 30W 3SG Rated FGCC Voltage: Watts 30 watt Driver Type: Roal Strato GE RSLD035-07 Infusion M2000/83 Rated CCT: Orientation VBU
More informationModel S471-LED Optometer
Model S471-LED Optometer For over 40 years UDT Instruments, a Gamma Scientific company, has been trusted by the world s leading organizations to provide accurate light measurement systems. UDT Instruments
More informationLM Test Report. for. Philips (China) Investment Co., Ltd. InstantFit LEDtube Model:
LM-79-08 Test Report for Philips (China) Investment Co., Ltd. Building 9, Lane 888, Tianlin Road Shanghai, China InstantFit LEDtube Model: 9290002840 Laboratory: Leading Testing Laboratories NVLAP CODE:
More informationNEW MEASUREMENT STANDARDS AND METHODS FOR PHOTOMETRY AND RADIOMETRY
TKK Dissertations 18 Espoo 2005 NEW MEASUREMENT STANDARDS AND METHODS FOR PHOTOMETRY AND RADIOMETRY Doctoral Dissertation Jari Hovila Helsinki University of Technology Department of Electrical and Communications
More informationMeasurement Services Optics Laboratory
Federal Institute of Metrology METAS CH-3003 Bern-Wabern, 1. February 2018 Measurement Services Optics Laboratory Valid from: 01.02.2018 In our laboratory we perform high accuracy calibrations of your
More informationLM Test Report. for Elec-Tech International Co.,Ltd. LED PAR30 Spot light Model: XX(XX:41-50)
LM-79-08 Test Report for Elec-Tech International Co.,Ltd No.1 Jinfeng Rd.,Tangjiawan Town, Xiangzhou District, Zhuhai City, Guangdong province, China LED PAR30 Spot light Model: 524142XX(XX:41-50) Laboratory:
More informationLM Test Report. for Elec-Tech International Co.,Ltd. LED CEILING LIGHT Model: XX(XX: 01-10)
LM-79-08 Test Report for Elec-Tech International Co.,Ltd No.1 Jinfeng Rd.,Tangjiawan Town, Xiangzhou District, Zhuhai City, Guangdong province, China LED CEILING LIGHT Model: 540751XX(XX: 01-10) Laboratory:
More informationLM Test Report. for Elec-Tech International Co.,Ltd. LED CEILING LIGHT Model: XX(XX: 61-70)
LM-79-08 Test Report for Elec-Tech International Co.,Ltd No.1 Jinfeng Rd.,Tangjiawan Town, Xiangzhou District, Zhuhai City, Guangdong province, China LED CEILING LIGHT Model: 540751XX(XX: 61-70) Laboratory:
More informationLM Test Report. for Elec-Tech International Co., Ltd. BR40 LAMP Model: XX(XX:01-10), WBR40L27S2A
LM-79-08 Test Report for Elec-Tech International Co., Ltd No.1 Jinfeng Rd.,Tangjiawan Town, Xiangzhou District, Zhuhai City, Guangdong province, China BR40 LAMP Model: 525021XX(XX:01-10), WBR40L27S2A Laboratory:
More informationLM Test Report. for Elec-Tech International Co.,Ltd. LED CEILING LIGHT Model: XX(XX: 01-10)
LM-79-08 Test Report for Elec-Tech International Co.,Ltd No.1 Jinfeng Rd.,Tangjiawan Town, Xiangzhou District, Zhuhai City, Guangdong province, China LED CEILING LIGHT Model: 540741XX(XX: 01-10) Laboratory:
More informationThe Art of Light Measurement. Avantes BV Apeldoorn, The Netherlands
The Art of Light Measurement Avantes BV Apeldoorn, The Netherlands Who am I? Ger Loop Product Manager Avantes BV Oude Apeldoornseweg 28 7333 NS APELDOORN The Netherlands Phone: (+31) 313 670170 Fax: (+31)
More informationOL Series 426 Low Light Level lntegrating Sphere Calibration Standards
OL Series 426 Low Light Level lntegrating Sphere Calibration Standards GENERAL The OL Series 426 Low-Light-Level Calibration Standard is designed for accurately calibrating very sensitive microphotometers,
More informationRadiometric Measurement Traceability Paths for Photovoltaic Calibrations. Howard W. Yoon Physical Measurement Laboratory NIST
Radiometric Measurement Traceability Paths for Photovoltaic Calibrations Howard W. Yoon Physical Measurement Laboratory NIST Solar energy and PV Solar radiation: free and abundant! Photovoltaics (PV):
More informationLI-193 Spherical Quantum Sensor
LI-193 Spherical Quantum Sensor The LI-193 Spherical Quantum Sensor measures PAR in air or underwater from all directions at depths up to 350 meters. This sensor is useful for studies of phytoplankton,
More informationRelation between the illuminance responsivity of a photometer and the spectral power distribution of a source
46 9, 093607 September 2007 Relation between the illuminance responsivity of a photometer and the spectral power distribution of a source Ferhat Sametoglu, MEMBER SPIE National Metrology Institute of Turkey
More informationCatalog Number UPC Number LED Spot Bullet Flood Light 42W 4204 Lumens White
Catalog Number 71694 UPC Number 60198671694 Description LED Spot Bullet Flood Light 42W 4204 Lumens White QPL ID # PWG6GTSQ Features Philips Lumiled LUXEON 3030 2D Voltage: 120-277VAC Isolated Driver Compartment
More informationWe bring quality to light. MAS 40 Mini-Array Spectrometer
MAS 40 Mini-Array Spectrometer Features at a glance Cost-effective and robust CCD spectrometer technology Standard USB interface Compatible with all Instrument Systems measuring adapters Different models
More informationTest Report. Report No.: EED35I Page 1 of 10. Customer : EIKO GLOBAL LLC Address : W 84TH ST SHAWNEE, KS USA
Report No.: EED35I000287-1 Page 1 of 10 Customer : EIKO GLOBAL LLC Address : 23220 W 84TH ST SHAWNEE, KS 66227 USA Description of the submitted sample(s): Sample Name : LED Tube Model/Type : LED9W4PH/830DR-G6
More informationDevelopement of the method for realization of spectral irradiance scale featuring system of
Home Search Collections Journals About Contact us My IOPscience Developement of the method for realization of spectral irradiance scale featuring system of spectral comparisons This article has been downloaded
More information\Ç à{x ÇtÅx Éy ALLAH à{x `xüv yâä
\Ç à{x ÇtÅx Éy ALLAH à{x `xüv yâä Ultraviolet Radiation from Some Types of Outdoor Lighting Lamps Dr.Essam El-Moghazy Photometry and Radiometry division, National Institute for Standards (NIS), Egypt.
More informationDetector-Based Traceability Chain for Spectral Irradiance using Tunable Laser-Based Facility at PTB
Detector-Based Traceability Chain for Spectral Irradiance using Tunable Laser-Based Facility at PTB Von der Fakultät für Elektrotechnik, Informationstechnik, Physik der Technischen Universität Carolo-Wilhelmina
More informationPart 1: New spectral stuff going on at NIST. Part 2: TSI Traceability of TRF to NIST
Part 1: New spectral stuff going on at NIST SIRCUS-type stuff (tunable lasers) now migrating to LASP Absolute Spectrally-Tunable Detector-Based Source Spectrally-programmable source calibrated via NIST
More informationMorris Products 53 Carey Road, Queensbury, NY 12804
Catalog Number 71462 UPC Number 60198671462 Description LED Vandal Resistant Canopy Light 40W Voltage: 120/277 Features Corrosion Die Cast Aluminum Housing Prismatic Polycarbonate Lens Superior Architectural
More informationSolid-State Lighting Photometry Issues
Les Industries Spectralux Inc. Spectralux Industries Inc. 2750 Sabourin, Saint-Laurent (Québec) H4S 1M2 Canada Tél.:(514) 332-0082 Fax : (514) 332-3590 www.spectralux.ca Solid-State Lighting Photometry
More informationRadiometry vs. Photometry. Radiometric and photometric units
Radiometry vs. Photometry Radiometry -- the measurement and specification of the power (energy) of a source of electromagnetic radiation. total energy or numbers of quanta Photometry -- the measurement
More informationCatalog Number UPC Number LED Spot Bullet Flood Light 15W 1620 Lumens Bronze
Catalog Number 71679 UPC Number 60198671679 Description LED Spot Bullet Flood Light 15W 1620 Lumens Bronze QPL ID # PJMW56XK Features Philips Lumiled LUXEON 3030 2D Voltage: 120-277VAC Isolated Driver
More informationRadiometric and Photometric Measurements with TAOS PhotoSensors
INTELLIGENT OPTO SENSOR DESIGNER S NUMBER 21 NOTEBOOK Radiometric and Photometric Measurements with TAOS PhotoSensors contributed by Todd Bishop March 12, 2007 ABSTRACT Light Sensing applications use two
More informationCatalog number Lens Option Very Narrow without diffuser Exterior200LED_VeryNarrow_withoutdiffuser_5600Kelvin.ies
Exterior 200 LED, Very Narrow (Balanced white 5600 Kelvin. Without diffuser) (13 ) Photometric Report Martin R&D Optics Laboratory, 31-Oct-2007 Data sheet conforms to American National Standard E1.9-2001
More informationBTS2048-VL-CP-ILED-B-IS-1.0-HL
BTS2048-VL-CP-ILED-B-IS-1.0-HL https://www.gigahertz-optik.de/en-us/product/bts2048-vl-cp-iled-b-is-1.0-hl Product tags: VIS, NIR, Spectral Data, LED Binning, CIE 127, ILED-B, LED Gigahertz-Optik GmbH
More informationLM Test Report for OSRAM SYLVANIA INC. 100 Endicott Street Danvers, MA 01923
LM-79-08 Test Report for OSRAM SYLVANIA INC. 100 Endicott Street Danvers, MA 01923 Model(s): Spectrum Lighting SG12SQLED21W35KE1/12SQGW/3.5LA 8 May 2011 This test report presents the results of measurements
More informationApplication Note (A19)
Application Note (A19) Reducing Uncertainty in Precision High-Brightness LED Measurements Revision: A December 2005 Gooch & Housego 4632 36 th Street, Orlando, FL 32811 Tel: 1 407 422 3171 Fax: 1 407 648
More informationCat# LED Classic Large Wall-Pack
Cat# 71440 LED Classic Large Wall-Pack OVERALL LAMP PARAMETERS LED DRIVER LED LIFESPAN & ENVIRONMENT SAFETY&EMC OTHERS Model: Input Voltage Input Current Input Power Power Factor Luminance Luminous Efficiency
More informationLED14.5WPAR30S/FL/827-DIM
LED14.5WPAR30S/FL/827-DIM TABLE OF CONTENTS 1 - GENERAL INFORMATION... 3 1.1 PRODUCT DESCRIPTION FOR EQUIPMENT UNDER TEST (EUT)... 3 1.2 OBJECTIVE... 3 1.3 TEST FACILITY DESCRIPTION... 4 1.4 TEST EQUIPMENT
More informationLI-192 Underwater Quantum Sensor
LI-192 Underwater Quantum Sensor The LI-192 Underwater Quantum Sensor measures PAR from all angles in one hemisphere. The LI-192 works in air or underwater at depths up to 560 meters. The measurements
More informationThe Standard for over 40 Years
Light Measurement The Standard for over 40 Years Introduction LI-COR radiation sensors measure the flux of radiant energy the energy that drives plant growth, warms the earth, and lights our world. The
More informationRadiometric Solar Telescope (RaST) The case for a Radiometric Solar Imager,
SORCE Science Meeting 29 January 2014 Mark Rast Laboratory for Atmospheric and Space Physics University of Colorado, Boulder Radiometric Solar Telescope (RaST) The case for a Radiometric Solar Imager,
More informationIESNA LM-79: Measurement and Test Report for Halco Lighting Technologies ATL 2940-A PACIFIC DRIVE. Dec 08, 2014
IESNA LM-79: 2008 Measurement and Test Report for Halco Lighting Technologies ATL 2940-A PACIFIC DRIVE Dec 08, 2014 Product Name: LED Model No: 82031 Test Engineer: David Zhang Report No.: BTR66.181.14.0039.06
More informationStatus of Low-Background Infrared Calibration Facility at NIST
Status of Low-Background Infrared Calibration Facility at NIST Simon G. Kaplan #, Solomon I. Woods #, Julia Scherschligt #, Joseph P. Rice #, Adriaan C. Carter *, and Timothy M. Jung * # National Institute
More informationIESNA LM-79: Measurement and Test Report for Jiawei Technology (USA) Limited Lincoln Ave.Hayward, CA USA.
IESNA LM-79: 2008 Measurement and Test Report for Jiawei Technology (USA) Limited 2305 Lincoln Ave.Hayward, CA 94545 USA Apr 03, 2015 Product Name: self-ballast LED Lamp Model No: DL-9PAR30-830-LN-25-D;ML-9PAR30-830-LN-25-D
More informationwww.greenelectricalsupply.com REPORT 3933 US ROUTE 11 CORTLAND, NEW YORK 13045 Project No. G100125842 Date: August 17, 2010 REPORT NO. 100125842CRT-010 TEST OF PAR 38 LED LAMPS MODEL NOS. S8854 / S8861
More informationwww.greenelectricalsupply.com REPORT 3933 US ROUTE 11 CORTLAND, NEW YORK 13045 Project No. G100125842 Date: August 17, 2010 REPORT NO. 100125842CRT-018 TEST OF PAR 30L LED LAMPS MODEL NOS. S8814 / S8821
More informationProduct Tripix 300 Catalog number Lens Option Standard. Tripix300_ ies
Tripix 300 and 1200, Tripix 300 Standard (Without diffuser) (20 ) Photometric Report Martin R&D Optics Laboratory, 28-Jun-2010 Data sheet conforms to American National Standard E1.9-2001 Product Tripix
More informationLINEARPYROMETER LP4. Technical Documentation KE November TN
1 LINEARPYROMETER LP4 Technical Documentation KE 256-6.2007 November 2010 5-TN-1622-100 2 1. General Description With the Linearpyrometer Type LP4 a measuring instrument has been made available for pyrometric
More informationNVLAP LAB CODE LM Test Report. For. Maxlite INC. (Brand Name: Maxlite) 12 York Ave., West Caldwell, NJ 07006
LM-79-08 Test Report For Maxlite INC (Brand Name: Maxlite) 12 York Ave., West Caldwell, NJ 07006 2x4 Luminaires for Ambient Lighting of Interior Commercial Spaces Model name(s): MLFP24EP50XX Remark: The
More informationwww.greenelectricalsupply.com REPORT 3933 US ROUTE 11 CORTLAND, NEW YORK 13045 Project No. G100125842 Date: August 17, 2010 REPORT NO. 100125842CRT-039 TEST OF LED CANDLE LAMPS MODEL NOS. S8899 / S8903
More informationwww.greenelectricalsupply.com REPORT 3933 US ROUTE 11 CORTLAND, NEW YORK 13045 Project No. G100125842 Date: August 17, 2010 REPORT NO. 100125842CRT-036 TEST OF LED CANDLE LAMPS MODEL NOS. S8891 / S8894
More informationReport No.: HZ k/R1. (Lumens) Stabilization Time CRI (K) (Light & Power) Table 1: Executive Data Summary
Test Summary Sample Tested: ML4LS10SOLBK Luminous Efficacy Total Luminous Flux Power (Lumens /Watt) (Lumens) (Watts) Power Factor 55.2 552.2 10.01 0.9797 CCT Stabilization Time CRI (K) (Light & Power)
More informationISS-30-VA. Product tags: Integrating Sphere Source. https://www.gigahertz-optik.de/en-us/product/iss-30-va. Gigahertz-Optik GmbH 1/5
ISS-30-VA https://www.gigahertz-optik.de/en-us/product/iss-30-va Product tags: Integrating Sphere Source Gigahertz-Optik GmbH 1/5 Description standards for spectral radiance Spectroradiometers and other
More informationIESNA LM-79: Measurement and Test Report for EiKO Global, LLC. Sep 09, 2013 LED8WMR16/38/830-DIM. David Zhang BTR
IESNA LM-79: 2008 Measurement and Test Report for EiKO Global, LLC Sep 09, 2013 Product Name: LED Model No: LED8WMR16/38/830-DIM Test Engineer: David Zhang Report No.: BTR66.181.13.1294.01-1 Sample Received
More informationThe Importance of Total Measured Optic Response in UV LED Measurement. Joe May, Jim Raymont, Mark Lawrence EIT Instrument Markets
The Importance of Total Measured Optic Response in UV LED Measurement Joe May, Jim Raymont, Mark Lawrence EIT Instrument Markets October 18, 2017 Measurement Expectations Temperature Industrial thermometry:
More informationPhotometric Test Report
Photometric Test Report Relevant Standards IES LM-79-2008, ANSI C82.77-2002, CIE 13.3-1995 CIE 15-2004, ANSI C78.377-2015, IES TM-30-2015 Prepared For Smart Grow Technologies 69319 Panoramic Dr Sisters,
More informationQuality Manual of Luminous Intensity Laboratory
Aalto University School of Electrical Engineering Metrology Research Institute Jari Hovila Pasi Manninen Tuomas Poikonen Tomi Pulli Quality Manual of Version 7.2 01/12/2015 Page 2 (21) 1. Table of contents
More informationCompany synopsis. Regulations and Standards
Goniophotometer Company synopsis Pleiades Instruments is an optoelectronic system maker, designing and manufacturing for you specific systems such as photometric measurement and customized systems. Designing
More informationLighting Sciences
Lighting Sciences www.lightingsciences.com Lighting Sciences Inc. 7826 E. Evans Road Scottsdale, Arizona 85260 USA Tel: 480-991-9260 Fax: 480-991-0375 INTENSITY(CANDLEPOWER) SUMMARY ANGLE MEAN CP LUMENS
More informationLM Test Report. For. Revolution Lighting Technologies, Inc. (Brand Name: ) 2280 Ward Ave. Simi Valley, CA
LM-79-08 Test Report For Revolution Lighting Technologies, Inc (Brand Name: ) 2280 Ward Ave. Simi Valley, CA. 93065 Outdoor Pole/Arm-Mounted Area and Roadway Luminaires Model name(s): 113XS8-3LT Remark:
More informationProperties of LED considering museum lighting
Properties of LED considering museum lighting Carsten Dam-Hansen, DTU Fotonik Contents Solid State Lighting (SSL) i.e. LED based lighting is a new lighting technology that may offer many advantages for
More informationNVLAP LAB CODE LM Test Report. For GREEN INOVA LIGHTING TECHNOLOGY (SHENZHEN) LTD. (Brand Name: GI LED LIGHTING)
LM-79-08 Test Report For GREEN INOVA LIGHTING TECHNOLOGY (SHENZHEN) LTD (Brand Name: GI LED LIGHTING) 4th floor, building 2, Zone 2, Hong Hua Ling Industrial Park, Liu xian Block,Nan Shan District, Shenzhen,
More information