Developement of the method for realization of spectral irradiance scale featuring system of
|
|
- Randolf Copeland
- 6 years ago
- Views:
Transcription
1 Home Search Collections Journals About Contact us My IOPscience Developement of the method for realization of spectral irradiance scale featuring system of spectral comparisons This article has been downloaded from IOPscience. Please scroll down to see the full text article. ( View the table of contents for this issue, or go to the journal homepage for more Download details: IP Address: The article was downloaded on 08/10/2010 at 08:55 Please note that terms and conditions apply.
2 PUBLISHED BY IOP PUBLISHING FOR SISSA RECEIVED: July 14, 2010 ACCEPTED: September 8, 2010 PUBLISHED: October 5, 2010 Developement of the method for realization of spectral irradiance scale featuring system of spectral comparisons V. Skerovic, a,1 V. Zarubica, a L. Zekovic, b I. Belca b and M. Aleksić a a Directorate of measures and precious metals, Optical radiation Metrology department, Mike Alasa 14, Belgrade, Serbia b Faculty of Physics, Department for Applied physics and metrology, Studentski trg 12-16, Belgrade, Serbia vladanskerovic@dmdm.rs ABSTRACT: Realization of the scale of spectral responsivity of the detectors in the Directorate of Measures and Precious Metals (DMDM) is based on silicon detectors traceable to LNE-INM. In order to realize the unit of spectral irradiance in the laboratory for photometry and radiometry of the Bureau of Measures and Precious Metals, the new method based on the calibration of the spectroradiometer by comparison with standard detector has been established. The development of the method included realization of the System of Spectral Comparisons (SSC), together with the detector spectral responsivity calibrations by means of a primary spectrophotometric system. The linearity testing and stray light analysis were preformed to characterize the spectroradiometer. Measurement of aperture diameter and calibration of transimpedance amplifier were part of the overall experiment. In this paper, the developed method is presented and measurement results with the associated measurement uncertainty budget are shown. KEYWORDS: Spectrometers; Photon detectors for UV, visible and IR photons (solid-state) 1 Corresponding author c 2010 IOP Publishing Ltd and SISSA doi: / /5/10/p10001
3 Contents 1 Introduction 1 2 System of Spectral Comparisons 1 3 Method of Derivation of the Value of Spectral Irradiance 3 4 Measurements Spectral responsivity measurements Spectral responsivity of filter-radiometer Determining the aperture size of the filter-radiometer Characterization of the transfer spectroradiometer 8 5 Measurement uncertainty budget 10 6 Conclusions 11 1 Introduction The traceability of radiometric quantities in the Directorate of measures and precious metals is established by means of standard detectors calibrated for the spectral responsivity at LNE-INM [1]. Realization of the spectral irradiance unit is based on the comparison method [2], where the value of spectral responsivity is transferred to filter radiometers and a spectroradiometer (used as a wavelength tuneable-filter radiometer). A filter radiometer is then used for spectral irradiance unit realization, whereas a spectroradiometer is used for derivation of the relative spectral irradiance of the FEL 1000 lamp [3]. The method of realization of the value of spectral irradiance is developed according to existing laboratory conditions, with respect to many already proved, detector based methods of realization of spectral irradiance scale [4 6]. The basic idea of this experiment was to use the transfer spectroradiometer to measure the relative spectral distribution of the standard lamp, instead of the set of filter-radiometers and interpolation based on adopted preliminary values of spectral emissivity of tungsten [5]. The overall experiment was realized within preparation for EURAMET.PR-K1A Comparison of Spectral Irradiance [7]. 2 System of Spectral Comparisons The system of spectral comparisons (SSC) was designed on the principle of spectral comparisons, where the value was assigned to the object by its comparison with reference standard. It means 1
4 Photometric bench Hg QHT D2 FEL 1000 Monochromator Aperture Standard Detector on Linear Stage Condenser Optical System Spectroradiometer Filter radiometer Figure 1. Measurement setup for realization of Spectral Irradiance Based on the System of Spectral Comparisons. that the system works in the same or similar way for measurements of various optical radiation quantities (spectral transmittance /reflectance, detector spectral responsivity, spectral irradiance). The SSC (figure 1) consists of a monochromatic radiation source, an optical system, a motion system for changing the object of measurement and a detection system consisting of a transimpedance amplifier and a scanner DVM(digital voltmeter) [8]. The light source, including QHT (quartz halogen tungsten) and D2 (deuterium source), has an optical system that projects the image of the source on the entrance slit of the monochromator. A mercury lamp is used for wavelength calibration of the system. A filter wheel with order sorting filters and shutter is positioned in front of the entrance slit. The monochromator (1/4 meter focal length) has three interchangeable diffraction gratings, covering a spectral range from 200 nm to 2500 nm. The exit slit is square (1 mm 1 mm, for spectral irradiance measurement). The exit optical system can be modified, depending on the application, and it can be either collimating (quartz lens objective) or condensing (quartz lens objective or combination of plane and spherical mirror). An aperture is used to define beam geometry and to reduce scattered light. The optical path of about 0.5 m enables beam geometry to be well defined. A linear stage is used to change the detectors in the focal plane of the system. 2
5 3 Method of Derivation of the Value of Spectral Irradiance The theoretical basis of the method for determining the spectral irradiance by means of absolute radiometers, used in this experiment, is the one presented by Boivin in 1980 [9]. The practical formula used in measurement is I = A d (r 0 /r) 2 S(λ) E(λ) dλ (3.1) where I is measured filter-radiometer photocurrent, A d is aperture area of the filter-radiometer, r and r 0 are the distance on which the measurement is made and the distance (0.5 m) for which the irradiance is defined respectively, the S(λ) is the spectral responsivity of the filter-radiometer, and E(λ) is spectral irradiance of the standard lamp [10]. The assumption is that the shape of spectral responsivity function of the filter-radiometer is spectrally symmetrical around λ max (peak wavelength of detector responsivity). When S(λ) = S(λ max ) S rel (λ) and E(λ) = E(λ max ) E rel (λ), the formula (3.1) becomes: I = A d (r 0 /r) 2 S(λ max ) E(λ max ) S rel (λ) E rel (λ) dλ, (3.2) from where the E(λ max ) can be determined. The values of E rel (λ) are obtained from measurements with the transfer spectroradiometer. The methodology of derivation of spectral irradiance is shown in a block diagram (figure 2). The SSC was used to calibrate the spectroradiometer for spectral responsivity by comparison with the standard detector at 23 wavelengths. The spectroradiometer wavelength scale was calibrated using a Mercury lamp. The slit function of the spectroradiometer was determined independently for each of the 23 wavelengths. The slit function was determined as relative spectral responsivity of the spectroradiometer in the range of ± 10 nm from center wavelength used for calibration. The data were used to calculate integral relative spectral responsivity of the spectroradiometer at each wavelength for spectral range with non-zero responsivity, as for the filter radiometer. The spectroradiometer was a modified EG&G 550 spectroradiometer. The entrance optics was a small integrating sphere and a quartz lens that projects the image on the entrance slit. The monochromator consisted of a filter wheel with order sorting filters and a 1200 g/mm concave diffraction grating. The bandwidth of the monochromator was 10 nm. The Hamamatsu H photomultiplier was used as the detector. The integrating sphere entrance was positioned in such a way that all the flux (irradiating the reference detector) entered the sphere. The spectroradiometer was then moved to the photometric bench for relative calibration of the FEL lamp. Absolute calibration of the lamp was performed with the filter radiometer (V(λ)), whose absolute spectral responsivity was determined by comparison with standard detectors in the primary spectrophotometric system (figure 2), [11]. The filter radiometer was a modified LMT Photometric detector, the diffuser replaced with a precision aperture. The aperture was calibrated by the method of direct comparison in the length department of the Directorate of Measures and Precious Metals [12]. Linearity measurements were performed by double aperture method [13, 14]. Cut-off filters were used at 600 nm and 400 nm to perform stray light analysis. 3
6 Figure 2. Methodology of derivation of spectral irradiance. Figure 3. Spectrophotometric system for detector comparisons. SISFRMS-, light source; SFRMS measurement station; RD, monitor detector; PN, detectors under test on translation stage. 4 Measurements 4.1 Spectral responsivity measurements The measurement of absolute responsivity of the filter radiometer was performed on the primary spectrophotometric system [11], (figure 3). 4
7 Figure 4. Simplified scheme of the transimpedance amplifier. Detectors were automatically exchanged in the path of the light beam. Each time, signals from the main and compensation (monitor) detectors were measured simultaneously, ten times per each detector. The average value of those readings was taken into account. At every wavelength, measurements were preformed for the sample and the reference detector, and the measurement of the dark current was made and correction applied before each. The photocurrent was measured by means of a specially constructed transimpedance amplifier [15, 16] and the scanner digital voltmeter. The transimpedance amplifier (figure 4) was calibrated by means of the KEITHLEY 263 low current source calibrator and the HP 3458 digital voltmeter [17]. The measurement uncertainty of the calibration of the transimpedance amplifier on 1MΩ range was relative, mainly resulting from the uncertainty of the low current calibration source. The reproducibility of measurements was determined on the sample of 30 independent measurements of each sample for every wavelength. In uncertainty evaluation, the uncertainty components due to beam divergence and bandwidth effect were taken as negligible [18]. Nonuniformity of the detectors was not analysed. Nonlinearity of the system was determined by the double aperture method [14]. The uncertainty budget of the comparison method of detector spectral responsivity is presented in table 1. The overall uncertainty of reproducing the value of spectral responsivity of detectors was evaluated from the uncertainty of the method of comparisons and the uncertainty of the LNE-INM calibration of the standard detector (which amounts 0.6%, k=2). Prior to that, mutual comparisons of reference detectors were made to prove measurement uncertainty of reproducing the value of spectral responsivity. The method of mutual comparisons is based on comparison measurements between calibrated standards, where one is used as a standard and the other is used as a calibration artefact. For example, results of the comparison of two detectors is calculated as the relative deviation of the results of measurements for the second detector s responsivity, using the first one as a standard, from its values given in certificate [2]. 5
8 Table 1. The uncertainty budget of the comparison method of detector spectral responsivity. Uncertainty component Uncertainty Uncertainty (Type A) (Type B) Reproducibility (detector No 1) Reproducibility (detector No 2) Calibration of transeimpedance amplifier (detector No 1) Calibration of transeimpedance amplifier (detector No 2) DVM uncertainty Nonlinearity Wavelength accuracy Stray light 10 8 Standard uncertainty (1σ) Combined standard uncertainty u m (1σ) Figure 5. Average deviations of the results for the detectors spectral responsivity, from the values given in LNE-INM certificate, in the function of wavelengths. Mutual comparisons of three standard detectors (SR1, SR2, SR3 - Hamamatsu S BQ photodiodes) were performed by comparing each to each, in both directions (measurement sequence), meaning that six different comparisons were made. For each of these comparisons, five measurement cycles were made at 23 wavelengths with ten readings at every wavelength. Calculations were made at each point of measurement. The key result of the measurements was the deviation of the results for the detectors spectral responsivity, obtained from measurements, from the values given in LNE-INM certificate [1]. The standard deviation of all mutual comparison measurements, for repeated measurements for each pair of detectors, was calculated and given as measure of reproducibility of measurements. The average reproducibility for all comparison measurements was below 0.1%, which complies with the results of evaluation of the measurement uncertainty of the comparisons method (table 1). In figure 5, the average deviations of the results for the detectors spectral responsivity (for all three detectors) from the values given in LNE-INM certificate after the latest calibration, in the function of wavelengths, are shown. Figure 5 shows that deviations of the results for the detectors spectral responsivity (obtained 6
9 Figure 6. Spectral responsivity of filter-radiometer. from measurements) from the values given in LNE-INM certificate are within the limits of measurement uncertainty of LNE-INM calibration of 0.6% (k = 2) and 1.0% for wavelengths above 900 nm. The average deviations (figure 5) show regularity in their spectral characteristics, which leads to the conclusion that some systematic effects have influence on the measurement results. The noticed systematic effects were not either explained or included in the measurement uncertainty of the method, evaluated and given in table 1. An additional component of measurement uncertainty due to noticed systematic effects was calculated from the results of mutual comparisons [19], as standard deviation of the results of mutual comparisons for the entire group of detectors and it is 0.25%, for k= Spectral responsivity of filter-radiometer The filter radiometer was a modified LMT Photometric detector, where the diffuser was replaced with a precision aperture. Measurements were made on the primary spectrophotometric system (figure 3), by comparisons with three reference detectors. The average values from these measurements were taken as the result for spectral responsivity. The measurement results are presented in figure 6, as a respective V(λ) curve. 4.2 Determining the aperture size of the filter-radiometer The aperture area was calculated from the size of the aperture diameter. Measurements of the aperture diameter were made with Carl Zeiss measuring microscope [12]. The measurements were made around the aperture space on angular positions of every 15 degrees. An average value was taken as the result for the aperture diameter. The standard deviation of the results from all angu- 7
10 lar positions was taken as a measurement uncertainty of determining the aperture diameter. The average diameter is d sr = mm, and σ sr= The measurement uncertainty of the aperture u A was obtained as the partial differential of the formula for aperture area size, and it equals 2σ sr. The aperture area size was A d = cm 2 and u A = (1σ). All type B uncertainty components in diameter measurements were too small compared to the calculated standard deviation and hence neglected. 4.3 Characterization of the transfer spectroradiometer The spectroradiometer was a modified EG&G 550 spectroradiometer. The diffuser on the entrance of the spectroradiometer was replaced with a small integrating sphere. The detector, together with the detection electronics, was changed, and the Photomultiplier Hamamatsu H was then used as the detector with transimpedance amplifier and a scanner digital voltmeter. The spectroradiometer wavelength scale was calibrated using a Mercury lamp together with the wavelength scale of the SSC in the setup shown in figure 1. The Mercury lamp was put in front of the entrance optics of the SSC, so that the wavelength calibrations of monochromator (SSC) and spectroradiometer were preformed simultaneously. In that way, both wavelength scales were matched. The uncertainty of the wavelength scale of the monochromator of the SSC was found to be 0.2 nm. The wavelength setting of the spectroradiometer was corrected to match the wavelength scale of the monochromator of the SSC. The uncertainty of setting of the wavelength of the spectroradiometer was 0.5 nm. This procedure was repeated for all wavelengths of interest for calibration. The slit function of the spectroradiometer was determined in the measurement setup shown in figure 1. The slit-function was measured as relative spectral responsivity of the spectroradiometer around central wavelength, by comparison with reference detector. In the measurement setup (figure 1), the reference detector was positioned in the light beam with step-motor driven linear stage. Spectroradiometer integrating sphere entrance was positioned in such a way that all the flux (irradiating the reference detector) entered the sphere. Measurements were made by setting the fixed wavelength of the spectroradiometer, and scanning with SSC over the range of ± 15 nm from central wavelength used for calibration. Results of measurements of the slit-function at the wavelength setting of 550 nm are shown in figure 7. The values represented as zero are of the order of Figure 7 shows that the slit function is almost perfectly symmetrical, and that existing nonsymmetry is the result of wavelength uncertainty of spectroradiometer of 0.5 nm. The general idea of making the measurements of the slit-function of spectroradiometer is to be able to calculate absolute spectral responsivity of the spectroradiometer by numerical integration over the bandpass. Also, the data can be used to make the corrections for bandwidth effects [20]. Due to the near perfect triangular bandpass function (figure 7), and considering the fact that, in this experiment, the spectroradiometer was used only for relative spectral distribution measurements, none of these corrections were made. The assumption is that, with symmetrical triangular bandpass function, the bandwidth effects in the measurement of relative spectral distribution with the spectroradiometer can be neglected in this experiment. 8
11 Figure 7. The slit function of the spectroradiometer at the wavelength of 550 nm. The spectroradiometer nonlinearity was determined by the adapted double aperture method [14]. The double aperture was placed in front of the entrance of the integrating sphere (a part of the entrance optics of the spectroradiometer). All together, they were placed on the photometric bench. The lamps of nominal electrical powers ranging from 1000 W to 60 W were used for illumination. The signal level was changed, according to the inverse square law, by changing the distance from the source on the photometric bench. The approximation was made that the use of the integrating sphere annulled any spatial differences in the illumination of the entrance slit of the spectroradiometer. In the signal range of interest, the measured nonlinearities were of the order of 0.2%. Stray light analysis was preformed using cut-off filters at 600 nm and 400 nm. The stray light level was found to be of the order of 0.2%, which is in compliance with manufacturer specifications of stray light 0.3% [21]. The results of the measurement of the spectroradiometer relative spectral responsivity in the SSC facility are presented in figure 8. At each wavelength, 20 measurements were made and the average value of those readings was taken into account. At each wavelength, the measurement of the dark current was made and correction applied. All the results were normalised at 555 nm. Discontinuities on the relative spectral responsivity characteristic were the result of the change in the order of the sorting filters. The spectroradiometer was then moved to the photometric bench for relative calibration of the FEL lamp. Absolute calibration of the lamp was preformed with filter radiometer (V(λ)). Positions on the photometric bench (distance from the lamp) for both the spectroradiometer and filter-radiometer were chosen in such a manner that the signal level was similar to the signal level in spectral responsivity measurements. This was done in order to minimize the effects of nonlinearity. The distance of filter radiometer from the lamp filament was 2.5 m. Measurements were repeated 30 times and standard deviation was calculated for every wavelength. This standard deviation was taken as a direct component of measurement uncertainty due to reproducibility of measurements. 9
12 Figure 8. Relative spectral responsivity of transfer spectroradiometer. Figure 9. Results of realization of spectral irradiance scale. The standard lamp was an old type of the FEL1000 lamp manufactured by Hoffman Engineering Corporation. The electrical parameters of the lamp were controlled with the state- of-the-art electrical equipment [22]. The DC source of the stability of was used for the lamp power supply. The measurement of the electrical current was preformed with a digital voltmeter (5 ppm accuracy) and a standard resistor. The lamp voltage was measured directly on the lamp to provide control of the lamp stability [22]. The results of the calibration of the lamp for spectral irradiance are shown in figure 9. The value of the irradiance at 550 nm was W m 2 nm 1. 5 Measurement uncertainty budget The measurement uncertainty analysis [10] was based on the measurement equation (3.2), and it included the analysis of both direct and indirect components of measurement uncertainty. The results and analysis presented previously in this paper were used for the uncertainty budget estimation [19]. The uncertainty budget is given in table 2. All uncertainties are relative values. 10
13 Table 2. The uncertainty budget of realization of the spectral irradiance scale. Uncertainty component Uncertainty [%] Type A d aperture area A r distance from the source B S(λ) responsivity of filter radiometer Signal measurement (standard detector) A Signal measurement (filter radiometer) A Gain (channal 1) B Gain (channal 2) B DVM B Nonlinearity B Wavelenght B Stray light 10 8 B Calibration of standard detector (LNE-INM) 0.3 B Mutual comparisons (deviation for the group of the detectors) 0.25 A Combined standard uncertainty u S E (λ) relative spectral distribution of the lamp Relative spectral responsivity of spectroradiometer A Signal measurement A Spectroradiometer nonlinearity 0.2 B Stray light in spectroradiometer 0.2 B Wavelenght 0.01 B Bandwidth neglected B Combined standard uncertainty u E Lamp parameters Power supply stability 0.02 B Electrical measurements 0.01 B Standard resistor 0.04 B Lamp stability 0.05 B Standard uncertainty u FEL Combined standard uncertainty (k=1) Conclusions This paper presents the current status of the research and development process in realization of the scale of spectral irradiance in the Directorate of measures and precious metals. The experiment was performed within the arrangement for Euramet K1.a key comparison of spectral irradiance measurement. All the results should be interpreted in that sense. The objective of this paper was to evaluate the adequacy of the method. The new method based on calibration of spectroradiometer by comparison with standard detector has been established to realize spectral irradiance scale. Within the process of development, also, the new methods for determination of the influential parameters have been established. 11
14 The modified double aperture method was realized in order to determine non-linearity of the spectroradiometer. The method of comparison has been used for determination the slit function of the spectroradiometer. In this paper, all the methods are explained and the results of measurements are presented. Results of spectral irradiance measurements presented in this paper are similar to the results reported in the CIPM Key Comparison K1-a Spectral Irradiance 250 nm to 2500 nm, [4]. The uncertainty analysis shows that the main components of uncertainty originate from the uncertainty of spectral responsivity scale. The evaluated measurement uncertainty corresponds to measurement uncertainties from the CIPM Key Comparison [23], which leads to a conclusion that the realized method is appropriate. The uncertainty analysis refers to main sources of uncertainties giving directions for some future improvements in the method. References [1] Calibration certificate N o 1180-Ra-07, LNE-INM, [2] V. Skerovic, P. Vukadin and V. Zarubica, Mutual comparison of detectors spectral responsivity to prove stated measurement uncertainty, in Newrad Conference 2005, October 2005, Davos, Switzerland. [3] J. Metzdorf, A Sperling, S Winter, K.-H. Raatz and W. Möller, A new FEL-type quartz-halogen lamp as an improved standard of spectral irradiance, Metrologia 35 (1998) 423. [4] CIPM Key Comparison K1-a Spectral Irradiance 250 nm to 2500 nm, Final Report, NPL, 23rd January [5] T Kubarsepp, P Karha, F Manoocheri, S Nevas, L. Ylianttila and E. Ikonen, Spectral irradiance measurements of tungsten lamps with filter radiometers in spectral range 290 nm to 900 nm, Metrologia 37 (2000) 305. [6] C. Carreras, A. Corrons, Absolute spectroradiometric and photometric scales based on an electrically calibrated pyroelectric radiometer, Applied Optics 20 (1981) [7] Draft Protocol for the Euramet K1.A Comparison of Spectral Irradiance, NPL, [8] V. Skerovic, V. Zarubica, Realization of Spectral Irradiance Scale Featuring System of Spectral Comparisons, in Newrad Conference 2008, October 2008, Daejeon, Korea. [9] L.P. Boivin, Calibration of incadescent lamps for spectral irradiance by means of absolute radiometers, Appl. Opt. 19 (1980) [10] B.K. Tsai, B.C Johnson, Evaluation of uncertainties in fundamental radiometric measurements, Metrologia 35 (1998) 587. [11] V. Skerovic, P. Vukadin, V. Zarubica and L. Zekovic, Realization of primary spectrophotometric system, Meas. Sci. Technol. 19 (2008) [12] C. Zeiss, User manual Universal-Messmikroskop, Jena. [13] L.P. Boivin, Automated absolute and relative spectral linearity measurements on photovoltaic detectors, Metrologia 30 (1993)
15 [14] K.D. Mielenz and K. Eckerle, Spectrophotometer linearity testing using the double-aperture method, Appl. Opt. 11 (1972) [15] G. Eppeldauer and J.E. Hardis, Fourteen-decade Photocurrent measurements with Large-area silicon photodiodes at room temperature, Appl. Opt. 30 (1991) [16] G. Eppeldauer, Optical Radiation Measurement with Selected Detectors and Matched Electronic Circuits between 200 nm and 20 µm, NIST Technical Note 1438, (2001). [17] G. Eppeldauer, Traceability of NIST photocurrent measurements to electrical standards, in Newrad Conference 2008, October 2008, Daejeon, Korea. [18] T.C. Larson, S.S. Bruce and A.C. Parr, Spectroradiometric Detector Measurements, NIST Special Publication , February [19] Guide of the Expression of Uncertainty in Measurements, International Organization for Standardization, Genève, Switzerland (1993). [20] E.R. Woolliams, M.G. Cox, P.M. Harris and H.M. Pegrum, Correcting for bandwidth effects in monochromator measurements, in Newrad Conference 2005, October 2005, Davos, Switzerland. [21] Instruction manual Model 550/555 spectroradiometer system, EG&G electro-optics, (1976). [22] P. Vukadin, V. Skerovic, V. Zarubica, Tracing the values of light quantities in Federal bureau of measures and precious metals, publication CIE x : proceedings of the CIE Expert Symposium 2001 on Uncertainty Evaluation Methods for Analysis of Uncertainties in Optical Radiation Measurement, January 2001, Vienna, Austria. [23] E.R. Woolliams, N.P. Fox, M.G. Cox, P.M. Harris and N.J. Harrison, The CCPR K1-a key comparison of spectral irradiance from 250 nm to 2500 nm: measurements, analysis and results, Metrologia 43 (2006)
Spectral responsivity uncertainty of silicon photodiodes due to calibration spectral bandwidth
INSTITUTE OF PHYSICS PUBLISHING Meas. Sci. Technol. 12 (2001) 1926 1931 MEASUREMENT SCIENCE AND TECHNOLOGY Spectral responsivity uncertainty of silicon photodiodes due to calibration spectral bandwidth
More informationCurrent and Future Realizations NRC Photometric and Spectroradiometric Calibration Chains
Measurement Science and Standards (MSS) Current and Future Realizations NRC Photometric and Spectroradiometric Calibration Chains Arnold A. Gaertner Photometry, Radiometry and Thermometry (PRT) CORM 30
More informationRADIOMETRIC AND PHOTOMETRIC MEASUREMENTS AT THE LNE-INM/CNAM
RADIOMETRIC AND PHOTOMETRIC MEASUREMENTS AT THE LNE-INM/CNAM Jean Bastie, Laura Patricia Gonzalez Galvan. Institut National de Métrologie Conservatoire National des Arts et Métiers 61 rue du Landy 93210
More informationA stray light corrected array spectroradiometer for complex high dynamic range measurements in the UV spectral range.
A stray light corrected array spectroradiometer for complex high dynamic range measurements in the UV spectral range Mike Clark Gigahertz-Optik GmbH m.clark@gigahertz-optik.de Array spectroradiometers
More informationUV-VIS-IR Spectral Responsivity Measurement System for Solar Cells
November 1998 NREL/CP-52-25654 UV-VIS-IR Spectral Responsivity Measurement System for Solar Cells H. Field Presented at the National Center for Photovoltaics Program Review Meeting, September 8 11, 1998,
More informationMeasurement of normalized spectral responsivity of digital imaging devices by using a LED-based tunable uniform source
See discussions, stats, and author profiles for this publication at: https://www.researchgate.net/publication/235716439 Measurement of normalized spectral responsivity of digital imaging devices by using
More informationSPECTRORADIOMETRY METHODS: A GUIDE TO PHOTOMETRY AND VISIBLE SPECTRORADIOMETRY
SPECTRORADIOMETRY METHODS: A GUIDE TO PHOTOMETRY AND VISIBLE SPECTRORADIOMETRY Written By William E. Schneider, Richard Young, Ph.D 1. SPECTRORADIOMETRY METHODS 1.1. Spectroradiometrics vs Photometric
More informationPhotometry for Traffic Engineers...
Photometry for Traffic Engineers... Workshop presented at the annual meeting of the Transportation Research Board in January 2000 by Frank Schieber Heimstra Human Factors Laboratories University of South
More informationInstruction Manual of Luminance and Spectral Radiance Calibrations
Aalto University School of Electrical Engineering Metrology Research Institute Jari Hovila Pasi Manninen Tuomas Poikonen Petri Kärhä Instruction Manual of Luminance and Version 2.6 07/12/2015 Page 2 (19)
More informationHigh Illuminance Calibration Facility and Procedures
Final manuscript for J. IES, 27-2, 132-140 (1998) High Illuminance Calibration Facility and Procedures Yoshi Ohno Optical Technology Division National Institute of Standards and Technology Metrology A320,
More informationIntercomparison of radiation temperature measurements over the temperature range from 1600 K to 3300 K
INSTITUTE OF PHYSICS PUBLISHING Metrologia 40 (2003) S39 S44 METROLOGIA PII: S0026-1394(03)56056-X Intercomparison of radiation temperature measurements over the temperature range from 1600 K to 3300 K
More informationKeysight Technologies Optical Power Meter Head Special Calibrations. Brochure
Keysight Technologies Optical Power Meter Head Special Calibrations Brochure Introduction The test and measurement equipment you select and maintain in your production and qualification setups is one of
More informationDevelopment of 2 Total Spectral Radiant Flux Standards at NIST
CIE/USA Annual Conference, October 7, 2014, Seattle, WA Development of 2 Total Spectral Radiant Flux Standards at NIST Yuqin Zong National Institute of Standards and Technology Gaithersburg, Maryland Outline
More informationAdvantages of white LED lamps and new detector technology in photometry
OPEN (2015) 4, e332; ß 2015 CIOMP. All rights reserved 2047-7538/15 www.nature.com/lsa ORIGINAL ARTICLE Advantages of white LED lamps and new detector technology in photometry Tomi Pulli 1, Timo Dönsberg
More informationINTERNATIONAL INTERCOMPARISON OF WAVELENGTH SCALE AND PHOTOMETRIC SCALE OF SPECTROPHOTOMETRY LABORATORIES CENAM - NRC - INMETRO - NIST.
INTERNATIONAL INTERCOMPARISON OF WAVELENGTH SCALE AND PHOTOMETRIC SCALE OF SPECTROPHOTOMETRY LABORATORIES - - - Arquímedes Ruiz 1, Joanne Zwinkels 2, Iakira Bougleux 3 and Yvonne Barnes 4 1 Optical and
More informationCapabilities of NIST SIRCUS for Calibrations of SSI Vis-IR Instruments
Capabilities of NIST SIRCUS for Calibrations of SSI Vis-IR Instruments Steve Brown National Institute of Standards & Technology Gaithersburg, MD steve.brown@nist.gov; 301.975.5167 Answer: Ask LASP folks
More informationBTS256-E WiFi - mobile light meter for photopic and scotopic illuminance, EVE factor, luminous color, color rendering index and luminous spectrum.
Page 1 BTS256-E WiFi - mobile light meter for photopic and scotopic illuminance, EVE factor, luminous color, color rendering index and luminous spectrum. The BTS256-E WiFi is a high-quality light meter
More informationComponents of Optical Instruments. Chapter 7_III UV, Visible and IR Instruments
Components of Optical Instruments Chapter 7_III UV, Visible and IR Instruments 1 Grating Monochromators Principle of operation: Diffraction Diffraction sources: grooves on a reflecting surface Fabrication:
More informationOL Series 426 Low Light Level lntegrating Sphere Calibration Standards
OL Series 426 Low Light Level lntegrating Sphere Calibration Standards GENERAL The OL Series 426 Low-Light-Level Calibration Standard is designed for accurately calibrating very sensitive microphotometers,
More informationNew automated laser facility for detector calibrations
CORM annual conference, NRC, Ottawa, CANADA June 1, 2012 New automated laser facility for detector calibrations Yuqin Zong National Institute of Standards and Technology Gaithersburg, Maryland USA Overview
More informationPhotometry for Traffic Engineers...
Photometry for Traffic Engineers... Workshop presented at the annual meeting of the Transportation Research Board in January 2000 by Frank Schieber Heimstra Human Factors Laboratories University of South
More informationSPECTRAL IRRADIANCE DATA
The radiometric data on the following pages was measured in our Standards Laboratory. The wavelength calibrations are based on our spectral calibration lamps. Irradiance data from 250 to 2500 nm is based
More informationApplication Note (A11)
Application Note (A11) Slit and Aperture Selection in Spectroradiometry REVISION: C August 2013 Gooch & Housego 4632 36 th Street, Orlando, FL 32811 Tel: 1 407 422 3171 Fax: 1 407 648 5412 Email: sales@goochandhousego.com
More informationRadiometric Measurement Traceability Paths for Photovoltaic Calibrations. Howard W. Yoon Physical Measurement Laboratory NIST
Radiometric Measurement Traceability Paths for Photovoltaic Calibrations Howard W. Yoon Physical Measurement Laboratory NIST Solar energy and PV Solar radiation: free and abundant! Photovoltaics (PV):
More informationNIST MEASUREMENT SERVICES:
NIST Special Publication 250-37 NIST MEASUREMENT SERVICES: PHOTOMETRIC CALIBRATIONS Yoshihiro Ohno Optical Technology Division Physics Laboratory National Institute of Standards and Technology Gaithersburg,
More informationSpectroradiometer characterisations for traceable solar radiation measurements
Spectroradiometer characterisations for traceable solar radiation measurements Julian Gröbner, Natalia Kouremeti, Stelios Kazadzis, Gregor Hülsen, Luca Egli Physikalisch-Meteorologisches Observatorium
More informationDevelopment of a New Reference Spectrophotometer. J.C. Zwinkels, D.S. Gignac
Development of a New Reference Spectrophotometer Development of a New Reference Spectrophotometer J.C. Zwinkels, D.S. Gignac J.C. Zwinkels, D.S. Gignac National Research Council of Canada, Division of
More informationGuide to SPEX Optical Spectrometer
Guide to SPEX Optical Spectrometer GENERAL DESCRIPTION A spectrometer is a device for analyzing an input light beam into its constituent wavelengths. The SPEX model 1704 spectrometer covers a range from
More information\Ç à{x ÇtÅx Éy ALLAH à{x `xüv yâä
\Ç à{x ÇtÅx Éy ALLAH à{x `xüv yâä Ultraviolet Radiation from Some Types of Outdoor Lighting Lamps Dr.Essam El-Moghazy Photometry and Radiometry division, National Institute for Standards (NIS), Egypt.
More informationISS-30-VA. Product tags: Integrating Sphere Source. https://www.gigahertz-optik.de/en-us/product/iss-30-va. Gigahertz-Optik GmbH 1/5
ISS-30-VA https://www.gigahertz-optik.de/en-us/product/iss-30-va Product tags: Integrating Sphere Source Gigahertz-Optik GmbH 1/5 Description standards for spectral radiance Spectroradiometers and other
More informationCIE 220:2016 Characterization and Calibration Method of UV Radiometers
CIE 220:2016 Characterization and Calibration Method of UV Radiometers Anton Gugg-Helminger Gigahertz-Optik GmbH, Germany www.gigahertz-optik.de Editor s note: This article has been reprinted from UV News,
More informationChemistry Instrumental Analysis Lecture 7. Chem 4631
Chemistry 4631 Instrumental Analysis Lecture 7 UV to IR Components of Optical Basic components of spectroscopic instruments: stable source of radiant energy transparent container to hold sample device
More informationInfluence of the light spectral distribution used in the radiometers calibration
Influence of the light spectral distribution used in the radiometers calibration J C Moraes and L O Guerra Allergisa Pesquisa Dermato-Cosmética Ltda GRUPO INVESTIGA, 452. Dr. Romeu Tórtima Avenue, Postcode:
More informationTRACEABLE SOLAR RADIATION MEASUREMENTS
TRACEABLE SOLAR RADIATION MEASUREMENTS Julian Gröbner, Natalia Kouremeti, Gregor Hülsen, Stelios Kazadzis, Luca Egli Physikalisch-Meteorologisches Observatorium Davos, World Radiation Center (PMOD/WRC)
More informationmetcon meteorologieconsultgmbh, Instruments for Atmospheric Research W1aa_Feb_2017_1.doc 1 -
metcon meteorologieconsultgmbh, Instruments for Atmospheric Research W1aa_Feb_2017_1.doc 1 - ACTINIC FLUX SPECTRAL RADIOMETERS Ultra-fast CCD-Detector Spectrometer, UVB enhanced Cooled CCD, 512 pixel *
More informationQuality Manual of Luminous Intensity Laboratory
Aalto University School of Electrical Engineering Metrology Research Institute Jari Hovila Pasi Manninen Tuomas Poikonen Tomi Pulli Quality Manual of Version 7.2 01/12/2015 Page 2 (21) 1. Table of contents
More informationBTS256-PAR. Product tags: VIS, Spectral Data, PAR, Waterproof, General lighting. https://www.gigahertz-optik.de/en-us/product/bts256-par
BTS256-PAR https://www.gigahertz-optik.de/en-us/product/bts256-par Product tags: VIS, Spectral Data, PAR, Waterproof, General lighting Gigahertz-Optik GmbH 1/8 Description Measurement and examination of
More informationMeasurement Services Optics Laboratory
Federal Institute of Metrology METAS CH-3003 Bern-Wabern, 1. February 2018 Measurement Services Optics Laboratory Valid from: 01.02.2018 In our laboratory we perform high accuracy calibrations of your
More informationLASP / University of Colorado
The SORCE SIM Instrument: Progress Toward Spectral Irradiance Time Series Throughout the 300-3000 nm Region. Jerald Harder, Juan Fontenla, Byron Smiley, Sean Davis, George Lawrence and Gary Rottman LASP
More informationCCT/10-13 MeP-K direct methods. Section 3.2: Absolute (spectral-band) radiometry (radiation thermometry)
CCT/10-13 MeP-K direct methods Section 3.2: Absolute (spectral-band) radiometry (radiation thermometry) Authors: Graham Machin (NPL, chair), Klaus Anhalt (PTB), Pieter Bloembergen (NIM, formerly NMIJ &
More informationJETI Specbos Instruments
Spectral measuring instruments for various applications JETI Specbos Instruments The new Specbos family offers compact, spectrometric instruments, designed to measure the color coordinates, spectral characteristics
More informationSection 1: SPECTRAL PRODUCTS
Section 1: Optical Non-dispersive Wavelength Selection Filter Based Filter Filter Fundamentals Filter at an Incidence Angle Filters and Environmental Conditions Dispersive Instruments Grating and Polychromators
More informationTechnical Notes. Integrating Sphere Measurement Part II: Calibration. Introduction. Calibration
Technical Notes Integrating Sphere Measurement Part II: Calibration This Technical Note is Part II in a three part series examining the proper maintenance and use of integrating sphere light measurement
More informationIntegrating Spheres. Why an Integrating Sphere? High Reflectance. How Do Integrating Spheres Work? High Damage Threshold
1354 MINIS Oriel Integrating Spheres Integrating spheres are ideal optical diffusers; they are used for radiometric measurements where uniform illumination or angular collection is essential, for reflectance
More informationA Laser-Driven Light Source (LDLS) as a portable spectral irradiance calibration source in the UV range and other radiometric applications.
A Laser-Driven Light Source (LDLS) as a portable spectral irradiance calibration source in the UV range and other radiometric applications. O. El Gawhary 1, S. van den Berg 1, N. van der Leden 1, P. Dekker
More informationObservational Astronomy
Observational Astronomy Instruments The telescope- instruments combination forms a tightly coupled system: Telescope = collecting photons and forming an image Instruments = registering and analyzing the
More informationThe PMOD/WRC Precision Spectroradiometer PSR
The PMOD/WRC Precision Spectroradiometer PSR Julian Gröbner, Natalia Kouremeti, Ricco Soder, Diego Wasser, Manfred Gyo, Fabian Dührig, Etienne De Coulon Physikalisch-Meteorologisches Observatorium Davos,
More informationTotal solar irradiance measurements with PREMOS/PICARD
Total solar irradiance measurements with PREMOS/PICARD Werner Schmutz, André Fehlmann, Wolfgang Finsterle, Greg Kopp, and Gerard Thuillier Citation: AIP Conf. Proc. 1531, 624 (2013); doi: 10.1063/1.4804847
More informationInstruction manual for Ocean Optics USB4000 and QE65 Pro spectroradiometers
Aalto University School of Electrical Engineering Metrology Research Institute Hans Baumgartner Instruction manual for Version 1.0 24/11/2016 Page 2 (9) 1. Table of contents 1. Table of contents... 2 2.
More informationSCCH 4: 211: 2015 SCCH
SCCH 211: Analytical Chemistry I Analytical Techniques Based on Optical Spectroscopy Atitaya Siripinyanond Office Room: C218B Email: atitaya.sir@mahidol.ac.th Course Details October 19 November 30 Topic
More informationGoals of the Lab: Photodetectors and Noise (Part 2) Department of Physics. Slide 1. PHYSICS6770 Laboratory 4
Slide 1 Goals of the Lab: Understand the origin and properties of thermal noise Understand the origin and properties of optical shot noise In this lab, You will qualitatively and quantitatively determine
More informationOptical and Quantum Electronics (2005) 37: Ó Springer 2005 DOI /s
Optical and Quantum Electronics (2005) 37:529 543 Ó Springer 2005 DOI 10.1007/s11082-005-3955-5 Cryogenic radiometer based absolute spectral power responsivity calibration of integrating sphere radiometer
More informationSpectrophotometric calibration system for DECam
Spectrophotometric calibration system for DECam J.-P. Rheault*, D. L. DePoy, T. W. Behm, E. W. Kylberg, Kris Cabral, Rick Allen, J. L. Marshall, Department of Physics and Astronomy, Texas A&M University,
More informationProgram for UV Intercomparison 2014 in Davos:
Program for UV Intercomparison 2014 in Davos: June 2014 Date: 7 16 July 2014 Location: PMOD/WRC Davos Switzerland. Information Update: http://projects.pmodwrc.ch/env03/index.php/8-emrp-uv/project/24- intercomparison-2014
More informationCalibration of a High Dynamic Range, Low Light Level Visible Source
Calibration of a High Dynamic Range, Low Light Level Visible Source Joe LaVeigne a, Todd Szarlan a, Nate Radtke a a Santa Barbara Infrared, Inc., 30 S. Calle Cesar Chavez, #D, Santa Barbara, CA 93103 ABSTRACT
More informationLI-193 Spherical Quantum Sensor
LI-193 Spherical Quantum Sensor The LI-193 Spherical Quantum Sensor measures PAR in air or underwater from all directions at depths up to 350 meters. This sensor is useful for studies of phytoplankton,
More informationRadiometric Solar Telescope (RaST) The case for a Radiometric Solar Imager,
SORCE Science Meeting 29 January 2014 Mark Rast Laboratory for Atmospheric and Space Physics University of Colorado, Boulder Radiometric Solar Telescope (RaST) The case for a Radiometric Solar Imager,
More informationNIST Agency Report May 2012 OUTLINE. The case for traceability NMI capabilities A view to the future the HIP Current/recent NIST activities
NIST Agency Report May 2012 OUTLINE The case for traceability NMI capabilities A view to the future the HIP Current/recent NIST activities The case for traceability Earth Radiation Budget: Solar irradiance
More informationRadiometric and Photometric Measurements with TAOS PhotoSensors
INTELLIGENT OPTO SENSOR DESIGNER S NUMBER 21 NOTEBOOK Radiometric and Photometric Measurements with TAOS PhotoSensors contributed by Todd Bishop March 12, 2007 ABSTRACT Light Sensing applications use two
More informationFully Portable Spectrophotometer for Transmission Measurement
November 2008 LCRT-2005-S Portable Spectrophotometer for Transmission Measurement Spectral Transmission in d/0 and 0/0 Measurement Geometries Transparency Measurement in real in-line Set-up Photometric
More informationThe University of Toledo R. Ellingson and M. Heben
focal length, f Spectral Measurement Using a Monochromator, Thermopile Detector, and Lock-In Amplifier September 18, 2012 The University of Toledo R. Ellingson and M. Heben Where are We, Where we are Going?
More informationCONFIGURING. Your Spectroscopy System For PEAK PERFORMANCE. A guide to selecting the best Spectrometers, Sources, and Detectors for your application
CONFIGURING Your Spectroscopy System For PEAK PERFORMANCE A guide to selecting the best Spectrometers, s, and s for your application Spectral Measurement System Spectral Measurement System Spectrograph
More informationQuality Manual of Luminous Flux Measurements
Aalto University School of Electrical Engineering Metrology Research Institute Jari Hovila Pasi Manninen Tuomas Poikonen Quality Manual of Luminous Version 3.0 02/12/2014 Page 2 (23) 1. Table of contents
More informationProduct tags: VIS, Spectral Data, Color Temperature, CRI, Waterproof, WiFi, Luminous Color, LED, Photometry, General lighting
BTS256-EF https://www.gigahertz-optik.de/en-us/product/bts256-ef Product tags: VIS, Spectral Data, Color Temperature, CRI, Waterproof, WiFi, Luminous Color, LED, Photometry, General lighting Gigahertz-Optik
More informationNIST EUVL Metrology Programs
NIST EUVL Metrology Programs S.Grantham, C. Tarrio, R.E. Vest, Y. Barad, S. Kulin, K. Liu and T.B. Lucatorto National Institute of Standards and Technology (NIST) Gaithersburg, MD USA L. Klebanoff and
More informationReport on the CCPR Pilot Comparison. Spectral Responsivity 10 nm to 20 nm
Page 1 of 30 Report on the CCPR Pilot Comparison Spectral Responsivity 10 nm to 20 nm reported by Frank Scholze Physikalisch-Technische Bundesanstalt Berlin Germany October 2009 Page 2 of 30 Page 3 of
More informationOriel Cornerstone 130 1/8 m Monochromator
1 Oriel Cornerstone 130 1/8 m Monochromator Cornerstone 130 1/8 m Monochromator The Cornerstone 130 family of Oriel Monochromators supports two gratings simultaneously, which can be easily interchanged,
More informationSensitive measurement of partial coherence using a pinhole array
1.3 Sensitive measurement of partial coherence using a pinhole array Paul Petruck 1, Rainer Riesenberg 1, Richard Kowarschik 2 1 Institute of Photonic Technology, Albert-Einstein-Strasse 9, 07747 Jena,
More informationThe Standard for over 40 Years
Light Measurement The Standard for over 40 Years Introduction LI-COR radiation sensors measure the flux of radiant energy the energy that drives plant growth, warms the earth, and lights our world. The
More informationApplication Note (A16)
Application Note (A16) Eliminating LED Measurement Errors Revision: A December 2001 Gooch & Housego 4632 36 th Street, Orlando, FL 32811 Tel: 1 407 422 3171 Fax: 1 407 648 5412 Email: sales@goochandhousego.com
More informationBTS2048-UV. Product tags: UV, Spectral Data, LED Binning, Industrial Applications, LED. https://www.gigahertz-optik.de/en-us/product/bts2048-uv
BTS2048-UV https://www.gigahertz-optik.de/en-us/product/bts2048-uv Product tags: UV, Spectral Data, LED Binning, Industrial Applications, LED Gigahertz-Optik GmbH 1/8 Description UV CCD spectroradiometer
More informationLEDGON 100. Goniophotometer. Two Global Leaders. One Complete Solution.
LEDGON 100 Goniophotometer Two Global Leaders. One Complete Solution. Our story Two Global Leaders. One Complete Solution. Konica Minolta Sensing Americas provides advanced optical technology that precisely
More informationStudy on a spectrophotometer system for measuring LED s photometric parameters
Study on a spectrophotometer system for measuring LED s photometric parameters XIAO-LI ZhOU 1, MU-QING LIU 1, YONG QIAN 2, HUI WANG 2, SHAO-LONG ZHU 1 1 Institute for Electric Light Sources, 2 Department
More informationLED Tester BTS256-LED
1 LED Tester BTS256-LED The BTS256-LED tester is one of the most compact light measurement devices with integrated integrating sphere for high accuracy measurement of luminous flux, spectral and color
More informationRelation between the illuminance responsivity of a photometer and the spectral power distribution of a source
46 9, 093607 September 2007 Relation between the illuminance responsivity of a photometer and the spectral power distribution of a source Ferhat Sametoglu, MEMBER SPIE National Metrology Institute of Turkey
More informationMeasurements of Infrared Sources with the Missile Defense Transfer Radiometer
Measurements of Infrared Sources with the Missile Defense Transfer Radiometer Simon G. Kaplan #, Solomon I. Woods #, Adriaan C. Carter, and Timothy M. Jung * # National Institute of Standards and Technology
More informationQuantum Efficiency Measurement System with Internal Quantum Efficiency Upgrade
Quantum Efficiency Measurement System with Internal Quantum Efficiency Upgrade QE / IPCE SYSTEM Upgraded with Advanced Features Includes IV Testing, Spectral Response, Quantum Efficiency System/ IPCE System
More informationUSER MANUAL FOR VISIBLE SPECTROPHOTOMETER
USER MANUAL FOR VISIBLE SPECTROPHOTOMETER 1 Table of Contents 1. MAIN USAGES...3 2. WORKING ENVIRONMENT...3 3. MAIN TECHNICAL DATA AND SPECIFICATIONS...4 4. WORKING PRINCIPLE...5 5. OPTICAL PRINCIPLE...6
More informationChemistry Instrumental Analysis Lecture 10. Chem 4631
Chemistry 4631 Instrumental Analysis Lecture 10 Types of Instrumentation Single beam Double beam in space Double beam in time Multichannel Speciality Types of Instrumentation Single beam Requires stable
More informationLI-192 Underwater Quantum Sensor
LI-192 Underwater Quantum Sensor The LI-192 Underwater Quantum Sensor measures PAR from all angles in one hemisphere. The LI-192 works in air or underwater at depths up to 560 meters. The measurements
More informationBTS256-EF. Product tags: VIS, Spectral Measurement, Waterproof, WiFi. Gigahertz-Optik GmbH 1/7
BTS256-EF http://www.gigahertz-optik.de/en-us/product/bts256-ef Product tags: VIS, Spectral Measurement, Waterproof, WiFi Gigahertz-Optik GmbH 1/7 Description Traditional lux meters are increasingly being
More informationTracQ. Basic Data Acquisition and Spectroscopy Software
Basic Data Acquisition and Spectroscopy Software TracQ Basic main application window. Many common spectroscopic measurements require coordinated operation of a detection instrument and light source, as
More informationLM Test Report for OSRAM SYLVANIA INC. 100 Endicott Street Danvers, MA 01923
LM-79-08 Test Report for OSRAM SYLVANIA INC. 100 Endicott Street Danvers, MA 01923 Model(s): Spectrum Lighting SG12SQLED21W35KE1/12SQGW/3.5LA 8 May 2011 This test report presents the results of measurements
More informationSpectrophotometer. An instrument used to make absorbance, transmittance or emission measurements is known as a spectrophotometer :
Spectrophotometer An instrument used to make absorbance, transmittance or emission measurements is known as a spectrophotometer : Spectrophotometer components Excitation sources Deuterium Lamp Tungsten
More informationCIPM and CCPR What are these organizations and how do they affect my testing results. Maria Nadal Photometry, Surface Color and Appearance NIST
CIPM and CCPR What are these organizations and how do they affect my testing results Maria Nadal Photometry, Surface Color and Appearance NIST CIE USA Annual Meeting October 6-7, 2014 Calibration Laboratory
More informationLINEARPYROMETER LP4. Technical Documentation KE November TN
1 LINEARPYROMETER LP4 Technical Documentation KE 256-6.2007 November 2010 5-TN-1622-100 2 1. General Description With the Linearpyrometer Type LP4 a measuring instrument has been made available for pyrometric
More informationSpectroscopy in the UV and Visible: Instrumentation. Spectroscopy in the UV and Visible: Instrumentation
Spectroscopy in the UV and Visible: Instrumentation Typical UV-VIS instrument 1 Source - Disperser Sample (Blank) Detector Readout Monitor the relative response of the sample signal to the blank Transmittance
More informationImprovement of the detection efficiency calibration and homogeneity measurement of Si SPAD detectors
DOI 10.1186/s40064-016-3735-7 RESEARCH Open Access Improvement of the detection efficiency calibration and homogeneity measurement of Si SPAD detectors Klodian Dhoska 1*, Helmuth Hofer 2, Beatrice Rodiek
More informationAn imaging device for multispectral analysis in the visible range. P. Fiorentin, E. Pedrotti, A. Scroccaro
An imaging device for multispectral analysis in the visible range P. Fiorentin, E. Pedrotti, A. Scroccaro University of Padova, Department of Electrical Engineering, via Gradenigo 6/a, Padov, Italy, +39498277914,
More informationPart 1: New spectral stuff going on at NIST. Part 2: TSI Traceability of TRF to NIST
Part 1: New spectral stuff going on at NIST SIRCUS-type stuff (tunable lasers) now migrating to LASP Absolute Spectrally-Tunable Detector-Based Source Spectrally-programmable source calibrated via NIST
More informationSpectroMaster. High Precision Automatic Spectrometer-Goniometer
SpectroMaster High Precision Automatic Spectrometer-Goniometer CONTENTS Overview 3 Measurement Principle 3 Error A naly sis and System Requirements 4 Goniometer Error...4 Ambient Conditions...6 Sample
More informationCHAPTER 7. Components of Optical Instruments
CHAPTER 7 Components of Optical Instruments From: Principles of Instrumental Analysis, 6 th Edition, Holler, Skoog and Crouch. CMY 383 Dr Tim Laurens NB Optical in this case refers not only to the visible
More informationOperating Manual. Model 721N. Visible Spectrophotometer
Operating Manual of Model 721N Visible Spectrophotometer 1 Table of Contents 1. Chief uses... 3 2. Working Conditions... 3 3. Main Specifications...3 4.Operating Principles...4 5. Optical design...4 6.
More informationPhotometry and Light Measurement
Photometry and Light Measurement Adrian Waltho, Analytik Ltd adrian.waltho@analytik.co.uk What is Light? What is Light? What is Light? Ultraviolet Light UV-C 180-280 nm UV-B 280-315 nm UV-A 315-400 nm
More informationSystems & Accessories
Light Source Sample Chambers Stepping Motor Controller Detector Variable Wavelength Fiber Optics Modules Tunable Light Sources For Applications In: Analytical Chemistry Physics Life Sciences Engineering
More informationChapter 7. Optical Measurement and Interferometry
Chapter 7 Optical Measurement and Interferometry 1 Introduction Optical measurement provides a simple, easy, accurate and reliable means for carrying out inspection and measurements in the industry the
More informationCalibration of ARM Spectral Shortwave Radiometers
Calibration of ARM Spectral Shortwave Radiometers J. J. Michalsky, J. L. Berndt, P. W. Kiedron, and L. C. Harrison Atmospheric Sciences Research Center State University of New York at Albany Albany, New
More informationExperiment 1: Fraunhofer Diffraction of Light by a Single Slit
Experiment 1: Fraunhofer Diffraction of Light by a Single Slit Purpose 1. To understand the theory of Fraunhofer diffraction of light at a single slit and at a circular aperture; 2. To learn how to measure
More informationPoS(PhotoDet 2012)058
Absolute Photo Detection Efficiency measurement of Silicon PhotoMultipliers Vincent CHAUMAT 1, Cyril Bazin, Nicoleta Dinu, Véronique PUILL 1, Jean-François Vagnucci Laboratoire de l accélérateur Linéaire,
More informationMiniaturized Spectroradiometer
Miniaturized Spectroradiometer Thomas Morgenstern, Gudrun Bornhoeft, Steffen Goerlich JETI Technische Instrumente GmbH, Jena, Germany Abstract This paper describes the basics of spectroradiometric instruments
More information