1 SR-5000N design: spectroradiometer's new performance improvements in FOV response uniformity (flatness) scan speed and other important features Dario Cabib *, Shmuel Shapira, Moshe Lavi, Amir Gil and Uri Milman CI Systems (Israel) Ltd., Ramat Gavriel, Migdal Haemek 10551, Israel, ABSTRACT As far as we know, CI has been the only manufacturer in the world of commercial visible/infrared spectroradiometers for remote sensing applications for many years. In this paper we describe the new design and some performance improvements that we are developing to renew and modernize the system. Simultaneous visible and infrared spectroradiometry, field of view flatness of response and scan speed are only some aspects of the system which have undergone significant improvement. The challenge is to achieve these functional improvements without losing any of the advantages of the traditional system as far as ruggedness for field use, interchangeability of spectral range and field of view and aiming and image recording facility. Important factors in the success of this endeavor are: i) the development of a new electronic signal processing package, ii) a modular optical concept that mechanically separates modules for small, medium and large field of view ranges, and iii) a compact overall shape for convenience of use. Actual instrument performance results will be reported in a future paper. Keywords: Spectroradiometer, spectral radiometers, infrared spectroradiometer 1. INTRODUCTION The first product developed and sold by CI Systems when the company was founded in the late 70 s was an infrared spectroradiometer, the SR 1000, a completely analog system, since at the time there were no personal computers and the measured spectra were plotted on millimetric paper. The company s know-how in the field of infrared spectroradiometry became during the years the basis for much of the later products in other fields of infrared applications. In the mid 80 s the spectroradiometric capability was extended to the visible, Near and Short Infrared (VIS/NIR and SWIR) wavelength ranges and the instrument was turned into PC based: spectra were since plotted on the computer screen and saved in digital form in its hard disk. Operated with DOS it was called SR A short historical review of CI s involvement in spectroradiometry is given in reference 1: in this paper, published in 2004, we described the transition of the system s software from DOS to Windows. Additional work was done in 2006 and described in reference 2, when CI developed a high sensitivity version of the SR 5000 for special applications of infrared equipment testing. Recently, CI decided to renew and restructure this product: motivated mostly by the obsolescence of many electronic components in the system, CI is taking the opportunity to introduce many opto-mechanical and user interface improvements, which will hopefully make the instrument more useful and easier to use. The instrument is called SR 5000N (SR for Spectroradiometer and N for New ), a prototype of which is now being built. In this paper we mention the improvement considerations, the new design, and the expected performance values of the important parameters. In a future paper we will report actual performance results based on system testing. 2. SYSTEM IMPROVEMENTS In order to understand the motivations for this work, let us summarize the main characteristics of the SR 5000, the traditional instrument being used now for more than two decades, first under DOS and then Windows operating systems. First a definition: A spectroradiometer is an instrument that measures the total spectral emission (self or reflected) of the objects within its field of view in absolute units of Watts/(steradian.cm 2.μ) as function of wavelength. * dario.cabib@ci-systems.com Infrared Imaging Systems: Design, Analysis, Modeling, and Testing XXII, edited by Gerald C. Holst, Keith A. Krapels, Proc. of SPIE Vol. 8014, SPIE CCC code: X/11/$18 doi: / Proc. of SPIE Vol

2 High performance in a list of critical features is most important and useful for most field applications. Most of the parameter values and the design details on the engineering solutions of the SR 5000 can be found in the cited references 1 and Measurement speed in both radiometric (chopped) and transient (unchopped) modes; the former will be improved by almost a factor of two from 30 to 50 scans/second, the latter by a factor five from 20 to 100Ksamples/second. 2. Uniformity of response as function of position of a point source (in short called FOV flatness ) within the field of view of the instrument (partially achieved in the SR 5000 with focal point blurring on the detector); it will be significantly improved from a deviation from flatness of about 11% of the SR 5000 down to a worst case of 5%, where the optics contribution is designed to be only 1%. 3. The ability to measure spectra in as wide wavelength range as possible in one scan (this is done in the SR 5000 with two pairs of stacked detectors, called sandwich detectors, with wide range reflective optics and with two interchangeable Circular Variable Filters or CVF s, one covering the visible and near infrared and one the mid and long wave infrared ranges; however, the visible to far infrared ranges cannot be covered simultaneously in one measurement); the additional visible/near infrared optical channel equipped with a visible spectrometer in the SR 5000N allows the two wide spectral ranges to be covered simultaneously in one measurement. 4. The total dynamic range of the SR 5000 is 4x10 6, covered with four x10 gain ranges and 14 bits digitization in each range (of which 12 significant). In order to get this full dynamic range in one measurement automatic gain control (AGC) must be used, but this acquisition mode has significant gain switch noise and limits the scan rate to 0.5 scan/second. In the SR 5000N the signal is sampled simultaneously at two gain ranges (automatically selected for best signal to noise ratio) with 16 bits digitization and a gain ratio of 100. The final spectrum is recorded and displayed by joining spectral sections obtained from the gain range giving the better signal to noise ratio of the two. As a result, taking into account that only 14 are significant out of the 16 bits, and the factor 100 between the two ranges, the final dynamic range of each measurement is 1.6x10 6 with lower noise and usable up to 20 scans/second. 5. A facility to accurately point the instrument on the target or object to be measured (done in the SR 5000 with a co-aligned optical channel for viewing the measured field of view and its background without parallax and a boresighted telescope); in the SR 5000N pointing, focusing and image recording is implemented with a boresighted CCD without parallax. 6. A user friendly software package that allows: i) automatic recognition of the instrument configuration, ii) measurement parameter settings, iii) data acquisition and simultaneous visualization, iv) calibration and other algorithms for spectral handling, v) spectral display, vi) spectral data saving and exporting to universally readable formats and vii) data files management. 7. Connectivity will be improved from IEEE-488 bus to single 1G LAN. Figure 1 shows the optical diagram of the instrument. 3. OPTICAL DESIGN OF THE SR 5000N Proc. of SPIE Vol

3 Focusing collecting mirror Flat 45 0 folding mirror Spectrometer for visible region See figure 2 CCD for viewing Chopper Incoming rays from field of view Dichroic filter Figure 1: General view of the optics layout of the SR 5000N for the 7.5 millirad. field. Region in dashed zone is enlarged in figure 2. The incoming radiation from the field of view is shown for convenience as if coming from the bottom of the figure but in practice the 45 0 folding mirror is rotated 90 0 around a horizontal axis passing from its center and parallel to the plane of the paper, soo that the radiation comes perpendicularly into the plane of the paper. Chopping and internal blackbody reference for high performance NET is maintained in this new design. However, a significant improvement in field of view uniformity of response as function of point source position is achieved by the detection optics concept adopted here. Referring to figure 1, the collection optics is an F#3 on-axis Newtonian system with 120 mm. clear aperture. The radiation to be measured from the field of view of the instrument (FOV) is first reflected by a large 45 0 flat folding mirror with a central aperture. The reflected rays travel towards the focusing collecting mirror for the infrared spectral measurement while the ones through the central aperture of the 45 0 flat mirror are collected by the spectrometer for the visible range measurement. This design allows spectra in the infrared and visible range (up to 1.1 microns) to be measured simultaneously. Before being focused on the field stop (see figure 2) the beam encounters a dichroic filter at 45 0 which reflects the infrared radiation to be measured toward the field stop and transmits the visible radiation towards a CCD for viewing and recording images of the measured FOV and its surroundings. The CCD channel is boresighted with the measurement channel. Proc. of SPIE Vol

4 Infrared detector window Detector Detector lens Focal plane with a number of different size interchangeable field stops Circular Variable Filter (CVF) Internal reference blackbody Dichroic filter Chopper CCD Figure 2: Detection optics: enlargement of dashed zone of figure 1. Figure 2 shows the enlarged detection system in the dashed zone of figure 1. After being chopped at high frequency (several thousand Hertz depending on the CVF scanning speed selected) by a chopper placed at 45 0, the beam is focused on the field stop, which is a motorized computer controlled plate with different size apertures: this allows the selection of the FOV size from 0.5 to 7.5 milliradians. The 45 0 chopper configuration allows the radiation from an internal floating temperature blackbody to be reflected towards the detector: this is the baseline blackbody radiation which will be used in the calibration procedure to provide the measured spectral radiance of the FOV in units of Watts/(sr.cm 2.μ). Figure 3 shows the exterior look of the instrument mounted on a tripod. The view of figure 3 shows the back side of the instrument: the measured radiation is incident horizontally from the opposite side of the figure. Laptop for functional control and data storage and calibration Interchangeable detector module CCD image of the scene on a built-in screen: central black circle is the measured FOV Interchangeable collection optics module (shown for 7.5 mrad. FOV) Electrical connectors to computer and mains Proc. of SPIE Vol

5 Referring to figure 3, both the detector and collection optics modules are interchangeable units without need of optical realignment. The detector can be a liquid nitrogen cooled sandwich InSb/MCT for medium and long infrared ranges, or an InGaAs for short infrared or a silicon detector for visible spectroradiometry. The collection optics module of figure 3 can be swapped with two different wider angle collection modules, one for a maximum of 5 degrees and one for 10 degrees. Figure 4 shows the optical elements used for the wider angles operation. All three optical collection modules (7.5 millirad., 5 0 and 10 0 ) are suitable for all wavelength ranges (visible, short wave infrared, medium and long wave infrared from 0.4 to 15 microns). However, whereas in each case the infrared channel has a selectable FOV, the visible spectrometer channel FOV is fixed to the nominal value. Spectrometer for visible region Detector Folding 45 0 mirror Interchangeable field stops CVF CCD for viewing Incoming radiation from FOV Wide angle collection lens for 5 0 or 10 0 CCD for viewing Figure 4: Optical configuration for the wide angle operation of 5 degrees and 10 degrees. The collection lenses have different focal lengths in the two situations. The folding mirror and lens make up an interchangeable module for each of the 5 0 and 10 0 situations. The dichroic filter, CCD, chopper, field stop, CVF and detector with its own lens are the same as the ones in figure 1 except that the CVF and detectors are interchanged with the ones suitable for the different spectral ranges. The FOV uniformity of response to signals from point sources is made significantly more constant than today s SR This is achieved by designing the detector lens to image the instrument entrance pupil on the detector element instead of the field stop. The result is that variations of response on the detector surface and variations of direction dependent optical transmission are averaged out and the calculated instrument response as function of point source position in the FOV for the 7.5 milliradians optics is as shown in figure 5. Proc. of SPIE Vol

6 Normalized response to point source at infinity Angle (milliradians) Figure 5: Response versus angle. The calculation is done by the Zemax program, as an energy diagram through all the optical elements in the optics train. Variations from constant behavior are in the order of 1% instead of 10% in the opresent SR Detector module CVF module being replaced Collection optics module for 7.5 millirad. FOV Optics back panel Figure 6: The instrument is shown here being opened for CVF interchange. 4. DESIGNED PARAMETER VALUES The design specifications of the important parameters of the system are as in the following table. In a future paper we will present the actually measured values. Specification Value Notes Narrow FOV (NFOV) 7.5 to 0.5 mrad Manual focus 3m to infinity. f/ 3.3, All reflective ±5% FOV uniformity (max). 5 diameter aperture Optics and FOV Spectral range and resolution Medium FOV (MFOV) 5 to for the infrared channel. Fixed to 5 0 for the spectrometer channel. Wide FOV (WFOV) 9.4 to for the infrared channel. Fixed to 10 0 for the spectrometer channel. 0.2 to 25 µm Fixed focus to infinity. Minimal working distance 10m. ±5% FOV uniformity (max). Fixed focus to infinity. Minimal working distance 5m. ±8% FOV uniformity (max). Integrated Spectrometer µm; spectral resolution < 3nm Proc. of SPIE Vol

7 Specification Value Notes CVF/Detector combinations µm spectral resolution < 2% of wavelength Bandpass Filter/Detector combinations throughout the entire spectral range Sampling and scan rates Aiming, focusing and FOV image recording Detector types Up to 20 scans/sec High sensitivity spectral mode Up to 50 scans/sec High speed spectral mode >4 Ksamples/sec High sensitivity radiometric mode >100 Ksamples/sec High speed radiometric mode Non parallax integrated CCD camera Si, InGaAs, InSb, MCT, Pyro With FOV designation overlay LN2 or Stirling Cycle cooling, TEC cooling Internal Blackbody reference Floating ambient temperature reference Monitoring accuracy > 0.1 C Chopping frequency Noise performance >4000 Hz Dynamic range 1:1000 Signal sampling digitization 16-bits. Wavelength and detector dependant. Internal or external (capable of locking onto a remote chopped source) Example: < 5 mdeg NET with InSb detector and CVF at 5µm, 1 Hz bandwidth, 100 C target filling 7.5 mrad FOV Simultaneous sampling of two detectors in two gain levels. 5. SPECTRAL RADIANCE CALIBRATION The calibration algorithm and procedure is as in section 3 of reference 1. The measured radiance spectrum is obtained by scaling the detector signal at each wavelength (which is linear in the incident radiance difference between the source and the internal blackbody) with the ratio of the signal received from a known temperature blackbody source filling the field of view of the instrument and the Planck function value at that wavelength and temperature; the result is then added wavelength by wavelength to the Planck function at the internal blackbody temperature. 6. CONCLUSIONS In this paper we have described the optical design, main characteristics and performance of a new visible/infrared spectroradiometer (SR 5000N) being developed at present. Actual performance values according to system testing will be presented in a future paper. REFERENCES [1] Cabib, Dario et al., New user interface and features of the SR 5000: revival of infrared CVF based spectroradiometry, Proc. SPIE 5431, p. 46, (2004). [2] Cabib, Dario et al., High performance spectroradiometer for very accurate radiometric calibrations and testing of blackbody sources and EO test equipment, Proc. SPIE 6207, 62070L, (2006). Proc. of SPIE Vol

Dario Cabib, Amir Gil, Moshe Lavi. Edinburgh April 11, 2011

Dario Cabib, Amir Gil, Moshe Lavi. Edinburgh April 11, 2011 New LWIR Spectral Imager with uncooled array SI-LWIR LWIR-UC Dario Cabib, Amir Gil, Moshe Lavi Edinburgh April 11, 2011 Contents BACKGROUND AND HISTORY RATIONALE FOR UNCOOLED CAMERA BASED SPECTRAL IMAGER

More information

Calibration of a High Dynamic Range, Low Light Level Visible Source

Calibration of a High Dynamic Range, Low Light Level Visible Source Calibration of a High Dynamic Range, Low Light Level Visible Source Joe LaVeigne a, Todd Szarlan a, Nate Radtke a a Santa Barbara Infrared, Inc., 30 S. Calle Cesar Chavez, #D, Santa Barbara, CA 93103 ABSTRACT

More information

New 3-5 µ wavelength range hyperspectral imager for ground and airborne use based on a single element interferometer

New 3-5 µ wavelength range hyperspectral imager for ground and airborne use based on a single element interferometer New 3-5 µ wavelength range hyperspectral imager for ground and airborne use based on a single element interferometer Dario Cabib, Amir Gil, Moshe Lavi, Robert A. Buckwald CI Systems Ltd., Industrial Park

More information

Evaluation of infrared collimators for testing thermal imaging systems

Evaluation of infrared collimators for testing thermal imaging systems OPTO-ELECTRONICS REVIEW 15(2), 82 87 DOI: 10.2478/s11772-007-0005-9 Evaluation of infrared collimators for testing thermal imaging systems K. CHRZANOWSKI *1,2 1 Institute of Optoelectronics, Military University

More information

Observational Astronomy

Observational Astronomy Observational Astronomy Instruments The telescope- instruments combination forms a tightly coupled system: Telescope = collecting photons and forming an image Instruments = registering and analyzing the

More information

LSST All-Sky IR Camera Cloud Monitoring Test Results

LSST All-Sky IR Camera Cloud Monitoring Test Results LSST All-Sky IR Camera Cloud Monitoring Test Results Jacques Sebag a, John Andrew a, Dimitri Klebe b, Ronald D. Blatherwick c a National Optical Astronomical Observatory, 950 N Cherry, Tucson AZ 85719

More information

Instructions for the Experiment

Instructions for the Experiment Instructions for the Experiment Excitonic States in Atomically Thin Semiconductors 1. Introduction Alongside with electrical measurements, optical measurements are an indispensable tool for the study of

More information

LINEARPYROMETER LP4. Technical Documentation KE November TN

LINEARPYROMETER LP4. Technical Documentation KE November TN 1 LINEARPYROMETER LP4 Technical Documentation KE 256-6.2007 November 2010 5-TN-1622-100 2 1. General Description With the Linearpyrometer Type LP4 a measuring instrument has been made available for pyrometric

More information

2.2 Wavefront Sensor Design. Lauren H. Schatz, Oli Durney, Jared Males

2.2 Wavefront Sensor Design. Lauren H. Schatz, Oli Durney, Jared Males Page: 1 of 8 Lauren H. Schatz, Oli Durney, Jared Males 1 Pyramid Wavefront Sensor Overview The MagAO-X system uses a pyramid wavefront sensor (PWFS) for high order wavefront sensing. The wavefront sensor

More information

Coherent Laser Measurement and Control Beam Diagnostics

Coherent Laser Measurement and Control Beam Diagnostics Coherent Laser Measurement and Control M 2 Propagation Analyzer Measurement and display of CW laser divergence, M 2 (or k) and astigmatism sizes 0.2 mm to 25 mm Wavelengths from 220 nm to 15 µm Determination

More information

NASTER System Definition Proposal

NASTER System Definition Proposal Remote Sensing Team NASTER System Definition Proposal All rights reserved. - 7/14/03 Page 1 Overview Review and comment the mid-ir requirements Presentation of ABB s current platform technology Proposed

More information

NIRCam optical calibration sources

NIRCam optical calibration sources NIRCam optical calibration sources Stephen F. Somerstein, Glen D. Truong Lockheed Martin Advanced Technology Center, D/ABDS, B/201 3251 Hanover St., Palo Alto, CA 94304-1187 ABSTRACT The Near Infrared

More information

VSR VERSATILE SPECTRO-RADIOMETER FOR INFRARED APPLICATIONS PERFORMANCE WITHOUT COMPROMISE

VSR VERSATILE SPECTRO-RADIOMETER FOR INFRARED APPLICATIONS PERFORMANCE WITHOUT COMPROMISE VSR VERSATILE SPECTRO-RADIOMETER FOR INFRARED APPLICATIONS LR Tech inc. 47 Saint-Joseph street Lévis, Qc, G6V 1A8 Canada lrtech.ca PERFORMANCE WITHOUT COMPROMISE DISCLAIMER This product description document

More information

PHY385H1F Introductory Optics Term Test 2 November 6, 2012 Duration: 50 minutes. NAME: Student Number:.

PHY385H1F Introductory Optics Term Test 2 November 6, 2012 Duration: 50 minutes. NAME: Student Number:. PHY385H1F Introductory Optics Term Test 2 November 6, 2012 Duration: 50 minutes NAME: Student Number:. Aids allowed: A pocket calculator with no communication ability. One 8.5x11 aid sheet, written on

More information

Using molded chalcogenide glass technology to reduce cost in a compact wide-angle thermal imaging lens

Using molded chalcogenide glass technology to reduce cost in a compact wide-angle thermal imaging lens Using molded chalcogenide glass technology to reduce cost in a compact wide-angle thermal imaging lens George Curatu a, Brent Binkley a, David Tinch a, and Costin Curatu b a LightPath Technologies, 2603

More information

MS260i 1/4 M IMAGING SPECTROGRAPHS

MS260i 1/4 M IMAGING SPECTROGRAPHS MS260i 1/4 M IMAGING SPECTROGRAPHS ENTRANCE EXIT MS260i Spectrograph with 3 Track Fiber on input and InstaSpec IV CCD on output. Fig. 1 OPTICAL CONFIGURATION High resolution Up to three gratings, with

More information

Radiometric Solar Telescope (RaST) The case for a Radiometric Solar Imager,

Radiometric Solar Telescope (RaST) The case for a Radiometric Solar Imager, SORCE Science Meeting 29 January 2014 Mark Rast Laboratory for Atmospheric and Space Physics University of Colorado, Boulder Radiometric Solar Telescope (RaST) The case for a Radiometric Solar Imager,

More information

PROCEEDINGS OF SPIE. Automated asphere centration testing with AspheroCheck UP

PROCEEDINGS OF SPIE. Automated asphere centration testing with AspheroCheck UP PROCEEDINGS OF SPIE SPIEDigitalLibrary.org/conference-proceedings-of-spie Automated asphere centration testing with AspheroCheck UP F. Hahne, P. Langehanenberg F. Hahne, P. Langehanenberg, "Automated asphere

More information

Horiba Jobin-Yvon LabRam Raman Confocal Microscope (GERB 120)

Horiba Jobin-Yvon LabRam Raman Confocal Microscope (GERB 120) Horiba Jobin-Yvon LabRam Raman Confocal Microscope (GERB 120) Please contact Dr. Amanda Henkes for training requests and assistance: 979-862-5959, amandahenkes@tamu.edu Hardware LN 2 FTIR FTIR camera 1

More information

UltraGraph Optics Design

UltraGraph Optics Design UltraGraph Optics Design 5/10/99 Jim Hagerman Introduction This paper presents the current design status of the UltraGraph optics. Compromises in performance were made to reach certain product goals. Cost,

More information

Guide to SPEX Optical Spectrometer

Guide to SPEX Optical Spectrometer Guide to SPEX Optical Spectrometer GENERAL DESCRIPTION A spectrometer is a device for analyzing an input light beam into its constituent wavelengths. The SPEX model 1704 spectrometer covers a range from

More information

Spatially Resolved Backscatter Ceilometer

Spatially Resolved Backscatter Ceilometer Spatially Resolved Backscatter Ceilometer Design Team Hiba Fareed, Nicholas Paradiso, Evan Perillo, Michael Tahan Design Advisor Prof. Gregory Kowalski Sponsor, Spectral Sciences Inc. Steve Richstmeier,

More information

Development and Applications of a Sample Compartment FTIR Microscope

Development and Applications of a Sample Compartment FTIR Microscope Application Note Development and Applications of a Sample Since the early to mid-1940 s, scientists using infrared spectroscopy have been trying to obtain spectral data from ever smaller samples. Starting

More information

ECEN. Spectroscopy. Lab 8. copy. constituents HOMEWORK PR. Figure. 1. Layout of. of the

ECEN. Spectroscopy. Lab 8. copy. constituents HOMEWORK PR. Figure. 1. Layout of. of the ECEN 4606 Lab 8 Spectroscopy SUMMARY: ROBLEM 1: Pedrotti 3 12-10. In this lab, you will design, build and test an optical spectrum analyzer and use it for both absorption and emission spectroscopy. The

More information

Supplementary Materials

Supplementary Materials Supplementary Materials In the supplementary materials of this paper we discuss some practical consideration for alignment of optical components to help unexperienced users to achieve a high performance

More information

Application Note (A11)

Application Note (A11) Application Note (A11) Slit and Aperture Selection in Spectroradiometry REVISION: C August 2013 Gooch & Housego 4632 36 th Street, Orlando, FL 32811 Tel: 1 407 422 3171 Fax: 1 407 648 5412 Email: sales@goochandhousego.com

More information

PROCEEDINGS OF SPIE. Measurement of low-order aberrations with an autostigmatic microscope

PROCEEDINGS OF SPIE. Measurement of low-order aberrations with an autostigmatic microscope PROCEEDINGS OF SPIE SPIEDigitalLibrary.org/conference-proceedings-of-spie Measurement of low-order aberrations with an autostigmatic microscope William P. Kuhn Measurement of low-order aberrations with

More information

Applications of Steady-state Multichannel Spectroscopy in the Visible and NIR Spectral Region

Applications of Steady-state Multichannel Spectroscopy in the Visible and NIR Spectral Region Feature Article JY Division I nformation Optical Spectroscopy Applications of Steady-state Multichannel Spectroscopy in the Visible and NIR Spectral Region Raymond Pini, Salvatore Atzeni Abstract Multichannel

More information

MR-i. Hyperspectral Imaging FT-Spectroradiometers Radiometric Accuracy for Infrared Signature Measurements

MR-i. Hyperspectral Imaging FT-Spectroradiometers Radiometric Accuracy for Infrared Signature Measurements MR-i Hyperspectral Imaging FT-Spectroradiometers Radiometric Accuracy for Infrared Signature Measurements FT-IR Spectroradiometry Applications Spectroradiometry applications From scientific research to

More information

1/8 m GRATING MONOCHROMATOR

1/8 m GRATING MONOCHROMATOR 1/8 m GRATING GRATING OUTPUT PORT INPUT PORT 77250 1/8 m Monochromator with 6025 Hg(Ar) Spectral Calibration Lamp. Low cost, compact size and high performance, ideal for OEM applications Very efficient

More information

MR-i. Hyperspectral Imaging FT-Spectroradiometers Radiometric Accuracy for Infrared Signature Measurements

MR-i. Hyperspectral Imaging FT-Spectroradiometers Radiometric Accuracy for Infrared Signature Measurements MR-i Hyperspectral Imaging FT-Spectroradiometers Radiometric Accuracy for Infrared Signature Measurements FT-IR Spectroradiometry Applications Spectroradiometry applications From scientific research to

More information

Oriel MS260i TM 1/4 m Imaging Spectrograph

Oriel MS260i TM 1/4 m Imaging Spectrograph Oriel MS260i TM 1/4 m Imaging Spectrograph MS260i Spectrograph with 3 Track Fiber on input and InstaSpec CCD on output. The MS260i 1 4 m Imaging Spectrographs are economical, fully automated, multi-grating

More information

Use of Mangin and aspheric mirrors to increase the FOV in Schmidt- Cassegrain Telescopes

Use of Mangin and aspheric mirrors to increase the FOV in Schmidt- Cassegrain Telescopes Use of Mangin and aspheric mirrors to increase the FOV in Schmidt- Cassegrain Telescopes A. Cifuentes a, J. Arasa* b,m. C. de la Fuente c, a SnellOptics, Prat de la Riba, 35 local 3, Interior Terrassa

More information

Current and Future Realizations NRC Photometric and Spectroradiometric Calibration Chains

Current and Future Realizations NRC Photometric and Spectroradiometric Calibration Chains Measurement Science and Standards (MSS) Current and Future Realizations NRC Photometric and Spectroradiometric Calibration Chains Arnold A. Gaertner Photometry, Radiometry and Thermometry (PRT) CORM 30

More information

LASP / University of Colorado

LASP / University of Colorado The SORCE SIM Instrument: Progress Toward Spectral Irradiance Time Series Throughout the 300-3000 nm Region. Jerald Harder, Juan Fontenla, Byron Smiley, Sean Davis, George Lawrence and Gary Rottman LASP

More information

Non-optically Combined Multi-spectral Source for IR, Visible, and Laser Testing

Non-optically Combined Multi-spectral Source for IR, Visible, and Laser Testing Non-optically Combined Multi-spectral Source for IR, Visible, and Laser Testing Joe LaVeigne a, Brian Rich a, Steve McHugh a, Peter Chua b a Santa Barbara Infrared, Inc., 30 S. Calle Cesar Chavez, #D,

More information

SPECTRAL SCANNER. Recycling

SPECTRAL SCANNER. Recycling SPECTRAL SCANNER The Spectral Scanner, produced on an original project of DV s.r.l., is an instrument to acquire with extreme simplicity the spectral distribution of the different wavelengths (spectral

More information

A collection of hyperspectral images for imaging systems research Torbjørn Skauli a,b, Joyce Farrell *a

A collection of hyperspectral images for imaging systems research Torbjørn Skauli a,b, Joyce Farrell *a A collection of hyperspectral images for imaging systems research Torbjørn Skauli a,b, Joyce Farrell *a a Stanford Center for Image Systems Engineering, Stanford CA, USA; b Norwegian Defence Research Establishment,

More information

Optoliner NV. Calibration Standard for Sighting & Imaging Devices West San Bernardino Road West Covina, California 91790

Optoliner NV. Calibration Standard for Sighting & Imaging Devices West San Bernardino Road West Covina, California 91790 Calibration Standard for Sighting & Imaging Devices 2223 West San Bernardino Road West Covina, California 91790 Phone: (626) 962-5181 Fax: (626) 962-5188 www.davidsonoptronics.com sales@davidsonoptronics.com

More information

Technical Synopsis and Discussion of:

Technical Synopsis and Discussion of: OPTI-521, Fall 2008 E.D. Fasse, Page 1 Technical Synopsis and Discussion of: Optical Alignment of a Pupil Imaging Spectrometer by Stephen Horchem and Richard Kohrman Proc. of SPIE Vol. 1167, Precision

More information

4DAD, a device to align angularly and laterally a high power laser using a conventional sighting telescope as metrology

4DAD, a device to align angularly and laterally a high power laser using a conventional sighting telescope as metrology 4DAD, a device to align angularly and laterally a high power laser using a conventional sighting telescope as metrology Christophe DUPUY, Thomas PFROMMER, Domenico BONACCINI CALIA European Southern Observatory,

More information

Compact Dual Field-of-View Telescope for Small Satellite Payloads

Compact Dual Field-of-View Telescope for Small Satellite Payloads Compact Dual Field-of-View Telescope for Small Satellite Payloads James C. Peterson Space Dynamics Laboratory 1695 North Research Park Way, North Logan, UT 84341; 435-797-4624 Jim.Peterson@sdl.usu.edu

More information

7x P/N C1601. General Description

7x P/N C1601. General Description METRICZOOM SWIR 7x METRIC ZOOM-SWIR ZOOM 7x P/N C1601 C General Description This family of high resolution METRIC ZOOM SWIR lenses image from 0.9 to 2.3 µm making them especially well-suited well for surveillance,

More information

NIST Agency Report May 2012 OUTLINE. The case for traceability NMI capabilities A view to the future the HIP Current/recent NIST activities

NIST Agency Report May 2012 OUTLINE. The case for traceability NMI capabilities A view to the future the HIP Current/recent NIST activities NIST Agency Report May 2012 OUTLINE The case for traceability NMI capabilities A view to the future the HIP Current/recent NIST activities The case for traceability Earth Radiation Budget: Solar irradiance

More information

Spectral Analysis of the LUND/DMI Earthshine Telescope and Filters

Spectral Analysis of the LUND/DMI Earthshine Telescope and Filters Spectral Analysis of the LUND/DMI Earthshine Telescope and Filters 12 August 2011-08-12 Ahmad Darudi & Rodrigo Badínez A1 1. Spectral Analysis of the telescope and Filters This section reports the characterization

More information

SR-80 EXTENDED AREA BLACKBODY

SR-80 EXTENDED AREA BLACKBODY SR-80 EXTENDED AREA BLACKBODY PRODUCT TECHNICAL DESCRIPTION Rev. C Prepared by CI Systems, Inc. 30961 West Agoura Road, Suite 109 Westlake Village, CA 91361-4618 Tel: 818-865-0402 Fax: 818-865-0403 Email:

More information

Filters for Dual Band Infrared Imagers

Filters for Dual Band Infrared Imagers Filters for Dual Band Infrared Imagers Thomas D. Rahmlow, Jr.* a, Jeanne E. Lazo-Wasem a, Scott Wilkinson b, and Flemming Tinker c a Rugate Technologies, Inc., 353 Christian Street, Oxford, CT 6478; b

More information

LTE. Tester of laser range finders. Integrator Target slider. Transmitter channel. Receiver channel. Target slider Attenuator 2

LTE. Tester of laser range finders. Integrator Target slider. Transmitter channel. Receiver channel. Target slider Attenuator 2 a) b) External Attenuators Transmitter LRF Receiver Transmitter channel Receiver channel Integrator Target slider Target slider Attenuator 2 Attenuator 1 Detector Light source Pulse gene rator Fiber attenuator

More information

Tailored bar concepts for 10 mm-mrad fiber coupled modules scalable to kw-class direct diode lasers

Tailored bar concepts for 10 mm-mrad fiber coupled modules scalable to kw-class direct diode lasers Tailored bar concepts for 1 mm-mrad fiber coupled modules scalable to kw-class direct diode lasers Andreas Unger*, Ross Uthoff, Michael Stoiber, Thomas Brand, Heiko Kissel, Bernd Köhler, Jens Biesenbach

More information

Modular Fourier Transform Infra-Red Spectrometer (FT-IR) Solution

Modular Fourier Transform Infra-Red Spectrometer (FT-IR) Solution FINE XTG MONITORS BRIGHTNESS CONTRAST Modular Fourier Transform Infra-Red Spectrometer (FT-IR) Solution The Oriel FTIR Building Block solution has been designed for routine analytical applications for

More information

Table of Content. Fiber-Coupled LED s Light-Guide-Coupled LED s LED Collimator Sources Low-cost LED Spot Lights...

Table of Content. Fiber-Coupled LED s Light-Guide-Coupled LED s LED Collimator Sources Low-cost LED Spot Lights... LIGHT SOURCES Table of Content Fiber-Coupled s... 40 -Guide-Coupled s... 41 Collimator... 42 Low-cost Spot s... 43 Precision Spot s... 45 Spectrum Synthesizing ( Cubic S )... 46 Spectrometers 39 sources

More information

Capabilities of NIST SIRCUS for Calibrations of SSI Vis-IR Instruments

Capabilities of NIST SIRCUS for Calibrations of SSI Vis-IR Instruments Capabilities of NIST SIRCUS for Calibrations of SSI Vis-IR Instruments Steve Brown National Institute of Standards & Technology Gaithersburg, MD steve.brown@nist.gov; 301.975.5167 Answer: Ask LASP folks

More information

DEFENSE APPLICATIONS IN HYPERSPECTRAL REMOTE SENSING

DEFENSE APPLICATIONS IN HYPERSPECTRAL REMOTE SENSING DEFENSE APPLICATIONS IN HYPERSPECTRAL REMOTE SENSING James M. Bishop School of Ocean and Earth Science and Technology University of Hawai i at Mānoa Honolulu, HI 96822 INTRODUCTION This summer I worked

More information

Norsk Elektro Optikk AS (NEO) HySpex Airborne Sensors System Overview

Norsk Elektro Optikk AS (NEO) HySpex Airborne Sensors System Overview Norsk Elektro Optikk AS (NEO) HySpex Airborne Sensors System Overview Trond Løke Research Scientist EUFAR meeting 14.04.2011 Outline Norsk Elektro Optikk AS (NEO) NEO company profile HySpex Optical Design

More information

Hyperspectral goes to UAV and thermal

Hyperspectral goes to UAV and thermal Hyperspectral goes to UAV and thermal Timo Hyvärinen, Hannu Holma and Esko Herrala SPECIM, Spectral Imaging Ltd, Finland www.specim.fi Outline Roadmap to more compact, higher performance hyperspectral

More information

COLOUR INSPECTION, INFRARED AND UV

COLOUR INSPECTION, INFRARED AND UV COLOUR INSPECTION, INFRARED AND UV TIPS, SPECIAL FEATURES, REQUIREMENTS LARS FERMUM, CHIEF INSTRUCTOR, STEMMER IMAGING THE PROPERTIES OF LIGHT Light is characterized by specifying the wavelength, amplitude

More information

PRODUCT OVERVIEW FOR THE. Corona 350 II FLIR SYSTEMS POLYTECH AB

PRODUCT OVERVIEW FOR THE. Corona 350 II FLIR SYSTEMS POLYTECH AB PRODUCT OVERVIEW FOR THE Corona 350 II FLIR SYSTEMS POLYTECH AB Table of Contents Table of Contents... 1 Introduction... 2 Overview... 2 Purpose... 2 Airborne Data Acquisition and Management Software (ADAMS)...

More information

Inverted-COR: Inverted-Occultation Coronagraph for Solar Orbiter

Inverted-COR: Inverted-Occultation Coronagraph for Solar Orbiter Inverted-COR: Inverted-Occultation Coronagraph for Solar Orbiter OATo Technical Report Nr. 119 Date 19-05-2009 by: Silvano Fineschi Release Date Sheet: 1 of 1 REV/ VER LEVEL DOCUMENT CHANGE RECORD DESCRIPTION

More information

Optical Design of an Off-axis Five-mirror-anastigmatic Telescope for Near Infrared Remote Sensing

Optical Design of an Off-axis Five-mirror-anastigmatic Telescope for Near Infrared Remote Sensing Journal of the Optical Society of Korea Vol. 16, No. 4, December 01, pp. 343-348 DOI: http://dx.doi.org/10.3807/josk.01.16.4.343 Optical Design of an Off-axis Five-mirror-anastigmatic Telescope for Near

More information

An integral eld spectrograph for the 4-m European Solar Telescope

An integral eld spectrograph for the 4-m European Solar Telescope Mem. S.A.It. Vol. 84, 416 c SAIt 2013 Memorie della An integral eld spectrograph for the 4-m European Solar Telescope A. Calcines 1,2, M. Collados 1,2, and R. L. López 1 1 Instituto de Astrofísica de Canarias

More information

Powerful DMD-based light sources with a high throughput virtual slit Arsen R. Hajian* a, Ed Gooding a, Thomas Gunn a, Steven Bradbury a

Powerful DMD-based light sources with a high throughput virtual slit Arsen R. Hajian* a, Ed Gooding a, Thomas Gunn a, Steven Bradbury a Powerful DMD-based light sources with a high throughput virtual slit Arsen R. Hajian* a, Ed Gooding a, Thomas Gunn a, Steven Bradbury a a Hindsight Imaging Inc., 233 Harvard St. #316, Brookline MA 02446

More information

Large format 17µm high-end VOx µ-bolometer infrared detector

Large format 17µm high-end VOx µ-bolometer infrared detector Large format 17µm high-end VOx µ-bolometer infrared detector U. Mizrahi, N. Argaman, S. Elkind, A. Giladi, Y. Hirsh, M. Labilov, I. Pivnik, N. Shiloah, M. Singer, A. Tuito*, M. Ben-Ezra*, I. Shtrichman

More information

Development of a new multi-wavelength confocal surface profilometer for in-situ automatic optical inspection (AOI)

Development of a new multi-wavelength confocal surface profilometer for in-situ automatic optical inspection (AOI) Development of a new multi-wavelength confocal surface profilometer for in-situ automatic optical inspection (AOI) Liang-Chia Chen 1#, Chao-Nan Chen 1 and Yi-Wei Chang 1 1. Institute of Automation Technology,

More information

Devices & Services Company

Devices & Services Company Devices & Services Company 10290 Monroe Drive, Suite 202 - Dallas, Texas 75229 USA - Tel. 214-902-8337 - Fax 214-902-8303 Web: www.devicesandservices.com Email: sales@devicesandservices.com D&S Technical

More information

OPTICS DIVISION B. School/#: Names:

OPTICS DIVISION B. School/#: Names: OPTICS DIVISION B School/#: Names: Directions: Fill in your response for each question in the space provided. All questions are worth two points. Multiple Choice (2 points each question) 1. Which of the

More information

Big League Cryogenics and Vacuum The LHC at CERN

Big League Cryogenics and Vacuum The LHC at CERN Big League Cryogenics and Vacuum The LHC at CERN A typical astronomical instrument must maintain about one cubic meter at a pressure of

More information

Laser Scanning 3D Display with Dynamic Exit Pupil

Laser Scanning 3D Display with Dynamic Exit Pupil Koç University Laser Scanning 3D Display with Dynamic Exit Pupil Kishore V. C., Erdem Erden and Hakan Urey Dept. of Electrical Engineering, Koç University, Istanbul, Turkey Hadi Baghsiahi, Eero Willman,

More information

Speed and Image Brightness uniformity of telecentric lenses

Speed and Image Brightness uniformity of telecentric lenses Specialist Article Published by: elektronikpraxis.de Issue: 11 / 2013 Speed and Image Brightness uniformity of telecentric lenses Author: Dr.-Ing. Claudia Brückner, Optics Developer, Vision & Control GmbH

More information

White Paper on SWIR Camera Test The New Swux Unit Austin Richards, FLIR Chris Durell, Joe Jablonski, Labsphere Martin Hübner, Hensoldt.

White Paper on SWIR Camera Test The New Swux Unit Austin Richards, FLIR Chris Durell, Joe Jablonski, Labsphere Martin Hübner, Hensoldt. White Paper on Introduction SWIR imaging technology based on InGaAs sensor products has been a staple of scientific sensing for decades. Large earth observing satellites have used InGaAs imaging sensors

More information

Be aware that there is no universal notation for the various quantities.

Be aware that there is no universal notation for the various quantities. Fourier Optics v2.4 Ray tracing is limited in its ability to describe optics because it ignores the wave properties of light. Diffraction is needed to explain image spatial resolution and contrast and

More information

Spotlight 150 and 200 FT-IR Microscopy Systems

Spotlight 150 and 200 FT-IR Microscopy Systems S P E C I F I C A T I O N S Spotlight 150 and 200 FT-IR Microscopy Systems FT-IR Microscopy Spotlight 200 with Frontier FT-IR Spectrometer Introduction PerkinElmer Spotlight FT-IR Microscopy Systems are

More information

WHITE PAPER MINIATURIZED HYPERSPECTRAL CAMERA FOR THE INFRARED MOLECULAR FINGERPRINT REGION

WHITE PAPER MINIATURIZED HYPERSPECTRAL CAMERA FOR THE INFRARED MOLECULAR FINGERPRINT REGION WHITE PAPER MINIATURIZED HYPERSPECTRAL CAMERA FOR THE INFRARED MOLECULAR FINGERPRINT REGION Denis Dufour, David Béland, Hélène Spisser, Loïc Le Noc, Francis Picard, Patrice Topart January 2018 Low-cost

More information

Compact High Resolution Imaging Spectrometer (CHRIS) siraelectro-optics

Compact High Resolution Imaging Spectrometer (CHRIS) siraelectro-optics Compact High Resolution Imaging Spectrometer (CHRIS) Mike Cutter (Mike_Cutter@siraeo.co.uk) Summary CHRIS Instrument Design Instrument Specification & Performance Operating Modes Calibration Plan Data

More information

Performance Comparison of Spectrometers Featuring On-Axis and Off-Axis Grating Rotation

Performance Comparison of Spectrometers Featuring On-Axis and Off-Axis Grating Rotation Performance Comparison of Spectrometers Featuring On-Axis and Off-Axis Rotation By: Michael Case and Roy Grayzel, Acton Research Corporation Introduction The majority of modern spectrographs and scanning

More information

THERMOGRAPHY. Courtesy of Optris. Fig1 : Thermographic image of steel slabs captured with PI1M

THERMOGRAPHY. Courtesy of Optris. Fig1 : Thermographic image of steel slabs captured with PI1M THERMOGRAPHY Non-contact sensing can provide the ability to evaluate the internal properties of objects without damage or disturbance by observing its shape, color, size, material or appearance. Non-contact

More information

Low Cost Earth Sensor based on Oxygen Airglow

Low Cost Earth Sensor based on Oxygen Airglow Assessment Executive Summary Date : 16.06.2008 Page: 1 of 7 Low Cost Earth Sensor based on Oxygen Airglow Executive Summary Prepared by: H. Shea EPFL LMTS herbert.shea@epfl.ch EPFL Lausanne Switzerland

More information

Part 1: New spectral stuff going on at NIST. Part 2: TSI Traceability of TRF to NIST

Part 1: New spectral stuff going on at NIST. Part 2: TSI Traceability of TRF to NIST Part 1: New spectral stuff going on at NIST SIRCUS-type stuff (tunable lasers) now migrating to LASP Absolute Spectrally-Tunable Detector-Based Source Spectrally-programmable source calibrated via NIST

More information

Multispectral. imaging device. ADVANCED LIGHT ANALYSIS by. Most accurate homogeneity MeasureMent of spectral radiance. UMasterMS1 & UMasterMS2

Multispectral. imaging device. ADVANCED LIGHT ANALYSIS by. Most accurate homogeneity MeasureMent of spectral radiance. UMasterMS1 & UMasterMS2 Multispectral imaging device Most accurate homogeneity MeasureMent of spectral radiance UMasterMS1 & UMasterMS2 ADVANCED LIGHT ANALYSIS by UMaster Ms Multispectral Imaging Device UMaster MS Description

More information

Wavelength Stabilization of HPDL Array Fast-Axis Collimation Optic with integrated VHG

Wavelength Stabilization of HPDL Array Fast-Axis Collimation Optic with integrated VHG Wavelength Stabilization of HPDL Array Fast-Axis Collimation Optic with integrated VHG C. Schnitzler a, S. Hambuecker a, O. Ruebenach a, V. Sinhoff a, G. Steckman b, L. West b, C. Wessling c, D. Hoffmann

More information

StarBright XLT Optical Coatings

StarBright XLT Optical Coatings StarBright XLT Optical Coatings StarBright XLT is Celestron s revolutionary optical coating system that outperforms any other coating in the commercial telescope market. Our most popular Schmidt-Cassegrain

More information

An Indian Journal FULL PAPER. Trade Science Inc. Parameters design of optical system in transmitive star simulator ABSTRACT KEYWORDS

An Indian Journal FULL PAPER. Trade Science Inc. Parameters design of optical system in transmitive star simulator ABSTRACT KEYWORDS [Type text] [Type text] [Type text] ISSN : 0974-7435 Volume 10 Issue 23 BioTechnology 2014 An Indian Journal FULL PAPER BTAIJ, 10(23), 2014 [14257-14264] Parameters design of optical system in transmitive

More information

Simultaneous Infrared-Visible Imager/Spectrograph a Multi-Purpose Instrument for the Magdalena Ridge Observatory 2.4-m Telescope

Simultaneous Infrared-Visible Imager/Spectrograph a Multi-Purpose Instrument for the Magdalena Ridge Observatory 2.4-m Telescope Simultaneous Infrared-Visible Imager/Spectrograph a Multi-Purpose Instrument for the Magdalena Ridge Observatory 2.4-m Telescope M.B. Vincent *, E.V. Ryan Magdalena Ridge Observatory, New Mexico Institute

More information

BEAM HALO OBSERVATION BY CORONAGRAPH

BEAM HALO OBSERVATION BY CORONAGRAPH BEAM HALO OBSERVATION BY CORONAGRAPH T. Mitsuhashi, KEK, TSUKUBA, Japan Abstract We have developed a coronagraph for the observation of the beam halo surrounding a beam. An opaque disk is set in the beam

More information

Why is There a Black Dot when Defocus = 1λ?

Why is There a Black Dot when Defocus = 1λ? Why is There a Black Dot when Defocus = 1λ? W = W 020 = a 020 ρ 2 When a 020 = 1λ Sag of the wavefront at full aperture (ρ = 1) = 1λ Sag of the wavefront at ρ = 0.707 = 0.5λ Area of the pupil from ρ =

More information

The University of Toledo R. Ellingson and M. Heben

The University of Toledo R. Ellingson and M. Heben focal length, f Spectral Measurement Using a Monochromator, Thermopile Detector, and Lock-In Amplifier September 18, 2012 The University of Toledo R. Ellingson and M. Heben Where are We, Where we are Going?

More information

TriVista. Universal Raman Solution

TriVista. Universal Raman Solution TriVista Universal Raman Solution Why choose the Princeton Instruments/Acton TriVista? Overview Raman Spectroscopy systems can be derived from several dispersive components depending on the level of performance

More information

Measurement and alignment of linear variable filters

Measurement and alignment of linear variable filters Measurement and alignment of linear variable filters Rob Sczupak, Markus Fredell, Tim Upton, Tom Rahmlow, Sheetal Chanda, Gregg Jarvis, Sarah Locknar, Florin Grosu, Terry Finnell and Robert Johnson Omega

More information

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer Page 1 of 11 Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer The Aramis Raman system is a software selectable multi-wavelength Raman system with mapping capabilities with a 400mm monochromator and

More information

Supplementary Figure 1. Effect of the spacer thickness on the resonance properties of the gold and silver metasurface layers.

Supplementary Figure 1. Effect of the spacer thickness on the resonance properties of the gold and silver metasurface layers. Supplementary Figure 1. Effect of the spacer thickness on the resonance properties of the gold and silver metasurface layers. Finite-difference time-domain calculations of the optical transmittance through

More information

DESIGN NOTE: DIFFRACTION EFFECTS

DESIGN NOTE: DIFFRACTION EFFECTS NASA IRTF / UNIVERSITY OF HAWAII Document #: TMP-1.3.4.2-00-X.doc Template created on: 15 March 2009 Last Modified on: 5 April 2010 DESIGN NOTE: DIFFRACTION EFFECTS Original Author: John Rayner NASA Infrared

More information

Tutorial Zemax 9: Physical optical modelling I

Tutorial Zemax 9: Physical optical modelling I Tutorial Zemax 9: Physical optical modelling I 2012-11-04 9 Physical optical modelling I 1 9.1 Gaussian Beams... 1 9.2 Physical Beam Propagation... 3 9.3 Polarization... 7 9.4 Polarization II... 11 9 Physical

More information

Test procedures Page: 1 of 5

Test procedures Page: 1 of 5 Test procedures Page: 1 of 5 1 Scope This part of document establishes uniform requirements for measuring the numerical aperture of optical fibre, thereby assisting in the inspection of fibres and cables

More information

HR2000+ Spectrometer. User-Configured for Flexibility. now with. Spectrometers

HR2000+ Spectrometer. User-Configured for Flexibility. now with. Spectrometers Spectrometers HR2000+ Spectrometer User-Configured for Flexibility HR2000+ One of our most popular items, the HR2000+ Spectrometer features a high-resolution optical bench, a powerful 2-MHz analog-to-digital

More information

A laser speckle reduction system

A laser speckle reduction system A laser speckle reduction system Joshua M. Cobb*, Paul Michaloski** Corning Advanced Optics, 60 O Connor Road, Fairport, NY 14450 ABSTRACT Speckle degrades the contrast of the fringe patterns in laser

More information

WITec Alpha 300R Quick Operation Summary October 2018

WITec Alpha 300R Quick Operation Summary October 2018 WITec Alpha 300R Quick Operation Summary October 2018 This document is frequently updated if you feel information should be added, please indicate that to the facility manager (currently Philip Carubia,

More information

EE119 Introduction to Optical Engineering Fall 2009 Final Exam. Name:

EE119 Introduction to Optical Engineering Fall 2009 Final Exam. Name: EE119 Introduction to Optical Engineering Fall 2009 Final Exam Name: SID: CLOSED BOOK. THREE 8 1/2 X 11 SHEETS OF NOTES, AND SCIENTIFIC POCKET CALCULATOR PERMITTED. TIME ALLOTTED: 180 MINUTES Fundamental

More information

Lens Design I. Lecture 5: Advanced handling I Herbert Gross. Summer term

Lens Design I. Lecture 5: Advanced handling I Herbert Gross. Summer term Lens Design I Lecture 5: Advanced handling I 2018-05-17 Herbert Gross Summer term 2018 www.iap.uni-jena.de 2 Preliminary Schedule - Lens Design I 2018 1 12.04. Basics 2 19.04. Properties of optical systems

More information

Hyperspectral MR-series FTIR Spectroradiometers Radiometric Accuracy for Infrared Signature Measurements

Hyperspectral MR-series FTIR Spectroradiometers Radiometric Accuracy for Infrared Signature Measurements Measurement & Analytics / Measurement made easy Hyperspectral MR-series FTIR Spectroradiometers Radiometric Accuracy for Infrared Signature Measurements The most accurate solution to capture the elusive

More information

Sintec Optronics Technology Pte Ltd 10 Bukit Batok Crescent #07-02 The Spire Singapore Tel: Fax:

Sintec Optronics Technology Pte Ltd 10 Bukit Batok Crescent #07-02 The Spire Singapore Tel: Fax: Sintec Optronics Technology Pte Ltd 10 Bukit Batok Crescent #07-02 The Spire Singapore 658079 Tel: +65 63167112 Fax: +65 63167113 High-power Nd:YAG Self-floating Laser Cutting Head We supply the laser

More information

Why select a BOS zoom lens over a COTS lens?

Why select a BOS zoom lens over a COTS lens? Introduction The Beck Optronic Solutions (BOS) range of zoom lenses are sometimes compared to apparently equivalent commercial-off-the-shelf (or COTS) products available from the large commercial lens

More information