Evaluation of 2D Imaging Schemes for Pulsed Arterial Spin Labeling of the Human Kidney Cortex

Size: px
Start display at page:

Download "Evaluation of 2D Imaging Schemes for Pulsed Arterial Spin Labeling of the Human Kidney Cortex"

Transcription

1 diagnostics Article Evaluation 2D Imaging Schemes Pulsed Arterial Spin Labeling Human Kidney Cortex Charlotte E. Buchanan, Eleanor F. Cox Susan T. Francis * ID Sir Peter Mansfield Imaging Centre, School Physics Astronomy, University Nottingham, Nottingham NG7 2RD, UK; charlotte.buchanan@nottingham.ac.uk (C.E.B.); eleanor.cox@nottingham.ac.uk (E.F.C.) * Correspondence: susan.francis@nottingham.ac.uk; Tel.: +44-(0) Received: 25 May 2018; Accepted: 25 June 2018; Published: 28 June 2018 Abstract: A number imaging readout s are proposed renal arterial spin labeling (ASL) to quantify kidney cortex, including gradient echo-based methods balanced fast field echo (bffe) gradient-echo echo-planar imaging (GE-EPI), or spin echo-based s spin-echo echo-planar imaging (SE-EPI) turbo spin-echo (TSE). Here, we compare se two-dimensional (2D) imaging s to evaluate optimal imaging pulsed ASL (PASL) assessment human kidney cortex at 3 T. Ten healthy volunteers with normal renal function were scanned using 2D multi-slice imaging, in combination with a respiratory triggered flow-sensitive alternating inversion recovery (FAIR) ASL on a 3 T Philips Achieva scanner. All volunteers returned a second identical scan session within two weeks first scan session. Comparisons were made between imaging s in terms -weighted image (PWI) signal-to-noise ratio (SNR) quantification, temporal SNR (tsnr), spatial coverage, repeatability. For imaging, renal cortex was calculated (bffe: 276 ± 29 ml/100g/min, GE-EPI: 222 ± 18 ml/100g/min, SE-EPI: 201 ± 36 ml/100g/min, TSE: 200 ± 20 ml/100g/min). Perfusion was found to be higher GE-based readouts when compared with SE-based readouts, with significantly higher measured bffe readout when compared with all or s (p < 0.05), attributed to greater vascular signal present. Despite PWI-SNR being significantly lower SE-EPI when compared with all or s (p < 0.05), SE-EPI readout gave highest tsnr, was found to be most reproducible assessment kidney cortex, with a coefficient variation (CoV) 17.2%, whilst minimizing variability -weighted signal across slices whole-kidney assessment. For assessment kidney cortex using 2D readout s, SE-EPI provides optimal tsnr, minimal variability across slices, repeatable data acquired in a short scan time with low specific absorption rate. Keywords: magnetic resonance imaging (MRI); arterial spin labeling; renal MRI; ; renal ASL 1. Introduction In clinical practice, renal function is typically determined via serum creatinine measurements to estimate glomerular filtration rate (GFR); however, this method is not highly sensitive, changes in GFR may develop relatively late in progression chronic kidney disease (CKD). Renal inms on delivery nutrients oxygen to tissue, is a key measure by which to monitor renal function. A method providing reliable repeatable assessment kidney, in conjunction with precise morphological inmation, would significantly improve assessment monitoring renal health. Arterial spin labeling (ASL) is a magnetic resonance imaging (MRI) technique that allows non-invasive quantitative assessment tissue, with advantage Diagnostics 2018, 8, 43; doi: /diagnostics

2 Diagnostics 2018, 8, not requiring any exogenous contrast agent, instead using magnetization endogenous-labeled blood to provide contrast. The majority renal ASL studies in literature employed a pulsed ASL (PASL) technique using flow-sensitive alternating inversion recovery (FAIR) [1 8]. In FAIR, two images are collected, a selective image which contains non-inverted arterial blood a non-selective image in which inflowing blood is magnetically inverted. By subtracting non-selective image from selective image, a -weighted image (PWI) is med, which with appropriate modeling can be quantified using a map in units ml/100g/min. A number different two-dimensional (2D) imaging readout s were implemented in literature renal ASL studies. To determine optimal readout renal ASL, a number factors must be considered. The optimal readout should have a short echo time (TE) in order to provide highest image signal-to-noise ratio (SNR), to reduce amount signal dephasing distortion. The short intrinsic T 2 T 2 * in abdomen lead to rapid signal dropout loss signal at longer TEs. The ideal readout should be collected in a short shot length to enable multiple slices through kidney to be acquired prior to decay ASL label, thus enabling whole-kidney assessment. Furrmore, if acquisition is respiratorily triggered, it is important to acquire all images within a respiratory cycle, ideally within flat component respiratory cycle at end expiration where motion is minimal. Finally, ideal readout should have a low specific absorption rate (SAR) so that a short temporal spacing between 2D images can be achieved when collecting a multi-slice dataset. Echo planar imaging (EPI) is one most commonly used readout techniques ASL brain due to its relatively short acquisition time making whole-head coverage feasible, its applicability is used discussed in References [9 16]. However, in body, larger field view (FOV) means EPI readouts typically have a longer TE (>10 ms, dependent on parallel imaging acceleration factor), high-spatial-resolution images, acquisition time can become very long resulting in poor image quality due to susceptibility-induced signal inhomogeneities, particularly close to geometrically irregular tissue air boundaries. Eir gradient echo (GE)- or spin-echo (SE)-based EPI can be used. GE-EPI is more influenced by any B 0 field homogeneity than SE-EPI, since phase shifts from field inhomogeneities, static tissue susceptibility gradients, chemical shifts are not cancelled. Since EPI has a short acquisition time, order 30 ms, multiple slices at peak ASL signal curve can be imaged; thus, low variance is expected in -weighted ASL signal across a multi-slice dataset. Sokolska et al. used a GE-EPI readout combined with pseudo-continuous ASL (pcasl) labeling to assess feasibility within-subject repeatability renal measures [17], whilst Gardener et al. employed a SE-EPI readout to acquire multiple slices across kidney, assessed different breathing strategies to overcome respiratory motion [4]. A balanced fast field echo (bffe), is widely used as image readout renal ASL [3,5,6,18]. It has advantage providing a very short TE high image SNR. However, shot length a bffe is long, at approximately 300 ms a typical abdominal FOV with 3 mm voxel resolution. Thus, a multi-slice acquisition, not all slices are collected at peak ASL signal curve, potentially resulting in greater variance in image -weighted signal across slices in comparison to EPI. In addition, long shot length limits number slices which can be collected when respiratorily triggering data acquisition. The bffe s are also limited by ir sensitivity to field inhomogeneity, with bing artefacts apparent in areas f-resonance in image. Gillis et al. measured inter-study reproducibility ASL at 3 T using a FAIR combined with bffe readout, concluded that this provides a repeatable method measuring renal [5]. Turbo spin-echo (TSE) imaging, also known as fast spin-echo (FSE) imaging, is anor alternative spin-echo-based 2D imaging. Here, time saved by scanning multiple lines k-space at once, when compared with stard spin-echo imaging, means it is possible to lengn time between excitation pulse, thus allowing more time T 1 recovery within readout,

3 Diagnostics 2018, 8, thus, resulting in improved image SNR. A higher number phase-encoding steps can also be used improved spatial resolution, susceptibility-induced signal losses are low. However, TE a TSE readout is typically 50 ms, so some blurring image will occur due to T 2 decay, whilst shot length is long at approximately 160 ms, resulting in slices being acquired at different points in ASL signal curve respiratory cycle, ree, potentially increasing variance in -weighted signal across slices. A furr limitation TSE is high SAR due to multiple refocusing pulses which can result in a long temporal spacing between multi-slice images to keep within SAR limits. It should be noted that three-dimensional (3D) readout s are applied to renal ASL. In References [1,19], a 3D GRASE was applied to assess whole-kidney in healthy volunteers. In a more recent publication, a TSE acquisition was used to evaluate 3D volumetric, isotropic-resolution renal ASL [20]. To date, re are no direct comparisons se 2D imaging readout s ASL in kidney. In this work, we compared GE-EPI, SE-EPI, bffe (a balanced gradient echo also termed TrueFISP or FIESTA), single-shot TSE (SSTSE) (a fast spin-echo sequence also termed fast SE (FSE)) readout s used in combination with a FAIR labeling to assess optimal (s) renal ASL. 2. Materials Methods 2.1. Subjects The study was approved by local ethics committee, all participants gave inmed, written consent. Ten healthy volunteers (age 27 ± 10 years; five female) were scanned approximately one hour on a 3 T Philips Achieva MRI scanner using dual-transmit a 16-channel XLTorso receive coil. To assess repeatability readout, volunteers returned a second visit, which comprised an identical scan session within two weeks first visit scan session. The MRI was permed at same time day on all subjects to minimize potential diurnal variations in renal physiologic function. Subjects fasted previous evening from 8 pm to enable a controlled hydration status all subjects. To ensure that all volunteers had normal kidney function, blood urine samples were collected evaluated by a clinician. Urea, electrolytes, urine protein creatinine ratio were assessed MRI Acquisition Initially, bffe localizer scans were acquired in three orthogonal planes to plan placement imaging ASL-labeling slabs relative to kidneys vessels. The FAIR labeling used a frequency-fset corrected inversion (FOCI) pulse to achieve a 45 mm selective (S) inversion slab (10 mm wider than imaging volume) a 400 mm non-selective (NS) inversion slab. Using such a labeling slab means that trailing edge non-selective label does not arrive within label delay time. Coronal-oblique imaging slices were collected through kidneys in descending order (lateral medial) whilst taking care that selective inversion slab avoided aorta ( 1). Identical readout geometry was acquired on subject all imaging s, with a mm FOV, in-plane spatial resolution 3 mm 5 mm slice thickness. All readout s were acquired with parallel acceleration with a SENSE factor 2, thus reducing achievable GE-EPI SE-EPI TE, minimizing readout duration, reby limiting susceptibility-related distortions signal drop out, allowing multiple slices to be acquired to sample peak ASL signal curve.

4 Diagnostics 2018, 8, Diagnostics 2018, 8, x FOR PEER REVIEW Flow-sensitive Flow-sensitive alternating alternating inversion inversion recovery recovery (FAIR) (FAIR).. Positioning Positioning selective selective non-selective non-selective labeling labeling slabs slabs relative relative to to imaging imaging volume volume kidneys kidneys aorta. aorta. To suppress any static tissue signal in -weighted images, in-plane water To suppress any static tissue signal in -weighted images, in-plane water suppression suppression enhanced through T1 effects (WET) presaturation pulses were applied immediately prior enhanced through T to S/NS pulse 1 effects (WET) presaturation pulses were applied immediately prior to a sinc post-saturation pulse was applied immediately after. A post-label delay S/NS pulse a sinc post-saturation pulse was applied immediately after. A post-label delay (PLD (PLD equivalent to inversion time, TI), defined to be time between inversion pulse to equivalent to inversion time, TI), defined to be time between inversion pulse to central central k-space first slice, 1300 ms was used bffe TSE readouts, 1800 ms GEk-space first slice, 1300 ms was used bffe TSE readouts, 1800 ms GE-EPI EPI SE-EPI readouts. This accounted different readout duration s SE-EPI readouts. This accounted different readout duration s (see Table 1), (see Table 1), ensuring maximum -weighted signal was sampled in central slice ensuring maximum -weighted signal was sampled in central slice.. All datasets were acquired respiratorily triggered on S/NS RF pulse, with a All datasets were acquired respiratorily triggered on S/NS RF pulse, with a minimum repetition minimum repetition time (TR) 3 s between S/NS RF pulse. In total, 25 S/NS image pairs were time (TR) 3 s between S/NS RF pulse. In total, 25 S/NS image pairs were acquired acquired readout. readout. Table 1. Imaging parameters balanced fast field echo (bffe), gradient-echo echo-planar Table 1. Imaging parameters balanced fast field echo (bffe), gradient-echo echo-planar imaging (GE-EPI), spin-echo echo-planar imaging (SE-EPI), turbo spin-echo (TSE) readout imaging (GE-EPI), spin-echo echo-planar imaging (SE-EPI), turbo spin-echo (TSE) readout s. s. The post-label delay (PLD) is defined as time to central k-space first slice. The post-label delay (PLD) is defined as time to central k-space first slice. Post-Label Echo Number Slices Slice Readout Readout Scheme Post-Label Echo Time Flip Angle ( ) Delay (ms) Time (ms) Flip Angle ( (Slice Gap (mm)) Spacing (ms) Scheme Delay (ms) (ms) Number Slices Slice Spacing ) (Slice Gap (mm)) (ms) bffe (0) 280 bffe (0) 280 GE-EPI (0) (0) SE-EPI (0) (0) TSE (5) 480 TSE (5) 480 Base magnetization (M0) 0 ) T1 T 1 relaxation time images were acquired with geometry readout s matched to ASL to allow quantification. Base magnetization (M0) 0 ) images were acquired at same point in respiratory cycle as ASL data using a trigger delay matched to ASL PLD time. A modified respiratorily triggered inversion-recovery sequence was implemented to map TT1 1 relaxation time in renal cortex. Images were acquired at multiple inversion times (TI) 200 ms to 1500 ms in 100 ms steps, but with all TIs collected in respiratory cycle at at same same time time as as ASL ASL data by data introducing by introducing an additional an additional delay (Tv) delay following (Tv) following respiratory trigger, respiratory trigger, prior to prior inversion to pulse inversion [21]. pulse T 1 data [21]. were T1 data collected were collected with a minimum with a minimum TR 8 str to allow 8 s to full allow signal full recovery signal recovery using ausing 400 mm a 400 NS mm inversion NS inversion slab. slab. For GE-EPI For GE-EPI SE-EPI SE-EPI readouts, readouts, multi-slice multi-slice T 1 dataset T1 dataset was acquired was acquired in descending descending order, while order, while bffe bffe TSE readout TSE s, readout s, multi-slice multi-slice T 1 dataset T1 dataset was acquired was acquired both ascending both ascending descending descending orders orders to increase to increase dynamic dynamic range range TI TI [22]. [22] Quantification Renal Cortex Perfusion TT1 1 Analysis was was permed permedusing usingcustom-written MATLAB MATLAB programs programs (Matlab (Matlab version version 8.1, The 8.1, The MathWorks, MathWorks, Inc., Inc., Natick, Natick, MA, MA, USA). USA). ASL ASL -weighted -weighted (PW) (PW) difference differenceimages imageswere med by subtracting non-selective images from selective images [21]. PW were inspected motion, misaligned pairs discarded, remaining PW difference images averaged to m an average PW difference image (ΔM) slice. These were n normalized to base

5 Diagnostics 2018, 8, inspected motion, misaligned pairs discarded, remaining PW difference images averaged to m an average PW difference image ( M) slice. These were n normalized to base M 0 image. T 1 maps were med by fitting inversion recovery data to a two-parameter model. M, T 1, base M 0 maps were used to generate a renal (f ) map, in units ml/100g/min, by fitting data to a kinetic model [23]. Each slice was fitted taking into account exact post-label delay at which slice was collected following labeling pulse (see Table 1; i.e., 1300 ms, 1580 ms, 1860 ms, 2140 ms, 2420 ms slices 1 to 5 bffe ), assuming an arterial transit delay 400 ms (as defined from previous healthy volunteer data [21]). To segment renal cortex, kidneys were manually segmented bee a histogram kidney T 1 was produced along with a threshold to create a cortex mask. This analysis procedure is outlined in 3 Reference [21] from which mode distribution within renal cortex could be computed. The renal cortex masks were compared across readout s to ensure approximately same number voxels were assessed; average DICE similarity coefficient between visits was calculated as 0.53 ± The mean stard deviation in renal cortex were calculated both left right kidneys across subjects Image Quality Assessment The following quantitative metrics were computed in renal cortex to assess quality data readout : (i) -weighted image (PWI) SNR, defined as mean PW signal divided by stard deviation in background noise PW image; (ii) temporal SNR (tsnr) -weighted image, defined as mean PW signal divided by stard deviation across 25 ASL pairs (average across slices); (iii) variance in PW signal across slices (var M ), defined as stard deviation in PW signal across slices divided by mean PW signal Statistical Analysis Statistical analysis was permed using SPSS stware version 21(IBM ). Quantitative variables are expressed as mean ± stard deviation (SD) or median interquartile range (IQR) depending on normality, with a Shapiro Wilk test used to test normality data. In all analyses, a p-value < 0.05 was considered as statistically significant. To assess differences between readout s, a repeated measures ANOVA test was used. For readout, between- within-subject variability measurements was assessed by coefficient variation (bcv, wcv). In addition, coefficient variation (CoV) (stard deviation divided by mean) was calculated to assess repeatability between scan sessions. 3. Results All healthy volunteers were confirmed to have normal kidney function, with egfr > 60 ml/min/1.73 m 2, with a creatinine level 76 ± 15 µmol/l, a urea level 4.2 ± 1.1 mmol/l ASL Image Quality Base M 0 images 2D readout are in 2, with good data quality all readouts minimal distortions, even EPI acquisitions. Note that TSE suffered from blurring, whilst vessels appeared brighter in bffe image. The TSE ASL had highest SAR at approximately 70% whole-body averaged SAR, whilst SE-EPI, GE-EPI, bffe ASL s all had SAR less than 35%. To minimize SAR, TSE implements a longer temporal spacing between slice acquisitions (see Table 1), limiting TSE acquisition to three slices acquired at peak ASL signal curve.

6 Diagnostics 2018, 8, Diagnostics 2018, 8, x FOR PEER REVIEW Example base magnetization (M0) images balanced fast field echo (bffe), gradientecho echo-planar imaging (GE-EPI), spin-echo echo-planar imaging (SE-EPI), turbo spin-echo 2. Example base magnetization (M 0 ) images balanced fast field echo (bffe), gradient-echo echo-planar imaging (GE-EPI), spin-echo echo-planar imaging (SE-EPI), turbo (TSE) readout s. spin-echo (TSE) readout s. The TSE ASL had highest SAR at approximately 70% whole-body averaged 3 shows multi-slice average -weighted images readout. The tsnr, SAR, whilst SE-EPI, GE-EPI, bffe ASL s all had SAR less than 35%. To PWI-SNR, variability -weighted signal (var minimize SAR, TSE implements a longer temporal spacing M ) are provided in Table 2. The TSE between slice acquisitions (see had highest PWI-SNR, whilst SE-EPI had lowest PWI-SNR. However, tsnr was Table 1), limiting TSE acquisition to three slices acquired at peak ASL signal curve. optimal SE-EPI, whilst GE-EPI had lowest t-snr. The variability 3 shows multi-slice average -weighted images readout. The -weighted signal across slices (var tsnr, PWI-SNR, variability -weighted M ) was found to be smallest SE-EPI, signal (var M) are provided in Table 2. The reflecting this to be a good multi-slice whole kidney assessment, with highest variability TSE had highest PWI-SNR, whilst SE-EPI had lowest PWI-SNR. However, tsnr recorded was Diagnostics optimal 2018, 8, bffe x FOR. SE-EPI PEER, REVIEW 4 shows example maps readout. whilst GE-EPI had lowest t-snr. The variability weighted signal across slices (varδm) was found to be smallest SE-EPI, reflecting this to be a good multi-slice whole kidney assessment, with highest variability recorded bffe. 4 shows example maps readout Example Example -weighted -weighted images images (PWI) (PWI) (bffe, (bffe, GE-EPI, GE-EPI, SE-EPI, SE-EPI, TSE) TSE) from a single subject. The post-label delay slice is indicated on image. from a single subject. The post-label delay slice is indicated on image.

7 Diagnostics 2018, 8, Table 2. Perfusion-weighted image signal-to-noise ratio (PWI-SNR), temporal SNR (tsnr), variability -weighted signal (var M ). Readout Scheme PWI-SNR tsnr var M (%) bffe 6.2 ± ± ± 11 GE-EPI 6.3 ± ± ± 5 3. Example SE-EPI -weighted 4.9 images ± 1.5(PWI) 2.6 ± 1.6 (bffe, 11 GE-EPI, ± 3 SE-EPI, TSE 8.5 ± ± ± 4 TSE) from a single subject. The post-label delay slice is indicated on image Example Example maps maps readout readout (bffe, (bffe, GE-EPI, GE-EPI, SE-EPI, SE-EPI, single-shot single-shot TSE) TSE) from from a single single subject subject central central slice. slice. Note Note higher higher signal signal bffe bffe due due to to contributions contributions from from arcuate arcuate arteries arteries kidney. kidney. Table 2. Perfusion-weighted image signal-to-noise ratio (PWI-SNR), temporal SNR (tsnr), 3.2. Perfusion Quantification variability -weighted signal (varδm). Mean renal cortex across all readout s was 223 ± 11 ml/100g/min; error Readout Scheme PWI-SNR tsnr var M indicates stard error across readout s. (%) For readout, measured renal cortex bffe 6.2 ± ± 2.0 first visit 26 ± are 11 in 5, were found to be as follows: GE-EPI bffe 276 ± ± ml/100g/min, ± 0.8 GE-EPI ± 5 ± 18 ml/100g/min, SE-EPI 201 ± 36 ml/100g/min, SE-EPI TSE ± ml/100g/min. 2.6 ± A ± repeated 3 measures ANOVA showed that bffe readout produced TSE consistently 8.5 ± 4.1 higher 2.4 ± ± 4 when compared with or three s (p = 0.03), but also had largest variance between subjects. The SE-EPI gave 3.2. Perfusion Quantification Diagnostics lowest 2018, 8, within-subject x FOR PEER REVIEW variance Mean renal cortex across all readout s was 223 ± 11 ml/100g/min; error indicates stard error across readout s. For readout, measured renal cortex first visit are in 5, were found to be as follows: bffe 276 ± 29 ml/100g/min, GE-EPI 222 ± 18 ml/100g/min, SE-EPI 201 ± 36 ml/100g/min, TSE 200 ± 20 ml/100g/min. A repeated measures ANOVA showed that bffe readout produced consistently higher when compared with or three s (p = 0.03), but also had largest variance between subjects. The SE-EPI gave lowest within-subject variance. The between-subject variation (bcv) s was 53.3%, 23.7%, 26.2%, 35.9% bffe, GE-EPI, SE-EPI, TSE, respectively. The within-subject variation (wcv) was 18.8% bffe, 23.9% GE-EPI, 15.1% SE-EPI, 17.2% TSE Renal Renal cortical cortical measured from from readout readout first first visit. visit. Values Values are are mean mean with with error error bars bars showing showing stard stard error error mean. mean. The The bffe readout gave significantly higher than or readouts (repeated measures bffe readout gave significantly higher than or readouts (repeated measures ANOVA, p = 0.03). A significant difference was observed between SE-EPI GE-EPI., with p- ANOVA, p = 0.03). A significant difference was observed between SE-EPI GE-EPI., with p- post-hoc paired t-tests. post-hoc paired t-tests. For readout, T1 renal cortex was also assessed computed to be 1186 ± 147 ms, 1259 ± 214 ms, 1294 ± 168 ms, 1135 ± 149 ms bffe, GE-EPI, SE-EPI, TSE readout s, respectively Repeatability

8 Diagnostics 2018, 8, The between-subject variation (bcv) s was 53.3%, 23.7%, 26.2%, 35.9% bffe, GE-EPI, SE-EPI, TSE, respectively. The within-subject variation (wcv) was 18.8% bffe, 23.9% GE-EPI, 15.1% SE-EPI, 17.2% TSE. For readout, T 1 renal cortex was also assessed computed to be 1186 ± 147 ms, 1259 ± 214 ms, 1294 ± 168 ms, 1135 ± 149 ms bffe, GE-EPI, SE-EPI, TSE readout s, respectively Repeatability The most repeatable readout was SE-EPI, which had a CoV 17.2%; least repeatable was GE-EPI with a CoV 28.3%. For readout, across 10 subjects, re were no significant differences in renal cortex between first second visit (p > 0.05). 4. Discussion Previous renal ASL studies used a variety readout s in combination with FAIR labeling. In this work, a comparison balanced fast field echo (bffe), gradient-echo EPI (GE-EPI), spin-echo EPI (SE-EPI), turbo spin-echo (TSE) s was made renal ASL. For all s, multi-slice coverage could be achieved with five contiguous slices collected bffe, GE-EPI, SE-EPI, three slices with a 5 mm slice gap TSE. The higher SAR led to wider readout spacing TSE. GE-EPI SE-EPI could achieve whole-kidney coverage in shortest amount time. In this work, cortical was higher gradient-echo s (GE-EPI bffe) in comparison to spin-echo based s (SE-EPI TSE). In particular, calculated from bffe readout data was found to be significantly higher than that from all or s which could be attributed to presence vascular signal in se images. The bffe had highest intravascular signal contribution all readout s due to high signal intensity from blood, which has an intrinsically high T 2 /T 1 ratio, as well as continuous replenishment fresh blood fact that longitudinal magnetization flowing blood is minimally disturbed by alpha pulse train in bffe readout [24]. When comparing SNR, SE-EPI gave a significantly lower PWI-SNR when compared with or s; however, this could, in part, be attributed to lower echo time this sequence or lower contribution from vascular signal in SE-EPI images. The SE-EPI also gave highest temporal SNR, suggesting lowest fluctuations from pulsatile vessels. The SE-EPI gave lowest variance in -weighted signal (var M ) across slices all readout s. This is unsurprising as SE-EPI has a short shot length per slice, so, all slices are acquired at almost same point on both ASL signal curve in respiratory cycle, resulting in a small variance in signal slice. Conversely, TSE readouts have a long shot length; this yielded highest variance in signal across slices. All 2D readout s were determined to be repeatable with a CoV 28.2% or less. SE-EPI was found to be optimal with a CoV 17.2%. The ASL quantitative measurements normal showed good within-subject variability repeatability. Note that differences in distortions between readout s are likely partly reflected in DICE coefficients between masks. In addition, in this study, T 1 maps were generated readout, used in quantification, with mean T 1 renal cortex also computed. As expected, bffe TSE readout s led to a shorter measured longitudinal recovery time when compared with EPI methods, likely due to influence T 2 */T 2 bffe readout [25], contribution magnetization transfer effects TSE readout [26]. A limitation this study is that data were not collected with multiple post-label delay times to compute individual subject arterial transit delay time maps use in quantification; instead, a fixed arterial transit delay time was assumed. However, all readout s, same selective non-selective label positions widths were used, slice coverage was small

9 Diagnostics 2018, 8, at 2.5 cm; thus, arterial transit delay effects were minimal equally affected readout s in quantification. These estimates m basis interpreting optimal readout s guiding future study design in assessing ASL renal. 5. Conclusions FAIR ASL measures were collected bffe, GE-EPI, SE-EPI, TSE readout s. All s were found to be repeatable with coefficients variation less than 29% all techniques. When comparing all four techniques, we conclude that SE-EPI provides optimal temporal SNR, consistency across slices, repeatability between sessions, has lowest specific absorption rate. Author Contributions: S.T.F. conceived designed experiments; C.E.B., S.T.F., E.F.C. permed experiments; C.E.B. analyzed data; C.E.B., E.F.C., S.T.F. wrote paper. Funding: This work was funded by Dr Hadwen Trust. The Dr Hadwen Trust (DHT) is UK s leading non-animal biomedical research charity that exclusively funds promotes human-relevant research that replaces use animals whilst supporting progress medicine. Conflicts Interest: The authors declare no conflict interest. References 1. Cutajar, M.; Thomas, D.L.; Hales, P.W.; Banks, T.; Clark, C.A.; Gordon, I. Comparison ASL DCE MRI non-invasive measurement renal blood flow: Quantification reproducibility. Eur. Radiol. 2014, 24, [CrossRef] 2. Dong, J.; Yang, L.; Su, T.; Yang, X.; Chen, B.; Zhang, J.; Wang, X.; Jiang, X. Quantitative assessment acute kidney injury by noninvasive arterial spin labeling MRI: A pilot study. Sci. China Life Sci. 2013, 56, [CrossRef] 3. Gardener, A.; Francis, S. Multi-slice kidney using SE-EPI FAIR: Optimised acquisition analysis strategies. In Proceedings 17th Scientific Meeting, International Society Magnetic Resonance in Medicine, Honolulu, HI, USA, April 2009; p Gardener, A.G.; Francis, S.T. Multislice kidneys using parallel imaging: Image acquisition analysis strategies. Magn. Reson. Med. 2010, 63, [CrossRef] 5. Gillis, K.A.; McComb, C.; Foster, J.E.; Taylor, A.H.M.; Patel, R.K.; Morris, S.T.W.; Jardine, A.G.; Schneider, M.P.; Roditi, G.H.; Delles, C.; et al. Inter-study reproducibility arterial spin labelling magnetic resonance imaging measurement renal in healthy volunteers at 3 Tesla. BMC Nephrol. 2014, 15, 23. [CrossRef] 6. Martirosian, P.; Klose, U.; Mader, I.; Schick, F. FAIR True-FISP Perfusion Imaging Kidneys. Magn. Reson. Med. 2004, 51, [CrossRef] 7. Pedrosa, I.; Rafatz, K.; Robson, P.; Wagner, A.A.; Atkins, M.B.; Rsky, N.M.; Alsop, D.C. Arterial spin labeling MR imaging characterisation renal masses in patients with impaired renal function: Initial experience. Eur. Radiol. 2012, 22, [CrossRef] 8. Robson, P.M.; Madhuranthakam, A.J.; Dai, W.; Pedrosa, I.; Rsky, N.M.; Alsop, D.C. Strategies reducing respiratory motion artifacts in renal imaging with arterial spin labeling. Magn. Reson. Med. 2009, 61, [CrossRef] 9. Alsop, D.C.; Detre, J.A.; Golay, X.; Günr, M.; Hendrikse, J.; Hernez-Garcia, L.; Lu, H.; Macintosh, B.J.; Parkes, L.M.; Smits, M.; et al. Recommended implementation arterial spin-labeled MRI clinical applications: A consensus ISMRM study group european consortium ASL in dementia. Magn. Reson. Med [CrossRef] 10. Brookes, M.J.; Morris, P.G.; Gowl, P.A.; Francis, S.T. Noninvasive measurement arterial cerebral blood volume using Look-Locker EPI arterial spin labeling. Magn. Reson. Med. 2007, 58, [CrossRef] 11. Gunr, M.; Bock, M.; Schad, L.R. Arterial spin labeling in combination with a look-locker sampling strategy: Inflow turbo-sampling EPI-FAIR (ITS-FAIR). Magn. Reson. Med. 2001, 46, [CrossRef]

10 Diagnostics 2018, 8, Hall, E.; Hall, E.; Wesolowski, R.; Wesolowski, R.; Gowl, P.; Gowl, P.; Francis, S.; Francis, S. Improved detection estimation using high spatial resolution ASL at 7T. In Proceedings 17th Scientific Meeting, International Society Magnetic Resonance in Medicine, Honolulu, HI, USA, April 2009; p Kim, S.G. Quantification relative cerebral blood flow change by flow-sensitive alternating inversion recovery (FAIR) technique: Application to functional mapping. Magn. Reson. Med. 1995, 34, [CrossRef] 14. Wong, E.C. Quantifying CBF with pulsed ASL: Technical pulse sequence factors. J. Magn. Reson. Imaging 2005, 22, [CrossRef] 15. Wong, E.C. An introduction to ASL labeling techniques. J. Magn. Reson. Imaging 2014, 40, [CrossRef] 16. Wong, E.C.; Buxton, R.B.; Frank, L.R. A oretical experimental comparison continuous pulsed arterial spin labeling techniques quantitative imaging. Magn. Reson. Med. 1998, 40, [CrossRef] 17. Sokolska, M.; Thomas, D.; Bainbridge, A.; Golay, X.; Taylor, S.; Punwani, S.; Pendse, D.; Uk, L. Renal Pseudo-continuous Arterial Spin Labelling ( pcasl) MRI: A Repeatability Study. Eur. Congr. Radiol. 2014, [CrossRef] 18. Park, S.H.; Wang, D.J.J.; Duong, T.Q. Balanced steady state free precession arterial spin labeling MRI: Initial experience blood flow mapping in human brain, retina, kidney. Magn. Reson. Imaging 2013, 31, [CrossRef] 19. Cutajar, M.; Thomas, D.L.; Banks, T.; Clark, C.A.; Golay, X.; Gordon, I. Repeatability renal arterial spin labelling MRI in healthy subjects. Magn. Reson. Mater. Phys. Biol. Med. 2012, 25, [CrossRef] 20. Robson, P.M.; Madhuranthakam, A.J.; Smith, M.P.; Sun, M.R.M.; Dai, W.; Rsky, N.M.; Pedrosa, I.; Alsop, D.C. Volumetric Arterial Spin-labeled Perfusion Imaging Kidneys with a Three-dimensional Fast Spin Echo Acquisition. Acad. Radiol. 2016, 23, [CrossRef] 21. Cox, E.F.; Buchanan, C.E.; Bradley, C.R.; Prestwich, B.; Mahmoud, H.; Taal, M.; Selby, N.M.; Francis, S.T. Multiparametric renal magnetic resonance imaging: Validation, interventions, alterations in chronic kidney disease. Front. Physiol. 2017, 8, [CrossRef] 22. Hoad, C.L.; Palaniyappan, N.; Kaye, P.; Chernova, Y.; James, M.W.; Costigan, C.; Austin, A.; Marciani, L.; Gowl, P.A.; Guha, I.N.; et al. A study T1 relaxation time as a measure liver fibrosis influence confounding histological factors. NMR Biomed. 2015, 28, [CrossRef] 23. Buxton, R.B.; Frank, L.R.; Wong, E.C.; Siewert, B.; Warach, S.; Edelman, R.R. A general kinetic model quantitative imaging with arterial spin labeling. Magn. Reson. Med. 1998, 40, [CrossRef] 24. Yan, L.; Li, C.; Kilroy, E.; Wehrli, F.W.; Wang, D.J.J. Quantification arterial cerebral blood volume using multiphase-balanced SSFP-based ASL. Magn. Reson. Med. 2012, 68, [CrossRef] 25. Schmitt, P.; Griswold, M.A.; Jakob, P.M.; Kotas, M.; Gulani, V.; Flentje, M.; Haase, A. Inversion recovery TrueFISP: quantification T(1), T(2), spin density. Magn. Reson. Med. 2004, 51, [CrossRef] 26. Turner, R.; Oros-Peusquens, A.M.; Romanzetti, S.; Zilles, K.; Shah, N.J. Optimised in vivo visualisation cortical structures in human brain at 3 T using IR-TSE. Magn. Reson. Imaging 2008, 26, [CrossRef] 2018 by authors. Licensee MDPI, Basel, Switzerl. This article is an open access article distributed under terms conditions Creative Commons Attribution (CC BY) license (

Noninvasive Blood Flow Mapping with Arterial Spin Labeling (ASL) Paul Kyu Han and Sung-Hong Park

Noninvasive Blood Flow Mapping with Arterial Spin Labeling (ASL) Paul Kyu Han and Sung-Hong Park Noninvasive Blood Flow Mapping with Arterial Spin Labeling (ASL) Paul Kyu Han and Sung-Hong Park Department of Bio and Brain Engineering, Korea Advanced Institute of Science and Technology (KAIST), Daejeon,

More information

Works-in-Progress package Version 1.0. For the SIEMENS Magnetom. Installation and User s Guide NUMARIS/4VA21B. January 22, 2003

Works-in-Progress package Version 1.0. For the SIEMENS Magnetom. Installation and User s Guide NUMARIS/4VA21B. January 22, 2003 Works-in-Progress package Version 1.0 For the Installation and User s Guide NUMARIS/4VA21B January 22, 2003 Section of Medical Physics, University Hospital Freiburg, Germany Contact: Klaus Scheffler PhD,

More information

Experience in implementing continuous arterial spin labeling on a commercial MR scanner

Experience in implementing continuous arterial spin labeling on a commercial MR scanner JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 6, NUMBER 1, WINTER 2005 Experience in implementing continuous arterial spin labeling on a commercial MR scanner Theodore R. Steger and Edward F. Jackson

More information

Cardiac MR. Dr John Ridgway. Leeds Teaching Hospitals NHS Trust, UK

Cardiac MR. Dr John Ridgway. Leeds Teaching Hospitals NHS Trust, UK Cardiac MR Dr John Ridgway Leeds Teaching Hospitals NHS Trust, UK Cardiac MR Physics for clinicians: Part I Journal of Cardiovascular Magnetic Resonance 2010, 12:71 http://jcmr-online.com/content/12/1/71

More information

Pulse Sequence Design and Image Procedures

Pulse Sequence Design and Image Procedures Pulse Sequence Design and Image Procedures 1 Gregory L. Wheeler, BSRT(R)(MR) MRI Consultant 2 A pulse sequence is a timing diagram designed with a series of RF pulses, gradients switching, and signal readout

More information

Pulse Sequence Design Made Easier

Pulse Sequence Design Made Easier Pulse Sequence Design Made Easier Gregory L. Wheeler, BSRT(R)(MR) MRI Consultant gurumri@gmail.com 1 2 Pulse Sequences generally have the following characteristics: An RF line characterizing RF Pulse applications

More information

Slice profile optimization in arterial spin labeling using presaturation and optimized RF pulses

Slice profile optimization in arterial spin labeling using presaturation and optimized RF pulses Magnetic Resonance Imaging 24 (2006) 1229 1240 Slice profile optimization in arterial spin labeling using presaturation and optimized RF pulses David Alberg Holm a,b, 4, Karam Sidaros a a Danish Research

More information

EPISTAR MRI: Multislice Mapping of Cerebral Blood Flow

EPISTAR MRI: Multislice Mapping of Cerebral Blood Flow EPISTAR MRI: Multislice Mapping of Cerebral Blood Flow Robert R. Edelman, Qun Chen A method is described for multislice EPISTAR that perfectly compensates magnetization transfer effects. lnflowing arterial

More information

BOLD fmri: signal source, data acquisition, and interpretation

BOLD fmri: signal source, data acquisition, and interpretation BOLD fmri: signal source, data acquisition, and interpretation Cheryl Olman 4 th year student, Department of Neuroscience and Center for Magnetic Resonance Research Discussion series Week 1: Biological

More information

NIH Public Access Author Manuscript Magn Reson Med. Author manuscript; available in PMC 2014 May 16.

NIH Public Access Author Manuscript Magn Reson Med. Author manuscript; available in PMC 2014 May 16. NIH Public Access Author Manuscript Published in final edited form as: Magn Reson Med. 2011 July ; 66(1): 168 173. doi:10.1002/mrm.22768. 3D GRASE PROPELLER: Improved Image Acquisition Technique for Arterial

More information

Background (~EE369B)

Background (~EE369B) Background (~EE369B) Magnetic Resonance Imaging D. Nishimura Overview of NMR Hardware Image formation and k-space Excitation k-space Signals and contrast Signal-to-Noise Ratio (SNR) Pulse Sequences 13

More information

1 Introduction. 2 The basic principles of NMR

1 Introduction. 2 The basic principles of NMR 1 Introduction Since 1977 when the first clinical MRI scanner was patented nuclear magnetic resonance imaging is increasingly being used for medical diagnosis and in scientific research and application

More information

Module 2. Artefacts and Imaging Optimisation for single shot methods. Content: Introduction. Phase error. Phase bandwidth. Chemical shift review

Module 2. Artefacts and Imaging Optimisation for single shot methods. Content: Introduction. Phase error. Phase bandwidth. Chemical shift review MRES 7005 - Fast Imaging Techniques Module 2 Artefacts and Imaging Optimisation for single shot methods Content: Introduction Phase error Phase bandwidth Chemical shift review Chemical shift in pixels

More information

Simultaneous Multi-Slice (Slice Accelerated) Diffusion EPI

Simultaneous Multi-Slice (Slice Accelerated) Diffusion EPI Simultaneous Multi-Slice (Slice Accelerated) Diffusion EPI Val M. Runge, MD Institute for Diagnostic and Interventional Radiology Clinics for Neuroradiology and Nuclear Medicine University Hospital Zurich

More information

SNR and functional sensitivity of BOLD and perfusion-based fmri using arterial spin labeling with spiral SENSE at 3 T

SNR and functional sensitivity of BOLD and perfusion-based fmri using arterial spin labeling with spiral SENSE at 3 T Available online at www.sciencedirect.com Magnetic Resonance Imaging 26 (2008) 513 522 SNR and functional sensitivity of BOLD and perfusion-based fmri using arterial spin labeling with spiral SENSE at

More information

Liver imaging beyond expectations with Ingenia

Liver imaging beyond expectations with Ingenia Publication for the Philips MRI Community Issue 47 2012/3 Liver imaging beyond expectations with Ingenia Contributed by John Penatzer, RT, MR clinical product specialist, Cleveland, OH, USA Publication

More information

Advanced MSK MRI Protocols at 3.0T. Garry E. Gold, M.D. Associate Professor Department of Radiology Stanford University

Advanced MSK MRI Protocols at 3.0T. Garry E. Gold, M.D. Associate Professor Department of Radiology Stanford University Advanced MSK MRI Protocols at 3.0T Garry E. Gold, M.D. Associate Professor Department of Radiology Stanford University Outline Why High Field for MSK? SNR and Relaxation Times Technical Issues Example

More information

Image Quality/Artifacts Frequency (MHz)

Image Quality/Artifacts Frequency (MHz) The Larmor Relation 84 Image Quality/Artifacts (MHz) 42 ω = γ X B = 2πf 84 0.0 1.0 2.0 Magnetic Field (Tesla) 1 A 1D Image Magnetic Field Gradients Magnet Field Strength Field Strength / Gradient Coil

More information

Pulse Sequences: Rapid Gradient Echo

Pulse Sequences: Rapid Gradient Echo Pulse Sequences: Rapid Gradient Echo M229 Advanced Topics in MRI Holden H. Wu, Ph.D. 2018.04.17 Department of Radiological Sciences David Geffen School of Medicine at UCLA Class Business Office hours -

More information

H 2 O and fat imaging

H 2 O and fat imaging H 2 O and fat imaging Xu Feng Outline Introduction benefit from the separation of water and fat imaging Chemical Shift definition of chemical shift origin of chemical shift equations of chemical shift

More information

MRI Metal Artifact Reduction

MRI Metal Artifact Reduction MRI Metal Artifact Reduction PD Dr. med. Reto Sutter University Hospital Balgrist Zurich University of Zurich OUTLINE Is this Patient suitable for MR Imaging? Metal artifact reduction Is this Patient suitable

More information

Cover Page. The handle holds various files of this Leiden University dissertation

Cover Page. The handle   holds various files of this Leiden University dissertation Cover Page The handle http://hdl.handle.net/1887/49562 holds various files of this Leiden University dissertation Author: Schmid, Sophie Title: Arterial spin labeling in space and time : new MRI sequences

More information

3T Unlimited. ipat on MAGNETOM Allegra The Importance of ipat at 3T. medical

3T Unlimited. ipat on MAGNETOM Allegra The Importance of ipat at 3T. medical 3T Unlimited ipat on MAGNETOM Allegra The Importance of ipat at 3T s medical ipat on MAGNETOM Allegra The Importance of ipat at 3T The rise of 3T MR imaging Ultra High Field MR (3T) has flourished during

More information

2014 M.S. Cohen all rights reserved

2014 M.S. Cohen all rights reserved 2014 M.S. Cohen all rights reserved mscohen@g.ucla.edu IMAGE QUALITY / ARTIFACTS SYRINGOMYELIA Source http://gait.aidi.udel.edu/res695/homepage/pd_ortho/educate/clincase/syrsco.htm Surgery is usually recommended

More information

ISSN X CODEN (USA): PCHHAX. The role of dual spin echo in increasing resolution in diffusion weighted imaging of brain

ISSN X CODEN (USA): PCHHAX. The role of dual spin echo in increasing resolution in diffusion weighted imaging of brain Available online at www.derpharmachemica.com ISSN 0975-413X CODEN (USA): PCHHAX Der Pharma Chemica, 2016, 8(17):15-20 (http://derpharmachemica.com/archive.html) The role of in increasing resolution in

More information

Gradient Spoiling. Average balanced SSFP magnetization Reduce sensitivity to off-resonance. FFE, FISP, GRASS, GRE, FAST, Field Echo

Gradient Spoiling. Average balanced SSFP magnetization Reduce sensitivity to off-resonance. FFE, FISP, GRASS, GRE, FAST, Field Echo Gradient Spoiling Average balanced SSFP magnetization Reduce sensitivity to off-resonance FFE, FISP, GRASS, GRE, FAST, Field Echo 1 Gradient-Spoiled Sequence (GRE, FFE, FISP, GRASS) RF TR G z G y G x Signal

More information

MR Advance Techniques. Flow Phenomena. Class II

MR Advance Techniques. Flow Phenomena. Class II MR Advance Techniques Flow Phenomena Class II Flow Phenomena In this class we will explore different phenomenona produced from nuclei that move during the acquisition of data. Flowing nuclei exhibit different

More information

Functional MRI with variable echo time acquisition

Functional MRI with variable echo time acquisition NeuroImage 20 (2003) 2062 2070 www.elsevier.com/locate/ynimg Functional MRI with variable echo time acquisition Nan-kuei Chen, Svetlana Egorova, Charles R.G. Guttmann, and Lawrence P. Panych* Center for

More information

HETERONUCLEAR IMAGING. Topics to be Discussed:

HETERONUCLEAR IMAGING. Topics to be Discussed: HETERONUCLEAR IMAGING BioE-594 Advanced MRI By:- Rajitha Mullapudi 04/06/2006 Topics to be Discussed: What is heteronuclear imaging. Comparing the hardware of MRI and heteronuclear imaging. Clinical applications

More information

(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010

(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010 (N)MR Imaging Lab Course Script FMP PhD Autumn School Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder Date: November 3rd, 2010 1 Purpose: Understanding the basic principles of MR imaging

More information

High Field MRI: Technology, Applications, Safety, and Limitations

High Field MRI: Technology, Applications, Safety, and Limitations High Field MRI: Technology, Applications, Safety, and Limitations R. Jason Stafford, Ph.D. The University of Texas M. D. Anderson Cancer Center, Houston, TX Introduction The amount of available signal

More information

Principles of MRI EE225E / BIO265. Lecture 21. Instructor: Miki Lustig UC Berkeley, EECS. M. Lustig, EECS UC Berkeley

Principles of MRI EE225E / BIO265. Lecture 21. Instructor: Miki Lustig UC Berkeley, EECS. M. Lustig, EECS UC Berkeley Principles of MRI Lecture 21 EE225E / BIO265 Instructor: Miki Lustig UC Berkeley, EECS Question What is the difference between the images? Answer Both T1-weighted spin-echo gradient-echo Lower SNR Meniscus

More information

Half-Pulse Excitation Pulse Design and the Artifact Evaluation

Half-Pulse Excitation Pulse Design and the Artifact Evaluation Half-Pulse Excitation Pulse Design and the Artifact Evaluation Phillip Cho. INRODUCION A conventional excitation scheme consists of a slice-selective RF excitation followed by a gradient-refocusing interval

More information

Supplementary Material

Supplementary Material Supplementary Material Orthogonal representation of sound dimensions in the primate midbrain Simon Baumann, Timothy D. Griffiths, Li Sun, Christopher I. Petkov, Alex Thiele & Adrian Rees Methods: Animals

More information

MR Basics: Module 6 Pulse Sequences

MR Basics: Module 6 Pulse Sequences Module 6 Transcript For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions.

More information

Passive Tracking Exploiting Local Signal Conservation: The White Marker Phenomenon

Passive Tracking Exploiting Local Signal Conservation: The White Marker Phenomenon Passive Tracking Exploiting Local Signal Conservation: The White Marker Phenomenon Jan-Henry Seppenwoolde,* Max A. Viergever, and Chris J.G. Bakker Magnetic Resonance in Medicine 50:784 790 (2003) This

More information

MR Basics: Module 8 Image Quality

MR Basics: Module 8 Image Quality Module 8 Transcript For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions.

More information

2015 Spin echoes and projection imaging

2015 Spin echoes and projection imaging 1. Spin Echoes 1.1 Find f0, transmit amplitudes, and shim settings In order to acquire spin echoes, we first need to find the appropriate scanner settings using the FID GUI. This was all done last week,

More information

MRI physics for SPM users

MRI physics for SPM users MRI physics for SPM users SPM course 11/2013 Antoine Lutti antoine.lutti@chuv.ch General principals Origin of the signal RF excitation Relaxation (T1, T2, ) Anatomical imaging Image contrast Outline Standard

More information

ACRIN 6686 / RTOG 0825

ACRIN 6686 / RTOG 0825 ACRIN 6686 (RTOG 0825) Advanced MRI Imaging Manual ACRIN 6686 / RTOG 0825 A phase III double blind placebo controlled trial of conventional chemoradiation and adjuvant temozolomide plus bevacizumab vs

More information

MRI Summer Course Lab 2: Gradient Echo T1 & T2* Curves

MRI Summer Course Lab 2: Gradient Echo T1 & T2* Curves MRI Summer Course Lab 2: Gradient Echo T1 & T2* Curves Experiment 1 Goal: Examine the effect caused by changing flip angle on image contrast in a simple gradient echo sequence and derive T1-curves. Image

More information

25 CP Generalize Concepts in Abstract Multi-dimensional Image Model Component Semantics Page 1

25 CP Generalize Concepts in Abstract Multi-dimensional Image Model Component Semantics Page 1 25 CP-1390 - Generalize Concepts in Abstract Multi-dimensional Image Model Component Semantics Page 1 1 STATUS Letter Ballot 2 Date of Last Update 2014/09/08 3 Person Assigned David Clunie 4 mailto:dclunie@dclunie.com

More information

NIH Public Access Author Manuscript Magn Reson Med. Author manuscript; available in PMC 2009 December 1.

NIH Public Access Author Manuscript Magn Reson Med. Author manuscript; available in PMC 2009 December 1. NIH Public Access Author Manuscript Published in final edited form as: Magn Reson Med. 2008 December ; 60(6): 1488 1497. doi:10.1002/mrm.21790. Continuous Flow Driven Inversion for Arterial Spin Labeling

More information

IMPROVING THE ROBUSTNESS OF PERFUSION IMAGING WITH ARTERIAL SPIN LABELING MAGNETIC RESONANCE IMAGING HUAN TAN

IMPROVING THE ROBUSTNESS OF PERFUSION IMAGING WITH ARTERIAL SPIN LABELING MAGNETIC RESONANCE IMAGING HUAN TAN IMPROVING THE ROBUSTNESS OF PERFUSION IMAGING WITH ARTERIAL SPIN LABELING MAGNETIC RESONANCE IMAGING BY HUAN TAN A Dissertation Submitted to the Graduate Faculty of WAKE FOREST UNIVERSITY GRADUATE SCHOOL

More information

Downloaded from by on 02/07/18 from IP address Copyright ARRS. For personal use only; all rights reserved

Downloaded from  by on 02/07/18 from IP address Copyright ARRS. For personal use only; all rights reserved Downloaded from www.ajronline.org by 46.3.192.5 on 02/07/18 from IP address 46.3.192.5. Copyright RRS. For personal use only; all rights reserved C oil sensitivity encoding (SENSE) is a new technique that

More information

The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging

The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging JOURNAL OF MAGNETIC RESONANCE IMAGING 20:1046 1051 (2004) Technical Note The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging James W. Goldfarb, PhD* Purpose: To describe a known (but undocumented)

More information

IR/SR TrueFISP. Works-in-Progress package Version 1.2. For the SIEMENS Magnetom. Installation and User s Guide NUMARIS/4VA21B.

IR/SR TrueFISP. Works-in-Progress package Version 1.2. For the SIEMENS Magnetom. Installation and User s Guide NUMARIS/4VA21B. Works-in-Progress package Version 1.2 For the Installation and User s Guide NUMARIS/4VA21B January 22, 2003 Section of Medical Physics, University Hospital Freiburg, Germany Contact: Klaus Scheffler PhD

More information

Hardware. MRI System. MRI system Multicoil Microstrip. Part1

Hardware. MRI System. MRI system Multicoil Microstrip. Part1 Hardware MRI system Multicoil Microstrip MRI System Part1 1 The MRI system is made up of a variety of subsystems. the Operator Workspace Gradient Driver subsystem The Physiological Acquisition Controller

More information

Numerical Evaluation of an 8-element Phased Array Torso Coil for Magnetic Resonance Imaging

Numerical Evaluation of an 8-element Phased Array Torso Coil for Magnetic Resonance Imaging Numerical Evaluation of an 8-element Phased Array Torso Coil for Magnetic Resonance Imaging Feng Liu, Joe Li, Ian Gregg, Nick Shuley and Stuart Crozier School of Information Technology and Electrical Engineering,

More information

Application Guide & Release Notes

Application Guide & Release Notes Application Guide & Release Notes Inner-volume-imaging (IVI) EPI C2P Release 002a 1 September 2015 TMII Translational and Molecular Imaging Institute Conditions of Use This package is provided to support

More information

MR in RTP. MR Data for Treatment Planning: Spatial Accuracy Issues, Protocol Optimization, and Applications (Preview of TG117 Report) Acknowledgements

MR in RTP. MR Data for Treatment Planning: Spatial Accuracy Issues, Protocol Optimization, and Applications (Preview of TG117 Report) Acknowledgements MR Data for Treatment Planning: Issues, Protocol Optimization, and s (Preview of TG117 Report) Debra H. Brinkmann Mayo Clinic, Rochester MN Acknowledgements TG-117 Use of MRI Data in Treatment Planning

More information

Steady-state sequences: Spoiled and balanced methods

Steady-state sequences: Spoiled and balanced methods Steady-state sequences: Spoiled and balanced methods Karla L Miller, FMRIB Centre, University of Oxford What is steady-state imaging? In the context of MRI pulse sequences, the term steady state typically

More information

NIH Public Access Author Manuscript Magn Reson Med. Author manuscript; available in PMC 2010 July 21.

NIH Public Access Author Manuscript Magn Reson Med. Author manuscript; available in PMC 2010 July 21. NIH Public Access Author Manuscript Published in final edited form as: Magn Reson Med. 2010 April ; 63(4): 1092 1097. doi:10.1002/mrm.22223. Spatially Varying Fat-Water Excitation Using Short 2DRF Pulses

More information

NIH Public Access Author Manuscript Int J Cardiovasc Imaging. Author manuscript; available in PMC 2008 May 26.

NIH Public Access Author Manuscript Int J Cardiovasc Imaging. Author manuscript; available in PMC 2008 May 26. NIH Public Access Author Manuscript Published in final edited form as: Int J Cardiovasc Imaging. 2001 August ; 17(4): 287 296. A comparison of prospective and retrospective respiratory navigator gating

More information

Diffusion and Functional MRI of the Spinal Cord Methods and Clinical Applications

Diffusion and Functional MRI of the Spinal Cord Methods and Clinical Applications Diffusion and Functional MRI of the Spinal Cord Methods and Clinical Applications Susceptibility artifacts in DTI of the spinal cord J. Cohen-Adad Q-space imaging and axon diameter measurements Functional

More information

High-Resolution, Spin-Echo BOLD, and CBF fmri at 4and7T

High-Resolution, Spin-Echo BOLD, and CBF fmri at 4and7T Magnetic Resonance in Medicine 48:589 593 (2002) High-Resolution, Spin-Echo BOLD, and CBF fmri at 4and7T Timothy Q. Duong,* Essa Yacoub, Gregory Adriany, Xiaoping Hu, Kamil Ugurbil, J. Thomas Vaughan,

More information

MR in Tx Planning. Acknowledgements. Outline. Overview MR in RTP

MR in Tx Planning. Acknowledgements. Outline. Overview MR in RTP MR Data for Treatment Planning and Stereotactic Procedures: Sources of Distortion, Protocol Optimization, and Assessment (Preview of TG117 Report) Debra H. Brinkmann Mayo Clinic, Rochester MN Acknowledgements

More information

10. Phase Cycling and Pulsed Field Gradients Introduction to Phase Cycling - Quadrature images

10. Phase Cycling and Pulsed Field Gradients Introduction to Phase Cycling - Quadrature images 10. Phase Cycling and Pulsed Field Gradients 10.1 Introduction to Phase Cycling - Quadrature images The selection of coherence transfer pathways (CTP) by phase cycling or PFGs is the tool that allows the

More information

RAD 229: MRI Signals and Sequences

RAD 229: MRI Signals and Sequences RAD 229: MRI Signals and Sequences Brian Hargreaves All notes are on the course website web.stanford.edu/class/rad229 Course Goals Develop Intuition Understand MRI signals Exposure to numerous MRI sequences

More information

TimTX TrueShape. The parallel transmit architecture of the future. Answers for life.

TimTX TrueShape.  The parallel transmit architecture of the future. Answers for life. www.siemens.com/trueshape TimTX TrueShape The parallel transmit architecture of the future. The product/feature (mentioned herein) is not commercially available. Due to regulatory reasons its future availability

More information

Supplementary Figure 1

Supplementary Figure 1 Supplementary Figure 1 Left aspl Right aspl Detailed description of the fmri activation during allocentric action observation in the aspl. Averaged activation (N=13) during observation of the allocentric

More information

Image quality evaluation of turbo-spin echo diffusion weighted image (TSE-DWI) : A phantom study

Image quality evaluation of turbo-spin echo diffusion weighted image (TSE-DWI) : A phantom study Image quality evaluation of turbo-spin echo diffusion weighted image (TSE-DWI) : A phantom study Poster No.: C-0631 Congress: ECR 2016 Type: Scientific Exhibit Authors: T. Yoshida, A. Urikura, K. Shirata,

More information

MRI Anatomy and Positioning Series Module 12: Fat Suppression Techniques

MRI Anatomy and Positioning Series Module 12: Fat Suppression Techniques MRI Anatomy and Positioning Series Module 12: Fat Suppression Techniques 1 Introduction... 3 RF FatSat... 4 HOAST... 4 FatSat... 5 Segment FS... 8 PhaseCycle... 9 Water Excitation... 10 STIR... 12 FatSep...

More information

Efficiency of Background Suppression for Arterial Spin Labeling. Dairon Garcia

Efficiency of Background Suppression for Arterial Spin Labeling. Dairon Garcia Efficiency of Background Suppression for Arterial Spin Labeling by Dairon Garcia Submitted to the Department of Electrical Engineering and Computer Science in partial fulfillment of the requirements for

More information

Multi-Slice Perfusion-Based Functional MRI using the FAIR Technique: Comparison of CBF and BOLD effects

Multi-Slice Perfusion-Based Functional MRI using the FAIR Technique: Comparison of CBF and BOLD effects NMR IN BIOMEDICINE, VOL. 10, 191 196 (1997) Multi-Slice Perfusion-Based Functional MRI using the FAIR Technique: Comparison of CBF and BOLD effects Seong-Gi Kim, 1 Nikolaos V. Tsekos 1 and James Ashe 2

More information

Personal Information. Contact information. Education / Training. Position

Personal Information. Contact information. Education / Training. Position Curriculum Vitae Personal Information Name Current Position Title Wanyong Shin Research Assistant (Ph.D candidate) Contact information Address 676 N. St. Clair 1400A, Chicago, IL, 60611 Phone Work: (312)

More information

Page 1 of 9. Protocol: adult_other_adni3_study_human_ge_3t_25w_ _ _1. 3 Plane Localizer. 3 Plane Localizer PATIENT POSITION

Page 1 of 9. Protocol: adult_other_adni3_study_human_ge_3t_25w_ _ _1. 3 Plane Localizer. 3 Plane Localizer PATIENT POSITION 3 Localizer FOV 26.0 Slice Thickness 5.0 Slice Spacing 0.0 Freq 256 Phase 128 3-PLANE 3 Localizer Unswap Phase Correction Gradient Echo Imaging Options Seq, Fast Recon All Images Contrast Yes/ 3 Localizer

More information

Enhancing Gray-to-White Matter Contrast in 3T T1 Spin-Echo Brain Scans by Optimizing Flip Angle

Enhancing Gray-to-White Matter Contrast in 3T T1 Spin-Echo Brain Scans by Optimizing Flip Angle AJNR Am J Neuroradiol 26:2000 2004, September 2005 Enhancing Gray-to-White Matter Contrast in 3T T1 Spin-Echo Brain Scans by Optimizing Flip Angle Bernd L. Schmitz, Georg Grön, Florian Brausewetter, Martin

More information

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia M R I Physics Course Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia M R I Physics Course Magnetic Resonance Imaging Spatial

More information

Magnetic Resonance Imaging Principles, Methods, and Techniques

Magnetic Resonance Imaging Principles, Methods, and Techniques Magnetic Resonance Imaging Principles, Methods, and Techniques Perry Sprawls Jr., Emory University Publisher: Medical Physics Publishing Corporation Publication Place: Madison, Wisconsin Publication Date:

More information

Improvement of the Image Quality of Black-blood Magnetic Resonance Imaging with the Subtraction Double Inversion Recovery Technique

Improvement of the Image Quality of Black-blood Magnetic Resonance Imaging with the Subtraction Double Inversion Recovery Technique J Radiol Sci 2014; 39: 105-110 Improvement of the Image Quality of Black-blood Magnetic Resonance Imaging with the Subtraction Double Inversion Recovery Technique Shih-Ming Huang 1,2 Shu-Hui Peng 1,2 Jan-Yang

More information

Improving high-field MRI using parallel excitation

Improving high-field MRI using parallel excitation review Improving high-field MRI using parallel excitation MRI at high magnetic field strengths promises to deliver clearer images of the body s structure and function. However, high-field MRI currently

More information

Magnetic Resonance Imaging

Magnetic Resonance Imaging Magnetic Resonance Imaging Principles, Methods, and Techniques Perry Sprawls, Ph.D., FACR, FAAPM, FIOMP Distinguished Emeritus Professor Department of Radiology Emory University Atlanta, Georgia Medical

More information

Standards for Imaging Endpoints in Clinical Trials: Standardization and Optimization of Image Acquisitions: Magnetic Resonance

Standards for Imaging Endpoints in Clinical Trials: Standardization and Optimization of Image Acquisitions: Magnetic Resonance FDA Workshop April 13, 2010 Standards for Imaging Endpoints in Clinical Trials: Standardization and Optimization of Image Acquisitions: Magnetic Resonance Edward F. Jackson, PhD Professor and Chief, Section

More information

functional MRI: A primer

functional MRI: A primer Activation Leads to: functional MRI: A primer CBF Increased +ΔR CBV Increased +ΔR (C+) O Utilization Increased slightly? Venous [O ] Increased -ΔR* Glucose Utilization Increased? Lactate BOLD R=/T R=/T

More information

Improve Image Quality of Transversal Relaxation Time PROPELLER and FLAIR on Magnetic Resonance Imaging

Improve Image Quality of Transversal Relaxation Time PROPELLER and FLAIR on Magnetic Resonance Imaging Journal of Physics: Conference Series PAPER OPEN ACCESS Improve Image Quality of Transversal Relaxation Time PROPELLER and FLAIR on Magnetic Resonance Imaging To cite this article: N Rauf et al 2018 J.

More information

Fast Field-Cycling Magnetic Resonance Imaging (FFC-MRI)

Fast Field-Cycling Magnetic Resonance Imaging (FFC-MRI) Fast Field-Cycling Magnetic Resonance Imaging (FFC-MRI) David J. Lurie Aberdeen Biomedical Imaging Centre University of Aberdeen Summary of talk Short introduction to MRI Physics Field-Cycling MRI Field-Cycling

More information

The feasibility of breath-hold high-resolution 3D-MRCP obtained with 32 channel torso cardiac coil and T2-prepBTFE

The feasibility of breath-hold high-resolution 3D-MRCP obtained with 32 channel torso cardiac coil and T2-prepBTFE The feasibility of breath-hold high-resolution 3D-MRCP obtained with 32 channel torso cardiac coil and T2-prepBTFE Poster No.: C-0022 Congress: ECR 2010 Type: Scientific Exhibit Topic: Abdominal Viscera

More information

Multiplexed Echo Planar Imaging for Sub-Second Whole Brain FMRI and Fast Diffusion Imaging

Multiplexed Echo Planar Imaging for Sub-Second Whole Brain FMRI and Fast Diffusion Imaging Multiplexed Echo Planar Imaging for Sub-Second Whole Brain FMRI and Fast Diffusion Imaging David A. Feinberg 1,2,3 *, Steen Moeller 4, Stephen M. Smith 5, Edward Auerbach 4, Sudhir Ramanna 1,MattF. Glasser

More information

Spiral MRI on a 9.4T Vertical-bore Superconducting Magnet Using Unshielded and Self-shielded Gradient Coils

Spiral MRI on a 9.4T Vertical-bore Superconducting Magnet Using Unshielded and Self-shielded Gradient Coils Magn Reson Med Sci doi:10.2463/mrms.tn.2016-0049 Published Online: March 27, 2017 TECHNICAL NOTE Spiral MRI on a 9.4T Vertical-bore Superconducting Magnet Using Unshielded and Self-shielded Gradient Coils

More information

Inherent Insensitivity to RF Inhomogeneity in FLASH Imaging

Inherent Insensitivity to RF Inhomogeneity in FLASH Imaging Inherent Insensitivity to RF Inhomogeneity in FLASH Imaging Danli Wang, Keith Heberlein, Stephen LaConte, and Xiaoping Hu* Magnetic Resonance in Medicine 52:927 931 (2004) Radiofrequency (RF) field inhomogeneity

More information

Saturated Double-Angle Method for Rapid B 1 Mapping

Saturated Double-Angle Method for Rapid B 1 Mapping Saturated Double-Angle Method for Rapid B 1 Mapping Charles H. Cunningham, 1 John M. Pauly, 1 and Krishna S. Nayak 2 * Magnetic Resonance in Medicine 55:1326 1333 (2006) For in vivo magnetic resonance

More information

MARP. MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005

MARP. MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005 ACR MRI accreditation program MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005 Carl R. Keener, Ph.D., DABMP, DABR keener@marpinc.com MARP Medical & Radiation Physics,

More information

Echo-Planar Imaging for a 9.4 Tesla Vertical-Bore Superconducting Magnet Using an Unshielded Gradient Coil

Echo-Planar Imaging for a 9.4 Tesla Vertical-Bore Superconducting Magnet Using an Unshielded Gradient Coil Magn Reson Med Sci, Vol. XX, No. X, pp. XXX XXX, 2015 2016 Japanese Society for Magnetic Resonance in Medicine TECHNICAL NOTE by J-STAGE doi:10.2463/mrms.tn.2015-0123 Echo-Planar Imaging for a 9.4 Tesla

More information

Signal Extraction Technology

Signal Extraction Technology Signal Extraction Technology Technical bulletin Introduction Masimo SET pulse oximetry is a new and fundamentally distinct method of acquiring, processing and reporting arterial oxygen saturation and pulse

More information

Imaging the brain at ultra-high resolution using 3D FatNavs

Imaging the brain at ultra-high resolution using 3D FatNavs Imaging the brain at ultra-high resolution using 3D FatNavs Daniel Gallichan Centre d Imagerie BioMédicale EPFL, Lausanne, Switzerland Overview Introduction How motion affects MRI scans Ways we can track

More information

Applications Guide. Spectral Editing with SVS. (Works-in-Progress) MAGNETOM TaTs and Verio Systems (3T)

Applications Guide. Spectral Editing with SVS. (Works-in-Progress) MAGNETOM TaTs and Verio Systems (3T) Applications Guide Spectral Editing with SVS (Works-in-Progress) MAGNETOM TaTs and Verio Systems (3T) syngo MR Numaris 4 VB17A June 2009 Version 1.1 WIP #529 Important Note This document provides a description

More information

Research Article Implementation and Application of PSF-Based EPI Distortion Correction to High Field Animal Imaging

Research Article Implementation and Application of PSF-Based EPI Distortion Correction to High Field Animal Imaging Hindawi Publishing Corporation International Journal of Biomedical Imaging Volume 2009, Article ID 946271, 7 pages doi:10.1155/2009/946271 Research Article Implementation and Application of PSF-Based EPI

More information

Encoding of inductively measured k-space trajectories in MR raw data

Encoding of inductively measured k-space trajectories in MR raw data Downloaded from orbit.dtu.dk on: Apr 10, 2018 Encoding of inductively measured k-space trajectories in MR raw data Pedersen, Jan Ole; Hanson, Christian G.; Xue, Rong; Hanson, Lars G. Publication date:

More information

Optimization of Energy Modulation Filter for Dual Energy CBCT Using Geant4 Monte-Carlo Simulation

Optimization of Energy Modulation Filter for Dual Energy CBCT Using Geant4 Monte-Carlo Simulation Original Article PROGRESS in MEDICAL PHYSICS 27(3), Sept. 2016 http://dx.doi.org/10.14316/pmp.2016.27.3.125 pissn 2508-4445, eissn 2508-4453 Optimization of Energy Modulation Filter for Dual Energy CBCT

More information

The development of the RF-pulse for the low level SAR used by the MRI.

The development of the RF-pulse for the low level SAR used by the MRI. The development of the RF-pulse for the low level SAR used by the MRI. Kojiro Yamaguchi a*, Eizo Umezawa a, Sachiko Ueoku b, Kazuhiro Katada c a Faculty of radiological technology, School of Health Science,

More information

Methods. Experimental Stimuli: We selected 24 animals, 24 tools, and 24

Methods. Experimental Stimuli: We selected 24 animals, 24 tools, and 24 Methods Experimental Stimuli: We selected 24 animals, 24 tools, and 24 nonmanipulable object concepts following the criteria described in a previous study. For each item, a black and white grayscale photo

More information

The promise of high-field MRI. High Field MRI Technology, Applications, Safety, and Limitations. High-field Scanners

The promise of high-field MRI. High Field MRI Technology, Applications, Safety, and Limitations. High-field Scanners High Field MRI Technology, Applications, Safety, and Limitations R. Jason Stafford, Ph.D. Department of Imaging Physics The University of Texas M. D. Anderson Cancer Center Houston, TX The promise of high-field

More information

The ACR magnetic resonance accreditation phantom (ACR MRAP) has been designed to examine a broad range of instrument parameters.

The ACR magnetic resonance accreditation phantom (ACR MRAP) has been designed to examine a broad range of instrument parameters. OVERVIEW OF THE ACR MRI ACCREDITATION PHANTOM Geoffrey D. Clarke, Ph.D. University of Texas Southwestern Medical Center at Dallas email: GEOFFREY.CLARKE@MAIL.SWMED.EDU INTRODUCTION The ACR magnetic resonance

More information

MRI at a Glance. Catherine Westbrook. Blackwell Science

MRI at a Glance. Catherine Westbrook. Blackwell Science MRI at a Glance Catherine Westbrook Blackwell Science MRI at a Glance MRI at a Glance CATHERINE WESTBROOK MSC DCRR CTC Director of Training and Education Lodestone Patient Care Ltd Blackwell Science 2002

More information

Correction of the local intensity nonuniformity artifact in high field MRI

Correction of the local intensity nonuniformity artifact in high field MRI Correction of the local intensity nonuniformity artifact in high field MRI Poster No.: C-0346 Congress: ECR 2012 Type: Authors: Keywords: DOI: Scientific Paper S. Kai, S. Kumazawa, H. Yabuuchi, F. Toyofuku;

More information

Development of a 1.0 T MR microscope using a Nd-Fe-B permanent magnet

Development of a 1.0 T MR microscope using a Nd-Fe-B permanent magnet Magnetic Resonance Imaging 19 (2001) 875 880 Development of a 1.0 T MR microscope using a Nd-Fe-B permanent magnet Tomoyuki Haishi, Takaaki Uematsu, Yoshimasa Matsuda, Katsumi Kose* Institute of Applied

More information

Variable-Rate Selective Excitation for Rapid MRI Sequences

Variable-Rate Selective Excitation for Rapid MRI Sequences Variable-Rate Selective Excitation for Rapid MRI Sequences Brian A. Hargreaves,* Charles H. Cunningham, Dwight G. Nishimura, and Steven M. Conolly Magnetic Resonance in Medicine 52:590 597 (2004) Balanced

More information

SIGNA Explorer Lift revives our MR

SIGNA Explorer Lift revives our MR Seiji Shiotani, MD, PhD Seirei Fuji Hospital in Fuji City, Shizuoka, Japan Masayoshi Sugimura Seirei Fuji Hospital in Fuji City, Shizuoka, Japan SIGN Explorer Lift revives our MR The clinical usefulness

More information

SUPPLEMENTARY INFORMATION

SUPPLEMENTARY INFORMATION a b STS IOS IOS STS c "#$"% "%' STS posterior IOS dorsal anterior ventral d "( "& )* e f "( "#$"% "%' "& )* Supplementary Figure 1. Retinotopic mapping of the non-lesioned hemisphere. a. Inflated 3D representation

More information