Fast Field-Cycling Magnetic Resonance Imaging (FFC-MRI)

Size: px
Start display at page:

Download "Fast Field-Cycling Magnetic Resonance Imaging (FFC-MRI)"

Transcription

1 Fast Field-Cycling Magnetic Resonance Imaging (FFC-MRI) David J. Lurie Aberdeen Biomedical Imaging Centre University of Aberdeen

2 Summary of talk Short introduction to MRI Physics Field-Cycling MRI Field-Cycling MRI Scanners Applications of Field-Cycling MRI

3 Physics of MRI

4 Central idea behind MRI:

5 Magnetic field gradient Magnetic field gradient is: linear variation of the strength of the applied magnetic field with position across the sample Without gradient: B z = B 0 everywhere With x-gradient (for example): B z (x) = B 0 + x.g x position (m) gradient (T/m)

6 Magnetic field gradient Main Field B z = B 0 PLUS Gradient Field B z (x) = x.g x EQUALS Total Field B z (x) = B 0 + x.g x 0 x (position)

7 Magnetic field gradient B z B 0 B z (x) = B 0 + x.g x 0 x ω ω 0 ω(x) = γ B z (x) = γ(b 0 + x.g x ) 0 x = ω 0 + γ.x.g x

8 Magnetic Resonance Imaging What is the effect of a gradient in the x-direction? 0 x (position)

9 Magnetic Resonance Imaging e.g. applied gradient G x = 0.5 mt/m Centre: B z = B 0 = 1.0 T f = 42.6 MHz 25 cm left: B z = T f = MHz 25 cm right: B z = T f = MHz 0 x (position)

10 Magnetic Resonance Imaging e.g. applied gradient G x = 0.5 mt/m 0 x (position)

11 Magnetic Resonance Imaging e.g. applied gradient G x = 0.5 mt/m Centre: B z = B 0 = 1.0 T f = 42.6 MHz 25 cm left: B z = T f = MHz 25 cm right: B z = T f = MHz 0 x (position)

12 Magnetic field Frequency encoding Observe NMR signal with gradient on 0 Position Bottles of water in the scanner NMR signal from bottle low frequency, large size medium frequency, small size high frequency, medium size total signal

13 Frequency encoding NMR Signals Amount of signal Frequency of signal This gives a one-dimensional projection of the objects

14 Frequency encoding NMR Signals Amount of signal Frequency of signal Moving a bottle changes its frequency & the projection

15 Frequency encoding NMR Signals Amount of signal Frequency of signal Moving a bottle changes its frequency & the projection

16 Magnetic Resonance Imaging Frequency analysis of NMR signals (Fourier transform) allows images to be produced Gradients are used in 3 dimensions

17 Main parts of an MRI scanner Very strong magnet (usually 1.5 or 3 tesla) Radiofrequency coil (a head coil in this case)

18 Main parts of an MRI scanner

19 Motivation for Field-Cycling MRI

20 Standard MRI vs Field-Cycling MRI Standard MRI The patient is placed inside a large magnet with a strong, fixed magnetic field (e.g. 1.5 T or 3.0 T) Field-Cycling MRI Instead of being constant, the magnetic field (B 0 ) is switched (cycled) between different levels during the collection of a scan The hypothesis is that this can provide extra information, especially about proteins in the body Could improve diagnosis of a wide range of diseases

21 T 1 based image contrast Standard MRI Much of the contrast in conventional MRI arises from differences in the T 1 relaxation time of tissues Typical T 1 -weighted proton MR image

22 Extra information: T 1 Dispersion Studies on small tissue samples have shown that the way in which T 1 changes with field strength ( T 1 dispersion ) could also be a marker of disease However, this information is completely hidden to conventional MRI scanners, because each scanner can only operate at its own fixed value of magnetic field, e.g. 1.5 tesla 3.0 tesla 7.0 tesla

23 T 1 Dispersion a new kind of contrast Our idea is to measure T 1 dispersion in patients, and use it as a completely new marker of disease FFC-MRI Normal Field strength

24 Motivation for FC-MRI So how can we measure T 1 vs field? (a) Buy 100 MRI scanners covering a large range of magnetic fields 0.01 T 0.02 T 0.03 T 0.04 T 0.05 T

25 Motivation for FC-MRI So how can we measure T 1 vs field? (b) Buy one MRI scanner with a variable field (if it was available) B 0

26 Motivation for FC-MRI So how can we measure T 1 vs field? (b) Buy one MRI scanner with a variable field (if it was available)

27 Motivation for FC-MRI So how can we measure T 1 vs field? (b) Buy one MRI scanner with a variable field (if it was available) But the RF coil(s) would have to be retuned at every field...

28 Motivation for FC-MRI So how can we measure T 1 vs field? (c) Use magnetic field cycling! and first, build the equipment

29 How does Field-Cycling work?

30 Introduction to Field-Cycling In Field-Cycling MRI, the magnetic field is switched between two (or more) levels during the MRI procedure

31 Low Field

32 High Field

33 Two fields in one experiment Prepare at low field Detect at higher field

34 Generalised FC method POLARISATION EVOLUTION DETECTION B 0 field B 0 P field B 0 D field B 0 E time Relaxation Relaxation occurs during the Evolution period, usually at low magnetic field The result is read out during the detection period, always at the same (higher) field

35 FFC-MRI imaging pulse sequence T ev B 0 B 0 typically 500 ms B 0 E D RF 90 Gradients Signal

36 FFC-MRI Scanners

37 FFC-MRI Scanners It is not (yet) possible to buy an FFC-MRI scanner from medical imaging companies So we have built our own scanners Most of the FFC-MRI scanner is the same as a standard scanner, except for the magnet, which is very different The magnet in a standard clinical MRI scanner cannot be switched to different magnetic fields (Though field-offset insert coils can be used, as presented by Ludovic de Rochefort yesterday)

38 Double magnet system (0.06 T) Primary Magnet (permanent) Vertical field 59 mt Secondary Magnet (resistive) Vertical field 55 mt Cancellation of magnetic fields: 4 mt at the patient

39 Whole-body FFC-MRI scanner (0.06 T)

40 Single magnet system (0.2 T) Higher detection field (better SNR) More flexible free choice of magnetic fields during the whole pulse sequence But field (in)stability can be problematic

41 Single magnet system (0.2 T) International Electric Company (Finland) 3 circuits, each supplied with 650 A

42 Images from 0.2 T FFC-MRI system

43 Information from T 1 dispersion

44 Remember the dispersion concept? The dispersion curve contains important disease-dependent information

45 The wiggles in the curves are Quadrupolar Dips Arise due to the coincidence of NMR and NQR interaction frequencies in immobile proteins T 1 (ms) T 1 NMR poly-l-alanine (-1 C) 14 N 1 H Frequency (MHz) Kimmich et al., J. Magn. Reson. 68, 263 (1986) H NMR 14 N Nuclear Quadrupole Resonance Magnetic Field (mt)

46 Quadrupolar Dips The quadrupolar dips are sensitive to protein concentration and motion By fitting the dispersion curves to theoretical models, we can extract parameters to use as quantitative biomarkers of disease

47 Biomedical applications

48 Biomedical applications Lionel Broche has led the biomedical applications work in our lab Lionel will give a talk on the applications tomorrow (14:30) I will give one example

49 Cartilage in Osteoarthritis Osteoarthritis is a debilitating disease of joints Can FFC-NMR or FFC-MRI detect disease at an early stage, and monitor treatment? First step was in vitro measurements of T 1 dispersion in normal and osteoarthritic cartilage

50 FFC-MRI results in osteoarthritis Patients undergoing joint-replacement surgery agreed to have FFC measurements made on their joints after they had been removed Relaxation rate (s -1 ) osteoarthritis OA cartilage data Lorentzian fit: R 2 = Healthy cartilage data Lorentzian fit: R 2 = healthy cartlilage Evolution field (MHz 1 H)

51 Conclusions

52 Conclusions Fast Field-cycling MRI can give new types of endogenous contrast, based on T 1 dispersion & Quadrupole peaks Initial results show that FFC-NMR relaxometry and FFC-MRI can give useful information about disease processes That information is invisible to conventional MRI, so represents a new type of image contrast for MRI We are currently investigating FFC-MRI applications in osteoarthritis, cancer, liver disease (fibrosis) and neurodegeneration (Alzheimer s and Parkinson s)

53 Members of the FFC-MRI group Dr. Lionel Broche Dr. Gareth Davies Dr. James Ross Nick Payne Vasilis Zampetoulas

54 Sources of Funding

55 The IDentIFY H2020 project

56 The IDentIFY H2020 project Improving Diagnosis by Fast Field-Cycling MRI Consortium of 7 universities/ research institutes and 2 SMEs in 6 countries Aberdeen is coordinating partner 4-year project started in January 2016 Total funding of 6.60m

57 IDentIFY Partners International Electric Company University of Aberdeen Technical University of Ilmenau University of Warmia and Mazury in Olsztyn INSERM, CEA, CNRS in Grenoble Stelar, S.r.l. University of Turin

58 IDentIFY: Aims & Objectives Improve the FFC-MRI technology Stabilise magnetic field Correct for environmental fields Speed up acquisition Develop FFC-tailored contrast agents Develop relaxation theory Understand the quadrupolar effects Fit data to reliable models Extract quantitative biomarkers of disease Demonstrate effectiveness of FFC-MRI On tissue samples (biobanks and from surgery) On small numbers of patients Bring FFC-MRI closer to the marketplace and to adoption in clinical radiology & research

59 Thanks for your attention! This project has received funding from the European Union s Horizon 2020 research and innovation programme under grant agreement No (project IDentIFY )

60

1 Introduction. 2 The basic principles of NMR

1 Introduction. 2 The basic principles of NMR 1 Introduction Since 1977 when the first clinical MRI scanner was patented nuclear magnetic resonance imaging is increasingly being used for medical diagnosis and in scientific research and application

More information

H 2 O and fat imaging

H 2 O and fat imaging H 2 O and fat imaging Xu Feng Outline Introduction benefit from the separation of water and fat imaging Chemical Shift definition of chemical shift origin of chemical shift equations of chemical shift

More information

MAGNETIC RESONANCE IMAGING

MAGNETIC RESONANCE IMAGING CSEE 4620 Homework 3 Fall 2018 MAGNETIC RESONANCE IMAGING 1. THE PRIMARY MAGNET Magnetic resonance imaging requires a very strong static magnetic field to align the nuclei. Modern MRI scanners require

More information

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia M R I Physics Course Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia M R I Physics Course Magnetic Resonance Imaging Spatial

More information

STELAR s.r.l. via E. Fermi, Mede (PV) - Italy tel Uni-Pavia 24/11/06

STELAR s.r.l. via E. Fermi, Mede (PV) - Italy tel Uni-Pavia 24/11/06 STELAR s.r.l. via E. Fermi,4 27035 Mede (PV) - Italy tel. +39 0384 820096 www.stelar.it info@stelar.it Uni-Pavia 24/11/06 From theory to practice... Gianni Ferrante Stelar School on Field Cycling NMR relaxometry

More information

Image Quality/Artifacts Frequency (MHz)

Image Quality/Artifacts Frequency (MHz) The Larmor Relation 84 Image Quality/Artifacts (MHz) 42 ω = γ X B = 2πf 84 0.0 1.0 2.0 Magnetic Field (Tesla) 1 A 1D Image Magnetic Field Gradients Magnet Field Strength Field Strength / Gradient Coil

More information

Off-resonance Magnetisation Transfer Contrast (MTC) MRI using Fast Field-Cycling (FFC)

Off-resonance Magnetisation Transfer Contrast (MTC) MRI using Fast Field-Cycling (FFC) Off-resonance Magnetisation Transfer Contrast (MTC) MRI using Fast Field-Cycling (FFC) Chang-Hoon Choi, Gareth R. Davies, and David J. Lurie Aberdeen Biomedical Imaging Centre, University of Aberdeen,

More information

Background (~EE369B)

Background (~EE369B) Background (~EE369B) Magnetic Resonance Imaging D. Nishimura Overview of NMR Hardware Image formation and k-space Excitation k-space Signals and contrast Signal-to-Noise Ratio (SNR) Pulse Sequences 13

More information

The development of the RF-pulse for the low level SAR used by the MRI.

The development of the RF-pulse for the low level SAR used by the MRI. The development of the RF-pulse for the low level SAR used by the MRI. Kojiro Yamaguchi a*, Eizo Umezawa a, Sachiko Ueoku b, Kazuhiro Katada c a Faculty of radiological technology, School of Health Science,

More information

HETERONUCLEAR IMAGING. Topics to be Discussed:

HETERONUCLEAR IMAGING. Topics to be Discussed: HETERONUCLEAR IMAGING BioE-594 Advanced MRI By:- Rajitha Mullapudi 04/06/2006 Topics to be Discussed: What is heteronuclear imaging. Comparing the hardware of MRI and heteronuclear imaging. Clinical applications

More information

2014 M.S. Cohen all rights reserved

2014 M.S. Cohen all rights reserved 2014 M.S. Cohen all rights reserved mscohen@g.ucla.edu IMAGE QUALITY / ARTIFACTS SYRINGOMYELIA Source http://gait.aidi.udel.edu/res695/homepage/pd_ortho/educate/clincase/syrsco.htm Surgery is usually recommended

More information

Portable MRI Scanner by

Portable MRI Scanner by Portable MRI Scanner by ENSC 305/440 Simon Fraser University Team J: Anterpal Singh Sandhu Barry Yim Evangeline Yee Gagandeep Kaur Robin Wisniewski Date: 22 December 2015 Team Profile Evangeline Yee, CEO

More information

MRI SYSTEM COMPONENTS Module One

MRI SYSTEM COMPONENTS Module One MRI SYSTEM COMPONENTS Module One 1 MAIN COMPONENTS Magnet Gradient Coils RF Coils Host Computer / Electronic Support System Operator Console and Display Systems 2 3 4 5 Magnet Components 6 The magnet The

More information

Standards for Imaging Endpoints in Clinical Trials: Standardization and Optimization of Image Acquisitions: Magnetic Resonance

Standards for Imaging Endpoints in Clinical Trials: Standardization and Optimization of Image Acquisitions: Magnetic Resonance FDA Workshop April 13, 2010 Standards for Imaging Endpoints in Clinical Trials: Standardization and Optimization of Image Acquisitions: Magnetic Resonance Edward F. Jackson, PhD Professor and Chief, Section

More information

Weber State University Radiologic Technology 4603

Weber State University Radiologic Technology 4603 Weber State University Radiologic Technology 4603 MRI Physics and Instrumentation Instructor: Rex T. Christensen MHA R.T. (R) (MR) (CT) (ARRT) CIIP Contact Info: E-mail: rexchristensen@weber.edu Phone:

More information

Experience in implementing continuous arterial spin labeling on a commercial MR scanner

Experience in implementing continuous arterial spin labeling on a commercial MR scanner JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 6, NUMBER 1, WINTER 2005 Experience in implementing continuous arterial spin labeling on a commercial MR scanner Theodore R. Steger and Edward F. Jackson

More information

(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010

(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010 (N)MR Imaging Lab Course Script FMP PhD Autumn School Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder Date: November 3rd, 2010 1 Purpose: Understanding the basic principles of MR imaging

More information

Chapter 11 Coherence Editing: Pulse-field Gradients and Phase Cycling

Chapter 11 Coherence Editing: Pulse-field Gradients and Phase Cycling Chapter 11 Coherence Editing: Pulse-field Gradients and Phase Cycling Coherence editing is used to remove unwanted signals from NMR spectra. For example, in the double quantum filtered COSY experiment,

More information

6.S02 MRI Lab Acquire MR signals. 2.1 Free Induction decay (FID)

6.S02 MRI Lab Acquire MR signals. 2.1 Free Induction decay (FID) 6.S02 MRI Lab 1 2. Acquire MR signals Connecting to the scanner Connect to VMware on the Lab Macs. Download and extract the following zip file in the MRI Lab dropbox folder: https://www.dropbox.com/s/ga8ga4a0sxwe62e/mit_download.zip

More information

Cardiac MR. Dr John Ridgway. Leeds Teaching Hospitals NHS Trust, UK

Cardiac MR. Dr John Ridgway. Leeds Teaching Hospitals NHS Trust, UK Cardiac MR Dr John Ridgway Leeds Teaching Hospitals NHS Trust, UK Cardiac MR Physics for clinicians: Part I Journal of Cardiovascular Magnetic Resonance 2010, 12:71 http://jcmr-online.com/content/12/1/71

More information

Potential Risks of MRI in Device Patients

Potential Risks of MRI in Device Patients Outline Potential Risks of MRI in Device Patients Redha Boubertakh r.boubertakh@qmul.ac.uk MRI and cardiac implantable electronic devices (CIED) Components of an MRI scanner MRI implant and device safety

More information

2015 Spin echoes and projection imaging

2015 Spin echoes and projection imaging 1. Spin Echoes 1.1 Find f0, transmit amplitudes, and shim settings In order to acquire spin echoes, we first need to find the appropriate scanner settings using the FID GUI. This was all done last week,

More information

Noninvasive Blood Flow Mapping with Arterial Spin Labeling (ASL) Paul Kyu Han and Sung-Hong Park

Noninvasive Blood Flow Mapping with Arterial Spin Labeling (ASL) Paul Kyu Han and Sung-Hong Park Noninvasive Blood Flow Mapping with Arterial Spin Labeling (ASL) Paul Kyu Han and Sung-Hong Park Department of Bio and Brain Engineering, Korea Advanced Institute of Science and Technology (KAIST), Daejeon,

More information

Numerical Evaluation of an 8-element Phased Array Torso Coil for Magnetic Resonance Imaging

Numerical Evaluation of an 8-element Phased Array Torso Coil for Magnetic Resonance Imaging Numerical Evaluation of an 8-element Phased Array Torso Coil for Magnetic Resonance Imaging Feng Liu, Joe Li, Ian Gregg, Nick Shuley and Stuart Crozier School of Information Technology and Electrical Engineering,

More information

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging Medical Imaging X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging From: Physics for the IB Diploma Coursebook 6th Edition by Tsokos, Hoeben and Headlee And Higher Level Physics 2 nd Edition

More information

Field Simulation Software to Improve Magnetic Resonance Imaging

Field Simulation Software to Improve Magnetic Resonance Imaging Field Simulation Software to Improve Magnetic Resonance Imaging a joint project with the NRI in South Korea CST Usergroup Meeting 2010 Darmstadt Institute for Biometry and Medicine Informatics J. Mallow,

More information

High Field MRI: Technology, Applications, Safety, and Limitations

High Field MRI: Technology, Applications, Safety, and Limitations High Field MRI: Technology, Applications, Safety, and Limitations R. Jason Stafford, Ph.D. The University of Texas M. D. Anderson Cancer Center, Houston, TX Introduction The amount of available signal

More information

Hardware. MRI System. MRI system Multicoil Microstrip. Part1

Hardware. MRI System. MRI system Multicoil Microstrip. Part1 Hardware MRI system Multicoil Microstrip MRI System Part1 1 The MRI system is made up of a variety of subsystems. the Operator Workspace Gradient Driver subsystem The Physiological Acquisition Controller

More information

Compact yet Sophisticated

Compact yet Sophisticated Compact yet Sophisticated Hitachi has brought Open MRI one step further in its evolution, to better assist medical professionals who work at the forefront of healthcare. AIRIS Light MSK offers radiologists

More information

Review of Progress in Quantitative Nondestructive Evaluation, Vol. 14 Edited by D.O. Thompson and D.E. Chimenti, Plenum Press, New Yorlc,l99S 1749

Review of Progress in Quantitative Nondestructive Evaluation, Vol. 14 Edited by D.O. Thompson and D.E. Chimenti, Plenum Press, New Yorlc,l99S 1749 MAGNETIC RESONANCE IMAGING AND SPECI'ROSCOPY USING SQUID DETECTION S. Kumar, W.F. Avrin, and B.D. Thorson Quantum Magnetics, Inc. 11578 Sorrento Valley Road San Diego, CA 92121 INTRODUCTION Magnetic Resonance

More information

Encoding of inductively measured k-space trajectories in MR raw data

Encoding of inductively measured k-space trajectories in MR raw data Downloaded from orbit.dtu.dk on: Apr 10, 2018 Encoding of inductively measured k-space trajectories in MR raw data Pedersen, Jan Ole; Hanson, Christian G.; Xue, Rong; Hanson, Lars G. Publication date:

More information

ISSN X CODEN (USA): PCHHAX. The role of dual spin echo in increasing resolution in diffusion weighted imaging of brain

ISSN X CODEN (USA): PCHHAX. The role of dual spin echo in increasing resolution in diffusion weighted imaging of brain Available online at www.derpharmachemica.com ISSN 0975-413X CODEN (USA): PCHHAX Der Pharma Chemica, 2016, 8(17):15-20 (http://derpharmachemica.com/archive.html) The role of in increasing resolution in

More information

Half-Pulse Excitation Pulse Design and the Artifact Evaluation

Half-Pulse Excitation Pulse Design and the Artifact Evaluation Half-Pulse Excitation Pulse Design and the Artifact Evaluation Phillip Cho. INRODUCION A conventional excitation scheme consists of a slice-selective RF excitation followed by a gradient-refocusing interval

More information

Correction of the local intensity nonuniformity artifact in high field MRI

Correction of the local intensity nonuniformity artifact in high field MRI Correction of the local intensity nonuniformity artifact in high field MRI Poster No.: C-0346 Congress: ECR 2012 Type: Authors: Keywords: DOI: Scientific Paper S. Kai, S. Kumazawa, H. Yabuuchi, F. Toyofuku;

More information

SYLLABUS. 1. Identification of Subject:

SYLLABUS. 1. Identification of Subject: SYLLABUS Date/ Revision : 30 January 2017/1 Faculty : Life Sciences Approval : Dean, Faculty of Life Sciences SUBJECT : Biophysics 1. Identification of Subject: Name of Subject : Biophysics Code of Subject

More information

MRI Summer Course Lab 2: Gradient Echo T1 & T2* Curves

MRI Summer Course Lab 2: Gradient Echo T1 & T2* Curves MRI Summer Course Lab 2: Gradient Echo T1 & T2* Curves Experiment 1 Goal: Examine the effect caused by changing flip angle on image contrast in a simple gradient echo sequence and derive T1-curves. Image

More information

3T Unlimited. ipat on MAGNETOM Allegra The Importance of ipat at 3T. medical

3T Unlimited. ipat on MAGNETOM Allegra The Importance of ipat at 3T. medical 3T Unlimited ipat on MAGNETOM Allegra The Importance of ipat at 3T s medical ipat on MAGNETOM Allegra The Importance of ipat at 3T The rise of 3T MR imaging Ultra High Field MR (3T) has flourished during

More information

MARP. MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005

MARP. MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005 ACR MRI accreditation program MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005 Carl R. Keener, Ph.D., DABMP, DABR keener@marpinc.com MARP Medical & Radiation Physics,

More information

Analysis of magnetic and electromagnetic field emissions produced by a MRI device

Analysis of magnetic and electromagnetic field emissions produced by a MRI device Sept. 8-1, 21, Kosice, Slovakia Analysis of magnetic and electromagnetic field emissions produced by a MRI device D. Giordano, M. Borsero, G. Crotti, M. ucca INRIM Istituto Nazionale di Ricerca Metrologica,

More information

MRI Metal Artifact Reduction

MRI Metal Artifact Reduction MRI Metal Artifact Reduction PD Dr. med. Reto Sutter University Hospital Balgrist Zurich University of Zurich OUTLINE Is this Patient suitable for MR Imaging? Metal artifact reduction Is this Patient suitable

More information

DICOM Conformance. DICOM Detailed Specification for Diagnostic Labs and Radiology Center Connectivity

DICOM Conformance. DICOM Detailed Specification for Diagnostic Labs and Radiology Center Connectivity DICOM Detailed Specification for Diagnostic Labs and Radiology Center Connectivity Authored by Global Engineering Team, Health Gorilla April 10, 2014 Table of Contents About Health Gorilla s Online Healthcare

More information

Downloaded from by on 02/07/18 from IP address Copyright ARRS. For personal use only; all rights reserved

Downloaded from  by on 02/07/18 from IP address Copyright ARRS. For personal use only; all rights reserved Downloaded from www.ajronline.org by 46.3.192.5 on 02/07/18 from IP address 46.3.192.5. Copyright RRS. For personal use only; all rights reserved C oil sensitivity encoding (SENSE) is a new technique that

More information

Multi-channel SQUID-based Ultra-Low Field Magnetic Resonance Imaging in Unshielded Environment

Multi-channel SQUID-based Ultra-Low Field Magnetic Resonance Imaging in Unshielded Environment Multi-channel SQUID-based Ultra-Low Field Magnetic Resonance Imaging in Unshielded Environment Andrei Matlashov, Per Magnelind, Shaun Newman, Henrik Sandin, Algis Urbaitis, Petr Volegov, Michelle Espy

More information

ISMRM weekend educational course, MR Systems Engineering, Console Electronics

ISMRM weekend educational course, MR Systems Engineering, Console Electronics ISMRM weekend educational course, MR Systems Engineering, Console Electronics. 2013-4-20 Declaration of Relevant Financial Interests or Relationships Speaker Name: Katsumi Kose, Ph.D. I have the following

More information

Human Subjects in fmri Research

Human Subjects in fmri Research HST.583: Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Harvard-MIT Division of Health Sciences and Technology Course Instructor: Dr. Randy Gollub. Human Subjects in fmri Research

More information

Magnetic Resonance Imaging Principles, Methods, and Techniques

Magnetic Resonance Imaging Principles, Methods, and Techniques Magnetic Resonance Imaging Principles, Methods, and Techniques Perry Sprawls Jr., Emory University Publisher: Medical Physics Publishing Corporation Publication Place: Madison, Wisconsin Publication Date:

More information

Inherent Insensitivity to RF Inhomogeneity in FLASH Imaging

Inherent Insensitivity to RF Inhomogeneity in FLASH Imaging Inherent Insensitivity to RF Inhomogeneity in FLASH Imaging Danli Wang, Keith Heberlein, Stephen LaConte, and Xiaoping Hu* Magnetic Resonance in Medicine 52:927 931 (2004) Radiofrequency (RF) field inhomogeneity

More information

Introduction to MR Hardware. RF Coils C M L C T. = g * B 0. Rotating magnetization produces alternating magnetic field

Introduction to MR Hardware. RF Coils C M L C T. = g * B 0. Rotating magnetization produces alternating magnetic field Introduction to MR Hardware RF Coils Dominik v. Elverfeldt Sep 5 th 2012 Courtesy of Hans Weber, Freiburg C M R = 50 Transmission = B 0 Reception L C T R Oscillating with Lamor frequency. B 1 field perpendicular

More information

Magnetic Resonance Imaging

Magnetic Resonance Imaging Magnetic Resonance Imaging Principles, Methods, and Techniques Perry Sprawls, Ph.D., FACR, FAAPM, FIOMP Distinguished Emeritus Professor Department of Radiology Emory University Atlanta, Georgia Medical

More information

Magnetization transfer attenuation of creatine resonances in localized proton MRS of human brain in vivo

Magnetization transfer attenuation of creatine resonances in localized proton MRS of human brain in vivo NMR IN BIOMEDICINE NMR Biomed. 1999;12:490 494 Magnetization transfer attenuation of creatine resonances in localized proton MRS of human brain in vivo Gunther Helms* and Jens Frahm Biomedizinische NMR

More information

Principios Básicos de RMN en sólidos destinado a usuarios. Gustavo Monti. Fa.M.A.F. Universidad Nacional de Córdoba Argentina

Principios Básicos de RMN en sólidos destinado a usuarios. Gustavo Monti. Fa.M.A.F. Universidad Nacional de Córdoba Argentina Principios Básicos de RMN en sólidos destinado a usuarios Gustavo Monti Fa.M.A.F. Universidad Nacional de Córdoba Argentina magnet 1 2 4 5 6 computer 3 Block diagrama of a traditional NMR spectrometer.

More information

Applications Guide. Spectral Editing with SVS. (Works-in-Progress) MAGNETOM TaTs and Verio Systems (3T)

Applications Guide. Spectral Editing with SVS. (Works-in-Progress) MAGNETOM TaTs and Verio Systems (3T) Applications Guide Spectral Editing with SVS (Works-in-Progress) MAGNETOM TaTs and Verio Systems (3T) syngo MR Numaris 4 VB17A June 2009 Version 1.1 WIP #529 Important Note This document provides a description

More information

MR in Tx Planning. Acknowledgements. Outline. Overview MR in RTP

MR in Tx Planning. Acknowledgements. Outline. Overview MR in RTP MR Data for Treatment Planning and Stereotactic Procedures: Sources of Distortion, Protocol Optimization, and Assessment (Preview of TG117 Report) Debra H. Brinkmann Mayo Clinic, Rochester MN Acknowledgements

More information

Pulse Sequence Design and Image Procedures

Pulse Sequence Design and Image Procedures Pulse Sequence Design and Image Procedures 1 Gregory L. Wheeler, BSRT(R)(MR) MRI Consultant 2 A pulse sequence is a timing diagram designed with a series of RF pulses, gradients switching, and signal readout

More information

ViP MRI: virtual phantom magnetic resonance imaging.

ViP MRI: virtual phantom magnetic resonance imaging. ViP MRI: virtual phantom magnetic resonance imaging. Hervé Saint-Jalmes, Pierre-Antoine Eliat, Johanne Bezy-Wendling, Alejandro Bordelois, Giulio Gambarota To cite this version: Hervé Saint-Jalmes, Pierre-Antoine

More information

nothing is more important than health

nothing is more important than health nothing is more important than health affidea affinity with doctors and patients affidea trust and fidelity in everything we do affidea constant progress through ideas and innovation Much has changed in

More information

Introduction. Parametric Imaging. The Ultrasound Research Interface: A New Tool for Biomedical Investigations

Introduction. Parametric Imaging. The Ultrasound Research Interface: A New Tool for Biomedical Investigations The Ultrasound Research Interface: A New Tool for Biomedical Investigations Shelby Brunke, Laurent Pelissier, Kris Dickie, Jim Zagzebski, Tim Hall, Thaddeus Wilson Siemens Medical Systems, Issaquah WA

More information

Magnetic Resonance Research Facility (MRRF) Resources

Magnetic Resonance Research Facility (MRRF) Resources Magnetic Resonance Research Facility (MRRF) Resources The Magnetic Resonance Research Facility (MRRF) has scanners located in both the hospital and research buildings on the campus of the University of

More information

Workshop on Rapid Scan EPR. University of Denver EPR Center and Bruker BioSpin July 28, 2013

Workshop on Rapid Scan EPR. University of Denver EPR Center and Bruker BioSpin July 28, 2013 Workshop on Rapid Scan EPR University of Denver EPR Center and Bruker BioSpin July 28, 2013 Direct detection Direct detected magnetic resonance that is, without modulation and phase-sensitive detection

More information

Gradients. Effects of B0 gradients on transverse magnetisation Similar to figure 10 of Sattler review Progr. NMR 34 (1999), 93

Gradients. Effects of B0 gradients on transverse magnetisation Similar to figure 10 of Sattler review Progr. NMR 34 (1999), 93 Gradients 1. What are gradients? Modern high-resolution NMR probes contain -besides the RF coils - additional coils that can be fed a DC current. The coils are built so that a pulse (~1 ms long) of DC

More information

Gradient Spoiling. Average balanced SSFP magnetization Reduce sensitivity to off-resonance. FFE, FISP, GRASS, GRE, FAST, Field Echo

Gradient Spoiling. Average balanced SSFP magnetization Reduce sensitivity to off-resonance. FFE, FISP, GRASS, GRE, FAST, Field Echo Gradient Spoiling Average balanced SSFP magnetization Reduce sensitivity to off-resonance FFE, FISP, GRASS, GRE, FAST, Field Echo 1 Gradient-Spoiled Sequence (GRE, FFE, FISP, GRASS) RF TR G z G y G x Signal

More information

2 Hardware for Magnetic Resonance Imaging

2 Hardware for Magnetic Resonance Imaging Hardware for Magnetic Resonance Imaging 13 2 Hardware for Magnetic Resonance Imaging Kenneth W. Fishbein, Joseph C. McGowan, and Richard G. Spencer CONTENTS 2.1 Introduction 13 2.2 Magnets 13 2.2.1 Permanent

More information

RF Amplifiers and RF Heating for Medical Applications FTF-IND-F0690

RF Amplifiers and RF Heating for Medical Applications FTF-IND-F0690 June 22, 2010 RF Amplifiers and RF Heating for Medical Applications FTF-IND-F0690 Leonard Pelletier, Jeannette Wilson Application Support, Product Marketing Agenda RF Power for Medical Applications RF

More information

ECSEL Smart Health Project Portfolio and Strategic Agenda. Patrick Vandenberghe Programme Officer

ECSEL Smart Health Project Portfolio and Strategic Agenda. Patrick Vandenberghe Programme Officer ECSEL Smart Health Project Portfolio and Strategic Agenda Patrick Vandenberghe Programme Officer ECSEL-IMI Joint Workshop July 5th 2017 1 Multi-Annual Strategic Plan MASP 2017 Strategic Thrusts Instruments

More information

Nikola Stikov website:

Nikola Stikov   website: Nikola Stikov e-mail: nikola.stikov@mcgill.ca website: http://www.bic.mni.mcgill.ca/personalstikovnikola Local Address: Permanent Address: 3555 Rue Berri Apt 910 bul. J. Sandanski 11-II/22 Montreal, H2L

More information

Cover Page. The handle holds various files of this Leiden University dissertation

Cover Page. The handle   holds various files of this Leiden University dissertation Cover Page The handle http://hdl.handle.net/1887/49562 holds various files of this Leiden University dissertation Author: Schmid, Sophie Title: Arterial spin labeling in space and time : new MRI sequences

More information

NMR Basics. Lecture 2

NMR Basics. Lecture 2 NMR Basics Lecture 2 Continuous wave (CW) vs. FT NMR There are two ways of tuning a piano: - key by key and recording each sound (or frequency). - or, kind of brutal, is to hit with a sledgehammer and

More information

Your first NMR measurement

Your first NMR measurement Your first NMR measurement Introduction Select 10mM water in D2O as NMR sample. The NMR spectrum of such sample consists of only two signals: the water signal and the peak of the reference (TSP). Follow

More information

PHY3902 PHY3904. Nuclear magnetic resonance Laboratory Protocol

PHY3902 PHY3904. Nuclear magnetic resonance Laboratory Protocol PHY3902 PHY3904 Nuclear magnetic resonance Laboratory Protocol PHY3902 PHY3904 Nuclear magnetic resonance Laboratory Protocol GETTING STARTED You might be tempted now to put a sample in the probe and try

More information

Towards new vistas in preamplifier design for MRI

Towards new vistas in preamplifier design for MRI Downloaded from orbit.dtu.dk on: Sep 18, 218 Towards new vistas in preamplifier design for MRI Johansen, Daniel Højrup; Sanchez, Juan Diego; Zhurbenko, Vitaliy; Ardenkjær-Larsen, Jan Henrik Published in:

More information

Mapping of the New IBA Superconducting Synchrocyclotron (S2C2) for Proton Therapy

Mapping of the New IBA Superconducting Synchrocyclotron (S2C2) for Proton Therapy Mapping of the New IBA Superconducting Synchrocyclotron (S2C2) for Proton Therapy J. Van de Walle, W. Kleeven, C. L'Abbate, Y. Paradis, V. Nuttens - IBA M. Conjat, J. Mandrillon, P. Mandrillon - AIMA Developpement

More information

The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging

The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging JOURNAL OF MAGNETIC RESONANCE IMAGING 20:1046 1051 (2004) Technical Note The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging James W. Goldfarb, PhD* Purpose: To describe a known (but undocumented)

More information

EPISTAR MRI: Multislice Mapping of Cerebral Blood Flow

EPISTAR MRI: Multislice Mapping of Cerebral Blood Flow EPISTAR MRI: Multislice Mapping of Cerebral Blood Flow Robert R. Edelman, Qun Chen A method is described for multislice EPISTAR that perfectly compensates magnetization transfer effects. lnflowing arterial

More information

Lab 8 6.S02 Spring 2013 MRI Projection Imaging

Lab 8 6.S02 Spring 2013 MRI Projection Imaging 1. Spin Echos 1.1 Find f0, TX amplitudes, and shim settings In order to acquire spin echos, we first need to find the appropriate scanner settings using the FID GUI. This was all done last week, but these

More information

RAD 229: MRI Signals and Sequences

RAD 229: MRI Signals and Sequences RAD 229: MRI Signals and Sequences Brian Hargreaves All notes are on the course website web.stanford.edu/class/rad229 Course Goals Develop Intuition Understand MRI signals Exposure to numerous MRI sequences

More information

Cook Medical Peripheral Intervention MRI Information

Cook Medical Peripheral Intervention MRI Information Page 1 of 13 Home My Account Help Careers Text size: A A A Home / Clinical / Peripheral Intervention / Products / I Information Browse by Product Family Disease State Alphabetical Cook Medical adheres

More information

Alae Tracker: Tracking of the Nasal Walls in MR-Imaging

Alae Tracker: Tracking of the Nasal Walls in MR-Imaging Alae Tracker: Tracking of the Nasal Walls in MR-Imaging Katharina Breininger 1, Andreas K. Maier 1, Christoph Forman 1, Wilhelm Flatz 2, Catalina Meßmer 3, Maria Schuster 3 1 Pattern Recognition Lab, Friedrich-Alexander-Universität

More information

MR Advance Techniques. Flow Phenomena. Class II

MR Advance Techniques. Flow Phenomena. Class II MR Advance Techniques Flow Phenomena Class II Flow Phenomena In this class we will explore different phenomenona produced from nuclei that move during the acquisition of data. Flowing nuclei exhibit different

More information

EE225E/BIOE265 Spring 2012 Principles of MRI. Assignment 7. Due March 16, 2012

EE225E/BIOE265 Spring 2012 Principles of MRI. Assignment 7. Due March 16, 2012 EE225E/BIOE265 Spring 2012 Principles of MRI Miki Lustig Assignment 7 Due March 16, 2012 1. From Midterm I 2010: You ve just programmed up your first 2DFT pulse sequence, and are trying it out on the scanner.

More information

Introduction, Review of Signals & Systems, Image Quality Metrics

Introduction, Review of Signals & Systems, Image Quality Metrics Introduction, Review of Signals & Systems, Image Quality Metrics Yao Wang Polytechnic University, Brooklyn, NY 11201 Based on Prince and Links, Medical Imaging Signals and Systems and Lecture Notes by

More information

MRI: From Signal to Image

MRI: From Signal to Image MRI: From Signal to Image Johannes Koch physics654 2013-05-06 1 / 27 Tomography Magnetic Resonance Tomography Tomography: tomos: section graphein: to write Signal measured as function of space 2 / 27 Tomography

More information

MR in RTP. MR Data for Treatment Planning: Spatial Accuracy Issues, Protocol Optimization, and Applications (Preview of TG117 Report) Acknowledgements

MR in RTP. MR Data for Treatment Planning: Spatial Accuracy Issues, Protocol Optimization, and Applications (Preview of TG117 Report) Acknowledgements MR Data for Treatment Planning: Issues, Protocol Optimization, and s (Preview of TG117 Report) Debra H. Brinkmann Mayo Clinic, Rochester MN Acknowledgements TG-117 Use of MRI Data in Treatment Planning

More information

Spiral MRI on a 9.4T Vertical-bore Superconducting Magnet Using Unshielded and Self-shielded Gradient Coils

Spiral MRI on a 9.4T Vertical-bore Superconducting Magnet Using Unshielded and Self-shielded Gradient Coils Magn Reson Med Sci doi:10.2463/mrms.tn.2016-0049 Published Online: March 27, 2017 TECHNICAL NOTE Spiral MRI on a 9.4T Vertical-bore Superconducting Magnet Using Unshielded and Self-shielded Gradient Coils

More information

A Complete Digital Magnetic Resonance Imaging (MRI) System at Low Magnetic Field (0.1 Tesla)

A Complete Digital Magnetic Resonance Imaging (MRI) System at Low Magnetic Field (0.1 Tesla) IEEE Instrumentation and Measurement Technology Conference Anchorage, AK, USA, 21-23 May 2002 A Complete igital Magnetic Resonance Imaging (MRI) System at Low Magnetic Field (0.1 Tesla) Kosai RAOOF*, IEEE

More information

Development of a 1.0 T MR microscope using a Nd-Fe-B permanent magnet

Development of a 1.0 T MR microscope using a Nd-Fe-B permanent magnet Magnetic Resonance Imaging 19 (2001) 875 880 Development of a 1.0 T MR microscope using a Nd-Fe-B permanent magnet Tomoyuki Haishi, Takaaki Uematsu, Yoshimasa Matsuda, Katsumi Kose* Institute of Applied

More information

Development of a new Q-meter module

Development of a new Q-meter module A. Berlin,, W. Meyer, G. Reicherz Experimentalphysik I, Ruhr-Universität Bochum E-mail: jonas.herick@rub.de In the research field of polarized target physics the Q-meter is a well established technique

More information

Analysis of spatial dependence of acoustic noise transfer function in magnetic resonance imaging

Analysis of spatial dependence of acoustic noise transfer function in magnetic resonance imaging Analysis of spatial dependence of acoustic noise transfer function in magnetic resonance imaging Award: Magna Cum Laude Poster No.: C-1988 Congress: ECR 2014 Type: Scientific Exhibit Authors: T. Hamaguchi,

More information

COMPUTED TOMOGRAPHY 1

COMPUTED TOMOGRAPHY 1 COMPUTED TOMOGRAPHY 1 Why CT? Conventional X ray picture of a chest 2 Introduction Why CT? In a normal X-ray picture, most soft tissue doesn't show up clearly. To focus in on organs, or to examine the

More information

Pre-amplifiers for a 15-Tesla magnetic resonance imager

Pre-amplifiers for a 15-Tesla magnetic resonance imager From the SelectedWorks of Chin-Leong Lim December, 2013 Pre-amplifiers for a 15-Tesla magnetic resonance imager Chin-Leong Lim Peter Serano, Massachusetts General Hospital Jerome L Ackerman, Massachusetts

More information

Qualitative analysis of Fruits and Vegetables using Earth s Field Nuclear Magnetic Resonance (EFNMR) and Magnetic Resonance Imaging (MRI)

Qualitative analysis of Fruits and Vegetables using Earth s Field Nuclear Magnetic Resonance (EFNMR) and Magnetic Resonance Imaging (MRI) RESEARCH ARTICLE OPEN ACCESS Qualitative analysis of Fruits and Vegetables using Earth s Field Nuclear Magnetic Resonance (EFNMR) and Magnetic Resonance Imaging (MRI) L. P. Deshmukh, K. R. Bagree and S.

More information

Supplementary Figure 1. Scanning Electron Microscopy images of the pristine electrodes. (a) negative electrode and (b) positive electrode.

Supplementary Figure 1. Scanning Electron Microscopy images of the pristine electrodes. (a) negative electrode and (b) positive electrode. a b Supplementary Figure 1. Scanning Electron Microscopy images of the pristine electrodes. (a) negative electrode and (b) positive electrode. Images were performed using a FEI/Philips XL4 microscope with

More information

used to diagnose and treat medical conditions. State the precautions necessary when X ray machines and CT scanners are used.

used to diagnose and treat medical conditions. State the precautions necessary when X ray machines and CT scanners are used. Page 1 State the properties of X rays. Describe how X rays can be used to diagnose and treat medical conditions. State the precautions necessary when X ray machines and CT scanners are used. What is meant

More information

RF Engineering: Live Construction of A Coil Let s build an RF human coil. Hiroyuki Fujita, Ph.D. 1,2,3,4,5 with Tsinghua Zheng, MSEE 1,2

RF Engineering: Live Construction of A Coil Let s build an RF human coil. Hiroyuki Fujita, Ph.D. 1,2,3,4,5 with Tsinghua Zheng, MSEE 1,2 RF Engineering: Live Construction of A Coil Let s build an RF human coil. Hiroyuki Fujita, Ph.D. 1,2,3,4,5 with Tsinghua Zheng, MSEE 1,2 1 Quality Electrodynamics (QED), LLC 2 eqed, LLC 700 Beta Drive,

More information

Pulse Sequence Design Made Easier

Pulse Sequence Design Made Easier Pulse Sequence Design Made Easier Gregory L. Wheeler, BSRT(R)(MR) MRI Consultant gurumri@gmail.com 1 2 Pulse Sequences generally have the following characteristics: An RF line characterizing RF Pulse applications

More information

The Pulsed Resistive Low-Field MR Scanner

The Pulsed Resistive Low-Field MR Scanner 39 Chapter 3 The Pulsed Resistive Low-Field MR Scanner 3.1 Background In the remaining part of this work we are going to describe hyperpolarized gas relaxation, diffusion and MR imaging experiments. These

More information

Pulse Sequences: Rapid Gradient Echo

Pulse Sequences: Rapid Gradient Echo Pulse Sequences: Rapid Gradient Echo M229 Advanced Topics in MRI Holden H. Wu, Ph.D. 2018.04.17 Department of Radiological Sciences David Geffen School of Medicine at UCLA Class Business Office hours -

More information

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 2 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 COMPUTERIZED IMAGING Section I: Chapter 2 RADT

More information

Determining acceptance levels for automatic daily image quality control in magnetic resonance imaging

Determining acceptance levels for automatic daily image quality control in magnetic resonance imaging Determining acceptance levels for automatic daily image quality control in magnetic resonance imaging Poster No.: C-1125 Congress: ECR 2016 Type: Authors: Keywords: DOI: Scientific Exhibit J. I. Peltonen,

More information

BOLD fmri: signal source, data acquisition, and interpretation

BOLD fmri: signal source, data acquisition, and interpretation BOLD fmri: signal source, data acquisition, and interpretation Cheryl Olman 4 th year student, Department of Neuroscience and Center for Magnetic Resonance Research Discussion series Week 1: Biological

More information

PULSED/CW NUCLEAR MAGNETIC RESONANCE

PULSED/CW NUCLEAR MAGNETIC RESONANCE PULSED/CW NUCLEAR MAGNETIC RESONANCE The Second Generation of TeachSpin s Classic Explore NMR for both Hydrogen (at 21 MHz) and Fluorine Nuclei Magnetic Field Stabilized to 1 part in 2 million Homogenize

More information