Improvement of the Image Quality of Black-blood Magnetic Resonance Imaging with the Subtraction Double Inversion Recovery Technique

Size: px
Start display at page:

Download "Improvement of the Image Quality of Black-blood Magnetic Resonance Imaging with the Subtraction Double Inversion Recovery Technique"

Transcription

1 J Radiol Sci 2014; 39: Improvement of the Image Quality of Black-blood Magnetic Resonance Imaging with the Subtraction Double Inversion Recovery Technique Shih-Ming Huang 1,2 Shu-Hui Peng 1,2 Jan-Yang Lin 1,2 Chao-Yu Shen 1,2 Jun-Cheng Weng 2 Yeu-Sheng Tyan 1,2,3 Teng-Fu Tsao 1,2,4 Jhih-Wei Chen 1,2 Department of Medical Imaging 1, Chung Shan Medical University Hospital, Taichung, Taiwan School of Medical Imaging and Radiological Sciences 2, School of Medicine 3,Chung Shan Medical University, Taichung, Taiwan Department of Veterinary Medicine 4, National Chung Hsing University, Taichung, Taiwan Abstract Black-blood images acquired with the double inversion recovery (IR) sequence in cardiovascular magnetic resonance imaging are commonly applied to distinguish the vessel wall from the lumen to evaluate vascular pathology. However, inadequate suppression of blood signals may introduce artifacts and cause mistakes. The objective of the current study was to determine whether subtraction double IR black-blood imaging improved the image quality. 11 scans from four rabbits were scanned using the conventional and subtraction T1 weighted double IR imaging. The subtraction method involved two scans with a third IR sequence with a series of inversion time (TI) settings in one of them. Each black-blood image was qualitatively evaluated based on contrast-to-noise ratio (CNR), signal-tonoise ratio (SNR), and noises. Using the best TI setting (575 ms), subtraction black-blood imaging yielded significantly better black-blood effects with a 48.6% increase in mean CNR between the area of the left carotid artery and adjacent paravertebral muscles (PM) and a 30.7% increase in mean CNR between the area of the left jugular vein and adjacent PM compared with the conventional method (5.65 vs and 6.8 vs. 5.21, both p values < 0.05). The subtraction black-blood imaging did not significantly deteriorate image quality regarding noises and mean SNRs in the area of PM. Subtraction double IR black-blood imaging is better than the conventional method in black-blood effects of blood signals. Cardiovascular magnetic resonance (MR) imaging without the need of contrast media is common in current clinical practice. Black-blood MR images are acquired with sequences designed to null the signal of flowing blood [1]. These images are essential for anatomic assessment of the heart and vascular structures to diagnose cardiovascular disease, such as myocardial disease, intramural hematoma, intraluminal dissection or vascular plaque [2, 3]. Several techniques have been used to null the bright-blood signal in cardiovascular MR imaging applications, such as double inversion recovery (IR) magnetization preparation [1, 3]. Conventionally, double IR sequences are designed specifically to null the signal from flowing blood. The prepulse portion is made of two 180 radiofrequency pulses: a nonselective 180 pulse, which inverts all of the protons in the field, immediately followed by a slice-selective 180 pulse, which reverts the protons in the imaging slice back to their original alignment, leaving all protons outside of the imaging slice inverted. Double IR sequences begin imaging when the magnetization vectors of the flowing blood crosses Correspondence Author to: Jhih-Wei Chen Department of Medical Imaging, Chung Shan Medical University Hospital,Taichung, Taiwan No. 110, Sec. 1, Chien-Kuo N. Road, Taichung 402, Taiwan J Radiol Sci December 2014 Vol.39 No.4 105

2 the null point the inversion time [2, 4]. However, an inadequate nulling of blood signals may introduce artifacts and cause mistakes. In this study, we established an animal model to evaluate a modified protocol of conventional single-shot double IR turbo fast low angle shot gradient echo (turbo- FLASH) T1 weighted imaging using a subtraction method. It involved two scans with a third IR sequence with a series of inversion time settings in one of them. Materials and Methods The animal model Four adult male New Zealand White rabbits, ranging in weight from 3.1 to 4.4 kg, were used in this study. The animal use protocol was reviewed and approved by the Institutional Animal Care and Use Committee (IACUC). Animals were housed in the laboratory animal resource center under standard conditions during the entire study period. Anesthesia was induced and maintained with 5% and 3% isoflurane (Halocarbon Laboratories, River Edge, NJ) gas anesthesia, respectively. MR scans were repeatedly performed on day 0, 7 and 14. The MR scan protocol All MR imaging examinations were performed with a 1.5-Tesla scanner (Magnetom Sonata, Siemens Medical Solutions, Erlangen, Germany) equipped with high-performance 3-axis gradient systems. Animals were placed in the prone position inside the magnet bore at the isocenter and a standard transmit/receive bird-cage human head coil was used. In addition to the triplane gradient echo localizer pulse sequence, the two-dimensional (2D) single-shot double IR turboflash T1 weighted imaging scan protocol included: (1) the conventional black-blood imaging with four excitations (NEX = 4), (2) the conventional blackblood imaging with two excitations (NEX = 2), and (3) 15 sets of two-excited bright-blood imaging owing to an additional non-slice-selective IR with a sequence of TI settings (225 ms, 250 ms, 275 ms, 575 ms). Except for the third IR pulses and number of excitations, all T1 weighted images covered same scan areas in entire animal necks from medullo-cervical junction to aortic arch with identical parameters as follows: TR / TE of 800 / 1.51 ms, flip angle of 8, receiver bandwidth of 470 Hz / Px, slice thickness of 4.0 mm, number of slices = 20, matrix size = 120 x 192, and field of view = 166 x 167 mm. Spectral fat saturation was used to minimize the high signals from perivascular fat. Image post-processing and data analysis The two-excited conventional black-blood images were subtracted from each set of two-excited bright-blood images on a pixel-by-pixel basis using the workstation provided by the MR manufacturer to produce the one-step post-processing bright-blood images. These images were subsequently subtracted from the two-excited conventional black-blood images to obtain the two-step post-processing black-blood images, the so-called two-excited subtraction black-blood images as shown in Fig. 1. These operations generated 15 sets of subtraction images with a different TI value. On the other hand, the scan time was identical between the four-excited conventional black-blood images and the two-excited subtraction black-blood images. The objective evaluation of the MR images was based on the signal intensity (SI) within manually drawn regions of interest (ROIs), which were placed by the radiologist (S.M.H.) who had 3 years of experience in cardiovascular imaging. The images were evaluated in a single section at the level of neck proximal to the bifurcation of left jugular vein. The inspected ROIs included the areas of the left carotid artery (LCA), left jugular vein (LJV), and the paravertebral muscles (PM) of each animal. The air outside the object was also taken. An effort was made to avoid circling the vascular wall when drawing ROIs. As the geometric patterns between the conventional black-blood images and 15 sets of subtraction black-blood images were exactly identical, the ROIs from the image with best anatomical information compared to others in the same scan time could be copied and applied to others. Noise was defined as the standard deviation of the SI within air. Signal-to-noise ratios (SNR) and contrastto-noise ratios (CNR) values were calculated with the following equations: SNR = SI PM /N, CNR LCA = (SI LCA SI PM ) /N and CNR LJV = (SI LJV - SI PM ) /N, where SI PM was the SI in the PM of the animal, SI LCA was the SI in the area of the LCA, SI LJV was the SI in the area of the LJV, and N was the noise. Statistical analysis In the objective evaluation of the two types of blackblood images, data were expressed as mean ± standard deviation. Further analysis was performed with the mean. A paired t test was used to analyze observed differences of noises, SNRs, and CNRs between the four-excited conventional black-blood images and the two-excited subtraction black-blood images. A p value less than 0.05 was considered statistically significant. Data documentation and statistical analyses were performed using Excel (version 2010, Microsoft). Results Twelve MR scans were performed on schedule. One scan showed motion artifacts owing to the inadequate anesthesia and resulted in 11 data subjected to statistical analysis. 106 J Radiol Sci December 2014 Vol.39 No.4

3 Figure 1 The first step: = 1a 1b 1c The second step: = 1b 1c 1d 1e Figure 1. Acquisition of two-excited subtraction black-blood image with post-processing, a. Two-excited bright-blood image with a third TI setting (TI = 575 ms), b. Two-excited conventional black-blood image, c. One-step post-processing bright-blood images, d. Two-excited subtraction black-blood image, e. Diagram of the two-dimensional single-shot double IR T1 black-blood MR sequence with additional non-slice-selective IR with a sequence of TI settings (225 ms, 250 ms, 275 ms, 575 ms). IR = inversion recovery; TI = inversion time. The optimal third TI setting for subtraction black-blood images The images had better quality with higher CNR LCA and CNR LJV as the third TI increased. Among 15 sets of subtraction black-blood imaging, 575 ms of TI had the best CNR LCA (5.65 ± 2.36) and CNR LJV (6.8 ± 2.22) (Fig. 2). As 575 ms was the upper limit of the pulse sequence of the machine, the image characteristics between the conventional black-blood images and subtraction black-blood images with a 575 ms of the third TI were compared (Fig. 3). Image characteristics between two types of black-blood images Using the optimal TI setting (575 ms), two-excited subtraction black-blood images yielded significantly better black-blood effects of blood signals with a 48.6% increase in mean CNR LCA and a 30.7% increase in mean CNR LJV compared with the four-excited conventional black-blood images (5.65 ± 2.36 vs ± 0.8, 6.8 ± 2.22 vs ± 1.33, respectively, both with p< 0.05) (Table 1). The subtraction black-blood imaging did not significantly deteriorate image J Radiol Sci December 2014 Vol.39 No.4 107

4 Figure 2 2a 2b Figure 2. CNRs of subtraction images increased as the third TI increasing. a. CNR of left carotid artery, b. CNR of left jugular vein. CNR = contrast-to-noise ratio Figure 3 3a 3b 3c 3d Figure 3. Comparison of the conventional black-blood images and subtraction black-blood images. a. Four-excited conventional black-blood image, b. Two-excited conventional black-blood image, c. Two-excited bright-blood image with a third TI setting (TI = 575 ms), d. Two-excited subtraction black-blood image. The CNRs of left carotid artery and left internal jugular vein were higher than that on the four-excited conventional image objectively. 108 J Radiol Sci December 2014 Vol.39 No.4

5 Table 1. MR pulse sequences Conventional black-blood images Subtraction black-blood images CNR LCA 3.81 ± ± 2.36 p< 0.05 CNR LJV 5.21 ± ± 2.22 p< 0.05 Noise ± ± 3.81 p = 0.06 SNR 7.65 ± ± 2.21 p = Note: Statistical significances were calculated using the paired student t test quality with more noise or lower mean SNR compared with the conventional method (13.58 ± 3.81 vs ± 1.74, p = 0.06 and 6.91 ± 2.21 vs ± 1.12, p = 0.188). The MR scan time and the post-processing time The MR scan time of the four-excited conventional black-blood imaging and the two-excited subtraction black-blood imaging was identical (80 seconds). The postprocessing time was about 1 minute per case in average. Discussion Black-blood imaging may be achieved using spatial saturation. However, it is susceptible to artifacts from slow flowing spins [5, 6]. Thus, a double IR pulse sequence was developed for blood signal suppression since more than 20 years ago [7]. It is available on most current commercial MR machines and can be combined with almost any scanning methods, such as fast spin echo (FSE), steady-state free precession (SSFP) and turboflash sequences [4, 7, 8]. Although black-blood imaging with the FSE sequence using a multislice acquisition method is feasible [4, 6], relatively long scan time and higher motion sensitivity are the disadvantages [1, 9]. To minimize motion related artifacts due to long scan time, fast breath-hold black-blood imaging with the SSFP or turboflash sequence using a single-shot technique was introduced [7, 9]. According to Lin et al, 2D double IR SSFP imaging has better image quality compared with the 2D double IR FSE sequence under fast heart rate conditions [9]. Bornstedt et al showed that three-dimensional (3D) stacks with the proposed motion sensitized segmented steadystate black-blood gradient echo technique (MSDS) enabled time-efficient coverage of large areas of the vessels without compromising wall-lumen CNR compared to conventional double IR turbo spin and gradient echo images [3]. The major disadvantage of SSFP is very sensitive to B0 inhomogeneities and small frequency shifts leading to substantial reduction in the transverse steady-state magnetization and, thus, to band-shaped signal losses [10]. In our study, black-blood imaging was achieved with a 2D single-shot double IR turboflash sequence, which has shorter acquisition time compared to the FSE and requires less of the hardware to maintain B0 homogeneities compared to the SSFP theoretically. Although double scans (two excitations in each) and post-processing are required for the subtraction black-blood method, the MR acquisition time itself was identical to the conventional black-blood method with four excitations, the default setting for the clinical practice at our institute. The repeated excitations are designed to compensate the poor SNR nature of turbo- FLASH sequences. Conventionally, triple IR pulse sequence means a pair of slice-selective and nonselective 180 inversion pulses to null the blood flow signal followed by a third inversion pulse to null the fat signal [11]. The inversion time of the third IR is around 120 to 180 ms at 1.5-Telsa field strength [12, 13]. In our study, the third IR was to invert the blood signal for acquiring later subtraction images, instead of suppression of signals from fat. After adding the third IR, the flowing blood revealed the high SI and the stationary tissue exhibited the low SI (Fig. 1a), although there was a slightly increased SI of the stationary tissue as the third TI increasing. The time delay between two scans in the subtraction method may increase the risk of motion-related misregistration artifacts as shown at one of our data. This may be the major disadvantage of our subtraction double IR black-blood imaging. To circumvent this, rearrangement of the two scans into one condensed sequence is planned for future studies. This study has limitations that should be addressed. Firstly, the subtraction double IR black-blood imaging was only compared with conventional one using turboflash sequences. The FSE and SSFP-based sequences were not addressed. Secondly, only 2D sequences were tested in this study. The efficiency of the subtraction method in 3D black-blood imaging is not yet clear. Thirdly, only objective evaluations based on SI measurements were performed. The subjective impacts of the imaging technique on the readers are not clear. Finally, the protocol was only tested on a limited number of animals. In conclusion, the subtraction turboflash sequence J Radiol Sci December 2014 Vol.39 No.4 109

6 is an alternative 2D double IR black-blood imaging. Compared to the conventional method, the new protocol showed better blood signal black-blood effects and CNR without compromising image quality in terms of noises and SNR under the same image acquisition time. Such improvements in imaging quality may promote the diagnostic capacity of cardiovascular magnetic resonance imaging. Acknowledgement This work was supported, in part, by a research grant (CSH-96-32) from the Chung Shan Medical University Hospital, Taichung, Taiwan. Reference 1. Edelman RR, Chien D, Kim D. Fast selective black blood MR imaging. Radiology 1991; 181: Vivian S. Lee. Cardiovascular MRI: Physical principles to practical protocols. Philadelphia, PA: Lippincott Williams & Wilkins, 2006; Bornstedt A, Burgmaier M, Hombach V, et al. Dual stack black blood carotid artery CMR at 3T: Application to wall thickness visualization. J Cardiov Magn Reson 2009; 11: Song HK, Wright AC, Wolf RL, et al. Multislice double inversion pulse sequence for efficient black-blood MRI. Magn Reson Med 2002; 47: Steinman DA, Rutt BK. On the nature and reduction of plaque-mimicking flow artifacts in black blood MRI of the carotid bifurcation. Magn Reson Med 1998; 39: Kuribayashi H, Tessier JJ, Checkley DR, et al. Effective blood signal suppression using double inversionrecovery and slice reordering for multislice fast spinecho MRI and its application in simultaneous proton density and T2 weighted imaging. J Magn Reson Imaging 2004; 20: Braendli M, Bongartz G. Combining two single-shot imaging techniques with slice-selective and non sliceselective inversion recovery pulses: New strategy for native MR angiography based on the long T1 relaxation time and inflow properties of blood. AJR Am J Roentgenol 2003; 180: Vlaardingerbroek MT, den Boer JA. Magnetic resonance imaging. New York: Springer 1999; Lin K, Bi XM, Liu Y, et al. Black-blood steady-state free precession (SSFP) coronary wall MRI for cardiac allografts: A feasibility study. J Magn Reson Imaging 2012; 35: Deshpande VS, Shea SM, Li D. Artifact reduction in true-fisp imaging of the coronary arteries by adjusting imaging frequency. Magn Reson Med 2003; 49: Payne AR, Casey M, McClure J, et al. Bright-blood T2-weighted MRI has higher diagnostic accuracy than dark-blood short tau inversion recovery MRI for detection of acute myocardial infarction and for assessment of the ischemic area at risk and myocardial salvage. Circ Cardiovasc Imaging 2011; 4: Francone M, Carbone I, Agati L, et al. Utility of T2-weighted short-tau inversion recovery (STIR) sequences in cardiac MRI: an overview of clinical applications in ischaemic and non-ischaemic heart disease. Radiol Med 2011; 116: Oh-Ici D, Ridgway JP, Kuehne T, et al. Cardiovascular magnetic resonance of myocardial edema using a short inversion time inversion recovery (STIR) black-blood technique: diagnostic accuracy of visual and semiquantitative assessment. J Cardiov Magn Reson 2012; 28: J Radiol Sci December 2014 Vol.39 No.4

Pulse Sequence Design and Image Procedures

Pulse Sequence Design and Image Procedures Pulse Sequence Design and Image Procedures 1 Gregory L. Wheeler, BSRT(R)(MR) MRI Consultant 2 A pulse sequence is a timing diagram designed with a series of RF pulses, gradients switching, and signal readout

More information

Cardiac MR. Dr John Ridgway. Leeds Teaching Hospitals NHS Trust, UK

Cardiac MR. Dr John Ridgway. Leeds Teaching Hospitals NHS Trust, UK Cardiac MR Dr John Ridgway Leeds Teaching Hospitals NHS Trust, UK Cardiac MR Physics for clinicians: Part I Journal of Cardiovascular Magnetic Resonance 2010, 12:71 http://jcmr-online.com/content/12/1/71

More information

Pulse Sequence Design Made Easier

Pulse Sequence Design Made Easier Pulse Sequence Design Made Easier Gregory L. Wheeler, BSRT(R)(MR) MRI Consultant gurumri@gmail.com 1 2 Pulse Sequences generally have the following characteristics: An RF line characterizing RF Pulse applications

More information

MRI Metal Artifact Reduction

MRI Metal Artifact Reduction MRI Metal Artifact Reduction PD Dr. med. Reto Sutter University Hospital Balgrist Zurich University of Zurich OUTLINE Is this Patient suitable for MR Imaging? Metal artifact reduction Is this Patient suitable

More information

H 2 O and fat imaging

H 2 O and fat imaging H 2 O and fat imaging Xu Feng Outline Introduction benefit from the separation of water and fat imaging Chemical Shift definition of chemical shift origin of chemical shift equations of chemical shift

More information

HETERONUCLEAR IMAGING. Topics to be Discussed:

HETERONUCLEAR IMAGING. Topics to be Discussed: HETERONUCLEAR IMAGING BioE-594 Advanced MRI By:- Rajitha Mullapudi 04/06/2006 Topics to be Discussed: What is heteronuclear imaging. Comparing the hardware of MRI and heteronuclear imaging. Clinical applications

More information

Gradient Spoiling. Average balanced SSFP magnetization Reduce sensitivity to off-resonance. FFE, FISP, GRASS, GRE, FAST, Field Echo

Gradient Spoiling. Average balanced SSFP magnetization Reduce sensitivity to off-resonance. FFE, FISP, GRASS, GRE, FAST, Field Echo Gradient Spoiling Average balanced SSFP magnetization Reduce sensitivity to off-resonance FFE, FISP, GRASS, GRE, FAST, Field Echo 1 Gradient-Spoiled Sequence (GRE, FFE, FISP, GRASS) RF TR G z G y G x Signal

More information

MR Advance Techniques. Flow Phenomena. Class II

MR Advance Techniques. Flow Phenomena. Class II MR Advance Techniques Flow Phenomena Class II Flow Phenomena In this class we will explore different phenomenona produced from nuclei that move during the acquisition of data. Flowing nuclei exhibit different

More information

Background (~EE369B)

Background (~EE369B) Background (~EE369B) Magnetic Resonance Imaging D. Nishimura Overview of NMR Hardware Image formation and k-space Excitation k-space Signals and contrast Signal-to-Noise Ratio (SNR) Pulse Sequences 13

More information

Principles of MRI EE225E / BIO265. Lecture 21. Instructor: Miki Lustig UC Berkeley, EECS. M. Lustig, EECS UC Berkeley

Principles of MRI EE225E / BIO265. Lecture 21. Instructor: Miki Lustig UC Berkeley, EECS. M. Lustig, EECS UC Berkeley Principles of MRI Lecture 21 EE225E / BIO265 Instructor: Miki Lustig UC Berkeley, EECS Question What is the difference between the images? Answer Both T1-weighted spin-echo gradient-echo Lower SNR Meniscus

More information

1 Introduction. 2 The basic principles of NMR

1 Introduction. 2 The basic principles of NMR 1 Introduction Since 1977 when the first clinical MRI scanner was patented nuclear magnetic resonance imaging is increasingly being used for medical diagnosis and in scientific research and application

More information

MR Basics: Module 8 Image Quality

MR Basics: Module 8 Image Quality Module 8 Transcript For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions.

More information

Simultaneous Multi-Slice (Slice Accelerated) Diffusion EPI

Simultaneous Multi-Slice (Slice Accelerated) Diffusion EPI Simultaneous Multi-Slice (Slice Accelerated) Diffusion EPI Val M. Runge, MD Institute for Diagnostic and Interventional Radiology Clinics for Neuroradiology and Nuclear Medicine University Hospital Zurich

More information

Advanced MSK MRI Protocols at 3.0T. Garry E. Gold, M.D. Associate Professor Department of Radiology Stanford University

Advanced MSK MRI Protocols at 3.0T. Garry E. Gold, M.D. Associate Professor Department of Radiology Stanford University Advanced MSK MRI Protocols at 3.0T Garry E. Gold, M.D. Associate Professor Department of Radiology Stanford University Outline Why High Field for MSK? SNR and Relaxation Times Technical Issues Example

More information

MR in RTP. MR Data for Treatment Planning: Spatial Accuracy Issues, Protocol Optimization, and Applications (Preview of TG117 Report) Acknowledgements

MR in RTP. MR Data for Treatment Planning: Spatial Accuracy Issues, Protocol Optimization, and Applications (Preview of TG117 Report) Acknowledgements MR Data for Treatment Planning: Issues, Protocol Optimization, and s (Preview of TG117 Report) Debra H. Brinkmann Mayo Clinic, Rochester MN Acknowledgements TG-117 Use of MRI Data in Treatment Planning

More information

MRI Summer Course Lab 2: Gradient Echo T1 & T2* Curves

MRI Summer Course Lab 2: Gradient Echo T1 & T2* Curves MRI Summer Course Lab 2: Gradient Echo T1 & T2* Curves Experiment 1 Goal: Examine the effect caused by changing flip angle on image contrast in a simple gradient echo sequence and derive T1-curves. Image

More information

Downloaded from by on 02/07/18 from IP address Copyright ARRS. For personal use only; all rights reserved

Downloaded from  by on 02/07/18 from IP address Copyright ARRS. For personal use only; all rights reserved Downloaded from www.ajronline.org by 46.3.192.5 on 02/07/18 from IP address 46.3.192.5. Copyright RRS. For personal use only; all rights reserved C oil sensitivity encoding (SENSE) is a new technique that

More information

High Field MRI: Technology, Applications, Safety, and Limitations

High Field MRI: Technology, Applications, Safety, and Limitations High Field MRI: Technology, Applications, Safety, and Limitations R. Jason Stafford, Ph.D. The University of Texas M. D. Anderson Cancer Center, Houston, TX Introduction The amount of available signal

More information

(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010

(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010 (N)MR Imaging Lab Course Script FMP PhD Autumn School Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder Date: November 3rd, 2010 1 Purpose: Understanding the basic principles of MR imaging

More information

MR in Tx Planning. Acknowledgements. Outline. Overview MR in RTP

MR in Tx Planning. Acknowledgements. Outline. Overview MR in RTP MR Data for Treatment Planning and Stereotactic Procedures: Sources of Distortion, Protocol Optimization, and Assessment (Preview of TG117 Report) Debra H. Brinkmann Mayo Clinic, Rochester MN Acknowledgements

More information

MR Basics: Module 6 Pulse Sequences

MR Basics: Module 6 Pulse Sequences Module 6 Transcript For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions.

More information

Module 2. Artefacts and Imaging Optimisation for single shot methods. Content: Introduction. Phase error. Phase bandwidth. Chemical shift review

Module 2. Artefacts and Imaging Optimisation for single shot methods. Content: Introduction. Phase error. Phase bandwidth. Chemical shift review MRES 7005 - Fast Imaging Techniques Module 2 Artefacts and Imaging Optimisation for single shot methods Content: Introduction Phase error Phase bandwidth Chemical shift review Chemical shift in pixels

More information

MRI Anatomy and Positioning Series Module 12: Fat Suppression Techniques

MRI Anatomy and Positioning Series Module 12: Fat Suppression Techniques MRI Anatomy and Positioning Series Module 12: Fat Suppression Techniques 1 Introduction... 3 RF FatSat... 4 HOAST... 4 FatSat... 5 Segment FS... 8 PhaseCycle... 9 Water Excitation... 10 STIR... 12 FatSep...

More information

IR/SR TrueFISP. Works-in-Progress package Version 1.2. For the SIEMENS Magnetom. Installation and User s Guide NUMARIS/4VA21B.

IR/SR TrueFISP. Works-in-Progress package Version 1.2. For the SIEMENS Magnetom. Installation and User s Guide NUMARIS/4VA21B. Works-in-Progress package Version 1.2 For the Installation and User s Guide NUMARIS/4VA21B January 22, 2003 Section of Medical Physics, University Hospital Freiburg, Germany Contact: Klaus Scheffler PhD

More information

Pulse Sequences: Rapid Gradient Echo

Pulse Sequences: Rapid Gradient Echo Pulse Sequences: Rapid Gradient Echo M229 Advanced Topics in MRI Holden H. Wu, Ph.D. 2018.04.17 Department of Radiological Sciences David Geffen School of Medicine at UCLA Class Business Office hours -

More information

Noninvasive Blood Flow Mapping with Arterial Spin Labeling (ASL) Paul Kyu Han and Sung-Hong Park

Noninvasive Blood Flow Mapping with Arterial Spin Labeling (ASL) Paul Kyu Han and Sung-Hong Park Noninvasive Blood Flow Mapping with Arterial Spin Labeling (ASL) Paul Kyu Han and Sung-Hong Park Department of Bio and Brain Engineering, Korea Advanced Institute of Science and Technology (KAIST), Daejeon,

More information

Image Quality/Artifacts Frequency (MHz)

Image Quality/Artifacts Frequency (MHz) The Larmor Relation 84 Image Quality/Artifacts (MHz) 42 ω = γ X B = 2πf 84 0.0 1.0 2.0 Magnetic Field (Tesla) 1 A 1D Image Magnetic Field Gradients Magnet Field Strength Field Strength / Gradient Coil

More information

Magnetic Resonance Imaging Principles, Methods, and Techniques

Magnetic Resonance Imaging Principles, Methods, and Techniques Magnetic Resonance Imaging Principles, Methods, and Techniques Perry Sprawls Jr., Emory University Publisher: Medical Physics Publishing Corporation Publication Place: Madison, Wisconsin Publication Date:

More information

2014 M.S. Cohen all rights reserved

2014 M.S. Cohen all rights reserved 2014 M.S. Cohen all rights reserved mscohen@g.ucla.edu IMAGE QUALITY / ARTIFACTS SYRINGOMYELIA Source http://gait.aidi.udel.edu/res695/homepage/pd_ortho/educate/clincase/syrsco.htm Surgery is usually recommended

More information

Magnetic Resonance Imaging

Magnetic Resonance Imaging Magnetic Resonance Imaging Principles, Methods, and Techniques Perry Sprawls, Ph.D., FACR, FAAPM, FIOMP Distinguished Emeritus Professor Department of Radiology Emory University Atlanta, Georgia Medical

More information

Works-in-Progress package Version 1.0. For the SIEMENS Magnetom. Installation and User s Guide NUMARIS/4VA21B. January 22, 2003

Works-in-Progress package Version 1.0. For the SIEMENS Magnetom. Installation and User s Guide NUMARIS/4VA21B. January 22, 2003 Works-in-Progress package Version 1.0 For the Installation and User s Guide NUMARIS/4VA21B January 22, 2003 Section of Medical Physics, University Hospital Freiburg, Germany Contact: Klaus Scheffler PhD,

More information

Experience in implementing continuous arterial spin labeling on a commercial MR scanner

Experience in implementing continuous arterial spin labeling on a commercial MR scanner JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 6, NUMBER 1, WINTER 2005 Experience in implementing continuous arterial spin labeling on a commercial MR scanner Theodore R. Steger and Edward F. Jackson

More information

3T Unlimited. ipat on MAGNETOM Allegra The Importance of ipat at 3T. medical

3T Unlimited. ipat on MAGNETOM Allegra The Importance of ipat at 3T. medical 3T Unlimited ipat on MAGNETOM Allegra The Importance of ipat at 3T s medical ipat on MAGNETOM Allegra The Importance of ipat at 3T The rise of 3T MR imaging Ultra High Field MR (3T) has flourished during

More information

NIH Public Access Author Manuscript Int J Cardiovasc Imaging. Author manuscript; available in PMC 2008 May 26.

NIH Public Access Author Manuscript Int J Cardiovasc Imaging. Author manuscript; available in PMC 2008 May 26. NIH Public Access Author Manuscript Published in final edited form as: Int J Cardiovasc Imaging. 2001 August ; 17(4): 287 296. A comparison of prospective and retrospective respiratory navigator gating

More information

Numerical Evaluation of an 8-element Phased Array Torso Coil for Magnetic Resonance Imaging

Numerical Evaluation of an 8-element Phased Array Torso Coil for Magnetic Resonance Imaging Numerical Evaluation of an 8-element Phased Array Torso Coil for Magnetic Resonance Imaging Feng Liu, Joe Li, Ian Gregg, Nick Shuley and Stuart Crozier School of Information Technology and Electrical Engineering,

More information

The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging

The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging JOURNAL OF MAGNETIC RESONANCE IMAGING 20:1046 1051 (2004) Technical Note The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging James W. Goldfarb, PhD* Purpose: To describe a known (but undocumented)

More information

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia M R I Physics Course Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia M R I Physics Course Magnetic Resonance Imaging Spatial

More information

MARP. MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005

MARP. MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005 ACR MRI accreditation program MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005 Carl R. Keener, Ph.D., DABMP, DABR keener@marpinc.com MARP Medical & Radiation Physics,

More information

functional MRI: A primer

functional MRI: A primer Activation Leads to: functional MRI: A primer CBF Increased +ΔR CBV Increased +ΔR (C+) O Utilization Increased slightly? Venous [O ] Increased -ΔR* Glucose Utilization Increased? Lactate BOLD R=/T R=/T

More information

Application Guide & Release Notes

Application Guide & Release Notes Application Guide & Release Notes Inner-volume-imaging (IVI) EPI C2P Release 002a 1 September 2015 TMII Translational and Molecular Imaging Institute Conditions of Use This package is provided to support

More information

Iterative Reconstruction in Image Space. Answers for life.

Iterative Reconstruction in Image Space. Answers for life. Iterative Reconstruction in Image Space Answers for life. Iterative Reconstruction in Image Space * (IRIS) * Please note: IRIS is used as an abbreviation for Iterative Reconstruction in Image Space throughout

More information

BOLD fmri: signal source, data acquisition, and interpretation

BOLD fmri: signal source, data acquisition, and interpretation BOLD fmri: signal source, data acquisition, and interpretation Cheryl Olman 4 th year student, Department of Neuroscience and Center for Magnetic Resonance Research Discussion series Week 1: Biological

More information

Inherent Insensitivity to RF Inhomogeneity in FLASH Imaging

Inherent Insensitivity to RF Inhomogeneity in FLASH Imaging Inherent Insensitivity to RF Inhomogeneity in FLASH Imaging Danli Wang, Keith Heberlein, Stephen LaConte, and Xiaoping Hu* Magnetic Resonance in Medicine 52:927 931 (2004) Radiofrequency (RF) field inhomogeneity

More information

Half-Pulse Excitation Pulse Design and the Artifact Evaluation

Half-Pulse Excitation Pulse Design and the Artifact Evaluation Half-Pulse Excitation Pulse Design and the Artifact Evaluation Phillip Cho. INRODUCION A conventional excitation scheme consists of a slice-selective RF excitation followed by a gradient-refocusing interval

More information

MAGNETIC RESONANCE IMAGING

MAGNETIC RESONANCE IMAGING CSEE 4620 Homework 3 Fall 2018 MAGNETIC RESONANCE IMAGING 1. THE PRIMARY MAGNET Magnetic resonance imaging requires a very strong static magnetic field to align the nuclei. Modern MRI scanners require

More information

Automatic Selection of Mask and Arterial Phase Images for Temporally-Resolved MR Digital Subtraction Angiography

Automatic Selection of Mask and Arterial Phase Images for Temporally-Resolved MR Digital Subtraction Angiography Automatic Selection of Mask and Arterial Phase Images for Temporally-Resolved MR Digital Subtraction Angiography 21 May 2002, ISMRM 2002 Junhwan Kim, Martin R. Prince, Ramin Zabih,, Jeff Bezanson, Richard

More information

Enhancing Gray-to-White Matter Contrast in 3T T1 Spin-Echo Brain Scans by Optimizing Flip Angle

Enhancing Gray-to-White Matter Contrast in 3T T1 Spin-Echo Brain Scans by Optimizing Flip Angle AJNR Am J Neuroradiol 26:2000 2004, September 2005 Enhancing Gray-to-White Matter Contrast in 3T T1 Spin-Echo Brain Scans by Optimizing Flip Angle Bernd L. Schmitz, Georg Grön, Florian Brausewetter, Martin

More information

ISSN X CODEN (USA): PCHHAX. The role of dual spin echo in increasing resolution in diffusion weighted imaging of brain

ISSN X CODEN (USA): PCHHAX. The role of dual spin echo in increasing resolution in diffusion weighted imaging of brain Available online at www.derpharmachemica.com ISSN 0975-413X CODEN (USA): PCHHAX Der Pharma Chemica, 2016, 8(17):15-20 (http://derpharmachemica.com/archive.html) The role of in increasing resolution in

More information

Improve Image Quality of Transversal Relaxation Time PROPELLER and FLAIR on Magnetic Resonance Imaging

Improve Image Quality of Transversal Relaxation Time PROPELLER and FLAIR on Magnetic Resonance Imaging Journal of Physics: Conference Series PAPER OPEN ACCESS Improve Image Quality of Transversal Relaxation Time PROPELLER and FLAIR on Magnetic Resonance Imaging To cite this article: N Rauf et al 2018 J.

More information

NIH Public Access Author Manuscript Magn Reson Med. Author manuscript; available in PMC 2010 July 21.

NIH Public Access Author Manuscript Magn Reson Med. Author manuscript; available in PMC 2010 July 21. NIH Public Access Author Manuscript Published in final edited form as: Magn Reson Med. 2010 April ; 63(4): 1092 1097. doi:10.1002/mrm.22223. Spatially Varying Fat-Water Excitation Using Short 2DRF Pulses

More information

TimTX TrueShape. The parallel transmit architecture of the future. Answers for life.

TimTX TrueShape.  The parallel transmit architecture of the future. Answers for life. www.siemens.com/trueshape TimTX TrueShape The parallel transmit architecture of the future. The product/feature (mentioned herein) is not commercially available. Due to regulatory reasons its future availability

More information

Hardware. MRI System. MRI system Multicoil Microstrip. Part1

Hardware. MRI System. MRI system Multicoil Microstrip. Part1 Hardware MRI system Multicoil Microstrip MRI System Part1 1 The MRI system is made up of a variety of subsystems. the Operator Workspace Gradient Driver subsystem The Physiological Acquisition Controller

More information

SIGNA Explorer Lift revives our MR

SIGNA Explorer Lift revives our MR Seiji Shiotani, MD, PhD Seirei Fuji Hospital in Fuji City, Shizuoka, Japan Masayoshi Sugimura Seirei Fuji Hospital in Fuji City, Shizuoka, Japan SIGN Explorer Lift revives our MR The clinical usefulness

More information

10. Phase Cycling and Pulsed Field Gradients Introduction to Phase Cycling - Quadrature images

10. Phase Cycling and Pulsed Field Gradients Introduction to Phase Cycling - Quadrature images 10. Phase Cycling and Pulsed Field Gradients 10.1 Introduction to Phase Cycling - Quadrature images The selection of coherence transfer pathways (CTP) by phase cycling or PFGs is the tool that allows the

More information

Delivering Better Patient Care with SIGNA Architect

Delivering Better Patient Care with SIGNA Architect Delivering etter Patient Care with SIGNA Architect As a regional leader in outpatient-based diagnostic imaging, the radiologists and staff at Inova Fairfax MRI Center are focused on one thing: the patient.

More information

The development of the RF-pulse for the low level SAR used by the MRI.

The development of the RF-pulse for the low level SAR used by the MRI. The development of the RF-pulse for the low level SAR used by the MRI. Kojiro Yamaguchi a*, Eizo Umezawa a, Sachiko Ueoku b, Kazuhiro Katada c a Faculty of radiological technology, School of Health Science,

More information

Page 1 of 9. Protocol: adult_other_adni3_study_human_ge_3t_25w_ _ _1. 3 Plane Localizer. 3 Plane Localizer PATIENT POSITION

Page 1 of 9. Protocol: adult_other_adni3_study_human_ge_3t_25w_ _ _1. 3 Plane Localizer. 3 Plane Localizer PATIENT POSITION 3 Localizer FOV 26.0 Slice Thickness 5.0 Slice Spacing 0.0 Freq 256 Phase 128 3-PLANE 3 Localizer Unswap Phase Correction Gradient Echo Imaging Options Seq, Fast Recon All Images Contrast Yes/ 3 Localizer

More information

Alae Tracker: Tracking of the Nasal Walls in MR-Imaging

Alae Tracker: Tracking of the Nasal Walls in MR-Imaging Alae Tracker: Tracking of the Nasal Walls in MR-Imaging Katharina Breininger 1, Andreas K. Maier 1, Christoph Forman 1, Wilhelm Flatz 2, Catalina Meßmer 3, Maria Schuster 3 1 Pattern Recognition Lab, Friedrich-Alexander-Universität

More information

1.5T HIGH FIELD SMALL ANIMAL MRI

1.5T HIGH FIELD SMALL ANIMAL MRI 1.5T HIGH FIELD SMALL ANIMAL MRI Designed Specifically for Veterinarians TECHNICAL GUIDE ADVANCING THE ART AND SCIENCE OF VETERINARY MRI The PetVet is the only high-field MRI system designed specifically

More information

MRI SYSTEM COMPONENTS Module One

MRI SYSTEM COMPONENTS Module One MRI SYSTEM COMPONENTS Module One 1 MAIN COMPONENTS Magnet Gradient Coils RF Coils Host Computer / Electronic Support System Operator Console and Display Systems 2 3 4 5 Magnet Components 6 The magnet The

More information

Clear delineation of optic radiation and very small vessels using phase difference enhanced imaging (PADRE)

Clear delineation of optic radiation and very small vessels using phase difference enhanced imaging (PADRE) Clear delineation of optic radiation and very small vessels using phase difference enhanced imaging (PADRE) Poster No.: C-2459 Congress: ECR 2010 Type: Scientific Exhibit Topic: Neuro Authors: T. Yoneda,

More information

MRI at a Glance. Catherine Westbrook. Blackwell Science

MRI at a Glance. Catherine Westbrook. Blackwell Science MRI at a Glance Catherine Westbrook Blackwell Science MRI at a Glance MRI at a Glance CATHERINE WESTBROOK MSC DCRR CTC Director of Training and Education Lodestone Patient Care Ltd Blackwell Science 2002

More information

Slice profile optimization in arterial spin labeling using presaturation and optimized RF pulses

Slice profile optimization in arterial spin labeling using presaturation and optimized RF pulses Magnetic Resonance Imaging 24 (2006) 1229 1240 Slice profile optimization in arterial spin labeling using presaturation and optimized RF pulses David Alberg Holm a,b, 4, Karam Sidaros a a Danish Research

More information

The promise of high-field MRI. High Field MRI Technology, Applications, Safety, and Limitations. High-field Scanners

The promise of high-field MRI. High Field MRI Technology, Applications, Safety, and Limitations. High-field Scanners High Field MRI Technology, Applications, Safety, and Limitations R. Jason Stafford, Ph.D. Department of Imaging Physics The University of Texas M. D. Anderson Cancer Center Houston, TX The promise of high-field

More information

PET/CT Instrumentation Basics

PET/CT Instrumentation Basics / Instrumentation Basics 1. Motivations for / imaging 2. What is a / Scanner 3. Typical Protocols 4. Attenuation Correction 5. Problems and Challenges with / 6. Examples Motivations for / Imaging Desire

More information

EPISTAR MRI: Multislice Mapping of Cerebral Blood Flow

EPISTAR MRI: Multislice Mapping of Cerebral Blood Flow EPISTAR MRI: Multislice Mapping of Cerebral Blood Flow Robert R. Edelman, Qun Chen A method is described for multislice EPISTAR that perfectly compensates magnetization transfer effects. lnflowing arterial

More information

MRI imaging in neuroscience Dr. Thom Oostendorp Lab class: 2 hrs

MRI imaging in neuroscience Dr. Thom Oostendorp Lab class: 2 hrs MRI imaging in neuroscience Dr. Thom Oostendorp Lab class: 2 hrs 1 Introduction In tomographic imaging techniques, such as MRI, a certain tissue property within a slice is imaged. For each voxel (volume

More information

Doppler in Obstetrics: book by K Nicolaides, G Rizzo, K Hecher. Chapter on Doppler ultrasound: principles and practice by Colin Deane

Doppler in Obstetrics: book by K Nicolaides, G Rizzo, K Hecher. Chapter on Doppler ultrasound: principles and practice by Colin Deane Doppler in Obstetrics: book by K Nicolaides, G Rizzo, K Hecher Chapter on Doppler ultrasound: principles and practice by Colin Deane INTRODUCTION Competent use of Doppler ultrasound techniques requires

More information

Passive Tracking Exploiting Local Signal Conservation: The White Marker Phenomenon

Passive Tracking Exploiting Local Signal Conservation: The White Marker Phenomenon Passive Tracking Exploiting Local Signal Conservation: The White Marker Phenomenon Jan-Henry Seppenwoolde,* Max A. Viergever, and Chris J.G. Bakker Magnetic Resonance in Medicine 50:784 790 (2003) This

More information

Applications Guide. Spectral Editing with SVS. (Works-in-Progress) MAGNETOM TaTs and Verio Systems (3T)

Applications Guide. Spectral Editing with SVS. (Works-in-Progress) MAGNETOM TaTs and Verio Systems (3T) Applications Guide Spectral Editing with SVS (Works-in-Progress) MAGNETOM TaTs and Verio Systems (3T) syngo MR Numaris 4 VB17A June 2009 Version 1.1 WIP #529 Important Note This document provides a description

More information

Multi-channel SQUID-based Ultra-Low Field Magnetic Resonance Imaging in Unshielded Environment

Multi-channel SQUID-based Ultra-Low Field Magnetic Resonance Imaging in Unshielded Environment Multi-channel SQUID-based Ultra-Low Field Magnetic Resonance Imaging in Unshielded Environment Andrei Matlashov, Per Magnelind, Shaun Newman, Henrik Sandin, Algis Urbaitis, Petr Volegov, Michelle Espy

More information

RAD 229: MRI Signals and Sequences

RAD 229: MRI Signals and Sequences RAD 229: MRI Signals and Sequences Brian Hargreaves All notes are on the course website web.stanford.edu/class/rad229 Course Goals Develop Intuition Understand MRI signals Exposure to numerous MRI sequences

More information

The feasibility of breath-hold high-resolution 3D-MRCP obtained with 32 channel torso cardiac coil and T2-prepBTFE

The feasibility of breath-hold high-resolution 3D-MRCP obtained with 32 channel torso cardiac coil and T2-prepBTFE The feasibility of breath-hold high-resolution 3D-MRCP obtained with 32 channel torso cardiac coil and T2-prepBTFE Poster No.: C-0022 Congress: ECR 2010 Type: Scientific Exhibit Topic: Abdominal Viscera

More information

Low-Cost MR-Compatible Moving Heart Phantom

Low-Cost MR-Compatible Moving Heart Phantom Journal of Cardiovascular Magnetic Resonance", 2(3), 18 1-187 (2000) Low-Cost MR-Compatible Moving Heart Phantom Michael E. Huber,'t4 Matthias Stuber,',* Rene M. Botnar,1'2 Peter B~esiger,~ and Warren

More information

ACRIN 6686 / RTOG 0825

ACRIN 6686 / RTOG 0825 ACRIN 6686 (RTOG 0825) Advanced MRI Imaging Manual ACRIN 6686 / RTOG 0825 A phase III double blind placebo controlled trial of conventional chemoradiation and adjuvant temozolomide plus bevacizumab vs

More information

Standards for Imaging Endpoints in Clinical Trials: Standardization and Optimization of Image Acquisitions: Magnetic Resonance

Standards for Imaging Endpoints in Clinical Trials: Standardization and Optimization of Image Acquisitions: Magnetic Resonance FDA Workshop April 13, 2010 Standards for Imaging Endpoints in Clinical Trials: Standardization and Optimization of Image Acquisitions: Magnetic Resonance Edward F. Jackson, PhD Professor and Chief, Section

More information

NIH Public Access Author Manuscript Magn Reson Med. Author manuscript; available in PMC 2009 December 1.

NIH Public Access Author Manuscript Magn Reson Med. Author manuscript; available in PMC 2009 December 1. NIH Public Access Author Manuscript Published in final edited form as: Magn Reson Med. 2008 December ; 60(6): 1488 1497. doi:10.1002/mrm.21790. Continuous Flow Driven Inversion for Arterial Spin Labeling

More information

Siemens AG, Healthcare Sector. syngo MR D Operator Manual - Breast 0.0.

Siemens AG, Healthcare Sector. syngo MR D Operator Manual - Breast 0.0. Siemens AG, Healthcare Sector Cs2 syngo Breast Operator 2010-2012 MR-05019 630 02 English 06/2012 n.a. Informatik, Manual D13 Cape syngo MR D13 Operator Manual - Breast syngo MR D13 www.siemens.com/healthcare

More information

Digital Imaging CT & MR

Digital Imaging CT & MR Digital Imaging CT & MR January 22, 2008 Digital Radiography, CT and MRI generate images in a digital format What is a Digital Image? A digital image is made up of picture elements, pixels row by column

More information

Photomultiplier Tube

Photomultiplier Tube Nuclear Medicine Uses a device known as a Gamma Camera. Also known as a Scintillation or Anger Camera. Detects the release of gamma rays from Radionuclide. The radionuclide can be injected, inhaled or

More information

INSTRUCTIONS FOR THE CAROTID MRI COMPLETION FORM, CMR, Version B, 12/08/2005

INSTRUCTIONS FOR THE CAROTID MRI COMPLETION FORM, CMR, Version B, 12/08/2005 INSTRUCTIONS FOR THE CAROTID MRI COMPLETION FORM, CMR, Version B, 12/08/2005 I. General Instructions: This form is used to document the conduct and completeness of the MRI examination. The MRI examination

More information

High-Field Surface-Coil MR Imaging of Localized Anatomy

High-Field Surface-Coil MR Imaging of Localized Anatomy 181 High-Field Surface-Coil MR Imaging of Localized Anatomy John F. Schenck,' Thomas H. Foster,' John l. Henkes,' William J. Adams,' Cecil Hayes,2 Howard R. Hart, Jr.,' William A. Edelstein,' Paul A. Bottomley,'

More information

Liver imaging beyond expectations with Ingenia

Liver imaging beyond expectations with Ingenia Publication for the Philips MRI Community Issue 47 2012/3 Liver imaging beyond expectations with Ingenia Contributed by John Penatzer, RT, MR clinical product specialist, Cleveland, OH, USA Publication

More information

Steady-state sequences: Spoiled and balanced methods

Steady-state sequences: Spoiled and balanced methods Steady-state sequences: Spoiled and balanced methods Karla L Miller, FMRIB Centre, University of Oxford What is steady-state imaging? In the context of MRI pulse sequences, the term steady state typically

More information

MRI Systems and Coil Technology

MRI Systems and Coil Technology MRI for Technologists MRI Systems and Coil Technology PROGRAM INFORMATION MRI for Technologists is a training program designed to meet the needs of radiologic technologists entering or working in the field

More information

Spiral MRI on a 9.4T Vertical-bore Superconducting Magnet Using Unshielded and Self-shielded Gradient Coils

Spiral MRI on a 9.4T Vertical-bore Superconducting Magnet Using Unshielded and Self-shielded Gradient Coils Magn Reson Med Sci doi:10.2463/mrms.tn.2016-0049 Published Online: March 27, 2017 TECHNICAL NOTE Spiral MRI on a 9.4T Vertical-bore Superconducting Magnet Using Unshielded and Self-shielded Gradient Coils

More information

Fast Field-Cycling Magnetic Resonance Imaging (FFC-MRI)

Fast Field-Cycling Magnetic Resonance Imaging (FFC-MRI) Fast Field-Cycling Magnetic Resonance Imaging (FFC-MRI) David J. Lurie Aberdeen Biomedical Imaging Centre University of Aberdeen Summary of talk Short introduction to MRI Physics Field-Cycling MRI Field-Cycling

More information

2015 Spin echoes and projection imaging

2015 Spin echoes and projection imaging 1. Spin Echoes 1.1 Find f0, transmit amplitudes, and shim settings In order to acquire spin echoes, we first need to find the appropriate scanner settings using the FID GUI. This was all done last week,

More information

Compact yet Sophisticated

Compact yet Sophisticated Compact yet Sophisticated Hitachi has brought Open MRI one step further in its evolution, to better assist medical professionals who work at the forefront of healthcare. AIRIS Light MSK offers radiologists

More information

A. SPECIFIC AIMS: phase graph (EPG) algorithms to cover a wide range of MRI

A. SPECIFIC AIMS: phase graph (EPG) algorithms to cover a wide range of MRI A. SPECIFIC AIMS: A.. Overview: The promise of improved MRI results at high field strength is compromised by the difficulties encountered at high field, including: i) Non-uniform excitation, due to the

More information

Encoding of inductively measured k-space trajectories in MR raw data

Encoding of inductively measured k-space trajectories in MR raw data Downloaded from orbit.dtu.dk on: Apr 10, 2018 Encoding of inductively measured k-space trajectories in MR raw data Pedersen, Jan Ole; Hanson, Christian G.; Xue, Rong; Hanson, Lars G. Publication date:

More information

Evaluation of a Subject specific dual-transmit approach for improving B 1 field homogeneity in cardiovascular magnetic resonance at 3T

Evaluation of a Subject specific dual-transmit approach for improving B 1 field homogeneity in cardiovascular magnetic resonance at 3T Krishnamurthy et al. Journal of Cardiovascular Magnetic Resonance 2013, 15:68 RESEARCH Open Access Evaluation of a Subject specific dual-transmit approach for improving B 1 field homogeneity in cardiovascular

More information

The Enlightened Choice for High-field MRI

The Enlightened Choice for High-field MRI The Enlightened Choice for High-field MRI ECHELON heralds the dawn of a new standard for 1.5T superconductive MRI. The ECHELON features a small footprint with economics that do not compromise diagnostic

More information

Software and Hardware in CCTA. Elly Castellano PhD

Software and Hardware in CCTA. Elly Castellano PhD Software and Hardware in CCTA Elly Castellano PhD Outline technical requirements for coronary CTA the modern cardiac CT scanner ECG-gating technology image reconstruction algorithms 2 Technical requirements

More information

Analysis of spatial dependence of acoustic noise transfer function in magnetic resonance imaging

Analysis of spatial dependence of acoustic noise transfer function in magnetic resonance imaging Analysis of spatial dependence of acoustic noise transfer function in magnetic resonance imaging Award: Magna Cum Laude Poster No.: C-1988 Congress: ECR 2014 Type: Scientific Exhibit Authors: T. Hamaguchi,

More information

Medical Imaging (EL582/BE620/GA4426)

Medical Imaging (EL582/BE620/GA4426) Medical Imaging (EL582/BE620/GA4426) Jonathan Mamou, PhD Riverside Research Lizzi Center for Biomedical Engineering New York, NY jmamou@riversideresearch.org On behalf of Prof. Daniel Turnbull Outline

More information

BACKGROUND. ** 78% of all MRI scanners have Image Quality problems. *** *** 25% of all Multi-Channel RF coils have at least one bad channel.

BACKGROUND. ** 78% of all MRI scanners have Image Quality problems. *** *** 25% of all Multi-Channel RF coils have at least one bad channel. Range of Results from over 534 ACR-mandated Annual MRI Performance Evaluations on over 204 Magnets from 8 Vendors Spanning a 10-year Period Moriel NessAiver, Ph.D. - Simply Physics - Baltimore, MD moriel@simplyphysics.com

More information

Difference in signal-to-noise-ratio (SNR) from vertical to horizontal scanner position using a 0,25 Tesla Weightbearing

Difference in signal-to-noise-ratio (SNR) from vertical to horizontal scanner position using a 0,25 Tesla Weightbearing Difference in signal-to-noise-ratio (SNR) from vertical to horizontal scanner position using a 0,25 Tesla Weightbearing scanner Poster No.: C-0672 Congress: ECR 2014 Type: Scientific Exhibit Authors: F.

More information

Research Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D.

Research Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D. CT Clinical Innovation Center Department of Radiology Mayo Clinic College of Medicine Rochester, MN Research Support National

More information

Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity)

Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity) Vascular Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO Medical Systems Division, Shimadzu Corporation Yoshiaki Miura 1. Introduction In recent years, digital cardiovascular

More information