Quality Control for X-Ray Systems A Tool Chain for NDT Applications
|
|
- Lorena McDaniel
- 6 years ago
- Views:
Transcription
1 4th International Symposium on NDT in Aerospace Tu.3.A.2 Quality Control for X-Ray Systems A Tool Chain for NDT Applications Stefan REISINGER *, Alexander ENNEN *, Thorsten WÖRLEIN *, Michael SCHMITT *, Virginia VOLAND * * Fraunhofer IIS/EZRT, Development Centre X-ray Technology (Dr.-Mack-Straße 81, Fürth, Germany / virginia.voland@iis.fraunhofer.de) Abstract. Within the last years, X-ray systems have been established in the field of NDT. Mainly caused by the following aspects the confidence in this technology is still inferior to the confidence in competitive NDT-techniques: Long-term degradation of the system status caused by aging process and abrasion of the system components Fluctuations of image quality during the measurement caused by instability of the X-ray source and possible breakdown of detector elements Poor image quality affected by defect detector elements or other detector characteristics, by a changing imaging geometry during the measurement and by geometric misalignments Insufficient repeatability and measurement accuracy In order to cope with these challenges, we implemented procedures to monitor the X-ray system. The monitoring of the system status is divided into several phases: Prior to use, a computed tomography (CT) system and all its components have to be adjusted to each other in order to prevent any kind of misalignments such as detector or axis tilt. The consequences of unadjusted CT Systems manifest in artifacts or blurred images and therefore result in lower image quality and worse analysis results. The stability of the system components should be controlled regularly during the system lifetime, e.g. weekly. Influencing factors, which are used to forecast the expected image quality and end of life of the X-ray components, are factors such as the spatial resolution, number and distribution of defect detector elements, the contrast sensitivity or size and stability of the focal spot. In preparation the measurement process should be simulated to find suitable parameters for the following measurement. Moreover, for each measurement the image acquisition must be monitored with regard to the image quality to be expected. We therefore defined a quantitative set of characteristics such as e.g. a constant illumination of the detector and present a way to monitor the data acquisition with respect to these characteristics. After the image acquisition is done, a wide range of algorithms can be applied if necessary in order to correct image artifacts and to further improve image quality before the reconstruction task and the final analysis steps occur. Algorithms for the automated adjustment, characterization and monitoring of an X- ray system and correction methods for the further improvement of image quality are presented. License: 1
2 1. Introduction Figure 1.1: Tool chain for NDT applications To ensure the quality of NDT-techniques different procedures have been established. For example the standard for tactile systems is to use calibration phantoms to ensure a stabile system status and resulting measurement inaccuracy. Though within the application area of industrial X-ray systems guidelines for the determination of measurement inaccuracy [7] exist, still no standard is committed in terms of a consistency check as in the medical application. In order to cope with these challenge, we found procedures to monitor the X-ray system. To ensure the stability of an X-ray system different steps are necessary (Figure 1.1). Once while bringing a system into service or after replacement of system components, the single components (X-ray detector, X-ray tube and manipulation system; Figure 1.2) must be aligned with respect to each other. Continuously (daily or weekly) the stability of the system components should be controlled. Besides parasitic drags on system quality has to be logged and corrected (during the measurement or in a post step). Therefore the aim is to develop a tool to monitor the system to allow conclusions over the system status. As a result an expert system can be established to determine suitable parameters and necessary corrections dependent on system status and monitoring of each measurement. Figure 1.2: Drawing of an X-ray system. 2. Tool chain for NDT applications This chapter describes the single steps within the tool chain for NDT applications. 2.1 System Adjustment Uniquely the system components must be aligned with respect to each other. Otherwise influences like tilts of detector and rotatory axis, imprecisely distances between X-ray source and X-ray detector or that the focal spot and the detector centre axis are not arranged 2
3 in the same plain lead to unsatisfactory image quality. Later corrections on the data cannot adjust all these missing alignments even if the maladjustment is completely known. 2.2 Monitoring of system status The stability of the system components should be controlled regularly during the system lifetime, e.g. weekly. Influencing factors, which are used to forecast the expected image quality and end of life of the X-ray components, are factors such as the spatial resolution, number and distribution of defect detector elements, the contrast sensitivity or size and stability of the focal spot. Especially with regard to performance of certification a logging of system status is essential. 2.3 Monitoring of image acquisition During each measurement task some characteristics are monitored to control the image acquisition. If a parameter (homogenous illumination of detector, mean attenuation, movement of optical focal spot) gets out of specified ranges steps can be defined to act on this behaviour. For example repeating single exposures, waiting for a specified time or even restart or abort of the whole measurement can be reasonable. 2.4 Just in time correction Some influences to image quality can be corrected synchronous during the measurement. These corrections are for example determining the position of the center of rotation and determining the position of the optical focal spot [4] or reduction of beam hardening effects [8]. Thus a online reconstruction with a syncronous correction can take place. 2.5 Post processing/ Correction of raw images After the image acquisition is done, a wide range of algorithms can be applied if necessary in order to reduce image artefacts [1] and to further improve image quality before the reconstruction task and final analysis steps occur [2]. These analysis steps are for example examination of dual energy measurements [5]. 3. Quality control for X-ray systems To simplify the use the tool chain of image processing all single steps have to be linked with each other. The challenge of the X-ray tool chain is monitoring of the time variable status of the X-ray system depending on its life time. The most important issue for a typical user of an X-ray system is the long-term management of best parameters (voltage, power, number of projections, integration time, etc.) for each measuring task. To use the full capacity of a system by means of determination and using of resolution limit and best quality of generated images is dependent on the selection of suitable parameters. Degradation of system status (such as basic spatial resolution (Figure 3.1) and detector efficiency (Figure 3.2)) or the state of single components brings up the demand to adapt the necessary parameters. Moreover these parameters are dependent on the measurement task and given requirements. 3
4 Figure 3.1: Determination of basic spatial resolution according to EN and ASTM Complete failures of single components come along with stealthy degradations. Thus the status of the X-ray system and the single components is logged over the time. With a certain empirical values it is possible to predict coming degradations and complete failures. This allows early replacement of single components especially because sourcing of new components is very time expensive due to long time of delivery. Hence down times of X- ray systems caused by breakdown of single components can be reduced. Figure 3.2: Determination of detector efficiency according to ASTM E [6]. Another part of quality control for X-ray systems is a just in time correction of influences like misalignments (Figure 3.3). Especially in the case of extensive measuring tasks or if the time for the reconstruction of the dataset which take some hours or days like described in [3] it is important to be able to react on parasitic drags instantly. In a extrem case these influences are not until noticed if the whole measurement and reconstrution task is finished and are irreparable. This would effect high costs and loss of time. Figure 3.4 shows a plot of logged values. It is observable that for a single projection no signal was triggered. Therefore a simple reiteration of this measurement fragment would lead to more usable results. 4
5 Figure 3.3: Observation of drift of focal spot. It is obvious, that the movement of the focal spot changes its direction during the long-term observation. Figure 3.4: Observation of influencing values during the measurement. At the marked instant of time all monitored values show anomalous values. This is an indication to insufficient image quality (In this case the tube was dropped out). If a synchronous correction is not possible at least a logging of system behaviour during the measurement to predict necessary corrections on the raw images or the reconstructed volume (Figure 3.5) is recommended. Figure 3.5: left: Ring artefacts with / without correction. Right: Detector shift with / without correction. 5
6 As a follow up, conclusions can be drawn from the final image quality to necessary parameters of coming up measurement tasks. Conclusion The main benefit of these tools is that the whole tool chain of image processing is available in one software solution. Therefore the possibility of data exchange between the single processing steps is given. The coming steps for this project are the generation of empirical values of the degradation of system components and involved impact to image quality. Having this data it is possible to specify necessary adaption of measurement parameters und give a forecast of system live time. References [1] S. Reisinger, V. Voland, S. Burtzlaff, M. Schmitt und N. Uhlmann, Rohbildkorrekturen für die Artefaktreduktion in der Computertomographie, Jahrestagung der DGZfP BB 127-CD, [2] V. Voland, J.-I. Tiemegni, T. Wörlein, N. Uhlmann und F. Kargl, Steuerung von Röntgenexperimenten, Jahrestagung der DGZfP BB 127-CD, [3] V. Voland, J. Freitag und N. Uhlmann, A CT System for the Analysis of Prehistoric Ice Cores, Elst (Hrsg): Microelectronic Systems: Circuits, Systems and Applications, Berlin, Springer, [4] V. Voland, M. Schmitt und S. Reisinger, Geometric Adjustment Methods to Improve Reconstruction Quality on Rotational Cone-beam Systems, in ICT Conference, Wels, [5] M. Firsching, F. Nachtrab, J. Mühlbauer und N. Uhlmann, Detection of Enclosed Diamonds using Dual Energy X-ray imaging, 18th World Conference on Nondestructive Testing, [6] ASTM E e1 Standard Practice for Manufacturing Characterization of Digital Detector Arrays, ASTM International, West Conshohocken, PA, 2003, DOI: /C A, [7] VDI/VDE Bestimmung der Messunsicherheit und der Prüfprozesseignung von Koordinatenmessgeräten mit CT-Sensoren. [8] M. Franz, S. Kasperl und M. Stamminger, Synchronous Arftefact Reduction in Industrial Computed Tomography, Technisches Messen TM 77, pp ,
v tome x m microfocus CT
GE Inspection Technologies v tome x m microfocus CT Uniting premium 3D metrology and inspection with quality and speed. gemeasurement.com/ct x plore precision CT line Inspect with precision, power, and
More informationPhilip Sperling. Sales Science and New Materials, YXLON International GmbH, Essener Bogen 15, Hamburg, Germany.
A new generation of x-ray computed tomography devices for quality inspection and metrology inspection in the field of additive manufacturing and other sciences Philip Sperling Sales Science and New Materials,
More informationInfluence of surface roughness on measurement uncertainty in Computed Tomography
Influence of surface roughness on measurement uncertainty in Computed Tomography More info about this article: http://www.ndt.net/?id=21910 Leonard Schild, Alexandra Kraemer, Doreen Reiling, Hanjue Wu,
More informationUncertainty in CT Metrology: Visualizations for Exploration and Analysis of Geometric Tolerances
Uncertainty in CT Metrology: Visualizations for Exploration and Analysis of Geometric Tolerances Artem Amirkhanov 1, Bernhard Fröhler 1, Michael Reiter 1, Johann Kastner 1, M. Eduard Grӧller 2, Christoph
More informationMigrating from traditional to Digital Radiography in Aerospace
Migrating from traditional to Digital Radiography in Aerospace More info about this article: http://www.ndt.net/?id=22663 Abstract Lennart Schulenburg VisiConsult X-ray System & Solutions GmbH Brandenbrooker
More informationPERFORMANCE CHARACTERIZATION OF AMORPHOUS SILICON DIGITAL DETECTOR ARRAYS FOR GAMMA RADIOGRAPHY
12 th A-PCNDT 2006 Asia-Pacific Conference on NDT, 5 th 10 th Nov 2006, Auckland, New Zealand PERFORMANCE CHARACTERIZATION OF AMORPHOUS SILICON DIGITAL DETECTOR ARRAYS FOR GAMMA RADIOGRAPHY Rajashekar
More informationAmorphous Selenium Direct Radiography for Industrial Imaging
DGZfP Proceedings BB 67-CD Paper 22 Computerized Tomography for Industrial Applications and Image Processing in Radiology March 15-17, 1999, Berlin, Germany Amorphous Selenium Direct Radiography for Industrial
More informationMaking use of phase-contrast in dimensional µct-measurements of weakly absorbing objects
Making use of phase-contrast in dimensional µct-measurements of weakly absorbing objects Andreas Staude 1, Karsten Ehrig 1, Bernhard Illerhaus 1, Martin Spreemann 2 1 BAM Bundesanstalt für Materialforschung
More informationTomographic 3D-Radiometry for the Visualisation and Measurement of the Defects of Girth Seams
ECNDT 2006 - We.3.2.3 Tomographic 3D-Radiometry for the Visualisation and Measurement of the Defects of Girth Seams Bernhard REDMER, Uwe EWERT Federal Institute of Materials Research and Testing (BAM),
More informationAlignment of the camera
Related topics Detector Alignment, Rotation axis, tilt, Principle Alignment of the detector and the rotation stage is very important to get optimal quality images of a CT scan. In this experiment, the
More informationAUTOMATED AND QUANTITATIVE METHOD FOR QUALITY ASSURANCE OF DIGITAL RADIOGRAPHY IMAGING SYSTEMS
International Workshop SMART MATERIALS, STRUCTURES & NDT in AEROSPACE Conference NDT in Canada 2011 2-4 November 2011, Montreal, Quebec, Canada AUTOMATED AND QUANTITATIVE METHOD FOR QUALITY ASSURANCE OF
More informationMammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over
Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over a lifetime Breast cancer screening programs rely on
More informationHIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE
HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE Takeyuki Hashimoto 1), Morio Onoe 2), Hiroshi Nakamura 3), Tamon Inouye 4), Hiromichi Jumonji 5), Iwao Takahashi 6); 1)Yokohama Soei
More informationHISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD
CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging Shawn D. Teague, MD DISCLOSURES 3DR- advisory committee CT PHYSICS WITH AN EMPHASIS ON APPLICATION IN THORACIC AND CARDIAC IMAGING
More informationOn Determination of Focal Laws for Linear Phased Array Probes as to the Active and Passive Element Size
19 th World Conference on Non-Destructive Testing 2016 On Determination of Focal Laws for Linear Phased Array Probes as to the Active and Passive Element Size Andreas GOMMLICH 1, Frank SCHUBERT 2 1 Institute
More informationX-RAY COMPUTED TOMOGRAPHY
X-RAY COMPUTED TOMOGRAPHY Bc. Jan Kratochvíla Czech Technical University in Prague Faculty of Nuclear Sciences and Physical Engineering Abstract Computed tomography is a powerful tool for imaging the inner
More informationDetector technology in simultaneous spectral imaging
Computed tomography Detector technology in simultaneous spectral imaging Philips IQon Spectral CT Z. Romman, I. Uman, Y. Yagil, D. Finzi, N. Wainer, D. Milstein; Philips Healthcare While CT has become
More informationFully Automated Inspection of Welded Seams with a New High Resolution Durable X-ray Camera System
ECNDT 2006 - Tu.2.4.4 Fully Automated Inspection of Welded Seams with a New High Resolution Durable X-ray Camera System Thomas WENZEL, Rolf BEHRENDT, Theobald FUCHS, Ulrich HÜTTEN, Günther KOSTKA, Peter
More informationRadionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT)
Radionuclide Imaging MII 3073 Single Photon Emission Computed Tomography (SPECT) Single Photon Emission Computed Tomography (SPECT) The successful application of computer algorithms to x-ray imaging in
More informationSECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS
RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 2 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 COMPUTERIZED IMAGING Section I: Chapter 2 RADT
More informationCOMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR
This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura New Technology https://www.youtube.com/watch?v=ptkzznazb 7U COMPUTED
More informationInstruction manual for T3DS software. Tool for THz Time-Domain Spectroscopy. Release 4.0
Instruction manual for T3DS software Release 4.0 Table of contents 0. Setup... 3 1. Start-up... 5 2. Input parameters and delay line control... 6 3. Slow scan measurement... 8 4. Fast scan measurement...
More informationHiding In Plain Sight. How Ultrasonics Can Help You Find the Smallest Bonded Wafer and Device Defects. A Sonix White Paper
Hiding In Plain Sight How Ultrasonics Can Help You Find the Smallest Bonded Wafer and Device Defects A Sonix White Paper If You Can See It, You Can Solve It: Understanding Ultrasonic Inspection of Bonded
More informationOperating Procedures for MICROCT1 Nikon XTH 225 ST
Operating Procedures for MICROCT1 Nikon XTH 225 ST Ensuring System is Ready (go through to ensure all windows and tasks below have been completed either by you or someone else prior to mounting and scanning
More informationDigital Radiographic Inspection replacing traditional RT and 3D RT Development
Digital Radiographic Inspection replacing traditional RT and 3D RT Development Iploca Novel Construction Meeting 27&28 March 2014 Geneva By Jan van der Ent Technical Authority International Contents Introduction
More informationQC Testing for Computed Tomography (CT) Scanner
QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including
More informationDiscover the difference in efficiency
Y.CT Compact Fan-beam computed tomography (CT) inspection system for high-density medium and large-sized parts Discover the difference in efficiency Technology with Passion Explore the art of detection
More informationMC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE
MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE A. Miceli ab, R. Thierry a, A. Flisch a, U. Sennhauser a, F. Casali b a Empa - Swiss Federal Laboratories for
More informationRobert Pagnanelli BSRT(R)(N), CNMT, NCT, FASNC Chief Technologist, Nuclear Imaging Duke University Medical Center. Thursday September 8, 2011
Robert Pagnanelli BSRT(R)(N), CNMT, NCT, FASNC Chief Technologist, Nuclear Imaging Duke University Medical Center Thursday September 8, 2011 Quality Control Quality control should be performed because:
More information2/14/2019. Nuclear Medicine Artifacts. Symmetric energy windows
Nuclear Medicine Artifacts SCPMG Medical Imaging Technology & Informatics Medical Physics Group Brian Helbig, MS, DABR 1 2 Symmetric energy windows 3 1 Dynamic clinical study Energy peak shift Electrical
More informationA positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 4, FALL 2009 A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup Huaiqun Guan,
More informationMore Info at Open Access Database by S. Dutta and T. Schmidt
More Info at Open Access Database www.ndt.net/?id=17657 New concept for higher Robot position accuracy during thermography measurement to be implemented with the existing prototype automated thermography
More informationDigital Radiography : Flat Panel
Digital Radiography : Flat Panel Flat panels performances & operation How does it work? - what is a sensor? - ideal sensor Flat panels limits and solutions - offset calibration - gain calibration - non
More informationTDI Imaging: An Efficient AOI and AXI Tool
TDI Imaging: An Efficient AOI and AXI Tool Yakov Bulayev Hamamatsu Corporation Bridgewater, New Jersey Abstract As a result of heightened requirements for quality, integrity and reliability of electronic
More informationInvestigation of the film Kodak MX 125 in Agfa NDT S eco processing
Investigation of the film Kodak MX 125 in Agfa NDT S eco processing Test Report-No.: BAM VIII.3 / 6568b Date: 26. 10. 2005 Name and address Kodak GmbH, of the customer: Hedelfinger Straße 60, 70327 Stuttgart,
More informationComputed Radiography
BAM Berlin Computed Radiography --INDE 2007, Kalpakkam, India -- Uwe Zscherpel, Uwe Ewert BAM Berlin, Division VIII.3 Requests Requests and and information information to: to: Dr. Dr. U. U. Zscherpel Zscherpel
More informationOPTIMIZING SPATIAL RESOLUTION WITH THE MECHANICAL DESIGN OF AN X-RAY COMPU1ED TOMOGRAPHY SCANNER
OPTIMIZING SPATIAL RESOLUTION WITH THE MECHANICAL DESIGN OF AN X-RAY COMPU1ED TOMOGRAPHY SCANNER Lowell D. Harris, RichardT. Bernardi, Simon H. C. Hughes, and Robert E. Slocum Bio-Imaging Research, Inc.
More informationDefining the Optimal Beam Hardening Correction Parameters for CT Dimensional Metrology Applications
International Conference on Competitive Manufacturing Defining the Optimal Beam Hardening Correction Parameters for CT Dimensional Metrology Applications Y. Tan 1,2, K. Kiekens 1,2, F. Welkenhuyzen 2,
More informationRadiographic Testing (RT) [10]
Radiographic Testing (RT) [10] Definition: An NDT method that utilizes x-rays or gamma radiation to detect discontinuities in materials, and to present their images on recording medium. 1> Electromagnetic
More informationTowards accurate measurements with synchrotron tomography Problems and pitfalls. Robert C. Atwood. Nghia T. Vo, Michael Drakopoulos, Thomas Connolley
Towards accurate measurements with synchrotron tomography Problems and pitfalls Robert C. Atwood Nghia T. Vo, Michael Drakopoulos, Thomas Connolley Artefacts in Synchrotron X-ray Tomography Rings Rings
More informationThermionic x-ray. Alternative technologies. Electron Field Emission. CNT Based Field Emission X-Ray Source
Energy Level (ev) Multi-beam x-ray source array based on carbon nanotube field emission O. Zhou, JP Lu, X. Calderon-Colon, X. Qian, G. Yang, G. Cao, E. Gidcumb, A. Tucker, J. Shan University of North Carolina
More informationDigital Radiology with Photon Counting Detectors
11th European Conference on Non-Destructive Testing (ECNDT 2014), October 6-10, 2014, Prague, Czech Republic Digital Radiology with Photon Counting Detectors More Info at Open Access Database www.ndt.net/?id=16709
More informationReal Time Linear Array Imaging. Brian Caccamise
Real Time Linear Array Imaging Brian Caccamise 1 Real Time Linear Array Imaging What is Real Time Linear Array Imaging? Or Real Time Radiography (RTR)? 2 Real Time Linear Array Imaging It s Not This! Shoe
More informationThe following article is a translation of parts of the original publication of Karl-Ludwig Bath in the german astronomical magazine:
The following article is a translation of parts of the original publication of Karl-Ludwig Bath in the german astronomical magazine: Sterne und Weltraum 1973/6, p.177-180. The publication of this translation
More informationIQI-Sensitivity and Applications of Flat Panel Detectors and X-Ray Image Intensifiers A Comparison
IQI-Sensitivity and Applications of Flat Panel Detectors and X-Ray Image Intensifiers A Comparison Dr. Matthias Purschke/ Ulf Reimer, Agfa NDT Pantak Seifert GmbH und Co. KG, Bogenstr. 4, 96 Ahrensburg,
More informationWaveMaster IOL. Fast and accurate intraocular lens tester
WaveMaster IOL Fast and accurate intraocular lens tester INTRAOCULAR LENS TESTER WaveMaster IOL Fast and accurate intraocular lens tester WaveMaster IOL is a new instrument providing real time analysis
More informationMicrowave Testing (µt): An Overview. Johann Hinken, FI Test- und Messtechnik GmbH Magdeburg, Germany, July 2016
Microwave Testing (µt): An Overview Johann Hinken, FI Test- und Messtechnik GmbH Magdeburg, Germany, July 2016 More info about this article: http://www.ndt.net/?id=21377 Content: 1. Introduction 2. Physical
More informationISO INTERNATIONAL STANDARD. Non-destructive testing of welds Radiographic testing Part 1: X- and gamma-ray techniques with film
INTERNATIONAL STANDARD ISO 17636-1 First edition 2013-01-15 Non-destructive testing of welds Radiographic testing Part 1: X- and gamma-ray techniques with film Contrôle non destructif des assemblages soudés
More informationPhotomultiplier Tube
Nuclear Medicine Uses a device known as a Gamma Camera. Also known as a Scintillation or Anger Camera. Detects the release of gamma rays from Radionuclide. The radionuclide can be injected, inhaled or
More informationA COMPARISON BETWEEN ASTM E588 AND SEP 1927 RELATING RESOLUTION LIMITS AT DETERMINATION OF THE PURITY GRADE
19 th World Conference on Non-Destructive Testing 2016 A COMPARISON BETWEEN ASTM E588 AND SEP 1927 RELATING RESOLUTION LIMITS AT DETERMINATION OF THE PURITY GRADE Daniel KOTSCHATE 1, Dirk GOHLKE 1, Rainer
More informationOperation and Sound Field of an Ultrasonic Biplane-Array
19 th World Conference on Non-Destructive Testing 2016 Operation and Sound Field of an Ultrasonic Biplane-Array R. HIPP 1, A. GOMMLICH 2, D. JONEIT 1, F. SCHUBERT 1, H. HEUER 1 1 Fraunhofer-Institut für
More informationAvailable online at ScienceDirect. Procedia CIRP 11 (2013 )
Available online at www.sciencedirect.com ScienceDirect Procedia CIRP 11 (213 ) 234 238 2 nd International Through-life Engineering Services Conference Applicability of industrial robots for machining
More informationScatter Correction by Modulation of Primary Radiation in Industrial X-ray CT: Beam-hardening Effects and their Correction
International Symposium on Digital Industrial Radiology and Computed Tomography - Mo.3.2 Scatter Correction by Modulation of Primary Radiation in Industrial X-ray CT: Beam-hardening Effects and their Correction
More informationLaser Technology Inc.
Field of View Determination for Shearography Nondestructive Evaluation John W. Newman, President Laser Technology Inc. Norristown, PA 19403 USA Introduction Shearography NDT is a highly developed, mature
More informationWaveMaster IOL. Fast and Accurate Intraocular Lens Tester
WaveMaster IOL Fast and Accurate Intraocular Lens Tester INTRAOCULAR LENS TESTER WaveMaster IOL Fast and accurate intraocular lens tester WaveMaster IOL is an instrument providing real time analysis of
More informationSEAMS DUE TO MULTIPLE OUTPUT CCDS
Seam Correction for Sensors with Multiple Outputs Introduction Image sensor manufacturers are continually working to meet their customers demands for ever-higher frame rates in their cameras. To meet this
More informationCOMPUTED TOMOGRAPHY 1
COMPUTED TOMOGRAPHY 1 Why CT? Conventional X ray picture of a chest 2 Introduction Why CT? In a normal X-ray picture, most soft tissue doesn't show up clearly. To focus in on organs, or to examine the
More informationEquivalent Penetrameter Sensitivity (EPS) for Performance Evaluation of Computed Radiography Systems Muzibur Khan * and Mike Brothers
Equivalent Penetrameter Sensitivity (EPS) for Performance Evaluation of Computed Radiography Systems Muzibur Khan * and Mike Brothers More info about this article: http://www.ndt.net/?id=22782 Abstract
More informationInvestigation of the line-pair pattern method for evaluating mammographic focal spot performance
Investigation of the line-pair pattern method for evaluating mammographic focal spot performance Mitchell M. Goodsitt, a) Heang-Ping Chan, and Bob Liu Department of Radiology, University of Michigan, Ann
More informationEvaluation of no-grid radiography using the digital scattered x-ray removal processing
Evaluation of no-grid radiography using the digital scattered x-ray removal processing Poster No.: C-0416 Congress: ECR 2016 Type: Authors: Scientific Exhibit R. Suzuki 1, T. Goto 1, H. Ogawa 2, N. Amimoto
More information2D Digital Radiography 3D Computed Tomography Inspection Services Laboratories. xrayinspectionservice.com
2D Digital Radiography 3D Computed Tomography Inspection Services Laboratories 4nsi.com xrayinspectionservice.com Celebrating over 28 years of Innovation North Star Imaging was founded in 1986 by Ken Ness
More informationStability of a Fiber-Fed Heterodyne Interferometer
Stability of a Fiber-Fed Heterodyne Interferometer Christoph Weichert, Jens Flügge, Paul Köchert, Rainer Köning, Physikalisch Technische Bundesanstalt, Braunschweig, Germany; Rainer Tutsch, Technische
More informationDISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter
DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter Revision Date: January 5th, 2017 www.disc-imaging.com Table of Contents Section A: Preliminary Setup Requirements... 4 Tools
More informationIntroduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image
Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology
More informationAn Activity in Computed Tomography
Pre-lab Discussion An Activity in Computed Tomography X-rays X-rays are high energy electromagnetic radiation with wavelengths smaller than those in the visible spectrum (0.01-10nm and 4000-800nm respectively).
More informationNovel Imaging Techniques for Defects Characterisation in Phased Array Inspection
Novel Imaging Techniques for Defects Characterisation in Phased Array Inspection P. Rioux 1, F. Lachance 1 and J. Turcotte 1 1 Sonatest, Québec, Canada Phone: +1 418 683 6222, e-mail: sales@sonatest.com
More informationmake it easy, with Ray
Lower dose - Quick scan times - Pulsed X-ray technology - Multiple scan modes 3 Dedicated detectors - Reliable performance - No damage - Long life span Easy upgrade - Ready to upgrade CBCT & Cephalometric
More informationProjection Data and Scanograms in X-ray Computed Tomography
DICTA2002: Digital Image Computing Techniques and Applications, 21 22 January 2002, Melbourne, Australia 1 Projection Data and Scanograms in X-ray Computed Tomography Russell Horney Imants Svalbe Peter
More informationStandard Practice for Qualification of Radioscopic Systems 1
Designation: 95 An American National Standard Standard Practice for Qualification of Radioscopic Systems 1 This standard is issued under the fixed designation ; the number immediately following the designation
More informationMaximum Performance, Minimum Space
TECHNOLOGY HISTORY For over 130 years, Toshiba has been a world leader in developing technology to improve the quality of life. Our 50,000 global patents demonstrate a long, rich history of leading innovation.
More informationNondestructive and Contactless Materials Characterization with the Help of Microwave Sensors
ECNDT 2006 - Tu.1.8.3 Nondestructive and Contactless Materials Characterization with the Help of Microwave Sensors Christoph SKLARCZYK, Roman PINCHUK, Valeri MELEV, Fraunhofer-Institut Zerstörungsfreie
More informationComputed Tomography. The Fundamentals of... THE FUNDAMENTALS OF... Jason H. Launders, MSc. Current Technology
The Fundamentals of... Computed Tomography Computed Tomography (CT) systems use x-rays to produce images of slices through a patient s anatomy. Despite having lower spatial resolution than other x-ray
More informationNovel laser power sensor improves process control
Novel laser power sensor improves process control A dramatic technological advancement from Coherent has yielded a completely new type of fast response power detector. The high response speed is particularly
More informationDürr NDT GmbH & Co. KG Höpfigheimer Straße 22 D Bietigheim-Bissingen Germany. Contract No. BAM ZBA Dürr
lndustrial Computed Radiography with storage phosphor imaging plates results of a classification of the system Dürr HD-CR 35 NDT scanner with blue HD- IP Plus imaging plates (HD-IP + ) BAM reference BAM
More informationHDR videos acquisition
HDR videos acquisition dr. Francesco Banterle francesco.banterle@isti.cnr.it How to capture? Videos are challenging: We need to capture multiple frames at different exposure times and everything moves
More informationX-ray technology for electronics inspection
X-ray technology for electronics inspection nikon metrology I vision beyond precision X-ray and CT inspection of electronic components Insight into the inside Get the inside picture of printed circuit
More informationAn experimental method for ripple minimization in transmission data for industrial X-ray computed tomography imaging system
Sādhanā Vol. 27, Part 3, June 2002, pp. 393 404. Printed in India An experimental method for ripple minimization in transmission data for industrial X-ray computed tomography imaging system UMESH KUMAR,
More informationVery short introduction to light microscopy and digital imaging
Very short introduction to light microscopy and digital imaging Hernan G. Garcia August 1, 2005 1 Light Microscopy Basics In this section we will briefly describe the basic principles of operation and
More informationNDT-PRO Services expands service offering
NDT-PRO Services expands service offering NDT-PRO Services announced the formal release of two advanced NDT methods, Phased Array (including TOFD) and Eddy Current. What are they and where are the used?
More informationAcceptance Testing of a Digital Breast Tomosynthesis Unit
Acceptance Testing of a Digital Breast Tomosynthesis Unit 2012 AAPM Spring Clinical Meeting Jessica Clements, M.S., DABR Objectives Review of technology and clinical advantages Acceptance Testing Procedures
More informationMulti-Lateral Shearing Interferometry: Principle and Application on X-ray Laboratory Sources
Multi-Lateral Shearing Interferometry: Principle and Application on X-ray Laboratory Sources International Symposium on Digital Industrial Radiology and Computed Tomography June 22-25, 2015 Adrien STOLIDI
More informationDigital Detector Array Image Quality for Various GOS Scintillators
Digital Detector Array Image Quality for Various GOS Scintillators More info about this article: http://www.ndt.net/?id=22768 Brian S. White 1, Mark E. Shafer 2, William H. Russel 3, Eric Fallet 4, Jacques
More informationOpto Engineering S.r.l.
TUTORIAL #1 Telecentric Lenses: basic information and working principles On line dimensional control is one of the most challenging and difficult applications of vision systems. On the other hand, besides
More informationDiscover new dimensions in flexibility
YXLON FF85 CT High-power and high-resolution computed tomography (CT) inspection system for a wide sample spectrum Discover new dimensions in flexibility Technology with Passion Explore the art of detection
More informationT h e P h a n t o m L a b o r a t o r y
T h e P h a n t o m L a b o r a t o r y 1 CCT228 ATCM Phantom Manual Copyright 2017 WARRANTY THE PHANTOM LABORATORY INCORPORATED ( Seller ) warrants that this product shall remain in good working order
More informationIDEAS+ WP3520 Calibration and data quality toolbox. July 2016 Steve Mackin James Warner
IDEAS+ WP3520 Calibration and data quality toolbox July 2016 Steve Mackin James Warner Proposition : Every image contains the same information Railroad Valley, Nevada London, UK Rationale for the project
More informationPitfalls and Remedies of MDCT Scanners as Quantitative Instruments
intensity m(e) m (/cm) 000 00 0 0. 0 50 0 50 Pitfalls and Remedies of MDCT Scanners as Jiang Hsieh, PhD GE Healthcare Technology University of Wisconsin-Madison Root-Causes of CT Number Inaccuracies Nature
More informationNOT FOR DISTRIBUTION JINST_128P_1010 v2
Pixel sensitivity variations in a CdTe-Medipix2 detector using poly-energetic x-rays R Aamir a, S P Lansley a, b,*, R Zainon a, M Fiederle c, A. Fauler c, D. Greiffenberg c, P H Butler a, d d, e, f, A
More informationDürr NDT GmbH & Co. KG Höpfigheimer Straße 22 D Bietigheim-Bissingen Germany. Contract No. BAM ZBA Dürr
Industrial Computed Radiography with storage phosphor imaging plates results of a classification of the system Dürr HD-CR 35 NDT Plus scanner using HD-IP Plus imaging plates (HD-IP + ) BAM reference BAM
More informationPrinciples of CT scan
Related topics Detector calibration, saturation, CT acquisition, CT reconstruction Principle X-ray computed tomography consists of using X-rays that are converted to a digital signal by a detector and
More informationON THE WAY TO DIGITAL RADIOGRAPHY
The 14 th International Conference of the Slovenian Society for Non-Destructive Testing»Application of Contemporary Non-Destructive Testing in Engineering«September 4-6, 2017, Bernardin, Slovenia More
More informationY11-DR Digital Radiography (DR) Image Quality
Y11-DR Digital Radiography (DR) Image Quality Image quality is stressed for all systems in Safety Code 35. In the relevant sections Health Canada s advice is the manufacturer s recommended test procedures
More informationExperiences of users in Digital Radiography
Computed Radiography Products & Applications Experiences of users in Digital Radiography Jimmy Opdekamp May Jimmy 2006Opdekamp Global Product Manager CR Int l Workshop Imaging NDT Chennai, 25-28 April
More informationCHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY
14 CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 2.1 INTRODUCTION kv-cbct integrated with linear accelerators as a tool for IGRT, was developed to
More informationAn EM Reconstruction with Improved Signal-to. to-noise Ratio for Coded Aperture Imaging
An EM Reconstruction with Improved Signal-to to-noise Ratio for Coded Aperture Imaging Cynthia Tozian, PhD UMass Lowell Bristol-Myers Squibb Medical Imaging Young Investigators Symposium April 12, 2006
More informationA machine vision system for scanner-based laser welding of polymers
A machine vision system for scanner-based laser welding of polymers Zelmar Echegoyen Fernando Liébana Laser Polymer Welding Recent results and future prospects for industrial applications in a European
More informationSynchrotron X-ray tomographic microscopy Theory vs. practice
Synchrotron X-ray tomographic microscopy Theory vs. practice Federica Marone Swiss Light Source, Paul Scherrer Institut, Villigen, Switzerland Theory Radon transform Rf x = Beer-Lambert law I E = I 0 (E)e
More informationSlide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC?
Slide 1 Multi-Slice CT Artifacts and Quality Control Dianna Cody, Ph.D. Chief, Radiologic Physics UT MD Anderson Cancer Center Houston, TX Slide 2 What are the rules or recommendations for CT QC? AAPM
More informationSuppression of metal artifacts using image-based monoenergetic DECT imaging
Suppression of metal artifacts using image-based monoenergetic DECT imaging Poster No.: C-0519 Congress: ECR 2011 Type: Scientific Paper Authors: B. Krauss, B. Schmidt, M. Sedlmair, T. Flohr; Forchheim/DE
More informationXT H Series. X-ray and CT technology for industrial applications NIKON METROLOGY I VISION BEYOND PRECISION
XT H Series X-ray and CT technology for industrial applications NIKON METROLOGY I VISION BEYOND PRECISION INSIGHT INTO THE INSIDE Get the inside picture of complex industrial parts, by looking into the
More information