Quality Control for X-Ray Systems A Tool Chain for NDT Applications

Size: px
Start display at page:

Download "Quality Control for X-Ray Systems A Tool Chain for NDT Applications"

Transcription

1 4th International Symposium on NDT in Aerospace Tu.3.A.2 Quality Control for X-Ray Systems A Tool Chain for NDT Applications Stefan REISINGER *, Alexander ENNEN *, Thorsten WÖRLEIN *, Michael SCHMITT *, Virginia VOLAND * * Fraunhofer IIS/EZRT, Development Centre X-ray Technology (Dr.-Mack-Straße 81, Fürth, Germany / virginia.voland@iis.fraunhofer.de) Abstract. Within the last years, X-ray systems have been established in the field of NDT. Mainly caused by the following aspects the confidence in this technology is still inferior to the confidence in competitive NDT-techniques: Long-term degradation of the system status caused by aging process and abrasion of the system components Fluctuations of image quality during the measurement caused by instability of the X-ray source and possible breakdown of detector elements Poor image quality affected by defect detector elements or other detector characteristics, by a changing imaging geometry during the measurement and by geometric misalignments Insufficient repeatability and measurement accuracy In order to cope with these challenges, we implemented procedures to monitor the X-ray system. The monitoring of the system status is divided into several phases: Prior to use, a computed tomography (CT) system and all its components have to be adjusted to each other in order to prevent any kind of misalignments such as detector or axis tilt. The consequences of unadjusted CT Systems manifest in artifacts or blurred images and therefore result in lower image quality and worse analysis results. The stability of the system components should be controlled regularly during the system lifetime, e.g. weekly. Influencing factors, which are used to forecast the expected image quality and end of life of the X-ray components, are factors such as the spatial resolution, number and distribution of defect detector elements, the contrast sensitivity or size and stability of the focal spot. In preparation the measurement process should be simulated to find suitable parameters for the following measurement. Moreover, for each measurement the image acquisition must be monitored with regard to the image quality to be expected. We therefore defined a quantitative set of characteristics such as e.g. a constant illumination of the detector and present a way to monitor the data acquisition with respect to these characteristics. After the image acquisition is done, a wide range of algorithms can be applied if necessary in order to correct image artifacts and to further improve image quality before the reconstruction task and the final analysis steps occur. Algorithms for the automated adjustment, characterization and monitoring of an X- ray system and correction methods for the further improvement of image quality are presented. License: 1

2 1. Introduction Figure 1.1: Tool chain for NDT applications To ensure the quality of NDT-techniques different procedures have been established. For example the standard for tactile systems is to use calibration phantoms to ensure a stabile system status and resulting measurement inaccuracy. Though within the application area of industrial X-ray systems guidelines for the determination of measurement inaccuracy [7] exist, still no standard is committed in terms of a consistency check as in the medical application. In order to cope with these challenge, we found procedures to monitor the X-ray system. To ensure the stability of an X-ray system different steps are necessary (Figure 1.1). Once while bringing a system into service or after replacement of system components, the single components (X-ray detector, X-ray tube and manipulation system; Figure 1.2) must be aligned with respect to each other. Continuously (daily or weekly) the stability of the system components should be controlled. Besides parasitic drags on system quality has to be logged and corrected (during the measurement or in a post step). Therefore the aim is to develop a tool to monitor the system to allow conclusions over the system status. As a result an expert system can be established to determine suitable parameters and necessary corrections dependent on system status and monitoring of each measurement. Figure 1.2: Drawing of an X-ray system. 2. Tool chain for NDT applications This chapter describes the single steps within the tool chain for NDT applications. 2.1 System Adjustment Uniquely the system components must be aligned with respect to each other. Otherwise influences like tilts of detector and rotatory axis, imprecisely distances between X-ray source and X-ray detector or that the focal spot and the detector centre axis are not arranged 2

3 in the same plain lead to unsatisfactory image quality. Later corrections on the data cannot adjust all these missing alignments even if the maladjustment is completely known. 2.2 Monitoring of system status The stability of the system components should be controlled regularly during the system lifetime, e.g. weekly. Influencing factors, which are used to forecast the expected image quality and end of life of the X-ray components, are factors such as the spatial resolution, number and distribution of defect detector elements, the contrast sensitivity or size and stability of the focal spot. Especially with regard to performance of certification a logging of system status is essential. 2.3 Monitoring of image acquisition During each measurement task some characteristics are monitored to control the image acquisition. If a parameter (homogenous illumination of detector, mean attenuation, movement of optical focal spot) gets out of specified ranges steps can be defined to act on this behaviour. For example repeating single exposures, waiting for a specified time or even restart or abort of the whole measurement can be reasonable. 2.4 Just in time correction Some influences to image quality can be corrected synchronous during the measurement. These corrections are for example determining the position of the center of rotation and determining the position of the optical focal spot [4] or reduction of beam hardening effects [8]. Thus a online reconstruction with a syncronous correction can take place. 2.5 Post processing/ Correction of raw images After the image acquisition is done, a wide range of algorithms can be applied if necessary in order to reduce image artefacts [1] and to further improve image quality before the reconstruction task and final analysis steps occur [2]. These analysis steps are for example examination of dual energy measurements [5]. 3. Quality control for X-ray systems To simplify the use the tool chain of image processing all single steps have to be linked with each other. The challenge of the X-ray tool chain is monitoring of the time variable status of the X-ray system depending on its life time. The most important issue for a typical user of an X-ray system is the long-term management of best parameters (voltage, power, number of projections, integration time, etc.) for each measuring task. To use the full capacity of a system by means of determination and using of resolution limit and best quality of generated images is dependent on the selection of suitable parameters. Degradation of system status (such as basic spatial resolution (Figure 3.1) and detector efficiency (Figure 3.2)) or the state of single components brings up the demand to adapt the necessary parameters. Moreover these parameters are dependent on the measurement task and given requirements. 3

4 Figure 3.1: Determination of basic spatial resolution according to EN and ASTM Complete failures of single components come along with stealthy degradations. Thus the status of the X-ray system and the single components is logged over the time. With a certain empirical values it is possible to predict coming degradations and complete failures. This allows early replacement of single components especially because sourcing of new components is very time expensive due to long time of delivery. Hence down times of X- ray systems caused by breakdown of single components can be reduced. Figure 3.2: Determination of detector efficiency according to ASTM E [6]. Another part of quality control for X-ray systems is a just in time correction of influences like misalignments (Figure 3.3). Especially in the case of extensive measuring tasks or if the time for the reconstruction of the dataset which take some hours or days like described in [3] it is important to be able to react on parasitic drags instantly. In a extrem case these influences are not until noticed if the whole measurement and reconstrution task is finished and are irreparable. This would effect high costs and loss of time. Figure 3.4 shows a plot of logged values. It is observable that for a single projection no signal was triggered. Therefore a simple reiteration of this measurement fragment would lead to more usable results. 4

5 Figure 3.3: Observation of drift of focal spot. It is obvious, that the movement of the focal spot changes its direction during the long-term observation. Figure 3.4: Observation of influencing values during the measurement. At the marked instant of time all monitored values show anomalous values. This is an indication to insufficient image quality (In this case the tube was dropped out). If a synchronous correction is not possible at least a logging of system behaviour during the measurement to predict necessary corrections on the raw images or the reconstructed volume (Figure 3.5) is recommended. Figure 3.5: left: Ring artefacts with / without correction. Right: Detector shift with / without correction. 5

6 As a follow up, conclusions can be drawn from the final image quality to necessary parameters of coming up measurement tasks. Conclusion The main benefit of these tools is that the whole tool chain of image processing is available in one software solution. Therefore the possibility of data exchange between the single processing steps is given. The coming steps for this project are the generation of empirical values of the degradation of system components and involved impact to image quality. Having this data it is possible to specify necessary adaption of measurement parameters und give a forecast of system live time. References [1] S. Reisinger, V. Voland, S. Burtzlaff, M. Schmitt und N. Uhlmann, Rohbildkorrekturen für die Artefaktreduktion in der Computertomographie, Jahrestagung der DGZfP BB 127-CD, [2] V. Voland, J.-I. Tiemegni, T. Wörlein, N. Uhlmann und F. Kargl, Steuerung von Röntgenexperimenten, Jahrestagung der DGZfP BB 127-CD, [3] V. Voland, J. Freitag und N. Uhlmann, A CT System for the Analysis of Prehistoric Ice Cores, Elst (Hrsg): Microelectronic Systems: Circuits, Systems and Applications, Berlin, Springer, [4] V. Voland, M. Schmitt und S. Reisinger, Geometric Adjustment Methods to Improve Reconstruction Quality on Rotational Cone-beam Systems, in ICT Conference, Wels, [5] M. Firsching, F. Nachtrab, J. Mühlbauer und N. Uhlmann, Detection of Enclosed Diamonds using Dual Energy X-ray imaging, 18th World Conference on Nondestructive Testing, [6] ASTM E e1 Standard Practice for Manufacturing Characterization of Digital Detector Arrays, ASTM International, West Conshohocken, PA, 2003, DOI: /C A, [7] VDI/VDE Bestimmung der Messunsicherheit und der Prüfprozesseignung von Koordinatenmessgeräten mit CT-Sensoren. [8] M. Franz, S. Kasperl und M. Stamminger, Synchronous Arftefact Reduction in Industrial Computed Tomography, Technisches Messen TM 77, pp ,

v tome x m microfocus CT

v tome x m microfocus CT GE Inspection Technologies v tome x m microfocus CT Uniting premium 3D metrology and inspection with quality and speed. gemeasurement.com/ct x plore precision CT line Inspect with precision, power, and

More information

Philip Sperling. Sales Science and New Materials, YXLON International GmbH, Essener Bogen 15, Hamburg, Germany.

Philip Sperling. Sales Science and New Materials, YXLON International GmbH, Essener Bogen 15, Hamburg, Germany. A new generation of x-ray computed tomography devices for quality inspection and metrology inspection in the field of additive manufacturing and other sciences Philip Sperling Sales Science and New Materials,

More information

Influence of surface roughness on measurement uncertainty in Computed Tomography

Influence of surface roughness on measurement uncertainty in Computed Tomography Influence of surface roughness on measurement uncertainty in Computed Tomography More info about this article: http://www.ndt.net/?id=21910 Leonard Schild, Alexandra Kraemer, Doreen Reiling, Hanjue Wu,

More information

Uncertainty in CT Metrology: Visualizations for Exploration and Analysis of Geometric Tolerances

Uncertainty in CT Metrology: Visualizations for Exploration and Analysis of Geometric Tolerances Uncertainty in CT Metrology: Visualizations for Exploration and Analysis of Geometric Tolerances Artem Amirkhanov 1, Bernhard Fröhler 1, Michael Reiter 1, Johann Kastner 1, M. Eduard Grӧller 2, Christoph

More information

Migrating from traditional to Digital Radiography in Aerospace

Migrating from traditional to Digital Radiography in Aerospace Migrating from traditional to Digital Radiography in Aerospace More info about this article: http://www.ndt.net/?id=22663 Abstract Lennart Schulenburg VisiConsult X-ray System & Solutions GmbH Brandenbrooker

More information

PERFORMANCE CHARACTERIZATION OF AMORPHOUS SILICON DIGITAL DETECTOR ARRAYS FOR GAMMA RADIOGRAPHY

PERFORMANCE CHARACTERIZATION OF AMORPHOUS SILICON DIGITAL DETECTOR ARRAYS FOR GAMMA RADIOGRAPHY 12 th A-PCNDT 2006 Asia-Pacific Conference on NDT, 5 th 10 th Nov 2006, Auckland, New Zealand PERFORMANCE CHARACTERIZATION OF AMORPHOUS SILICON DIGITAL DETECTOR ARRAYS FOR GAMMA RADIOGRAPHY Rajashekar

More information

Amorphous Selenium Direct Radiography for Industrial Imaging

Amorphous Selenium Direct Radiography for Industrial Imaging DGZfP Proceedings BB 67-CD Paper 22 Computerized Tomography for Industrial Applications and Image Processing in Radiology March 15-17, 1999, Berlin, Germany Amorphous Selenium Direct Radiography for Industrial

More information

Making use of phase-contrast in dimensional µct-measurements of weakly absorbing objects

Making use of phase-contrast in dimensional µct-measurements of weakly absorbing objects Making use of phase-contrast in dimensional µct-measurements of weakly absorbing objects Andreas Staude 1, Karsten Ehrig 1, Bernhard Illerhaus 1, Martin Spreemann 2 1 BAM Bundesanstalt für Materialforschung

More information

Tomographic 3D-Radiometry for the Visualisation and Measurement of the Defects of Girth Seams

Tomographic 3D-Radiometry for the Visualisation and Measurement of the Defects of Girth Seams ECNDT 2006 - We.3.2.3 Tomographic 3D-Radiometry for the Visualisation and Measurement of the Defects of Girth Seams Bernhard REDMER, Uwe EWERT Federal Institute of Materials Research and Testing (BAM),

More information

Alignment of the camera

Alignment of the camera Related topics Detector Alignment, Rotation axis, tilt, Principle Alignment of the detector and the rotation stage is very important to get optimal quality images of a CT scan. In this experiment, the

More information

AUTOMATED AND QUANTITATIVE METHOD FOR QUALITY ASSURANCE OF DIGITAL RADIOGRAPHY IMAGING SYSTEMS

AUTOMATED AND QUANTITATIVE METHOD FOR QUALITY ASSURANCE OF DIGITAL RADIOGRAPHY IMAGING SYSTEMS International Workshop SMART MATERIALS, STRUCTURES & NDT in AEROSPACE Conference NDT in Canada 2011 2-4 November 2011, Montreal, Quebec, Canada AUTOMATED AND QUANTITATIVE METHOD FOR QUALITY ASSURANCE OF

More information

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over a lifetime Breast cancer screening programs rely on

More information

HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE

HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE Takeyuki Hashimoto 1), Morio Onoe 2), Hiroshi Nakamura 3), Tamon Inouye 4), Hiromichi Jumonji 5), Iwao Takahashi 6); 1)Yokohama Soei

More information

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging Shawn D. Teague, MD DISCLOSURES 3DR- advisory committee CT PHYSICS WITH AN EMPHASIS ON APPLICATION IN THORACIC AND CARDIAC IMAGING

More information

On Determination of Focal Laws for Linear Phased Array Probes as to the Active and Passive Element Size

On Determination of Focal Laws for Linear Phased Array Probes as to the Active and Passive Element Size 19 th World Conference on Non-Destructive Testing 2016 On Determination of Focal Laws for Linear Phased Array Probes as to the Active and Passive Element Size Andreas GOMMLICH 1, Frank SCHUBERT 2 1 Institute

More information

X-RAY COMPUTED TOMOGRAPHY

X-RAY COMPUTED TOMOGRAPHY X-RAY COMPUTED TOMOGRAPHY Bc. Jan Kratochvíla Czech Technical University in Prague Faculty of Nuclear Sciences and Physical Engineering Abstract Computed tomography is a powerful tool for imaging the inner

More information

Detector technology in simultaneous spectral imaging

Detector technology in simultaneous spectral imaging Computed tomography Detector technology in simultaneous spectral imaging Philips IQon Spectral CT Z. Romman, I. Uman, Y. Yagil, D. Finzi, N. Wainer, D. Milstein; Philips Healthcare While CT has become

More information

Fully Automated Inspection of Welded Seams with a New High Resolution Durable X-ray Camera System

Fully Automated Inspection of Welded Seams with a New High Resolution Durable X-ray Camera System ECNDT 2006 - Tu.2.4.4 Fully Automated Inspection of Welded Seams with a New High Resolution Durable X-ray Camera System Thomas WENZEL, Rolf BEHRENDT, Theobald FUCHS, Ulrich HÜTTEN, Günther KOSTKA, Peter

More information

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT)

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT) Radionuclide Imaging MII 3073 Single Photon Emission Computed Tomography (SPECT) Single Photon Emission Computed Tomography (SPECT) The successful application of computer algorithms to x-ray imaging in

More information

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 2 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 COMPUTERIZED IMAGING Section I: Chapter 2 RADT

More information

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura New Technology https://www.youtube.com/watch?v=ptkzznazb 7U COMPUTED

More information

Instruction manual for T3DS software. Tool for THz Time-Domain Spectroscopy. Release 4.0

Instruction manual for T3DS software. Tool for THz Time-Domain Spectroscopy. Release 4.0 Instruction manual for T3DS software Release 4.0 Table of contents 0. Setup... 3 1. Start-up... 5 2. Input parameters and delay line control... 6 3. Slow scan measurement... 8 4. Fast scan measurement...

More information

Hiding In Plain Sight. How Ultrasonics Can Help You Find the Smallest Bonded Wafer and Device Defects. A Sonix White Paper

Hiding In Plain Sight. How Ultrasonics Can Help You Find the Smallest Bonded Wafer and Device Defects. A Sonix White Paper Hiding In Plain Sight How Ultrasonics Can Help You Find the Smallest Bonded Wafer and Device Defects A Sonix White Paper If You Can See It, You Can Solve It: Understanding Ultrasonic Inspection of Bonded

More information

Operating Procedures for MICROCT1 Nikon XTH 225 ST

Operating Procedures for MICROCT1 Nikon XTH 225 ST Operating Procedures for MICROCT1 Nikon XTH 225 ST Ensuring System is Ready (go through to ensure all windows and tasks below have been completed either by you or someone else prior to mounting and scanning

More information

Digital Radiographic Inspection replacing traditional RT and 3D RT Development

Digital Radiographic Inspection replacing traditional RT and 3D RT Development Digital Radiographic Inspection replacing traditional RT and 3D RT Development Iploca Novel Construction Meeting 27&28 March 2014 Geneva By Jan van der Ent Technical Authority International Contents Introduction

More information

QC Testing for Computed Tomography (CT) Scanner

QC Testing for Computed Tomography (CT) Scanner QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including

More information

Discover the difference in efficiency

Discover the difference in efficiency Y.CT Compact Fan-beam computed tomography (CT) inspection system for high-density medium and large-sized parts Discover the difference in efficiency Technology with Passion Explore the art of detection

More information

MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE

MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE A. Miceli ab, R. Thierry a, A. Flisch a, U. Sennhauser a, F. Casali b a Empa - Swiss Federal Laboratories for

More information

Robert Pagnanelli BSRT(R)(N), CNMT, NCT, FASNC Chief Technologist, Nuclear Imaging Duke University Medical Center. Thursday September 8, 2011

Robert Pagnanelli BSRT(R)(N), CNMT, NCT, FASNC Chief Technologist, Nuclear Imaging Duke University Medical Center. Thursday September 8, 2011 Robert Pagnanelli BSRT(R)(N), CNMT, NCT, FASNC Chief Technologist, Nuclear Imaging Duke University Medical Center Thursday September 8, 2011 Quality Control Quality control should be performed because:

More information

2/14/2019. Nuclear Medicine Artifacts. Symmetric energy windows

2/14/2019. Nuclear Medicine Artifacts. Symmetric energy windows Nuclear Medicine Artifacts SCPMG Medical Imaging Technology & Informatics Medical Physics Group Brian Helbig, MS, DABR 1 2 Symmetric energy windows 3 1 Dynamic clinical study Energy peak shift Electrical

More information

A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup

A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 4, FALL 2009 A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup Huaiqun Guan,

More information

More Info at Open Access Database by S. Dutta and T. Schmidt

More Info at Open Access Database  by S. Dutta and T. Schmidt More Info at Open Access Database www.ndt.net/?id=17657 New concept for higher Robot position accuracy during thermography measurement to be implemented with the existing prototype automated thermography

More information

Digital Radiography : Flat Panel

Digital Radiography : Flat Panel Digital Radiography : Flat Panel Flat panels performances & operation How does it work? - what is a sensor? - ideal sensor Flat panels limits and solutions - offset calibration - gain calibration - non

More information

TDI Imaging: An Efficient AOI and AXI Tool

TDI Imaging: An Efficient AOI and AXI Tool TDI Imaging: An Efficient AOI and AXI Tool Yakov Bulayev Hamamatsu Corporation Bridgewater, New Jersey Abstract As a result of heightened requirements for quality, integrity and reliability of electronic

More information

Investigation of the film Kodak MX 125 in Agfa NDT S eco processing

Investigation of the film Kodak MX 125 in Agfa NDT S eco processing Investigation of the film Kodak MX 125 in Agfa NDT S eco processing Test Report-No.: BAM VIII.3 / 6568b Date: 26. 10. 2005 Name and address Kodak GmbH, of the customer: Hedelfinger Straße 60, 70327 Stuttgart,

More information

Computed Radiography

Computed Radiography BAM Berlin Computed Radiography --INDE 2007, Kalpakkam, India -- Uwe Zscherpel, Uwe Ewert BAM Berlin, Division VIII.3 Requests Requests and and information information to: to: Dr. Dr. U. U. Zscherpel Zscherpel

More information

OPTIMIZING SPATIAL RESOLUTION WITH THE MECHANICAL DESIGN OF AN X-RAY COMPU1ED TOMOGRAPHY SCANNER

OPTIMIZING SPATIAL RESOLUTION WITH THE MECHANICAL DESIGN OF AN X-RAY COMPU1ED TOMOGRAPHY SCANNER OPTIMIZING SPATIAL RESOLUTION WITH THE MECHANICAL DESIGN OF AN X-RAY COMPU1ED TOMOGRAPHY SCANNER Lowell D. Harris, RichardT. Bernardi, Simon H. C. Hughes, and Robert E. Slocum Bio-Imaging Research, Inc.

More information

Defining the Optimal Beam Hardening Correction Parameters for CT Dimensional Metrology Applications

Defining the Optimal Beam Hardening Correction Parameters for CT Dimensional Metrology Applications International Conference on Competitive Manufacturing Defining the Optimal Beam Hardening Correction Parameters for CT Dimensional Metrology Applications Y. Tan 1,2, K. Kiekens 1,2, F. Welkenhuyzen 2,

More information

Radiographic Testing (RT) [10]

Radiographic Testing (RT) [10] Radiographic Testing (RT) [10] Definition: An NDT method that utilizes x-rays or gamma radiation to detect discontinuities in materials, and to present their images on recording medium. 1> Electromagnetic

More information

Towards accurate measurements with synchrotron tomography Problems and pitfalls. Robert C. Atwood. Nghia T. Vo, Michael Drakopoulos, Thomas Connolley

Towards accurate measurements with synchrotron tomography Problems and pitfalls. Robert C. Atwood. Nghia T. Vo, Michael Drakopoulos, Thomas Connolley Towards accurate measurements with synchrotron tomography Problems and pitfalls Robert C. Atwood Nghia T. Vo, Michael Drakopoulos, Thomas Connolley Artefacts in Synchrotron X-ray Tomography Rings Rings

More information

Thermionic x-ray. Alternative technologies. Electron Field Emission. CNT Based Field Emission X-Ray Source

Thermionic x-ray. Alternative technologies. Electron Field Emission. CNT Based Field Emission X-Ray Source Energy Level (ev) Multi-beam x-ray source array based on carbon nanotube field emission O. Zhou, JP Lu, X. Calderon-Colon, X. Qian, G. Yang, G. Cao, E. Gidcumb, A. Tucker, J. Shan University of North Carolina

More information

Digital Radiology with Photon Counting Detectors

Digital Radiology with Photon Counting Detectors 11th European Conference on Non-Destructive Testing (ECNDT 2014), October 6-10, 2014, Prague, Czech Republic Digital Radiology with Photon Counting Detectors More Info at Open Access Database www.ndt.net/?id=16709

More information

Real Time Linear Array Imaging. Brian Caccamise

Real Time Linear Array Imaging. Brian Caccamise Real Time Linear Array Imaging Brian Caccamise 1 Real Time Linear Array Imaging What is Real Time Linear Array Imaging? Or Real Time Radiography (RTR)? 2 Real Time Linear Array Imaging It s Not This! Shoe

More information

The following article is a translation of parts of the original publication of Karl-Ludwig Bath in the german astronomical magazine:

The following article is a translation of parts of the original publication of Karl-Ludwig Bath in the german astronomical magazine: The following article is a translation of parts of the original publication of Karl-Ludwig Bath in the german astronomical magazine: Sterne und Weltraum 1973/6, p.177-180. The publication of this translation

More information

IQI-Sensitivity and Applications of Flat Panel Detectors and X-Ray Image Intensifiers A Comparison

IQI-Sensitivity and Applications of Flat Panel Detectors and X-Ray Image Intensifiers A Comparison IQI-Sensitivity and Applications of Flat Panel Detectors and X-Ray Image Intensifiers A Comparison Dr. Matthias Purschke/ Ulf Reimer, Agfa NDT Pantak Seifert GmbH und Co. KG, Bogenstr. 4, 96 Ahrensburg,

More information

WaveMaster IOL. Fast and accurate intraocular lens tester

WaveMaster IOL. Fast and accurate intraocular lens tester WaveMaster IOL Fast and accurate intraocular lens tester INTRAOCULAR LENS TESTER WaveMaster IOL Fast and accurate intraocular lens tester WaveMaster IOL is a new instrument providing real time analysis

More information

Microwave Testing (µt): An Overview. Johann Hinken, FI Test- und Messtechnik GmbH Magdeburg, Germany, July 2016

Microwave Testing (µt): An Overview. Johann Hinken, FI Test- und Messtechnik GmbH Magdeburg, Germany, July 2016 Microwave Testing (µt): An Overview Johann Hinken, FI Test- und Messtechnik GmbH Magdeburg, Germany, July 2016 More info about this article: http://www.ndt.net/?id=21377 Content: 1. Introduction 2. Physical

More information

ISO INTERNATIONAL STANDARD. Non-destructive testing of welds Radiographic testing Part 1: X- and gamma-ray techniques with film

ISO INTERNATIONAL STANDARD. Non-destructive testing of welds Radiographic testing Part 1: X- and gamma-ray techniques with film INTERNATIONAL STANDARD ISO 17636-1 First edition 2013-01-15 Non-destructive testing of welds Radiographic testing Part 1: X- and gamma-ray techniques with film Contrôle non destructif des assemblages soudés

More information

Photomultiplier Tube

Photomultiplier Tube Nuclear Medicine Uses a device known as a Gamma Camera. Also known as a Scintillation or Anger Camera. Detects the release of gamma rays from Radionuclide. The radionuclide can be injected, inhaled or

More information

A COMPARISON BETWEEN ASTM E588 AND SEP 1927 RELATING RESOLUTION LIMITS AT DETERMINATION OF THE PURITY GRADE

A COMPARISON BETWEEN ASTM E588 AND SEP 1927 RELATING RESOLUTION LIMITS AT DETERMINATION OF THE PURITY GRADE 19 th World Conference on Non-Destructive Testing 2016 A COMPARISON BETWEEN ASTM E588 AND SEP 1927 RELATING RESOLUTION LIMITS AT DETERMINATION OF THE PURITY GRADE Daniel KOTSCHATE 1, Dirk GOHLKE 1, Rainer

More information

Operation and Sound Field of an Ultrasonic Biplane-Array

Operation and Sound Field of an Ultrasonic Biplane-Array 19 th World Conference on Non-Destructive Testing 2016 Operation and Sound Field of an Ultrasonic Biplane-Array R. HIPP 1, A. GOMMLICH 2, D. JONEIT 1, F. SCHUBERT 1, H. HEUER 1 1 Fraunhofer-Institut für

More information

Available online at ScienceDirect. Procedia CIRP 11 (2013 )

Available online at  ScienceDirect. Procedia CIRP 11 (2013 ) Available online at www.sciencedirect.com ScienceDirect Procedia CIRP 11 (213 ) 234 238 2 nd International Through-life Engineering Services Conference Applicability of industrial robots for machining

More information

Scatter Correction by Modulation of Primary Radiation in Industrial X-ray CT: Beam-hardening Effects and their Correction

Scatter Correction by Modulation of Primary Radiation in Industrial X-ray CT: Beam-hardening Effects and their Correction International Symposium on Digital Industrial Radiology and Computed Tomography - Mo.3.2 Scatter Correction by Modulation of Primary Radiation in Industrial X-ray CT: Beam-hardening Effects and their Correction

More information

Laser Technology Inc.

Laser Technology Inc. Field of View Determination for Shearography Nondestructive Evaluation John W. Newman, President Laser Technology Inc. Norristown, PA 19403 USA Introduction Shearography NDT is a highly developed, mature

More information

WaveMaster IOL. Fast and Accurate Intraocular Lens Tester

WaveMaster IOL. Fast and Accurate Intraocular Lens Tester WaveMaster IOL Fast and Accurate Intraocular Lens Tester INTRAOCULAR LENS TESTER WaveMaster IOL Fast and accurate intraocular lens tester WaveMaster IOL is an instrument providing real time analysis of

More information

SEAMS DUE TO MULTIPLE OUTPUT CCDS

SEAMS DUE TO MULTIPLE OUTPUT CCDS Seam Correction for Sensors with Multiple Outputs Introduction Image sensor manufacturers are continually working to meet their customers demands for ever-higher frame rates in their cameras. To meet this

More information

COMPUTED TOMOGRAPHY 1

COMPUTED TOMOGRAPHY 1 COMPUTED TOMOGRAPHY 1 Why CT? Conventional X ray picture of a chest 2 Introduction Why CT? In a normal X-ray picture, most soft tissue doesn't show up clearly. To focus in on organs, or to examine the

More information

Equivalent Penetrameter Sensitivity (EPS) for Performance Evaluation of Computed Radiography Systems Muzibur Khan * and Mike Brothers

Equivalent Penetrameter Sensitivity (EPS) for Performance Evaluation of Computed Radiography Systems Muzibur Khan * and Mike Brothers Equivalent Penetrameter Sensitivity (EPS) for Performance Evaluation of Computed Radiography Systems Muzibur Khan * and Mike Brothers More info about this article: http://www.ndt.net/?id=22782 Abstract

More information

Investigation of the line-pair pattern method for evaluating mammographic focal spot performance

Investigation of the line-pair pattern method for evaluating mammographic focal spot performance Investigation of the line-pair pattern method for evaluating mammographic focal spot performance Mitchell M. Goodsitt, a) Heang-Ping Chan, and Bob Liu Department of Radiology, University of Michigan, Ann

More information

Evaluation of no-grid radiography using the digital scattered x-ray removal processing

Evaluation of no-grid radiography using the digital scattered x-ray removal processing Evaluation of no-grid radiography using the digital scattered x-ray removal processing Poster No.: C-0416 Congress: ECR 2016 Type: Authors: Scientific Exhibit R. Suzuki 1, T. Goto 1, H. Ogawa 2, N. Amimoto

More information

2D Digital Radiography 3D Computed Tomography Inspection Services Laboratories. xrayinspectionservice.com

2D Digital Radiography 3D Computed Tomography Inspection Services Laboratories. xrayinspectionservice.com 2D Digital Radiography 3D Computed Tomography Inspection Services Laboratories 4nsi.com xrayinspectionservice.com Celebrating over 28 years of Innovation North Star Imaging was founded in 1986 by Ken Ness

More information

Stability of a Fiber-Fed Heterodyne Interferometer

Stability of a Fiber-Fed Heterodyne Interferometer Stability of a Fiber-Fed Heterodyne Interferometer Christoph Weichert, Jens Flügge, Paul Köchert, Rainer Köning, Physikalisch Technische Bundesanstalt, Braunschweig, Germany; Rainer Tutsch, Technische

More information

DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter

DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter Revision Date: January 5th, 2017 www.disc-imaging.com Table of Contents Section A: Preliminary Setup Requirements... 4 Tools

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

An Activity in Computed Tomography

An Activity in Computed Tomography Pre-lab Discussion An Activity in Computed Tomography X-rays X-rays are high energy electromagnetic radiation with wavelengths smaller than those in the visible spectrum (0.01-10nm and 4000-800nm respectively).

More information

Novel Imaging Techniques for Defects Characterisation in Phased Array Inspection

Novel Imaging Techniques for Defects Characterisation in Phased Array Inspection Novel Imaging Techniques for Defects Characterisation in Phased Array Inspection P. Rioux 1, F. Lachance 1 and J. Turcotte 1 1 Sonatest, Québec, Canada Phone: +1 418 683 6222, e-mail: sales@sonatest.com

More information

make it easy, with Ray

make it easy, with Ray Lower dose - Quick scan times - Pulsed X-ray technology - Multiple scan modes 3 Dedicated detectors - Reliable performance - No damage - Long life span Easy upgrade - Ready to upgrade CBCT & Cephalometric

More information

Projection Data and Scanograms in X-ray Computed Tomography

Projection Data and Scanograms in X-ray Computed Tomography DICTA2002: Digital Image Computing Techniques and Applications, 21 22 January 2002, Melbourne, Australia 1 Projection Data and Scanograms in X-ray Computed Tomography Russell Horney Imants Svalbe Peter

More information

Standard Practice for Qualification of Radioscopic Systems 1

Standard Practice for Qualification of Radioscopic Systems 1 Designation: 95 An American National Standard Standard Practice for Qualification of Radioscopic Systems 1 This standard is issued under the fixed designation ; the number immediately following the designation

More information

Maximum Performance, Minimum Space

Maximum Performance, Minimum Space TECHNOLOGY HISTORY For over 130 years, Toshiba has been a world leader in developing technology to improve the quality of life. Our 50,000 global patents demonstrate a long, rich history of leading innovation.

More information

Nondestructive and Contactless Materials Characterization with the Help of Microwave Sensors

Nondestructive and Contactless Materials Characterization with the Help of Microwave Sensors ECNDT 2006 - Tu.1.8.3 Nondestructive and Contactless Materials Characterization with the Help of Microwave Sensors Christoph SKLARCZYK, Roman PINCHUK, Valeri MELEV, Fraunhofer-Institut Zerstörungsfreie

More information

Computed Tomography. The Fundamentals of... THE FUNDAMENTALS OF... Jason H. Launders, MSc. Current Technology

Computed Tomography. The Fundamentals of... THE FUNDAMENTALS OF... Jason H. Launders, MSc. Current Technology The Fundamentals of... Computed Tomography Computed Tomography (CT) systems use x-rays to produce images of slices through a patient s anatomy. Despite having lower spatial resolution than other x-ray

More information

Novel laser power sensor improves process control

Novel laser power sensor improves process control Novel laser power sensor improves process control A dramatic technological advancement from Coherent has yielded a completely new type of fast response power detector. The high response speed is particularly

More information

Dürr NDT GmbH & Co. KG Höpfigheimer Straße 22 D Bietigheim-Bissingen Germany. Contract No. BAM ZBA Dürr

Dürr NDT GmbH & Co. KG Höpfigheimer Straße 22 D Bietigheim-Bissingen Germany. Contract No. BAM ZBA Dürr lndustrial Computed Radiography with storage phosphor imaging plates results of a classification of the system Dürr HD-CR 35 NDT scanner with blue HD- IP Plus imaging plates (HD-IP + ) BAM reference BAM

More information

HDR videos acquisition

HDR videos acquisition HDR videos acquisition dr. Francesco Banterle francesco.banterle@isti.cnr.it How to capture? Videos are challenging: We need to capture multiple frames at different exposure times and everything moves

More information

X-ray technology for electronics inspection

X-ray technology for electronics inspection X-ray technology for electronics inspection nikon metrology I vision beyond precision X-ray and CT inspection of electronic components Insight into the inside Get the inside picture of printed circuit

More information

An experimental method for ripple minimization in transmission data for industrial X-ray computed tomography imaging system

An experimental method for ripple minimization in transmission data for industrial X-ray computed tomography imaging system Sādhanā Vol. 27, Part 3, June 2002, pp. 393 404. Printed in India An experimental method for ripple minimization in transmission data for industrial X-ray computed tomography imaging system UMESH KUMAR,

More information

Very short introduction to light microscopy and digital imaging

Very short introduction to light microscopy and digital imaging Very short introduction to light microscopy and digital imaging Hernan G. Garcia August 1, 2005 1 Light Microscopy Basics In this section we will briefly describe the basic principles of operation and

More information

NDT-PRO Services expands service offering

NDT-PRO Services expands service offering NDT-PRO Services expands service offering NDT-PRO Services announced the formal release of two advanced NDT methods, Phased Array (including TOFD) and Eddy Current. What are they and where are the used?

More information

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Acceptance Testing of a Digital Breast Tomosynthesis Unit Acceptance Testing of a Digital Breast Tomosynthesis Unit 2012 AAPM Spring Clinical Meeting Jessica Clements, M.S., DABR Objectives Review of technology and clinical advantages Acceptance Testing Procedures

More information

Multi-Lateral Shearing Interferometry: Principle and Application on X-ray Laboratory Sources

Multi-Lateral Shearing Interferometry: Principle and Application on X-ray Laboratory Sources Multi-Lateral Shearing Interferometry: Principle and Application on X-ray Laboratory Sources International Symposium on Digital Industrial Radiology and Computed Tomography June 22-25, 2015 Adrien STOLIDI

More information

Digital Detector Array Image Quality for Various GOS Scintillators

Digital Detector Array Image Quality for Various GOS Scintillators Digital Detector Array Image Quality for Various GOS Scintillators More info about this article: http://www.ndt.net/?id=22768 Brian S. White 1, Mark E. Shafer 2, William H. Russel 3, Eric Fallet 4, Jacques

More information

Opto Engineering S.r.l.

Opto Engineering S.r.l. TUTORIAL #1 Telecentric Lenses: basic information and working principles On line dimensional control is one of the most challenging and difficult applications of vision systems. On the other hand, besides

More information

Discover new dimensions in flexibility

Discover new dimensions in flexibility YXLON FF85 CT High-power and high-resolution computed tomography (CT) inspection system for a wide sample spectrum Discover new dimensions in flexibility Technology with Passion Explore the art of detection

More information

T h e P h a n t o m L a b o r a t o r y

T h e P h a n t o m L a b o r a t o r y T h e P h a n t o m L a b o r a t o r y 1 CCT228 ATCM Phantom Manual Copyright 2017 WARRANTY THE PHANTOM LABORATORY INCORPORATED ( Seller ) warrants that this product shall remain in good working order

More information

IDEAS+ WP3520 Calibration and data quality toolbox. July 2016 Steve Mackin James Warner

IDEAS+ WP3520 Calibration and data quality toolbox. July 2016 Steve Mackin James Warner IDEAS+ WP3520 Calibration and data quality toolbox July 2016 Steve Mackin James Warner Proposition : Every image contains the same information Railroad Valley, Nevada London, UK Rationale for the project

More information

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments intensity m(e) m (/cm) 000 00 0 0. 0 50 0 50 Pitfalls and Remedies of MDCT Scanners as Jiang Hsieh, PhD GE Healthcare Technology University of Wisconsin-Madison Root-Causes of CT Number Inaccuracies Nature

More information

NOT FOR DISTRIBUTION JINST_128P_1010 v2

NOT FOR DISTRIBUTION JINST_128P_1010 v2 Pixel sensitivity variations in a CdTe-Medipix2 detector using poly-energetic x-rays R Aamir a, S P Lansley a, b,*, R Zainon a, M Fiederle c, A. Fauler c, D. Greiffenberg c, P H Butler a, d d, e, f, A

More information

Dürr NDT GmbH & Co. KG Höpfigheimer Straße 22 D Bietigheim-Bissingen Germany. Contract No. BAM ZBA Dürr

Dürr NDT GmbH & Co. KG Höpfigheimer Straße 22 D Bietigheim-Bissingen Germany. Contract No. BAM ZBA Dürr Industrial Computed Radiography with storage phosphor imaging plates results of a classification of the system Dürr HD-CR 35 NDT Plus scanner using HD-IP Plus imaging plates (HD-IP + ) BAM reference BAM

More information

Principles of CT scan

Principles of CT scan Related topics Detector calibration, saturation, CT acquisition, CT reconstruction Principle X-ray computed tomography consists of using X-rays that are converted to a digital signal by a detector and

More information

ON THE WAY TO DIGITAL RADIOGRAPHY

ON THE WAY TO DIGITAL RADIOGRAPHY The 14 th International Conference of the Slovenian Society for Non-Destructive Testing»Application of Contemporary Non-Destructive Testing in Engineering«September 4-6, 2017, Bernardin, Slovenia More

More information

Y11-DR Digital Radiography (DR) Image Quality

Y11-DR Digital Radiography (DR) Image Quality Y11-DR Digital Radiography (DR) Image Quality Image quality is stressed for all systems in Safety Code 35. In the relevant sections Health Canada s advice is the manufacturer s recommended test procedures

More information

Experiences of users in Digital Radiography

Experiences of users in Digital Radiography Computed Radiography Products & Applications Experiences of users in Digital Radiography Jimmy Opdekamp May Jimmy 2006Opdekamp Global Product Manager CR Int l Workshop Imaging NDT Chennai, 25-28 April

More information

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 14 CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 2.1 INTRODUCTION kv-cbct integrated with linear accelerators as a tool for IGRT, was developed to

More information

An EM Reconstruction with Improved Signal-to. to-noise Ratio for Coded Aperture Imaging

An EM Reconstruction with Improved Signal-to. to-noise Ratio for Coded Aperture Imaging An EM Reconstruction with Improved Signal-to to-noise Ratio for Coded Aperture Imaging Cynthia Tozian, PhD UMass Lowell Bristol-Myers Squibb Medical Imaging Young Investigators Symposium April 12, 2006

More information

A machine vision system for scanner-based laser welding of polymers

A machine vision system for scanner-based laser welding of polymers A machine vision system for scanner-based laser welding of polymers Zelmar Echegoyen Fernando Liébana Laser Polymer Welding Recent results and future prospects for industrial applications in a European

More information

Synchrotron X-ray tomographic microscopy Theory vs. practice

Synchrotron X-ray tomographic microscopy Theory vs. practice Synchrotron X-ray tomographic microscopy Theory vs. practice Federica Marone Swiss Light Source, Paul Scherrer Institut, Villigen, Switzerland Theory Radon transform Rf x = Beer-Lambert law I E = I 0 (E)e

More information

Slide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC?

Slide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC? Slide 1 Multi-Slice CT Artifacts and Quality Control Dianna Cody, Ph.D. Chief, Radiologic Physics UT MD Anderson Cancer Center Houston, TX Slide 2 What are the rules or recommendations for CT QC? AAPM

More information

Suppression of metal artifacts using image-based monoenergetic DECT imaging

Suppression of metal artifacts using image-based monoenergetic DECT imaging Suppression of metal artifacts using image-based monoenergetic DECT imaging Poster No.: C-0519 Congress: ECR 2011 Type: Scientific Paper Authors: B. Krauss, B. Schmidt, M. Sedlmair, T. Flohr; Forchheim/DE

More information

XT H Series. X-ray and CT technology for industrial applications NIKON METROLOGY I VISION BEYOND PRECISION

XT H Series. X-ray and CT technology for industrial applications NIKON METROLOGY I VISION BEYOND PRECISION XT H Series X-ray and CT technology for industrial applications NIKON METROLOGY I VISION BEYOND PRECISION INSIGHT INTO THE INSIDE Get the inside picture of complex industrial parts, by looking into the

More information