Development of a New Reference Spectrophotometer. J.C. Zwinkels, D.S. Gignac
|
|
- Norman Hood
- 5 years ago
- Views:
Transcription
1 Development of a New Reference Spectrophotometer Development of a New Reference Spectrophotometer J.C. Zwinkels, D.S. Gignac J.C. Zwinkels, D.S. Gignac National Research Council of Canada, Division of Physics Ottawa, Ontario, Canada K1A OR6 National Research Council of Canada, Division of Physics Ottawa, Ontario, Canada K1A OR6 ABSTRACT ABSTRACT A new reference Spectrophotometer is being developed at the National Research Council of Canada (NRCC) for high-accuracy transmittance measurements over the spectral range 200 to 2500 nm. This computerized instrument is a single-beam design based upon a servomotor-driven double monochromator with a wavelength resolution of nm. The other main components are: (1) two interchangeable sources (deuterium and tungsten-halogen), (2) two computer-selectable TE-cooled detectors (GaAs photomultiplier tube and PbS cell), (3) all-reflective input and exit optics, and (4) a large sample compartment. The significant feature of the optical system is a large, well-collimated measurement beam: the angle of convergence is 0.7 for a slit height of 7 mm, and the maximal beam size is 37 X 20 mm. This beam geometry eliminates the need for polarization corrections (using linearly polarized light) or compensation for spatially non-uniform detector responsivity (using averaging sphere). The paper describes the instrument design and presents data from preliminary performance tests. The systematic and random sources of error that have been investigated include: wavelength accuracy and reproducibility, bandpass, beam uniformity, polarization, stray light, system drift and linearity. A new linearity tester has been developed for checking the photometric accuracy. This automated device is based upon the double-aperture method but takes advantage of high precision piezoelectric motors to give a single adjustable aperture. Transmittance measurements of several neutral-density glass filters at 546 nm demonstrate that the photometric precision is better than 0.01% of the measured value and that the photometric accuracy is a few parts in The wavelength scale is accurate to better than ±0.1 nm from 300 to 2500 nm, and is reproducible within ±0.03 nm. A new reference spectrophotometer is being developed at the National Research Council of Canada (NRCC) for high- accuracy transmittance measurements over the spectral range 200 to 2500 nm. This computerized instrument is a single -beam design based upon a servomotor- driven double monochromator with a wavelength resolution of nm. The other main components are: (1) two interchangeable sources (deuterium and tungsten -halogen), (2) two computer -selectable TE- cooled detectors (GaAs photomultiplier tube and PbS cell), (3) all- reflective input and exit optics, and (4) a large sample compartment. The significant feature of the optical system is a large, well -collimated measurement beam: the angle of convergence is 0.7 for a slit height of 7 mm, and the maximal beam size is 37x20 mm. This beam geometry eliminates the need for polarization corrections (using linearly polarized light) or compensation for spatially non -uniform detector responsivity (using averaging sphere). The paper describes the instrument design and presents data from preliminary performance tests. The systematic and random sources of error that have been investigated include: wavelength accuracy and reproducibility, bandpass, beam uniformity, polarization, stray light, system drift and linearity. A new linearity tester has been developed for checking the photometric accuracy. This automated device is based upon the double- aperture method but takes advantage of high precision piezoelectric motors to give a single adjustable aperture. Transmittance measurements of several neutral- density glass filters at 546 nm demonstrate that the photometric precision is better than 0.01% of the measured value and that the photometric accuracy is a few parts in 104. The wavelength scale is accurate to better than ±0.1 nm from 300 to 2500 nm, and is reproducible within ±0.03 nm. 1. INTRODUCTION 1. INTRODUCTION Spectrophotometry involves the ratio measurement of two radiometric quantities. The accuracy of the measurement is defined not by a fundamental standard but by the design and calibration of the measuring instrument, and by the sample itself. However, most commercial spectrophotometers are designed more for ease of operation and reliability rather than the highest accuracy. Their double-beam design and restricted source, sample and detector compartments do not readily permit an investigation of their systematic errors. This limits their measurement uncertainty to typically a few parts in In order to achieve higher accuracy spectrophotometric measurements at the level of 1 part in 104, it is necessary to build a reference instrument which is sufficiently simple in design and versatile in function that all systematic errors can be identified, separated, and accurately characterized. Spectrophotometry involves the ratio measurement of two radiometric quantities. The accuracy of the measurement is defined not by a fundamental standard but by the design and calibration of the measuring instrument, and by the sample itself. However, most commercial spectrophotometers are designed more for ease of operation and reliability rather than the highest accuracy. Their double -beam design and restricted source, sample and detector compartments do not readily permit an investigation of their systematic errors. This limits their measurement uncertainty to typically a few parts in 103. In order to achieve higher accuracy spectrophotometric measurements at the level of 1 part in 104, it is necessary to build a reference instrument which is sufficiently simple in design and versatile in function that all systematic errors can be identified, separated, and accurately characterized. At the NRCC, such a reference Spectrophotometer has been recently designed and constructed to perform regular transmittance measurements over the wavelength range 200 to 2500 nm. Regular transmittance is the undeviated component of transmitted flux, as contrasted to the diffuse component. This mode of measurement is used for transparent, nonscattering samples. However, a modular design approach has been used to allow ready extension to other measurement modes at a future date. This paper presents details of the instrument design, calibration procedure, and current performance capabilities. At the NRCC, such a reference spectrophotometer has been recently designed and constructed to perform regular transmittance measurements over the wavelength range 200 to 2500 nm. Regular transmittance is the undeviated component of transmitted flux, as contrasted to the diffuse component. This mode of measurement is used for transparent, nonscattering samples. However, a modular design approach has been used to allow ready extension to other measurement modes at a future date. This paper presents details of the instrument design, calibration procedure, and current performance capabilities. 2. INSTRUMENT DESIGN 2. INSTRUMENT DESIGN A schematic drawing of the instrument is shown in Pig. 1. The double monochromator is a commercial item, but the rest of the system was developed at NRCC. The main features of the instrument are: A schematic drawing of the instrument is shown in Fig. 1. The double monochromator is a commercial item, but the rest of the system was developed at NRCC. The main features of the instrument are: (1) a single-beam design which allows a clear separation of its various systematic errors; (2) a well-collimated measurement beam which, for normal-incidence transmittance of a plane-parallel sample, unambiguously defines the path length through the sample, and eliminates the effects of instrument polarization and beam geometry errors such as beam displacement and defocussing; and (3) a large sample compartment to accommodate a variety of samples and sampling accessories. The large sample space, in combination with off-axis mirrors in the exit optics and good imaging properties, also reduces the probability of interreflections between the sample and the monochromator and/or detector. (1) a single -beam design which allows a clear separation of its various systematic errors; (2) a well -collimated measurement beam which, for normal- incidence transmittance of a plane- parallel sample, unambiguously defines the path length through the sample, and eliminates the effects of instrument polarization and beam geometry errors such as beam displacement and defocussing; and (3) a large sample compartment to accommodate a variety of samples and sampling accessories. The large sample space, in combination with off -axis mirrors in the exit optics and good imaging properties, also reduces the probability of interreflections between the sample and the monochromator and /or detector. SPIE Vol 1109 Optical Radiation Measurements 11 (1989) / 89 SP/E Vol Optical Radiation Measurements II (1989) / 89
2 Sample Compartment Sample Compartment PbS Cary 14 Monochromator Gary 14 Monochromator He -Ne Loser I Sources 2.1 Sources Figure 1. Optical layout of NRCC Reference Spectrophotometer The UV source ( nm) is a deuterium lamp with a 1 mm circular aperture (Cathodeon D902). It is operated at a current of -300 ma (30W) and its power supply is stable to 0.01 %. The VIS -NIR source ( nm) is a tungsten -halogen lamp (Q6.6A /T4 /1CL), operated under constant current regulation at 6.5A (200W) to %. The coiled -coil filament is 10 mm long by 3 mm wide. 2.2 Input optics The condenser optics comprise a plane mirror, an 80 mm diameter spherical concave mirror (f =125 mm) and a 50 mm diameter cylindrical convex mirror (f = -500 mm) to image the source onto the monochromator entrance slit with a magnification of x2. This arrangement provides adequate illumination of the grating aperture with both the deuterium and tungsten- halogen sources. The overfilling in the case of the tungsten -halogen lamp is desirable in that it minimizes sensitivity to misalignment. The spherical concave mirror is oversize with an aperture stop to prevent the beam from catching the edge of the optics and producing stray light. The purpose of the cylindrical mirror is to correct for astigmatism. All mirrors are coated with aluminum and SiO. A computer -controlled shutter, immediately before the monochromator, is used to determine the dark current signal. There is also provision in the compact condenser optics for adding a light chopper for NIR operation and a polarizing prism for special polarization studies. 2.3 Monochromator Figure 1. Optical layout of NRCC Reference Spectrophotometer The UV source ( nm) is a deuterium lamp with a 1 mm circular aperture (Cathodeon D902). It is operated at a current of -300 ma (30W) and its power supply is stable to 0.01$. The VIS-NIR source ( nm) is a tungsten-halogen lamp (Q6.6A/T4/1CL), operated under constant current regulation at 6.5A (200W) to 0.005%. The coiled-coil filament is 10 mm long by 3 mm wide. 2.2 Input optics The condenser optics comprise a plane mirror, an 80 mm diameter spherical concave mirror (f=125 mm) and a 50 mm diameter cylindrical convex mirror (f=-500 mm) to image the source onto the monochromator entrance slit with a magnification of x2. This arrangement provides adequate illumination of the grating aperture with both the deuterium and tungsten-halogen sources. The overfilling in the case of the tungsten-halogen lamp is desirable in that it minimizes sensitivity to misalignment. The spherical concave mirror is oversize with an aperture stop to prevent the beam from catching the edge of the optics and producing stray light. The purpose of the cylindrical mirror is to correct for astigmatism. All mirrors are coated with aluminum and SiO. A computer-controlled shutter, immediately before the monochromator, is used to determine the dark current signal. There is also provision in the compact condenser optics for adding a light chopper for NIR operation and a polarizing prism for special polarization studies. 2.3 Monochromator The monochromator is a commercial item (Gary Model 14) 1 with a prism predisperser (f=30 cm) for spectral order sorting and a 600 line/mm echelette grating (f=40 cm), which both operate with an aperture ratio of f/8. This monochromator is noteworthy because of its wide spectral range, high resolution, low stray light, and wavelength scale reproducibility. The entrance, intermediate, and exit slits are simultaneously adjustable from to 3-0 mm and the slit height is fixed at either 7 or 20 mm. To reduce the effects of stray light and slit curvature mismatch, a slit height of 7 mm is routinely used. The reciprocal linear dispersion is 3.55 nm/mm at 550 nm, ie. giving a nominal bandpass of 3-55 nm with a 1 mm slit width. The monochromator is a commercial item (Cary Model 14)1 with a prism predisperser (f =30 cm) for spectral order sorting and a 600 line /mm echelette grating (f =40 cm), which both operate with an aperture ratio of f /8. This monochromator is noteworthy because of its wide spectral range, high resolution, low stray light, and wavelength scale reproducibility. The entrance, intermediate, and exit slits are simultaneously adjustable from to 3.0 mm and the slit height is fixed at either 7 or 20 mm. To reduce the effects of stray light and slit curvature mismatch, a slit height of 7 mm is routinely used. The reciprocal linear dispersion is 3.55 nm /mm at 550 nm, ie. giving a nominal bandpass of 3.55 nm with a 1 mm slit width. 90 / SPIE Vol Optical Radiation Measurements II (1989) 90 / SPIE Vol Optical Radiation Measurements II (1989)
3 The monochrornator monochromator sine-bar mechanism is driven by a leadscrew coupled to a servomotor (PMI). The wavelength position is read by an absolute encoder (Litton, 17 bit) which is controlled by two anti-backlash gears with a gear ratio of 3.5:1. The larger gear is mounted on the shaft of the leadscrew and is rotated 360 about every 20 nm. This arrangement gives a wavelength resolution of nm. A coupling plate also mounted on the shaft of the leadscrew provides thermal isolation from the motor drive, thereby minimizing instrumental wavelength shifts due to leadscrew expansion. 2.4 Exit optics The light beam emerging from the monochromator exit slit is diverted by a plane mirror to a 60 mm diameter spherical concave mirror tilted at at (f(f=300 mm). This produces a large, highly-collimated beam. The maximum beam dimensions are 37 mm high by 20 mm wide, and the degree of convergence for a 7 mm monochromator slit height is is A different size and shape of beam irradiating the sample are obtained with a precision limiting aperture. The sample compartment is 25 cm long. The usual sample holder is an automated stepper motor-driven filter wheel with six positions and a resolution of of This can accommodate either 50 mm square or circular filters, from 1 to 10 mm in thickness. A novel automated linearity tester, described in Section 3.6, can also be inserted into the measurement beam. At the far end of the sample compartment, the light beam is diverted by a plane mirror to a second spherical concave mirror (f=300 mm) also tilted at 15 and to a stepper motor-driven plane mirror which automatically selects the appropriate detector during the spectral scan. 2.5 Detectors The UV-VIS detector ( nm) nm) is is a a photomultiplier () with a GaAs cathode (RCA 4832; 12.7x5 mm cathode with 9 circular-cage dynodes) used behind a ground Suprasil diffusing plate (50 mm diameter). Because the monochromator exit slit, not the sample, is imaged onto the detector, and the beam is well-collimated, a diffuser- combination compensates for slight beam displacement due to insertion of imperfect samples. This particular was chosen for its stability, its flat spectral response over a wide wavelength range, and its very low dark emission. To further increase the sensitivity and lower the dark current noise, the is TE-cooled to 258 K. To improve linearity of response, 110 V Zener diodes are placed between the cathode and first dynode and between the last dynode and ground, and the anode current is is taken directly to to a a current-to-voltage converter with a 10 MO Mft feedback resistor. The high voltage power supply is regulated to 0.001?. %. A typical maximum operating signal is 8V (0.8 ya) pa) for a bias voltage of 1000 V. Measurement precision is improved by integrating the signal with a 5 1/2 digit digital voltmeter (DVM). The average of 600 readings corresponds to to a a 1010-sec integration period. For the NIR range ( nm), a lead sulfide (PbS) detector (Quantum Detector Technology, 8x8 mm), operated with a 100V bias voltage, is used. It is TE-cooled to 258 K and the source is chopped at 90 Hz, to reduce the influence of the PbS's high dark level. The modulated signal, across a 1MS2 IMfl load resistor, is amplified by a wide-ranging (10 nv to 500 mv) lock-in amplifier. The lock-in or or the DVM perform analogue or digital signal averaging, respectively, under computer control. 2.6 Automation A block diagram of the signal processing in the NRCC Reference Spectrophotometer is shown in Fig. Pig. 2. The instrument operation is automated by by a a microcomputer (Hewlett-Packard 9817) programmed in BASIC. Software has been written to evaluate the various instrumental systematic errors, and to perform transmittance measurements, data storage on 3 1/2" diskettes, and statistical analysis of results. Measurement parameters that are under software-control include: wavelength range, wavelength step for pause and measure mode, detector settling time before measurement, number of readings, and measurement sequence for dark, 100% 100? and sample readings. SPIE Vol Optical Radiation Measurements II ll (1989) / / 91
4 Thermoelectric Cooler Regulated H.V. Supply UV/VIS Detection Printer Plotter GPIB Current to Voltage Converter Shutter Drive 5 f'z digit DVM GPIB Micro Computer GPIB 17 Bit Abs Encoder Interface Encoder I Wavelength I Drive Wavelength Drive Thermoelectric Cooler Lock -In Amplifier GPIB (1M RAM) Servo Drive Amplifier Bios Voltage Supply Preamplifier NIR Detection GPIB Motor Control Motor Motor Filter Wheel Detector Interchange Mirror Detector Interchange Mirror Figure 2. Chopper Reference Motion GAx s 1 Signal Controller Motor Linearity Tester Axis 2 Linearity Tester Figure 2. Schematic diagram of signal processing in NRCC Reference Spectrophotometer Schematic diagram of signal processing in NRCC Reference Spectrophotometer 3. ERROR EVALUATION In order to determine the accuracy and precision attainable with the new reference spectrophotometer, the following instrument errors were evaluated: wavelength accuracy and reproducibility, bandpass, beam uniformity, polarization, stray light, system drift, noise and non -linearity. The details of this characterization are presented below. 3.1 Wavelength scale and bandpass 3- ERROR EVALUATION In order to determine the accuracy and precision attainable with the new reference spectrophotometer, the following instrument errors were evaluated: wavelength accuracy and reproducibility, bandpass, beam uniformity, polarization, stray light, system drift, noise and non-linearity. The details of this characterization are presented below. 3.1 Wavelength scale and bandpass The wavelength scale was calibrated with 31 atomic lines from Hg, Cd, Cs and He spectrum lamps. The intensity data for each atomic line was recorded at small wavelength steps (0.03 to 0.1 nm) and with a narrow instrument bandpass (<0.3 nm). The positions of the peak maxima were estimated to the nearest 0.01 nm by fitting a triangular line shape to the data. The reproducibility of this wavelength calibration procedure was typically better than ±0.02 nm from nm and better than ±0.03 nm from nm. The worst reproducibility of ±0.055 nm occurred in the region nm. The wavelength calibration data recorded with the detector and with the PbS detector were separately fitted to first degree polynomial functions. These results are shown in Fig. 3- In both cases, the average agreement between calculated and measured wavelength readings is better than ±0.1 nm. The largest discrepancy of 0.20 nm occurs at 250 nm. The wavelength scale was calibrated with 31 atomic lines from Hg, Cd, Cs and He spectrum lamps. The intensity data for each atomic line was recorded at small wavelength steps (0.03 to 0.1 nm) and with a narrow instrument bandpass (<0.3 nm). The positions of the peak maxima were estimated to the nearest 0.01 nm by fitting a triangular line shape to the data. The reproducibility of this wavelength calibration procedure was typically better than ±0.02 nm from nm and better than ±0.03 nm from nm. The worst reproducibility of ±0.055 run occurred in the region nm. The wavelength calibration data recorded with the detector and with the PbS detector were separately fitted to first degree polynomial functions. These results are shown in Fig. 3. In both cases, the average agreement between calculated and measured wavelength readings is better than ±0.1 nm. The largest discrepancy of 0.20 nm occurs at 250 nm. The spectral resolution or bandpass was determined by measuring the full-width at half-maximum height of three Hg atomic lines (253.6, , and nm) as a function of monochromator slit width. The measured bandpass differs from the nominal bandpass (given by the product of the reciprocal linear dispersion and slit width) because of slit curvature mismatch, diffraction, and optical aberrations. However, at larger slit widths (0.1 to 2.0 mm) the relationship between nominal and actual bandpass was linear. At very small slit widths (<0.1 mm), limiting spectral resolutions of 0.2, 0.1 and 0.1 nm were obtained at 253.6, and nm, respectively. The spectral resolution or bandpass was determined by measuring the full -width at half- maximum height of three Hg atomic lines (253.6, , and nm) as a function of monochromator slit width. The measured bandpass differs from the nominal bandpass (given by the product of the reciprocal linear dispersion and slit width) because of slit curvature mismatch, diffraction, and optical aberrations. However, at larger slit widths (0.1 to 2.0 mm) the relationship between nominal and actual bandpass was linear. At very small slit widths (<0.1 mm), limiting spectral resolutions of 0.2, 0.1 and 0.1 nm were obtained at 253.6, and nm, respectively. The temperature dependence of the wavelength scale was checked by monitoring the peak position of the nm Hg line as a function of monochromator temperature. Over a typical temperature range of G, there was no detectable wavelength shift. The temperature dependence of the wavelength scale was checked by monitoring the peak position of the nm Hg line as a function of monochromator temperature. Over a typical temperature range of C, there was no detectable wavelength shift. 92 / SPIE Vol 1109 Optical Radiation Measurements ll (1989) 92 / SPIE Vol Optical Radiation Measurements II (1989)
5 T T T T I I T~ Wavelength Calibration for UV-VIS Range Wavelength Calibration for UV -VIS Range Wavelength Calibration for NIR Range Wavelength Calibration for NIR Range O) c r-» T3 (0 (D GC cn C.,y co a) en c T CO ~ Hg ocd nhe ACs I T Wavelength (nm) Wavelength (nm) Hg QHe ACs I I Wavelength (nm) Wavelength (nm) (a) (a) (b) (b) Figure 3. Least-square fits to wavelength calibration data: ; (b) NIR range with PbS. Figure 3. Least- square fits to wavelength calibration data: (a) UV -VIS range with ; (b) NIR range with PbS. 3.2 Beam uniformity 3.2 Beam uniformity (a) UV-VIS range with The uniformity of the beam in the sample compartment was determined by mapping the signal intensity as a 2.5 mm circular aperture was scanned across the beam in a horizontal and vertical direction. To ensure a uniform detector response, a Si photodiode with an opal glass'diffuser was used. A typical result is shown in Fig. 4. Beam uniformity was checked at other wavelengths and slit width settings and found to exhibit similar profiles. In all cases the uniformity was better in the vertical than horizontal direction with a maximum variation of -10% along a 32 mm beam height. The dip in the middle of the horizontal beam profile is most likely due to some blemish, eg. scratch on the input or monochromator optics. The uniformity of the beam in the sample compartment was determined by mapping the signal intensity as a 2.5 mm circular aperture was scanned across the beam in a horizontal and vertical direction. To ensure a uniform detector response, a Si photodiode with an opal glass'diffuser was used. A typical result is shown in Fig. 4. Beam uniformity was checked at other wavelengths and slit width settings and found to exhibit similar profiles. In all cases, the uniformity was better in the vertical than horizontal direction with a maximum variation of -10% along a 32 mm beam height. The dip in the middle of the horizontal beam profile is most likely due to some blemish, eg. scratch on the input or monochromator optics. 7 (0 O) rt C/D Figure 4. Beam uniformity in sample compartment at 500 nm (slit horizontal, and vertical, o to instrument optical axis, Figure 4. Beam uniformity in sample compartment at 500 nm (slit horizontal, and vertical, o to instrument optical axis r -2 Position (mm) Position (mm) I 10 I SPIE Vol Optical Radiation Measurements II (1989) / 93 SPIE Vol Optical Radiation Measurements II (1989) / 93
6 3.3 Instrument polarization 3.3 Instrument polarization The degree of polarization, 2 P was determined over the spectral range 400 to 800 nm with a linear polarizing filter (Oriel model 27300) in the sample compartment and with a polarization-insensitive detector (Si diode with opal glass diffuser). The value of P varied from.18 to 0.46, depending upon the wavelength. Calculations show that the transmittance error for a nonabsorbing glass filter (n=1.5) due to P=0.50 and an angle of incidence of 0.7 (convergence angle for 7 mm slit height), is less than 2x10~5 transmittance units. Thus, in the visible range, it is not necessary to perform transmittance measurements in linearly polarized light to correct for the effects of instrument polarization. The degree of polarization,2 P was determined over the spectral range 400 to 800 nm with a linear polarizing filter (Oriel model 27300) in the sample compartment and with a polarization- insensitive detector (Si diode with opal glass diffuser). The value of P varied from.18 to 0.46, depending upon the wavelength. Calculations show that the transmittance error for a nonabsorbing glass filter (n =1.5) due to P =0.50 and an angle of incidence of 0.7 (convergence angle for 7 mm slit height), is less than 2x10-5 transmittance units. Thus, in the visible range, it is not necessary to perform transmittance measurements in linearly polarized light to correct for the effects of instrument polarization. 3.4 Stray light 3.4 Stray light The isochromatic stray light error, ie. the transmitted radiation which bypasses the sample, was determined by measuring the transmittance of an opaque aluminum mirror (50 mm square) covering the sample beam. This test should exacerbate the error by introducing additional scattering from interreflections between the monochromator optics and the mirror sample. For a typical 100% level of 8V (0.8 ya), the isochromatic stray light error was a few parts in a million, although dependent on the detector settling time (eg. 3«7X 10~6, 2.6xlO" 6 and 2.2xlO~ 6 for 10, 20, and 40 sec delays, respectively, from 100% reading). The isochromatic stray light error, ie. the transmitted radiation which bypasses the sample, was determined by measuring the transmittance of an opaque aluminum mirror (50 mm square) covering the sample beam. This test should exacerbate the error by introducing additional scattering from interreflections between the monochromator optics and the mirror sample. For a typical 100% level of 8V (0.8 RA), the isochromatic stray light error was a few parts in a million, although dependent on the detector settling time (eg. 3.7x10-6, 2.6x10-6 and 2.2x10-6 for 10, 20, and 40 sec delays, respectively, from 100% reading). The heterochromatic stray light (HSL) error, ie. the transmitted radiation different from the nominal instrument bandpass, was determined by measuring the transmittance through one and two thicknesses of a red filter (Schott RG665) at wavelengths below its cutoff. Since the HSL is a function of the source and detector as well as the monochromator, the sample, and the operating conditions, the results are only valid for the particular conditions of the test. 3 In the case of the Si detector (1 mm slit width), the far-hsl error (at 450, 500 and 550 nm) was found to be 1 to 2 parts in 10 6, but the near-hsl error (at 600 nm) increased to several parts in With the detector (0.15 mm slit width), the far- and near-hsl errors were both 4 to 5 parts in The heterochromatic stray light (HSL) error, ie. the transmitted radiation different from the nominal instrument bandpass, was determined by measuring the transmittance through one and two thicknesses of a red filter (Schott RG665) at wavelengths below its cutoff. Since the HSL is a function of the source and detector as well as the monochromator, the sample, and the operating conditions, the results are only valid for the particular conditions of the test.3 In the case of the Si detector (1 mm slit width), the far -HSL error (at 450, 500 and 550 nm) was found to be 1 to 2 parts in 106, but the near -HSL error (at 600 nm) increased to several parts in 105. With the detector (0.15 mm slit width), the far- and near -HSL errors were both 4 to 5 parts in System drift 3.5 System drift The principle disadvantage of a single-beam instrument design is that fluctuations in the source output and detector response are not dynamically compensated. However, if the system drift exhibits a linear systematic bias, it can be effectively eliminated by performing' the measurements in a time-symmetrical sequence, eg. 100% reading, sample, sample, 100% reading, and averaging the results. To check if this assumption was valid, the system drift was measured for the various source-detector combinations used to cover the range nm. The signal was monitored every minute (10-sec integration) for a period of one hour. In all cases, the data could be satisfactorily fitted to a straight line. The slope gives a measure of the systematic drift and the residual least-squares error gives a measure of the random noise. Typical results for the tungsten-halogen lamp and PbS detector are shown in Fig. 5- The principle disadvantage of a single -beam instrument design is that fluctuations in the source output and detector response are not dynamically compensated. However, if the system drift exhibits a linear systematic bias, it can be effectively eliminated by performing the measurements in a time -symmetrical sequence, eg. 100% reading, sample, sample, 100% reading, and averaging the results. To check if this assumption was valid, the system drift was measured for the various source -detector combinations used to cover the range nm. The signal was monitored every minute (10 -sec integration) for a period of one hour. In all cases, the data could be satisfactorily fitted to a straight line. The slope gives a measure of the systematic drift and the residual least- squares error gives a measure of the random noise. Typical results for the tungsten- halogen lamp and PbS detector are shown in Fig Non-linearity 3.6 Non -linearity Photometric accuracy requires linearity of the photometric scale, ie. that the output quantity of the detector is exactly proportional to the input quantity over a specified range. However, all spectrophotometers exhibit nonlinearity errors. Accuracy is therefore limited by the precise Photometric accuracy requires linearity of the photometric scale, ie. that the output quantity of the detector is exactly proportional to the input quantity over a specified range. However, all spectrophotometers exhibit non - linearity errors. Accuracy is therefore limited by the precise > E 4.2 > Figure 5. I I I I Straight line fit (DRIFT % /min) ^Straight line fit (DRIFT 0.005%/min) Time dv 0.01% Time (minutes) (minutes) > E `-* Figure 5. System drift and noise as a function of time for tungsten-halogen lamp and PbS detector (258 K) at 1300 nm. System drift and noise as a function of time for tungsten -halogen lamp and PbS detector (258 K) at 1300 nm. 94 / SPIE Vol Optical Radiation Measurements ll (1989) 94 / SPIE Vol Optical Radiation Measurements II (1989)
7 evaluation of this error. error. To non-linearity To determine the non -linearity correction, correction, we we built built aa new completely completely automated automated linearity tester4. tester4. The The device device is is aa modified modified version of of the the double-aperture > 6 which double- aperturemethod methodof of light light addition5 addition516 which uses uses aa single single adjustable adjustable aperture aperture instead instead of apertures to to measure measure 2:1 2:1 flux(4)) flux(<j>) ratios. ratios. This of two two fixed apertures is made possible possible by by This procedure is the the large, large, uniform uniform nature nature of of our our sample sample beam beam and and by by the the recent recent availability availability of of high high precision precision piezoelectric piezoelectric (PZT) (PZT) motors motors to to precisely precisely define define the the measurement measurement apertures apertures A, A, B, B, and and A+B A +Bfor forarbitrary arbitrary sizes sizes and and therefore therefore flux levels. levels. This, turn, eliminates eliminates the the need need of This, in turn, of external to characterize characterize different different flux external optical optical attenuation to flux levels. levels. AA photograph of of the new NRCC NRCC linearity linearity tester tester is Is shown shown in in Fig. Fig It It consists of of two two blades and closed closed independently independently by blades which which are are opened opened and by directly directly coupled coupled PZT motors motors driven by aa driven by dedicated controller (Burleigh dedicated programmable controller (Burlelgh Instruments). Instruments). The The motors are clamped clamped to to aa mounting one above above the the other centrally with mounting plate plate and and located located one other and and centrally with respect respect to to the the instrument Instrument optical axis. axis. The The blades blades can can be be adjusted adjusted in in ym pm Increments increments for for aa total total travel of mm travel of each. An each. the drive shaft of An Incremental incremental encoder encoder on the drive shaft of each each motor motor reads an reads the the position to to an accuracy of of 11 um. pm. The The accuracy accuracy and and precision precision of of the the new new linearity linearity tester tester were were checked checked with with aa highly highly linear linear and SI photodiode photodiode (EG (EG&G and stable Si &G UV4444 BQ).7 BQ). 7 At 546 nm, nm, over a photometric range At 546 range of of 100% 100$ to to 10JE (100$ T=4.1 10% TT (100% T =4.1pA), pa), the the diode diode was was found found to to be be linear linear to to 22 parts parts in in 101* 104 with with a repreductibility 5, which reproducibility of of aa few few parts parts In in , which is is consistent consistent with with the the reported reported linearity linearity of of thi s de this detector tec tor in in the the visible v i s Ib 1 e.7.7 The non-linearity factors, 5 aa%%, 9 for, determined The non -linearity correction correction factors,5 for AA == BB == 1/2 1/2 $4), determined for for the the detector at 546 nm nm and and for for aa full full-scale detector at -scale (T-1.0) (T =1.0)anode anode current current of of pa, pa, are are shown shown graphically in In Fig. Pig. 77 versus versus transmittance transmlttancet(4)). T(«f>). A smooth curve curve has has been been drawn drawn through through the the measured had an average reproducibility reproduclblllty of 11 to measured points points which had to 22 parts in ** b ' T Figure 7. J. Figure Figure Photograph of Photograph of adjustable adjustableaperture linearity tester aperture with PZT PZT-controlled -controlled blades (0) Measured non non-linearity error, Measured -linearity error, of versus versus transmittance, transmittance, a% of T(<j> ) for for T(1.0) T( 1.0) =0.8 T(4) =0.8 pa and nm. Data obtained obtained with, XX=546 =546 nm. with, x and x and without, without, o N.D.1.0 filter in in beam. beam. filter TRANSMITTANCE MEASUREMENTS the preceding error evaluation, evaluation, the the combined combined effects From the effects of of wavelength wavelength scale scale error, error, beam non-uniformity, non -uniformity, instrument instrument polarization, polarization, and and stray stray light, light, should should contribute contribute less part less than than 11 part in 104 I0 k to the total measurement uncertainty. uncertainty. Therefore, Therefore, no no correction correction is is applied applied for for these these errors. However, system specific errors. system drift, drift, noise, noise, and -linearity are and non non-linearity are significant. significant. The The procedures for correcting for procedures for these these latter latter errors errors are outlined below. below. SPIE Vol. SPIE Vol Optical OpticalRadiation RadiationMeasurements MeasurementsII Il/19891 (1989)/ / 95 95
8 4.1 Measurement procedure 4.1 Measurement procedure In order to remove the influence of linear drift on the results, the measurements are performed in a time-symmetrical sequence. For example, for a four-filter measurement on the automated 6-position filter wheel, positions 1 and 4 are used as clear spaces, and the following series of readings are taken: In order to remove the influence of linear drift on the results, the measurements are performed in a time -symmetrical sequence. For example, for a four -filter measurement on the automated 6- position filter wheel, positions 1 and 4 are used as clear spaces, and the following series of readings are taken: Pass 1 : 0, Fo, F1, F2, 0, Fo, F3, Fy, Fo, 0 Pass 2 : F0, 0, F4, F3, Fo, 0, F2, F1, FD, 0 where Pass 1 involves rotation of the filter wheel 360 clockwise and Pass 2 involves rotation 360 counter -clockwise; 0, Fo, F1, F9, F,, and F4 are signals for the dark current, clear space, filter 1, filter 2, filter 3, and filter 4, respectively. The data is processed by taking the average value of 0 and Fo on either side of an individual filter reading, F (j =1 to 4), giving 0 and Fo, then computing the filter transmittance, Ti according to: T = j Fo -0 The corresponding filter transmittance values for the two passes are then averaged. 4.2 Reproducibility The effectiveness of this time -symmetrical measurement procedure was evaluated by repeat transmission measurements on four neutral- density glass filters of nominal transmittances 10, 25, 50 and 70% (Schott NG4 and NG11, 50x50 mm, 1-5 mm thick). These results are reported in Table 1. The measurements were performed at 546 nm with a spectral bandpass of 0.5 nm and a temperature of 23 ±1 C. The filters were aligned normal to the beam and measured with the maximum beam size of 37x20 mm. Four independent measurements were performed on each filter, where each run represents the average transmittance of the two -pass measurement sequence described in Section 4.1 with a 10 -sec integration interval for each individual reading. The average % relative standard deviation for the four filters is %. Table 1. Reproducibility of Transmittance Measurements on Four Glass Filters at 546 nm Run 1 Run 2 Run 3 Run 4 Pass 1 Average % Standard Deviation Pass 2 0, P., F,, P n, 0, F9, P., Pn, where Pass 1 involves rotation of the filter wheel 360 clockwise and Pass 2 involves rotation 360 counter-clockwise; 0, FQ, F X, F^, F-, and F^ are signals for the dark current, clear space, filter 1, filter 2, filter 3> and filter 4, respectively. The data is processed by taking the average value of 0 and FQ on either side of an individual filter reading, Pj(J=l to 4), giving 0 and FQ, then computing the filter transmittance, T according to: The corresponding filter transmittance values for the two passes are then averaged. 4.2 Reproducibility The effectiveness of this time-symmetrical measurement procedure was evaluated by repeat transmission measurements on four neutral-density glass filters of nominal transmittances 10, 25, 50 and 70$ (Schott NG4 and NG11, 50x50 mm, 1-5 mm thick). These results are reported in Table 1. The measurements were performed at 546 nm with a spectral bandpass of 0.5 nm and a temperature of 23 ± 1 C. The filters were aligned normal to the beam and measured with the maximum beam size of 37x20 mm. Pour Independent measurements were performed on each filter, where each run represents the average transmittance of the two-pass measurement sequence described in Section 4.1 with a 10-sec integration interval for each individual reading. The average % relative standard deviation for the four filters is 0.006$. Table 1. Reproducibility of Transmittance Measurements on Pour Glass Filters at 546 nm Run 1 Run 2 Run 3 Run 4 Average % Standard Deviation Nominal % Transmittance Nominal % Transmittance Non-linearity error correction 4.3 Non -linearity error correction The transmittance error correction, AT which must be added to the measured apparent transmittance, T was computed from the non-linearity error correction factors, a% in Pig. 7 according to the procedure of Clarke. 6 This yielded the smooth curve shown in Pig. 8. A maximum correction value of 3.64xlO~ 3 transmittance units occurs at a transmittance level of 40$. The transmittance error correction, AT which must be added to the measured apparent transmittance, T was computed from the non -linearity error correction factors, a% in Fig. 7 according to the procedure of Clarke.6 This yielded the smooth curve shown in Fig. 8. A maximum correction value of 3.64x10-3 transmittance units occurs at a transmittance level of 40 %. 4.4 Results 4.4 Results The accuracy of the non-linearity correction curve, AT(<j>) derived for the (Fig. 8), was independently checked by comparison with transmittance measurements using a highly linear Si diode (EG&G UV444 BQ with ground Suprasil diffuser). Six neutral-density glass filters (combinations of Schott NG4 and NG11), ranging from 10 to 70$ transmittance, and 1 to 5 mm in thickness, were measured on the reference spectrophotometer using the and the Si diode. In both cases, the measurements were performed at 546 nm with a 0.5 nm spectral bandpass, and a beam size of 37x20 mm. The was operated at the conditions corresponding The accuracy of the non -linearity correction curve, AT($) derived for the (Fig. 8), was independently checked by comparison with transmittance measurements using a highly linear Si diode (EG &G UV444 BQ with ground Suprasil diffuser). Six neutral- density glass filters (combinations of Schott NG4 and NG11), ranging from 10 to 70% transmittance, and 1 to 5 mm in thickness, were measured on the reference spectrophotometer using the and the Si diode. In both cases, the measurements were performed at 546 nm with a 0.5 nm spectral bandpass, and a beam size of 37x20 mm. The was operated at the conditions corresponding 96 / SPIE Vol Optical Radiation Measurements Il (1989) 96 / SPIE Vol Optical Radiation Measurements II (1989)
9 I I I I I I I r Figure 8. Transmittance correction, AT computed from non-linearity data of Pig. 7. Figure 8. Transmittance correction, AT computed from non -linearity data of Fig to Pig. 7. These results are summarized in Table 2, where both the raw and linearity-corrected are included. The average agreement between the Si diode and linearity-corrected values for the six filters is 2.5 parts in This low overall figure indicates that the new adjustable aperture linearity tester and the -diffuser combination provide reliable and accurate results. to Fig. 7. These results are summarized in Table 2, where both the raw and linearity- corrected are included. The average agreement between the Si diode and linearity- corrected values for the six filters is 2.5 parts in 104. This low overall figure indicates that the new adjustable aperture linearity tester and the -diffuser combination provide reliable and accurate results. Table 2. Comparison of Percent Transmittance (W) for Six Glass Filters at 546 run Using Si diode and detectors %T Si diode (measured) (corrected)* Difference Table 2. Comparison of Percent Transmittance ( %T) for Six Glass Filters at 546 nm Using Si diode and detectors %T Si diode (measured) (corrected)* Difference ^corrected for non-linearity transmittance error, AT, in-fig. 8 *corrected for non -linearity transmittance error, AT, in Fig CONCLUSIONS 5. CONCLUSIONS A new high-accuracy spectrophotometer for performing regular transmittance measurements over the nm spectral range has been developed. This automated single-beam instrument possesses a highly-collimated beam geometry which eliminates the need of polarization correction and of an averaging sphere. Error evaluation is well-advanced, particularly in the visible region, where the overall photometric accuracy is estimated to be a few parts in lo 4, with an average reproducibility of 0.006$ of the value for four repeat measurements. The wavelength scale is accurate to better than ±0.1 nm from 300 to 2500 nm, ±0.2 nm from 200 to 300 nm, and is reproducible within ±0.03 nm. A new high- accuracy spectrophotometer for performing regular transmittance measurements over the nm spectral range has been developed. This automated single -beam instrument possesses a highly -collimated beam geometry which eliminates the need of polarization correction and of an averaging sphere. Error evaluation is well- advanced, particularly in the visible region, where the overall photometric accuracy is estimated to be a few parts in 104, with an average reproducibility of 0.006% of the value for four repeat measurements. The wavelength scale is accurate to better than ±0.1 nm from 300 to 2500 nm, ±0.2 nm from 200 to 300 nm, and is reproducible within ±0.03 nm. A new automated linearity tester with an adjustable, precisely-defined aperture has been designed and built. Its reliability and accuracy have been checked with a highly linear Si diode. In the visible, the largest systematic error limiting the accuracy of the reference spectrophotometer is non-linearity of the which is of the order of several parts in 10 3 for a 1 11A photocurrent. A new automated linearity tester with an adjustable, precisely- defined aperture has been designed and built. Its reliability and accuracy have been checked with a highly linear Si diode. In the visible, the largest systematic error limiting the accuracy of the reference spectrophotometer is non -linearity of the which is of the order of several parts in 103 for a 1 ua photocurrent. Further work is underway to refine the linearity tester design and to select a more linear that will provide a measurement accuracy of better than 1 part in 10 4 transmittance units in the VIS range. In the UV and NIR, error evaluation is continuing. The extension of the instrument's measurement capabilities to diffuse reflectance in the UV-VIS range has also begun. Further work is underway to refine the linearity tester design and to select a more linear that will provide a measurement accuracy of better than 1 part in 104 transmittance units in the VIS range. In the UV and NIR, error evaluation is continuing. The extension of the instrument's measurement capabilities to diffuse reflectance in the UV -VIS range has also begun. 6. ACKNOWLEDGEMENTS 6. ACKNOWLEDGEMENTS The authors wish to thank Dr. I. Powell for helpful discussions and for the design of the input and exit optics. We also wish to acknowledge the contributions of M. Kottler to the design of several of the mechanical components of the system. The authors wish to thank Dr. I. Powell for helpful discussions and for the design of the input and exit optics. We also wish to acknowledge the contributions of M. Kottler to the design of several of the mechanical components of the system. SPIE Vol Optical Radiation Measurements ll (1989) / 97 SPIE Vol Optical Radiation Measurements II (1989) / 97
Spectrophotometer. An instrument used to make absorbance, transmittance or emission measurements is known as a spectrophotometer :
Spectrophotometer An instrument used to make absorbance, transmittance or emission measurements is known as a spectrophotometer : Spectrophotometer components Excitation sources Deuterium Lamp Tungsten
More informationMS260i 1/4 M IMAGING SPECTROGRAPHS
MS260i 1/4 M IMAGING SPECTROGRAPHS ENTRANCE EXIT MS260i Spectrograph with 3 Track Fiber on input and InstaSpec IV CCD on output. Fig. 1 OPTICAL CONFIGURATION High resolution Up to three gratings, with
More informationMeasuring optical filters
Measuring optical filters Application Note Author Don Anderson and Michelle Archard Agilent Technologies, Inc. Mulgrave, Victoria 3170, Australia Introduction Bandpass filters are used to isolate a narrow
More informationLambda 1050 / 950 UV/Vis/NIR
www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Lambda 1050 / 950 UV/Vis/NIR LAMBDA 1050 Choose the LAMBDA 1050 with its triple
More informationChemistry Instrumental Analysis Lecture 10. Chem 4631
Chemistry 4631 Instrumental Analysis Lecture 10 Types of Instrumentation Single beam Double beam in space Double beam in time Multichannel Speciality Types of Instrumentation Single beam Requires stable
More informationChemistry Instrumental Analysis Lecture 7. Chem 4631
Chemistry 4631 Instrumental Analysis Lecture 7 UV to IR Components of Optical Basic components of spectroscopic instruments: stable source of radiant energy transparent container to hold sample device
More informationAgilent Cary 7000 Universal Measurement Spectrophotometer (UMS)
Agilent Cary 7000 Universal Measurement Spectrophotometer (UMS) Specifications Introduction The Agilent Cary 7000 Universal Measurement Spectrophotometer (UMS) is designed for superior performance, flexibility
More informationAutomated high precision variable aperture for spectrophotometer linearity testing
Automated high precision variable aperture for spectrophotometer linearity testing Joanne C. Zwinkels and D. S. Gignac A new automated linearity tester with a single variable aperture has been designed
More informationPerformance Comparison of Spectrometers Featuring On-Axis and Off-Axis Grating Rotation
Performance Comparison of Spectrometers Featuring On-Axis and Off-Axis Rotation By: Michael Case and Roy Grayzel, Acton Research Corporation Introduction The majority of modern spectrographs and scanning
More informationGCMS-3 GONIOSPECTROPHOTOMETER SYSTEM
MURAKAMI Color Research Laboratory 11-3 Kachidoki 3-Chome Chuo-Ku Tokyo 104 Japan Tel: +81 3 3532 3011 Fax: +81 3 3532 2056 GCMS-3 GONIOSPECTROPHOTOMETER SYSTEM GSP-1 Main System Overview The colour and
More informationApplication Note (A11)
Application Note (A11) Slit and Aperture Selection in Spectroradiometry REVISION: C August 2013 Gooch & Housego 4632 36 th Street, Orlando, FL 32811 Tel: 1 407 422 3171 Fax: 1 407 648 5412 Email: sales@goochandhousego.com
More informationINTERNATIONAL INTERCOMPARISON OF WAVELENGTH SCALE AND PHOTOMETRIC SCALE OF SPECTROPHOTOMETRY LABORATORIES CENAM - NRC - INMETRO - NIST.
INTERNATIONAL INTERCOMPARISON OF WAVELENGTH SCALE AND PHOTOMETRIC SCALE OF SPECTROPHOTOMETRY LABORATORIES - - - Arquímedes Ruiz 1, Joanne Zwinkels 2, Iakira Bougleux 3 and Yvonne Barnes 4 1 Optical and
More informationCHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES
CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES The current multiplication mechanism offered by dynodes makes photomultiplier tubes ideal for low-light-level measurement. As explained earlier, there
More informationGuide to SPEX Optical Spectrometer
Guide to SPEX Optical Spectrometer GENERAL DESCRIPTION A spectrometer is a device for analyzing an input light beam into its constituent wavelengths. The SPEX model 1704 spectrometer covers a range from
More informationOriel MS260i TM 1/4 m Imaging Spectrograph
Oriel MS260i TM 1/4 m Imaging Spectrograph MS260i Spectrograph with 3 Track Fiber on input and InstaSpec CCD on output. The MS260i 1 4 m Imaging Spectrographs are economical, fully automated, multi-grating
More information1/8 m GRATING MONOCHROMATOR
1/8 m GRATING GRATING OUTPUT PORT INPUT PORT 77250 1/8 m Monochromator with 6025 Hg(Ar) Spectral Calibration Lamp. Low cost, compact size and high performance, ideal for OEM applications Very efficient
More informationOriel Cornerstone 130 1/8 m Monochromator
1 Oriel Cornerstone 130 1/8 m Monochromator Cornerstone 130 1/8 m Monochromator The Cornerstone 130 family of Oriel Monochromators supports two gratings simultaneously, which can be easily interchanged,
More informationSpectroscopy in the UV and Visible: Instrumentation. Spectroscopy in the UV and Visible: Instrumentation
Spectroscopy in the UV and Visible: Instrumentation Typical UV-VIS instrument 1 Source - Disperser Sample (Blank) Detector Readout Monitor the relative response of the sample signal to the blank Transmittance
More informationUV-VIS-IR Spectral Responsivity Measurement System for Solar Cells
November 1998 NREL/CP-52-25654 UV-VIS-IR Spectral Responsivity Measurement System for Solar Cells H. Field Presented at the National Center for Photovoltaics Program Review Meeting, September 8 11, 1998,
More informationComponents of Optical Instruments. Chapter 7_III UV, Visible and IR Instruments
Components of Optical Instruments Chapter 7_III UV, Visible and IR Instruments 1 Grating Monochromators Principle of operation: Diffraction Diffraction sources: grooves on a reflecting surface Fabrication:
More informationCONFIGURING. Your Spectroscopy System For PEAK PERFORMANCE. A guide to selecting the best Spectrometers, Sources, and Detectors for your application
CONFIGURING Your Spectroscopy System For PEAK PERFORMANCE A guide to selecting the best Spectrometers, s, and s for your application Spectral Measurement System Spectral Measurement System Spectrograph
More informationCHAPTER 7. Components of Optical Instruments
CHAPTER 7 Components of Optical Instruments From: Principles of Instrumental Analysis, 6 th Edition, Holler, Skoog and Crouch. CMY 383 Dr Tim Laurens NB Optical in this case refers not only to the visible
More informationSCCH 4: 211: 2015 SCCH
SCCH 211: Analytical Chemistry I Analytical Techniques Based on Optical Spectroscopy Atitaya Siripinyanond Office Room: C218B Email: atitaya.sir@mahidol.ac.th Course Details October 19 November 30 Topic
More informationCornerstone 260 1/4 m Monochromators
Cornerstone /4 m Monochromators The Oriel Cornerstone is a high performance, economical and user-friendly monochromator an ideal instrument for research and OEM applications. Oriel has made it easy to
More informationPerformance Comparisons of InGaAs, extended InGaAs, and Short-wave HgCdTe Detectors between 1 µm and 2.5 µm
Performance Comparisons of, extended, and Short-wave HgCdTe Detectors between 1 µm and 2.5 µm Howard W. Yoon, Matt C. Dopkiss, and George P. Eppeldauer Optical Technology Division National Institute of
More informationPolarization Experiments Using Jones Calculus
Polarization Experiments Using Jones Calculus Reference http://chaos.swarthmore.edu/courses/physics50_2008/p50_optics/04_polariz_matrices.pdf Theory In Jones calculus, the polarization state of light is
More informationSpectral Analysis of the LUND/DMI Earthshine Telescope and Filters
Spectral Analysis of the LUND/DMI Earthshine Telescope and Filters 12 August 2011-08-12 Ahmad Darudi & Rodrigo Badínez A1 1. Spectral Analysis of the telescope and Filters This section reports the characterization
More informationThe FTNIR Myths... Misinformation or Truth
The FTNIR Myths... Misinformation or Truth Recently we have heard from potential customers that they have been told that FTNIR instruments are inferior to dispersive or monochromator based NIR instruments.
More informationMeasuring photometric accuracy using the double aperture method
Measuring photometric accuracy using the double aperture method Application Note Author Robert Francis Agilent Technologies, Inc. Mulgrave, Victoria 3170, Australia. Introduction Photometric accuracy is
More informationIntegrating Spheres. Why an Integrating Sphere? High Reflectance. How Do Integrating Spheres Work? High Damage Threshold
1354 MINIS Oriel Integrating Spheres Integrating spheres are ideal optical diffusers; they are used for radiometric measurements where uniform illumination or angular collection is essential, for reflectance
More informationUnique Scattering Measurements Using the Agilent Universal Measurement Accessory (UMA)
Click to edit Master title style Unique Scattering Measurements Using the Agilent Universal Measurement Accessory (UMA) mark.fisher@agilent.com Click to edit Master title style Rapid, Automated, Quality
More informationDETECTORS Important characteristics: 1) Wavelength response 2) Quantum response how light is detected 3) Sensitivity 4) Frequency of response
DETECTORS Important characteristics: 1) Wavelength response 2) Quantum response how light is detected 3) Sensitivity 4) Frequency of response (response time) 5) Stability 6) Cost 7) convenience Photoelectric
More informationAutomated Spectrophotometric Spatial Profiling of Coated Optical Wafers
Automated Spectrophotometric Spatial Profiling of Coated Optical Wafers Application note Materials testing and research Authors Travis Burt Fabian Zieschang Agilent Technologies, Inc. Parts of this work
More informationAutotracker III. Applications...
Autotracker III Harmonic Generation System Model AT-III Applications... Automatic Second Harmonic and Third Harmonic Generation of UV Wavelengths Automatic Production of IR Wavelengths by Difference Frequency
More informationRadial Polarization Converter With LC Driver USER MANUAL
ARCoptix Radial Polarization Converter With LC Driver USER MANUAL Arcoptix S.A Ch. Trois-portes 18 2000 Neuchâtel Switzerland Mail: info@arcoptix.com Tel: ++41 32 731 04 66 Principle of the radial polarization
More informationSpectro p photomete p r V-700 series
Spectrophotometer p V-700 series V-700 Series UV-Vis/NIR Spectrophotometers V-730 SBW=1.0 nm Class-leading high S/N V-730BIO New irm & Spectra Manager V-750/760/770 Wavelength-independent dynamic range
More informationAutomated Double Aperture Accessory
For the Cary 1, 3, 100, 300, 4, 5, 400, 500, 500i, 4000, 5000, 6000i, Deep UV Installation Category II Pollution Degree 2 Equipment Class I Table of Contents Introduction Theory Operation Installation
More informationHistorical. McPherson 15 Mount
McPherson 15 Mount Normal incidence designs include the McPherson 15 (classical 1.0 meter focal length) and modern NIM units. The latter features smaller included angles, longer focal lengths (e.g. 3,
More informationLINEARPYROMETER LP4. Technical Documentation KE November TN
1 LINEARPYROMETER LP4 Technical Documentation KE 256-6.2007 November 2010 5-TN-1622-100 2 1. General Description With the Linearpyrometer Type LP4 a measuring instrument has been made available for pyrometric
More informationThe equipment used share any common features regardless of the! being measured. Electronic detection was not always available.
The equipment used share any common features regardless of the! being measured. Each will have a light source sample cell! selector We ll now look at various equipment types. Electronic detection was not
More informationDevelopement of the method for realization of spectral irradiance scale featuring system of
Home Search Collections Journals About Contact us My IOPscience Developement of the method for realization of spectral irradiance scale featuring system of spectral comparisons This article has been downloaded
More informationLaser Speckle Reducer LSR-3000 Series
Datasheet: LSR-3000 Series Update: 06.08.2012 Copyright 2012 Optotune Laser Speckle Reducer LSR-3000 Series Speckle noise from a laser-based system is reduced by dynamically diffusing the laser beam. A
More informationComponents of Optical Instruments
Components of Optical Instruments General Design of Optical Instruments Sources of Radiation Wavelength Selectors (Filters, Monochromators, Interferometers) Sample Containers Radiation Transducers (Detectors)
More informationCHAPTER 5 FINE-TUNING OF AN ECDL WITH AN INTRACAVITY LIQUID CRYSTAL ELEMENT
CHAPTER 5 FINE-TUNING OF AN ECDL WITH AN INTRACAVITY LIQUID CRYSTAL ELEMENT In this chapter, the experimental results for fine-tuning of the laser wavelength with an intracavity liquid crystal element
More informationBasic Components of Spectroscopic. Instrumentation
Basic Components of Spectroscopic Ahmad Aqel Ifseisi Assistant Professor of Analytical Chemistry College of Science, Department of Chemistry King Saud University P.O. Box 2455 Riyadh 11451 Saudi Arabia
More informationProperties of Structured Light
Properties of Structured Light Gaussian Beams Structured light sources using lasers as the illumination source are governed by theories of Gaussian beams. Unlike incoherent sources, coherent laser sources
More informationCornerstone 130 1/8 m Monochromators
Cornerstone 130 1/8 m Monochromators The Oriel Cornerstone 130 is a high performance, economical and user-friendly monochromator an ideal instrument for research and OEM applications. Oriel has made it
More informationInstructions for the Experiment
Instructions for the Experiment Excitonic States in Atomically Thin Semiconductors 1. Introduction Alongside with electrical measurements, optical measurements are an indispensable tool for the study of
More informationLAMBDA 800 UV/VIS and 900 UV/VIS/NIR Spectrophotometer Systems
Inorganic Analysis Chromatography Molecular Spectroscopy Thermal / Elemental Analysis Informatics UV-Vis Spectroscopy LAMBDA 800 UV/VIS and 900 UV/VIS/NIR Spectrophotometer Systems customer satisfaction
More information06SurfaceQuality.nb Optics James C. Wyant (2012) 1
06SurfaceQuality.nb Optics 513 - James C. Wyant (2012) 1 Surface Quality SQ-1 a) How is surface profile data obtained using the FECO interferometer? Your explanation should include diagrams with the appropriate
More informationSP-8001 UV/Visible Spectrophotometer
SP-8001 UV/Visible Spectrophotometer Provide Accurate Results High performance UV/Visible Spectrophotometer for Research, and Laboratory work Product Features The Metertech SP-8001 UV/Visible Spectrophotometer
More informationUV / VIS Spectrophotometer EMCLAB Instruments GmbH
A Professional Manufacturer UV / VIS Spectrophotometer EMCLAB Instruments GmbH EMCLAB Instruments GmbH Bismarckstrasse 120 47057 Duisburg Germany Phone: +49 203 3064042 Fax: +49 203 3064044 E-mail: info@emc-lab.de
More informationOL Series 426 Low Light Level lntegrating Sphere Calibration Standards
OL Series 426 Low Light Level lntegrating Sphere Calibration Standards GENERAL The OL Series 426 Low-Light-Level Calibration Standard is designed for accurately calibrating very sensitive microphotometers,
More informationMeasurement Method of High Absorbance (Low Transmittance) Samples by UH4150 INTRODUCTION
INTRODUCTION With UH4150, a detector can be selected depending on the analysis purpose. When analyzing a solid sample which doesn t contain any diffuse components, by selecting the direct light detector,
More informationECEN. Spectroscopy. Lab 8. copy. constituents HOMEWORK PR. Figure. 1. Layout of. of the
ECEN 4606 Lab 8 Spectroscopy SUMMARY: ROBLEM 1: Pedrotti 3 12-10. In this lab, you will design, build and test an optical spectrum analyzer and use it for both absorption and emission spectroscopy. The
More informationStarBright XLT Optical Coatings
StarBright XLT Optical Coatings StarBright XLT is Celestron s revolutionary optical coating system that outperforms any other coating in the commercial telescope market. Our most popular Schmidt-Cassegrain
More informationSPECTRAL SCANNER. Recycling
SPECTRAL SCANNER The Spectral Scanner, produced on an original project of DV s.r.l., is an instrument to acquire with extreme simplicity the spectral distribution of the different wavelengths (spectral
More informationSystems & Accessories
Light Source Sample Chambers Stepping Motor Controller Detector Variable Wavelength Fiber Optics Modules Tunable Light Sources For Applications In: Analytical Chemistry Physics Life Sciences Engineering
More informationPulsed Laser Power Measurement Systems
Pulsed Laser Power Measurement Systems Accurate, reproducible method of determining total laser and laser diode power Ideal for Beam Power Measurement Labsphere s Pulsed Laser Power Measurement Systems
More informationARCoptix. Radial Polarization Converter. Arcoptix S.A Ch. Trois-portes Neuchâtel Switzerland Mail: Tel:
ARCoptix Radial Polarization Converter Arcoptix S.A Ch. Trois-portes 18 2000 Neuchâtel Switzerland Mail: info@arcoptix.com Tel: ++41 32 731 04 66 Radially and azimuthally polarized beams generated by Liquid
More informationIntroduction to the operating principles of the HyperFine spectrometer
Introduction to the operating principles of the HyperFine spectrometer LightMachinery Inc., 80 Colonnade Road North, Ottawa ON Canada A spectrometer is an optical instrument designed to split light into
More informationADVANCED OPTICS LAB -ECEN Basic Skills Lab
ADVANCED OPTICS LAB -ECEN 5606 Basic Skills Lab Dr. Steve Cundiff and Edward McKenna, 1/15/04 Revised KW 1/15/06, 1/8/10 Revised CC and RZ 01/17/14 The goal of this lab is to provide you with practice
More informationQE65000 Spectrometer. Scientific-Grade Spectroscopy in a Small Footprint. now with. Spectrometers
QE65000 Spectrometer Scientific-Grade Spectroscopy in a Small Footprint QE65000 The QE65000 Spectrometer is the most sensitive spectrometer we ve developed. Its Hamamatsu FFT-CCD detector provides 90%
More informationUniversity of Wisconsin Chemistry 524 Spectroscopic Components *
University of Wisconsin Chemistry 524 Spectroscopic Components * In journal articles, presentations, and textbooks, chemical instruments are often represented as block diagrams. These block diagrams highlight
More informationChemistry 524--"Hour Exam"--Keiderling Mar. 19, pm SES
Chemistry 524--"Hour Exam"--Keiderling Mar. 19, 2013 -- 2-4 pm -- 170 SES Please answer all questions in the answer book provided. Calculators, rulers, pens and pencils permitted. No open books allowed.
More informationObservational Astronomy
Observational Astronomy Instruments The telescope- instruments combination forms a tightly coupled system: Telescope = collecting photons and forming an image Instruments = registering and analyzing the
More informationThe designs for a high resolution Czerny-Turner spectrometer are presented. The results of optical
ARTICLE High Resolution Multi-grating Spectrometer Controlled by an Arduino Karl Haebler, Anson Lau, Jackson Qiu, Michal Bajcsy University of Waterloo, Waterloo, Ontario, Canada Abstract The designs for
More informationOptoliner NV. Calibration Standard for Sighting & Imaging Devices West San Bernardino Road West Covina, California 91790
Calibration Standard for Sighting & Imaging Devices 2223 West San Bernardino Road West Covina, California 91790 Phone: (626) 962-5181 Fax: (626) 962-5188 www.davidsonoptronics.com sales@davidsonoptronics.com
More informationDevices & Services Company
Devices & Services Company 10290 Monroe Drive, Suite 202 - Dallas, Texas 75229 USA - Tel. 214-902-8337 - Fax 214-902-8303 Web: www.devicesandservices.com Email: sales@devicesandservices.com D&S Technical
More informationUSB. Part No Wavelength range. Spectral bandwidth 5 nm 4 nm Optical system
Visible range spectrophotometers V-1100 and VR-2000 V-1100 MODEL WITH MANUAL WAVELENGTH SETTINGS AND AUTOMATIC BLANK. VR-2000 MODEL WITH AUTOMATIC WAVELENGTH SETTINGS AND BLANK. V-1100 Part no. 4120025
More informationMicroscope-Spectrometer
20 Micro-spectrometer ToupTek s spectrometer is applicable for spectral detection within the wavelength range between 200nm and 1100nm. Due to their high stability and performance, these portable instruments
More informationR. J. Jones Optical Sciences OPTI 511L Fall 2017
R. J. Jones Optical Sciences OPTI 511L Fall 2017 Semiconductor Lasers (2 weeks) Semiconductor (diode) lasers are by far the most widely used lasers today. Their small size and properties of the light output
More informationUpgrade of the ultra-small-angle scattering (USAXS) beamline BW4
Upgrade of the ultra-small-angle scattering (USAXS) beamline BW4 S.V. Roth, R. Döhrmann, M. Dommach, I. Kröger, T. Schubert, R. Gehrke Definition of the upgrade The wiggler beamline BW4 is dedicated to
More informationSPECTRAL IRRADIANCE DATA
The radiometric data on the following pages was measured in our Standards Laboratory. The wavelength calibrations are based on our spectral calibration lamps. Irradiance data from 250 to 2500 nm is based
More informationPowerful DMD-based light sources with a high throughput virtual slit Arsen R. Hajian* a, Ed Gooding a, Thomas Gunn a, Steven Bradbury a
Powerful DMD-based light sources with a high throughput virtual slit Arsen R. Hajian* a, Ed Gooding a, Thomas Gunn a, Steven Bradbury a a Hindsight Imaging Inc., 233 Harvard St. #316, Brookline MA 02446
More informationSPECTRORADIOMETRY METHODS: A GUIDE TO PHOTOMETRY AND VISIBLE SPECTRORADIOMETRY
SPECTRORADIOMETRY METHODS: A GUIDE TO PHOTOMETRY AND VISIBLE SPECTRORADIOMETRY Written By William E. Schneider, Richard Young, Ph.D 1. SPECTRORADIOMETRY METHODS 1.1. Spectroradiometrics vs Photometric
More informationFully Portable Spectrophotometer for Transmission Measurement
November 2008 LCRT-2005-S Portable Spectrophotometer for Transmission Measurement Spectral Transmission in d/0 and 0/0 Measurement Geometries Transparency Measurement in real in-line Set-up Photometric
More informationLaser Beam Analysis Using Image Processing
Journal of Computer Science 2 (): 09-3, 2006 ISSN 549-3636 Science Publications, 2006 Laser Beam Analysis Using Image Processing Yas A. Alsultanny Computer Science Department, Amman Arab University for
More informationExperiment 1: Fraunhofer Diffraction of Light by a Single Slit
Experiment 1: Fraunhofer Diffraction of Light by a Single Slit Purpose 1. To understand the theory of Fraunhofer diffraction of light at a single slit and at a circular aperture; 2. To learn how to measure
More informationLaser Telemetric System (Metrology)
Laser Telemetric System (Metrology) Laser telemetric system is a non-contact gauge that measures with a collimated laser beam (Refer Fig. 10.26). It measure at the rate of 150 scans per second. It basically
More informationGEOMETRICAL OPTICS Practical 1. Part I. BASIC ELEMENTS AND METHODS FOR CHARACTERIZATION OF OPTICAL SYSTEMS
GEOMETRICAL OPTICS Practical 1. Part I. BASIC ELEMENTS AND METHODS FOR CHARACTERIZATION OF OPTICAL SYSTEMS Equipment and accessories: an optical bench with a scale, an incandescent lamp, matte, a set of
More informationYOUNGS MODULUS BY UNIFORM & NON UNIFORM BENDING OF A BEAM
YOUNGS MODULUS BY UNIFORM & NON UNIFORM BENDING OF A BEAM RECTANGULAR BEAM PLACED OVER TWO KNIFE EDGES & DISTANCE BETWEEN KNIFE EDGES IS KEPT CONSTANT AS l= 50cm UNIFORM WEIGHT HANGERS ARE SUSPENDED WITH
More informationCharacteristics of point-focus Simultaneous Spatial and temporal Focusing (SSTF) as a two-photon excited fluorescence microscopy
Characteristics of point-focus Simultaneous Spatial and temporal Focusing (SSTF) as a two-photon excited fluorescence microscopy Qiyuan Song (M2) and Aoi Nakamura (B4) Abstracts: We theoretically and experimentally
More informationLI-193 Spherical Quantum Sensor
LI-193 Spherical Quantum Sensor The LI-193 Spherical Quantum Sensor measures PAR in air or underwater from all directions at depths up to 350 meters. This sensor is useful for studies of phytoplankton,
More informationPage 1 BIOIMAGER. Spectrophotometers. Toll Free: (855) BIO- IMAG / (855) , E: S:
Page 1 Comparison of all models Page 2 Model BK- V1000 BK- UV1000 BK- V1200 BK- UV1200 BK- V1600 BK- UV1600 BK- V1800 BK- UV1800 BK- V1900 BK- UV1900 BK- S360 BK- S380 BK- S390 BK- D560 BK- D580 BK- D590
More informationBrewer Ozone Spectrophotometer Mechanical Design Principles #2. C.T. McElroy York University Canada
Brewer Ozone Spectrophotometer Mechanical Design Principles #2 C.T. McElroy York University Canada Optics of the Brewer Spectrophotometer A Little History Fore-optics (1, 2) Spectrometer layout Grating
More informationSection 1: SPECTRAL PRODUCTS
Section 1: Optical Non-dispersive Wavelength Selection Filter Based Filter Filter Fundamentals Filter at an Incidence Angle Filters and Environmental Conditions Dispersive Instruments Grating and Polychromators
More informationMultispectral. imaging device. ADVANCED LIGHT ANALYSIS by. Most accurate homogeneity MeasureMent of spectral radiance. UMasterMS1 & UMasterMS2
Multispectral imaging device Most accurate homogeneity MeasureMent of spectral radiance UMasterMS1 & UMasterMS2 ADVANCED LIGHT ANALYSIS by UMaster Ms Multispectral Imaging Device UMaster MS Description
More informationSouthern African Large Telescope. RSS CCD Geometry
Southern African Large Telescope RSS CCD Geometry Kenneth Nordsieck University of Wisconsin Document Number: SALT-30AM0011 v 1.0 9 May, 2012 Change History Rev Date Description 1.0 9 May, 2012 Original
More informationSupplementary Figure 1
Supplementary Figure 1 Technical overview drawing of the Roadrunner goniometer. The goniometer consists of three main components: an inline sample-viewing microscope, a high-precision scanning unit for
More informationUSER MANUAL FOR VISIBLE SPECTROPHOTOMETER
USER MANUAL FOR VISIBLE SPECTROPHOTOMETER 1 Table of Contents 1. MAIN USAGES...3 2. WORKING ENVIRONMENT...3 3. MAIN TECHNICAL DATA AND SPECIFICATIONS...4 4. WORKING PRINCIPLE...5 5. OPTICAL PRINCIPLE...6
More informationBGS400 Lamp & Luminaire Goniophotometer (C-γ, B-ß Plane)
BGS400 Lamp & Luminaire Goniophotometer (C-γ, B-ß Plane) BGS400 Luminaire Goniophotometer (C-γ, B-ß Plane ) The BGS400 moving lamp and luminaire gonio-photometer is an economical, turnkey solution for
More informationImaging Fourier transform spectrometer
Rochester Institute of Technology RIT Scholar Works Theses Thesis/Dissertation Collections 2001 Imaging Fourier transform spectrometer Eric Sztanko Follow this and additional works at: http://scholarworks.rit.edu/theses
More informationContent Spectrophotometers
Spectroph Content Spectrophotometers Selection table Spectrophotometers Page 163 PRIM Page 164 UviLine Page 166 162 Selection table Spectrophotometers PRIM Light/ PRIM Advanced UviLine 9100/ UviLine 9400
More informationHigh Energy Non - Collinear OPA
High Energy Non - Collinear OPA Basics of Operation FEATURES Pulse Duration less than 10 fs possible High Energy (> 80 microjoule) Visible Output Wavelength Tuning Computer Controlled Tuning Range 250-375,
More informationTECHNICAL MEMORANDUM
TECHNICAL MEMORANDUM NOTICE: This document is version controlled and was produced as a part of the GEX Information Program which requires that all Series 100 documents be reviewed periodically to maintain
More informationPROCEEDINGS OF SPIE. Automated asphere centration testing with AspheroCheck UP
PROCEEDINGS OF SPIE SPIEDigitalLibrary.org/conference-proceedings-of-spie Automated asphere centration testing with AspheroCheck UP F. Hahne, P. Langehanenberg F. Hahne, P. Langehanenberg, "Automated asphere
More informationT92+ UV-VIS SPECTROPHOTOMETER
T92+ UV-VIS SPECTROPHOTOMETER The T92+ is a high performance double beam spectrophotometer with a variable spectral bandwidth from 0.1-5nm, selected by a continuous variable slit. The instrument is fitted
More informationTracQ. Basic Data Acquisition and Spectroscopy Software
Basic Data Acquisition and Spectroscopy Software TracQ Basic main application window. Many common spectroscopic measurements require coordinated operation of a detection instrument and light source, as
More informationEE119 Introduction to Optical Engineering Spring 2002 Final Exam. Name:
EE119 Introduction to Optical Engineering Spring 2002 Final Exam Name: SID: CLOSED BOOK. FOUR 8 1/2 X 11 SHEETS OF NOTES, AND SCIENTIFIC POCKET CALCULATOR PERMITTED. TIME ALLOTTED: 180 MINUTES Fundamental
More information