Southern African Large Telescope. RSS CCD Geometry

Size: px
Start display at page:

Download "Southern African Large Telescope. RSS CCD Geometry"

Transcription

1 Southern African Large Telescope RSS CCD Geometry Kenneth Nordsieck University of Wisconsin Document Number: SALT-30AM0011 v May, 2012 Change History Rev Date Description May, 2012 Original Table of Contents 1 Scope Mosaicing calibration method requirements 2.1History Technique description Current results Continuum traces Arc line traces Final results Systematic errors...7

2 RSS CCD Geometry May 9, Scope This document describes a technique for measuring the RSS CCD mosaicing geometry, and gives results which apply to data taken up to present. 2 Mosaicing calibration method requirements 2.1 History The relative rotation and lateral location of the three RSS CCD chips is described by six numbers. These are the displacement x,y (in unbinned pixels) and rotation ρ (in degrees) of the "outer" CCD's (1 and 3) about their centers relative to the center one (CCD 2), starting from the nominal position with a gap of 90 pixels in the x direction, and a y displacement and ρ rotation of zero. These numbers, which are stored in the RSSGeom.dat file, are used to define a linear transformation compensating for these offsets in the saltclean pipeline process by resampling the raw data from the 6 amplifiers into a single FITS image, through the IRAF imlintran task. Initial values for the 6 transformation numbers (table 1) were supplied on delivery in 2006, and are documented in SALT-3120AS0023 "PFIS Distortion and Alignment Model", section 2. These are the negatives of x, y, and ρ. Date gap xshift(1) yshift(1) rot(1) xshift(3) yshift(3) rot(3) 1/1/ Stated errors in these numbers, which were based on laboratory imaging of a laser-machined slitmask grid, were 0.3 pixels and 4E. Subsequent on-sky spectral data showed spectral traces with ~1 pixel jumps at the CCD gaps, which was clearly insufficient. A better technique was clearly required. Errors of better than pixel (.005E) are required in order that wavelength calibration errors are dominated by centroid errors rather than systematic geometry errors. 2.2 Technique description Table 1. Delivered CCD Mosaic Transformation The original technique suffered from several limitations: Most important, the laser machining errors in producing the calibration mask are on the order of 10 microns, or 1/3 pixel at the detector (4 arcsec on the sky). Secondly, the reimaging of the mask onto the detector is affected by the rather large (2%) distortion in the collimator, which must be modeled out of the re-imaged data in the same step as the evaluation of the mosaic geometry. Errors in the distortion coefficients and the mosaicing parameters are coupled, so that errors in the distortion model are reflected in the mosaicing numbers. This situation would not be improved by using on-sky astrometry, since astrometry errors are likely to be as high as the laser-machining errors, and the collimator distortion is still a factor. The proposed improved technique uses laboratory spectral data, taking advantage of the discovery that very small mosaicing errors were easily seen in spectral traces. Calibration images are formed by lamps via the grating equation, a very simple and repeatable process. A less obvious advantage is that the calibration image is formed at the grating, so does not suffer the collimator distortion in being imaged onto the detector. There is a very small distortion in

3 RSS CCD Geometry May 9, the camera, about one order of magnitude down from the collimator, but this is easily taken out, and its effects are likewise an order of magnitude down. This technique is implemented by two different lamp spectra, a continuum spectrum through very small holes to give continuum traces, and high dispersion spectra through the narrowest long-slit to give curved spectral lines which can be made to fall across a CCD gap. The first pins down the "spectrum jump" at the gap, which provides a relation between y and ρ of the outer chip. The second establishes the x gap error and the chip rotation through comparison of the straddling trace with traces of lines on either side of the gap. Below we apply this technique to lamp spectra from , yielding numbers which appear to be good to better than 5 pixels and.002e. Continuum traces from 2005 are consistent with these numbers, suggesting that the new results may be used for all RSS data taken since delivery. A regular application of the method is suggested to check for changes in the mosaicing, especially when the detector needs to be let up to room temperature for maintenance. 3 Current results Image Grating Gr Angle Art Angle Filter 3.1 Continuum traces PC04600 Data for this step (table 2) were obtained PC04600 with the QTH lamp using the PC04600 "geometry" MOS slitmask P00000N06, which has minimal laser-drilled holes PC00000 along the vertical axis, spaced 10 mm apart. Unbinned images with 1-3 Table 2 Continuum trace configurations minute exposures are required, using both QTH1 and QTH2 lamps with no filter, to obtain adequate signal/noise through the very small holes. These data were processed through the pipeline using the delivered transformation of table 1, so that we are evaluating dx, dy and dρ from this geometry. On the left of figure 2 on the following page is shown the IRAF aptrace of the center three spectra from each of these 4 images, offset by rows. Apart from the obvious remaining jumps at the gaps, an overall curved shape is apparent which changes with vertical spectral position and with grating. This is governed by the camera distortion and by the misalignments of the gratings in their holders. On the right in figure 2 is the residual to a fit of a fourth-order polynomial, with gap error dy and dρ for each gap left as a free parameter. Figure 1 shows these 12 solutions plotted as dy vs dρ for each gap. Clearly, this measurement does not give an adequate measure of these parameters individually, but rather a relation between dy and dρ where the spectra meet at the gap. dy (pix) CCD 1 fit1 Intermediate 1 Final 1 CCD 3 fit3 Intermediate 3 Final d (deg) Figure 1. Continuum trace results (CCD1, blue, CCD 3, red)

4 RSS CCD Geometry May 9, PG1300 PG00 PG2300 PG3000 Figure 2. Continuum spectrum traces for the central three spectra of each of four configurations (left). Residuals to power law plus CCD mosaic fit (right, 10 mag).

5 RSS CCD Geometry May 9, The angle the spectra make with each other at the gap is less well defined, mainly because of the imperfect fit of the curved spectra, plus artifacts at the intersection of the two amplifiers in each CCD. (These can be seen in the residuals, which are magnified by 10 from the traces). 3.2 Arc line traces Data for this step (table 3) were obtained with high dispersion Ne Arc spectra through the narrow 0.6 arcsec longslit. The grating angle was carefully adjusted to an off- Littrow setting that would cause a bright line to straddle a gap, while there would be sufficient neighboring continuous traces to establish the form of the trace at the gap. Images were processed using the SALT pipeline with an intermediate guess for the geometry shown in figure 1, so that the large y error would not couple into the results. Table 4 shows this intermediate geometry. Image Grating Gr Angle Art Angle CCD Table 3 Arc trace configurations Date gap xshift(1) yshift(1) rot(1) xshift(3) yshift(3) rot(3) 1/1/ Table 4. Intermediate CCD Mosaic Transformation Figures 3 and 4 show the details of the analysis for one CCD1 ( ) and one CCD3 ( ) observation. On the left (3,4a) is shown the spectrum around the target gap, with the cross-dispersion scale compressed by 2 to emphasize the curvature. The labeled lines were traced perpendicular to the dispersion using the IRAF aptrace task. Top center (3,4b) shows these traces superposed, to illustrate the change in trace curvature with wavelength. Each trace was then fit with a fourth-order polynomial and the residuals are shown superposed in 3,4c. There is a common structure to these residuals which is due to the slight roughness in the lasermachined longslit. The mean of these residuals ("slit correction") was removed from the traces, and they were re-fit. To establish the change in trace shape with wavelength, the four polynomial coefficients for each (non-gap) individual line fit are plotted vs the on-axis column of each line in figures 3,4d and 3,4e. A linear change with column appears to be justified. To estimate the expected trace shape for the line which straddles the gap, a "global" least-squares fit was performed on all non-gap lines simultaneously, forcing the polynomial coefficients into a linear dependence on column. The coefficient fit is shown in 3,4d and 3,4e, and the residuals of the traces for each line are show in 3,4f, with an arbitrary offset between each line for clarity. Finally, the coefficients for the gap line were evaluated and the deviation of the gap line trace is shown in 3,4g ("uncorrected"). A few tenths of a pixel jump at the gap and (for CCD 3) a slight remaining rotation is apparent. Finally, the x gap and rotation were allowed to vary, and the resulting "corrected" deviation of the gap line is shown.

6 RSS CCD Geometry May 9, Gap Line Slit Cor'n column deviation (pixels) trace fit error (pixels) global fit error (pixels) + arbitrary offset A A1 Global Fit A2 Gap E 05 6.E 05 A2 5.E 05 4.E gap line fit error (pixels) Column (pixels) E 11 6.E 11 A3 4.E 11 2.E 11 0.E+00 A3 Global Fit A E E A4 1.0E E E 13 uncorrected corrected column (pixels) Ar Ar Ne Ne Ne Ne Ar Ne Ar 3a 3b 3c 3f 3d 3e 3g Figure 3 X Gap analysis for CCD1 observation

7 RSS CCD Geometry May 9, Gap Line column deviation (pixels) trace fit error (pixels) Slit Cor'n global fit error (pixels) + arbitrary offset A A1 Global Fit A2 Gap gap line fit error (pixels) Column (pixels) E 05 7.E 05 A2 6.E E E 11 A3 6.0E E E E 1.0E 13 A4 1.5E E E 13 uncorrected corrected column (pixels) Ne Ne Ne Ne Ne Ne 3a 3b 3c 3f 3d 3e A3 Global Fit A4 3g Figure 4 X Gap analysis for CCD 3 Observation

8 RSS CCD Geometry May 9, A plot of the dx vs dρ results for each of the 8 configurations in Table 3, together with the weighted mean for each CCD, is shown in figure Final results For a final geometry estimation, the mean dρ in each CCD as shown in figure 5 was used with the dy-dρ fit as shown in figure 1 to arrive a final dy estimate for each CCD, giving the results shown in table 5. Adopted errors are shown in the second line. 4 Systematic errors CCD 1 CCD1 mean CCD CCD3 mean In general, the errors seen here are larger than what is expected from centroiding errors. Some possible sources of systematic error, in order of judged level of importance, are: Change in geometry with time. This is not seen to be important. First, the geometry is consistent with being unchanged since the detector was first operated in 2004, judging by analysis of the y gap in QTH spectral traces from that era. Second, there is no evidence of time-dependent effects in individual dy and dx estimates in figures 1 and 5. CCD layout errors. Small persistent jumps at the boundary between amplifiers within a CCD are apparent in figure 2 (right), but these are at the 1 pixel level, and probably do not contribute to the analysis. Change of image quality across the detector (with wavelength, and with position in the camera field of view). The optics is designed to deliver images better than 2 unbinned pixels (5 arcsec). Below that level, there are comatic and astigmatic aberrations which are wavelength and FOV dependent. This probably accounts for smooth deviations in the centroiding residuals see in figures 2, 3, and 4, at the < 3 pixel level, and is controlled here by using more than one grating and wavelength interval for each analysis. The largest error in this analysis is likely the coupling of the above imaging effect with imperfect coplanarity of the chips. A systematic focus change across the gap will, in the presence of aberrations, give rise to a systematic change in centroiding, which will be the same for all field-of-view dependent aberrations, so will not cancel out. This probably accounts for the systematic change in residual shape across the gaps seen in figures 3,4f, which has a peak-to-peak variation of 5 pixel dx (pix) d (deg) Figure 5. Arc line trace results (CCD1, blue, CCD 3, red) Date gap xshift(1) yshift(1) rot(1) xshift(3) yshift(3) rot(3) 1/1/ Err Table 5. Final CCD Mosaic Transformation

Southern African Large Telescope. RSS UW Commissioning Activities,

Southern African Large Telescope. RSS UW Commissioning Activities, Southern African Large Telescope RSS UW Commissioning Activities, 2014-1 Kenneth Nordsieck University of Wisconsin v 1.1 5 Nov, 2014 Change History Rev Date Description 1.0 3 Nov, 2014 Original 1.1 5 Nov,

More information

Southern African Large Telescope. Prime Focus Imaging Spectrograph. Instrument Acceptance Testing Plan

Southern African Large Telescope. Prime Focus Imaging Spectrograph. Instrument Acceptance Testing Plan Southern African Large Telescope Prime Focus Imaging Spectrograph Instrument Acceptance Testing Plan Eric B. Burgh University of Wisconsin Document Number: SALT-3160AP0003 Revision 2.2 29 April 2004 1

More information

Descriptions for Each Test

Descriptions for Each Test Descriptions for Each Test 1. Image Field Size: a. The image field size is determined by the slitmask frame, which has a machined aperture of 109mm. The plate scale of the SALT focal plane has been determined

More information

Eric B. Burgh University of Wisconsin. 1. Scope

Eric B. Burgh University of Wisconsin. 1. Scope Southern African Large Telescope Prime Focus Imaging Spectrograph Optical Integration and Testing Plan Document Number: SALT-3160BP0001 Revision 5.0 2007 July 3 Eric B. Burgh University of Wisconsin 1.

More information

SPACE TELESCOPE SCIENCE INSTITUTE Operated for NASA by AURA

SPACE TELESCOPE SCIENCE INSTITUTE Operated for NASA by AURA SPACE TELESCOPE SCIENCE INSTITUTE Operated for NASA by AURA Instrument Science Report WFC3 2010-08 WFC3 Pixel Area Maps J. S. Kalirai, C. Cox, L. Dressel, A. Fruchter, W. Hack, V. Kozhurina-Platais, and

More information

Performance Comparison of Spectrometers Featuring On-Axis and Off-Axis Grating Rotation

Performance Comparison of Spectrometers Featuring On-Axis and Off-Axis Grating Rotation Performance Comparison of Spectrometers Featuring On-Axis and Off-Axis Rotation By: Michael Case and Roy Grayzel, Acton Research Corporation Introduction The majority of modern spectrographs and scanning

More information

Wavelength Calibration Accuracy of the First-Order CCD Modes Using the E1 Aperture

Wavelength Calibration Accuracy of the First-Order CCD Modes Using the E1 Aperture Wavelength Calibration Accuracy of the First-Order CCD Modes Using the E1 Aperture Scott D. Friedman August 22, 2005 ABSTRACT A calibration program was carried out to determine the quality of the wavelength

More information

Information for users of the SOAR Goodman Spectrograph Multi-Object Slit (MOS) mode. César Briceño and Sean Points

Information for users of the SOAR Goodman Spectrograph Multi-Object Slit (MOS) mode. César Briceño and Sean Points Information for users of the SOAR Goodman Spectrograph Multi-Object Slit (MOS) mode César Briceño and Sean Points CTIO, June 2014 The Goodman Spectrograph has been offered for use in MOS mode starting

More information

Southern African Large Telescope. Prime Focus Imaging Spectrograph. Polarimetric Optics Design Study

Southern African Large Telescope. Prime Focus Imaging Spectrograph. Polarimetric Optics Design Study Southern African Large Telescope Prime Focus Imaging Spectrograph Polarimetric Optics Design Study Kenneth Nordsieck University of Wisconsin Revision 1.1 5 Oct 2001 SALT PFIS/IMPALAS Polarimetric Optics

More information

Observation Data. Optical Images

Observation Data. Optical Images Data Analysis Introduction Optical Imaging Tsuyoshi Terai Subaru Telescope Imaging Observation Measure the light from celestial objects and understand their physics Take images of objects with a specific

More information

Presented by Jerry Hubbell Lake of the Woods Observatory (MPC I24) President, Rappahannock Astronomy Club

Presented by Jerry Hubbell Lake of the Woods Observatory (MPC I24) President, Rappahannock Astronomy Club Presented by Jerry Hubbell Lake of the Woods Observatory (MPC I24) President, Rappahannock Astronomy Club ENGINEERING A FIBER-FED FED SPECTROMETER FOR ASTRONOMICAL USE Objectives Discuss the engineering

More information

AgilEye Manual Version 2.0 February 28, 2007

AgilEye Manual Version 2.0 February 28, 2007 AgilEye Manual Version 2.0 February 28, 2007 1717 Louisiana NE Suite 202 Albuquerque, NM 87110 (505) 268-4742 support@agiloptics.com 2 (505) 268-4742 v. 2.0 February 07, 2007 3 Introduction AgilEye Wavefront

More information

ECEN 4606, UNDERGRADUATE OPTICS LAB

ECEN 4606, UNDERGRADUATE OPTICS LAB ECEN 4606, UNDERGRADUATE OPTICS LAB Lab 2: Imaging 1 the Telescope Original Version: Prof. McLeod SUMMARY: In this lab you will become familiar with the use of one or more lenses to create images of distant

More information

Cerro Tololo Inter-American Observatory. CHIRON manual. A. Tokovinin Version 2. May 25, 2011 (manual.pdf)

Cerro Tololo Inter-American Observatory. CHIRON manual. A. Tokovinin Version 2. May 25, 2011 (manual.pdf) Cerro Tololo Inter-American Observatory CHIRON manual A. Tokovinin Version 2. May 25, 2011 (manual.pdf) 1 1 Overview Calibration lamps Quartz, Th Ar Fiber Prism Starlight GAM mirror Fiber Viewer FEM Guider

More information

F/48 Slit Spectroscopy

F/48 Slit Spectroscopy 1997 HST Calibration Workshop Space Telescope Science Institute, 1997 S. Casertano, et al., eds. F/48 Slit Spectroscopy R. Jedrzejewski & M. Voit Space Telescope Science Institute, Baltimore, MD 21218

More information

ENGINEERING CHANGE ORDER ECO No. COS-070 Center for Astrophysics & Space Astronomy Date 6 February 2002 University of Colorado, Boulder Sheet 1 of 3

ENGINEERING CHANGE ORDER ECO No. COS-070 Center for Astrophysics & Space Astronomy Date 6 February 2002 University of Colorado, Boulder Sheet 1 of 3 University of Colorado, Boulder Sheet 1 of 3 Description of Change: 1. Page 11 (Rev. 16), Sec. 1.3.3: Replace the opening paragraph with the following text, which reflects the decision to fly a 225-type

More information

VATTSpec Instructions Rev. 10/23/2015

VATTSpec Instructions Rev. 10/23/2015 VATTSpec Instructions Rev. 10/23/2015 Introduction VATTSpec is a medium resolution CCD range spectrograph with a skinny chip having excellent cosmetics. Its UA ITL chip, Serial Number 8228, has a gain

More information

Southern African Large Telescope. RSS Throughput Test Plan

Southern African Large Telescope. RSS Throughput Test Plan Southern African Large Telescope RSS Throughput Test Plan Kenneth Nordsieck University of Wisconsin Document Number: SALT-3160AP0005 Revision 1.0 27 June, 2006 Change History Rev Date Description 1.0 27

More information

Application Note (A11)

Application Note (A11) Application Note (A11) Slit and Aperture Selection in Spectroradiometry REVISION: C August 2013 Gooch & Housego 4632 36 th Street, Orlando, FL 32811 Tel: 1 407 422 3171 Fax: 1 407 648 5412 Email: sales@goochandhousego.com

More information

Camera Resolution and Distortion: Advanced Edge Fitting

Camera Resolution and Distortion: Advanced Edge Fitting 28, Society for Imaging Science and Technology Camera Resolution and Distortion: Advanced Edge Fitting Peter D. Burns; Burns Digital Imaging and Don Williams; Image Science Associates Abstract A frequently

More information

Optical design of a high resolution vision lens

Optical design of a high resolution vision lens Optical design of a high resolution vision lens Paul Claassen, optical designer, paul.claassen@sioux.eu Marnix Tas, optical specialist, marnix.tas@sioux.eu Prof L.Beckmann, l.beckmann@hccnet.nl Summary:

More information

Southern African Large Telescope SALTICAM Preliminary Design Review. Document Number 3360AE0001: Detector Document

Southern African Large Telescope SALTICAM Preliminary Design Review. Document Number 3360AE0001: Detector Document 3360AE0001: Detector Document 1 Southern African Large Telescope SALTICAM Preliminary Design Review Document Number 3360AE0001: Detector Document Darragh O Donoghue Dave Carter Geoff Evans Willie Koorts

More information

SOAR Integral Field Spectrograph (SIFS): Call for Science Verification Proposals

SOAR Integral Field Spectrograph (SIFS): Call for Science Verification Proposals Published on SOAR (http://www.ctio.noao.edu/soar) Home > SOAR Integral Field Spectrograph (SIFS): Call for Science Verification Proposals SOAR Integral Field Spectrograph (SIFS): Call for Science Verification

More information

III. Descriptions for Each Test and Results

III. Descriptions for Each Test and Results III. Descriptions for Each Test and Results 1. Image Field Size: a. The image field size is determined by the slitmask frame, which has a machined aperture of 109mm. The plate scale of the SALT focal plane

More information

Master sky images for the WFC3 G102 and G141 grisms

Master sky images for the WFC3 G102 and G141 grisms Master sky images for the WFC3 G102 and G141 grisms M. Kümmel, H. Kuntschner, J. R. Walsh, H. Bushouse January 4, 2011 ABSTRACT We have constructed master sky images for the WFC3 near-infrared G102 and

More information

Light gathering Power: Magnification with eyepiece:

Light gathering Power: Magnification with eyepiece: Telescopes Light gathering Power: The amount of light that can be gathered by a telescope in a given amount of time: t 1 /t 2 = (D 2 /D 1 ) 2 The larger the diameter the smaller the amount of time. If

More information

Notes on the VPPEM electron optics

Notes on the VPPEM electron optics Notes on the VPPEM electron optics Raymond Browning 2/9/2015 We are interested in creating some rules of thumb for designing the VPPEM instrument in terms of the interaction between the field of view at

More information

GPI INSTRUMENT PAGES

GPI INSTRUMENT PAGES GPI INSTRUMENT PAGES This document presents a snapshot of the GPI Instrument web pages as of the date of the call for letters of intent. Please consult the GPI web pages themselves for up to the minute

More information

Be aware that there is no universal notation for the various quantities.

Be aware that there is no universal notation for the various quantities. Fourier Optics v2.4 Ray tracing is limited in its ability to describe optics because it ignores the wave properties of light. Diffraction is needed to explain image spatial resolution and contrast and

More information

Spectral Analysis of the LUND/DMI Earthshine Telescope and Filters

Spectral Analysis of the LUND/DMI Earthshine Telescope and Filters Spectral Analysis of the LUND/DMI Earthshine Telescope and Filters 12 August 2011-08-12 Ahmad Darudi & Rodrigo Badínez A1 1. Spectral Analysis of the telescope and Filters This section reports the characterization

More information

Solar Optical Telescope (SOT)

Solar Optical Telescope (SOT) Solar Optical Telescope (SOT) The Solar-B Solar Optical Telescope (SOT) will be the largest telescope with highest performance ever to observe the sun from space. The telescope itself (the so-called Optical

More information

Southern African Large Telescope. Prime Focus Imaging Spectrograph. Grating and Filter Specification Document

Southern African Large Telescope. Prime Focus Imaging Spectrograph. Grating and Filter Specification Document Southern African Large Telescope Prime Focus Imaging Spectrograph Grating and Filter Specification Document Chip Kobulnicky University of Wisconsin Kenneth Nordsieck University of Wisconsin Revision 2.1

More information

NIRSPEC Data Reduction Pipeline Data Products Specification

NIRSPEC Data Reduction Pipeline Data Products Specification NIRSPEC Data Reduction Pipeline Data Products Specification Table of Contents 1 Introduction... 2 2 Data Products... 2 2.1 Tables...2 2.1.1 Table Format...2 2.1.2 Flux Table...3 2.1.3 Profile Table...4

More information

6. Very low level processing (radiometric calibration)

6. Very low level processing (radiometric calibration) Master ISTI / PARI / IV Introduction to Astronomical Image Processing 6. Very low level processing (radiometric calibration) André Jalobeanu LSIIT / MIV / PASEO group Jan. 2006 lsiit-miv.u-strasbg.fr/paseo

More information

Waves & Oscillations

Waves & Oscillations Physics 42200 Waves & Oscillations Lecture 33 Geometric Optics Spring 2013 Semester Matthew Jones Aberrations We have continued to make approximations: Paraxial rays Spherical lenses Index of refraction

More information

Observational Astronomy

Observational Astronomy Observational Astronomy Instruments The telescope- instruments combination forms a tightly coupled system: Telescope = collecting photons and forming an image Instruments = registering and analyzing the

More information

Flux Calibration Monitoring: WFC3/IR G102 and G141 Grisms

Flux Calibration Monitoring: WFC3/IR G102 and G141 Grisms Instrument Science Report WFC3 2014-01 Flux Calibration Monitoring: WFC3/IR and Grisms Janice C. Lee, Norbert Pirzkal, Bryan Hilbert January 24, 2014 ABSTRACT As part of the regular WFC3 flux calibration

More information

Unit 1: Image Formation

Unit 1: Image Formation Unit 1: Image Formation 1. Geometry 2. Optics 3. Photometry 4. Sensor Readings Szeliski 2.1-2.3 & 6.3.5 1 Physical parameters of image formation Geometric Type of projection Camera pose Optical Sensor

More information

Southern African Large Telescope. RSS Observer s Guide

Southern African Large Telescope. RSS Observer s Guide Southern African Large Telescope RSS Observer s Guide Eric B. Burgh Kenneth Nordsieck University of Wisconsin Document Number: SALT-3170AM0007 Version 0.5 23 Jan, 2009 Change History Rev Date Description

More information

3 General Principles of Operation of the S7500 Laser

3 General Principles of Operation of the S7500 Laser Application Note AN-2095 Controlling the S7500 CW Tunable Laser 1 Introduction This document explains the general principles of operation of Finisar s S7500 tunable laser. It provides a high-level description

More information

Big League Cryogenics and Vacuum The LHC at CERN

Big League Cryogenics and Vacuum The LHC at CERN Big League Cryogenics and Vacuum The LHC at CERN A typical astronomical instrument must maintain about one cubic meter at a pressure of

More information

Improved Spectra with a Schmidt-Czerny-Turner Spectrograph

Improved Spectra with a Schmidt-Czerny-Turner Spectrograph Improved Spectra with a Schmidt-Czerny-Turner Spectrograph Abstract For years spectra have been measured using traditional Czerny-Turner (CT) design dispersive spectrographs. Optical aberrations inherent

More information

Computer Generated Holograms for Optical Testing

Computer Generated Holograms for Optical Testing Computer Generated Holograms for Optical Testing Dr. Jim Burge Associate Professor Optical Sciences and Astronomy University of Arizona jburge@optics.arizona.edu 520-621-8182 Computer Generated Holograms

More information

Aberrations of a lens

Aberrations of a lens Aberrations of a lens 1. What are aberrations? A lens made of a uniform glass with spherical surfaces cannot form perfect images. Spherical aberration is a prominent image defect for a point source on

More information

UV/Optical/IR Astronomy Part 2: Spectroscopy

UV/Optical/IR Astronomy Part 2: Spectroscopy UV/Optical/IR Astronomy Part 2: Spectroscopy Introduction We now turn to spectroscopy. Much of what you need to know about this is the same as for imaging I ll concentrate on the differences. Slicing the

More information

COS: NUV and FUV Detector Flat Field Status

COS: NUV and FUV Detector Flat Field Status The 2005 HST Calibration Workshop Space Telescope Science Institute, 2005 A. M. Koekemoer, P. Goudfrooij, and L. L. Dressel, eds. COS: NUV and FUV Detector Flat Field Status Steven V. Penton Center for

More information

The predicted performance of the ACS coronagraph

The predicted performance of the ACS coronagraph Instrument Science Report ACS 2000-04 The predicted performance of the ACS coronagraph John Krist March 30, 2000 ABSTRACT The Aberrated Beam Coronagraph (ABC) on the Advanced Camera for Surveys (ACS) has

More information

The IRAF Mosaic Data Reduction Package

The IRAF Mosaic Data Reduction Package Astronomical Data Analysis Software and Systems VII ASP Conference Series, Vol. 145, 1998 R. Albrecht, R. N. Hook and H. A. Bushouse, eds. The IRAF Mosaic Data Reduction Package Francisco G. Valdes IRAF

More information

DESIGN NOTE: DIFFRACTION EFFECTS

DESIGN NOTE: DIFFRACTION EFFECTS NASA IRTF / UNIVERSITY OF HAWAII Document #: TMP-1.3.4.2-00-X.doc Template created on: 15 March 2009 Last Modified on: 5 April 2010 DESIGN NOTE: DIFFRACTION EFFECTS Original Author: John Rayner NASA Infrared

More information

ISIS A beginner s guide

ISIS A beginner s guide ISIS A beginner s guide Conceived of and written by Christian Buil, ISIS is a powerful astronomical spectral processing application that can appear daunting to first time users. While designed as a comprehensive

More information

ECEN. Spectroscopy. Lab 8. copy. constituents HOMEWORK PR. Figure. 1. Layout of. of the

ECEN. Spectroscopy. Lab 8. copy. constituents HOMEWORK PR. Figure. 1. Layout of. of the ECEN 4606 Lab 8 Spectroscopy SUMMARY: ROBLEM 1: Pedrotti 3 12-10. In this lab, you will design, build and test an optical spectrum analyzer and use it for both absorption and emission spectroscopy. The

More information

Nonlinearity in the Detector used in the Subaru Telescope High Dispersion Spectrograph

Nonlinearity in the Detector used in the Subaru Telescope High Dispersion Spectrograph Nonlinearity in the Detector used in the Subaru Telescope High Dispersion Spectrograph Akito Tajitsu Subaru Telescope, National Astronomical Observatory of Japan, 650 North A ohoku Place, Hilo, HI 96720,

More information

This experiment is under development and thus we appreciate any and all comments as we design an interesting and achievable set of goals.

This experiment is under development and thus we appreciate any and all comments as we design an interesting and achievable set of goals. Experiment 7 Geometrical Optics You will be introduced to ray optics and image formation in this experiment. We will use the optical rail, lenses, and the camera body to quantify image formation and magnification;

More information

Camera Calibration Certificate No: DMC III 27542

Camera Calibration Certificate No: DMC III 27542 Calibration DMC III Camera Calibration Certificate No: DMC III 27542 For Peregrine Aerial Surveys, Inc. #201 1255 Townline Road Abbotsford, B.C. V2T 6E1 Canada Calib_DMCIII_27542.docx Document Version

More information

NGAO NGS WFS design review

NGAO NGS WFS design review NGAO NGS WFS design review Caltech Optical 1 st April2010 1 Presentation outline Requirements (including modes of operation and motion control) Introduction NGSWFS input feed (performance of the triplet

More information

PROCEEDINGS OF SPIE. Measurement of low-order aberrations with an autostigmatic microscope

PROCEEDINGS OF SPIE. Measurement of low-order aberrations with an autostigmatic microscope PROCEEDINGS OF SPIE SPIEDigitalLibrary.org/conference-proceedings-of-spie Measurement of low-order aberrations with an autostigmatic microscope William P. Kuhn Measurement of low-order aberrations with

More information

arxiv: v1 [astro-ph.im] 26 Mar 2012

arxiv: v1 [astro-ph.im] 26 Mar 2012 The image slicer for the Subaru Telescope High Dispersion Spectrograph arxiv:1203.5568v1 [astro-ph.im] 26 Mar 2012 Akito Tajitsu The Subaru Telescope, National Astronomical Observatory of Japan, 650 North

More information

UltraGraph Optics Design

UltraGraph Optics Design UltraGraph Optics Design 5/10/99 Jim Hagerman Introduction This paper presents the current design status of the UltraGraph optics. Compromises in performance were made to reach certain product goals. Cost,

More information

m Copernico Telescope, Ekar User Manual

m Copernico Telescope, Ekar User Manual Afosc @1.82 m Copernico Telescope, Ekar User Manual L. Tomasella, S. Benetti, V. Chiomento, L. Traverso, M. Fiaschi Version 2.0 - March 2012 Note: In 2011 the old detectors (cooling system based on liquid

More information

Reference and User Manual May, 2015 revision - 3

Reference and User Manual May, 2015 revision - 3 Reference and User Manual May, 2015 revision - 3 Innovations Foresight 2015 - Powered by Alcor System 1 For any improvement and suggestions, please contact customerservice@innovationsforesight.com Some

More information

First steps with a Slit Spectroscope

First steps with a Slit Spectroscope First steps with a Slit Spectroscope Ken M Harrison September 2012 Ok, you ve taken the quantum leap and moved into your first slit spectroscope. Your initial experiences may have been with a grating (Star

More information

Exoplanet transit, eclipse, and phase curve observations with JWST NIRCam. Tom Greene & John Stansberry JWST NIRCam transit meeting March 12, 2014

Exoplanet transit, eclipse, and phase curve observations with JWST NIRCam. Tom Greene & John Stansberry JWST NIRCam transit meeting March 12, 2014 Exoplanet transit, eclipse, and phase curve observations with JWST NIRCam Tom Greene & John Stansberry JWST NIRCam transit meeting March 12, 2014 1 Scope of Talk NIRCam overview Suggested transit modes

More information

Cameras. CSE 455, Winter 2010 January 25, 2010

Cameras. CSE 455, Winter 2010 January 25, 2010 Cameras CSE 455, Winter 2010 January 25, 2010 Announcements New Lecturer! Neel Joshi, Ph.D. Post-Doctoral Researcher Microsoft Research neel@cs Project 1b (seam carving) was due on Friday the 22 nd Project

More information

Sensors and Sensing Cameras and Camera Calibration

Sensors and Sensing Cameras and Camera Calibration Sensors and Sensing Cameras and Camera Calibration Todor Stoyanov Mobile Robotics and Olfaction Lab Center for Applied Autonomous Sensor Systems Örebro University, Sweden todor.stoyanov@oru.se 20.11.2014

More information

Paper Synopsis. Xiaoyin Zhu Nov 5, 2012 OPTI 521

Paper Synopsis. Xiaoyin Zhu Nov 5, 2012 OPTI 521 Paper Synopsis Xiaoyin Zhu Nov 5, 2012 OPTI 521 Paper: Active Optics and Wavefront Sensing at the Upgraded 6.5-meter MMT by T. E. Pickering, S. C. West, and D. G. Fabricant Abstract: This synopsis summarized

More information

APPROVAL SHEET. TITLE : Prime Focus Image Spectrograph (PFIS) ICD of the Southern African Large Telescope (SALT)

APPROVAL SHEET. TITLE : Prime Focus Image Spectrograph (PFIS) ICD of the Southern African Large Telescope (SALT) APPROVAL SHEET TITLE : Prime Focus Image Spectrograph (PFIS) ICD of the Southern African Large Telescope (SALT) DOCUMENT NUMBER : 1520AS0002 ISSUE: 3 SYNOPSIS : This document describes the Interface between

More information

Epic Detector Matrices

Epic Detector Matrices Epic Detector Matrices June 2013 - SAS workshop Why do we need detector matrices? Initial source spectrum Source flux How do we use them? Spectral fitting: - xspec, CIAO Try a spectral model and see if

More information

1.6 Beam Wander vs. Image Jitter

1.6 Beam Wander vs. Image Jitter 8 Chapter 1 1.6 Beam Wander vs. Image Jitter It is common at this point to look at beam wander and image jitter and ask what differentiates them. Consider a cooperative optical communication system that

More information

XTcalc: MOSFIRE Exposure Time Calculator v2.3

XTcalc: MOSFIRE Exposure Time Calculator v2.3 XTcalc: MOSFIRE Exposure Time Calculator v2.3 by Gwen C. Rudie gwen@astro.caltech.edu July 2, 2012 1 Installation using IDL Virtual Machine This is the default way to run the code. It does not require

More information

This release contains deep Y-band images of the UDS field and the extracted source catalogue.

This release contains deep Y-band images of the UDS field and the extracted source catalogue. ESO Phase 3 Data Release Description Data Collection HUGS_UDS_Y Release Number 1 Data Provider Adriano Fontana Date 22.09.2014 Abstract HUGS (an acronym for Hawk-I UDS and GOODS Survey) is a ultra deep

More information

A collection of hyperspectral images for imaging systems research Torbjørn Skauli a,b, Joyce Farrell *a

A collection of hyperspectral images for imaging systems research Torbjørn Skauli a,b, Joyce Farrell *a A collection of hyperspectral images for imaging systems research Torbjørn Skauli a,b, Joyce Farrell *a a Stanford Center for Image Systems Engineering, Stanford CA, USA; b Norwegian Defence Research Establishment,

More information

The DSI for Autostar Suite

The DSI for Autostar Suite An Introduction To DSI Imaging John E. Hoot President Software Systems Consulting 1 The DSI for Autostar Suite Meade Autostar Suite Not Just A Project, A Mission John E. Hoot System Architect 2 1 DSI -

More information

DBSP Observing Manual

DBSP Observing Manual DBSP Observing Manual I. Arcavi, P. Bilgi, N.Blagorodnova, K.Burdge, A.Y.Q.Ho June 18, 2018 Contents 1 Observing Guides 2 2 Before arrival 2 2.1 Submit observing setup..................................

More information

Goodman Cookbook. Goodman Spectrograph. Adapted by D. Sanmartim from L. Fraga's Guide. Sep SOAR Telescope

Goodman Cookbook. Goodman Spectrograph. Adapted by D. Sanmartim from L. Fraga's Guide. Sep SOAR Telescope Goodman Spectrograph 1 Goodman Spectrograph Documentation Goodman HTS Manual http://www.ctio.noao.edu/soar/content/goodman-hts-manual Goodman Overview http://www.ctio.noao.edu/soar/content/goodman-spectrograph-overview

More information

Measurement and alignment of linear variable filters

Measurement and alignment of linear variable filters Measurement and alignment of linear variable filters Rob Sczupak, Markus Fredell, Tim Upton, Tom Rahmlow, Sheetal Chanda, Gregg Jarvis, Sarah Locknar, Florin Grosu, Terry Finnell and Robert Johnson Omega

More information

CHARA Collaboration Review New York 2007 CHARA Telescope Alignment

CHARA Collaboration Review New York 2007 CHARA Telescope Alignment CHARA Telescope Alignment By Laszlo Sturmann Mersenne (Cassegrain type) Telescope M2 140 mm R= 625 mm k = -1 M1/M2 provides an afocal optical system 1 m input beam and 0.125 m collimated output beam Aplanatic

More information

JCMT HETERODYNE DR FROM DATA TO SCIENCE

JCMT HETERODYNE DR FROM DATA TO SCIENCE JCMT HETERODYNE DR FROM DATA TO SCIENCE https://proposals.eaobservatory.org/ JCMT HETERODYNE - SHANGHAI WORKSHOP OCTOBER 2016 JCMT HETERODYNE INSTRUMENTATION www.eaobservatory.org/jcmt/science/reductionanalysis-tutorials/

More information

PhysicsAndMathsTutor.com 1

PhysicsAndMathsTutor.com 1 PhysicsAndMathsTutor.com 1 Q1. Just over two hundred years ago Thomas Young demonstrated the interference of light by illuminating two closely spaced narrow slits with light from a single light source.

More information

Modern UV-curing technology

Modern UV-curing technology Spectroradiometry in UV Curing By Lawrence E. Schmutz Figure 1 Absorption spectra for two common photoinitiator families (Spectra reproduced by permission of Sigma-Aldrich Corporation) Modern UV-curing

More information

Phase-2 Preparation Tool

Phase-2 Preparation Tool Gran Telescopio Canarias Phase-2 Preparation Tool Valid from period 2014A Updated: 5 December 2013 1 Contents 1. The GTC Phase-2 System... 3 1.1. Introduction... 3 1.2. Logging in... 3 2. Defining an observing

More information

CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES

CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES The current multiplication mechanism offered by dynodes makes photomultiplier tubes ideal for low-light-level measurement. As explained earlier, there

More information

Imaging Systems Laboratory II. Laboratory 8: The Michelson Interferometer / Diffraction April 30 & May 02, 2002

Imaging Systems Laboratory II. Laboratory 8: The Michelson Interferometer / Diffraction April 30 & May 02, 2002 1051-232 Imaging Systems Laboratory II Laboratory 8: The Michelson Interferometer / Diffraction April 30 & May 02, 2002 Abstract. In the last lab, you saw that coherent light from two different locations

More information

GMT Instruments and AO. GMT Science Meeting - March

GMT Instruments and AO. GMT Science Meeting - March GMT Instruments and AO GMT Science Meeting - March 2008 1 Instrument Status Scientific priorities have been defined Emphasis on: Wide-field survey science (cosmology) High resolution spectroscopy (abundances,

More information

AP Physics Problems -- Waves and Light

AP Physics Problems -- Waves and Light AP Physics Problems -- Waves and Light 1. 1974-3 (Geometric Optics) An object 1.0 cm high is placed 4 cm away from a converging lens having a focal length of 3 cm. a. Sketch a principal ray diagram for

More information

Geometry of Aerial Photographs

Geometry of Aerial Photographs Geometry of Aerial Photographs Aerial Cameras Aerial cameras must be (details in lectures): Geometrically stable Have fast and efficient shutters Have high geometric and optical quality lenses They can

More information

Instruction Manual for HyperScan Spectrometer

Instruction Manual for HyperScan Spectrometer August 2006 Version 1.1 Table of Contents Section Page 1 Hardware... 1 2 Mounting Procedure... 2 3 CCD Alignment... 6 4 Software... 7 5 Wiring Diagram... 19 1 HARDWARE While it is not necessary to have

More information

Image Formation: Camera Model

Image Formation: Camera Model Image Formation: Camera Model Ruigang Yang COMP 684 Fall 2005, CS684-IBMR Outline Camera Models Pinhole Perspective Projection Affine Projection Camera with Lenses Digital Image Formation The Human Eye

More information

CHARA AO Calibration Process

CHARA AO Calibration Process CHARA AO Calibration Process Judit Sturmann CHARA AO Project Overview Phase I. Under way WFS on telescopes used as tip-tilt detector Phase II. Not yet funded WFS and large DM in place of M4 on telescopes

More information

VS7550 VUV/UV Mini Spectrograph Operating Manual

VS7550 VUV/UV Mini Spectrograph Operating Manual Document RD 15 11 No: VS7550 VUV/UV Mini Spectrograph Operating Manual VS7550 Operating Manual 1 Table of Contents Table of Contents Overview Specifications Vacuum Interface Software and Drivers Packing

More information

Y11-DR Digital Radiography (DR) Image Quality

Y11-DR Digital Radiography (DR) Image Quality Y11-DR Digital Radiography (DR) Image Quality Image quality is stressed for all systems in Safety Code 35. In the relevant sections Health Canada s advice is the manufacturer s recommended test procedures

More information

Performance comparison of aperture codes for multimodal, multiplex spectroscopy

Performance comparison of aperture codes for multimodal, multiplex spectroscopy Performance comparison of aperture codes for multimodal, multiplex spectroscopy Ashwin A. Wagadarikar, Michael E. Gehm, and David J. Brady* Duke University Fitzpatrick Institute for Photonics, Box 90291,

More information

WITec Alpha 300R Quick Operation Summary October 2018

WITec Alpha 300R Quick Operation Summary October 2018 WITec Alpha 300R Quick Operation Summary October 2018 This document is frequently updated if you feel information should be added, please indicate that to the facility manager (currently Philip Carubia,

More information

Better Imaging with a Schmidt-Czerny-Turner Spectrograph

Better Imaging with a Schmidt-Czerny-Turner Spectrograph Better Imaging with a Schmidt-Czerny-Turner Spectrograph Abstract For years, images have been measured using Czerny-Turner (CT) design dispersive spectrographs. Optical aberrations inherent in the CT design

More information

University of Wisconsin Chemistry 524 Spectroscopic Components *

University of Wisconsin Chemistry 524 Spectroscopic Components * University of Wisconsin Chemistry 524 Spectroscopic Components * In journal articles, presentations, and textbooks, chemical instruments are often represented as block diagrams. These block diagrams highlight

More information

ECEN 4606, UNDERGRADUATE OPTICS LAB

ECEN 4606, UNDERGRADUATE OPTICS LAB ECEN 4606, UNDERGRADUATE OPTICS LAB Lab 3: Imaging 2 the Microscope Original Version: Professor McLeod SUMMARY: In this lab you will become familiar with the use of one or more lenses to create highly

More information

The diffraction of light

The diffraction of light 7 The diffraction of light 7.1 Introduction As introduced in Chapter 6, the reciprocal lattice is the basis upon which the geometry of X-ray and electron diffraction patterns can be most easily understood

More information

Corner Rafts LSST Camera Workshop SLAC Sept 19, 2008

Corner Rafts LSST Camera Workshop SLAC Sept 19, 2008 Corner Rafts LSST Camera Workshop SLAC Sept 19, 2008 Scot Olivier LLNL 1 LSST Conceptual Design Review 2 Corner Raft Session Agenda 1. System Engineering 1. Tolerance analysis 2. Requirements flow-down

More information

SPECTRAL SCANNER. Recycling

SPECTRAL SCANNER. Recycling SPECTRAL SCANNER The Spectral Scanner, produced on an original project of DV s.r.l., is an instrument to acquire with extreme simplicity the spectral distribution of the different wavelengths (spectral

More information

WFC3 Thermal Vacuum Testing: UVIS Broadband Flat Fields

WFC3 Thermal Vacuum Testing: UVIS Broadband Flat Fields WFC3 Thermal Vacuum Testing: UVIS Broadband Flat Fields H. Bushouse June 1, 2005 ABSTRACT During WFC3 thermal-vacuum testing in September and October 2004, a subset of the UVIS20 test procedure, UVIS Flat

More information

Simulation team in Vienna. Joao Alves, Werner Zeilinger, Rainer Köhler, Michael Mach

Simulation team in Vienna. Joao Alves, Werner Zeilinger, Rainer Köhler, Michael Mach The Simulation team in Vienna Kieran Leschinski and Oliver Czoske Joao Alves, Werner Zeilinger, Rainer Köhler, Michael Mach What is SimCADO? SimCADO is a python package which allows one to simulate mock

More information