Urmi Ray (Qualcomm) Sesh Ramaswami (Applied Materials) Richard Allen (SEMATECH) Chris Moore (Semilab) Standards Staff:

Size: px
Start display at page:

Download "Urmi Ray (Qualcomm) Sesh Ramaswami (Applied Materials) Richard Allen (SEMATECH) Chris Moore (Semilab) Standards Staff:"

Transcription

1 Background Statement for SEMI Draft Document 5409 NEW STANDARD: GUIDE FOR METROLOGY FOR MEASURING THICKNESS, TOTAL THICKNESS VARIATION (TTV), BOW, WARP/SORI, AND FLATNESS OF BONDED WAFER STACKS Notice: This background statement is not part of the balloted item. It is provided solely to assist the recipient in reaching an informed decision based on the rationale of the activity that preceded the creation of this Document. Notice: Recipients of this Document are invited to submit, with their comments, notification of any relevant patented technology or copyrighted items of which they are aware and to provide supporting documentation. In this context, patented technology is defined as technology for which a patent has issued or has been applied for. In the latter case, only publicly available information on the contents of the patent application is to be provided. Background This Guide, which is targeted at the 3DS-IC community, is intended to assist in the selection and use of Bonded Wafer Stack (BWS) metrology tools and to provide guidance in performing BWS measurements such as total thickness variation, bow, warp, and sori. In addition, the Guide provides examples of BWS measurements for these metrology tools and can assist wafer producers and users of BWS metrology to develop products and conduct meaningful evaluations. This document was developed in the Inspection and Metrology TF of the N.A. 3DS-IC Committee. The SNARF for this was approved April 3, Draft Document 5409 was approved for yellow ballot in Cycle 1 in CY2013, by the 3DS-IC Global Coordinating Subcommittee (GCS). Review and Adjudication Information Task Force Review Committee Adjudication Group: Inspection & Metrology Task Force North America 3DS-IC Committee Date: April 2, 2013 April 2, 2013 Time & Time zone: 8:00 AM to 10:00 AM, Pacific Time 3:00 PM to 5:00 PM, Pacific Time Location: SEMI Headquarters SEMI Headquarters City, State/Country: San Jose, California San Jose, California Leader(s): Victor Vartanian (SEMATECH) David Read (NIST) Yi-Shao Lai (ASE) Urmi Ray (Qualcomm) Sesh Ramaswami (Applied Materials) Richard Allen (SEMATECH) Chris Moore (Semilab) Standards Staff: Paul Trio (SEMI NA) /ptrio@semi.org Paul Trio (SEMI NA) / ptrio@semi.org This meeting s details are subject to change, and additional review sessions may be scheduled if necessary. Contact Standards staff for confirmation. Telephone and web information will be distributed to interested parties as the meeting date approaches. If you will not be able to attend these meetings in person but would like to participate by telephone/web, please contact Standards staff.

2 SEMI Draft Document 5409 NEW STANDARD: GUIDE FOR METROLOGY FOR MEASURING THICKNESS, TOTAL THICKNESS VARIATION (TTV), BOW, WARP/SORI, AND FLATNESS OF BONDED WAFER STACKS 1 Purpose 1.1 Control of parameters, such as bonded wafer stack (BWS) thickness, total thickness variation (TTV), bow, warp/sori, and flatness metrology, is essential to successful implementation of a wafer bonding process. These parameters provide meaningful information about the quality of the wafer thinning process (if used), the uniformity of the bonding process, and the amount of deformation induced to the wafer stack by the bonding process. Total thickness variation is also critical in certain bonded wafer manufacturing process steps, since non-planarity can lead to problems in subsequent processing steps, including lithographic overlay and intermittent electrical contact between metal layers in the bonded wafers. This Guide provides a description of tools that can be used to determine these key parameters before, during, and after the process steps involved in wafer bonding. 2 Scope 2.1 This Guide provides examples of the capabilities and limitations of various measurement technologies applicable to bonded wafer stacks (BWS) as well as their suitability for different applications. 2.2 The Guide describes metrology techniques that are applicable to both temporary and permanently bonded wafer stacks. 2.3 This Guide is complementary to existing SEMI Test Methods for measuring these parameters on single wafers, in some cases extending existing metrology techniques to a bonded wafer stack and in other cases describing metrology techniques specific to a bonded wafer stack. 2.4 The Guide focuses on general measurement techniques including IR laser profiling, white light confocal microscopy, visible and IR interferometry, capacitance, and back-pressure metrology. Each technology has unique strengths and weaknesses some rely on front-side illumination, others on back-side illumination. Some techniques can measure the thicknesses of individual layers in the bonded wafer stack, and some are additionally capable of measuring surface nanotopography. 2.5 The metrology examples provided in this Guide originated from industry experts and are believed to be representative of tool performance as of the year However, as tool and measurement techniques continue to evolve and improve, BWS measurement performance may surpass what is contained in this Guide. The user should investigate metrology suppliers current capabilities. 2.6 The measurements described in this Guide are on bonded wafer stacks with thickness in the range of 50 to 1550 µm. 2.7 The stacks considered include carrier and device wafers and bonding layers, including cases where there are more than two wafers in a stack. Bonded wafers may be classified as either temporarily bonded (i.e. a device to a carrier wafer) or permanently bonded. Temporary bonding uses a temporary adhesive; permanent bonding could be adhesive, oxide, metal-metal (e.g. Cu-Cu), or hybrid bonding. Two representative two-wafer stacks are depicted in Figures 1 and 2. The first stack (Figure 1) is a bonded pair of 775 µm thick wafers following TSV formation and the bonding operation. The second stack (Figure 2) is a bonded wafer stack with a top wafer thinned to ~50 µm, and bonded on top of a 775 µm wafer using a temporary adhesive. Page 1 Doc SEMI

3 Figure 1 Bonded Wafer Pair Following TSV Formation and Wafer Bonding Operation Figure 2 Temporarily Bonded Wafer Pair--Device Wafer is Edge-Trimmed and Thinned NOTICE: SEMI Standards and Safety Guidelines do not purport to address all safety issues associated with their use. It is the responsibility of the users of the Documents to establish appropriate safety and health practices, and determine the applicability of regulatory or other limitations prior to use. 3 Limitations 3.1 While this Guide provides an overview on the use of several available techniques, it does not provide the level of detail typically available in Test Methods. For this purpose, SEMI documents are referenced that the user may find of interest. In addition, all suppliers may not follow SEMI standard measurement methods in their metrology procedures. 3.2 The information in this guide does not encompass the establishment of specific a methodology for such measurements but is only a guide to the user describing the various types of metrology including their capabilities and limitations. 3.3 This Guide describes metrologies that have been applied to a wafer stack composed of, at most, two wafers. Thus, the user is cautioned that the methods described may be limited with applied to wafer stacks consisting of more than two wafers. 4 Referenced Standards and Documents 4.1 SEMI Standards and Safety Guidelines SEMI M23 Specification for Polished Monocrystalline Indium Phosphide Wafers SEMI M43 Guide for Reporting Wafer Nanotopography SEMI M49 Guide for Specifying Geometry Measurement Systems for Silicon Wafers for the 130 nm to 22 nm Technology Generations SEMI M59 Terminology for Silicon Technology SEMI M65 Specifications for Sapphire Substrates to Use for Compound Semiconductor Epitaxial Wafers SEMI MF533 Test Method for Thickness and Thickness Variation of Si Wafers SEMI MF534 Test Method for Bow of Silicon Wafers Page 2 Doc SEMI

4 SEMI MF657 Standard Test Method for Measuring Warp and Total Thickness Variation on Silicon Slices and Wafers by a Non-contact Scanning Method SEMI MF1390 Standard Test Method for Measuring Warp on Silicon Wafers by Automated Non-Contact Scanning SEMI MF1451 Test Method for Measuring Sori on Silicon Wafers by Automated Non-Contact Scanning SEMI MF1530 Standard test method for Measuring Flatness, Thickness, and Thickness Variation on Silicon Wafers by Automated Non-Contact Scanning NOTICE: Unless otherwise indicated, all documents cited shall be the latest published versions. 5 Terminology NOTE 1: Refer to the SEMI Standards Compilation of Terms (COTs) for a list of the current Abbreviations, Acronyms, Definitions, and Symbols. 5.1 Abbreviations and Acronyms AM acoustic microscopy BWS bonded wafer stack CD critical dimension CWS chromatic white light sensor CSI coherence scanning interferometry FAMM focus/acquire/measure/move FPD focal plane deviation HVM high volume manufacturing IR infra-red ISO international organization for standardization RPD reference plane deviation SLED superluminescent light emitting diode TIR total indicated runout TS through-silicon via TTV total thickness variation WLI white light interferometry 5.2 Bonded Wafer Metrology Overview explanation of wafer metrology terms 1 the parameters primarily measured, wafer bow, warp, and sori need some explanation. Figure 3 shows schematic diagrams representing these terms bow the deviation of the center point of the median surface of a free, unclamped wafer from a median surface reference plane established by three points equally spaced on a circle with diameter a specified amount less than the nominal diameter of the wafer; SEMI MF23 provides a standard test method for measuring the bow of a single wafer. Note that bow is a signed value. The method can be adapted to a bonded wafer stack warp the difference between the most positive and most negative distances of the median surface of a free, unclamped wafer from a reference plane; SEMI M59 provides a standard test method for measuring the warp of a single wafer. Warp can be zero, even for a wafer with curvature, if the curves are mirror images of each other. SEMI MF657 measures median surface warp using a three-point back-surface reference plane, resulting in thickness variation being included in the warp value. The use of a median surface reference plane in SEMI MF SEMI International Standards: Compilation of Terms (COT) Page 3 Doc SEMI

5 eliminates this effect. SEMI MF1390 provides a standard test method for measuring the warp of a single wafer. The method can be adapted to a bonded wafer stack sori the algebraic difference between the most positive and most negative deviations of the front surface of a wafer that is not chucked from a reference plane that is a least squares fit to the front surface within the fixed quality area flatness the deviation of the front surface, expressed in TIR or maximum FPD relative to a specified reference plane when the back surface of the wafer is ideally flat, as when pulled down by a vacuum onto an ideally flat chuck non-contact metrology that allows a wafer to be measured without physical contact to the wafer surface, preventing contamination or damage to the wafer substrate chromatic white light sensor a chromatic white light sensor (CWS) is based on the principal of confocal optics and relies on chromatic scanning. A lens is used that refracts white light differentially based on its wavelength in order to carry out distance measurements. Resolution depends on the intensity of reflected light interferometry a technique that relies on the principal of superposition of multiple beams of light to determine the effect that a material has on the state (phase and amplitude) of the original light beam. It is this introduced phase difference that creates the interference pattern between the initially identical waves. If a single beam has been split along two paths, then the phase difference is diagnostic of anything that changes the phase along the paths. This could be a physical change in the path length itself or a change in the refractive index along the path gravity compensation for horizontally-supported wafers unlike thickness and TTV measurements which are independent of how the wafer is held, bow and warp measurements are complicated not only by wafer stress but by how the wafer is supported and by gravity. Gravity causes substantial deformation of large diameter or thin wafers, whether they are supported at the edge or in the middle. Unless compensated, gravity will induce a large error in warp measurements. SEMI MF1390 describes three compensation approaches. One approach is to correct for the gravitational effect of warp measurements by inverting the wafer and measuring both the top and bottom surfaces of the wafer. Any differences between two values at the same site are due to the effect of gravity and can be used to correct for single-side measurements. Another approach is to use an analytical expression for gravitational deformation and subtract it from a single-side warp measurement 2. Measurements obtained on representative wafers can also be performed and the gravity value determined In addition, the wafer can either rest on a chuck or be supported by three points (MF657) gravity compensation for vertically supported-wafers on tools that support the wafer vertically, effects due to gravity are negligible and therefore do not require compensation. Bow is only measured on one side so there is no need to invert the wafer. The manufacturing tolerance for sori is very low, such that a 300 mm diameter silicon wafer must achieve sori in the nanometer range in order to achieve the maximum yields in semiconductor device processing. Figure 3 Wafer Warp, Bow, and Sori Depictions 2 2 W. R. Runyan, T. J. Schaffner, Semiconductor Measurements and Instrumentation, McGraw-Hill, 1998, pp Page 4 Doc SEMI

6 5.3 Measurement Considerations thickness and total thickness variation (SEMI MF533) Wafers that are excessively thin may break during processing or have inadequate thermal mass or electrical resistance during certain processing steps. Wafers that are too thick may jam during wafer handling. Excessive thickness variation may encounter problems during mechanical handling. Excessive deviation from flatness may cause focus problems in photolithography steps warp (SEMI MF1390) Warp can adversely affect the yield of semiconductor devices and can affect wafer handling and processing. Warp may be caused by unequal stresses on the two exposed surfaces of a wafer. Warp cannot be determined from measurements on a single exposed surface. The median surface may contain regions with upward or downward curvature or both. In some cases, the median surface may be flat. In all cases, warp is a zero or positive quantity. Warp is a measure of the distortion of a median surface of a wafer. SEMI MF657 measures median surface warp using a three-point back-surface reference plane. The back-surface reference plane results in thickness variation being included in the reported warp value. The use of a median surface reference plane in SEMI MF1390 eliminates this effect bow (SEMI MF534) If the median surface of a free, unclamped wafer has a curvature that is everywhere the same, bow is a measure of its concave (dished) or convex (mounded) deformation, independent of any thickness variation that may be present. Measurement of a semiconductor wafer, the deviation of the center point of the median surface of a free, unclamped wafer from a median-surface reference plane established by three hemispherical points equally spaced on a circle with diameter a specified amount less than the nominal diameter of the wafer sori (SEMI MF1451) Sori solves the ambiguity of warp by measuring the difference between the maximum and minimum distances from the front surface of the wafer to a reference plane outside the surface of the wafer. SEMI documents MF59, M65, and MF1451 provide a standard test method for measuring the sori of a single wafer. A wafer is supported on a small-area chuck with the front surface facing up. Both external surfaces of a wafer are simultaneously measured by an opposed set of probes to obtain a set of values at the same x and y coordinates of the distances between each surface and the nearest probe. The paired distances are used to construct the median surface. Gravity correction can be performed. One half of the thickness at each point is added to the corrected median surface to construct the corrected front surface. A least-squares reference plane is constructed from the corrected front surface. The reference plane deviation (RPD) is calculated at each pair of measurement points. Sori is then reported as the algebraic difference between the most positive RPD and the most negative RPD. 6 Bonded Wafer Metrology Techniques 6.1 Optical Techniques Chromatic White Light Chromatic white light sensor metrology utilizes the principle of wavelength-dependent focal length to determine distance (Figure 4). The spectrum of light reflected from a surface generates an intensity profile as a function of focal length that is used to determine distance to the sample surface. The peak in intensity occurs at the optimal focal point for each wavelength. This technique is useful for determination of 2D profile, 3D topography, planarity, roughness, and wafer contour (bow and warp). Figure 5 shows a typical set of measurement locations used in an analysis. The resulting 2D measurement profiles for thickness, TTV, and warp for an 825 µm thick bonded wafer are shown in Figure 6, and a wafer thickness contour map is shown in Figure 7. Page 5 Doc SEMI

7 Figure 4 Chromatic White Light Sensor Used for Distance Measurements for Bow, Warp, and Topographical Measurements NOTE: Focal lengths are wavelength dependent, providing distance measurements. Figure 5 Data Collection Locations Showing Profile (a-d), and Points (A-I) Page 6 Doc SEMI

8 Figure 6 Profile Measurements for Thickness, TTV, and Warp for 825 µm Thin Bonded Pair Wafer Figure 7 Wafer Map Results for 825 µm Thin Bonded Pair Wafer Infra-red Interferometry Interferometry is a technique in which light waves are superimposed to obtain information on surface profiles or displacements between two surfaces. Infra-red interferometry is useful for thickness measurements of substrates that are transparent in near IR light. Thickness measurements can be performed on single or multi-layer films, including adhesion layers in BWS. At each film interface, some incident light is reflected back to a detector When two light waves combine, the resulting pattern is determined by the phase difference between the two waves waves that are in phase undergo constructive interference while waves that are out-of-phase undergo destructive interference. Typically, a beam of coherent light is split into two beams by a beam splitter. Each beam travels a different path and the two are recombined before arriving at the detector. The path difference, the difference in the distance traveled by each beam, creates a phase difference between them For a thin transparent medium, the reflection from the top surface interferes with the reflection from the bottom surface. For some wavelengths, the interference is constructive, and for others it is destructive. In general, the reflected intensity follows the Fabry-Perot equation (eq. 1): Page 7 Doc SEMI

9 δ δ Where r 1 is the reflection coefficient at the first surface of the cavity, r 2 is the reflection coefficient at the second surface, and δ = 2 π n l / λ, n is the index of refraction, l is the cavity thickness, and λ is the wavelength. Because of the complex exponential, the reflected intensity is periodic in δ. Since δ is proportional to optical frequency (c / λ), the reflected intensity is periodic in optical frequency with period c / 2nl, where n is the index of refraction, l is the layer thickness, and c is the speed of light in a vacuum. Then the basic procedure to measure the thickness is to: 1) measure the reflectance for a spectrum of wavelengths; 2) analyze the measured spectrum for a periodic intensity as a function of optical frequency; 3) calculate thickness from the period and index of refraction The light source has a wavelength in the neighborhood of 1.3 µm, and is placed facing the back of the wafer. The sensor measures both surfaces of the wafer interferometrically and returns a thickness measurement (Figure 8). The fast (millisecond) and non-destructive measurements based on backside IR illumination allow immediate HVM process feedback. The sensor is used for measuring thickness of thick wafers, thinned wafers, and etch depth of TSVs and other features. (1) Figure 8 Infra-Red Interferometric Sensor Used for Distance Measurements for Bow, Warp, and Topographical Measurements NOTE: Focal lengths are wavelength-dependent, providing distance measurements Some tools utilize a dual-sided interferometer to measure wafer geometry parameters such as thickness, shape and flatness. The wafer is held vertically between 2 sets of optics so that each surface is measured simultaneously. The tool is mainly used by bare silicon wafer suppliers to improve their process for wafer geometry and to qualify wafers before shipment to customers Wafer geometry data acquired in 3D is capable of providing nanotopography and wafer edge roll-off. Wafer manufacturing issues can affect geometry at the edge of the wafer, causing die defocus and CMP removal non-uniformity The thickness and warp of the two individual wafers in a bonded pair can be measured using interferometry utilizing the interference between the light reflected from the wafer and the light reflected from a moving mirror. The general scheme of such an interferometer is shown in Figure 9. Page 8 Doc SEMI

10 Figure 9 Schematic View of an Interferometer System NOTE: Beam 2, which is reflected from the bottom surface of the wafer, is delayed with respect to beam 1 reflected from the top surface. The delay between beam 1 and beam 2 is proportional to the thickness of the wafer. In a real experiment, the incident beam is perpendicular to the surface of the wafer A SLED (superluminescent light-emitting diode) with a broad line width was used and retro-reflecting mirrors were mounted in a nano-motor stage. The dual-probe scheme for total thickness and warp measurement is shown in Figure 10. Fiber Reference Mirror 10X Microscope Objective Si Substrate Reference Mirror Bottom Probe Figure 10 Dual Probe Scheme for the Measurement of the Thickness and Warp Fiber In the dual-probe setup, each probe has a reference and a measurement arm. Light from each arm interferes with light from the scanning mirror to generate the first two interference signals in the upper probe. Similarly, light from the reference arm and the back surface of wafer produces the first two interference signals in the lower probe. The distance between the two probes can be determined using the known thickness of a block gauge. This information in conjunction with the two signals from the upper and lower probes makes it possible to measure the total thickness of the wafer. If the wafer surfaces are rough and/or the substrate is opaque to the incident light, the Page 9 Doc SEMI

11 dual probe is a good way to measure its thickness. A typical signal collected by an interferometer system on a bare Si wafer is shown in Figure 11. For warp measurement, the distance between the reference signal and the signal from the front surface of a wafer into the top probe is measured. In a warp measurement, a 300 mm blank wafer is used as the un-warped reference. Figure 11 Typical Signal from an Interferometer System In Figure 12, TTV maps are shown with measurement locations (left) as well as 2D TTV (center) and 3D TTV (right) maps for an 825 µm two wafer stack. Typical wafer thickness results are shown in Figures for a 1550 µm thick two wafer stack. Figure Point 2D and 3D Image Files of TTV for 825 µm Two Wafer Stack Page 10 Doc SEMI

12 Figure µm Two Wafer Stack Results Showing Bottom Wafer Thickness Map Figure 14 Two Wafer Stack Results Showing Top Wafer Warp Map Figure µm Two Wafer Stack Results Showing Top (Thinned) Wafer Thickness Map Page 11 Doc SEMI

13 6.1.3 Coherence Scanning Interferometry (CSI) or White Light Interferometry (WLI) White light interferometry 3 is a non-contact optical profiling system for measuring step heights and surface roughness in precision engineering applications. The technology utilizes a white light beam which passes through a filter and then a microscope objective lens to the surface of the wafer. The light reflecting back from the surface is combined with the reference beam and captured for software analysis. After obtaining data for each point, the system can generate a 3D image (topography) of the surface. The vertical (i.e. height) resolution of this technique is extremely good, better than 0.01 nm (0.1 Angstrom), which makes it a potentially practical tool for assessing wafer surfaces for roughness and nanotopography. However, the lateral resolution can be limited to the spot size, in the range of 0.35 µm. With its broad capability, it is possible to measure local step height, critical dimensions (CD), overlay, multilayer film thickness and optical properties, combined topography and film thickness, and wafer bow. The technology has particular strength in new advanced packaging applications for process control of throughsilicon vias, microbumps, redistribution layers, and copper pillars Figure 16 illustrates the principle of CSI. A broadband light source produces a narrow region of optical interference co-located with the focus of the interferometer microscope objective. As the objective scans in Z, orthogonal to the surface, interference is detected by the imaging camera as the measured sample comes into focus. At focus, each pixel detects the interference peak. Analysis software produces a map of surface height variation from the pixel-by-pixel interference peak detection. A key CSI advantage is that Z resolution is independent of field-of-view and lateral resolution. This is not the case for confocal microscopes, where Z resolution decreases with increasing field-of-view due to decreased numerical aperture. The minimum surface feature height measurable by CSI depends on the manufacturer s measurement algorithm, and ranges from ~10 nm to <0.1 nm. CSI imaging resolution depends on the numerical aperture (NA) of the objective used and the number of pixels in the camera array. The area, or field-of-view measured depends on the objective magnification, which may range from 2.5X- 100X. Nominal resolution and field-of-view combinations are 0.7 µm with 500 µm field-of-view to 5 µm with 2.5 mm field-of-view. Nominal resolution for CSI is two times worse than optical imaging microscopes as expected for a given NA. CSI systems report height, whereas image microscopes report contrast. It is possible to resolve features in an imaging microscope until contrast equals the background noise. CSI must separate height features, which is only possible at twice imaging microscope resolution. Figure 16 Coherence Scanning Interferometer for the Measurement of Surface Topography NOTE: Z height is measured at each pixel as the interferometric objective is scanned in Z and the camera detects the peak of the coherence envelope as shown. 3 Blunt, R.T., White light interferometry a production worthy technique for measuring surface roughness on semiconductor wafers, CS MANTECH Conference, April 24-27, 2006, Vancouver, British Columbia, Canada. Page 12 Doc SEMI

14 CSI can measure polished and rough surfaces, as well as flatness to nm to mm high stepped and sloped surfaces. When combined with image analysis software, CSI can measure and position three-dimensional features to sub-micron repeatability in X and Y, combined with sub-nanometer repeatability in Z. Typical applications are surface roughness (commercial tools provide industry standard results, but often lack ISO specified software bandpass filters), nanotopography, local step height, critical dimensions (CD) and overlay. The technology has particular strength in new advanced packaging applications for process control of through-silicon vias (TSVs), microbumps, redistribution layers, copper studs and pillars A major strength of CSI is ease-of-use and speed of measurement. Advanced tools incorporate autofocus and automation capabilities to speed the measurement process. Typically no special preparation is required of the measurement sample, if the sample is homogeneous and without film layers. Microbump/pillar measurements have been demonstrated in production with 0.5 second Focus/Acquire/Measure/Move (FAMM) times, while maintaining nanometer Z performance. Laboratory tools can experience one to five-minute FAMM s primarily due to instrument set up time The presence of films and films stacks are both an opportunity and technology limitation of CSI. Present CSI algorithms are limited to reliably measuring in the presence of films to thicknesses greater than 1.5µm optical thickness (thickness times index of refraction). Some manufacturers claim this capability down to 1-µm optical thickness. Reported measurement results are film thickness (if the index of refraction is known), surface profile of the top and also the bottom of the film. Advanced work is occurring to measure features in thin film structures, improved CD and other edge of resolution features. As these technologies emerge new applications will be possible. When films are less than the measurable optical thickness unpredictable errors occur in the measured data, and the data is unreliable in Z height Measuring non-homogeneous materials limits CSI accuracy. CSI measures Z height by measuring interferometric phase. Phase is dependent on the material measured. Insulators experience a constant 180 phase shift, whereas conductors and semi-conductors induce varying phase shifts. These phase shifts appears as surface topography errors up to 10 s of nanometers. It is possible to correct for these by identifying material types by regions and region-by-region apply a correction factor. Some manufacturers have seen success correcting algorithmically but this has seen little commercial utilization. For process control, phase change on reflection is a constant offset and therefore can be ignored when process variation is being tracked when absolute values are not critical CSI requires minimal to no surface preparation. Rapid measurements, nanometer level Z resolution and optical level image resolution make it a preferred tool for many applications. Measurement errors due to films and phase change of reflection of non-homogeneous materials are major limitations for applications based on semiconductor processes Five sets of 49 wafer bow measurements (top and bottom) for a 1550 µm bonded wafer pair are averaged in Figure 17, and the 3 sigma standard deviation is shown in Figure 18. Page 13 Doc SEMI

15 Figure 17 Results for 49 Point Dynamic Repeat Bow Measurement on a 1550 µm Bonded Wafer Pair (3σ, 5 Repeats) Average Bow of Top and Bottom Measurement (3 Point Suspension) Figure 18 Results for 49 Point Dynamic Repeat Bow Measurement on a 1550 µm Bonded Wafer Pair (3σ, 5 Repeats) Average Standard Deviation of Top and Bottom Measurement (3 Point Suspension) Laser Profiling The diagram in Figure 19 demonstrates the principal of laser profiling with a single optical head where the laser is focused first on the top and then on the bottom of the wafer. The laser peak positions that are used to determine the wafer thickness are also shown Two focused laser beams can also be used, one focused on the top and the other on the bottom of the wafer stack (Figure 20). The measurement is made by moving the focus position from the surface of the top wafer to the top surface of the bottom wafer. The laser can be focused in the objective to a spot as small as 1 µm. The dual optical measurement system provides accurate wafer thickness measurements independent of material properties, especially useful for patterned or bumped wafers, GaAs and other wafer types, and after back grinding and dicing. Laser profiling provides thickness resolution of approximately 0.1 µm and has also been applied to TTV, bow, warp, and surface roughness. Page 14 Doc SEMI

16 SIGNAL DETECTION MODULE OPTICAL THICKNESS MODULE DETECTOR TOP OF WAFER BOTTOM OF WAFER OPTICAL FIBER PINHOLE FOCUSING LENS BEAM EXPANDER BEAM SPLITTER OBJECTIVE MEASUREMENT SAMPLE Figure 19 Principle of Single Beam Wafer Thickness Measurement System Showing Wafer Top and Bottom Signal Peak Positions Figure 20 Dual-Beam Thickness Description The plot in Figure 21 is a plot of the laser focused on the wafer surface as well as a glass surface. The difference in the in-focus locations equals the thickness of the glass. Page 15 Doc SEMI

17 Figure 21 Example Plot Thickness Measurement Figures illustrate measurements for a set of five static repeatability runs obtained on an 825 µm thick bonded wafer pair stack using laser profiling. A 2D thickness map is shown for each set of measurements in Figure 22, as well as 3D thickness maps for the five runs (average and 3σ sigma standard deviation) in Figures Figure 23 and 24. Two-dimensional bow and warp maps are shown in Figures 25 and 26. Page 16 Doc SEMI

18 Figure 22 Bonded Pair Wafer Thickness Map (Five Repeats) Figure 23 Bonded Wafer Pair 25 Site Average Thickness Map (Five Repeats) Page 17 Doc SEMI

19 Semiconductor Equipment and Materials International Figure 24 Bonded Wafer Pair 25 Site Thickness Map (3 σ Std. Dev., Five Repeats) Figure 25 Bonded Pair Wafer Bow Map (Five Repeats) Page 18 Doc SEMI

20 Figure µm Bonded Pair Wafer Warp Maps (Five Repeats) 6.2 Electrical Technique for Measuring Proximity Capacitance Displacement Metrology Capacitive sensors can be divided into two categories based upon their performance and intended use. High resolution sensors are typically used in displacement and position monitoring applications where high accuracy, stability and low temperature drift are required. Quite frequently, these sensors are used in process monitoring and closed-loop feedback control systems. Proximity type capacitive sensors are typically used to detect the presence of a part or used in counting applications. The following describes characteristics of high resolution systems, their operating principle and application The capacitance sensor shown in Figure 27 is used throughout a variety of industries to provide highly stable, accurate measurements of displacement, vibration, position, thickness and runout. Capacitance probes are typically modeled as a parallel plate capacitor. If two conductive surfaces are separated by a distance and a voltage is applied to one of the surfaces, an electric field is created. This occurs due to the different charges stored on each of the surfaces. Page 19 Doc SEMI

21 Figure 27 Typical Capacitance Sensor Capacitance refers to the ability of the surfaces to hold a charge. In a typical sensor system the probe is one of the plates and the target being measured is the other plate. If a constant current is applied, the capacitance change can be monitored as a linear voltage charge related to the distance between the plates (Figure 28). Figure 28 Typical Capacitance Sensor Measurement Parameters This distance, or gap, is a function of the area of the capacitance sensor according to the following equation (Eq. 2): (2) Where capacitance is the area of sensor, A = πr 2 times the dielectric constant of air, ε, divided by the gap, d From this relationship, capacitance is directly proportional to the area of the sensor and the dielectric property of the material between the sensor and target (typically air). The greater the area of the capacitance sensor the larger the measurement range, or gap. If it is assumed that the area and the dielectric constant between the plates remain constant for a specific probe, any change in capacitance is inversely proportional to the change in distance between the probe and target being measured and this change is converted to a voltage for monitoring. The amount of voltage output change for a given distance change is commonly referred to as the sensitivity of the system. For example, if a distance change of 1 mm corresponds to a voltage output change of 10 V the sensitivity would be 1 mm/10 V, or 0.1 mm/v. Page 20 Doc SEMI

22 6.2.2 Capacitance Sensors Electric Field A typical high performance capacitance sensor consists of three basic elements: the sensor tip, the guard and the ground shell. Figure 29 illustrates a typical capacitance probe. When a voltage is applied to the sensor tip, an electric field is established between it and any other local conductive material. To maintain accuracy and linearity it is essential that the electric field in the measurement area be linear, directed toward the target, and not distorted. To protect this field, each capacitive sensor has a guard electrode. This creates an additional field around the probe sensing tip that is driven at the same phase and voltage potential as the sensor. By being equal, the auxiliary field protects the area from becoming warped and cancels any stray capacitance between the two elements Because the ground shell is at a different potential than the guard, a partial distortion of the guard field to the ground shell occurs. Although undesirable, the distortion of the guard is acceptable as long as the sensor tip field remains linear. For best performance, the width of the guard should be at least 2X the system measurement range. Figure 29 Capacitance Sensor Field Capacitance probes should be designed with sufficient guard electrodes to protect the sensing area under normal operating conditions. This probe s range should not be greater than the width of the guard electrode or poor linearity will result In addition to improving linearity and accuracy, the guard is also used to reduce noise and external interference. Each capacitance probe is driven by a low noise coaxial cable. The shield of the cable is used to deliver the voltage to the guard, at the same voltage and phase. This eliminates any stray capacitance that might be created between the center conductor and the shield of the cable, or any other part that may be close to the cable. By design, this protection significantly reduces external influences from RFI and EMI. It is important to note that ordinary coaxial cable usually does not provide adequate protection or shielding for the system and special cable is generally required Characteristics of Capacitive Sensors Non-Contact Capacitive displacement sensors are non-contact by design. That is, they are able to precisely measure the position or displacement of an object without touching it. Because of this the object being measured will not be distorted or damaged and target motions will not be dampened. Additionally, they can measure high frequency motions because no part of the sensor needs to stay in contact with the object, making them ideal for vibration measurements or high speed production line applications Range/Standoff Distance As mentioned above, the range of a capacitance sensor is dictated by the diameter, or area, of the sensor. The larger the area, the larger the measurement range. Measurement range is typically specified starting when the probe is touching the target. At this point the output from the amplifier is 0 V. When the gap is increased to equal the full scale measurement range of the capacitive system the amplifier output is 10 V (V dc ). In theory, the probe can operate anywhere between these two extremes, however, it is not recommended to operate below 10% of the gap. Thus, the ideal operating or standoff distance is between 5 V and 7 V dc, permitting the target to move closer to or Page 21 Doc SEMI

23 further away from the probe without going out of range. Figure 30 is a simplified diagram showing range, output voltage and recommended standoff for a typical capacitance sensor Resolution Figure 30 Capacitance Probe Operating Range The resolution of a displacement sensor is defined as the smallest amount of distance change that can be reliably measured by a specific system. Capacitance sensors offer extremely high resolution and stability, often exceeding that of complex laser interferometer systems. Because of their ability to detect such small motions, they have been successfully used in many demanding measurement applications including computer disk drive runout, microscope focusing and nano-positioning within highly complex photolithography tools The primary factor in determining resolution is the system s electrical noise. If the distance between the sensor and target is constant, the voltage output will still fluctuate slightly due to the white noise of the system. It is assumed that, without external signal processing, one cannot detect a shift in the voltage output of less than the random noise of the instrument. Because of this, most resolution values are presented based on the peak-to-peak value of noise and can be represented by the following formula: Resolution = Sensitivity Noise Sensitivity is simply the measurement range divided by the voltage output swing of the capacitance amplifier, that for a fixed sensitivity the resolution is solely dependent upon the noise of the system, so that the lower the noise, the better the resolution It is important to note that some manufacturers specify resolution based on peak or rms noise, resulting in claims that are 2X and 6X respectively better than peak-to-peak. Although an acceptable method, it is somewhat misleading as most users do not have the ability to measure voltage changes less than the peak-to-peak noise value Bandwidth The bandwidth, or cutoff frequency, of a system is typically defined as the point where the output is dampened by 3 db. This is approximately equal to an output voltage drop of 30% of the actual value. In other words, if a target is vibrating with an amplitude of 1 mm at 5 khz and the bandwidth of the capacitance sensor is 5 khz the actual sensor output would be 1 mm 70% = 0.7 mm Push or Range Extension Typical capacitive amplifier systems operate over a specific capacitance range, limiting their ability to measure large motions or operate at comfortable standoff distances. To overcome this problem some suppliers use a proprietary circuit that, with minor component modifications, can be adjusted to change the range and meet a wider variety of customer requirements. For example, a small diameter probe with a ½ mm measurement range can be pushed to have a measurement range of 1 mm or even 2 mm. This allows capacitive probes to be used in applications where space is limited or the target being measured is small. It is important to note, however, that a pushed probe should have a guard width sufficient enough to maintain the performance required, as mentioned Page 22 Doc SEMI

24 above. Additionally, pushing a sensor also amplifies system noise, reducing probe resolution. Noise increase is proportional to push; 2 Push = 2 Noise Spatial Resolution The field established between a capacitance probe tip and measured object is typically larger than the diameter of the probe tip (Figure 31). This is because there is an epoxy gap between the tip and guard elements. The field diameter is equal to D plus 1X the epoxy gap. When obtaining measurements, capacitance probes provide a distance equal to the average surface location within the spot area. They are not capable of accurately detecting the position of features smaller than the size of the spot. However, they can repeatedly measure rough surfaces. Because of this, the probe tip should always be 25% smaller than the smallest feature targeted for measurement. Smaller sensors can distinguish smaller features on an object. Figure 31 Effective Spot Size of a Capacitive Sensor Linearity Capacitance sensors may have an output of 0 10 V dc over the full scale measurement range (FSR). In an ideal world this output would be perfectly linear and not deviate from a straight line at any point. However, in reality there are slight deviations from linearity. Typically, linearity is specified as a percentage of the Full Scale Measurement range. During calibration the output from the amplifier is compared to the output of a highly precise standard and differences are noted. Some capacitance systems exceed ±0.05% FSR with some achieving ±0.01% or better Accuracy is a function of linearity, resolution, temperature stability and drift, with linearity being the majority contributor. Calibration reports provide data that can be used to correct for the non-linearity of a system with inexpensive computers and correction software. Digital correction typically yields a linearity of ±0.01% or better Stability Stability is a function of a variety of different internal and external factors. For short term or relative measurement applications stability is typically not an issue. However, if high accuracy is required over a long period of time, care must be taken when designing fixtures, selecting components and specifying materials of construction Temperature is typically the biggest factor that affects stability. Temperature swings not only cause electronic drift but can also cause fixture and probe expansion and contraction. For critical applications high-quality capacitors, resistors and inductors specifically designed for stability to minimize the electronic drift should be used. To minimize mechanical drift, probes can be manufactured from special low thermal coefficient materials such as Invar. Thermal correction coefficients can also be provided and used for real-time compensation Active capacitance probe systems should never be used in high stability applications because any localized temperature change surrounding the sensor will result in drift. Page 23 Doc SEMI

25 Calibration For low-end proximity sensors calibration is typically not important because linearity of the sensor is not critical. Most high performance systems are, by design, inherently linear to approximately ±0.1% of the full scale measurement range. Some capacitance manufacturers offer sensors with this performance, however, they are typically not suitable for high precision applications. During calibration, the manufacturer adjusts circuit gain, offset, and typically performs a proprietary linearity adjustment. Improved performance can be obtained by adding adjustable break point linearization circuitry within the amplifier circuit or by providing digital correction During calibration the output of the amplifier verses position of a target is recorded. A best fit straight line is generated based on this data. Each recorded point is then compared to the generated straight line and the percent deviation is calculated and plotted. Based upon the results adjustments can be made to improve the deviation to within acceptable limits Applying Capacitive Probes Target Material and Grounding A capacitance measurement system mimics a parallel plate capacitor with the sensor as one plate and the target being measured the other. To create the electric field between the two plates the target must be made of a conductive material. The composition or thickness of the target is not important, allowing them to be used in many applications not suitable for eddy current type sensors. In fact, the surface can even be a few hundred ohm-cm To complete the capacitance circuit the target should be grounded back to the amplifier. For optimal performance a conductive path is required, however, capacitive coupled targets can work well if the capacitance is 0.01 µf or higher. An example of a capacitively-coupled target is a shaft rotating on air bearings. In theory, air bearings are non-contact but the gap between elements is small, and their area is relatively large, creating a high capacitance path. Thousands of successful applications world-wide have been installed with this type of ground If the target is poorly grounded, the system is susceptible to external noise and interference. Care should be taken when designing the ground return path Target Size The lines of flux in the electric field established between the probe and target always leave the capacitance sensor normal (90 ) to its surface and always enter the target normal to its surface. If the target being measured is large enough, and the sensor is within range, the field within the sensing area will be consistent and linear. If the target is not large enough to support the field it will tend to wrap around the edge and enter normal to the target side (Figure 32). This field distortion will create measurement errors by degrading the sensor linearity and changing its measurement range. Because of this the original factory calibration can no longer be used and an inplace calibration is required, however, accuracy may still be compromised. Figure 32 Field Distortion from an Insufficient Target Size Page 24 Doc SEMI

26 Targets that are too small to support the electric field can also provide false displacement signals from lateral motions. Capacitive sensors are typically used to measure the gap or movement in the direction of the sensor axis. If the target is large enough any lateral motion will not distort the electric field, however, if the field is wrapped around the target any lateral motion will change the shape of the field causing a change in output even if the gap remains constant. As a general rule, the target should be 30-50% larger than the capacitance sensor Target Shape Capacitance probes will measure the average distance to the target under the area of the sensor. If a tilted or curved target is being measured the electric field will be distorted and the accuracy compromised. When sensors are calibrated to a flat target the output voltage is theoretically zero when the probe is in contact with the target. This is not true when measuring curved or tilted surfaces because the surface prevents the probe from full target contact. The result will be a shift in the zero point from its original calibration which will be reflected as an offset in the measurement, not a sensitivity change. To overcome both issues an in-place calibration is possible to correct for the sensitivity change, however, the sensor measurement range may be reduced. As a general rule, a curved target should be 10 times larger in diameter than the sensing element of the capacitance sensor Spatial Resolution Capacitance sensors have a relatively large sensing area in relationship to their measurement range. As mentioned above, these types of sensors take an average measurement to the surface in question. If this surface has features that are smaller than the sensing element the feature may not be detected or the sensor output may not respond accordingly. Figure 33 shows how the probe size can affect the sensor output when measuring a stepped object, demonstrating sharper voltage changes from a smaller diameter sensor (Sensor C). Figure 33 Spatial Resolution Similarly, the output will depend on the surface roughness. If the roughness changes over an area the output from the capacitance sensor will change when the target translates beneath the probe because the average distance to the surface has changed. The amount of sensor output shift will depend on the magnitude of the surface roughness Environmental Conditions Capacitance changes as the distance to the target changes and also depends upon the dielectric property of the material in the gap. Because of this, it is important that there is a homogeneous, non-conductive material between the probe and target. In most applications, this material is air, however, many times oil or some other dielectric fluid is successfully used. If it is not homogeneous, or if the dielectric properties in the gap change, then accuracy will be affected. This is typically not a problem for changing air properties because the effects are small. Page 25 Doc SEMI

27 For instance, the dielectric constant of air changes by ~1.4 ppm/1% of relative humidity. This represents a potential offset in the senor output of less than 1 mv with a relative humidity change of 50%. Care must be taken to avoid dielectric changes from other materials and ensure that dirt and debris do not accumulate in the capacitive probe gap The most common environmental problem that can affect the accuracy of a capacitive sensor is temperature. Not only do the electronics exhibit temperature drift but also expansion and contraction of the probe and fixturing physically changes the probe gap. Custom probes manufactured of highly stable materials, such as Invar, are available for extreme stability applications. It is also possible to adjust the temperature coefficient of components for custom high-stability applications Advantages and Disadvantages Advantages As with any sensing technology, capacitive systems have both advantages and disadvantages. Perhaps their greatest attribute is their ability to resolve measurements below one micro-inch (< 25 nm), at a fraction of the cost of other high performance technologies. Most are "passive" by design allowing them to be used in extreme environments while still maintaining stability. Sensors can be easily customized, allowing them to be adapted into a variety of applications or settings. They are immune to target composition and work equally well on all conductive targets, unlike eddy current probes. They are largely immune to ultrasonic noise, electromagnetic fields, lighting conditions, humidity and temperature Disadvantages Capacitance technology dictates that the probe be mounted close to the target. This increases the probability of crashing the sensor or damaging the material being measured. Some suppliers have provisions to extend the measurement range and standoff of the sensor, however, this distance is rarely greater than 15 mm Capacitance sensors should also be kept clean. Dirt or other foreign debris can cause an offset in the measurement so frequent cleaning may be required depending on the application Applications Thickness Measurements Thickness quality control monitoring is better applied on line during the manufacturing process instead of periodic sampling after a product has been manufactured. This way process adjustments can be made on the fly, reducing or eliminating the continued production of product that does not meet specification. In some applications, contact methods can be utilized. However, they are slow, can damage the product and are subject to wear. Noncontact sensors are commonly used in these applications A typical application consists of two capacitance probes, one on either side of the material being measured. The difference between the output of each sensor is directly related to the thickness of the material being measured. By taking a differential measurement, any positional movement of the material within the probe gap is cancelled Figure 34 shows a typical thickness setup with A and B representing the sensor outputs. The gap between the probes, G, is equal to A + B + T. If an initial sample of known thickness, T, is placed within the gap, G can be determined and used in future calculations. Since thickness = G (A + B), and G is constant, the thickness can be calculated by simply subtracting the two outputs. Page 26 Doc SEMI

28 Figure 34 Typical Thickness Measurement Using Dual Capacitance Probes Single-sided thickness measurements can also be successfully made using capacitive sensors if the back side of the material being measured can be referenced to some fixed plane. Figure 35 shows a typical single-sided measurement. From this figure it is apparent that the product thickness is directly proportional to the gap between the probe and the surface of the material. Figures 36 and 37 show a typical bonded pair wafer bow map and raw thickness data from edge (point 0) to center of the wafer. Figure 35 Single Point Thickness Measurements Page 27 Doc SEMI

29 Figure µm Bonded Pair Wafer Bow Map Figure µm Bonded Pair Raw Thickness Data 6.3 Pneumatic Technique for Measuring Proximity The pneumatic technique for measuring proximity relies on the dependence of backpressure in an orifice conducting flowing gas toward a flat surface on the proximity of the surface to the orifice. Dual sensors allow differential measurements Differential Backpressure Metrology With a dual backpressure sensor, one can measure thickness (similar to a white light confocal sensor), as well as bow and warp. The advantage of differential backpressure sensing technology is that it works nearly independent of any surface condition (smooth or rough) or material property (conductive or non-conductive) requiring only a low supply pressure (10 psi or less) and only CDA or N 2. In this technique, back-pressure is converted to voltage; a calibration curve is generated and the operating point is determined on the curve. A Page 28 Doc SEMI

30 reference backpressure measurement is made on a gauge block prior to every wafer measurement to compensate for ambient temperature variation effects An air gauge is a broader terminology used for this technology, but in a broader sense, it includes any kind of sensing set up using a pressure sensor to detect a backpressure. Usually it uses higher pressure, has poor resolution, and is not stable. Differential backpressure technology is part of the air gauging technology but requires the use of specific sensor layout (Wheatstone bridge equivalent) to meet the characteristics needed for its use in the wafer industry Differential backpressure sensing technology can be used to achieve non-contact, high-resolution wafer thickness measurements. The differential backpressure sensors (PEL sensors) use clean, dry, compressed air or an inert gas, such as nitrogen The PEL sensors utilize a pneumatic Wheatstone bridge arrangement (Figure 38) with equal flow restriction and a low-pressure differential sensor. The sensing orifice uses a calibrated high-precision sapphire nozzle. Any small variation of the distance between the sensing nozzle and the target is detected by the PEL sensor and converted into a corresponding output voltage. The PEL sensor includes built-in temperature compensation and voltage regulation to provide extremely high stability over time and resolution down to 0.02 µm. Figure 38 Backpressure Sensor Configuration The nominal distance between the sensing nozzle and the target is set at approximately 100 µm. This is the optimal distance to operate within the linear portion of the backpressure response curve with the highest sensitivity. Under normal operating conditions, the distance between the sensing nozzle and the target may vary by ±20 µm around the preset nominal distance. With a response time of 5 ms the sensor is ideally suited for relatively fast measurement operations of fixed targets Source gas is supplied to the sensor at 10 to 12 psi (P S ). Any variation in the backpressure (P B ) resulting from the changes in the distance (X) between the sensing nozzle (R x ) and the gauged object (Figure 38) will generate a change in the output voltage at a rate of 50 mv/µm. Adjusting the setting of the valve (R V ) provides a wide range of settings and sensitivities that can vary the pressure reference (P R ) Measurement System Figure 39 shows a block diagram of the sensor subassembly and the associated interconnecting electronic and pneumatic controls. The servomotor in the measurement head is a voice coil that moves the probe relative to the sample such that the backpressure reading falls on the calibration curve in the linear working region. Page 29 Doc SEMI

31 Figure 39 Measurement System Block Diagram The Encoder Interface conditions the signal from the Encoder, which determines the position of the probe. The Backpressure Sensor converts the backpressure to a voltage output to be supplied to the computer. The computer processes all data from the measurement head as well as from the robotic handling system and the Y- Theta or X-Y stage to control the three component subassemblies Thickness Calculation The system employs a measurement sensor on the underside of the workstation as well as the topside. From the drawing in Figure 40 the thickness, t, of a sample is calculated in Equation 3 as: t = t 0 + (D-D 0 ) (a-a 0 ) (b-b 0 ) (3) Where: t 0 is the thickness of the gauge block, D and D 0 are the relative positions of the probe, a 0 is the distance from the top probe tip to the gauge block, a is the distance from the top probe tip to the sample, b 0 is the distance from the bottom probe tip to the backside of the gauge block, and b is the distance from the bottom probe tip to the back of the sample. Page 30 Doc SEMI

32 Figure 40 Probe Measurement Technique An example of a thickness map using differential back-pressure monitoring is shown in Figure 41. The mean thickness is 872 µm. Figure 41 Thickness Map Obtained by Differential Back Pressure Monitoring An example of a warp map using differential back-pressure monitoring is shown in Figure 42. The warp values are for an 875 µm two wafer stack. Page 31 Doc SEMI

33 Figure 42 Warp Map Obtained by Differential Back Pressure Monitoring 6.4 Acoustic Microscopy for Measuring Bonded Wafer Thickness A-Mode Scanning Acoustic data can be collected at the smallest X-Y-Z region defined by the limitations of the given acoustic microscope. An A-mode display contains amplitude and phase/polarity information (Figure 43 on left) as a function of time of flight at a single point in the X-Y plane, as shown in Figure 43 (Figure 43 on right). Figure 43 Example of an A-Mode Scan and Spectrum B-Mode Scanning Acoustic data can also be collected along an X-Z or Y-Z plane versus depth using a reflective acoustic microscope, as shown in Figure 44 on the left. A B-mode scan, shown in Figure 44, contains amplitude and phase/polarity information as a function of time of flight at each point along the scan line. A B-mode scan furnishes a two-dimensional (cross-sectional) description along a scan line (X or Y), as shown in Figure 44 on the right. Page 32 Doc SEMI

34 Semiconductor Equipment and Materials International Figure 44 Example of a B-Mode Scan and its Resulting Image C-Mode Scanning Acoustic data collected in an X-Y plane at depth (Z) using a reflective acoustic microscope is shown in Figure 45 on the left. A C-mode scan contains amplitude and phase/polarity information at each point in the scan plane. A C-mode scan furnishes a two-dimensional (area) image of echoes arising from reflections at a particular depth (Z), as shown in Figure 45 on the right. Figure 45 Example of a C-Mode Scan and its Resulting Image While acoustic microscopes are typically used to image and assess the bond quality between bonded wafers, the intrinsic data used to image the bond quality also contains time/distance information (Figure 46 on left). In order to image at a particular interface of interest, the time/distance to that interface is also known at each X-Y location and is displayed in an A-Scan waveform format. The A-Scan at each X-Y location can be processed to determine the thickness of each layer at that point and the surface contours of the front, internal interface or back of the wafer stack (Figure 46 on right). Page 33 Doc SEMI

35 Figure 46 Typical A-Scan Waveform Display (Left) and C-Scan Showing the Bond Quality at Several Layers of a Wafer Bond Stack Thickness Measurements with AM NOTE: White and light gray areas indicate delamination present at several layers As mentioned above, the A-Scan waveform consists of time and material interface data at an X-Y location on a bonded wafer. Knowing the speed of sound within a material converts the time data to a distance between one material interface and another, as shown in Figure 47 on the left. The time/distance data can be processed to color code the defects at different layers, as shown in Figure 47 on the right. Figure 47 An Example of a Color Coded Profile Image That Maps the Distance from the Surface of a Part to Anomalies at a Specific Depth Within a Stacked Wafer NOTE: All anomalies of the same color are at the same depth and the color map indicates that the blue anomalies are closer to the top surface, deeper green and orange being the deepest in the stack In addition, thickness measurement of a layer within a wafer stack can be obtained, as shown in Figure 48. In this image, the C-scan image is shown on the left, while the A-Scan waveforms showing the full bonded wafer thickness, the delamination region, and defective regions within the stack are shown on the right. The data shown in Table 1 shows the time and distance measurements, based on the speed of sound of the material, made on a sample at five points of interest. Page 34 Doc SEMI

36 Figure 48 C-Scan Image of a Stacked Wafer Showing 3 A-Scan Waveforms NOTE: Waveform #1 shows the complete thickness of the stacked wafer. Waveforms #2 and #3 show the distance from the top surface to the delamination and anomalous regions within the wafer stack. Table 1 Example Table of Various Thicknesses Measured at 5 Positions on a Sample. Waveform Thickness (ns) Thickness (µm) Bonded Wafer Thickness, TTV, Warp, Bow, Sori, and Flatness Metrology with Acoustic Microscopy (AM) Acoustic microscopy (AM) collects data similar to a depth finder in that it can measure the transit time for sound from a reference point, in this case the transducer, to the first surface it sees. The contour of that surface over the X-Y area scanned can then easily be displayed in a simple color coded map or process further to provide specific warp, bow, sori or global flatness data. In Figures 49 and 50, examples are provided of both types of data displays for 3D measurement of wafer stack surface contours. Page 35 Doc SEMI

37 Figure 49 Example of an AM Surface Flatness Color-Coded Display with Data Overlaid NOTE: In this case the wafer stack is bowed 140 µm from center to edge. Figure 50 Example of a 3D AM Surface Flatness Map with the Data Processed and Displayed. Page 36 Doc SEMI

Optical Characterization and Defect Inspection for 3D Stacked IC Technology

Optical Characterization and Defect Inspection for 3D Stacked IC Technology Minapad 2014, May 21 22th, Grenoble; France Optical Characterization and Defect Inspection for 3D Stacked IC Technology J.Ph.Piel, G.Fresquet, S.Perrot, Y.Randle, D.Lebellego, S.Petitgrand, G.Ribette FOGALE

More information

Technical Explanation for Displacement Sensors and Measurement Sensors

Technical Explanation for Displacement Sensors and Measurement Sensors Technical Explanation for Sensors and Measurement Sensors CSM_e_LineWidth_TG_E_2_1 Introduction What Is a Sensor? A Sensor is a device that measures the distance between the sensor and an object by detecting

More information

A Laser-Based Thin-Film Growth Monitor

A Laser-Based Thin-Film Growth Monitor TECHNOLOGY by Charles Taylor, Darryl Barlett, Eric Chason, and Jerry Floro A Laser-Based Thin-Film Growth Monitor The Multi-beam Optical Sensor (MOS) was developed jointly by k-space Associates (Ann Arbor,

More information

Bringing Answers to the Surface

Bringing Answers to the Surface 3D Bringing Answers to the Surface 1 Expanding the Boundaries of Laser Microscopy Measurements and images you can count on. Every time. LEXT OLS4100 Widely used in quality control, research, and development

More information

LION PRECISION. TechNote LT February, Capacitive Sensor Operation and Optimization

LION PRECISION. TechNote LT February, Capacitive Sensor Operation and Optimization LION PRECISION TechNote LT03-0020 February, 2009 Capacitive Sensor Operation and Optimization Contents Capacitance and Distance 2 Focusing the Electric Field 3 Effects of Target Size 3 Range of Measurement

More information

Measurement of Microscopic Three-dimensional Profiles with High Accuracy and Simple Operation

Measurement of Microscopic Three-dimensional Profiles with High Accuracy and Simple Operation 238 Hitachi Review Vol. 65 (2016), No. 7 Featured Articles Measurement of Microscopic Three-dimensional Profiles with High Accuracy and Simple Operation AFM5500M Scanning Probe Microscope Satoshi Hasumura

More information

Background Statement for SEMI Draft Document 5293C NEW STANDARD: TEST METHODS FOR POSITIONAL ACCURACY OF CAPACITIVE TOUCHSCREEN PANEL

Background Statement for SEMI Draft Document 5293C NEW STANDARD: TEST METHODS FOR POSITIONAL ACCURACY OF CAPACITIVE TOUCHSCREEN PANEL Background Statement for SEMI Draft Document 5293C NEW STANDARD: TEST METHODS FOR POSITIONAL ACCURACY OF CAPACITIVE TOUCHSCREEN PANEL Notice: This background statement is not part of the balloted item.

More information

TechNote. T001 // Precise non-contact displacement sensors. Introduction

TechNote. T001 // Precise non-contact displacement sensors. Introduction TechNote T001 // Precise non-contact displacement sensors Contents: Introduction Inductive sensors based on eddy currents Capacitive sensors Laser triangulation sensors Confocal sensors Comparison of all

More information

OPSENS WHITE-LIGHT POLARIZATION INTERFEROMETRY TECHNOLOGY

OPSENS WHITE-LIGHT POLARIZATION INTERFEROMETRY TECHNOLOGY OPSENS WHITE-LIGHT POLARIZATION INTERFEROMETRY TECHNOLOGY 1. Introduction Fiber optic sensors are made up of two main parts: the fiber optic transducer (also called the fiber optic gauge or the fiber optic

More information

OPSENS WHITE-LIGHT POLARIZATION INTERFEROMETRY TECHNOLOGY

OPSENS WHITE-LIGHT POLARIZATION INTERFEROMETRY TECHNOLOGY OPSENS WHITE-LIGHT POLARIZATION INTERFEROMETRY TECHNOLOGY 1. Introduction Fiber optic sensors are made up of two main parts: the fiber optic transducer (also called the fiber optic gauge or the fiber optic

More information

Fast Optical Form Measurements of Rough Cylindrical and Conical Surfaces in Diesel Fuel Injection Components

Fast Optical Form Measurements of Rough Cylindrical and Conical Surfaces in Diesel Fuel Injection Components Fast Optical Form Measurements of Rough Cylindrical and Conical Surfaces in Diesel Fuel Injection Components Thomas J. Dunn, Robert Michaels, Simon Lee, Mark Tronolone, and Andrew Kulawiec; Corning Tropel

More information

The End of Thresholds: Subwavelength Optical Linewidth Measurement Using the Flux-Area Technique

The End of Thresholds: Subwavelength Optical Linewidth Measurement Using the Flux-Area Technique The End of Thresholds: Subwavelength Optical Linewidth Measurement Using the Flux-Area Technique Peter Fiekowsky Automated Visual Inspection, Los Altos, California ABSTRACT The patented Flux-Area technique

More information

Advanced Features of InfraTec Pyroelectric Detectors

Advanced Features of InfraTec Pyroelectric Detectors 1 Basics and Application of Variable Color Products The key element of InfraTec s variable color products is a silicon micro machined tunable narrow bandpass filter, which is fully integrated inside the

More information

Difrotec Product & Services. Ultra high accuracy interferometry & custom optical solutions

Difrotec Product & Services. Ultra high accuracy interferometry & custom optical solutions Difrotec Product & Services Ultra high accuracy interferometry & custom optical solutions Content 1. Overview 2. Interferometer D7 3. Benefits 4. Measurements 5. Specifications 6. Applications 7. Cases

More information

SENSOR+TEST Conference SENSOR 2009 Proceedings II

SENSOR+TEST Conference SENSOR 2009 Proceedings II B8.4 Optical 3D Measurement of Micro Structures Ettemeyer, Andreas; Marxer, Michael; Keferstein, Claus NTB Interstaatliche Hochschule für Technik Buchs Werdenbergstr. 4, 8471 Buchs, Switzerland Introduction

More information

Applications of Optics

Applications of Optics Nicholas J. Giordano www.cengage.com/physics/giordano Chapter 26 Applications of Optics Marilyn Akins, PhD Broome Community College Applications of Optics Many devices are based on the principles of optics

More information

attosnom I: Topography and Force Images NANOSCOPY APPLICATION NOTE M06 RELATED PRODUCTS G

attosnom I: Topography and Force Images NANOSCOPY APPLICATION NOTE M06 RELATED PRODUCTS G APPLICATION NOTE M06 attosnom I: Topography and Force Images Scanning near-field optical microscopy is the outstanding technique to simultaneously measure the topography and the optical contrast of a sample.

More information

Background Statement for SEMI Draft Document 5728 New Standard: Test Method for the Wire Tension of Multi-wire Saws

Background Statement for SEMI Draft Document 5728 New Standard: Test Method for the Wire Tension of Multi-wire Saws Background Statement for SEMI Draft Document 5728 ew Standard: Test Method for the Wire Tension of Multi-wire Saws otice: This background statement is not part of the balloted item. It is provided solely

More information

Surface Finish Measurement Methods and Instrumentation

Surface Finish Measurement Methods and Instrumentation 125 years of innovation Surface Finish Measurement Methods and Instrumentation Contents Visual Inspection Surface Finish Comparison Plates Contact Gauges Inductive / Variable Reluctance (INTRA) Piezo Electric

More information

1.6 Beam Wander vs. Image Jitter

1.6 Beam Wander vs. Image Jitter 8 Chapter 1 1.6 Beam Wander vs. Image Jitter It is common at this point to look at beam wander and image jitter and ask what differentiates them. Consider a cooperative optical communication system that

More information

200mm and 300mm Test Patterned Wafers for Bonding Process Applications SKW ASSOCIATES, INC.

200mm and 300mm Test Patterned Wafers for Bonding Process Applications SKW ASSOCIATES, INC. C M P C h a r a c t e r I z a t I o n S o l u t I o n s 200mm and 300mm Test Patterned Wafers for Bonding Process Applications SKW ASSOCIATES, INC. 2920 Scott Blvd., Santa Clara, CA 95054 Tel: 408-919-0094,

More information

Background Statement for SEMI Draft Document 5530 NEW STANDARD: SPECIFICATION FOR ORIENTATION FIDUCIAL MARKS FOR PV SILICON WAFERS

Background Statement for SEMI Draft Document 5530 NEW STANDARD: SPECIFICATION FOR ORIENTATION FIDUCIAL MARKS FOR PV SILICON WAFERS Background Statement for SEMI Draft Document 5530 NEW STANDARD: SPECIFICATION FOR ORIENTATION FIDUCIAL MARKS FOR PV SILICON WAFERS Notice: This background statement is not part of the balloted item. It

More information

NEW LASER ULTRASONIC INTERFEROMETER FOR INDUSTRIAL APPLICATIONS B.Pouet and S.Breugnot Bossa Nova Technologies; Venice, CA, USA

NEW LASER ULTRASONIC INTERFEROMETER FOR INDUSTRIAL APPLICATIONS B.Pouet and S.Breugnot Bossa Nova Technologies; Venice, CA, USA NEW LASER ULTRASONIC INTERFEROMETER FOR INDUSTRIAL APPLICATIONS B.Pouet and S.Breugnot Bossa Nova Technologies; Venice, CA, USA Abstract: A novel interferometric scheme for detection of ultrasound is presented.

More information

BMC s heritage deformable mirror technology that uses hysteresis free electrostatic

BMC s heritage deformable mirror technology that uses hysteresis free electrostatic Optical Modulator Technical Whitepaper MEMS Optical Modulator Technology Overview The BMC MEMS Optical Modulator, shown in Figure 1, was designed for use in free space optical communication systems. The

More information

The spectral colours of nanometers

The spectral colours of nanometers Reprint from the journal Mikroproduktion 3/2005 Berthold Michelt and Jochen Schulze The spectral colours of nanometers Precitec Optronik GmbH Raiffeisenstraße 5 D-63110 Rodgau Phone: +49 (0) 6106 8290-14

More information

Application Note #548 AcuityXR Technology Significantly Enhances Lateral Resolution of White-Light Optical Profilers

Application Note #548 AcuityXR Technology Significantly Enhances Lateral Resolution of White-Light Optical Profilers Application Note #548 AcuityXR Technology Significantly Enhances Lateral Resolution of White-Light Optical Profilers ContourGT with AcuityXR TM capability White light interferometry is firmly established

More information

attocfm I for Surface Quality Inspection NANOSCOPY APPLICATION NOTE M01 RELATED PRODUCTS G

attocfm I for Surface Quality Inspection NANOSCOPY APPLICATION NOTE M01 RELATED PRODUCTS G APPLICATION NOTE M01 attocfm I for Surface Quality Inspection Confocal microscopes work by scanning a tiny light spot on a sample and by measuring the scattered light in the illuminated volume. First,

More information

Micro-sensors - what happens when you make "classical" devices "small": MEMS devices and integrated bolometric IR detectors

Micro-sensors - what happens when you make classical devices small: MEMS devices and integrated bolometric IR detectors Micro-sensors - what happens when you make "classical" devices "small": MEMS devices and integrated bolometric IR detectors Dean P. Neikirk 1 MURI bio-ir sensors kick-off 6/16/98 Where are the targets

More information

Specifying and Measuring Nanometer Surface Properties. Alson E. Hatheway

Specifying and Measuring Nanometer Surface Properties. Alson E. Hatheway Specifying and Measuring Nanometer Surface Properties a seminar prepared for the American Society of Mechanical Engineers 93663a.p65(1 Alson E. Hatheway Alson E. Hatheway Inc. 787 West Woodbury Road Unit

More information

Non-Contact Capacitance Gauging Instrument & Series 2800 Capacitive Probes

Non-Contact Capacitance Gauging Instrument & Series 2800 Capacitive Probes 4810 Non-Contact Capacitance Gauging Instrument & Series 2800 Capacitive Probes Sub nanometer resolution for ultra-precise measurements Exceptional temperature stability Wide variety of precision capacitive

More information

Lithography. 3 rd. lecture: introduction. Prof. Yosi Shacham-Diamand. Fall 2004

Lithography. 3 rd. lecture: introduction. Prof. Yosi Shacham-Diamand. Fall 2004 Lithography 3 rd lecture: introduction Prof. Yosi Shacham-Diamand Fall 2004 1 List of content Fundamental principles Characteristics parameters Exposure systems 2 Fundamental principles Aerial Image Exposure

More information

Chapter 7. Optical Measurement and Interferometry

Chapter 7. Optical Measurement and Interferometry Chapter 7 Optical Measurement and Interferometry 1 Introduction Optical measurement provides a simple, easy, accurate and reliable means for carrying out inspection and measurements in the industry the

More information

Optical design of a high resolution vision lens

Optical design of a high resolution vision lens Optical design of a high resolution vision lens Paul Claassen, optical designer, paul.claassen@sioux.eu Marnix Tas, optical specialist, marnix.tas@sioux.eu Prof L.Beckmann, l.beckmann@hccnet.nl Summary:

More information

Laser Telemetric System (Metrology)

Laser Telemetric System (Metrology) Laser Telemetric System (Metrology) Laser telemetric system is a non-contact gauge that measures with a collimated laser beam (Refer Fig. 10.26). It measure at the rate of 150 scans per second. It basically

More information

Development of innovative fringe locking strategies for vibration-resistant white light vertical scanning interferometry (VSI)

Development of innovative fringe locking strategies for vibration-resistant white light vertical scanning interferometry (VSI) Development of innovative fringe locking strategies for vibration-resistant white light vertical scanning interferometry (VSI) Liang-Chia Chen 1), Abraham Mario Tapilouw 1), Sheng-Lih Yeh 2), Shih-Tsong

More information

Measurement of Surface Profile and Layer Cross-section with Wide Field of View and High Precision

Measurement of Surface Profile and Layer Cross-section with Wide Field of View and High Precision Hitachi Review Vol. 65 (2016), No. 7 243 Featured Articles Measurement of Surface Profile and Layer Cross-section with Wide Field of View and High Precision VS1000 Series Coherence Scanning Interferometer

More information

7 CHAPTER 7: REFRACTIVE INDEX MEASUREMENTS WITH COMMON PATH PHASE SENSITIVE FDOCT SETUP

7 CHAPTER 7: REFRACTIVE INDEX MEASUREMENTS WITH COMMON PATH PHASE SENSITIVE FDOCT SETUP 7 CHAPTER 7: REFRACTIVE INDEX MEASUREMENTS WITH COMMON PATH PHASE SENSITIVE FDOCT SETUP Abstract: In this chapter we describe the use of a common path phase sensitive FDOCT set up. The phase measurements

More information

Review and Adjudication Information. Group: Inspection & Metrology Task Force North America 3DS-IC Committee Date: April 2, 2013 April 2, 2013

Review and Adjudication Information. Group: Inspection & Metrology Task Force North America 3DS-IC Committee Date: April 2, 2013 April 2, 2013 Background Statement for SEMI Draft Document 5410 NEW STANDARD: GUIDE FOR METROLOGY TECHNIQUES TO BE USED IN MEASUREMENT OF GEOMETRICAL PARAMETERS OF THROUGH-SILICON VIAS (TSVs) IN 3DS-IC STRUCTURES Notice:

More information

Large-Area Interference Lithography Exposure Tool Development

Large-Area Interference Lithography Exposure Tool Development Large-Area Interference Lithography Exposure Tool Development John Burnett 1, Eric Benck 1 and James Jacob 2 1 Physical Measurements Laboratory, NIST, Gaithersburg, MD, USA 2 Actinix, Scotts Valley, CA

More information

Silicon Light Machines Patents

Silicon Light Machines Patents 820 Kifer Road, Sunnyvale, CA 94086 Tel. 408-240-4700 Fax 408-456-0708 www.siliconlight.com Silicon Light Machines Patents USPTO No. US 5,808,797 US 5,841,579 US 5,798,743 US 5,661,592 US 5,629,801 US

More information

Fabrication of Probes for High Resolution Optical Microscopy

Fabrication of Probes for High Resolution Optical Microscopy Fabrication of Probes for High Resolution Optical Microscopy Physics 564 Applied Optics Professor Andrès La Rosa David Logan May 27, 2010 Abstract Near Field Scanning Optical Microscopy (NSOM) is a technique

More information

Chapter 18 Optical Elements

Chapter 18 Optical Elements Chapter 18 Optical Elements GOALS When you have mastered the content of this chapter, you will be able to achieve the following goals: Definitions Define each of the following terms and use it in an operational

More information

Advanced Stepper Lithography Technology to Enable Flexible AMOLED Displays. Keith Best Roger McCleary Elvino M da Silveira 5/19/17

Advanced Stepper Lithography Technology to Enable Flexible AMOLED Displays. Keith Best Roger McCleary Elvino M da Silveira 5/19/17 Advanced Stepper Lithography Technology to Enable Flexible AMOLED Displays Keith Best Roger McCleary Elvino M da Silveira 5/19/17 Agenda About Rudolph JetStep G System overview and performance Display

More information

Sensors and Metrology - 2 Optical Microscopy and Overlay Measurements

Sensors and Metrology - 2 Optical Microscopy and Overlay Measurements Sensors and Metrology - 2 Optical Microscopy and Overlay Measurements 1 Optical Metrology Optical Microscopy What is its place in IC production? What are the limitations and the hopes? The issue of Alignment

More information

Confocal Imaging Through Scattering Media with a Volume Holographic Filter

Confocal Imaging Through Scattering Media with a Volume Holographic Filter Confocal Imaging Through Scattering Media with a Volume Holographic Filter Michal Balberg +, George Barbastathis*, Sergio Fantini % and David J. Brady University of Illinois at Urbana-Champaign, Urbana,

More information

06SurfaceQuality.nb Optics James C. Wyant (2012) 1

06SurfaceQuality.nb Optics James C. Wyant (2012) 1 06SurfaceQuality.nb Optics 513 - James C. Wyant (2012) 1 Surface Quality SQ-1 a) How is surface profile data obtained using the FECO interferometer? Your explanation should include diagrams with the appropriate

More information

2. Pulsed Acoustic Microscopy and Picosecond Ultrasonics

2. Pulsed Acoustic Microscopy and Picosecond Ultrasonics 1st International Symposium on Laser Ultrasonics: Science, Technology and Applications July 16-18 2008, Montreal, Canada Picosecond Ultrasonic Microscopy of Semiconductor Nanostructures Thomas J GRIMSLEY

More information

Lab Report 3: Speckle Interferometry LIN PEI-YING, BAIG JOVERIA

Lab Report 3: Speckle Interferometry LIN PEI-YING, BAIG JOVERIA Lab Report 3: Speckle Interferometry LIN PEI-YING, BAIG JOVERIA Abstract: Speckle interferometry (SI) has become a complete technique over the past couple of years and is widely used in many branches of

More information

Sensing. Autonomous systems. Properties. Classification. Key requirement of autonomous systems. An AS should be connected to the outside world.

Sensing. Autonomous systems. Properties. Classification. Key requirement of autonomous systems. An AS should be connected to the outside world. Sensing Key requirement of autonomous systems. An AS should be connected to the outside world. Autonomous systems Convert a physical value to an electrical value. From temperature, humidity, light, to

More information

Real-Time Scanning Goniometric Radiometer for Rapid Characterization of Laser Diodes and VCSELs

Real-Time Scanning Goniometric Radiometer for Rapid Characterization of Laser Diodes and VCSELs Real-Time Scanning Goniometric Radiometer for Rapid Characterization of Laser Diodes and VCSELs Jeffrey L. Guttman, John M. Fleischer, and Allen M. Cary Photon, Inc. 6860 Santa Teresa Blvd., San Jose,

More information

Instructions for the Experiment

Instructions for the Experiment Instructions for the Experiment Excitonic States in Atomically Thin Semiconductors 1. Introduction Alongside with electrical measurements, optical measurements are an indispensable tool for the study of

More information

Development of a new multi-wavelength confocal surface profilometer for in-situ automatic optical inspection (AOI)

Development of a new multi-wavelength confocal surface profilometer for in-situ automatic optical inspection (AOI) Development of a new multi-wavelength confocal surface profilometer for in-situ automatic optical inspection (AOI) Liang-Chia Chen 1#, Chao-Nan Chen 1 and Yi-Wei Chang 1 1. Institute of Automation Technology,

More information

Photonic device package design, assembly and encapsulation.

Photonic device package design, assembly and encapsulation. Photonic device package design, assembly and encapsulation. Abstract. A.Bos, E. Boschman Advanced Packaging Center. Duiven, The Netherlands Photonic devices like Optical transceivers, Solar cells, LED

More information

Test procedures Page: 1 of 5

Test procedures Page: 1 of 5 Test procedures Page: 1 of 5 1 Scope This part of document establishes uniform requirements for measuring the numerical aperture of optical fibre, thereby assisting in the inspection of fibres and cables

More information

Advanced 3D Optical Profiler using Grasshopper3 USB3 Vision camera

Advanced 3D Optical Profiler using Grasshopper3 USB3 Vision camera Advanced 3D Optical Profiler using Grasshopper3 USB3 Vision camera Figure 1. The Zeta-20 uses the Grasshopper3 and produces true color 3D optical images with multi mode optics technology 3D optical profiling

More information

Large Field of View, High Spatial Resolution, Surface Measurements

Large Field of View, High Spatial Resolution, Surface Measurements Large Field of View, High Spatial Resolution, Surface Measurements James C. Wyant and Joanna Schmit WYKO Corporation, 2650 E. Elvira Road Tucson, Arizona 85706, USA jcwyant@wyko.com and jschmit@wyko.com

More information

MINIATURE X-RAY SOURCES AND THE EFFECTS OF SPOT SIZE ON SYSTEM PERFORMANCE

MINIATURE X-RAY SOURCES AND THE EFFECTS OF SPOT SIZE ON SYSTEM PERFORMANCE 228 MINIATURE X-RAY SOURCES AND THE EFFECTS OF SPOT SIZE ON SYSTEM PERFORMANCE D. CARUSO, M. DINSMORE TWX LLC, CONCORD, MA 01742 S. CORNABY MOXTEK, OREM, UT 84057 ABSTRACT Miniature x-ray sources present

More information

PROCEEDINGS OF SPIE. Automated asphere centration testing with AspheroCheck UP

PROCEEDINGS OF SPIE. Automated asphere centration testing with AspheroCheck UP PROCEEDINGS OF SPIE SPIEDigitalLibrary.org/conference-proceedings-of-spie Automated asphere centration testing with AspheroCheck UP F. Hahne, P. Langehanenberg F. Hahne, P. Langehanenberg, "Automated asphere

More information

ADALAM Sensor based adaptive laser micromachining using ultrashort pulse lasers for zero-failure manufacturing D2.2. Ger Folkersma (Demcon)

ADALAM Sensor based adaptive laser micromachining using ultrashort pulse lasers for zero-failure manufacturing D2.2. Ger Folkersma (Demcon) D2.2 Automatic adjustable reference path system Document Coordinator: Contributors: Dissemination: Keywords: Ger Folkersma (Demcon) Ger Folkersma, Kevin Voss, Marvin Klein (Demcon) Public Reference path,

More information

Stability of a Fiber-Fed Heterodyne Interferometer

Stability of a Fiber-Fed Heterodyne Interferometer Stability of a Fiber-Fed Heterodyne Interferometer Christoph Weichert, Jens Flügge, Paul Köchert, Rainer Köning, Physikalisch Technische Bundesanstalt, Braunschweig, Germany; Rainer Tutsch, Technische

More information

PHYS 3153 Methods of Experimental Physics II O2. Applications of Interferometry

PHYS 3153 Methods of Experimental Physics II O2. Applications of Interferometry Purpose PHYS 3153 Methods of Experimental Physics II O2. Applications of Interferometry In this experiment, you will study the principles and applications of interferometry. Equipment and components PASCO

More information

Dynamic Phase-Shifting Electronic Speckle Pattern Interferometer

Dynamic Phase-Shifting Electronic Speckle Pattern Interferometer Dynamic Phase-Shifting Electronic Speckle Pattern Interferometer Michael North Morris, James Millerd, Neal Brock, John Hayes and *Babak Saif 4D Technology Corporation, 3280 E. Hemisphere Loop Suite 146,

More information

Application Bulletin 240

Application Bulletin 240 Application Bulletin 240 Design Consideration CUSTOM CAPABILITIES Standard PC board fabrication flexibility allows for various component orientations, mounting features, and interconnect schemes. The starting

More information

queensgate a brand of Elektron Technology

queensgate a brand of Elektron Technology NanoSensors NX/NZ NanoSensor The NanoSensor is a non-contact position measuring system based on the principle of capacitance micrometry. Two sensor plates, a Target and a Probe, form a parallel plate capacitor.

More information

Vixar High Power Array Technology

Vixar High Power Array Technology Vixar High Power Array Technology I. Introduction VCSELs arrays emitting power ranging from 50mW to 10W have emerged as an important technology for applications within the consumer, industrial, automotive

More information

Kit for building your own THz Time-Domain Spectrometer

Kit for building your own THz Time-Domain Spectrometer Kit for building your own THz Time-Domain Spectrometer 16/06/2016 1 Table of contents 0. Parts for the THz Kit... 3 1. Delay line... 4 2. Pulse generator and lock-in detector... 5 3. THz antennas... 6

More information

Adaptive Optics for LIGO

Adaptive Optics for LIGO Adaptive Optics for LIGO Justin Mansell Ginzton Laboratory LIGO-G990022-39-M Motivation Wavefront Sensor Outline Characterization Enhancements Modeling Projections Adaptive Optics Results Effects of Thermal

More information

IMAGE SENSOR SOLUTIONS. KAC-96-1/5" Lens Kit. KODAK KAC-96-1/5" Lens Kit. for use with the KODAK CMOS Image Sensors. November 2004 Revision 2

IMAGE SENSOR SOLUTIONS. KAC-96-1/5 Lens Kit. KODAK KAC-96-1/5 Lens Kit. for use with the KODAK CMOS Image Sensors. November 2004 Revision 2 KODAK for use with the KODAK CMOS Image Sensors November 2004 Revision 2 1.1 Introduction Choosing the right lens is a critical aspect of designing an imaging system. Typically the trade off between image

More information

Collimation Tester Instructions

Collimation Tester Instructions Description Use shear-plate collimation testers to examine and adjust the collimation of laser light, or to measure the wavefront curvature and divergence/convergence magnitude of large-radius optical

More information

3D Optical Motion Analysis of Micro Systems. Heinrich Steger, Polytec GmbH, Waldbronn

3D Optical Motion Analysis of Micro Systems. Heinrich Steger, Polytec GmbH, Waldbronn 3D Optical Motion Analysis of Micro Systems Heinrich Steger, Polytec GmbH, Waldbronn SEMICON Europe 2012 Outline Needs and Challenges of measuring Micro Structure and MEMS Tools and Applications for optical

More information

nanovea.com PROFILOMETERS 3D Non Contact Metrology

nanovea.com PROFILOMETERS 3D Non Contact Metrology PROFILOMETERS 3D Non Contact Metrology nanovea.com PROFILOMETER INTRO Nanovea 3D Non-Contact Profilometers are designed with leading edge optical pens using superior white light axial chromatism. Nano

More information

visibility values: 1) V1=0.5 2) V2=0.9 3) V3=0.99 b) In the three cases considered, what are the values of FSR (Free Spectral Range) and

visibility values: 1) V1=0.5 2) V2=0.9 3) V3=0.99 b) In the three cases considered, what are the values of FSR (Free Spectral Range) and EXERCISES OF OPTICAL MEASUREMENTS BY ENRICO RANDONE AND CESARE SVELTO EXERCISE 1 A CW laser radiation (λ=2.1 µm) is delivered to a Fabry-Pérot interferometer made of 2 identical plane and parallel mirrors

More information

Wavelength Control and Locking with Sub-MHz Precision

Wavelength Control and Locking with Sub-MHz Precision Wavelength Control and Locking with Sub-MHz Precision A PZT actuator on one of the resonator mirrors enables the Verdi output wavelength to be rapidly tuned over a range of several GHz or tightly locked

More information

EE119 Introduction to Optical Engineering Fall 2009 Final Exam. Name:

EE119 Introduction to Optical Engineering Fall 2009 Final Exam. Name: EE119 Introduction to Optical Engineering Fall 2009 Final Exam Name: SID: CLOSED BOOK. THREE 8 1/2 X 11 SHEETS OF NOTES, AND SCIENTIFIC POCKET CALCULATOR PERMITTED. TIME ALLOTTED: 180 MINUTES Fundamental

More information

StarBright XLT Optical Coatings

StarBright XLT Optical Coatings StarBright XLT Optical Coatings StarBright XLT is Celestron s revolutionary optical coating system that outperforms any other coating in the commercial telescope market. Our most popular Schmidt-Cassegrain

More information

Copyright 2000 Society of Photo Instrumentation Engineers.

Copyright 2000 Society of Photo Instrumentation Engineers. Copyright 2000 Society of Photo Instrumentation Engineers. This paper was published in SPIE Proceedings, Volume 4043 and is made available as an electronic reprint with permission of SPIE. One print or

More information

R.B.V.R.R. WOMEN S COLLEGE (AUTONOMOUS) Narayanaguda, Hyderabad.

R.B.V.R.R. WOMEN S COLLEGE (AUTONOMOUS) Narayanaguda, Hyderabad. R.B.V.R.R. WOMEN S COLLEGE (AUTONOMOUS) Narayanaguda, Hyderabad. DEPARTMENT OF PHYSICS QUESTION BANK FOR SEMESTER III PAPER III OPTICS UNIT I: 1. MATRIX METHODS IN PARAXIAL OPTICS 2. ABERATIONS UNIT II

More information

1393 DISPLACEMENT SENSORS

1393 DISPLACEMENT SENSORS 1393 DISPLACEMENT SENSORS INTRODUCTION While regular sensors detect the existence of objects, displacement sensors detect the amount of displacement when objects move from one position to another. Detecting

More information

Evaluation of Scientific Solutions Liquid Crystal Fabry-Perot Etalon

Evaluation of Scientific Solutions Liquid Crystal Fabry-Perot Etalon Evaluation of Scientific Solutions Liquid Crystal Fabry-Perot Etalon Testing of the etalon was done using a frequency stabilized He-Ne laser. The beam from the laser was passed through a spatial filter

More information

POCKET DEFORMABLE MIRROR FOR ADAPTIVE OPTICS APPLICATIONS

POCKET DEFORMABLE MIRROR FOR ADAPTIVE OPTICS APPLICATIONS POCKET DEFORMABLE MIRROR FOR ADAPTIVE OPTICS APPLICATIONS Leonid Beresnev1, Mikhail Vorontsov1,2 and Peter Wangsness3 1) US Army Research Laboratory, 2800 Powder Mill Road, Adelphi Maryland 20783, lberesnev@arl.army.mil,

More information

Chapter Ray and Wave Optics

Chapter Ray and Wave Optics 109 Chapter Ray and Wave Optics 1. An astronomical telescope has a large aperture to [2002] reduce spherical aberration have high resolution increase span of observation have low dispersion. 2. If two

More information

Laser Speckle Reducer LSR-3000 Series

Laser Speckle Reducer LSR-3000 Series Datasheet: LSR-3000 Series Update: 06.08.2012 Copyright 2012 Optotune Laser Speckle Reducer LSR-3000 Series Speckle noise from a laser-based system is reduced by dynamically diffusing the laser beam. A

More information

Lecture 20: Optical Tools for MEMS Imaging

Lecture 20: Optical Tools for MEMS Imaging MECH 466 Microelectromechanical Systems University of Victoria Dept. of Mechanical Engineering Lecture 20: Optical Tools for MEMS Imaging 1 Overview Optical Microscopes Video Microscopes Scanning Electron

More information

(51) Int Cl.: G01B 9/02 ( ) G01B 11/24 ( ) G01N 21/47 ( )

(51) Int Cl.: G01B 9/02 ( ) G01B 11/24 ( ) G01N 21/47 ( ) (19) (12) EUROPEAN PATENT APPLICATION (11) EP 1 939 581 A1 (43) Date of publication: 02.07.2008 Bulletin 2008/27 (21) Application number: 07405346.3 (51) Int Cl.: G01B 9/02 (2006.01) G01B 11/24 (2006.01)

More information

Properties of Structured Light

Properties of Structured Light Properties of Structured Light Gaussian Beams Structured light sources using lasers as the illumination source are governed by theories of Gaussian beams. Unlike incoherent sources, coherent laser sources

More information

Absolute distance interferometer in LaserTracer geometry

Absolute distance interferometer in LaserTracer geometry Absolute distance interferometer in LaserTracer geometry Corresponding author: Karl Meiners-Hagen Abstract 1. Introduction 1 In this paper, a combination of variable synthetic and two-wavelength interferometry

More information

Practical Flatness Tech Note

Practical Flatness Tech Note Practical Flatness Tech Note Understanding Laser Dichroic Performance BrightLine laser dichroic beamsplitters set a new standard for super-resolution microscopy with λ/10 flatness per inch, P-V. We ll

More information

Sensors & Applications Glass Industry. More Precision

Sensors & Applications Glass Industry. More Precision Sensors & Applications Glass Industry More Precision Sensors and measuring systems for glass production Modern glass production is increasingly determined by maximum efficiency. Therefore, rapid access

More information

Lecture 04: Solar Imaging Instruments

Lecture 04: Solar Imaging Instruments Hale COLLAGE (NJIT Phys-780) Topics in Solar Observation Techniques Lecture 04: Solar Imaging Instruments Wenda Cao New Jersey Institute of Technology Valentin M. Pillet National Solar Observatory SDO

More information

Application Instruction 002. Superluminescent Light Emitting Diodes: Device Fundamentals and Reliability

Application Instruction 002. Superluminescent Light Emitting Diodes: Device Fundamentals and Reliability I. Introduction II. III. IV. SLED Fundamentals SLED Temperature Performance SLED and Optical Feedback V. Operation Stability, Reliability and Life VI. Summary InPhenix, Inc., 25 N. Mines Road, Livermore,

More information

Radial Polarization Converter With LC Driver USER MANUAL

Radial Polarization Converter With LC Driver USER MANUAL ARCoptix Radial Polarization Converter With LC Driver USER MANUAL Arcoptix S.A Ch. Trois-portes 18 2000 Neuchâtel Switzerland Mail: info@arcoptix.com Tel: ++41 32 731 04 66 Principle of the radial polarization

More information

Section 2: Lithography. Jaeger Chapter 2 Litho Reader. The lithographic process

Section 2: Lithography. Jaeger Chapter 2 Litho Reader. The lithographic process Section 2: Lithography Jaeger Chapter 2 Litho Reader The lithographic process Photolithographic Process (a) (b) (c) (d) (e) (f) (g) Substrate covered with silicon dioxide barrier layer Positive photoresist

More information

Instruction manual and data sheet ipca h

Instruction manual and data sheet ipca h 1/15 instruction manual ipca-21-05-1000-800-h Instruction manual and data sheet ipca-21-05-1000-800-h Broad area interdigital photoconductive THz antenna with microlens array and hyperhemispherical silicon

More information

Fiber-optic Michelson Interferometer Sensor Fabricated by Femtosecond Lasers

Fiber-optic Michelson Interferometer Sensor Fabricated by Femtosecond Lasers Sensors & ransducers 2013 by IFSA http://www.sensorsportal.com Fiber-optic Michelson Interferometer Sensor Fabricated by Femtosecond Lasers Dong LIU, Ying XIE, Gui XIN, Zheng-Ying LI School of Information

More information

Section 2: Lithography. Jaeger Chapter 2 Litho Reader. EE143 Ali Javey Slide 5-1

Section 2: Lithography. Jaeger Chapter 2 Litho Reader. EE143 Ali Javey Slide 5-1 Section 2: Lithography Jaeger Chapter 2 Litho Reader EE143 Ali Javey Slide 5-1 The lithographic process EE143 Ali Javey Slide 5-2 Photolithographic Process (a) (b) (c) (d) (e) (f) (g) Substrate covered

More information

Absorption: in an OF, the loss of Optical power, resulting from conversion of that power into heat.

Absorption: in an OF, the loss of Optical power, resulting from conversion of that power into heat. Absorption: in an OF, the loss of Optical power, resulting from conversion of that power into heat. Scattering: The changes in direction of light confined within an OF, occurring due to imperfection in

More information

Copyright 1997 by the Society of Photo-Optical Instrumentation Engineers.

Copyright 1997 by the Society of Photo-Optical Instrumentation Engineers. Copyright 1997 by the Society of Photo-Optical Instrumentation Engineers. This paper was published in the proceedings of Microlithographic Techniques in IC Fabrication, SPIE Vol. 3183, pp. 14-27. It is

More information

Outline: Introduction: What is SPM, history STM AFM Image treatment Advanced SPM techniques Applications in semiconductor research and industry

Outline: Introduction: What is SPM, history STM AFM Image treatment Advanced SPM techniques Applications in semiconductor research and industry 1 Outline: Introduction: What is SPM, history STM AFM Image treatment Advanced SPM techniques Applications in semiconductor research and industry 2 Back to our solutions: The main problem: How to get nm

More information

Optical coherence tomography

Optical coherence tomography Optical coherence tomography Peter E. Andersen Optics and Plasma Research Department Risø National Laboratory E-mail peter.andersen@risoe.dk Outline Part I: Introduction to optical coherence tomography

More information

Computer Generated Holograms for Optical Testing

Computer Generated Holograms for Optical Testing Computer Generated Holograms for Optical Testing Dr. Jim Burge Associate Professor Optical Sciences and Astronomy University of Arizona jburge@optics.arizona.edu 520-621-8182 Computer Generated Holograms

More information