Realization of a Liquid Atomic Force Microscope

Size: px
Start display at page:

Download "Realization of a Liquid Atomic Force Microscope"

Transcription

1 Realization of a Liquid Atomic Force Microscope Ivo de Rijk DCT Traineeship report Supervisor: prof. dr. H. Kawakatsu prof. dr. ir. M. Steinbuch Technische Universiteit Eindhoven Department Mechanical Engineering Dynamics and Control Technology Group Eindhoven, February, 2008

2 2 Acknowledgements This report is the result of a research project of four months at Tokyo University, Japan. The project could not have been successful without the help of my team members Kentaro Minemura and Yuki Nishimori. Furthermore, I would like to thank the laboratory members in general for the extremely educational and pleasant time. More specific I would like to thank Yasuo Hoshi for his effort on the software, Dai Kobayashi for his tremendous help with the control part, Shuhei Nishida for sharing his knowledge of the liquid AFM, Kazuhisa Nakagawa for the, always available, helping hand and of course professor Kawakatsu for giving me the opportunity to do this project and his inspiring help during the project.

3 Contents 1 Introduction 4 2 Liquid AFM 5 3 Mechanical head Cantilever positioning Cantilever fixation Sample positioning Tube piezo Optics Heterodyne Doppler interferometer Photothermal excitation Control Approach Results 24 7 Conclusions & Recommendations Conclusions Recommendations A Mechanical drawings LAFM mechanical head 30 A.1 Mechanical drawings lower part mechanical head A.2 Mechanical drawings upper part mechanical head B Piezo actuated stages 55 C Test results sample step size 57 3

4 Chapter 1 Introduction Since the introduction of the Atomic Force Microscope (AFM) in 1986 by Binnig et al.[1] the AFM has shown to be a versatile tool to gain knowledge about surfaces, down to atomic scale. One of the most promising applications of the AFM is imaging of surfaces in a liquid. For example, cells at atomic resolution in water or a real time observation of the behavior of water molecules on a surface with changing temperature. A liquid AFM (LAFM) was designed and build at Kawakatsu laboratory and completed in 2005 [7]. This LAFM has produced atomic scale resolution images in a liquid e.g. mica in water, which is shown in figure 2.1. The objective of this research project is to realize a new version of the LAFM, improving the mechanics where possible and realizing a home-built optical setup to measure the cantilever speed and excite the cantilever by photothermal means. This project is performed at Tokyo University, Institute of Industrial Science, Kawakatsu laboratory in Japan as a part of the Mechanical Engineering Master s program at Eindhoven University of technology in The Netherlands. The report starts with a basic introduction to the LAFM, describing its working principles. The mechanical part of the LAFM, the head, is discussed in chapter 3. The mechanical head performs the actuation and positioning of the sample and cantilever. The optics are explained in chapter 4, which incorporates the measuring and excitation of the cantilever together with camera view on the cantilever. The control circuitry is roughly explained in chapter 5, followed by the overall result in chapter 6 and the conclusions & recommendations in chapter 7. 4

5 Chapter 2 Liquid AFM The Frequency Modulated Liquid Atomic Force Microscope (FM-LAFM) enables imaging of a surface at high resolution in a liquid, for example biological molecules in their native environment. In a liquid, the highest resolution has been obtained in constant-force mode. In constant-force mode the tip-sample interaction force is kept constant by adjusting the distance between the tip and sample. However, scanning a sample with a constant vertical force implies lateral friction forces which can damage or detach molecules from the sample. Furthermore, in dynamic mode the (high frequency) signal can easily be distinguished from (low frequency) drift. (Hoogenboom 2006[2]) Figure 2.1: Constant frequency image of mica in water, size= 5x5 nm, f 0 = KHz, f= -8 KHz, scan speed= 48.8 nm/s, amplitude= 0.18 nm p p The LAFM in this report uses Frequency Modulation (FM). In the FM detection system a cantilever oscillates at resonance frequency in the vicinity of 5

6 6 CHAPTER 2. LIQUID AFM a sample surface, typically 900 KHz in the 2 nd bending mode. The gradient of the conservative tip-sample interaction force leads to a shift, f, of the eigenfrequency, f 0, of the cantilever. Dissipative forces damp the amplitude of the oscillation, A, but do not change f. The conservative tip-sample interaction varies over the oscillation cycle of the cantilever. The interaction force can be quantitatively determined, provided that f is known as a function of the tip-sample distance and that A is kept constant by adjusting the cantilever excitation. (Albrecht 1991[3], Giessibl 1995[4]) A typical result of a surface scan by a FM-LAFM is shown in figure 2.1. The scan is performed on mica in water at the 2 nd bending mode of the cantilever. The cantilever is photothermally excited by an excitation laser and the speed is measured by a heterodyne Doppler interferometer. The excitation of cantilevers in a liquid obviously has its limitations compared to operation in air. Resonant vibration of the cantilever is impeded by the large damping and added inertial mass of the liquid. These effects result in a low quality factor (Q factor) and a smaller shift of the resonance frequency, resulting in degradation of the sensitivity. (Ramos 2006[5]) However, if piezoelectric actuation is chosen to excite the cantilever, this will result in coupling of the support vibration to acoustic modes of the liquid. (Ratcliff 1998[6]) Furthermore, photothermal excitation enables the use of the torsion mode, in which lateral forces between the cantilever and the sample can be measured. A schematic drawing of the realized LAFM is shown in figure 2.2. In this figure the LAFM is divided into three parts: 1. Mechanical head, to fixate and position: 2. Optics 3. Control The objective lens Cantilever Sample Heterodyne Doppler interferometer to measure cantilever velocity Laser Diode for photothermal excitation of the cantilever Camera view for vision on the cantilever and the laser spots Approach control Phase Locked Forced Oscillation Method for imaging The three parts will be discussed in detail in Chapter 3, 4 and 5 respectively.

7 Figure 2.2: Schematic drawing of the LAFM, adjusted from [7] 7

8 Chapter 3 Mechanical head The mechanical head of the LAFM enables the positioning and fixation of the objective lens, cantilever and sample. Figure 3.1 shows the realized mechanical head. Figure 3.1: Mechanical head The design is made with the use of CAD software (Unigraphics NX3). The mechanical design is a changed version of the first version of the LAFM [7]. Technical drawings are included in Appendix A. The AFM is of the inverted type, which means that the probe is oriented upwards and the sample approaches from the upside. The mechanical head is mainly constructed from magnetic stainless steel (SUS304). For parts that do not need to be magnetic, duralumin or brass is used to save manufacturing-time. Figure 3.3 shows the lower part of the mechanical head in which the reference, measurement and excitation laser, together with the green LED light 8

9 3.1. CANTILEVER POSITIONING 9 for the CCD camera, enter the mechanical head in x-direction from the optics. The reference part of the Doppler interferometer passes the 90 -mirror and hits a normal mirror in the lower part of the mechanical head. The other beams need to be positioned by manipulation of mirrors in the optics-part to hit the 90 -mirror and go over the central axis of the objective lens, in z- direction. The objective lens (Olympus LUCPLFLN 20X) is designed for observation through a substrate and can be manually adjusted to get best focus depending on the thickness of the substrate. Furthermore, this lens has a relatively large working distance ( mm), where a minimum distance of 3 mm is needed to have the possibility to focus on the sample in the approach procedure (explained in more detail in section 5.1). This lens can easily be replaced by the 40X version instead, if needed in the future, which has a working distance of mm and can increase the measurement and excitation laser performance. The objective lens can be moved in z-direction to focus by manually moving the commercial x-stage in the lower part of the mechanical head. The x-stage moves the mounted wedge in x-direction, which moves the objectiveholder in z-direction. The movement of the objective lens is guided by a commercial guide fixed on the lower base of the mechanical head. 3.1 Cantilever positioning On top of the lower base two stainless steel (SUS304) stages are stacked to position the cantilever relatively to the objective central axis. One plate to guide the cantilever x-stage is fixed to the lower base and two stages for the x- and y- direction movement are stacked on top (figure 3.2). The two stages are moved by shear piezo s (Fuji-ceramics C-22). The walking principle of the shear piezo s is explained in Appendix B. The stages and lower base have a hole around the laser path for the objective lens to approach the probe. A glass substrate is placed over the hole in the cantilever y-stage, on which the cantilever is fixed. The fixation is further explained in section Cantilever fixation The silicon chip of the cantilever is fixed by anodic bonding. By applying a voltage across the glass substrate and the silicon chip, alkali ions in the glass move towards the negative electrode, leaving a thin negatively charged region at the interface to silicon. In other words, all voltage applied is applied to the

10 10 CHAPTER 3. MECHANICAL HEAD Figure 3.2: Schematic drawing upper part mechanical head, adjusted from [7] Figure 3.3: Schematic drawing lower part mechanical head, adjusted from [7]

11 3.2. SAMPLE POSITIONING 11 glass-silicon interface. While the voltage is applied, the glass is heated up to several hundred degrees Celsius. Due to the dielectric attraction force, added to the interface, the alkali atoms of glass (which are at the interface) make covalent bonds with silicon atoms in the silicon chip of the cantilever. The anodic bonding setup is shown in figure 3.4. Pyrex #7740 is used as glass substrate, because it contains the Alkali ions and its thermal expansion coefficient is close to that of silicon, which leaves less stress in the silicon chip and substrate. The glass substrate is fixed by placing magnets on top of the substrate, which attract the cantilever y-stage under the substrate. Furthermore, a rubber ring is positioned on the substrate, around the silicon chip, to surround the liquid applied. Figure 3.4: Anodic bonding setup 3.2 Sample positioning The rough x- and y-movement of the sample is also realized with shear piezo s. This can be seen in figure 3.2. The sample x-stage is guided by two stainless steel bars in the top base and the y-stage is stacked on top of the x-stage. The tube piezo holder is on top of the wedge stage and is pushed up, in z-direction,when the wedge stage moves horizontally, in y-direction. The movement of the sample relatively to the cantilever can be monitored by a CCD camera (Sony-DXC107A), which has view on the cantilever through a hole in de sample stages. The camera is equipped with a variable zoom lens (Suruga seiki - V20-501, variable zoom), the end of the lens can be seen in the left side of figure 3.5(a).

12 12 CHAPTER 3. MECHANICAL HEAD (a) (b) (c) Figure 3.5: Mechanical head; a) front view; b) left side view; c) right side view Tube piezo The tube piezo holder is shown in figure 3.6. There are three shear piezo s between the wedge stage and tube piezo holder and four between the Z-base and tube piezo holder. The magnets are placed to: - Generate enough normal force between the piezo s and the surface, to increase the friction. - Keep the tube piezo holder attached to the Z-base when the wedge stage moves.

13 3.2. SAMPLE POSITIONING 13 The seven piezo s attached to the tube piezo holder and the four piezo s under the wedge stage are fed by the same signal. This means that slip occurs between the wedge stage and tube piezo holder and between the Z-base and tube piezo holder, when the tube piezo holder is pushed up. The construction to make the rough z-movement of the sample is over constrained, with the two guides and seven piezo s with line and surface contacts. The reason that this construction is used is the choice to use an existing well working construction: a custom designed piezoelectric stage at Kawakatsu laboratory, which uses the same construction on a smaller scale. But, there is room for improvement, considering that this rough sample positioning in z-direction is used in the approach procedure. This procedure will be further explained in section 5.1. Figure 3.6: Tube piezo holder To test the rough z-movement of the sample, the step size of the tube piezo holder is measured. A triangular wave input signal is applied to the wedge stage and the movement of the tube piezo holder is measured by a capacitance sensor (MESS-TEK M-2213). The test results are shown in table 3.1. More details can be found in Appendix C. Table 3.1: Rough test results sample step size Amplitude [V ] Step size [nm] The tube piezo connected to the holder is glued to a steel plug on both sides. This can be seen in figure 3.7. The top plug is bolted to the tube piezo holder and the bottom plug is connected to the sample holder by a magnet. The magnet is located in a hole between the lower plug and the sample holder. A sample is glued to the sample holder.

14 14 CHAPTER 3. MECHANICAL HEAD The working principle of the tube piezo is shown in figure 3.7. By applying a voltage on the x- and y piezo s the scanning movements are performed. The z-piezo is used for probe to sample height adjustment. Figure 3.7: Tube piezo working principle; GND= ground

15 Chapter 4 Optics The optics of the LAFM consists of: A heterodyne Doppler interferometer to measure the cantilever deflection velocity. λ= nm; red laser Photo thermal excitation of the cantilever. λ= 405 nm; blue laser CCD camera with green LED light The three parts will be discussed in more detail in the next sections. A overview of the optical system is shown in figure 4.1. The setup is fixed on the same base plate as the mechanical head. To position and focus the measurement and excitation beam on the cantilever, a CCD camera is added. The camera is in upright position at the end of the dichroic mirror-line. A normal mirror under the camera enables the positioning of the camera view through the dichroic mirrors and over the central axis of the objective lens. In the (temporary) setup of figure 4.1 an extra dichroic mirror is added to introduce the green LED light. However, in the final setup the LED will be combined in the upright camera construction. Figure 4.2 shows the camera view of the cantilever, with the measurement- (red) and excitation (blue) laser. 4.1 Heterodyne Doppler interferometer A heterodyne Doppler interferometer is implemented. (Kawakatsu 2002[8]) The optical setup is depicted in figure 4.3 To analyze the difference in performance, two types of He-Ne laser were used as source: 15

16 16 CHAPTER 4. OPTICS Figure 4.1: Overview of the optics, with light paths; red laser= measurement; blue laser= excitation; PBS= Polarizing Beam Splitter; AOM= Acousto Optic Modulator; APD= Avalanche Photo Diode; LED= Light Emitting Diode; CCD= Charge-Coupled Device - Power stabilized (Neoark - 2MSS), λ= 633nm - Frequency stabilized (Neoark - Model 430), λ= 633nm The laser beam is first split into a reference and measurement beam at the first polarizing beam splitter (PBS). The beat frequency of the laser source reference beam is shifted with 80 MHz (f 1 ), on the reference path, by an acousto optic modulator (AOM, Panasonic EFLM) and continues towards the mechanical head to hit the reference mirror. The measurement beam directly goes to the objective lens (under the mechanical head) and hits the cantilever, getting a frequency shift, f 2, due to the Doppler effect. At the point where the reference and measurement beam come back from the mechanical head and are recombined (f 0 +f 1 ), interference takes place by which the signal contains f 1 + f 2 and f 1 f 2 peaks. The signal is detected by the two Avalanche Photo Diodes (APD s) after which one peak is selected by a band path filter and frequency demodulated by a delay line descriminator. A schematic example of the input and output of the APD s is shown in figure 4.4. Figure 4.1 also

17 4.1. HETERODYNE DOPPLER INTERFEROMETER 17 Figure 4.2: Camera view of the cantilever s backside Figure 4.3: Schematic of the heterodyne laser Doppler interferometer shows a lens in front of the APD s to focus the beam on the diodes after it is scattered mainly due to the cantilever. To get the best result from the Doppler interferometer it is preferable to locate the reference mirror next to the cantilever. However, this could not be realized in this setup, which means that the difference between the measurement en reference beam causes an error. Amongst other things this difference is caused by the difference in speed of the laser beam due to temperature/ pressure- gradients. To compensate that difference, the reference mirror is actuated by a PI-controlled piezo.(nishida 2008[9]) By feeding back the phase difference between the AOM and APD, the length of the light path is actively adjusted to compensate the difference in light speed. Tests on the Doppler interferometer were performed to determine the noise density of the setup using a He-Ne laser source and a laser diode source. These results will be shown in chapter 6.

18 18 CHAPTER 4. OPTICS Figure 4.4: top) carrier, AOM and signal, Cantilever ; bottom) resulting output 4.2 Photothermal excitation The most common method to generate cantilever oscillations uses a piezoelectric transducer. However, in a liquid the use of a piezoelectric transducer leads to coupling of the support vibration to acoustic modes of the liquid. This produces spectra that contain many features unrelated to the cantilever vibrational modes.(ratcliff 1998 [6]). Photothermal excitation does not have vibration coupling. To excite the cantilever with photothermal excitation, a laser is aimed at the cantilever s backside. A laser with λ= 405 nm is selected to focus on the silicon cantilever. A significant part of this light is absorbed by the cantilever and converted to thermal energy. The intensity-modulated light focused onto a region of the cantilever produces a time-dependent temperature distribution due to the absorption of optical energy.(ramos 2006 [5]). Furthermore, the spot location can be chosen to excite both bending and torsional modes of the cantilever. In figure 4.1 it can be seen that the optical fiber enters the optical setup and goes through a collimator to make the beam from the fiber parallel. After the collimator a variable lens is located to focus the laser beam on the cantilever.

19 Chapter 5 Control The project focussed on the design and realization of the mechanical head and optics. The control-part is implemented, but designed and realized by other members of Kawakatsu laboratory, for that reason the control part will only be explained shortly. The control of a conventional FM-AFM can be divided into two parts: 1. The cantilever oscillation is generated by self-oscillation. The detected vibration signal is fed back to the actuation through filtering, amplitude adjustment and phase adjustment. In self-oscillation the cantilever s frequency automatically corresponds with the cantilever s resonance frequency. 2. The oscillation frequency is monitored by a FM demodulator and z-axis distance is controlled in order to keep the frequency constant. These parts are shown schematically in figure 5.1. In a liquid the Q-factor degrades to 10 or less, which results in large fluctuations of the oscillation frequency around the center frequency, a poor carrier to noise ratio and it is difficult to keep small amplitude oscillations stable. This is why a different method, the Phase Locked Forced Oscillation Method (PLFO) is used. In this method the cantilever is driven by an oscillating signal from a stable signal generator and the phase change between the driving signal and the resulting vibration is used as feedback signal, which is shown in figure 5.2.(Kobayashi 2007 [10]) The lower Q-factor has less effect on the phase change with changing oscillation frequency compared with the amplitude change, thus using the phase change as feedback signal gives better results at low Q factors. 19

20 20 CHAPTER 5. CONTROL Figure 5.1: Conventional FM-AFM control scheme; self-oscillation and z- distance control loop[10] 5.1 Approach While discussing the control of the LAFM it is assumed that the LAFM is operating in imaging-mode. However, before imaging can take place an approach procedure is needed. In the first step of the approach procedure the wedge stage is actuated while watching the approach with use of the camera at the side of the mechanical head (section 3.2). The second camera, from the optics, is focused somewhat closer to the sample, above and next to the cantilever, by which the further approach can be monitored by watching the sample getting into focus. The final approach step uses the piezo walkers of the wedge stage together with the tube piezo: 1. The tube piezo is maximal retracted 2. One approaching step is performed with the piezo walker of the wedge stage 3. The tube piezo is extended and the oscillation frequency is monitored 4. If a change in frequency is detected the approach is finished, otherwise step 1-3 is repeated

21 5.1. APPROACH 21 Figure 5.2: Control scheme of the forced oscillation method at imaging[10]; LD= Laser Diode When applying the total approach procedure, the control scheme shown in figure 5.2 can not be used since the Z distance control loop doesn t work before the tip-sample distance comes into a range which the Z-tube can cover. Therefore the control scheme shown in figure 5.3 is used in the approach procedure. In the approach procedure, the frequency of the driving signal is controlled, similar to the self-oscillation method. The total control scheme is shown in figure 5.4. A soft switch is implemented to switch between the approach and imaging control scheme.

22 22 CHAPTER 5. CONTROL Figure 5.3: Control scheme of the forced oscillation method at approach[10]; LD= Laser Diode

23 5.1. APPROACH 23 Figure 5.4: Schematic overview control scheme[10]; PLFO= Phase Locked Forced Oscillation; PLFO-1= Imaging path; PLFO-2 Approach path

24 Chapter 6 Results At the end of the project the LAFM is completely realized. Some practical imperfections are left, therefore imaging is not yet possible. Furthermore, the mechanical head and optics need testing and optimization. The software, which is developed in parallel at Kawakatsu laboratory, by laboratory staffmember Y. Hoshi, has not been tested with the DSP (Digital Signal Processor) and the LAFM. Figure 6.1 shows the total LAFM setup. On the isolation table an aluminum base plate is located, on which the mechanical head and the optics are mounted. Around the table control devices and monitoring equipment can be seen. The noise density of the optics with the mechanical head is investigated to compare the home-build system with a commercial system. A schematic of the test setup is shown in figure 6.2, no liquid is applied on the cantilever in the test setup. The excitation frequency applied to the cantilever is variated from 1 Hz to 2 MHz by a network analyzer via the excitation laser. The cantilever velocity-measurement is performed using a output power stabilized He-Ne laser(neoark - 2MSS). The resulting signal on the APD s is connected to the FM-demodulator and fed back to the network analyzer, without any filtering. The sensitivity of the FM-demodulator can be chosen at 0.1, 1, 10, 100 or 1000 mm/s/v. The resulting noise density is shown in figure 6.3. At the frequency-range of interest, approximately MHz, the noise density lies between fm/ Hz. This is already better then commercial Doppler interferometers, which offer typical values in de range fm/ Hz in air. (Fukuma 2005[11]) Furthermore, it should be noted that the presented noise density data is obtained before optimization of the optics. 24

25 25 Figure 6.1: Picture of the realized LAFM The noise density is also measured using a Laser Diode as laser source. The resulting noise density, with a sensitivity of 0.1 mm/s/v, is also shown in figure 6.3. The Laser Diode data shows good coherence with the He-Ne data and more research will reveal if the He-Ne laser source can be replaced by the laser diode. Figure 6.2: Schematic overview of the noise density test setup

26 26 CHAPTER 6. RESULTS Figure 6.3: Noise density of the LAFM excitation and measurement system

27 Chapter 7 Conclusions & Recommendations 7.1 Conclusions In this report a new design of a LAFM has been presented and the results of it s realization have been shown. The mechanical head of the LAFM handles the actuation and positioning of the cantilever and sample, these functions are performed by shear piezo s. The performance of the rough sample approach has been shown, with step sizes down to 2 nm. The measurement of the cantilever s frequency has been performed by a home made Doppler interferometer, which has a lower noise density level compared to commercial available systems. The excitation of the cantilever is realized by photothermal actuation, which is combined with the measurement optics and camera view on the cantilever into an all-in-one optical setup. The control circuitry for phase locked oscillation is implemented. The mechanical head, optics and control, together with the software left a fully equipped LAFM. It should be noted that debugging and optimization is needed before atomic resolution images of surfaces in a liquid can be obtained. 27

28 28 CHAPTER 7. CONCLUSIONS & RECOMMENDATIONS 7.2 Recommendations In the quest for ever increasing image quality the following recommendations should be considered. Experience with the first version of the LAFM at Kawakatsu laboratory has shown that an important limiting factor in getting high resolution images is the influence of temperature gradients. Redesign the mechanical head in such a way that the probe tip of the cantilever is as close as possible to the thermal center. This way temperature changes have minimal effect on the image. Furthermore this also gives opportunities for surface scans in a liquid with changing temperature. Low expansion coefficient and high conductivity materials should be selected to minimize temperature influence on the construction. To be able to predict the movement of the mechanical head under the thermal changes (and loads) the over constrained stages, with four piezo s each, should be replaces by a statically determined construction. More specific, the current rough sample positioning, which is a critical component in the approach procedure, should be replaced. The current wedge-construction is significantly over constrained, which leads to uncertainty in the rough steps. In general the mechanical head should be scaled down. The main reason for the current size is the accessibility to the cantilever to apply the liquid. The accessibility should be solved, since scaling down the construction can significantly decrease geometric and thermal errors.

29 Bibliography [1] G. Binnig, C. F. Quate, C. Gerber, Physical Review Letters, 56, (1986) [2] B. W. Hoogenboom, H. J. Hug, Y. Pellmount, S. Martin, P. L. T. M. Frederix, D. Fotiadis and A. Engel, Applied physical letters, 88, (2006) [3] T. R. Albrecht, P. Grütter, D. Horne and D. Rugar, Applied physical letters, 69, 668 (1991) [4] F. J. Giessibl, Science, 267 pp (1995) [5] D. Ramos, J. Tamayo, J. Mertens and M. Calleja, Journal of applied physics, 99, (2006) [6] G. C. Ratcliff, A. E. Dorothy and R. Superfine, Applied physical letters, 72, 15 (1998) [7] S. Nishida, D. Kobayashi, T. Sakurada, T. Nakazawa, High-resolution imaging in liquid by Frequency Modulation Atomic Force Microscope (2005) [8] H. Kawakatsu, S. Kawai, D. Saya, M. Nagashio, D. Kobayashi, H. Toshiyoshi and H. Fujitao, Review of scientific instruments, 73, 6 (2002) [9] S. Nishida, D. Kobayashi, T. Sakurada and H. Kawakatsu, The 14 th International Colloquium on scanning Probe Microscopy, S7-5, [10] D. Kobayashi, presentation Control of a LAFM (2007) [11] T. Fukuma and S. P. Jarvis, Review of scientific instruments, 77 (2006) 29

30 Appendix A Mechanical drawings LAFM mechanical head A.1 Mechanical drawings lower part mechanical head 30

31 A.1. MECHANICAL DRAWINGS LOWER PART MECHANICAL HEAD

32 32APPENDIX A. MECHANICAL DRAWINGS LAFM MECHANICAL HEAD 1 2 Dural

33 A.1. MECHANICAL DRAWINGS LOWER PART MECHANICAL HEAD Brass

34 34APPENDIX A. MECHANICAL DRAWINGS LAFM MECHANICAL HEAD 3 1 Brass

35 A.1. MECHANICAL DRAWINGS LOWER PART MECHANICAL HEAD Dural

36 36APPENDIX A. MECHANICAL DRAWINGS LAFM MECHANICAL HEAD 6 1 SUS304

37 A.1. MECHANICAL DRAWINGS LOWER PART MECHANICAL HEAD SUS304

38 38APPENDIX A. MECHANICAL DRAWINGS LAFM MECHANICAL HEAD 8 1 SUS304

39 A.1. MECHANICAL DRAWINGS LOWER PART MECHANICAL HEAD SUS304

40 40APPENDIX A. MECHANICAL DRAWINGS LAFM MECHANICAL HEAD 10 1 SUS304

41 A.1. MECHANICAL DRAWINGS LOWER PART MECHANICAL HEAD SUS304

42 42APPENDIX A. MECHANICAL DRAWINGS LAFM MECHANICAL HEAD 12 1 Dural

43 A.1. MECHANICAL DRAWINGS LOWER PART MECHANICAL HEAD Dural

44 44APPENDIX A. MECHANICAL DRAWINGS LAFM MECHANICAL HEAD A.2 Mechanical drawings upper part mechanical head

45 A.2. MECHANICAL DRAWINGS UPPER PART MECHANICAL HEAD

46 46APPENDIX A. MECHANICAL DRAWINGS LAFM MECHANICAL HEAD 14 1 SUS304

47 A.2. MECHANICAL DRAWINGS UPPER PART MECHANICAL HEAD SUS304

48 48APPENDIX A. MECHANICAL DRAWINGS LAFM MECHANICAL HEAD 16 1 SUS304

49 A.2. MECHANICAL DRAWINGS UPPER PART MECHANICAL HEAD SUS304

50 50APPENDIX A. MECHANICAL DRAWINGS LAFM MECHANICAL HEAD SUS304

51 A.2. MECHANICAL DRAWINGS UPPER PART MECHANICAL HEAD SUS304

52 52APPENDIX A. MECHANICAL DRAWINGS LAFM MECHANICAL HEAD 18 1 SUS304

53 A.2. MECHANICAL DRAWINGS UPPER PART MECHANICAL HEAD SUS304

54 54APPENDIX A. MECHANICAL DRAWINGS LAFM MECHANICAL HEAD 20 3 SUS304

55 Appendix B Piezo actuated stages The working principal of the shear piezo s is shown in figure B.1. By applying a trapezoidal input voltage with variable frequency to the piezo s, they walk with variable speed over the surface. Figure B.1: Working principal of shear piezo s Each stage rests on four shear piezo s. Two piezo s walk over the flat stainless steel surface and the other two are equipped with a bearing and walk over two bars. The two bars function as guide for the stage. The bottom view of a stage is shown in figure B.2. 55

56 56 APPENDIX B. PIEZO ACTUATED STAGES Figure B.2: Bottom view of a stage

57 Appendix C Test results sample step size The tests to determine the step size are performed at 1 Hz. The test setup is shown in figure C.1. The capacitance sensor used in the test setup has a sensitivity of 10µm/V. The resulting oscilloscope output, for 100, 150 and 200 V input amplitude, is shown below for the approach and retract situation. Figure C.1: Schematic overview sample z-movement test setup 57

58 58 APPENDIX C. TEST RESULTS SAMPLE STEP SIZE Figure C.2: V = 100V, approach Figure C.3: V = 100V, retract Figure C.4: V = 150V, approach Figure C.5: V = 150V, retract Figure C.6: V = 200V, approach Figure C.7: V = 200V, retract

attosnom I: Topography and Force Images NANOSCOPY APPLICATION NOTE M06 RELATED PRODUCTS G

attosnom I: Topography and Force Images NANOSCOPY APPLICATION NOTE M06 RELATED PRODUCTS G APPLICATION NOTE M06 attosnom I: Topography and Force Images Scanning near-field optical microscopy is the outstanding technique to simultaneously measure the topography and the optical contrast of a sample.

More information

Advanced Nanoscale Metrology with AFM

Advanced Nanoscale Metrology with AFM Advanced Nanoscale Metrology with AFM Sang-il Park Corp. SPM: the Key to the Nano World Initiated by the invention of STM in 1982. By G. Binnig, H. Rohrer, Ch. Gerber at IBM Zürich. Expanded by the invention

More information

SENSOR+TEST Conference SENSOR 2009 Proceedings II

SENSOR+TEST Conference SENSOR 2009 Proceedings II B8.4 Optical 3D Measurement of Micro Structures Ettemeyer, Andreas; Marxer, Michael; Keferstein, Claus NTB Interstaatliche Hochschule für Technik Buchs Werdenbergstr. 4, 8471 Buchs, Switzerland Introduction

More information

- Near Field Scanning Optical Microscopy - Electrostatic Force Microscopy - Magnetic Force Microscopy

- Near Field Scanning Optical Microscopy - Electrostatic Force Microscopy - Magnetic Force Microscopy - Near Field Scanning Optical Microscopy - Electrostatic Force Microscopy - Magnetic Force Microscopy Yongho Seo Near-field Photonics Group Leader Wonho Jhe Director School of Physics and Center for Near-field

More information

OPTICS IN MOTION. Introduction: Competing Technologies: 1 of 6 3/18/2012 6:27 PM.

OPTICS IN MOTION. Introduction: Competing Technologies:  1 of 6 3/18/2012 6:27 PM. 1 of 6 3/18/2012 6:27 PM OPTICS IN MOTION STANDARD AND CUSTOM FAST STEERING MIRRORS Home Products Contact Tutorial Navigate Our Site 1) Laser Beam Stabilization to design and build a custom 3.5 x 5 inch,

More information

Outline: Introduction: What is SPM, history STM AFM Image treatment Advanced SPM techniques Applications in semiconductor research and industry

Outline: Introduction: What is SPM, history STM AFM Image treatment Advanced SPM techniques Applications in semiconductor research and industry 1 Outline: Introduction: What is SPM, history STM AFM Image treatment Advanced SPM techniques Applications in semiconductor research and industry 2 Back to our solutions: The main problem: How to get nm

More information

3D Optical Motion Analysis of Micro Systems. Heinrich Steger, Polytec GmbH, Waldbronn

3D Optical Motion Analysis of Micro Systems. Heinrich Steger, Polytec GmbH, Waldbronn 3D Optical Motion Analysis of Micro Systems Heinrich Steger, Polytec GmbH, Waldbronn SEMICON Europe 2012 Outline Needs and Challenges of measuring Micro Structure and MEMS Tools and Applications for optical

More information

Measurement of Microscopic Three-dimensional Profiles with High Accuracy and Simple Operation

Measurement of Microscopic Three-dimensional Profiles with High Accuracy and Simple Operation 238 Hitachi Review Vol. 65 (2016), No. 7 Featured Articles Measurement of Microscopic Three-dimensional Profiles with High Accuracy and Simple Operation AFM5500M Scanning Probe Microscope Satoshi Hasumura

More information

Study of shear force as a distance regulation mechanism for scanning near-field optical microscopy

Study of shear force as a distance regulation mechanism for scanning near-field optical microscopy Study of shear force as a distance regulation mechanism for scanning near-field optical microscopy C. Durkan a) and I. V. Shvets Department of Physics, Trinity College Dublin, Ireland Received 31 May 1995;

More information

NSOM (SNOM) Overview

NSOM (SNOM) Overview NSOM (SNOM) Overview The limits of far field imaging In the early 1870s, Ernst Abbe formulated a rigorous criterion for being able to resolve two objects in a light microscope: d > ë / (2sinè) where d

More information

Stability of a Fiber-Fed Heterodyne Interferometer

Stability of a Fiber-Fed Heterodyne Interferometer Stability of a Fiber-Fed Heterodyne Interferometer Christoph Weichert, Jens Flügge, Paul Köchert, Rainer Köning, Physikalisch Technische Bundesanstalt, Braunschweig, Germany; Rainer Tutsch, Technische

More information

Atomic Force Microscopy (Bruker MultiMode Nanoscope IIIA)

Atomic Force Microscopy (Bruker MultiMode Nanoscope IIIA) Atomic Force Microscopy (Bruker MultiMode Nanoscope IIIA) This operating procedure intends to provide guidance for general measurements with the AFM. For more advanced measurements or measurements with

More information

Lecture 20: Optical Tools for MEMS Imaging

Lecture 20: Optical Tools for MEMS Imaging MECH 466 Microelectromechanical Systems University of Victoria Dept. of Mechanical Engineering Lecture 20: Optical Tools for MEMS Imaging 1 Overview Optical Microscopes Video Microscopes Scanning Electron

More information

Nanoscale Material Characterization with Differential Interferometric Atomic Force Microscopy

Nanoscale Material Characterization with Differential Interferometric Atomic Force Microscopy Nanoscale Material Characterization with Differential Interferometric Atomic Force Microscopy F. Sarioglu, M. Liu, K. Vijayraghavan, A. Gellineau, O. Solgaard E. L. Ginzton Laboratory University Tip-sample

More information

1.6 Beam Wander vs. Image Jitter

1.6 Beam Wander vs. Image Jitter 8 Chapter 1 1.6 Beam Wander vs. Image Jitter It is common at this point to look at beam wander and image jitter and ask what differentiates them. Consider a cooperative optical communication system that

More information

Akiyama-Probe (A-Probe) guide

Akiyama-Probe (A-Probe) guide Akiyama-Probe (A-Probe) guide This guide presents: what is Akiyama-Probe, how it works, and what you can do Dynamic mode AFM Version: 2.0 Introduction NANOSENSORS Akiyama-Probe (A-Probe) is a self-sensing

More information

1393 DISPLACEMENT SENSORS

1393 DISPLACEMENT SENSORS 1393 DISPLACEMENT SENSORS INTRODUCTION While regular sensors detect the existence of objects, displacement sensors detect the amount of displacement when objects move from one position to another. Detecting

More information

Cutting-edge Atomic Force Microscopy techniques for large and multiple samples

Cutting-edge Atomic Force Microscopy techniques for large and multiple samples Cutting-edge Atomic Force Microscopy techniques for large and multiple samples Study of up to 200 mm samples using the widest set of AFM modes Industrial standards of automation A unique combination of

More information

Akiyama-Probe (A-Probe) guide

Akiyama-Probe (A-Probe) guide Akiyama-Probe (A-Probe) guide This guide presents: what is Akiyama-Probe, how it works, and its performance. Akiyama-Probe is a patented technology. Version: 2009-03-23 Introduction NANOSENSORS Akiyama-Probe

More information

la. Smith and C.P. Burger Department of Mechanical Engineering Texas A&M University College Station Tx

la. Smith and C.P. Burger Department of Mechanical Engineering Texas A&M University College Station Tx INJECTION LOCKED LASERS AS SURF ACE DISPLACEMENT SENSORS la. Smith and C.P. Burger Department of Mechanical Engineering Texas A&M University College Station Tx. 77843 INTRODUCTION In an age where engineered

More information

NEW LASER ULTRASONIC INTERFEROMETER FOR INDUSTRIAL APPLICATIONS B.Pouet and S.Breugnot Bossa Nova Technologies; Venice, CA, USA

NEW LASER ULTRASONIC INTERFEROMETER FOR INDUSTRIAL APPLICATIONS B.Pouet and S.Breugnot Bossa Nova Technologies; Venice, CA, USA NEW LASER ULTRASONIC INTERFEROMETER FOR INDUSTRIAL APPLICATIONS B.Pouet and S.Breugnot Bossa Nova Technologies; Venice, CA, USA Abstract: A novel interferometric scheme for detection of ultrasound is presented.

More information

Low Vibration, Low Thermal Fluctuation System for Pulse Tube and Gifford- McMahon Cryocoolers

Low Vibration, Low Thermal Fluctuation System for Pulse Tube and Gifford- McMahon Cryocoolers Low Vibration, Low Thermal Fluctuation System for Pulse Tube and Gifford- McMahon Cryocoolers L. Mauritsen, D. Snow, A. Woidtke, M. Chase, and I. Henslee S2 Corporation Bozeman, MT ABSTRACT A compact,

More information

Prepare Sample 3.1. Place Sample in Stage. Replace Probe (optional) Align Laser 3.2. Probe Approach 3.3. Optimize Feedback 3.4. Scan Sample 3.

Prepare Sample 3.1. Place Sample in Stage. Replace Probe (optional) Align Laser 3.2. Probe Approach 3.3. Optimize Feedback 3.4. Scan Sample 3. CHAPTER 3 Measuring AFM Images Learning to operate an AFM well enough to get an image usually takes a few hours of instruction and practice. It takes 5 to 10 minutes to measure an image if the sample is

More information

Module 5: Experimental Modal Analysis for SHM Lecture 36: Laser doppler vibrometry. The Lecture Contains: Laser Doppler Vibrometry

Module 5: Experimental Modal Analysis for SHM Lecture 36: Laser doppler vibrometry. The Lecture Contains: Laser Doppler Vibrometry The Lecture Contains: Laser Doppler Vibrometry Basics of Laser Doppler Vibrometry Components of the LDV system Working with the LDV system file:///d /neha%20backup%20courses%2019-09-2011/structural_health/lecture36/36_1.html

More information

Park NX-Hivac: Phase-lock Loop for Frequency Modulation Non-Contact AFM

Park NX-Hivac: Phase-lock Loop for Frequency Modulation Non-Contact AFM Park Atomic Force Microscopy Application note #21 www.parkafm.com Hosung Seo, Dan Goo and Gordon Jung, Park Systems Corporation Romain Stomp and James Wei Zurich Instruments Park NX-Hivac: Phase-lock Loop

More information

Laser stabilization and frequency modulation for trapped-ion experiments

Laser stabilization and frequency modulation for trapped-ion experiments Laser stabilization and frequency modulation for trapped-ion experiments Michael Matter Supervisor: Florian Leupold Semester project at Trapped Ion Quantum Information group July 16, 2014 Abstract A laser

More information

INDIAN INSTITUTE OF TECHNOLOGY BOMBAY

INDIAN INSTITUTE OF TECHNOLOGY BOMBAY IIT Bombay requests quotations for a high frequency conducting-atomic Force Microscope (c-afm) instrument to be set up as a Central Facility for a wide range of experimental requirements. The instrument

More information

Akinori Mitani and Geoff Weiner BGGN 266 Spring 2013 Non-linear optics final report. Introduction and Background

Akinori Mitani and Geoff Weiner BGGN 266 Spring 2013 Non-linear optics final report. Introduction and Background Akinori Mitani and Geoff Weiner BGGN 266 Spring 2013 Non-linear optics final report Introduction and Background Two-photon microscopy is a type of fluorescence microscopy using two-photon excitation. It

More information

1. Introduction. 2. Concept. reflector. transduce r. node. Kraftmessung an verschiedenen Fluiden in akustischen Feldern

1. Introduction. 2. Concept. reflector. transduce r. node. Kraftmessung an verschiedenen Fluiden in akustischen Feldern 1. Introduction The aim of this Praktikum is to familiarize with the concept and the equipment of acoustic levitation and to measure the forces exerted by an acoustic field on small spherical objects.

More information

Active Vibration Control in Ultrasonic Wire Bonding Improving Bondability on Demanding Surfaces

Active Vibration Control in Ultrasonic Wire Bonding Improving Bondability on Demanding Surfaces Active Vibration Control in Ultrasonic Wire Bonding Improving Bondability on Demanding Surfaces By Dr.-Ing. Michael Brökelmann, Hesse GmbH Ultrasonic wire bonding is an established technology for connecting

More information

Surface Finish Measurement Methods and Instrumentation

Surface Finish Measurement Methods and Instrumentation 125 years of innovation Surface Finish Measurement Methods and Instrumentation Contents Visual Inspection Surface Finish Comparison Plates Contact Gauges Inductive / Variable Reluctance (INTRA) Piezo Electric

More information

Absolute distance interferometer in LaserTracer geometry

Absolute distance interferometer in LaserTracer geometry Absolute distance interferometer in LaserTracer geometry Corresponding author: Karl Meiners-Hagen Abstract 1. Introduction 1 In this paper, a combination of variable synthetic and two-wavelength interferometry

More information

photolithographic techniques (1). Molybdenum electrodes (50 nm thick) are deposited by

photolithographic techniques (1). Molybdenum electrodes (50 nm thick) are deposited by Supporting online material Materials and Methods Single-walled carbon nanotube (SWNT) devices are fabricated using standard photolithographic techniques (1). Molybdenum electrodes (50 nm thick) are deposited

More information

Diode Laser Control Electronics. Diode Laser Locking and Linewidth Narrowing. Rudolf Neuhaus, Ph.D. TOPTICA Photonics AG

Diode Laser Control Electronics. Diode Laser Locking and Linewidth Narrowing. Rudolf Neuhaus, Ph.D. TOPTICA Photonics AG Appl-1012 Diode Laser Control Electronics Diode Laser Locking and Linewidth Narrowing Rudolf Neuhaus, Ph.D. TOPTICA Photonics AG Introduction Stabilized diode lasers are well established tools for many

More information

Kit for building your own THz Time-Domain Spectrometer

Kit for building your own THz Time-Domain Spectrometer Kit for building your own THz Time-Domain Spectrometer 16/06/2016 1 Table of contents 0. Parts for the THz Kit... 3 1. Delay line... 4 2. Pulse generator and lock-in detector... 5 3. THz antennas... 6

More information

Wavelength Control and Locking with Sub-MHz Precision

Wavelength Control and Locking with Sub-MHz Precision Wavelength Control and Locking with Sub-MHz Precision A PZT actuator on one of the resonator mirrors enables the Verdi output wavelength to be rapidly tuned over a range of several GHz or tightly locked

More information

Chapter 30: Principles of Active Vibration Control: Piezoelectric Accelerometers

Chapter 30: Principles of Active Vibration Control: Piezoelectric Accelerometers Chapter 30: Principles of Active Vibration Control: Piezoelectric Accelerometers Introduction: Active vibration control is defined as a technique in which the vibration of a structure is reduced or controlled

More information

Basic methods in imaging of micro and nano structures with atomic force microscopy (AFM)

Basic methods in imaging of micro and nano structures with atomic force microscopy (AFM) Basic methods in imaging of micro and nano P2538000 AFM Theory The basic principle of AFM is very simple. The AFM detects the force interaction between a sample and a very tiny tip (

More information

R. J. Jones College of Optical Sciences OPTI 511L Fall 2017

R. J. Jones College of Optical Sciences OPTI 511L Fall 2017 R. J. Jones College of Optical Sciences OPTI 511L Fall 2017 Active Modelocking of a Helium-Neon Laser The generation of short optical pulses is important for a wide variety of applications, from time-resolved

More information

Near-field Optical Microscopy

Near-field Optical Microscopy Near-field Optical Microscopy R. Fernandez, X. Wang, N. Li, K. Parker, and A. La Rosa Physics Department Portland State University Portland, Oregon Near-Field SPIE Optics Microscopy East 2005 Group PSU

More information

Characterization of Silicon-based Ultrasonic Nozzles

Characterization of Silicon-based Ultrasonic Nozzles Tamkang Journal of Science and Engineering, Vol. 7, No. 2, pp. 123 127 (24) 123 Characterization of licon-based Ultrasonic Nozzles Y. L. Song 1,2 *, S. C. Tsai 1,3, Y. F. Chou 4, W. J. Chen 1, T. K. Tseng

More information

INTEGRATED ACOUSTO-OPTICAL HETERODYNE INTERFEROMETER FOR DISPLACEMENT AND VIBRATION MEASUREMENT

INTEGRATED ACOUSTO-OPTICAL HETERODYNE INTERFEROMETER FOR DISPLACEMENT AND VIBRATION MEASUREMENT INTEGRATED ACOUSTO-OPTICAL HETERODYNE INTERFEROMETER FOR DISPLACEMENT AND VIBRATION MEASUREMENT AGUS RUBIYANTO Abstract A complex, fully packaged heterodyne interferometer has been developed for displacement

More information

attocube systems Probe Stations for Extreme Environments CRYOGENIC PROBE STATION fundamentals principles of cryogenic probe stations

attocube systems Probe Stations for Extreme Environments CRYOGENIC PROBE STATION fundamentals principles of cryogenic probe stations PAGE 88 & 2008 2007 PRODUCT CATALOG CRYOGENIC PROBE STATION fundamentals...................... 90 principles of cryogenic probe stations attocps I.......................... 92 ultra stable cryogenic probe

More information

FLUTTER CONTROL OF WIND TUNNEL MODEL USING A SINGLE ELEMENT OF PIEZO-CERAMIC ACTUATOR

FLUTTER CONTROL OF WIND TUNNEL MODEL USING A SINGLE ELEMENT OF PIEZO-CERAMIC ACTUATOR 24 TH INTERNATIONAL CONGRESS OF THE AERONAUTICAL SCIENCES FLUTTER CONTROL OF WIND TUNNEL MODEL USING A SINGLE ELEMENT OF PIEZO-CERAMIC ACTUATOR Naoki Kawai Department of Mechanical Engineering, University

More information

Constant Frequency / Lock-In (AM-AFM) Constant Excitation (FM-AFM) Constant Amplitude (FM-AFM)

Constant Frequency / Lock-In (AM-AFM) Constant Excitation (FM-AFM) Constant Amplitude (FM-AFM) HF2PLL Phase-locked Loop Connecting an HF2PLL to a Bruker Icon AFM / Nanoscope V Controller Zurich Instruments Technical Note Keywords: AM-AFM, FM-AFM, AFM control Release date: February 2012 Introduction

More information

visibility values: 1) V1=0.5 2) V2=0.9 3) V3=0.99 b) In the three cases considered, what are the values of FSR (Free Spectral Range) and

visibility values: 1) V1=0.5 2) V2=0.9 3) V3=0.99 b) In the three cases considered, what are the values of FSR (Free Spectral Range) and EXERCISES OF OPTICAL MEASUREMENTS BY ENRICO RANDONE AND CESARE SVELTO EXERCISE 1 A CW laser radiation (λ=2.1 µm) is delivered to a Fabry-Pérot interferometer made of 2 identical plane and parallel mirrors

More information

레이저의주파수안정화방법및그응용 박상언 ( 한국표준과학연구원, 길이시간센터 )

레이저의주파수안정화방법및그응용 박상언 ( 한국표준과학연구원, 길이시간센터 ) 레이저의주파수안정화방법및그응용 박상언 ( 한국표준과학연구원, 길이시간센터 ) Contents Frequency references Frequency locking methods Basic principle of loop filter Example of lock box circuits Quantifying frequency stability Applications

More information

UNIVERSITY OF WATERLOO Physics 360/460 Experiment #2 ATOMIC FORCE MICROSCOPY

UNIVERSITY OF WATERLOO Physics 360/460 Experiment #2 ATOMIC FORCE MICROSCOPY UNIVERSITY OF WATERLOO Physics 360/460 Experiment #2 ATOMIC FORCE MICROSCOPY References: http://virlab.virginia.edu/vl/home.htm (University of Virginia virtual lab. Click on the AFM link) An atomic force

More information

Electric polarization properties of single bacteria measured with electrostatic force microscopy

Electric polarization properties of single bacteria measured with electrostatic force microscopy Electric polarization properties of single bacteria measured with electrostatic force microscopy Theoretical and practical studies of Dielectric constant of single bacteria and smaller elements Daniel

More information

Optodevice Data Book ODE I. Rev.9 Mar Opnext Japan, Inc.

Optodevice Data Book ODE I. Rev.9 Mar Opnext Japan, Inc. Optodevice Data Book ODE-408-001I Rev.9 Mar. 2003 Opnext Japan, Inc. Section 1 Operating Principles 1.1 Operating Principles of Laser Diodes (LDs) and Infrared Emitting Diodes (IREDs) 1.1.1 Emitting Principles

More information

Part 2: Second order systems: cantilever response

Part 2: Second order systems: cantilever response - cantilever response slide 1 Part 2: Second order systems: cantilever response Goals: Understand the behavior and how to characterize second order measurement systems Learn how to operate: function generator,

More information

Technical Explanation for Displacement Sensors and Measurement Sensors

Technical Explanation for Displacement Sensors and Measurement Sensors Technical Explanation for Sensors and Measurement Sensors CSM_e_LineWidth_TG_E_2_1 Introduction What Is a Sensor? A Sensor is a device that measures the distance between the sensor and an object by detecting

More information

MEMS for RF, Micro Optics and Scanning Probe Nanotechnology Applications

MEMS for RF, Micro Optics and Scanning Probe Nanotechnology Applications MEMS for RF, Micro Optics and Scanning Probe Nanotechnology Applications Part I: RF Applications Introductions and Motivations What are RF MEMS? Example Devices RFIC RFIC consists of Active components

More information

Heisenberg) relation applied to space and transverse wavevector

Heisenberg) relation applied to space and transverse wavevector 2. Optical Microscopy 2.1 Principles A microscope is in principle nothing else than a simple lens system for magnifying small objects. The first lens, called the objective, has a short focal length (a

More information

SUPPLEMENTARY INFORMATION

SUPPLEMENTARY INFORMATION SUPPLEMENTARY INFORMATION Piezoresistive AFM cantilevers surpassing standard optical beam detection in low noise topography imaging Maja Dukic, Jonathan D. Adams and Georg E. Fantner Contents I Dependence

More information

CO2 laser heating system for thermal compensation of test masses in high power optical cavities. Submitted by: SHUBHAM KUMAR to Prof.

CO2 laser heating system for thermal compensation of test masses in high power optical cavities. Submitted by: SHUBHAM KUMAR to Prof. CO2 laser heating system for thermal compensation of test masses in high power optical cavities. Submitted by: SHUBHAM KUMAR to Prof. DAVID BLAIR Abstract This report gives a description of the setting

More information

NON-TRADITIONAL MACHINING PROCESSES ULTRASONIC, ELECTRO-DISCHARGE MACHINING (EDM), ELECTRO-CHEMICAL MACHINING (ECM)

NON-TRADITIONAL MACHINING PROCESSES ULTRASONIC, ELECTRO-DISCHARGE MACHINING (EDM), ELECTRO-CHEMICAL MACHINING (ECM) NON-TRADITIONAL MACHINING PROCESSES ULTRASONIC, ELECTRO-DISCHARGE MACHINING (EDM), ELECTRO-CHEMICAL MACHINING (ECM) A machining process is called non-traditional if its material removal mechanism is basically

More information

7. Michelson Interferometer

7. Michelson Interferometer 7. Michelson Interferometer In this lab we are going to observe the interference patterns produced by two spherical waves as well as by two plane waves. We will study the operation of a Michelson interferometer,

More information

D.C. Emmony, M.W. Godfrey and R.G. White

D.C. Emmony, M.W. Godfrey and R.G. White A MINIATURE OPTICAL ACOUSTIC EMISSION TRANSDUCER ABSTRACT D.C. Emmony, M.W. Godfrey and R.G. White Department of Physics Loughborough University of Technology Loughborough, Leicestershire LEll 3TU United

More information

Supplementary Figure 1

Supplementary Figure 1 Supplementary Figure 1 Technical overview drawing of the Roadrunner goniometer. The goniometer consists of three main components: an inline sample-viewing microscope, a high-precision scanning unit for

More information

Scanning Tunneling Microscopy

Scanning Tunneling Microscopy EMSE-515 02 Scanning Tunneling Microscopy EMSE-515 F. Ernst 1 Scanning Tunneling Microscope: Working Principle 2 Scanning Tunneling Microscope: Construction Principle 1 sample 2 sample holder 3 clamps

More information

Development of a Package for a Triaxial High-G Accelerometer Optimized for High Signal Fidelity

Development of a Package for a Triaxial High-G Accelerometer Optimized for High Signal Fidelity Development of a Package for a Triaxial High-G Accelerometer Optimized for High Signal Fidelity R. Langkemper* 1, R. Külls 1, J. Wilde 2, S. Schopferer 1 and S. Nau 1 1 Fraunhofer Institute for High-Speed

More information

Installation and Characterization of the Advanced LIGO 200 Watt PSL

Installation and Characterization of the Advanced LIGO 200 Watt PSL Installation and Characterization of the Advanced LIGO 200 Watt PSL Nicholas Langellier Mentor: Benno Willke Background and Motivation Albert Einstein's published his General Theory of Relativity in 1916,

More information

ADALAM Sensor based adaptive laser micromachining using ultrashort pulse lasers for zero-failure manufacturing D2.2. Ger Folkersma (Demcon)

ADALAM Sensor based adaptive laser micromachining using ultrashort pulse lasers for zero-failure manufacturing D2.2. Ger Folkersma (Demcon) D2.2 Automatic adjustable reference path system Document Coordinator: Contributors: Dissemination: Keywords: Ger Folkersma (Demcon) Ger Folkersma, Kevin Voss, Marvin Klein (Demcon) Public Reference path,

More information

MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Electrical Engineering and Computer Science

MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Electrical Engineering and Computer Science Student Name Date MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Electrical Engineering and Computer Science 6.161 Modern Optics Project Laboratory Laboratory Exercise No. 6 Fall 2010 Solid-State

More information

POCKET DEFORMABLE MIRROR FOR ADAPTIVE OPTICS APPLICATIONS

POCKET DEFORMABLE MIRROR FOR ADAPTIVE OPTICS APPLICATIONS POCKET DEFORMABLE MIRROR FOR ADAPTIVE OPTICS APPLICATIONS Leonid Beresnev1, Mikhail Vorontsov1,2 and Peter Wangsness3 1) US Army Research Laboratory, 2800 Powder Mill Road, Adelphi Maryland 20783, lberesnev@arl.army.mil,

More information

NANOSCOPIC EVALUATION OF MICRO-SYSTEMS

NANOSCOPIC EVALUATION OF MICRO-SYSTEMS NANOSCOPIC EVALUATION OF MICRO-SYSTEMS A. Altes 1, L.J. Balk 1, H.L. Hartnagel 2, R. Heiderhoff 1, K. Mutamba 2, and Ch. Thomas 1 1 Bergische Universität Wuppertal, Lehrstuhl für Elektronik, Wuppertal,

More information

Spectrometer using a tunable diode laser

Spectrometer using a tunable diode laser Spectrometer using a tunable diode laser Ricardo Vasquez Department of Physics, Purdue University, West Lafayette, IN April, 2000 In the following paper the construction of a simple spectrometer using

More information

Zeeman Shifted Modulation Transfer Spectroscopy in Atomic Cesium

Zeeman Shifted Modulation Transfer Spectroscopy in Atomic Cesium Zeeman Shifted Modulation Transfer Spectroscopy in Atomic Cesium Modulation transfer spectroscopy (MTS) is a useful technique for locking a laser on one of the closed cesium D transitions. We have focused

More information

Standard Operating Procedure of Atomic Force Microscope (Anasys afm+)

Standard Operating Procedure of Atomic Force Microscope (Anasys afm+) Standard Operating Procedure of Atomic Force Microscope (Anasys afm+) The Anasys Instruments afm+ system incorporates an Atomic Force Microscope which can scan the sample in the contact mode and generate

More information

Modal analysis: a comparison between Finite Element Analysis (FEA) and practical Laser Doppler Vibrometer (LDV) testing.

Modal analysis: a comparison between Finite Element Analysis (FEA) and practical Laser Doppler Vibrometer (LDV) testing. 2017 UKSim-AMSS 19th International Conference on Modelling & Simulation Modal analysis: a comparison between Finite Element Analysis (FEA) and practical Laser Doppler Vibrometer (LDV) testing. Luca Pagano

More information

Multi-Probe Atomic Force Microscopy Using Piezo-Resistive Cantilevers and Interaction between Probes

Multi-Probe Atomic Force Microscopy Using Piezo-Resistive Cantilevers and Interaction between Probes e-journal of Surface Science and Nanotechnology 26 January 2013 e-j. Surf. Sci. Nanotech. Vol. 11 (2013) 13-17 Regular Paper Multi-Probe Atomic Force Microscopy Using Piezo-Resistive Cantilevers and Interaction

More information

PACS Nos v, Fc, Yd, Fs

PACS Nos v, Fc, Yd, Fs A Shear Force Feedback Control System for Near-field Scanning Optical Microscopes without Lock-in Detection J. W. P. Hsu *,a, A. A. McDaniel a, and H. D. Hallen b a Department of Physics, University of

More information

Manufacturing Metrology Team

Manufacturing Metrology Team The Team has a range of state-of-the-art equipment for the measurement of surface texture and form. We are happy to discuss potential measurement issues and collaborative research Manufacturing Metrology

More information

Radial Polarization Converter With LC Driver USER MANUAL

Radial Polarization Converter With LC Driver USER MANUAL ARCoptix Radial Polarization Converter With LC Driver USER MANUAL Arcoptix S.A Ch. Trois-portes 18 2000 Neuchâtel Switzerland Mail: info@arcoptix.com Tel: ++41 32 731 04 66 Principle of the radial polarization

More information

plasmonic nanoblock pair

plasmonic nanoblock pair Nanostructured potential of optical trapping using a plasmonic nanoblock pair Yoshito Tanaka, Shogo Kaneda and Keiji Sasaki* Research Institute for Electronic Science, Hokkaido University, Sapporo 1-2,

More information

Instruction manual and data sheet ipca h

Instruction manual and data sheet ipca h 1/15 instruction manual ipca-21-05-1000-800-h Instruction manual and data sheet ipca-21-05-1000-800-h Broad area interdigital photoconductive THz antenna with microlens array and hyperhemispherical silicon

More information

S.No Description/Specifications Qty 01. Post office box Trainer.

S.No Description/Specifications Qty 01. Post office box Trainer. Specification of Equipments for Physics lab S.No Description/Specifications Qty 01. Post office box Trainer. 06 The trainer should have: On Board DC Power Supply : 5V Galvanometer ; Deflection : 30 0 30

More information

SA210-Series Scanning Fabry Perot Interferometer

SA210-Series Scanning Fabry Perot Interferometer 435 Route 206 P.O. Box 366 PH. 973-579-7227 Newton, NJ 07860-0366 FAX 973-300-3600 www.thorlabs.com technicalsupport@thorlabs.com SA210-Series Scanning Fabry Perot Interferometer DESCRIPTION: The SA210

More information

Chemistry Instrumental Analysis Lecture 10. Chem 4631

Chemistry Instrumental Analysis Lecture 10. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 10 Types of Instrumentation Single beam Double beam in space Double beam in time Multichannel Speciality Types of Instrumentation Single beam Requires stable

More information

Figure for the aim4np Report

Figure for the aim4np Report Figure for the aim4np Report This file contains the figures to which reference is made in the text submitted to SESAM. There is one page per figure. At the beginning of the document, there is the front-page

More information

A Narrow-Band Tunable Diode Laser System with Grating Feedback

A Narrow-Band Tunable Diode Laser System with Grating Feedback A Narrow-Band Tunable Diode Laser System with Grating Feedback S.P. Spirydovich Draft Abstract The description of diode laser was presented. The tuning laser system was built and aligned. The free run

More information

Applications of Optics

Applications of Optics Nicholas J. Giordano www.cengage.com/physics/giordano Chapter 26 Applications of Optics Marilyn Akins, PhD Broome Community College Applications of Optics Many devices are based on the principles of optics

More information

:... resolution is about 1.4 μm, assumed an excitation wavelength of 633 nm and a numerical aperture of 0.65 at 633 nm.

:... resolution is about 1.4 μm, assumed an excitation wavelength of 633 nm and a numerical aperture of 0.65 at 633 nm. PAGE 30 & 2008 2007 PRODUCT CATALOG Confocal Microscopy - CFM fundamentals :... Over the years, confocal microscopy has become the method of choice for obtaining clear, three-dimensional optical images

More information

A Project Report Submitted to the Faculty of the Graduate School of the University of Minnesota By

A Project Report Submitted to the Faculty of the Graduate School of the University of Minnesota By Observation and Manipulation of Gold Clusters with Scanning Tunneling Microscopy A Project Report Submitted to the Faculty of the Graduate School of the University of Minnesota By Dogukan Deniz In Partial

More information

A Thermal Compensation System for the gravitational wave detector Virgo

A Thermal Compensation System for the gravitational wave detector Virgo A Thermal Compensation System for the gravitational wave detector Virgo M. Di Paolo Emilio University of L Aquila and INFN Roma Tor Vergata On behalf of the Virgo Collaboration Index: 1) Thermal Lensing

More information

ATOMIC FORCE MICROSCOPY

ATOMIC FORCE MICROSCOPY B47 Physikalisches Praktikum für Fortgeschrittene Supervision: Prof. Dr. Sabine Maier sabine.maier@physik.uni-erlangen.de ATOMIC FORCE MICROSCOPY Version: E1.4 first edit: 15/09/2015 last edit: 05/10/2018

More information

Universal and compact laser stabilization electronics

Universal and compact laser stabilization electronics top-of-fringe LaseLock LaseLock Universal and compact laser stabilization electronics Compact, stand-alone locking electronics for diode lasers, dye lasers, Ti:Sa lasers, or optical resonators Side-of-fringe

More information

DEVELOPMENT OF HEAT-RESISTANT OPTICAL FIBER AE SENSOR

DEVELOPMENT OF HEAT-RESISTANT OPTICAL FIBER AE SENSOR DEVELOPMENT OF HEAT-RESISTANT OPTICAL FIBER AE SENSOR PORNTHEP CHIVAVIBUL 1, HIROYUKI FUKUTOMI 1, SHIN TAKAHASHI 2 and YUICHI MACHIJIMA 2 1) Central Research Institute of Electric Power Industry (CRIEPI),

More information

LOS 1 LASER OPTICS SET

LOS 1 LASER OPTICS SET LOS 1 LASER OPTICS SET Contents 1 Introduction 3 2 Light interference 5 2.1 Light interference on a thin glass plate 6 2.2 Michelson s interferometer 7 3 Light diffraction 13 3.1 Light diffraction on a

More information

DEVELOPMENT OF STABILIZED AND HIGH SENSITIVE OPTICAL FI- BER ACOUSTIC EMISSION SYSTEM AND ITS APPLICATION

DEVELOPMENT OF STABILIZED AND HIGH SENSITIVE OPTICAL FI- BER ACOUSTIC EMISSION SYSTEM AND ITS APPLICATION DEVELOPMENT OF STABILIZED AND HIGH SENSITIVE OPTICAL FI- BER ACOUSTIC EMISSION SYSTEM AND ITS APPLICATION HIDEO CHO, RYOUHEI ARAI and MIKIO TAKEMOTO Faculty of Mechanical Engineering, Aoyama Gakuin University,

More information

A Review of MEMS Based Piezoelectric Energy Harvester for Low Frequency Applications

A Review of MEMS Based Piezoelectric Energy Harvester for Low Frequency Applications Available Online at www.ijcsmc.com International Journal of Computer Science and Mobile Computing A Monthly Journal of Computer Science and Information Technology IJCSMC, Vol. 3, Issue. 9, September 2014,

More information

Development of Shock Acceleration Calibration Machine in NMIJ

Development of Shock Acceleration Calibration Machine in NMIJ IMEKO 20 th TC3, 3 rd TC16 and 1 st TC22 International Conference Cultivating metrological knowledge 27 th to 30 th November, 2007. Merida, Mexico. Development of Shock Acceleration Calibration Machine

More information

Theory and Applications of Frequency Domain Laser Ultrasonics

Theory and Applications of Frequency Domain Laser Ultrasonics 1st International Symposium on Laser Ultrasonics: Science, Technology and Applications July 16-18 2008, Montreal, Canada Theory and Applications of Frequency Domain Laser Ultrasonics Todd W. MURRAY 1,

More information

Micro-sensors - what happens when you make "classical" devices "small": MEMS devices and integrated bolometric IR detectors

Micro-sensors - what happens when you make classical devices small: MEMS devices and integrated bolometric IR detectors Micro-sensors - what happens when you make "classical" devices "small": MEMS devices and integrated bolometric IR detectors Dean P. Neikirk 1 MURI bio-ir sensors kick-off 6/16/98 Where are the targets

More information

Laser Locking with Doppler-free Saturated Absorption Spectroscopy

Laser Locking with Doppler-free Saturated Absorption Spectroscopy Laser Locking with Doppler-free Saturated Absorption Spectroscopy Paul L. Stubbs, Advisor: Irina Novikova W&M Quantum Optics Group May 12, 2010 Abstract The goal of this project was to lock the frequency

More information

Angular Drift of CrystalTech (1064nm, 80MHz) AOMs due to Thermal Transients. Alex Piggott

Angular Drift of CrystalTech (1064nm, 80MHz) AOMs due to Thermal Transients. Alex Piggott Angular Drift of CrystalTech 38 197 (164nm, 8MHz) AOMs due to Thermal Transients Alex Piggott July 5, 21 1 .1 General Overview of Findings The AOM was found to exhibit significant thermal drift effects,

More information

Optical Microscope. Active anti-vibration table. Mechanical Head. Computer and Software. Acoustic/Electrical Shield Enclosure

Optical Microscope. Active anti-vibration table. Mechanical Head. Computer and Software. Acoustic/Electrical Shield Enclosure Optical Microscope On-axis optical view with max. X magnification Motorized zoom and focus Max Field of view: mm x mm (depends on zoom) Resolution : um Working Distance : mm Magnification : max. X Zoom

More information

Initial Results from the C-Mod Prototype Polarimeter/Interferometer

Initial Results from the C-Mod Prototype Polarimeter/Interferometer Initial Results from the C-Mod Prototype Polarimeter/Interferometer K. R. Smith, J. Irby, R. Leccacorvi, E. Marmar, R. Murray, R. Vieira October 24-28, 2005 APS-DPP Conference 1 Abstract An FIR interferometer-polarimeter

More information

SPRAY DROPLET SIZE MEASUREMENT

SPRAY DROPLET SIZE MEASUREMENT SPRAY DROPLET SIZE MEASUREMENT In this study, the PDA was used to characterize diesel and different blends of palm biofuel spray. The PDA is state of the art apparatus that needs no calibration. It is

More information