INDIAN INSTITUTE OF TECHNOLOGY BOMBAY
|
|
- Stephany Clark
- 5 years ago
- Views:
Transcription
1 IIT Bombay requests quotations for a high frequency conducting-atomic Force Microscope (c-afm) instrument to be set up as a Central Facility for a wide range of experimental requirements. The instrument must have a robust design as it will be used daily by students of all levels of experience. A major fraction of the usage will be related to imaging and conductance measurements on thin film substrates and advanced nanoscale devices. As this will be a shared facility instrument, an upgrade path for additional advanced modes, heated chuck, etc is preferred. The high frequency c-afm must meet all of the requirements listed below: Atomic Force Microscope 1. Operating Modes and Environmental Controls The AFM system must be able to image samples and perform measurements in air and in liquid using the same cantilever holder. The cantilever holder must be compatible with most commercial cantilevers. The AFM system must include the following modes: Scanning modes: Contact, AC (tapping), Multi-frequency (able to drive and detect the cantilever motion at more than one frequency simultaneously), frequency modulation (FM), force modulation, electric force microscopy (EFM), magnetic force (MFM), Kelvin Probe (SKPM), Conducting AFM (CAFM), and Piezoresponse Force Microscopy (PFM) Point measurement modes:nanolithography and Nanomanipulation, Force vs Displacement (single point and mapping), IV measurements, and pulse measurements. Conductive AFM The system must allow measurement of current flowing through the sample in the scanning mode, at user-programmable locations and at applied bias voltages.
2 It must be possible to apply complex voltage wave forms such as square, sine, triangle, pulse, or other arbitrary user defined shapes, while concurrently monitoring the current flow through the tip. For DC measurements, it must be possible to sense currents as low as 1pA. It must be possible to apply voltage pulses with rise times, hold times and fall times, all between 5 and 15 ns to the tip and measure current flowing through the tip during the pulse. It should be possible to measure low currents flowing through the sample, at least 10µA or lower, concurrently while the tip is in contact with the sample and a voltage pulse or waveform is applied. It must be possible to apply bias voltages to the sample in the range of -10V to +10V It must be possible to apply an AC bias/signal either to the tip or the sample through software control. The software must provide easy analysis and plotting so that I-V curves can be visualized during measurement in a straight-forward manner. It must be possible to obtain conductivity maps of the sample, which could be overlaid with topographic or imaging maps, in real-time. 2. AFM Scanner and Optical Lever Detection Systems Instrument Resolution It must be possible to obtain atomic lattice resolution in AC mode and contact mode imaging. The same scanner should also be capable of imaging large areas: at least 50µm 50µm 10 µm. Scanner
3 A single closed-loop scanning system with XY range of at least 50 µm and Z range of at least 10 µm is required. The XY scanner must be separate from the Z scanner. In order to conduct tip based measurements, sample scanning must be possible. It should be possible to accept samples up to 50mm lateral dimensions and at least 10mm thickness. The noise in the X and Y sensors must be less than 100pm ADev in a 0.1Hz to 1kHz bandwidth. Z sensor noise must be less than 500pm ADev in a 0.1Hz to 1kHz bandwidth, with sensor nonlinearity <0.1% at full scan. Optical Lever Arm: Light Source and Photodetector The cantilever holder and the optical lever assembly (laser, optics, and detector) must move together on a single rigid frame to eliminate relative motion between components which leads to errors and artifacts during imaging. The optical lever arm must use a low coherence light source (such as an SLD) to reduce artifacts from optical interference effects. The instrument must use an infrared SLD (or equivalent) for the optical lever arm to eliminate optical crosstalk. The SLD (or equivalent) must preferably be incident on the cantilever at an angle (20-25 degrees) to the sample normal to reduce interference and minimize noise. The instrument s DC Detector Noise must be less than 50pm ADev in a 0.1Hz to 1kHz BW. DC Height Noise (deflection noise with the tip on the surface and the gains turned off) must be less than 50pm ADev in a 0.1Hz to 1kHz bandwidth.
4 The photodetector must have an optical sensing bandwidth of at least DC to 1 MHz. 3. AFM Control System Electronics All the control must be 100% digital operation. The sensing bandwidth of the controller must be at least 1 MHz. The overall thermal tunes of the cantilever must be up to at least 1 MHz. There should be BNC access to all major input and output signals from the electronic controller. The AFM instrument must include auto-configuration of external hardware and accessories (ie. plug and play). The instrument must include a user programmable control knob that can be used to fine tune and adjust all scan parameters during any advanced operation. Software The system software must be based in Windows or Linux for compatibility with university network and other laboratory instrument software systems. The software must be capable of driving the cantilever simultaneously at two or more arbitrarily chosen excitation frequencies in AC (dynamic) mode, while simultaneously collecting and displaying the amplitude and phase signals and images from each of these frequencies, along with the height or Z-sensor data. All software must be in open-source software programming language so that users can easily add their control and measurement programs. The system software must include a macro building graphical user interface to automate complete multiple-step experiments and data processing sequences.
5 The system s software must support one-click configuration for most standard measurements. It must be possible to render imaged surfaces in 3D in real-time, during imaging. The AFM control software must include optical image navigation using any realtime optical input. The software must have an image overlay function to combine AFM images with optical images or other plots (such as conductivity maps) for direct feature comparison. The software must allow image files to be exported in standard formats as JPEG, PNG, BMP and TIFF. Must include built-in nanolithography and nanomanipulation software. Must include drift compensation software so that a region of interest can be tracked in real time (<5 nm precision) thereby eliminating any distortion in the image. Drift compensation should be possible during any advanced characterization mode (imaging, pulsed measurement, etc). Must include free offline software for analysis of experimental data. The software must be compatible with PC and Mac for offline image processing. Free software upgrades must be available for the life of the instrument. System Computer The System must include a current commercial computer model suitable for controlling the AFM system and performing data analysis simultaneously. Dual monitors at least 20 must be included. The Computer-to-Controller communication must be via USB2 for easy upgrades.
6 4. Instrument Environmental Isolation The system must include a thermally- and acoustically-isolating enclosure. The enclosure must provide at least 20dB of acoustic isolation. An air temperature control system should be optionally available for the acoustic hood to minimize thermal drift. The system must include an active vibration isolation table. 6. Guarantee, Warranty, Support and Service Must include at least three (3) year warranty on all parts and labor. Must include free AFM software upgrades for the life of the instrument. Must provide biannual training sessions for students during the warranty period.
Cutting-edge Atomic Force Microscopy techniques for large and multiple samples
Cutting-edge Atomic Force Microscopy techniques for large and multiple samples Study of up to 200 mm samples using the widest set of AFM modes Industrial standards of automation A unique combination of
More informationAsylum Research. MFP-3D Infinity. Endless Applications. Unlimited Potential. Performance / Versatility / Support
MFP-3D Infinity AFM Asylum Research Endless Applications. Unlimited Potential. Performance / Versatility / Support MFP-3D Infinity AFM Asylum Research Endless applications. Unlimited potential. The Asylum
More informationPark NX20 The leading nano metrology tool for failure analysis and large sample research.
The Most Accurate Atomic Force Microscope Park NX20 The leading nano metrology tool for failure analysis and large sample research www.parkafm.com The Most Accurate Atomic Force Microscope Park NX20 The
More informationPark XE7 The most affordable research grade AFM with flexible sample handling.
Park XE7 The most affordable research grade AFM with flexible sample handling www.parkafm.com Park Systems The Most Accurate Atomic Force Microscope Park XE7 The economical choice for innovative research
More informationThe Most Accurate Atomic Force Microscope. Park NX20 The leading nano metrology tool for failure analysis and large sample research.
The Most Accurate Atomic Force Microscope Park NX20 The leading nano metrology tool for failure analysis and large sample research www.parkafm.com Park Systems The Most Accurate Atomic Force Microscope
More informationIndian Institute of Technology Bombay
Specifications for High Resolution Scanning Probe Microscope Item Essential measuring modes with complete hardware and software. All the modes should be demonstrated during installation & training Scanners
More informationMeasurement of Microscopic Three-dimensional Profiles with High Accuracy and Simple Operation
238 Hitachi Review Vol. 65 (2016), No. 7 Featured Articles Measurement of Microscopic Three-dimensional Profiles with High Accuracy and Simple Operation AFM5500M Scanning Probe Microscope Satoshi Hasumura
More informationMaterials Science Devices Life Science Advanced Applications
TM Head XY Scanner Base Technical Innovations TM Sensored, closed loop positioning for high resolution imaging, accuracy, and reproducibility. Pioneering all-digital controller for open software adaptability,
More information- Near Field Scanning Optical Microscopy - Electrostatic Force Microscopy - Magnetic Force Microscopy
- Near Field Scanning Optical Microscopy - Electrostatic Force Microscopy - Magnetic Force Microscopy Yongho Seo Near-field Photonics Group Leader Wonho Jhe Director School of Physics and Center for Near-field
More informationElectronic Characterization of Materials Using Conductive AFM
Electronic Characterization of Materials Using Conductive AFM Amir Moshar Electrical Measurements SKPM EFM CAFM PFM SCM Non-Contact Electrical Techniques Scanning Kelvin Probe Microscopy Electric Force
More informationAdvanced Nanoscale Metrology with AFM
Advanced Nanoscale Metrology with AFM Sang-il Park Corp. SPM: the Key to the Nano World Initiated by the invention of STM in 1982. By G. Binnig, H. Rohrer, Ch. Gerber at IBM Zürich. Expanded by the invention
More informationThe Most Accurate Atomic Force Microscope. Park XE15 Power and versatility, brilliantly combined.
The Most Accurate Atomic Force Microscope Park XE15 Power and versatility, brilliantly combined. www.parkafm.com Park XE15 Increase your productivity with our powerfully versatile atomic force microscope
More informationStandard Operating Procedure of Atomic Force Microscope (Anasys afm+)
Standard Operating Procedure of Atomic Force Microscope (Anasys afm+) The Anasys Instruments afm+ system incorporates an Atomic Force Microscope which can scan the sample in the contact mode and generate
More informationSPM The Industry s Performance Leader High Resolution Closed-loop System Fast, Easy Tip & Sample Exchange Versatility and Value Powerful Research
SPM The Industry s Performance Leader High Resolution Closed-loop System Fast, Easy Tip & Sample Exchange Versatility and Value Powerful Research Flexibility Atomic resolution STM image of highly-oriented
More informationOutline: Introduction: What is SPM, history STM AFM Image treatment Advanced SPM techniques Applications in semiconductor research and industry
1 Outline: Introduction: What is SPM, history STM AFM Image treatment Advanced SPM techniques Applications in semiconductor research and industry 2 Back to our solutions: The main problem: How to get nm
More informationOptions and Accessories for Asylum Research MFP-3D AFMs
ACCESSORIES Options and Accessories for Asylum Research MFP-3D AFMs Empower your research with powerful, innovative new capabilities Go beyond topography with advanced modes and environmental control Powerful
More informationInvestigate in magnetic micro and nano structures by Magnetic Force Microscopy (MFM)
Investigate in magnetic micro and nano 5.3.85- Related Topics Magnetic Forces, Magnetic Force Microscopy (MFM), phase contrast imaging, vibration amplitude, resonance shift, force Principle Caution! -
More informationUniversity of MN, Minnesota Nano Center Standard Operating Procedure
Equipment Name: Atomic Force Microscope Badger name: afm DI5000 PAN Revisionist Paul Kimani Model: Dimension 5000 Date: October 6, 2017 Location: Bay 1 PAN Revision: 1 A. Description i. Enhanced Motorized
More informationOptical Microscope. Active anti-vibration table. Mechanical Head. Computer and Software. Acoustic/Electrical Shield Enclosure
Optical Microscope On-axis optical view with max. X magnification Motorized zoom and focus Max Field of view: mm x mm (depends on zoom) Resolution : um Working Distance : mm Magnification : max. X Zoom
More informationAkiyama-Probe (A-Probe) guide
Akiyama-Probe (A-Probe) guide This guide presents: what is Akiyama-Probe, how it works, and what you can do Dynamic mode AFM Version: 2.0 Introduction NANOSENSORS Akiyama-Probe (A-Probe) is a self-sensing
More informationPark NX-Hivac: Phase-lock Loop for Frequency Modulation Non-Contact AFM
Park Atomic Force Microscopy Application note #21 www.parkafm.com Hosung Seo, Dan Goo and Gordon Jung, Park Systems Corporation Romain Stomp and James Wei Zurich Instruments Park NX-Hivac: Phase-lock Loop
More informationMEMS for RF, Micro Optics and Scanning Probe Nanotechnology Applications
MEMS for RF, Micro Optics and Scanning Probe Nanotechnology Applications Part I: RF Applications Introductions and Motivations What are RF MEMS? Example Devices RFIC RFIC consists of Active components
More informationKeysight 9500 AFM. Data Sheet
Keysight 9500 AFM Data Sheet System Overview The Keysight Technologies, Inc. 9500 AFM seamlessly integrates revolutionary new software, a new high-bandwidth digital controller, and a state-of-the-art mechanical
More informationPark NX10. The most accurate and easiest to use Atomic Force Microscope.
The most accurate and easiest to use Atomic Force Microscope www.parkafm.com Park Systems Enabling Nanoscale Advances The premiere choice for nanotechnology research Better data Park NX10 produces data
More informationModule 5: Experimental Modal Analysis for SHM Lecture 36: Laser doppler vibrometry. The Lecture Contains: Laser Doppler Vibrometry
The Lecture Contains: Laser Doppler Vibrometry Basics of Laser Doppler Vibrometry Components of the LDV system Working with the LDV system file:///d /neha%20backup%20courses%2019-09-2011/structural_health/lecture36/36_1.html
More informationNanosurf easyscan 2 FlexAFM
Nanosurf easyscan 2 FlexAFM Your Versatile AFM System for Materials and Life Science www.nanosurf.com The new Nanosurf easyscan 2 FlexAFM scan head makes measurements in liquid as simple as measuring in
More informationattosnom I: Topography and Force Images NANOSCOPY APPLICATION NOTE M06 RELATED PRODUCTS G
APPLICATION NOTE M06 attosnom I: Topography and Force Images Scanning near-field optical microscopy is the outstanding technique to simultaneously measure the topography and the optical contrast of a sample.
More informationMicroscopic Structures
Microscopic Structures Image Analysis Metal, 3D Image (Red-Green) The microscopic methods range from dark field / bright field microscopy through polarisation- and inverse microscopy to techniques like
More informationAkiyama-Probe (A-Probe) guide
Akiyama-Probe (A-Probe) guide This guide presents: what is Akiyama-Probe, how it works, and its performance. Akiyama-Probe is a patented technology. Version: 2009-03-23 Introduction NANOSENSORS Akiyama-Probe
More informationBasic methods in imaging of micro and nano structures with atomic force microscopy (AFM)
Basic methods in imaging of micro and nano P2538000 AFM Theory The basic principle of AFM is very simple. The AFM detects the force interaction between a sample and a very tiny tip (
More informationMFP-3D-BIO. Asylum Research. The only full-capability AFM on an inverted optical microscope
MFP-3D-BIO AFM Asylum Research The only full-capability AFM on an inverted optical microscope MFP-3D-BIO AFM Full-capability AFM integrated with optical microscopy The Asylum Research MFP-3D-BIO sets the
More informationHAMEG Programmable Measuring Instruments Series 8100
HAMEG Programmable Measuring Instruments Series 8100 HAMEG Programmable Measuring Instruments Series 8100 are ideally suited for test installations in production and automated tests in laboratories. They
More informationPark NX-Hivac The world s most accurate and easy to use high vacuum AFM for failure analysis.
Park NX-Hivac The world s most accurate and easy to use high vacuum AFM for failure analysis www.parkafm.com Park NX-Hivac High vacuum scanning for failure analysis applications 4 x 07 / Cm3 Current (µa)
More informationAtomic Force Microscopy (Bruker MultiMode Nanoscope IIIA)
Atomic Force Microscopy (Bruker MultiMode Nanoscope IIIA) This operating procedure intends to provide guidance for general measurements with the AFM. For more advanced measurements or measurements with
More informationSuivie de résonance: méthodes à fréquences multiples. Romain Stomp Application Scientist, Zurich Instruments AG. ZI Applications
Suivie de résonance: méthodes à fréquences multiples Romain Stomp Application Scientist, Zurich Instruments AG Slide 1 Sommaire 1. Un peu de traitement du signal pour le SPM Détection synchrone pour le
More informationNanosurf Nanite. Automated AFM for Industry & Research.
Nanosurf Nanite Automated AFM for Industry & Research www.nanosurf.com Multiple Measurements Automated Got work? Nanosurf has the solution! The Swiss-based innovator and manufacturer of the most compact
More informationPrecision Non-Linear Materials Testers. New Precision PiezoMEMS Analyzer
Precision Non-Linear Materials Testers Since its inception in 1988, Radiant Technologies, Inc. has been dedicated to innovating characterization equipment for non-volatile memory technologies, non-linear
More informationArbStudio Arbitrary Waveform Generators. Powerful, Versatile Waveform Creation
ArbStudio Arbitrary Waveform Generators Powerful, Versatile Waveform Creation UNMATCHED WAVEFORM UNMATCHED WAVEFORM GENERATION GENERATION Key Features 125 MHz bandwidth 1 GS/s maximum sample rate Long
More informationFlex-Axiom. Nanosurf. The Most Versatile AFM System for Materials Research
Flex-Axiom The Most Versatile AFM System for Materials Research Measurement capabilities in air and liquid Versatility in applications and Compatibility with inverted microscopes High precision scanning
More informationFigure for the aim4np Report
Figure for the aim4np Report This file contains the figures to which reference is made in the text submitted to SESAM. There is one page per figure. At the beginning of the document, there is the front-page
More informationKeysight Technologies 5500 AFM Controller Upgrade. Data Sheet
Keysight Technologies 5500 AFM Controller Upgrade Data Sheet 02 Keysight 5500 AFM Controller Upgrade - Data Sheet Upgrade Overview The Keysight 5500 AFM Controller Upgrade offers a tremendously convenient
More informationLecture 20: Optical Tools for MEMS Imaging
MECH 466 Microelectromechanical Systems University of Victoria Dept. of Mechanical Engineering Lecture 20: Optical Tools for MEMS Imaging 1 Overview Optical Microscopes Video Microscopes Scanning Electron
More informationAtomic Force Microscopes
Nanoscale Surface Characterization tomic Force Microscopes www.witec.de WITec tomic Force Microscopes Nanoscale Surface Characterization The WITec tomic Force Microscope (FM) module integrated with a research-grade
More informationRebirth of Force Spectroscopy: Advanced Nanomechanical, Electrical, Optical, Thermal and Piezoresponse Studies
HybriD Mode Rebirth of Force Spectroscopy: Advanced Nanomechanical, Electrical, Optical, Thermal and Piezoresponse Studies Fast Quantitative Nanomechanical Measurements and Force Volume Simultaneous Electrostatic
More information3D Optical Motion Analysis of Micro Systems. Heinrich Steger, Polytec GmbH, Waldbronn
3D Optical Motion Analysis of Micro Systems Heinrich Steger, Polytec GmbH, Waldbronn SEMICON Europe 2012 Outline Needs and Challenges of measuring Micro Structure and MEMS Tools and Applications for optical
More informationArbStudio Arbitrary Waveform Generators
ArbStudio Arbitrary Waveform Generators Key Features Outstanding performance with 16-bit, 1 GS/s sample rate and 2 Mpts/Ch 2 and 4 channel models Digital pattern generator PWM mode Sweep and burst modes
More informationPrepare Sample 3.1. Place Sample in Stage. Replace Probe (optional) Align Laser 3.2. Probe Approach 3.3. Optimize Feedback 3.4. Scan Sample 3.
CHAPTER 3 Measuring AFM Images Learning to operate an AFM well enough to get an image usually takes a few hours of instruction and practice. It takes 5 to 10 minutes to measure an image if the sample is
More informationLarge Signal Displacement Measurement with an Asylum SA Atomic Force Microscope Rev B
Radiant Technologies, Inc. 2835D Pan American Freeway NE Albuquerque, NM 87107 Tel: 505-842-8007 Fax: 505-842-0366 e-mail: radiant@ferrodevices.com www.ferrodevices.com Large Signal Displacement Measurement
More informationPFM Experiments with High Voltage DC/AC Bias
PFM Experiments with High Voltage DC/AC Bias Support Note Shijie Wu and John Alexander Agilent Technologies Introduction Piezoelectric force microscopy (PFM) has found major applications in the study of
More informationHAMEG Modular System Series 8000
HAMEG Modular System Series 8000 In many years of practical application the HAMEG Modular System Series 8000 has proven its value to the customer. The advantages of this Modular System have been demonstrated
More informationManufacturing Metrology Team
The Team has a range of state-of-the-art equipment for the measurement of surface texture and form. We are happy to discuss potential measurement issues and collaborative research Manufacturing Metrology
More informationOPTICS IN MOTION. Introduction: Competing Technologies: 1 of 6 3/18/2012 6:27 PM.
1 of 6 3/18/2012 6:27 PM OPTICS IN MOTION STANDARD AND CUSTOM FAST STEERING MIRRORS Home Products Contact Tutorial Navigate Our Site 1) Laser Beam Stabilization to design and build a custom 3.5 x 5 inch,
More informationNanoscale Material Characterization with Differential Interferometric Atomic Force Microscopy
Nanoscale Material Characterization with Differential Interferometric Atomic Force Microscopy F. Sarioglu, M. Liu, K. Vijayraghavan, A. Gellineau, O. Solgaard E. L. Ginzton Laboratory University Tip-sample
More informationAgilent Technologies Gli analizzatori di reti della serie-x
Agilent Technologies Gli analizzatori di reti della serie-x Luigi Fratini 1 Introducing the PNA-X Performance Network Analyzer For Active Device Test 500 GHz & beyond! 325 GHz 110 GHz 67 GHz 50 GHz 43.5
More informationPICO MASTER 200. UV direct laser writer for maskless lithography
PICO MASTER 200 UV direct laser writer for maskless lithography 4PICO B.V. Jan Tinbergenstraat 4b 5491 DC Sint-Oedenrode The Netherlands Tel: +31 413 490708 WWW.4PICO.NL 1. Introduction The PicoMaster
More informationLOW TEMPERATURE STM/AFM
* CreaTec STM of Au(111) using a CO-terminated tip, 20mV bias, 0.6nA* LOW TEMPERATURE STM/AFM High end atomic imaging, spectroscopy and manipulation Designed and manufactured in Germany by CreaTec Fischer
More informationattocube systems Probe Stations for Extreme Environments CRYOGENIC PROBE STATION fundamentals principles of cryogenic probe stations
PAGE 88 & 2008 2007 PRODUCT CATALOG CRYOGENIC PROBE STATION fundamentals...................... 90 principles of cryogenic probe stations attocps I.......................... 92 ultra stable cryogenic probe
More informationNanoFocus Inc. Next Generation Scanning Probe Technology. Tel : Fax:
NanoFocus Inc. Next Generation Scanning Probe Technology www.nanofocus.kr Tel : 82-2-864-3955 Fax: 82-2-864-3956 Albatross SPM is Multi functional research grade system Flexure scanner and closed-loop
More informationArbitrary/Function Generator AFG1000 Series Datasheet
Arbitrary/Function Generator AFG1000 Series Datasheet 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com Compatible with TekSmartLab
More informationBruker Dimension Icon AFM Quick User s Guide
Bruker Dimension Icon AFM Quick User s Guide August 8 2014 GLA Contacts Jingjing Jiang (jjiang2@caltech.edu 626-616-6357) Xinghao Zhou (xzzhou@caltech.edu 626-375-0855) Bruker Tech Support (AFMSupport@bruker-nano.com
More informationFirst results of a high performance optically-pumped cesium beam clock
First results of a high performance optically-pumped cesium beam clock Berthoud Patrick, Chief Scientist Time & Frequency Workshop on Synchronization and Timing Systems, WSTS 2016, San Jose CA, USA, June
More informationNOVA S12. Compact and versatile high performance camera system. 1-Megapixel CMOS Image Sensor: 1024 x 1024 pixels at 12,800fps
NOVA S12 1-Megapixel CMOS Image Sensor: 1024 x 1024 pixels at 12,800fps Maximum Frame Rate: 1,000,000fps Class Leading Light Sensitivity: ISO 12232 Ssat Standard ISO 64,000 monochrome ISO 16,000 color
More informationAbsolute distance interferometer in LaserTracer geometry
Absolute distance interferometer in LaserTracer geometry Corresponding author: Karl Meiners-Hagen Abstract 1. Introduction 1 In this paper, a combination of variable synthetic and two-wavelength interferometry
More informationHA-2600, HA Features. 12MHz, High Input Impedance Operational Amplifiers. Applications. Pinouts. Ordering Information
HA26, HA26 September 998 File Number 292.3 2MHz, High Input Impedance Operational Amplifiers HA26/26 are internally compensated bipolar operational amplifiers that feature very high input impedance (MΩ,
More informationMRI & NMR spectrometer
AMOS MRI & NMR spectrometer The AMOS Spectrometer is a highly modular and flexible unit that provides the ability to customize synchronized configurations for preclinical and clinical MR applications.
More informationTender Document. National Centre of Excellence in Analytical Chemistry, ISSUED BY: University of Sindh, Jamshoro.
National Centre of Excellence in Analytical Chemistry, University of Sindh, Jamshoro. Tender Document Potentiostat ISSUED TO M/s. ------------------------------------------------ Purchase Date: ----------------------------------
More informationMPI TS300-SE 300 mm Manual Probe System with ShielDEnvironment TM For accurate and reliable DC/CV, RF and mmw measurements
MPI TS300-SE 300 mm Manual Probe System with ShielDEnvironment TM For accurate and reliable DC/CV, RF and mmw measurements FEATURES / BENEFITS Universal Use Designed for wide variety of applications such
More informationA miniature all-optical photoacoustic imaging probe
A miniature all-optical photoacoustic imaging probe Edward Z. Zhang * and Paul C. Beard Department of Medical Physics and Bioengineering, University College London, Gower Street, London WC1E 6BT, UK http://www.medphys.ucl.ac.uk/research/mle/index.htm
More informationWindfreak Technologies SynthHD v1.4 Preliminary Data Sheet v0.2b
Windfreak Technologies SynthHD v1.4 Preliminary Data Sheet v0.2b $1299.00US 54 MHz 13.6 GHz Dual Channel RF Signal Generator Features Open source Labveiw GUI software control via USB Run hardware functions
More informationP a g e 1 ST985. TDR Cable Analyzer Instruction Manual. Analog Arts Inc.
P a g e 1 ST985 TDR Cable Analyzer Instruction Manual Analog Arts Inc. www.analogarts.com P a g e 2 Contents Software Installation... 4 Specifications... 4 Handling Precautions... 4 Operation Instruction...
More informationLaboratory Experiment #1 Introduction to Spectral Analysis
J.B.Francis College of Engineering Mechanical Engineering Department 22-403 Laboratory Experiment #1 Introduction to Spectral Analysis Introduction The quantification of electrical energy can be accomplished
More informationNSOM (SNOM) Overview
NSOM (SNOM) Overview The limits of far field imaging In the early 1870s, Ernst Abbe formulated a rigorous criterion for being able to resolve two objects in a light microscope: d > ë / (2sinè) where d
More informationSENSOR+TEST Conference SENSOR 2009 Proceedings II
B8.4 Optical 3D Measurement of Micro Structures Ettemeyer, Andreas; Marxer, Michael; Keferstein, Claus NTB Interstaatliche Hochschule für Technik Buchs Werdenbergstr. 4, 8471 Buchs, Switzerland Introduction
More information; A=4π(2m) 1/2 /h. exp (Fowler Nordheim Eq.) 2 const
Scanning Tunneling Microscopy (STM) Brief background: In 1981, G. Binnig, H. Rohrer, Ch. Gerber and J. Weibel observed vacuum tunneling of electrons between a sharp tip and a platinum surface. The tunnel
More informationFlexDDS-NG DUAL. Dual-Channel 400 MHz Agile Waveform Generator
FlexDDS-NG DUAL Dual-Channel 400 MHz Agile Waveform Generator Excellent signal quality Rapid parameter changes Phase-continuous sweeps High speed analog modulation Wieserlabs UG www.wieserlabs.com FlexDDS-NG
More informationLateral Force: F L = k L * x
Scanning Force Microscopy (SFM): Conventional SFM Application: Topography measurements Force: F N = k N * k N Ppring constant: Spring deflection: Pieo Scanner Interaction or force dampening field Contact
More informationFlex-Bio. Nanosurf. Versatile Research AFM System for Life Science
Flex-Bio Versatile Research AFM System for Life Science Measurement capabilities in air and in liquid Versatility in applications and Compatibility with inverted microscopes High precision scanning and
More informationCharacterisation of the Montana Instruments Cryostation C2 for low temperature Magneto-Optical Kerr Effect measurements using the NanoMOKE 3
Technical Report TR16711rev3 Characterisation of the Montana Instruments Cryostation C2 for low temperature Magneto-Optical Kerr Effect measurements using the NanoMOKE 3 EXECUTIVE SUMMARY This technical
More informationOptical cesium beam clock for eprtc telecom applications
Optical cesium beam clock for eprtc telecom applications Michaud Alain, Director R&D and PLM Time & Frequency, Oscilloquartz Dr. Patrick Berthoud, Chief Scientist Time & Frequency, Oscilloquartz Workshop
More informationA scanning tunneling microscopy based potentiometry technique and its application to the local sensing of the spin Hall effect
A scanning tunneling microscopy based potentiometry technique and its application to the local sensing of the spin Hall effect Ting Xie 1, a), Michael Dreyer 2, David Bowen 3, Dan Hinkel 3, R. E. Butera
More informationDepartment of Physics
UNIVERSITY OF CALICUT Calicut University PO Thenjipalam, Malappuram Dist Kerala 673 635, India Department of Physics No. DP/46/FIST-II/2011 05/11/2011 TENDER NOTICE Sealed competitive tenders are invited
More informationImaging and Spectroscopy Applications Guide
SPM Applications Guide USER GUIDE 3 Imaging and Spectroscopy Applications Guide User Guide including beta (complete, reviewed) chapters, including draft (nearly complete, not reviewed) chapters, Versionα,
More informationLarge Signal Displacement Measurement with an MTI Photonic Sensor Rev B
Radiant Technologies, Inc. 2835D Pan American Freeway NE Albuquerque, NM 8717 Tel: 55-842-87 Fax: 55-842-366 e-mail: radiant@ferrodevices.com www.ferrodevices.com Large Signal Displacement Measurement
More informationInvestigate the characteristics of PIN Photodiodes and understand the usage of the Lightwave Analyzer component.
PIN Photodiode 1 OBJECTIVE Investigate the characteristics of PIN Photodiodes and understand the usage of the Lightwave Analyzer component. 2 PRE-LAB In a similar way photons can be generated in a semiconductor,
More informationGSM Transmitter Modulation Quality Measurement Option
Performs all required measurements for GSM transmitters Outputs multiple time mask parameters for process control analysis Obtains frequency error, rms phase error, and peak phase error with one command
More informationqueensgate a brand of Elektron Technology
NanoSensors NX/NZ NanoSensor The NanoSensor is a non-contact position measuring system based on the principle of capacitance micrometry. Two sensor plates, a Target and a Probe, form a parallel plate capacitor.
More informationOptical Monitoring System Enables Greater Accuracy in Thin-Film Coatings. Line Scan Cameras What Do They Do?
November 2017 Optical Monitoring System Enables Greater Accuracy in Thin-Film Coatings Line Scan Cameras What Do They Do? Improved Surface Characterization with AFM Imaging Supplement to Tech Briefs CONTENTS
More informationTF TF Analyzer 2000 Measurement System
TF Analyzer 2000 Measurement System The TF Analyzer 2000 is the most sophisticated analyzer of electroceramic material and devices. The test equipment is based on a modular idea, where four different probe
More informationInspector Data Sheet. EM-FI Transient Probe. High speed pulsed EM fault injection probe for localized glitches. Riscure EM-FI Transient Probe 1/8
Inspector Data Sheet EM-FI Transient Probe High speed pulsed EM fault injection probe for localized glitches. Riscure EM-FI Transient Probe 1/8 Introduction With increasingly challenging chip packages
More informationNON-AMPLIFIED PHOTODETECTOR USER S GUIDE
NON-AMPLIFIED PHOTODETECTOR USER S GUIDE Thank you for purchasing your Non-amplified Photodetector. This user s guide will help answer any questions you may have regarding the safe use and optimal operation
More informationIonscope SICM. About Ionscope. Scanning Ion Conductance Microscopy. Ionscope A brand of Openiolabs Limited
SICM About is a brand of openiolabs Ltd, headquartered in Cambridge UK, is the world-leader in (SICM), a rapidly emerging Scanning Probe Microscopy (SPM) technique which allows nanoscale topographical
More informationMSO Supplied with a full SDK including example programs Software compatible with Windows XP, Windows Vista and Windows 7 Free Technical Support
PicoScope 2205 MSO USB-POWERED MIXED SIGNAL OSCILLOSCOPE Think logically... 25 MHz analog bandwidth 100 MHz max. digital input frequency 200 MS/s mixed signal sampling Advanced digital triggers SDK and
More informationUsing Frequency Diversity to Improve Measurement Speed Roger Dygert MI Technologies, 1125 Satellite Blvd., Suite 100 Suwanee, GA 30024
Using Frequency Diversity to Improve Measurement Speed Roger Dygert MI Technologies, 1125 Satellite Blvd., Suite 1 Suwanee, GA 324 ABSTRACT Conventional antenna measurement systems use a multiplexer or
More informationVibration Fundamentals Training System
Vibration Fundamentals Training System Hands-On Turnkey System for Teaching Vibration Fundamentals An Ideal Tool for Optimizing Your Vibration Class Curriculum The Vibration Fundamentals Training System
More informationNanotechnology Solutions Partner
Nanotechnology Solutions Partner Park Systems Corp. KANC 4F, Iui-Dong 6-10, Suwon, Korea 443-270 Tel. +82-31-546-6800 Fax. +82-31-546-6805 www.parkafm.co.kr Park Systems Inc. 3040 Olcott St. Santa Clara,
More informationPicoMaster 100. Unprecedented finesse in creating 3D micro structures. UV direct laser writer for maskless lithography
UV direct laser writer for maskless lithography Unprecedented finesse in creating 3D micro structures Highest resolution in the market utilizing a 405 nm diode laser Structures as small as 300 nm 375 nm
More information2520 Pulsed Laser Diode Test System
Complete pulse test of laser diode bars and chips with dual photocurrent measurement channels 0 Pulsed Laser Diode Test System Simplifies laser diode L-I-V testing prior to packaging or active temperature
More informationIsolated, Frequency Input 5B45 / 5B46 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM
Isolated, Frequency Input 5B45 / 5B46 FEATURES Isolated Frequency Input. Amplifies, Protects, Filters, and Isolates Analog Input. Generates an output of 0 to +5V proportional to input frequency. Model
More informationMULTI-CHANNEL ULTRASONIC FLAW DETECTOR ОКО-22М-UT THE BEST INDUSTRIAL OEM SOLUTION FOR IN-LINE AND IN-SERVICE SYSTEMS
MULTI-CHANNEL ULTRASONIC FLAW DETECTOR ОКО-22М-UT THE BEST INDUSTRIAL OEM SOLUTION FOR IN-LINE AND IN-SERVICE SYSTEMS www.ndt.com.ua PURPOSE OKO-22M-UT PRO WIRELESS INTERFACE WI-FI The ОКО-22М-UT ultrasonic
More informationBitScope Micro - a mixed signal test & measurement system for Raspberry Pi
BitScope Micro - a mixed signal test & measurement system for Raspberry Pi BS BS05U The BS05U is a fully featured mixed signal test & measurement system. A mixed signal scope in a probe! 20 MHz Bandwidth.
More information