Threshold contrast visibility of micro calcifications in digital mammography

Size: px
Start display at page:

Download "Threshold contrast visibility of micro calcifications in digital mammography"

Transcription

1 Threshold contrast visibility of micro calcifications in digital mammography Ann-Katherine Carton*, Hilde Bosmans*, Dirk Vandenbroucke, Chantal Van Ongeval*, Geert Souverijns*, Frank Rogge*, Guy Marchal* * Dept. Of Radiology, University Hospitals of KUL, Herestraat 49, Leuven 3000, Belgium Agfa-Gevaert, Septestraat 27, 2640 Mortsel, Belgium ABSTRACT The purpose of this study is to describe a method that allows the calculation of a contrast-detail curve for a particular system configuration using simulated micro calcifications into clinical mammograms. We made use of simulated templates of micro calcifications and adjusted their x-ray transmission coefficients and resolution to the properties of the mammographic system under consideration (4). We expressed the thickness of the simulated micro calcifications in terms of Al equivalence. In a first step we validated that the thickness of very small Al particles with well known size and thickness can be calculated from their x-ray transmission characteristics at a particular X-ray beam energy. Then, micro calcifications with equivalent diameters in the plane of the detector ranging from 300 to 800 µm and thicknesses, expressed in Al equivalent, covering 77 to 800 µm were simulated into the raw data of real clinical images. The procedure was tested on 2 system configurations: the GE Senographe 2000 D and the Se based Agfa Embrace DM1000 system. We adapted the X-ray transmissions and spatial characteristics of the simulated micro calcifications such that the same physical micro calcification could be simulated into images with the specific exposure parameters (Senographe 2000D: 28 kvp-rh/rh, Embrace DM1000: 28 kvp-mo/rh), compressed breast thickness (42+/-5mm) and detector under consideration. After processing and printing, 3 observers scored the visibility of the micro calcifications. We derived contrast-detail curves. This psychophysical method allows to summarize the performance of a digital mammography detector including processing and visualization. Keywords: simulation, micro calcifications, phantom, digital mammography, observer study 1. INTRODUCTION Characterization of digital mammography systems is often performed by means of contrast detail curves using a phantom with inserts of different sizes and thicknesses (1) (2). As the success of digital mammography also depends on the optimal selection of image processing protocols for display and printing we aimed in this study for a threshold contrast-detail analysis with closer link to clinical practice. To do so, we simulated micro calcifications with different sizes and thicknessess into clinical images and we scored their visibility. The calculation of the size of the micro calcification that corresponds with a self-created template, required a validation phase first: we proved that the x-ray transmission factors of small well known Al particles can be used to estimate their true thickness. A similar procedure was used to deduce size parameters for the templates of the micro calcifications. Templates of micro calcifications were created for 2 different detector systems. We started from templates in a given detector. These were then corrected for differences in x-ray transmission coefficients (as other beam qualities were used) and spatial characteristics. These templates were then inserted in the raw data of clinical mammograms and these composed images were then processed and viewed as routine mammograms. Then the visibility of the micro calcifications was compared for the 2 digital mammography systems. SPIE USE, V (p.1 of 10) / Color: No / Format: A4/ AF: A4 / Date: :41:11

2 2. MATERIAL AND METHODS 2.1. Validation of Al-equivalent thickness Very small rectangular-like Al particles with a well known size (200 µm to 3600 µm) and nominal thickness (100 and 500 µm) were imaged with a Senographe 2000D detector (GE Medical Systems, Milwaukee). The Al particles were exposed on top of 10 mm PMMA at 27 kvp, Mo/Mo (half value layer (HVL) with compression plate=0.308 mm Al). Magnification factor 2 and small focus were used. Measurements were performed on the images with signal intensities proportional to the x-ray exposure (the raw data). The image segments including Al particles were extracted. The signal intensities in these segments were normalized to the average signal intensities of the background, i.e. the region without micro calcification. This made the pixel values adjacent to the Al particles close to one. The pixel values at the location of the Al particles represent the fraction of transmitted x-rays as detected by the detector. The measured x-ray transmissions I 1 were transformed to the x-ray transmissions I 2 of the Al particles as would be obtained in an ideally sharp detector (i.e. Modulation Transfer Function (MTF) = 1 at all frequencies) using: 1 2DFFT [] ([] 1 I ) 1 1 I = 2DFFT 2 MTF 2GE 2 X where [1]-I 2 and [1]-I 1 are the image segments with the signal profiles pixel by pixel. The function 2DFFT is the 2- dimensional discrete Fourier transform, DFFT -1 is the inverse 2-dimensional discrete Fourier transform. MTF2 GE2x is the 2-dimensional MTF of the GE detector in magnification mode. The MTF2 GE2x was computed from the horizontal and vertical MTFs derived with an edge phantom (3). Our edge was exposed under similar conditions as those used for the acquisition of the Al particles (on top of 10mm PMMA at 27kVp, Mo/Mo, small focus, magnification = 2x). The calculated 1 dimensional MTFs were then modelled by Lorentzian-type functions: a 1 a MTF ( f ) = mod 2 + el i 2 f f i i b c where a, b, c are parameters with a [0,1], b>0 and c>0. Linear interpolation of the curves for the horizontal and vertical direction resulted in the MTF2 GE2x. Then, we exposed large Al sheets with thickness from 0.1 mm to 0.7mm using the same Senographe 2000D, identical geometry and exposure parameters as those used for the small Al particles. We calculated their x-ray transmissions. We made a log fit through the nominal thickness of the Al-sheets along the measured x-ray transmissions Al-thickness = ln(x-ray transmission) We used the above relation to compute the thickness of the small Al particles from their x-ray transmissions as calculated for the ideally sharp detector. The calculated thickness of the small Al particles were found to be very similar to the real Al thickness. Figure1 illustrates this match. The minimum and maximum calculated values had deviations of less than 11% of the nominal value for the 100µm thick particles and less than 8.5% of the nominal value for the 500µm thick particles. Fig. 1: Calculated thickness of Al particles with nominal thickness of 100 µm (left) and 500 µm (right). The horizontal axes represent the average diagonal size of the small Al particles as measured under a stereomicroscope. The solid lines show the nominal Al thickness SPIE USE, V (p.2 of 10) / Color: No / Format: A4/ AF: A4 / Date: :41:11

3 2.2. Description of imaging systems The procedure was tested on 2 system configurations: the GE Senographe 2000D (Milwaukee, USA) and the Agfa Embrace DM1000 (Mortsel, Belgium). Figure 2 illustrates the resolution and noise of the two detectors. Both detectors have an almost radial symmetric behavior. Therefore, we only plotted their vertical Normalized NPS (NNPS) and MTF. Both NNPS and MTF of these systems were measured under conditions that are as close as possible to those used in clinical practice for the acquisition of a standard breast. For the GE Senographe 2000D the NNPS was calculated from a flat field image of 45mm PMMA obtained at 28kVp, Rh/Rh under the Automatic Exposure Parameter (AOP) mode. The STD setting was chosen. For the Embrace DM1000, 45mm PMMA was exposed using a manual look-up table for the exposure as a function of breast thickness and composition. Twenty-eight kvp, Mo/Rh and 55.2 mas were used. The MTF was computed from an exposure of an edge sandwiched between 40mm PMMA. Table 1 summarizes the properties of the two detectors. With these settings, we calculated Mean Glandular doses of 1.24 mgy for the Senographe 2000D and 1.40mGy for the Embrace DM1000. Notwithstanding this 11.4% higher MGD for the Embrace DM1000, its NNPS performance is only better up to 1.5 lp/mm. At 5 lp/mm its NNPS is 2.3 x higher than the NNPS of the Senographe 2000D. This may influence detectability at the higher frequency levels. Fig. 2: The normalized NPS (left) and MTF (right) of the GE Senographe 2000D (gray ) and the Agfa Embrace DM1000 (black). Table 1: Properties of the Senographe 2000D and the Embrace DM1000 Senographe 2000D Embrace DM1000 Vendor GE Agfa Detector Type Indirect Direct Manufactor GE Lorad Detector Material a-si Se Digital Matrix 1914x x3328 Pixel Size [µm] Image Depth [bits] Acquisition of mammograms Thirty-five mammograms of patients with a compressed breast thickness of 4.2 +/- 0.5 cm were acquired with the GE Senographe 2000D. Thirty-three mammograms were acquired with the Agfa Embrace DM1000 in the same thickness class. The mammograms had been exposed with the routine clinically used parameters (i.e. 28 kvp and Rh/Rh for the SPIE USE, V (p.3 of 10) / Color: No / Format: A4/ AF: A4 / Date: :41:11

4 GE Senographe 2000D 28 kvp and Mo/Rh for the Agfa Embrace DM1000). The mammograms did not show any pathology Simulation of micro calcifications All micro calcifications were simulated starting from the x-ray transmissions of templates of real micro calcifications exposed at 27kVp, Mo/Mo on top of 40mm PMMA with a prototype of a CR plate of Agfa and using magnification (4). An overview of the different steps is shown in Figure 3. First, we computed the x-ray transmissions of the templates at 27kVp, Mo/Mo on top of 40mm PMMA as would be obtained for an ideally sharp digital detector. These templates were then expressed in their Al-equivalent thickness for the ideally sharp detector. The x-ray transmission properties of these (ideal) templates were then computed for the x-ray beam energy and compressed breast thickness under consideration in this study (4). We calculated the x-ray transmissions of the templates as if they had been exposed with an ideally sharp detector for 28 kvp, Rh/Rh (GE Senographe 2000D) and 28kVp Mo/Rh (Agfa Embrace DM1000) on top of 40 mm PMMA. To do so, radiographs of large Al sheets with different thicknessess on top of 40 mm at 28 kvp, Rh/Rh and 28kVp, Mo/Rh were acquired. The average x-ray transmissions of the different Al thicknesses were calculated from the raw images. Exponential fits were applied through the data points (Figure 4).. It shows the x-ray transmission as function of Al-thickness for an ideally sharp detector, as relatively large sheets of Al had been exposed. These relationships were used to adjust the templates of the ideally sharp digital detector to the exposure settings. Corrections for the differences in the spatial resolution were made using the method described in (4). Our templates of the ideally sharp detector had a pixel size of 50 µm (4). Therefore, the templates of the Embrace DM1000 (pixel size = 70µm) and the GE Senographe 2000D (pixel size = 100µm) had to be adapted for the difference in pixel size. The templates for both FFDM were finally corrected based on their pre-sampled MTFs. Figure 5 shows the median, 25- percentile and 75-percentile of the minimum x-ray transmissions of the final templates (i.e. after correction for beam energy and resolution characteristics of the 2 systems under consideration). x-ray Transmission of Template 27 kvp Mo/Mo 40 mm PMMA Prototype CR plate of Agfa x-ray Transmission of Template 27 kvp Mo/Mo 40 mm PMMA Ideally Sharp Digital Detector Al-equivalent of Template 27 kvp Mo/Mo 40 mm PMMA Ideally Sharp Digital Detector x-ray Transmission of Template X-ray Beam Energy Ideally Sharp Digital Detector x-ray Transmission of Template x-ray Beam Energy Digital Detector 2 Fig. 3: Overview of the different steps necessary to simulate a particular micro calcification. The procedure starts from templates acquired at 27 kvp, Mo/Mo on top of 40 mm PMMA with a prototype CR plate of Agfa (4). From these templates, new templates are calculated for a particular x-ray beam energy (i.e. 28kVp, Rh/Rh and 28kVp, MoRh on top of 40mm PMMA in this study) with a particular digital detector 2 (i.e. Senographe 2000D & Embrace DM1000 in this study). SPIE USE, V (p.4 of 10) / Color: No / Format: A4/ AF: A4 / Date: :41:11

5 Fig. 4: x-ray transmission as function of the Al-equivalent thickness of the large Al-sheets exposed on top of 40 mm PMMA at 27 kvp and Mo/Mo using the prototype CR plate of Agfa, at 28 kvp and Mo/Rh using the Agfa Embrace DM1000, at 28 kvp and Rh/Rh using the Senographe 2000D flat panel detector. The size of the simulated micro calcifications in the detector plane, expressed in equivalent diameter d eq ( d 2 # pixels eq = pixelsize ), as they would be obtained with an ideally sharp detector ranged from 200 µm to 800 µm. π The peak contrast, expressed in Al equivalent thickness, as they would be obtained with an ideally sharp detector, varied from 77µm to 800µm. In total 600 micro calcifications were simulated for each FFDM. Software phantoms were then created as follows: 0 to 8 simulated micro calcifications were randomly distributed in 2cmx2cm frames. They were then embedded in the linear data of the real mammograms in the parts with a constant breast thickness (5). In each mammogram, different types of background were chosen: fatty tissue, tissue that is a nearly homogeneous mixture of fat and glandular tissue and purely glandular tissue. Slow and medium varying anatomical variations were used. Different phantom configurations were made to avoid memorization. To the best of our knowledge, the phantoms were positioned in regions without any real micro calcification. Fig. 5: Minimum x-ray transmissions of the final templates, i.e. at 28 kvp, Rh/Rh corrected for spatial resolution of the Senographe 2000D (left) and at 28 kvp, Mo/Rh corrected for the spatial resolution of the Embrace DM1000 (right) for the different equivalent diameter groups as function of their peak Al-equivalent thickness that would be obtained in an ideally sharp digital detector. Error bars are 25-percentile and 75-percentile minimum x-ray transmissions. SPIE USE, V (p.5 of 10) / Color: No / Format: A4/ AF: A4 / Date: :41:11

6 The composed raw images were compressed, processed and printed as routinely obtained images. The Embrace DM1000 images were processed with MUSICA (6). The dedicated GE processing was used for the Senographe 2000D images. The Dryview 8610 (Eastman, Kodak) high resolution laser printer was used to print the GE, Senographe 2000D images. The Agfa Embrace DM1000 mammograms were printed with the high resolution Drystar 4500M (Agfa, Mortsel, Belgium) Observer experiment Three radiologists with experience in digital mammography viewed the full set of images. It was explained to the observers that each phantom contained between 0 to 8 micro calcifications. The observers were asked to indicate the locations of the micro calcifications and rate their confidence about their presence using a discrete 4-point scale (with 4=very likely to be a micro calcification, 1=probably not a micro calcification). There were no constraints on reading time and reading distance. The observers were encouraged to use a magnifying glass. Prior to the study, the observers first viewed a training set of 10 phantoms and got feed-back after rating these phantoms. A True Positive (TP) result was assumed when a detected micro calcification was within the contours of a simulated micro calcification. The False Negatives (FN) were given the score Data analysis The results of the observer experiments were analyzed using the AFROC methodology (7). The analyses were performed using ROCKIT (8). We calculated the areas under the AFROC curves A l which give a scalar summary of detectability. The confidence intervals on the area parameters were also computed. The analyses were performed for both systems, for all micro calcification sizes and for each individual observer. For each observer, the fractions of detected micro calcifications for each Al-equivalent thickness group were plotted along each equivalent diameter group. To do so, we used a binary score to calculate the fraction of detected micro calcifications for each of the 20 size groups. The scores less than or equal to 1 were set equally to 0. The scores 2, 3 and 4 were set equal to 1. The fraction of the binary score 1 was calculated for each of the 20 size groups. As AFROC analysis only takes account of the highest rated FP event in each phantom, and all other FP responses on that phantoms are neglected, we plotted the number of FP along with their scores (i.e. 1, 2, 3, 4) for each observer. 3. RESULTS The results of the AFROC can be found in table 2. No significant results are seen between the A l of the Senographe 2000D and the Embrace DM1000. Figure 6 summarizes for both FFDM the detected fraction of simulated micro calcifications for each Al-equivalent thickness along their equivalent diameter. For the micro calcifications with equivalent diameter >500µm and Alequivalent thickness of >400µm both detectors perform very similarly. For the micro calcifications with an equivalent diameter >500µm, and Al-equivalent thickness group > µm, the average detected fraction is overall larger with the Embrace DM1000: the detected fraction of micro calcifications is on average more than 50%. The Senographe 2000D detector wins for the > µm equivalent diameter group with Al-equivalent thickness >400µm. The detected fraction of the micro calcifications in the Senographe 2000D system is poor compared to the Embrace DM1000 for micro calcifications with an equivalent diameter > µm, independently of their Alequivalent thickness. Figure 7 shows the scores assigned to the FP. Each observer assigned less FP to the Senographe 2000D detector than to the Embrace DM1000: r1 assigned 3.03x less FP, r2 assigned 1.96x less FP and r3 assigned 1.14x less FP. SPIE USE, V (p.6 of 10) / Color: No / Format: A4/ AF: A4 / Date: :41:11

7 Fig. 6: Fractions of detected micro calcifications as a function of the equivalent diameter for the different Al-equivalent thicknessess. r1=reader 1, r2=reader 2, r3=reader3. Left curves represent the GE Senographe 2000D, Right curves represent the Agfa Embrace DM1000 SPIE USE, V (p.7 of 10) / Color: No / Format: A4/ AF: A4 / Date: :41:11

8 Fig. 7: Number of FP along their scores for each observer, r1= reader1, r2=reader2, r3=reader3. Left curves represent the GE Senographe 2000D, Right curves represent the Agfa Embrace DM1000 Table 2: A l and corresponding 95% confidence intervals (between brackets) under the AFROC curves for the Senographe 2000D and the Embrace DM1000 Senographe 2000D Embrace DM1000 Reader [0.465,0.558] [0.470,0.552] Reader [0.508,0.681] [0.525,0.654] Reader [0.485,0.569] [0.501,0.585] 4. DISCUSSION AND CONCLUSIONS Present paper compares the detectability of micro calcifications using simulated templates in 2 systems as they are operated in clinical routine today. The templates had been carefully constructed taking into account the detected differences in x-ray transmissions and differences in spatial characteristics of the FFDM. The observer performance test that we performed is very close to the clinical task for detection of micro calcifications. The observers had to scrutinize the images and locate the specific image regions that were suspicious for containing micro calcifications. Different from clinical practice, where it is searched for clusters, was the fact that the micro calcifications were distributed in uncorrelated locations. This restriction was made because AFROC analysis requires independent observations. The two different flat panel detectors under consideration have markedly different image physical characteristics at clinical used exposures. The Embrace DM1000 performs better for resolution (Figure 2) whereas the Senographe 2000D has superior noise characteristics for > 1.5 lp/mm (Figure 2). That are not the only parameters that determine which detector is better for clinical tasks. The clinical used processing parameters and viewing may also have their influence. The effect of processing and printing are included in this study. But, it was impossible to derive their impact on the visibility of micro calcifications. Comparison of different processing or viewing conditions would be straightforward with current procedure and is under study. In this preliminary study a conflict is seen between the overall AFROC results on one hand and the detected fraction for the different micro calcification sizes and the number of FP on the other hand. Whereas the AFROC results look to SPIE USE, V (p.8 of 10) / Color: No / Format: A4/ AF: A4 / Date: :41:11

9 promise no significant difference between both mammography systems, the performance of both systems is different when splitting up for the different micro calcifications sizes and when taking into account all indicated FP. We believe that the detected fraction for each micro calcification size and the number of the number of FP reflects much more the clinical reality than the AFROC results. We tried to perform AFROC analyses for the different micro calcification sizes but a too small statistical power for the analyses has been observed. The results of our observer study suggest that, despite the higher noise level and lower x-ray transmission of the micro calcifications, the Embrace DM1000 allows overall for greater detectability of micro calcifications than the Senographe 2000D. The higher resolution (pixel size + MTF) of the Embrace DM1000 very probably explains this performance difference for the greater part. The overall larger number of FP for the Embrace DM1000 may be due to the higher noise content of the Embrace DM1000. This indicates that readers were not sure about the true nature of the indicated artifacts, and would hesitate to interpret them as micro calcification in real practice. This preliminary study shows that there is a more or less constant detectability level for the micro calcifications of the two smallest equivalent diameter groups (> µm and > µm) in the Embrace DM1000 images. In the images of the Senographe 2000D, an increased detectability is observed for the micro calcifications of the > µm equivalent diameter group compared to these of the > µm equivalent diameter group (except for the smallest Alequivalent thickness group). The larger pixel size of the Senographe 2000D seems to dissolve the smallest micro calcifications in the background. This result could have been predicted by a simple viewing of the templates. Again, the clinically used processing parameters, printer and film for the two FFDM could also partly have determined the results. Further research of the separate parts of the imaging system, rather than a global analysis is needed here. Present study quantifies the sizes of detected micro calcifications for 2 FFDM system using physical methods, clinical mammograms and human reading. The results may be important and straightforward information for radiologists in charge of assessing the role of digital mammography in their (screening) practice or of comparing FFDM systems. It remains to be seen in larger studies whether clinical practice will confirm the present observations regarding the detectability of lesions. It goes without saying that also more different systems, such as computed radiography plates or slot scan systems, could be evaluated in the same way. Finally, noise, contrast and spatial resolution are the very fundamental properties with an influence on image quality. The digital nature of digital images adds post processing and viewing to this list. In this study, 2 detectors were involved with different properties and studied as they are used in the clinical routine. The detector has certainly a big influence on the final result. This preliminary study suggests that the influence of resolution on the detectability of small lesions like micro calcifications can compensate for a higher noise level. It must be recognized however that the chosen exposure settings determine image contrast and the dose level influences the noise in the images. Post processing can further enlarge or decrease the sharpness or the global appearance of the lesions. Further studies will unravel the complete link between the detectability of particular lesions and the noise, contrast and resolution properties of an imaging system and, in this way, guide the user towards optimal exposure settings. Present procedure may add complementary information to more established approaches. 5. ACKNOWLEDGMENT We would like to thank A. Chèvrement of GE for all practical help with the printing of the Senographe 2000D images. Octavian Dragusin is greatly acknowledged for his help with the acquisition of the results. This work was partly sponsored by DIMONDIII, an EC funded research project. REFERENCES 1. R.E. Van Engen, W.J.H. Veldkamp, L.J. Oostveen, M.A.O. Thijssen, N. Karssemeijer, Image Quality of a Dual Side Reading Fuji Computed Mammography System Compared to the GE FFDM System and a Film/Screen Mammography System, IWDM th International Workshop on Digital Mammography, p X.J. Rong, C.C. Shaw, D.A. Johnston, M.R. Lemacks, X. Liu, G.J. Whitman, M.J. Dryden, T.W. Stephens, S.K. Thompson, K.T. Krugh and C-J. Lai, Microcalcification detectability for four mammographic detectors: Flatpanel, CCD, CR and screen/film, Med.Phys., 29 (9), September 2002, SPIE USE, V (p.9 of 10) / Color: No / Format: A4/ AF: A4 / Date: :41:11

10 3. E. Samei, M. J. Flynn, and D. A. Reimann, A method for measuring the presampled MTF of digital radiographic systems using an edge test device, Med. Phys. 25, 102 (1998) 4. A.-K. Carton, H. Bosmans, C. Vanongeval, G. Souverijns, F. Rogge, A. Van Steen, G. Marchal, Development and validation of a simulation procedure to study the visibility of micro calcifications in digital mammograms, Med. Phys., 2003, August 2003, A.E. Burgess, Mammographic structure: Data preparation and spatial statistics analysis, Medical Imaging 1999: Image Processing, K. Hanson (ed.), Vol.3661, , SPIE, Bellingham WA P. Vuylsteke, P. Dewaele, E. Schoeters, "Optimizing Computed Radiography Imaging Performance", in "The Expanding Role of Medical Physics in Diagnostic Imaging", G.D. Frey, P. Sprawls. Eds., Proceedings of the 1997 AAPM Summer School, pp , D. P. Chakraborty, The FROC, AFROC and DROC variants of the ROC Analysis, Chapter 16 in Handbook of medical imaging. Vol. 1: Physics and Psychophysics. SPIE Editors: J. Beutel, H.L. Kundel, R.L. Van Metter 8. SPIE USE, V (p.10 of 10) / Color: No / Format: A4/ AF: A4 / Date: :41:11

Tailoring automatic exposure control toward constant detectability in digital mammography

Tailoring automatic exposure control toward constant detectability in digital mammography Tailoring automatic exposure control toward constant detectability in digital mammography Elena Salvagnini a) Department of Imaging and Pathology, Medical Physics and Quality Assessment, KUL, Herestraat

More information

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis

More information

Imaging Technique Optimization of Tungsten Anode FFDM System

Imaging Technique Optimization of Tungsten Anode FFDM System Imaging Technique Optimization of Tungsten Anode FFDM System Biao Chen a*, Andrew P. Smith b, Zhenxue Jing a, Elena Ingal a a Hologic, Inc. 600 Technology Drive, DE 1970 b Hologic, Inc. 35 Crosby Drive,

More information

Mammography: Physics of Imaging

Mammography: Physics of Imaging Mammography: Physics of Imaging Robert G. Gould, Sc.D. Professor and Vice Chair Department of Radiology and Biomedical Imaging University of California San Francisco, California Mammographic Imaging: Uniqueness

More information

Published text: Institute of Cancer Research Repository Please direct all s to:

Published text: Institute of Cancer Research Repository   Please direct all  s to: This is an author produced version of an article that appears in: MEDICAL PHYSICS The internet address for this paper is: https://publications.icr.ac.uk/1316/ Copyright information: http://www.aip.org/pubservs/web_posting_guidelines.html

More information

Quality Control of Full Field Digital Mammography Units

Quality Control of Full Field Digital Mammography Units Quality Control of Full Field Digital Mammography Units Melissa C. Martin, M.S., FACMP, FACR, FAAPM Melissa@TherapyPhysics.com 310-612-8127 ACMP Annual Meeting Virginia Beach, VA May 2, 2009 History of

More information

Investigation of the line-pair pattern method for evaluating mammographic focal spot performance

Investigation of the line-pair pattern method for evaluating mammographic focal spot performance Investigation of the line-pair pattern method for evaluating mammographic focal spot performance Mitchell M. Goodsitt, a) Heang-Ping Chan, and Bob Liu Department of Radiology, University of Michigan, Ann

More information

Collimation Assessment Using GAFCHROMIC XR-M2

Collimation Assessment Using GAFCHROMIC XR-M2 Collimation Assessment Using GAFCHROMIC XR-M2 I. Introduction A method of collimation assessment for GE Senographe full-field digital mammography (FFDM) systems is described that uses a self-developing

More information

Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities

Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities Melissa Martin, M.S., FAAPM, FACR, FACMP AAPM Annual Meeting - Philadelphia July 19, 2010 MO-B-204C-1 Educational Objectives

More information

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Acceptance Testing of a Digital Breast Tomosynthesis Unit Acceptance Testing of a Digital Breast Tomosynthesis Unit 2012 AAPM Spring Clinical Meeting Jessica Clements, M.S., DABR Objectives Review of technology and clinical advantages Acceptance Testing Procedures

More information

Predicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities

Predicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities Journal of Physics: Conference Series PAPER OPEN ACCESS Predicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities Recent citations - Resolution Properties of a

More information

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS JAMES A. TOMLINSON, M.S. Diagnostic Radiological Physicist American Board of Radiology Certified Medical Physics Consultants, Inc. Bio 28 yrs experience 100%

More information

Optimization of Digital Mammography Resolution Using Magnification Technique in Computed Radiography 1

Optimization of Digital Mammography Resolution Using Magnification Technique in Computed Radiography 1 Optimization of Digital Mammography Resolution Using Magnification Technique in Computed Radiography 1 Gham Hur, M.D., Yoon Joon Hwang, M.D., Soon Joo Cha, M.D., Su Young Kim, M.D., Yong Hoon Kim, M.D.

More information

Introduction. Digital Mammography QA: Comparing the Manufacturers Recommendations. What is QC and why is it important? Review & compare QC tests

Introduction. Digital Mammography QA: Comparing the Manufacturers Recommendations. What is QC and why is it important? Review & compare QC tests Slide 1 Digital Mammography QA: Comparing the Manufacturers Recommendations Eric A. Berns, Ph.D. Slide 2 Introduction What is QC and why is it important? Review & compare QC tests Key take home points

More information

Features and Weaknesses of Phantoms for CR/DR System Testing

Features and Weaknesses of Phantoms for CR/DR System Testing Physics testing of image detectors Parameters to test Features and Weaknesses of Phantoms for CR/DR System Testing Spatial resolution Contrast resolution Uniformity/geometric distortion Dose response/signal

More information

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Definition and Validation of Exposure Indices Ingrid Reiser, PhD DABR Department of Radiology University of Chicago

More information

Evaluation of a quality control phantom for digital chest radiography

Evaluation of a quality control phantom for digital chest radiography JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 2, NUMBER 2, SPRING 2001 Evaluation of a quality control phantom for digital chest radiography Eugene Mah* Department of Radiology, Medical University

More information

Display of mammograms on a CRT

Display of mammograms on a CRT Display of mammograms on a CRT Hans Roehrig, Ph.D. William J. Dallas, Ph.D. Elizabeth Krupinski, Ph.D. Jiahua Fan, M.S. University of Arizona This work was supported by 2 Grants from NIH In most radiological

More information

Digital Quality Control Phantom for X-ray Mammography Image

Digital Quality Control Phantom for X-ray Mammography Image Digital Quality Control Phantom for X-ray Mammography Image Yuichi Nagai, Atsuko Takada National: Cancer Hospital East Department of Radiology Collaborated by Nobuyuki Niwa : Kyoto Kagaku Co., Ltd. Introduction

More information

Digital Breast Tomosynthesis

Digital Breast Tomosynthesis Digital Breast Tomosynthesis OLIVE PEART MS, RT(R) (M) HTTP://WWW.OPEART.COM 2D Mammography Not 100% effective Limited by tissue superimposition Overlapping tissue can mask tumors False negative Overlapping

More information

Objective Evaluation of Radiographic Contrast- Enhancement Masks

Objective Evaluation of Radiographic Contrast- Enhancement Masks Chapter 8 Objective Evaluation of Radiographic Contrast- Enhancement Masks The development and application of radiographic contrast-enhancement masks (RCMs) in digital radiography (DR) were discussed in

More information

TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know. JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist

TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know. JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist Topics Image Uniformity and Artifacts Image Quality - Detail

More information

DRAFT Technical evaluation of Philips Microdose SI digital mammography system

DRAFT Technical evaluation of Philips Microdose SI digital mammography system DRAFT Technical evaluation of Philips Microdose SI digital mammography system NHSBSP Equipment Report 1310 August 2013 About the NHS Cancer Screening Programmes The national office of the NHS Cancer Screening

More information

Estimation of signal transfer property for wireless digital detector in different measurement schemes

Estimation of signal transfer property for wireless digital detector in different measurement schemes Estimation of signal transfer property for wireless digital detector in different measurement schemes Anatoli Vladimirov, Kalle Kepler Training Centre of Medical Physics, University of Tartu, Estonia 11

More information

radiography detector

radiography detector Clinical evaluation of a full field digital projection radiography detector Gary S. Shaber'1, Denny L. Leeb, Jeffrey Belib, Gregory Poweii1', Andrew D.A. Maidment'1 a Thomas Jefferson University Hospital,

More information

Digital Imaging Considerations Computed Radiography

Digital Imaging Considerations Computed Radiography Digital Imaging Considerations Digital Radiography Computed Radiography o Cassette based Direct or Indirect Digital Radiography o Cassetteless Computed Radiography 1 CR Image Acquisition Most like conventional

More information

Testing a wavelet based noise reduction method using computersimulated

Testing a wavelet based noise reduction method using computersimulated Testing a wavelet based noise reduction method using computersimulated mammograms Christoph Hoeschen 1, Oleg Tischenko 1, David R Dance 2, Roger A Hunt 2, Andrew DA Maidment 3, Predrag R Bakic 3 1 GSF-

More information

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation AAPM Annual Clinical Meeting Indianapolis, IN August 5, 2013 Learning Objectives Become familiar with the recommendations and

More information

Investigation of Effective DQE (edqe) parameters for a flat panel detector

Investigation of Effective DQE (edqe) parameters for a flat panel detector Investigation of Effective DQE (edqe) parameters for a flat panel detector Poster No.: C-1892 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Exhibit D. Bor 1, S. Cubukcu 1, A. Yalcin 1, O.

More information

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893 Digital Radiography D. J. Hall, Ph.D. x20893 djhall@ucsd.edu Background Common Digital Modalities Digital Chest Radiograph - 4096 x 4096 x 12 bit CT - 512 x 512 x 12 bit SPECT - 128 x 128 x 8 bit MRI -

More information

Amorphous Selenium Direct Radiography for Industrial Imaging

Amorphous Selenium Direct Radiography for Industrial Imaging DGZfP Proceedings BB 67-CD Paper 22 Computerized Tomography for Industrial Applications and Image Processing in Radiology March 15-17, 1999, Berlin, Germany Amorphous Selenium Direct Radiography for Industrial

More information

Improved Tomosynthesis Reconstruction using Super-resolution and Iterative Techniques

Improved Tomosynthesis Reconstruction using Super-resolution and Iterative Techniques Improved Tomosynthesis Reconstruction using Super-resolution and Iterative Techniques Wataru FUKUDA* Junya MORITA* and Masahiko YAMADA* Abstract Tomosynthesis is a three-dimensional imaging technology

More information

ADVANCED MEDICAL SYSTEMS PTE LTD Singapore Malaysia India Australia

ADVANCED MEDICAL SYSTEMS PTE LTD Singapore Malaysia India Australia Innovative design is combined with cutting-edge technology to yield a definitive diagnosis and never before seen ergonomics GIOTTO CLASS is the result of 25 years of experience in the research and development

More information

Differentiation of Malignant and Benign Masses on Mammograms Using Radial Local Ternary Pattern

Differentiation of Malignant and Benign Masses on Mammograms Using Radial Local Ternary Pattern Differentiation of Malignant and Benign Masses on Mammograms Using Radial Local Ternary Pattern Chisako Muramatsu 1, Min Zhang 1, Takeshi Hara 1, Tokiko Endo 2,3, and Hiroshi Fujita 1 1 Department of Intelligent

More information

Patient-Assisted Compression Impact on Image Quality and Workflow

Patient-Assisted Compression Impact on Image Quality and Workflow Patient-Assisted Compression Impact on Image Quality and Workflow Senographe Pristina In 2017, GE Healthcare s Senographe Pristina ( Pristina ) was approved by the FDA using the standard technologist-controlled

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

Observer Performance of Reduced X-Ray Images on Liquid Crystal Displays

Observer Performance of Reduced X-Ray Images on Liquid Crystal Displays Original Paper Forma, 29, S45 S51, 2014 Observer Performance of Reduced X-Ray Images on Liquid Crystal Displays Akiko Ihori 1, Chihiro Kataoka 2, Daigo Yokoyama 2, Naotoshi Fujita 3, Naruomi Yasuda 4,

More information

Mammography Solution. AMULET Innovality. The new leader in the AMULET series. Tomosynthesis, 3D mammography and biopsy are all available.

Mammography Solution. AMULET Innovality. The new leader in the AMULET series. Tomosynthesis, 3D mammography and biopsy are all available. Mammography Solution AMULET Innovality The new leader in the AMULET series. Tomosynthesis, 3D mammography and biopsy are all available. FUJIFILM supports the Pink Ribbon Campaign for early detection of

More information

SECTION I - CHAPTER 1 DIGITAL RADIOGRAPHY: AN OVERVIEW OF THE TEXT. Exam Content Specifications 8/22/2012 RADT 3463 COMPUTERIZED IMAGING

SECTION I - CHAPTER 1 DIGITAL RADIOGRAPHY: AN OVERVIEW OF THE TEXT. Exam Content Specifications 8/22/2012 RADT 3463 COMPUTERIZED IMAGING RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 1 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 1 DIGITAL RADIOGRAPHY: AN OVERVIEW OF THE TEXT RADT 3463 COMPUTERIZED IMAGING Section I: Chapter

More information

2 nd generation TOMOSYNTHESIS

2 nd generation TOMOSYNTHESIS 2 nd generation TOMOSYNTHESIS 2 nd generation DBT true innovation in breast imaging synthesis graphy Combo mode Stereotactic Biopsy Works in progress: Advanced Technology, simplicity and ergonomics Raffaello

More information

Phase Imaging Using Focused Polycapillary Optics

Phase Imaging Using Focused Polycapillary Optics Phase Imaging Using Focused Polycapillary Optics Sajid Bashir, Sajjad Tahir, Jonathan C. Petruccelli, C.A. MacDonald Dept. of Physics, University at Albany, Albany, New York Abstract Contrast in conventional

More information

30 lesions. 30 lesions. false positive fraction

30 lesions. 30 lesions. false positive fraction Solutions to the exercises. 1.1 In a patient study for a new test for multiple sclerosis (MS), thirty-two of the one hundred patients studied actually have MS. For the data given below, complete the two-by-two

More information

Digital radiography (DR) post processing techniques for pediatric radiology

Digital radiography (DR) post processing techniques for pediatric radiology Digital radiography (DR) post processing techniques for pediatric radiology St Jude Children s Research Hospital Samuel Brady, MS PhD DABR samuel.brady@stjude.org Purpose Review common issues and solutions

More information

Unit thickness. Unit area. σ = NΔX = ΔI / I 0

Unit thickness. Unit area. σ = NΔX = ΔI / I 0 Unit thickness I 0 ΔI I σ = ΔI I 0 NΔX = ΔI / I 0 NΔX Unit area Δx Average probability of reaction with atom for the incident photons at unit area with the thickness of Delta-X Atom number at unit area

More information

AUTOMATED AND QUANTITATIVE METHOD FOR QUALITY ASSURANCE OF DIGITAL RADIOGRAPHY IMAGING SYSTEMS

AUTOMATED AND QUANTITATIVE METHOD FOR QUALITY ASSURANCE OF DIGITAL RADIOGRAPHY IMAGING SYSTEMS International Workshop SMART MATERIALS, STRUCTURES & NDT in AEROSPACE Conference NDT in Canada 2011 2-4 November 2011, Montreal, Quebec, Canada AUTOMATED AND QUANTITATIVE METHOD FOR QUALITY ASSURANCE OF

More information

Abstract. ZEIGLER, GARY BOYCE, II. Direct Detection of Microcalcification Pairs in Simulated

Abstract. ZEIGLER, GARY BOYCE, II. Direct Detection of Microcalcification Pairs in Simulated Abstract ZEIGLER, GARY BOYCE, II. Direct Detection of Microcalcification Pairs in Simulated Digital Mammograms. (Under the Direction of Professor Kuruvilla Verghese.) Using the MCMIS (Monte Carlo for Mammography

More information

GE Healthcare. Senographe 2000D Full-field digital mammography system

GE Healthcare. Senographe 2000D Full-field digital mammography system GE Healthcare Senographe 2000D Full-field digital mammography system Digital has arrived. The Senographe 2000D Full-Field Digital Mammography (FFDM) system gives you a unique competitive advantage. That

More information

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over a lifetime Breast cancer screening programs rely on

More information

New spectral benefi ts, proven low dose

New spectral benefi ts, proven low dose New spectral benefi ts, proven low dose Philips MicroDose mammography SI, technical data sheet Philips MicroDose SI with single-shot spectral imaging is a fullfi eld digital mammography solution that delivers

More information

The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System

The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System Manus

More information

Digital mammography: DQE versus optimized image quality in clinical environment an on site study

Digital mammography: DQE versus optimized image quality in clinical environment an on site study Digital mammography: DQE versus optimized image quality in clinical environment an on site study Nadia Oberhofer* a, Alessandro Fracchetti a, Margareth Springeth a, Ehrenfried Moroder a a Health Service

More information

Attikon, Rimini 1, , Athens, Greece , Athens, Greece , Athens, Greece

Attikon, Rimini 1, , Athens, Greece , Athens, Greece , Athens, Greece Radiation Protection Dosimetry (2005), Vol. 117, No. 1 3, pp. 291 297 doi:10.1093/rpd/nci742 Advance Access published on February 7, 2006 QUALITY ASSURANCE (QA) PROCEDURES FOR SOFTWARE: EVALUATION OF AN

More information

10/3/2012. Study Harder

10/3/2012. Study Harder This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura Study Harder CR detection is inefficient, inferior to film screen

More information

Use of Wavelet Multiresolution Analysis to Reduce Radiation Dose in Digital Mammography

Use of Wavelet Multiresolution Analysis to Reduce Radiation Dose in Digital Mammography 2015 IEEE 28th International Symposium on Computer-Based Medical Systems Use of Wavelet Multiresolution Analysis to Reduce Radiation Dose in Digital Mammography Helder C. R. de Oliveira*, Lucas R. Borges*,

More information

10/26/2015. Study Harder

10/26/2015. Study Harder This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura Study Harder CR detection is inefficient, inferior to film screen

More information

Aquilion Precision Ultra-High Resolution CT: Quantifying diagnostic image quality

Aquilion Precision Ultra-High Resolution CT: Quantifying diagnostic image quality Aquilion Precision Ultra-High CT: Quantifying diagnostic image quality Kirsten Boedeker, PhD, DABR Senior Manager, Quantitative Image Quality Canon Medical Systems Corporation Introduction Over the last

More information

7/24/2014. Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation. Disclosures. Outline

7/24/2014. Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation. Disclosures. Outline Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation Image Quality Review I: Basics and Image Quality TH-A-16A-1 Thursday 7:30AM - 9:30AM Room: 16A J. Anthony

More information

Acquisition, Processing and Display

Acquisition, Processing and Display Acquisition, Processing and Display Terri L. Fauber, R.T. (R)(M) Department of Radiation Sciences School of Allied Health Professions Virginia Commonwealth University Topics Image Characteristics Image

More information

FFDM in the Field: Physicist's Role in the QC of Mammography Laser Printers May Carl R. Keener, Ph.D., DABMP, DABR

FFDM in the Field: Physicist's Role in the QC of Mammography Laser Printers May Carl R. Keener, Ph.D., DABMP, DABR FFDM in the Field: Physicist's Role in the QC of Mammography Laser Printers May 2010 Carl R. Keener, Ph.D., DABMP, DABR keener@marpinc.com MARP Medical & Radiation Physics, Inc. Physicist's Role in the

More information

DIGITAL RADIOGRAPHY. Digital radiography is a film-less technology used to record radiographic images.

DIGITAL RADIOGRAPHY. Digital radiography is a film-less technology used to record radiographic images. DIGITAL RADIOGRAPHY Digital radiography is a film-less technology used to record radiographic images. 1 The purpose of digital imaging is to generate images that can be used in the diagnosis and assessment

More information

KODAK DIRECTVIEW CR Mammography Feature User s Guide

KODAK DIRECTVIEW CR Mammography Feature User s Guide KODAK DIRECTVIEW CR Mammography Feature User s Guide 17 September 2010 9G3741 Version 1.0 Carestream Health, Inc. 150 Verona Street Rochester, NY 14608 CARESTREAM, DIRECTVIEW, and DRYVIEW are trademarks

More information

Comparison of computed radiography and filmõscreen combination using a contrast-detail phantom

Comparison of computed radiography and filmõscreen combination using a contrast-detail phantom JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 1, WINTER 2003 Comparison of computed radiography and filmõscreen combination using a contrast-detail phantom Z. F. Lu,* E. L. Nickoloff, J.

More information

1.1.Clinical Artefacts

1.1.Clinical Artefacts 1.1.Clinical Artefacts While the incidence of artefact on digital mammographic images 1 is typically less than with film based mammography, artefacts can be produced on digital systems. This section provides

More information

X-ray light valve (XLV): a novel detectors technology for digital mammography*

X-ray light valve (XLV): a novel detectors technology for digital mammography* X-ray light valve (XLV): a novel detectors technology for digital mammography* Sorin Marcovici, Vlad Sukhovatkin, Peter Oakham XLV Diagnostics Inc., Thunder Bay, ON P7A 7T1, Canada ABSTRACT A novel method,

More information

Applications of a Pixellated Detection System to Digital Mammography

Applications of a Pixellated Detection System to Digital Mammography Applications of a Pixellated Detection System to Digital Mammography Valeria Rosso Dipartimento di Fisica, Universita di Pisa and Sezione INFN Pisa, Italy + valeria.rosso@pi.infn.it Outline The detection

More information

Disclosures. Outline 7/31/2017. Current Implementation Status of IEC Standard : Exposure Index (EI) for Digital Radiography

Disclosures. Outline 7/31/2017. Current Implementation Status of IEC Standard : Exposure Index (EI) for Digital Radiography Current Implementation Status of IEC Standard 62494-1: Exposure Index (EI) for Digital Radiography July 31, 2017 Ryan Fisher, PhD, DABR Katie Hulme, MS, DABR None Disclosures Outline Review of IEC Standard

More information

Image Quality Artifacts in Digital Imaging

Image Quality Artifacts in Digital Imaging MAHIDOL UNIVERSITY Wisdom of the Land Image Quality Artifacts in Digital Imaging Napapong Pongnapang, Ph.D. Department of Radiological Technology Faculty of Medical Technology Mahidol University, Bangkok,

More information

Comparison of new digital mammography systems: physical characterisation and image quality evaluation

Comparison of new digital mammography systems: physical characterisation and image quality evaluation Comparison of new digital mammography systems: physical characterisation and image quality evaluation Poster No.: C-2321 Congress: ECR 2011 Type: Scientific Exhibit Authors: N. Oberhofer; Bolzano/IT DOI:

More information

8/2/2017. Radiologist Responsibilities. Radiologist Responsibilities. Medical Physicist Mammography Equipment Evaluation and Annual Survey

8/2/2017. Radiologist Responsibilities. Radiologist Responsibilities. Medical Physicist Mammography Equipment Evaluation and Annual Survey Implementation of the 2016 ACR Digital Mammography QC Manual Medical Physicist Mammography Equipment Evaluation and Annual Survey Eric A Berns, PhD, FACR Radiologist Responsibilities Radiologist Responsibilities

More information

Full-field digital mammography with grid-less acquisition and software-based scatter correction: Investigation of dose saving and image quality

Full-field digital mammography with grid-less acquisition and software-based scatter correction: Investigation of dose saving and image quality Reprint Full-field digital mammography with grid-less acquisition and software-based scatter correction: Investigation of dose saving and image quality Andreas Fieselmann, Daniel Fischer, Ghani Hilal,

More information

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 2 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 COMPUTERIZED IMAGING Section I: Chapter 2 RADT

More information

IBEX MATERIALS DETECTION TECHNOLOGY

IBEX MATERIALS DETECTION TECHNOLOGY WHITE PAPER: IBEX MATERIALS DETECTION TECHNOLOGY IBEX Innovations Ltd. Registered in England and Wales: 07208355 Address: Discovery 2, NETPark, William Armstrong Way, Sedgefield, TS21 3FH, UK Patents held

More information

Image Enhancement in Spatial Domain

Image Enhancement in Spatial Domain Image Enhancement in Spatial Domain 2 Image enhancement is a process, rather a preprocessing step, through which an original image is made suitable for a specific application. The application scenarios

More information

COMPUTER-AIDED DETECTION OF CLUSTERED CALCIFICATION USING IMAGE MORPHOLOGY

COMPUTER-AIDED DETECTION OF CLUSTERED CALCIFICATION USING IMAGE MORPHOLOGY COMPUTER-AIDED DETECTION OF CLUSTERED CALCIFICATION USING IMAGE MORPHOLOGY Ariya Namvong Department of Information and Communication Technology, Rajamangala University of Technology Isan, Nakhon Ratchasima,

More information

ADC COMPACT FULL-LEG/FULL-SPINE APPLICATION SOFTWARE USER MANUAL

ADC COMPACT FULL-LEG/FULL-SPINE APPLICATION SOFTWARE USER MANUAL ADC COMPACT FULL-LEG/FULL-SPINE APPLICATION SOFTWARE USER MANUAL TABLE OF CONTENTS Full-leg/full-spine image stitching: a new and accurate CR-based imaging software 1 The conventional approach 1 CR and

More information

History of digital imaging

History of digital imaging CR/QA RADCHEX History of digital imaging Early, crude digital detectors were developed in the 1970 s Image quality was problematic Processing time of digital images was untenable Viewing, transfer and

More information

Do you have any other questions? Please call us at (Toll Free) or , or

Do you have any other questions? Please call us at (Toll Free) or , or INSTRUCTIONS Read the appropriate course/ textbook. This is an open book test. A score of 75% or higher is needed to receive CE credit. You will have a maximum of three attempts to pass this course. Please

More information

3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development

3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development Teaching Radiographic Technique in a Digital Imaging Paradigm Objectives 1. Discuss the historical development of digital imaging. Dawn Couch Moore, M.M.Sc., RT(R) Assistant Professor and Director Emory

More information

A Computer-Based Cascaded Modeling and Experimental Approach to the Physical Characterization of a Clinical Full-Field Digital Mammography System

A Computer-Based Cascaded Modeling and Experimental Approach to the Physical Characterization of a Clinical Full-Field Digital Mammography System A Computer-Based Cascaded Modeling and Experimental Approach to the Physical Characterization of a Clinical Full-Field Digital Mammography System by Hetal Ved A Thesis submitted to the Faculty of the WORCESTER

More information

Basis of Computed Radiography & PACS

Basis of Computed Radiography & PACS Basis of Computed Radiography & PACS Slavik Tabakov Computed Radiography (CR) refers to new types of X-ray detectors (i.e. replaces the X-ray Film) The CR output media is a digital image, which can be

More information

Current technology in digital image production (CR/DR and other modalities) Jaroonroj Wongnil 25 Mar 2016

Current technology in digital image production (CR/DR and other modalities) Jaroonroj Wongnil 25 Mar 2016 Current technology in digital image production (CR/DR and other modalities) Jaroonroj Wongnil 25 Mar 2016 Current technology in digital image production (CR/DR and other modalities) 2/ Overview Digital

More information

A Study On Preprocessing A Mammogram Image Using Adaptive Median Filter

A Study On Preprocessing A Mammogram Image Using Adaptive Median Filter A Study On Preprocessing A Mammogram Image Using Adaptive Median Filter Dr.K.Meenakshi Sundaram 1, D.Sasikala 2, P.Aarthi Rani 3 Associate Professor, Department of Computer Science, Erode Arts and Science

More information

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura New Technology https://www.youtube.com/watch?v=ptkzznazb 7U COMPUTED

More information

Performance and care. all in one

Performance and care. all in one Performance and care all in one INNOVATION IS WHAT DRIVES US THINKING ABOUT THE FUTURE Preventive diagnostics remains an essential weapon in defeating breast cancer. Metaltronica s forward-thinking design

More information

Digital Mammography: From Physical Performance Evaluation To Image Quality Analysis

Digital Mammography: From Physical Performance Evaluation To Image Quality Analysis Digital Mammography: From Physical Performance Evaluation To Image Quality Analysis Anastasios C. Konstantinidis Department of Medical Physics and Bioengineering, University College London, UK. taskon25@yahoo.gr

More information

Exposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations

Exposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations Exposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations S. Baechler 1, P. Monnin 1, A. Aroua 1, J.F. Valley 1, M. Perrier, P. Trueb 3, F.R. Verdun 1 1 University

More information

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division

More information

MILADY. Product Data. Page 1 of 8

MILADY. Product Data. Page 1 of 8 Page 1 of 8 The MILADY Mammographic Unit offers the best quality-to-price ratio to our customers worldwide. The unit advanced technology together with the application of industrial production standards,

More information

Protocol for the Quality Control of the Physical and Technical Aspects of Digital Breast Tomosynthesis Systems

Protocol for the Quality Control of the Physical and Technical Aspects of Digital Breast Tomosynthesis Systems Protocol for the Quality Control of the Physical and Technical Aspects of Digital Breast Tomosynthesis Systems Draft version 0.10 February 2013 European Reference Organisation for Quality Assured Breast

More information

While digital techniques have the potential to reduce patient doses, they also have the potential to significantly increase them.

While digital techniques have the potential to reduce patient doses, they also have the potential to significantly increase them. In press 2004 1 2 Guest Editorial (F. Mettler, H. Ringertz and E. Vano) Guest Editorial (F. Mettler, H. Ringertz and E. Vano) Digital radiology An appropriate analogy that is easy for most people to understand

More information

RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II

RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II APPROVED 12/O2/2011 EFFECTIVE SPRING 2013-14 Prefix & Number RAD 150 Course Title: Radiologic Exposure Technique II & Lab Purpose of this submission: New Change/Updated

More information

Comparison between film-screen and computed radiography systems in Brazilian mammography

Comparison between film-screen and computed radiography systems in Brazilian mammography X Congreso Regional Latinoamericano IRPA de Protección y Seguridad Radiológica Radioprotección: Nuevos Desafíos para un Mundo en Evolución Buenos Aires, al 7 de abril, 05 SOCIEDAD ARGENTINA DE RADIOPROTECCIÓN

More information

Pseudocolour Image Processing in Digital Mammography

Pseudocolour Image Processing in Digital Mammography Pseudocolour Image Processing in Digital Mammography M.Y. Sanavullah 1 and R. Samson Ravindran 2 1 Principal and Research Guide, VMKV Engineering College, V.M.R.F Deemed University, Salem, TN 2 Research

More information

TITLE: Automatic Exposure Control Device for Digital Mammography

TITLE: Automatic Exposure Control Device for Digital Mammography AD Award Number: DAMD17-99-1-9429 TITLE: Automatic Exposure Control Device for Digital Mammography PRINCIPAL INVESTIGATOR: Laurie L. Fajardo, M.D. CONTRACTING ORGANIZATION: Johns Hopkins University Baltimore,

More information

ACPSEM Position Paper RECOMMENDATIONS FOR A DIGITAL MAMMOGRAPHY QUALITY ASSURANCE PROGRAM V4.0

ACPSEM Position Paper RECOMMENDATIONS FOR A DIGITAL MAMMOGRAPHY QUALITY ASSURANCE PROGRAM V4.0 Heggie et al ACPSEM Position Paper: Digital Mammography V4.0 ACPSEM Position Paper RECOMMENDATIONS FOR A DIGITAL MAMMOGRAPHY QUALITY ASSURANCE PROGRAM V4.0 JCP Heggie 1, P Barnes 2, L Cartwright 3, J Diffey

More information

Introduction of Computed Radiography in Two Mammography Services: Image Quality and Dose Analysis

Introduction of Computed Radiography in Two Mammography Services: Image Quality and Dose Analysis Introduction of Computed Radiography in Two Mammography Services: Image Quality and Dose Analysis Rosangela Requi Jakubiak* a, Humberto Remigio Gamba a, Maria Manuela Ramos a, Gislene Gabrielle Faversani

More information

Computed Radiography

Computed Radiography BAM Berlin Computed Radiography --INDE 2007, Kalpakkam, India -- Uwe Zscherpel, Uwe Ewert BAM Berlin, Division VIII.3 Requests Requests and and information information to: to: Dr. Dr. U. U. Zscherpel Zscherpel

More information

CR Basics and FAQ. Overview. Historical Perspective

CR Basics and FAQ. Overview. Historical Perspective Page: 1 of 6 CR Basics and FAQ Overview Computed Radiography is a term used to describe a system that electronically records a radiographic image. Computed Radiographic systems use unique image receptors

More information

Digital Detector Array Image Quality for Various GOS Scintillators

Digital Detector Array Image Quality for Various GOS Scintillators Digital Detector Array Image Quality for Various GOS Scintillators More info about this article: http://www.ndt.net/?id=22768 Brian S. White 1, Mark E. Shafer 2, William H. Russel 3, Eric Fallet 4, Jacques

More information

Modeling detector performance in digital mammography using the linear cascaded systems approach

Modeling detector performance in digital mammography using the linear cascaded systems approach 51 Modeling detector performance in digital mammography using the linear cascaded systems approach V. Spyropoulou a,b, N. Kalivas a,b, A. Gaitanis a,b, C. Michail a,b, G. Panayiotakis b, I. Kandarakis

More information