Display of mammograms on a CRT

Size: px
Start display at page:

Download "Display of mammograms on a CRT"

Transcription

1 Display of mammograms on a CRT Hans Roehrig, Ph.D. William J. Dallas, Ph.D. Elizabeth Krupinski, Ph.D. Jiahua Fan, M.S. University of Arizona This work was supported by 2 Grants from NIH

2 In most radiological imaging workstations the monochrome cathode ray tube (CRT) is the electronic display of choice. The CRT offers the best performance; it is the most highly developed, mature and reliable display in common use. Unfortunately, because of limited spatial resolution and because of limited dynamic range, the efficacy of soft-copy versus hard-copy diagnosis has been challenged.

3

4 It has been shown that, when the images are displayspecifically processed before the display to compensate for the reduced MTF of the CRT and its reduced dynamic range, current high-resolution monochrome CRT display systems will present all the information in mammograms without interaction by the reader 1,2. The basic idea is to perform an operation on the image before it is displayed that will counteract the degradation that will come from the action of the display system s MTF.

5 This paper discusses display specific processing with respect to spatial resolution (MTF) for the specific case of Mammography. The display specific processing is based on the evaluation of the CRT image quality with the aid of a CCD camera.

6 The CCD Camera was used to image test-stimuli like squarewave patterns and single lines to determine the MTF It was also used to take images of some of the processed and unprocessed images as they appeared on the CRT for the purpose of close inspection and analysis.

7 We use the Wiener-Helstrom filtering method to prepare the image for display. Denoting the original image u(x,y), its Fourier transform is U(ξ,η). Modeling the display as an approximately linear shift-invariant system, the output will be degraded on the display by the optical transfer function OTF(ξ,η) according to the equation below, where N represents the Fourier spectral density of the noise. V ( ξ, η) = U ( ξ, η)[ OTF( ξ, η) + N( ξ, η)] The use of the Wiener filter in this form comes from the fact that the actual restoration we are attempting is variations in the point spread function, not in the image.

8 Because the OTF is positive over the entire region of frequency space, we may replace it by its absolute value, the modulation transfer function (MTF). In actual practice, we replace the noise power spectral density by a constant that we adjust as a parameter in the filter. We make an additional simplification and assume the MTF to be separable so that we can actually perform the measurements and the filtering as 1-D operations. MTFξ ( ξ ) MTFη ( η) F( ξ, η) = MTF ( ξ ) N MTF ( η) N ξ + Caveat: There is subtle ringing in the processed image resulting from discontinuities in the unprocessed image, especially at the borders. The MTF of the CRT reduces ringing. Total removal of ringing requires extra work. ξ η + η

9 Processing and ringing Image with single line: unprocessed Image with single line: processed Ringing Profile of the above unprocessed single line image Profile of the above processed single line image with ringing

10 Processing and ringing (Continued) Single line image: unprocessed and displayed on the CRT Single line image: processed and displayed on the CRT Ringing is reduced because of the CRT MTF Profile of the above unprocessed single line image Profile of the above processed single line image with subtle ringing

11 MTFs(from the Square Wave response) of a5m-pixel Monitor with a P45 Phosphor for Different Mean Luminances (Nyquist Frequency: 3.47 lp/mm) Vertical MTFs cd/m 2 (ADU 55) Compensated Horizontal MTFs of a 5M-pixel Monitor with a P45 Phosphor for Different Estimates(n) of Noise (Luminance: 64 cd/m 2, Nyquist Frequency: 3.47 lp/mm) MTF cd/m 2 (ADU 55) Horizontal MTFs cd/m 2 (120 ADU) cd/m 2 (120 ADU) cd/m 2 (ADU 210) cd/m 2 (ADU 210) Spatial Frequency(lp/mm) MTF Unprocessed Vertical MTF 1 2 Unprocessed Horizontal MTF Spatial Frequency (lp/mm) 12 Note that n is a parameter for noise we adjust in the Wiener filter. The larger n, the larger the estimate of the noise, and the lower the boost on the high frequency.

12 Typical Mammogram Right MLO Right CC Microcalcifications, which need to be displayed on a CRT with high fidelity.

13 Pixel Magnitude (CCD Camera ADU) Profiles of a Micro-Calcification before and after processing with n = 1 to compensate for the monitor MTF The slopes of the Micro_Calcification are noticeably steeper after the processing.the Micro-Calcification appears now much sharper. Processed Unprocessed Pixel Magnitude (CCD Camera ADU) Profiles of a Micro-Calcification after processing with n = 1 and n = 10 The slopes of the Micro_Calcification after processing with n = 1 are slightly steeper than after processing with n = 10.The Micro-Calcification appears slightly more sharp for n = 1 Processed with n = 1 Processed with n = Pixel Position Pixel Position

14 Examples of Processing of Microcalcifications Unprocessed image from the database Unprocessed image as displayed on CRT Processed image as displayed on CRT Difference between processed image and unprocessed image

15 Profiles of the calcification in the center of the previous images Pixel Magnitude (CCD Camera ADU) Profiles of a Micro-Calcification before and after processing with n = 1 to compensate for the monitor MTF The slopes of the Micro_Calcification are noticeably steeper after the processing.the Micro-Calcification appears now much sharper. Processed Unprocessed Pixel Magnitude (CCD Camera ADU) Profiles of a Micro-Calcification after processing with n = 1 and n = 10 The slopes of the Micro_Calcification after processing with n = 1 are slightly steeper than after processing with n = 10.The Micro-Calcification appears slightly more sharp for n = 1 Processed with n = 1 Processed with n = Pixel Position Pixel Position

16 Example of Processing of a Benign Microcalcifications Unprocessed image from the database Unprocessed image as displayed on the CRT Image on the film Processed image as displayed on the CRT Difference between processed image and unprocessed image

17 Results of a Pilot ROC Study Mean ROC Az % 50% 75% 100% Overall Unprocessed Processed Lesion Contrast Performance as measured by ROC Az (Receiver Operating Characteristic Area Under the Curve) was higher at all contrast levels for the images processed for the monitor MTF compared to the unprocessed versions, but the difference was not significant.

18 The data are based on 2 observers viewing sets of 250 images in both the unprocessed and MTF-correction processed images. Lesion contrasts are at 25%, 50%, 75% and 100% (original images). Performance as measured by ROC Az (Receiver Operating Characteristic Area Under the Curve) was higher at all contrast levels for the MTF-correction processed images compared to the unprocessed versions. With only 2 observers at this time, none of the conditions reached statistical significance, although the early trends suggest that with more readers some of the contrast level conditions (50%, 75% and overall) are likely to reach statistical significance.

19 Conclusion Image processing by using the Wiener-Helstrom filtering method compensates for the blurring caused by the MTF of the CRT. Initial results indicate that this compensation is clinically useful. Display specific processing must also include consideration for the reduced dynamic range, which is not included in this study but will be implemented in the future.

20 References 1. Reiker GG, Blume H, Slone RM, Woodard PM, Gierada DS, Sagel SS, Jost RG, Blaine GJ: Film less Digital Chest Radiography within the Radiology Department; SPIE Vol 3035, , Roehrig H, The Monochrome Cathode Ray Tube Display and its Performance ; Volume 3, Chapter 4 of Yongmin Kim and Steven C. Horii, eds, Display and PACS ; in Jacob Beutel, Harold L. Kundel and Richard L. Van Metter, eds, Handbook of Medical Imaging, SPIE Press, Bellingham,WA, June 2000.

SUSPENSION CRITERIA FOR IMAGE MONITORS AND VIEWING BOXES.

SUSPENSION CRITERIA FOR IMAGE MONITORS AND VIEWING BOXES. SUSPENSION CRITERIA FOR IMAGE MONITORS AND VIEWING BOXES. Tingberg, Anders Published in: Radiation Protection Dosimetry DOI: 10.1093/rpd/ncs302 Published: 2013-01-01 Link to publication Citation for published

More information

Defense Technical Information Center Compilation Part Notice

Defense Technical Information Center Compilation Part Notice UNCLASSIFIED Defense Technical Information Center Compilation Part Notice ADPO 11345 TITLE: Measurement of the Spatial Frequency Response [SFR] of Digital Still-Picture Cameras Using a Modified Slanted

More information

Optimization of Digital Mammography Resolution Using Magnification Technique in Computed Radiography 1

Optimization of Digital Mammography Resolution Using Magnification Technique in Computed Radiography 1 Optimization of Digital Mammography Resolution Using Magnification Technique in Computed Radiography 1 Gham Hur, M.D., Yoon Joon Hwang, M.D., Soon Joo Cha, M.D., Su Young Kim, M.D., Yong Hoon Kim, M.D.

More information

Observer Performance of Reduced X-Ray Images on Liquid Crystal Displays

Observer Performance of Reduced X-Ray Images on Liquid Crystal Displays Original Paper Forma, 29, S45 S51, 2014 Observer Performance of Reduced X-Ray Images on Liquid Crystal Displays Akiko Ihori 1, Chihiro Kataoka 2, Daigo Yokoyama 2, Naotoshi Fujita 3, Naruomi Yasuda 4,

More information

LCD handheld displays characterization by means of the MTF measurement

LCD handheld displays characterization by means of the MTF measurement MSc in Photonics Universitat Politècnica de Catalunya (UPC) Universitat Autònoma de Barcelona (UAB) Universitat de Barcelona (UB) Institut de Ciències Fotòniques (ICFO) PHOTONICSBCN http://www.photonicsbcn.eu

More information

Threshold contrast visibility of micro calcifications in digital mammography

Threshold contrast visibility of micro calcifications in digital mammography Threshold contrast visibility of micro calcifications in digital mammography Ann-Katherine Carton*, Hilde Bosmans*, Dirk Vandenbroucke, Chantal Van Ongeval*, Geert Souverijns*, Frank Rogge*, Guy Marchal*

More information

Preliminary Modulation Transfer Function Study on Amorphous Silicon Flat Panel System for Industrial Digital Radiography

Preliminary Modulation Transfer Function Study on Amorphous Silicon Flat Panel System for Industrial Digital Radiography ECNDT 26 - Poster 17 Preliminary Modulation Transfer Function Study on Amorphous Silicon Flat Panel System for Industrial Digital Radiography Khairul Anuar MOHD SALLEH, Ab. Razak HAMZAH and Mohd Ashhar

More information

Influence of the Display Monitor on Observer Performance in Detection of Dental Caries

Influence of the Display Monitor on Observer Performance in Detection of Dental Caries Received 25 July 2007; Accepted 1 September 2007 Influence of the Display Monitor on Observer Performance in Detection of Dental Caries Farzad Esmaeili 1 Esrafil Balaei 2 Firoz Pouralibaba 3 * Farzaneh

More information

Digital Detector Array Image Quality for Various GOS Scintillators

Digital Detector Array Image Quality for Various GOS Scintillators Digital Detector Array Image Quality for Various GOS Scintillators More info about this article: http://www.ndt.net/?id=22768 Brian S. White 1, Mark E. Shafer 2, William H. Russel 3, Eric Fallet 4, Jacques

More information

Amorphous Selenium Direct Radiography for Industrial Imaging

Amorphous Selenium Direct Radiography for Industrial Imaging DGZfP Proceedings BB 67-CD Paper 22 Computerized Tomography for Industrial Applications and Image Processing in Radiology March 15-17, 1999, Berlin, Germany Amorphous Selenium Direct Radiography for Industrial

More information

Performance of Image Intensifiers in Radiographic Systems

Performance of Image Intensifiers in Radiographic Systems DOE/NV/11718--396 LA-UR-00-211 Performance of Image Intensifiers in Radiographic Systems Stuart A. Baker* a, Nicholas S. P. King b, Wilfred Lewis a, Stephen S. Lutz c, Dane V. Morgan a, Tim Schaefer a,

More information

MTF Analysis and its Measurements for Digital Still Camera

MTF Analysis and its Measurements for Digital Still Camera MTF Analysis and its Measurements for Digital Still Camera Yukio Okano*, Minolta Co., Ltd. Takatsuki Laboratory, Takatsuki, Japan *present address Sharp Company, Nara, Japan Abstract MTF(Modulation Transfer

More information

DIGITAL IMAGE PROCESSING IN X-RAY IMAGING

DIGITAL IMAGE PROCESSING IN X-RAY IMAGING DIGITAL IMAGE PROCESSING IN X-RAY IMAGING Shalini Kumari 1, Bachan Prasad 2,Aliya Nasim 3 Department of Electronics And Communication Engineering R.V.S College of Engineering & Technology, Jamshedpur,

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

A comparison between medical-grade liquid crystal display (LCD) and ipad color imaging

A comparison between medical-grade liquid crystal display (LCD) and ipad color imaging A comparison between medical-grade liquid crystal display (LCD) and ipad color imaging Poster No.: C-1377 Congress: ECR 2014 Type: Scientific Exhibit Authors: K. Yoshimura 1, K. Shibata 2, T. Nihashi 1,

More information

Module 3: Video Sampling Lecture 18: Filtering operations in Camera and display devices. The Lecture Contains: Effect of Temporal Aperture:

Module 3: Video Sampling Lecture 18: Filtering operations in Camera and display devices. The Lecture Contains: Effect of Temporal Aperture: The Lecture Contains: Effect of Temporal Aperture: Spatial Aperture: Effect of Display Aperture: file:///d /...e%20(ganesh%20rana)/my%20course_ganesh%20rana/prof.%20sumana%20gupta/final%20dvsp/lecture18/18_1.htm[12/30/2015

More information

Digital Quality Control Phantom for X-ray Mammography Image

Digital Quality Control Phantom for X-ray Mammography Image Digital Quality Control Phantom for X-ray Mammography Image Yuichi Nagai, Atsuko Takada National: Cancer Hospital East Department of Radiology Collaborated by Nobuyuki Niwa : Kyoto Kagaku Co., Ltd. Introduction

More information

A No Reference Image Blur Detection using CPBD Metric and Deblurring of Gaussian Blurred Images using Lucy-Richardson Algorithm

A No Reference Image Blur Detection using CPBD Metric and Deblurring of Gaussian Blurred Images using Lucy-Richardson Algorithm A No Reference Image Blur Detection using CPBD Metric and Deblurring of Gaussian Blurred Images using Lucy-Richardson Algorithm Suresh S. Zadage, G. U. Kharat Abstract This paper addresses sharpness of

More information

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 2 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 COMPUTERIZED IMAGING Section I: Chapter 2 RADT

More information

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893 Digital Radiography D. J. Hall, Ph.D. x20893 djhall@ucsd.edu Background Common Digital Modalities Digital Chest Radiograph - 4096 x 4096 x 12 bit CT - 512 x 512 x 12 bit SPECT - 128 x 128 x 8 bit MRI -

More information

Investigation of the line-pair pattern method for evaluating mammographic focal spot performance

Investigation of the line-pair pattern method for evaluating mammographic focal spot performance Investigation of the line-pair pattern method for evaluating mammographic focal spot performance Mitchell M. Goodsitt, a) Heang-Ping Chan, and Bob Liu Department of Radiology, University of Michigan, Ann

More information

The Effect of Intensity Windowing on the Detection of Simulated Masses Embedded in Dense Portions of Digitized Mammograms in a Laboratory Setting

The Effect of Intensity Windowing on the Detection of Simulated Masses Embedded in Dense Portions of Digitized Mammograms in a Laboratory Setting The Effect of Intensity Windowing on the Detection of Simulated Masses Embedded in Dense Portions of Digitized Mammograms in a Laboratory Setting Etta D. Pisano, MD* # Jayanthi Chandramouli` Bradley M.

More information

Neonatal Chest Computed Radiography: Image Processing and Optimal Image Display

Neonatal Chest Computed Radiography: Image Processing and Optimal Image Display Neonatal Computed Radiography Pediatric Imaging Clinical Observations Steven Don 1,2 Bruce R. Whiting 2 Jacquelyn S. Ellinwood 3 David H. Foos 3 Keith A. Kronemer 1 Richard A. Kraus 1 Don S, Whiting BR,

More information

1.1.Clinical Artefacts

1.1.Clinical Artefacts 1.1.Clinical Artefacts While the incidence of artefact on digital mammographic images 1 is typically less than with film based mammography, artefacts can be produced on digital systems. This section provides

More information

Pseudocolour Image Processing in Digital Mammography

Pseudocolour Image Processing in Digital Mammography Pseudocolour Image Processing in Digital Mammography M.Y. Sanavullah 1 and R. Samson Ravindran 2 1 Principal and Research Guide, VMKV Engineering College, V.M.R.F Deemed University, Salem, TN 2 Research

More information

A Practical Overview of the Clinical and Operational Impact of Computed Radiography(CR) Implementations. Shirley Weddle, RT(R)(M), CIIP, BBA

A Practical Overview of the Clinical and Operational Impact of Computed Radiography(CR) Implementations. Shirley Weddle, RT(R)(M), CIIP, BBA A Practical Overview of the Clinical and Operational Impact of Computed Radiography(CR) Implementations Shirley Weddle, RT(R)(M), CIIP, BBA OBJECTIVES Define Computed Radiography (CR) Discuss CR vendor

More information

Chapter 2 Fourier Integral Representation of an Optical Image

Chapter 2 Fourier Integral Representation of an Optical Image Chapter 2 Fourier Integral Representation of an Optical This chapter describes optical transfer functions. The concepts of linearity and shift invariance were introduced in Chapter 1. This chapter continues

More information

Refined Slanted-Edge Measurement for Practical Camera and Scanner Testing

Refined Slanted-Edge Measurement for Practical Camera and Scanner Testing Refined Slanted-Edge Measurement for Practical Camera and Scanner Testing Peter D. Burns and Don Williams Eastman Kodak Company Rochester, NY USA Abstract It has been almost five years since the ISO adopted

More information

A Study of Slanted-Edge MTF Stability and Repeatability

A Study of Slanted-Edge MTF Stability and Repeatability A Study of Slanted-Edge MTF Stability and Repeatability Jackson K.M. Roland Imatest LLC, 2995 Wilderness Place Suite 103, Boulder, CO, USA ABSTRACT The slanted-edge method of measuring the spatial frequency

More information

SECTION I - CHAPTER 1 DIGITAL RADIOGRAPHY: AN OVERVIEW OF THE TEXT. Exam Content Specifications 8/22/2012 RADT 3463 COMPUTERIZED IMAGING

SECTION I - CHAPTER 1 DIGITAL RADIOGRAPHY: AN OVERVIEW OF THE TEXT. Exam Content Specifications 8/22/2012 RADT 3463 COMPUTERIZED IMAGING RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 1 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 1 DIGITAL RADIOGRAPHY: AN OVERVIEW OF THE TEXT RADT 3463 COMPUTERIZED IMAGING Section I: Chapter

More information

Features and Weaknesses of Phantoms for CR/DR System Testing

Features and Weaknesses of Phantoms for CR/DR System Testing Physics testing of image detectors Parameters to test Features and Weaknesses of Phantoms for CR/DR System Testing Spatial resolution Contrast resolution Uniformity/geometric distortion Dose response/signal

More information

Comparison of Reconstruction Algorithms for Images from Sparse-Aperture Systems

Comparison of Reconstruction Algorithms for Images from Sparse-Aperture Systems Published in Proc. SPIE 4792-01, Image Reconstruction from Incomplete Data II, Seattle, WA, July 2002. Comparison of Reconstruction Algorithms for Images from Sparse-Aperture Systems J.R. Fienup, a * D.

More information

SYLLABUS. TITLE: Equipment Operation I. DEPARTMENT: Radiologic Technology

SYLLABUS. TITLE: Equipment Operation I. DEPARTMENT: Radiologic Technology CODE: RADT 156 INSTITUTE: Health Science TITLE: Equipment Operation I DEPARTMENT: Radiologic Technology COURSE DESCRIPTION: This course covers the principles of equipment operation and maintenance of radiographic

More information

30 lesions. 30 lesions. false positive fraction

30 lesions. 30 lesions. false positive fraction Solutions to the exercises. 1.1 In a patient study for a new test for multiple sclerosis (MS), thirty-two of the one hundred patients studied actually have MS. For the data given below, complete the two-by-two

More information

Extraction of Lesions and Micro calcifications from Mammograms of Breast Images: A survey

Extraction of Lesions and Micro calcifications from Mammograms of Breast Images: A survey RESEARCH ARTICLE OPEN ACCESS Extraction of Lesions and Micro calcifications from Mammograms of Breast Images: A survey Abhay Goyal Abstract: Images taken from different scans have always been a method

More information

Image Display and Perception

Image Display and Perception Image Display and Perception J. Anthony Seibert, Ph.D. Department of Radiology UC Davis Medical Center Sacramento, California, USA Image acquisition, display, & interpretation X-rays kvp mas Tube filtration

More information

Digital Images & Image Quality

Digital Images & Image Quality Introduction to Medical Engineering (Medical Imaging) Suetens 1 Digital Images & Image Quality Ho Kyung Kim Pusan National University Radiation imaging DR & CT: x-ray Nuclear medicine: gamma-ray Ultrasound

More information

Appropriate Inspection Distance of Digital X-Ray Imaging Equipment for Diagnosis

Appropriate Inspection Distance of Digital X-Ray Imaging Equipment for Diagnosis Indian Journal of Science and Technology Vol 8(S8), 380-386, April 2015 ISSN (Print) : 0974-6846 ISSN (Online) : 0974-5645 DOI: 10.17485/ijst/2015/v8iS8/70528 Appropriate Inspection Distance of Digital

More information

Comparison of computed radiography and filmõscreen combination using a contrast-detail phantom

Comparison of computed radiography and filmõscreen combination using a contrast-detail phantom JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 1, WINTER 2003 Comparison of computed radiography and filmõscreen combination using a contrast-detail phantom Z. F. Lu,* E. L. Nickoloff, J.

More information

Evaluation of a quality control phantom for digital chest radiography

Evaluation of a quality control phantom for digital chest radiography JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 2, NUMBER 2, SPRING 2001 Evaluation of a quality control phantom for digital chest radiography Eugene Mah* Department of Radiology, Medical University

More information

Influence of Image Enhancement Processing on SFR of Digital Cameras

Influence of Image Enhancement Processing on SFR of Digital Cameras IS&T s 998 PICS Conference Copyright 998, IS&T Influence of Image Processing on SFR of Digital Cameras Yukio Okano Sharp Corporation, Information Systems Labs. Yamatokoriyama, Nara, JAPAN Abstract The

More information

Defocusing and Deblurring by Using with Fourier Transfer

Defocusing and Deblurring by Using with Fourier Transfer Defocusing and Deblurring by Using with Fourier Transfer AKIRA YANAGAWA and TATSUYA KATO 1. Introduction Image data may be obtained through an image system, such as a video camera or a digital still camera.

More information

What is an image? Bernd Girod: EE368 Digital Image Processing Pixel Operations no. 1. A digital image can be written as a matrix

What is an image? Bernd Girod: EE368 Digital Image Processing Pixel Operations no. 1. A digital image can be written as a matrix What is an image? Definition: An image is a 2-dimensional light intensity function, f(x,y), where x and y are spatial coordinates, and f at (x,y) is related to the brightness of the image at that point.

More information

An Improved Phantom for Quality Control of Laser Scanner Digitizers in Picture Archival and Communications Systems

An Improved Phantom for Quality Control of Laser Scanner Digitizers in Picture Archival and Communications Systems An Improved Phantom for Quality Control of Laser Scanner Digitizers in Picture Archival and Communications Systems Ethan J. Halpern and Peter D. Esser Various quality control (QC) procedures may be used

More information

ISO INTERNATIONAL STANDARD. Photography Electronic still-picture cameras Resolution measurements

ISO INTERNATIONAL STANDARD. Photography Electronic still-picture cameras Resolution measurements INTERNATIONAL STANDARD ISO 12233 First edition 2000-09-01 Photography Electronic still-picture cameras Resolution measurements Photographie Appareils de prises de vue électroniques Mesurages de la résolution

More information

Evaluating clinical performance of the new Coronis Fusion 6MP DL display

Evaluating clinical performance of the new Coronis Fusion 6MP DL display White Paper Evaluating clinical performance of the new Coronis Fusion 6MP DL display What s inside? Clinical performance of Coronis Fusion 6MP DL for lung nodule screening ROC curves for Coronis Fusion

More information

Ayaz Jhanorwala, Rishikesh Kumar, Satish Tilva. L & T MHPS Boilers Pvt Ltd., Hazira, Surat , India

Ayaz Jhanorwala, Rishikesh Kumar, Satish Tilva. L & T MHPS Boilers Pvt Ltd., Hazira, Surat , India National Seminar & Exhibition on Non-Destructive Evaluation, NDE 2014, Pune, December 4-6, 2014 (NDE-India 2014) Vol.20 No.6 (June 2015) - The e-journal of Nondestructive Testing - ISSN 1435-4934 www.ndt.net/?id=17842

More information

Indirect Method of Measurement and Evaluating the Noise Power Spectrum of A Medical X-ray Imaging System

Indirect Method of Measurement and Evaluating the Noise Power Spectrum of A Medical X-ray Imaging System Indirect Method of Measurement and Evaluating the Noise Power Spectrum of A Medical X-ray Imaging System Zeki Ahmed Darwish National University of Malaysia (UKM) Bangi 43600, Selangor, Malaysia Abstract

More information

Phase Contrast Imaging with X-ray tube

Phase Contrast Imaging with X-ray tube Phase Contrast Imaging with X-ray tube Institute for Roentgen Optics /IRO/, Moscow Vladimir Shovkun and Muradin Kumakhov Proc. SPIE v.5943, 2005 Institute for Roentgen Optics. Vladimir Ya. Shovkun. E-mail:

More information

Do you have any other questions? Please call us at (Toll Free) or , or

Do you have any other questions? Please call us at (Toll Free) or , or INSTRUCTIONS Read the appropriate course/ textbook. This is an open book test. A score of 75% or higher is needed to receive CE credit. You will have a maximum of three attempts to pass this course. Please

More information

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Acceptance Testing of a Digital Breast Tomosynthesis Unit Acceptance Testing of a Digital Breast Tomosynthesis Unit 2012 AAPM Spring Clinical Meeting Jessica Clements, M.S., DABR Objectives Review of technology and clinical advantages Acceptance Testing Procedures

More information

Visual Requirements for High-Fidelity Display 1

Visual Requirements for High-Fidelity Display 1 Michael J Flynn, PhD Visual Requirements for High-Fidelity Display 1 The digital radiographic process involves (a) the attenuation of x rays along rays forming an orthographic projection, (b) the detection

More information

Comparison between film-screen and computed radiography systems in Brazilian mammography

Comparison between film-screen and computed radiography systems in Brazilian mammography X Congreso Regional Latinoamericano IRPA de Protección y Seguridad Radiológica Radioprotección: Nuevos Desafíos para un Mundo en Evolución Buenos Aires, al 7 de abril, 05 SOCIEDAD ARGENTINA DE RADIOPROTECCIÓN

More information

Image acquisition. In both cases, the digital sensing element is one of the following: Line array Area array. Single sensor

Image acquisition. In both cases, the digital sensing element is one of the following: Line array Area array. Single sensor Image acquisition Digital images are acquired by direct digital acquisition (digital still/video cameras), or scanning material acquired as analog signals (slides, photographs, etc.). In both cases, the

More information

High Resolution Imaging for Inspection of Laser Beam Melting Systems

High Resolution Imaging for Inspection of Laser Beam Melting Systems High Resolution Imaging for Inspection of Laser Beam Melting Systems a), Stefan Kleszczynski b), Dorian Schneider a), and Gerd Witt b) a) Institute of Imaging and Computer Vision RWTH Aachen University

More information

Small Animal Radiographic Techniques and Positioning COPYRIGHTED MATERIAL

Small Animal Radiographic Techniques and Positioning COPYRIGHTED MATERIAL Small Animal Radiographic Techniques and Positioning COPYRIGHTED MATERIAL Section 1 Theory and Equipment 1 Introduction to Digital Imaging Small animal radiography has changed dramatically in the past

More information

PRACTICE GUIDELINE FOR DETERMINANTS OF IMAGE QUALITY IN DIGITAL MAMMOGRAPHY

PRACTICE GUIDELINE FOR DETERMINANTS OF IMAGE QUALITY IN DIGITAL MAMMOGRAPHY The American College of Radiology, with more than 30,000 members, is the principal organization of radiologists, radiation oncologists, and clinical medical physicists in the United States. The College

More information

radiography detector

radiography detector Clinical evaluation of a full field digital projection radiography detector Gary S. Shaber'1, Denny L. Leeb, Jeffrey Belib, Gregory Poweii1', Andrew D.A. Maidment'1 a Thomas Jefferson University Hospital,

More information

Criteria for Optical Systems: Optical Path Difference How do we determine the quality of a lens system? Several criteria used in optical design

Criteria for Optical Systems: Optical Path Difference How do we determine the quality of a lens system? Several criteria used in optical design Criteria for Optical Systems: Optical Path Difference How do we determine the quality of a lens system? Several criteria used in optical design Computer Aided Design Several CAD tools use Ray Tracing (see

More information

Published text: Institute of Cancer Research Repository Please direct all s to:

Published text: Institute of Cancer Research Repository   Please direct all  s to: This is an author produced version of an article that appears in: MEDICAL PHYSICS The internet address for this paper is: https://publications.icr.ac.uk/1316/ Copyright information: http://www.aip.org/pubservs/web_posting_guidelines.html

More information

Sharpness, Resolution and Interpolation

Sharpness, Resolution and Interpolation Sharpness, Resolution and Interpolation Introduction There are a lot of misconceptions about resolution, camera pixel count, interpolation and their effect on astronomical images. Some of the confusion

More information

digital film technology Resolution Matters what's in a pattern white paper standing the test of time

digital film technology Resolution Matters what's in a pattern white paper standing the test of time digital film technology Resolution Matters what's in a pattern white paper standing the test of time standing the test of time An introduction >>> Film archives are of great historical importance as they

More information

Image Deblurring. This chapter describes how to deblur an image using the toolbox deblurring functions.

Image Deblurring. This chapter describes how to deblur an image using the toolbox deblurring functions. 12 Image Deblurring This chapter describes how to deblur an image using the toolbox deblurring functions. Understanding Deblurring (p. 12-2) Using the Deblurring Functions (p. 12-5) Avoiding Ringing in

More information

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis

More information

REAL-TIME X-RAY IMAGE PROCESSING; TECHNIQUES FOR SENSITIVITY

REAL-TIME X-RAY IMAGE PROCESSING; TECHNIQUES FOR SENSITIVITY REAL-TIME X-RAY IMAGE PROCESSING; TECHNIQUES FOR SENSITIVITY IMPROVEMENT USING LOW-COST EQUIPMENT R.M. Wallingford and J.N. Gray Center for Aviation Systems Reliability Iowa State University Ames,IA 50011

More information

PIAS -II. Print Quality Measurements anytime, anywhere objective, reliable, easy. Innovative measurement instruments from

PIAS -II. Print Quality Measurements anytime, anywhere objective, reliable, easy. Innovative measurement instruments from Print Quality Measurements anytime, anywhere objective, reliable, easy PIS -II is QE s cutting-edge portable measurement device for objective image quality analysis. With the PIS -II, image evaluation

More information

PLD5600A High Frequency Digital Gastrointestinal &DR System(630mA)

PLD5600A High Frequency Digital Gastrointestinal &DR System(630mA) PLD5600A High Frequency Digital Gastrointestinal &DR System(630mA) Application: Full support perspective, gastrointestinal spot film, GI (barium meal, barium enema), orthopedic photography, pediatrics

More information

Improved Tomosynthesis Reconstruction using Super-resolution and Iterative Techniques

Improved Tomosynthesis Reconstruction using Super-resolution and Iterative Techniques Improved Tomosynthesis Reconstruction using Super-resolution and Iterative Techniques Wataru FUKUDA* Junya MORITA* and Masahiko YAMADA* Abstract Tomosynthesis is a three-dimensional imaging technology

More information

Optical Performance of Nikon F-Mount Lenses. Landon Carter May 11, Measurement and Instrumentation

Optical Performance of Nikon F-Mount Lenses. Landon Carter May 11, Measurement and Instrumentation Optical Performance of Nikon F-Mount Lenses Landon Carter May 11, 2016 2.671 Measurement and Instrumentation Abstract In photographic systems, lenses are one of the most important pieces of the system

More information

SPECT Reconstruction & Filtering

SPECT Reconstruction & Filtering SPECT Reconstruction & Filtering Goals Understand the basics of SPECT Reconstruction Filtered Backprojection Iterative Reconstruction Make informed choices on filter selection and settings Pre vs. Post

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 16371-1 First edition 2011-10-01 Non-destructive testing Industrial computed radiography with storage phosphor imaging plates Part 1: Classification of systems Essais non destructifs

More information

The Appearance of Images Through a Multifocal IOL ABSTRACT. through a monofocal IOL to the view through a multifocal lens implanted in the other eye

The Appearance of Images Through a Multifocal IOL ABSTRACT. through a monofocal IOL to the view through a multifocal lens implanted in the other eye The Appearance of Images Through a Multifocal IOL ABSTRACT The appearance of images through a multifocal IOL was simulated. Comparing the appearance through a monofocal IOL to the view through a multifocal

More information

On spatial resolution

On spatial resolution On spatial resolution Introduction How is spatial resolution defined? There are two main approaches in defining local spatial resolution. One method follows distinction criteria of pointlike objects (i.e.

More information

Acquisition, Processing and Display

Acquisition, Processing and Display Acquisition, Processing and Display Terri L. Fauber, R.T. (R)(M) Department of Radiation Sciences School of Allied Health Professions Virginia Commonwealth University Topics Image Characteristics Image

More information

Abstract. ZEIGLER, GARY BOYCE, II. Direct Detection of Microcalcification Pairs in Simulated

Abstract. ZEIGLER, GARY BOYCE, II. Direct Detection of Microcalcification Pairs in Simulated Abstract ZEIGLER, GARY BOYCE, II. Direct Detection of Microcalcification Pairs in Simulated Digital Mammograms. (Under the Direction of Professor Kuruvilla Verghese.) Using the MCMIS (Monte Carlo for Mammography

More information

Image evaluation method for digital mammography with soft-copy reading using a digital phantom

Image evaluation method for digital mammography with soft-copy reading using a digital phantom Image evaluation method for digital mammography with soft-copy reading using a digital phantom Norimitsu Shinohara 1, Katsuhei Horita 2, Tokiko Endo 3 1 Department of Radiological Technology, Faculty of

More information

Open. the Digitized world. Fuji Computed Radiography

Open. the Digitized world. Fuji Computed Radiography Open the Digitized world Fuji Computed Radiography If just one of these applies to you... Managing developing fluid is hard and darkroom work is a hassle... Images are not stable... Isn t digitalization

More information

Use of Wavelet Multiresolution Analysis to Reduce Radiation Dose in Digital Mammography

Use of Wavelet Multiresolution Analysis to Reduce Radiation Dose in Digital Mammography 2015 IEEE 28th International Symposium on Computer-Based Medical Systems Use of Wavelet Multiresolution Analysis to Reduce Radiation Dose in Digital Mammography Helder C. R. de Oliveira*, Lucas R. Borges*,

More information

A pre-processing technique to improve the performance of a computer-aided detection scheme in mammography

A pre-processing technique to improve the performance of a computer-aided detection scheme in mammography A pre-processing technique to improve the performance of a computer-aided detection scheme in mammography Larissa C. S. Romualdo, Marcelo A. C. Vieira, Homero Schiabel Laboratório de Análise e Processamento

More information

Attikon, Rimini 1, , Athens, Greece , Athens, Greece , Athens, Greece

Attikon, Rimini 1, , Athens, Greece , Athens, Greece , Athens, Greece Radiation Protection Dosimetry (2005), Vol. 117, No. 1 3, pp. 291 297 doi:10.1093/rpd/nci742 Advance Access published on February 7, 2006 QUALITY ASSURANCE (QA) PROCEDURES FOR SOFTWARE: EVALUATION OF AN

More information

Commercial Scanners and Science

Commercial Scanners and Science Commercial Scanners and Science Specs vs Reality Ian Shelton - DDO Bob Simcoe - Harvard 4/28/2008 RJS Starting with Pixels Photosensitive area on the CCD chip This pixel would often be called a 4um pixel

More information

Using Optics to Optimize Your Machine Vision Application

Using Optics to Optimize Your Machine Vision Application Expert Guide Using Optics to Optimize Your Machine Vision Application Introduction The lens is responsible for creating sufficient image quality to enable the vision system to extract the desired information

More information

Effect of Backscattered Radiation on X-Ray Image Contrast

Effect of Backscattered Radiation on X-Ray Image Contrast Applied Physics Research; Vol. 9, No. 1; 2017 ISSN 1916-9639 E-ISSN 1916-9647 Published by Canadian Center of Science and Education Effect of Backscattered Radiation on X-Ray Image Contrast A. T. Naji

More information

Optical basics for machine vision systems. Lars Fermum Chief instructor STEMMER IMAGING GmbH

Optical basics for machine vision systems. Lars Fermum Chief instructor STEMMER IMAGING GmbH Optical basics for machine vision systems Lars Fermum Chief instructor STEMMER IMAGING GmbH www.stemmer-imaging.de AN INTERNATIONAL CONCEPT STEMMER IMAGING customers in UK Germany France Switzerland Sweden

More information

Appearance Match between Soft Copy and Hard Copy under Mixed Chromatic Adaptation

Appearance Match between Soft Copy and Hard Copy under Mixed Chromatic Adaptation Appearance Match between Soft Copy and Hard Copy under Mixed Chromatic Adaptation Naoya KATOH Research Center, Sony Corporation, Tokyo, Japan Abstract Human visual system is partially adapted to the CRT

More information

Applied Sciences at Winterthur)

Applied Sciences at Winterthur) Comparison of image quality between a digital panorama X-ray unit with a CdTe-CMOS detector and panorama X- ray units with other types of digital detectors Stephan Scheidegger 1 1 Zürcher Hochschule für

More information

X-RAY FLUOROSCOPY IMAGING SYSTEMS. Dr Slavik Tabakov. Luminescence: Dept. Medical Eng. & Physics King s College London

X-RAY FLUOROSCOPY IMAGING SYSTEMS. Dr Slavik Tabakov. Luminescence: Dept. Medical Eng. & Physics King s College London X-RAY FLUOROSCOPY IMAGING SYSTEMS Dr Slavik Tabakov OBJECTIVES - Image Intensifier construction - Input window - Accelerating and focusing electrodes - Output window - Conversion factor - II characteristics

More information

Edge-Raggedness Evaluation Using Slanted-Edge Analysis

Edge-Raggedness Evaluation Using Slanted-Edge Analysis Edge-Raggedness Evaluation Using Slanted-Edge Analysis Peter D. Burns Eastman Kodak Company, Rochester, NY USA 14650-1925 ABSTRACT The standard ISO 12233 method for the measurement of spatial frequency

More information

Image Restoration and Super- Resolution

Image Restoration and Super- Resolution Image Restoration and Super- Resolution Manjunath V. Joshi Professor Dhirubhai Ambani Institute of Information and Communication Technology, Gandhinagar, Gujarat email:mv_joshi@daiict.ac.in Overview Image

More information

Mammography Solution. AMULET Innovality. The new leader in the AMULET series. Tomosynthesis, 3D mammography and biopsy are all available.

Mammography Solution. AMULET Innovality. The new leader in the AMULET series. Tomosynthesis, 3D mammography and biopsy are all available. Mammography Solution AMULET Innovality The new leader in the AMULET series. Tomosynthesis, 3D mammography and biopsy are all available. FUJIFILM supports the Pink Ribbon Campaign for early detection of

More information

ABOUT RESOLUTION. pco.knowledge base

ABOUT RESOLUTION. pco.knowledge base The resolution of an image sensor describes the total number of pixel which can be used to detect an image. From the standpoint of the image sensor it is sufficient to count the number and describe it

More information

Image Enhancement in Spatial Domain

Image Enhancement in Spatial Domain Image Enhancement in Spatial Domain 2 Image enhancement is a process, rather a preprocessing step, through which an original image is made suitable for a specific application. The application scenarios

More information

Multispectral. imaging device. ADVANCED LIGHT ANALYSIS by. Most accurate homogeneity MeasureMent of spectral radiance. UMasterMS1 & UMasterMS2

Multispectral. imaging device. ADVANCED LIGHT ANALYSIS by. Most accurate homogeneity MeasureMent of spectral radiance. UMasterMS1 & UMasterMS2 Multispectral imaging device Most accurate homogeneity MeasureMent of spectral radiance UMasterMS1 & UMasterMS2 ADVANCED LIGHT ANALYSIS by UMaster Ms Multispectral Imaging Device UMaster MS Description

More information

Focus-Aid Signal for Super Hi-Vision Cameras

Focus-Aid Signal for Super Hi-Vision Cameras Focus-Aid Signal for Super Hi-Vision Cameras 1. Introduction Super Hi-Vision (SHV) is a next-generation broadcasting system with sixteen times (7,680x4,320) the number of pixels of Hi-Vision. Cameras for

More information

KODAK VISION Expression 500T Color Negative Film / 5284, 7284

KODAK VISION Expression 500T Color Negative Film / 5284, 7284 TECHNICAL INFORMATION DATA SHEET TI2556 Issued 01-01 Copyright, Eastman Kodak Company, 2000 1) Description is a high-speed tungsten-balanced color negative camera film with color saturation and low contrast

More information

Measurement of Texture Loss for JPEG 2000 Compression Peter D. Burns and Don Williams* Burns Digital Imaging and *Image Science Associates

Measurement of Texture Loss for JPEG 2000 Compression Peter D. Burns and Don Williams* Burns Digital Imaging and *Image Science Associates Copyright SPIE Measurement of Texture Loss for JPEG Compression Peter D. Burns and Don Williams* Burns Digital Imaging and *Image Science Associates ABSTRACT The capture and retention of image detail are

More information

CS148: Introduction to Computer Graphics and Imaging. Displays. Topics. Spatial resolution Temporal resolution Tone mapping. Display technologies

CS148: Introduction to Computer Graphics and Imaging. Displays. Topics. Spatial resolution Temporal resolution Tone mapping. Display technologies CS148: Introduction to Computer Graphics and Imaging Displays Topics Spatial resolution Temporal resolution Tone mapping Display technologies Resolution World is continuous, digital media is discrete Three

More information

TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know. JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist

TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know. JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist Topics Image Uniformity and Artifacts Image Quality - Detail

More information

Investigation of Effective DQE (edqe) parameters for a flat panel detector

Investigation of Effective DQE (edqe) parameters for a flat panel detector Investigation of Effective DQE (edqe) parameters for a flat panel detector Poster No.: C-1892 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Exhibit D. Bor 1, S. Cubukcu 1, A. Yalcin 1, O.

More information

Image Enhancement in spatial domain. Digital Image Processing GW Chapter 3 from Section (pag 110) Part 2: Filtering in spatial domain

Image Enhancement in spatial domain. Digital Image Processing GW Chapter 3 from Section (pag 110) Part 2: Filtering in spatial domain Image Enhancement in spatial domain Digital Image Processing GW Chapter 3 from Section 3.4.1 (pag 110) Part 2: Filtering in spatial domain Mask mode radiography Image subtraction in medical imaging 2 Range

More information