A CMOS image sensor method of focal spot size measurement*

Size: px
Start display at page:

Download "A CMOS image sensor method of focal spot size measurement*"

Transcription

1 Australasian Physical & Engineering Sciences in Medicine Volume 27 Number 2, 2004 TECHNICAL REPORT A CMOS image sensor method of focal spot size measurement* T. Tuchyna 1 and D. Paix 2 1 Department of Medical Technology and Physics, Sir Charles Gairdner Hospital, Perth, Australia 2 School of Information & Electrical Engineering, University of South Australia, Mawson Lakes, Australia Abstract A phosphor opto-coupled monochrome CMOS image sensor with a slit diaphragm was used to investigate focal spot characteristics. Images were captured during x-ray exposure with a triggered frame grabber and subsequently enhanced. Dimensions of the focal spot width (1.39mm) and length (1.92mm) were determined from the focal spot intensity profiles and their corresponding Full Width at Half Maxima (FWHM) in two orthogonal orientations. The CMOS image sensor measurements demonstrated differences in the measured width and length dimensions when compared to film measurements. The obtained nominal focal spot values however showed that image-sensor determined focal spot dimensions agreed with the direct film and film-screen methods when based on the AS/NZS defined nominal focal spot values. The CMOS image sensor tested appears to lack the measurement accuracy required for the measurement of small focal spot sizes due in part to its limited camera sensitivity. Key words focal spot, radiography, CMOS image sensor, electronic imaging, quality assurance Introduction Charge coupled devices (CCDs) and their variant complimentary metal oxide semiconductor (CMOS) devices are finding an increasing number of image capture roles, having many research and industrial applications 1-5. Most small field radiological imaging systems use the proven and reliable CCD technology. CCD systems have found routine application in small field mammography for stereotactic localisation. Speller et. al. 6 used a CCD based camera for the production of pinhole focal spot images, whereas recently Rong et. al. 7 evaluated computed radiography and flat panel amorphous silicon plates in their ability to measure focal spot size. CCDs must however be fabricated in special CCD chip foundries, they are expensive and do not integrate easily with other circuit and image system components. The CMOS process of design and fabrication results in an imaging system with a greatly improved dynamic range, *Presented in part at the Engineering & Physical Sciences in Medicine Conference, Hobart, Australia, November 1998 Corresponding author: T. Tuchyna, Department of Medical Technology & Physics, Sir Charles Gairdner Hospital, Perth, Hospital Avenue, Nedlands WA 6009, Australia Tel: (08) , Fax: (08) , thomas.tuchyna@health.wa.gov.au Received: 28 January 2004; Accepted: 24 June 2004 Copyright 2004 ACPSEM/EA lower noise, easy circuit integration and reduced cost. In a preliminary assessment of the suitability of a CMOS image sensor for the capture and analysis of x-ray images, we evaluated a technique for measurement of the focal spot size for a radiographic tube, using a CMOS image sensor and the slit test tool. The performance of this technique is characterised and compared to the direct film and film/screen methods of focal spot size measurement. The three basic methods for the measurement of the focal spot in a radiography tube employ the capture and measurement of a pinhole, slit or star resolution pattern images 8. The slit method is the preferred choice of manufacturers and is endorsed in the Australian/New Zealand Standard (AS/NZS) 4274:1995 report 9. The size and shape of the focal spot is critical to the viewing quality of the resultant image, where viewing quality is interpreted as visibility and resolution of small details. In general radiography the focal spot is usually small enough to have little effect on the clinical value of most radiographs. However the focal spot can significantly degrade the image sharpness where resolution is critical, i.e. mammography 10 and magnification work 11. Further, the determination of the dimensions of the focal spot is useful since it ensures that the tube is operating correctly at installation and is continuing to work correctly. Materials The Image Sensor Module consists of an image sensor chip (Model: VVL 1011C-001, VLSI Vision, Edinburgh, Scotland) mounted on a printed circuit board (PCB). The chip 63

2 is essentially a photodiode array with each photodiode forming a part of a MOS transistor. The charge collected by each photodiode is read out through the switching of the transistor. The photodiode array consists of 312 x 287 pixels 2 forming a field of view of 6.1 x 4.6 mm 2 with a pixel size of 19.6 x 16.0 m. The photodiodes are read out one row at a time. A vertical register successively activates the row lines while a horizontal register controls the sequential read-out within each column line. The image sensor module requires only a 5V DC power input and produces a 1.0V p-p composite monochrome video output. In order to utilise the image sensor capabilities, a fluorescent phosphor was opto-coupled to the photodiode array using silicone optical grease to minimise reflections between the phosphor and photodiodes. A literature search along with a qualitative assessment of several commercially available phosphors suggested the use of gadolinium oxysulphide (Gd 2 O 2 S:Tb), the phosphor used in the construction of the KODAK LANEX Regular screens. A small piece of the actual Kodak Lanex regular screen (7 x 5 mm) with a stated phosphor density of approximately 67 mg/cm 2 was used. During irradiation, light that is generated in the phosphor screen is converted to charge within the photodiode, causing a current to flow between the photodiode and an amplifier. The charge is proportional to the light reaching the photodiode during irradiation. A clock crystal resident on the PCB governs the row and column readout rates, as well as controlling the time over which charge can be collected. The minimum integration time for the image sensor was 40ms. In order to capture the image sensor s monochrome video output signal a low cost PCI based frame grabber (Model: DT3155, Data Translation, Hants, England) was used with a Pentium 100MHZ, Personal Computer, (PC). The frame grabber requires both horizontal and vertical sync signals. The Sync circuitry provides a means of informing the frame grabber where in the video signal to expect a new frame acquisition. The DT3155 can accept an external trigger so that image acquisition can be synchronised with an event external to the PC. This was achieved by designing and constructing an electronic circuit where the initial pulse was taken from the X-ray On panel light of the x-ray generator, thus matching frame capture to the x-ray 'switch on' period. The circuit produced a TTL trigger pulse with a variable time delay required for the initiation of the frame grabber. The delay was used so that high frequency noise signals during the kv build up of the x-ray tube did not cause pre-triggering of the frame grabber. An Odel Polaire OT 1201, fully rectified, single phase x-ray generator was used (1973 model). The x-ray machine selectable ma settings were 50, 100, 200 and 300 ma. An indicator on the panel showed that the small focal spot was used at 50 and 100 ma, while the large focal spot was used at 200 and 300 ma. Due to the age of the x-ray machine, it was not possible to obtain manufacturer focal spot data for the CGR tube. Owing to the small detector area of the photodiode array, accurate alignment of the image sensor with the central axis of the beam is important. To enable correct positioning of the image sensor to the central axis of the x- ray beam, an adjustable x-y bench jig using a microscope stage was made. This permitted reasonably accurate two dimensional movement in 1mm graduations. The slit was manufactured from two tungsten pieces 1.90 mm thick in order to fully attenuate the primary x-ray beam and machined to dimensions specified in the AS/NZS 4274:1995 report, see figure 1. The blade-shaped tungsten pieces were placed into a specially machined aluminium holder with a tightly fitting channel to accommodate the blades, allowing movement in one dimension only. A 10 m spacer of mica crystal, chosen for its hardness, was obtained by splitting a mica block. The two tungsten blades were then brought together and secured in place by screws, separated only by the spacer, which was subsequently removed. Focal Spot // Dimensions in millimetres Figure 1. Dimensions of the manufactured slit diaphragm based on the AS/NZS 4274: 1995 report. Methods 8.4 o The need for a scaling factor The captured image was displayed on an NEC 16" monitor with a pixel pitch of 0.28 mm and capable of displaying 256 grey values. Due to the different number of pixels between the camera, (312 x 276), and frame grabber (768 x 512), as well as different aspect ratios (camera formatted to 4:3, frame grabber 3:2) it was necessary to determine a scaling factor. With a known camera-monitor scaling factor and radiographic magnification factor, actual focal spot dimensions could be calculated. A simple method based on measuring the image of a known object size was used. 64

3 Focal Spot Projected cross-hairs from Light Beam Diaphragm Image Sensor Assembly Adjustable x-y Bench (a) Reference Axis Axis of symmetry of the slit Slit Assembly (b) Figure 2. a) System set up for electronic image capture. The slit assembly is seen below the light beam assembly located along the reference axis of the beam, with the image sensor placed below, with its plane perpendicular to the same axis, with the use of the adjustable x-y table. b) Alignment of the slit diaphragm. A small circular hole of 0.6 mm diameter was made with a high speed drill in the Kodak phosphor. The phosphor was coupled to the photodiode array using silicone grease and exposed at 75 kv and 200 ma. The phosphor was then exposed several times at various focalcamera distances. Exposure time was limited to the standard video frame length of 40ms at 50Hz. The resultant image was sharpened by use of a high pass filter. Edge detection was performed by application of an edge detection algorithm available in Paint Shop Pro. The resultant image was greatly magnified on the computer screen to enable the counting of individual pixels and the number of pixels from edge to edge averaged over two orthogonal dimensions were counted for each image and hence the scale factor calculated. This yielded, 80 pixels = 0.6 mm, so that a scale factor of 1 pixel = mm was determined. The images were found to be circular to within 3 or 4 pixels, translating to an error of 5%, with the discrepancy possibly attributed to the edge detection algorithm. Although this is not an extremely accurate method to determine the scale factor, it is very simple and fast to perform. Image sensor measurements The measurement of the x-ray focal spot required that the slit be positioned accurately in the central beam. The centre of the projected cross-hairs provided on the light beam diaphragm was used as one end of the reference axis. The slit assembly rested in a purpose built stainless steel paddle, lined with 1mm of lead and supported on a retort stand. The paddle contained a circular opening for the slit assembly, which allowed for easy rotation between images. The slit was positioned at a distance of 30 cm from the focal spot, with a slit detector distance of 42 cm and a corresponding magnification of M = 1.4, along the reference axis so that the reference axis passed through the centre of the slit diaphragm, as shown in Figure 2. A spirit level was used to check that the face of the slit diaphragm was normal to the reference axis. The image sensor also required precise positioning, with its plane normal to the reference axis. The specially constructed x-y bench jig was used to position the image sensor assembly in directions parallel and perpendicular to the cathode-anode axis. The digital images obtained were difficult to see on the monitor due to the low exposure time of 40 ms, and hence the low light output of the phosphor. Exposures were made on both the large and small foci at settings of 75 kvp and various ma values, to test the effect of ma. Finally, the focal spot slit images were enhanced to facilitate analysis and observation. Dimensions of the focal spot were determined from the focal spot intensity profile and its corresponding FWHM in orthogonal orientations. The image dimensions being magnified were converted to the effective focal spot size width and length, by dividing by the magnification factor defined above. The AS/NZS 4274 standard defines the focal spot value f as a dimensionless quantity with reference to permissible values of width and length when measured with the slit method. An excerpt from the standard reproduced as Table 1, illustrates the focal spot value f along with its range of focal spot width and length values. Image enhancement With the camera-frame grabber configuration used, each photodiode is capable of collecting only a limited number of photons during the x-ray exposure, hence generating a small photocurrent per detector pixel. This limitation is primarily dictated by phosphor efficiency and frame length. Since the acquired images did not have a brightness range covering the full range of the digitiser, image enhancement techniques 65

4 Nominal focal spot value Focal spot dimensions Permissible value (mm) Width Length Table 1. Permissible values of focal spot dimensions for nominal focal spot values according to AS/NZS 4274:1995. were required to improve visual discrimination of small variations in sample density Appearance of the captured images was improved using established image enhancement techniques in succession, namely: frame averaging, histogram expansion and neighbourhood averaging Frame averaging was performed by acquiring and adding together N image frames. For images obtained at 50 and 100 ma, 16 frames were averaged. For images obtained at 200 and 300 ma, 8 frames were averaged. The added images were divided by N to re-scale the pixel data in order that brightness values did not exceed the maximum grey level value of 255. Frame averaging was accomplished using KHOROS, a specialist image analysis program. The averaging of frames in this way improved the signal to noise ratio. Due to the narrow brightness range of the averaged image, typically less than 7% of the full range of 0-255, the image contrast was poor. Linear expansion of contrast was carried out by assigning the darkest pixel value to black and the highest pixel value to white. To further improve the signal to noise ratio, a neighbourhood averaging operation was performed. This involved the computation of each output pixel as a function of a 5 x 5 Gaussian approximated kernel with = pixels. This kernel was implemented by the image analysis program, KHOROS. Although these processes improved the appearance of the captured images, such transformations are likely to distort raw data, hence the focal spot intensity profile obtained from each frame was used to determine focal spot dimensions. The combined effect of image processing could shift the apparent size of an image structure. A shift of 14 pixels would be required to cause an error of 0.1mm (from the scaling factor of 80 pixels = 0.6 mm). For a typical diagnostic x-ray machine of a nominal focal spot size of 0.4, the range of focal spot dimensions may vary by more than 0.2 mm without affecting the nominal focal spot size. Although the results obtained demonstrate that the possible shift of several pixels does not appear to be critical in determining the nominal focal spot size at the range measured, the accuracy of this technique is limited and more suited to the measurement of larger focal spot sizes at higher ma settings. Direct film measurements Focal spot slit images were also obtained using the direct film method with the radiographic film, Fuji RX-U, placed at a distance of cm from the slit diaphragm. The slit was positioned at a distance of 30 cm from the focal spot with a slit-film distance of 76.5 cm, giving a magnification of M = 2.55, along the reference axis so that the reference axis passed through the centre of the slit diaphragm, as shown in Figure 2. Exposures were made on both the large and small foci at settings of 75 kvp and various ma values as for the image sensor measurements. The films were processed (Protec M45 processor and Agfa developing chemicals at 34 o C). The focal spot width and length films were examined on a light box and measured using a 10x magnifying glass with a built in graticule having a scale with divisions of 0.1 mm which were read to the nearest 0.05 mm. All measurements were scaled by the appropriate magnification factor to determine the actual focal spot size. Film-screen measurements The radiographs for film-screen measurements were obtained by using the same set up as for the direct film method described above; however the Fuji RX-U film was used in conjunction with a cassette and Kodak X-Omatic fine screens. The viewing and measurement method was exactly the same as for the direct film. Results Focal spot size The dimensions of the focal spot were determined from the focal spot intensity profiles of the captured electronic images, and their corresponding FWHMs. Examples of captured focal spot slit images and the corresponding focal spot intensity profiles are shown in figure 3. Images were obtained at exposure settings of 75 kvp, 40 ms over the available range of ma values and a magnification of Table 2 gives the focal spot sizes determined from the captured electronic images. Note that the focal spot value f, referred to in tables 2 and 3 is a dimensionless quantity with reference to permissible values of width and length (see table 1). The analysed focal spot dimensions for the Odel Polaire x-ray machine show a nominal focal spot of 1.0 at 50 ma, for measured dimensions of 1.37 mm and 1.82 mm, for the width and length respectively. At 100 ma, a nominal focal spot of 1.0 was also determined, although the dimensions of width and length are somewhat greater at 1.39 mm and 1.92 mm, respectively. At the 200 ma and 300 ma tube current settings a nominal focal spot of 2.2 is obtained and again it is 66

5 Brightness level Figure 3. Cross sectional images of a focal spot determined using the image sensor showing a) width and b) length as well as the corresponding focal spot intensity profile (FSIP) for exposure settings of 75 kvp, 100 ma and 40 ms. The displayed images were enhanced with frame averaging, histogram expansion and neighbourhood averaging as described in the text. Note, the included scale is approximate only. Tube current (ma) a) Width b) Length FSIP mm Approximate Scale Focal-spot width (mm) FSIP Distance across image (mm) Focal-spot length (mm) Nominal ƒ Table 2. Focal spot sizes determined from images captured by the image sensor. Dimensions for the images were obtained from the FWHM of corresponding focal spot intensity profiles. noted, that the dimensions at 300 ma are somewhat greater than at 200 ma. The dimensions of the focal spot, and hence the nominal focal spot size, determined from the X-ray images for the direct film and screen-film methods are shown in Table 3. Images were obtained at 75 kv and tube current settings of 100 and 300 ma. Significantly increased exposure times were necessary in order to obtain visible images. Increased exposure times were obtained by repeated exposures. A magnification value of 2.55 was used as opposed to the magnification of 1.40 that was used with the image sensor device. The increased magnification value was obtained for a longer focus-to-film distance, which was not possible with the image sensor due to the small (4.6 mm x 6.1 mm) image sensor area. The measurements and subsequent analysis showed that nominal focal spot sizes of 1.0 and 2.2 are obtained for the direct film and screen-film methods. A comparison of the measured focal spot sizes for the three measurement methods are shown in Table 3. It was noted that the dimensions are smallest for the direct film method, increasing slightly for the screen-film and image sensorphosphor methods. Due to the range of permissible focal spot size widths and lengths used to define the nominal focal spot size (see Table 1), excellent agreement of the nominal focal spot size was obtained for the slit tool in combination with direct film, film-screen or the CMOS image sensor. The accuracy of the image sensor system appears to be about 10% when the actual width and length dimensions are compared to the direct film method. Discussion In conclusion, a method based on the slit tool was developed and tested under laboratory conditions for the measurement of focal spot size. We have demonstrated that a digital system based on an opto-coupled phosphor and VVL 1011C CMOS monochrome image sensor can acquire x-ray images of a radiographic focal spot. The acquired images were however of relatively poor quality and required enhancement to achieve satisfactory visual observation. Enhancement of the captured images substantially improved the image quality and enabled focal spot dimensions to be determined. However, the process of enhancement limits the measurement accuracy of the method by introducing image distortion such as pixel shifting. The greater dimensions obtained for the same machine-selected focal spot at the higher ma values is possibly attributable to the greater photon flux and hence a greater amount of light output from the phosphor at the higher ma setting. Although the dimensions are slightly greater, the nominal focal spot size is unaffected. Several improvements to the described method could be incorporated in a further model. Firstly, the camera appears to be a limiting component in the set-up, mainly due to its synchronous signals resulting in short frame grab times. Either programming of the frame grabber or purchase of an asynchronous camera would eliminate this major problem. Further, a larger photodiode array would permit the use of greater magnifications, allowing the imaging of small focal spot sizes. A reduced pixel size would also improve the system resolution. CMOS image sensors with a larger photodiode array and a reduced pixel size of 7.5 m x 7.5 m have recently become available. In addition, a reducing taper fibre optic coupler could be used, which would also improve resolution. The choice of phosphor appears to be good, although system resolution could be improved by using a pure Gd 2 O 2 S:Tb coating instead of using it in its screen form. For routine use, integration of the trigger circuit to the HV generator, requiring only a BNC type connection, would simplify the model. Finally, the use of a dedicated package capable of image enhancement with minimal image distortion and generation of focal spot intensity profiles would enhance the user interface. About 300 exposures of 0.5s average duration were made over the course of the project. No adverse radiation damage effects were noted. 67

6 Tube current (ma) Focal spot width (mm) Focal spot length (mm) Nominal focal spot size Method Direct film Screen film Image sensor a) Small focal spot Tube surrent (ma) Focal spot width (mm) Focal spot length (mm) NominalfFocal spot size Method Direct film Screen film Image sensor b) Large focal spot Table 3. Comparison of focal spot size slit image measurements obtained under direct, screen-film and image sensor conditions for a) the small focal spot at 100 ma and b) the large focal spot at 300 ma. Acknowledgements We would like to thank Dr R I Price and Dr R A Fox for comments on the manuscript. References 1. Lu, T., Udpa, S. S. and Udpa, L. Tomographic reconstruction using optoelectronic architecture, IEEE Volume 4, Muirhead, I. T., Developments in CMOS camera technology, IEE Lake, D., Why CMOS imagers are the rage: A revolution in the making, Advanced Imaging Magazine, 12-18, Fossum, E. R., CMOS image sensors: Electronic camera on a chip, IEEE Denyer, P. B., Renshaw, D., Wang Guoyu and Lu Mingying, A single chip sensor and image processor for fingerprint verification, Proceedings of IEEE Custom Integrated Circuits Conference, San Diego, California, May 9-12, Speller, R. D., Martinez-Davalos, A., and Farquharson, M., A CCD based focal spot camera, Phys. Med. Biol. 40, , Rong, X. J., Krugh, K. T., Shepard, S. J. and Geiser, W. R. Measurement of focal spot size with slit camera using computed radiography and flat-panel based digital detectors, Med. Phys. 30(7): , Everson, J. D. and Gray, J. E., Focal spot measurement: Comparison of slit, pinhole and star resolution pattern techniques, Radiology, 165, , Australian/New Zealand Standard (AS/NZS), X-ray tube assemblies for medical diagnosis-characteristics of focal spots, AS/NZS 4274, Nickoloff, E. L., Donnelly, E., Eve, L. and Atherton, J.V., Mammographic resolution: Influence of focal spot intensity distribution and geometry, Med. Phys., 17(3), , Schiabel, H., Ventura, A. and Frere, A. F., A formal study of lateral magnification and its influence on mammographic imaging sharpness, Med. Phys., 21(2), , Yaffe, M. J., and Rowlands, J. A., X-ray detectors for digital radiography, Phys. Med. Biol. 42, 1-39, Kandarakis, I., Cavouras, D., Panayiotakis, G., Agelis, T., Nomicos, C. D., Giakoumakis G. E, X-ray induced luminescence and spatial resolution of La 2 O 2 S:Tb phosphor screens, Phys. Med. Biol, 41, , Giakoumakis, G.E., Nomicos, C.D., Yiakoumakis, E.N., Katsarioti, M.C., Kalikatsos, J.A., Rovithi, M., Panayiotakis, G. S., Evangelou, E. K., Y 2 O 2 S:Eu phosphor screens evaluation, Med. Phys. 20(1), 79-83, Giakoumakis, G. E., Nomicos, C. D., and Sandilos, P. X., Absolute efficiency of Gd 2 O 2 S:Tb screens under fluoroscopy conditions, Phys. Med. Biol, 34, 673-8, Giakoumakis, G. E. and Nomicos, C. D., Absolute efficiency of Y 2 O 2 S:Tb screens under fluoroscopy conditions, J.Appl.Phys. 58(7), , Russ, J., The image processing handbook, 2 nd Ed., CRC Press, Boca Raton, Florida, Bankman, I. N., Handbook of Medical Imaging, Academic Press, San Diego, California, de Graff, C. N., and Viergever, M.A., Information processing in medical imaging, Plenum 20. Press, New York, Frei, W., Image enhancement by image hyperbolization, Comp. Graph. Image Process, 6, , Hall, E. H., Almost uniform distributions from image enhancement, IEEE Trans. Comp. C-23(2), ,

Investigation of the line-pair pattern method for evaluating mammographic focal spot performance

Investigation of the line-pair pattern method for evaluating mammographic focal spot performance Investigation of the line-pair pattern method for evaluating mammographic focal spot performance Mitchell M. Goodsitt, a) Heang-Ping Chan, and Bob Liu Department of Radiology, University of Michigan, Ann

More information

Amorphous Selenium Direct Radiography for Industrial Imaging

Amorphous Selenium Direct Radiography for Industrial Imaging DGZfP Proceedings BB 67-CD Paper 22 Computerized Tomography for Industrial Applications and Image Processing in Radiology March 15-17, 1999, Berlin, Germany Amorphous Selenium Direct Radiography for Industrial

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

Observer Performance of Reduced X-Ray Images on Liquid Crystal Displays

Observer Performance of Reduced X-Ray Images on Liquid Crystal Displays Original Paper Forma, 29, S45 S51, 2014 Observer Performance of Reduced X-Ray Images on Liquid Crystal Displays Akiko Ihori 1, Chihiro Kataoka 2, Daigo Yokoyama 2, Naotoshi Fujita 3, Naruomi Yasuda 4,

More information

17th World Conference on Nondestructive Testing, Oct 2008, Shanghai, China

17th World Conference on Nondestructive Testing, Oct 2008, Shanghai, China 17th World Conference on Nondestructive Testing, 25-28 Oct 2008, Shanghai, China Real-time Radiographic Non-destructive Inspection for Aircraft Maintenance Xin Wang 1, B. Stephen Wong 1, Chen Guan Tui

More information

Phase Imaging Using Focused Polycapillary Optics

Phase Imaging Using Focused Polycapillary Optics Phase Imaging Using Focused Polycapillary Optics Sajid Bashir, Sajjad Tahir, Jonathan C. Petruccelli, C.A. MacDonald Dept. of Physics, University at Albany, Albany, New York Abstract Contrast in conventional

More information

Fig.2: Scanner VistaScan for image plates

Fig.2: Scanner VistaScan for image plates RADIOGRAPHIC INSPECTION OF WELDINGS BY DIGITAL SENSORS H. Thiele, H.-J. Friemel RADIS GmbH, Johanniskirchen, Germany Abstract: The newly available digital sensors for radiographic inspection are suitable

More information

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 2 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 COMPUTERIZED IMAGING Section I: Chapter 2 RADT

More information

Mammography: Physics of Imaging

Mammography: Physics of Imaging Mammography: Physics of Imaging Robert G. Gould, Sc.D. Professor and Vice Chair Department of Radiology and Biomedical Imaging University of California San Francisco, California Mammographic Imaging: Uniqueness

More information

The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System

The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System Manus

More information

X-RAY COMPUTED TOMOGRAPHY

X-RAY COMPUTED TOMOGRAPHY X-RAY COMPUTED TOMOGRAPHY Bc. Jan Kratochvíla Czech Technical University in Prague Faculty of Nuclear Sciences and Physical Engineering Abstract Computed tomography is a powerful tool for imaging the inner

More information

10/26/2015. Study Harder

10/26/2015. Study Harder This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura Study Harder CR detection is inefficient, inferior to film screen

More information

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over a lifetime Breast cancer screening programs rely on

More information

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893 Digital Radiography D. J. Hall, Ph.D. x20893 djhall@ucsd.edu Background Common Digital Modalities Digital Chest Radiograph - 4096 x 4096 x 12 bit CT - 512 x 512 x 12 bit SPECT - 128 x 128 x 8 bit MRI -

More information

Acquisition, Processing and Display

Acquisition, Processing and Display Acquisition, Processing and Display Terri L. Fauber, R.T. (R)(M) Department of Radiation Sciences School of Allied Health Professions Virginia Commonwealth University Topics Image Characteristics Image

More information

Setting up digital imaging department!

Setting up digital imaging department! Outline Setting up digital imaging department! From screen/film to digital radiography PACS/Tele radiology Setting up digital department Digital Imaging Napapong Pongnapang, Ph.D. Department of Radiological

More information

10/3/2012. Study Harder

10/3/2012. Study Harder This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura Study Harder CR detection is inefficient, inferior to film screen

More information

FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION

FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION Revised November 15, 2017 INTRODUCTION The simplest and most commonly described examples of diffraction and interference from two-dimensional apertures

More information

Hardware for High Energy Applications 30 October 2009

Hardware for High Energy Applications 30 October 2009 Paper No. 003 09 Hardware for High Energy Applications 30 October 2009 This document was created by the Federal Working Group on Industrial Digital Radiography. Reproduction is authorized. Federal Working

More information

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS JAMES A. TOMLINSON, M.S. Diagnostic Radiological Physicist American Board of Radiology Certified Medical Physics Consultants, Inc. Bio 28 yrs experience 100%

More information

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE

More information

Collimation Assessment Using GAFCHROMIC XR-M2

Collimation Assessment Using GAFCHROMIC XR-M2 Collimation Assessment Using GAFCHROMIC XR-M2 I. Introduction A method of collimation assessment for GE Senographe full-field digital mammography (FFDM) systems is described that uses a self-developing

More information

ANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER

ANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 325 ANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER ABSTRACT William Chang, Jonathan Kerner, and Edward

More information

SYLLABUS. TITLE: Equipment Operation I. DEPARTMENT: Radiologic Technology

SYLLABUS. TITLE: Equipment Operation I. DEPARTMENT: Radiologic Technology CODE: RADT 156 INSTITUTE: Health Science TITLE: Equipment Operation I DEPARTMENT: Radiologic Technology COURSE DESCRIPTION: This course covers the principles of equipment operation and maintenance of radiographic

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 16371-1 First edition 2011-10-01 Non-destructive testing Industrial computed radiography with storage phosphor imaging plates Part 1: Classification of systems Essais non destructifs

More information

HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE

HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE Takeyuki Hashimoto 1), Morio Onoe 2), Hiroshi Nakamura 3), Tamon Inouye 4), Hiromichi Jumonji 5), Iwao Takahashi 6); 1)Yokohama Soei

More information

MINIATURE X-RAY SOURCES AND THE EFFECTS OF SPOT SIZE ON SYSTEM PERFORMANCE

MINIATURE X-RAY SOURCES AND THE EFFECTS OF SPOT SIZE ON SYSTEM PERFORMANCE 228 MINIATURE X-RAY SOURCES AND THE EFFECTS OF SPOT SIZE ON SYSTEM PERFORMANCE D. CARUSO, M. DINSMORE TWX LLC, CONCORD, MA 01742 S. CORNABY MOXTEK, OREM, UT 84057 ABSTRACT Miniature x-ray sources present

More information

Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC

Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC R. Bellazzini a,b, G. Spandre a*, A. Brez a, M. Minuti a, M. Pinchera a and P. Mozzo b a INFN Pisa

More information

X-ray light valve (XLV): a novel detectors technology for digital mammography*

X-ray light valve (XLV): a novel detectors technology for digital mammography* X-ray light valve (XLV): a novel detectors technology for digital mammography* Sorin Marcovici, Vlad Sukhovatkin, Peter Oakham XLV Diagnostics Inc., Thunder Bay, ON P7A 7T1, Canada ABSTRACT A novel method,

More information

Evaluating Commercial Scanners for Astronomical Images. The underlying technology of the scanners: Pixel sizes:

Evaluating Commercial Scanners for Astronomical Images. The underlying technology of the scanners: Pixel sizes: Evaluating Commercial Scanners for Astronomical Images Robert J. Simcoe Associate Harvard College Observatory rjsimcoe@cfa.harvard.edu Introduction: Many organizations have expressed interest in using

More information

Digital Detector Array Image Quality for Various GOS Scintillators

Digital Detector Array Image Quality for Various GOS Scintillators Digital Detector Array Image Quality for Various GOS Scintillators More info about this article: http://www.ndt.net/?id=22768 Brian S. White 1, Mark E. Shafer 2, William H. Russel 3, Eric Fallet 4, Jacques

More information

CR Basics and FAQ. Overview. Historical Perspective

CR Basics and FAQ. Overview. Historical Perspective Page: 1 of 6 CR Basics and FAQ Overview Computed Radiography is a term used to describe a system that electronically records a radiographic image. Computed Radiographic systems use unique image receptors

More information

Charged Coupled Device (CCD) S.Vidhya

Charged Coupled Device (CCD) S.Vidhya Charged Coupled Device (CCD) S.Vidhya 02.04.2016 Sensor Physical phenomenon Sensor Measurement Output A sensor is a device that measures a physical quantity and converts it into a signal which can be read

More information

Digital radiography: Practical advantages of Digital Radiography. Practical Advantages in image quality

Digital radiography: Practical advantages of Digital Radiography. Practical Advantages in image quality Digital radiography: Digital radiography is set to become the most common form of processing radiographic images in the next 10 years. This is due to a number of practical and image quality issues. Practical

More information

CMS Note Mailing address: CMS CERN, CH-1211 GENEVA 23, Switzerland

CMS Note Mailing address: CMS CERN, CH-1211 GENEVA 23, Switzerland Available on CMS information server CMS NOTE 1998/16 The Compact Muon Solenoid Experiment CMS Note Mailing address: CMS CERN, CH-1211 GENEVA 23, Switzerland January 1998 Performance test of the first prototype

More information

Preliminary Modulation Transfer Function Study on Amorphous Silicon Flat Panel System for Industrial Digital Radiography

Preliminary Modulation Transfer Function Study on Amorphous Silicon Flat Panel System for Industrial Digital Radiography ECNDT 26 - Poster 17 Preliminary Modulation Transfer Function Study on Amorphous Silicon Flat Panel System for Industrial Digital Radiography Khairul Anuar MOHD SALLEH, Ab. Razak HAMZAH and Mohd Ashhar

More information

PERFORMANCE CHARACTERIZATION OF AMORPHOUS SILICON DIGITAL DETECTOR ARRAYS FOR GAMMA RADIOGRAPHY

PERFORMANCE CHARACTERIZATION OF AMORPHOUS SILICON DIGITAL DETECTOR ARRAYS FOR GAMMA RADIOGRAPHY 12 th A-PCNDT 2006 Asia-Pacific Conference on NDT, 5 th 10 th Nov 2006, Auckland, New Zealand PERFORMANCE CHARACTERIZATION OF AMORPHOUS SILICON DIGITAL DETECTOR ARRAYS FOR GAMMA RADIOGRAPHY Rajashekar

More information

IMAGE SENSOR SOLUTIONS. KAC-96-1/5" Lens Kit. KODAK KAC-96-1/5" Lens Kit. for use with the KODAK CMOS Image Sensors. November 2004 Revision 2

IMAGE SENSOR SOLUTIONS. KAC-96-1/5 Lens Kit. KODAK KAC-96-1/5 Lens Kit. for use with the KODAK CMOS Image Sensors. November 2004 Revision 2 KODAK for use with the KODAK CMOS Image Sensors November 2004 Revision 2 1.1 Introduction Choosing the right lens is a critical aspect of designing an imaging system. Typically the trade off between image

More information

LECTURE 1 The Radiographic Image

LECTURE 1 The Radiographic Image LECTURE 1 The Radiographic Image Prepared by:- KAMARUL AMIN ABDULLAH @ ABU BAKAR UiTM Faculty of Health Sciences Medical Imaging Department 11/23/2011 KAMARUL AMIN (C) 1 Lesson Objectives At the end of

More information

Do you have any other questions? Please call us at (Toll Free) or , or

Do you have any other questions? Please call us at (Toll Free) or , or INSTRUCTIONS Read the appropriate course/ textbook. This is an open book test. A score of 75% or higher is needed to receive CE credit. You will have a maximum of three attempts to pass this course. Please

More information

DIGITAL RADIOGRAPHY. Digital radiography is a film-less technology used to record radiographic images.

DIGITAL RADIOGRAPHY. Digital radiography is a film-less technology used to record radiographic images. DIGITAL RADIOGRAPHY Digital radiography is a film-less technology used to record radiographic images. 1 The purpose of digital imaging is to generate images that can be used in the diagnosis and assessment

More information

Copyright 2000 Society of Photo Instrumentation Engineers.

Copyright 2000 Society of Photo Instrumentation Engineers. Copyright 2000 Society of Photo Instrumentation Engineers. This paper was published in SPIE Proceedings, Volume 4043 and is made available as an electronic reprint with permission of SPIE. One print or

More information

RADIOGRAPHY TERMS TO KNOW SELF STUDY DENTALELLE TUTORING

RADIOGRAPHY TERMS TO KNOW SELF STUDY DENTALELLE TUTORING RADIOGRAPHY TERMS TO KNOW SELF STUDY DENTALELLE TUTORING PLEASE NOTE You DO NOT need to study these for the board exam if this is why you bought our Radiography course, however if you come across any terms

More information

Detectors for microscopy - CCDs, APDs and PMTs. Antonia Göhler. Nov 2014

Detectors for microscopy - CCDs, APDs and PMTs. Antonia Göhler. Nov 2014 Detectors for microscopy - CCDs, APDs and PMTs Antonia Göhler Nov 2014 Detectors/Sensors in general are devices that detect events or changes in quantities (intensities) and provide a corresponding output,

More information

Shad-o-Box HS Product Family

Shad-o-Box HS Product Family Shad-o-Box HS Product Family DATASHEET Overview Key Features Large active area up to 10x15 cm Up to 10 lp/mm resolution Gigabit Ethernet interface (Camera Link optional) 14-bit digital video output Energy

More information

Image acquisition. In both cases, the digital sensing element is one of the following: Line array Area array. Single sensor

Image acquisition. In both cases, the digital sensing element is one of the following: Line array Area array. Single sensor Image acquisition Digital images are acquired by direct digital acquisition (digital still/video cameras), or scanning material acquired as analog signals (slides, photographs, etc.). In both cases, the

More information

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura New Technology https://www.youtube.com/watch?v=ptkzznazb 7U COMPUTED

More information

Rad-icon Imaging Corp A Division of DALSA Corporation

Rad-icon Imaging Corp A Division of DALSA Corporation Rad-icon Imaging Corp A Division of DALSA Corporation el: 408-486-0886 Fax: 408-486-0882 www.rad-icon.com PRELIMINARY DAA SHEE SkiaGraph 8 Very Large Area X-Ray Camera Key Features: Active area of 20 cm

More information

In-line measurements of rolling stock macro-geometry

In-line measurements of rolling stock macro-geometry Optical measuring systems for plate mills Advances in camera technology have enabled a significant enhancement of dimensional measurements in plate mills. Slabs and as-rolled and cut-to-size plates can

More information

High Energy Digital Radiography & 3D-CT for Industrial Systems

High Energy Digital Radiography & 3D-CT for Industrial Systems DIR 2007 - International Symposium on Digital industrial Radiology and Computed Tomography, June 25-27, 2007, Lyon, France High Energy Digital Radiography & 3D-CT for Industrial Systems Non-Destructive

More information

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question.

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question. RA110 test 3 Multiple Choice Identify the letter of the choice that best completes the statement or answers the question. 1. An object 35 cm in width is radiographed at 100 cm SID and at a 50 cm SOD. What

More information

Keysight Technologies Why Magnification is Irrelevant in Modern Scanning Electron Microscopes. Application Note

Keysight Technologies Why Magnification is Irrelevant in Modern Scanning Electron Microscopes. Application Note Keysight Technologies Why Magnification is Irrelevant in Modern Scanning Electron Microscopes Application Note Introduction From its earliest inception, the Scanning Electron Microscope (SEM) has been

More information

Very short introduction to light microscopy and digital imaging

Very short introduction to light microscopy and digital imaging Very short introduction to light microscopy and digital imaging Hernan G. Garcia August 1, 2005 1 Light Microscopy Basics In this section we will briefly describe the basic principles of operation and

More information

X-ray investigation of crystal structures / Laue method with digital X-ray detector (XRIS) (Item No.: P )

X-ray investigation of crystal structures / Laue method with digital X-ray detector (XRIS) (Item No.: P ) X-ray investigation of crystal structures / Laue method with digital X-ray detector (XRIS) (Item No.: P2541602) Curricular Relevance Area of Expertise: Physik Education Level: Hochschule Topic: Moderne

More information

ME 6406 MACHINE VISION. Georgia Institute of Technology

ME 6406 MACHINE VISION. Georgia Institute of Technology ME 6406 MACHINE VISION Georgia Institute of Technology Class Information Instructor Professor Kok-Meng Lee MARC 474 Office hours: Tues/Thurs 1:00-2:00 pm kokmeng.lee@me.gatech.edu (404)-894-7402 Class

More information

Dental Radiography. One of the problems of dental radiography is having different dimensions than normal.

Dental Radiography. One of the problems of dental radiography is having different dimensions than normal. The prototype receptor (the recording medium) most commonly used in dental radiography is the radiographic film. However, there are many other new more efficient receptors than the formed one that can

More information

3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development

3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development Teaching Radiographic Technique in a Digital Imaging Paradigm Objectives 1. Discuss the historical development of digital imaging. Dawn Couch Moore, M.M.Sc., RT(R) Assistant Professor and Director Emory

More information

Outline. Digital Radiography. Understanding Digital Modalities: Image Quality and Dose. Image Quality. Dose Control

Outline. Digital Radiography. Understanding Digital Modalities: Image Quality and Dose. Image Quality. Dose Control Understanding Digital Modalities: Image Quality and Dose S. Jeff Shepard, M.S. University of Texas M. D. Anderson Cancer Center Houston, Texas Special Acknowledgement: Stephen K. Thompson, M.S. William

More information

Criteria for Optical Systems: Optical Path Difference How do we determine the quality of a lens system? Several criteria used in optical design

Criteria for Optical Systems: Optical Path Difference How do we determine the quality of a lens system? Several criteria used in optical design Criteria for Optical Systems: Optical Path Difference How do we determine the quality of a lens system? Several criteria used in optical design Computer Aided Design Several CAD tools use Ray Tracing (see

More information

X-RAYS - NO UNAUTHORISED ENTRY

X-RAYS - NO UNAUTHORISED ENTRY Licencing of premises Premises Refer Guidelines A radiation warning sign and warning notice, X-RAYS - NO UNAUTHORISED ENTRY must be displayed at all entrances leading to the rooms where x-ray units are

More information

Experimental Analysis of Luminescence in Printed Materials

Experimental Analysis of Luminescence in Printed Materials Experimental Analysis of Luminescence in Printed Materials A. D. McGrath, S. M. Vaezi-Nejad Abstract - This paper is based on a printing industry research project nearing completion [1]. While luminescent

More information

HIGH SPEED, HIGH RESOLUTION AND LOW COST DIGITAL RADIOGRAPHY

HIGH SPEED, HIGH RESOLUTION AND LOW COST DIGITAL RADIOGRAPHY HIGH SPEED, HIGH RESOLUTION AND LOW COST DIGITAL RADIOGRAPHY AND COMPUTED TOMOGRAPHY SYSTEM Kasiviswanathan Rangarajan1,2 and T. Jensen 1 Department of Computer Engineering 2 Center for Nondestructive

More information

PROCEEDINGS OF A SYMPOSIUM HELD AT THE CAVENDISH LABORATORY, CAMBRIDGE, Edited by

PROCEEDINGS OF A SYMPOSIUM HELD AT THE CAVENDISH LABORATORY, CAMBRIDGE, Edited by X - R A Y M I C R O S C O P Y A N D M I C R O R A D I O G R A P H Y PROCEEDINGS OF A SYMPOSIUM HELD AT THE CAVENDISH LABORATORY, CAMBRIDGE, 1956 Edited by V. E. COSSLETT Cavendish Laboratory, University

More information

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 14 CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 2.1 INTRODUCTION kv-cbct integrated with linear accelerators as a tool for IGRT, was developed to

More information

Tomographic 3D-Radiometry for the Visualisation and Measurement of the Defects of Girth Seams

Tomographic 3D-Radiometry for the Visualisation and Measurement of the Defects of Girth Seams ECNDT 2006 - We.3.2.3 Tomographic 3D-Radiometry for the Visualisation and Measurement of the Defects of Girth Seams Bernhard REDMER, Uwe EWERT Federal Institute of Materials Research and Testing (BAM),

More information

SPRINGFIELD TECHNICAL COMMUNITY COLLEGE ACADEMIC AFFAIRS

SPRINGFIELD TECHNICAL COMMUNITY COLLEGE ACADEMIC AFFAIRS SPRINGFIELD TECHNICAL COMMUNITY COLLEGE ACADEMIC AFFAIRS Course Number: RADG 112 Department: Radiography Course Title: Image Production & Eval. Semester: Spring Year: 1997 Objectives/ Unit One: Introduction

More information

Optical Coherence: Recreation of the Experiment of Thompson and Wolf

Optical Coherence: Recreation of the Experiment of Thompson and Wolf Optical Coherence: Recreation of the Experiment of Thompson and Wolf David Collins Senior project Department of Physics, California Polytechnic State University San Luis Obispo June 2010 Abstract The purpose

More information

A new Photon Counting Detector: Intensified CMOS- APS

A new Photon Counting Detector: Intensified CMOS- APS A new Photon Counting Detector: Intensified CMOS- APS M. Belluso 1, G. Bonanno 1, A. Calì 1, A. Carbone 3, R. Cosentino 1, A. Modica 4, S. Scuderi 1, C. Timpanaro 1, M. Uslenghi 2 1- I.N.A.F.-Osservatorio

More information

White Paper High Dynamic Range Imaging

White Paper High Dynamic Range Imaging WPE-2015XI30-00 for Machine Vision What is Dynamic Range? Dynamic Range is the term used to describe the difference between the brightest part of a scene and the darkest part of a scene at a given moment

More information

White Paper: Modifying Laser Beams No Way Around It, So Here s How

White Paper: Modifying Laser Beams No Way Around It, So Here s How White Paper: Modifying Laser Beams No Way Around It, So Here s How By John McCauley, Product Specialist, Ophir Photonics There are many applications for lasers in the world today with even more on the

More information

Investigation of Effective DQE (edqe) parameters for a flat panel detector

Investigation of Effective DQE (edqe) parameters for a flat panel detector Investigation of Effective DQE (edqe) parameters for a flat panel detector Poster No.: C-1892 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Exhibit D. Bor 1, S. Cubukcu 1, A. Yalcin 1, O.

More information

Current technology in digital image production (CR/DR and other modalities) Jaroonroj Wongnil 25 Mar 2016

Current technology in digital image production (CR/DR and other modalities) Jaroonroj Wongnil 25 Mar 2016 Current technology in digital image production (CR/DR and other modalities) Jaroonroj Wongnil 25 Mar 2016 Current technology in digital image production (CR/DR and other modalities) 2/ Overview Digital

More information

Film Replacement in Radiographic Weld Inspection The New ISO Standard

Film Replacement in Radiographic Weld Inspection The New ISO Standard BAM Berlin Film Replacement in Radiographic Weld Inspection The New ISO Standard 17636-2 Uwe Ewert, Uwe Zscherpel, Mirko Jechow Requests and information to: uwez@bam.de 1 Outline - The 3 essential parameters

More information

A new Photon Counting Detector: Intensified CMOS- APS

A new Photon Counting Detector: Intensified CMOS- APS A new Photon Counting Detector: Intensified CMOS- APS M. Belluso 1, G. Bonanno 1, A. Calì 1, A. Carbone 3, R. Cosentino 1, A. Modica 4, S. Scuderi 1, C. Timpanaro 1, M. Uslenghi 2 1-I.N.A.F.-Osservatorio

More information

Examination of Pipe Welds by Image Plate Based Computed Radiography System

Examination of Pipe Welds by Image Plate Based Computed Radiography System Examination of Pipe Welds by Image Plate Based Computed Radiography System Sanjoy Das, M.S.Rana, Benny Sebastian, D. Mukherjee and K.K. Abdulla Atomic Fuels Division Bhabha Atomic Research Centre Mumbai

More information

A Laser-Based Thin-Film Growth Monitor

A Laser-Based Thin-Film Growth Monitor TECHNOLOGY by Charles Taylor, Darryl Barlett, Eric Chason, and Jerry Floro A Laser-Based Thin-Film Growth Monitor The Multi-beam Optical Sensor (MOS) was developed jointly by k-space Associates (Ann Arbor,

More information

Seminar 8. Radiology S8 1

Seminar 8. Radiology S8 1 Seminar 8 Radiology Medical imaging. X-ray image formation. Energizing and controlling the X-ray tube. Image detectors. The acquisition of analog and digital images. Digital image processing. Selected

More information

Photons and solid state detection

Photons and solid state detection Photons and solid state detection Photons represent discrete packets ( quanta ) of optical energy Energy is hc/! (h: Planck s constant, c: speed of light,! : wavelength) For solid state detection, photons

More information

Real Time Linear Array Imaging. Brian Caccamise

Real Time Linear Array Imaging. Brian Caccamise Real Time Linear Array Imaging Brian Caccamise 1 Real Time Linear Array Imaging What is Real Time Linear Array Imaging? Or Real Time Radiography (RTR)? 2 Real Time Linear Array Imaging It s Not This! Shoe

More information

Small Animal Radiographic Techniques and Positioning COPYRIGHTED MATERIAL

Small Animal Radiographic Techniques and Positioning COPYRIGHTED MATERIAL Small Animal Radiographic Techniques and Positioning COPYRIGHTED MATERIAL Section 1 Theory and Equipment 1 Introduction to Digital Imaging Small animal radiography has changed dramatically in the past

More information

MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE

MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE A. Miceli ab, R. Thierry a, A. Flisch a, U. Sennhauser a, F. Casali b a Empa - Swiss Federal Laboratories for

More information

Veraviewepocs 2D High Speed Panoramic X-Ray Crystal Clear Images with Reduced Radiation

Veraviewepocs 2D High Speed Panoramic X-Ray Crystal Clear Images with Reduced Radiation Diagnostic and Imaging Equipment Treatment Units Handpieces and Instruments Endodontic Systems Laser Equipment Laboratory Devices Veraviewepocs 2D High Speed Panoramic X-Ray Crystal Clear Images with Reduced

More information

Bias errors in PIV: the pixel locking effect revisited.

Bias errors in PIV: the pixel locking effect revisited. Bias errors in PIV: the pixel locking effect revisited. E.F.J. Overmars 1, N.G.W. Warncke, C. Poelma and J. Westerweel 1: Laboratory for Aero & Hydrodynamics, University of Technology, Delft, The Netherlands,

More information

Digital Radiography : Flat Panel

Digital Radiography : Flat Panel Digital Radiography : Flat Panel Flat panels performances & operation How does it work? - what is a sensor? - ideal sensor Flat panels limits and solutions - offset calibration - gain calibration - non

More information

GE Healthcare. Senographe 2000D Full-field digital mammography system

GE Healthcare. Senographe 2000D Full-field digital mammography system GE Healthcare Senographe 2000D Full-field digital mammography system Digital has arrived. The Senographe 2000D Full-Field Digital Mammography (FFDM) system gives you a unique competitive advantage. That

More information

Attikon, Rimini 1, , Athens, Greece , Athens, Greece , Athens, Greece

Attikon, Rimini 1, , Athens, Greece , Athens, Greece , Athens, Greece Radiation Protection Dosimetry (2005), Vol. 117, No. 1 3, pp. 291 297 doi:10.1093/rpd/nci742 Advance Access published on February 7, 2006 QUALITY ASSURANCE (QA) PROCEDURES FOR SOFTWARE: EVALUATION OF AN

More information

Moving from film to digital: A study of digital x-ray benefits, challenges and best practices

Moving from film to digital: A study of digital x-ray benefits, challenges and best practices Moving from film to digital: A study of digital x-ray benefits, challenges and best practices H.U. Pöhler 1 and N. D Ademo 2 DÜRR NDT GmbH & Co. KG, Höpfigheimer Straße 22, Bietigheim-Bissingen, 74321,

More information

Computed Radiography of Resistance Temperature Sensor for Indian PHWR

Computed Radiography of Resistance Temperature Sensor for Indian PHWR National Seminar & Exhibition on Non-Destructive Evaluation, NDE 2014, Pune, December 4-6, 2014 (NDE-India 2014) Vol.20 No.6 (June 2015) - The e-journal of Nondestructive Testing - ISSN 1435-4934 www.ndt.net/?id=17831

More information

Photomultiplier Tube

Photomultiplier Tube Nuclear Medicine Uses a device known as a Gamma Camera. Also known as a Scintillation or Anger Camera. Detects the release of gamma rays from Radionuclide. The radionuclide can be injected, inhaled or

More information

Development of the LBNL Positron Emission Mammography Camera

Development of the LBNL Positron Emission Mammography Camera Development of the LBNL Positron Emission Mammography Camera J.S. Huber, Member, IEEE, W.S. Choong, Member, IEEE, J. Wang, Member, IEEE, J.S. Maltz, Member, IEEE, J. Qi, Member, IEEE, E. Mandelli, Member,

More information

Gas scintillation Glass GEM detector for high-resolution X-ray imaging and CT

Gas scintillation Glass GEM detector for high-resolution X-ray imaging and CT Gas scintillation Glass GEM detector for high-resolution X-ray imaging and CT Takeshi Fujiwara 1, Yuki Mitsuya 2, Hiroyuki Takahashi 2, and Hiroyuki Toyokawa 2 1 National Institute of Advanced Industrial

More information

MILADY. Product Data. Page 1 of 8

MILADY. Product Data. Page 1 of 8 Page 1 of 8 The MILADY Mammographic Unit offers the best quality-to-price ratio to our customers worldwide. The unit advanced technology together with the application of industrial production standards,

More information

ROBOT VISION. Dr.M.Madhavi, MED, MVSREC

ROBOT VISION. Dr.M.Madhavi, MED, MVSREC ROBOT VISION Dr.M.Madhavi, MED, MVSREC Robotic vision may be defined as the process of acquiring and extracting information from images of 3-D world. Robotic vision is primarily targeted at manipulation

More information

ISO INTERNATIONAL STANDARD. Non-destructive testing of welds Radiographic testing Part 1: X- and gamma-ray techniques with film

ISO INTERNATIONAL STANDARD. Non-destructive testing of welds Radiographic testing Part 1: X- and gamma-ray techniques with film INTERNATIONAL STANDARD ISO 17636-1 First edition 2013-01-15 Non-destructive testing of welds Radiographic testing Part 1: X- and gamma-ray techniques with film Contrôle non destructif des assemblages soudés

More information

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Acceptance Testing of a Digital Breast Tomosynthesis Unit Acceptance Testing of a Digital Breast Tomosynthesis Unit 2012 AAPM Spring Clinical Meeting Jessica Clements, M.S., DABR Objectives Review of technology and clinical advantages Acceptance Testing Procedures

More information

X-ray Tube and Generator Basic principles and construction

X-ray Tube and Generator Basic principles and construction X-ray Tube and Generator Basic principles and construction Dr Slavik Tabakov - Production of X-rays and Patient Dose OBJECTIVES - X-ray tube construction - Anode - types, efficiency - Classical X-ray generator

More information

Optimization of Digital Mammography Resolution Using Magnification Technique in Computed Radiography 1

Optimization of Digital Mammography Resolution Using Magnification Technique in Computed Radiography 1 Optimization of Digital Mammography Resolution Using Magnification Technique in Computed Radiography 1 Gham Hur, M.D., Yoon Joon Hwang, M.D., Soon Joo Cha, M.D., Su Young Kim, M.D., Yong Hoon Kim, M.D.

More information

FMT18 FLOOR MOUNTED SYSTEM

FMT18 FLOOR MOUNTED SYSTEM mas Time AEC 320 kvp 64 mas 320 ma 320 ma 320 DEN 0.0 mm Cu 17 in X 17 in 72.0 in FMT18 FLOOR MOUNTED SYSTEM with Synchronized Tracking System Overview Clinical Efficiency The FMT18 System was designed

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 07-649 CDRH Fluoroscopic Phantom Users Manual March 2005 Manual No. 07-649-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks

More information

Beam-Restricting Devices

Beam-Restricting Devices Beam-Restricting Devices Three factors contribute to an increase in scatter radiation: Increased kvp Increased Field Size Increased Patient or Body Part Size. X-ray Interactions a some interact with the

More information