1 \OL. 117, No. A PHANTOM FOR THE EVALUATION OF TECHNIQUES AND EQUIPMENT USED FOR ROENTGENOGRAPHY OF THE CHEST* By E. DALE TROUT, D.SC., and JOHN P. KELLEY, B.S. CORVALLIS, P UBLIC Law passed by the Congress in 1969 and often referred to as the black lung law established provisions for roentgenographic examinations of approximately ioo,ooo active coal miners. These examinations were to be carried out in approved facilities of which there are presently about 270. If there is to be any uniformity of roentgenographic quality, some test object that can be sent to these facilities is needed. It seems reasonable to assume that as such programs move into other areas of industrial medicine, there will be a sustained need for some means of evaluating roentgenologic procedures used in such studies. Surely, there must be some common denominator for roentgenographic quality if any long-time, generalized effects are to be reported for epidemiologic purposes. A roentgenogram of less than good quality may be usable for a single, local determination made by a single physician. It is not adequate for a mass study where current findings are to be reported by a number of readers. It is even less acceptable for inclusion in a system subject to periodic review over a period of years. Test objects for purposes such as that considered here have great value. Patients are not suitable, for it is virtually impossible to select at each facility a patient having common characteristics. Then, too, there is always the charge that such patients are receiving unproductive exposure. In addition, a reproducible test object can be used over a period of years to evaluate and incorporate new technology and new materials since the original conditions can * Presented at the Seventy-third Annual Meeting of the American Roentgen Ray Society, Washington, D. C., October 3-6, OREGON always be reproduced at any time and at any location. Since aluminum test blocks have been widely used, first thoughts naturally are given to that material. For many applications, one would like to use a material providing scattering comparable to that of tissue, but aluminum does not meet this requirement. Water or some of the tissue equivalent materials could be used. A stepped water phantom would involve a complicated container. The tissue equivalent plastics are sometimes difficult to obtain and are not inexpensive. For that reason, a readily available plastic with scattering characteristics not too different from tissue was selected. DEVISE OF TEST PHANTOM The test phantom that we have devised consists of 10 sheets of inch cast thermoplastic acrylic resin, Plexiglas-G. The Plexiglas-G that we have used has what the manufacturer calls Class A tolerance, inches, -o.o8o inches and a nominal weight of 4.62 pounds per square foot. The sheets are 6 inches wide and, since each step is i inches long, the total length is i5 inches. When the phantom is exposed employing the factors routinely used for a chest patient measuring cm., the phantom provides the range of film densities found in a chest roentgenogram. A readout of the densities of the steps can be plotted against the step number to provide a graph that can be compared with another graph made under known and controlled conditions. Such a comparison provides information indicative of processing The work upon which this publication is based was performed pursuant to contract FDA with the Bureau of Radiological Health, Public Health Service, Food and Drug Administration, and the National Institute for Occupational Safety and Health, Department of Health, Education and Welfare. 77

2 772 E. Dale Trout and John P. Kellev AIRIL, 1973 SHARP FACE PB SER NO ALL DIMENSIONS IN INCHES tic. 1. I)rawing of phantom, front view. procedures as well as exposure factors. One of the major problems in the black lung program lies in the fact that many of 3RASS HANDLE - DIA NO 10 EYEBOLT HANDLE BRACKET NO.8* I MACHINE SCREW PLEXIGLAS ROO-2 PCS the facilities involved are used for general roentgenography. As a result, the geometry that results from re-positioning the tube leads to some pretty terrible geometry due to the fact that rigid stops, markers, etc., have never been provided. To get an evaluation of the geometry used, a lead pin is embedded in the top of the phantom and a ring is embedded in tile bottom of the thickest step. The positions of the pin and ring are such that a 72 inch source film distance and with i inch space between the film and tile face of tile cassette holder, the image of tile pin will be centered in the ring if the source is centered to the film and at right ALL DIMENSIONS IN INCHES 11G. 2. Drawing of phantom, side view. FIG. 3. Photograph of phantom.

3 \OL. 117, No. 4 Evaluation of Roentgenographv of the Chest I I..) angles in both planes. The use of correct geometry is very critical when grids are used, especially a grid of high ratio. In facilities where equipment is of limited capacity, there may be a tendency to cheat I ic.. Photograph of phantom and cassette holder. on the required 72 inch source-film distance. In the top of the thickest step are 2 lead bars, 3/32 mdl square and 3/8 inch long. When a 72 inch distance is used tile spacing between the inner images of the FiG. 5. Roentgenograms of phantom: 70 kvp, i mas, no grid; 100 kvp, 10 mas, 6:1 grid; iig kvp, 7.5 mas, 8 :i grid; 130 kyp, 5 mas, 10:1 grid.

4 774 E. Dale Trout and John P. Kellev APRiL, d 2.! U, z w I I I I DENSITY VS STEP PHANTOM A MEDIUM SPEED FILM AND SCREENS 72 INCH SOURCE FILM DISTANCE - 7OkVp.I5mAs, No Grid x.iookvp,iomas, 61 GrId kvp,7.5mas,&i Grid kvp, 5.0 mas, 01 Grid g I STEP IL NUMBER 0,0 9 B ( b DI 3 Fic. 6. 1)ensity c s. step number for phantom roentgenograms. bars is 10 cm. A measurement of this distance can be used to confirm compliance and if necessary to calculate the actual distance used. To record gross lack of roentgenographic detail usually associated with poor screen contact, pieces of brass wire mesh of o and 6o mesh are mounted on steps, 6 and 7. No great significance is attached to their use for other than detection of a gross condition such as poor screen contact or movement of some mechanical structure during exposure. Since the phantom weighs is pounds and has a very smooth surface, a hinged handle has been provided. Figures i and 2 provide information as to construction and dimensions. Figure 3 is a close up photograph and Figure shows the phantom in place in front of a cassette holder. The Plexiglas-G sheets have lengths of i4, 3, 4, 6, 74, 9, io4, 12, 134, and 15 inches. The 4 inch sharp face lead numbers for step identification and serial number and the 4 inch hair line lead numbers for brass mesh identification are placed in holes drilled in the Plexiglas and the holes filled with a clear polyester casting resin. The 2 lead bars, the lead pin and the ring are also placed in holes drilled in the Plexiglas and embedded in clear polyester casting resin. The brass wire mesh is placed on the surface of the Plexiglas and held in place by a thin coat of the casting resin. After all parts have been assembled on or in the Plexiglas, the sheets are assembled and clamped in place, and a small amount of Plexiglas solvent cement (dichloroethane) is placed along the junction of each of the sheets. EXPERIMEN IAL STUDIES In addition to providing a fast and simple evaluation of a facility, the phantom is useful for experimental studies.

5 VOL. 117, No. Evaluation of Roentgenography of tile Chest G. 8. Roentgenograms of phantom: 105 kvp, 10 mas, 8:i grid; 115 kvp, 7.5 mas, 8 :i grid. An example of one such study is that made on the use of grids of different ratios. There is, of course, general agreement that 100 line grids should be used. The agreement as to the value of a grid and the ratio to be used has been the subject of a great deal of discussion. The purpose of a grid is to reduce scattered radiation. There is a tendency in chest roentgenography when using a grid to increase kilovoltage in order to retain a short exposure time. If the kilovoltage is increased, the scattered radiation rate and energy increase, and, be\ond a certain point, the contrast with a grid is less than would have been provided b a non-grid technique. To develop some information on what this kilovoltage might be, we studied 100 line grids having ratios of 6:1, 8:1 and 10:1. The roentgenograms and techniques without a grid and with the 3 grids are shown in Figure. The exposure in milliampere-seconds lies within the ranges generally found in use with single-phase equipment. It will be seen that roentgenograms using 70 kvp without grid are comparable with those using a grid when 100 kvp is used with a 6:1 grid, 115 kvp when an 8:i grid is used and 130 kvp when a 10:1 grid is used. When the roentgenographic density of the steps is plotted against the step number (i.e., thickness) as shown in Figure 6, it will be seen that there is good agreement. Figure 7 shows roentgenograms made on the same individual using the techniques employed with the phantom. Again it will be seen that the results are comparable. It seems reasonable to assume that grids of any ratio can be used to produce acceptable results if the kilovoltage is adjusted for the grid ratio. In the examples shown, the contrast is comparable with that obtained without grid. If contrast is to be increased through the use of a grid, some lower kilovoltage with an increase in milliampere-seconds must be used. To illustrate this, Figure 8 shows for comparison a roentgenogram madeusingan 8:i grid at 115 kvp and 7.5

6 776 E. Dale Trout and John P. Kellev APRIL, 1973 I I I I DENSITY VS STEP PHANTOM A IO5kVp, lonias, 8IGrid o----o 115 kvp,7.5mas,8igrjd / / STEP NUMBER 105 kvp 10mM / 8 I GRID.-.-_ I - I /0/ I milliampere-seconds and a roentgenogram 3.. made with the same grid reducing the kilovoltage to 105 kyp and increasing the ex- MEDIUM SPEED FILM AND SCREENS 72 INCH SOURCE-FILM DISTANCE posure to 10 milliampere-seconds. As would 3. be expected, the reduction in kilovoltage leads to an increase in contrast as shown in Figure 9, where density is plotted as a function of step number. There is, of course, a 2.. slight reduction in latitude. Figure 10 shows roentgenograms made on the same individual using the techniques employed with the phantom. Similar studies are currently underway using three-phase, 12 pulse and battery powered equipment. SUMMARY - A test pilantom that is useful for evaluation of procedures in chest roentgenography has been made. It is not expensive, the or materials are generally available and its use to evaluate a technique or a facility is 115 kvp 7.5 mm 8 l GRID tic. 10. Chest roentgenograms: 105 kvp, 10 mas, 8:i grid; 115 kvp, 7.5 mas, 8:i grid. not complicated E. Dale Trout, D. Sc. Oregon State University tic.. 1)ensitv : s. step number for X-Ray Science and Engineering Laboratory phantom roentgenograms. Corvallis, Oregon 97331

Nuclear Associates

Nuclear Associates Nuclear Associates 07-649 CDRH Fluoroscopic Phantom Users Manual March 2005 Manual No. 07-649-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 07-706 Patient Phantom/Penetrometer System Users Manual March 2005 Manual No. 07-706-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are

More information

Exposure System Selection

Exposure System Selection Principles of Imaging Science II (RAD120) Exposure Systems Exposure System Selection Radiographic exposure is a very complex process Best technique systems manipulate one variable while holding others

More information

RADIOGRAPHIC EXPOSURE

RADIOGRAPHIC EXPOSURE RADIOGRAPHIC EXPOSURE Receptor Exposure Receptor Exposure the that interacts with the receptor. Computed Radiography ( ) requires a. Direct Digital Radiography (DR) requires a. Exposure Indicators Exposure

More information

BEAM QUALITY MEASUREMENTS IN DIAGNOSTIC ROENTGENOLOGY*

BEAM QUALITY MEASUREMENTS IN DIAGNOSTIC ROENTGENOLOGY* JULY, 1970 BEAM QUALITY MEASUREMENTS IN DIAGNOSTIC ROENTGENOLOGY* By E 1)ALE TROUT, l)sc, JOHN P KELLEY, BS, and KENNETH E WEAVER, MS CORVALLIS, T THE Annual Meeting of this Society in 1967 we reported

More information

Veterinary Science Preparatory Training for the Veterinary Assistant. Floron C. Faries, Jr., DVM, MS

Veterinary Science Preparatory Training for the Veterinary Assistant. Floron C. Faries, Jr., DVM, MS Veterinary Science Preparatory Training for the Veterinary Assistant Floron C. Faries, Jr., DVM, MS Radiology Floron C. Faries, Jr., DVM, MS Objectives Determine the appropriate machine settings for making

More information

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question.

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question. RA110 test 3 Multiple Choice Identify the letter of the choice that best completes the statement or answers the question. 1. An object 35 cm in width is radiographed at 100 cm SID and at a 50 cm SOD. What

More information

Nuclear Associates , , , , , ,

Nuclear Associates , , , , , , Nuclear Associates 57-411, 57-412, 57-413 57-426, 57-431, 57-432 57-433, 57-435, 57-436 CLEAR-Pb Transparent X-Ray Compensation Filters Users Manual March 2005 Manual No. 57-XXX-1 Rev. 2 2003, 2005 Fluke

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 76-700 Digital Subtraction Angiography Phantom Users Manual March 2005 Manual No. 76-700-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names

More information

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Acceptance Testing of a Digital Breast Tomosynthesis Unit Acceptance Testing of a Digital Breast Tomosynthesis Unit 2012 AAPM Spring Clinical Meeting Jessica Clements, M.S., DABR Objectives Review of technology and clinical advantages Acceptance Testing Procedures

More information

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE

More information

Beam-Restricting Devices

Beam-Restricting Devices Beam-Restricting Devices Three factors contribute to an increase in scatter radiation: Increased kvp Increased Field Size Increased Patient or Body Part Size. X-ray Interactions a some interact with the

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 07-644 Grid Alignment Test Tool Users Manual March 2005 Manual No. 07-644-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks

More information

Ludlum Medical Physics

Ludlum Medical Physics Ludlum Medical Physics Medical Imaging Radiology QA Test Tools NEW LUDLUM PRODUCT LINE Medical Physics Products Medical Physics Products What are they? Products used to measure radiation output and to

More information

Nuclear Associates EZ CR-DIN Phantoms

Nuclear Associates EZ CR-DIN Phantoms Nuclear Associates 07-605-7777 EZ CR-DIN Phantoms Users Manual August 2006 Manual No. 07-605-7777-1 Rev. 4 2006 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks

More information

1-1. GENERAL 1-2. DISCOVERY OF X-RAYS

1-1. GENERAL 1-2. DISCOVERY OF X-RAYS 1-1. GENERAL Radiography is a highly technical field, indispensable to the modern dental practice, but presenting many potential hazards. The dental radiographic specialist must be thoroughly familiar

More information

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura New Technology https://www.youtube.com/watch?v=ptkzznazb 7U COMPUTED

More information

QC Testing for Computed Tomography (CT) Scanner

QC Testing for Computed Tomography (CT) Scanner QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including

More information

Safelight Fog does what to contrast and density on film?

Safelight Fog does what to contrast and density on film? Terri Jurkiewicz Safelight Fog does what to contrast and density on film? ANSWER INCREASES DENSITY DECREASES CONTRAST Explain how you determine if the focal spot size is within appropriate limits.

More information

Y11-DR Digital Radiography (DR) Image Quality

Y11-DR Digital Radiography (DR) Image Quality Y11-DR Digital Radiography (DR) Image Quality Image quality is stressed for all systems in Safety Code 35. In the relevant sections Health Canada s advice is the manufacturer s recommended test procedures

More information

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Studies on reduction of exposure dose using digital scattered X-ray removal processing Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.

More information

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Studies on reduction of exposure dose using digital scattered X-ray removal processing Studies on reduction of exposure dose using digital scattered X-ray removal processing Poster No.: C-1834 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Kashiyama, M. Funahashi, T. Nakaoka, T.

More information

STUDENT REVIEW QUESTION SET K CR/DR CONTENT AREA

STUDENT REVIEW QUESTION SET K CR/DR CONTENT AREA STUDENT REVIEW QUESTION SET K CR/DR CONTENT AREA RADT 2913 COMPREHENSIVE REVIEW 1 The CR cassette is backed by aluminum that: A. reflects x-rays B. absorbs x-rays C. captures the image D. transmits x-rays

More information

Radiology. Radiograph: Is the image of an object made with use of X- ray instead of light.

Radiology. Radiograph: Is the image of an object made with use of X- ray instead of light. Radiology د. اريج Lec. 3 X Ray Films Radiograph: Is the image of an object made with use of X- ray instead of light. Dental x- ray film: Is a recording media on which image of the object was made by exposing

More information

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging Medical Imaging X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging From: Physics for the IB Diploma Coursebook 6th Edition by Tsokos, Hoeben and Headlee And Higher Level Physics 2 nd Edition

More information

The effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography

The effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography Journal of Physics: Conference Series OPEN ACCESS The effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography To cite this article: N A A Daud et al 2014 J. Phys.:

More information

Physical characteristics of four roentgenographic chains proposed for chest radiography

Physical characteristics of four roentgenographic chains proposed for chest radiography 1982, British Journal of Radiology, 55, 1-6 DECEMBER 1982 Physical characteristics of four roentgenographic chains proposed for chest radiography By H. Manninen, M.D., M.Sc, M. Pitkanen, M.Sc. and M. Wiljasalo,

More information

X-RAYS - NO UNAUTHORISED ENTRY

X-RAYS - NO UNAUTHORISED ENTRY Licencing of premises Premises Refer Guidelines A radiation warning sign and warning notice, X-RAYS - NO UNAUTHORISED ENTRY must be displayed at all entrances leading to the rooms where x-ray units are

More information

Equivalent Penetrameter Sensitivity (EPS) for Performance Evaluation of Computed Radiography Systems Muzibur Khan * and Mike Brothers

Equivalent Penetrameter Sensitivity (EPS) for Performance Evaluation of Computed Radiography Systems Muzibur Khan * and Mike Brothers Equivalent Penetrameter Sensitivity (EPS) for Performance Evaluation of Computed Radiography Systems Muzibur Khan * and Mike Brothers More info about this article: http://www.ndt.net/?id=22782 Abstract

More information

Title: A COMPARISON OF Cs-137 AND X-RAY SOURCES AS CALIBRATION REFERENCES FOR THERMOLUMINESCENT DOSIMETER CHIPS

Title: A COMPARISON OF Cs-137 AND X-RAY SOURCES AS CALIBRATION REFERENCES FOR THERMOLUMINESCENT DOSIMETER CHIPS Title: A COMPARISON OF Cs-137 AND X-RAY SOURCES AS CALIBRATION REFERENCES FOR THERMOLUMINESCENT DOSIMETER CHIPS By Aravind Ravichandran arr192@mail.usask.ca University of Saskatchewan Address: 2424 Cumberland

More information

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto X-ray Imaging PHYS Lecture Carlos Vinhais Departamento de Física Instituto Superior de Engenharia do Porto cav@isep.ipp.pt Overview Projection Radiography Anode Angle Focal Spot Magnification Blurring

More information

Investigation of the line-pair pattern method for evaluating mammographic focal spot performance

Investigation of the line-pair pattern method for evaluating mammographic focal spot performance Investigation of the line-pair pattern method for evaluating mammographic focal spot performance Mitchell M. Goodsitt, a) Heang-Ping Chan, and Bob Liu Department of Radiology, University of Michigan, Ann

More information

APRIL 2009 / NEW-100 / PAGE 1 OF 13

APRIL 2009 / NEW-100 / PAGE 1 OF 13 APRIL 2009 / NEW-100 / PAGE 1 OF 13 The standard UNIDEX line covers reaming applications from 5/16 to 1 1/4 diameter. The single indexable blade and high wear resistant carbide or cermet pads provide a

More information

ISO INTERNATIONAL STANDARD. Non-destructive testing of welds Radiographic testing of fusionwelded

ISO INTERNATIONAL STANDARD. Non-destructive testing of welds Radiographic testing of fusionwelded INTERNATIONAL STANDARD ISO 17636 First edition 2003-09-15 Non-destructive testing of welds Radiographic testing of fusionwelded joints Contrôle non destructif des assemblages soudés Contrôle par radiographie

More information

Spokane Industries Computed Radiography Implementation Experience. David Jolin Technical Sales

Spokane Industries Computed Radiography Implementation Experience. David Jolin Technical Sales Spokane Industries Computed Radiography Implementation Experience David Jolin Technical Sales djolin@spokaneindustries.com Outline Background Our Decision for Computed Radiography Implementation of CR

More information

Amorphous Selenium Direct Radiography for Industrial Imaging

Amorphous Selenium Direct Radiography for Industrial Imaging DGZfP Proceedings BB 67-CD Paper 22 Computerized Tomography for Industrial Applications and Image Processing in Radiology March 15-17, 1999, Berlin, Germany Amorphous Selenium Direct Radiography for Industrial

More information

Implementation of the Theory of Terri L. Fauber on Various Tube Voltage (KV) and Current (mas) on Radiographic Density

Implementation of the Theory of Terri L. Fauber on Various Tube Voltage (KV) and Current (mas) on Radiographic Density Implementation of the Theory of Terri L. Fauber on Various Tube Voltage (KV) and Current (mas) on Radiographic I. Musdalifah 1,*, S. Suryani 2, N. Rauf 2 1 Academy of Radiodiagnostic Engineering and Radiotherapy

More information

Quality Control for Stereotactic Breast Biopsy. Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc

Quality Control for Stereotactic Breast Biopsy. Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc Quality Control for Stereotactic Breast Biopsy Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc. 716-924-0350 Methods of Imaging Guided Breast Biopsy Ultrasound guided, hand-held needle

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 16371-1 First edition 2011-10-01 Non-destructive testing Industrial computed radiography with storage phosphor imaging plates Part 1: Classification of systems Essais non destructifs

More information

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over a lifetime Breast cancer screening programs rely on

More information

Dicing Through Hard and Brittle Materials in the Micro Electronic Industry By Gideon Levinson, Dicing Tools Product Manager

Dicing Through Hard and Brittle Materials in the Micro Electronic Industry By Gideon Levinson, Dicing Tools Product Manager Dicing Through Hard and Brittle Materials in the Micro Electronic Industry By Gideon Levinson, Dicing Tools Product Manager A high percentage of micro electronics dicing applications require dicing completely

More information

ISO INTERNATIONAL STANDARD. Non-destructive testing of welds Radiographic testing Part 1: X- and gamma-ray techniques with film

ISO INTERNATIONAL STANDARD. Non-destructive testing of welds Radiographic testing Part 1: X- and gamma-ray techniques with film INTERNATIONAL STANDARD ISO 17636-1 First edition 2013-01-15 Non-destructive testing of welds Radiographic testing Part 1: X- and gamma-ray techniques with film Contrôle non destructif des assemblages soudés

More information

Teaching Digital Radiography and Fluoroscopic Radiation Protection

Teaching Digital Radiography and Fluoroscopic Radiation Protection Teaching Digital Radiography and Fluoroscopic Radiation Protection WCEC 20 th Student Educator Radiographer Conference Dennis Bowman, RT(R), CRT (R)(F) Community Hospital of the Monterey Peninsula (CHOMP)

More information

Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities

Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities Melissa Martin, M.S., FAAPM, FACR, FACMP AAPM Annual Meeting - Philadelphia July 19, 2010 MO-B-204C-1 Educational Objectives

More information

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Definition and Validation of Exposure Indices Ingrid Reiser, PhD DABR Department of Radiology University of Chicago

More information

Contrast. Contrast: the difference in density on adjacent areas of a radiograph or other image receptor. Subjective. Long Scale (Low Contrast)

Contrast. Contrast: the difference in density on adjacent areas of a radiograph or other image receptor. Subjective. Long Scale (Low Contrast) Contrast Contrast: the difference in density on adjacent areas of a radiograph or other image receptor. Subject Subjective Radiographic Long Scale (Low Contrast) Short Scale (High Contrast) Factors affecting

More information

Test Equipment for Radiology and CT Quality Control Contents

Test Equipment for Radiology and CT Quality Control Contents Test Equipment for Radiology and CT Quality Control Contents Quality Control Testing...2 Photometers for Digital Clinical Display QC...3 Primary Workstations...3 Secondary Workstations...3 Testing of workstations...3

More information

SUBCHAPTER 14. THERAPEUTIC INSTALLATIONS

SUBCHAPTER 14. THERAPEUTIC INSTALLATIONS SUBCHAPTER 14. THERAPEUTIC INSTALLATIONS 7:28-14.1 Scope (a) This subchapter covers therapeutic installations used in the healing arts. These therapeutic installations include x-ray, accelerator and teletherapy

More information

DOE Project: Resist Characterization

DOE Project: Resist Characterization DOE Project: Resist Characterization GOAL To achieve high resolution and adequate throughput, a photoresist must possess relatively high contrast and sensitivity to exposing radiation. The objective of

More information

Film Replacement in Radiographic Weld Inspection The New ISO Standard

Film Replacement in Radiographic Weld Inspection The New ISO Standard BAM Berlin Film Replacement in Radiographic Weld Inspection The New ISO Standard 17636-2 Uwe Ewert, Uwe Zscherpel, Mirko Jechow Requests and information to: uwez@bam.de 1 Outline - The 3 essential parameters

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

Fabrication and Application of Transparent Polycarbonate Sockets 1

Fabrication and Application of Transparent Polycarbonate Sockets 1 Fabrication and Application of Transparent Polycarbonate Sockets 1 by Vert Mooney, M.D. 2 and Roy Snelson, C.P.O. 2 The potential usefulness of transparent sockets in both research and clinical practice

More information

Investigation of Effective DQE (edqe) parameters for a flat panel detector

Investigation of Effective DQE (edqe) parameters for a flat panel detector Investigation of Effective DQE (edqe) parameters for a flat panel detector Poster No.: C-1892 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Exhibit D. Bor 1, S. Cubukcu 1, A. Yalcin 1, O.

More information

Effect of patient support pads on image quality and dose in fluoroscopy a)

Effect of patient support pads on image quality and dose in fluoroscopy a) Effect of patient support pads on image quality and dose in fluoroscopy a) William R. Geiser, Walter Huda, b) and Nikolaos A. Gkanatsios Department of Radiology, University of Florida, Gainesville, Florida

More information

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS JAMES A. TOMLINSON, M.S. Diagnostic Radiological Physicist American Board of Radiology Certified Medical Physics Consultants, Inc. Bio 28 yrs experience 100%

More information

NEXT Protocol for the Survey of Pediatric Chest Radiography

NEXT Protocol for the Survey of Pediatric Chest Radiography NEXT 1998 Protocol for the Survey of Pediatric Chest Radiography Prepared by David C. Spelic Division of Mammography Quality and Radiation Programs Center for Devices and Radiological Health Office of

More information

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of radiation to the population due to Medical Imaging

More information

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893 Digital Radiography D. J. Hall, Ph.D. x20893 djhall@ucsd.edu Background Common Digital Modalities Digital Chest Radiograph - 4096 x 4096 x 12 bit CT - 512 x 512 x 12 bit SPECT - 128 x 128 x 8 bit MRI -

More information

Engineering Policy & Procedure

Engineering Policy & Procedure FPD > Engineering > Global Standards Engineering Policy & Procedure Revision History Number: G2-4 Section: G Subject: Radiographic Examination Procedure 1.0 SCOPE This procedure specifies the requirements

More information

NJDEP Medical Physicist s Radiographic QC Survey Registration Number:

NJDEP Medical Physicist s Radiographic QC Survey Registration Number: Facility Name NJDEP ID # NJDEP Medical Physicist s Radiographic QC Survey PLEASE PRINT Facility Information Unit Information Manufacturer Model Console Model # Console serial # Tube serial # Location (room)

More information

3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development

3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development Teaching Radiographic Technique in a Digital Imaging Paradigm Objectives 1. Discuss the historical development of digital imaging. Dawn Couch Moore, M.M.Sc., RT(R) Assistant Professor and Director Emory

More information

Visualization of sources of scattered radiation from x-ray equipment used for interventional radiology

Visualization of sources of scattered radiation from x-ray equipment used for interventional radiology Visualization of sources of scattered radiation from x-ray equipment used for interventional radiology Poster No.: C-1190 Congress: ECR 2011 Type: Scientific Exhibit Authors: K. Chida, T. Takahashi, D.

More information

Nuclear Associates , , CT Head and Body Dose Phantom

Nuclear Associates , , CT Head and Body Dose Phantom Nuclear Associates 76-414,76-414-4150,76-415 CT Head and Body Dose Phantom Users Manual March 2005 Manual No. 76-414-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product

More information

THE ART OF THE IMAGE: IDENTIFICATION AND REMEDIATION OF IMAGE ARTIFACTS IN MAMMOGRAPHY

THE ART OF THE IMAGE: IDENTIFICATION AND REMEDIATION OF IMAGE ARTIFACTS IN MAMMOGRAPHY THE ART OF THE IMAGE: IDENTIFICATION AND REMEDIATION OF IMAGE ARTIFACTS IN MAMMOGRAPHY William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org INTRODUCTION

More information

CR Basics and FAQ. Overview. Historical Perspective

CR Basics and FAQ. Overview. Historical Perspective Page: 1 of 6 CR Basics and FAQ Overview Computed Radiography is a term used to describe a system that electronically records a radiographic image. Computed Radiographic systems use unique image receptors

More information

Introduction. Sam R. Kottamasu Lawrence R. Kuhns

Introduction. Sam R. Kottamasu Lawrence R. Kuhns Pediatr Radiol (1997) 27: 119 123 Springer-Verlag 1997 Sam R. Kottamasu Lawrence R. Kuhns Musculoskeletal computed radiography in children: scatter reduction and improvement in bony trabecular sharpness

More information

GE Healthcare. Senographe 2000D Full-field digital mammography system

GE Healthcare. Senographe 2000D Full-field digital mammography system GE Healthcare Senographe 2000D Full-field digital mammography system Digital has arrived. The Senographe 2000D Full-Field Digital Mammography (FFDM) system gives you a unique competitive advantage. That

More information

Ch. 223 VETERINARY MEDICINE CHAPTER 223. VETERINARY MEDICINE GENERAL PROVISIONS X-RAYS RADIOACTIVE MATERIAL. Authority

Ch. 223 VETERINARY MEDICINE CHAPTER 223. VETERINARY MEDICINE GENERAL PROVISIONS X-RAYS RADIOACTIVE MATERIAL. Authority Ch. 223 VETERINARY MEDICINE 25 223.1 CHAPTER 223. VETERINARY MEDICINE Sec. 223.1. Purpose and scope. 223.2. [Reserved]. 223.2a. Definitions. 223.3 223.6. [Reserved]. 223.7. Structural shielding. 223.8.

More information

Open. the Digitized world. Fuji Computed Radiography

Open. the Digitized world. Fuji Computed Radiography Open the Digitized world Fuji Computed Radiography If just one of these applies to you... Managing developing fluid is hard and darkroom work is a hassle... Images are not stable... Isn t digitalization

More information

Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity)

Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity) Vascular Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO Medical Systems Division, Shimadzu Corporation Yoshiaki Miura 1. Introduction In recent years, digital cardiovascular

More information

A Folding 5-Element Yagi for 144 MHz

A Folding 5-Element Yagi for 144 MHz A Folding 5-Element Yagi for 144 MHz Steve Kavanagh, VE3SMA, April 2017 1. Introduction I have found antennas which fold up quickly to take less space in the car to be useful in VHF/UHF portable operating.

More information

Digital Imaging Considerations Computed Radiography

Digital Imaging Considerations Computed Radiography Digital Imaging Considerations Digital Radiography Computed Radiography o Cassette based Direct or Indirect Digital Radiography o Cassetteless Computed Radiography 1 CR Image Acquisition Most like conventional

More information

diagnostic examination

diagnostic examination RADIOLOGICAL PHYSICS 2011 Raphex diagnostic examination Adel A. Mustafa, Ph.D., Editor PUBLISHED FOR: RAMPS (Radiological and Medical Physics Society of New York) preface The RAPHEX Diagnostic exam 2011

More information

DIGITAL RADIOGRAPHY ARTIFACTS

DIGITAL RADIOGRAPHY ARTIFACTS IMAGING LAB MPHY 487 DIGITAL RADIOGRAPHY ARTIFACTS Mohammad Esmael Alsulimane B.Sc, M.Sc Medical Physics Lecturer - Physics Department All Rights Reserved: Some information and figures in this presentation

More information

Testing. Material testing will be according to applicable AASHTO, ASTM or Department methods as specified.

Testing. Material testing will be according to applicable AASHTO, ASTM or Department methods as specified. 907.01 Section 907. FENCING MATERIALS 907.01 General Requirements. Materials for use in fencing property, right-of-way and other installations must comply with this section. 907.02 Testing. Material testing

More information

SPRINGFIELD TECHNICAL COMMUNITY COLLEGE ACADEMIC AFFAIRS

SPRINGFIELD TECHNICAL COMMUNITY COLLEGE ACADEMIC AFFAIRS SPRINGFIELD TECHNICAL COMMUNITY COLLEGE ACADEMIC AFFAIRS Course Number: RADG 112 Department: Radiography Course Title: Image Production & Eval. Semester: Spring Year: 1997 Objectives/ Unit One: Introduction

More information

RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II

RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II APPROVED 12/O2/2011 EFFECTIVE SPRING 2013-14 Prefix & Number RAD 150 Course Title: Radiologic Exposure Technique II & Lab Purpose of this submission: New Change/Updated

More information

Mammography: Physics of Imaging

Mammography: Physics of Imaging Mammography: Physics of Imaging Robert G. Gould, Sc.D. Professor and Vice Chair Department of Radiology and Biomedical Imaging University of California San Francisco, California Mammographic Imaging: Uniqueness

More information

crease in exposure when the image intensifier fluoroscope was used Abstract approved

crease in exposure when the image intensifier fluoroscope was used Abstract approved AN ABSTRACT OF THE THESIS OF Kenneth Roy Envall for the M. S. in Radiological Physics (Name) (Degree) (Major) Date thesis is presented August 18, 1966 Title MEASUREMENT OF GASTRO- INTESTINAL AND GONADAL

More information

Plastic Welding Experiences

Plastic Welding Experiences Plastic Welding Experiences In the past I have advocated plastic solvent welding with ABS solvent to rejoin ABS plastic parts without repainting. Solvent is inexpensive but it does takes considerable time

More information

Digital radiography (DR) post processing techniques for pediatric radiology

Digital radiography (DR) post processing techniques for pediatric radiology Digital radiography (DR) post processing techniques for pediatric radiology St Jude Children s Research Hospital Samuel Brady, MS PhD DABR samuel.brady@stjude.org Purpose Review common issues and solutions

More information

Appropriate Inspection Distance of Digital X-Ray Imaging Equipment for Diagnosis

Appropriate Inspection Distance of Digital X-Ray Imaging Equipment for Diagnosis Indian Journal of Science and Technology Vol 8(S8), 380-386, April 2015 ISSN (Print) : 0974-6846 ISSN (Online) : 0974-5645 DOI: 10.17485/ijst/2015/v8iS8/70528 Appropriate Inspection Distance of Digital

More information

MILADY. Product Data. Page 1 of 8

MILADY. Product Data. Page 1 of 8 Page 1 of 8 The MILADY Mammographic Unit offers the best quality-to-price ratio to our customers worldwide. The unit advanced technology together with the application of industrial production standards,

More information

Comparison of computed radiography and filmõscreen combination using a contrast-detail phantom

Comparison of computed radiography and filmõscreen combination using a contrast-detail phantom JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 1, WINTER 2003 Comparison of computed radiography and filmõscreen combination using a contrast-detail phantom Z. F. Lu,* E. L. Nickoloff, J.

More information

Evaluation of a quality control phantom for digital chest radiography

Evaluation of a quality control phantom for digital chest radiography JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 2, NUMBER 2, SPRING 2001 Evaluation of a quality control phantom for digital chest radiography Eugene Mah* Department of Radiology, Medical University

More information

ELECTRONIC CONTROL CONCEPTS 160 Partition Street Saugerties, NY or local phone

ELECTRONIC CONTROL CONCEPTS 160 Partition Street Saugerties, NY or local phone ELECTRONIC CONTROL CONCEPTS 160 Partition Street Saugerties, NY 12477 (800)VIP-XRAY (845)247-9028 Fax or 800-847-9729 local phone 845-246-9013 http://www.eccxray.com sales@eccxray.com INSTRUCTION MANUAL

More information

Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC

Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC R. Bellazzini a,b, G. Spandre a*, A. Brez a, M. Minuti a, M. Pinchera a and P. Mozzo b a INFN Pisa

More information

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis

More information

User Manual Thank you for purchasing our product. If there are any questions, or requests, please do not hesitate to contact the dealer.

User Manual Thank you for purchasing our product. If there are any questions, or requests, please do not hesitate to contact the dealer. User Manual Thank you for purchasing our product. If there are any questions, or requests, please do not hesitate to contact the dealer.. This manual may contain several technical incorrect places or printing

More information

CHAPTER 6 QC Test For Fluoroscopic Equipment. Prepared by:- Kamarul Amin bin Abu Bakar School of Medical Imaging KLMUC

CHAPTER 6 QC Test For Fluoroscopic Equipment. Prepared by:- Kamarul Amin bin Abu Bakar School of Medical Imaging KLMUC CHAPTER 6 QC Test For Fluoroscopic Equipment Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC Lesson Outcomes Describe the objectives of each QC test done. Identify QC

More information

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division

More information

Your worldwide source of grids. What a difference a grid makes

Your worldwide source of grids. What a difference a grid makes Your worldwide source of grids What a difference a grid makes The right assembly to meet your every need Trust the experts A high degree of expertise is required to design and manufacture quality X-ray

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 07-591 Focal Spot Test Tool Users Manual February 2005 Manual No. 07-591-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks

More information

M M Assemblies Including Over-Molded Light Pipe, Positive Locating Mechanism I We Sell Solutions

M M Assemblies Including Over-Molded Light Pipe, Positive Locating Mechanism I We Sell Solutions M Assemblies Including Over-Molded Light Pipe, Positive Locating Mechanism Modern Molding Inc. 796 7 th Street South Delano, MN 55328 Phone (763) 972 6761 Fax (763) 972 6764 Contact: Sales@modernmolding.com

More information

NON-DESTRUCTIVE EVALUATION UTILIZING IMAGING PLATES FOR FIELD RADIOGRAPHY APPLICATIONS

NON-DESTRUCTIVE EVALUATION UTILIZING IMAGING PLATES FOR FIELD RADIOGRAPHY APPLICATIONS 19 th World Conference on Non-Destructive Testing 2016 NON-DESTRUCTIVE EVALUATION UTILIZING IMAGING PLATES FOR FIELD RADIOGRAPHY APPLICATIONS Brian S. WHITE 1 1 Carestream NDT, 1049 Ridge Road West, Rochester,

More information

Testing and Performance Validation of a Shielded Waste Segregation and Clearance Monitor Designed for the Measurement of Low Level Waste-13043

Testing and Performance Validation of a Shielded Waste Segregation and Clearance Monitor Designed for the Measurement of Low Level Waste-13043 Testing and Performance Validation of a Shielded Waste Segregation and Clearance Monitor Designed for the Measurement of Low Level Waste-13043 John A. Mason*, Graham Beaven**, Kevin J. Burke*, Robert Spence**

More information

NATIONWIDE EVALUATION OF X-RAY TRENDS (NEXT) TABULATION AND GRAPHICAL SUMMARY OF THE 1999 DENTAL RADIOGRAPHY SURVEY

NATIONWIDE EVALUATION OF X-RAY TRENDS (NEXT) TABULATION AND GRAPHICAL SUMMARY OF THE 1999 DENTAL RADIOGRAPHY SURVEY CRCPD Publication E-3-6-a Available Online at No Charge $15. for a Computer-Generated Copy NATIONWIDE EVALUATION OF X-RAY TRENDS (NEXT) TABULATION AND GRAPHICAL SUMMARY OF THE 1999 DENTAL RADIOGRAPHY SURVEY

More information

Photons interaction with matter

Photons interaction with matter ب س م هللا الر ح من الر حیم Photons interaction with matter Ionization Ionization is the process of removing an electron from an electrically neutral atom to produce an ion pair. An ion is an atom or subatomic

More information

COMPUTED RADIOGRAPHY (CR)

COMPUTED RADIOGRAPHY (CR) COMPUTED RADIOGRAPHY (CR) Moving with the time Avi Avner BVSc BSc CVR DVDI MRCVS CR-Basics A five step process: 1. X-ray image received on phosphor plate 2. Image extracted from phosphor plate by Laser

More information

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah X-RAY IMAGING EE 472 F2017 Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook Stewart C. Bushong, Radiologic Science for Technologists: Physics, Biology, and Protection, 10 th ed., Mosby,

More information