PACS. Optimum detector bias settings for Ge:Ga detectors, Time constant: bias change spectrometer IMT 509

Size: px
Start display at page:

Download "PACS. Optimum detector bias settings for Ge:Ga detectors, Time constant: bias change spectrometer IMT 509"

Transcription

1 Test Analysis Report FM-ILT/IST Page 1 Optimum detector bias settings for Ge:Ga detectors, Time constant: bias change spectrometer IMT 509 J. Schreiber 1, U. Klaas 1, H. Dannerbauer 1, M. Nielbock 1, J. Bouwman 1 1 Max-Planck-Institut für Astronomie, Königstuhl 17, D Heidelberg, Germany

2 Test Analysis Report FM-ILT/IST Page 2 Req. IMT 509 Optimum detector bias settings for Ge:Ga detectors, Time constant: bias change spectrometer IMT A. History Version Date Author(s) Change description 1.0 June 1, 2007 Juergen Schreiber, Ulrich Klaas, First issue Helmut Dannerbauer, Markus Nielbock, Jeroen Bouwman IMT B. Summary The minimum NEP and therefore the corresponding optimum bias is reached between 60 and 70 mv bias voltage for the red array and 169 and 199 mv for the blue array which is slightly lower than found in the module level tests. Some more pixels than the already known bad pixels show an extraordinarily deviating NEP behaviour and are identified. For some bias steps signal transient behaviour was found with decay times between 1/2 to 4 minutes. The decay times setting the optimum bias of the red array were determined to be between 1 and 2 minutes. The recommendation is therefore to insert wait times of about 8 minutes should a bias change be necessary. No transient behaviour could be detected for the optimum bias voltage setting of the blue array. Of course, these statements are only valid for the used constant bias steps of 10 mv for the red and 30 mv for the blue array, respectively. IMT C. Data Reference Sheet Ref Date Archive filename TM file 1 10/04/07 FILT IST SPEC detector imt509 OBS 45.tm RD-1 13/02/06 DEC/MEC User Manual, -CL-SR-002 RD-2 01/03/05 CQM-ILT Analysis Report, Part II, PICC-KL-TR-001 RD-3 22/02/06 Test Analysis Report EQM-IMT: Optimum Bias Settings for Ge:Ga Detectors (Test ID: 523) RD-4 01/03/07 PTD RD-5 05/01/07 Calibration Document -MA-GS-001, draft 8 RD-6 10/08/06 Cold Performance Tests on FM High Stress Ge:Ga Detector Modules, -ME-TR-063 RD-7 16/03/06 Summary of the cold performance tests on LS-FM Ge:Ga detector modules at MPIA, -MA-TR-030 RD-8 12/02/07 Wavelength Calibration of the Spectrometer FM ILT part I -Calibration Analysis Report PTD 4.2.1

3 Test Analysis Report FM-ILT/IST Page 3 IMT D. Test Description IMT D.1. Introduction This FM-ILT/IST test was a preparation and reference test for the FM-IST tests at Astrium/Friedrichshafen. The goal of this test is to verify the optimum bias voltage at which the Ge:Ga detectors work stably (e.g. do not show spiking pixels) and the NEP shows a minimum. Transient effects of the signal after a bias change should be quantified. The resulting values should be compared to the corresponding results of the module level tests. IMT D.2. Test Overview The FM-IL/IST bias scan measurements were carried out on 10 April 2007 during the final phase of the FM ILT. The temperature of the blue array was set to 2.5 K which is for thermal budget reasons the maximum temperature close to the optimum derived during the nominal FM-ILT Ge:Ga optimization tests. The chopper was held at a constant position of staring to the internal calibration source 2 which was at a temperature of 60 K. The grating position was set constant at corresponding to a wavelength of µmfor the red array and 87.8 µmfor the blue array. The smallest capacitance of 0.14 pf (effective) was used. Data acquisition was done in default mode, i.e. averaging 32 readouts each in 1/2 s ramps, consequently this results in 4 samples per ramp that were fitted linearily to get the signal in readouts per second. The whole 16 bit ADC range was assumed to correspond to 6.26 V. For the red array we scanned through a bias voltage range between 20 and 90 mv in steps of 10 mv, while for the blue array a range between 80 and 290 mv in steps of 30 mv was sampled with a time length of about 120 s for each step. In Fig. 1 the course of the signal of one pixel of the red and blue array is shown overplotted by the applied bias voltage and the errors of the fit of the averaged ramps, which is a measure of the noise. The increase in signal with increasing bias voltage is clearly visible (bias voltages are negative against ground). The blue pixel clearly shows a transient behaviour at high bias voltages after a bias change. Also the increase of noise with increasing bias can be recognized. The known dead pixels described in RD-8 (4 in each array), the open and the dummy channels are not considered in the following calculations.

4 Test Analysis Report FM-ILT/IST Page 4 Figure 1: signals, fit errors, and bias voltages of red and blue pixel 10, 10

5 Test Analysis Report FM-ILT/IST Page 5 IMT E. Results IMT E.1. Optimum Bias Voltage Since the absolute flux of the calibration sources is not accurately known at the moment, the standard deviation (stddev) divided by the signal is used as a representative measure for the NEP (which is actually proportional to this measure). In Fig. 2 the mean and median NEP measures of all pixels of the red array is plotted against the bias, the error bars are the calculated standard deviation of this measure over all pixels hence reflect the dispersion. The minimum of the mean and median NEP measure is at a bias voltage of 70 mv, where also the dispersion shows a minimum. The difference of the NEP measure at 60 mv and 70 mv bias voltage is very small and the dispersion of the 60 mv NEP measure includes the whole 70 mv range. Consequently, the whole bias range between 60 mv and 70 mv could be considered as the optimum bias voltage. In Fig. 3 the NEP measures of the whole red array at 70 mv are displayed except the open and dummy channel. The 4 known dead pixels ([11,3], [11,5], [10, 19], [5, 22]) show up in black, while the 2 known low response pixels ([12, 1], [7, 12], see RD-8) are in white. But there are 2 more deviating pixels with high NEP at the positions [11, 2] and [16, 2] which were not conspicuous at module level tests. The boarder modules at y-position 4, 9, 14, 24 show up brighter on average than the other modules. This is very likely due to a remaining mis-alignment which leads to a reduced flux falling onto these boarder modules, thus increasing the NEP. Figure 2: mean (left panel) and median (right panel) stddev/signal vs. bias for all pixels of the red array. The error bars reflect the dispersion of the measurement. The mean and median NEP measures of the blue array and their dependence on the bias voltage are plotted in Fig. 4. The minimum of the mean NEP measure is at a bias voltage of 169 mv (measured housekeeping value while 170 mv were commanded). For the median NEP measure the minimum is found for the bias voltage of 199 mv (measured housekeeping value while 200 mv were commanded), but this is only slightly lower than the median NEP measure at 169 mv. For both measures the error bars are minimum for the 169 mv bias voltage. The differences of the NEP measures at 169 and 199 mv are very small, consequently, the whole bias range between 169 and 199 mv could be considered as the optimum bias voltage. In Fig. 5 the NEP measures of the whole blue array at 169 mv are displayed except for the open and dummy channels. The 4 known dead pixels ([6, 15], [3, 16], [5, 22], [2, 23]) show up in black. In the blue array all boarder modules at y-position 4, 9, 14, 19 and 24 are affected by poor alignment and show up brighter than the other modules. Additionally, there are some deviating pixels showing up exceedingly bright in the left-most module and in module 23. Also the pixels [1, 19] and [9, 23] show up extraordinarily bright.

6 Test Analysis Report FM-ILT/IST Page 6 Figure 3: mean stddev/signal at 70 mv bias of the red array. Modules (pixel columns in y-direction) are counted from 0 to 24, pixel rows are counted from 0 to 15. Figure 4: mean (left panel) and median (right panel) stddev/signal vs. bias for all pixels of the blue array. The error bars reflect the dispersion of the measurement.

7 Test Analysis Report FM-ILT/IST Page 7 Figure 5: mean stddev/signal at 169 mv bias of the blue array. Modules (pixel columns in y-direction) are counted from 0 to 24, pixel rows are counted from 0 to 15.

8 Test Analysis Report FM-ILT/IST Page 8 IMT E.2. Signal Transient Behaviour after Bias Change In order to estimate the stabilisation times of the signals after a bias change we fitted the signal course of each bias setting by a simple exponential function after subtracting the mean of the last few signals of the plateaux (see fit examples in Fig. 6). Practically, a linear fit was applied to the logarithm of the signal values. Therefore, the exponential function is offset-free, but has a multiplicative offset as free fit parameter. The other free fit parameter was the decay time of the exponential function. Figure 6: Examples of exponential fits to signals after bias change, left panel: red pixel, right panel: blue pixel. The resulting median exponential decay times of all pixels of both arrays are depicted in Fig. 7. The error bars show the dispersion of the decay times over all pixels. The resulting median multiplicative factor of all pixels is shown in Fig. 8, where the error bars indicate the standard deviation. The red array has only negative offsets for all biases (although not significant for the 2 highest bias voltages), i.e. the signals rise with time up to the final plateau value. The resulting decay times are not significant for 30 and 40 mv bias voltage and the multiplicative offsets are very small, i.e for these biases a transient behaviour cannot be confirmed. The fastest decays (decay time about 1 minute) was found at the smallest bias voltage level (20 mv). There is a minimum of decay times at the optimum bias voltage of 69 mv with decay times of about 1 to 2 minutes. The significant decay times span a range between 1 minute to 4 minutes. Therefore, we recommend to include a wait time of about 8 minutes after each bias change before restarting data acquisition with the red array. The transient behaviour of the blue array looks a bit different. The multiplicative offset is negative for the smallest bias and changes to positive values for the highest biases. This indicates rising signals with time up to the plateau value for the smallest bias (80 mv) and decreasing signals down to the plateau values for the highest biases (200 to 290 mv, although not significant for 290 mv). The resulting multiplicative offsets of the intermediate biases (110, 140 and 170 mv) are close to zero and therefore not significant. Additionally, the decay times at these biases are very high and not significant due to the high dispersion, i.e. a transient behaviour at these bias voltages cannot be really confirmed. The significant decay times span a range between 1/2 minute to 4 minutes, therefore, we recommend to include a wait time of about 8 minutes after each bias change before restarting data acquisition with the blue array.

9 Test Analysis Report FM-ILT/IST Page 9 Figure 7: Median decay time of signals after a bias change from the next lower level for all pixels, left panel: red arry, right panel: blue array. The error bars indicate the dispersion over the whole array. Figure 8: Median fit result of the offset factor to the exponential fit function for all pixels, left panel: red array, right panel: blue array. The error bars indicate the dispersion over the whole array.

10 Test Analysis Report FM-ILT/IST Page 10 IMT F. Conclusions The bias scan tests were carried out successfully and allow a verification of the optimum bias setting for both arrays. The minimum NEP and therefore the optimum bias is reached between 60 and 70 mv bias voltage for the red array and 169 and 199 mv for the blue array. This result differs from the bias of minimum NEP found in the module level tests where slightly higher optimum bias voltages were established (which is between 70 and 80 mv for the red array (see RD-6) and around 200 mv for the blue array (see RD-7)). Some more pixels than the already known bad pixels show extraordinarily deviating high NEP behaviour. Future tests of this type should provide a check whether the number of affected pixels remains stable. For some bias change steps a transient behaviour was found with decay times between 1/2 to 4 minutes. The decay times at the optimum bias of the red array were determined to be between 1 and 2 minutes while no transient behaviour could be detected at the optimum bias voltage of the blue array. However, from the general transient behaviour wait times of 8 minutes after a bias change are recommended. IMT G. IA scripts used / remarks on PCSS CAP ist.py spg java function fitramps() to fit all default mode ramps. IMT H. Lessons learned for IMT/IST/PV The measurement time after the bias changes should be extended from 2 minutes to about 8 minutes (if possible) to be sure to reach a stable signal plateau. This would also allow a more accurate determination of the transient behaviour. The right-most modules at the boarder of the field of view seem to receive less radiation than the others. It is known that the alignment was not completely perfect by the repair before ILT period 2. Another alignment improvement was done after ILT period 2, so that changes are expected for the upcoming IST tests.

PACS. Photometer SPT during Commissioning Phase. Herschel. PACS Commissioning Phase FFT/SPT report Page 1

PACS. Photometer SPT during Commissioning Phase. Herschel. PACS Commissioning Phase FFT/SPT report Page 1 PACS Commissioning Phase FFT/SPT report Page 1 Photometer SPT during Commissioning Phase PACS Commissioning Phase FFT/SPT report Page 2 Req. 11.6.19 Staring Measurement on Calibration Source for Low Frequency

More information

PACS. PACS Photometer glitch analysis on the electronics. Herschel. PACS Photometer glitch analysis on the electronics Page 1. K.

PACS. PACS Photometer glitch analysis on the electronics. Herschel. PACS Photometer glitch analysis on the electronics Page 1. K. PACS Photometer glitch analysis on the electronics Page 1 PACS Photometer glitch analysis on the electronics K. Okumura PACS Photometer glitch analysis on the electronics Page 2 Req. 1.1.9bis Glitches

More information

PACS SED and large range scan AOT release note PACS SED and large range scan AOT release note

PACS SED and large range scan AOT release note PACS SED and large range scan AOT release note Page: 1 of 16 PACS SED and large range scan AOT PICC-KL-TN-039 Prepared by Bart Vandenbussche Alessandra Contursi Helmut Feuchtgruber Ulrich Klaas Albrecht Poglitsch Pierre Royer Roland Vavrek Approved

More information

Flux Calibration Monitoring: WFC3/IR G102 and G141 Grisms

Flux Calibration Monitoring: WFC3/IR G102 and G141 Grisms Instrument Science Report WFC3 2014-01 Flux Calibration Monitoring: WFC3/IR and Grisms Janice C. Lee, Norbert Pirzkal, Bryan Hilbert January 24, 2014 ABSTRACT As part of the regular WFC3 flux calibration

More information

WFC3 SMOV Program 11433: IR Internal Flat Field Observations

WFC3 SMOV Program 11433: IR Internal Flat Field Observations Instrument Science Report WFC3 2009-42 WFC3 SMOV Program 11433: IR Internal Flat Field Observations B. Hilbert 27 October 2009 ABSTRACT We have analyzed the internal flat field behavior of the WFC3/IR

More information

Characterisation of the VELO High Voltage System

Characterisation of the VELO High Voltage System Characterisation of the VELO High Voltage System Public Note Reference: LHCb-2008-009 Created on: July 18, 2008 Prepared by: Barinjaka Rakotomiaramanana a, Chris Parkes a, Lars Eklund a *Corresponding

More information

Temperature Dependent Dark Reference Files: Linear Dark and Amplifier Glow Components

Temperature Dependent Dark Reference Files: Linear Dark and Amplifier Glow Components Instrument Science Report NICMOS 2009-002 Temperature Dependent Dark Reference Files: Linear Dark and Amplifier Glow Components Tomas Dahlen, Elizabeth Barker, Eddie Bergeron, Denise Smith July 01, 2009

More information

Some plots from March 2007 tests related to bolometer PSF

Some plots from March 2007 tests related to bolometer PSF Some plots from March 2007 tests related to bolometer PSF D.Lutz May 3, 2007 1 Introduction Document number PICC-ME-TN-020 This is a collection of sparsely commented plots from a quick analysis of some

More information

Mini Workshop Interferometry. ESO Vitacura, 28 January Presentation by Sébastien Morel (MIDI Instrument Scientist, Paranal Observatory)

Mini Workshop Interferometry. ESO Vitacura, 28 January Presentation by Sébastien Morel (MIDI Instrument Scientist, Paranal Observatory) Mini Workshop Interferometry ESO Vitacura, 28 January 2004 - Presentation by Sébastien Morel (MIDI Instrument Scientist, Paranal Observatory) MIDI (MID-infrared Interferometric instrument) 1st generation

More information

WFC3 SMOV Proposal 11422/ 11529: UVIS SOFA and Lamp Checks

WFC3 SMOV Proposal 11422/ 11529: UVIS SOFA and Lamp Checks WFC3 SMOV Proposal 11422/ 11529: UVIS SOFA and Lamp Checks S.Baggett, E.Sabbi, and P.McCullough November 12, 2009 ABSTRACT This report summarizes the results obtained from the SMOV SOFA (Selectable Optical

More information

Instruction manual for T3DS software. Tool for THz Time-Domain Spectroscopy. Release 4.0

Instruction manual for T3DS software. Tool for THz Time-Domain Spectroscopy. Release 4.0 Instruction manual for T3DS software Release 4.0 Table of contents 0. Setup... 3 1. Start-up... 5 2. Input parameters and delay line control... 6 3. Slow scan measurement... 8 4. Fast scan measurement...

More information

Digital Radiography : Flat Panel

Digital Radiography : Flat Panel Digital Radiography : Flat Panel Flat panels performances & operation How does it work? - what is a sensor? - ideal sensor Flat panels limits and solutions - offset calibration - gain calibration - non

More information

NHSC/PACS Web Tutorials Running the PACS Spectrometer pipeline for CHOP/NOD Mode. PACS-301 Level 0 to 1 processing

NHSC/PACS Web Tutorials Running the PACS Spectrometer pipeline for CHOP/NOD Mode. PACS-301 Level 0 to 1 processing NHSC/PACS s Running the PACS Spectrometer pipeline for CHOP/NOD Mode page 1 PACS-301 Level 0 to 1 processing Prepared by Dario Fadda September 2012 Introduction This tutorial will guide you through the

More information

Low Drift Thrust Balance with High Resolution

Low Drift Thrust Balance with High Resolution Low Drift Thrust Balance with High Resolution IEPC-2015-257/ISTS-2015-b-257 Presented at Joint Conference of 30th International Symposium on Space Technology and Science, 34th International Electric Propulsion

More information

NHSC/PACS Web Tutorials Running the PACS Spectrometer pipeline for unchopped line mode. PACS-303 Level 0 to 2 processing

NHSC/PACS Web Tutorials Running the PACS Spectrometer pipeline for unchopped line mode. PACS-303 Level 0 to 2 processing NHSC/PACS s Running the PACS Spectrometer pipeline for unchopped line mode PACS-303 Level 0 to 2 processing Prepared by Dario Fadda April 2014 page 1 Introduction This tutorial will guide you through the

More information

ScanArray Overview. Principle of Operation. Instrument Components

ScanArray Overview. Principle of Operation. Instrument Components ScanArray Overview The GSI Lumonics ScanArrayÒ Microarray Analysis System is a scanning laser confocal fluorescence microscope that is used to determine the fluorescence intensity of a two-dimensional

More information

NHSC/PACS Web Tutorials Running PACS photometer pipelines. PACS 201 (for Hipe 5.0) Level 0 to Level 1 processing: From raw to calibrated data cubes

NHSC/PACS Web Tutorials Running PACS photometer pipelines. PACS 201 (for Hipe 5.0) Level 0 to Level 1 processing: From raw to calibrated data cubes NHSC/PACS Web Tutorials Running PACS photometer pipelines PACS 201 (for Hipe 5.0) Level 0 to Level 1 processing: From raw to calibrated data cubes Prepared by Nicolas Billot and Roberta Paladini August

More information

Performance characteristics of a new wide range, fast settling electrometer design for a residual gas analysis mass spectrometer

Performance characteristics of a new wide range, fast settling electrometer design for a residual gas analysis mass spectrometer Performance characteristics of a new wide range, fast settling electrometer design for a residual gas analysis mass spectrometer MKS Spectra Products, January 2010 Design considerations for RGA components

More information

Stitching MetroPro Application

Stitching MetroPro Application OMP-0375F Stitching MetroPro Application Stitch.app This booklet is a quick reference; it assumes that you are familiar with MetroPro and the instrument. Information on MetroPro is provided in Getting

More information

CCD reductions techniques

CCD reductions techniques CCD reductions techniques Origin of noise Noise: whatever phenomena that increase the uncertainty or error of a signal Origin of noises: 1. Poisson fluctuation in counting photons (shot noise) 2. Pixel-pixel

More information

High-end CMOS Active Pixel Sensor for Hyperspectral Imaging

High-end CMOS Active Pixel Sensor for Hyperspectral Imaging R11 High-end CMOS Active Pixel Sensor for Hyperspectral Imaging J. Bogaerts (1), B. Dierickx (1), P. De Moor (2), D. Sabuncuoglu Tezcan (2), K. De Munck (2), C. Van Hoof (2) (1) Cypress FillFactory, Schaliënhoevedreef

More information

Presented by Jerry Hubbell Lake of the Woods Observatory (MPC I24) President, Rappahannock Astronomy Club

Presented by Jerry Hubbell Lake of the Woods Observatory (MPC I24) President, Rappahannock Astronomy Club Presented by Jerry Hubbell Lake of the Woods Observatory (MPC I24) President, Rappahannock Astronomy Club ENGINEERING A FIBER-FED FED SPECTROMETER FOR ASTRONOMICAL USE Objectives Discuss the engineering

More information

NHSC/PACS Web Tutorials Running PACS photometer pipelines PACS-201

NHSC/PACS Web Tutorials Running PACS photometer pipelines PACS-201 NHSC/PACS Web Tutorials Running PACS photometer pipelines PACS 201 Level 0 to Level 1 processing: From raw to calibrated data cubes Prepared by Nicolas Billot February 2011 - page 1 Introduc=on NHSC PACS

More information

Spectral Analysis of the LUND/DMI Earthshine Telescope and Filters

Spectral Analysis of the LUND/DMI Earthshine Telescope and Filters Spectral Analysis of the LUND/DMI Earthshine Telescope and Filters 12 August 2011-08-12 Ahmad Darudi & Rodrigo Badínez A1 1. Spectral Analysis of the telescope and Filters This section reports the characterization

More information

GPI INSTRUMENT PAGES

GPI INSTRUMENT PAGES GPI INSTRUMENT PAGES This document presents a snapshot of the GPI Instrument web pages as of the date of the call for letters of intent. Please consult the GPI web pages themselves for up to the minute

More information

Improvements in Operating the Raytheon 320 # 240 Pixel Si:As Impurity Band Conduction Mid-Infrared Array

Improvements in Operating the Raytheon 320 # 240 Pixel Si:As Impurity Band Conduction Mid-Infrared Array Publications of the Astronomical Society of the Pacific, 115:1407 1418, 2003 December 2003. The Astronomical Society of the Pacific. All rights reserved. Printed in U.S.A. Improvements in Operating the

More information

Technical Notes. Integrating Sphere Measurement Part II: Calibration. Introduction. Calibration

Technical Notes. Integrating Sphere Measurement Part II: Calibration. Introduction. Calibration Technical Notes Integrating Sphere Measurement Part II: Calibration This Technical Note is Part II in a three part series examining the proper maintenance and use of integrating sphere light measurement

More information

Camera Test Protocol. Introduction TABLE OF CONTENTS. Camera Test Protocol Technical Note Technical Note

Camera Test Protocol. Introduction TABLE OF CONTENTS. Camera Test Protocol Technical Note Technical Note Technical Note CMOS, EMCCD AND CCD CAMERAS FOR LIFE SCIENCES Camera Test Protocol Introduction The detector is one of the most important components of any microscope system. Accurate detector readings

More information

Processing ACA Monitor Window Data

Processing ACA Monitor Window Data Processing ACA Monitor Window Data CIAO 3.4 Science Threads Processing ACA Monitor Window Data 1 Table of Contents Processing ACA Monitor Window Data CIAO 3.4 Background Information Get Started Obtaining

More information

Angular Drift of CrystalTech (1064nm, 80MHz) AOMs due to Thermal Transients. Alex Piggott

Angular Drift of CrystalTech (1064nm, 80MHz) AOMs due to Thermal Transients. Alex Piggott Angular Drift of CrystalTech 38 197 (164nm, 8MHz) AOMs due to Thermal Transients Alex Piggott July 5, 21 1 .1 General Overview of Findings The AOM was found to exhibit significant thermal drift effects,

More information

Intermediate and Advanced Labs PHY3802L/PHY4822L

Intermediate and Advanced Labs PHY3802L/PHY4822L Intermediate and Advanced Labs PHY3802L/PHY4822L Torsional Oscillator and Torque Magnetometry Lab manual and related literature The torsional oscillator and torque magnetometry 1. Purpose Study the torsional

More information

WFC3/UVIS TV3 Post-flash Results

WFC3/UVIS TV3 Post-flash Results Technical Instrument Report WFC3 2012-01 WFC3/UVIS TV3 Post-flash Results S. Baggett and T. Wheeler March 29, 2012 Abstract Given recent interest in potentially reviving the WFC3 post-flash capability,

More information

A Test of non-standard Gain Settings for the NICMOS Detectors

A Test of non-standard Gain Settings for the NICMOS Detectors Instrument Science Report NICMOS 23-6 A Test of non-standard Gain Settings for the NICMOS Detectors Chun Xu & Torsten Böker 2 May, 23 ABSTRACT We report on the results of a test program to explore the

More information

Chemistry Instrumental Analysis Lecture 10. Chem 4631

Chemistry Instrumental Analysis Lecture 10. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 10 Types of Instrumentation Single beam Double beam in space Double beam in time Multichannel Speciality Types of Instrumentation Single beam Requires stable

More information

ARRAY CONTROLLER REQUIREMENTS

ARRAY CONTROLLER REQUIREMENTS ARRAY CONTROLLER REQUIREMENTS TABLE OF CONTENTS 1 INTRODUCTION...3 1.1 QUANTUM EFFICIENCY (QE)...3 1.2 READ NOISE...3 1.3 DARK CURRENT...3 1.4 BIAS STABILITY...3 1.5 RESIDUAL IMAGE AND PERSISTENCE...4

More information

WFC3 TV3 Testing: IR Channel Nonlinearity Correction

WFC3 TV3 Testing: IR Channel Nonlinearity Correction Instrument Science Report WFC3 2008-39 WFC3 TV3 Testing: IR Channel Nonlinearity Correction B. Hilbert 2 June 2009 ABSTRACT Using data taken during WFC3's Thermal Vacuum 3 (TV3) testing campaign, we have

More information

READOUT TECHNIQUES FOR DRIFT AND LOW FREQUENCY NOISE REJECTION IN INFRARED ARRAYS

READOUT TECHNIQUES FOR DRIFT AND LOW FREQUENCY NOISE REJECTION IN INFRARED ARRAYS READOUT TECHNIQUES FOR DRIFT AND LOW FREQUENCY NOISE REJECTION IN INFRARED ARRAYS Finger 1, G, Dorn 1, R.J 1, Hoffman, A.W. 2, Mehrgan, H. 1, Meyer, M. 1, Moorwood A.F.M. 1 and Stegmeier, J. 1 1) European

More information

Interpixel Capacitance in the IR Channel: Measurements Made On Orbit

Interpixel Capacitance in the IR Channel: Measurements Made On Orbit Interpixel Capacitance in the IR Channel: Measurements Made On Orbit B. Hilbert and P. McCullough April 21, 2011 ABSTRACT Using high signal-to-noise pixels in dark current observations, the magnitude of

More information

Cerro Tololo Inter-American Observatory. CHIRON manual. A. Tokovinin Version 2. May 25, 2011 (manual.pdf)

Cerro Tololo Inter-American Observatory. CHIRON manual. A. Tokovinin Version 2. May 25, 2011 (manual.pdf) Cerro Tololo Inter-American Observatory CHIRON manual A. Tokovinin Version 2. May 25, 2011 (manual.pdf) 1 1 Overview Calibration lamps Quartz, Th Ar Fiber Prism Starlight GAM mirror Fiber Viewer FEM Guider

More information

Calibrating Radioactive Monitors

Calibrating Radioactive Monitors 1 Calibrating Radioactive Monitors William Hackeman, Todd Averett April 18, 2008 1. Introduction This research will focus on the calibration of five radiation monitors. Two of the monitors are made by

More information

A flexible compact readout circuit for SPAD arrays ABSTRACT Keywords: 1. INTRODUCTION 2. THE SPAD 2.1 Operation 7780C - 55

A flexible compact readout circuit for SPAD arrays ABSTRACT Keywords: 1. INTRODUCTION 2. THE SPAD 2.1 Operation 7780C - 55 A flexible compact readout circuit for SPAD arrays Danial Chitnis * and Steve Collins Department of Engineering Science University of Oxford Oxford England OX13PJ ABSTRACT A compact readout circuit that

More information

CS 445 HW#2 Solutions

CS 445 HW#2 Solutions 1. Text problem 3.1 CS 445 HW#2 Solutions (a) General form: problem figure,. For the condition shown in the Solving for K yields Then, (b) General form: the problem figure, as in (a) so For the condition

More information

VS7550 VUV/UV Mini Spectrograph Operating Manual

VS7550 VUV/UV Mini Spectrograph Operating Manual Document RD 15 11 No: VS7550 VUV/UV Mini Spectrograph Operating Manual VS7550 Operating Manual 1 Table of Contents Table of Contents Overview Specifications Vacuum Interface Software and Drivers Packing

More information

Comparing Aperture Photometry Software Packages

Comparing Aperture Photometry Software Packages Comparing Aperture Photometry Software Packages V. Bajaj, H. Khandrika April 6, 2017 Abstract Multiple software packages exist to perform aperture photometry on HST data. Three of the most used softwares

More information

Application Note (A13)

Application Note (A13) Application Note (A13) Fast NVIS Measurements Revision: A February 1997 Gooch & Housego 4632 36 th Street, Orlando, FL 32811 Tel: 1 407 422 3171 Fax: 1 407 648 5412 Email: sales@goochandhousego.com In

More information

First and second order systems. Part 1: First order systems: RC low pass filter and Thermopile. Goals: Department of Physics

First and second order systems. Part 1: First order systems: RC low pass filter and Thermopile. Goals: Department of Physics slide 1 Part 1: First order systems: RC low pass filter and Thermopile Goals: Understand the behavior and how to characterize first order measurement systems Learn how to operate: function generator, oscilloscope,

More information

SPIRE Broad-Band Photometry Extraction

SPIRE Broad-Band Photometry Extraction SPIRE Broad-Band Photometry Extraction Bernhard Schulz (NHSC/IPAC) on behalf of the SPIRE ICC, the HSC and the NHSC Contents Point Source Photometry Choices Extended gain correction factors Zero-point

More information

Cell Management Module (CMM)

Cell Management Module (CMM) Cell Management Module (CMM) 2V CMM and 4V CMM versions Monitoring every 2 seconds of cell voltage & temperature 3W of passive balancing configurable for any cell chemistry Amount of balancing coulombs

More information

MAOP-702. CCD 47 Characterization

MAOP-702. CCD 47 Characterization Doc # : MAOP702 Date: 2013Apr03 Page: 1 of 14 MAOP702 Prepared By: Name(s) and Signature(s) Date Jared R. Males Approved By Name and Signature Title Laird Close PI Victor Gasho Program Manager Date Revision

More information

WFC3 Thermal Vacuum Testing: UVIS Broadband Flat Fields

WFC3 Thermal Vacuum Testing: UVIS Broadband Flat Fields WFC3 Thermal Vacuum Testing: UVIS Broadband Flat Fields H. Bushouse June 1, 2005 ABSTRACT During WFC3 thermal-vacuum testing in September and October 2004, a subset of the UVIS20 test procedure, UVIS Flat

More information

WFC3 TV2 Testing: UVIS Filtered Throughput

WFC3 TV2 Testing: UVIS Filtered Throughput WFC3 TV2 Testing: UVIS Filtered Throughput Thomas M. Brown Oct 25, 2007 ABSTRACT During the most recent WFC3 thermal vacuum (TV) testing campaign, several tests were executed to measure the UVIS channel

More information

X-RAY COMPUTED TOMOGRAPHY

X-RAY COMPUTED TOMOGRAPHY X-RAY COMPUTED TOMOGRAPHY Bc. Jan Kratochvíla Czech Technical University in Prague Faculty of Nuclear Sciences and Physical Engineering Abstract Computed tomography is a powerful tool for imaging the inner

More information

AD596/AD597 SPECIFICATIONS +60 C and V S = 10 V, Type J (AD596), Type K (AD597) Thermocouple,

AD596/AD597 SPECIFICATIONS +60 C and V S = 10 V, Type J (AD596), Type K (AD597) Thermocouple, AD597 SPECIFICATIONS (@ +60 C and V S = 10 V, Type J (AD596), Type K (AD597) Thermocouple, unless otherwise noted) Model AD596AH AD597AH AD597AR Min Typ Max Min Typ Max Min Typ Max Units ABSOLUTE MAXIMUM

More information

A 1.3 Megapixel CMOS Imager Designed for Digital Still Cameras

A 1.3 Megapixel CMOS Imager Designed for Digital Still Cameras A 1.3 Megapixel CMOS Imager Designed for Digital Still Cameras Paul Gallagher, Andy Brewster VLSI Vision Ltd. San Jose, CA/USA Abstract VLSI Vision Ltd. has developed the VV6801 color sensor to address

More information

UltraGraph Optics Design

UltraGraph Optics Design UltraGraph Optics Design 5/10/99 Jim Hagerman Introduction This paper presents the current design status of the UltraGraph optics. Compromises in performance were made to reach certain product goals. Cost,

More information

Technical Note How to Compensate Lateral Chromatic Aberration

Technical Note How to Compensate Lateral Chromatic Aberration Lateral Chromatic Aberration Compensation Function: In JAI color line scan cameras (3CCD/4CCD/3CMOS/4CMOS), sensors and prisms are precisely fabricated. On the other hand, the lens mounts of the cameras

More information

Monoblock Management Module (MMM)

Monoblock Management Module (MMM) Monoblock Management Module (MMM) 6V MMM and 12V MMM versions Monitoring every 2 seconds of monoblock voltage & temperature 3W of passive balancing configurable for desired float Amount of balancing coulombs

More information

A Readout ASIC for CZT Detectors

A Readout ASIC for CZT Detectors A Readout ASIC for CZT Detectors L.L.Jones a, P.Seller a, I.Lazarus b, P.Coleman-Smith b a STFC Rutherford Appleton Laboratory, Didcot, OX11 0QX, UK b STFC Daresbury Laboratory, Warrington WA4 4AD, UK

More information

NIRSpec Technical Note NTN Author(s): S. Birkmann Date of Issue: September 27, 2012 Version: 1.2

NIRSpec Technical Note NTN Author(s): S. Birkmann Date of Issue: September 27, 2012 Version: 1.2 NIRSpec Technical Note NTN-2012-002 Author(s): S. Birkmann Date of Issue: September 27, 2012 Version: 1.2 estec European Space Research and Technology Centre Keplerlaan 1 2201 AZ Noordwijk The Netherlands

More information

Global Erratum for Kepler Q0-Q17 & K2 C0-C5 Short-Cadence Data

Global Erratum for Kepler Q0-Q17 & K2 C0-C5 Short-Cadence Data Global Erratum for Kepler Q0-Q17 & K2 C0-C5 Short-Cadence Data KSCI-19080-002 23 March 2016 NASA Ames Research Center Moffett Field, CA 94035 Prepared by: Date Douglas Caldwell, Instrument Scientist Prepared

More information

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer Page 1 of 11 Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer The Aramis Raman system is a software selectable multi-wavelength Raman system with mapping capabilities with a 400mm monochromator and

More information

Stability of IR-arrays for robotized observations at dome C

Stability of IR-arrays for robotized observations at dome C Stability of IR-arrays for robotized observations at dome C 27.3.2007, Tenerife Page Nr. 1 IR wide field imaging MPIA IR projects and studies OMEGA2000: NIR WFI Calar Alto NACO: NIR AO-supported Imager

More information

TRIANGULATION-BASED light projection is a typical

TRIANGULATION-BASED light projection is a typical 246 IEEE JOURNAL OF SOLID-STATE CIRCUITS, VOL. 39, NO. 1, JANUARY 2004 A 120 110 Position Sensor With the Capability of Sensitive and Selective Light Detection in Wide Dynamic Range for Robust Active Range

More information

Thermocouple Conditioner and Setpoint Controller AD596*/AD597*

Thermocouple Conditioner and Setpoint Controller AD596*/AD597* a FEATURES Low Cost Operates with Type J (AD596) or Type K (AD597) Thermocouples Built-In Ice Point Compensation Temperature Proportional Operation 10 mv/ C Temperature Setpoint Operation ON/OFF Programmable

More information

Wavelength Calibration Accuracy of the First-Order CCD Modes Using the E1 Aperture

Wavelength Calibration Accuracy of the First-Order CCD Modes Using the E1 Aperture Wavelength Calibration Accuracy of the First-Order CCD Modes Using the E1 Aperture Scott D. Friedman August 22, 2005 ABSTRACT A calibration program was carried out to determine the quality of the wavelength

More information

GenePix Application Note

GenePix Application Note GenePix Application Note Determining the Signal-to-Noise Ratio and Optimal Photomultiplier gain setting in the GenePix 4000B Siobhan Pickett, M.S., Sean Carriedo, Ph.D. and Chang Wang, Ph.D. Axon Instruments,

More information

XIS 2003 Jun23. K. Hayashida (Osaka University)

XIS 2003 Jun23. K. Hayashida (Osaka University) XIS Meeting @MIT 2003 Jun23 K. Hayashida (Osaka University) FM-DE FM-DE /EM-AE Connection Test @MHI June3-9 9 / 2003 MHI (Mitsubishi Heavy Industry) finished fabrication at the beginning of 2003. Mechanical

More information

SUMMARY/DIALOGUE 2 PRESHAPE PIXEL OVERVIEW 3 BRIEF OPERATING INSTRUCTIONS 3 PRESHAPE PIXEL SIMULATION: EXAMPLE OPERATION 4 PRESHAPE PIXEL SIMULATION:

SUMMARY/DIALOGUE 2 PRESHAPE PIXEL OVERVIEW 3 BRIEF OPERATING INSTRUCTIONS 3 PRESHAPE PIXEL SIMULATION: EXAMPLE OPERATION 4 PRESHAPE PIXEL SIMULATION: SUMMARY/DIALOGUE 2 PRESHAPE PIXEL OVERVIEW 3 BRIEF OPERATING INSTRUCTIONS 3 PRESHAPE PIXEL SIMULATION: EXAMPLE OPERATION 4 PRESHAPE PIXEL SIMULATION: SMALL SIGNALS AROUND THRESHOLD 5 PRESHAPE PIXEL SIMULATION:

More information

CCD Image Processing of M15 Images Estimated time: 4 hours

CCD Image Processing of M15 Images Estimated time: 4 hours CCD Image Processing of M15 Images Estimated time: 4 hours For this part of the astronomy lab, you will use the astronomy software package IRAF (Image Reduction and Analysis Facility) to perform the basic

More information

Observational Astronomy

Observational Astronomy Observational Astronomy Instruments The telescope- instruments combination forms a tightly coupled system: Telescope = collecting photons and forming an image Instruments = registering and analyzing the

More information

Performance status of IASI on MetOp-A and MetOp-B

Performance status of IASI on MetOp-A and MetOp-B Performance status of IASI on MetOp-A and MetOp-B E. Jacquette (1), E. Péquignot (1), J. Chinaud (1), C. Maraldi (1), D. Jouglet (1), S. Gaugain (1), L. Buffet (1), C. Villaret (1), C. Larigauderie (1),

More information

Introduction Test results standard tests Test results extended tests Conclusions

Introduction Test results standard tests Test results extended tests Conclusions Production and Tests of Hybrid Photon Detectors for the LHCb RICH Detectors, University of Edinburgh On behalf of the LHCb experiment Introduction Test results standard tests Test results extended tests

More information

X-ray Spectroscopy Laboratory Suresh Sivanandam Dunlap Institute for Astronomy & Astrophysics, University of Toronto

X-ray Spectroscopy Laboratory Suresh Sivanandam Dunlap Institute for Astronomy & Astrophysics, University of Toronto X-ray Spectroscopy Laboratory Suresh Sivanandam, 1 Introduction & Objectives At X-ray, ultraviolet, optical and infrared wavelengths most astronomical instruments employ the photoelectric effect to convert

More information

Manual CBT current preamplifier PA1 and CBT Reader software Version 1.5

Manual CBT current preamplifier PA1 and CBT Reader software Version 1.5 Manual CBT current preamplifier PA1 and CBT Reader software Version 1.5 Aivon Oy Tietotie 3, FI-02150 Finland tel. +358-400-265501 email: info@aivon.fi Updates to this manual found at: www.aivon.fi This

More information

Tutors Dominik Dannheim, Thibault Frisson (CERN, Geneva, Switzerland)

Tutors Dominik Dannheim, Thibault Frisson (CERN, Geneva, Switzerland) Danube School on Instrumentation in Elementary Particle & Nuclear Physics University of Novi Sad, Serbia, September 8 th 13 th, 2014 Lab Experiment: Characterization of Silicon Photomultipliers Dominik

More information

Control of Noise and Background in Scientific CMOS Technology

Control of Noise and Background in Scientific CMOS Technology Control of Noise and Background in Scientific CMOS Technology Introduction Scientific CMOS (Complementary metal oxide semiconductor) camera technology has enabled advancement in many areas of microscopy

More information

LOGARITHMIC PROCESSING APPLIED TO NETWORK POWER MONITORING

LOGARITHMIC PROCESSING APPLIED TO NETWORK POWER MONITORING ARITHMIC PROCESSING APPLIED TO NETWORK POWER MONITORING Eric J Newman Sr. Applications Engineer in the Advanced Linear Products Division, Analog Devices, Inc., email: eric.newman@analog.com Optical power

More information

Amplifier Luminescence and RBI. Richard Crisp May 21,

Amplifier Luminescence and RBI. Richard Crisp May 21, Amplifier Luminescence and RBI Richard Crisp May 21, 2013 rdcrisp@earthlink.net www.narrowbandimaging.com Outline What is amplifier luminescence? What mechanism causes amplifier luminescence at the transistor

More information

Wide-field Infrared Survey Explorer (WISE)

Wide-field Infrared Survey Explorer (WISE) Wide-field Infrared Survey Explorer (WISE) Latent Image Characterization Version 1.0 12-July-2009 Prepared by: Deborah Padgett Infrared Processing and Analysis Center California Institute of Technology

More information

1 A1 PROs. Ver0.1 Ai9943. Complete 10-bit, 25MHz CCD Signal Processor. Features. General Description. Applications. Functional Block Diagram

1 A1 PROs. Ver0.1 Ai9943. Complete 10-bit, 25MHz CCD Signal Processor. Features. General Description. Applications. Functional Block Diagram 1 A1 PROs A1 PROs Ver0.1 Ai9943 Complete 10-bit, 25MHz CCD Signal Processor General Description The Ai9943 is a complete analog signal processor for CCD applications. It features a 25 MHz single-channel

More information

PASA Production Acceptance Test

PASA Production Acceptance Test PASA Production Acceptance Test Uwe Bonnes, Helmut Oeschler, Simon Lang Institut für Kernphysik, Technische Universität Darmstadt, Germany Luciano Musa CERN, Geneva, CH-1211 Switzerland Lennart Österman

More information

a simple optical imager

a simple optical imager Imagers and Imaging a simple optical imager Here s one on our 61-Inch Telescope Here s one on our 61-Inch Telescope filter wheel in here dewar preamplifier However, to get a large field we cannot afford

More information

NIRSpec Technical Note NTN / ESA-JWST-TN Authors: G. Giardino, S. Birkmann, M. Sirianni Date of Issue: 9 Nov Version: 1.

NIRSpec Technical Note NTN / ESA-JWST-TN Authors: G. Giardino, S. Birkmann, M. Sirianni Date of Issue: 9 Nov Version: 1. NIRSpec Technical Note NTN-2011-005 / ESA-JWST-TN-18258 Authors: G. Giardino, S. Birkmann, M. Sirianni Date of Issue: 9 Nov. 2011 Version: 1.1 estec European Space Research and Technology Centre Keplerlaan

More information

Pulse Shape Analysis for a New Pixel Readout Chip

Pulse Shape Analysis for a New Pixel Readout Chip Abstract Pulse Shape Analysis for a New Pixel Readout Chip James Kingston University of California, Berkeley Supervisors: Daniel Pitzl and Paul Schuetze September 7, 2017 1 Table of Contents 1 Introduction...

More information

Epic Detector Matrices

Epic Detector Matrices Epic Detector Matrices June 2013 - SAS workshop Why do we need detector matrices? Initial source spectrum Source flux How do we use them? Spectral fitting: - xspec, CIAO Try a spectral model and see if

More information

ENGINEERING CHANGE ORDER ECO No. COS-070 Center for Astrophysics & Space Astronomy Date 6 February 2002 University of Colorado, Boulder Sheet 1 of 3

ENGINEERING CHANGE ORDER ECO No. COS-070 Center for Astrophysics & Space Astronomy Date 6 February 2002 University of Colorado, Boulder Sheet 1 of 3 University of Colorado, Boulder Sheet 1 of 3 Description of Change: 1. Page 11 (Rev. 16), Sec. 1.3.3: Replace the opening paragraph with the following text, which reflects the decision to fly a 225-type

More information

A NEW DC MEASUREMENT PRINCIPLE TO FULLY COVER DEVICE SELF-HEATING

A NEW DC MEASUREMENT PRINCIPLE TO FULLY COVER DEVICE SELF-HEATING -1- A NEW DC MEASUREMENT PRINCIPLE TO FULLY COVER DEVICE SELF-HEATING Keywords: Device Self-Heating During DC Measurement, Data Sampling After Self-Heating Reached Final Value, IC-CAP Demo File Contents:

More information

Camera Calibration Certificate No: DMC III 27542

Camera Calibration Certificate No: DMC III 27542 Calibration DMC III Camera Calibration Certificate No: DMC III 27542 For Peregrine Aerial Surveys, Inc. #201 1255 Townline Road Abbotsford, B.C. V2T 6E1 Canada Calib_DMCIII_27542.docx Document Version

More information

Computerized Data Acquisition Systems. Chapter 4

Computerized Data Acquisition Systems. Chapter 4 Computerized Data Acquisition Systems Chapter 4 Data Acquisition - Objectives State and discuss in terms a bright high school student would understand the following definitions related to data acquisition

More information

Noise Analysis of AHR Spectrometer Author: Andrew Xiang

Noise Analysis of AHR Spectrometer Author: Andrew Xiang 1. Introduction Noise Analysis of AHR Spectrometer Author: Andrew Xiang The noise from Spectrometer can be very confusing. We will categorize different noise and analyze them in this document from spectrometer

More information

ENGINEERING CHANGE ORDER ECO No. COS-058 Center for Astrophysics & Space Astronomy Date 2 March 2001 University of Colorado, Boulder Sheet 1 of 16

ENGINEERING CHANGE ORDER ECO No. COS-058 Center for Astrophysics & Space Astronomy Date 2 March 2001 University of Colorado, Boulder Sheet 1 of 16 University of Colorado, Boulder Sheet 1 of 16 Description of Change: 1. Section 2.1.1.2, p. 30 of Rev 10: The LSHUTINI macro was deleted some time ago, so the only remaining macro is to Move External Shutter.

More information

RAZTEC LINK CURRENT SENSOR TECHNICAL INFORMATION

RAZTEC LINK CURRENT SENSOR TECHNICAL INFORMATION RAZTEC LINK CURRENT SENSOR TECHNICAL INFORMATION DESCRIPTION The Raztec Link current sensor looks rather like a fuse or even a shunt but offers some very significant technical advantages over shunts when

More information

Cross-Talk in the ACS WFC Detectors. II: Using GAIN=2 to Minimize the Effect

Cross-Talk in the ACS WFC Detectors. II: Using GAIN=2 to Minimize the Effect Cross-Talk in the ACS WFC Detectors. II: Using GAIN=2 to Minimize the Effect Mauro Giavalisco August 10, 2004 ABSTRACT Cross talk is observed in images taken with ACS WFC between the four CCD quadrants

More information

OPTICAL BACKSCATTER REFLECTOMETER TM (Model OBR 5T-50)

OPTICAL BACKSCATTER REFLECTOMETER TM (Model OBR 5T-50) OPTICAL BACKSCATTER REFLECTOMETER TM (Model OBR 5T-50) The Luna OBR 5T-50 delivers fast, accurate return loss, insertion loss, and length measurements with 20 micron spatial resolution. PERFORMANCE HIGHLIGHTS

More information

Analogue to Digital Conversion

Analogue to Digital Conversion Analogue to Digital Conversion Turns electrical input (voltage/current) into numeric value Parameters and requirements Resolution the granularity of the digital values Integral NonLinearity proportionality

More information

VU And PPM Audio Metering

VU And PPM Audio Metering Introduction VU And PPM Audio Metering Rod Elliott (ESP) VU (Volume Unit) meters are still the mainstay of audio metering systems. The Peak Programme Meter (PPM) was originally developed by the BBC to

More information

Use of the Shutter Blade Side A for UVIS Short Exposures

Use of the Shutter Blade Side A for UVIS Short Exposures Instrument Science Report WFC3 2014-009 Use of the Shutter Blade Side A for UVIS Short Exposures Kailash Sahu, Sylvia Baggett, J. MacKenty May 07, 2014 ABSTRACT WFC3 UVIS uses a shutter blade with two

More information

Application Notes: Discrete Amplification Photon Detector 5x5 Array Including Pre- Amplifiers Board

Application Notes: Discrete Amplification Photon Detector 5x5 Array Including Pre- Amplifiers Board Application Notes: Discrete Amplification Photon Detector 5x5 Array Including Pre- Amplifiers Board March 2015 General Description The 5x5 Discrete Amplification Photon Detector (DAPD) array is delivered

More information

Science Detectors for E-ELT Instruments. Mark Casali

Science Detectors for E-ELT Instruments. Mark Casali Science Detectors for E-ELT Instruments Mark Casali 1 The Telescope Nasmyth telescope with a segmented primary mirror. Novel 5 mirror design to include adaptive optics in the telescope. Classical 3mirror

More information

STIS CCD Saturation Effects

STIS CCD Saturation Effects SPACE TELESCOPE SCIENCE INSTITUTE Operated for NASA by AURA Instrument Science Report STIS 2015-06 (v1) STIS CCD Saturation Effects Charles R. Proffitt 1 1 Space Telescope Science Institute, Baltimore,

More information