PACS. Photometer SPT during Commissioning Phase. Herschel. PACS Commissioning Phase FFT/SPT report Page 1
|
|
- Amelia Chapman
- 6 years ago
- Views:
Transcription
1 PACS Commissioning Phase FFT/SPT report Page 1 Photometer SPT during Commissioning Phase
2 PACS Commissioning Phase FFT/SPT report Page 2 Req Staring Measurement on Calibration Source for Low Frequency Noise A. History Version Date Author(s) Change description JUN 2009 K.Okumura First issue B. Summary The low frequency noise is measured using a spectral density of the measurements on a stable source over 30 minutes. The source used here is one of the PACS calibration sources (CSs) C. Data Reference Sheet OBSID Date Archive filename comment Phot LongBatch tm on VRL FIST SPT Phot tm DDCS FIST SPT Phot tm direct FIST SPT StaringMeasurementCsLowFreqPhot tm OD0026 OBSID PACS Calibration CPFFT 3- on VRL CPFFT IST414 nstdphotsetup nominal na 0001.tm OD0027 OBSID PACS Calibration CPSPTPHOT 2- DDCS CPSPTPHOT IST520 nstdlowfreq ddcs na 0001.tm OD0027 OBSID PACS Calibration CPSPTPHOT 2- direct CPSPTPHOT IST520 nstdlowfreq direct na 0001.tm D. Test Description The relatively well calibrated source available during IST at ESTEC is the PACS calibration sources (CSs). The noise spectrum is measured on CS2 during 30 minutes. The stability of the responsivity and offset is measured before and after the noise measurement using the chopper frequency of 0.5 Hz between CS1 and CS E. Results E.1. Obtained signal E.2. Spectral distribution of noise In order to assess the noise from the electronic chain without the bolometer, 1 hour data were measured in the configuration where the input power is only from the voltage VRL. Any input comes from the bolometer arrays. Figs 2 show the spectral density of the noise of these measurements. Except for the matrix9 of the red channel with VRL of 0.3V, which goes out of the AD converter s upper limit of 250mV, the other measurements provide meaningfull results. Two peaks around and Hz correspond to 30 and 60 seconds, but the origin of these components are presently unknown. The noise level around 3Hz reaches as low as 3µV/ Hz which is, as seen below, well below the noise level measured with the bolometer. However, this noise level is at the limit of the signal digitization, as shown in req (page??), and its value itsef should be taken with caution.
3 PACS Commissioning Phase FFT/SPT report Page 3 Figure 1: Signal of the pixel [8,8] of the blue [left column] and red detector [right column]. [top] Measurement of electronics (on VRL=0.45V) was obtained right after the PACS photometer switch-on and during the internal calibration sources (CSs) stabilisation. [middle]. DDCS mode and [bottom] direct mode with the standard bias setting.
4 PACS Commissioning Phase FFT/SPT report Page 4 Figure 2: Noise spectrum of blue detectors [left column] and red detector [right column]. The measurements during ILT (2007) with reference voltage VRL = 0.3 [top], 0.4 [middle]. During the commissioning phase VRL = 0.45V [bottom].
5 PACS Commissioning Phase FFT/SPT report Page 5 Figs. 3 comapare the noise spectra in direct mode and processed with and without the deglitching. Figs. 4 show the same for the DDCS mode. The comparison shows clearly the part of contribution of glitches in the noise spectra. The deglitching has been performed with photmmtdeglitching algorhithme with the parameters: scales=3, nsigma=8. It is seen that the contribution of glitches is seen on large frequency range below abour 3Hz. This increases of course the noise level, but the deglitching implemented in IA allows to recover quite well the noise level we had before on ground. Figure 3: Noise spectral density in direct mode, showing a comparison between [left column] with and [right column] without glitches Figs. 5 comapare the noise spectra in direct mode measured during IST (2008 Aug 27), those measured during TVTB (2008 Dec 4) both at ESTEC and those measured during the commissioning phase (2009 Jun 11). Figs. 6
6 PACS Commissioning Phase FFT/SPT report Page 6 Figure 4: Noise spectral density in DDCS mode, showing a comparison between [left column] with and [right column] without glitches
7 PACS Commissioning Phase FFT/SPT report Page 7 show the same for the DDCS mode. These results show that the signal after the glitch removal has a comparable noise spectra as on ground data before launch E.3. Stability of the response and the drift of the offset The chopped measurements before and after the noise measurement allows to assess the stability of the response and the offset within the time scale of the noise measurement. The variations measured during the commissioning phase is comparable with those before the launch. Table 1: Stability assessment of the response and offset Mode Blue [%] Blue [µv] Red [%] Red [µv] response offset response offset direct DDCS F. Conclusions G. IA scripts used / remarks on PCSS onetmtosavefile.py lowfreqmean.py lowfreqmeanplot.py ampli CSs.py
8 PACS Commissioning Phase FFT/SPT report Page 8 Figure 5: Noise spectral density in direct mode. Comparison between [Top] IST, [Middle] TVTB both measured at ESTEC and [Bottom] deglitched Commissioning Phase data
9 PACS Commissioning Phase FFT/SPT report Page 9 Figure 6: Noise spectral density in DDCS mode. Comparison between [Top] IST, [Middle] TVTB both measured at ESTEC and [Bottom] deglitched Commissioning Phase data
10 PACS Commissioning Phase FFT/SPT report Page 10 Req Time constant for flux change in photometry A. History Version Date Author(s) Change description JUN 2009 K.Okumura First issue B. Summary The flux change is produced by chopper movements between the PACS calibration sources (CSs) which have different temperatures. The chopper is moved at different frequencies to explore the different amplitude of the signal modulation as a function of the chopper frequency. The general behaviour during TVTB was found here to be comparable to the results of IST at ESTEC on August However, the measured amplitude of the signal ascillation is between that of TVTB and IST. The reason is currently unknown C. Data Reference Sheet OBSID Date Archive filename FIST SPT Phot tm FIST SPT TimeConstantsFluxChangesPhot tm OD0027 OBSID PACS Calibration CPSPTPHOT 2- CPSPTPHOT IST521 nstdtimeconst fluxchange na 0001.tm D. Test Description During these measurements the chopper moves between 2 positions corresponding to the 2 calibration sources (CSs). Different chopper frequencies are set to analyse the effect of the flux change. The left column of Figs. 7 and Figs. 8 show some typical signals. The 4 different frequencies are repeated twice for 2 different filters of the blue side. The 4th frequency for each filter is 4Hz. This means that there are 5 readouts on each chopper plateau before the average of 4 succesive readouts by SPU. The 3 plateaus every 5 contain 4 succesive readouts which mixes the different plateau levels giving the SPU output without any obvious plateau anymore. With a relatively low bias biases (VH-VL) used here (2.0V for the blue and 1.5V for the red), the time constant is longer than with the nominal bias. The effect of the average of 4 readouts with these biases decreases the modulation amplitude as seen in the figures E. Results For each chopper frequency the amplitude of the signal modulation is measured by the minimum and maximum in the corresponding intervall for each pixel. The pixel to pixel dispersion is then computed as the standard deviation of the amplitudes over all the valid pixel of each matrix. Figs. 9 and Figs. 10 show the results for 3 wavelengths. The amplitude decreases slightly from 0.03Hz to 1Hz and then it drops at 2Hz as expected due to the global time constant of the bolometer and the electronics. This trend was also seen in both ILT and IST data. It is however noticeable that the modulation amplitude of the blue detectors at 3 low frequencies is systematically lower in TVTB data than IST data. The measurements from the commissioning phase show however the regain in amplitude. This amplitude fluctuation is unexplained so far F. Conclusions
11 PACS Commissioning Phase FFT/SPT report Page 11 Figure 7: [Left] Typical blue signal modulation over all the test sequence, [Right] Zoom on the signal with 1 Hz chopper movement. [Top] IST at ESTEC (August 2008) [Middle] TVTB at ESTEC (December 2008) [Bottom] Commissioning Phase (June 2009).
12 PACS Commissioning Phase FFT/SPT report Page 12 Figure 8: [Left] Typical blue signal modulation over all the test sequence, [Right] Zoom on the signal with 1 Hz chopper movement. [Top] IST at ESTEC (August 2008), [Middle] TVTB at ESTEC (December 2008), [Bottom] Commissioning Phase (June 2009).
13 PACS Commissioning Phase FFT/SPT report Page 13 Figure 9: Measured amplitudes of the blue signal modulation : [Left] 70µm filter, and [Right] 100µm filter. The error bars represent the pixel-to-pixel dispersion on each matrix. [Top] IST at ESTEC (August 2008), [Middle] TVTB at ESTEC (December 2008), [Bottom] Commissioning Phase (June 2009)
14 PACS Commissioning Phase FFT/SPT report Page 14 Figure 10: Measured amplitudes of the red signal modulation. The error bars represent the pixel-to-pixel dispersion on each matrix. [Top] IST at ESTEC (August 2008), [Middle] TVTB at ESTEC (December 2008), [Bottom] Commissioning Phase (June 2009)
15 PACS Commissioning Phase FFT/SPT report Page 15 The measurements during the commissioning phase are globally consistent with those of TVTB measurements at ESTEC in Decembre 2008 and those of IST at ESTEC in August However, the signal amplitude is varying from one test to another. The reason for this is currently unknown G. IA scripts used / remarks on PCSS timeconstist.py timeconstist py timeconstist py timeconstistod27.py
PACS. PACS Photometer glitch analysis on the electronics. Herschel. PACS Photometer glitch analysis on the electronics Page 1. K.
PACS Photometer glitch analysis on the electronics Page 1 PACS Photometer glitch analysis on the electronics K. Okumura PACS Photometer glitch analysis on the electronics Page 2 Req. 1.1.9bis Glitches
More informationPACS. Optimum detector bias settings for Ge:Ga detectors, Time constant: bias change spectrometer IMT 509
Test Analysis Report FM-ILT/IST Page 1 Optimum detector bias settings for Ge:Ga detectors, Time constant: bias change spectrometer IMT 509 J. Schreiber 1, U. Klaas 1, H. Dannerbauer 1, M. Nielbock 1, J.
More informationNHSC/PACS Web Tutorials Running PACS photometer pipelines PACS-201
NHSC/PACS Web Tutorials Running PACS photometer pipelines PACS 201 Level 0 to Level 1 processing: From raw to calibrated data cubes Prepared by Nicolas Billot February 2011 - page 1 Introduc=on NHSC PACS
More informationSome plots from March 2007 tests related to bolometer PSF
Some plots from March 2007 tests related to bolometer PSF D.Lutz May 3, 2007 1 Introduction Document number PICC-ME-TN-020 This is a collection of sparsely commented plots from a quick analysis of some
More informationNHSC/PACS Web Tutorials Running PACS photometer pipelines. PACS 201 (for Hipe 5.0) Level 0 to Level 1 processing: From raw to calibrated data cubes
NHSC/PACS Web Tutorials Running PACS photometer pipelines PACS 201 (for Hipe 5.0) Level 0 to Level 1 processing: From raw to calibrated data cubes Prepared by Nicolas Billot and Roberta Paladini August
More informationNHSC/PACS Web Tutorials Running the PACS Spectrometer pipeline for CHOP/NOD Mode. PACS-301 Level 0 to 1 processing
NHSC/PACS s Running the PACS Spectrometer pipeline for CHOP/NOD Mode page 1 PACS-301 Level 0 to 1 processing Prepared by Dario Fadda September 2012 Introduction This tutorial will guide you through the
More informationBasic principles of PACS photometer data reduc7on
Basic principles of PACS photometer data reduc7on B. Altieri, HSC/ESAC November 2012 B. Altieri HSC/ESAC November 2012 Slide 1 PACS photometer Two filled bolometer arrays in dual band imaging: 64x32 pixels
More informationIssues with Photometer Data & How to Resolve them with HIPE Tools
Issues with Photometer Data & How to Resolve them with HIPE Tools Kevin Xu NHSC/IPAC on behalf of the SPIRE ICC Known Issues New Flux Calibration in HIPE 11 Cooler Burps Glitches Missed Thermistor Signal
More information2.996/6.971 Biomedical Devices Design Laboratory Lecture 7: OpAmps
2.996/6.971 Biomedical Devices Design Laboratory Lecture 7: OpAmps Instructor: Dr. Hong Ma Oct. 3, 2007 Fundamental Circuit: Source and Load Sources Power supply Signal Generator Sensor Amplifier output
More informationPACS photometry on extended sources
PACS photometry on extended sources Total flux experiments Bruno Altieri on behalf of Marc Sauvage 1. Point-source photometry status 2. Prospect on extended emission photometry from theory 3. Results from
More informationPACS SED and large range scan AOT release note PACS SED and large range scan AOT release note
Page: 1 of 16 PACS SED and large range scan AOT PICC-KL-TN-039 Prepared by Bart Vandenbussche Alessandra Contursi Helmut Feuchtgruber Ulrich Klaas Albrecht Poglitsch Pierre Royer Roland Vavrek Approved
More informationSPIRE Broad-Band Photometry Extraction
SPIRE Broad-Band Photometry Extraction Bernhard Schulz (NHSC/IPAC) on behalf of the SPIRE ICC, the HSC and the NHSC Contents Point Source Photometry Choices Extended gain correction factors Zero-point
More informationSpectral Analysis of the LUND/DMI Earthshine Telescope and Filters
Spectral Analysis of the LUND/DMI Earthshine Telescope and Filters 12 August 2011-08-12 Ahmad Darudi & Rodrigo Badínez A1 1. Spectral Analysis of the telescope and Filters This section reports the characterization
More informationQUICK-START GUIDE TO HERSCHEL PACS THE PHOTOMETER. Katrina Exter HERSCHEL-HSC-DOC-2151, version 1.0, February 28, 2017
QUICK-START GUIDE TO HERSCHEL PACS THE PHOTOMETER Katrina Exter HERSCHEL-HSC-DOC-2151, version 1.0, February 28, 2017 Contents The PACS photometer 2 Photometer beams 4 Photometer observing modes (AOTs)
More informationChemistry Instrumental Analysis Lecture 10. Chem 4631
Chemistry 4631 Instrumental Analysis Lecture 10 Types of Instrumentation Single beam Double beam in space Double beam in time Multichannel Speciality Types of Instrumentation Single beam Requires stable
More informationExtracting Photometry from SPIRE Maps
NHSC SPIRE Point Source Spectroscopy Webinar 21 March 2012 Extracting Photometry from SPIRE Maps David Shupe, Bernhard Schulz, Kevin Xu on behalf of the SPIRE ICC PACS NHSC SPIRE Point Source Photometry
More informationPACS data reduction for the PEP deep extragalactic survey
PACS data reduction for the PEP deep extragalactic survey D. Lutz, P. Popesso, S. Berta and the PEP reduction team Herschel map making workshop Jan 28-31 2013 Ugly! Boring! how do we detect yet more of
More informationPUV3402 LED multiwave photometer A new approach to online process photometry
ABB MEASUREMENT & ANALYTICS WHITE PAPER PUV3402 LED multiwave photometer A new approach to online process photometry The UV LED photometer with a design concept advantage. Measurement made easy PUV3402
More informationSchedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK
2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK 21-23 Campus Road Contact: Mr K G Dove Listerhills Science Park Tel: +44 (0)1274 393857 Bradford Fax: +44 (0)1274 393336 BD7 1HR E-Mail: sales@otc.co.uk
More informationJ. E. Brau, N. B. Sinev, D. M. Strom University of Oregon, Eugene. C. Baltay, H. Neal, D. Rabinowitz Yale University, New Haven
Chronopixe status J. E. Brau, N. B. Sinev, D. M. Strom University of Oregon, Eugene C. Baltay, H. Neal, D. Rabinowitz Yale University, New Haven EE work is contracted to Sarnoff Corporation 1 Outline of
More informationThe Herschel Legacy Data Products
The Herschel Legacy Data Products D. Teyssier, on behalf of the Calibration Scientists at the HSC, NHSC and the Instrument Control Centres (past and present) Instrument and Calibration Scientists Team
More informationColor Measurement with the LSS-100P
Color Measurement with the LSS-100P Color is complicated. This paper provides a brief overview of color perception and measurement. XYZ and the Eye We can model the color perception of the eye as three
More informationWFC3 Thermal Vacuum Testing: UVIS Science Performance Monitor
WFC3 Thermal Vacuum Testing: UVIS Science Performance Monitor H. Bushouse and O. Lupie May 24, 2005 ABSTRACT During WFC3 thermal-vacuum testing in September and October 2004, the UVIS28 test procedure,
More informationFlux Calibration Monitoring: WFC3/IR G102 and G141 Grisms
Instrument Science Report WFC3 2014-01 Flux Calibration Monitoring: WFC3/IR and Grisms Janice C. Lee, Norbert Pirzkal, Bryan Hilbert January 24, 2014 ABSTRACT As part of the regular WFC3 flux calibration
More informationThe CMS ECAL Laser Monitoring System
The CMS ECAL Laser Monitoring System CALOR 2006 XII INTERNATIONAL CONFERENCE on CALORIMETRY in HIGH ENERGY PHYSICS Adi Bornheim California Institute of Technology Chicago, June 8, 2006 Introduction CMS
More informationLow Cost, DC to 500 MHz, 92 db Logarithmic Amplifier AD8307
Low Cost, DC to 500 MHz, 9 db Logarithmic Amplifier AD807 FEATURES Complete multistage logarithmic amplifier 9 db dynamic range: 75 dbm to +7 dbm to 90 dbm using matching network Single supply of.7 V minimum
More informationRadiation Test Report Paul Scherer Institute Proton Irradiation Facility
the Large Hadron Collider project CERN CH-2 Geneva 23 Switzerland CERN Div./Group RadWG EDMS Document No. xxxxx Radiation Test Report Paul Scherer Institute Proton Irradiation Facility Responsibility Tested
More informationHighly Miniaturised Radiation Monitor (HMRM) Status Report. Yulia Bogdanova, Nicola Guerrini, Ben Marsh, Simon Woodward, Rain Irshad
Highly Miniaturised Radiation Monitor (HMRM) Status Report Yulia Bogdanova, Nicola Guerrini, Ben Marsh, Simon Woodward, Rain Irshad HMRM programme aim Aim of phase A/B: Develop a chip sized prototype radiation
More informationDetection Beyond 100µm Photon detectors no longer work ("shallow", i.e. low excitation energy, impurities only go out to equivalent of
Detection Beyond 100µm Photon detectors no longer work ("shallow", i.e. low excitation energy, impurities only go out to equivalent of 100µm) A few tricks let them stretch a little further (like stressing)
More informationGEOMETRIC RECTIFICATION OF EUROPEAN HISTORICAL ARCHIVES OF LANDSAT 1-3 MSS IMAGERY
GEOMETRIC RECTIFICATION OF EUROPEAN HISTORICAL ARCHIVES OF LANDSAT -3 MSS IMAGERY Torbjörn Westin Satellus AB P.O.Box 427, SE-74 Solna, Sweden tw@ssc.se KEYWORDS: Landsat, MSS, rectification, orbital model
More informationStudy of the HD target spin rotations during G14
Study of the HD target spin rotations during G14 A. Deur, deurpam@jlab.org April 18, 2013 1 Introduction During the G14 run, the target spin was reversed using either magnetic field rotations or RF spin
More informationPerformance of H Maser During the EOC Week 29 July to 03 August
Performance of H Maser During the EOC Week 29 July to 03 August ALMA Technical Note Number: 6 Status: FINAL Prepared by: Organization: Date: Anthony Remijan (EOC Program Scientist for Extension and Optimization
More informationExperiment No. 3 Pre-Lab Phase Locked Loops and Frequency Modulation
Experiment No. 3 Pre-Lab Phase Locked Loops and Frequency Modulation The Pre-Labs are informational and although they follow the procedures in the experiment, they are to be completed outside of the laboratory.
More informationT = 4.2 K T = 300 K Drain Current (A) Drain-Source Voltage (V) Drain-Source Voltage (V)
The Institute of Space and Astronautical Science Report SP No.14, December 2000 Evaluation of Cryogenic Readout Circuits with GaAs JFETs for Far-Infrared Detectors By Kenichi Okumura Λ and Norihisa Hiromoto
More informationTowards lower Uncooled IR-FPA system integration cost
Towards lower Uncooled IR-FPA system integration cost Benoit DUPONT 1,2,3, Michel VILAIN 1 1 ULIS, Veurey-Voroise, FRANCE 2 Laboratoire d'electronique de Technologie de l'information, Commissariat à l
More informationFunction Guide for the Fourier Transformation Package SPIRE-UOL-REP
Function Guide for the Fourier Transformation Package Prepared by: Peter Davis (University of Lethbridge) peter.davis@uleth.ca Andres Rebolledo (University of Lethbridge) andres.rebolledo@uleth.ca Trevor
More informationTRANSFER RADIATION THERMOMETER
HEITRONICS Infrarot Messtechnik TRANSFER RADIATION THERMOMETER COVERING THE TEMPERATURE RANGE OF -50 C TO 1000 C Contents General Information Temperature Ranges and Spectral Ranges Field of Views and SSE
More informationOP400A. Quad Low-Offset, Low-Power Operational Amplifier FEATURES: DESCRIPTION: Memory. Logic Diagram (One Amplifier)
Quad Low-Offset, Low-Power Operational Amplifier v+ BIAS OUT VOLTAGE LIMITING NETWORK +IN -IN v- Logic Diagram (One Amplifier) FEATURES: RAD-PAK technology-hardened against natural space radiation Total
More informationUse of the Shutter Blade Side A for UVIS Short Exposures
Instrument Science Report WFC3 2014-009 Use of the Shutter Blade Side A for UVIS Short Exposures Kailash Sahu, Sylvia Baggett, J. MacKenty May 07, 2014 ABSTRACT WFC3 UVIS uses a shutter blade with two
More informationWavelength Calibration Accuracy of the First-Order CCD Modes Using the E1 Aperture
Wavelength Calibration Accuracy of the First-Order CCD Modes Using the E1 Aperture Scott D. Friedman August 22, 2005 ABSTRACT A calibration program was carried out to determine the quality of the wavelength
More informationUpdate on SPIRE Photometer Pipeline
Update on SPIRE Photometer Pipeline Kevin Xu NHSC/IPAC on behalf of the SPIRE ICC page 1 What s New in HIPE 9.1 Improved map-making, with destriper in standard processing Electrical crosstalk correction
More informationTiming accuracy of the GEO 600 data acquisition system
INSTITUTE OF PHYSICS PUBLISHING Class. Quantum Grav. 1 (4) S493 S5 CLASSICAL AND QUANTUM GRAVITY PII: S64-9381(4)6861-X Timing accuracy of the GEO 6 data acquisition system KKötter 1, M Hewitson and H
More informationM Hewitson, K Koetter, H Ward. May 20, 2003
A report on DAQ timing for GEO 6 M Hewitson, K Koetter, H Ward May, Introduction The following document describes tests done to try and validate the timing accuracy of GEO s DAQ system. Tests were done
More informationNoise Measurements Using a Teledyne LeCroy Oscilloscope
Noise Measurements Using a Teledyne LeCroy Oscilloscope TECHNICAL BRIEF January 9, 2013 Summary Random noise arises from every electronic component comprising your circuits. The analysis of random electrical
More informationEpic Detector Matrices
Epic Detector Matrices June 2013 - SAS workshop Why do we need detector matrices? Initial source spectrum Source flux How do we use them? Spectral fitting: - xspec, CIAO Try a spectral model and see if
More informationNHSC/PACS Web Tutorials Running the PACS Spectrometer pipeline for unchopped line mode. PACS-303 Level 0 to 2 processing
NHSC/PACS s Running the PACS Spectrometer pipeline for unchopped line mode PACS-303 Level 0 to 2 processing Prepared by Dario Fadda April 2014 page 1 Introduction This tutorial will guide you through the
More informationEXAM Amplifiers and Instrumentation (EE1C31)
DELFT UNIVERSITY OF TECHNOLOGY Faculty of Electrical Engineering, Mathematics and Computer Science EXAM Amplifiers and Instrumentation (EE1C31) April 18, 2017, 9.00-12.00 hr This exam consists of four
More informationMini Workshop Interferometry. ESO Vitacura, 28 January Presentation by Sébastien Morel (MIDI Instrument Scientist, Paranal Observatory)
Mini Workshop Interferometry ESO Vitacura, 28 January 2004 - Presentation by Sébastien Morel (MIDI Instrument Scientist, Paranal Observatory) MIDI (MID-infrared Interferometric instrument) 1st generation
More informationExtra slides. 10/05/2011 SAC meeting IRAM Grenoble 1
Extra slides 10/05/2011 SAC meeting IRAM Grenoble 1 New NIKA spectral responses Bands spectral response obtained with a Martin-Puplett interferometer 10/05/2011 SAC meeting IRAM Grenoble 2 New NIKA backend
More informationCerro Tololo Inter-American Observatory. CHIRON manual. A. Tokovinin Version 2. May 25, 2011 (manual.pdf)
Cerro Tololo Inter-American Observatory CHIRON manual A. Tokovinin Version 2. May 25, 2011 (manual.pdf) 1 1 Overview Calibration lamps Quartz, Th Ar Fiber Prism Starlight GAM mirror Fiber Viewer FEM Guider
More informationVATTSpec Instructions Rev. 10/23/2015
VATTSpec Instructions Rev. 10/23/2015 Introduction VATTSpec is a medium resolution CCD range spectrograph with a skinny chip having excellent cosmetics. Its UA ITL chip, Serial Number 8228, has a gain
More informationEVLA System Commissioning Results
EVLA System Commissioning Results EVLA Advisory Committee Meeting, March 19-20, 2009 Rick Perley EVLA Project Scientist t 1 Project Requirements EVLA Project Book, Chapter 2, contains the EVLA Project
More informationInterpixel Capacitance in the IR Channel: Measurements Made On Orbit
Interpixel Capacitance in the IR Channel: Measurements Made On Orbit B. Hilbert and P. McCullough April 21, 2011 ABSTRACT Using high signal-to-noise pixels in dark current observations, the magnitude of
More informationProgram for UV Intercomparison 2014 in Davos:
Program for UV Intercomparison 2014 in Davos: June 2014 Date: 7 16 July 2014 Location: PMOD/WRC Davos Switzerland. Information Update: http://projects.pmodwrc.ch/env03/index.php/8-emrp-uv/project/24- intercomparison-2014
More informationDetection of the mm-wave radiation using a low-cost LWIR microbolometer camera from a multiplied Schottky diode based source
Detection of the mm-wave radiation using a low-cost LWIR microbolometer camera from a multiplied Schottky diode based source Basak Kebapci 1, Firat Tankut 2, Hakan Altan 3, and Tayfun Akin 1,2,4 1 METU-MEMS
More informationTerraSAR-X Calibration Status 2 Years in Flight
2 Years in Flight Dirk Schrank, Marco Schwerdt, Markus Bachmann, Björn Döring, Clemens Schulz November 2009 CEOS 09 VG 1 Calibration Tasks Performed 2009 Introduction Challenge Schedule Re-Calibration
More informationFast Solar Polarimeter
Fast Solar Polarimeter A. Feller, F. Iglesias, K. Nagaraju, S. K. Solanki Max Planck Institute for Solar System Research and colleagues from the Max Planck semiconductor lab A. Feller FSP IAUS 305 1 /
More informationInstruction manual for T3DS software. Tool for THz Time-Domain Spectroscopy. Release 4.0
Instruction manual for T3DS software Release 4.0 Table of contents 0. Setup... 3 1. Start-up... 5 2. Input parameters and delay line control... 6 3. Slow scan measurement... 8 4. Fast scan measurement...
More informationFirst Summary of Neptune Beam Measurements
First Summary of Neptune Beam Measurements Background subtraction with modified telescope normalization for asymmetric chopping sans nod works very well; offset in Gaussian fit is
More informationFully depleted, thick, monolithic CMOS pixels with high quantum efficiency
Fully depleted, thick, monolithic CMOS pixels with high quantum efficiency Andrew Clarke a*, Konstantin Stefanov a, Nicholas Johnston a and Andrew Holland a a Centre for Electronic Imaging, The Open University,
More informationI1A Magnetic Field Transducers
DESCRIPTION: The I1A denotes a range of SENIS Magnetic Fieldto-Voltage Transducers with integrated 1-axis Hall Probe. It measures magnetic fields perpendicular to the probe plane (By). The Hall Probe contains
More informationIRACSIM: Simula.ng IRAC Data
IRACSIM: Simula.ng IRAC Data Jim Ingalls (Spitzer Science Center) K. Mighell (NOAO) Components of a Data Simulator Components of a Data Simulator Source Variation Components of a Data Simulator Source
More informationAbstract. Preface. Acknowledgments
Contents Abstract Preface Acknowledgments iv v vii 1 Introduction 1 1.1 A Very Brief History of Visible Detectors in Astronomy................ 1 1.2 The CCD: Astronomy s Champion Workhorse......................
More informationSOAR Integral Field Spectrograph (SIFS): Call for Science Verification Proposals
Published on SOAR (http://www.ctio.noao.edu/soar) Home > SOAR Integral Field Spectrograph (SIFS): Call for Science Verification Proposals SOAR Integral Field Spectrograph (SIFS): Call for Science Verification
More informationNHSC/PACS Webinar PACS Point Source Photometry. Aperture Photometry and Photometric Uncertain5es. Roberta Paladini NHSC PACS.
NHSC/PACS Webinar PACS Point Source Photometry Aperture Photometry and Photometric Uncertain5es Roberta Paladini July 2012 1 Content I Point- source aperture photometry in HIPE II Calibra=on errors III
More informationVERY LARGE TELESCOPE
EUROPEAN SOUTHERN OBSERVATORY VERY LARGE TELESCOPE NAOS-CONICA Calibration Plan Doc. No. VLT-PLA-ESO-14200-2664 Issue 80 March 03, 2007 N. Ageorges, C. Lidman Prepared..........................................
More informationWFC3 SMOV Program 11433: IR Internal Flat Field Observations
Instrument Science Report WFC3 2009-42 WFC3 SMOV Program 11433: IR Internal Flat Field Observations B. Hilbert 27 October 2009 ABSTRACT We have analyzed the internal flat field behavior of the WFC3/IR
More informationFringe Parameter Estimation and Fringe Tracking. Mark Colavita 7/8/2003
Fringe Parameter Estimation and Fringe Tracking Mark Colavita 7/8/2003 Outline Visibility Fringe parameter estimation via fringe scanning Phase estimation & SNR Visibility estimation & SNR Incoherent and
More informationRFI Monitoring and Analysis at Decameter Wavelengths. RFI Monitoring and Analysis
Observatoire de Paris-Meudon Département de Radio-Astronomie CNRS URA 1757 5, Place Jules Janssen 92195 MEUDON CEDEX " " Vincent CLERC and Carlo ROSOLEN E-mail adresses : Carlo.rosolen@obspm.fr Vincent.clerc@obspm.fr
More informationSTIS CCD Saturation Effects
SPACE TELESCOPE SCIENCE INSTITUTE Operated for NASA by AURA Instrument Science Report STIS 2015-06 (v1) STIS CCD Saturation Effects Charles R. Proffitt 1 1 Space Telescope Science Institute, Baltimore,
More informationDInGaAs Series. Spectroscopy Detectors PROFESSIONAL DINGAAS SERIES DETECTORS DINGAAS SERIES TE COOLED INGAAS DETECTORS
DInGaAs Series DINGAAS SERIES COOLED INGAAS DECTORS cooled InGaAs detectors, spectral range from 0.8 μm to 2.6 μm. DInGaAs(x) detectors have the same construction. And x-1700/ 1900/ 2200/ 2400/ 2600 all
More informationXMM OM Serendipitous Source Survey Catalogue (XMM-SUSS2.1)
XMM OM Serendipitous Source Survey Catalogue (XMM-SUSS2.1) 1 Introduction The second release of the XMM OM Serendipitous Source Survey Catalogue (XMM-SUSS2) was produced by processing the XMM-Newton Optical
More information771 Series LASER SPECTRUM ANALYZER. The Power of Precision in Spectral Analysis. It's Our Business to be Exact! bristol-inst.com
771 Series LASER SPECTRUM ANALYZER The Power of Precision in Spectral Analysis It's Our Business to be Exact! bristol-inst.com The 771 Series Laser Spectrum Analyzer combines proven Michelson interferometer
More informationPhotometry. Variable Star Photometry
Variable Star Photometry Photometry One of the most basic of astronomical analysis is photometry, or the monitoring of the light output of an astronomical object. Many stars, be they in binaries, interacting,
More informationLINEARPYROMETER LP4. Technical Documentation KE November TN
1 LINEARPYROMETER LP4 Technical Documentation KE 256-6.2007 November 2010 5-TN-1622-100 2 1. General Description With the Linearpyrometer Type LP4 a measuring instrument has been made available for pyrometric
More informationComparison of the Josephson Voltage Standards of the NIMT and the BIPM
Comparison of the Josephson Voltage Standards of the NIMT and the BIPM (part of the ongoing BIPM key comparison BIPM.EM-K10.b) S. Solve, R. Chayramy, M. Stock, S. Pimsut* and N. Rujirat* Bureau International
More informationIT FR R TDI CCD Image Sensor
4k x 4k CCD sensor 4150 User manual v1.0 dtd. August 31, 2015 IT FR 08192 00 R TDI CCD Image Sensor Description: With the IT FR 08192 00 R sensor ANDANTA GmbH builds on and expands its line of proprietary
More informationThe CMS ECAL Laser Monitoring System
The CMS ECAL Laser Monitoring System IPRD 2008 11th Topical Seminar On Innovative Particle and Radiation Detectors Adi Bornheim California Institute of Technology On behalf of the CMS ECAL Collaboration
More informationPre-Launch Radiometric Calibration of the S-NPP and JPSS-1 VIIRS Day/Night Bands
Pre-Launch Radiometric Calibration of the S-NPP and JPSS-1 VIIRS Day/Night Bands Thomas Schwarting Science Systems and Applications, Lanham, MD Jeff McIntire, Science Systems and Applications, Lanham,
More informationAdvanced Test Equipment Rentals ATEC (2832)
Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) Electric and Magnetic Field Measurement For Isotropic Measurement of Magnetic and Electric Fields Evaluation of Field
More informationUsers Manual. DIGITAL MICROVOLTMETER Model: DMV-001
Users Manual DIGITAL MICROVOLTMETER Model: DMV-001 Manufactured by.. Scientific Equipment & Services 358/1, New Adarsh Nagar, Roorkee - 247 667, UA, INDIA Ph.: +91-1332-272852, 277118 Fax: +91-1332-274831
More informationWHITE PAPER. Programmable narrow-band filtering using the WaveShaper 1000S and WaveShaper 4000S. Abstract. 2. WaveShaper Optical Design
WHITE PAPER Programmable narrow-band filtering using the WaveShaper 1S and WaveShaper 4S Abstract The WaveShaper family of Programmable Optical Processors provide unique capabilities for the manipulation
More informationSpur Detection, Analysis and Removal Stable32 W.J. Riley Hamilton Technical Services
Introduction Spur Detection, Analysis and Removal Stable32 W.J. Riley Hamilton Technical Services Stable32 Version 1.54 and higher has the capability to detect, analyze and remove discrete spectral components
More informationThree Axis Magnetic Field Transducer x-h3x-xx_e3d-2.5khz-0.1-2t
Three Axis Magnetic Field Transducer x-h3x-xx_e3d-2.5khz-.1-2t Test results Version 1. S.D. E3D 2T test results 26.7.6 Page 1 of 8 1. Table of contents 1. Table of contents... 2 2. Noise power spectral
More informationDetermination of Uncertainty for Dielectric Properties Determination of Printed Circuit Board Material
Determination of Uncertainty for Dielectric Properties Determination of Printed Circuit Board Material Marko Kettunen, Kare-Petri Lätti, Janne-Matti Heinola, Juha-Pekka Ström and Pertti Silventoinen Lappeenranta
More informationNON-AMPLIFIED HIGH SPEED PHOTODETECTOR USER S GUIDE
NON-AMPLIFIED HIGH SPEED PHOTODETECTOR USER S GUIDE Thank you for purchasing your Non-amplified High Speed Photodetector. This user s guide will help answer any questions you may have regarding the safe
More informationGA GHz. Digital Spectrum Analyzer
Digital Spectrum Analyzer GA4063 3GHz Professional Performance Robust Measurement features High frequency stability Easy- to-use User Interface Compact size, Light weight, Portable design www.attenelectronics.com
More informationGlobal Erratum for Kepler Q0-Q17 & K2 C0-C5 Short-Cadence Data
Global Erratum for Kepler Q0-Q17 & K2 C0-C5 Short-Cadence Data KSCI-19080-002 23 March 2016 NASA Ames Research Center Moffett Field, CA 94035 Prepared by: Date Douglas Caldwell, Instrument Scientist Prepared
More informationOrder Overlap. A single wavelength constructively interferes in several directions A given direction can receive multiple wavelengths.
Order Overlap A single wavelength constructively interferes in several directions A given direction can receive multiple wavelengths. Spectral Calibration TripleSpec Users Guide Spectral Calibration TripleSpec
More informationBetter Absolute Accuracy and Repeatability, Shorter Test Time With New Servo Improvement Package
Better Absolute Accuracy and Repeatability, Shorter Test Time With New Servo Improvement Package Increase servo positioning accuracy to.4 nm (.16 µinch) *, 1 σ Achieve higher accuracy and repeatability
More informationYour first NMR measurement
Your first NMR measurement Introduction Select 10mM water in D2O as NMR sample. The NMR spectrum of such sample consists of only two signals: the water signal and the peak of the reference (TSP). Follow
More informationCARMA Memorandum Series #14 1
CARMA Memorandum Series #14 1 Stability of BIMA antenna position solutions J. R. Forster Hat Creek Observatory, University of California, Berkeley, CA, 94720 September 25, 2003 ABSTRACT We review the stability
More informationARRAY CONTROLLER REQUIREMENTS
ARRAY CONTROLLER REQUIREMENTS TABLE OF CONTENTS 1 INTRODUCTION...3 1.1 QUANTUM EFFICIENCY (QE)...3 1.2 READ NOISE...3 1.3 DARK CURRENT...3 1.4 BIAS STABILITY...3 1.5 RESIDUAL IMAGE AND PERSISTENCE...4
More informationmetcon meteorologieconsultgmbh, Instruments for Atmospheric Research W1aa_Feb_2017_1.doc 1 -
metcon meteorologieconsultgmbh, Instruments for Atmospheric Research W1aa_Feb_2017_1.doc 1 - ACTINIC FLUX SPECTRAL RADIOMETERS Ultra-fast CCD-Detector Spectrometer, UVB enhanced Cooled CCD, 512 pixel *
More informationGA GHz. Digital Spectrum Analyzer
Digital Spectrum Analyzer GA4063 3GHz Professional Performance Robust Measurement features High frequency stability Easy- to-use User Interface Compact size, Light weight, Portable design www.attenelectronics.com
More informationNON-AMPLIFIED PHOTODETECTOR USER S GUIDE
NON-AMPLIFIED PHOTODETECTOR USER S GUIDE Thank you for purchasing your Non-amplified Photodetector. This user s guide will help answer any questions you may have regarding the safe use and optimal operation
More informationProcessing ACA Monitor Window Data
Processing ACA Monitor Window Data CIAO 3.4 Science Threads Processing ACA Monitor Window Data 1 Table of Contents Processing ACA Monitor Window Data CIAO 3.4 Background Information Get Started Obtaining
More informationPerformance status of IASI on MetOp-A and MetOp-B
Performance status of IASI on MetOp-A and MetOp-B E. Jacquette (1), E. Péquignot (1), J. Chinaud (1), C. Maraldi (1), D. Jouglet (1), S. Gaugain (1), L. Buffet (1), C. Villaret (1), C. Larigauderie (1),
More informationPhotometry from Herschel maps Ivan Valtchanov
Photometry from Herschel maps Ivan Valtchanov SPIRE Instrument and Calibration Scientist Herschel Science Centre, ESAC, ESA Herschel and ALMA: PACS 1. Two broad-band photometers with Herschel: a. PACS:
More informationIDEAS+ WP3520 Calibration and data quality toolbox. July 2016 Steve Mackin James Warner
IDEAS+ WP3520 Calibration and data quality toolbox July 2016 Steve Mackin James Warner Proposition : Every image contains the same information Railroad Valley, Nevada London, UK Rationale for the project
More information