RF Pulse Design. Multi-dimensional Excitation II. M229 Advanced Topics in MRI Kyung Sung, Ph.D Class Business

Size: px
Start display at page:

Download "RF Pulse Design. Multi-dimensional Excitation II. M229 Advanced Topics in MRI Kyung Sung, Ph.D Class Business"

Transcription

1 RF Pulse Design Multi-dimensional Excitation II M229 Advanced Topics in MRI Kyung Sung, Ph.D Class Business - Homework 1 will be due on 4/26 - Office hours Instructors: Fri noon TAs: Thursday 3-5pm s beforehand would be helpful - Papers and Slides

2 Today s Topics - Recap of excitation k-space - 2D excitation pulses 2D EPI pulse design Spatial-spectral pulse design - Matlab exercise Summary for Excitation k-space

3 Small-Tip Approximation B1(t) T Gz(t) k(t,t0) t1 t2 t3 t4 t5 t6 t7 Small-Tip-Angle Solution as a k-space Integral k-space weighting k-space sampling

4 2D Pulse Design 1. Choose a k-space trajectory 2. Choose a weighting function 3. Design the RF pulse 2D Spatial Pulse Design - EPI Non-isotropic resolution Sidelobes in one dimension Spectral-spatial pulses - Spiral Unity aspect ratio Minimum length Circular sidelobe

5 2D EPI Pulse Design Designing EPI k-space Trajectory - Ideally, an EPI trajectory scans a 2D raster in k- space Resolution? / FOV?

6 Designing EPI k-space Trajectory - Resolution: - FOV = 1/ ky - Ghost FOV = FOV/2 Eddy currents & delays produce this Designing EPI k-space Trajectory - Refocusing gradients Returns to origin at the end of pulse

7 Designing EPI Gradients - Designing readout lobes and blips Flat-top only design RF only played during flat part (simpler) To the board...

8 Designing EPI Gradients - Easy to get k-space coverage in ky - Hard to get k-space coverage in kx - We can get more k-space coverage by making blips narrower playing RF during part of ramps Blipped EPI - Rectilinear scan of k-space - Most efficient EPI trajectory - Common choice for spatial pulses - Sensitive to eddy currents and gradient delays

9 Blipped EPI G x,g y Gradient Waveforms k y k x 2k y,max k-space Trajectory 2k x,max Continuous EPI - Non-uniform k-space coverage - Need to oversample to avoid side lobes Less efficient than blipped - Sensitive to eddy currents and gradient delays Only choice for spectral-spatial pulses

10 Continuous EPI G x,g y Gradient Waveforms k y k x 2k y,max k-space Trajectory 2k x,max Flyback EPI - Can be blipped or continuous - Less efficient since retraces not used (depends on gradient system) - Almost completely immune to eddy currents and gradient delays

11 Flyback EPI G x,g y Gradient Waveforms k y Retrace k x 2k y,max k-space Trajectory 2k x,max Designing 2D EPI Spatial Pulses - Two major options General approach, same as 2D spiral pulses Seperable, product design (easier) - General approach Choose EPI k-space trajectory Design gradient waveforms Design W(k), k-space weighting Design B1(t)

12 Separable, Product Design - Assume, AS(ky): weighting in the slow, blipped direction AF(kx): weighting in the fast oscillating direction - Each impulse corresponds to a pulse in the fast direction, AF(kx) Separable, Product Design

13

14 Spectral Spatial Pulse Spatial-Spectral Pulses - 2D pulses selective in space and frequency - Excite a slice at a limited band of frequencies - Clinical applications High speed imaging (spiral/epi) Robust lipid suppression Spectroscopic imaging (MRSI)

15 Basic Idea - Gy gradient simply establishes a linear relationship between position and frequency Spatial selectivity in y Frequency selectivity Basic Idea

16 Basic Idea f kf Note that kf is time! Spectral Pulses

17 Add Spatial Selectivity to a Spectral Pulse - Hard pulses Hard pulses B1(t) B1(t) t Slice selective subpulses t Gz(t) Each sub pulse starts and ends at kz = 0 To the board...

18 Flyback Design 1.1 ms sublobes 8 sublobes 250 Hz spectral passband 15 ms length Water Lipid Opposed Null Design 2.2 ms sublobes 8 sublobes 250 Hz spectral passband 13.2 ms length Water Lipid

19 True Null Design 1.1 ms sublobes 16 sublobes 250 Hz spectral passband 15ms length Water Lipid Source of Bipolar Sidelobes Interference between excitations from positive and negative gradient lobes position [mm] position [mm] position [mm] resonant frequency [Hz] resonant frequency [Hz] resonant frequency [Hz]

20 Matlab Exercise Windowed Sinc RF Pulse

21 RF Pulse Scaling RF Pulse Scaling

22 Bloch Simulation Slice Thickness - Pulse duration = 1 ms - TBW = 4 - Gz = 1 G/cm ɣ/2π = khz/g

23 Summary - Adiabatic Pulse Design - 2D Pulse Design Examples: - EPI pulse design - Spatial-Spectral Pulses - Matlab Exercise - Homework 2-2D EPI design - SPSP design - Next time: Pulse sequences Thanks! Kyung Sung, PhD ksung@mednet.ucla.edu

EE469B: Assignment 4 Solutions

EE469B: Assignment 4 Solutions EE469B Fall 26-7 RF Pulse Design for MRI EE469B: Assignment 4 Solutions Due Thursday Oct 27. True Null/Flyback Spectral-Spatial Pulses True null and flyback designs are very closely related. In this problem

More information

EE225E/BIOE265 Spring 2012 Principles of MRI. Assignment 7. Due March 16, 2012

EE225E/BIOE265 Spring 2012 Principles of MRI. Assignment 7. Due March 16, 2012 EE225E/BIOE265 Spring 2012 Principles of MRI Miki Lustig Assignment 7 Due March 16, 2012 1. From Midterm I 2010: You ve just programmed up your first 2DFT pulse sequence, and are trying it out on the scanner.

More information

EE469B: Assignment 1 Solutions

EE469B: Assignment 1 Solutions EE469B Fall 26-7 RF Pulse Design for MRI EE469B: Assignment Solutions Due Thursday Oct 6 Introduction This assignment concerns typical Fourier transform designs of excitation pulses. This includes designing

More information

EE225E/BIOE265 Spring 2014 Principles of MRI. Assignment 6. Due Friday March 7th, 2014, Self Grading Due Monday March 10th, 2014

EE225E/BIOE265 Spring 2014 Principles of MRI. Assignment 6. Due Friday March 7th, 2014, Self Grading Due Monday March 10th, 2014 EE225E/BIOE265 Spring 2014 Principles of MRI Miki Lustig 1. Read Nishimura Ch. 6 Assignment 6 Due Friday March 7th, 2014, Self Grading Due Monday March 10th, 2014 2. Nishimura assignment 6.5 3. Mimimum-Phase

More information

RF Pulse Toolkit: Application Specific Design

RF Pulse Toolkit: Application Specific Design RF Pulse Toolkit: Application Specific Design William A Grissom Department of Biomedical Engineering, Vanderbilt University, Nashville, TN, USA will.grissom@vanderbilt.edu Introduction RF excitation is

More information

Background (~EE369B)

Background (~EE369B) Background (~EE369B) Magnetic Resonance Imaging D. Nishimura Overview of NMR Hardware Image formation and k-space Excitation k-space Signals and contrast Signal-to-Noise Ratio (SNR) Pulse Sequences 13

More information

EE225E/BIOE265 Spring 2011 Principles of MRI. Assignment 6 Solutions. (y 0 + vt) dt. 2 y 0T + 3 )

EE225E/BIOE265 Spring 2011 Principles of MRI. Assignment 6 Solutions. (y 0 + vt) dt. 2 y 0T + 3 ) EE225E/BIOE265 Spring 211 Principles of MRI Miki Lustig Handout Assignment 6 Solutions 1. Nishimura 6.7 (Thanks Galen!) a) After the 9 y pulse, the spin is in the ˆx direction (using left-handed rotations).

More information

Half-Pulse Excitation Pulse Design and the Artifact Evaluation

Half-Pulse Excitation Pulse Design and the Artifact Evaluation Half-Pulse Excitation Pulse Design and the Artifact Evaluation Phillip Cho. INRODUCION A conventional excitation scheme consists of a slice-selective RF excitation followed by a gradient-refocusing interval

More information

k y 2k y,max k x 2k x,max

k y 2k y,max k x 2k x,max EE225E/BIOE265 Spring 2012 Principles of MRI Miki Lustig Assignment 5 Due Feb 26, 2012 1. Finish reading Nishimura Ch. 5. 2. For the 16 turn spiral trajectory, plotted below, what is the a) Spatial resolution,

More information

Magnetic Resonance Imaging (MRI)

Magnetic Resonance Imaging (MRI) C. A. Bouman: Digital Image Processing - February 15, 2 1 Magnetic Resonance Imaging (MRI) Can be very high resolution No radiation exposure Very flexible and programable Tends to be expensive, noisy,

More information

NIH Public Access Author Manuscript Magn Reson Med. Author manuscript; available in PMC 2010 July 21.

NIH Public Access Author Manuscript Magn Reson Med. Author manuscript; available in PMC 2010 July 21. NIH Public Access Author Manuscript Published in final edited form as: Magn Reson Med. 2010 April ; 63(4): 1092 1097. doi:10.1002/mrm.22223. Spatially Varying Fat-Water Excitation Using Short 2DRF Pulses

More information

EE469B: Assignment 2 Solutions

EE469B: Assignment 2 Solutions EE469B Fall 26-7 RF Pulse Design for MRI EE469B: Assignment 2 s Due Thursday Oct 3 Introduction This assignment concerns the design of small-tip-angle 2D excitation pulses based on spiral k-space trajectories.

More information

Encoding of inductively measured k-space trajectories in MR raw data

Encoding of inductively measured k-space trajectories in MR raw data Downloaded from orbit.dtu.dk on: Apr 10, 2018 Encoding of inductively measured k-space trajectories in MR raw data Pedersen, Jan Ole; Hanson, Christian G.; Xue, Rong; Hanson, Lars G. Publication date:

More information

Pulse Sequences: Rapid Gradient Echo

Pulse Sequences: Rapid Gradient Echo Pulse Sequences: Rapid Gradient Echo M229 Advanced Topics in MRI Holden H. Wu, Ph.D. 2018.04.17 Department of Radiological Sciences David Geffen School of Medicine at UCLA Class Business Office hours -

More information

RAD 229: MRI Signals and Sequences

RAD 229: MRI Signals and Sequences RAD 229: MRI Signals and Sequences Brian Hargreaves All notes are on the course website web.stanford.edu/class/rad229 Course Goals Develop Intuition Understand MRI signals Exposure to numerous MRI sequences

More information

Echo-Planar Imaging for a 9.4 Tesla Vertical-Bore Superconducting Magnet Using an Unshielded Gradient Coil

Echo-Planar Imaging for a 9.4 Tesla Vertical-Bore Superconducting Magnet Using an Unshielded Gradient Coil Magn Reson Med Sci, Vol. XX, No. X, pp. XXX XXX, 2015 2016 Japanese Society for Magnetic Resonance in Medicine TECHNICAL NOTE by J-STAGE doi:10.2463/mrms.tn.2015-0123 Echo-Planar Imaging for a 9.4 Tesla

More information

RF pulse design and the Small Tip Angle Approximation

RF pulse design and the Small Tip Angle Approximation RF pulse design and the Small Tip Angle Approximation Dr Shaihan J Malik Lecturer in Imaging Sciences Division of Imaging Sciences & Biomedical Engineering King s College London shaihan.malik@kcl.ac.uk

More information

2015 Spin echoes and projection imaging

2015 Spin echoes and projection imaging 1. Spin Echoes 1.1 Find f0, transmit amplitudes, and shim settings In order to acquire spin echoes, we first need to find the appropriate scanner settings using the FID GUI. This was all done last week,

More information

Module 2. Artefacts and Imaging Optimisation for single shot methods. Content: Introduction. Phase error. Phase bandwidth. Chemical shift review

Module 2. Artefacts and Imaging Optimisation for single shot methods. Content: Introduction. Phase error. Phase bandwidth. Chemical shift review MRES 7005 - Fast Imaging Techniques Module 2 Artefacts and Imaging Optimisation for single shot methods Content: Introduction Phase error Phase bandwidth Chemical shift review Chemical shift in pixels

More information

H 2 O and fat imaging

H 2 O and fat imaging H 2 O and fat imaging Xu Feng Outline Introduction benefit from the separation of water and fat imaging Chemical Shift definition of chemical shift origin of chemical shift equations of chemical shift

More information

Reconstruction of Non-Cartesian MRI Data

Reconstruction of Non-Cartesian MRI Data G16.448 Practical Magnetic Resonance Imaging II Sacler Institute of Biomedical Sciences New Yor Universit School of Medicine Reconstruction of Non-Cartesian MRI Data Ricardo Otazo PhD ricardo.otazo@numc.org

More information

Hadamard Slice Encoding for Reduced-FOV Diffusion- Weighted Imaging

Hadamard Slice Encoding for Reduced-FOV Diffusion- Weighted Imaging FULL PAPER Magnetic Resonance in Medicine 00:00 00 (2013) Hadamard Slice Encoding for Reduced-FOV Diffusion- Weighted Imaging Emine Ulku Saritas, 1,2,3 * Daeho Lee, 1 Tolga Çukur, 1,2,3 Ajit Shankaranarayanan,

More information

Spiral MRI on a 9.4T Vertical-bore Superconducting Magnet Using Unshielded and Self-shielded Gradient Coils

Spiral MRI on a 9.4T Vertical-bore Superconducting Magnet Using Unshielded and Self-shielded Gradient Coils Magn Reson Med Sci doi:10.2463/mrms.tn.2016-0049 Published Online: March 27, 2017 TECHNICAL NOTE Spiral MRI on a 9.4T Vertical-bore Superconducting Magnet Using Unshielded and Self-shielded Gradient Coils

More information

Liver imaging beyond expectations with Ingenia

Liver imaging beyond expectations with Ingenia Publication for the Philips MRI Community Issue 47 2012/3 Liver imaging beyond expectations with Ingenia Contributed by John Penatzer, RT, MR clinical product specialist, Cleveland, OH, USA Publication

More information

MR in RTP. MR Data for Treatment Planning: Spatial Accuracy Issues, Protocol Optimization, and Applications (Preview of TG117 Report) Acknowledgements

MR in RTP. MR Data for Treatment Planning: Spatial Accuracy Issues, Protocol Optimization, and Applications (Preview of TG117 Report) Acknowledgements MR Data for Treatment Planning: Issues, Protocol Optimization, and s (Preview of TG117 Report) Debra H. Brinkmann Mayo Clinic, Rochester MN Acknowledgements TG-117 Use of MRI Data in Treatment Planning

More information

A k-space Analysis of Small-Tip-Angle Excitation

A k-space Analysis of Small-Tip-Angle Excitation JOURNAL OF MAGNETIC RESONANCE 81,43-56 ( 1989) A k-space Analysis of Small-Tip-Angle Excitation JOHNPAULY,DWIGHTNISHIMURA,ANDALBERTMACOVSKI Information Systems Laboratory. Stanford University, Stanford,

More information

MR in Tx Planning. Acknowledgements. Outline. Overview MR in RTP

MR in Tx Planning. Acknowledgements. Outline. Overview MR in RTP MR Data for Treatment Planning and Stereotactic Procedures: Sources of Distortion, Protocol Optimization, and Assessment (Preview of TG117 Report) Debra H. Brinkmann Mayo Clinic, Rochester MN Acknowledgements

More information

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia M R I Physics Course Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia M R I Physics Course Magnetic Resonance Imaging Spatial

More information

Improving high-field MRI using parallel excitation

Improving high-field MRI using parallel excitation review Improving high-field MRI using parallel excitation MRI at high magnetic field strengths promises to deliver clearer images of the body s structure and function. However, high-field MRI currently

More information

System/Imaging Imperfections

System/Imaging Imperfections System/Imaging Imperfections B0 variations: Shim, Susceptibility B1 variations: Transmit, Receive Gradient Imperfections: Non-linearities Delays and Eddy currents Concomitant terms 1 B0 Variations - Off-Resonance

More information

MARP. MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005

MARP. MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005 ACR MRI accreditation program MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005 Carl R. Keener, Ph.D., DABMP, DABR keener@marpinc.com MARP Medical & Radiation Physics,

More information

(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010

(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010 (N)MR Imaging Lab Course Script FMP PhD Autumn School Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder Date: November 3rd, 2010 1 Purpose: Understanding the basic principles of MR imaging

More information

MRI & NMR spectrometer

MRI & NMR spectrometer AMOS MRI & NMR spectrometer The AMOS Spectrometer is a highly modular and flexible unit that provides the ability to customize synchronized configurations for preclinical and clinical MR applications.

More information

Noninvasive Blood Flow Mapping with Arterial Spin Labeling (ASL) Paul Kyu Han and Sung-Hong Park

Noninvasive Blood Flow Mapping with Arterial Spin Labeling (ASL) Paul Kyu Han and Sung-Hong Park Noninvasive Blood Flow Mapping with Arterial Spin Labeling (ASL) Paul Kyu Han and Sung-Hong Park Department of Bio and Brain Engineering, Korea Advanced Institute of Science and Technology (KAIST), Daejeon,

More information

Fast Joint design of RF and Gradient waveforms for MRI parallel excitation

Fast Joint design of RF and Gradient waveforms for MRI parallel excitation Fast Joint design of RF and Gradient waveforms for MRI parallel excitation by Daehyun Yoon A dissertation submitted in partial fulfillment of the requirements for the degree of Doctor of Philosophy (Electrical

More information

2014 M.S. Cohen all rights reserved

2014 M.S. Cohen all rights reserved 2014 M.S. Cohen all rights reserved mscohen@g.ucla.edu IMAGE QUALITY / ARTIFACTS SYRINGOMYELIA Source http://gait.aidi.udel.edu/res695/homepage/pd_ortho/educate/clincase/syrsco.htm Surgery is usually recommended

More information

Cardiac MR. Dr John Ridgway. Leeds Teaching Hospitals NHS Trust, UK

Cardiac MR. Dr John Ridgway. Leeds Teaching Hospitals NHS Trust, UK Cardiac MR Dr John Ridgway Leeds Teaching Hospitals NHS Trust, UK Cardiac MR Physics for clinicians: Part I Journal of Cardiovascular Magnetic Resonance 2010, 12:71 http://jcmr-online.com/content/12/1/71

More information

Lab 8 6.S02 Spring 2013 MRI Projection Imaging

Lab 8 6.S02 Spring 2013 MRI Projection Imaging 1. Spin Echos 1.1 Find f0, TX amplitudes, and shim settings In order to acquire spin echos, we first need to find the appropriate scanner settings using the FID GUI. This was all done last week, but these

More information

Antennas 1. Antennas

Antennas 1. Antennas Antennas Antennas 1! Grading policy. " Weekly Homework 40%. " Midterm Exam 30%. " Project 30%.! Office hour: 3:10 ~ 4:00 pm, Monday.! Textbook: Warren L. Stutzman and Gary A. Thiele, Antenna Theory and

More information

RF PULSE DESIGN FOR PARALLEL TRANSMISSION IN ULTRA HIGH FIELD MAGNETIC RESONANCE IMAGING. Hai Zheng. B.S., Xi an JiaoTong University, 2005

RF PULSE DESIGN FOR PARALLEL TRANSMISSION IN ULTRA HIGH FIELD MAGNETIC RESONANCE IMAGING. Hai Zheng. B.S., Xi an JiaoTong University, 2005 RF PULSE DESIGN FOR PARALLEL TRANSMISSION IN ULTRA HIGH FIELD MAGNETIC RESONANCE IMAGING by Hai Zheng B.S., Xi an JiaoTong University, 2005 Submitted to the Graduate Faculty of the Swanson School of Engineering

More information

Reduced Field-of-View Excitation Using Second-Order Gradients and Spatial-Spectral Radiofrequency Pulses

Reduced Field-of-View Excitation Using Second-Order Gradients and Spatial-Spectral Radiofrequency Pulses Magnetic Resonance in Medicine 69:503 508 (2013) Reduced Field-of-View Excitation Using Second-Order Gradients and Spatial-Spectral Radiofrequency Pulses Chao Ma, 1,2* Dan Xu, 3 Kevin F. King, 3 and Zhi-Pei

More information

VERTICAL POWER AMPLIFIER POWER AMPLIFIER

VERTICAL POWER AMPLIFIER POWER AMPLIFIER VERTICAL POWER AMPLIFIER POWER AMPLIFIER A vertical power amplifier is very much related to an audio power amplifier. Audio amplifiers are voltage amplifiers (voltage in voltage out). Vertical amplifiers

More information

Principles of MRI EE225E / BIO265. Lecture 21. Instructor: Miki Lustig UC Berkeley, EECS. M. Lustig, EECS UC Berkeley

Principles of MRI EE225E / BIO265. Lecture 21. Instructor: Miki Lustig UC Berkeley, EECS. M. Lustig, EECS UC Berkeley Principles of MRI Lecture 21 EE225E / BIO265 Instructor: Miki Lustig UC Berkeley, EECS Question What is the difference between the images? Answer Both T1-weighted spin-echo gradient-echo Lower SNR Meniscus

More information

Magnetic Resonance in Medicine. Root-flipped multiband radiofrequency pulse design. For Peer Review. Journal: Magnetic Resonance in Medicine

Magnetic Resonance in Medicine. Root-flipped multiband radiofrequency pulse design. For Peer Review. Journal: Magnetic Resonance in Medicine Root-flipped multiband radiofrequency pulse design Journal: Manuscript ID: Draft Wiley - Manuscript type: Full Paper Date Submitted by the Author: n/a Complete List of Authors: Sharma, Anuj; Vanderbilt

More information

MRI: From Signal to Image

MRI: From Signal to Image MRI: From Signal to Image Johannes Koch physics654 2013-05-06 1 / 27 Tomography Magnetic Resonance Tomography Tomography: tomos: section graphein: to write Signal measured as function of space 2 / 27 Tomography

More information

Pulse Compression. Since each part of the pulse has unique frequency, the returns can be completely separated.

Pulse Compression. Since each part of the pulse has unique frequency, the returns can be completely separated. Pulse Compression Pulse compression is a generic term that is used to describe a waveshaping process that is produced as a propagating waveform is modified by the electrical network properties of the transmission

More information

Multiband Excitation Pulses for Hyperpolarized 13 C Dynamic Chemical Shift Imaging

Multiband Excitation Pulses for Hyperpolarized 13 C Dynamic Chemical Shift Imaging Multiband Excitation Pulses for Hyperpolarized 13 C Dynamic Chemical Shift Imaging Peder E. Z. Larson 1, Adam B. Kerr 2, Albert P. Chen 1, Michael S. Lustig 2, Matthew L. Zierhut 1, Simon Hu 1, Charles

More information

SIEMENS MAGNETOM Skyra syngo MR D13

SIEMENS MAGNETOM Skyra syngo MR D13 Page 1 of 12 SIEMENS MAGNETOM Skyra syngo MR D13 \\USER\CIND\StudyProtocols\PTSA\*ep2d_M0Map_p2_TE15 TA:7.9 s PAT:2 Voxel size:2.5 2.5 3.0 mm Rel. SNR:1.00 :epfid Properties Routine Contrast Prio Recon

More information

MRI Metal Artifact Reduction

MRI Metal Artifact Reduction MRI Metal Artifact Reduction PD Dr. med. Reto Sutter University Hospital Balgrist Zurich University of Zurich OUTLINE Is this Patient suitable for MR Imaging? Metal artifact reduction Is this Patient suitable

More information

Pulse Sequence Design and Image Procedures

Pulse Sequence Design and Image Procedures Pulse Sequence Design and Image Procedures 1 Gregory L. Wheeler, BSRT(R)(MR) MRI Consultant 2 A pulse sequence is a timing diagram designed with a series of RF pulses, gradients switching, and signal readout

More information

Image Quality/Artifacts Frequency (MHz)

Image Quality/Artifacts Frequency (MHz) The Larmor Relation 84 Image Quality/Artifacts (MHz) 42 ω = γ X B = 2πf 84 0.0 1.0 2.0 Magnetic Field (Tesla) 1 A 1D Image Magnetic Field Gradients Magnet Field Strength Field Strength / Gradient Coil

More information

Additive Angle Method for Fast Large-Tip-Angle RF Pulse Design in Parallel Excitation

Additive Angle Method for Fast Large-Tip-Angle RF Pulse Design in Parallel Excitation Magnetic Resonance in Medicine 59:779 787 (2008) Additive Angle Method for Fast Large-Tip-Angle RF Pulse Design in Parallel Excitation William A. Grissom, 1 Chun-Yu Yip, 2 Steven M. Wright, 3 Jeffrey A.

More information

A. SPECIFIC AIMS: phase graph (EPG) algorithms to cover a wide range of MRI

A. SPECIFIC AIMS: phase graph (EPG) algorithms to cover a wide range of MRI A. SPECIFIC AIMS: A.. Overview: The promise of improved MRI results at high field strength is compromised by the difficulties encountered at high field, including: i) Non-uniform excitation, due to the

More information

Exercise 1-3. Radar Antennas EXERCISE OBJECTIVE DISCUSSION OUTLINE DISCUSSION OF FUNDAMENTALS. Antenna types

Exercise 1-3. Radar Antennas EXERCISE OBJECTIVE DISCUSSION OUTLINE DISCUSSION OF FUNDAMENTALS. Antenna types Exercise 1-3 Radar Antennas EXERCISE OBJECTIVE When you have completed this exercise, you will be familiar with the role of the antenna in a radar system. You will also be familiar with the intrinsic characteristics

More information

Technician License Course Chapter 2. Lesson Plan Module 3 Modulation and Bandwidth

Technician License Course Chapter 2. Lesson Plan Module 3 Modulation and Bandwidth Technician License Course Chapter 2 Lesson Plan Module 3 Modulation and Bandwidth The Basic Radio Station What Happens During Radio Communication? Transmitting (sending a signal): Information (voice, data,

More information

Pulse Sequence Design Made Easier

Pulse Sequence Design Made Easier Pulse Sequence Design Made Easier Gregory L. Wheeler, BSRT(R)(MR) MRI Consultant gurumri@gmail.com 1 2 Pulse Sequences generally have the following characteristics: An RF line characterizing RF Pulse applications

More information

Lesson 06: Pulse-echo Imaging and Display Modes. These lessons contain 26 slides plus 15 multiple-choice questions.

Lesson 06: Pulse-echo Imaging and Display Modes. These lessons contain 26 slides plus 15 multiple-choice questions. Lesson 06: Pulse-echo Imaging and Display Modes These lessons contain 26 slides plus 15 multiple-choice questions. These lesson were derived from pages 26 through 32 in the textbook: ULTRASOUND IMAGING

More information

Simultaneous Multi-Slice (Slice Accelerated) Diffusion EPI

Simultaneous Multi-Slice (Slice Accelerated) Diffusion EPI Simultaneous Multi-Slice (Slice Accelerated) Diffusion EPI Val M. Runge, MD Institute for Diagnostic and Interventional Radiology Clinics for Neuroradiology and Nuclear Medicine University Hospital Zurich

More information

March, 2003 IEEE P /131r0. IEEE P Wireless Personal Area Networks

March, 2003 IEEE P /131r0. IEEE P Wireless Personal Area Networks Project Title IEEE P802.15 Wireless Personal rea Networks IEEE P802.15 Working Group for Wireless Personal rea Networks (WPNs) PHY Proposal Using Dual Independent Single Sideband, Non-coherent M and Defined

More information

SIGNA Pioneer: Ultra High Efficiency Gradient System Advancing the gradient technology curve

SIGNA Pioneer: Ultra High Efficiency Gradient System Advancing the gradient technology curve GE Healthcare SIGNA Pioneer: Ultra High Efficiency Gradient System Advancing the gradient technology curve NEW TECHNOLOGY 40W Watts spec is irrelevant. 4W LED bulb delivers same brightness as 40W incandescent

More information

Fundamentals of Digital Communication

Fundamentals of Digital Communication Fundamentals of Digital Communication Network Infrastructures A.A. 2017/18 Digital communication system Analog Digital Input Signal Analog/ Digital Low Pass Filter Sampler Quantizer Source Encoder Channel

More information

Advanced MSK MRI Protocols at 3.0T. Garry E. Gold, M.D. Associate Professor Department of Radiology Stanford University

Advanced MSK MRI Protocols at 3.0T. Garry E. Gold, M.D. Associate Professor Department of Radiology Stanford University Advanced MSK MRI Protocols at 3.0T Garry E. Gold, M.D. Associate Professor Department of Radiology Stanford University Outline Why High Field for MSK? SNR and Relaxation Times Technical Issues Example

More information

Compensation in 3T Cardiac Imaging Using Short 2DRF Pulses

Compensation in 3T Cardiac Imaging Using Short 2DRF Pulses Magnetic Resonance in Medicine 59:441 446 (2008) B + 1 Compensation in 3T Cardiac Imaging Using Short 2DRF Pulses Kyunghyun Sung and Krishna S. Nayak The purpose of this study was to determine if tailored

More information

MR Basics: Module 6 Pulse Sequences

MR Basics: Module 6 Pulse Sequences Module 6 Transcript For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions.

More information

WaveSurfer. Basic acoustics part 2 Spectrograms, resonance, vowels. Spectrogram. See Rogers chapter 7 8

WaveSurfer. Basic acoustics part 2 Spectrograms, resonance, vowels. Spectrogram. See Rogers chapter 7 8 WaveSurfer. Basic acoustics part 2 Spectrograms, resonance, vowels See Rogers chapter 7 8 Allows us to see Waveform Spectrogram (color or gray) Spectral section short-time spectrum = spectrum of a brief

More information

a. Use (at least) window lengths of 256, 1024, and 4096 samples to compute the average spectrum using a window overlap of 0.5.

a. Use (at least) window lengths of 256, 1024, and 4096 samples to compute the average spectrum using a window overlap of 0.5. 1. Download the file signal.mat from the website. This is continuous 10 second recording of a signal sampled at 1 khz. Assume the noise is ergodic in time and that it is white. I used the MATLAB Signal

More information

FCC and ETSI Requirements for Short-Range UHF ASK- Modulated Transmitters

FCC and ETSI Requirements for Short-Range UHF ASK- Modulated Transmitters From December 2005 High Frequency Electronics Copyright 2005 Summit Technical Media FCC and ETSI Requirements for Short-Range UHF ASK- Modulated Transmitters By Larry Burgess Maxim Integrated Products

More information

Supplementary Material

Supplementary Material Supplementary Material Orthogonal representation of sound dimensions in the primate midbrain Simon Baumann, Timothy D. Griffiths, Li Sun, Christopher I. Petkov, Alex Thiele & Adrian Rees Methods: Animals

More information

Human Subjects in fmri Research

Human Subjects in fmri Research HST.583: Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Harvard-MIT Division of Health Sciences and Technology Course Instructor: Dr. Randy Gollub. Human Subjects in fmri Research

More information

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893 Digital Radiography D. J. Hall, Ph.D. x20893 djhall@ucsd.edu Background Common Digital Modalities Digital Chest Radiograph - 4096 x 4096 x 12 bit CT - 512 x 512 x 12 bit SPECT - 128 x 128 x 8 bit MRI -

More information

A Spatial Domain Method for the Design of RF Pulses in Multi-Coil Parallel Excitation

A Spatial Domain Method for the Design of RF Pulses in Multi-Coil Parallel Excitation A Spatial Domain Method for the Design of RF Pulses in Multi-Coil Parallel Excitation Revised Draft, March 8, 2006 William Grissom 1, Chun-yu Yip 2, Zhenghui Zhang 3, V. Andrew Stenger 4, Jeffrey A. Fessler

More information

Magnetic Resonance Spectroscopic Imaging using Parallel Transmission at 7T

Magnetic Resonance Spectroscopic Imaging using Parallel Transmission at 7T Magnetic Resonance Spectroscopic Imaging using Parallel Transmission at 7T by Borjan Aleksandar Gagoski Submitted to the Department of Electrical Engineering and Computer Science in partial fulfillment

More information

Electronically Steerable planer Phased Array Antenna

Electronically Steerable planer Phased Array Antenna Electronically Steerable planer Phased Array Antenna Amandeep Kaur Department of Electronics and Communication Technology, Guru Nanak Dev University, Amritsar, India Abstract- A planar phased-array antenna

More information

ELT Receiver Architectures and Signal Processing Fall Mandatory homework exercises

ELT Receiver Architectures and Signal Processing Fall Mandatory homework exercises ELT-44006 Receiver Architectures and Signal Processing Fall 2014 1 Mandatory homework exercises - Individual solutions to be returned to Markku Renfors by email or in paper format. - Solutions are expected

More information

Photonic Signal Processing(PSP) of Microwave Signals

Photonic Signal Processing(PSP) of Microwave Signals Photonic Signal Processing(PSP) of Microwave Signals 2015.05.08 김창훈 R. A. Minasian, Photonic signal processing of microwave signals, IEEE Trans. Microw. Theory Tech., vol. 54, no. 2, pp. 832 846, Feb.

More information

ELT COMMUNICATION THEORY

ELT COMMUNICATION THEORY ELT 41307 COMMUNICATION THEORY Project work, Fall 2017 Experimenting an elementary single carrier M QAM based digital communication chain 1 ASSUMED SYSTEM MODEL AND PARAMETERS 1.1 SYSTEM MODEL In this

More information

Chapter 11 Coherence Editing: Pulse-field Gradients and Phase Cycling

Chapter 11 Coherence Editing: Pulse-field Gradients and Phase Cycling Chapter 11 Coherence Editing: Pulse-field Gradients and Phase Cycling Coherence editing is used to remove unwanted signals from NMR spectra. For example, in the double quantum filtered COSY experiment,

More information

ISMRM weekend educational course, MR Systems Engineering, Console Electronics

ISMRM weekend educational course, MR Systems Engineering, Console Electronics ISMRM weekend educational course, MR Systems Engineering, Console Electronics. 2013-4-20 Declaration of Relevant Financial Interests or Relationships Speaker Name: Katsumi Kose, Ph.D. I have the following

More information

Variable-Rate Selective Excitation for Rapid MRI Sequences

Variable-Rate Selective Excitation for Rapid MRI Sequences Variable-Rate Selective Excitation for Rapid MRI Sequences Brian A. Hargreaves,* Charles H. Cunningham, Dwight G. Nishimura, and Steven M. Conolly Magnetic Resonance in Medicine 52:590 597 (2004) Balanced

More information

Implementation of parallel search algorithms using spatial encoding by nuclear magnetic resonance

Implementation of parallel search algorithms using spatial encoding by nuclear magnetic resonance Implementation of parallel search algorithms using spatial encoding by nuclear magnetic resonance Rangeet Bhattacharyya, 1 Ranabir Das, 1 K. V. Ramanathan, 2 and Anil Kumar 1,2, * 1 Department of Physics,

More information

Implementing Orthogonal Binary Overlay on a Pulse Train using Frequency Modulation

Implementing Orthogonal Binary Overlay on a Pulse Train using Frequency Modulation Implementing Orthogonal Binary Overlay on a Pulse Train using Frequency Modulation As reported recently, overlaying orthogonal phase coding on any coherent train of identical radar pulses, removes most

More information

Understanding The HA2500's Dynamic METER Tests

Understanding The HA2500's Dynamic METER Tests Understanding The HA2500's Dynamic METER Tests The HA2500's METER Dynamic Tests are designed to analyze parameters of the horizontal output stage with the chassis powered on. Horizontal output stage measurements

More information

Application Guide & Release Notes

Application Guide & Release Notes Application Guide & Release Notes Inner-volume-imaging (IVI) EPI C2P Release 002a 1 September 2015 TMII Translational and Molecular Imaging Institute Conditions of Use This package is provided to support

More information

MR Advance Techniques. Flow Phenomena. Class II

MR Advance Techniques. Flow Phenomena. Class II MR Advance Techniques Flow Phenomena Class II Flow Phenomena In this class we will explore different phenomenona produced from nuclei that move during the acquisition of data. Flowing nuclei exhibit different

More information

The Old Cat and Mouse Game Continues

The Old Cat and Mouse Game Continues The Old Cat and Mouse Game Continues or, How Advances in Radar Development Drive Testing Requirements for Next Generation EW Systems by: Walt Schulte Agilent Technologies Microwave and Communications Division

More information

UNDERSTANDING HORIZONTAL OUTPUT STAGES OF COMPUTER MONITORS

UNDERSTANDING HORIZONTAL OUTPUT STAGES OF COMPUTER MONITORS UNDERSTANDING HORIZONTAL OUTPUT STAGES OF COMPUTER MONITORS Today's computer, medical, security, design and industrial video display monitors operate at a host of different horizontal resolutions or scanning

More information

FM Superheterodyne Receiver

FM Superheterodyne Receiver EE321 Final Project Chun-Hao Lo XiaoKai Sun Background: FM Superheterodyne Receiver Superheterodyne Receiver is the receiver that convert a received signal from the transmitter to an intermediate frequency.

More information

Acoustic noise reduction of MRI systems by means of magnetic shielding

Acoustic noise reduction of MRI systems by means of magnetic shielding Acoustic noise reduction of MRI systems by means of magnetic shielding D. Biloen, N.B. Roozen Philips Applied Technologies, P.O.Box 218/Bldg. SAQ 2121, 56MD Eindhoven, The Netherlands {david.biloen, n.b.roozen}@philips.com,

More information

Technician License Course Chapter 2. Lesson Plan Module 2 Radio Signals and Waves

Technician License Course Chapter 2. Lesson Plan Module 2 Radio Signals and Waves Technician License Course Chapter 2 Lesson Plan Module 2 Radio Signals and Waves The Basic Radio Station What Happens During Radio Communication? Transmitting (sending a signal): Information (voice, data,

More information

Newsletter 5.4. New Antennas. The profiled horns. Antenna Magus Version 5.4 released! May 2015

Newsletter 5.4. New Antennas. The profiled horns. Antenna Magus Version 5.4 released! May 2015 Newsletter 5.4 May 215 Antenna Magus Version 5.4 released! Version 5.4 sees the release of eleven new antennas (taking the total number of antennas to 277) as well as a number of new features, improvements

More information

MR Basics: Module 8 Image Quality

MR Basics: Module 8 Image Quality Module 8 Transcript For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions.

More information

Objectives. Presentation Outline. Digital Modulation Revision

Objectives. Presentation Outline. Digital Modulation Revision Digital Modulation Revision Professor Richard Harris Objectives To identify the key points from the lecture material presented in the Digital Modulation section of this paper. What is in the examination

More information

NIH Public Access Author Manuscript J Magn Reson. Author manuscript; available in PMC 2012 May 1.

NIH Public Access Author Manuscript J Magn Reson. Author manuscript; available in PMC 2012 May 1. NIH Public Access Author Manuscript Published in final edited form as: J Magn Reson. 2011 May ; 210(1): 38 43. doi:10.1016/j.jmr.2011.02.004. Fast fat-suppressed reduced field-of-view temperature mapping

More information

ANTENNA INTRODUCTION / BASICS

ANTENNA INTRODUCTION / BASICS ANTENNA INTRODUCTION / BASICS RULES OF THUMB: 1. The Gain of an antenna with losses is given by: 2. Gain of rectangular X-Band Aperture G = 1.4 LW L = length of aperture in cm Where: W = width of aperture

More information

EE 422G - Signals and Systems Laboratory

EE 422G - Signals and Systems Laboratory EE 422G - Signals and Systems Laboratory Lab 3 FIR Filters Written by Kevin D. Donohue Department of Electrical and Computer Engineering University of Kentucky Lexington, KY 40506 September 19, 2015 Objectives:

More information

ADAPTIVE ANTENNAS. TYPES OF BEAMFORMING

ADAPTIVE ANTENNAS. TYPES OF BEAMFORMING ADAPTIVE ANTENNAS TYPES OF BEAMFORMING 1 1- Outlines This chapter will introduce : Essential terminologies for beamforming; BF Demonstrating the function of the complex weights and how the phase and amplitude

More information

MAGNETIC RESONANCE IMAGING

MAGNETIC RESONANCE IMAGING CSEE 4620 Homework 3 Fall 2018 MAGNETIC RESONANCE IMAGING 1. THE PRIMARY MAGNET Magnetic resonance imaging requires a very strong static magnetic field to align the nuclei. Modern MRI scanners require

More information

Sampling the World in 3D by Airborne LIDAR Assessing the Information Content of LIDAR Point Clouds

Sampling the World in 3D by Airborne LIDAR Assessing the Information Content of LIDAR Point Clouds Sampling the World in 3D by Airborne LIDAR Assessing the Information Content of LIDAR Point Clouds PhoWo 2013 September 11 th, 2013 Stuttgart, Germany Andreas Ullrich RIEGL LMS GmbH sequential data acquisition

More information

Analysis of magnetic and electromagnetic field emissions produced by a MRI device

Analysis of magnetic and electromagnetic field emissions produced by a MRI device Sept. 8-1, 21, Kosice, Slovakia Analysis of magnetic and electromagnetic field emissions produced by a MRI device D. Giordano, M. Borsero, G. Crotti, M. ucca INRIM Istituto Nazionale di Ricerca Metrologica,

More information

Ultrasound Bioinstrumentation. Topic 2 (lecture 3) Beamforming

Ultrasound Bioinstrumentation. Topic 2 (lecture 3) Beamforming Ultrasound Bioinstrumentation Topic 2 (lecture 3) Beamforming Angular Spectrum 2D Fourier transform of aperture Angular spectrum Propagation of Angular Spectrum Propagation as a Linear Spatial Filter Free

More information