Optimal Pupil Design for Confocal Microscopy

Size: px
Start display at page:

Download "Optimal Pupil Design for Confocal Microscopy"

Transcription

1 Optimal Pupil Design for Confocal Microscopy Yogesh G. Patel 1, Milind Rajadhyaksha 3, and Charles A. DiMarzio 1,2 1 Department of Electrical and Computer Engineering, 2 Department of Mechanical and Industrial Engineering, Northeastern University, 440 Dana Building, 360 Huntington Avenue, Boston, MA Department of Dermatology, Memorial Sloan-Kettering Cancer Center, 160 East 53 rd Street, New York, NY ABSTRACT Confocal reflectance microscopy may enable screening and diagnosis of skin cancers noninvasively and in real-time, as an adjunct to biopsy and pathology. Current instruments are large, complex, and expensive. A simpler, confocal line-scanning microscope may accelerate the translation of confocal microscopy in clinical and surgical dermatology. A confocal reflectance microscope may use a beamsplitter, transmitting and detecting through the pupil, or a divided pupil, or theta configuration, with half used for transmission and half for detection. The divided pupil may offer better sectioning and contrast. We present a Fourier optics model and compare the on-axis irradiance of a confocal point-scanning microscope in both pupil configurations, optimizing the profile of a Gaussian beam in a circular or semicircular aperture. We repeat both calculations with a cylindrical lens which focuses the source to a line. The variable parameter is the fillfactor, h, the ratio of the 1/e 2 diameter of the Gaussian beam to the diameter of the full aperture. The optimal values of h, for point scanning are 0.90 (full) and 0.66 for the half-aperture. For line-scanning, the fill-factors are 1.02 (full) and 0.52 (half). Additional parameters to consider are the optimal location of the point-source beam in the divided-pupil configuration, the optimal line width for the line-source, and the width of the aperture in the divided-pupil configuration. Additional figures of merit are field-of-view and sectioning. Use of optimal designs is critical in comparing the experimental performance of the different configurations. Keywords: confocal microscopy, Fourier optics, point-scanning, line-scanning, full-pupil, divided-pupil 1. INTRODUCTION In the United States, over one million new cases of skin cancer are diagnosed per year, but many more are suspected. Because biopsy is the only currently accepted method of confirming a diagnosis, over 5 million biopsies are performed to detect these one million cancers [1-3]. These biopsies cost the U.S. healthcare system over $1.5 billion [1-3]. Reducing the need for biopsy in the 80% of cases which are found to be benign could eliminate most of this cost. Confocal reflectance microscopy may enable early noninvasive screening and comparable diagnosis of skin cancers in real time with and minimize the need for biopsies. The confocal reflectance point-scanning microscope [Figure 1] was developed by Marvin Minsky to create an image of a plane within a turbid specimen while blocking out-of-focus light [4-5]. In Minsky's invention the sample was moved to accomplish scanning. This system achieved the ability to image sections throughout a specimen noninvasively. A confocal microscope uses a point source of light to illuminate a point within a specimen. The scattered light from this focused point is then collected onto a detector through a point aperture. The collected light forms an image from that point with sharp axial and lateral resolution. The size of the detector and illuminated point Three-Dimensional and Multidimensional Microscopy: Image Acquisition and Processing XVII, edited by Jose-Angel Conchello, Carol J. Cogswell, Tony Wilson, Thomas G. Brown, Proc. of SPIE Vol. 7570, 75700P 2010 SPIE CCC code: /10/$18 doi: / Proc. of SPIE Vol P-1

2 are matched such that light not from the point of focus is blocked and the detector receives light only from the plane of focus. Figure 1: Confocal reflectance point-scanning microscope optical system layout. The confocal reflectance line-scanning system [Figure 2] is similar to a point-scanning system expect instead of a focused point within a specimen, a focused line is illuminated onto a specimen. Instead of a point illumination and detection through a pinhole, the plane being imaged is scanned with a focused line and detected through a confocal slit (i.e. a line slit). The line source is formed by a cylindrical lens between the laser source and objective lens. The line is formed in one direction and an image is formed in real-time using a scanner. Figure 2: Confocal reflectance line-scanning microscope optical system layout. There are two types of confocal reflectance line-scanning, the divided-pupil objective lens aperture, pioneered by Charles Koester [6-7], and the full-pupil confocal reflectance line-scanning system introduced by GJ Brakenhoff and K. Visscher [8-10]. Proc. of SPIE Vol P-2

3 We present a Fourier-analysis computational model and compare the on-axis irradiance of a confocal point-scanning microscope in both full-pupil and divided-pupil configurations, optimizing the profile of a Gaussian beam. We repeat both calculations with for a confocal line-scanning microscope, which focuses the source to a line, in fullpupil and divided-pupil configurations. 2. FOURIER-ANALYSIS COMPUTATIONAL MODEL The amplitudes and phases of all the light rays reaching the image plane can be characterized using Fourier methods. Fourier analysis allows for treating optical processes in terms of spatial frequencies [11]. Therefore this type of analysis can be used for modeling the aperture and objective lens in a confocal scanning system. A pictorial representation of Fourier analysis in a lens system is shown in Figure 3. The field in the pupil plane (Σ p ) is the Fourier transform of the field in the front focal plane of the lens, which is the object (Σ o ) or image plane. Figure 3: Illustration of Fourier analysis in a lens system. The pupil plane (Σ p ) points become the Fourier transform at the back focal plane of a lens, which is the object (Σ o ) or image plane. Fourier analysis schematic shows the pupil plane (Σ p ) where a uniform plane wave is diffracted at the front focal point of the lens and then converges to form the far-field diffraction pattern at the back focal plane, or object/image plane (Σ o ) of the lens (depending on the direction of light and how the imaging system is being used). The object plane, Σ o, also called the transform plane, is where the far-field diffraction electric-field pattern (distribution) is displayed. Since optical systems generally involve two-dimensional components and light, the Fourier transform pair can be generalized in two dimensions as Equation (1) and its FT as Equation (2), where the quantities k x and k y are the angular spatial frequencies along the two corresponding axes. The object scatters waves, which are collected by the lens, and the parallel bundles of rays are brought to convergence at the back focal plane [12]. The pupil plane mask is known as the aperture function: (1) and the object plane distribution is the Fourier transform of the pupil field distribution: Proc. of SPIE Vol P-3

4 (2) 2.1 SYSTEM CONFIGURATION The Fourier analysis computational model evaluated two scanning modes: (1) a point-scanning confocal reflectance microscope and (2) a line-scanning confocal reflectance microscope. In addition, the model evaluated two pupil configurations: (1) full-pupil configuration [Figure 4] and (2) divided-pupil configuration [Figure 5]. Figure 4: Full-Pupil Configuration Figure 5: Divided-Pupil Configuration The full-pupil configuration allows for illumination and detection with the entire pupil of the objective lens. This configuration leads to good resolution; however, there is greater acceptance of multiply scattered light [13-15]. The divided-pupil configuration illuminates half of the objective lens and collects from the other half. This configuration looses optical sectioning due to using only ½ the NA of the objective lens; however, it regains performance with ability to filter out multiply scattered light [13-15]. 2.2 MODELING PARAMETERS In addition to the scanning modes and pupil configurations, there are several parameters that are accounted for in the model. The first of these parameters is the wavelength, λ, which we stipulated to be 632nm and 830nm, corresponding to the 632nm HeNe laser (Melles Griot), and 830nm laser diode (Micro Laser Systems, Inc.) of our system. The lens diameter for the full-pupil or divided-pupil is d and the Gaussian beam at 1/e 2 is h. d. The variable parameter, the fill-factor, h, is the ratio of the 1/e 2 diameter of the Gaussian beam to the diameter (d) of the full-pupil or divided-pupil, depending on the configuration modeled. In the divided-pupil configuration, aw, is the width of the divider, where aw = width/d. The center position of the light source, x c, is given as x c = position/d. Proc. of SPIE Vol P-4

5 We present a Fourier-analysis optics model and compare the on-axis irradiance of a confocal point-scanning or linescanning microscope in both pupil configurations, optimizing the profile of a Gaussian beam in a full-pupil or divided-pupil. 3. RESULTS We will show that the optimal value of fill-factor, h, is 0.90 for full-pupil point-scanning and 0.66 for the dividedpupil. For line-scanning, the fill-factors are 1.02 for full-pupil and 0.52 for the divided-pupil. 3.1 FULL-PUPIL POINT-SCANNING CONFIGURATION We compare the computed point-scan on-axis image irradiance values to a calculated Gaussian source and a uniform source at varying h values, [Figure 7]. For fill factor h=0.3, [(O), Figure 7], the pupil irradiance and image irradiance is shown in Figure 8. The Gaussian approximation is good for small values of h as Figure 7 shows good correlation of on-axis image irradiance for the Gaussian source (. ) and the computed point-scan on-axis image irradiance curve ( ) for fill factor values, h < 1. As h becomes larger, (h>1), the Gaussian approximation does not hold because there is diffraction due to truncation of the Gaussian source. The uniform source, ( === ), shows good correlation of on-axis image irradiance to the computed point-scan on-axis image irradiance curve ( ) for fill factor values, h > 2 [Figure 7]. For fill factor h=3.0, [(Δ), Figure 7], the pupil irradiance and image irradiance is shown in Figure 9. For large values of h, the uniform source correlates well to the point-scan image irradiance, however, there is a great loss of power in the image plane. Figure 7: Graph shows the on-axis image irradiance for computed line-scan, Gaussian, and uniform sources. Proc. of SPIE Vol P-5

6 Figure 8: Pupil & Irradiance for h = 0.3 Image Figure 9: Pupil & Irradiance for h = 3.0 Image The optimum fill factor, h, value seems to be near the intersection of the computed Gaussian source and Uniform source. We show the pupil plane irradiance and image plane irradiance for the optimum fill factor, h = 0.9, [( ), Figure 7], and the horizontal slice through the respective plane showing the beam profile and peak irradiance, [Figure 10]. Figure 10: Pupil & Image Irradiance for h = 0.9, and the horizontal slice through the respective plane showing the beam profile and peak irradiance. Proc. of SPIE Vol P-6

7 3.2 FULL-PUPIL LINE-SCANNING CONFIGURATION In the full-pupil line-scanning configuration, the optimum fill factor is h=1.02. The pupil is slightly overfilled to produce the highest on-axis irradiance in the image plane. Figure 11 shows the pupil & image plane irradiance and a horizontal slice through the respective plane. The plot of the maximum on-axis image irradiance of the computed line-scan ( ) for varying fill factor values is shown in Figure 12, with Gaussian source (. ) and Uniform sources ( === ) as comparison. Figure 11: Pupil & Image Plane Irradiance, with corresponding horizontal slice through plane for fullpupil line-scanning configuration. Figure 12: Graph shows the on-axis image irradiance for computed line-scan, Gaussian, and uniform sources. 3.3 DIVIDED-PUPIL POINT-SCANNING CONFIGURATION The Fourier analysis computational model was repeated for the divided-pupil configuration for a point-scanning and line-scanning system. Figure 13 shows the pupil and image plane irradiance and the horizontal slice through the respective plane for h=0.66. The plot of the on-axis image irradiance signal for the divided-pupil point-scan is shown in Figure 14. Proc. of SPIE Vol P-7

8 Figure 13: Pupil & Image Plane Irradiance, with corresponding horizontal slice through plane for fullpupil line-scanning configuration. Figure 14: Graph shows the on-axis image irradiance for computed line-scan, Gaussian, and uniform sources. 3.4 DIVIDED-PUPIL LINE-SCANNING CONFIGURATION The Fourier analysis computational model evaluation for the divided-pupil configuration line-scanning system determined the optimum fill factor value to be h=0.52. Figure 15 shows the pupil and image plane irradiance and the horizontal slice through the respective plane for h=0.52. The plot of the on-axis image irradiance signal for the divided-pupil point-scan is shown in Figure 16. Figure 15: Pupil & Image Plane Irradiance, with corresponding horizontal slice through plane for fullpupil line-scanning configuration. Figure 16: Graph shows the on-axis image irradiance for computed line-scan, Gaussian, and uniform sources. 4. CONCLUSIONS We have presented a Fourier-analysis computational model in optimal pupil design for confocal microscopy. We presented optimum parameters for four confocal microscopy configurations: (1) full-pupil point-scanning, (2) fullpupil line-scanning, (3) divided-pupil point-scanning, and (4) divided-pupil line-scanning. We have shown the Proc. of SPIE Vol P-8

9 optimal value of fill-factor, h, is 0.90 for full-pupil point-scanning and 0.66 for the divided-pupil. For line-scanning, the fill-factors are 1.02 for full-pupil and 0.52 for the divided-pupil. In future work, additional parameters we will consider are the optimal location of the point-source beam in the divided-pupil configuration, the optimal line width for the line-source, and the width of the aperture in the dividedpupil configuration. Use of optimal designs is critical in comparing the experimental performance of the different configurations. ACKNOWLEDGEMENTS This work was supported in part by Gordon-CenSSIS, the Center for Subsurface Sensing and Imaging Systems, under the Engineering Research Centers Program of the National Science Foundation (Award Number EEC ). Special thanks to Dr. John Haller & Dr. Yantian Zhang, for the support of the National Institute of Biomedical Imaging and Bioengineering (NIBIB) at the National Institute of Health (NIH). The primary author received funding through the Integrative Graduate Education and Research Traineeship (IGERT) in the area of Nanomedicine at Northeastern University. REFERENCES [1] National Cancer Institute, American Cancer Society, 2008 statistics [2] [3] [4] Confocal scanning microscope, M Minsky Rapport technique, Patent, (1955). [5] Memoir on inventing the confocal scanning microscope, M Minsky, SPIE Milestone Series MS, VOL 131, pages 7-17, (1996). [6] CJ Koester, Scanning mirror microscope with optical sectioning characteristics: applications to ophthalmology, Appl.Opt. 19, (1980). [7] CJ Koester, High efficiency optical sectioning with confocal slits, Trans. R. Microsc. Soc. 90, (1990). [8] GJ Brakenhoff and K Visscher, Confocal imaging with bilateral scanning and array detectors, J. Micros. 165, (1992). [9] GJ Brakenhoff and K Visscher, "Novel confocal imaging and visualization techniques," in Micro 90: Proceedings of the Royal Microscopical Society Conference, H. Y. Elder, ed. Chapter 9, (Hilger, London, 1990). [10] Visscher, K., G. J. Brakenhoff, and J. J. Krol. Micromanipulation by "multiple" optical traps created by a single fast scanning trap integrated with the bilateral confocal scanning laser microscope. Cytometry. 14: , [11] M. Born, E. Wolf. Principles of optics. 7 th Edition. Cambridge University Press, Cambridge [12] E. Hecht. Optics. 4 th Ed. Addison-Wesley, New York [13] PJ Dwyer, CA DiMarzio, JM Zavislan, WJ Fox, and M Rajadhyaksha, "Confocal reflectance theta line scanning microscope for imaging human skin in vivo," Optics Letters 31, (2006). [14] PJ Dwyer, CA DiMarzio, and M Rajadhyaksha, "Confocal theta line-scanning microscope for imaging human tissues," Applied Optics 46, (2007). [15] DS Gareau, S Abeytunge, and M Rajadhyaksha, "Line-scanning reflectance confocal microscopy of human skin: comparison of full-pupil and divided-pupil configurations," Optics Letters 34, (2009). *ypatel@ece.neu.edu, Phone: (617) , Fax: (617) Proc. of SPIE Vol P-9

Development of a new multi-wavelength confocal surface profilometer for in-situ automatic optical inspection (AOI)

Development of a new multi-wavelength confocal surface profilometer for in-situ automatic optical inspection (AOI) Development of a new multi-wavelength confocal surface profilometer for in-situ automatic optical inspection (AOI) Liang-Chia Chen 1#, Chao-Nan Chen 1 and Yi-Wei Chang 1 1. Institute of Automation Technology,

More information

Optimization of pupil design for point-scanning and line-scanning confocal microscopy

Optimization of pupil design for point-scanning and line-scanning confocal microscopy Optimization o pupil design or point-scanning and line-scanning conocal microscopy Yogesh G. Patel, 1,* Milind Rajadhyaksha, 2 and Charles A. DiMarzio 1,3 1 Electrical & Computer Engineering Department,

More information

Reflecting optical system to increase signal intensity. in confocal microscopy

Reflecting optical system to increase signal intensity. in confocal microscopy Reflecting optical system to increase signal intensity in confocal microscopy DongKyun Kang *, JungWoo Seo, DaeGab Gweon Nano Opto Mechatronics Laboratory, Dept. of Mechanical Engineering, Korea Advanced

More information

Confocal Imaging Through Scattering Media with a Volume Holographic Filter

Confocal Imaging Through Scattering Media with a Volume Holographic Filter Confocal Imaging Through Scattering Media with a Volume Holographic Filter Michal Balberg +, George Barbastathis*, Sergio Fantini % and David J. Brady University of Illinois at Urbana-Champaign, Urbana,

More information

Study of Graded Index and Truncated Apertures Using Speckle Images

Study of Graded Index and Truncated Apertures Using Speckle Images Study of Graded Index and Truncated Apertures Using Speckle Images A. M. Hamed Department of Physics, Faculty of Science, Ain Shams University, Cairo, 11566 Egypt amhamed73@hotmail.com Abstract- In this

More information

Beam Profiling. Introduction. What is Beam Profiling? by Michael Scaggs. Haas Laser Technologies, Inc.

Beam Profiling. Introduction. What is Beam Profiling? by Michael Scaggs. Haas Laser Technologies, Inc. Beam Profiling by Michael Scaggs Haas Laser Technologies, Inc. Introduction Lasers are ubiquitous in industry today. Carbon Dioxide, Nd:YAG, Excimer and Fiber lasers are used in many industries and a myriad

More information

VISUAL PHYSICS ONLINE DEPTH STUDY: ELECTRON MICROSCOPES

VISUAL PHYSICS ONLINE DEPTH STUDY: ELECTRON MICROSCOPES VISUAL PHYSICS ONLINE DEPTH STUDY: ELECTRON MICROSCOPES Shortly after the experimental confirmation of the wave properties of the electron, it was suggested that the electron could be used to examine objects

More information

III. United States Patent (19) Zavislan et al. 35 Claims, 2 Drawing Sheets

III. United States Patent (19) Zavislan et al. 35 Claims, 2 Drawing Sheets United States Patent (19) Zavislan et al. 54 75 73 21 22 60 (51) (52) 58 56) CONFOCAL MAGING THROUGH THCK DERMAL TSSUE Inventors: James M. Zavislan; Jay M. Eastman, both of Pittsford, N.Y. Assignee: Lucid

More information

Shaping light in microscopy:

Shaping light in microscopy: Shaping light in microscopy: Adaptive optical methods and nonconventional beam shapes for enhanced imaging Martí Duocastella planet detector detector sample sample Aberrated wavefront Beamsplitter Adaptive

More information

Physics 431 Final Exam Examples (3:00-5:00 pm 12/16/2009) TIME ALLOTTED: 120 MINUTES Name: Signature:

Physics 431 Final Exam Examples (3:00-5:00 pm 12/16/2009) TIME ALLOTTED: 120 MINUTES Name: Signature: Physics 431 Final Exam Examples (3:00-5:00 pm 12/16/2009) TIME ALLOTTED: 120 MINUTES Name: PID: Signature: CLOSED BOOK. TWO 8 1/2 X 11 SHEET OF NOTES (double sided is allowed), AND SCIENTIFIC POCKET CALCULATOR

More information

Point Spread Function. Confocal Laser Scanning Microscopy. Confocal Aperture. Optical aberrations. Alternative Scanning Microscopy

Point Spread Function. Confocal Laser Scanning Microscopy. Confocal Aperture. Optical aberrations. Alternative Scanning Microscopy Bi177 Lecture 5 Adding the Third Dimension Wide-field Imaging Point Spread Function Deconvolution Confocal Laser Scanning Microscopy Confocal Aperture Optical aberrations Alternative Scanning Microscopy

More information

Laser Surface Profiler

Laser Surface Profiler 'e. * 3 DRAFT 11-02-98 Laser Surface Profiler An-Shyang Chu and M. A. Butler Microsensor R & D Department Sandia National Laboratories Albuquerque, New Mexico 87185-1425 Abstract By accurately measuring

More information

Sensitive measurement of partial coherence using a pinhole array

Sensitive measurement of partial coherence using a pinhole array 1.3 Sensitive measurement of partial coherence using a pinhole array Paul Petruck 1, Rainer Riesenberg 1, Richard Kowarschik 2 1 Institute of Photonic Technology, Albert-Einstein-Strasse 9, 07747 Jena,

More information

Be aware that there is no universal notation for the various quantities.

Be aware that there is no universal notation for the various quantities. Fourier Optics v2.4 Ray tracing is limited in its ability to describe optics because it ignores the wave properties of light. Diffraction is needed to explain image spatial resolution and contrast and

More information

Some of the important topics needed to be addressed in a successful lens design project (R.R. Shannon: The Art and Science of Optical Design)

Some of the important topics needed to be addressed in a successful lens design project (R.R. Shannon: The Art and Science of Optical Design) Lens design Some of the important topics needed to be addressed in a successful lens design project (R.R. Shannon: The Art and Science of Optical Design) Focal length (f) Field angle or field size F/number

More information

Physics 3340 Spring Fourier Optics

Physics 3340 Spring Fourier Optics Physics 3340 Spring 011 Purpose Fourier Optics In this experiment we will show how the Fraunhofer diffraction pattern or spatial Fourier transform of an object can be observed within an optical system.

More information

FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION

FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION Revised November 15, 2017 INTRODUCTION The simplest and most commonly described examples of diffraction and interference from two-dimensional apertures

More information

Confocal theta line-scanning microscope for imaging human tissues

Confocal theta line-scanning microscope for imaging human tissues Confocal theta line-scanning microscope for imaging human tissues Peter J. Dwyer, Charles A. DiMarzio, and Milind Rajadhyaksha A confocal reflectance theta line-scanning microscope demonstrates imaging

More information

EE119 Introduction to Optical Engineering Spring 2003 Final Exam. Name:

EE119 Introduction to Optical Engineering Spring 2003 Final Exam. Name: EE119 Introduction to Optical Engineering Spring 2003 Final Exam Name: SID: CLOSED BOOK. THREE 8 1/2 X 11 SHEETS OF NOTES, AND SCIENTIFIC POCKET CALCULATOR PERMITTED. TIME ALLOTTED: 180 MINUTES Fundamental

More information

Circular Dichroism Microscopy Free from Commingling Linear Dichroism via Discretely Modulated Circular Polarization

Circular Dichroism Microscopy Free from Commingling Linear Dichroism via Discretely Modulated Circular Polarization Supplementary information Circular Dichroism Microscopy Free from Commingling Linear Dichroism via Discretely Modulated Circular Polarization Tetsuya Narushima AB and Hiromi Okamoto A* A Institute for

More information

PHY 431 Homework Set #5 Due Nov. 20 at the start of class

PHY 431 Homework Set #5 Due Nov. 20 at the start of class PHY 431 Homework Set #5 Due Nov. 0 at the start of class 1) Newton s rings (10%) The radius of curvature of the convex surface of a plano-convex lens is 30 cm. The lens is placed with its convex side down

More information

Adaptive optics two-photon fluorescence microscopy

Adaptive optics two-photon fluorescence microscopy Adaptive optics two-photon fluorescence microscopy Yaopeng Zhou 1, Thomas Bifano 1 and Charles Lin 2 1. Manufacturing Engineering Department, Boston University 15 Saint Mary's Street, Brookline MA, 02446

More information

Optical Coherence: Recreation of the Experiment of Thompson and Wolf

Optical Coherence: Recreation of the Experiment of Thompson and Wolf Optical Coherence: Recreation of the Experiment of Thompson and Wolf David Collins Senior project Department of Physics, California Polytechnic State University San Luis Obispo June 2010 Abstract The purpose

More information

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009 Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009 Introduction of Fluoresence Confocal Microscopy The first confocal microscope was invented by Princeton

More information

Single-shot depth-section imaging through chromatic slit-scan confocal microscopy

Single-shot depth-section imaging through chromatic slit-scan confocal microscopy Single-shot depth-section imaging through chromatic slit-scan confocal microscopy Paul C. Lin, Pang-Chen Sun, Lijun Zhu, and Yeshaiahu Fainman A chromatic confocal microscope constructed with a white-light

More information

1 Laboratory 7: Fourier Optics

1 Laboratory 7: Fourier Optics 1051-455-20073 Physical Optics 1 Laboratory 7: Fourier Optics 1.1 Theory: References: Introduction to Optics Pedrottis Chapters 11 and 21 Optics E. Hecht Chapters 10 and 11 The Fourier transform is an

More information

Chapter Ray and Wave Optics

Chapter Ray and Wave Optics 109 Chapter Ray and Wave Optics 1. An astronomical telescope has a large aperture to [2002] reduce spherical aberration have high resolution increase span of observation have low dispersion. 2. If two

More information

Development of Laser Confocal Microscopy for Internal Defect Measurement

Development of Laser Confocal Microscopy for Internal Defect Measurement Development of Laser Confocal Microscopy for Internal Defect Measurement Chia-Liang Yeh*, Fu-Cheng Yang, Wei-Hsiung Tsai, and Keng-Li Lin Center for Measurement Standards, Industrial Technology Research

More information

3D light microscopy techniques

3D light microscopy techniques 3D light microscopy techniques The image of a point is a 3D feature In-focus image Out-of-focus image The image of a point is not a point Point Spread Function (PSF) 1D imaging 1 1 2! NA = 0.5! NA 2D imaging

More information

attocfm I for Surface Quality Inspection NANOSCOPY APPLICATION NOTE M01 RELATED PRODUCTS G

attocfm I for Surface Quality Inspection NANOSCOPY APPLICATION NOTE M01 RELATED PRODUCTS G APPLICATION NOTE M01 attocfm I for Surface Quality Inspection Confocal microscopes work by scanning a tiny light spot on a sample and by measuring the scattered light in the illuminated volume. First,

More information

The extended-focus, auto-focus and surface-profiling techniques of confocal microscopy

The extended-focus, auto-focus and surface-profiling techniques of confocal microscopy JOURNAL OF MODERN OPTICS, 1988, voi,. 35, NO. 1, 145-154 The extended-focus, auto-focus and surface-profiling techniques of confocal microscopy C. J. R. SHEPPARD and H. J. MATTHEWS University of Oxford,

More information

Basics of confocal imaging (part I)

Basics of confocal imaging (part I) Basics of confocal imaging (part I) Swiss Institute of Technology (EPFL) Faculty of Life Sciences Head of BIOIMAGING AND OPTICS BIOP arne.seitz@epfl.ch Lateral resolution BioImaging &Optics Platform Light

More information

A wavefront generator for complex pupil function synthesis and point spread function engineering

A wavefront generator for complex pupil function synthesis and point spread function engineering Journal of Microscopy, Vol. 197, Pt 3, March 2000, pp. 219±223. Received 27 September 1999; accepted 30 November 1999 SHORT COMMUNICATION A wavefront generator for complex pupil function synthesis and

More information

Mutually Optimizing Resolution Enhancement Techniques: Illumination, APSM, Assist Feature OPC, and Gray Bars

Mutually Optimizing Resolution Enhancement Techniques: Illumination, APSM, Assist Feature OPC, and Gray Bars Mutually Optimizing Resolution Enhancement Techniques: Illumination, APSM, Assist Feature OPC, and Gray Bars Bruce W. Smith Rochester Institute of Technology, Microelectronic Engineering Department, 82

More information

Optical coherence tomography

Optical coherence tomography Optical coherence tomography Peter E. Andersen Optics and Plasma Research Department Risø National Laboratory E-mail peter.andersen@risoe.dk Outline Part I: Introduction to optical coherence tomography

More information

1 Introduction. Research Article

1 Introduction. Research Article dv. Opt. Techn. 214; 3(4): 425 433 Research rticle Hiroki Yokozeki, Ryota Kudo, Satoru Takahashi* and Kiyoshi Takamasu Lateral resolution improvement of laser-scanning imaging for nano defects detection

More information

Zero Focal Shift in High Numerical Aperture Focusing of a Gaussian Laser Beam through Multiple Dielectric Interfaces. Ali Mahmoudi

Zero Focal Shift in High Numerical Aperture Focusing of a Gaussian Laser Beam through Multiple Dielectric Interfaces. Ali Mahmoudi 1 Zero Focal Shift in High Numerical Aperture Focusing of a Gaussian Laser Beam through Multiple Dielectric Interfaces Ali Mahmoudi a.mahmoudi@qom.ac.ir & amahmodi@yahoo.com Laboratory of Optical Microscopy,

More information

A 3D Profile Parallel Detecting System Based on Differential Confocal Microscopy. Y.H. Wang, X.F. Yu and Y.T. Fei

A 3D Profile Parallel Detecting System Based on Differential Confocal Microscopy. Y.H. Wang, X.F. Yu and Y.T. Fei Key Engineering Materials Online: 005-10-15 ISSN: 166-9795, Vols. 95-96, pp 501-506 doi:10.408/www.scientific.net/kem.95-96.501 005 Trans Tech Publications, Switzerland A 3D Profile Parallel Detecting

More information

Chapter 4: Fourier Optics

Chapter 4: Fourier Optics Chapter 4: Fourier Optics P4-1. Calculate the Fourier transform of the function rect(2x)rect(/3) The rectangular function rect(x) is given b 1 x 1/2 rect( x) when 0 x 1/2 P4-2. Assume that ( gx (, )) G

More information

Stereoscopic Hologram

Stereoscopic Hologram Stereoscopic Hologram Joonku Hahn Kyungpook National University Outline: 1. Introduction - Basic structure of holographic display - Wigner distribution function 2. Design of Stereoscopic Hologram - Optical

More information

Lab4 Hanbury Brown and Twiss Setup. Photon Antibunching

Lab4 Hanbury Brown and Twiss Setup. Photon Antibunching Lab4 Hanbury Brown and Twiss Setup. Photon Antibunching Shule Li Abstract Antibunching is a purely quantum effect and cannot be realized from the classical theory of light. By observing the antibunching

More information

BEAM HALO OBSERVATION BY CORONAGRAPH

BEAM HALO OBSERVATION BY CORONAGRAPH BEAM HALO OBSERVATION BY CORONAGRAPH T. Mitsuhashi, KEK, TSUKUBA, Japan Abstract We have developed a coronagraph for the observation of the beam halo surrounding a beam. An opaque disk is set in the beam

More information

Microscopy. Matti Hotokka Department of Physical Chemistry Åbo Akademi University

Microscopy. Matti Hotokka Department of Physical Chemistry Åbo Akademi University Microscopy Matti Hotokka Department of Physical Chemistry Åbo Akademi University What s coming Anatomy of a microscope Modes of illumination Practicalities Special applications Basic microscope Ocular

More information

Guide to SPEX Optical Spectrometer

Guide to SPEX Optical Spectrometer Guide to SPEX Optical Spectrometer GENERAL DESCRIPTION A spectrometer is a device for analyzing an input light beam into its constituent wavelengths. The SPEX model 1704 spectrometer covers a range from

More information

Enhancement of the lateral resolution and the image quality in a line-scanning tomographic optical microscope

Enhancement of the lateral resolution and the image quality in a line-scanning tomographic optical microscope Summary of the PhD thesis Enhancement of the lateral resolution and the image quality in a line-scanning tomographic optical microscope Author: Dudás, László Supervisors: Prof. Dr. Szabó, Gábor and Dr.

More information

Supplementary Figure 1. Effect of the spacer thickness on the resonance properties of the gold and silver metasurface layers.

Supplementary Figure 1. Effect of the spacer thickness on the resonance properties of the gold and silver metasurface layers. Supplementary Figure 1. Effect of the spacer thickness on the resonance properties of the gold and silver metasurface layers. Finite-difference time-domain calculations of the optical transmittance through

More information

Fastest high definition Raman imaging. Fastest Laser Raman Microscope RAMAN

Fastest high definition Raman imaging. Fastest Laser Raman Microscope RAMAN Fastest high definition Raman imaging Fastest Laser Raman Microscope RAMAN - 11 www.nanophoton.jp Observation A New Generation in Raman Observation RAMAN-11 developed by Nanophoton was newly created by

More information

BASICS OF CONFOCAL IMAGING (PART I)

BASICS OF CONFOCAL IMAGING (PART I) BASICS OF CONFOCAL IMAGING (PART I) INTERNAL COURSE 2012 LIGHT MICROSCOPY Lateral resolution Transmission Fluorescence d min 1.22 NA obj NA cond 0 0 rairy 0.61 NAobj Ernst Abbe Lord Rayleigh Depth of field

More information

Experimental demonstration of polarization-assisted transverse and axial optical superresolution

Experimental demonstration of polarization-assisted transverse and axial optical superresolution Optics Communications 241 (2004) 315 319 www.elsevier.com/locate/optcom Experimental demonstration of polarization-assisted transverse and axial optical superresolution Jason B. Stewart a, *, Bahaa E.A.

More information

(12) United States Patent (10) Patent No.: US 6,525,828 B1

(12) United States Patent (10) Patent No.: US 6,525,828 B1 USOO6525828B1 (12) United States Patent (10) Patent No.: US 6,525,828 B1 Grosskopf (45) Date of Patent: *Feb. 25, 2003 (54) CONFOCAL COLOR 5,978,095 A 11/1999 Tanaami... 356/445 6,031,661. A 2/2000 Tanaami...

More information

Parallel Mode Confocal System for Wafer Bump Inspection

Parallel Mode Confocal System for Wafer Bump Inspection Parallel Mode Confocal System for Wafer Bump Inspection ECEN5616 Class Project 1 Gao Wenliang wen-liang_gao@agilent.com 1. Introduction In this paper, A parallel-mode High-speed Line-scanning confocal

More information

Polarization Experiments Using Jones Calculus

Polarization Experiments Using Jones Calculus Polarization Experiments Using Jones Calculus Reference http://chaos.swarthmore.edu/courses/physics50_2008/p50_optics/04_polariz_matrices.pdf Theory In Jones calculus, the polarization state of light is

More information

Supplementary Figure 1. GO thin film thickness characterization. The thickness of the prepared GO thin

Supplementary Figure 1. GO thin film thickness characterization. The thickness of the prepared GO thin Supplementary Figure 1. GO thin film thickness characterization. The thickness of the prepared GO thin film is characterized by using an optical profiler (Bruker ContourGT InMotion). Inset: 3D optical

More information

Laser Beam Analysis Using Image Processing

Laser Beam Analysis Using Image Processing Journal of Computer Science 2 (): 09-3, 2006 ISSN 549-3636 Science Publications, 2006 Laser Beam Analysis Using Image Processing Yas A. Alsultanny Computer Science Department, Amman Arab University for

More information

AgilOptics mirrors increase coupling efficiency into a 4 µm diameter fiber by 750%.

AgilOptics mirrors increase coupling efficiency into a 4 µm diameter fiber by 750%. Application Note AN004: Fiber Coupling Improvement Introduction AgilOptics mirrors increase coupling efficiency into a 4 µm diameter fiber by 750%. Industrial lasers used for cutting, welding, drilling,

More information

Supplementary Materials

Supplementary Materials Supplementary Materials In the supplementary materials of this paper we discuss some practical consideration for alignment of optical components to help unexperienced users to achieve a high performance

More information

Lab Report 3: Speckle Interferometry LIN PEI-YING, BAIG JOVERIA

Lab Report 3: Speckle Interferometry LIN PEI-YING, BAIG JOVERIA Lab Report 3: Speckle Interferometry LIN PEI-YING, BAIG JOVERIA Abstract: Speckle interferometry (SI) has become a complete technique over the past couple of years and is widely used in many branches of

More information

Test procedures Page: 1 of 5

Test procedures Page: 1 of 5 Test procedures Page: 1 of 5 1 Scope This part of document establishes uniform requirements for measuring the numerical aperture of optical fibre, thereby assisting in the inspection of fibres and cables

More information

NSERC Summer Project 1 Helping Improve Digital Camera Sensors With Prof. Glenn Chapman (ENSC)

NSERC Summer Project 1 Helping Improve Digital Camera Sensors With Prof. Glenn Chapman (ENSC) NSERC Summer 2016 Digital Camera Sensors & Micro-optic Fabrication ASB 8831, phone 778-782-319 or 778-782-3814, Fax 778-782-4951, email glennc@cs.sfu.ca http://www.ensc.sfu.ca/people/faculty/chapman/ Interested

More information

Characteristics of point-focus Simultaneous Spatial and temporal Focusing (SSTF) as a two-photon excited fluorescence microscopy

Characteristics of point-focus Simultaneous Spatial and temporal Focusing (SSTF) as a two-photon excited fluorescence microscopy Characteristics of point-focus Simultaneous Spatial and temporal Focusing (SSTF) as a two-photon excited fluorescence microscopy Qiyuan Song (M2) and Aoi Nakamura (B4) Abstracts: We theoretically and experimentally

More information

High resolution extended depth of field microscopy using wavefront coding

High resolution extended depth of field microscopy using wavefront coding High resolution extended depth of field microscopy using wavefront coding Matthew R. Arnison *, Peter Török #, Colin J. R. Sheppard *, W. T. Cathey +, Edward R. Dowski, Jr. +, Carol J. Cogswell *+ * Physical

More information

PROCEEDINGS OF SPIE. Three-dimensional transfer functions

PROCEEDINGS OF SPIE. Three-dimensional transfer functions PROCEEDINGS OF SPIE SPIEDigitalLibrary.org/conference-proceedings-of-spie Three-dimensional transfer functions Colin J. R. Sheppard Colin J. R. Sheppard, "Three-dimensional transfer functions," Proc. SPIE

More information

Coherent Laser Measurement and Control Beam Diagnostics

Coherent Laser Measurement and Control Beam Diagnostics Coherent Laser Measurement and Control M 2 Propagation Analyzer Measurement and display of CW laser divergence, M 2 (or k) and astigmatism sizes 0.2 mm to 25 mm Wavelengths from 220 nm to 15 µm Determination

More information

Coherence of Light and Generation of Speckle Patterns in Photobiology and Photomedicine

Coherence of Light and Generation of Speckle Patterns in Photobiology and Photomedicine Coherence of Light and Generation of Speckle Patterns in Photobiology and Photomedicine Zeev Zalevsky 1* and Michael Belkin 1 Faculty of Engineering, Bar-Ilan University, Ramat-Gan 5900, Israel, Goldshleger

More information

Criteria for Optical Systems: Optical Path Difference How do we determine the quality of a lens system? Several criteria used in optical design

Criteria for Optical Systems: Optical Path Difference How do we determine the quality of a lens system? Several criteria used in optical design Criteria for Optical Systems: Optical Path Difference How do we determine the quality of a lens system? Several criteria used in optical design Computer Aided Design Several CAD tools use Ray Tracing (see

More information

Fluorescence Imaging of Single Spins in Nitrogen-Vacancy centers using a Confocal Microscope. Advanced Lab Course University of Basel

Fluorescence Imaging of Single Spins in Nitrogen-Vacancy centers using a Confocal Microscope. Advanced Lab Course University of Basel Fluorescence Imaging of Single Spins in Nitrogen-Vacancy centers using a Confocal Microscope Advanced Lab Course University of Basel October 6, 2015 Contents 1 Introduction 2 2 The Confocal Microscope

More information

Aberrations and adaptive optics for biomedical microscopes

Aberrations and adaptive optics for biomedical microscopes Aberrations and adaptive optics for biomedical microscopes Martin Booth Department of Engineering Science And Centre for Neural Circuits and Behaviour University of Oxford Outline Rays, wave fronts and

More information

Shreyash Tandon M.S. III Year

Shreyash Tandon M.S. III Year Shreyash Tandon M.S. III Year 20091015 Confocal microscopy is a powerful tool for generating high-resolution images and 3-D reconstructions of a specimen by using point illumination and a spatial pinhole

More information

TRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO

TRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO TRAINING MANUAL Multiphoton Microscopy LSM 510 META-NLO September 2010 Multiphoton Microscopy Training Manual Multiphoton microscopy is only available on the LSM 510 META-NLO system. This system is equipped

More information

Fast Laser Raman Microscope RAMAN

Fast Laser Raman Microscope RAMAN Fast Laser Raman Microscope RAMAN - 11 www.nanophoton.jp Fast Raman Imaging A New Generation of Raman Microscope RAMAN-11 developed by Nanophoton was created by combining confocal laser microscope technology

More information

APPLICATION NOTE

APPLICATION NOTE THE PHYSICS BEHIND TAG OPTICS TECHNOLOGY AND THE MECHANISM OF ACTION OF APPLICATION NOTE 12-001 USING SOUND TO SHAPE LIGHT Page 1 of 6 Tutorial on How the TAG Lens Works This brief tutorial explains the

More information

Option G 4:Diffraction

Option G 4:Diffraction Name: Date: Option G 4:Diffraction 1. This question is about optical resolution. The two point sources shown in the diagram below (not to scale) emit light of the same frequency. The light is incident

More information

Design Description Document

Design Description Document UNIVERSITY OF ROCHESTER Design Description Document Flat Output Backlit Strobe Dare Bodington, Changchen Chen, Nick Cirucci Customer: Engineers: Advisor committee: Sydor Instruments Dare Bodington, Changchen

More information

High-speed 1-frame ms scanning confocal microscope with a microlens and Nipkow disks

High-speed 1-frame ms scanning confocal microscope with a microlens and Nipkow disks High-speed 1-framems scanning confocal microscope with a microlens and Nipkow disks Takeo Tanaami, Shinya Otsuki, Nobuhiro Tomosada, Yasuhito Kosugi, Mizuho Shimizu, and Hideyuki Ishida We have developed

More information

Investigation of an optical sensor for small angle detection

Investigation of an optical sensor for small angle detection Investigation of an optical sensor for small angle detection usuke Saito, oshikazu rai and Wei Gao Nano-Metrology and Control Lab epartment of Nanomechanics Graduate School of Engineering, Tohoku University

More information

Optical design of a high resolution vision lens

Optical design of a high resolution vision lens Optical design of a high resolution vision lens Paul Claassen, optical designer, paul.claassen@sioux.eu Marnix Tas, optical specialist, marnix.tas@sioux.eu Prof L.Beckmann, l.beckmann@hccnet.nl Summary:

More information

A novel solution for various monitoring applications at CERN

A novel solution for various monitoring applications at CERN A novel solution for various monitoring applications at CERN F. Lackner, P. H. Osanna 1, W. Riegler, H. Kopetz CERN, European Organisation for Nuclear Research, CH-1211 Geneva-23, Switzerland 1 Department

More information

Thermal tuning of volume Bragg gratings for high power spectral beam combining

Thermal tuning of volume Bragg gratings for high power spectral beam combining Thermal tuning of volume Bragg gratings for high power spectral beam combining Derrek R. Drachenberg, Oleksiy Andrusyak, Ion Cohanoschi, Ivan Divliansky, Oleksiy Mokhun, Alexei Podvyaznyy, Vadim Smirnov,

More information

Programmable array microscopy with a ferroelectric liquid-crystal spatial light modulator

Programmable array microscopy with a ferroelectric liquid-crystal spatial light modulator Programmable array microscopy with a ferroelectric liquid-crystal spatial light modulator Patrick J. Smith, Cian M. Taylor, Alan J. Shaw, and Eithne M. McCabe We present a programmable array microscope

More information

The DCS-120 Confocal Scanning FLIM System

The DCS-120 Confocal Scanning FLIM System he DCS-120 Confocal Scanning FLIM System he bh DCS-120 confocal scanning FLIM system converts a conventional microscope into a high-performance fluorescence lifetime imaging system. he system is based

More information

Development of a High-speed Super-resolution Confocal Scanner

Development of a High-speed Super-resolution Confocal Scanner Development of a High-speed Super-resolution Confocal Scanner Takuya Azuma *1 Takayuki Kei *1 Super-resolution microscopy techniques that overcome the spatial resolution limit of conventional light microscopy

More information

Variable microinspection system. system125

Variable microinspection system. system125 Variable microinspection system system125 Variable micro-inspection system Characteristics Large fields, high NA The variable microinspection system mag.x system125 stands out from conventional LD inspection

More information

Laser Scanning 3D Display with Dynamic Exit Pupil

Laser Scanning 3D Display with Dynamic Exit Pupil Koç University Laser Scanning 3D Display with Dynamic Exit Pupil Kishore V. C., Erdem Erden and Hakan Urey Dept. of Electrical Engineering, Koç University, Istanbul, Turkey Hadi Baghsiahi, Eero Willman,

More information

Laser Scanning for Surface Analysis of Transparent Samples - An Experimental Feasibility Study

Laser Scanning for Surface Analysis of Transparent Samples - An Experimental Feasibility Study STR/03/044/PM Laser Scanning for Surface Analysis of Transparent Samples - An Experimental Feasibility Study E. Lea Abstract An experimental investigation of a surface analysis method has been carried

More information

CREATING ROUND AND SQUARE FLATTOP LASER SPOTS IN MICROPROCESSING SYSTEMS WITH SCANNING OPTICS Paper M305

CREATING ROUND AND SQUARE FLATTOP LASER SPOTS IN MICROPROCESSING SYSTEMS WITH SCANNING OPTICS Paper M305 CREATING ROUND AND SQUARE FLATTOP LASER SPOTS IN MICROPROCESSING SYSTEMS WITH SCANNING OPTICS Paper M305 Alexander Laskin, Vadim Laskin AdlOptica Optical Systems GmbH, Rudower Chaussee 29, 12489 Berlin,

More information

Fiber Optic Communications

Fiber Optic Communications Fiber Optic Communications ( Chapter 2: Optics Review ) presented by Prof. Kwang-Chun Ho 1 Section 2.4: Numerical Aperture Consider an optical receiver: where the diameter of photodetector surface area

More information

Using Stock Optics. ECE 5616 Curtis

Using Stock Optics. ECE 5616 Curtis Using Stock Optics What shape to use X & Y parameters Please use achromatics Please use camera lens Please use 4F imaging systems Others things Data link Stock Optics Some comments Advantages Time and

More information

Optical Components for Laser Applications. Günter Toesko - Laserseminar BLZ im Dezember

Optical Components for Laser Applications. Günter Toesko - Laserseminar BLZ im Dezember Günter Toesko - Laserseminar BLZ im Dezember 2009 1 Aberrations An optical aberration is a distortion in the image formed by an optical system compared to the original. It can arise for a number of reasons

More information

Applications of Optics

Applications of Optics Nicholas J. Giordano www.cengage.com/physics/giordano Chapter 26 Applications of Optics Marilyn Akins, PhD Broome Community College Applications of Optics Many devices are based on the principles of optics

More information

Modulation Transfer Function

Modulation Transfer Function Modulation Transfer Function The resolution and performance of an optical microscope can be characterized by a quantity known as the modulation transfer function (MTF), which is a measurement of the microscope's

More information

Optics and Lasers. Matt Young. Including Fibers and Optical Waveguides

Optics and Lasers. Matt Young. Including Fibers and Optical Waveguides Matt Young Optics and Lasers Including Fibers and Optical Waveguides Fourth Revised Edition With 188 Figures Springer-Verlag Berlin Heidelberg New York London Paris Tokyo Hong Kong Barcelona Budapest Contents

More information

FDTD Analysis of Readout Characteristics in a near-field MAMMOS recording system. Matthew Manfredonia Paul Nutter & David Wright

FDTD Analysis of Readout Characteristics in a near-field MAMMOS recording system. Matthew Manfredonia Paul Nutter & David Wright FDTD Analysis of Readout Characteristics in a near-field MAMMOS recording system Matthew Manfredonia Paul Nutter & David Wright Electronic & Information Storage Systems Research Group School of Computer

More information

Properties of Structured Light

Properties of Structured Light Properties of Structured Light Gaussian Beams Structured light sources using lasers as the illumination source are governed by theories of Gaussian beams. Unlike incoherent sources, coherent laser sources

More information

LEOK-3 Optics Experiment kit

LEOK-3 Optics Experiment kit LEOK-3 Optics Experiment kit Physical optics, geometrical optics and fourier optics Covering 26 experiments Comprehensive documents Include experiment setups, principles and procedures Cost effective solution

More information

Parallel Digital Holography Three-Dimensional Image Measurement Technique for Moving Cells

Parallel Digital Holography Three-Dimensional Image Measurement Technique for Moving Cells F e a t u r e A r t i c l e Feature Article Parallel Digital Holography Three-Dimensional Image Measurement Technique for Moving Cells Yasuhiro Awatsuji The author invented and developed a technique capable

More information

Focus detection in digital holography by cross-sectional images of propagating waves

Focus detection in digital holography by cross-sectional images of propagating waves Focus detection in digital holography by cross-sectional images of propagating waves Meriç Özcan Sabancı University Electronics Engineering Tuzla, İstanbul 34956, Turkey STRCT In digital holography, computing

More information

Akinori Mitani and Geoff Weiner BGGN 266 Spring 2013 Non-linear optics final report. Introduction and Background

Akinori Mitani and Geoff Weiner BGGN 266 Spring 2013 Non-linear optics final report. Introduction and Background Akinori Mitani and Geoff Weiner BGGN 266 Spring 2013 Non-linear optics final report Introduction and Background Two-photon microscopy is a type of fluorescence microscopy using two-photon excitation. It

More information

X-ray generation by femtosecond laser pulses and its application to soft X-ray imaging microscope

X-ray generation by femtosecond laser pulses and its application to soft X-ray imaging microscope X-ray generation by femtosecond laser pulses and its application to soft X-ray imaging microscope Kenichi Ikeda 1, Hideyuki Kotaki 1 ' 2 and Kazuhisa Nakajima 1 ' 2 ' 3 1 Graduate University for Advanced

More information

Three-dimensional behavior of apodized nontelecentric focusing systems

Three-dimensional behavior of apodized nontelecentric focusing systems Three-dimensional behavior of apodized nontelecentric focusing systems Manuel Martínez-Corral, Laura Muñoz-Escrivá, and Amparo Pons The scalar field in the focal volume of nontelecentric apodized focusing

More information

Tutorial Zemax 9: Physical optical modelling I

Tutorial Zemax 9: Physical optical modelling I Tutorial Zemax 9: Physical optical modelling I 2012-11-04 9 Physical optical modelling I 1 9.1 Gaussian Beams... 1 9.2 Physical Beam Propagation... 3 9.3 Polarization... 7 9.4 Polarization II... 11 9 Physical

More information