Quality Assurance Results for a Commercial Radiosurgery System: A Communication

Size: px
Start display at page:

Download "Quality Assurance Results for a Commercial Radiosurgery System: A Communication"

Transcription

1 Technology in Cancer Research and Treatment ISSN Volume 14 Number 5 October June 16. Epub ahead of print. Quality Assurance Results for a Commercial Radiosurgery System: A Communication DOI: /tcrt The purpose of this communication is to inform the radiosurgery community of quality assurance (QA) results requiring attention in a commercial FDA-approved linac-based cone stereotactic radiosurgery (SRS) system. Standard published QA guidelines as per the American Association of Physics in Medicine (AAPM) were followed during the SRS system s commissioning process including end-to-end testing, cone concentricity testing, image transfer verification, and documentation. Several software and hardware deficiencies that were deemed risky were uncovered during the process and QA processes were put in place to mitigate these risks during clinical practice. In particular, the present work focuses on daily cone concentricity testing and commissioning-related findings associated with the software. Cone concentricity/alignment is measured daily using both optical light field inspection, as well as quantitative radiation field tests with the electronic portal imager. In 10 out of 36 clini- cal treatments, adjustments to the cone position had to be made to align the cone with the collimator axis to less than 0.5 mm and on two occasions the pre-adjustment measured offset was 1.0 mm. Software-related errors discovered during commissioning included incorrect transfer of the isocentre in DICOM coordinates, improper handling of non-axial image sets, and complex handling of beam data, especially for multi-target treatments. QA processes were established to mitigate the occurrence of the software errors. With proper QA processes, the reported SRS system complies with tolerances set out in established guidelines. Discussions with the vendor are ongoing to address some of the hardware issues related to cone alignment. Mark Ruschin, Ph.D. 1,2 * Alexander Lightstone, Ph.D. 1,2 David Beachey, Ph.D. 1,2 Matt Wronski, Ph.D. 1 Steven Babic, Ph.D. 1 Collins Yeboah, Ph.D. 1,2 Young Lee, Ph.D. 1 Hany Soliman, M.D. 2,3 Arjun Sahgal, M.D. 2,3 1 Department of Medical Physics, Sunnybrook Odette Cancer Centre, Toronto, ON, Canada 2 Department of Radiation Oncology, University of Toronto, Toronto, ON, Canada 3 Department of Radiation Oncology, Sunnybrook Odette Cancer Centre, Toronto, ON, Canada Key words: Image-guidance; Radiosurgery; Stereotactic; Stereotactic cones. Introduction The AAPM Report 54 (Task Group 42) provides a detailed description of stereotactic radiosurgery (SRS) systems, sources of error, and quality assurance (QA) tolerances (1). For example, according to Report 54, all machine axes should coincide within a 1 mm radius sphere for all possible gantry, collimator and couch angles (1). More recent practice guidelines such as AAPM TG-142 and ACR guidelines have updated some tests, such as those related to cone-beam CT image guidance and frameless radiosurgery (2). Additionally, Canadian guidelines from the Canadian Abbreviations: AAPM: American Associate of Physicists in Medicine; ACR: American College of Radiology; CAPCA: Canadian Association of Provincial Cancer Agencies; CT: Computed Tomography; DICOM: Digital Imaging and Communications in Medicine; EPID: Electronic Portal Imaging Device; FDA: Food and Drug Administration; MR: Magnetic Resonance; MU: Monitor Unit; QA: Quality Assurance; R&V: Record and Verify; SRS: Stereotactic Radiosurgery; TPS: Treatment Planning System. *Corresponding author: Mark Ruschin, Ph.D. Phone: X Mark.Ruschin@sunnybrook.ca 601

2 602 Ruschin et al. Association of Provincial Cancer Agencies (CAPCA) have been developed with more specific tolerances and action levels for the various areas within radiation oncology. For example, based on CAPCA report on radiosurgery/radiotherapy, the tolerance for cone-alignment is 0.5 mm (3). In June of 2013 a new SRS system was installed at our institution and consisted of a set of stereotactic cones and an associated treatment planning system (TPS) from a major linac vendor (Elekta AB, Crawley, England). FDA clearance for the TPS system (ERGO++) and stereotactic cones was received in 2008 and 2001 respectively. Between June 2013 and October 2013, we performed a series of standard commissioning tests based on published guidelines, including independent dose verification by the Radiological Physics Center in Houston, Texas. From October 2013 to February 2014, we have used the system to treat 36 patients, comprising 82 brain metastases. The purpose of the present report is to inform the SRS community to look for areas of potential risks that were identified during the commissioning process and provide QA processes that can be utilized to mitigate those risks. Materials and Methods Stereotactic Cone Tray As shown in Figure 1, the cones are screwed into an aluminum tray that is mounted onto the linac head. The tray is on a translational stage that can be adjusted with two perpendicular micrometers. The cone tray is attached to the linac head with a set of four thumb screws, and held in place by friction. During commissioning, the stage was adjusted to centre the cone aperture on the mechanical isocentre under collimator rotation. A daily QA test was then implemented in which the Figure 1: Photograph of the adjustable cone tray. The cone tray is attached to the linac head with four thumb screws (one near each corner) and the central tray containing the cone can be translated using the micrometers (indicated with white arrows). electronic portal imaging device (EPID) captures two images of the circular cone separated by a collimator rotation of 180 degrees, as shown in Figure 2. The measured displacement of the cone aperture between the two images indicates the precession of the cone under collimator rotation. The alignment error between the collimator axis and the cone axis is approximately half the measured distance. Since the flat panel detector remains stationary between the two measurements, the difference between the two images is purely due to collimator precession, resulting from the cone long axis not being aligned with the collimator axis of rotation. The resolution of the images is 0.25 mm projected at the isocentre plane. A measured displacement of greater than 0.4 mm Figure 2: Day-of-treatment cone concentricity test. Left: EPID image of cone at Collimator Right: EPID image of cone at Collimator The distance from the digital cross-hair in purple (not indicative of the mechanical isocentre) to the edge of the collimator radiation field is measured on both images and the difference between the two measurements should be 0.4 mm, otherwise the cone tray may be adjusted. In this case, the distance is 0.0 mm.

3 Errors in Commercial Radiosurgery System 603 (i.e., 2 pixels) indicates a cone misalignment of greater than 0.2 mm and requires an investigation and potential adjustment in order to ensure that we are safely within the tolerance recommended in the CAPCA guidelines (3). We report the measured displacement for all 36 treatment days and the frequency of adjustments. Treatment Planning System The treatment planning system (TPS) consists of four modules: (1) Image management, (2) Localization, (3) Image fusion, and (4) Radiosurgery planning. Image datasets: During the commissioning process, we encountered several issues (i.e., software bugs) in terms of how the TPS handles image sets in the image management, image fusion, and radiosurgery planning modules. We report these bugs and how we have mitigated their effect. Transfer of beam data: Once a treatment plan has been generated in the radiosurgery planning module, a TPS report can be generated and the beam data is exported to the record and verify (R&V) system. Several errors/bugs and complications were encountered during the commissioning process that we will report and how we have attempted to mitigate their effects. It should be noted these types of errors tended to be exacerbated when planning for multiple lesions for which the TPS is limited in terms of its ability to handle. Isocentre coordinate for cone-beam CT image guidance: Along with the beam data, the isocentre coordinates in DICOM space are also sent to the R&V system for use with cone-beam CT image guidance. A software bug was encountered during commissioning, which bears reporting as a separate issue apart from the transfer of beam data. Results Stereotactic Cone Tray The displacement of the cone between collimator 5 90 and collimator as a function of treatment day is shown in Figure 3. On 10 of 36 clinical days, an adjustment to the cone tray was made using the micrometers to bring the test within a 0.4 mm tolerance. The physical distance to adjust the tray is roughly half of the measured displacement, i.e., if the displacement was 0.5 mm at the isocentre, a shift of approximately 0.25 mm was applied to correct it. The standard deviation in the pre-adjustment displacement value was 0.25 mm and 0.1 mm in the gun-target and left-right positions, respectively. While the measurement of cone displacement with collimator displacement is part of the standard patient QA test, we suggest that there should be mechanical pins to assist in re-locating this cone tray on the linac Figure 3: Plot of cone collimator precession each clinical day. G-T is Gun-Target direction and L-R is left-right. The dashed arrows point to the final value after an adjustment was made to the cone tray to bring the test within tolerance. Of 36 clinical days, adjustments were made on 10 days. head with accuracy better than 0.1 mm. We also recommend a plate to better protect the stage micrometers from accidental manipulation. Treatment Planning System Image datasets: A software bug was encountered in which the TPS accepts images that are tilted with respect to the true axial plane, but does not properly link contours or fusion transformations to this orientation. As a result, if a contour is drawn on a tilted MR image set the contour will appear shifted relative to its initial position the next time the image set is opened, as shown in Figure 4. Similarly, if the tilted MR image set is properly fused to the CT dataset in the Fusion module, the MR image set will appear shifted relative to the CT set the next time the MR is loaded. To mitigate this problem, we have taken steps to ensure that the primary images used for planning are not tilted. The Orientation tag in the DICOM header (0020, 0037) should be 1\0\0\0\1\0 for true axial images. If secondary images (such as MR) are tilted but the primary images (e.g., CT) are not, then make sure to assign all contours to the primary, non-tilted dataset and verify that they are in the correct location within the Fusion module. Transfer of beam data: The treatment plan report generated by the TPS includes (for each arc) the average depth, MU, and the dose at isocentre. Several issues have complicated the process for performing QA of the beam data transferred to the R&V system. First of all, the TPS report truncates the values of MUs and dose values, cutting off anything after the decimal place. As a result, all of MUs loaded up in the R&V differ from the TPS report by up to 0.9 MU per arc. Additional issues arise when multiple targets/lesions are planned. At our institution, planning for two or more brain metastases is common.

4 604 Ruschin et al. Figure 4: Example of re-opening a tilted image set on contour position. Image on the left is the original contour. The next two images are subsequent saves and re-opening of the images with continual displacement of the contour. Furthermore, each lesion may involve multiple isocentres. However, as the number of isocentres being used increases, it becomes increasingly challenging to safely track these data. The current TPS version assigns only one dose prescription per plan, linked to one isocentre. If a multiple isocentre plan is generated and exported, there is no identification given in the R&V as to which beam corresponds to which isocentre, or what prescription is assigned to any target. To mitigate the risk of inaccurate data transfer, all of our multi-isocentre plans are manually split up into individual plans, i.e., one plan for each isocentre. The composite plan is copied once for each iso, and in each copy all but one isocentre (the treated one) is deleted. The prescriptions are re-normalized to that isocentre. Since the dose per arc is also truncated, when the doses are summed up for a given target, the truncation error per arc also gets summed. Re-prescribing in the individual plans based on the erroneous summed dose has lead to discrepancies in MU of up to 5 MU per arc. Therefore, manual inspection and multiple checks by planning dosimetrist, physicist, and radiation therapist(s) are required. We have therefore developed an extensive and explicit protocol for handling the beam transfer, which involves independent checks that all of the plan data has been properly transferred to the R&V. The independent checks include: MU, jaw-size, couch and gantry angles, site setup (DICOM coordinate transfer), visual verification of isocentre position and distribution on CT/MR images, plan revision number verification, and setup-plates cross-verification with stereotactic coordinates. We also have implemented a secondary MU calculation for each isocentre using measured tissue-phantom-ratio and relative output factor look-up tables, with a tolerance of 2% for the agreement between planning and calculated MUs. Isocentre coordinate for cone-beam CT image guidance: As part of the published guidelines for quality assurance of cone-beam CT it is necessary to verify the transfer of information from the TPS to the treatment unit (4). Since cone-beam CT uses a different coordinate system than stereotactic, verification of the isocentre coordinates is vital. At our institution, we discovered that the conversion of stereotactic coordinates to DICOM coordinates involves a rounding in the longitudinal coordinate to the nearest CT slice, resulting in a potential error in isocentre position of up to half a CT slice thickness. At our institution, we use 1 mm slices for SRS planning, so the maximal error is 0.5 mm. This is small, but represents a systematic deviation that would otherwise go undetected. Institutions using larger CT slices would yield even greater errors. To mitigate this error, we were able to use a function within the TPS that can calculate a stereotactic coordinate from a user-entered DICOM coordinate using the stereotactic localization for that patient. We therefore verify that the DICOM coordinate being sent to the R&V system corresponds to the planned isocentre in stereotactic coordinates. Clinically, we almost always have to modify the exported longitudinal DICOM coordinate to recapitulate the true planned stereotactic coordinate. Multiple independent checks and communication is vital for this process. Discussion The present paper reports results of QA tests during commissioning and routine usage of a commercial FDA-approved SRS system. The software errors and bugs may not get addressed by the manufacturer since the ERGO++ TPS will be discontinued. However, the issues that were discovered in the present TPS should still be assessed and evaluated for any given TPS. In the meanwhile, the vendor has agreed to provide a small patch that would change the way the beams are labelled for easier identification, but each isocentre coordinate and prescription would still have to be manually entered and second-checked. Although the TPS may be replaced, the hardware (i.e., stereotactic cone assembly) will continue to be used and therefore

5 Errors in Commercial Radiosurgery System 605 represents a more long-term problem. As demonstrated in Figure 3, the cone axis has been shifted from the collimator mechanical axis by 0.5 mm (i.e., a measured displacement of 1 mm) in two clinical days. If all of the mechanical axes should align within 1 mm as per AAPM guidelines, the cone alignment with the collimator alone should be within 0.5 mm, and this is also the recommended tolerance of the CAPCA guidelines (3). The misalignment could be due to an accidental adjustment of the micrometers and/or due to the tray not being pinned or located to the linac head. We are already in communication with the vendor about potential redesigns of the current hardware. For the meanwhile, we have implemented daily EPID-based QA, which is performed immediately prior to the patient coming in for treatment. Conflict of Interest All authors certify that this manuscript has not been published in whole or in part nor is it being considered for publication elsewhere. Dr. Sahgal has received honoraria for educational seminars from Elekta AB, otherwise, the authors have no conflicts of interest to declare. References 1. Schell, M. C., Bova, F. J., Larson, D., Leavitt, D. D., Lutz, W., Podgorsak, E. B., Wu, A. Stereotactic Radiosurgery. American Association of Physicists in Medicine (AAPM) Report 54 (1995). 2. Seung, S. K., Larson, D. A., Galvin, J. M., Mehta, M. P., Potters, L., Schultz, C. J., Yajnik, S. V., Hartford, A. C., Rosenthal, S. A. American college of radiology (ACR) and American society for radiation oncology (ASTRO) practice guideline for the performance of stereotactic radiosurgery (SRS). American Journal of Clinical Oncology 36, (2013). DOI: /COC.0b013e31826e053d 3. Arsenault, C., Bissonnette, J. P., Dunscombe, P., Gallet, J., Mawko, G., Seuntjens, J. P. Canadian association of provincial cancer agencies (CAPCA): stereotactic radiosurgery/radiotherapy. Standards for Quality Control at Canadian Radiation Treatment Centres (2006). 4. Bissonnette, J. P., Balter, P. A., Dong, L., Langen, K. M., Lovelock, D. M., Miften, M., Moseley, D. J., Pouliot, J., Sonke, J. J., Yoo, S. Quality assurance for image-guided radiation therapy utilizing CTbased technologies: a report of the AAPM TG-179. Medical Physics 39, (2012). DOI: / Received: February 7, 2014; Accepted: February 26, 2014

Nathan Childress, Ph.D., DABR

Nathan Childress, Ph.D., DABR Nathan Childress, Ph.D., DABR Introduction TG-142 is a comprehensive QA protocol Covers nearly every aspect of machine and safety QA Recommends quantitative results Recommends high testing frequencies

More information

Mobius3D. Software based IMRT QA

Mobius3D. Software based IMRT QA Mobius3D Software based IMRT QA What is Mobius Medical Systems? Clinical Expertise Software Expertise Nathan Childress, Ph.D., Founder Eli Stevens, Chief Technical Officer Support Expertise Physicists

More information

Emerging Technology: Real-Time Monitoring of Treatment Delivery EPID Exit Dose QA

Emerging Technology: Real-Time Monitoring of Treatment Delivery EPID Exit Dose QA Emerging Technology: Real-Time Monitoring of Treatment Delivery EPID Exit Dose QA Arthur Olch, PhD, FAAPM AAPM Spring Clinical Meeting, March 21, 2017 Or.. What Dose are the Patients Really Getting???

More information

COMPREHENSIVE TG-142 IMAGING AND MACHINE QA

COMPREHENSIVE TG-142 IMAGING AND MACHINE QA QA SOFTWARE COMPREHENSIVE TG-142 IMAGING AND MACHINE QA Automate the analysis of over thirty TG-142 recommended QA tasks The rapid progress of Radiation Therapy has created the need for Quality Assurance

More information

7/23/2014. Acknowledgements. Implementing a new digital medical accelerator. New Generation of Medical Accelerators

7/23/2014. Acknowledgements. Implementing a new digital medical accelerator. New Generation of Medical Accelerators Implementing a new digital medical accelerator John Wong Johns Hopkins University AAPM, Austin, 2014 Acknowledgements Yin Zhang, Ken Wang, Kai Ding (Commissioning - JHU) Esteban Velarde, Joe Moore (QA

More information

The Current State of EPID-Based Linear Accelerator Quality Assurance. Disclosures. Purpose of this First Talk 8/3/2017

The Current State of EPID-Based Linear Accelerator Quality Assurance. Disclosures. Purpose of this First Talk 8/3/2017 The Current State of EPID-Based Linear Accelerator Quality Assurance Timothy Ritter, PhD, DABR, FAAPM 1 Disclosures Employed by the Veterans Health Administration Faculty appointment with the University

More information

A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup

A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 4, FALL 2009 A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup Huaiqun Guan,

More information

QUALITY CONTROL PHANTOMS FOR RADIOTHERAPY AND MEDICAL IMAGING

QUALITY CONTROL PHANTOMS FOR RADIOTHERAPY AND MEDICAL IMAGING 1 QUALITY CONTROL PHANTOMS FOR RADIOTHERAPY AND MEDICAL IMAGING QualiFormeD Phantoms A selection of test objects facilitating regulatory quality controls in radiation therapy and medical imaging Practical,

More information

Performance evaluation of the RITG148 + set of TomoTherapy quality assurance tools using RTQA 2 radiochromic film

Performance evaluation of the RITG148 + set of TomoTherapy quality assurance tools using RTQA 2 radiochromic film JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 4, 2016 Performance evaluation of the RITG148 + set of TomoTherapy quality assurance tools using RTQA 2 radiochromic film Eric C. Lobb Department

More information

PHYSICS QUESTIONNAIRE FORM

PHYSICS QUESTIONNAIRE FORM PHYSICS QUESTIONNAIRE FORM Institution Name: Date: Contact Information (name, address, phone, fax, email): Physicist: Radiation Oncologist: Dosimetrist (if applicable): Study Coordinator (if applicable):

More information

IQM Detector Characteristics: Signal reproducibility

IQM Detector Characteristics: Signal reproducibility The Integral Quality Monitor (IQM) System is a real-time beam verification system that monitors the accuracy of radiation delivery throughout each patient treatment without any user interaction. IQM continuously

More information

8/3/2017. Use of EPIDs for Non-Routine Linac QA. Disclosure. Learning Objectives. Parts of this project received support from Varian Medical System.

8/3/2017. Use of EPIDs for Non-Routine Linac QA. Disclosure. Learning Objectives. Parts of this project received support from Varian Medical System. Use of EPIDs for Non-Routine Linac QA Bin Cai PhD Disclosure Parts of this project received support from Varian Medical System. Learning Objectives Learn the recent development of EPID based Non-routine

More information

Sub-mm accuracy of accelerators: How manufacturers achieve it, how physicists verify it

Sub-mm accuracy of accelerators: How manufacturers achieve it, how physicists verify it Sub-mm accuracy of accelerators: How manufacturers achieve it, how physicists verify it Presented by Ivan A. Brezovich, PhD, at the 2015 Annual meeting of the SEAAPM, Raleigh, NC, April 24, 2015 Purpose

More information

DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE TG-148 RECOMMENDED TESTS 1. V.B.1.C. - Y-JAW DIVERGENCE/BEAM CENTERING

DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE TG-148 RECOMMENDED TESTS 1. V.B.1.C. - Y-JAW DIVERGENCE/BEAM CENTERING DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE Rev. 1.0 DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE DoseLab users may reference the following instructions to perform Tomotherapy Quality Assurance tests as recommended

More information

X3D in Radiation Therapy Procedure Planning. Felix G. Hamza-Lup, Ph.D. Computer Science Armstrong Atlantic State University Savannah, Georgia USA

X3D in Radiation Therapy Procedure Planning. Felix G. Hamza-Lup, Ph.D. Computer Science Armstrong Atlantic State University Savannah, Georgia USA X3D in Radiation Therapy Procedure Planning Felix G. Hamza-Lup, Ph.D. Computer Science Armstrong Atlantic State University Savannah, Georgia USA Outline 1. What is radiation therapy? 2. Treatment planning

More information

ISO Cube Daily QA Package

ISO Cube Daily QA Package ISO Cube Daily QA Package Model 023-05 AFFORDABLE TURNKEY SOLUTION FOR DAILY MACHINE QA POWERED BY AQUILAB 2428 Almeda Avenue Suite 316 Norfolk, Virginia 23513 USA Tel: 757-855-2765 WWW.CIRSINC.COM CAPABILITIES

More information

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 14 CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 2.1 INTRODUCTION kv-cbct integrated with linear accelerators as a tool for IGRT, was developed to

More information

A device for precision positioning and alignment of room lasers to diminish their contribution to patient setup errors

A device for precision positioning and alignment of room lasers to diminish their contribution to patient setup errors JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 8, NUMBER 4, FALL 2007 A device for precision positioning and alignment of room lasers to diminish their contribution to patient setup errors Ivan A.

More information

Evaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array

Evaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array Evaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array José A. Bencomo, * Geoffrey Ibbott, Seungsoo Lee, and Joao A. Borges Department of Radiation Physics.

More information

3D Diode Array Commissioning: Building Confidence in 3D QA Technology

3D Diode Array Commissioning: Building Confidence in 3D QA Technology 3D Diode Array Commissioning: Building Confidence in 3D QA Technology Caroline Yount, MS CANCER CENTER 3D QA The complex three-dimensional (3D) shapes of intensity modulated radiation therapy (IMRT) dose

More information

Volumetric Modulated Arc Therapy. David Shepard Swedish Cancer Institute Seattle, WA

Volumetric Modulated Arc Therapy. David Shepard Swedish Cancer Institute Seattle, WA Volumetric Modulated Arc Therapy David Shepard Swedish Cancer Institute Seattle, WA Disclaimer Our VMAT work has been sponsored in part by Elekta. Outline David Shepard VMAT Basics and VMAT Plan Quality

More information

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Acceptance Testing of a Digital Breast Tomosynthesis Unit Acceptance Testing of a Digital Breast Tomosynthesis Unit 2012 AAPM Spring Clinical Meeting Jessica Clements, M.S., DABR Objectives Review of technology and clinical advantages Acceptance Testing Procedures

More information

GafChromic QuiCk Phantom with EBT3P/3+P Film and FilmQA Pro for Radiation Therapy Dosimetry Applications

GafChromic QuiCk Phantom with EBT3P/3+P Film and FilmQA Pro for Radiation Therapy Dosimetry Applications GafChromic QuiCk Phantom with EBT3P/3+P Film and FilmQA Pro for Radiation Therapy Dosimetry Applications I. SCOPE The protocol applies to GafChromic EBT3P and EBT3+P films exposed in GafChromic QuiCk Phantom

More information

QC Testing for Computed Tomography (CT) Scanner

QC Testing for Computed Tomography (CT) Scanner QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including

More information

CyberKnife Iris Beam QA using Fluence Divergence

CyberKnife Iris Beam QA using Fluence Divergence CyberKnife Iris Beam QA using Fluence Divergence Ronald Berg, Ph.D., Jesse McKay, M.S. and Brett Nelson, M.S. Erlanger Medical Center and Logos Systems, Scotts Valley, CA Introduction The CyberKnife radiosurgery

More information

ISPFILMQATM STATE-OF-THE-ART RADIOTHERAPY VERIFICATION SOFTWARE. Supports all major radiotherapy technologies! FilmQA TM

ISPFILMQATM STATE-OF-THE-ART RADIOTHERAPY VERIFICATION SOFTWARE. Supports all major radiotherapy technologies! FilmQA TM FILMQA STATE-OF-THE-ART RADIOTHERAPY VERIFICATION SOFTWARE Supports all major radiotherapy technologies! FilmQA is optimized for use with Gafchromic film products, including EBT2 and RTQA2. FILMQA helps

More information

SUBCHAPTER 14. THERAPEUTIC INSTALLATIONS

SUBCHAPTER 14. THERAPEUTIC INSTALLATIONS SUBCHAPTER 14. THERAPEUTIC INSTALLATIONS 7:28-14.1 Scope (a) This subchapter covers therapeutic installations used in the healing arts. These therapeutic installations include x-ray, accelerator and teletherapy

More information

ArcCHECK, ein neuartiger QS-Ansatz bei der Rotationsbestrahlung

ArcCHECK, ein neuartiger QS-Ansatz bei der Rotationsbestrahlung ArcCHECK, ein neuartiger QS-Ansatz bei der Rotationsbestrahlung Treffen des Arbeitskreises IMRT der DGMP Würzburg, 26 + 27.03.2009 Salih Arican Sun Nuclear Corporation QA Challenge for Rotational Beams

More information

DORADOnova MR3T L ASER SYSTEM FOR RADIATION ONCOLOGY MRI

DORADOnova MR3T L ASER SYSTEM FOR RADIATION ONCOLOGY MRI EN DORADOnova MR3T L ASER SYSTEM FOR RADIATION ONCOLOGY MRI UNIQUE DORADOnova TECHNOLOGY A FAILSAFE SYSTEM The built-in microprocessor and linear encoder checks for possible stepper motor misses which,

More information

CT Scanner Dose Survey

CT Scanner Dose Survey CT Scanner Dose Survey Measurement Protocol Version 5.0 July 1997 Co-ordinated by ImPACT and The Medical Physics Department,, London SW17, UK. 0181-725-3366 CT Scanner Dose Survey: Measurement Protocol

More information

A Beam-Level Delivery Accuracy Study of the Robotic Image Guided Radiosurgery System Using a Scintillator/CCD Phantom

A Beam-Level Delivery Accuracy Study of the Robotic Image Guided Radiosurgery System Using a Scintillator/CCD Phantom A Beam-Level Delivery Accuracy Study of the Robotic Image Guided Radiosurgery System Using a Scintillator/CCD Phantom Lei Wang 1, Shi Liu 1, Brett Nelson 2 1. Department of Radiation Oncology, Stanford

More information

Eric B. Burgh University of Wisconsin. 1. Scope

Eric B. Burgh University of Wisconsin. 1. Scope Southern African Large Telescope Prime Focus Imaging Spectrograph Optical Integration and Testing Plan Document Number: SALT-3160BP0001 Revision 5.0 2007 July 3 Eric B. Burgh University of Wisconsin 1.

More information

SRS MapCHECK. SRS Patient QA, No Film. Your Most Valuable QA and Dosimetry Tools

SRS MapCHECK. SRS Patient QA, No Film. Your Most Valuable QA and Dosimetry Tools SRS MapCHECK SRS Patient QA, No Film Your Most Valuable QA and Dosimetry Tools SRS Patient QA, No Film With improvements in targeting and localization, stereotactic treatments have become prevalent. To

More information

Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA

Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA Gena M.A.H 1, Ahmed L.El-Attar 2, Elbadry M. Zahran 3, Hany El-Gamal

More information

A new approach to measure dwell position inaccuracy in HDR ring applicators quantification and corrective QA

A new approach to measure dwell position inaccuracy in HDR ring applicators quantification and corrective QA JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 12, NUMBER 1, WINTER 2010 A new approach to measure dwell position inaccuracy in HDR ring applicators quantification and corrective QA Abdul Qadir Jangda,

More information

Geometric Dimensioning and Tolerancing

Geometric Dimensioning and Tolerancing Geometric Dimensioning and Tolerancing (Known as GDT) What is GDT Helps ensure interchangeability of parts. Use is dictated by function and relationship of the part feature. It does not take the place

More information

C a t p h a n. T h e P h a n t o m L a b o r a t o r y. Ordering Information

C a t p h a n. T h e P h a n t o m L a b o r a t o r y. Ordering Information Ordering Information Please contact us if you have any questions or if you would like a quote or delivery schedule regarding the Catphan phantom. phone 800-525-1190, or 518-692-1190 fax 518-692-3329 mail

More information

Isocenter and Field of View Accuracy Measurement Software for Linear Accelerator

Isocenter and Field of View Accuracy Measurement Software for Linear Accelerator Isocenter and Field of View Accuracy Measurement Software for Linear Accelerator Aleksei E. Zhdanov 1 and Leonid G. Dorosinskiy 1 Ural Federal University named after the first President of Russia B. N.

More information

IMRT Delivery System QA. IMRT Dose Delivery. Acceptance testing. Why: specific tests for IMRT? Accuracy of leaf positioning (gaps) MLC Alignment

IMRT Delivery System QA. IMRT Dose Delivery. Acceptance testing. Why: specific tests for IMRT? Accuracy of leaf positioning (gaps) MLC Alignment 1 IMRT Delivery System Q Thomas LoSasso, PhD Memorial Sloan Kettering Cancer Center IMRT Dose Delivery cceptance testing Commissioning Quality assurance Verification Q Why: specific tests for IMRT? 2.

More information

Tomophan TSP004 Manual

Tomophan TSP004 Manual T h e P h a n t o m L a b o r a t o r y 1 Tomophan TSP004 Manual Copyright 2016 WARRANTY THE PHANTOM LABORATORY INCORPORATED ( Seller ) warrants that this product shall remain in good working order and

More information

QUALITY CONTROL SOFTWARE FOR RADIOTHERAPY AND MEDICAL IMAGING

QUALITY CONTROL SOFTWARE FOR RADIOTHERAPY AND MEDICAL IMAGING 1 QUALITY CONTROL SOFTWARE FOR RADIOTHERAPY AND MEDICAL IMAGING The QUALIMAGIQ platform and its modules An optimal software solution for automatic analyses and traceability of quality controls in radiation

More information

LINEAR ACCELERATOR. Buyer's Guide. Version 1.1

LINEAR ACCELERATOR. Buyer's Guide. Version 1.1 PRE-OWNED LINEAR ACCELERATOR Buyer's Guide Version 1.1 Pre-Owned Linear Accelerator Buyer's Guide TABLE OF CONTENTS Considerations For Buying A Used Linear Accelerator... 3 Linear Accelerators Overview...

More information

Accuracy of SRS dose delivery using the TomoTherapy Hi-Art System

Accuracy of SRS dose delivery using the TomoTherapy Hi-Art System Louisiana State University LSU Digital Commons LSU Master's Theses Graduate School 2010 Accuracy of SRS dose delivery using the TomoTherapy Hi-Art System Catherine Leigh Batte Louisiana State University

More information

MOVING LASER SYSTEMS IN RT

MOVING LASER SYSTEMS IN RT MOVING LASER SYSTEMS IN RT ESSENTIAL FOR PRECISE PATIENT ALIGNMENT NEXT GENERATION LAP IS YOUR IDEAL PARTNER FOR YOUR EXTERNAL LASER SYSTEMS OUR MOVING LASER SYSTEMS ARE RECOMMENDED BY LEADING CT SCANNER

More information

A proposed method for linear accelerator photon beam steering using EPID

A proposed method for linear accelerator photon beam steering using EPID Received: 13 January 2018 Revised: 11 May 2018 Accepted: 29 June 2018 DOI: 10.1002/acm2.12419 RADIATION ONCOLOGY PHYSICS A proposed method for linear accelerator photon beam steering using EPID Michael

More information

Comparative performance evaluation of a new a-si EPID that exceeds quad high-definition resolution

Comparative performance evaluation of a new a-si EPID that exceeds quad high-definition resolution JBUON 2018; 23(2): 507-513 ISSN: 1107-0625, online ISSN: 2241-6293 www.jbuon.com E-mail: editorial_office@jbuon.com ORIGINAL ARTICLE Comparative performance evaluation of a new a-si EPID that exceeds quad

More information

SUN NUCLEAR. EPIDose : An Overview of EPIDose and the EPIDose Process and Algorithm. corporation. Your Most Valuable QA and Dosimetry Tools

SUN NUCLEAR. EPIDose : An Overview of EPIDose and the EPIDose Process and Algorithm. corporation. Your Most Valuable QA and Dosimetry Tools EPIDose : An Overview of EPIDose and the EPIDose Process and Algorithm SUN NUCLEAR corporation Your Most Valuable QA and Dosimetry Tools introduction Pre-treatment dose QA is an important process required

More information

QA Considerations. QA for LGK Perfexion : : Follow NRC licensing guidelines (10( CFR ) Leksell Gamma Knife Perfexion

QA Considerations. QA for LGK Perfexion : : Follow NRC licensing guidelines (10( CFR ) Leksell Gamma Knife Perfexion Leksell Gamma Knife Perfexion QA Considerations Paula L. Petti, Ph.D. Taylor McAdam Bell Neuroscience Institute Washington Hospital Healthcare System Fremont, CA 1 QA for LGK Perfexion : : Follow NRC licensing

More information

Slide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC?

Slide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC? Slide 1 Multi-Slice CT Artifacts and Quality Control Dianna Cody, Ph.D. Chief, Radiologic Physics UT MD Anderson Cancer Center Houston, TX Slide 2 What are the rules or recommendations for CT QC? AAPM

More information

Gantry angle determination during arc IMRT: evaluation of a simple EPID-based technique and two commercial inclinometers

Gantry angle determination during arc IMRT: evaluation of a simple EPID-based technique and two commercial inclinometers JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 6, 2012 Gantry angle determination during arc IMRT: evaluation of a simple EPID-based technique and two commercial inclinometers Pejman Rowshanfarzad,

More information

T h e P h a n t o m L a b o r a t o r y

T h e P h a n t o m L a b o r a t o r y T h e P h a n t o m L a b o r a t o r y 1 CCT228 ATCM Phantom Manual Copyright 2017 WARRANTY THE PHANTOM LABORATORY INCORPORATED ( Seller ) warrants that this product shall remain in good working order

More information

Uncertainty in CT Metrology: Visualizations for Exploration and Analysis of Geometric Tolerances

Uncertainty in CT Metrology: Visualizations for Exploration and Analysis of Geometric Tolerances Uncertainty in CT Metrology: Visualizations for Exploration and Analysis of Geometric Tolerances Artem Amirkhanov 1, Bernhard Fröhler 1, Michael Reiter 1, Johann Kastner 1, M. Eduard Grӧller 2, Christoph

More information

diagnostic examination

diagnostic examination RADIOLOGICAL PHYSICS 2011 Raphex diagnostic examination Adel A. Mustafa, Ph.D., Editor PUBLISHED FOR: RAMPS (Radiological and Medical Physics Society of New York) preface The RAPHEX Diagnostic exam 2011

More information

Impact of energy variation on Cone Ratio, PDD10, TMR20 10 and IMRT doses for flattening filter free (FFF) beam of TomoTherapy Hi-Art TM machines

Impact of energy variation on Cone Ratio, PDD10, TMR20 10 and IMRT doses for flattening filter free (FFF) beam of TomoTherapy Hi-Art TM machines JBUON 2014; 19(4): 1105-1110 ISSN: 1107-0625, online ISSN: 2241-6293 www.jbuon.com E-mail: editorial_office@jbuon.com ORIGINAL ARTICLE Impact of energy variation on Cone Ratio, PDD10, TMR20 10 and IMRT

More information

COMMON SYMBOLS/ ISO SYMBOL ASME Y14.5M ISO FEATURE CONTROL FRAME DIAMETER/ SPHERICAL DIAMETER/ AT MAXIMUM MATERIAL CONDITION

COMMON SYMBOLS/ ISO SYMBOL ASME Y14.5M ISO FEATURE CONTROL FRAME DIAMETER/ SPHERICAL DIAMETER/ AT MAXIMUM MATERIAL CONDITION 1 82 COMMON SYMBOLS/ Shown below are the most common symbols that are used with geometric tolerancing and other related dimensional requirements on engineering drawings. Note the comparison with the ISO

More information

Sensitivity study of an automated system for daily patient QA using EPID exit dose images

Sensitivity study of an automated system for daily patient QA using EPID exit dose images Received: 27 June 2017 Revised: 8 December 2017 Accepted: 27 January 2018 DOI: 10.1002/acm2.12303 RADIATION ONCOLOGY PHYSICS Sensitivity study of an automated system for daily patient QA using EPID exit

More information

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS JAMES A. TOMLINSON, M.S. Diagnostic Radiological Physicist American Board of Radiology Certified Medical Physics Consultants, Inc. Bio 28 yrs experience 100%

More information

Published in: Radiation Oncology. Document Version: Publisher's PDF, also known as Version of record

Published in: Radiation Oncology. Document Version: Publisher's PDF, also known as Version of record An EPID-based method for comprehensive verification of gantry, EPID and the MLC carriage positional accuracy in Varian linacs during arc treatments Rowshanfarzad, P., McGarry, C., Barnes, M. P., Sabet,

More information

Engineering Working Drawings Basics

Engineering Working Drawings Basics Engineering Working Drawings Basics Engineering graphics is an effective way of communicating technical ideas and it is an essential tool in engineering design where most of the design process is graphically

More information

Aperture Based Inverse Planning AAPM Summer School 2003

Aperture Based Inverse Planning AAPM Summer School 2003 Aperture Based Inverse Planning AAPM Summer School 003 D.M. Shepard, M.A. Earl, Y. Xiao, C.X. Yu Acknowledgements Ziping Jiang Allen Li Shahid Naqvi James Galvin Di Yan Prowess, Inc. University of Maryland

More information

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging Shawn D. Teague, MD DISCLOSURES 3DR- advisory committee CT PHYSICS WITH AN EMPHASIS ON APPLICATION IN THORACIC AND CARDIAC IMAGING

More information

A Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images

A Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images A Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images Aiping Ding, Bin Han, Lei Wang, Lei Xing Department of Radiation Oncology, Stanford University School of

More information

Initial setup and subsequent temporal position monitoring using implanted RF transponders

Initial setup and subsequent temporal position monitoring using implanted RF transponders Initial setup and subsequent temporal position monitoring using implanted RF transponders James Balter, Ph.D. University of Michigan Has financial interest in Calypso Medical Technologies Acknowledgements

More information

An Introduction to TG-142 Imaging QA Using Standard Imaging Products. Mark Wiesmeyer, PhD, DABR Technical Product Manager Standard Imaging, Inc.

An Introduction to TG-142 Imaging QA Using Standard Imaging Products. Mark Wiesmeyer, PhD, DABR Technical Product Manager Standard Imaging, Inc. An Introduction to TG-142 Imaging QA Using Standard Imaging Products Mark Wiesmeyer, PhD, DABR Technical Product Manager Standard Imaging, Inc. Goals Understand the nature and intent of TG 142 imaging

More information

Quality Control for Stereotactic Breast Biopsy. Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc

Quality Control for Stereotactic Breast Biopsy. Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc Quality Control for Stereotactic Breast Biopsy Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc. 716-924-0350 Methods of Imaging Guided Breast Biopsy Ultrasound guided, hand-held needle

More information

Elementary Dimensioning

Elementary Dimensioning Elementary Dimensioning Standards Institutions ANSI - American National Standards Institute - creates the engineering standards for North America. ISO - International Organization for Standardization -

More information

Calibration of KAP meters

Calibration of KAP meters Calibration of KAP meters Alexandr Malusek! Division of Radiological Sciences Department of Medical and Health Sciences Linköping University! 2014-04-15 1 Outline 1. KAP meter construction 2. Air kerma-area

More information

SRS MapCHECK. SRS Patient QA, No Film. Your Most Valuable QA and Dosimetry Tools

SRS MapCHECK. SRS Patient QA, No Film. Your Most Valuable QA and Dosimetry Tools SRS MapCHECK SRS Patient QA, No Film Your Most Valuable QA and Dosimetry Tools SRS Patient QA, No Film As clinics strive to treat with the smallest field possible, the demand for SRS/SBRT QA grows as well.

More information

Measurement of table feed speed in modern CT

Measurement of table feed speed in modern CT JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 3, 2014 Measurement of table feed speed in modern CT Atsushi Fukuda, 1,2a Pei-Jan P. Lin, 3 Kosuke Matsubara, 2 Tosiaki Miyati 2 Department

More information

Investigation of the mechanical performance of Siemens linacs components during arc: gantry, MLC, and electronic portal imaging device

Investigation of the mechanical performance of Siemens linacs components during arc: gantry, MLC, and electronic portal imaging device Medical Devices: Evidence and Research open access to scientific and medical research Open Access Full Text Article Original Research Investigation of the mechanical performance of Siemens linacs components

More information

Module-4 Lecture-2 Perpendicularity measurement. (Refer Slide Time: 00:13)

Module-4 Lecture-2 Perpendicularity measurement. (Refer Slide Time: 00:13) Metrology Prof. Dr. Kanakuppi Sadashivappa Department of Industrial and Production Engineering Bapuji Institute of Engineering and Technology-Davangere Module-4 Lecture-2 Perpendicularity measurement (Refer

More information

Imaging system QA of a medical accelerator, Novalis Tx, for IGRT per TG 142: our 1 year experience

Imaging system QA of a medical accelerator, Novalis Tx, for IGRT per TG 142: our 1 year experience JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 4, 2012 Imaging system QA of a medical accelerator, Novalis Tx, for IGRT per TG 142: our 1 year experience Zheng Chang, a James Bowsher, Jing

More information

DICOM Correction Item

DICOM Correction Item DICOM Correction Item Correction Number CP- 617 Log Summary: Type of Modification Addition Name of Standard PS 3.3 2006 Rationale for Correction: The motion of modern patient support devices is no longer

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 07-647 R/F QC Phantom Operators Manual March 2005 Manual No. 07-647-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. All product names are trademarks of their respective companies

More information

Distributed source x-ray tube technology for tomosynthesis imaging

Distributed source x-ray tube technology for tomosynthesis imaging Distributed source x-ray tube technology for tomosynthesis imaging Authors: F. Sprenger a*, X. Calderon-Colon b, Y. Cheng a, K. Englestad a, J. Lu b, J. Maltz c, A. Paidi c, X. Qian b, D. Spronk a, S.

More information

Characterization of an in vivo diode dosimetry system for clinical use

Characterization of an in vivo diode dosimetry system for clinical use JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 2, SPRING 2003 Characterization of an in vivo diode dosimetry system for clinical use Kai Huang, 1, * William S. Bice, Jr., 2, and Oscar Hidalgo-Salvatierra

More information

Test Answers and Exam Booklet. Geometric Tolerancing

Test Answers and Exam Booklet. Geometric Tolerancing Test Answers and Exam Booklet Geometric Tolerancing iii Contents ANSWERS TO THE GEOMETRIC TOLERANCING TEST............. 1 Part 1. Questions Part 2. Calculations SAMPLE ANSWERS TO THE GEOMETRIC TOLERANCING

More information

Answers to Questions and Problems

Answers to Questions and Problems Fundamentals of Geometric Dimensioning and Tolerancing Using Critical Thinking Skills 3 rd Edition By Alex Krulikowski Answers to Questions and Problems Second Printing Product #: 1103 Price: $25.00 Copyright

More information

Characterization, Commissioning and Evaluation of Delta 4 IMRT QA System. Ram Sadagopan 1 UTMD Anderson Cancer Center Houston, TX.

Characterization, Commissioning and Evaluation of Delta 4 IMRT QA System. Ram Sadagopan 1 UTMD Anderson Cancer Center Houston, TX. Characterization, Commissioning and Evaluation of Delta 4 IMRT QA System Ram Sadagopan 1 UTMD Anderson Cancer Center Houston, TX. 1 Acknowledgements Collaborators: Jose Bencomo, Rafael. M. Landrove, Peter

More information

Diagnostic X-Ray Shielding

Diagnostic X-Ray Shielding Diagnostic X-Ray Shielding Multi-Slice CT Scanners Using NCRP 147 Methodology Melissa C. Martin, M.S., FAAPM, FACR Therapy Physics Inc., Bellflower, CA AAPM Annual Meeting, Orlando, FL FL Refresher Course

More information

Addressing Limitations of a Spatially Sensitive Large-Area Ion Chamber for Real-Time Verification of Intensity Modulated Radiation Therapy.

Addressing Limitations of a Spatially Sensitive Large-Area Ion Chamber for Real-Time Verification of Intensity Modulated Radiation Therapy. Addressing Limitations of a Spatially Sensitive Large-Area Ion Chamber for Real-Time Verification of Intensity Modulated Radiation Therapy. by Xun Lin A thesis submitted in conformity with the requirements

More information

A feasibility study of using conventional jaws to deliver IMRT plans in the treatment of prostate cancer *

A feasibility study of using conventional jaws to deliver IMRT plans in the treatment of prostate cancer * IOP PUBLISHING Phys. Med. Biol. 52 (7) 2147 2156 PHYSICS IN MEDICINE AND BIOLOGY doi:1.188/31-9155/52/8/7 A feasibility study of using conventional jaws to deliver IMRT plans in the treatment of prostate

More information

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation AAPM Annual Clinical Meeting Indianapolis, IN August 5, 2013 Learning Objectives Become familiar with the recommendations and

More information

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London Automated dose control in multi-slice CT Nicholas Keat Formerly ImPACT, St George's Hospital, London Introduction to presentation CT contributes ~50+ % of all medical radiation dose Ideally all patients

More information

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions TOPICS: Computed Tomography Quick Overview CT Dosimetry Effects of CT

More information

SUSPENSION CRITERIA FOR IMAGE MONITORS AND VIEWING BOXES.

SUSPENSION CRITERIA FOR IMAGE MONITORS AND VIEWING BOXES. SUSPENSION CRITERIA FOR IMAGE MONITORS AND VIEWING BOXES. Tingberg, Anders Published in: Radiation Protection Dosimetry DOI: 10.1093/rpd/ncs302 Published: 2013-01-01 Link to publication Citation for published

More information

Laser Trackers for Production of Automotive Tooling

Laser Trackers for Production of Automotive Tooling Case Study Laser Trackers for Production of Automotive Tooling PICO EUROPE designs, manufactures, installs and commissions automotive production lines worldwide for many automotive manufacturers including

More information

Clinical experience with EPID dosimetry for prostate IMRT pre-treatment dose verification

Clinical experience with EPID dosimetry for prostate IMRT pre-treatment dose verification Clinical experience with EPID dosimetry for prostate IMRT pre-treatment dose verification L. N. McDermott, M. Wendling, B. van Asselen, J. Stroom, J.-J. Sonke, M. van Herk, and B. J. Mijnheer a Department

More information

Development of a Virtual Simulation Environment for Radiation Treatment Planning

Development of a Virtual Simulation Environment for Radiation Treatment Planning Journal of Medical and Biological Engineering, 25(2): 61-66 61 Development of a Virtual Simulation Environment for Radiation Treatment Planning Tai-Sin Su De- Kai Chen Wen-Hsu Sung Ching-Fen Jiang * Shuh-Ping

More information

Computed Tomography. The Fundamentals of... THE FUNDAMENTALS OF... Jason H. Launders, MSc. Current Technology

Computed Tomography. The Fundamentals of... THE FUNDAMENTALS OF... Jason H. Launders, MSc. Current Technology The Fundamentals of... Computed Tomography Computed Tomography (CT) systems use x-rays to produce images of slices through a patient s anatomy. Despite having lower spatial resolution than other x-ray

More information

FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION

FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION Revised November 15, 2017 INTRODUCTION The simplest and most commonly described examples of diffraction and interference from two-dimensional apertures

More information

Evaluation of a diode array for QA measurements on a helical tomotherapy unit

Evaluation of a diode array for QA measurements on a helical tomotherapy unit Evaluation of a diode array for QA measurements on a helical tomotherapy unit K. M. Langen, a S. L. Meeks, D. O. Poole, T. H. Wagner, T. R. Willoughby, O. A. Zeidan, and P. A. Kupelian Department of Radiation

More information

END-TO-END TESTING OF THE LEKSELL GAMMA KNIFE ICON SYSTEM

END-TO-END TESTING OF THE LEKSELL GAMMA KNIFE ICON SYSTEM END-TO-END TESTING OF THE LEKSELL GAMMA KNIFE ICON SYSTEM A Thesis Presented to The Academic Faculty by Krista R. Burton In Partial Fulfillment of the Requirements for the Degree Master of Science in Medical

More information

Optical Engineering 421/521 Sample Questions for Midterm 1

Optical Engineering 421/521 Sample Questions for Midterm 1 Optical Engineering 421/521 Sample Questions for Midterm 1 Short answer 1.) Sketch a pechan prism. Name a possible application of this prism., write the mirror matrix for this prism (or any other common

More information

Aim. Images for this section: Page 2 of 13

Aim. Images for this section: Page 2 of 13 Changes in CT number of high atomic number materials with field of view when using an extended CT number to electron density curve and a metal artifact reduction reconstruction algorithm Poster No.: R-0094

More information

Commissioning. Basic machine performance MLC Dose rate control Gantry speed control End-to-end tests

Commissioning. Basic machine performance MLC Dose rate control Gantry speed control End-to-end tests Acknowledgements David Shepard, Ph.D. Daliang Cao, Ph.D. Muhammad K. N. Afghan, Ph.D. Jinsong Ye, M.S. Tony P. Wong, Ph.D. Fan Chen, Ph.D. Min Rao, Ph.D. Vivek Mehta, M.D. Igor Gomola, Ph.D. David Housley

More information

THE ART OF THE IMAGE: IDENTIFICATION AND REMEDIATION OF IMAGE ARTIFACTS IN MAMMOGRAPHY

THE ART OF THE IMAGE: IDENTIFICATION AND REMEDIATION OF IMAGE ARTIFACTS IN MAMMOGRAPHY THE ART OF THE IMAGE: IDENTIFICATION AND REMEDIATION OF IMAGE ARTIFACTS IN MAMMOGRAPHY William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org INTRODUCTION

More information

UNIT 5a STANDARD ORTHOGRAPHIC VIEW DRAWINGS

UNIT 5a STANDARD ORTHOGRAPHIC VIEW DRAWINGS UNIT 5a STANDARD ORTHOGRAPHIC VIEW DRAWINGS 5.1 Introduction Orthographic views are 2D images of a 3D object obtained by viewing it from different orthogonal directions. Six principal views are possible

More information

Understanding Optical Specifications

Understanding Optical Specifications Understanding Optical Specifications Optics can be found virtually everywhere, from fiber optic couplings to machine vision imaging devices to cutting-edge biometric iris identification systems. Despite

More information

DICOM Correction Proposal

DICOM Correction Proposal Tracking Information - Administration Use Only DICOM Correction Proposal Correction Proposal Number Status CP-1713 Letter Ballot Date of Last Update 2018/01/23 Person Assigned Submitter Name David Clunie

More information