TEP Principles of Digital X-ray Imaging

Size: px
Start display at page:

Download "TEP Principles of Digital X-ray Imaging"

Transcription

1 Related topics exposure time, detector saturation, full well capacity, detector calibration, beam intensity, detector offset, pixel specific gain, attenuation and transmission, image processing. Principle With digital X-ray imaging, X-ray photons that interact with the detector are converted to a digital signal. This permits to record digital radiographies. With this experiment, the principles of digital detectors for X- ray imaging are laid out. Equipment 1 XRE 4.0 X-ray expert set XRCT 4.0 X-ray Computed Tomography upgrade set XR 4.0 Accessories for CT Additional equipment PC, Windows 7 or higher Fig. 1: P P PHYWE Systeme GmbH & Co. KG All rights reserved 1

2 Principles of Digital X-ray Imaging Tasks 1. Define a good exposure time. 2. Calibrate the detector 3. Take some radiographies of an object and process the images with the image viewer Set-up Attach the XRIS to its stage. Place the Digital X-ray detector XRIS on the rail at position 25 cm. The back side of the XRIS stage corresponds to its position on the rail. This position is called the 'source to detector distance' SDD (mm). Fig. 2: Set-up of the XRIS 2 PHYWE Systeme GmbH & Co. KG All rights reserved P

3 Note Details concerning the operation of the X-ray unit and Detector as well as information on how to handle the detector can be found in the respective operating instructions. Procedure - Connect the X-ray unit via USB cable to the USB port of your computer (the correct port of the X-ray unit is marked in Fig. 3). - Connect the usb cable of the detector to the computer Fig. 3: Connection of the computer - Start the measurect program. A virtual X-ray unit, rotation stage and Detector will be displayed on the screen. The green indication LED on the left of each components indicates that its presence has been detected (Fig. 4) - You can change the High Voltage and current of the X-ray tube in the corresponding input windows or manually on the unit. (Fig.4) - When clicking on the unit pictogram additional information concerning the unit can be retrieved( Fig.4) - The status pictogram indicate the status of the unit and can also be used to control the unit such as switching on and off the light or the X- rays (Fig4.) - The position of the digital detector can be adjusted to its real position either by moving the XRIS pictogram or by filling in the correct value in the input window. (Fig.4) - The settings of the XRIS can be adjusted using the input windows. The exposure time controls the time between two frames are retrieved from the detector, the number of frames defines how many frames are averaged and with the binning mode the charge of neighbouring pixels is averaged to reduce the total amount of pixels in one frame. Fig. 4: Part of the user interface of the software P PHYWE Systeme GmbH & Co. KG All rights reserved 3

4 Principles of Digital X-ray Imaging Experiment execution 1. Exposure time and saturation - Adjust the XRIS settings and X-ray unit settings according to fig.5 or load the configuration from the predefined CTO file 'Experiment 1' (see Fig. 5). Overview of the settings of the XRIS and X- ray unit: - 35kV, 1.00mA - exposure time 0.5 sec - Number of frames: 1 - Binning mode 500x500 - SDD= 250 Fig. 5: The settings for this experiment (left panel) and the method load and adjust the settings (right panel) - Start a new experiment, give it a unique name and fill in your details (fig.6). Alternatively it is also possible to load this experiment with pre-recorded images and open this manual. The correct configuration will be loaded automatically as well but the functionalities of the software will be limited to avoid overwriting the existing data. Fig. 6: How to create a new or open an existing experiment 4 PHYWE Systeme GmbH & Co. KG All rights reserved P

5 - Switch on the X-rays and activate the 'Live view' (see fig.7). When the Live view is activated, every new image that is retrieved from the X-ray detector is displayed. The Detector exposure load bar (see fig.7) indicates the average degree of fill for each pixel. It is very important to remain below the maximal fill degree of the detector. Otherwise the detector will be saturated and won't work properly. If the saturation level is reached, the load bar will turn red. (see theory for more explanation). Fig. 7: Activation of the Live View and the detector exposure fill degree Increase and decrease the exposure time of the detector between 0.23 sec and 2 sec while watching the load bar. Note: after the adjustment of the exposure time, leave sufficient time for the detector to physically adjust itself to the new settings. The fill degree of the detector is linearly dependant on the exposure time but is also influenced by the intensity of the beam. This beam intensity can be modified in three ways: 1. Change of maximal kv 2. Change of current 3. Variation in SDD - Change the maximal kv between 10 and 35 kv for a given current and exposure time while watching the Load bar. - Change the current between 0.1 and 1.00 ma for a given kv and exposure time while watching the Load bar - - The increase in fill degree is not linearly proportional to the adjustment of the kv range, see experiment 3 for more explanation. - - The fill degree of the detector is also influenced by the distance of the detector to the source. The fill degree is proportional to the intensity of the beam that falls on the detector which itself is proportional to its opening angle. The intensity is governed by the inverse square law (see experiment 3). - Vary the SDD by manually placing the detector closer or further from the source and watch the load bar: Open the Unit P PHYWE Systeme GmbH & Co. KG All rights reserved 5

6 Principles of Digital X-ray Imaging Change the SDD Switch on the X-rays and wait 10 sec Look at the load bar Note: For all the further experiments it is important to remain under the saturation level, watch the load bar closely after each modification of one of these three parameters. 2. Detector calibration After having defined the desired and adequate configuration for the experiment (fig.5), the detector needs to be calibrated. Each digital detector has a different and variable offset and pixel-specific gain (see theory). During the calibration these variations will be measured and stored. Such corrections are performed automatically by the software: Automatic correction: - Make sure there is no object between the source and the detector, otherwise the calibration will be incorrect. - deactivate the 'Live view' - click on calibrate: Fig. 8: Calibrate the detector - After successful calibration, the red LED in the calibrate button will turn green. From now on the detector is calibrated and the displayed images will be corrected for offset and pixel gain - The Load bar will not be visible anymore while the intensity bar is made visible. With the intensity bar the contrast in the displayed frame can be adjusted. Fig. 9: After calibration, the fill degree bar disappears and the contrast/intensity bar is shown Note: if the configuration is altered, the calibration will be incorrect, this will result in the deactivation of the calibration status. Calibration will have to be performed again. 6 PHYWE Systeme GmbH & Co. KG All rights reserved P

7 3. Applications Once the detector is calibrated, good quality images can be displayed, recorded and saved. - Place one of the demo objects between the source and the detector, switch on the X-rays and activate the live view. To place the object in front of the detector, a stand can be used. Adjust the contrast if desired. Fig. 10: Adjusting the contrast/intensity - The colours of the displayed image can also be altered with three levels (Low, Middle and High). The 'High' colour corresponds with the pixels having a value higher than the upper contrast/intensity handle. The 'Low' colour corresponds with the pixels having a value lower than the lower contrast/intensity handle. The 'Middle' colour corresponds with the pixels having a value between than the upper and lower contrast/intensity handle. Fig. 11: Adjusting the colours - To save the resulting image, stop the live view and click on save. Four formats of images can be saved (tif, png, jpg and bmp). For the three later image formats, the contrast adjustments are retained during saving while this is not the case for the tif format. For tif, the images are saved in a raw format. If the images will be used for further processing or measuring in the measurect software it is important to save them as tif. P PHYWE Systeme GmbH & Co. KG All rights reserved 7

8 Principles of Digital X-ray Imaging - To save a sequence of image, stop the live view and click on record sequence. Four formats of images can be saved (tif, png, jpg and bmp). For the three later image formats, the contrast adjustments are retained during saving while this is not the case for the tif format. For tif, the images are saved in a raw format. If the images will be used for further processing or measuring in the measurect software it is important to save them as tif. Fig. 13: Recording a series of images To further process an image you need to open the image viewer. When the viewer is opened, the image that is present in the frame display will be transferred automatically to the image viewer. The image viewer can be opened in two ways, either from the taskbar or using the shortcut button. When you are finished with the viewer, close it, otherwise the next time you open it the remaining image will still be present. Fig. 14: Open the Image Viewer To process images that were saved as tif, open the destination directory (fig 15). All the tif images that are present in the folder will be displayed in the list. The viewer will typically open while listing the current operational folder and the current operational image for ease of use. Fig. 15: Open a directory with tif- images 8 PHYWE Systeme GmbH & Co. KG All rights reserved P

9 The image viewer has several functions available. The tif files that are present in the current operational directory are listed (fig.16.1) and the current operational image is displayed in the image viewer display (fig.16.2). In this display, the line profile rulers are present (fig.16.3). Additional functionalities can be accessed through the menu (fig.16.4). The histogram of the image and the contrast settings are listed on the right (fig.16.6) as well as the line profile data (fig.16.7) and the region parameters (fig.16.8). Saving and exporting options are available through several buttons (fig.16.9) and the colour of the image levels van be adjusted ((fig.16.10). Fig. 16: The Image Viewer functionalities Using the image buttons (fig.16..5) it is possible: - To select the ruler function (fig.17.1) which allows to change the line profile rulers from position (fig.16.3). - To select several zoom functions (fig.17.2 and fig.17.4) to zoom in on a region of the image - To select the move function (fig.17.3) to change the position of the image once it is zoomed in on. Fig. 17: The image buttons options P PHYWE Systeme GmbH & Co. KG All rights reserved 9

10 Principles of Digital X-ray Imaging Histogram The histogram of the current image is displayed in the histogram display (fig18.1). Such a histogram refers to the number of pixels (Yscale) with a certain grey-value (X-scale) that are present in the image. By adjusting the lower and/or upper level handle (fig18.2), the image contrast can be altered and with the autoscale button(fig18.3), the lower and upper contrast level are reset to optimal conditions. The histogram can be saved as an excel file (fig18.4). Fig. 18: Histogram options Line profile The line profile function calculates the gray-value of each pixel along a line from the centre of one ruler to the centre of the other ruler (fig 16.3). The result is displayed in the line profile display (fig19.1). The line profile can be saved as an excel file (fig19.2). Fig. 19: Line profile options Regions parameters The rulers of the line profile also define a square region. By clicking on the calculate button from the region parameters (fig20), the average grey value and the spread of that region are calculated. Play, Save, AVI and Convert Fig. 20: Region parameters With the play button (fig21.1), the images that are listed in the directory are subsequently displayed, starting from the first selected image. It is possible to save one single image under a different formats or with other colours (fig21.2). A series of images that are selected can be saved as an avi-movie using the AVI-button (fig21.3) and a series of images can be converted to a different file format using the convert button(fig21.4). Fig. 21: Play, save, AVI functions 10 PHYWE Systeme GmbH & Co. KG All rights reserved P

11 Image viewer menu Apart from opening a new directory, the image viewer menu (figxxx.4) has also several other functionalities. - set play speed: changes the speed for playing the images P PHYWE Systeme GmbH & Co. KG All rights reserved 11

12 Principles of Digital X-ray Imaging Theory Detector saturation With digital X-ray imaging, X-ray photons that interact with the detector are converted to a digital signal. Such a digital detector is composed of a raster of pixels (picture elements) and each pixel can be considered as bucket. For each interaction of an X-ray photon with the detector, a series of electrons are produced in the pixel corresponding with the location of the photon interaction. These electrons are stored in the pixel, gradually filling up the bucket. After a set time interval, "exposure time", the electron content of the pixel is measured by emptying it. For the same intensity of X-ray's, a longer exposure time will results in a larger number of pixels in the bucket. Each digital detector has a limited bucket size which is called the 'full well capacity' of the detector. When this level off fill is reached, additional electrons are thrown away because the detector is saturated. A saturated detector will cause inconsistent measurements and has thus to be avoided. Detector calibration Each digital detector has a different and variable offset and pixel-specific output. During the calibration these variations will be measured and used in the subsequent imaging. Even without the X-rays on, the detector will generate a read-out value that is different from 0, called 'dark image'. This has several reasons from which the main reasons are an electronic offset and readout noise. When determining the beam intensity I 0, it is important to subtract this offset (I D ) from the measured read-out (I 0,M ). Another issue that exists, is that the sensitivity of every detector pixel is slightly different, resulting in a wide variation of I 0 values for every pixel. During calibration, these variations are recorded. After calibration, the Transmission value for every pixel (I) is recalculated based on the beam intensity of that pixel at that time (I M ), the beam intensity during calibration(i 0,M ) and the dark current intensity (I D ) during calibration. If the calibration was successfully performed, after calibration the images have grey-values between 0 and PHYWE Systeme GmbH & Co. KG All rights reserved P

13 Transmission and Attenuation In X-ray imaging, the transmission or attenuation of X-rays through a certain object are measured. Depending on the settings of the source, a beam with a certain intensity I 0 is measured by the detector when no object is placed between the source and the detector. When an object is placed in the path of the beam, this object will attenuate the beam so that the detector measures a smaller intensity I instead of I 0. The remaining intensity I compared to the original I 0 is called transmission (T), which is the opposite of the attenuation (A) of the object. For a calibrated detector, the beam intensity is rescaled to a value between 1 and 0. With T= 1 for the beam without an object in front of the detector (I 0 ). P PHYWE Systeme GmbH & Co. KG All rights reserved 13

Principles of CT scan

Principles of CT scan Related topics Detector calibration, saturation, CT acquisition, CT reconstruction Principle X-ray computed tomography consists of using X-rays that are converted to a digital signal by a detector and

More information

TEP Optimization of the CT scan quality. Related topics Number of projections, stop angle, exposure time, image statistics, binning

TEP Optimization of the CT scan quality. Related topics Number of projections, stop angle, exposure time, image statistics, binning Optimization of the CT scan quality TEP Related topics Number of projections, stop angle, exposure time, image statistics, binning Principle The quality of the image will improve with a number of parameters.

More information

Related topics Beam hardening, cupping effect, Beam hardening correction, metal artefacts, photon starvation

Related topics Beam hardening, cupping effect, Beam hardening correction, metal artefacts, photon starvation Beam hardening and metal artefacts TEP Related topics Beam hardening, cupping effect, Beam hardening correction, metal artefacts, photon starvation Principle X-ray sources produce a polychromatic spectrum

More information

Alignment of the camera

Alignment of the camera Related topics Detector Alignment, Rotation axis, tilt, Principle Alignment of the detector and the rotation stage is very important to get optimal quality images of a CT scan. In this experiment, the

More information

X-ray investigation of crystal structures / Laue method with digital X-ray detector (XRIS) (Item No.: P )

X-ray investigation of crystal structures / Laue method with digital X-ray detector (XRIS) (Item No.: P ) X-ray investigation of crystal structures / Laue method with digital X-ray detector (XRIS) (Item No.: P2541602) Curricular Relevance Area of Expertise: Physik Education Level: Hochschule Topic: Moderne

More information

ThermaViz. Operating Manual. The Innovative Two-Wavelength Imaging Pyrometer

ThermaViz. Operating Manual. The Innovative Two-Wavelength Imaging Pyrometer ThermaViz The Innovative Two-Wavelength Imaging Pyrometer Operating Manual The integration of advanced optical diagnostics and intelligent materials processing for temperature measurement and process control.

More information

Digital Portable Overhead Document Camera LV-1010

Digital Portable Overhead Document Camera LV-1010 Digital Portable Overhead Document Camera LV-1010 Instruction Manual 1 Content I Product Introduction 1.1 Product appearance..3 1.2 Main functions and features of the product.3 1.3 Production specifications.4

More information

MIF ZEISS LSM510 CONFOCAL USER PROTOCOL

MIF ZEISS LSM510 CONFOCAL USER PROTOCOL MIF ZEISS LSM510 CONFOCAL USER PROTOCOL START-UP Turn on the Mercury Bulb Power Supply (if needed). Power-on the Control Box. Turn on the computer. Open the LSM 510 software. Choose Scan New Images and

More information

Fig.2: Scanner VistaScan for image plates

Fig.2: Scanner VistaScan for image plates RADIOGRAPHIC INSPECTION OF WELDINGS BY DIGITAL SENSORS H. Thiele, H.-J. Friemel RADIS GmbH, Johanniskirchen, Germany Abstract: The newly available digital sensors for radiographic inspection are suitable

More information

X-RAY COMPUTED TOMOGRAPHY

X-RAY COMPUTED TOMOGRAPHY X-RAY COMPUTED TOMOGRAPHY Bc. Jan Kratochvíla Czech Technical University in Prague Faculty of Nuclear Sciences and Physical Engineering Abstract Computed tomography is a powerful tool for imaging the inner

More information

GUI - DLD Software. Manual

GUI - DLD Software. Manual GUI - DLD Software Manual 2 GUI - DLD Software All rights reserved. No part of this manual may be reproduced without the prior permission of Surface Concept GmbH. Surface Concept GmbH Am Sägewerk 23a 55124

More information

DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter

DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter Revision Date: January 5th, 2017 www.disc-imaging.com Table of Contents Section A: Preliminary Setup Requirements... 4 Tools

More information

CONTENTS. Chapter I Introduction Package Includes Appearance System Requirements... 1

CONTENTS. Chapter I Introduction Package Includes Appearance System Requirements... 1 User Manual CONTENTS Chapter I Introduction... 1 1.1 Package Includes... 1 1.2 Appearance... 1 1.3 System Requirements... 1 1.4 Main Functions and Features... 2 Chapter II System Installation... 3 2.1

More information

User Manual for HoloStudio M4 2.5 with HoloMonitor M4. Phase Holographic Imaging

User Manual for HoloStudio M4 2.5 with HoloMonitor M4. Phase Holographic Imaging User Manual for HoloStudio M4 2.5 with HoloMonitor M4 Phase Holographic Imaging 1 2 HoloStudio M4 2.5 Software instruction manual 2013 Phase Holographic Imaging AB 3 Contact us: Phase Holographic Imaging

More information

See what you need to see, and see it anywhere.

See what you need to see, and see it anywhere. See what you need to see, and see it anywhere. Fujifilm Computed Radiography (FCR) that makes the best use of Fujifilm s unique image processing technology has become even more convenient. In addition

More information

ROTATING SYSTEM T-12, T-20, T-50, T- 150 USER MANUAL

ROTATING SYSTEM T-12, T-20, T-50, T- 150 USER MANUAL ROTATING SYSTEM T-12, T-20, T-50, T- 150 USER MANUAL v. 1.11 released 12.02.2016 Table of contents Introduction to the Rotating System device 3 Device components 4 Technical characteristics 4 Compatibility

More information

Cropping And Sizing Information

Cropping And Sizing Information and General The procedures and techniques described herein are intended to provide a means of modifying digital images for use in projection situations. This includes images being displayed on a screen

More information

Zeiss 780 Training Notes

Zeiss 780 Training Notes Zeiss 780 Training Notes Turn on Main Switch, System PC and Components Switches 780 Start up sequence Do you need the argon laser (458, 488, 514 nm lines)? Yes Turn on the laser s main power switch and

More information

1. What is SENSE Batch

1. What is SENSE Batch 1. What is SENSE Batch 1.1. Introduction SENSE Batch is processing software for thermal images and sequences. It is a modern software which automates repetitive tasks with thermal images. The most important

More information

GXCapture 8.1 Instruction Manual

GXCapture 8.1 Instruction Manual GT Vision image acquisition, managing and processing software GXCapture 8.1 Instruction Manual Contents of the Instruction Manual GXC is the shortened name used for GXCapture Square brackets are used to

More information

User Manual for PROGRES GRYPHAX software

User Manual for PROGRES GRYPHAX software User Manual for PROGRES GRYPHAX software PROGRES GRYPHAX software JENOPTIK l Healthcare & Industry JENOPTIK Optical Systems GmbH Göschwitzer Straße 25 07745 Jena, Germany Dear user, Please read the instructions

More information

Leica DMi8A Quick Guide

Leica DMi8A Quick Guide Leica DMi8A Quick Guide 1 Optical Microscope Quick Start Guide The following instructions are provided as a Quick Start Guide for powering up, running measurements, and shutting down Leica s DMi8A Inverted

More information

User Guide for TWAIN / DirectX interface for GRYPHAX USB 3.0 cameras

User Guide for TWAIN / DirectX interface for GRYPHAX USB 3.0 cameras User Guide for TWAIN / DirectX interface for GRYPHAX USB 3.0 cameras The TWAIN & DirectX driver for PROGRES GRYPHAX USB 3.0 cameras enables user to operate with TWAIN and DirectX supported 3 rd party software

More information

GlobiScope Analysis Software for the Globisens QX7 Digital Microscope. Quick Start Guide

GlobiScope Analysis Software for the Globisens QX7 Digital Microscope. Quick Start Guide GlobiScope Analysis Software for the Globisens QX7 Digital Microscope Quick Start Guide Contents GlobiScope Overview... 1 Overview of home screen... 2 General Settings... 2 Measurements... 3 Movie capture...

More information

Digital Dental Quality Assurance Manual Initial Baseline and Longitudinal Assessment

Digital Dental Quality Assurance Manual Initial Baseline and Longitudinal Assessment Digital Dental Quality Assurance Manual Initial Baseline and Longitudinal Assessment Initial Baseline Assessment 1. Place the detector to be evaluated directly under the central portion of the Digital

More information

Wound Healing Analysis in the Automated Cellular Analysis System

Wound Healing Analysis in the Automated Cellular Analysis System Wound Healing Analysis in the Automated Cellular Analysis System 1 ibidi GmbH, Version 1.0, 2017-07-14 Table of Content Specifications... 3 Step-by-Step Guide... 4 Analysis... 6 Example Report Job... 8

More information

Image Processing Tutorial Basic Concepts

Image Processing Tutorial Basic Concepts Image Processing Tutorial Basic Concepts CCDWare Publishing http://www.ccdware.com 2005 CCDWare Publishing Table of Contents Introduction... 3 Starting CCDStack... 4 Creating Calibration Frames... 5 Create

More information

Handbook. Antenna Rotator Controller

Handbook. Antenna Rotator Controller Handbook Controller 3S design GmbH 2. Südwieke 231 26817 Rhauderfehn Germany www.3sdesign.de Tel. +49-4952-921950 Contents I Table of content Part I 1 1 Operating... 1 2 Installation... 2 1 1 1.1 Operating

More information

We recommend downloading the latest core installer for our software from our website. This can be found at:

We recommend downloading the latest core installer for our software from our website. This can be found at: Dusk Getting Started Installing the Software We recommend downloading the latest core installer for our software from our website. This can be found at: https://www.atik-cameras.com/downloads/ Locate and

More information

The ideal K-12 science microscope solution. User Guide. for use with the Nova5000

The ideal K-12 science microscope solution. User Guide. for use with the Nova5000 The ideal K-12 science microscope solution User Guide for use with the Nova5000 NovaScope User Guide Information in this document is subject to change without notice. 2009 Fourier Systems Ltd. All rights

More information

Machinery HDR Effects 3

Machinery HDR Effects 3 1 Machinery HDR Effects 3 MACHINERY HDR is a photo editor that utilizes HDR technology. You do not need to be an expert to achieve dazzling effects even from a single image saved in JPG format! MACHINERY

More information

Appendix A ACE exam objectives map

Appendix A ACE exam objectives map A 1 Appendix A ACE exam objectives map This appendix covers these additional topics: A ACE exam objectives for Photoshop CS6, with references to corresponding coverage in ILT Series courseware. A 2 Photoshop

More information

CCD User s Guide SBIG ST7E CCD camera and Macintosh ibook control computer with Meade flip mirror assembly mounted on LX200

CCD User s Guide SBIG ST7E CCD camera and Macintosh ibook control computer with Meade flip mirror assembly mounted on LX200 Massachusetts Institute of Technology Department of Earth, Atmospheric, and Planetary Sciences Handout 8 /week of 2002 March 18 12.409 Hands-On Astronomy, Spring 2002 CCD User s Guide SBIG ST7E CCD camera

More information

Spatial intensity distribution analysis Matlab user guide

Spatial intensity distribution analysis Matlab user guide Spatial intensity distribution analysis Matlab user guide August 2011 Guide on how to use the SpIDA graphical user interface. This little tutorial provides a step by step tutorial explaining how to get

More information

CHAPTER1: QUICK START...3 CAMERA INSTALLATION... 3 SOFTWARE AND DRIVER INSTALLATION... 3 START TCAPTURE...4 TCAPTURE PARAMETER SETTINGS... 5 CHAPTER2:

CHAPTER1: QUICK START...3 CAMERA INSTALLATION... 3 SOFTWARE AND DRIVER INSTALLATION... 3 START TCAPTURE...4 TCAPTURE PARAMETER SETTINGS... 5 CHAPTER2: Image acquisition, managing and processing software TCapture Instruction Manual Key to the Instruction Manual TC is shortened name used for TCapture. Help Refer to [Help] >> [About TCapture] menu for software

More information

User Manual. Copyright 2010 Lumos. All rights reserved

User Manual. Copyright 2010 Lumos. All rights reserved User Manual The contents of this document may not be copied nor duplicated in any form, in whole or in part, without prior written consent from Lumos. Lumos makes no warranties as to the accuracy of the

More information

NanEye in Awaiba Viewer

NanEye in Awaiba Viewer NanEye in Awaiba Viewer Table of Contents 1 Introduction...3 2 NanEye in Awaiba Viewer...4 2.1 NanEye Sensor control...5 2.1.1 Manual Control Tab...5 2.1.2 Supply Voltage...6 2.1.3 Automatic Control...6

More information

An Introduction to Histograms in Photography

An Introduction to Histograms in Photography An Introduction to Histograms in Photography Histograms are a graphical representation of all the pixels that make up an image, and are plotted by 'Luminance' or brightness. Every pixel, regardless of

More information

Tribometrics. Version 2.11

Tribometrics. Version 2.11 Tribometrics Version 2.11 Table of Contents Tribometrics... 1 Version 2.11... 1 1. About This Document... 4 1.1. Conventions... 4 2. Introduction... 5 2.1. Software Features... 5 2.2. Tribometrics Overview...

More information

Optika ISview. Image acquisition and processing software. Instruction Manual

Optika ISview. Image acquisition and processing software. Instruction Manual Optika ISview Image acquisition and processing software Instruction Manual Key to the Instruction Manual IS is shortened name used for OptikaISview Square brackets are used to indicate items such as menu

More information

Full-screen mode Popup controls. Overview of the microscope user interface, TEM User Interface and TIA on the left and EDS on the right

Full-screen mode Popup controls. Overview of the microscope user interface, TEM User Interface and TIA on the left and EDS on the right Quick Guide to Operating FEI Titan Themis G2 200 (S)TEM: TEM mode Susheng Tan Nanoscale Fabrication and Characterization Facility, University of Pittsburgh Office: M104/B01 Benedum Hall, 412-383-5978,

More information

ISCapture User Guide. advanced CCD imaging. Opticstar

ISCapture User Guide. advanced CCD imaging. Opticstar advanced CCD imaging Opticstar I We always check the accuracy of the information in our promotional material. However, due to the continuous process of product development and improvement it is possible

More information

CR Basics and FAQ. Overview. Historical Perspective

CR Basics and FAQ. Overview. Historical Perspective Page: 1 of 6 CR Basics and FAQ Overview Computed Radiography is a term used to describe a system that electronically records a radiographic image. Computed Radiographic systems use unique image receptors

More information

ToupSky Cameras Quick-guide

ToupSky Cameras Quick-guide ToupSky Cameras Quick-guide ToupSky is a capture and processing software offered by Touptek, the original manufacturer of the Toupcamera series. These are video cameras that offer live image capture for

More information

Using Adobe Photoshop to enhance the image quality. Assistant course web site:

Using Adobe Photoshop to enhance the image quality. Assistant course web site: Using Adobe Photoshop to enhance the image quality Assistant course web site: http://www.arches.uga.edu/~skwang/edit6170/course.htm Content Introduction 2 Unit1: Scan images 3 Lesson 1-1: Preparations

More information

Applying mathematics to digital image processing using a spreadsheet

Applying mathematics to digital image processing using a spreadsheet Jeff Waldock Applying mathematics to digital image processing using a spreadsheet Jeff Waldock Department of Engineering and Mathematics Sheffield Hallam University j.waldock@shu.ac.uk Introduction When

More information

AgilEye Manual Version 2.0 February 28, 2007

AgilEye Manual Version 2.0 February 28, 2007 AgilEye Manual Version 2.0 February 28, 2007 1717 Louisiana NE Suite 202 Albuquerque, NM 87110 (505) 268-4742 support@agiloptics.com 2 (505) 268-4742 v. 2.0 February 07, 2007 3 Introduction AgilEye Wavefront

More information

Gentec-EO USA. T-RAD-USB Users Manual. T-Rad-USB Operating Instructions /15/2010 Page 1 of 24

Gentec-EO USA. T-RAD-USB Users Manual. T-Rad-USB Operating Instructions /15/2010 Page 1 of 24 Gentec-EO USA T-RAD-USB Users Manual Gentec-EO USA 5825 Jean Road Center Lake Oswego, Oregon, 97035 503-697-1870 voice 503-697-0633 fax 121-201795 11/15/2010 Page 1 of 24 System Overview Welcome to the

More information

MC3 Motion Control System Shutter Stream Quickstart

MC3 Motion Control System Shutter Stream Quickstart MC3 Motion Control System Shutter Stream Quickstart Revised 7/6/2016 Carousel USA 6370 N. Irwindale Rd. Irwindale, CA 91702 www.carousel-usa.com Proprietary Information Carousel USA has proprietary rights

More information

LACERTA M-GEN Stand-Alone AutoGuider

LACERTA M-GEN Stand-Alone AutoGuider LACERTA M-GEN Stand-Alone AutoGuider Changes from Firmware 01.22 to 01.99 (pre-release of FW 02.00) Created by: Zoltán Tobler 13 February 2011 1 New features Hardware binning operating modes: Binning mode

More information

ECOVIEW 9 / ECOVIEW 9 PLUS Digital Radiographic System

ECOVIEW 9 / ECOVIEW 9 PLUS Digital Radiographic System Co., LTD. 3F, Urbanlight B/D, 630, Eonju-ro, Gangnam-gu, Seoul, Korea 135-832 Factory at 621-14, Dochun-dong, Gwangsan-gu, Gwangju, Korea 506-301 TEL. : +82-70-7510-3400 FAX. : +82-70-8630-3420 sales@ecoray.kr

More information

September CoroCAM 6D. Camera Operation Training. Copyright 2012

September CoroCAM 6D. Camera Operation Training. Copyright 2012 CoroCAM 6D Camera Operation Training September 2012 CoroCAM 6D Body Rubber cover on SD Card slot & USB port Lens Cap retention loop Charging port, video & audio output, audio input Laser pointer CoroCAM

More information

LIGHT STUDIO 485. Software Instrrctions Mancal. Read trhe mancal tarefclly before operaing trhis software

LIGHT STUDIO 485. Software Instrrctions Mancal. Read trhe mancal tarefclly before operaing trhis software LIGHT STUDIO 485 Software Instrrctions Mancal Read trhe mancal tarefclly before operaing trhis software PSI, spol. s r. o., Drásov 470, 664 4 Drásov, Czeth Repcblit FAX: +40 5 440 90, TEL: +40 5 440 0,

More information

4.5.1 Mirroring Gain/Offset Registers GPIO CMV Snapshot Control... 14

4.5.1 Mirroring Gain/Offset Registers GPIO CMV Snapshot Control... 14 Thank you for choosing the MityCAM-C8000 from Critical Link. The MityCAM-C8000 MityViewer Quick Start Guide will guide you through the software installation process and the steps to acquire your first

More information

Fly Elise-ng Grasstrook HG Eindhoven The Netherlands Web: elise-ng.net Tel: +31 (0)

Fly Elise-ng Grasstrook HG Eindhoven The Netherlands Web:  elise-ng.net Tel: +31 (0) Fly Elise-ng Grasstrook 24 5658HG Eindhoven The Netherlands Web: http://fly.elise-ng.net Email: info@elise elise-ng.net Tel: +31 (0)40 7114293 Fly Elise-ng Immersive Calibration PRO Step-By Single Camera

More information

Motic Live Imaging Module. Windows OS User Manual

Motic Live Imaging Module. Windows OS User Manual Motic Live Imaging Module Windows OS User Manual Motic Live Imaging Module Windows OS User Manual CONTENTS (Linked) Introduction 05 Menus, bars and tools 06 Title bar 06 Menu bar 06 Status bar 07 FPS 07

More information

Visioneer OneTouch Scanner. Installation Guide FOR WINDOWS

Visioneer OneTouch Scanner. Installation Guide FOR WINDOWS Visioneer OneTouch Scanner Installation Guide FOR WINDOWS TABLE OF CONTENTS i TABLE OF CONTENTS Getting Started with your new Scanner....................... 1 Step 1: Installing the Scanner Software.......................

More information

Information & Instructions

Information & Instructions KEY FEATURES 1. USB 3.0 For the Fastest Transfer Rates Up to 10X faster than regular USB 2.0 connections (also USB 2.0 compatible) 2. High Resolution 4.2 MegaPixels resolution gives accurate profile measurements

More information

For customers in USA This device complies with Part 15 of the FCC rules. Operation is subject to the following two conditions:

For customers in USA This device complies with Part 15 of the FCC rules. Operation is subject to the following two conditions: User manual For customers in North and South America For customers in USA This device complies with Part 15 of the FCC rules. Operation is subject to the following two conditions: (1) This device may not

More information

Digital Radiographic Inspection replacing traditional RT and 3D RT Development

Digital Radiographic Inspection replacing traditional RT and 3D RT Development Digital Radiographic Inspection replacing traditional RT and 3D RT Development Iploca Novel Construction Meeting 27&28 March 2014 Geneva By Jan van der Ent Technical Authority International Contents Introduction

More information

ECOVIEW 9 / ECOVIEW 9 PLUS Digital Radiographic System

ECOVIEW 9 / ECOVIEW 9 PLUS Digital Radiographic System ECORAY Co., LTD. 3F, Urbanlight B/D, 630, Eonju-ro, Gangnam-gu, Seoul, Korea 135-832 Factory at 621-14, Dochun-dong, Gwangsan-gu, Gwangju, Korea 506-301 TEL. : +82-70-7510-3400 FAX. : +82-70-8630-3420

More information

BEAMAGE-3.0 KEY FEATURES BEAM DIAGNOSTICS AVAILABLE MODELS MAIN FUNCTIONS SEE ALSO ACCESSORIES. CMOS Beam Profiling Cameras

BEAMAGE-3.0 KEY FEATURES BEAM DIAGNOSTICS AVAILABLE MODELS MAIN FUNCTIONS SEE ALSO ACCESSORIES. CMOS Beam Profiling Cameras BEAM DIAGNOSTICS BEAM DIAGNOSTICS SPECIAL PRODUCTS OEM DETECTORS THZ DETECTORS PHOTO DETECTORS HIGH POWER DETECTORS POWER DETECTORS ENERGY DETECTORS MONITORS CMOS Beam Profiling Cameras AVAILABLE MODELS

More information

Magnetic induction with Cobra3

Magnetic induction with Cobra3 Principle A magnetic field of variable frequency and varying strength is produced in a long coil. The voltages induced across thin coils which are pushed into the long coil are determined as a function

More information

Tube Formation Analysis in the Automated Cellular Analysis System

Tube Formation Analysis in the Automated Cellular Analysis System Tube Formation Analysis in the Automated Cellular Analysis System 1 ibidi GmbH, Version 1.0, 2017-07-14 Table of Content Specifications... 3 Step-by-Step Guide... 4 Analysis... 6 Example Report Job...

More information

!! Select Professional Mode. !! Click on Other then Select Scratch.

!! Select Professional Mode. !! Click on Other then Select Scratch. SCANNING BASICS FOR EPSON SCANNERS When using EpsonScan software with any of the Epson flatbed scanners in Labs 1, 2, 3 or 4, the following steps must be followed in order to properly save files. Select

More information

ERS KEY FEATURES BEAM DIAGNOSTICS MAIN FUNCTIONS AVAILABLE MODEL. CMOS Beam Profiling Camera. 1 USB 3.0 for the Fastest Transfer Rates

ERS KEY FEATURES BEAM DIAGNOSTICS MAIN FUNCTIONS AVAILABLE MODEL. CMOS Beam Profiling Camera. 1 USB 3.0 for the Fastest Transfer Rates POWER DETECTORS ENERGY DETECTORS MONITORS SPECIAL PRODUCTS OEM DETECTORS THZ DETECTORS PHOTO DETECTORS HIGH POWER DETECTORS CAMERA PROFIL- CMOS Beam Profiling Camera KEY FEATURES ERS 1 USB 3.0 for the

More information

Manual. Cell Border Tracker. Jochen Seebach Institut für Anatomie und Vaskuläre Biologie, WWU Münster

Manual. Cell Border Tracker. Jochen Seebach Institut für Anatomie und Vaskuläre Biologie, WWU Münster Manual Cell Border Tracker Jochen Seebach Institut für Anatomie und Vaskuläre Biologie, WWU Münster 1 Cell Border Tracker 1. System Requirements The software requires Windows XP operating system or higher

More information

Planmeca Romexis. quick guide. Viewer EN _2

Planmeca Romexis. quick guide. Viewer EN _2 Planmeca Romexis Viewer quick guide EN 10029550_2 TABLE OF CONTENTS 1 START-UP OF PLANMECA ROMEXIS VIEWER...1 1.1 Selecting the interface language... 1 1.2 Selecting images...1 1.3 Starting the Planmeca

More information

Basic Tutorials Series: Import A Photograph. RenoWorks Support Team Document #HWPRO0003

Basic Tutorials Series: Import A Photograph. RenoWorks Support Team Document #HWPRO0003 Basic Tutorials Series: Import A Photograph RenoWorks Support Team Document #HWPRO0003 Import A Photograph 2 1 Import Your Own Photograph The Photo Import Wizard The Photo Import Wizard is the first tool

More information

Version 1.0. TechnicVR. Student Guide

Version 1.0. TechnicVR. Student Guide Version 1.0 TechnicVR s h a d e r w a r e. c o m Student Guide TechnicVR s h a d e r w a r e. c o m Student Guide shaderware 2008 PO Box 103 Saltburn Cleveland TS12 1WP w w w. s h a d e r w a r e. c o

More information

PANalytical X pert Pro Gazing Incidence X-ray Reflectivity User Manual (Version: )

PANalytical X pert Pro Gazing Incidence X-ray Reflectivity User Manual (Version: ) University of Minnesota College of Science and Engineering Characterization Facility PANalytical X pert Pro Gazing Incidence X-ray Reflectivity User Manual (Version: 2012.10.17) The following instructions

More information

Table of Contents 1. Image processing Measurements System Tools...10

Table of Contents 1. Image processing Measurements System Tools...10 Introduction Table of Contents 1 An Overview of ScopeImage Advanced...2 Features:...2 Function introduction...3 1. Image processing...3 1.1 Image Import and Export...3 1.1.1 Open image file...3 1.1.2 Import

More information

DigiScope II v3 TM Aperture Scope User s Manual

DigiScope II v3 TM Aperture Scope User s Manual DigiScope II v3 TM Aperture Scope User s Manual Welcome Thank you for choosing DigiScope II v3 TM Aperture scope! The DigiScope II v3 TM Aperture Scope is an exciting new device to Capture and record the

More information

Instruction manual for T3DS software. Tool for THz Time-Domain Spectroscopy. Release 4.0

Instruction manual for T3DS software. Tool for THz Time-Domain Spectroscopy. Release 4.0 Instruction manual for T3DS software Release 4.0 Table of contents 0. Setup... 3 1. Start-up... 5 2. Input parameters and delay line control... 6 3. Slow scan measurement... 8 4. Fast scan measurement...

More information

WALLY ROTARY ENCODER. USER MANUAL v. 1.0

WALLY ROTARY ENCODER. USER MANUAL v. 1.0 WALLY ROTARY ENCODER USER MANUAL v. 1.0 1.MEASUREMENTS ANGULAR POSITIONING a. General Description The angular positioning measurements are performed with the use of the Wally rotary encoder. This measurement

More information

PHOTOSHOP. Introduction to Adobe Photoshop

PHOTOSHOP. Introduction to Adobe Photoshop PHOTOSHOP You will; 1. Learn about some of Photoshop s Tools. 2. Learn how Layers work. 3. Learn how the Auto Adjustments in Photoshop work. 4. Learn how to adjust Colours. 5. Learn how to measure Colours.

More information

Operating Procedures for MICROCT1 Nikon XTH 225 ST

Operating Procedures for MICROCT1 Nikon XTH 225 ST Operating Procedures for MICROCT1 Nikon XTH 225 ST Ensuring System is Ready (go through to ensure all windows and tasks below have been completed either by you or someone else prior to mounting and scanning

More information

Cutting out in GIMP. Navigation click to go to a section

Cutting out in GIMP. Navigation click to go to a section Cutting out in GIMP Navigation click to go to a section Before you start Cutting out an element Using several elements Using backgrounds with cut out elements Adjusting the colour of elements Creating

More information

ISIS A beginner s guide

ISIS A beginner s guide ISIS A beginner s guide Conceived of and written by Christian Buil, ISIS is a powerful astronomical spectral processing application that can appear daunting to first time users. While designed as a comprehensive

More information

BEAMAGE KEY FEATURES AVAILABLE MODELS. CMOS Beam Profiling Cameras

BEAMAGE KEY FEATURES AVAILABLE MODELS. CMOS Beam Profiling Cameras BEAM DIAGNOS TICS Beam Profiling Cameras KEY FEATURES SPECIAL PRODUCTS OEM DETECTORS THZ DETECTORS PHOTO DETECTORS HIGH POWER SOLUTIONS POWER DETECTORS ENERGY DETECTORS MONITORS AVAILABLE MODELS Beamage-3.0

More information

3DExplorer Quickstart. Introduction Requirements Getting Started... 4

3DExplorer Quickstart. Introduction Requirements Getting Started... 4 Page 1 of 43 Table of Contents Introduction... 2 Requirements... 3 Getting Started... 4 The 3DExplorer User Interface... 6 Description of the GUI Panes... 6 Description of the 3D Explorer Headbar... 7

More information

1 ImageBrowser Software User Guide 5.1

1 ImageBrowser Software User Guide 5.1 1 ImageBrowser Software User Guide 5.1 Table of Contents (1/2) Chapter 1 What is ImageBrowser? Chapter 2 What Can ImageBrowser Do?... 5 Guide to the ImageBrowser Windows... 6 Downloading and Printing Images

More information

The Spot Colors module in ZePrA 3.5

The Spot Colors module in ZePrA 3.5 The Spot Colors module in ZePrA 3.5 A new module for high-quality conversion of spot colors to the target color space has been integrated in Version 3.5 of our ZePrA color server. The module is chargeable

More information

Lettering Fabric Preparation deco 340 aurora 430E & 440QEE NAME artista 630E, 635LE & 640E

Lettering Fabric Preparation deco 340 aurora 430E & 440QEE NAME artista 630E, 635LE & 640E Lettering Fabric Preparation Deco 340 Hoop a piece of firm woven fabric backed with tear-away stabilizer in the Large hoop. aurora, artista, and 8 Series Hoop a piece of firm woven fabric backed with tear-away

More information

Magnetic induction with Cobra3

Magnetic induction with Cobra3 Magnetic induction with Cobra3 LEP Related Topics Maxwell s equations, electrical eddy field, magnetic field of coils, coil, magnetic flux, induced voltage. Principle A magnetic field of variable frequency

More information

Imaging Beyond the Basics: Optimizing Settings on the Leica SP8 Confocal

Imaging Beyond the Basics: Optimizing Settings on the Leica SP8 Confocal Imaging Beyond the Basics: Optimizing Settings on the Leica SP8 Confocal Todays Goal: Introduce some additional functionalities of the Leica SP8 confocal HyD vs. PMT detectors Dye Assistant Scanning By

More information

Contents Technical background II. RUMBA technical specifications III. Hardware connection IV. Set-up of the instrument Laboratory set-up

Contents Technical background II. RUMBA technical specifications III. Hardware connection IV. Set-up of the instrument Laboratory set-up RUMBA User Manual Contents I. Technical background... 3 II. RUMBA technical specifications... 3 III. Hardware connection... 3 IV. Set-up of the instrument... 4 1. Laboratory set-up... 4 2. In-vivo set-up...

More information

CT parameter studies for porous metal samples. Sören R. Lindemann Daimler AG Werk Untertürkheim

CT parameter studies for porous metal samples. Sören R. Lindemann Daimler AG Werk Untertürkheim CT parameter studies for porous metal samples Sören R. Lindemann Daimler AG Werk Untertürkheim Where do we stand and what are we looking for? small material samples (high absorption coefficient, low porosity)

More information

Contents Chapter One- Introduction

Contents Chapter One- Introduction Contents Chapter One- Introduction... 1 1.1 Applications supported... 1 1.2 Hardware... 1 1.2.1 Specifications... 1 1.2.2 System Components... 2 1.2.2.1 Darkroom... 2 1.2.2.2 UV transilluminator... 1.2.2.

More information

KoPa Scanner. User's Manual A99. Ver 1.0. SHENZHEN OSTEC OPTO-ELECTRONIC TECHNOLOGY CO.,LTD.

KoPa Scanner. User's Manual A99. Ver 1.0. SHENZHEN OSTEC OPTO-ELECTRONIC TECHNOLOGY CO.,LTD. KoPa Scanner A99 User's Manual Ver 1.0 SHENZHEN OSTEC OPTO-ELECTRONIC TECHNOLOGY CO.,LTD. http://www.ostec.com.cn Content Chapter 1 Start... 1 1.1 Safety Warnings and Precautions... 1 1.2 Installation

More information

ARTRAY Camera / Converter Viewer Software. ART-VIEWER v1370. Manual

ARTRAY Camera / Converter Viewer Software. ART-VIEWER v1370. Manual ARTRAY Camera / Converter Viewer Software ART-VIEWER v1370 Manual Table of Contents TABLE OF CONTENTS... 2 1. SUMMARY... 4 2. INSTRUCTIONS FOR CAMERAS... 5 2.1. USB2.0 CAMERA ARTCNVIIDEVICE DRIVER INSTALLATION...

More information

The Basics of Focus Stacking. by Michael K. Miller

The Basics of Focus Stacking. by Michael K. Miller The Basics of Focus Stacking by Michael K. Miller Introduction Focus (or image) stacking is a method to increase the depth of field (DOF) by combining a series of images taken at either different focus

More information

REMEMBER: You have 5GB of disk space on this microscope. Check before you start if you have room for your experiment. If not delete your old data.

REMEMBER: You have 5GB of disk space on this microscope. Check before you start if you have room for your experiment. If not delete your old data. 1 Use of the Zeiss LSM 510 Inverted Firstly please be aware that this microscope should be treated with respect and care at all times. Rules of use: This Microscope can only be used by Masters by Research

More information

User Manual. 3 MegaPixel CMOS Color Camera for Light Microscopy SC30

User Manual. 3 MegaPixel CMOS Color Camera for Light Microscopy SC30 User Manual 3 MegaPixel CMOS Color Camera for Light Microscopy SC30 Any copyrights relating to this manual shall belong to Olympus Soft Imaging Solutions GmbH. We at Olymnpus Soft Imaging Solutions GmbH

More information

First English edition for Ulead COOL 360 version 1.0, February 1999.

First English edition for Ulead COOL 360 version 1.0, February 1999. First English edition for Ulead COOL 360 version 1.0, February 1999. 1992-1999 Ulead Systems, Inc. All rights reserved. No part of this publication may be reproduced or transmitted in any form or by any

More information

KM-4800w. Copy/Scan Operation Manual

KM-4800w. Copy/Scan Operation Manual KM-4800w Copy/Scan Operation Manual NOTE: This Operation Manual contains information that corresponds to using both the metric and inch versions of these machines. The metric versions of these machines

More information

ALPHASHOT 360 AUTOMATED PHOTO STUDIO FOR SMALL-SIZED PRODUCTS CUT COSTS INCREASE SALES SPEED UP WORKFLOW

ALPHASHOT 360 AUTOMATED PHOTO STUDIO FOR SMALL-SIZED PRODUCTS CUT COSTS INCREASE SALES SPEED UP WORKFLOW ALPHASHOT 360 AUTOMATED PHOTO STUDIO FOR SMALL-SIZED PRODUCTS CUT COSTS INCREASE SALES SPEED UP WORKFLOW Put your products in the spotlight! IN-HOUSE AUTOMATED PHOTO STUDIO FOR SMALL-SIZED PRODUCTS VIEW

More information

Demetra User documentation DC0029A

Demetra User documentation DC0029A Demetra User documentation DC0029A Rev Date Who What 1.0 25/09/2017 F. Cochard First version (corrections from T. Rodda) Table of content 1 2 3 4 5 Introduction...2 Requirements...3 Installation...3 General

More information

Micro-Image Capture 8 Installation Instructions & User Guide

Micro-Image Capture 8 Installation Instructions & User Guide Micro-Image Capture 8 Installation Instructions & User Guide Software installation: Micro-Image Capture Software 1. Load Micro-Image Capture software CD onto host PC. Auto Run should start driver/software

More information

REV.C. SmartScope 5M

REV.C. SmartScope 5M REV.C SmartScope 5M The 200X digital microscope provides a 10~200X adjustable magnification range, and the 500X microscope provides 500X magnification. The built-in high-performance LEDs can illuminate

More information