Diabetic Retinopathy Clinical Research Network (DRCR.net) UWF Optos 200Tx Imaging Protocol. Version 3.0 9/19/16

Size: px
Start display at page:

Download "Diabetic Retinopathy Clinical Research Network (DRCR.net) UWF Optos 200Tx Imaging Protocol. Version 3.0 9/19/16"

Transcription

1 Diabetic Retinopathy Clinical Research Network (DRCR.net) UWF Optos 200Tx Imaging Protocol Version 3.0 9/19/16 DRCR.net UWF 200 Tx Imaging Protocol V Final Page 1 of 14

2 Table of Contents Background... 3 P200Tx Device... 3 Imaging Protocol... 4 How will the Study Eye be determined?... 5 Image Capture Sequence... 5 Certification Procedures... 8 Image Transfer Procedures... 8 Image Retakes... 9 Examples of Well Captured images... 9 Contact Information: Cleaning the mirror Process Description DRCR.net UWF 200 Tx Imaging Protocol V Final Page 2 of 14

3 Version Author Effective Date Revision Description 2.0 Adam Glassman 02/25/2015 Initial Implemented Version 3.0 Adam Glassman 09/19/2016 Removed nasal and temporal steered images from FA procedure Background This imaging protocol will provide you detailed information regarding imaging your diabetic retinopathy study using the Optos system, sending images to the Coordinating Center, image storage and system maintenance. P200Tx Device The Optos P200Tx system is a scanning laser ophthalmoscopes (SLO) containing an ellipsoidal mirror; that scans through one focal point with a SLO, a virtual scanner is created inside the eye allowing majority of retina (82%) including the retinal periphery to be imaged. This results in a 200 image that can be captured through a non-dilated pupil with a resolution of 3000 X 3000 pixels with its smallest resolvable feature at 14 microns. Images can be captured in color composite mode, which uses a red laser (633nm) and a green laser (532nm) to visualize the choroid and the RPE respectively, fluorescein angiography (FA) and AF (FAF) images can also be acquired. FAF images are captured using the green laser wavelength and are not averaged like other systems. The blue laser is used when capturing angiographic images. A series of images is captured as the fluorescein flows through the retinal vessels. Optos technology is designed to operate through a minimum pupil diameter of 2mm. The composite image supports analysis of the retinal image by allowing individual review of the green and red channels information. The Green Channel image contains information from the sensory retina through the pigment epithelium layers of the retina. The Red Channel image contains information from the deeper structures of the retina, from the pigment epithelium through the choroid. DRCR.net UWF 200 Tx Imaging Protocol V Final Page 3 of 14

4 Imaging Protocol Following informed consent, each participant will have the following images captured: Color Optomap Plus Images (12 total images - 6 images per eye): 2200 color central fixation Optomap plus images 1 superior steered 200 color Optomap plus image 1 inferior steered 200 color Optomap plus image 1 nasal steered 200 color Optomap plus image* 1 temporal steered 200 color Optomap plus image* * only at baseline visit FA Optomap Images (12 images: 6 per eye*) Early phase (starting at 15 sec) central fixation images study eye central fixation images fellow eye Late phase (starting at 4 min) DRCR.net UWF 200 Tx Imaging Protocol V Final Page 4 of 14

5 1 200 central fixation image study eye superior steered image study eye inferior steered image study eye followed by central fixation image fellow eye superior steered image fellow eye inferior steered image fellow eye A total of 12 images per subject will be saved and uploaded to Reading Center. *If there is only one study eye, this eye should always be imaged first. If both eyes are study eyes, the right eye is photographed first and then the left eye is imaged. Each digital image must be evaluated to be of adequate quality for submission, according to the study procedures. If photograph quality is judged substandard by the operator, then the imaging should be repeated until a good quality image is obtained. How will the Study Eye be determined? The study eye will be determined by eligibility criteria. A patient may have 2 study eyes. In this case, the right eye should be imaged first and the left eye imaged next Image Capture Sequence 1. Select the Patient List button. 2. Select the New Patient button. 3. Enter in subject information as follows to mask patient information: a) Last Name = STUDY PROTOCOL b) First Name = Participant Study #, Name Code (e.g , ABC) c) ID = Participant Study # (e.g ) d) Date of Birth = 01/01/1899 All images MUST be acquired using the masked subject information as detailed above. Images should NOT be acquired under MRN# or patient name. 4. Ensure the patient s name is highlighted and then choose OK. 5. Select Optomap Plus and then choose OK. If this patient will also be receiving FA imaging as a part of this protocol select DRCR.net UWF 200 Tx Imaging Protocol V Final Page 5 of 14

6 Optomap FA this will start the session with color images 6. Select the Study Eye for the first image. For each subsequent image you will need to select which eye you will image. 7. Position participant with their feet flat on the floor and facing the device 8. Adjust the table height using the bottom button on the hand control. Then align the canthus with the marker on the chinrest using the chinrest adjustment knob. 9. Ensure the chinrest is on the left side to image the Right Eye or the right side to image the Left Eye. 10. Have the participant lean in towards the device with their face at a 45 angle; they should have one eye lined up with the center of the hole and their nose should be facing to the left of the hole. 11. Ask the participant to fixate on the central Green LED. 12. Using the hand controller center alignment and +/- buttons to adjust the instrument until the edge of the ring on the capture monitor is at the edge of the limbus, the central target should be at the center of the pupil. 13. When the participant is positioned correctly, the ring on the monitor will turn from red to green, when the ring is red the central target will indicate whether the patient needs to be closer to (indicated by a + sign in the center) or further away from the device (indicated by a - sign in the center). The green ring is just a guide. It is more important to make sure the iris is centered within the limbus ring correctly. 14. Adjust participant position using the hand controller until satisfied, then capture the image by pressing the button on top of the hand controller. Clicking the capture button again will return you to the capture screen for the same eye, if you use the touchscreen you will need to once again select the eye. DRCR.net UWF 200 Tx Imaging Protocol V Final Page 6 of 14

7 15. For the next image switch on the superior eye-steering LED (this LED will be red) and ask the participant to fixate on it. 16. Using the hand controller to move the limbus ring up until it is once again positioned properly around the iris. 17. When the participant s eye is in the proper position press the capture button. 18. After acquisition, press the capture button again to return or select the eye you wish to image and select reset position button. 19. Then switch on the lower inferior eye-steering LED (this LED will be red) and ask the patient to fixate on it. 20. Adjust the device using the hand controller to move the limbus ring down until it is once again positioned properly around the iris. 21. When the participant s eye is in the proper position press the capture button. 22. After acquisition, press the capture button again to return or select the eye you wish to image and select reset position. 23. When the color images have been reviewed, choose the complete session button in the top right hand corner. This will end the session or start the FA session. 24. At this point the Operator should set up for the FA injection. When ready to start the session, select the Study Eye. This will bring up the FA imaging window. 25. Make sure the contrast is set to 15 to begin the imaging session. 26. Position and align the patient to image the study eye. 27. Select the Right Eye or Left Eye button for the study eye. 28. During the transit phase, capture central 200 Optomap fa images. 29. As you begin the injection, press the capture button once. This will start the timer. 30. When the injection is complete, press the capture button again to record injection time. DRCR.net UWF 200 Tx Imaging Protocol V Final Page 7 of 14

8 31. Begin taking images at the rate of about an image per second beginning about 15 seconds after the start of the injection. A total of 3 images should be submitted to the Coordinating Center for each eye. 32. As more dye enters the eye the contrast slider may need to be decreased to avoid overexposure to preserve image detail. 33. Move to the non-study eye, adjust the chinrest, and select the appropriate eye. 34. Capture 3 central 200 Optomap fa images of the fellow eye. 35. Return to the study eye, adjust the chinrest, and select the appropriate eye 36. At 4 minutes, Capture 1 central 200 Optomap fa image of the study eye. 37. Capture 1 superior 200 Optomap fa image of the study eye 38. Capture 1 inferior 200 Optomap fa image of the study eye 39. Return to the fellow eye, adjust the chinrest, and select the appropriate eye 40. Capture 1 superior 200 Optomap fa image of the fellow eye 41. Capture 1 inferior 200 Optomap fa image of the fellow eye 42. It is important to remember only 75 images can be captured in a session. After 75 images you will need to start a new session, which will also restart your clock, so this is not recommended. 43. At the end of the session you can review images on the device and edit the session or you can review and edit at the review station. 44. When finished choose the complete session button in the upper right hand corner. This will save the images and they can be viewed at the review station. Certification Procedures A. Optos Training and Certification 1. Technician training and certification will be performed by Optos. Only certified imagers should obtain photos. If you have any questions about certification or ongoing training please contact Dana Keane at Optos (dkeane@optos.com). 2. B. Submitting Certification Scans Certification images should be submitted via directly to Dana Keane at Optos (dkeane@optos.com). Due to large file size, certification images should be submitted as screenshots of the actual image. Please contact Dana if you require further assistance. Image Transfer Procedures To collect images for upload: 1. Go to C: Optos Data/Images/Primary 2. All images are named Patient ID Date Time Eye Image # 3. Images can be copied directly from here for upload to the Coordinating Center. 4. Images will be uploaded to the Coordinating Center through the DRCR.net Image Uploader application and may be uploaded by operators and/or coordinators. DRCR.net UWF 200 Tx Imaging Protocol V Final Page 8 of 14

9 Image Retakes Images that do not meet quality standards will be flagged and you will be contacted to re-image the patient. Additional images can be taken to obtain the best quality and only the best uploaded. Examples of Well Captured images Images will need to have a full 200 view with no clipping of the iris or lens. Lid and lash should be minimized to obscure less than 10 of the field of view. For central images the reader should be able to visualize an unobscured view of the disc and macula with a field of view out to the vortex veins. Central Fixation Images Steered Images Superior field Inferior field DRCR.net UWF 200 Tx Imaging Protocol V Final Page 9 of 14

10 Contact Information: Director of Clinical Dana Keane Office: Cell: Sr Clinical Research Associate Jessica Carnevale Office: Cell: DRCR.net UWF 200 Tx Imaging Protocol V Final Page 10 of 14

11 Appendix A Cleaning the mirror Tools Required Suitable powder-free gloves. Flashlight (optional). Q-tips (cotton buds). Opto-Wipe cloths. De-ionized water. Process Description Safety The edges of the main mirror are sharp. Do not touch them. Only clean the area of the main mirror that you can see. Do not put your hand or any cleaning material into parts of the device you cannot see. Dust Do not use lint cloths, tissues or other materials that may create dust, near the scan head. General Cleaning Process This process may be carried out by customers who have been trained by Optos. The surface of the main mirror is very delicate, and must be handled with extreme care. Follow the steps below to remove dust from the mirror surface: 1. Switch on the device and take a normal eye image to assess image quality. Note: When switched on from cold, allow time for the lasers to reach the temperature required for optimum performance. P200MA and P200C devices will require 10 minutes, the P200 will require 15 minutes. 2. Switch off the device. 3. Remove any rings, watches, bracelets, jewelry or other items which may come into contact with the mirror surface. 4. Access the main mirror: i. Detach the head rest from the front of the eye aperture by pulling the pin at the bottom of the head rest down and sliding it completely off of the supporting bracket. ii. Remove the soft face pad from the eye aperture. 5. Put on the powder-free gloves. 6. Using the flashlight, look through the eye aperture and inspect the surface of the mirror for visible dust. Take notice of any smudges or spots. 9. Gather the corners of a clean dry Opto-wipe cloth into a mushroom shape. Use the mushroom ball to gently wipe the mirror. Reach into the eye aperture and carefully wipe the DRCR.net UWF 200 Tx Imaging Protocol V Final Page 11 of 14

12 visible surface of the mirror. Starting at one side of the mirror, using single downwards strokes, gently wipe the mirror. Always start from the top of the area you can see. Do not rub the mirror. Dispose of the dirty cloth. 10. Using the flashlight check the mirror again for any remaining smudges or spots. If it appears clean go to step 14 of this procedure. 11. If the main mirror still shows smudges or spots dip a Q-tip or clean dry Opto-wipe in deionized water and rub gently. CAUTION: Take extreme care not to touch anything inside the eye aperture other than the visible mirror surface, and not to drip any water inside the device. 12. Use a clean Opto-wipe cloth to lightly polish out any streaks. NOTE: GENTLE rubbing is essential. It is better to gently rub the spot several times than to rub the spot too hard as this may damage the mirror. 13. Inspect the mirror surface with the flashlight and confirm all contaminants have been removed. Repeat cleaning with the Q-tip if necessary. 14. Remove the gloves. 15. Replace the face pad and head rest. 16. Switch on the device and wait at least 10 minutes for device to warm up to the operating temperature required for optimum performance. Note: An error message may be displayed if capture is attempted before the required temperature has been reached. 17. Take a test eye images to determine if the image quality has been sufficiently improved. DRCR.net UWF 200 Tx Imaging Protocol V Final Page 12 of 14

13 Appendix B Operator Competency Assessment Operator: Optos Trainer: Site Name: Device Serial Number: Key Skills YES NO Remarks Does the user understand the safety guidelines? Can the operator use the Capture application and capture good images? Can the operator understand and explain what the device does? Does the user understand the imaging protocol? Can the user capture clinical trial quality images? Can the user archive the system? Does the user know where to get help? Does the user understand how to upload and transmit images to the reading center? Image Quality: Review of 3-5 patient sessions Image Type Acquired Excellent Average Poor Central Gaze Superior Steered Inferior Steered DRCR.net UWF 200 Tx Imaging Protocol V Final Page 13 of 14

14 Nasal Steered Temporal Steered Appropriate FA timing Pass Retraining Needed Operator Signature & Date: Optos Trainer Signature & Date: DRCR.net UWF 200 Tx Imaging Protocol V Final Page 14 of 14

Diabetic Retinopathy Clinical Research Network (DRCR.net) UWF Optos Imaging Protocol. Version /14/14

Diabetic Retinopathy Clinical Research Network (DRCR.net) UWF Optos Imaging Protocol. Version /14/14 Diabetic Retinopathy Clinical Research Network (DRCR.net) UWF Optos Imaging Protocol Version 1.0 10/14/14 DRCR.net UWF Imaging Protocol FINAL 10-14-14 Page 1 of 14 Table of Contents Background... 3 P200Tx

More information

DRCR.net Image Acquisition Protocol

DRCR.net Image Acquisition Protocol DRCR.net Image Acquisition Protocol Optical Coherence Tomography Angiography (OCT-A) Using: Optovue AngioVue Version 3.0 August 14, 2017 DRCR.net OCT-A Optovue AngioVue Procedure Manual 3.0 8-14-17 Table

More information

SPECTRALIS Training Guide

SPECTRALIS Training Guide SPECTRALIS Training Guide SPECTRALIS Diagram 1 SPECTRALIS Training Guide Table of Contents 1. Entering Patient Information & Aligning the Patient a. Start Up/Shut Down the System... 4 b. Examine a New

More information

Version 1.0. th March 2011

Version 1.0. th March 2011 Optical Coherence Tomography Scan and Retinal Imagingg Version 1.0 http://www.ukbiobank.ac.uk/ 5 th March 2011 This manual details the procedure for Scan and Retinal Imagingg at an Assessment Centre of

More information

Fundus Photograph Reading Center

Fundus Photograph Reading Center Autofluorescence Using Confocal Scanning Laser Ophthalmoscope (cslo) Instruments (AF-D) 8010 Excelsior Drive, Suite 100, Madison WI 53717 Telephone: (608) 410-0560 Fax: (608) 410-0566 Table of Contents

More information

RETINOPATHY SCREENING GUIDE

RETINOPATHY SCREENING GUIDE RETINOPATHY SCREENING GUIDE WHAT IS DIABETIC RETINOPATHY and RETINOPATHY SCREENING? Retinopathy is a disease of the retina. The retina is the nerve layer at the back of the eye. It is the part of the eye

More information

Macula centred, giving coverage of the temporal retinal. Disc centred. Giving coverage of the nasal retina.

Macula centred, giving coverage of the temporal retinal. Disc centred. Giving coverage of the nasal retina. 3. Field positions, clarity and overall quality For retinopathy screening purposes in England two images are taken of each eye. These have overlapping fields of view and between them cover the main area

More information

Zeiss LSM 780 Protocol

Zeiss LSM 780 Protocol Zeiss LSM 780 Protocol 1) System Startup F Please note the sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.

More information

Going beyond the surface of your retina

Going beyond the surface of your retina Going beyond the surface of your retina OCT-HS100 Optical Coherence Tomography Canon s expertise in optics and innovative technology have resulted in a fantastic 3 μm optical axial resolution for amazing

More information

Training Eye Instructions

Training Eye Instructions Training Eye Instructions Using the Direct Ophthalmoscope with the Model Eye The Model Eye uses a single plastic lens in place of the cornea and crystalline lens of the real eye (Fig. 20). The lens is

More information

The First True Color Confocal Scanner on the Market

The First True Color Confocal Scanner on the Market The First True Color Confocal Scanner on the Market White color and infrared confocal images: the advantages of white color and confocality together for better fundus images. The infrared to see what our

More information

Zeiss LSM 880 Protocol

Zeiss LSM 880 Protocol Zeiss LSM 880 Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.

More information

What s Fundus photography s purpose? Why do we take them? Why do we do it? Why do we do it? Why do we do it? 11/3/2014. To document the retina

What s Fundus photography s purpose? Why do we take them? Why do we do it? Why do we do it? Why do we do it? 11/3/2014. To document the retina What s Fundus photography s purpose? To document the retina Photographers role to show the retina Document other ocular structures Why do we take them? Why do we do it? We as photographers help the MD

More information

The TRC-NW8F Plus: As a multi-function retinal camera, the TRC- NW8F Plus captures color, red free, fluorescein

The TRC-NW8F Plus: As a multi-function retinal camera, the TRC- NW8F Plus captures color, red free, fluorescein The TRC-NW8F Plus: By Dr. Beth Carlock, OD Medical Writer Color Retinal Imaging, Fundus Auto-Fluorescence with exclusive Spaide* Filters and Optional Fluorescein Angiography in One Single Instrument W

More information

Impressive Wide Field Image Quality with Small Pupil Size

Impressive Wide Field Image Quality with Small Pupil Size Impressive Wide Field Image Quality with Small Pupil Size White color and infrared confocal images: the advantages of white color and confocality together for better fundus images. The infrared to see

More information

Zeiss AxioImager.Z2 Brightfield Protocol

Zeiss AxioImager.Z2 Brightfield Protocol Zeiss AxioImager.Z2 Brightfield Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge

More information

Diabetic Retinopathy Clinical Research Network

Diabetic Retinopathy Clinical Research Network Diabetic Retinopathy Clinical Research Network Photography and OCT Testing Procedures Manual Version 3.0 April 8, 2005 Table of Contents Modified 7-Standard Field/Modified 3-Standard Field Photographer

More information

Sets distance refraction. Moves camera forward / backward. Moves camera up / down. Moves camera left / right. Starts image acquisition (HRT 3 only)

Sets distance refraction. Moves camera forward / backward. Moves camera up / down. Moves camera left / right. Starts image acquisition (HRT 3 only) The perfect Image General workflow Do not conduct any examination beforehand that can disturb the tear film (e.g., examination using a contact glass, applanation tonometry). Explain the examination process

More information

The First True-Color Wide-Field Confocal Scanner

The First True-Color Wide-Field Confocal Scanner The First True-Color Wide-Field Confocal Scanner 2 Company Profile CenterVue designs and manufactures highly automated medical devices for the diagnosis and management of ocular pathologies, including

More information

Service Manual for XLE/XLT Series Laser Engravers

Service Manual for XLE/XLT Series Laser Engravers Service Manual for XLE/XLT Series Laser Engravers Table of Contents Maintenance...1 Beam alignment...3 Auto focus alignment...8 Bridge alignment...10 Electronics panel replacement...11 X motor change...12

More information

Standard Operating Procedure

Standard Operating Procedure Standard Operating Procedure Nanosurf Atomic Force Microscopy Operation Facility NCCRD Nanotechnology Center for Collaborative Research and Development Department of Chemistry and Engineering Physics The

More information

Revision 2, March Retinopathy Screening Manual

Revision 2, March Retinopathy Screening Manual Retinopathy Screening Manual 1 CONTENTS Introduction Page 3 Screeners - General Information Page 3 Items & Equipment Needed for Screening Page 3 Camera Set-Up & Screening Area Set-Up Page 3-5 -Fuses, Location

More information

Aeon ALR Series. EDGE FREE Ambient Light Rejecting Fixed Frame Screen USER S GUIDE

Aeon ALR Series. EDGE FREE Ambient Light Rejecting Fixed Frame Screen USER S GUIDE Aeon ALR Series EDGE FREE Ambient Light Rejecting Fixed Frame Screen USER S GUIDE Product Description: The Aeon ALR Series is a fixed frame projection screen with an EDGE FREE design The EDGE FREE design

More information

Sheep Eye Dissection

Sheep Eye Dissection Sheep Eye Dissection Question: How do the various parts of the eye function together to make an image appear on the retina? Materials and Equipment: Preserved sheep eye Scissors Dissection tray Tweezers

More information

Basic Users Manual for Tecnai-F20 TEM

Basic Users Manual for Tecnai-F20 TEM Basic Users Manual for Tecnai-F20 TEM NB: This document contains my personal notes on the operating procedure of the Tecnai F20 and may be used as a rough guide for those new to the microscope. It may

More information

The First True Color Confocal Scanner

The First True Color Confocal Scanner The First True Color Confocal Scanner White color and infrared confocal images: the advantages of white color and confocality together for better fundus images. The infrared to see what our eye is not

More information

Sable Frame CineGrey 3D

Sable Frame CineGrey 3D Sable Frame CineGrey 3D Ambient Light Rejecting Fixed Frame USER S GUIDE (spring type) Congratulations on your new Sable Frame purchase! The screen material included is our ISF Certified CineGrey 3D which

More information

Retinal Imaging System User Manual

Retinal Imaging System User Manual Retinal Imaging System User Manual 2014 Brien Holden Vision Diagnostics, Inc. All rights are reserved. No part of this manual may be reproduced or distributed in any form without written permission from

More information

Aeon CineGrey 3D Series

Aeon CineGrey 3D Series Aeon CineGrey 3D Series EDGE FREE Ceiling/Ambient Light Rejecting Fixed Frame Screen USER S GUIDE Version 2 (Spring type) Product Description: The Aeon Series is a fixed frame projection screen that uses

More information

2. Raise HT to 200kVby following the procedure explained in 1.6.

2. Raise HT to 200kVby following the procedure explained in 1.6. JEOL 2100 MANUAL Quick check list 1. If needed, fill the reservoir with LN2 2. Raise HT to 200kVby following the procedure explained in 1.6. 3. Insert specimen holder into TEM (Insert holder in airlock,

More information

Eyes. Inspection Visual Acuity Visual Fields Pupillary Response Fundoscopic Exam

Eyes. Inspection Visual Acuity Visual Fields Pupillary Response Fundoscopic Exam Eyes Inspection Visual Acuity Visual Fields Pupillary Response Fundoscopic Exam Eye Examination Inspection 11.Inspects external ocular (eye) structures (lids, conjunctiva, iris, cornea, pupils) 12.Gently

More information

Catt Follow-Up Study Scheie Image Reading Center. Digital Retinal Images Web Submission Application User's Guide

Catt Follow-Up Study Scheie Image Reading Center. Digital Retinal Images Web Submission Application User's Guide Catt Follow-Up Study Scheie Image Reading Center Digital Retinal Images Web Submission Application User's Guide Version 1 01/06/2014 Table of Contents Introduction...1 System Requirements...1 Accessing

More information

Wide Angle Ophthalmoscope Instructions

Wide Angle Ophthalmoscope Instructions Wide Angle Ophthalmoscope Instructions PLEASE READ AND FOLLOW THESE INSTRUCTIONS CAREFULLY Contents 1. Symbols 2. Warnings & Cautions 3. Description of Product 4. Getting Started 5. Apertures & Filters

More information

Sable Frame Series. Fixed Frame Projection Screen. Featuring CineGrey 3D Ambient Light Rejecting Material. User s Guide

Sable Frame Series. Fixed Frame Projection Screen. Featuring CineGrey 3D Ambient Light Rejecting Material. User s Guide Sable Frame Series Fixed Frame Projection Screen Featuring CineGrey 3D Ambient Light Rejecting Material User s Guide Congratulations on your new Sable Frame purchase! The screen material included is our

More information

CLARUS 500 from ZEISS HD ultra-widefield fundus imaging

CLARUS 500 from ZEISS HD ultra-widefield fundus imaging CLARUS 500 from ZEISS HD ultra-widefield fundus imaging Imaging ultra-wide without compromise. ZEISS CLARUS 500 // INNOVATION MADE BY ZEISS Compromising image quality may leave some pathology unseen. Signs

More information

DASH KIT INSTALLATION

DASH KIT INSTALLATION ÿÿÿÿÿ ÿÿÿÿÿÿÿÿÿÿÿÿ ÿÿÿ ÿÿÿÿÿ ÿÿ ÿÿÿ ÿÿÿÿÿÿÿÿÿÿÿÿÿÿ ÿÿÿÿÿÿÿÿÿÿ ÿÿÿÿ ÿÿÿÿ ÿÿ ÿÿÿÿÿ ÿÿÿÿÿÿÿÿ PROUDLY MADE IN THE USA DASH KIT INSTALLATION These instructions are aimed at the do-it-yourself installer. Read

More information

CLARUS 500 from ZEISS HD ultra-widefield fundus imaging

CLARUS 500 from ZEISS HD ultra-widefield fundus imaging CLARUS 500 from ZEISS HD ultra-widefield fundus imaging Imaging ultra-wide without compromise. ZEISS CLARUS 500 // INNOVATION MADE BY ZEISS Compromising image quality may leave some pathology unseen. Signs

More information

Going beyond the surface of your retina OCT-HS100 OPTICAL COHERENCE TOMOGRAPHY

Going beyond the surface of your retina OCT-HS100 OPTICAL COHERENCE TOMOGRAPHY Going beyond the surface of your retina OCT-HS100 OPTICAL COHERENCE TOMOGRAPHY Automatic functions make examinations short and simple. Perform the examination with only two simple mouse clicks! 1. START

More information

Going beyond the surface of your retina OCT-HS100 OPTICAL COHERENCE TOMOGRAPHY

Going beyond the surface of your retina OCT-HS100 OPTICAL COHERENCE TOMOGRAPHY Going beyond the surface of your retina OCT-HS100 OPTICAL COHERENCE TOMOGRAPHY Full Auto OCT High specifications in a very compact design Automatic functions make examinations short and simple. Perform

More information

THE DIPPING TECHNIQUE

THE DIPPING TECHNIQUE THE DIPPING TECHNIQUE An Alternative Finishing Technique By B. L. Herm Ely INTRODUCTION: A good finish should excite the senses. When touched with the finger tips, the finish should feel silky smooth.

More information

WITec Alpha 300R Quick Operation Summary October 2018

WITec Alpha 300R Quick Operation Summary October 2018 WITec Alpha 300R Quick Operation Summary October 2018 This document is frequently updated if you feel information should be added, please indicate that to the facility manager (currently Philip Carubia,

More information

TABLE OF CONTENTS Missouri Over There / Traveling Exhibit

TABLE OF CONTENTS Missouri Over There / Traveling Exhibit TABLE OF CONTENTS 1 Packing List 2 Floor Plans 3 Setup Directions 4 Case Contents PACKING LIST PACKING LIST case description weight dimensions 1A 5 - Graphic Panels 4 - Magnetic Struts 1 - Setup Book (in

More information

ezframe CineGrey 5D Ambient Light Rejecting Fixed Frame Screen

ezframe CineGrey 5D Ambient Light Rejecting Fixed Frame Screen 1 ezframe CineGrey 5D Ambient Light Rejecting Fixed Frame USER S GUIDE Congratulations on your new ezframe purchase! The screen material included is our award winning CineGrey 5D which is a front projection

More information

The SureID Registration Station. Fingerprint Services

The SureID Registration Station. Fingerprint Services The SureID Registration Station Fingerprint Services Handbook Overview Overview 3 What are FBI Fingerprinting Services? Why Do People Need This Service? What Will I Do? Fingerprints 4 Type of Fingerprints

More information

JEOL 6500 User Manual

JEOL 6500 User Manual LOG IN to your session on the computer to the left of the microscope. Starting Conditions 1. Press Ctrl-Alt-Del and log on to the microscope computer. Click on JEOL PC SEM 6500 icon. Click yes if message

More information

User Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com

User Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com User Manual Digital Compound Binocular LED Microscope Model MD82ES10 MicroscopeNet.com Table of Contents i. Caution... 1 ii. Care and Maintenance... 2 1. Components Illustration... 3 2. Installation...

More information

NOTES ON PRINTING 08/06

NOTES ON PRINTING 08/06 NOTES ON PRINTING 08/06 1-Photographic materials are sensitive to most light sources. Never open your box of paper unless you are working under an appropriate yellow or red safelight. (Color photographic

More information

Practical work no. 3: Confocal Live Cell Microscopy

Practical work no. 3: Confocal Live Cell Microscopy Practical work no. 3: Confocal Live Cell Microscopy Course Instructor: Mikko Liljeström (MIU) 1 Background Confocal microscopy: The main idea behind confocality is that it suppresses the signal outside

More information

PRINTING SETUP with OMEGA ENLARGER

PRINTING SETUP with OMEGA ENLARGER PRINTING SETUP with OMEGA ENLARGER Follow this procedure when beginning a work session in the Darkroom. Use this page as an actual checklist as you prepare to work. Get a key from the Monitor in the Cage

More information

How-to guide. Working with a pre-assembled THz system

How-to guide. Working with a pre-assembled THz system How-to guide 15/06/2016 1 Table of contents 0. Preparation / Basics...3 1. Input beam adjustment...4 2. Working with free space antennas...5 3. Working with fiber-coupled antennas...6 4. Contact details...8

More information

Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope

Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope The procedures given below were written specifically for the FEI Tecnai G 2 Sphera microscope. Modifications will need to

More information

Microscopy. The dichroic mirror is an important component of the fluorescent scope: it reflects blue light while transmitting green light.

Microscopy. The dichroic mirror is an important component of the fluorescent scope: it reflects blue light while transmitting green light. Microscopy I. Before coming to lab Read this handout and the background. II. Learning Objectives In this lab, you'll investigate the physics of microscopes. The main idea is to understand the limitations

More information

Yokohama City University lecture INTRODUCTION TO HUMAN VISION Presentation notes 7/10/14

Yokohama City University lecture INTRODUCTION TO HUMAN VISION Presentation notes 7/10/14 Yokohama City University lecture INTRODUCTION TO HUMAN VISION Presentation notes 7/10/14 1. INTRODUCTION TO HUMAN VISION Self introduction Dr. Salmon Northeastern State University, Oklahoma. USA Teach

More information

Biology 29 Cell Structure and Function Spring, 2009 Springer LABORATORY 1: THE LIGHT MICROSCOPE

Biology 29 Cell Structure and Function Spring, 2009 Springer LABORATORY 1: THE LIGHT MICROSCOPE Biology 29 Cell Structure and Function Spring, 2009 Springer LABORATORY 1: THE LIGHT MICROSCOPE Prior to lab: 1) Read these instructions (p 1-6) 2) Go through the online tutorial, the microscopy pre-lab

More information

HP Color LaserJet CP3525 Series Manage and maintain

HP Color LaserJet CP3525 Series Manage and maintain Load paper and print media Load Tray 1 1 Open Tray 1. CAUTION: To avoid jams, never add or remove paper from Tray 1 during printing. 2 Fold out the tray extension to support the paper and set the side

More information

ML7520 ML7530 DIOPTER ADJUSTMENT RING BINOCULAR BODY, INCLINED 30. (a) Field Iris Control Lever. (c) Filter Slots EYEPIECES, KHW10X

ML7520 ML7530 DIOPTER ADJUSTMENT RING BINOCULAR BODY, INCLINED 30. (a) Field Iris Control Lever. (c) Filter Slots EYEPIECES, KHW10X JAPAN DIOPTER ADJUSTMENT RING BINOCULAR BODY, INCLINED 30 (a) Field Iris Control Lever (c) Filter Slots EYEPIECES, KHW10X ANALYZER CONTROL LEVER (b) Aperture Iris Control Lever LIGHT SOURCE HOUSING VERTICAL

More information

Marine Invertebrate Zoology Microscope Introduction

Marine Invertebrate Zoology Microscope Introduction Marine Invertebrate Zoology Microscope Introduction Introduction A laboratory tool that has become almost synonymous with biology is the microscope. As an extension of your eyes, the microscope is one

More information

Renishaw InVia Raman microscope

Renishaw InVia Raman microscope Laser Spectroscopy Labs Renishaw InVia Raman microscope Operation instructions 1. Turn On the power switch, system power switch is located towards the back of the system on the right hand side. Wait ~10

More information

Fundus Photograph Reading Center

Fundus Photograph Reading Center Spectral Domain Optical Coherence Tomography (SD-OCT) Heidelberg Spectralis 8010 Excelsior Drive, Suite 100, Madison WI 53717 Telephone: (608) 410-0560 Fax: (608) 410-0566 Table of Contents 1. Heidelberg

More information

General Help. Last revised: Winter When I try to print something on the computer, it appears to work, but nothing comes out of the printer.

General Help. Last revised: Winter When I try to print something on the computer, it appears to work, but nothing comes out of the printer. General Help Last revised: Winter 2015 Problem Solution When I try to print something on the computer, it appears to work, but nothing comes out of the printer. See the next item. When I try to print something

More information

ADVANCED USER S GUIDE

ADVANCED USER S GUIDE ADVANCED USER S GUIDE DCP-J125 DCP-J315W Version 0 ARL/ASA/NZ/SAF User's Guides and where do I find it? Which manual? What's in it? Where is it? Safety and Legal Quick Setup Guide Basic User's Guide Advanced

More information

Ambient Light Rejecting Fixed Frame Screen. (3.9 width frame) User s Guide

Ambient Light Rejecting Fixed Frame Screen. (3.9 width frame) User s Guide Ambient Light Rejecting Fixed Frame Screen (3.9 width frame) User s Guide Attention: The DarkStar 9 is composed of high quality sensitive multi-layer projection material. Please follow the precautions

More information

INSTRUCTION MANUAL INF Fax: (503)

INSTRUCTION MANUAL INF Fax: (503) INSTRUCTION MANUAL INF151 1-800-547-5740 Fax: (503) 643-6322 www.ueiautomotive.com email: info@ueitest.com Introduction Congratulations on your purchase of the INF151 infrared thermometer. Like all UEi

More information

OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop

OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop 1. Log in usage using the SMIF web site 2. Turn power

More information

Point Calibration. July 3, 2012

Point Calibration. July 3, 2012 Point Calibration July 3, 2012 The purpose of the Point Calibration process is to generate a map of voltages (for galvos) or motor positions of the pointing device to the voltages or pixels of the reference

More information

Ox-RAC-08 Ribbon Angled Fiber Cleaver User Manual

Ox-RAC-08 Ribbon Angled Fiber Cleaver User Manual Ox-RAC-08 Ribbon Angled Fiber Cleaver User Manual Issue 2.0 Contents Introduction... 2 Contents of Cleaving Kit & Unpacking... 3 Cleaving Problems... 8 Blade damage:... 9 Cleaver Maintenance... 10 Cleaning

More information

Guide to Confocal 5. Starting session

Guide to Confocal 5. Starting session Guide to Confocal 5 Remember that when booking and before starting session you can check for any problems at https://www.bris.ac.uk/biochemistry/uobonly/cif/index.html Starting session Switch on microscope

More information

Image and Data Acquisition

Image and Data Acquisition Image and Data Acquisition LCP Image Acquisition Procedures This section provides guidelines for scanning images that will be added to the LCP image archive. By scanning the image, we obtain a digital

More information

User Manual. For MAC

User Manual. For MAC User Manual For MAC A3 Scanner with The TWAIN Driver (For Macintosh) English User Manual Contents Preface Chapter One Setting Up the Scanner 1 System Requirements... 1 Unpacking the Scanner... 2 The Scanner

More information

Laser LA-4P. Operating instructions

Laser LA-4P. Operating instructions Laser LA-4P GB Operating instructions A 1 2 3a 5 3c 3b 3a 4 11 11 6 10 7 14a 14b 14c 12 9 8 B C 2. 1. D E F Ø 50mm - 115 mm Ø 2-4,5 G I K s > 6m > 20ft L M N P1 Q O 13 P2 GB Operating instructions The

More information

Stitching MetroPro Application

Stitching MetroPro Application OMP-0375F Stitching MetroPro Application Stitch.app This booklet is a quick reference; it assumes that you are familiar with MetroPro and the instrument. Information on MetroPro is provided in Getting

More information

INTRODUCTION. SO-2200LED Binocular Indirect Ophthalmoscope

INTRODUCTION. SO-2200LED Binocular Indirect Ophthalmoscope INTRODUCTION Please read the following information carefully before installing and using the Scan Optics SO-2200LED binocular indirect ophthalmoscope. Scan Optics is responsible for the safety, reliability

More information

Using a Microscope. Year Group: BVSc1 + Document number: CSL_L07

Using a Microscope. Year Group: BVSc1 + Document number: CSL_L07 Year Group: BVSc1 + Document number: CSL_L07 Equipment list: Equipment for this station: Microscope Power supply and a level surface to work on Gloves The sample to examine Marker or pencil for labelling

More information

3. are adherent cells (ie. cells in suspension are too far away from the coverslip)

3. are adherent cells (ie. cells in suspension are too far away from the coverslip) Before you begin, make sure your sample... 1. is seeded on #1.5 coverglass (thickness = 0.17) 2. is an aqueous solution (ie. fixed samples mounted on a slide will not work - not enough difference in refractive

More information

ifs 150 khz IntraLase Technology

ifs 150 khz IntraLase Technology How to create a perfect LASIK flap with Intralase Dr Lim Li, FRCS Head(Clinical & Education) Senior Consultant Corneal and External Eye Disease Service Singapore National Eye Centre Financial Interest:

More information

BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions

BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions Starting up: Schematic of Olympus BX-61. 1. Turn on Olympus microscope power box (left of microscope) with toggle switch on

More information

IDEXX-PACS * 4.0. Imaging Software. Quick Reference Guide

IDEXX-PACS * 4.0. Imaging Software. Quick Reference Guide 4 IDEXX-PACS * 4.0 Imaging Software Quick Reference Guide Capturing Images Before you begin: Adjust the collimation properly. Make sure the body part you are imaging matches the exam type you have selected.

More information

CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE

CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE University of Virginia ITC Academic Computing Health Sciences CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE Introduction The Olympus BH-2 microscope in ACHS s microscope lab has objectives from

More information

Scanning Archival Images

Scanning Archival Images Scanning Archival Images A Guide for Community Heritage Projects A Project of the Gimli Municipal Heritage Advisory Committee Scanning Archival Images A Guide for Community Heritage Projects THIS GUIDE

More information

Micropulse Duty Cycle. # of eyes (20 ms) Total spots (200 ms)

Micropulse Duty Cycle. # of eyes (20 ms) Total spots (200 ms) Micropulse Duty Cycle Total spots (2 ms) # of eyes (2 ms) Total spots (2 ms) % 269 44 3 47% 9 4 4 25% 3 5 4 4 5% 2 4 3 5 2% 5 2 NA NA 9% 2 4 6% NA NA 57 2 5% 4 5 6 3 3% 39 5 35 5 # of eyes (2 ms) Supplemental

More information

NYSITELL SCANNING PROCESS FOR NYC SCHOOLS. Version 1.2 Updated February 2014

NYSITELL SCANNING PROCESS FOR NYC SCHOOLS. Version 1.2 Updated February 2014 NYSITELL SCANNING POCESS FO NYC SCHOOLS Version 1.2 Updated February 2014 2 P a g e Table of Contents Overview... 3 Section 1: Printing Answer Document from ATS... 4 Section 1A: The Trailing Test Page...

More information

CLEANING THE GALVO MIRRORS

CLEANING THE GALVO MIRRORS CLEANING THE GALVO MIRRORS Formlabs Customer Support can provide additional information on mirror cleaning and in some cases, provide the customer with the appropriate cleaning supplies. Tools: 2.5mm hex

More information

JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES

JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES RULES All users must go through a series of standard operation procedure training. For more information contact: Longlong Liao Teaching

More information

b. Turn the power switch and key to on position for blue laser.

b. Turn the power switch and key to on position for blue laser. OLYMPUS FLUOVIEW 300 CONFOCAL MICOSCOPE OPERATION PROCEDURE 1. Turn ON microscope in this order: 1) Turn on mercury lamp (Note: once the mercury lamp is turned off, DO NOT turn it back on for at least

More information

Advanced User s Guide

Advanced User s Guide dvanced User s Guide DCP-J40W Version 0 UK/IRE/GEN User s Guides and where do I find them? Which Guide? What's in it? Where is it? Product Safety Guide Quick Setup Guide Basic User s Guide dvanced User

More information

Motorized Axio Observer Start-up instructions

Motorized Axio Observer Start-up instructions Start-up instructions 1. If using fluorescence turn on Fluorescent light source. TL light Source (Hal 100) 2. Turn on microscope using switch on lower left side of the microscope. 3. If imaging, turn on

More information

Photometer Maintenance Guide

Photometer Maintenance Guide Photometer Maintenance Guide Product Support Information This guide is designed to help ensure your instrument remains in a good condition and continues to deliver accurate results. It includes a series

More information

Kulicke & Soffa 4700AD Instructions A.Atalar (April 2017)

Kulicke & Soffa 4700AD Instructions A.Atalar (April 2017) Kulicke & Soffa 4700AD Instructions A.Atalar (April 2017) 4700AD model wire bonder can act like a wedge or ball-bonder. In the user manual refer to 4523AD for wedge bonding and 4524AD for ball-bonding.

More information

TECHNICAL INSTRUCTIONS FOR DataTrace CALIBRATION PROCEDURES

TECHNICAL INSTRUCTIONS FOR DataTrace CALIBRATION PROCEDURES TECHNICAL INSTRUCTIONS FOR DataTrace CALIBRATION PROCEDURES HUMIDITY CALIBRATION The following procedures provide instructions for performing field calibrations on DataTrace equipment. There are two types

More information

Scratch Assay Device for Wound Healing Studies

Scratch Assay Device for Wound Healing Studies Scratch Assay Device for Wound Healing Studies By Team Cell Scratcher: Hannah Morgan, Charles Merchant, and Zachariah Cribbin 1 Quick Start Guide 2 Comments: _ Table of Contents Cover Page 1 1.0 Quick

More information

SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014

SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014 CIAN LSM1 or LSM2 short instructions, version 1.4, September 2014 page 1 of 6 SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014 Before starting To work with LSM1

More information

Non-contact Forehead IR Thermometer

Non-contact Forehead IR Thermometer User s Manual Non-contact Forehead IR Thermometer Model IR200 Introduction Congratulations on your purchase of the Model IR200 IR Thermometer. This thermometer is intended for scanning groups of individuals

More information

Advanced User s Guide

Advanced User s Guide Advanced User s Guide DCP-J72W Version 0 ARL/NZ User's Guides and where do I find them? Which Guide? What's in it? Where is it? Product Safety Guide Quick Setup Guide Basic User s Guide Advanced User s

More information

RegionFinder The perfect Use. SPECTRALIS HRA, HRA+OCT and SPECTRALIS OCT

RegionFinder The perfect Use. SPECTRALIS HRA, HRA+OCT and SPECTRALIS OCT RegionFinder The perfect Use SPECTRALIS HRA, HRA+OCT and SPECTRALIS OCT RegionFinder - The perfect Use SPECTRALIS HRA, HRA+OCT and SPECTRALIS OCT General workflow to acquire the perfect AF image Adjust

More information

LASER ENHANCED REVOLVER GRIP OWNER S MANUAL RED LASER GREEN LASER

LASER ENHANCED REVOLVER GRIP OWNER S MANUAL RED LASER GREEN LASER LASER ENHANCED RED LASER GREEN LASER REVOLVER GRIP OWNER S MANUAL LASER ENHANCED GRIP Installation Instructions Caution... 3 Safety Labels... 4 Installation...5-7 Programming...8-10 Batteries (Red Laser)...

More information

ILFORD SPORTSVIEW PROJECTOR INSTRUCTION BOOK

ILFORD SPORTSVIEW PROJECTOR INSTRUCTION BOOK ILFORD SPORTSVIEW PROJECTOR INSTRUCTION BOOK Now that you're the owner of a new Sportsview Projector, you'll want to begin using it right away. The Sportsview Projector is extremely simple to operate,

More information

Lecture 2 Slit lamp Biomicroscope

Lecture 2 Slit lamp Biomicroscope Lecture 2 Slit lamp Biomicroscope 1 Slit lamp is an instrument which allows magnified inspection of interior aspect of patient s eyes Features Illumination system Magnification via binocular microscope

More information

VISULAS Trion. Treatment flexibility to the power of three. Multicolor Photocoagulation Laser

VISULAS Trion. Treatment flexibility to the power of three. Multicolor Photocoagulation Laser VISULAS Trion Treatment flexibility to the power of three Multicolor Photocoagulation Laser Carl Zeiss: A pioneer in retinal therapy For many years, Carl Zeiss has fostered a culture of highest precision,

More information

Scanner Parts. Scanner Specifications. General. Mechanical. Epson Perfection V200 Photo. P Power/ ] Start button Status light. { Copy button.

Scanner Parts. Scanner Specifications. General. Mechanical. Epson Perfection V200 Photo. P Power/ ] Start button Status light. { Copy button. Scanner Parts { Copy button Scan to E-mail button a b P Power/ ] Start button Status light c e d a Scan to PDF button Film holder Scanner Specifications General Scanner type Flatbed color Photoelectric

More information

LSOCA. Fundus Photography Reading Center. Procedures for standard field color fundus images using film and digital imaging

LSOCA. Fundus Photography Reading Center. Procedures for standard field color fundus images using film and digital imaging LSOCA Fundus Photography Reading Center Procedures for standard field color fundus images using film and digital imaging University of Wisconsin - Madison Department of Ophthalmology and Visual Sciences

More information