Phase Drift Budget Analysis for 12 GeV 1497 MHz LLRF System

Size: px
Start display at page:

Download "Phase Drift Budget Analysis for 12 GeV 1497 MHz LLRF System"

Transcription

1 Phase Drift Budget Analysis for 12 GeV 1497 MHz LLRF System John Musson 28-Sept-7 Introduction The 12 GeV upgrade effort included the creation of LLRF Requirements, directed at achieving.4% gradient regulation,.5 degrees of fast phase regulation, and 3 degrees of slow phase regulation [1]. Although the requirements are more demanding than those for the initial 6 GeV CEBAF machine, a digital LLRF Control Module has been designed which conforms to the new specifications. An abbreviated Requirements List appears in Appendix A. During the 12 GeV LLRF design, strict attention was given to phase-versus-temperature performance, in an effort to improve current operating performance by at least an order of magnitude. Original CEBAF LLRF performance was limited to ~1 degree / C (at 499 MHz), and although great attempts were made to physically temperature-stabilize the injector service building, +/- 2 C was the best achievable ambient regulation. In addition, a 5th-order polynomial correction was included in the initial CEBAF LLRF module, but attempts to apply the correction to improve stability were inconsistent and ultimately abandoned. The 12 GeV upgrade specifies a slow, diurnal RF drift of < 3 degrees over reasonable seasonal temperature ranges. A 5 C peak-to-peak change is currently considered reasonable. As part of 6 GeV accelerator hardening, the warm cavity 499 MHz LLRF components were replaced with interim-design modules, in an effort to improve the temperature stability to < 1 ps/ C. The 499 MHz injector upgrade LLRF module provided.1 degree/ C (~73 fs/ C) performance, an order of magnitude improvement. Appendix B provides supporting data for both 499 MHz LLRF systems. Overall system performance involves not only the LLRF Control Module, but also the cabling and coupling elements. Figure 1 is a system diagram, highlighting the LLRF as well as the passive components contained in the control loop. The feedback cable actually consists of 55' of LDF4 Heliax, and a 4' length of FSJ1 Heliax, serving as a jumper. Figure 1. Schematic description of LLRF control loop, highlighting major passive components.

2 Component Characterization Of particular concern are those elements in the feedback path, which ultimately corrupt the desired cavity field probe measurement. HP 858A Vector Voltmeters (VVM) were applied to the injector LLRF system as a slow-drift phase monitor / correction system, and also provided in-situ system performance data. HP858A specifications are provided in Appendix C. In addition to Injector phase monitoring, the VVM was used to characterize the Heliax and components. In order to verify the published specifications, of the VVM, a test was performed, whereby an identical 1427 MHz Master Oscillator signal was applied to both A and B inputs for several days. The setup and resulting baseline drift are shown in Figure 2. Figure 2. Depiction of setup used to test intrinsic phase drift of HP 858A Vector Voltmeter, in situ, and associated data. Improvements to the CEBAF Injector Driveline in 1998 led to a battery of temperature studies, in order to completely characterize the behavior of the 1427 MHz LO distribution system. Thermal data was subjected to Monte Carlo simulation, and limits were determined for best-and-worst-case ambient temperature fluctuations (with and without active driveline regulation). Subsequently, the temperature data has proven to be useful for Heliax selection, and for phase budget calculations. Figure 3 shows the experimental test setup, with an environmental chamber, used to gather the component thermal performance data. The setup was capable of single-frequency measurements, as well as IF measurements involving frequency translation. Figure 3. Test setup used to perform environmental studies of various cables and components. Ramp and soak cycles are performed, and data collected by hand or GPIB interface. Single-frequency or IF measurements are possible, as shown.

3 All common-mode items were co-located outside the environmental chamber, so as to limit the effects of room temperature on the the measurements. The DUT was placed inside the oven, with careful attention placed on lengths and typs of interconnect cables within the enclosure (again, to isolate the DUT from any testbed effects due to thermal and/or frequency variations). Straight-through measurements were performed, to verify that the ambient effects were at or below.1-degrees. Typically, the temperature was ramped up in several degree increments, after experiencing a 15-3 minute soak at each point, and then ramped down, but landing on different points. Residual hysteresis provided an indication as to whether soak times were adequate, and also provided information on the stability and repeatability of the DUT. It was discovered that Heliax requires a bakeout process, whereby the cable is allowed to sit for several days at 55 C. The inner dielectric appears to have a mechanical relaxation process, which stabilizes after being subjected to higher-than-expected operating temperatures. Results for several common Heliax products are shown in Figure 4, demonstrating the degree of stabilization, as well as suggesting possible optimum operating temperatures [2]. LDF4 Heliax LDF2 Heliax y =.71x x R2 = y =.172x x R2 = y =.15219x x R2 = y =.183x x R2 = Conformable Cable FSJ4 Heliax, Composite, Baked FSJ2 Heliax, 55' FSJ1 Heliax, Composite y =.126x x R2 = y = x x R2 = Figure 4. Measured delay response vs. temperature for various Heliax products, and associated polynomial regression data.

4 Table 1 is a summary of extrapolated temperature coefficients for the various cables, at a given temperature, normalized to ps/ft/ C. These slopes were derived from the resulting second-order curve fit. In application, the proposed ambient temperature is chosen, and the slope for the specific cable is used to determine the excess time delay in the neighborhood of temperatures. Large temperature ranges should refer to the entire polynomial. Table 1. Extrapolated temperature slopes for given cables, at specific temperatures. Units are normalized to ps/ft/ C. Temperature, C LDF2 LDF4 FSJ1 FSJ2 FSJ Conformable Table 2 is a tabulation of various passive components, also subjected to thermal study. The coupler slopes had much larger error bars, and were very close to the measureable limits of the HP858A. Table 2. Tabulated temperature slopes for given cables and passive components, at specific temperatures. Units are degrees/ft/ C (cable) and degrees/ C (passives), measured at 1427 Mhz. To convert to ps/ C, multiply by 1.95 ps / degree.

5 An in-situ measurement of a LDF4 probe cable at 499 MHz was performed by setting up a standing wave in the 55' cable at 1.6 GHz, while under full operation. Since the 499 MHz Chopper cavities fully reflect the 1.6 GHz signal, high resolution phase drift measurements were possible over a several day period. Figure 5 demonstrates the procedure, as well as providing data, confirming the ~6 fs / ft / C as given by the static oven measurements (nominal temperature = C, variation = ~2 C). Figure 5. Dynamic, in-situ measurement of 55' LDF4 probe cable for interim-upgrade 499 MHz RFCM. A 2-degree variaiton (measured at 1.6 GHz), at ~ C (+/- 2 C) confirmed the oven measurement and regression prediction of ~6 fs/ft/ C. Data courtesy of Tomasz Plawski. Finally, the a 1497 MHz LLRF receiver protoype for the 12 GeV upgrade was evaluated, and measurements were compared to the predicted performance of the system, as well as to previous 499 MHz module performance. Figure 6 demonstrates the degree / C slope, measured at 1497 MHz. The resulting ~4 fs/ C temperature coefficient is below the.5ps/ C receiver specification of the 12 GeV LLRF Upgrade Requirements. Figure 6. Relative phase drift vs. temperature for 1497 MHz prototype receiver, demonstrating a 4 fs / C tempco.

6 Tabulated System Performance Although the klystron gallery service buildings are temperature controlled, it is not uncommon to see several degrees of diurnal fluctuation. Presuming a +/- 3 C change, it it possible to construct a phase drift budget for the entire system, shown in Table 3: Table 3. Tabulation of phase drifts for 1497 Mhz LLRF system. Ambient temperature = C, range = 6 C (+/- 3 C). Component LLRF Module Feedback Coupler Feedback Heliax, LDF4, 55', C Probe Heliax Jumper, FSJ1, 4', C Total and RMS +/- 3 C peak / 1497 MHz ps / C (ps/ C) ps / C.54 ps / C -5.17ps, peak 2.87 degrees, peak 3.24 ps, RMS 1.8 degrees, RMS Summary Although the LLRF receiver and feedback path represent only a few elements, a great deal of temperature characterization was required to quantify each component, in order to obtain an accurate perfomance prediction. The ~2 degree RMS (3 degree, peak) diurnal prediction represents a value leading to minimal impact on accelerator operations. Several degrees can actually be automatically removed using the MO Modulation system, which gently krests the Linac energy based on a precise mass-spectrometric measurement of beam energy. Operators are also able to easily remove the day-today degree variations by hand. Ultimately, preliminary engineering efforts, as well as prototype measurements, demonstrate the abillity to achieve the short and long-term stability requirements of.5 ps / C, or 15 ps over 1 C to 4 C range, thus satisfying control requirements for the future 12 GeV accelerator. Acknowledgements The author would like to thank student intern, Guy Randall, for his eager participation in painstaking data collection during the thermal testing. Also, to colleagues Tomasz Plawski for sharing his data taken while performing the in-situ probe cable measurement, and Mark Wissmann for assistance with the antique environmental chamber. References 1. H. Dong, A. Hofler, C. Hovater, J. Musson, Performance and Technical Requirements for the CEBAF Energy Upgrade Low Level RF System, M:\ees\RF\12 GeV_LLRF\LLRF Documentation and Specifications 2. Actual spreadshhet is located in the JLAB Operations Electronic Logbook, LLRFLOG, Entry #

7 Appendix A General Requirements List Frequency 1497 MHz G*BW 1 MHz IF 7 MHz LO input 1427 MHz Phase Ctrl.5 degrees Amplitude Control.45% Dynamic Range 2 db Setpoint Resolution.1%.1 degree Diagnostics Loopback Cable Fault Thermal Stability +/-.5 db, 15 ps; 1 C < T < 4 C.2 db / C;.5 ps / C Power Use less energy than the sun

8 Appendix B Thermal Drift of Old 499 MHz CEBAF LLRFCM vs. 12 Gev Digital LLRFCM Bronze Age 1 degree / C > 5.6 ps per C Digital LLRF Module ~.13 degree / C

9 Appendix C HP 858A Vector Voltmeter Specifications

Digital LLRF Test on the Renascence Cryomodule

Digital LLRF Test on the Renascence Cryomodule Digital LLRF Test on the Renascence Cryomodule Trent Allison, Rama Bachimanchi, Curt Hovater, John Musson and Tomasz Plawski Introduction The Renascence cryomodule was the first opportunity for testing

More information

Direct Digital Down/Up Conversion for RF Control of Accelerating Cavities

Direct Digital Down/Up Conversion for RF Control of Accelerating Cavities Direct Digital Down/Up Conversion for RF Control of Accelerating Cavities C. Hovater, T. Allison, R. Bachimanchi, J. Musson and T. Plawski Introduction As digital receiver technology has matured, direct

More information

Low-Level RF. S. Simrock, DESY. MAC mtg, May 05 Stefan Simrock DESY

Low-Level RF. S. Simrock, DESY. MAC mtg, May 05 Stefan Simrock DESY Low-Level RF S. Simrock, DESY Outline Scope of LLRF System Work Breakdown for XFEL LLRF Design for the VUV-FEL Cost, Personpower and Schedule RF Systems for XFEL RF Gun Injector 3rd harmonic cavity Main

More information

Design considerations for the RF phase reference distribution system for X-ray FEL and TESLA

Design considerations for the RF phase reference distribution system for X-ray FEL and TESLA Design considerations for the RF phase reference distribution system for X-ray FEL and TESLA Krzysztof Czuba *a, Henning C. Weddig #b a Institute of Electronic Systems, Warsaw University of Technology,

More information

C100 Cryomodule. Seven cell Cavity, 0.7 m long (high Q L ) 8 Cavities per Cryomodule Fits the existing Cryomodule footprint

C100 Cryomodule. Seven cell Cavity, 0.7 m long (high Q L ) 8 Cavities per Cryomodule Fits the existing Cryomodule footprint 1 new module C100 Cryomodule Seven cell Cavity, 0.7 m long (high Q L ) 8 Cavities per Cryomodule Fits the existing Cryomodule footprint Fundamental frequency f 0 Accelerating gradient E acc 1497 MHz >

More information

ABSTRACT 1 CEBAF UPGRADE CAVITY/CRYOMODULE

ABSTRACT 1 CEBAF UPGRADE CAVITY/CRYOMODULE Energy Content (Normalized) SC Cavity Resonance Control System for the 12 GeV Upgrade Cavity: Requirements and Performance T. Plawski, T. Allison, R. Bachimanchi, D. Hardy, C. Hovater, Thomas Jefferson

More information

R.Bachimanchi, IPAC, May 2015, Richmond, VA

R.Bachimanchi, IPAC, May 2015, Richmond, VA 1 new module C100 Cryomodule Seven cell Cavity, 0.7 m long (high Q L ) 8 Cavities per Cryomodule Fits the existing Cryomodule footprint Fundamental frequency f 0 Accelerating gradient E acc 1497 MHz >

More information

New apparatus for precise synchronous phase shift measurements in storage rings 1

New apparatus for precise synchronous phase shift measurements in storage rings 1 New apparatus for precise synchronous phase shift measurements in storage rings 1 Boris Podobedov and Robert Siemann Stanford Linear Accelerator Center, Stanford University, Stanford, CA 94309 Measuring

More information

A Sub-Picosecond Phase Stability Frequency Multiplier

A Sub-Picosecond Phase Stability Frequency Multiplier A Sub-Picosecond Phase Stability Frequency Multiplier Charles L. Wenzel, Richard Koehler, Reed Irion Abstract ---- A UHF to S-band (476 MHz to 2.856 GHz) frequency multiplier exhibiting sub-picosecond

More information

Energy Recovering Linac Issues

Energy Recovering Linac Issues Energy Recovering Linac Issues L. Merminga Jefferson Lab EIC Accelerator Workshop Brookhaven National Laboratory February 26-27, 2002 Outline Energy Recovery RF Stability in Recirculating, Energy Recovering

More information

RF System LSD Work. William Merz

RF System LSD Work. William Merz RF System LSD Work William Merz LSD Re-Baseline Review Jefferson Lab Thomas Jefferson National Accelerator Facility Page 1 Outline What I will talk about 12 GEV RF power system installation and commissioning

More information

Digital Self Excited Loop Implementation and Experience. Trent Allison Curt Hovater John Musson Tomasz Plawski

Digital Self Excited Loop Implementation and Experience. Trent Allison Curt Hovater John Musson Tomasz Plawski Digital Self Excited Loop Implementation and Experience Trent Allison Curt Hovater John Musson Tomasz Plawski Overview Why Self Excited Loop? Algorithm Building Blocks Hardware and Sampling Digital Signal

More information

Review on Progress in RF Control Systems. Cornell University. Matthias Liepe. M. Liepe, Cornell U. SRF 2005, July 14

Review on Progress in RF Control Systems. Cornell University. Matthias Liepe. M. Liepe, Cornell U. SRF 2005, July 14 Review on Progress in RF Control Systems Matthias Liepe Cornell University 1 Why this Talk? As we all know, superconducting cavities have many nice features one of which is very high field stability. Why?

More information

INSTALLATION AND FIRST COMMISSIONING OF THE LLRF SYSTEM

INSTALLATION AND FIRST COMMISSIONING OF THE LLRF SYSTEM INSTALLATION AND FIRST COMMISSIONING OF THE LLRF SYSTEM FOR THE EUROPEAN XFEL Julien Branlard, for the LLRF team TALK OVERVIEW 2 Introduction Brief reminder about the XFEL LLRF system Commissioning goals

More information

Beam Diagnostics, Low Level RF and Feedback for Room Temperature FELs. Josef Frisch Pohang, March 14, 2011

Beam Diagnostics, Low Level RF and Feedback for Room Temperature FELs. Josef Frisch Pohang, March 14, 2011 Beam Diagnostics, Low Level RF and Feedback for Room Temperature FELs Josef Frisch Pohang, March 14, 2011 Room Temperature / Superconducting Very different pulse structures RT: single bunch or short bursts

More information

State of the Art in RF Control

State of the Art in RF Control State of the Art in RF Control S. Simrock, DESY LINAC 2004, Lübeck Stefan Simrock DESY Outline RF System Architecture Requirements for RF Control RF Control Design Considerations Design Efforts Worldwide

More information

Design Considerations for Phase Reference Distribution

Design Considerations for Phase Reference Distribution Design Considerations for Reference Distribution Rihua Zeng, Anders J Johansson September 13, 2012 Abstract Coaxial cable based solution and optical fibre based solution are discussed in this note for

More information

SNS LLRF Design Experience and its Possible Adoption for the ILC

SNS LLRF Design Experience and its Possible Adoption for the ILC SNS LLRF Design Experience and its Possible Adoption for the ILC Brian Chase SNS - Mark Champion Fermilab International Linear Collider Workshop 11/28/2005 1 Why Consider the SNS System for ILC R&D at

More information

Software Requirements Specification for LLRF Applications at FLASH Version 1.0 Prepared by Zheqiao Geng MSK, DESY Nov. 06, 2009

Software Requirements Specification for LLRF Applications at FLASH Version 1.0 Prepared by Zheqiao Geng MSK, DESY Nov. 06, 2009 Software Specification for LLRF Applications at FLASH Version 1.0 Prepared by Zheqiao Geng MSK, DESY Nov. 06, 2009 Copyright 2009 by Zheqiao Geng. Any change of this document should be agreed by the development

More information

EVLA Memo 105. Phase coherence of the EVLA radio telescope

EVLA Memo 105. Phase coherence of the EVLA radio telescope EVLA Memo 105 Phase coherence of the EVLA radio telescope Steven Durand, James Jackson, and Keith Morris National Radio Astronomy Observatory, 1003 Lopezville Road, Socorro, NM, USA 87801 ABSTRACT The

More information

ALICE SRF SYSTEM COMMISSIONING EXPERIENCE A. Wheelhouse ASTeC, STFC Daresbury Laboratory

ALICE SRF SYSTEM COMMISSIONING EXPERIENCE A. Wheelhouse ASTeC, STFC Daresbury Laboratory ALICE SRF SYSTEM COMMISSIONING EXPERIENCE A. Wheelhouse ASTeC, STFC Daresbury Laboratory ERL 09 8 th 12 th June 2009 ALICE Accelerators and Lasers In Combined Experiments Brief Description ALICE Superconducting

More information

CEBAF waveguide absorbers. R. Rimmer for JLab SRF Institute

CEBAF waveguide absorbers. R. Rimmer for JLab SRF Institute CEBAF waveguide absorbers R. Rimmer for JLab SRF Institute Outline Original CEBAF HOM absorbers Modified CEBAF loads for FEL New materials for replacement loads High power loads for next generation FELs

More information

Predictions of LER-HER limits

Predictions of LER-HER limits Predictions of LER-HER limits PEP-II High Current Performance T. Mastorides, C. Rivetta, J.D. Fox, D. Van Winkle Accelerator Technology Research Div., SLAC 2e 34 Meeting, May 2, 27 Contents In this presentation

More information

- RF Master-Reference Update (F.Ludwig, H.Weddig - DESY, K.Czuba - TU Warsaw) - Beam Stability Update (C.Gerth, F.Ludwig, G.

- RF Master-Reference Update (F.Ludwig, H.Weddig - DESY, K.Czuba - TU Warsaw) - Beam Stability Update (C.Gerth, F.Ludwig, G. FLASH Meeting, 21/04/09 Beam Stability at FLASH - update F.Ludwig - DESY Content : - Motivation - RF Master-Reference Update (F.Ludwig, H.Weddig - DESY, K.Czuba - TU Warsaw) - Beam Stability Update (C.Gerth,

More information

Femtosecond-stability delivery of synchronized RFsignals to the klystron gallery over 1-km optical fibers

Femtosecond-stability delivery of synchronized RFsignals to the klystron gallery over 1-km optical fibers FEL 2014 August 28, 2014 THB03 Femtosecond-stability delivery of synchronized RFsignals to the klystron gallery over 1-km optical fibers Kwangyun Jung 1, Jiseok Lim 1, Junho Shin 1, Heewon Yang 1, Heung-Sik

More information

EUROFEL-Report-2006-DS EUROPEAN FEL Design Study

EUROFEL-Report-2006-DS EUROPEAN FEL Design Study EUROFEL-Report-2006-DS3-034 EUROPEAN FEL Design Study Deliverable N : D 3.8 Deliverable Title: RF Amplitude and Phase Detector Task: Author: DS-3 F.Ludwig, M.Hoffmann, M.Felber, Contract N : 011935 P.Strzalkowski,

More information

Microphonics. T. Powers

Microphonics. T. Powers Microphonics T. Powers What is microphonics? Microphonics is the time domain variation in cavity frequency driven by external vibrational sources. A 1.5 GHz structure 0.5 m long will change in frequency

More information

THE ORION PHOTOINJECTOR: STATUS and RESULTS

THE ORION PHOTOINJECTOR: STATUS and RESULTS THE ORION PHOTOINJECTOR: STATUS and RESULTS Dennis T. Palmer SLAC / ARDB ICFA Sardinia 4 July 2002 1. Introduction 2. Beam Dynamics Simulations 3. Photoinjector 1. RF Gun 2. Solenoidal Magnet 3. Diagnostics

More information

Digital Logic, Algorithms, and Functions for the CEBAF Upgrade LLRF System Hai Dong, Curt Hovater, John Musson, and Tomasz Plawski

Digital Logic, Algorithms, and Functions for the CEBAF Upgrade LLRF System Hai Dong, Curt Hovater, John Musson, and Tomasz Plawski Digital Logic, Algorithms, and Functions for the CEBAF Upgrade LLRF System Hai Dong, Curt Hovater, John Musson, and Tomasz Plawski Introduction: The CEBAF upgrade Low Level Radio Frequency (LLRF) control

More information

Effects of Intensity and Position Modulation On Switched Electrode Electronics Beam Position Monitor Systems at Jefferson Lab*

Effects of Intensity and Position Modulation On Switched Electrode Electronics Beam Position Monitor Systems at Jefferson Lab* JLAB-ACT--9 Effects of Intensity and Position Modulation On Switched Electrode Electronics Beam Position Monitor Systems at Jefferson Lab* Tom Powers Thomas Jefferson National Accelerator Facility Newport

More information

R100 Microphonics. Kirk Davis, Mike Drury, Leigh Harwood, Mark Wiseman, etc. Andrew Hutton

R100 Microphonics. Kirk Davis, Mike Drury, Leigh Harwood, Mark Wiseman, etc. Andrew Hutton R100 Microphonics Andrew Hutton Reporting on work by Kirk Davis, Mike Drury, Leigh Harwood, John Hogan, Kurt Hovater, Thomas Plawski, Mark Wiseman, etc. The Problem Vibrations of the superconducting cavities

More information

Commissioning of the ALICE SRF Systems at Daresbury Laboratory Alan Wheelhouse, ASTeC, STFC Daresbury Laboratory ESLS RF 1 st 2 nd October 2008

Commissioning of the ALICE SRF Systems at Daresbury Laboratory Alan Wheelhouse, ASTeC, STFC Daresbury Laboratory ESLS RF 1 st 2 nd October 2008 Commissioning of the ALICE SRF Systems at Daresbury Laboratory Alan Wheelhouse, ASTeC, STFC Daresbury Laboratory ESLS RF 1 st 2 nd October 2008 Overview ALICE (Accelerators and Lasers In Combined Experiments)

More information

LLRF Operation and Performance of the European XFEL. An overview

LLRF Operation and Performance of the European XFEL. An overview LLRF Operation and Performance of the European XFEL. An overview Mathieu Omet LLRF, Barcelona, 16.10.2017 Contents > Introduction > LLRF commissioning > Energy Reach > LLRF performance > Summary / Outlook

More information

Tutorial on Design of RF system for Indus Accelerator. Maherdra Lad Head, Radio Frequency Systems Division RRCAT, Indore

Tutorial on Design of RF system for Indus Accelerator. Maherdra Lad Head, Radio Frequency Systems Division RRCAT, Indore Tutorial on Design of RF system for Indus Accelerator Maherdra Lad Head, Radio Frequency Systems Division RRCAT, Indore Basic principle of RF Acceleration RF Power Amplifier The RF source supplies power

More information

Measurement and Analysis for Switchmode Power Design

Measurement and Analysis for Switchmode Power Design Measurement and Analysis for Switchmode Power Design Switched Mode Power Supply Measurements AC Input Power measurements Safe operating area Harmonics and compliance Efficiency Switching Transistor Losses

More information

W-band vector network analyzer based on an audio lock-in amplifier * Abstract

W-band vector network analyzer based on an audio lock-in amplifier * Abstract SLAC PUB 7884 July 1998 W-band vector network analyzer based on an audio lock-in amplifier * R. H. Siemann Stanford Linear Accelerator Center, Stanford University, Stanford CA 94309 Abstract The design

More information

AN ENVIRONMENTALLY HARDENED PRECISION QUARTZ

AN ENVIRONMENTALLY HARDENED PRECISION QUARTZ AN ENVIRONMENTALLY HARDENED PRECISION QUARTZ OSCILLATOR S.M. Bass, B.T. Milliren, and R.M. Garvey Frequency and Time Systems, Incorporated Beverly, Massachusetts 01 915 ABSTRACT Frequency and Time Systems

More information

Performance of the Prototype NLC RF Phase and Timing Distribution System *

Performance of the Prototype NLC RF Phase and Timing Distribution System * SLAC PUB 8458 June 2000 Performance of the Prototype NLC RF Phase and Timing Distribution System * Josef Frisch, David G. Brown, Eugene Cisneros Stanford Linear Accelerator Center, Stanford University,

More information

Wafer-Level Calibration & Verification up to 750 GHz. Choon Beng Sia, Ph.D. Mobile:

Wafer-Level Calibration & Verification up to 750 GHz. Choon Beng Sia, Ph.D.   Mobile: Wafer-Level Calibration & Verification up to 750 GHz Choon Beng Sia, Ph.D. Email: Choonbeng.sia@cmicro.com Mobile: +65 8186 7090 2016 Outline LRRM vs SOLT Calibration Verification Over-temperature RF calibration

More information

Cavity Field Control - RF Field Controller. LLRF Lecture Part3.3 S. Simrock, Z. Geng DESY, Hamburg, Germany

Cavity Field Control - RF Field Controller. LLRF Lecture Part3.3 S. Simrock, Z. Geng DESY, Hamburg, Germany Cavity Field Control - RF Field Controller LLRF Lecture Part3.3 S. Simrock, Z. Geng DESY, Hamburg, Germany Content Introduction to the controller Control scheme selection In-phase and Quadrature (I/Q)

More information

SWR/Return Loss Measurements Using System IIA

SWR/Return Loss Measurements Using System IIA THE GLOBAL SOURCE FOR PROVEN TEST SWR/Return Loss Measurements Using System IIA SWR/Return Loss Defined Both SWR and Return Loss are a measure of the divergence of a microwave device from a perfect impedance

More information

Agilent 8703B Lightwave Component Analyzer Technical Specifications. 50 MHz to GHz modulation bandwidth

Agilent 8703B Lightwave Component Analyzer Technical Specifications. 50 MHz to GHz modulation bandwidth Agilent 8703B Lightwave Component Analyzer Technical Specifications 50 MHz to 20.05 GHz modulation bandwidth 2 The 8703B lightwave component analyzer is a unique, general-purpose instrument for testing

More information

Installation and Characterization of the Advanced LIGO 200 Watt PSL

Installation and Characterization of the Advanced LIGO 200 Watt PSL Installation and Characterization of the Advanced LIGO 200 Watt PSL Nicholas Langellier Mentor: Benno Willke Background and Motivation Albert Einstein's published his General Theory of Relativity in 1916,

More information

Overview of ERL Projects: SRF Issues and Challenges. Matthias Liepe Cornell University

Overview of ERL Projects: SRF Issues and Challenges. Matthias Liepe Cornell University Overview of ERL Projects: SRF Issues and Challenges Matthias Liepe Cornell University Overview of ERL projects: SRF issues and challenges Slide 1 Outline Introduction: SRF for ERLs What makes it special

More information

Making sense of electrical signals

Making sense of electrical signals Making sense of electrical signals Our thanks to Fluke for allowing us to reprint the following. vertical (Y) access represents the voltage measurement and the horizontal (X) axis represents time. Most

More information

Borut Baricevic. Libera LLRF. 17 September 2009

Borut Baricevic. Libera LLRF. 17 September 2009 Borut Baricevic Libera LLRF borut.baricevic@i-tech.si 17 September 2009 Outline Libera LLRF introduction Libera LLRF system topology Signal processing structure GUI and signal acquisition RF system diagnostics

More information

RF-Based Detector for Measuring Fiber Length Changes with Sub-5 Femtosecond Long-Term Stability.

RF-Based Detector for Measuring Fiber Length Changes with Sub-5 Femtosecond Long-Term Stability. RF-Based Detector for Measuring Fiber Length Changes with Sub-5 Femtosecond Long-Term Stability. J. Zemella 1, V. Arsov 1, M. K. Bock 1, M. Felber 1, P. Gessler 1, K. Gürel 3, K. Hacker 1, F. Löhl 1, F.

More information

CERN (The European Laboratory for Particle Physics)

CERN (The European Laboratory for Particle Physics) 462 IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, VOL. 48, NO. 2, APRIL 1999 The Measurement Challenge of the LHC Project Gunnar Fernqvist Abstract In 2005, CERN is planning to commission its next

More information

Update 5/19/11. Kirk Davis, Mike Drury, Leigh Harwood, Mark Wiseman, etc. Andrew Hutton

Update 5/19/11. Kirk Davis, Mike Drury, Leigh Harwood, Mark Wiseman, etc. Andrew Hutton R100 Microphonics i Update 5/19/11 Andrew Hutton Reporting on work by Kirk Davis, Mike Drury, Leigh Harwood, John Hogan, Curt Hovater, Thomas Plawski, Mark Wiseman, etc. The Problem Vibrations of the superconducting

More information

Electron Beam Properties and Instrumentation MOLLER Director s Review, Jan. 14, 2010 Mark Pitt, Virginia Tech

Electron Beam Properties and Instrumentation MOLLER Director s Review, Jan. 14, 2010 Mark Pitt, Virginia Tech Electron Beam Properties and Instrumentation MOLLER Director s Review, Jan. 14, 2010 Mark Pitt, Virginia Tech This talk will focus on the electron beam properties and beam instrumentation requirements

More information

RF-based Synchronization of the Seed and Pump-Probe Lasers to the Optical Synchronization System at FLASH

RF-based Synchronization of the Seed and Pump-Probe Lasers to the Optical Synchronization System at FLASH RF-based Synchronization of the Seed and Pump-Probe Lasers to the Optical Synchronization System at FLASH Introduction to the otical synchronization system and concept of RF generation for locking of Ti:Sapphire

More information

Vibration studies of a superconducting accelerating

Vibration studies of a superconducting accelerating Vibration studies of a superconducting accelerating module at room temperature and at 4.5 K Ramila Amirikas, Alessandro Bertolini, Wilhelm Bialowons Vibration studies on a Type III cryomodule at room temperature

More information

EVLA Memo # 194 EVLA Ka-band Receiver Down Converter Module Harmonics: The Mega-Birdie at MHz

EVLA Memo # 194 EVLA Ka-band Receiver Down Converter Module Harmonics: The Mega-Birdie at MHz EVLA Memo # 194 EVLA Ka-band Receiver Down Converter Module Harmonics: The Mega-Birdie at 29440 MHz R. Selina, E. Momjian, W. Grammer, J. Jackson NRAO February 5, 2016 Abstract Observations carried out

More information

Status of the European XFEL Accelerator Construction Project. Reinhard Brinkmann, DESY

Status of the European XFEL Accelerator Construction Project. Reinhard Brinkmann, DESY Status of the European XFEL Accelerator Construction Project Reinhard Brinkmann, DESY European XFEL Introduction Some specifications Photon energy 0.3-24 kev Pulse duration ~ 10-100 fs Pulse energy few

More information

A Synchrotron Phase Detector for the Fermilab Booster

A Synchrotron Phase Detector for the Fermilab Booster FERMILAB-TM-2234 A Synchrotron Phase Detector for the Fermilab Booster Xi Yang and Rene Padilla Fermi National Accelerator Laboratory Box 5, Batavia IL 651 Abstract A synchrotron phase detector is diagnostic

More information

SPHERICAL NEAR-FIELD SELF-COMPARISON MEASUREMENTS

SPHERICAL NEAR-FIELD SELF-COMPARISON MEASUREMENTS SPHERICAL NEAR-FIELD SELF-COMPARISON MEASUREMENTS Greg Hindman, Allen C. Newell Nearfield Systems Inc. 1973 Magellan Dr. Torrance, CA 952 ABSTRACT Spherical near-field measurements require an increased

More information

2302 Battery Simulator 2306, 2306-PJ Battery/Charger Simulators

2302 Battery Simulator 2306, 2306-PJ Battery/Charger Simulators Ultrafast response to transient load currents Choice of single- or dualchannel supplies Optimized for development and testing of battery-powered devices Variable output resistance for simulating battery

More information

PLL Synchronizer User s Manual / Version 1.0.6

PLL Synchronizer User s Manual / Version 1.0.6 PLL Synchronizer User s Manual / Version 1.0.6 AccTec B.V. Den Dolech 2 5612 AZ Eindhoven The Netherlands phone +31 (0) 40-2474321 / 4048 e-mail AccTecBV@tue.nl Contents 1 Introduction... 3 2 Technical

More information

MASSACHUSETTS INSTITUTE OF TECHNOLOGY HAYSTACK OBSERVATORY WESTFORD, MASSACHUSETTS

MASSACHUSETTS INSTITUTE OF TECHNOLOGY HAYSTACK OBSERVATORY WESTFORD, MASSACHUSETTS To: From: EDGES MEMO #104 MASSACHUSETTS INSTITUTE OF TECHNOLOGY HAYSTACK OBSERVATORY WESTFORD, MASSACHUSETTS 01886 January 14, 2013 Telephone: 781-981-5400 Fax: 781-981-0590 EDGES Group Alan E.E. Rogers

More information

Physics Requirements Document Document Title: SCRF 1.3 GHz Cryomodule Document Number: LCLSII-4.1-PR-0146-R0 Page 1 of 7

Physics Requirements Document Document Title: SCRF 1.3 GHz Cryomodule Document Number: LCLSII-4.1-PR-0146-R0 Page 1 of 7 Document Number: LCLSII-4.1-PR-0146-R0 Page 1 of 7 Document Approval: Originator: Tor Raubenheimer, Physics Support Lead Date Approved Approver: Marc Ross, Cryogenic System Manager Approver: Jose Chan,

More information

Progress in High Gradient Accelerator Research at MIT

Progress in High Gradient Accelerator Research at MIT Progress in High Gradient Accelerator Research at MIT Presented by Richard Temkin MIT Physics and Plasma Science and Fusion Center May 23, 2007 MIT Accelerator Research Collaborators MIT Plasma Science

More information

Baseband simulation model of the vector rf voltage control system for the J-PARC RCS

Baseband simulation model of the vector rf voltage control system for the J-PARC RCS Journal of Physics: Conference Series PAPER OPEN ACCESS Baseband simulation model of the vector rf voltage control system for the J-PARC RCS To cite this article: Fumihiko Tamura et al 2018 J. Phys.: Conf.

More information

INTRA-TRAIN LONGITUDINAL FEEDBACK FOR BEAM STABILIZATION AT FLASH

INTRA-TRAIN LONGITUDINAL FEEDBACK FOR BEAM STABILIZATION AT FLASH INTRA-TRAIN LONGITUDINAL FEEDBACK FOR BEAM STABILIZATION AT FLASH W. Koprek*, C. Behrens, M. K. Bock, M. Felber, P. Gessler, K. Hacker, H. Schlarb, C. Schmidt, B. Steffen, S. Wesch, DESY, Hamburg, Germany

More information

A TECHNIQUE TO EVALUATE THE IMPACT OF FLEX CABLE PHASE INSTABILITY ON mm-wave PLANAR NEAR-FIELD MEASUREMENT ACCURACIES

A TECHNIQUE TO EVALUATE THE IMPACT OF FLEX CABLE PHASE INSTABILITY ON mm-wave PLANAR NEAR-FIELD MEASUREMENT ACCURACIES A TECHNIQUE TO EVALUATE THE IMPACT OF FLEX CABLE PHASE INSTABILITY ON mm-wave PLANAR NEAR-FIELD MEASUREMENT ACCURACIES Daniël Janse van Rensburg Nearfield Systems Inc., 133 E, 223rd Street, Bldg. 524,

More information

Waveguide Arc Restrike Test Results Abstract Background

Waveguide Arc Restrike Test Results Abstract Background Waveguide Arc Restrike Test Results Tom Powers, Doug Curry, Kirk Davis, Larry King, and Mike Tiefenback Thomas Jefferson National Accelerator Facility (Test dates July 6, 2004 through September 2, 2004)

More information

A HIGH PRECISION QUARTZ OSCILLATOR WITH PERFORMANCE COMPARABLE TO RUBIDIUM OSCILLATORS IN MANY RESPECTS

A HIGH PRECISION QUARTZ OSCILLATOR WITH PERFORMANCE COMPARABLE TO RUBIDIUM OSCILLATORS IN MANY RESPECTS A HIGH PRECISION QUARTZ OSCILLATOR WITH PERFORMANCE COMPARABLE TO RUBIDIUM OSCILLATORS IN MANY RESPECTS Manish Vaish MTI-Milliren Technologies, Inc. Two New Pasture Road Newburyport, MA 195 Abstract An

More information

LLRF Plans for SMTF. Ruben Carcagno (Fermilab) Nigel Lockyer (University of Pennsylvania) Thanks to DESY, PISA, KEK, Fermilab, SLAC Colleagues

LLRF Plans for SMTF. Ruben Carcagno (Fermilab) Nigel Lockyer (University of Pennsylvania) Thanks to DESY, PISA, KEK, Fermilab, SLAC Colleagues LLRF Plans for SMTF Ruben Carcagno (Fermilab) Nigel Lockyer (University of Pennsylvania) Thanks to DESY, PISA, KEK, Fermilab, SLAC Colleagues Outline Near-term (< 1.5 years) SMTF LLRF plan Long-term (>

More information

Preliminary Users Manual for the Self Contained Return Loss and Cable Fault Test Set with Amplified Wideband Noise Source Copyright 2001 Bryan K.

Preliminary Users Manual for the Self Contained Return Loss and Cable Fault Test Set with Amplified Wideband Noise Source Copyright 2001 Bryan K. Preliminary Users Manual for the Self Contained Return Loss and Cable Fault Test Set with Amplified Wideband Noise Source Copyright 2001 Bryan K. Blackburn Self Contained Test Set Test Port Regulated 12

More information

HOM Based Diagnostics at the TTF

HOM Based Diagnostics at the TTF HOM Based Diagnostics at the TTF Nov 14, 2005 Josef Frisch, Nicoleta Baboi, Linda Hendrickson, Olaf Hensler, Douglas McCormick, Justin May, Olivier Napoly, Rita Paparella, Marc Ross, Claire Simon, Tonee

More information

2005 Modelithics Inc.

2005 Modelithics Inc. Precision Measurements and Models You Trust Modelithics, Inc. Solutions for RF Board and Module Designers Introduction Modelithics delivers products and services to serve one goal accelerating RF/microwave

More information

Communication Circuit Lab Manual

Communication Circuit Lab Manual German Jordanian University School of Electrical Engineering and IT Department of Electrical and Communication Engineering Communication Circuit Lab Manual Experiment 3 Crystal Oscillator Eng. Anas Alashqar

More information

Characterization of Magnet Noise in Superconducting Magnets When Charging the Magnetic Field in Unidirectional Steps

Characterization of Magnet Noise in Superconducting Magnets When Charging the Magnetic Field in Unidirectional Steps Overview MPMS Service Note 11- Characterization of Magnet Noise in Superconducting Magnets When Charging the Magnetic Field in Unidirectional Steps This service note outlines the effects of magnet noise

More information

A Method for Gain over Temperature Measurements Using Two Hot Noise Sources

A Method for Gain over Temperature Measurements Using Two Hot Noise Sources A Method for Gain over Temperature Measurements Using Two Hot Noise Sources Vince Rodriguez and Charles Osborne MI Technologies: Suwanee, 30024 GA, USA vrodriguez@mitechnologies.com Abstract P Gain over

More information

Synchronization Overview

Synchronization Overview Synchronization Overview S. Simrock, DESY ERL Workshop 2005 Stefan Simrock DESY What is Synchronization Outline Synchronization Requirements for RF, Laser and Beam Timing stability RF amplitude and phase

More information

AUTOMOTIVE CURRENT TRANSDUCER HAH3DR 700-S02

AUTOMOTIVE CURRENT TRANSDUCER HAH3DR 700-S02 AUTOMOTIVE CURRENT TRANSDUCER Introduction The HAH3DR family, a tri-phase tranducer is for the electronic measurement of DC, AC or pulsed s in high power automotive applications with galvanic isolation

More information

Linac Coherent Light Source (LCLS) Low Level RF Status LCLS FAC. October 30, 2007

Linac Coherent Light Source (LCLS) Low Level RF Status LCLS FAC. October 30, 2007 Linac Coherent Light Source (LCLS) Low Level RF Status LCLS Emma LCLS RF Gun, L0, and L1 Emma Dual Feed L0A L0B L0A 57MV 19MV/m L0B 72MV 24MV/m Off Axis Injector Vault Injector Transverse Accelerator 55cm

More information

First results of a high performance optically-pumped cesium beam clock

First results of a high performance optically-pumped cesium beam clock First results of a high performance optically-pumped cesium beam clock Berthoud Patrick, Chief Scientist Time & Frequency Workshop on Synchronization and Timing Systems, WSTS 2016, San Jose CA, USA, June

More information

Slide Title. Bulleted Text

Slide Title. Bulleted Text Slide Title 1 Slide Outline Title Brief view of the C-AD Complex Review of the RHIC LLRF Upgrade Platform Generic Implementation of a Feedback Loop RHIC Bunch by Bunch Longitudinal Damper Cavity Controller

More information

Linear Particle Accelerator Control Performance

Linear Particle Accelerator Control Performance Linear Particle Accelerator Control Performance 2007 ExpertTune-TiPS Conference April 17-19, 2007 Austin, TX Johnny Tang Overview of the Spallation Neutron Source Accelerator J. Tang 2 Overview of the

More information

Timing Noise Measurement of High-Repetition-Rate Optical Pulses

Timing Noise Measurement of High-Repetition-Rate Optical Pulses 564 Timing Noise Measurement of High-Repetition-Rate Optical Pulses Hidemi Tsuchida National Institute of Advanced Industrial Science and Technology 1-1-1 Umezono, Tsukuba, 305-8568 JAPAN Tel: 81-29-861-5342;

More information

Design, Development and Testing of RF Window for C band 250 kw CW Power Klystron

Design, Development and Testing of RF Window for C band 250 kw CW Power Klystron Available online www.ejaet.com European Journal of Advances in Engineering and Technology, 2016, 3(6): 26-30 Research Article ISSN: 2394-658X Design, Development and Testing of RF Window for C band 250

More information

325 to 500 GHz Vector Network Analyzer System

325 to 500 GHz Vector Network Analyzer System 325 to 500 GHz Vector Network Analyzer System By Chuck Oleson, Tony Denning and Yuenie Lau OML, Inc. Abstract - This paper describes a novel and compact WR-02.2 millimeter wave frequency extension transmission/reflection

More information

Making sense of electrical signals

Making sense of electrical signals APPLICATION NOTE Making sense of electrical signals Devices that convert electrical power to mechanical power run the industrial world, including pumps, compressors, motors, conveyors, robots and more.

More information

RF Characterization Report

RF Characterization Report SMA-J-P-H-ST-MT1 Mated with: RF316-01SP1-01BJ1-0305 Description: 50-Ω SMA Board Mount Jack, Mixed Technology Samtec, Inc. 2005 All Rights Reserved Table of Contents Introduction...1 Product Description...1

More information

Experiment-4 Study of the characteristics of the Klystron tube

Experiment-4 Study of the characteristics of the Klystron tube Experiment-4 Study of the characteristics of the Klystron tube OBJECTIVE To study the characteristics of the reflex Klystron tube and to determine the its electronic tuning range EQUIPMENTS Klystron power

More information

AUTOMOTIVE CURRENT TRANSDUCER HAH3DR 700-S00

AUTOMOTIVE CURRENT TRANSDUCER HAH3DR 700-S00 AUTOMOTIVE CURRENT TRANSDUCER HAH3DR 700-S00 Introduction The HAH3DR family, a tri-phase tranducer is for the electronic measurement of DC, AC or pulsed s in high power automotive applications with galvanic

More information

Note on the LCLS Laser Heater Review Report

Note on the LCLS Laser Heater Review Report Note on the LCLS Laser Heater Review Report P. Emma, Z. Huang, C. Limborg, J. Schmerge, J. Wu April 15, 2004 1 Introduction This note compiles some initial thoughts and studies motivated by the LCLS laser

More information

Chapter IX Using Calibration and Temperature Compensation to improve RF Power Detector Accuracy By Carlos Calvo and Anthony Mazzei

Chapter IX Using Calibration and Temperature Compensation to improve RF Power Detector Accuracy By Carlos Calvo and Anthony Mazzei Chapter IX Using Calibration and Temperature Compensation to improve RF Power Detector Accuracy By Carlos Calvo and Anthony Mazzei Introduction Accurate RF power management is a critical issue in modern

More information

Battery Simulator Battery/Charger Simulators

Battery Simulator Battery/Charger Simulators Test Equipment Depot - 800.517.8431-99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com 2302, 2302-PJ, Ultrafast response to transient load currents Choice of single- or dualchannel supplies

More information

Continuous-Wave (CW) Single-Frequency IR Laser. NPRO 125/126 Series

Continuous-Wave (CW) Single-Frequency IR Laser. NPRO 125/126 Series Continuous-Wave (CW) Single-Frequency IR Laser NPRO 125/126 Series www.lumentum.com Data Sheet The Lumentum NPRO 125/126 diode-pumped lasers produce continuous-wave (CW), singlefrequency output at either

More information

MIMO-LTI Feedback Controller Design -Status report-

MIMO-LTI Feedback Controller Design -Status report- MIMO-LTI Feedback Controller Design -Status report- Christian Schmidt Deutsches Elektronen Synchrotron Technische Universitaet Hamburg Harburg FLASH Seminar 4/1/28 Outline Current RF Feedback System MIMO

More information

ENGLISH. General features 2. Technical features 3. Installation 10. Preliminary operations 12. Front panel description 13.

ENGLISH. General features 2. Technical features 3. Installation 10. Preliminary operations 12. Front panel description 13. LDI 35 OPERATING INSTRUCTIONS General features 2 Technical features 3 Installation 10 Preliminary operations 12 Front panel description 13 Operating mode 15 Important: We suggest you keep the original

More information

Vector Network Analyzer Application note

Vector Network Analyzer Application note Vector Network Analyzer Application note Version 1.0 Vector Network Analyzer Introduction A vector network analyzer is used to measure the performance of circuits or networks such as amplifiers, filters,

More information

MAKING TRANSIENT ANTENNA MEASUREMENTS

MAKING TRANSIENT ANTENNA MEASUREMENTS MAKING TRANSIENT ANTENNA MEASUREMENTS Roger Dygert, Steven R. Nichols MI Technologies, 1125 Satellite Boulevard, Suite 100 Suwanee, GA 30024-4629 ABSTRACT In addition to steady state performance, antennas

More information

RF STATUS OF SUPERCONDUCTING MODULE DEVELOPMENT SUITABLE FOR CW OPERATION: ELBE CRYOSTATS

RF STATUS OF SUPERCONDUCTING MODULE DEVELOPMENT SUITABLE FOR CW OPERATION: ELBE CRYOSTATS RF STATUS OF SUPERCONDUCTING MODULE DEVELOPMENT SUITABLE FOR CW OPERATION: ELBE CRYOSTATS J. Teichert, A. Büchner, H. Büttig, F. Gabriel, P. Michel, K. Möller, U. Lehnert, Ch. Schneider, J. Stephan, A.

More information

Evaluation of HOM Coupler Probe Heating by HFSS Simulation

Evaluation of HOM Coupler Probe Heating by HFSS Simulation G. Wu, H. Wang, R. A. Rimmer, C. E. Reece Abstract: Three different tip geometries in a HOM coupler on a CEBAF Upgrade Low Loss cavity have been evaluated by HFSS simulation to understand the tip surface

More information

Chapter 6. Temperature Effects

Chapter 6. Temperature Effects Chapter 6. Temperature Effects 6.1 Introduction This chapter documents the investigation into temperature drifts that can cause a receiver clock bias even when a stable reference is used. The first step

More information

EVLA Memo 108 LO/IF Phase Dependence on Antenna Elevation

EVLA Memo 108 LO/IF Phase Dependence on Antenna Elevation EVLA Memo 108 LO/IF Phase Dependence on Antenna Elevation Abstract K. Morris, J. Jackson, V. Dhawan June 18, 2007 EVLA test observations revealed interferometric phase changes that track EVLA antenna elevation

More information

An Explanation of GP:50 Performance Specifications

An Explanation of GP:50 Performance Specifications Introduction: The performance specifications of GP:50 pressure transducers are typically identified on the sales literature or datasheet for the particular model type. Performance specifications include

More information