DEFIBRILLATORS often use a small-signal ac measurement

Size: px
Start display at page:

Download "DEFIBRILLATORS often use a small-signal ac measurement"

Transcription

1 1858 IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, VOL. 61, NO. 6, JUNE 214 Bioimpedance-Based Respiration Monitoring With a Defibrillator Ørjan G. Martinsen, Senior Member, IEEE, Bernt Nordbotten, Sverre Grimnes, Helge Fossan, and Joar Eilevstjønn, Member, IEEE Abstract Defibrillators often use an ac measurement to ensure safe electrode placement. Some defibrillators also utilize impedance measurements for monitoring. This paper investigates how such measurements can be optimized for high respiration sensitivity and finds that it is beneficial to add two extra electrodes in addition to the two defibrillator electrodes. This four electrode setup increases sensitivity and also allows respiration measurements at lower frequencies than the two electrode setup. Index Terms Bioimpedance, defibrillator, respiration, sensitivity,. I. INTRODUCTION DEFIBRILLATORS often use a small-signal ac measurement current in the range khz in order to ensure good electrode placement. In this frequency range, the impedance between defibrillator pads on a human body is predominantly resistive and normally matches the impedance opposing the defibrillator current quite well. Typical impedances of an Apex-Sternum placement of defibrillator pads are in the 25 2 Ω range, and defibrillation is normally restricted to an impedance range close to this. Some defibrillators have implemented extra functionality based on the measured impedance; most common is monitoring which is done by measuring the variation of the base impedance caused by. The modulation is normally below ±3 Ω and the signal can be obscured by motion artifacts. It is believed that detection/monitoring can be improved if the impedance measurement system was more optimized for this task. It is the purpose of this study to find whether the respirational plethysmographic signal can be increased by using a Manuscript received December, 213; revised February 9, 214; accepted February 2, 214. Date of publication February 28, 214; date of current version May 15, 214. Asterisk indicates corresponding author. Ø. G. Martinsen is with the Department of Physics, University of Oslo, Oslo 316, Norway, and also with the Department of Clinical and Biomedical Engineering, Oslo University Hospital, Oslo 424, Norway ( ogm@fys.uio.no). B. Nordbotten is with the Department of Physics, University of Oslo, Oslo 316, Norway ( berntn@gmail.com). S. Grimnes is with the Department of Physics, University of Oslo, Oslo 316, Norway, and also with the Department of Clinical and Biomedical Engineering, Oslo University Hospital, Oslo 424, Norway ( sverre.grimnes@fys.uio.no) H. Fossan and J. Eilevstjønn are with Laerdal Medical, Stavanger 42, Norway ( helge.fossan@laerdal.no; joar.eilevstjonn@laerdal.no). Digital Object Identifier.19/TBME four-electrode system [1], and to compare the admittance and impedance parameters. II. MATERIALS AND METHODS Ten healthy male volunteers participated in this study. Average age was 3 years (21 4). They all gave informed consent and the study was approved by the regional ethics committee. All measurements were conducted in a controlled laboratory with 22 C and 3% RH. A. Impedance Measurements All impedance measurements were done on healthy male volunteers, using a Solartron 126/1294 combination with a controlled ac of 3 mv rms. At maximum inspiration or maximum expiration, respectively, one complex spectrum took about 45 s during which the test held his breath. (Note that this time could have been reduced, e.g., by utilizing time domain measurements with multisine or chirp signal excitation [2].) B. Defibrillator Measurements A Philips HeartStart MRx, which has a controlled ac of 25 μa rms at 32 khz for impedance measurement, was used for the defibrillator measurements. The MRx is an advanced monitor/defibrillator device with 12 lead ECG, blood pressure, ETCO2, pulse-oximetry, networking capability, and CPR feedback (QCPR) functionality. C. Electrode Positioning The Apex-Sternum electrode placement is shown in Fig. 1. The defibrillator electrodes Philips Medical DP2/DP6 and the small ECG electrodes are Ambu Blue Sensor Q A. The defibrillator electrodes were only used for the defibrillator measurements and for the Solatron two-electrode setup. The Blue Sensor electrodes were used as voltage pick-up electrodes in the four-electrode setup. The edge-to-edge distance between current injecting and voltage pick-up electrodes is about 1 cm. D. Forced Breathing and Generation of Motion Artifacts For the time-series measurements, it was measured both during spontaneous breathing where the test subjects breathed freely, and also during forced breathing where the subjects respiration was controlled with an even rate. For the forced breathing a manually operated calibration syringe with a volume of 75 ml, together with a valve system which allowed the IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission. See standards/publications/rights/index.html for more information.

2 MARTINSEN et al.: BIOIMPEDANCE-BASED RESPIRATION MONITORING WITH A DEFIBRILLATOR 1859 TABLE I DEFIBRILLATOR MEASUREMENT RESULTS Fig. 1. Electrode placement for respiration monitoring. Large defibrillator electrodes used in the two-electrode setup and as current injecting electrodes in the four-electrode setup. Small ECG electrodes used as potential pick-up electrodes in the four-electrode setup. inhalation to be controlled while allowing for free exhalation, was used. The pump was fitted with a single use mouth piece for each test subject. Motion artifacts can be due to compression-induced changes in either the electrode materials or in the measured tissue [3], and different electrodes and electrode systems may be more or less prone to producing motion artifacts [4]. In this study, motion artifacts were generated by pushing on the electrodes; either by hand on the lower electrode or using the MRx CPR sensor as an interface between the hand and upper electrode. E. Protocol The following protocol was followed for all the ten test subjects: 1) The sat comfortably in a chair with no clothing on the upper body. After a few minutes of acclimatization, the electrodes were positioned. 2) The defibrillator was then connected to the two defibrillator electrodes and impedance measurements were done continuously through the following sequence: 3 s of spontaneous breathing, 6 s of forced breathing, a few seconds of spontaneous breathing, and then 6 s of forced breathing while repeatedly pressing on one of the defibrillator electrodes. 3) The was then instead connected to the Solartron 126/1294 system in a two-electrode setup, and the above sequence was repeated, using a frequency of 32 khz. The was then asked to breathe maximally in and keep his breath during a frequency scan from Hz to 1 MHz. Subsequently, the breathed maximally out and kept his breath while a new frequency scan was conducted. 4) Lastly the same procedure, including the frequency scans, was repeated with a four-electrode setup % % % % % % % % % % % TABLE II TWO-ELECTRODE RESISTANCE TIME SERIES RESULTS % % % % % % % % % % % III. RESULTS A. Defibrillator Measurements The results from the measured time-series with the defibrillator can be seen in Table I. The table shows the total number of s, the mean of all measured values in the time series (mean ), the mean of the impedance change for all s (mean ) with standard deviation, and the relative. The mean relative was found to be 1.8%. B. Solatron 126/1294 Measurements For the time series measurements, only the 6-s time period with forced breathing was since during this period the test subject has a fairly even and reproducible respiration rate. 1) Two-Electrode Setup: The results from the two-electrode time series measurements are shown in Table II. Only resistance values are. The mean relative was found to be 2.5%. A frequency scan with the two-electrode setup is shown as mean values in Fig. 2, both at maximum inspiration and expiration. The relative difference between maximum expiration and inspiration is shown in Figs. 3 and 4 as percent change in impedance parameters (resistance and reactance) and admittance parameters (conductance and susceptance), respectively.

3 186 IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, VOL. 61, NO. 6, JUNE 214 TABLE III FOUR-ELECTRODE RESISTANCE TIME SERIES RESULTS Fig. 2. Impedance modulus and phase spectra (mean values) for maximum inspiration measured with the two-electrode set-up shown in Fig. 1. Dashed lines are maximum expiration values. Error bars show standard deviation. Impedance % % % % % % % % % % % Impedance Phase angle Phase angle [deg.] Fig. 3. Two-electrode impedance measurements. Percent impedance increase from maximum expiration to maximum inspiration held breath values. Dashed line indicates extreme values due to measurement uncertainty/error. Admi ance increase [%] Conductance Susceptance 1 1 Fig. 4. Percent admittance increase (or decrease when negative values) from maximum expiration to maximum inspiration held breath values for twoelectrode measurements. Dashed line indicates extreme values due to measurement uncertainty/error. Fig. 5. Transfer impedance, mean values. Modulus and phase spectra measured with the four-electrode method. Continuous lines are maximum inspiration values and dashed lines are maximum expiration values. Error bars show standard deviation. C. Four-Electrode Setup The results from the four-electrode time series measurements are shown in Table III. Only resistance values are. The mean relative was found to be 4.%. The results from the frequency scans using a four-electrode system [1] are shown as mean values in Figs Data from one of the ten test subjects have been removed, since the lowfrequency reactance values for maximum expiration were more than ten times higher than for any of the other subjects, and the corresponding resistance about four times higher, hence indicating loose electrodes. 1) Noise Measurements: Fig. 8 shows a typical time course of the measured resistance using the two-electrode system. The last part of the time series shows noise caused by repeatedly pressing one of the defibrillator electrodes. By visual inspection of the curves, no significant difference in the noise contribution between the two- and four-electrode measurements was found. IV. DISCUSSION Figs. 2 and 3 support the choice of 32 khz for the impedance measurements in the defibrillator. However, from the difference

4 MARTINSEN et al.: BIOIMPEDANCE-BASED RESPIRATION MONITORING WITH A DEFIBRILLATOR 1861 Impedance increase [%] Resistance Reactance 1 1 Fig. 6. Percent transfer impedance increase from maximum expiration to maximum inspiration held breath values. Four-electrode method. Dashed line indicates extreme values due to measurement uncertainty/error. Fig. 7. Percent transfer admittance increase from maximum expiration to maximum inspiration held breath values. Four-electrode method. Dashed line indicates extreme values due to measurement uncertainty/error. Fig. 8. Typical time course of resistance at 32 khz measured with twoelectrode setup. in the mean relative between the twoand four-electrode setup using the Solartron 126/1294 system, there is an indication that the four-electrode setup is better suited for picking up the small resistance changes due to respiration. The differences between the two- and four-electrode setup will now be further discussed. In the two-electrode setup, the reactance increase diverges to a very high value (359%) at 46 khz. This is due to a very small phase angle as the phase is actually crossing zero in that frequency range. At this frequency, the data did not meet the normality criterion and a one-sample Sign for the distribution of relative reactance increase at 46 khz revealed that the median increase was not significantly different from zero (p = 1.). Hence, what appears as very high sensitivity for the reactance and susceptance at high frequencies is actually an artifact due to small (random noise) changes in very small numbers. The two-electrode results are clearly dominated by skin impedance (23 Ω at Hz) [5] in the frequency range Hz, Fig. 2, which of course is not very dependent on respiration, Figs. 3 and 4. At higher frequencies, the results are determined by deeper tissue layers and the sensitivity to respiration increases. The highest sensitivity can be attained by measuring resistance (or equally well impedance since the phase angle is very small) at frequencies around khz or higher where a relative increase of around 18% can be achieved, or alternatively by measuring susceptance in the frequency range 5 khz where the decrease is 2%. The same problem with low-phase angle as described with the two-electrode system is present for low frequencies with the four-electrode system. Extremely low ( 122%) and high (57%) average relative increases were found in the reactance at the two lowest frequencies, respectively. At these frequencies, the data did not meet the normality criterion and a one-sample Sign for the distribution of relative reactance increase at and 215 Hz revealed that the median increase was not significantly different from zero (p =.51 for both). In fact the reactance increase was not significantly different from zero for any frequency below khz. For the susceptance, the extreme value at Hz ( 1%) is not significantly different from zero (p =.18) whereas values at 215 Hz (2%) and 464 Hz ( 21%) are marginally different from zero (p =.4). The four-electrode method measures transfer impedance [6], resulting in very low values around 4 Ω, which are clearly not influenced by skin impedance. An appreciable respiration sensitivity of around 18% (see Figs. 6 and 7) is therefore possible even at Hz. As for the two-electrode measurements, it is also found here that resistance or impedance measurements at high frequencies (above khz) yield good sensitivity (3% 35%). The highest sensitivity, however, is found for susceptance measurements between 2 and khz, where a relative increase of 35% was achieved. For the four-electrode measurements, using the impedance modulus yields basically the same results as using resistance, since the phase angle is very small. For the two-electrode measurements, the results for using only the impedance modulus give a mean relative that is about 1% lower than for the resistance throughout the frequency range. Only male subjects were measured in this study. This may possibly influence the conclusions about the sensitivity of the monitoring of the respiration, since the thoracic impedance is influenced by the female chest [7]. No significant difference in the noise contribution between the two- and four-electrode measurements was found. However, it should be noted that a four-electrode system in some cases

5 1862 IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, VOL. 61, NO. 6, JUNE 214 may produce more noise than a two-electrode system, e.g., in the case of drying voltage pick-up electrodes in combination with a very high input impedance differential amplifier. V. CONCLUSION It has been shown that the addition of two extra electrodes is beneficial for the measurement of respiration. With the two electrode setup, the increase in resistance due to respiration is only 1.38% at Hz and increases to 18.18% at 1 MHz, while for the four electrode setup the increase in resistance is 18.4% at Hz and increases to 34.16% at 1 MHz. It is furthermore concluded that with the four-electrode setup, it is also possible to measure the respiration at low frequencies, whereas for the two-electrode setup the measurements at low frequencies are dominated by the skin impedance. No significant difference was found between using impedance modulus and resistance values, but a small improvement of sensitivity could be achieved by using the susceptance. [2] J. Ojarand and M. Min, Simple and efficient excitation signals for fast impedance spectroscopy, Electron. Elect. Eng.,vol.19,pp.49 52,213. [3] B. Belmont, R. E. Dodde, and A. J. Shih, Impedance of tissue-mimicking phantom material under compression, J. Electr. Bioimp.,vol.4,pp.2 12, 213. [4] S. Luo, V. X. Afonso, J. G. Webster, and W. J. Tompkins, The electrode system in impedance-based measurement, IEEE Trans. Biomed. Eng., vol. 39, no. 11, pp , Nov [5] Ø. G. Martinsen, S. Grimnes, and E. Haug, Measuring depth depends on frequency in electrical skin impedance measurements, Skin Res. Technol., vol. 5, pp , [6] S. Grimnes and Ø. G. Martinsen, Sources of error in tetrapolar impedance measurements on biomaterials and other ionic conductors, J. Phys. D: Appl. Phys., vol. 4, pp. 9 14, 27. [7] J. H. Meijer, J. P. H. Reulen, P. L. Oe, W. Allon, L. G. Thijs, and H. Schneider, Differential impedance plethysmography for measuring thoracic impedances, Med. Biol. Eng. Comput., vol. 2, pp , REFERENCES [1] H. P. Schwan, Determination of biological impedances, in Physical Techniques in Biological Research, W. L. Nastuk, Ed. New York, NY, USA: Academic Press, 1963, vol. 6, pp Authors photographs and biographies not available at the time of publication.

BIOMEDICAL INSTRUMENTATION PROBLEM SHEET 1

BIOMEDICAL INSTRUMENTATION PROBLEM SHEET 1 BIOMEDICAL INSTRUMENTATION PROBLEM SHEET 1 Dr. Gari Clifford Hilary Term 2013 1. (Exemplar Finals Question) a) List the five vital signs which are most commonly recorded from patient monitors in high-risk

More information

A Comprehensive Model for Power Line Interference in Biopotential Measurements

A Comprehensive Model for Power Line Interference in Biopotential Measurements IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, VOL. 49, NO. 3, JUNE 2000 535 A Comprehensive Model for Power Line Interference in Biopotential Measurements Mireya Fernandez Chimeno, Member, IEEE,

More information

Implementation of wireless ECG measurement system in ubiquitous health-care environment

Implementation of wireless ECG measurement system in ubiquitous health-care environment Implementation of wireless ECG measurement system in ubiquitous health-care environment M. C. KIM 1, J. Y. YOO 1, S. Y. YE 2, D. K. JUNG 3, J. H. RO 4, G. R. JEON 4 1 Department of Interdisciplinary Program

More information

FEATURES OF VOLTAGE PULSE PLETHYSMOGRAPHY

FEATURES OF VOLTAGE PULSE PLETHYSMOGRAPHY FEATURES OF VOLTAGE PULSE PLETHYSMOGRAPHY Martina Melinščak, B.Sc., Polytechnic of Karlovac, 7 Karlovac, I. Meštrovića, Croatia, martina.melinscak@vuka.hr Ante Šantić, Prof. D.Sc., Faculty of Electrical

More information

Electrode comparison for textile-integrated electrocardiogram and impedance pneumography measurement

Electrode comparison for textile-integrated electrocardiogram and impedance pneumography measurement Tampere University of Technology Electrode comparison for textile-integrated electrocardiogram and impedance pneumography measurement Citation Tuohimäki, K., Mahdiani, S., Jeyhani, V., Vehkaoja, A., Iso-Ketola,

More information

Quick Check of EIS System Performance

Quick Check of EIS System Performance Quick Check of EIS System Performance Introduction The maximum frequency is an important specification for an instrument used to perform Electrochemical Impedance Spectroscopy (EIS). The majority of EIS

More information

Non-invasive Bio-impedance Measurement Using Voltage-Current Pulse Technique

Non-invasive Bio-impedance Measurement Using Voltage-Current Pulse Technique International Conference on Electrical, Electronics and Biomedical Engineering (ICEEBE'01) Penang (Malaysia) May 19-0, 01 Non-invasive Bio-impedance Measurement Using Voltage-Current Pulse Technique Sagar

More information

PHOTOPLETHYSMOGRAPHIC DETECTOR FOR PERIPHERAL PULSE REGISTRATION

PHOTOPLETHYSMOGRAPHIC DETECTOR FOR PERIPHERAL PULSE REGISTRATION PHOTOPLETHYSMOGRAPHIC DETECTOR FOR PERIPHERAL PULSE REGISTRATION Tatyana Dimitrova Neycheva, Dobromir Petkov Dobrev Centre of Biomedical Engineering Ivan Daskalov Bulgarian Academy of Sciences, Bl. 105

More information

A SIMPLE METHOD TO COMPARE THE SENSITIVITY OF DIFFERENT AE SENSORS FOR TANK FLOOR TESTING

A SIMPLE METHOD TO COMPARE THE SENSITIVITY OF DIFFERENT AE SENSORS FOR TANK FLOOR TESTING A SIMPLE METHOD TO COMPARE THE SENSITIVITY OF DIFFERENT AE SENSORS FOR TANK FLOOR TESTING HARTMUT VALLEN, JOCHEN VALLEN and JENS FORKER Vallen-Systeme GmbH, 82057 Icking, Germany Abstract AE testing of

More information

A HIGH OUTPUT IMPEDANCE CURRENT SOURCE FOR WIDEBAND BIOIMPEDANCE SPECTROSCOPY USING 0.35µM TSMC CMOS TECHNOLOGY

A HIGH OUTPUT IMPEDANCE CURRENT SOURCE FOR WIDEBAND BIOIMPEDANCE SPECTROSCOPY USING 0.35µM TSMC CMOS TECHNOLOGY A HIGH OUTPUT IMPEDANCE CURRENT SOURCE FOR WIDEBAND BIOIMPEDANCE SPECTROSCOPY USING 0.35µM TSMC CMOS TECHNOLOGY ENGR. ANGELITO A. SILVERIO, M.S., ENGR. ANGELINA A. SILVERIO, M.A., M.S. ELECTRONICS ENGINEERING,

More information

CONDUCTIVITY sensors are required in many application

CONDUCTIVITY sensors are required in many application IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, VOL. 54, NO. 6, DECEMBER 2005 2433 A Low-Cost and Accurate Interface for Four-Electrode Conductivity Sensors Xiujun Li, Senior Member, IEEE, and Gerard

More information

Electrical Impedance of Stainless Steel Needle Electrodes

Electrical Impedance of Stainless Steel Needle Electrodes Annals of Biomedical Engineering, Vol. 38, No. 7, July 2010 (Ó 2010) pp. 2371 2382 DOI: 10.1007/s10439-010-9989-2 Electrical Impedance of Stainless Steel Needle Electrodes HÅVARD KALVØY, 1,2 CHRISTIAN

More information

Technical data ESA614 Electrical Safety Analyzer

Technical data ESA614 Electrical Safety Analyzer Technical data ESA614 Electrical Safety Analyzer The ESA614 Electrical Safety Analyzer brings fast and simple automated testing in the form of a portable analyzer to healthcare technology professionals

More information

CDTE and CdZnTe detector arrays have been recently

CDTE and CdZnTe detector arrays have been recently 20 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 44, NO. 1, FEBRUARY 1997 CMOS Low-Noise Switched Charge Sensitive Preamplifier for CdTe and CdZnTe X-Ray Detectors Claudio G. Jakobson and Yael Nemirovsky

More information

Optimizing the Measurement Frequency in Electrical Impedance Tomography

Optimizing the Measurement Frequency in Electrical Impedance Tomography POSTER 2017, PRAGUE MAY 23 1 Optimizing the Measurement Frequency in Electrical Impedance Tomography Jakob ORSCHULIK 1, Tobias MENDEN 1 1 Philips Chair for Medical Information Technology, Helmholtz Institute

More information

BME 3113, Dept. of BME Lecture on Introduction to Biosignal Processing

BME 3113, Dept. of BME Lecture on Introduction to Biosignal Processing What is a signal? A signal is a varying quantity whose value can be measured and which conveys information. A signal can be simply defined as a function that conveys information. Signals are represented

More information

A Superior Current Source with Improved Bandwidth and Output Impedance for Bioimpedance Spectroscopy

A Superior Current Source with Improved Bandwidth and Output Impedance for Bioimpedance Spectroscopy International Journal of Engineering Research and Development e-issn: 2278-067X, p-issn: 2278-800X, www.ijerd.com Volume 12, Issue 12 (December 2016), PP.24-29 A Superior Current Source with Improved Bandwidth

More information

An impedance-based integrated biosensor for suspended DNA characterisation

An impedance-based integrated biosensor for suspended DNA characterisation An impedance-based integrated biosensor for suspended DNA characterisation Hanbin Ma, Richard W.R. Wallbank, Reza Chaji, Jiahao Li, Yuji Suzuki, Chris Jiggins and Arokia Nathan Supplementary Item Title

More information

Resolution Enhancement and Frequency Compounding Techniques in Ultrasound.

Resolution Enhancement and Frequency Compounding Techniques in Ultrasound. Resolution Enhancement and Frequency Compounding Techniques in Ultrasound. Proposal Type: Innovative Student PI Name: Kunal Vaidya PI Department: Chester F. Carlson Center for Imaging Science Position:

More information

FATIGUE INDEPENDENT AMPLITUDE-FREQUENCY CORRELATIONS IN EMG SIGNALS

FATIGUE INDEPENDENT AMPLITUDE-FREQUENCY CORRELATIONS IN EMG SIGNALS Fatigue independent amplitude-frequency correlations in emg signals. Adam SIEMIEŃSKI 1, Alicja KEBEL 1, Piotr KLAJNER 2 1 Department of Biomechanics, University School of Physical Education in Wrocław

More information

Laboratory Activities Handbook

Laboratory Activities Handbook Laboratory Activities Handbook Answer Key 0 P a g e Contents Introduction... 2 Optical Heart Rate Monitor Overview... 2 Bare Board Preparation... 3 Light Indicator... 5 Low Pass Filter... 7 Amplifier...

More information

Biomedical Instrumentation B2. Dealing with noise

Biomedical Instrumentation B2. Dealing with noise Biomedical Instrumentation B2. Dealing with noise B18/BME2 Dr Gari Clifford Noise & artifact in biomedical signals Ambient / power line interference: 50 ±0.2 Hz mains noise (or 60 Hz in many data sets)

More information

Electrical noise in the OR

Electrical noise in the OR Electrical noise in the OR Chris Thompson Senior Staff Specialist Royal Prince Alfred Hospital SYDNEY SOUTH WEST AREA HEALTH SERVICE NSW HEALTH Electrical noise in the OR Root causes Tiny little signals

More information

Electrical Safety. Principles and examples of safety. History of (electrical) safety. Rules and regulations. Safety in the Hospital

Electrical Safety. Principles and examples of safety. History of (electrical) safety. Rules and regulations. Safety in the Hospital Electrical Safety History of (electrical) safety Safety in the Hospital Role of Clinical Engineer Rules and regulations Food and Drug Administration (FDA), Underwriter s Laboratory (UL), Federal Communication

More information

AC-Coupled Front-End for Biopotential Measurements

AC-Coupled Front-End for Biopotential Measurements IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, VOL. 50, NO. 3, MARCH 2003 391 AC-Coupled Front-End for Biopotential Measurements Enrique Mario Spinelli 3, Student Member, IEEE, Ramon Pallàs-Areny, Fellow,

More information

Lab E5: Filters and Complex Impedance

Lab E5: Filters and Complex Impedance E5.1 Lab E5: Filters and Complex Impedance Note: It is strongly recommended that you complete lab E4: Capacitors and the RC Circuit before performing this experiment. Introduction Ohm s law, a well known

More information

DC feedback for wide band frequency fixed current source

DC feedback for wide band frequency fixed current source DC feedback for wide band frequency fixed current source Aoday H. Al-Rawi 1, W. M. A. Ibrahim 1, 2 and Eraj Humayun Mirza 1 1. Department of Biomedical Engineering, University of Malaya, Kuala Lumpur,

More information

Instrumentation amplifier

Instrumentation amplifier Instrumentationamplifieris a closed-loop gainblock that has a differential input and an output that is single-ended with respect to a reference terminal. Application: are intended to be used whenever acquisition

More information

Fundamentals of Pacing Therapy

Fundamentals of Pacing Therapy 1 4 7 2 5 8 3 6 9 Fundamentals of Pacing Therapy This presentation is provided with the understanding that the slide content must not be altered in any manner as the content is subject to FDA regulations.

More information

Suitability of the INPHAZE impedance analyzer for Bioimpedance

Suitability of the INPHAZE impedance analyzer for Bioimpedance Suitability of the INPHAZE impedance analyzer for Bioimpedance and EIT Sugashine Jeganathan 1,2 and Alistair McEwan 1, 1 School of Electrical and Information Engineering, The University of Sydney, NSW,

More information

Biomedical Sensor Systems Laboratory. Institute for Neural Engineering Graz University of Technology

Biomedical Sensor Systems Laboratory. Institute for Neural Engineering Graz University of Technology Biomedical Sensor Systems Laboratory Institute for Neural Engineering Graz University of Technology 2017 Bioinstrumentation Measurement of physiological variables Invasive or non-invasive Minimize disturbance

More information

PORTABLE ECG MONITORING APPLICATION USING LOW POWER MIXED SIGNAL SOC ANURADHA JAKKEPALLI 1, K. SUDHAKAR 2

PORTABLE ECG MONITORING APPLICATION USING LOW POWER MIXED SIGNAL SOC ANURADHA JAKKEPALLI 1, K. SUDHAKAR 2 PORTABLE ECG MONITORING APPLICATION USING LOW POWER MIXED SIGNAL SOC ANURADHA JAKKEPALLI 1, K. SUDHAKAR 2 1 Anuradha Jakkepalli, M.Tech Student, Dept. Of ECE, RRS College of engineering and technology,

More information

Low-cost photoplethysmograph solutions using the Raspberry Pi

Low-cost photoplethysmograph solutions using the Raspberry Pi Low-cost photoplethysmograph solutions using the Raspberry Pi Tamás Nagy *, Zoltan Gingl * * Department of Technical Informatics, University of Szeged, Hungary nag.tams@gmail.com, gingl@inf.u-szeged.hu

More information

Ques on (2): [18 Marks] a) Draw the atrial synchronous Pacemaker block diagram and explain its operation. Benha University June 2013

Ques on (2): [18 Marks] a) Draw the atrial synchronous Pacemaker block diagram and explain its operation. Benha University June 2013 Benha University June 2013 Benha Faculty of Engineering Electrical Department Hospital Instrumentations (E472) 4 Th year (control) Dr.Waleed Abdel Aziz Salem Time: 3 Hrs Answer the following questions.

More information

Biomedical Instrumentation (BME420 ) Chapter 6: Biopotential Amplifiers John G. Webster 4 th Edition

Biomedical Instrumentation (BME420 ) Chapter 6: Biopotential Amplifiers John G. Webster 4 th Edition Biomedical Instrumentation (BME420 ) Chapter 6: Biopotential Amplifiers John G. Webster 4 th Edition Dr. Qasem Qananwah BME 420 Department of Biomedical Systems and Informatics Engineering 1 Biopotential

More information

Standoff Human Life Sign Detection using High Sensitivity Pulsed Laser Vibrometer

Standoff Human Life Sign Detection using High Sensitivity Pulsed Laser Vibrometer Standoff Human Life Sign Detection using High Sensitivity Pulsed Laser Vibrometer Chen-Chia Wang, Sudhir Trivedi, and Feng Jin Brimrose Corp. of America, 7720 Belair Road, Baltimore, MD 21236 V. Swaminathan,

More information

Robust Wrist-Type Multiple Photo-Interrupter Pulse Sensor

Robust Wrist-Type Multiple Photo-Interrupter Pulse Sensor Robust Wrist-Type Multiple Photo-Interrupter Pulse Sensor TOSHINORI KAGAWA, NOBUO NAKAJIMA Graduate School of Informatics and Engineering The University of Electro-Communications Chofugaoka 1-5-1, Chofu-shi,

More information

Acoustic resolution. photoacoustic Doppler velocimetry. in blood-mimicking fluids. Supplementary Information

Acoustic resolution. photoacoustic Doppler velocimetry. in blood-mimicking fluids. Supplementary Information Acoustic resolution photoacoustic Doppler velocimetry in blood-mimicking fluids Joanna Brunker 1, *, Paul Beard 1 Supplementary Information 1 Department of Medical Physics and Biomedical Engineering, University

More information

A Tri-Mode Coupled Coil with Tunable Focal Point Adjustment for Bio-Medical Applications

A Tri-Mode Coupled Coil with Tunable Focal Point Adjustment for Bio-Medical Applications > REPLACE THIS LINE WITH YOUR PAPER IDENTIFICATION NUMBER (DOUBLE-CLICK HERE TO EDIT) < A Tri-Mode Coupled Coil with Tunable Focal Point Adjustment for Bio-Medical Applications Raunaq Pradhan, Student

More information

Lecture 4 Biopotential Amplifiers

Lecture 4 Biopotential Amplifiers Bioinstrument Sahand University of Technology Lecture 4 Biopotential Amplifiers Dr. Shamekhi Summer 2016 OpAmp and Rules 1- A = (gain is infinity) 2- Vo = 0, when v1 = v2 (no offset voltage) 3- Rd = (input

More information

potentiostat/galvanostat/impedance analyser

potentiostat/galvanostat/impedance analyser potentiostat/galvanostat/impedance analyser Rev. 9-2014 potentiostat/galvanostat/impedance PalmSens3 is a battery-powered, handheld instrument which allows the application of most of the relevant voltammetric,

More information

Design on Electrocardiosignal Detection Sensor

Design on Electrocardiosignal Detection Sensor Sensors & Transducers 203 by IFSA http://www.sensorsportal.com Design on Electrocardiosignal Detection Sensor Hao ZHANG School of Mathematics and Computer Science, Tongling University, 24406, China E-mail:

More information

Real Time Deconvolution of In-Vivo Ultrasound Images

Real Time Deconvolution of In-Vivo Ultrasound Images Paper presented at the IEEE International Ultrasonics Symposium, Prague, Czech Republic, 3: Real Time Deconvolution of In-Vivo Ultrasound Images Jørgen Arendt Jensen Center for Fast Ultrasound Imaging,

More information

Transient Response of Low-Power ECG Recoding Amplifiers for Use with Un-gelled Electrodes

Transient Response of Low-Power ECG Recoding Amplifiers for Use with Un-gelled Electrodes MATEC Web of Conferences 5, 000 (07) DOI: 0.05/ matecconf/075000 CSCC 07 Transient esponse of Low-Power ECG ecoding Amplifiers for Use with Un-gelled s Martin J. Burke,a Oscar Tuohy Dept. of Electronic

More information

HEART RATE DETECTION FROM IMPEDANCE PLETHYSMOGRAPHY BASED ON CONCEALED CAPACITIVE ELECTRODES

HEART RATE DETECTION FROM IMPEDANCE PLETHYSMOGRAPHY BASED ON CONCEALED CAPACITIVE ELECTRODES XIX IMEKO World Congress Fundamental and Applied Metrology September 6 11, 2009, Lisbon, Portugal HEART RATE DETECTION FROM IMPEDANCE PLETHYSMOGRAPHY BASED ON CONCEALED CAPACITIVE ELECTRODES Pablo S. Luna-Lozano,

More information

Voltage Sensors URV5-Z

Voltage Sensors URV5-Z Data sheet Version 05.00 Voltage Sensors URV5-Z May 2005 Universal voltage measurements from RF to microwaves The voltage sensors of the URV5-Z series are indispensable tools in RF and microwave laboratories,

More information

* Notebook is excluded. Features KL-720 contains nine modules, including Electrocardiogram Measurement, E lectromyogram Measurement,

* Notebook is excluded. Features KL-720 contains nine modules, including Electrocardiogram Measurement, E lectromyogram Measurement, KL-720 Biomedical Measurement System Supplied by: 011 683 4365 This equipment is intended for students to learn how to design specific measuring circuits and detect the basic physiological signals with

More information

Bio-Impedance Excitation System: A Comparison of Voltage Source and Current Source Designs

Bio-Impedance Excitation System: A Comparison of Voltage Source and Current Source Designs Available online at www.sciencedirect.com ScienceDirect APCBEE Procedia 7 (2013 ) 42 47 ICBET 2013: May 19-20, 2013, Copenhagen, Denmark Bio-Impedance Excitation System: A Comparison of Voltage Source

More information

D. Impedance probe fabrication and characterization

D. Impedance probe fabrication and characterization D. Impedance probe fabrication and characterization This section summarizes the fabrication process of the MicroCard bioimpedance probes. The characterization process is also described and the main electrical

More information

*Notebook is excluded

*Notebook is excluded Biomedical Measurement Training System This equipment is designed for students to learn how to design specific measuring circuits and detect the basic physiological signals with practical operation. Moreover,

More information

150Hz to 1MHz magnetic field coupling to a typical shielded cable above a ground plane configuration

150Hz to 1MHz magnetic field coupling to a typical shielded cable above a ground plane configuration 150Hz to 1MHz magnetic field coupling to a typical shielded cable above a ground plane configuration D. A. Weston Lowfreqcablecoupling.doc 7-9-2005 The data and information contained within this report

More information

Stability and Dynamic Performance of Current-Sharing Control for Paralleled Voltage Regulator Modules

Stability and Dynamic Performance of Current-Sharing Control for Paralleled Voltage Regulator Modules 172 IEEE TRANSACTIONS ON POWER ELECTRONICS, VOL. 17, NO. 2, MARCH 2002 Stability Dynamic Performance of Current-Sharing Control for Paralleled Voltage Regulator Modules Yuri Panov Milan M. Jovanović, Fellow,

More information

Design Considerations for Wrist- Wearable Heart Rate Monitors

Design Considerations for Wrist- Wearable Heart Rate Monitors Design Considerations for Wrist- Wearable Heart Rate Monitors Wrist-wearable fitness bands and smart watches are moving from basic accelerometer-based smart pedometers to include biometric sensing such

More information

Non-invasive Assessment of EIS (Electrical Impedance Spectroscopy) based on cell Equivalent Electrical circuit Model

Non-invasive Assessment of EIS (Electrical Impedance Spectroscopy) based on cell Equivalent Electrical circuit Model International Journal of Advancements in Research & Technology, Volume 2, Issue3, March-2013 1 Non-invasive Assessment of EIS (Electrical Impedance Spectroscopy) based on cell Equivalent Electrical circuit

More information

Transcutaneous Energy Transmission Based Wireless Energy Transfer to Implantable Biomedical Devices

Transcutaneous Energy Transmission Based Wireless Energy Transfer to Implantable Biomedical Devices Transcutaneous Energy Transmission Based Wireless Energy Transfer to Implantable Biomedical Devices Anand Garg, Lakshmi Sridevi B.Tech, Dept. of Electronics and Instrumentation Engineering, SRM University

More information

Testing Electrochemical Capacitors Part 3: Electrochemical Impedance Spectroscopy

Testing Electrochemical Capacitors Part 3: Electrochemical Impedance Spectroscopy Testing Electrochemical Capacitors Part 3: Electrochemical Impedance Spectroscopy Introduction Part 1 of this series of notes discusses basic theory of capacitors and describes several techniques to investigate

More information

Homework Assignment 03

Homework Assignment 03 Homework Assignment 03 Question 1 (Short Takes), 2 points each unless otherwise noted. 1. Two 0.68 μf capacitors are connected in series across a 10 khz sine wave signal source. The total capacitive reactance

More information

Hardware. MRI System. MRI system Multicoil Microstrip. Part1

Hardware. MRI System. MRI system Multicoil Microstrip. Part1 Hardware MRI system Multicoil Microstrip MRI System Part1 1 The MRI system is made up of a variety of subsystems. the Operator Workspace Gradient Driver subsystem The Physiological Acquisition Controller

More information

Medical Electronics Dr. Neil Townsend Michaelmas Term 2001 ( Pulse Oximetry: The story so far

Medical Electronics Dr. Neil Townsend Michaelmas Term 2001 (  Pulse Oximetry: The story so far Medical Electronics Dr. Neil Townsend Michaelmas Term 2001 (www.robots.ox.ac.uk/~neil/teaching/lectures/med_elec) Oxygen is carried in the blood by haemoglobin which has two forms: Hb and HbO 2. These

More information

Superconducting single-photon detectors as photon-energy and polarization resolving devices. Roman Sobolewski

Superconducting single-photon detectors as photon-energy and polarization resolving devices. Roman Sobolewski Superconducting single-photon detectors as photon-energy and polarization resolving devices Roman Sobolewski Departments of Electrical and Computing Engineering Physics and Astronomy, Materials Science

More information

Short Interfacial Antennas for Medical Microwave Imaging

Short Interfacial Antennas for Medical Microwave Imaging Short Interfacial Antennas for Medical Microwave Imaging J. Sachs; M. Helbig; S. Ley; P. Rauschenbach Ilmenau University of Technology M. Kmec; K. Schilling Ilmsens GmbH Folie 1 Copyright The use of this

More information

Medical Imaging (EL582/BE620/GA4426)

Medical Imaging (EL582/BE620/GA4426) Medical Imaging (EL582/BE620/GA4426) Jonathan Mamou, PhD Riverside Research Lizzi Center for Biomedical Engineering New York, NY jmamou@riversideresearch.org On behalf of Prof. Daniel Turnbull Outline

More information

SIMPLE WIDE FREQUENCY RANGE IMPEDANCE METER BASED ON AD5933 INTEGRATED CIRCUIT. Konrad Chabowski, Tomasz Piasecki, Andrzej Dzierka, Karol Nitsch

SIMPLE WIDE FREQUENCY RANGE IMPEDANCE METER BASED ON AD5933 INTEGRATED CIRCUIT. Konrad Chabowski, Tomasz Piasecki, Andrzej Dzierka, Karol Nitsch Metrol. Meas. Syst., Vol. XXII (2015), No. 1, pp. 13 24. METROLOGY AND MEASUREMENT SYSTEMS Index 330930, ISSN 0860-8229 www.metrology.pg.gda.pl SIMPLE WIDE FREQUENCY RANGE IMPEDANCE METER BASED ON AD5933

More information

Removal of Power-Line Interference from Biomedical Signal using Notch Filter

Removal of Power-Line Interference from Biomedical Signal using Notch Filter ISSN:1991-8178 Australian Journal of Basic and Applied Sciences Journal home page: www.ajbasweb.com Removal of Power-Line Interference from Biomedical Signal using Notch Filter 1 L. Thulasimani and 2 M.

More information

Available online at ScienceDirect. Procedia Engineering 120 (2015 ) EUROSENSORS 2015

Available online at   ScienceDirect. Procedia Engineering 120 (2015 ) EUROSENSORS 2015 Available online at www.sciencedirect.com ScienceDirect Procedia Engineering 120 (2015 ) 511 515 EUROSENSORS 2015 Inductive micro-tunnel for an efficient power transfer T. Volk*, S. Stöcklin, C. Bentler,

More information

First steps towards an implantable electromyography (EMG) sensor powered and controlled by galvanic coupling

First steps towards an implantable electromyography (EMG) sensor powered and controlled by galvanic coupling First steps towards an implantable electromyography (EMG) sensor powered and controlled by galvanic coupling Laura Becerra-Fajardo 1[0000-0002-5414-8380] and Antoni Ivorra 1,2[0000-0001-7718-8767] 1 Department

More information

TBM - Tone Burst Measurement (CEA 2010)

TBM - Tone Burst Measurement (CEA 2010) TBM - Tone Burst Measurement (CEA 21) Software of the R&D and QC SYSTEM ( Document Revision 1.7) FEATURES CEA21 compliant measurement Variable burst cycles Flexible filtering for peak measurement Monitor

More information

Sensing Local Temperature and Conductivity. Microwave Radiometry

Sensing Local Temperature and Conductivity. Microwave Radiometry Forum for Electromagnetic Research Methods and Application Technologies (FERMAT) Sensing Local Temperature and Conductivity Changes in a Brain Phantom Using Near-Field Microwave Radiometry Evangelos Groumpas,

More information

Comparative Testing of Synchronized Phasor Measurement Units

Comparative Testing of Synchronized Phasor Measurement Units Comparative Testing of Synchronized Phasor Measurement Units Juancarlo Depablos Student Member, IEEE Virginia Tech Virgilio Centeno Member, IEEE Virginia Tech Arun G. Phadke Life Fellow, IEEE Virginia

More information

Transmit filter designs for ADSL modems

Transmit filter designs for ADSL modems Transmit filter designs for ADSL modems 1. OBJECTIVES... 2 2. REFERENCE... 2 3. CIRCUITS... 2 4. COMPONENTS AND SPECIFICATIONS... 3 5. DISCUSSION... 3 6. PRE-LAB... 4 6.1 RECORDING SPECIFIED OPAMP PARAMETERS

More information

Bio-Impedance Spectroscopy (BIS) Measurement System for Wearable Devices

Bio-Impedance Spectroscopy (BIS) Measurement System for Wearable Devices Bio-Impedance Spectroscopy (BIS) Measurement System for Wearable Devices Bassem Ibrahim*, Drew A. Hall, Roozbeh Jafari* * Embedded Signal Processing (ESP) Lab, Texas A&M University, TX, USA BioSensors

More information

EFFECT OF INTEGRATION ERROR ON PARTIAL DISCHARGE MEASUREMENTS ON CAST RESIN TRANSFORMERS. C. Ceretta, R. Gobbo, G. Pesavento

EFFECT OF INTEGRATION ERROR ON PARTIAL DISCHARGE MEASUREMENTS ON CAST RESIN TRANSFORMERS. C. Ceretta, R. Gobbo, G. Pesavento Sept. 22-24, 28, Florence, Italy EFFECT OF INTEGRATION ERROR ON PARTIAL DISCHARGE MEASUREMENTS ON CAST RESIN TRANSFORMERS C. Ceretta, R. Gobbo, G. Pesavento Dept. of Electrical Engineering University of

More information

Name Kyla Jackson, Todd Germeroth, Jake Spooler Date May 5, 2010 Lab 3E Group 3 Experiment Title Project Deliverable 3

Name Kyla Jackson, Todd Germeroth, Jake Spooler Date May 5, 2010 Lab 3E Group 3 Experiment Title Project Deliverable 3 Name Kyla Jackson, Todd Germeroth, Jake Spooler Date May 5, 2010 Lab 3E Group 3 Experiment Title Project Deliverable 3 Objective The objective of this project was to design and construct an ECG measurement

More information

CS114 + CS115 + CS116

CS114 + CS115 + CS116 System description Test Setup for MIL-STD-461 D, E&F CS114 + CS115 + CS116 1. MONTENA EMC... 2 1.1 PRODUCTS... 3 1.2 TURN KEY MIL STD 461 TEST INSTALLATIONS... 3 2. TEST SETUP DESCRIPTION... 4 2.1 TEST

More information

A Wrist-Worn Strap with an Array of Electrodes for Robust Physiological Sensing

A Wrist-Worn Strap with an Array of Electrodes for Robust Physiological Sensing A Wrist-Worn Strap with an Array of s for Robust Physiological Sensing Bassem Ibrahim, IEEE Student Member, Justin McMurray, IEEE Student Member, and Roozbeh Jafari, IEEE Senior Member Abstract Robust

More information

Setting up a Multi sine impedance measurement

Setting up a Multi sine impedance measurement Setting up a Multi sine impedance measurement Case study: how do I setup a Multi Sine impedance measurement? 1 Single sine vs Multi sine Traditional electrochemical impedance spectroscopy measurements

More information

EXPERIMENT 5 Bioelectric Measurements

EXPERIMENT 5 Bioelectric Measurements Objectives EXPERIMENT 5 Bioelectric Measurements 1) Generate periodic signals with a Signal Generator and display on an Oscilloscope. 2) Investigate a Differential Amplifier to see small signals in a noisy

More information

EHRLUND MICROPHONES EAP MANUAL. for. Ehrlund Acoustic Pickup Ehrlund Preamp Portable Ehrlund Preamp Phantom

EHRLUND MICROPHONES EAP MANUAL. for. Ehrlund Acoustic Pickup Ehrlund Preamp Portable Ehrlund Preamp Phantom EHRLUND MICROPHONES EAP MANUAL for Ehrlund Acoustic Pickup Ehrlund Preamp Portable Ehrlund Preamp Phantom 1 CONTENTS QUICKSTART... 3 EHRLUND ACOUSTIC PICKUP... 4 PLACING THE PICKUP... 5 KEEPING YOUR PICKUP

More information

EIS Measurement of a Very Low Impedance Lithium Ion Battery

EIS Measurement of a Very Low Impedance Lithium Ion Battery EIS Measurement of a Very Low Impedance Lithium Ion Battery Introduction Electrochemical Impedance Spectroscopy, EIS, is a very powerful way to gain information about electrochemical systems. It is often

More information

Non-Contact Ultrasound Characterization of Paper Substrates

Non-Contact Ultrasound Characterization of Paper Substrates ECNDT 006 - Poster 04 Non-Contact Ultrasound Characterization of Paper Substrates María HELGUERA, J. ARNEY, N. TALLAPALLY, D. ZOLLO., CFC Center for Imaging Science, Rochester Institute of Technology,

More information

Measurement Of The Magnitude And Direction Of The Electric Field Of A Mobile Phone In The Near Field

Measurement Of The Magnitude And Direction Of The Electric Field Of A Mobile Phone In The Near Field Measurement Of The Magnitude And Direction Of The Electric Field Of A Mobile Phone In The Near Field Pantinakis A. and E. Batsaki E. Department of Sciences, Physics Section, Technical University of Crete,

More information

easypll UHV Preamplifier Reference Manual

easypll UHV Preamplifier Reference Manual easypll UHV Preamplifier Reference Manual 1 Table of Contents easypll UHV-Pre-Amplifier for Tuning Fork 2 Theory... 2 Wiring of the pre-amplifier... 4 Technical specifications... 5 Version 1.1 BT 00536

More information

THE TREND toward implementing systems with low

THE TREND toward implementing systems with low 724 IEEE JOURNAL OF SOLID-STATE CIRCUITS, VOL. 30, NO. 7, JULY 1995 Design of a 100-MHz 10-mW 3-V Sample-and-Hold Amplifier in Digital Bipolar Technology Behzad Razavi, Member, IEEE Abstract This paper

More information

CONDITION ASSESSMENT OF XLPE MV CABLE JOINTS BY USING AN INSULATION TESTER

CONDITION ASSESSMENT OF XLPE MV CABLE JOINTS BY USING AN INSULATION TESTER CONDITION ASSESSMENT OF XLPE MV CABLE JOINTS BY USING AN INSULATION TESTER Henrik ENOKSEN Espen EBERG Sverre HVIDSTEN SINTEF Energy Research - Norway SINTEF Energy Research - Norway SINTEF Energy Research

More information

A GTEM BEST PRACTICE GUIDE APPLYING IEC TO THE USE OF GTEM CELLS

A GTEM BEST PRACTICE GUIDE APPLYING IEC TO THE USE OF GTEM CELLS - 27-39 H1 A BEST PRACTICE GUIDE APPLYING IEC 61-4-2 TO THE USE OF CELLS A. Nothofer, M.J. Alexander, National Physical Laboratory, Teddington, UK, D. Bozec, D. Welsh, L. Dawson, L. McCormack, A.C. Marvin,

More information

CHAPTER 6 SIGNAL PROCESSING TECHNIQUES TO IMPROVE PRECISION OF SPECTRAL FIT ALGORITHM

CHAPTER 6 SIGNAL PROCESSING TECHNIQUES TO IMPROVE PRECISION OF SPECTRAL FIT ALGORITHM CHAPTER 6 SIGNAL PROCESSING TECHNIQUES TO IMPROVE PRECISION OF SPECTRAL FIT ALGORITHM After developing the Spectral Fit algorithm, many different signal processing techniques were investigated with the

More information

Electrical Characterization of OLED s Using Solartron Instrumentation

Electrical Characterization of OLED s Using Solartron Instrumentation MTSAP2 Electrical Characterization of OLED s Using Solartron Instrumentation Introduction An OLED is a light emitting diode with an organic emissive electro-luminescent layer. The organic layer can be

More information

EFFECT OF SHIELDING ON CABLE RF INGRESS MEASUREMENTS LARRY COHEN

EFFECT OF SHIELDING ON CABLE RF INGRESS MEASUREMENTS LARRY COHEN EFFECT OF SHIELDING ON CABLE RF INGRESS MEASUREMENTS LARRY COHEN OVERVIEW Purpose: Examine the common-mode and differential RF ingress levels of 4-pair UTP, F/UTP, and F/FTP cables at an (RJ45) MDI port

More information

EMG feature extraction for tolerance of white Gaussian noise

EMG feature extraction for tolerance of white Gaussian noise EMG feature extraction for tolerance of white Gaussian noise Angkoon Phinyomark, Chusak Limsakul, Pornchai Phukpattaranont Department of Electrical Engineering, Faculty of Engineering Prince of Songkla

More information

HETERONUCLEAR IMAGING. Topics to be Discussed:

HETERONUCLEAR IMAGING. Topics to be Discussed: HETERONUCLEAR IMAGING BioE-594 Advanced MRI By:- Rajitha Mullapudi 04/06/2006 Topics to be Discussed: What is heteronuclear imaging. Comparing the hardware of MRI and heteronuclear imaging. Clinical applications

More information

THERMAL NOISE. Advanced Laboratory, Physics 407, University of Wisconsin. Madison, Wisconsin 53706

THERMAL NOISE. Advanced Laboratory, Physics 407, University of Wisconsin. Madison, Wisconsin 53706 (revised 1/25/07) THERMAL NOISE Advanced Laboratory, Physics 407, University of Wisconsin Madison, Wisconsin 53706 Abstract The aim of this experiment is to observe the thermal noise in a resistor, to

More information

Analog Circuits and Systems

Analog Circuits and Systems Analog Circuits and Systems Prof. K Radhakrishna Rao Lecture 21: Filters 1 Review Integrators as building blocks of filters Frequency compensation in negative feedback systems Opamp and LDO frequency compensation

More information

Electrical Characterisation of Dry Electrodes for ECG Recording

Electrical Characterisation of Dry Electrodes for ECG Recording th WSEAS International Conference on CIRCUITS, Heraklion, Greece, July -, 8 Electrical Characterisation of Dry Electrodes for ECG Recording Baba A., Burke M. J. Department of Electronic and Electrical

More information

Validation of the Happify Breather Biofeedback Exercise to Track Heart Rate Variability Using an Optical Sensor

Validation of the Happify Breather Biofeedback Exercise to Track Heart Rate Variability Using an Optical Sensor Phyllis K. Stein, PhD Associate Professor of Medicine, Director, Heart Rate Variability Laboratory Department of Medicine Cardiovascular Division Validation of the Happify Breather Biofeedback Exercise

More information

ICAM. Electronics & Software. Industrial Charge Amplifier for Applications in Manufacturing. Type 5073A...

ICAM. Electronics & Software. Industrial Charge Amplifier for Applications in Manufacturing. Type 5073A... Electronics & Software ICAM Type 5073A... Industrial Charge Amplifier for Applications in Manufacturing The ICAM charge amplifier (Industrial Charge Amplifier Manufacturing) converts the piezoelectric

More information

Advanced Test Equipment Rentals ATEC (2832)

Advanced Test Equipment Rentals ATEC (2832) Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) A.H. Systems Model Active Monopole Antennas Active Monopole Antenna Series Operation Manual 1 TABLE OF CONTENTS INTRODUCTION

More information

CENTRE OF TESTING SERVICE INTERNATIONAL

CENTRE OF TESTING SERVICE INTERNATIONAL CENTRE OF TESTING SERVICE INTERNATIONAL OPERATE ACCORDING TO ISO/IEC 17025 FCC TEST REPORT TEST REPORT NUMBER : CGZ3150202-00097-E A101,No.65,Zhuji Highway,Tianhe District,Guangzhou, Guangdong, China Report

More information

UNIT 2. Q.1) Describe the functioning of standard signal generator. Ans. Electronic Measurements & Instrumentation

UNIT 2. Q.1) Describe the functioning of standard signal generator. Ans.   Electronic Measurements & Instrumentation UNIT 2 Q.1) Describe the functioning of standard signal generator Ans. STANDARD SIGNAL GENERATOR A standard signal generator produces known and controllable voltages. It is used as power source for the

More information

Biosignal filtering and artifact rejection, Part II. Biosignal processing, S Autumn 2017

Biosignal filtering and artifact rejection, Part II. Biosignal processing, S Autumn 2017 Biosignal filtering and artifact rejection, Part II Biosignal processing, 521273S Autumn 2017 Example: eye blinks interfere with EEG EEG includes ocular artifacts that originates from eye blinks EEG: electroencephalography

More information

Measuring Batteries using the Right Setup: Dual-cell CR2032 and Battery Holder

Measuring Batteries using the Right Setup: Dual-cell CR2032 and Battery Holder Measuring Batteries using the Right Setup: Dual-cell CR2032 and 18650 Battery Holder Introduction Knowing the exact specifications when testing batteries or any other energy-storage device is crucial.

More information