1 PFC/JA Response of X-UV Photodiodes to kev X Rays and MeV Alpha Particles K. W. Wenzel, C. K. Li, R. D. Petrasso, D. H. Lo, M. W. Bautz*, G. R. Picker, Jr.*, and E. Hsieht October 1992 To be published in: Review of Scientific Instruments MIT Plasma Fusion Center * MIT Center for Space Research Cambridge, MA 2139 t Lawrence Livermore National Laboratory Livermore, CA 9455 This work was supported in part by NASA Contract No. NASW-4372, U.S. DOE Grant No. DE-FG2-91ER5419, and LLNL Subcontract No. B

2 Response of X-UV Photodiodes to kev X Rays and MeV Alpha Particles Kevin W. Wenzel, Chi-Kang Li, Richard D. Petrasso, and Daniel H. Lo MIT Plasma Fusion Center, Cambridge, MA 2139 Marshall W. Bautz and George R. Ricker, Jr. MIT Center for Space Research, Cambridge, MA 2139 Ed Hsieh Lawrence Livermore National Laboratory, Livermore, CA 9455 Abstract The absolute x-ray response of three X-UV photodiodes was measured over an energy range of 1.5 to 17.5 kev so that they could be used to calibrate x-ray imaging systems for the Astro-D satellite mission. An intense electronbeam x-ray generator was used to test both the DC and AC x-ray response at 1.5, 4.5, 8., and 17.5 kev, and an 55 Fe source was used to examine one of the photodiodes at 5.9 kev. The x-ray response was determined by comparing the X-UV diode signal to that of a previously calibrated silicon surface barrier diode (SBD). The X-UV detector response was similar to the SBD response at low energies (1.5 and 4.5 kev). At 8 kev, the X-UV detectors exhibited about 7% of the SBD response, and at 17.5 kev, about 5%. This result is surprising, because the X-UV diodes actually have a greater silicon thickness than the SBD. In contrast to our findings for SBDs in the past, this implies that not the entire physical volume of these detectors comprises the active volume. The X-UV detector x-ray response was also examined as a function of the applied bias voltage. No significant bias voltage dependence of the x-ray signal was found, which indicates that the depletion layer thickness does not determine the active volume either. However, the detector noise was found to decrease substantially as the applied bias was raised from zero to a few volts. Response of these detectors, operated in pulse mode, to 22eRa a particles indicated large charge carrier recombination in the bulk silicon. This feature renders these detectors unsuitable for most chargedparticle spectroscopy applications. 1

3 Introduction Silicon photodiodes with stable, high quantum efficiency in the soft x- ray and ultraviolet wavelength regions have recently become available. 1 3 These detectors are referred to as X-UV photodiodes because of the spectral region over which they are sensitive. In addition to their high efficiency and stability, X-UV detectors have the added advantages of large dynamic range, low noise, low cost, little or no bias voltage requirements, and large sensitive areas. These advantages make X-UV detectors quite attractive for measuring radiation emitted by both space and laboratory plasmas. These detectors can be used as radiation intensity monitors if their absolute response is known as a function of wavelength. The absolute response to ultraviolet photons in the 1 ev to 15 ev region has previously been measured.- 3 The absolute x-ray response from a radioactive 55 Fe source emitting Mn K x rays at 5.9 and 6.5 kev has also been measured. 1 Most x-ray emission of importance to astrophysical and fusion plasmas lies in the.5 to 1 kev regions (especially from plasmas with electron temperatures of.5-2 kev). It was therefore the goal of this work to extend the region of absolute response measurements to x-ray energies from about 1 to 1 kev. In this paper, we describe measurements of the absolute x-ray response of three X-UV detectors at 1.5, 4.5, 5.9, 8., and 17.5 kev. These measurements were made primarily so that the X-UV diodes could be used as a standard to calibrate the x-ray imaging CCD for the Astro-D satellite mission. 4 The particular detectors examined in this work comprised two X-UV35 (.35 cm 2 area) and one X-UV1C (1. cm 2 area), all of which were manufactured by United Detector Technology 5 (see Table 1). In addition to the absolute response measurements, we also explored changes in the x-ray response and detector/preamp noise as the applied bias voltage was varied. Bias effects are important for at least three reasons. First, these X-UV detectors are claimed to have high quantum efficiency with zero applied bias. Thus it is important to determine whether the efficiency could be increased simply by applying a bias voltage. Second, we have previously seen distortions of large x-ray signals (~ 1 PA) from silicon surface barrier diodes (SBD) when their applied bias was reduced below about one volt. 7 Third, by observing the detector performance with different bias voltages, we can improve our overall understanding of these detectors. For example, it has long been believed that the depletion layer, which has 2

4 a thickness proportional to the square root of the bias voltage 8, comprises the x-ray sensitive portion of silicon detectors. 8 ' In contrast, we found for SBDs that the x-ray sensitive volume (as well as the charged-particle sensitive volume) is determined primarily by the physical silicon thickness, not the depletion layer thickness."" Thus it is of general interest to observe any bias dependence in the X-UV x-ray response. In this paper, we therefore describe the x-ray response of X-UV photodiodes as the bias voltage was varied from to -3 V. No dependence of the x-ray response was observed over this range of voltages. We conclude that the x-ray response is not determined primarily by the depletion layer. We did observe a decrease in the electronic noise as the bias was increased. For this reason, it may be desirable to operate these detectors with some low applied bias voltage (a few volts is the maximum suggested by the manufacturer). To further understand what constitutes the sensitive thicknesses of the X-UV detectors, their responses were compared to the SBD response to a particles from a 1Ra source. This isotope emits a particles with energies from 4.7 to 7.6 MeV, corresponding to ranges in silicon of 22 to 45 pm. Of greatest interest to us was to observe any degradation in the highest energy a line (7.6 MeV) which would allow us to estimate the active thickness of the X-UV detector. Although this effort was not successful, we were able to conclude that these detectors were unsuitable for most charged-particle spectroscopy applications due to larger charge recombination in the bulk silicon. However, they may be useful for the less demanding task of chargedparticle counting experiments.' This paper is organized as follows. Section I describes the electron-beam x-ray source used to characterize the X-UV detectors (and a radioactive 55 Fe source used to measure the relative X-UV response to Mn K line radiation at 5.9 and 6.5 kev). Section II describes DC x-ray measurements with zero applied bias voltage. Section III describes measurements of the detector responses to modulated x-rays using an AC-coupled current-to-voltage amplifier. The data analysis is discussed in Section IV, and the absolute response results are summarized. Section V describes the effect of varying the bias voltage on each detector. Section VI describes the a-particle response of the X-UV detectors, and Section VII contains a summary of all the results and a discussion of their implications. 3

5 I. X-Ray Sources The main x-ray source was an intense electron-beam x-ray generator (illustrated in Fig. 1).14,15 The electron beam was generated by a bare tungsten filament operated at ground potential. The beam was accelerated over a distance of less than 3 cm to an anode (target) maintained at positive high voltage. This configuration was found to be preferable to applying a negative high voltage to the cathode (and grounding the anode), because it eliminated spurious signals from electrons that were accelerated from the filament to the detector at ground potential. The anode potential could be varied between zero and +4 kv. The emission current could be varied from zero to 2 ma by changing the power input to the filament. The generator ran quite stably when the power deposited on the target was less than about 2 W. However, with higher power levels, the x-ray source would only run for a short time (typically less than two minutes) before a discharge breakdown occurred between the filament and the anode. Consequently, the DC data described below were acquired while running below the nominal 2 W threshold in order to maintain stable signal levels over several minutes. The AC data were taken over a range of operating conditions, mostly with higher emission current in order to maximize the signal level. It was possible to obtain the AC data at high power levels, because the data could be obtained in a short time period. The maximum power run was about 1 kw, during the molybdenum series, with 25 kv and 4 ma emission current. In addition to the electron-beam x-ray generator, we examined the response of one of the X-UV diodes (# 2 in Table 1) to Mn K x rays at 5.9 and 6.5 kev from a radioactive 55 Fe source. This source was well calibrated, with an activity of 2.8 mci. II. DC X-Ray Response Measurements The DC x-ray response measurements were performed by simultaneously measuring the X-UV photodiode current and the current from a standard silicon surface barrier diode (SBD) when both detectors viewed the same anode (both detectors were unbiased). For our comparison diode, we used a partially depleted SBD from EG&G ORTEC 16, serial number B, for which the x-ray response from.3 kev to 17.5 kev has been absolutely measured

6 We checked to make sure that the x-ray beam uniformly illuminated both detectors. This was done by rotating the detectors about an axis located onehalf way between them (they were separated by about 4 cm). Furthermore, with the exception of the 55 Fe case, the measurements were performed with the detectors located about 92 cm from the source anode (see Fig. 1). Both detectors were apertured with brass collimators of.32 cm diameter (except for the "Fe measurements, where a smaller aperture was used). This was done to avoid spurious signals due to x-ray absorption in the detector edge region, 7 and to make their active areas equal. The detectors were also covered, for each wavelength, with thin filters of the same composition as the anode to, first, increase the intensity of characteristic line radiation relative to the bremsstrahlung continuum, and second, to eliminate the visible light response from the x-ray source filament. We hereafter refer to this filter as the "Line Filter" (see Table 2). In addition, a separate, movable aluminum absorption filter of known thickness was used to test the spectral purity of the source by inserting the filter between the source and detectors and checking that the detector signals dropped by the predicted amount (see Fig. 1 and Table 2). Hereafter, we refer to this filter as the "Test Filter". The positive signal current from both detectors was measured with a Keithley 617 digital electrometer. 8 The background current, due to leakage current and slight visible light response, was -1 pa for the X-UV diode, and ~4 pa for the SBD. It was monitored and subtracted from the total signal current to obtain the net x-ray signal current. We found it was important to monitor the background signal over the duration of a single calibration experiment (i.e., one x-ray wavelength), because there was noticeable drift in the background leakage current. Because the leakage current increased with time, we believe this drift was due to heating of the detectors by the x-ray source filament. II.A. DC Response to 1.5 kev Al K X Rays The Al K x-ray response at 1.5 kev was measured while running the x-ray generator with an aluminum anode at potentials of 5 to 8 kv with emission currents of 6 to 14 ma. The Test Filter was 17.3 Pm Al foil (thickness measured with 55 Fe), which is predicted to transmit 15% of the 1.5 kev Al K, line (using the cross sections of Ref. 19). Both detectors also had 17 pm aluminum foils as the Line Filters for visible light and bremsstrahlung 5

7 rejection. The response currents without the aluminum Test Filter ranged from 14 pa to 44 pa for both the X-UV and SBD detectors. The response currents with the Test Filter in place ranged from 1 pa to 11 pa. 2 The Test Filter transmission varied from 17% for the 5 kv anode voltage to 3 % for the 8 kv anode voltage. The 17% result agrees well with the predicted 15% transmission, which strongly supports the conclusion that the detectors were both responding to essentially pure Al K radiation. On the other hand, the 8 kv result indicates that the spectrum may have been contaminated with some higher energy bremsstrahlung continuum. However, the same relative X-UV response was obtained with the anode at 5 kv and at 8 kv. The measured spectrum (shown in Fig. 2) also indicated quite clean line radiation at 1.5 kev. II.B. DC Response to 4.5 kev Ti K X Rays The Ti K x-ray response at 4.5 kev was measured while running the x-ray generator with a titanium anode at potentials of 7 to 1 kv with emission currents of 2 to 14 ma. The Test Filter was 17.3 pm Al foil, which is predicted to transmit 31% of the 4.5 kev Ti line (Ref. 19). Both detectors also had 25 pm titanium foils as the Line Filters for visible light and bremsstrahlung rejection. This Line Filter was predicted to transmit 3% of the Ti K" line. The response currents without the aluminum Test Filter ranged from 7 pa to 2 pa for both the X-UV and SBD detectors. The response currents with the Test Filter in place ranged from 25 pa to 7 pa. 2 The Test Filter transmission varied from 32% for the 7 kv anode voltage to 35 % for the 1 kv anode voltage. These results agree well with the predicted 31% transmission, which strongly supports the conclusion that the detectors were both responding to essentially pure Ti K radiation. The measured spectrum (shown in Fig. 3) also indicated quite clean line radiation at 4.5 kev. II.C. DC Response to 5.9 kev Mn K X Rays The Mn K x-ray response of X-UV #2 was measured using an 5 -Fe radioactive source (2.8 mci). Both detectors were collimated with.24 cm diameter brass apertures so they would view the same geometry. Furthermore, they were shielded with 17 pm aluminum foil to reject the room light. The x-ray response currents were Ix-uv=1 4 pa and ISBD= 1 8 pa, so the 6

8 relative response at this energy was IX-UV/ISBD= We also measured the spectrum with the Si(Li) spectrometer, which had an additional 12.5 pm copper foil to further reduce the x-ray signal (otherwise the geometry was very similar to the diode measurements). Fig. 4 shows the 55 Fe spectrum. To determine the uncertainty in the relative X-UV response, we estimated the uncertainty in the individual current measurements to be of the order 2 pa. Then adding the X-UV and SBD uncertainty in quadrature gives the total uncertainty to be about 18%. Then at 5.9 kev, Ix-Uv/ISBD=. 7 8±.14 (Table 1). Note that this response is lower than reported in Ref. 1. II.D. DC Response to 8. kev Cu K X Rays The Cu K x-ray response at 8. kev was measured while running the x-ray generator with a copper anode at potentials of 11 to 14 kv and with emission currents of 3 to 13 ma. The Test Filter was 86.4 pm Al foil, which is predicted to transmit 34% of the 8 kev Cu line (Ref. 19). Both detectors also had 5 pm copper foils as the Line Filters for visible light and bremsstrahlung rejection (thinner Line Filters were used for the AC measurements to obtain larger x-ray signals; compare Tables 2 and 3). This Line Filter was predicted to transmit 11% of the Cu K, line. The response currents without the aluminum Test Filter ranged from about 6 pa to about 3 pa for both the X-UV and SBD detectors. The response currents with the Test Filter in place ranged from 1 pa to 1 pa. 2 At this line energy, the X-UV diodes always exhibited lower signal currents than the SBD. The Test Filter transmission varied from 33% for the 11 kv anode voltage to 37% for the 14 kv anode voltage. These results agree well with the predicted 34% transmission, which strongly support the conclusion that the detectors were both responding to essentially pure Cu K radiation. The spectrum shown in Fig. 5 was taken with the thinner copper Line Filter used for the AC measurements (Table 3). The DC measurement used a thicker copper filter (Table 2), so the corresponding spectrum was cleaner than that shown in Fig. 5. II.E. DC Response to 17.5 kev Mo K X Rays The Mo K x-ray response at 17.5 kev was measured while running the x-ray generator with a molybdenum anode at potentials of >25 kv with emission currents of 8 to 11 ma. The Test Filter was 86.4 pm Al foil, which 7

9 is predicted to transmit 89% of the 17.5 kev Mo line (using the cross sections of Ref. 21). This Test Filter was not attenuating enough to adequately confirm the spectral purity of the Mo K line. Both detectors also had 124 pm molybdenum foils as the Line Filters for visible light and bremsstrahlung rejection. This Line Filter was predicted to transmit 9.7% of the Mo K, line. The response currents with and without the aluminum Test Filter ranged from about 1 pa to about 15 pa for the SBD and from 6 to 14 pa for the X-UV detectors. At this line energy, the X-UV diodes always exhibited lower signal currents than the SBD. The measured spectrum (shown in Fig. 6) indicated quite clean line radiation at 17.5 and 19.6 kev. III. AC X-Ray Measurements In order to test the effect of different bias voltages, we had initially planned to use the internal voltage source of the Keithley 617 digital electrometer, and measure the signal current directly with the electrometer. However, when the bias was applied, the leakage current increased dramatically (by at least an order of magnitude, depending on the specific applied bias). Moreover, this "background" current drifted over a time scale of minutes, and continued to do so over several hours. It was therefore impossible to establish a DC background current to subtract from the total x-ray current. We therefore used a simple AC-coupled transresistance current-to-voltage amplifier 22 (see Fig. 7). With this circuit, we examined the x-ray signal level and the noise level as a function of the applied bias voltage. This section describes the AC x-ray measurements. The x-ray signal was mechanically chopped by a motor-driven aluminum disk with eight rectangular apertures. This "chopper wheel" was located in the vacuum chamber, directly in front of the X-UV detector and the SBD. In this case, the Test Filter covered every other aperture in the chopper wheel, so alternate x-ray pulses were filtered. This provided a quick, in-situ determination of the x-ray spectral purity. Only those data that exhibited the expected line transmission through the Test Filter were used to determine the absolute x-ray response. Table 3 lists the Line and Test Filters used during these AC measurements. For these measurements, the detectors were moved to 71 cm from the anode, and their brass apertures were enlarged to about.52 cm diameter to maximize the signal levels. Fig. 8 shows representative signals from each anode material. 8

10 III.A. AC Response to 1.5 kev Al K X Rays The modulated Al K x-ray response at 1.5 kev was measured while running the x-ray generator with an aluminum anode at potentials of 5 to 8 kv with emission currents up to 4 ma. The Test Filter was 17.3 Pm Al foil, the same as used for the DC measurements. Both detectors also had 17 pm aluminum foils as the Line Filters for visible light and bremsstrahlung rejection. The response currents without the aluminum Test Filter ranged from 2, pa to 21, pa for both the X-UV and SBD detectors. The response currents with the Test Filter in place ranged from 1, pa to 7, pa. 2 The Test Filter transmission was 2% for both the 5 kv and the 8 kv anode voltage. These results agree reasonably well with the predicted 15% transmission, which supports the conclusion that the detectors were both responding to essentially pure Al K radiation. The measured spectrum (shown in Fig. 2) also indicated quite clean line radiation at 1.5 kev. III.B. AC Response to 4.5 kev Ti K X Rays The modulated Ti K x-ray response at 4.5 kev was measured while running the x-ray generator with a titanium anode at potentials of 8 to 16 kv with emission currents of 1 to 3 ma. The Test Filter was 17.3 pm Al foil, the same as for the DC measurement. Both detectors also had 25 Pm titanium foils as the Line Filters for visible light and bremsstrahlung rejection. The response currents without the aluminum Test Filter ranged from 5, pa to 53, pa for both the X-UV and SBD detectors. The response currents with the Test Filter in place ranged from 2, pa to 21, pa. 2 O The Test Filter transmission varied from 32% for the 8 kv anode voltage to 37% for the 16 kv anode voltage. These results agree reasonably well with the predicted 31% transmission, which supports the conclusion that the detectors were both responding to essentially pure Ti K radiation. The measured spectrum (shown in Fig. 3) also indicated quite clean line radiation at 4.5 kev. III.C. AC Response to 8. kev Cu K X Rays The modulated Cu K x-ray response at 8. kev was measured while running the x-ray generator with a copper anode at potentials of 15 to 2 kv with emission currents of 2 to 48 ma. The Test Filter was 12.5 pm Cu foil, 9

11 which was different than the Test Filter used for the DC measurements. This Test Filter was predicted to transmit 57% of the 8 kev Cu line (Ref. 19). Both detectors also had 25 pm copper foils as the Line Filters for visible light and bremsstrahlung rejection. This Line Filter was predicted to transmit 33% of the Cu K, line. The response currents without the copper Test Filter ranged from about 4,5 pa to about 5, pa for both the X-UV and SBD detectors. The response currents with the Test Filter in place ranged from 3, pa to about 28, pa. At this line energy, the X-UV diodes always exhibited lower signal currents than the SBD. The Test Filter transmission was 55% for both the 15 kv and the 2 kv anode voltage. These results agree well with the predicted 57% transmission, which strongly support the conclusion that the detectors were both responding to essentially pure Cu K radiation. The measured spectrum (shown in Fig. 5) also indicated quite clean line radiation at 8.5 and 9.5 kev. III.D. AC Response to 17.5 kev Mo K X Rays The modulated Mo K x-ray response at 17.5 kev was measured while running the x-ray generator with a molybdenum anode at potentials of Z2 5 kv with emission currents of 2 to 4 ma. The Test Filter was 62 Mm Mo foil, which is predicted to transmit 31% of the 17.5 kev Mo line (using the cross sections of Ref. 21). The response currents without the molybdenum Test Filter were in the range of 1,5 to 7,5 pa for the X-UV diode and 4, to 12, pa for the SBD. At this line energy, the X-UV diodes always exhibited lower signal currents than the SBD. The signal-to-noise ratio was too small to accurately determine the Mo Test Filter transmission, but it was about 3%, close to the predicted value. The measured spectrum (shown in Fig. 6) also indicated quite clean line radiation at 17.5 and 19. kev. IV. Data Analysis and Absolute Response Summary To determine the X-UV absolute response, we simultaneously measured the X-UV and SBD photocurrents as described above. The net x-ray response of each diode was determined by subtracting the "background" current (due to leakage current and visible light) from the total current measured when the x-ray generator was on; I = IT - IB. (1) 1

12 The uncertainty in this net current was obtained by adding in quadrature the uncertainty in IT and IB. The relative X-UV response, px, was then calculated for each measurement; PX = Op, (2) Is where Ix is the net X-UV signal current and Is is the net SBD signal current. The uncertainty in this relative response, up was obtained by propagating the uncertainties in Ix and Is, assuming they were uncorrelated 23, so that IX Xi O (s3) S Is (23 s The most likely value of the relative response was then calculated by taking a weighted average 23 of all the individual relative response values. This average is given by A = Eipx,i/Op,i Px = ~j/, (4) Eil/u,, which weights most heavily those measurements with the smallest individual uncertainty. The uncertainty in the relative response was then assumed to be the standard deviation of all the individual measurements, since the systematic variation from one measurement to another tended to be larger than the uncertainty in any single measurement of px. Finally, the absolute X-UV response was calculated from the response relative to the SBD and the previously measured SBD absolute response, Rs, taken from Refs. 17 and 7; where Rx = Pxs k ax, (5) = Sx 2 r2 ) 1/2 (6) The results from the DC measurements are sumarized in Table 4. X-UV detectors #1 and #2 have nearly the same response at 1.5 and 4.5 kev. Their responses are also very close to the "ideal" response of 4.5 x 17 C/keV (this corresponds to 3.6 ev per charge pair). The response of both diodes decreases with energy faster than the SBD response. X-UV #2 exhibited a faster roll off with energy than X-UV #1. 11

13 The results from the AC measurements are summarized in Table 5. All three diodes were again found to have similar responses at 1.5 and 4.5 kev. However, the AC measurements yielded consistently lower values (up to about 15%) for the X-UV responses than the DC measurements. On the other hand, the uncertainties in the AC and DC results overlap. At 8. kev, the AC results were nearly identical to the DC results. X-UV #2 was again found to have a faster roll off at high energy than both X-UV #1 and X-UV #3 (for which DC measurements were not performed). The measurements at 17.5 kev were consistent with these observations; X-UV #2 had the lowest response of the three detectors, and X-UV #1 and X-UV #3 were very similar to each other. The DC and AC measurements are summarized in Fig. 9. The data are plotted as a function of photon energy. The solid line in each graph represents the calculated response assuming the entire physical silicon thickness forms the active thickness (347 pm for the SBD and 375 pm for the X-UV diodes 2 ). It is clear in these data that the X-UV responses drop with increasing photon energy faster than the physical-thickness model would predict. V. Bias Voltage Effects We examined the effect of varying the bias voltage for three main reasons. First, one of the advertised advantages of silicon X-UV detectors is their ability to run without an applied bias. Second, we have previously seen signal distortions in SBDs when the bias was lowered below about 1 V and the signal levels were quite high (of order 1 pa). 7 Third, changes in the signal level with bias can indicate the influence of the depletion region on the detector response. It was previously believed that the depletion region comprised the active region of silicon detectors. 8 - However, we found in the past that the physical thickness of silicon actually determines the x-ray sensitive region. 7 It was therefore of great interest to determine if varying the bias voltage had any effect on the x-ray signal level. A representative bias scan is shown in Fig. 1. This figure illustrates the X-UV #2 and SBD responses to Cu K x rays as the bias voltage was varied from to -3 V. The net x-ray signal did not change significantly as the reverse bias voltage was changed. This fact supports our earlier contention that it is not the depletion layer thickness that determines the x-ray response of silicon diodes. However, the lack of any bias dependence is somewhat 12

14 surprising, because we have observed, in the past, distortions in SBD x-ray response when the bias was reduced below about 1 V. 7 These distortions were most severe when the x-ray signal was in the 1 to 1 pa range, and they occurred at lower bias voltage as the signal level decreased. We therefore conclude that the bias-dependent distortions were caused by a non-linear, signal-level dependent effect in the detector at high signal levels. Thus, these effects do not hinder unbiased operation at signal levels up to at least tens of na. Although it shows no effect of the bias voltage on the signal level, Fig. 1 does illustrate a different significant effect of the bias voltage. That is that the noise in the system was substantially decreased as the bias voltage was increased. Specifically, for the X-UV data shown in Fig. 1, the noise (peak-to-peak) decreased from about 6.4 na at bias down to 3.2 na at -3 V bias. Similarly, the SBD noise also decreased from about 2.4 na down to 1.6 na as the bias was increased from to -3 V. We always observed larger noise on the X-UV than on the SBD detector. This change in noise characteristics with bias voltage is probably a manifestation of the decreased diode capacitance as the bias voltage is raised. This effect points out that, while zero bias operation is feasible, it may be desirable to operate with some low level of bias to reduce the overall system noise. VI. a-particle Response Measurements We had hoped to determine the sensitive thickness of the X-UV photodiodes by examining their response to a particles, which have a short range in silicon. Specifically, we examined the a-particle response of the X-UV diodes by exposing them to a.13 pci 226la source in air and in vacuum. This source emits four main groups of a particles with energies of 4.7, 5.4, 5.9 and 7.6 MeV. These particles have ranges in silicon of about 22, 26, 3, and 44 pm respectively." The purpose of this experiment was to estimate the "active" thickness of the X-UV detector by observing degraded response to the higher energy a particles. These detectors ostensibly consist of ~375 pm thick silicon 2 5, sufficient to stop all of the a particles. For comparison, we also measured the same a particles with an unbiased SBD, #16-675B (the same detector used for x rays). The SBD is 347 pm thick, also sufficient to completely stop all the a particles. The a-particle spectrum obtained in vacuum with the SBD operated with a bias voltage of -5 V is shown in Fig. 11. This bias voltage is enough to 13

15 create a depletion layer of -1 pm in this detector, approximately twice the range of the most energetic a particle. Charge pairs created by the interaction of the a particle with the silicon are therefore swept to the detector electrodes, rather than diffusing to the electrodes. Since the charge collection time in a depleted region ( ; 1-8 s) is much shorter than a typical minority carrier lifetime (.1 to 1 ms) 6 ' 2 ", essentially all of the charge pairs are collected. The channel numbers where the particles appear in Fig. 11 therefore represent their full energies. The a-particle spectra obtained with X-UV #2 with zero and -3 V bias are shown in Figs. 12a and 12b. The corresponding spectra, obtained with the SBD operated in the same manner are shown in Figs. 13a and 13b. The a particles are barely resolved by the X-UV detector operated with no bias; in contrast, the SBD is able to resolve all of the a particles. In both detectors, however, the particles are degraded from their full energy (by about 35% for the X-UV and by about 13% for the SBD). With a bias voltage of -3 V, the a particles were still only barely resolved by the X-UV detector, and their energies were still significantly degraded. Based on these observations, we conclude that there is significantly more recombination of the charge carriers in the bulk silicon of the X-UV detectors than in the SBD. This is because at zero or low bias voltage, since there is no electric field in the device, the primary mechanism for charge collection is diffusion. 27 The primary mechanism for charge carrier loss is recombination. Since the a particles will have about the same range in the X-UV and SBD detectors, they should deposit their energy at the same depth. Subsequently, the charge carriers created by the a must diffuse the same distance in both devices before they are collected. However, the X-UV detector consistently provided smaller pulses than the SBD, which implies that a larger fraction of the charge carriers was lost to recombination. Because of this effect, we also conclude here that the X-UV photodiodes are not suitable for charged-particle spectroscopy. These diodes may be useful, however, as simple charged particle counters, because as the particles interact they should create a pulse within some range of detectable voltages. VII. Discussion The x-ray response of silicon surface barrier detectors appears to be determined by their physical thickness. 7 At first look, therefore, it is somewhat 14

16 surprising to find that the X-UV photodiodes (nominally 375 pm thick) have lower x-ray sensitivity than the comparison SBD (nominally 347 pm thick) at 8 and 17.5 kev. If the x-ray response were determined solely by the silicon x- ray absorption, as in the case of the SBD, the signal should be approximately proportional to (1 - e"pt), where p is the silicon x-ray mass attenuation coefficient, and p and t are the silicon density and thickness. The X-UV diode and the SBD would then be predicted to exhibit nearly the same x-ray response, with the X-UV response being slightly larger at higher energies (in fact, the relative X-UV response should scale as tx-uv/tsbd for high-energy x rays where ppt is small). The a-particle response of the X-UV photodiode helps to explain its smaller x-ray response at 8 and 17.5 kev. The charged particle data imply that there is enhanced charge carrier recombination in the X-UV as compared to the SBD. This effect will be especially important at low bias voltage, when the charge carriers are collected exclusively by a diffusive process. Recombination will also reduce the relative x-ray response, especially for x rays that deposit much of their energy far from the p - n junction (e.g., 17.5 kev Mo K, radiation). In conclusion, X-UV detectors exhibit x-ray responses similar to an SBD from 1.5 to about 6 kev. At energies above 8 kev, the X-UV x-ray response drops faster than the response of an SBD with similar silicon thickness. We conclude that this drop in high-energy x-ray response is due to (relatively) higher recombination in the bulk silicon of the X-UV diode. This recombination renders this detector unsuitable for charged-particle spectroscopy applications. However, these detectors may still be useful for more simple charged-particle counting. Acknowledgments We thank David Gilbert for his contribution to improving the electronbeam x-ray generator. K. W. also appreciated useful conversations with Raj Korde of International Radiation Detectors. This work was supported in part by NASA Contract No. NASW-4372 and DOE Grant No. DE-FG2-91ER

17 References 1. R. Korde and L. R. Canfield, SPIE (1989). 2. L R. Canfield, J. Kerner, and R. Korde, Applied Optics (1989). 3. L. R. Canfield, J. Kerner, and R. Korde, SPIE (199). 4. G. R. Ricker, M. W. Bautz, et al., in preparation for IEEE Trans. in Nuclear Sciences. 5. United Detector Technology, Hawthorne CA Private communication, Dean Ligocki, UDT Sensors, October (1992). 7. K. W. Wenzel and R. D. Petrasso, Rev. Sci. Instrum (1988). 8. G. F. Knoll, Radiation Detection and Measurement, Wiley, New York (1989). 9. M. Malacarne and I. H. Hutchinson, Culham Report CLM-R227, (1982). 1. J. Kiraly, M. Bitter, P. Efthimion, S. von Goeler, B. Grek, K. W. Hill, D. Johnson, K. McGuire, N. Sauthoff, S. Sesnic, F. Stauffer, G. Tait, G. Taylor, Nucl. Fusion (1987). 11. D. H. Lo, R. D. Petrasso, K. W. Wenzel, J. W. Coleman, C. K. Li, J. R. Lierzer, and T. Wei, Bull. Am. Phys. Soc (1991) and Rev. Sci. Instrum (1992). 12. In general the importance of the depletion region in determining detector response depends on the mode of operation. Radiation can interact in a partially-depleted detector either with depleted or undepleted silicon. Charge pairs created in the depletion region and accelerated to the electrodes will contribute to the signal. Charge pairs created in the undepleted region may or may not contribute to the signal, depending on the rate of diffusion, the rate of carrier recombination, and the integration time of the collection circuit (e.g., whether the system is operated in current or pulse mode). 16

18 13. D. H. Lo, K. W. Wenzel, R. D. Petrasso, J. Doty, and G. Ricker in preparation for Rev. Sci. Instrum. 14. C. K. Li, K. W. Wenzel, R. D. Petrasso, J. W. Coleman, and D. H. Lo, Bull. Am. Phys. Soc (199). 15. K. W. Wenzel, J. W. Coleman, R. D. Petrasso, C. K. Li, and D. H. Lo, Bull. Am. Phys. Soc (199). 16. EG&G ORTEC, Oak Ridge, TN R. D. Petrasso, M. Gerassimenko, F. H. Seguin, R. Krogstad, P. Gauthier, W. Hamilton, A. T. Ramsey, P. Burstein, and R. Granetz, Rev. Sci. Instrum., (198). 18. Keithley Instruments, Cleveland, OH B. L. Henke, P. Lee, T. J. Tanaka, R. L. Shimabukuro, and B. K. Fujikawa, Atomic Data and Nucl. Data Tables, 27 1 (1982). 2. In all cases, the response currents quoted with and without the Test Filter are given only to indicate typical signal levels. The high and low response currents were not necessarily obtained under the same source conditions, so the Test Filter transmission cannot be calculated from these representative values. The filter transmission results were obtained by comparing signal levels under identical source conditions. 21. W. H. McMaster, N. Kerr Del Grande, J. H. Mallet, and J. H. Hubbell, UCRL-5174-SEC-2-R1, (1969). 22. P. Horowitz and W. Hill, The Art of Electronics, Cambridge University Press (198). 23. P. R. Bevington, Data Reduction and Error Analysis for the Physical Sciences, McGraw-Hill, New York (1969). 24. The SBD response at 4.5 and 5.9 kev has not been measured. It is assumed to be close to the "ideal" response (3.6 ev per charge pair) since both the 1.5 and 8 kev measurements are close to this value (see Tables 4 and 5). 17

19 25. Private communication with Matthew Rabdau, United Detector Technology, March (1992). 26. F. S. Goulding, private communication (1988). 27. R. F. Pierret, Semiconductor Fundamentals, Addison-Wesley, Reading MA (1983). 18

20 Table 1: Three X-UV detectors were examined in this work. Detectors #1 and #2 were both Model X-UV35, and detector #3 was Model X-UV1C. All three detectors were manufactured by United Detector Technology, and some of their nominal characteristics are shown here. 5 '" Because this work examined x-ray responses for photon energies >1.5 kev, the resistivity and minority carrier lifetime values are for the bulk, p-type silicon where the majority of the x-rays are stopped. The characteristics of the SBD are also listed for comparison. Detector Model Active Capacitance Resistivity Minority Carrier I I Area (cm2) (nf) (Q-cm) Lifetime (ps) 1 X-UV X-UV X-UV1C SBD R Table 2: This table lists the materials and thicknesses used for the moveable Test Filter and the Line Filter during the DC measurements. In order to check the spectral purity of the source, the Test Filter was moved in between the source and detectors. The Line Filters, which were the same material as the anode, were attached to the front of the detectors as light and bremsstrahlung filters. Anode K, Line Test Line Material I Energy Filter Filter Al 1.5 kev 17 pm Al 17 pm Al Ti 4.5 kev 17 pm Al 25 pm Ti Cu 8. kev 86 pm Al 5 pm Cu Mo 17.5 kev 86 pm Al 124 pm Mo 19

21 Table 3: This table lists the materials and thicknesses used as the Test and Line Filters during the AC measurements. For the AC measurements, the Test Filters covered every other aperture in the eight-apertured chopper wheel, so that alternate x-ray pulses were filtered. The Line Filters func- tioned as light and bremsstrahlung filters. fanode K, Line Test Line Material Energy Filter Filter Al 1.5 kev 17 pm Al 17 pm Al Ti 4.5 kev 17 pm Al 25 pm Ti Cu 8. kev 13 pm Cu 25 pm Cu Mo 17.5 kev 62 pm Mo 124 pm Mo Table 4: This table summarizes the DC x-ray response measurements for X-UV diodes #1 and #2. Rs is the absolute SBD response (previously measured',) and R1, 2 is the absolute response of X-UV diode #1 and #2 respectively [from Eq. (5)]. All absolute response (R) values are listed in the units of 1-l 7 C/keV; 4.5x1-7 C/keV corresponds to 3.6 ev per charge pair. The uncertainties were calculated as described in Section IV. E IpV P2 Rs (kev) ± ±.1 3.8± ± ± ±.2 1.5±.5 4.5± ± ± ± ± ± ±.1.54±.6 3.7± ±.73 2.± ±.3.44± :1, 1.28it:2.88'45 2

22 Table 5: This table summarizes the AC x-ray response measurements for the three X-UV diodes listed in Table 1. Rs is the absolute SBD response (previously measured 7 ' 5 ) and R 1, 2, 3 is the absolute response of X-UV diode #1, #2, and #3 respectively [from Eq. (5)]. All absolute response (R) values are listed in the units of 1- C/keV; 4.5x1-1 7 C/keV corresponds to 3.6 ev per charge pair. The uncertainties were calculated as described in Section IV. (kev) P P2 P3 Rs R 1 R 2 R ±.9.96±.9.9±.1 3.8± ± ± ± ±.7.85±.6.89± ± ± ± ± ±.7.56±.3.73±.6 3.7±.8 2.7± ± ± ±.1.5±.1.7±.1 2.+g:g 1.2+g:l 1.-:4 1.4Ig: 21

23 k#- 1M FILAMENT POWER MorNITOR SBD HIGH VOLTAGE P1\ WER SUPPLY TEST FILTER TARGET UU txx-u FEEDTHROUGH BREAK f TFILAMEPNT PUM GATE VALVE TURBO PUMP ROUGHING PUMP 1 ROUGHING PUMP 2 Fig. 1. This schematic shows the experimental setup, including the moveable "Test Filter". Table 1 lists specific materials and thicknesses used for both the Line and the Test Filters for each calibration energy (i.e., anode material). The anode target runs at positive high voltage and the filament at ground potential. A Si(Li) spectrometer could be installed at the feedthrough position after the X-UV and SBD detectors were removed. This drawing is not to scale; the dimension is meant to indicate the overall size of the device. 22

24 C : Energy (kev) Fig. 2. The spectrum from an aluminum anode (8 kv, 1 ma emission current) was obtained with an EG&G ORTEC" 6 Si(Li) spectrometer. The spectrometer viewed the source through a 17 pm Al filter, which is the same thickness as the Line Filter used on the X-UV diode and the SBD. The emission current of the x-ray source was lower than for the diode measurements to reduce the count rate to a desirable level. 23

25 C Energy (kev) Fig. 3. The spectrum from a titanium anode (16 kv, 1 ma) was obtained through a 25 pm Ti filter, which is the same thickness as the Line Filter used on the X-UV diode and the SBD. The emission current was lower than for the diode measurements to reduce the count rate to a desirable level. 24

26 n 2 - o Energy (kev) Fig. 4. Manganese K x rays (from a 55 Fe source) were measured with the Si(Li) detector. The detector viewed the source through a.24 cm diameter aperture and 12.5 pm copper foil to reduce the count rate to desirable levels. 25

27 C Energy (kev) Fig. 5. The spectrum from a copper anode (2 kv, 1 ma) was obtained with the Si(Li) spectrometer. The spectrometer viewed the source through a 25 pm Cu filter, which is the same as the Line Filter used on the X-UV diode and the SBD during the AC measurements. For the DC measurements, we used a thicker copper Line filter, which would create a cleaner line spectrum (compare Tables 2 and 3). 26

28 4 I I I I I I I I I I 1 1 1,,,,,, I 3 C Energy (kev) Fig. 6. The spectrum from a molybdenum anode (25 kv, 1 ma) was obtained through a 124 pm Mo filter, which is the same as the Line Filter used on the X-UV diode and the SBD. The emission current was lower than for the diode measurements to reduce the count rate to a desirable level. 27

29 5 pf Bias 1 kq 1 MQ 2 jx 1 MQ F x-ray op Fig. 7. The simple AC-coupled transresistance current-to-voltage amplifier was used to measure the x-ray response while various bias voltages could be applied to the detectors. The 1 MQ feedback resistor resulted in a gain of 1 mv/na, and the 5 pf capacitor filtered out noise with frequencies above about 2 khz. 28

30 a) Al 1 11 Ib) Ti 1 I I 1 l l ) c)cuv - (~\IBD ~XfNFYUf C) CU X-Uv - SBD d) Mo - X-UV- SBD - SBD a I I I I I Fig. 8. Data from each of the anode materials is shown. (Table 3 lists the Line and Test Filters used for each measurement, and Figs. 3, 4, 6, and 7 show the corresponding spectra.) a) These aluminum data were taken with X-UV #1 while the generator operated at 8 kv and 4 ma; the scale is 1 mv and 5 ms per division. b) These titanium data were taken with X-UV #1 while the generator operated at 12 kv and 28 ma; the scale is 1 mv and 5 ms per division. c) These copper data were taken with X-UV #1 while the generator operated at 2 kv and 2 ma; the scale is 1 mv and 5 ms per division. d) These molybdenum data were taken with X-UV #1 while the generator operated at about 25 kv and 4 ma; the scale is 1 mv and 5 ms per division. 29

31 1, a) SBD b) XUV #1 4 -j-. a-.1 *C) XUV #2 TV ; 7T d) XUV #3 C Photon Energy (kev). I 1 1 Photon Energy (kev) Fig. 9. The measured responses of the SBD" (from Refs. 17 and 7) and the X-UV detector are plotted as a function of photon energy. In the X-UV data, squares with narrow error bars represent the DC measurements (Table 4), and circles with wider error bars represent the AC measurements (Table 5). The solid lines are the calculated responses assuming the entire silicon thickness forms the sensitive thickness (347 Mm for the SBD 7 and 375 pm for the X-UV 2 "). The low energy response is determined by the front surface of the devices. The SBDs have 15 A aluminum front surface, and were assumed to have an additional 2 A SiO 2 dead layer. The X-UV diodes were assumed to have only a 2 A SiO 2 dead layer. 3

32 a) I I I b) -1V 1V -SBD - SBD c) -2 V I III I li- P d) -3 V 2WWMJKI SBD a- SBD -. I, II I II I Fig. 1. The SBD and X-UV #2 detector responses to Cu K x rays are shown as a function of bias voltage. For all four sets of data, the x-ray generator was run with the anode at 2 kv and with 2 ma of emission current. The Line Filter used to obtain these data was 25 pim Cu, and the Test Filter was 12.5 pm Cu. Plots a) through d) illustrate the x-ray response obtained using the circuit of Fig. 7, as the bias was varied from V to -3 V. Neither the X-UV nor the SBD x-ray response changed significantly as the bias voltage was raised. However, the noise in the X-UV response decreased from about 6.4 na (peak-to-peak) to 3.2 na. Similarly, the noise in the SBD signal also decreased from about 2.4 na at zero bias to less than 1.6 na at 3 V bias. 31

33 35 SBD 3 - V IUS C :3 o Channel Fig. 11. The a-particle spectrum from a 22 a source was obtained with a partially-depleted SBD (#16-675B, the same SBD used for the x- ray measurements) biased at -5V bias voltage. Because this bias voltage creates a depletion layer thicker than the longest range of the a particles, the channel locations of the particles represents their full energy. 32

34 25 - a) XUV #2 V bias 2 - C = b) XUV #2-3 V bias L Channel Fig. 12. These spectra of the 22 6 Ra a emission were obtained with X-UV #2. a) With zero bias applied to the detector, the a particles with different energy could not be resolved. b) With 3 volts bias applied to the detector, the individual a particles could be barely resolved, but the collected charge from each particle was significantly degraded as compared to the SBD data shown in Fig

35 a) SBD V bias 2 - C b) SBD -3 V bias 2 - C Channel Fig. 13. These spectra of the PRa a emission were obtained with the same SBD used to obtained the spectra in Fig. 11. a) With zero bias applied to the detector, the full energy and resolution of the a particles was somewhat degraded, but the individual particles could still be resolved. b) With 3 volts bias applied to the detector, the a-particle spectrum is slightly improved over the spectrum taken with zero bias. The important feature is that even with low bias, all the a particles could be distinguished, whereas the X-UV diodes could not resolve the particles with a small or zero bias voltage appplied. 34

Soft X-Ray Silicon Photodiodes with 100% Quantum Efficiency

Soft X-Ray Silicon Photodiodes with 100% Quantum Efficiency PFC/JA-94-4 Soft X-Ray Silicon Photodiodes with 1% Quantum Efficiency K. W. Wenzel, C. K. Li, D. A. Pappas, Raj Kordel MIT Plasma Fusion Center Cambridge, Massachusetts 2139 USA March 1994 t Permanent

More information

Title detector with operating temperature.

Title detector with operating temperature. Title Radiation measurements by a detector with operating temperature cryogen Kanno, Ikuo; Yoshihara, Fumiki; Nou Author(s) Osamu; Murase, Yasuhiro; Nakamura, Masaki Citation REVIEW OF SCIENTIFIC INSTRUMENTS

More information

RANDY W. ALKIRE, GEROLD ROSENBAUM AND GWYNDAF EVANS

RANDY W. ALKIRE, GEROLD ROSENBAUM AND GWYNDAF EVANS S-94,316 PATENTS-US-A96698 BEAM POSITION MONITOR RANDY W. ALKIRE, GEROLD ROSENBAUM AND GWYNDAF EVANS CONTRACTUAL ORIGIN OF THE INVENTION The United States Government has rights in this invention pursuant

More information

Measurements of MeV Photon Flashes in Petawatt Laser Experiments

Measurements of MeV Photon Flashes in Petawatt Laser Experiments UCRL-JC-131359 PREPRINT Measurements of MeV Photon Flashes in Petawatt Laser Experiments M. J. Moran, C. G. Brown, T. Cowan, S. Hatchett, A. Hunt, M. Key, D.M. Pennington, M. D. Perry, T. Phillips, C.

More information

Physics Laboratory Scattering of Photons from Electrons: Compton Scattering

Physics Laboratory Scattering of Photons from Electrons: Compton Scattering RR Oct 2001 SS Dec 2001 MJ Oct 2009 Physics 34000 Laboratory Scattering of Photons from Electrons: Compton Scattering Objective: To measure the energy of high energy photons scattered from electrons in

More information

Energy Measurements with a Si Surface Barrier Detector and a 5.5-MeV 241 Am α Source

Energy Measurements with a Si Surface Barrier Detector and a 5.5-MeV 241 Am α Source Energy Measurements with a Si Surface Barrier Detector and a 5.5-MeV 241 Am α Source October 18, 2017 The goals of this experiment are to become familiar with semiconductor detectors, which are widely

More information

Partial Replication of Storms/Scanlan Glow Discharge Radiation

Partial Replication of Storms/Scanlan Glow Discharge Radiation Partial Replication of Storms/Scanlan Glow Discharge Radiation Rick Cantwell and Matt McConnell Coolescence, LLC March 2008 Introduction The Storms/Scanlan paper 1 presented at the 8 th international workshop

More information

Today s Outline - January 25, C. Segre (IIT) PHYS Spring 2018 January 25, / 26

Today s Outline - January 25, C. Segre (IIT) PHYS Spring 2018 January 25, / 26 Today s Outline - January 25, 2018 C. Segre (IIT) PHYS 570 - Spring 2018 January 25, 2018 1 / 26 Today s Outline - January 25, 2018 HW #2 C. Segre (IIT) PHYS 570 - Spring 2018 January 25, 2018 1 / 26 Today

More information

14.2 Photodiodes 411

14.2 Photodiodes 411 14.2 Photodiodes 411 Maximum reverse voltage is specified for Ge and Si photodiodes and photoconductive cells. Exceeding this voltage can cause the breakdown and severe deterioration of the sensor s performance.

More information

ANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER

ANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 325 ANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER ABSTRACT William Chang, Jonathan Kerner, and Edward

More information

Interpixel crosstalk in a 3D-integrated active pixel sensor for x-ray detection

Interpixel crosstalk in a 3D-integrated active pixel sensor for x-ray detection Interpixel crosstalk in a 3D-integrated active pixel sensor for x-ray detection The MIT Faculty has made this article openly available. Please share how this access benefits you. Your story matters. Citation

More information

Role of guard rings in improving the performance of silicon detectors

Role of guard rings in improving the performance of silicon detectors PRAMANA c Indian Academy of Sciences Vol. 65, No. 2 journal of August 2005 physics pp. 259 272 Role of guard rings in improving the performance of silicon detectors VIJAY MISHRA, V D SRIVASTAVA and S K

More information

Citation X-Ray Spectrometry (2011), 40(4): 2. Right final form at

Citation X-Ray Spectrometry (2011), 40(4): 2.   Right final form at TitleSi PIN X-ray photon counter Author(s) Nakaye, Yasukazu; Kawai, Jun Citation X-Ray Spectrometry (2011), 40(4): 2 Issue Date 2011-03-24 URL http://hdl.handle.net/2433/197743 This is the peer reviewed

More information

Optodevice Data Book ODE I. Rev.9 Mar Opnext Japan, Inc.

Optodevice Data Book ODE I. Rev.9 Mar Opnext Japan, Inc. Optodevice Data Book ODE-408-001I Rev.9 Mar. 2003 Opnext Japan, Inc. Section 1 Operating Principles 1.1 Operating Principles of Laser Diodes (LDs) and Infrared Emitting Diodes (IREDs) 1.1.1 Emitting Principles

More information

Doppler-Free Spetroscopy of Rubidium

Doppler-Free Spetroscopy of Rubidium Doppler-Free Spetroscopy of Rubidium Pranjal Vachaspati, Sabrina Pasterski MIT Department of Physics (Dated: April 17, 2013) We present a technique for spectroscopy of rubidium that eliminates doppler

More information

Gamma Spectrometer Initial Project Proposal

Gamma Spectrometer Initial Project Proposal Gamma Spectrometer Initial Project Proposal Group 9 Aman Kataria Johnny Klarenbeek Dean Sullivan David Valentine Introduction There are currently two main types of gamma radiation detectors used for gamma

More information

LAB V. LIGHT EMITTING DIODES

LAB V. LIGHT EMITTING DIODES LAB V. LIGHT EMITTING DIODES 1. OBJECTIVE In this lab you will measure the I-V characteristics of Infrared (IR), Red and Blue light emitting diodes (LEDs). Using a photodetector, the emission intensity

More information

X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage

X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage X-rays Ouch! 1 X-rays X-rays are produced when electrons are accelerated and collide with a target Bremsstrahlung x-rays Characteristic x-rays X-rays are sometimes characterized by the generating voltage

More information

Silicon Photodiodes - SXUV Series with Platinum Silicide Front Entrance Windows

Silicon Photodiodes - SXUV Series with Platinum Silicide Front Entrance Windows Silicon Photodiodes - SXUV Series with Platinum Silicide Front Entrance Windows SXUV Responsivity Stability It is known that the UV photon exposure induced instability of common silicon photodiodes is

More information

Author(s) Osamu; Nakamura, Tatsuya; Katagiri,

Author(s) Osamu; Nakamura, Tatsuya; Katagiri, TitleCryogenic InSb detector for radiati Author(s) Kanno, Ikuo; Yoshihara, Fumiki; Nou Osamu; Nakamura, Tatsuya; Katagiri, Citation REVIEW OF SCIENTIFIC INSTRUMENTS (2 2533-2536 Issue Date 2002-07 URL

More information

LAB V. LIGHT EMITTING DIODES

LAB V. LIGHT EMITTING DIODES LAB V. LIGHT EMITTING DIODES 1. OBJECTIVE In this lab you are to measure I-V characteristics of Infrared (IR), Red and Blue light emitting diodes (LEDs). The emission intensity as a function of the diode

More information

Report on BLP Spectroscopy Experiments Conducted on October 6, 2017: M. Nansteel

Report on BLP Spectroscopy Experiments Conducted on October 6, 2017: M. Nansteel Report on BLP Spectroscopy Experiments Conducted on October 6, 2017: M. Nansteel Summary Several spectroscopic measurements were conducted on October 6, 2017 at BLP to characterize the radiant power of

More information

Department of Electrical Engineering IIT Madras

Department of Electrical Engineering IIT Madras Department of Electrical Engineering IIT Madras Sample Questions on Semiconductor Devices EE3 applicants who are interested to pursue their research in microelectronics devices area (fabrication and/or

More information

Instrumentation Development for a Novel Local Electric and Magnetic Field Fluctuation Diagnostic

Instrumentation Development for a Novel Local Electric and Magnetic Field Fluctuation Diagnostic Instrumentation Development for a Novel Local Electric and Magnetic Field Fluctuation Diagnostic Mindy Bakken On behalf of: R.J. Fonck, M.G. Burke, B.T. Lewicki, A.T. Rhodes, G.R. Winz 58 th Annual Meeting

More information

Detection and application of Doppler and motional Stark features in the DNB emission spectrum in the high magnetic field of the Alcator C-Mod tokamak

Detection and application of Doppler and motional Stark features in the DNB emission spectrum in the high magnetic field of the Alcator C-Mod tokamak Detection and application of Doppler and motional Stark features in the DNB emission spectrum in the high magnetic field of the Alcator C-Mod tokamak I. O. Bespamyatnov a, W. L. Rowan a, K. T. Liao a,

More information

Semiconductor Detector Systems

Semiconductor Detector Systems Semiconductor Detector Systems Helmuth Spieler Physics Division, Lawrence Berkeley National Laboratory OXFORD UNIVERSITY PRESS ix CONTENTS 1 Detector systems overview 1 1.1 Sensor 2 1.2 Preamplifier 3

More information

CHAPTER 11 HPD (Hybrid Photo-Detector)

CHAPTER 11 HPD (Hybrid Photo-Detector) CHAPTER 11 HPD (Hybrid Photo-Detector) HPD (Hybrid Photo-Detector) is a completely new photomultiplier tube that incorporates a semiconductor element in an evacuated electron tube. In HPD operation, photoelectrons

More information

Initial Results from a Cryogenic Proton Irradiation of a p-channel CCD

Initial Results from a Cryogenic Proton Irradiation of a p-channel CCD Centre for Electronic Imaging Initial Results from a Cryogenic Proton Irradiation of a p-channel CCD Jason Gow Daniel Wood, David Hall, Ben Dryer, Simeon Barber, Andrew Holland and Neil Murray Jason P.

More information

Photodiode Characteristics and Applications

Photodiode Characteristics and Applications Photodiode Characteristics and Applications Silicon photodiodes are semiconductor devices responsive to highenergy particles and photons. Photodiodes operate by absorption of photons or charged particles

More information

ABSTRACT. Supported by U.S. DoE grant No. DE-FG02-96ER54375

ABSTRACT. Supported by U.S. DoE grant No. DE-FG02-96ER54375 ABSTRACT A CCD imaging system is currently being developed for T e (,t) and bolometric measurements on the Pegasus Toroidal Experiment. Soft X-rays (E

More information

Real Time Pulse Pile-up Recovery in a High Throughput Digital Pulse Processor

Real Time Pulse Pile-up Recovery in a High Throughput Digital Pulse Processor Real Time Pulse Pile-up Recovery in a High Throughput Digital Pulse Processor Paul A. B. Scoullar a, Chris C. McLean a and Rob J. Evans b a Southern Innovation, Melbourne, Australia b Department of Electrical

More information

ALMY Stability. Kevan S Hashemi and James R Bensinger Brandeis University January 1998

ALMY Stability. Kevan S Hashemi and James R Bensinger Brandeis University January 1998 ATLAS Internal Note MUON-No-221 ALMY Stability Kevan S Hashemi and James R Bensinger Brandeis University January 1998 Introduction An ALMY sensor is a transparent, position-sensitive, optical sensor made

More information

D. F. Spencer R. Aryaeinejad E. L. Reber. October 2001

D. F. Spencer R. Aryaeinejad E. L. Reber. October 2001 INEEL/CON-01-01424 PREPRINT Using The Cockroft-Walton Voltage Multiplier Design In Handheld Devices D. F. Spencer R. Aryaeinejad E. L. Reber October 2001 Nuclear Science & Medical Imaging Symposium This

More information

The Absolute Detection Efficiency of Vacuum Electron Multipliers, to kev Protons and Ar + Ions.

The Absolute Detection Efficiency of Vacuum Electron Multipliers, to kev Protons and Ar + Ions. The Absolute Detection Efficiency of Vacuum Electron Multipliers, to kev Protons and Ar + Ions. S. Shchemelinin 1, S. Pszona 2, G. Garty 1,, A. Breskin 1, R. Chechik 1 1 Department of Particle Physics,

More information

High collection efficiency MCPs for photon counting detectors

High collection efficiency MCPs for photon counting detectors High collection efficiency MCPs for photon counting detectors D. A. Orlov, * T. Ruardij, S. Duarte Pinto, R. Glazenborg and E. Kernen PHOTONIS Netherlands BV, Dwazziewegen 2, 9301 ZR Roden, The Netherlands

More information

PERFORMANCE OF THE CMS ECAL LASER MONITORING SOURCE IN THE TEST BEAM

PERFORMANCE OF THE CMS ECAL LASER MONITORING SOURCE IN THE TEST BEAM PERFORMANCE OF THE CMS ECAL LASER MONITORING SOURCE IN THE TEST BEAM A. BORNHEIM CALTECH 2 E. California Blvd., Pasadena, CA 925, USA E-mail: bornheim@hep.caltech.edu On behalf of the CMS ECAL Collaboration.

More information

Detailed Characterisation of a New Large Area CCD Manufactured on High Resistivity Silicon

Detailed Characterisation of a New Large Area CCD Manufactured on High Resistivity Silicon Detailed Characterisation of a New Large Area CCD Manufactured on High Resistivity Silicon Mark S. Robbins *, Pritesh Mistry, Paul R. Jorden e2v technologies Ltd, 106 Waterhouse Lane, Chelmsford, Essex

More information

Performance characteristics of a new wide range, fast settling electrometer design for a residual gas analysis mass spectrometer

Performance characteristics of a new wide range, fast settling electrometer design for a residual gas analysis mass spectrometer Performance characteristics of a new wide range, fast settling electrometer design for a residual gas analysis mass spectrometer MKS Spectra Products, January 2010 Design considerations for RGA components

More information

A New Single-Photon Avalanche Diode in 90nm Standard CMOS Technology

A New Single-Photon Avalanche Diode in 90nm Standard CMOS Technology A New Single-Photon Avalanche Diode in 90nm Standard CMOS Technology Mohammad Azim Karami* a, Marek Gersbach, Edoardo Charbon a a Dept. of Electrical engineering, Technical University of Delft, Delft,

More information

INFRARED MEASUREMENTS OF THE SYNTHETIC DIAMOND WINDOW OF A 110 GHz HIGH POWER GYROTRON

INFRARED MEASUREMENTS OF THE SYNTHETIC DIAMOND WINDOW OF A 110 GHz HIGH POWER GYROTRON GA A23723 INFRARED MEASUREMENTS OF THE SYNTHETIC DIAMOND WINDOW by I.A. GORELOV, J. LOHR, R.W. CALLIS, W.P. CARY, D. PONCE, and M.B. CONDON JULY 2001 This report was prepared as an account of work sponsored

More information

Amorphous Selenium Direct Radiography for Industrial Imaging

Amorphous Selenium Direct Radiography for Industrial Imaging DGZfP Proceedings BB 67-CD Paper 22 Computerized Tomography for Industrial Applications and Image Processing in Radiology March 15-17, 1999, Berlin, Germany Amorphous Selenium Direct Radiography for Industrial

More information

ORTEC. Research Applications. Pulse-Height, Charge, or Energy Spectroscopy. Detectors. Processing Electronics

ORTEC. Research Applications. Pulse-Height, Charge, or Energy Spectroscopy. Detectors. Processing Electronics ORTEC Spectroscopy systems for ORTEC instrumentation produce pulse height distributions of gamma ray or alpha energies. MAESTRO-32 (model A65-B32) is the software included with most spectroscopy systems

More information

photolithographic techniques (1). Molybdenum electrodes (50 nm thick) are deposited by

photolithographic techniques (1). Molybdenum electrodes (50 nm thick) are deposited by Supporting online material Materials and Methods Single-walled carbon nanotube (SWNT) devices are fabricated using standard photolithographic techniques (1). Molybdenum electrodes (50 nm thick) are deposited

More information

Application of CMOS sensors in radiation detection

Application of CMOS sensors in radiation detection Application of CMOS sensors in radiation detection S. Ashrafi Physics Faculty University of Tabriz 1 CMOS is a technology for making low power integrated circuits. CMOS Complementary Metal Oxide Semiconductor

More information

Chemistry Instrumental Analysis Lecture 10. Chem 4631

Chemistry Instrumental Analysis Lecture 10. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 10 Types of Instrumentation Single beam Double beam in space Double beam in time Multichannel Speciality Types of Instrumentation Single beam Requires stable

More information

Radiation Detection Instrumentation

Radiation Detection Instrumentation Radiation Detection Instrumentation Principles of Detection and Gas-filled Ionization Chambers Neutron Sensitive Ionization Chambers Detection of radiation is a consequence of radiation interaction with

More information

Evaluating the Performance of a Commercial Silicon Drift Detector for X-ray Microanalysis

Evaluating the Performance of a Commercial Silicon Drift Detector for X-ray Microanalysis Evaluating the Performance of a Commercial Silicon Drift Detector for X-ray Microanalysis Edward A. Kenik Materials Science & Technology Division, Oak Ridge National Laboratory, Oak Ridge, TN 37831 kenikea@ornl.gov

More information

XRF Instrumentation. Introduction to spectrometer

XRF Instrumentation. Introduction to spectrometer XRF Instrumentation Introduction to spectrometer AMPTEK, INC., Bedford, MA 01730 Ph: +1 781 275 2242 Fax: +1 781 275 3470 sales@amptek.com 1 Instrument Excitation source Sample X-ray tube or radioisotope

More information

Chapter Semiconductor Electronics

Chapter Semiconductor Electronics Chapter Semiconductor Electronics Q1. p-n junction is said to be forward biased, when [1988] (a) the positive pole of the battery is joined to the p- semiconductor and negative pole to the n- semiconductor

More information

Chapter 3 OPTICAL SOURCES AND DETECTORS

Chapter 3 OPTICAL SOURCES AND DETECTORS Chapter 3 OPTICAL SOURCES AND DETECTORS 3. Optical sources and Detectors 3.1 Introduction: The success of light wave communications and optical fiber sensors is due to the result of two technological breakthroughs.

More information

LEDs, Photodetectors and Solar Cells

LEDs, Photodetectors and Solar Cells LEDs, Photodetectors and Solar Cells Chapter 7 (Parker) ELEC 424 John Peeples Why the Interest in Photons? Answer: Momentum and Radiation High electrical current density destroys minute polysilicon and

More information

Week 9: Chap.13 Other Semiconductor Material

Week 9: Chap.13 Other Semiconductor Material Week 9: Chap.13 Other Semiconductor Material Exam Other Semiconductors and Geometries -- Why --- CZT properties -- Silicon Structures --- CCD s Gamma ray Backgrounds The MIT Semiconductor Subway (of links

More information

Spectrophotometer. An instrument used to make absorbance, transmittance or emission measurements is known as a spectrophotometer :

Spectrophotometer. An instrument used to make absorbance, transmittance or emission measurements is known as a spectrophotometer : Spectrophotometer An instrument used to make absorbance, transmittance or emission measurements is known as a spectrophotometer : Spectrophotometer components Excitation sources Deuterium Lamp Tungsten

More information

Components of Optical Instruments

Components of Optical Instruments Components of Optical Instruments General Design of Optical Instruments Sources of Radiation Wavelength Selectors (Filters, Monochromators, Interferometers) Sample Containers Radiation Transducers (Detectors)

More information

Atomic and nuclear physics

Atomic and nuclear physics Atomic and nuclear physics X-ray physics Physics of the atomic shell LEYBOLD Physics Leaflets Investigating the energy spectrum of an x-ray tube as a function of the high voltage and the emission current

More information

How Does One Obtain Spectral/Imaging Information! "

How Does One Obtain Spectral/Imaging Information! How Does One Obtain Spectral/Imaging Information! How do we measure the position, energy, and arrival time of! an X-ray photon?! " What we observe depends on the instruments that one observes with!" In

More information

Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC

Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC R. Bellazzini a,b, G. Spandre a*, A. Brez a, M. Minuti a, M. Pinchera a and P. Mozzo b a INFN Pisa

More information

Physics of Waveguide Photodetectors with Integrated Amplification

Physics of Waveguide Photodetectors with Integrated Amplification Physics of Waveguide Photodetectors with Integrated Amplification J. Piprek, D. Lasaosa, D. Pasquariello, and J. E. Bowers Electrical and Computer Engineering Department University of California, Santa

More information

Recent Development and Study of Silicon Solid State Photomultiplier (MRS Avalanche Photodetector)

Recent Development and Study of Silicon Solid State Photomultiplier (MRS Avalanche Photodetector) Recent Development and Study of Silicon Solid State Photomultiplier (MRS Avalanche Photodetector) Valeri Saveliev University of Obninsk, Russia Vienna Conference on Instrumentation Vienna, 20 February

More information

Figure Responsivity (A/W) Figure E E-09.

Figure Responsivity (A/W) Figure E E-09. OSI Optoelectronics, is a leading manufacturer of fiber optic components for communication systems. The products offer range for Silicon, GaAs and InGaAs to full turnkey solutions. Photodiodes are semiconductor

More information

New Detectors for X-Ray Metal Thickness Measuring

New Detectors for X-Ray Metal Thickness Measuring ECNDT 2006 - Poster 132 New Detectors for X-Ray Metal Thickness Measuring Boris V. ARTEMIEV, Alexander I. MASLOV, Association SPEKTR- GROUP, Moscow, Russia Abstract. X-ray thickness measuring instruments

More information

236 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 59, NO. 1, FEBRUARY 2012

236 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 59, NO. 1, FEBRUARY 2012 236 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 59, NO. 1, FEBRUARY 2012 Characterization of the H3D ASIC Readout System and 6.0 cm 3-D Position Sensitive CdZnTe Detectors Feng Zhang, Cedric Herman, Zhong

More information

Electronic devices-i. Difference between conductors, insulators and semiconductors

Electronic devices-i. Difference between conductors, insulators and semiconductors Electronic devices-i Semiconductor Devices is one of the important and easy units in class XII CBSE Physics syllabus. It is easy to understand and learn. Generally the questions asked are simple. The unit

More information

UNIT VIII-SPECIAL PURPOSE ELECTRONIC DEVICES. 1. Explain tunnel Diode operation with the help of energy band diagrams.

UNIT VIII-SPECIAL PURPOSE ELECTRONIC DEVICES. 1. Explain tunnel Diode operation with the help of energy band diagrams. UNIT III-SPECIAL PURPOSE ELECTRONIC DEICES 1. Explain tunnel Diode operation with the help of energy band diagrams. TUNNEL DIODE: A tunnel diode or Esaki diode is a type of semiconductor diode which is

More information

Bias errors in PIV: the pixel locking effect revisited.

Bias errors in PIV: the pixel locking effect revisited. Bias errors in PIV: the pixel locking effect revisited. E.F.J. Overmars 1, N.G.W. Warncke, C. Poelma and J. Westerweel 1: Laboratory for Aero & Hydrodynamics, University of Technology, Delft, The Netherlands,

More information

ORTEC Experiment 13. Gamma-Gamma Coincidence with Angular Correlation. Equipment Required

ORTEC Experiment 13. Gamma-Gamma Coincidence with Angular Correlation. Equipment Required ORTEC Experiment 13 Equipment Required Two 905-3 2-in. x 2-in. NaI(Tl) Scintillation Detector Assemblies. Two 266 Photomultiplier Tube Bases. Two 113 Scintillation Preamplifiers. Two 556 High Voltage Power

More information

Photons and solid state detection

Photons and solid state detection Photons and solid state detection Photons represent discrete packets ( quanta ) of optical energy Energy is hc/! (h: Planck s constant, c: speed of light,! : wavelength) For solid state detection, photons

More information

UV/EUV CONTINUOUS POSITION SENSOR

UV/EUV CONTINUOUS POSITION SENSOR UV/EUV CONTINUOUS POSITION SENSOR ODD-SXUV-DLPSD FEATURES Submicron position resolution Stable response after exposure to UV/EUV 5 mm x 5 mm active area TO-8 windowless package RoHS ELECTRO-OPTICAL CHARACTERISTICS

More information

1. Introduction X-ray absorption fine structure (XAFS) is an element-specific powerful technique for chemical analysis. In general, XAFS spectra are o

1. Introduction X-ray absorption fine structure (XAFS) is an element-specific powerful technique for chemical analysis. In general, XAFS spectra are o Installation of the soft X-ray quick XAFS system in the SR Center of Ritsumeikan University Masashi Yoshimura 1, Kohji Nakanishi 1, Kei Mitsuhara 2, Toshiaki Ohta 1 1) The SR Center, Ritsumeikan University,

More information

Spectral Analysis of the LUND/DMI Earthshine Telescope and Filters

Spectral Analysis of the LUND/DMI Earthshine Telescope and Filters Spectral Analysis of the LUND/DMI Earthshine Telescope and Filters 12 August 2011-08-12 Ahmad Darudi & Rodrigo Badínez A1 1. Spectral Analysis of the telescope and Filters This section reports the characterization

More information

PFC/JA X-Ray Response of Silicon Surface Barrier Diodes at. 8 and 17.5 kev: Evidence that the X-Ray Sensitive

PFC/JA X-Ray Response of Silicon Surface Barrier Diodes at. 8 and 17.5 kev: Evidence that the X-Ray Sensitive PFC/JA-88-3 X-Ray Response of Silicon Surface Barrier Diodes at 8 and 17.5 kev: Evidence that the X-Ray Sensitive Depth is not Generally the Depletion Depth Kevin W. Wenzel, Richard D. Petrasso March 1988

More information

UV-VIS-IR Spectral Responsivity Measurement System for Solar Cells

UV-VIS-IR Spectral Responsivity Measurement System for Solar Cells November 1998 NREL/CP-52-25654 UV-VIS-IR Spectral Responsivity Measurement System for Solar Cells H. Field Presented at the National Center for Photovoltaics Program Review Meeting, September 8 11, 1998,

More information

Active Pixel Sensors Fabricated in a Standard 0.18 um CMOS Technology

Active Pixel Sensors Fabricated in a Standard 0.18 um CMOS Technology Active Pixel Sensors Fabricated in a Standard.18 um CMOS Technology Hui Tian, Xinqiao Liu, SukHwan Lim, Stuart Kleinfelder, and Abbas El Gamal Information Systems Laboratory, Stanford University Stanford,

More information

Silicon sensors for radiant signals. D.Sc. Mikko A. Juntunen

Silicon sensors for radiant signals. D.Sc. Mikko A. Juntunen Silicon sensors for radiant signals D.Sc. Mikko A. Juntunen 2017 01 16 Today s outline Introduction Basic physical principles PN junction revisited Applications Light Ionizing radiation X-Ray sensors in

More information

Review Energy Bands Carrier Density & Mobility Carrier Transport Generation and Recombination

Review Energy Bands Carrier Density & Mobility Carrier Transport Generation and Recombination Review Energy Bands Carrier Density & Mobility Carrier Transport Generation and Recombination Current Transport: Diffusion, Thermionic Emission & Tunneling For Diffusion current, the depletion layer is

More information

EDC Lecture Notes UNIT-1

EDC Lecture Notes UNIT-1 P-N Junction Diode EDC Lecture Notes Diode: A pure silicon crystal or germanium crystal is known as an intrinsic semiconductor. There are not enough free electrons and holes in an intrinsic semi-conductor

More information

DETECTORS Important characteristics: 1) Wavelength response 2) Quantum response how light is detected 3) Sensitivity 4) Frequency of response

DETECTORS Important characteristics: 1) Wavelength response 2) Quantum response how light is detected 3) Sensitivity 4) Frequency of response DETECTORS Important characteristics: 1) Wavelength response 2) Quantum response how light is detected 3) Sensitivity 4) Frequency of response (response time) 5) Stability 6) Cost 7) convenience Photoelectric

More information

PERFORMANCE OF PHOTODIGM S DBR SEMICONDUCTOR LASERS FOR PICOSECOND AND NANOSECOND PULSING APPLICATIONS

PERFORMANCE OF PHOTODIGM S DBR SEMICONDUCTOR LASERS FOR PICOSECOND AND NANOSECOND PULSING APPLICATIONS PERFORMANCE OF PHOTODIGM S DBR SEMICONDUCTOR LASERS FOR PICOSECOND AND NANOSECOND PULSING APPLICATIONS By Jason O Daniel, Ph.D. TABLE OF CONTENTS 1. Introduction...1 2. Pulse Measurements for Pulse Widths

More information

OPTI510R: Photonics. Khanh Kieu College of Optical Sciences, University of Arizona Meinel building R.626

OPTI510R: Photonics. Khanh Kieu College of Optical Sciences, University of Arizona Meinel building R.626 OPTI510R: Photonics Khanh Kieu College of Optical Sciences, University of Arizona kkieu@optics.arizona.edu Meinel building R.626 Photodetectors Introduction Most important characteristics Photodetector

More information

Design and Simulation of N-Substrate Reverse Type Ingaasp/Inp Avalanche Photodiode

Design and Simulation of N-Substrate Reverse Type Ingaasp/Inp Avalanche Photodiode International Refereed Journal of Engineering and Science (IRJES) ISSN (Online) 2319-183X, (Print) 2319-1821 Volume 2, Issue 8 (August 2013), PP.34-39 Design and Simulation of N-Substrate Reverse Type

More information

Photops. Photodiode-Amplifier Hybrids

Photops. Photodiode-Amplifier Hybrids Photops Photodiode-Amplifier Hybrids The Photop Series, combines a photodiode with an operational amplifier in the same package. Photops general-purpose detectors have a spectral range from either 350

More information

ECEN 4606, UNDERGRADUATE OPTICS LAB

ECEN 4606, UNDERGRADUATE OPTICS LAB ECEN 4606, UNDERGRADUATE OPTICS LAB Lab 10: Photodetectors Original: Professor McLeod SUMMARY: In this lab, you will characterize the fundamental low-frequency characteristics of photodiodes and the circuits

More information

Gamma-ray spectral imaging using a single-shutter radiation camera

Gamma-ray spectral imaging using a single-shutter radiation camera Nuclear Instruments and Methods in Physics Research A299 (1990) 495-500 North-Holland 495 Gamma-ray spectral imaging using a single-shutter radiation camera T.A. DeVol, D.K. Wehe and G.F. Knoll The University

More information

HOW DIODES WORK CONTENTS. Solder plated Part No. Lot No Cathode mark. Solder plated 0.

HOW DIODES WORK CONTENTS.  Solder plated Part No. Lot No Cathode mark. Solder plated 0. www.joeknowselectronics.com Joe Knows, Inc. 1930 Village Center Circle #3-8830 Las Vegas, NV 89134 How Diodes Work Copyright 2013 Joe Knows Electronics HOW DIODES WORK Solder plated 0.4 1.6 There are several

More information

Lecture 2. Part 2 (Semiconductor detectors =sensors + electronics) Segmented detectors with pn-junction. Strip/pixel detectors

Lecture 2. Part 2 (Semiconductor detectors =sensors + electronics) Segmented detectors with pn-junction. Strip/pixel detectors Lecture 2 Part 1 (Electronics) Signal formation Readout electronics Noise Part 2 (Semiconductor detectors =sensors + electronics) Segmented detectors with pn-junction Strip/pixel detectors Drift detectors

More information

Analysis of 1=f Noise in CMOS Preamplifier With CDS Circuit

Analysis of 1=f Noise in CMOS Preamplifier With CDS Circuit IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 49, NO. 4, AUGUST 2002 1819 Analysis of 1=f Noise in CMOS Preamplifier With CDS Circuit Tae-Hoon Lee, Gyuseong Cho, Hee Joon Kim, Seung Wook Lee, Wanno Lee, and

More information

Gamma Ray Spectroscopy with NaI(Tl) and HPGe Detectors

Gamma Ray Spectroscopy with NaI(Tl) and HPGe Detectors Nuclear Physics #1 Gamma Ray Spectroscopy with NaI(Tl) and HPGe Detectors Introduction: In this experiment you will use both scintillation and semiconductor detectors to study γ- ray energy spectra. The

More information

Advanced Features of InfraTec Pyroelectric Detectors

Advanced Features of InfraTec Pyroelectric Detectors 1 Basics and Application of Variable Color Products The key element of InfraTec s variable color products is a silicon micro machined tunable narrow bandpass filter, which is fully integrated inside the

More information

High sensitive photodiodes

High sensitive photodiodes epc200 High sensitive photodiodes General Description The epc200 is a high-sensitive, high-speed, low-cost photo diode for light-barriers, light-curtains, and similar applications. These photo diodes are

More information

OPERATING CHARACTERISTICS OF THE GEIGER COUNTER

OPERATING CHARACTERISTICS OF THE GEIGER COUNTER OPERATING CHARACTERISTICS OF THE GEIGER COUNTER OBJECTIVE The objective of this laboratory is to determine the operating voltage for a Geiger tube and to calculate the effect of the dead time and recovery

More information

DEVICE APPLICATION OF NON-EQUILIBRIUM MOS CAPACITORS FABRICATED ON HIGH RESISTIVITY SILICON

DEVICE APPLICATION OF NON-EQUILIBRIUM MOS CAPACITORS FABRICATED ON HIGH RESISTIVITY SILICON INTERNATIONAL JOURNAL ON SMART SENSING AND INTELLIGENT SYSTEMS, VOL. 4, NO. 4, DECEMBER 2011 DEVICE APPLICATION OF NON-EQUILIBRIUM MOS CAPACITORS FABRICATED ON HIGH RESISTIVITY SILICON O. Malik, F. J.

More information

CADMIUM Telluride (CdTe) and Cadmium Zinc Telluride

CADMIUM Telluride (CdTe) and Cadmium Zinc Telluride Evaluation of 5 mm-thick CdTe Detectors from the Company Acrorad Alfred Garson III 1, Ira V. Jung 1, Jeremy Perkins 1, and Henric Krawczynski 1 arxiv:astro-ph/511577v1 18 Nov 25 Abstract Using 2 2.5 cm

More information

ECE 340 Lecture 29 : LEDs and Lasers Class Outline:

ECE 340 Lecture 29 : LEDs and Lasers Class Outline: ECE 340 Lecture 29 : LEDs and Lasers Class Outline: Light Emitting Diodes Lasers Semiconductor Lasers Things you should know when you leave Key Questions What is an LED and how does it work? How does a

More information

Development of Solid-State Detector for X-ray Computed Tomography

Development of Solid-State Detector for X-ray Computed Tomography Proceedings of the Korea Nuclear Society Autumn Meeting Seoul, Korea, October 2001 Development of Solid-State Detector for X-ray Computed Tomography S.W Kwak 1), H.K Kim 1), Y. S Kim 1), S.C Jeon 1), G.

More information

CDTE and CdZnTe detector arrays have been recently

CDTE and CdZnTe detector arrays have been recently 20 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 44, NO. 1, FEBRUARY 1997 CMOS Low-Noise Switched Charge Sensitive Preamplifier for CdTe and CdZnTe X-Ray Detectors Claudio G. Jakobson and Yael Nemirovsky

More information

CHAPTER 5 FINE-TUNING OF AN ECDL WITH AN INTRACAVITY LIQUID CRYSTAL ELEMENT

CHAPTER 5 FINE-TUNING OF AN ECDL WITH AN INTRACAVITY LIQUID CRYSTAL ELEMENT CHAPTER 5 FINE-TUNING OF AN ECDL WITH AN INTRACAVITY LIQUID CRYSTAL ELEMENT In this chapter, the experimental results for fine-tuning of the laser wavelength with an intracavity liquid crystal element

More information

Key Questions. What is an LED and how does it work? How does a laser work? How does a semiconductor laser work? ECE 340 Lecture 29 : LEDs and Lasers

Key Questions. What is an LED and how does it work? How does a laser work? How does a semiconductor laser work? ECE 340 Lecture 29 : LEDs and Lasers Things you should know when you leave Key Questions ECE 340 Lecture 29 : LEDs and Class Outline: What is an LED and how does it How does a laser How does a semiconductor laser How do light emitting diodes

More information

High energy X-ray emission driven by high voltage circuit system

High energy X-ray emission driven by high voltage circuit system Journal of Physics: Conference Series OPEN ACCESS High energy X-ray emission driven by high voltage circuit system To cite this article: M Di Paolo Emilio and L Palladino 2014 J. Phys.: Conf. Ser. 508

More information

Lecture 18: Photodetectors

Lecture 18: Photodetectors Lecture 18: Photodetectors Contents 1 Introduction 1 2 Photodetector principle 2 3 Photoconductor 4 4 Photodiodes 6 4.1 Heterojunction photodiode.................... 8 4.2 Metal-semiconductor photodiode................

More information

High Breakdown Voltage, Fully Depleted Series Large Active Area Photodiodes

High Breakdown Voltage, Fully Depleted Series Large Active Area Photodiodes High Breakdown Voltage, Fully Depleted Series Photodiodes The High Speed Detectors can be fully depleted to achieve the lowest possible junction capacitance for fast response times. They may be operated

More information