Ultrascan 1000 CCD Camera with CM12

Size: px
Start display at page:

Download "Ultrascan 1000 CCD Camera with CM12"

Transcription

1 Chen Xu $BrandeisEM: ~/emdoc/en_us.iso8859-1/articles/us1000-ccd-with-cm12/article.sgml,v2.01 Thu Nov 29 14:29:47 EST 2012 xuchen Exp$ A Gatan Ultrascan CCD camera has been installed on CM12 since beginning of December, Similar to the US4000 on F30, this camera is also a 4-port readout optical stack. The MTF property of this camera on 120keV is significantly better than that of US4000 on F30 under 300keV. It is expected that it can replace electron films in most of applications that use CM12, since the image quality is comparable to that of film. Unfortunately, due to the outdated version of CM12 firmware, DigitalMicrograph doesn t communicate with our CM12. Therefore, when using DigitalMicrograph to take image, one has to manually input magnification information from scope, if you want to have accurate scale bar shown in the image. Beside DigitalMicrograph, a recent version of SerialEM has also been installed on this system. Although SerialEM is designed for Tecnai or JEOL system and to control them, it runs just fine on our CM12. It does provide flexibility and more options to drive the camera. It is recommended to use SerialEM to collect the data which is going to be processed, since, beside several other advantages, it also saves image directly into MRC format. In this document, I first give you a couple of notes for using DigitalMicrograph and SerialEM. Then, it lists a table for calibration of magnifications and exact pixel size on CCD. Finally, there is a section regarding electron dose. You can also get pdf version of this document here (article.pdf). Table of Contents 1. Some Properties of this US1000 CCD Things that are needed to pay attention to Note about DigitalMicrograph Note about Shutter control Note about SerialEM Monitor Drift Rate using SerialEM Mag & Pixel Size Calibration table Note about electron dose Access CM12 Data Remotely Remove X-ray and Unstack...8 1

2 1. Some Properties of this US1000 CCD Ultrascan 1000 CCD Camera with CM12 This is a 2k x 2k camera, with U type scintillator. Pixel size is 2048 x ports readout. MTF at 1/2 Nyquist frequency is about 0.16 at 120keV. Electron to counts Conversion Efficiency is 24 Counts/Prim e-. This is a 16bit camera. 2. Things that are needed to pay attention to WarningDo not use focused Diffraction spots on CCD! Otherwise, the central spot is so strong that it will damage the Scintillator material that faces electron directly. One may use diffraction mode, as some of people do for their Search mode setup. Just make sure you lower large screen down first before turning on Diff mode and spreading central beam by adjusting the "Focus" knob in Diff mode. Tip: The CCD should be in retracted position before one inserts the specimen holder into column and before vacuum is ready. So in case of vacuum crash, the back stream oil from ODP pump won t directly drop onto CCD phosphorous surface. Important: On camera controller, the temperature toggle switch should be at OFF position while P3 is not ready! Normally, the switch is always at "COOL" position to reduce the readout noise. When the vacuum of the scope is not ready, specifically, when P3 is not ready, the camera should not be "COOL"ed or "WARM"ed up for baking. Therefore, one should manually switch to "OFF" position when to vent the camera chamber, or to shutdown the scope. Note: Obtain gain reference ONLY when CCD is completely cooled down. The final temperature of our system is about -25C. 3. Note about DigitalMicrograph If you want to have scale bar in your image, follow the two steps here. Similarly, you can turn off the scale bar by undoing those two steps. Turn on image magnification questions. DigitalMicrograph -> Microscope -> Configure Microscope, Click on left side "session -> Microscope", and on the right side select "Always" or "Once" for Magnification. 2

3 Now, turn on scale bar. From the same "Global Info" windows as above, on left side, click on Data Bar -> Page Mode. On right side, make sure on the bottom the box "Add Scale Marker After Acquisition" being checked. 3

4 4. Note about Shutter control Ultrascan 1000 CCD Camera with CM12 Currently, there is only one shutter control cable connecting from US1000 FirstLight controller to CM12 electronics board. Therefore, there is no way to specify the offset between two shutters when taking a CCD shot, normally one is above specimen and one below. So, for plastic sections, it might cause problem of "drifting" due to charging. Fortunately, the single cable is connecting to pre-specimnen shutter. To be more specific, it is using Gun-UX coil to shift beam away as a blanker. It is comforting to know at least cryo sample is protected, IF the camera is configured correctly. The correct way to configure this is from DM -> Camera -> Configuration, to make sure the Shutter 1 is checked for "shutter idle status". The large screen of CM12 is now working as shutter switch - beam is there when screen is Down and beam goes away when large screen is Up. One can easily check to see if the shutter is configured correctly by lifting the large screen handle slightly - the beam should go away. The value for current change related to large screen Up and Down should be noticeable on CM12 -> Parameter -> Display Current. Interestingly, the beam blanker in Low-Dose mode uses the very same coil to shift beam away. It adds a large value on top of the value from camera shutter configuration. 5. Note about SerialEM New version of SerialEM has been installed on CM12. This "NoScope" version doesn t give any error message when starting up. And we can align two images like a normal version, except it reports the pixel instead of nm. This makes it possible for us to accurately monitor the drift rate before taking final shot. Now, you can use serialem normally like on F30, except for controlling microscope part. When SerialEM main window is activated, the keyboard SPACE bar does two nice tricks for CCD: it stops continuous shot, or it repeats the very last action. WarningBefore taking film picture, one should retract CCD. There are two reasons for it. First, if CCD not retracted, the shutter control with DM by large screen will not be automatically overridden by low-dose software on CM12. In other word, you will have blank picture except text area. Second, the small fragments won t fall on phosphor surface. There is no default button to do that, like in Gatan DM interface. However, you can make a very simple Macro and make it clickable in Camera Setup Control. MacroName CCD ## retract CCD before taking film shot RetractCamera Note: This macro to retract CCD can integrate with drifting control macro in next section, so that CCD can be retracted automatically when drifting rate is below certain threshold. 4

5 6. Monitor Drift Rate using SerialEM Ultrascan 1000 CCD Camera with CM12 After checking Focus and Stigmation at S1 or S2 mode, we usually watch the drifting visually on analog TV monitor. When we feel it is stable, then we take the shot. This watching is not quantitative at all. With CCD only system, because of the jiggling due to reading out speed of CCD camera, this watching is even more problematic. Luckily, with SerialEM s macro ability, we can get the drifting rate quantitatively. We do this by running the following macro: MacroName Drift #### Editing ########################### # Set to type of picture to take shot = Focus # unbinned pixel size in nm (related with mag, check Chen s table) pixel = 0.18 # binning of shot in CCD setup bin = 2 # Set to interval between pictures in seconds interval = 15 # Set to maximum number of pictures times = 10 # Set to criterion drift in A/sec crit = 1 # End of parameters ##### End of Editing ##################### # actual pixelsize ps = $pixel * $bin SuppressReports $shot Delay $interval Loop $times index $shot AlignTo B ReportAlignShift ClearAlignment dx = $reportedvalue1 dy = $reportedvalue2 dist = sqrt $dx * $dx + $dy * $dy #echo Distance = $dist pixels dist_in_nm = $dist * $ps #echo Distance = $dist_in_nm nm drift = $dist_in_nm / $interval * 10 echo echo Drift = $drift A/sec if $drift < $crit echo Drift is low enough after shot $index ## for film, uncomment the next line to retract CCD #Call CCD break endif if $index < $times 5

6 Delay $interval else Pause Drift never got below $crit: Continue anyway? endif EndLoop The idea of this macro is very clear: By aligning two images taken at certain interval, the distance is figured out. Dividing the distance by the interval, we get the drifting rate. It is worth mentioning that there are two conditions needed to be satisfied. 1. There is no continuous imaging currently taken. 2. The shot defined in the macro needs to in single mode. E.g., in the above macro, Focus is used for this job, therefore, Focus has to be in single mode, selected in Camera Control panel. Pay attention to the line #Call CCD Uncomment this line will have CCD automatically being retracted when the drifting rate is low enough. 7. Mag & Pixel Size Calibration table The calibration was done using standard 0.463um distance grating grid and home made Tropomyosin paracrystal which has 39.5nm repeat. The later gives more accurate calculation for higher range of magnifications. Mag_ScreenDown(X) Mag_ScreenUp(X) Mag_OnCCD(X)

7 Mag_ScreenDown(X) Mag_ScreenUp(X) Mag_OnCCD(X) Note about electron dose Since we know the conversion rate of electron to counts and pixel size for each magnifications, we could use these two parameter to figure out the relationship between the mean counts on CCD and dose on specimen. It is as following: Mean_Counts = Dose * Conversion_Rate * Pixel_Size * Pixel_Size For example, if you take images at mag of 19500X(ScreenDown), the pixel size at specimen level is then 0.488nm. If you want 20e/A2 as target dose, you can then adjust your beam condition and exposure time to get your CCD image with Mean Counts of Mean_Counts = 20 * 24 * 4.88 * 4.88 = It is important to point out that in default setup for SerialEM the readout count on your screen is only half of the original counts. This is done by checking "divided by 2 for 16 bits camera" under Camera menu. It is believed that this will remove noise more effectively, in the case of dark image background drifts sometimes. Quoted here is from SerialEM help file "Data can be divided by 2 as soon as they are acquired, by selecting Divide 16-bit By 2 in the Camera menu. They will thus never appear to be unsigned integers, making the options for treatment of 16-bit data irrelevant. If the camera noise level is high enough to make this an acceptable procedure, then it has the additional advantage that it will preserve negative values in gain-normalized images. Such values can occur and be meaningful in images with very low counts if the bias of the camera drifts downwards after taking a dark reference.". Therefore, if you measure your count in serialem for dose purpose, you either uncheck "divided by 2 for 16 bits camera" and read the counts directly or otherwise use half of the counts in above equations. To make it clear, if the menu "divided by 2 for 16 bits camera" is being checked in SerialEM, one should use the following equations: Mean_Counts = 0.5 * Dose * Conversion_Rate * Pixel_Size * Pixel_Size Mean_Counts = 0.5 * 20 * 24 * 4.88 * 4.88 =

8 9. Access CM12 Data Remotely Ultrascan 1000 CCD Camera with CM12 All the data should be directly saved in local computer s hard drive. It is recommended to put them into "C:\CM12-Data". For convenience, this folder is now configured as shared and can be accessed from other computers remotely. Thanks to Lynn for setting this up. This folder can now be mounted on various platforms as the detailed below. LINUX To start up the ftp-like smbclient and connect to the share from the command line, enter: % smbclient //cm12.rose2.brandeis.edu/cm12-data -W EMLAB -U emlab The user will be prompted for the password for the emlab account. Enter help <return> for information on using smbclient. WINDOWS To mount the share on windows, Start->Run, then enter the following in the "Open:" field or in the Search box in Windows 7 computer: \\cm12.rose2.brandeis.edu\cm12-data When prompted, the user should authenticate using User name emlab and the password for the emlab account. A window into the share should open on the desktop. MAC OS X Use CMD-K to pull up the Connect to Server dialog box, then enter the following in the Server Address field: smb://cm12.rose2.brandeis.edu/cm12-data The user will be prompted for authentication information and should enter: Workgroup or Domain: EMLAB [if not already filled in] Name: emlab Password: password for emlab account An icon for the share should appear on the desktop. 10. Remove X-ray and Unstack Due to X-ray, a few pixels may have very high number. These pixels should be replaced by reasonable value, usually a local mean. We call this process "remove x-ray". This process is very simple with IMOD package. One only needs to run pre-process step from etomo interface. One might need to unstack the X-ray removed stack into multiple single images to pick particles. IMOD has a standalone command "newstack" that unstacks an image, but itself doesn t automatically unstack all the images(sections) in a stack file at once. Fortunately, this can be done easily using a script like this example as below: 8

9 Example 1. unstack.sh #!/bin/sh for file in $1 ; do root= basename ${file}.st num_file= header ${file} grep sections awk {printf ($9)} for (( i=1; i<=${num_file}; i++ )) do f= expr $i - 1 index= echo $f awk {printf("%03d",$1)} ## for simple number index #newstack -secs ${f} ${file} ${root}.${f}.mrc ## for 3 digit index, like 001,002 newstack -secs ${f} ${file} ${root}_${index}.mrc done done exit Download the above script unstack.sh. Run script like this: %./unstack.sh <yourinputstack.st> 9

Operating F20/F30 with SerialEM

Operating F20/F30 with SerialEM Chen Xu xuchen@brandeis.ede $BrandeisEM: ~emdoc-xml/en_us.iso8859-1/articles/operating-f20-or-f30/article.xml, 1 2013-01-19 01:42:20 xuchen Exp$ This is a quick check list for the Tecnai F20 or Tecnai

More information

XTEM. --Software for Complex Transmission Electron Microscopy. Version 1.0

XTEM. --Software for Complex Transmission Electron Microscopy. Version 1.0 XTEM --Software for Complex Transmission Electron Microscopy Version 1.0 1. Introduction XTEM is the software for complex microscopy on JEOL 3100 electron microscopes. The XTEM software consists of a suite

More information

SerialEM Help Index. ABC Amber HLP Converte r T rial version,

SerialEM Help Index. ABC Amber HLP Converte r T rial version, SerialEM Help Index Genera l Topics Introduction to SerialEM Mouse and Keyboard Controls Control Panels Image Acquisition Acquiring Tilt Series Montaging Low Dose Mode Energy Filtering Using the Navigator

More information

FEI Tecnai G 2 F20 Operating Procedures

FEI Tecnai G 2 F20 Operating Procedures FEI Tecnai G 2 F20 Operating Procedures 1. Startup (1) Sign-up in the microscope log-sheet. Please ensure you have written an account number for billing. (2) Log in to the computer: Login to your account

More information

FEI Falcon Direct Electron Detector. Best Practice Document

FEI Falcon Direct Electron Detector. Best Practice Document FEI Falcon Direct Electron Detector Best Practice Document 2 1. Introduction FEI Falcon Direct Electron Detector Best Practice Application Guide The FEI Falcon Detector is based on direct electron detection

More information

Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope

Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope The procedures given below were written specifically for the FEI Tecnai G 2 Sphera microscope. Modifications will need to

More information

This document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the digital camera.

This document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the digital camera. FEI Tecnai F20 S/TEM: acquiring diffraction patterns Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 (352) 392-3077 Last updated: 10/18/17 This document assumes the user is already familiar with basic

More information

User Operation of JEOL 1200 EX II

User Operation of JEOL 1200 EX II **Log onto Computer** Open item program Start Up Procedure User Operation of JEOL 1200 EX II 1. If scope is not running, locate an electron microscopy technician (EMT) to find out why not. 2. Turn up brightness

More information

Instructions for Tecnai a brief start up manual

Instructions for Tecnai a brief start up manual Instructions for Tecnai a brief start up manual Version 3.0, 8.12.2015 Manual of Tecnai 12 transmission electron microscope located at Aalto University's Nanomicroscopy Center. More information of Nanomicroscopy

More information

Jeol JEM Responsible personell: Endy ( ) Online booking is compulsory!

Jeol JEM Responsible personell: Endy ( ) Online booking is compulsory! Jeol JEM 1230 Responsible personell: Endy (45279377) Online booking is compulsory! After training you will have access to working alone on the instrument. All insertion of samples is done by responsible

More information

1.2. Make sure the viewing screen is covered (exposure to liquid N 2 may cause it to crack).

1.2. Make sure the viewing screen is covered (exposure to liquid N 2 may cause it to crack). FEI Tecnai F20 S/TEM: imaging in TEM mode Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 (352) 392-3077 Last updated: 01/21/18 1. Filling the cold trap (if needed) 1.1. Prior to use, the cold trap needs

More information

LowDose for JEOL. --Software LowDose and Advanced Data Acquisition. Version 1.1

LowDose for JEOL. --Software LowDose and Advanced Data Acquisition. Version 1.1 LowDose for JEOL --Software LowDose and Advanced Data Acquisition Version 1.1 1. Introduction This is the LowDose software for the user of JEOL electron microscopes. The LowDose software is a Plug-In for

More information

2 How to operate the microscope/obtain an image

2 How to operate the microscope/obtain an image Morgagni Operating Instructions 50079 010912 2-1 2 ow to operate the microscope/obtain an image 2.1 Starting the microscope 2.1.1 Starting the microscope with several manually-operated steps 1. Turn on

More information

1.3. Before loading the holder into the TEM, make sure the X tilt is set to zero and the goniometer locked in place (this will make loading easier).

1.3. Before loading the holder into the TEM, make sure the X tilt is set to zero and the goniometer locked in place (this will make loading easier). JEOL 200CX operating procedure Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 1. Specimen loading 1.1. Unlock the TUMI system. 1.2. Load specimen(s) into the holder. If using the double tilt holder, ensure

More information

Overview. About other software. Administrator password. 58. UltraVIEW VoX Getting Started Guide

Overview. About other software. Administrator password. 58. UltraVIEW VoX Getting Started Guide Operation 58. UltraVIEW VoX Getting Started Guide Overview This chapter outlines the basic methods used to operate the UltraVIEW VoX system. About other software Volocity places great demands on the computer

More information

Operating Instructions Pocket Pictor For use with Pocket Pc s

Operating Instructions Pocket Pictor For use with Pocket Pc s Introduction Operating Instructions Pocket Pictor For use with Pocket Pc s The compact size and low power consumption of Pocket PC s make them ideal for use in the field. Pocket Pictor is designed for

More information

Operating Checklist for using the Scanning Electron. Microscope, JEOL JSM 6400.

Operating Checklist for using the Scanning Electron. Microscope, JEOL JSM 6400. Smith College August 2009 Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400. CONTENT, page no. Pre-Check 1 Startup 1 Specimen Insertion 2 Filament Saturation 2 Beam Alignment

More information

Scanning Electron Microscope FEI INSPECT F50. Step by step operation manual

Scanning Electron Microscope FEI INSPECT F50. Step by step operation manual Scanning Electron Microscope FEI INSPECT F50 Step by step operation manual Scanning Electron Microscope, FEI Inspect F50 FE-SEM-F Observation Flow Saving Data And Analysis Specimen preparation Error check

More information

OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop

OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop 1. Log in usage using the SMIF web site 2. Turn power

More information

1. Preliminary sample preparation

1. Preliminary sample preparation FEI Helios NanoLab 600 standard operating procedure Nicholas G. Rudawski ngr@ufl.edu (352) 392 3077 (office) (805) 252-4916 (cell) Last updated: 03/02/18 What this document provides: an overview of basic

More information

Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah

Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah Follow the procedures below when you use the Hitachi 7100 TEM. Starting Session 1. Turn on the cold

More information

ORIUS SC 1000 CCD Camera User s Guide

ORIUS SC 1000 CCD Camera User s Guide ORIUS SC 1000 CCD Camera User s Guide Gatan, Inc. 5933 Coronado Lane Pleasanton, CA 94588 Tel (925) 463-0200 FAX (925) 463-0204 March 2006 832.MANUAL Revision 1.03 Preface About this Guide This ORIUS User

More information

SEM Training Notebook

SEM Training Notebook SEM Training Notebook Lab Manager: Dr. Perry Cheung MSE Fee-For-Service Facility Materials Science and Engineering University of California, Riverside March 8, 2018 (rev. 3.5) 1 Before you begin Complete

More information

LEO 912 TEM Short Manual. Prepared/copyrighted by RH Berg Danforth Plant Science Center

LEO 912 TEM Short Manual. Prepared/copyrighted by RH Berg Danforth Plant Science Center LEO 912 TEM Short Manual Prepared/copyrighted by RH Berg Danforth Plant Science Center Specimen holder [1] Never touch the holder (outside of the O-ring, double-headed arrow) because finger oils will contaminate

More information

Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM

Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM Warning Always adjust your specimen height before closing the chamber door to make sure your specimen will not hit the bottom of the lens;

More information

Figure 1 The Raith 150 TWO

Figure 1 The Raith 150 TWO RAITH 150 TWO SOP Figure 1 The Raith 150 TWO LOCATION: Raith 150 TWO room, Lithography area, NanoFab PRIMARY TRAINER: SECONDARY TRAINER: 1. OVERVIEW The Raith 150 TWO is an ultra high resolution, low voltage

More information

SEM Training Notebook

SEM Training Notebook SEM Training Notebook Lab Manager: Dr. Perry Cheung MSE Fee-For-Service Facility Materials Science and Engineering University of California, Riverside December 21, 2017 (rev. 3.4) 1 Before you begin Complete

More information

Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400.

Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400. Smith College August 2005 Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400. CONTENT, page no. Pre-Check, 1 Specimen Insertion, 1 Startup, 2 Filament Saturation, 2 Beam Alignment,

More information

Fire CR Calibration Guide

Fire CR Calibration Guide 1 Fire CR Calibration Guide This reference guide will guide you through the steps to complete the calibration for the Fire CR.. Getting Started: 1. Click on the Opal Icon on the Desktop. Figure 1 2. Once

More information

Instruction Manual for HyperScan Spectrometer

Instruction Manual for HyperScan Spectrometer August 2006 Version 1.1 Table of Contents Section Page 1 Hardware... 1 2 Mounting Procedure... 2 3 CCD Alignment... 6 4 Software... 7 5 Wiring Diagram... 19 1 HARDWARE While it is not necessary to have

More information

JEOL 2010 FasTEM & DigitalMicrograph User's Guide

JEOL 2010 FasTEM & DigitalMicrograph User's Guide JEOL 2010 FasTEM & DigitalMicrograph User's Guide Electron Microscopy Laboratory Instititute of Materials Science University of Connecticut The purpose of this manual is to remind you of the essential

More information

Simplified Instructions: Olympus Widefield Microscope S1230

Simplified Instructions: Olympus Widefield Microscope S1230 Contents General Microscope Operation Simple Image Capture Multi-Wavelength Capture Z-Series Timelapse Combining Capture Modes Synopsis of Other Functions Pages 2-23 24-40 41-47 48-56 57-59 60-68 69-83

More information

This procedure assumes the user is already familiar with basic operation of the SEM and the MiraTC interface.

This procedure assumes the user is already familiar with basic operation of the SEM and the MiraTC interface. Tescan MIRA3 SEM: EDS using EDAX TEAM Nicholas G. Rudawski ngr@ufl.edu Cell: (805) 252-4916 Office: (352) 392-3077 Last updated: 12/04/17 This procedure assumes the user is already familiar with basic

More information

STEM Spectrum Imaging Tutorial

STEM Spectrum Imaging Tutorial STEM Spectrum Imaging Tutorial Gatan, Inc. 5933 Coronado Lane, Pleasanton, CA 94588 Tel: (925) 463-0200 Fax: (925) 463-0204 April 2001 Contents 1 Introduction 1.1 What is Spectrum Imaging? 2 Hardware 3

More information

1 ImageBrowser Software User Guide 5.1

1 ImageBrowser Software User Guide 5.1 1 ImageBrowser Software User Guide 5.1 Table of Contents (1/2) Chapter 1 What is ImageBrowser? Chapter 2 What Can ImageBrowser Do?... 5 Guide to the ImageBrowser Windows... 6 Downloading and Printing Images

More information

Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below.

Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below. Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below. http://staff.washington.edu/wpchan/if/cm100_inst.shtml Instructions for the Philips CM100 TEM and peripherals

More information

K2 Summit TM and K2 Base TM Direct Detection Camera User s Guide

K2 Summit TM and K2 Base TM Direct Detection Camera User s Guide K2 Summit TM and K2 Base TM Direct Detection Camera User s Guide Model Number 1000, 1000.S, 1000.B and 1000.U Document Part Number: 1000.40004 Revision 01 Gatan, Inc. 5794 W. Las Positas Blvd Pleasanton,

More information

LSM 510 Training Notes

LSM 510 Training Notes LSM 510 Training Notes Turning on the system Turn on the arc lamp, found on the bench top left of the microscope. This supplies light for epifluorescence for viewing your samples through the microscope.

More information

Last updated 6/12/18. F20 User Manual at the Simons Electron Microscopy Center

Last updated 6/12/18. F20 User Manual at the Simons Electron Microscopy Center F20 User Manual at the Simons Electron Microscopy Center 1 Table of Contents F20 Information Sheet 2 F20 User Guide (starting your session) 3 F20 User Guide (ending your session) 5 Cryo Screening with

More information

Digital Microscope. User Manual

Digital Microscope. User Manual Digital Microscope User Manual Features The digital microscope provides 10~200X adjustable magnification range. The build-in high-performance white LED can illuminate the object without using any auxiliary

More information

RENISHAW INVIA RAMAN SPECTROMETER

RENISHAW INVIA RAMAN SPECTROMETER STANDARD OPERATING PROCEDURE: RENISHAW INVIA RAMAN SPECTROMETER Purpose of this Instrument: The Renishaw invia Raman Spectrometer is an instrument used to analyze the Raman scattered light from samples

More information

COMPACT MANUAL FOR GI USERS OF THE JEM 1400 FLASH BEGINNERS (For internal use only) Gray means additional information at the end of this mini-manual

COMPACT MANUAL FOR GI USERS OF THE JEM 1400 FLASH BEGINNERS (For internal use only) Gray means additional information at the end of this mini-manual 1 COMPACT MANUAL FOR GI USERS OF THE JEM 1400 FLASH BEGINNERS (For internal use only) Gray means additional information at the end of this mini-manual ABOUT THIS MICROSCOPE (room HG01.240) The JEM-1400Flash

More information

Tecnai on-line help User interface 1 Tecnai F20 Tecnai F30 User interface Software version 2.1.8/3.0

Tecnai on-line help User interface 1 Tecnai F20 Tecnai F30 User interface Software version 2.1.8/3.0 Tecnai on-line help User interface 1 Tecnai on-line help manual -- User interface Table of Contents 1 User Interface...5 1.1 View modes...6 1.2 Toolbar...6 1.3 Workset tabs...7 1.4 Control panels...7 1.5

More information

MSE 460 TEM Lab 2: Basic Alignment and Operation of Microscope

MSE 460 TEM Lab 2: Basic Alignment and Operation of Microscope MSE 460 TEM Lab 2: Basic Alignment and Operation of Microscope Last updated on 1/8/2018 Jinsong Wu, jinsong-wu@northwestern.edu Aims: The aim of this lab is to familiarize you with basic TEM alignment

More information

LSM 510 Meta Training Notes

LSM 510 Meta Training Notes LSM 510 Meta Training Notes Turning on the system Turn on X-Cite power supply. This supplies light for epifluorescence for viewing your samples through the microscope. Turn on the remote control switch.

More information

Goodman Cookbook. Goodman Spectrograph. Adapted by D. Sanmartim from L. Fraga's Guide. Sep SOAR Telescope

Goodman Cookbook. Goodman Spectrograph. Adapted by D. Sanmartim from L. Fraga's Guide. Sep SOAR Telescope Goodman Spectrograph 1 Goodman Spectrograph Documentation Goodman HTS Manual http://www.ctio.noao.edu/soar/content/goodman-hts-manual Goodman Overview http://www.ctio.noao.edu/soar/content/goodman-spectrograph-overview

More information

Widefield-NikonEclipseTE200-ORCA Nikon Eclipse TE200 Inverted Microscope with Hamamatsu 1394 Orca-ER Cooled CCD Camera and Micromanager Software

Widefield-NikonEclipseTE200-ORCA Nikon Eclipse TE200 Inverted Microscope with Hamamatsu 1394 Orca-ER Cooled CCD Camera and Micromanager Software Widefield-NikonEclipseTE200-ORCA Nikon Eclipse TE200 Inverted Microscope with Hamamatsu 1394 Orca-ER Cooled CCD Camera and Micromanager Software September 2007 Check website for most current User Guide

More information

Basic Users Manual for Tecnai-F20 TEM

Basic Users Manual for Tecnai-F20 TEM Basic Users Manual for Tecnai-F20 TEM NB: This document contains my personal notes on the operating procedure of the Tecnai F20 and may be used as a rough guide for those new to the microscope. It may

More information

1.1. In regular TEM imaging mode, find a region of interest and set it at eucentric height.

1.1. In regular TEM imaging mode, find a region of interest and set it at eucentric height. JEOL 2010F operating procedure Covers operation in STEM mode (See separate procedures for operation in TEM mode and operation of EDS system) Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 NOTE: this operating

More information

FEI Quanta 200 ESEM Basic instructions

FEI Quanta 200 ESEM Basic instructions FEI Quanta 200 ESEM Basic instructions Desktop and then start the UI. If the computer has restarted and you need to login, Username: supervisor and Password: supervisor Log-in to the Microscope using the

More information

Progeny Imaging. User Guide V x and Higher. Part Number: ECN: P1808 REV. F

Progeny Imaging. User Guide V x and Higher. Part Number: ECN: P1808 REV. F Progeny Imaging User Guide V. 1.6.0.x and Higher Part Number: 00-02-1598 ECN: P1808 REV. F Contents 1 About This Manual... 5 How to Use this Guide... 5 Text Conventions... 5 Getting Assistance... 6 2 Overview...

More information

A Stony Brook Student s Guide to Using CCDSoft By Stephanie Zajac Last Updated: 3 February 2012

A Stony Brook Student s Guide to Using CCDSoft By Stephanie Zajac Last Updated: 3 February 2012 A Stony Brook Student s Guide to Using CCDSoft By Stephanie Zajac Last Updated: 3 February 2012 This document is meant to serve as a quick start guide to using CCDSoft to take data using the Mt. Stony

More information

How to use the Jeol 1010 TEM of GI (Liesbeth own GI version)

How to use the Jeol 1010 TEM of GI (Liesbeth own GI version) How to use the Jeol 1010 TEM of GI (Liesbeth own GI version) 1.Load the specimen Load a grid into the rod holder: USE ONLY THE TOP POSITION (blue arrow), Specimen selection on 1 (The rear one is only a

More information

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling Zeiss LSM 510 Confocor III Training Notes Center for Cell Analysis & Modeling Confocor 3 Start Up Go to System Module Turn on Main Switch, System/ PC, and Components Switches Do you need the arc lamp?

More information

1.1. Log on to the TUMI system (you cannot proceed further until this is done).

1.1. Log on to the TUMI system (you cannot proceed further until this is done). FEI DB235 SEM mode operation Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 1. Sample loading 1.1. Log on to the TUMI system (you cannot proceed further until this is done). 1.2. The FIB software (xp)

More information

Characterization Microscope Nikon LV150

Characterization Microscope Nikon LV150 Characterization Microscope Nikon LV150 Figure 1: Microscope Nikon LV150 Introduction This upright optical microscope is designed for investigating up to 150 mm (6 inch) semiconductor wafers but can also

More information

RAITH e-line OPERATING INSTRUCTIONS

RAITH e-line OPERATING INSTRUCTIONS RAITH e-line OPERATING INSTRUCTIONS 1) LOADING A SAMPLE a. Start the system i. On the Column PC (Right side monitor [R]), select the SmartSEM icon to on the desktop to begin the column software. ii. On

More information

Faculty Lecture Capture Guide

Faculty Lecture Capture Guide Faculty Lecture Capture Guide If you have never used Panopto before, follow this first part. Log into your Blackboard Account and open the course you wish to capture: Open your Course Management Control

More information

CCD User s Guide SBIG ST7E CCD camera and Macintosh ibook control computer with Meade flip mirror assembly mounted on LX200

CCD User s Guide SBIG ST7E CCD camera and Macintosh ibook control computer with Meade flip mirror assembly mounted on LX200 Massachusetts Institute of Technology Department of Earth, Atmospheric, and Planetary Sciences Handout 8 /week of 2002 March 18 12.409 Hands-On Astronomy, Spring 2002 CCD User s Guide SBIG ST7E CCD camera

More information

START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7

START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7 Leica DMI AF6000LX Table of contents START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7 ACQUIRE MODULE 6 SETTING THE LIGHTPATH 6

More information

Brief description of GIRAFFE

Brief description of GIRAFFE Brief description of The SAAO Grating Instrument for Radiation Analysis with a Fibre Fed Échelle - - consists of two components: (i) The head which is mounted at the Cassegrain focus to collect light from

More information

Micro-Image Capture 8 Installation Instructions & User Guide

Micro-Image Capture 8 Installation Instructions & User Guide Micro-Image Capture 8 Installation Instructions & User Guide Software installation: Micro-Image Capture Software 1. Load Micro-Image Capture software CD onto host PC. Auto Run should start driver/software

More information

IBIL setup operation manual for SynerJY software version

IBIL setup operation manual for SynerJY software version IBIL setup operation manual for SynerJY software version 1.8.5.0 Manual version 1.0, 31/10/2008 Author: Carlos Marques Equipment Managers: Carlos Marques, +351219946084, cmarques@itn.pt Luís Alves, +351219946112,

More information

Camera Test Protocol. Introduction TABLE OF CONTENTS. Camera Test Protocol Technical Note Technical Note

Camera Test Protocol. Introduction TABLE OF CONTENTS. Camera Test Protocol Technical Note Technical Note Technical Note CMOS, EMCCD AND CCD CAMERAS FOR LIFE SCIENCES Camera Test Protocol Introduction The detector is one of the most important components of any microscope system. Accurate detector readings

More information

FLIR Tools for PC 7/21/2016

FLIR Tools for PC 7/21/2016 FLIR Tools for PC 7/21/2016 1 2 Tools+ is an upgrade that adds the ability to create Microsoft Word templates and reports, create radiometric panorama images, and record sequences from compatible USB and

More information

05/20/14 1. Philips CM200T. Standby Condition

05/20/14 1. Philips CM200T. Standby Condition 05/20/14 1 Philips CM200T Standby Condition HT and filament off, HT setting at 200kV. RESET HOLDER, center sample tilt knobs, and remove sample. Mag ~ 5-10kX Objective and SA apertures out, C2 aperture

More information

Before you start, make sure that you have a properly calibrated system to obtain high-quality images.

Before you start, make sure that you have a properly calibrated system to obtain high-quality images. CONTENT Step 1: Optimizing your Workspace for Acquisition... 1 Step 2: Tracing the Region of Interest... 2 Step 3: Camera (& Multichannel) Settings... 3 Step 4: Acquiring a Background Image (Brightfield)...

More information

The user should already be familiar with operation of the instrument in STEM mode, use of the Microscope Control interface, and TIA.

The user should already be familiar with operation of the instrument in STEM mode, use of the Microscope Control interface, and TIA. FEI Tecnai F20 S/TEM: EDS system operation Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 (352) 392-3077 Last updated: 01/22/18 The user should already be familiar with operation of the instrument in

More information

Introduction: Why electrons?

Introduction: Why electrons? Introduction: Why electrons? 1 Radiations Visible light X-rays Electrons Neutrons Advantages Not very damaging Easily focused Eye wonderful detector Small wavelength (Angstroms) Good penetration Small

More information

JEOL 6700 User Manual 05/18/2009

JEOL 6700 User Manual 05/18/2009 JEOL 6700 User Manual 05/18/2009 LOG IN to your session on the computer to the right of the microscope. Starting Conditions 1. Click the button and read the Penning Gauge to ensure that the microscope

More information

DiaInspect-P. Operating manual. Version Page 1

DiaInspect-P. Operating manual. Version Page 1 DiaInspect-P Operating manual Version 8.1.9 Page 1 1. Application of the device... 3 2. Operating concept... 3 2.1. Hardware... 4 2.2. Software... 5 2.3. Data archiving... 6 3. Functions in operators mode...

More information

Dickinson College Department of Geology

Dickinson College Department of Geology Dickinson College Department of Geology Title: Equipment: BASIC OPERATION OF THE SCANNING ELECTRON MICROSCOPE (SEM) JEOL JSM-5900 SCANNING ELECTRON MICROSCOPE Revision: 2.2 Effective Date: 1/29/2003 Author(s):

More information

Progeny Imaging Veterinary

Progeny Imaging Veterinary Progeny Imaging Veterinary User Guide V1.14 and higher 00-02-1605 Rev. K1 ECN: ECO052875 Revision Date: 5/17/2017 Contents 1. About This Manual... 6 How to Use this Guide... 6 Text Conventions... 6 Getting

More information

Training Guide for Leica SP8 Confocal/Multiphoton Microscope

Training Guide for Leica SP8 Confocal/Multiphoton Microscope Training Guide for Leica SP8 Confocal/Multiphoton Microscope LAS AF v3.3 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn ON power switch for epifluorescence

More information

LSM 710 Confocal Microscope Standard Operation Protocol

LSM 710 Confocal Microscope Standard Operation Protocol LSM 710 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Switch on Main power switch 2. Switch on System / PC power button 3. Switch on Components power button 4.

More information

JEOL JEM-1400 Transmission Electron Microscope Operating Instructions

JEOL JEM-1400 Transmission Electron Microscope Operating Instructions JEOL JEM-1400 Transmission Electron Microscope Operating Instructions Anti-contamination device Objective aperture Objective aperture translation knobs Specimen holder Pump/air switch Left hand control

More information

Check that the pneumatic hose is disconnected!!!! (unless your using the BSE detector, of course)

Check that the pneumatic hose is disconnected!!!! (unless your using the BSE detector, of course) JEOL 7000F BASIC OPERATING INSTRUCTIONS-Ver.-2.0 Note: This is minimal operation checklist and does not replace the other reference manuals. Read the manual for Specimen Exchange (JEOL 7000 Specimen Exchange

More information

Olympus xcellence Software - basic user guide

Olympus xcellence Software - basic user guide Olympus xcellence Software - basic user guide This is a basic overview of setting up time lapse experiments using Olympus's xcellence software on BIU's IX81 inverted phase contrast system - the software

More information

CM20 USER GUIDE. Duncan Alexander, CIME 2010

CM20 USER GUIDE. Duncan Alexander, CIME 2010 CM20 USER GUIDE Duncan Alexander, CIME 2010 CM20 START UP AND CHECK LIST 2 SPECIMEN EXCHANGE 5 - REMOVING SAMPLE HOLDER 6 - INSERTING SAMPLE HOLDER 7 TURNING ON HT 8 STARTING THE FILAMENT 9 GUN TILT ALIGNMENT

More information

FEI Titan Image Corrected STEM

FEI Titan Image Corrected STEM 05/03/16 1 FEI Titan 60-300 Image Corrected STEM Standby Condition HT setting at 300kV, Col. Valves Closed RESET Holder and remove sample. Mag ~ 5-10kX Objective and SA apertures out, C2 aperture at 150µm

More information

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope AIM 4.2 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Verify that main power switches on the

More information

1. Specimen Holder Removal, Loading, and Insertion

1. Specimen Holder Removal, Loading, and Insertion OPERATION OF THE PHILIPS CM-200 FEG-TEM When not in use, the CM-200 should be in the MICROSCOPE ON configuration with the HIGH TENSION ON (illuminates green when the high tension is on).. The microscope

More information

ALTURA EDS. Rev. 0915

ALTURA EDS. Rev. 0915 ALTURA EDS Rev. 0915 Enable the Oxford PC Enable the Altura-EDS under Dual Beam Tools in Coral. Or enter your NETID and password directly into the Oxford PC. Warning: Ion-milling, GISs/microprobe, and

More information

Titan on-line help manual -- User Interface

Titan on-line help manual -- User Interface 1 Titan on-line help manual -- User Interface Table of Contents 1 User Interface... 6 1.1 View modes... 7 1.2 Workset tabs... 7 1.3 Control panels... 7 1.4 Popup panels... 9 1.5 Display... 12 1.5.1 Binding

More information

Full-screen mode Popup controls. Overview of the microscope user interface, TEM User Interface and TIA on the left and EDS on the right

Full-screen mode Popup controls. Overview of the microscope user interface, TEM User Interface and TIA on the left and EDS on the right Quick Guide to Operating FEI Titan Themis G2 200 (S)TEM: TEM mode Susheng Tan Nanoscale Fabrication and Characterization Facility, University of Pittsburgh Office: M104/B01 Benedum Hall, 412-383-5978,

More information

JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES

JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES RULES All users must go through a series of standard operation procedure training. For more information contact: Longlong Liao Teaching

More information

Things to check before start-up.

Things to check before start-up. Byeong Cha Page 1 11/24/2009 Manual for Leica SP2 Confocal Microscope Enter you name, the date, the time, and the account number in the user log book. Things to check before start-up. Make sure that your

More information

Indiana University JEM-3200FS

Indiana University JEM-3200FS Indiana University JEM-3200FS Installation Specification Model: JEM 3200FS Serial Number: EM 15000013 Objective Lens Configuration: High Resolution Pole Piece (HRP) JEOL Engineer: Michael P. Van Etten

More information

JEOL 6500 User Manual

JEOL 6500 User Manual LOG IN to your session on the computer to the left of the microscope. Starting Conditions 1. Press Ctrl-Alt-Del and log on to the microscope computer. Click on JEOL PC SEM 6500 icon. Click yes if message

More information

MSE 595T Transmission Electron Microscopy. Laboratory III TEM Imaging - I

MSE 595T Transmission Electron Microscopy. Laboratory III TEM Imaging - I MSE 595T Basic Transmission Electron Microscopy TEM Imaging - I Purpose The purpose of this lab is to: 1. Make fine adjustments to the microscope alignment 2. Obtain a diffraction pattern 3. Obtain an

More information

Renishaw InVia Raman microscope

Renishaw InVia Raman microscope Laser Spectroscopy Labs Renishaw InVia Raman microscope Operation instructions 1. Turn On the power switch, system power switch is located towards the back of the system on the right hand side. Wait ~10

More information

JEM-2100F. Field Emission. Transmission Electron Microscope. (JEOL Ltd. / Japan)

JEM-2100F. Field Emission. Transmission Electron Microscope. (JEOL Ltd. / Japan) JEM-2100F Field Emission Transmission Electron Microscope (JEOL Ltd. / Japan) Contents Introduction Features of JEM-2100F JADAS software SerialEM software Digital Image (CCD) and Cryo Transfer holders

More information

LEICA TCS SP5 AOBS TANDEM USER MANUAL

LEICA TCS SP5 AOBS TANDEM USER MANUAL LEICA TCS SP5 AOBS TANDEM USER MANUAL STARTING THE SYSTEM...2 THE LAS AF SOFTWARE...3 THE «ACQUIRE» MENU...5 CHOOSE AND CREATE A SETTING...6 THE CONTROL PANEL...8 THE DMI6000B MICROSCOPE...10 ACQUIRE ONE

More information

General information. If you see the instrument turned off, notify MIC personnel. MIC personnel will help you insert your samples into the instrument.

General information. If you see the instrument turned off, notify MIC personnel. MIC personnel will help you insert your samples into the instrument. JEOL JSM-7400F Table of contents General information.. 3 The operation panel. 4 The different sample holders and inserting the samples.. 5 Turning on the beam... 6 Stage map control... 8 Correcting astigmatism...

More information

SARG: The Graphical User Interface Manual

SARG: The Graphical User Interface Manual 1/28 SARG: The Graphical User Interface Manual Document: TNG-SARG-001 Issue: 1.0 Prepared by : Name: S. Scuderi Institute: INAF-Osservatorio Astrofisico di Catania Date : Approved by : Name: R. Cosentino

More information

STEM alignment procedures

STEM alignment procedures STEM alignment procedures Step 1. ASID alignment mode 1. Write down STD for TEM, and then open the ASID control window from dialogue. Also, start Simple imager viewer program on the Desktop. 2. Click on

More information

Kigamo Scanback which fits in your view camera in place of conventional film.

Kigamo Scanback which fits in your view camera in place of conventional film. What's included Kigamo Scanback which fits in your view camera in place of conventional film. SCSI Cable to connect your Scanback to the host computer. A 3-meter SCSI cable is standard. Kigamo also has

More information

4.5.1 Mirroring Gain/Offset Registers GPIO CMV Snapshot Control... 14

4.5.1 Mirroring Gain/Offset Registers GPIO CMV Snapshot Control... 14 Thank you for choosing the MityCAM-C8000 from Critical Link. The MityCAM-C8000 MityViewer Quick Start Guide will guide you through the software installation process and the steps to acquire your first

More information

Model SU3500 Scanning Electron Microscope

Model SU3500 Scanning Electron Microscope Model SU3500 Scanning Electron Microscope Modified and Parts taken from Hitachi Easy Operation Guide. Before using the Model SU3500 SEM, be sure to read the [GENERAL SAFETY GUIDELINES] in the instruction

More information

Nanoscale Fabrication & Characterization Facility. Raith e-line EBL Users Guide (updated:aug 2 nd, 2017)

Nanoscale Fabrication & Characterization Facility. Raith e-line EBL Users Guide (updated:aug 2 nd, 2017) Nanoscale Fabrication & Characterization Facility Raith e-line EBL Users Guide (updated:aug 2 nd, 2017) Overview: The Raith e-line EBL system is designed to write features with critical dimensions as small

More information