SPECTRONIC Standards User Guide

Size: px
Start display at page:

Download "SPECTRONIC Standards User Guide"

Transcription

1 SPECTRONIC Standards User Guide

2 The information in this publication is provided for reference only. All information contained in this publication is believed to be correct and complete. Thermo Fisher Scientific shall not be liable for errors contained herein nor for incidental or consequential damages in connection with the furnishing, performance or use of this material. All product specifications, as well as the information contained in this publication, are subject to change without notice. This publication may contain or reference information and products protected by copyrights or patents and does not convey any license under our patent rights, nor the rights of others. We do not assume any liability arising out of any infringements of patents or other rights of third parties. We make no warranty of any kind with regard to this material, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Customers are ultimately responsible for validation of their systems Thermo Fisher Scientific Inc. All rights reserved. No part of this publication may be stored in a retrieval system, transmitted, or reproduced in any way, including but not limited to photocopy, photograph, magnetic or other record, without our prior written permission. For Technical Support, please contact: Thermo Fisher Scientific 5225 Verona Road Madison, WI U.S.A. Telephone: us.techsupport.analyze@thermofisher.com World Wide Web: For International Support, please contact: Thermo Fisher Scientific Telephone: support.madison@thermofisher.com World Wide Web: Microsoft, Windows and Windows NT are either trademarks or registered trademarks of Microsoft Corporation in the United States and/or other countries. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries , Rev. C

3 Contents Introduction...1 Conventions used in this manual... 2 Questions or concerns... 2 Description % Transmittance Standard... 3 Wavelength Standard... 4 Stray Light Standards... 4 Photometric Accuracy Standards... 5 Empty Filter Holder... 5 Recertification and repair... 5 Handling and Care of Standards... 6 Instrument Setup... 8 Operating technique... 8 Adapter and accessory filters... 8 Recording and Interpreting Test Results Recording test results Interpreting test results Use of Absorbance mode Test Procedures % transmittance and stray light test Wavelength accuracy test Wavelength peak selection SPECTRONIC 20 + and SPECTRONIC 20D Photometric accuracy/linearity test SPECTRONIC Standards User Guide i

4

5 Introduction SPECTRONIC Standards are an affordable, accurate and stable set of filters for performance validation purposes. SPECTRONIC Standards enable the spectrophotometer user to quickly and reliably evaluate the major performance parameters of Thermo Scientific's spectrophotometers. As a part of the laboratory's normal quality control program, SPECTRONIC Standards are an invaluable aid in the detection of incipient instrument problems before errors occur. SPECTRONIC Standards User Guide 1

6 Conventions used in this manual This manual includes safety precautions and other important information presented in the following format: Note Notice Caution Warning Danger Notes contain helpful supplementary information. Follow instructions labeled Notice to avoid damaging the system hardware or losing data. Indicates a hazardous situation which, if not avoided, could result in minor or moderate injury. Indicates a hazardous situation which, if not avoided, could result in death or serious injury. Indicates a hazardous situation which, if not avoided, will result in death or serious injury. Questions or concerns In case of emergency, follow the procedures established by your facility. If you have questions or concerns about safety or need assistance with operation, repairs or replacement parts, you can contact our sales or service representative in your area or use the information at the beginning of this document to contact us. 2 SPECTRONIC Standards User Guide

7 Description SPECTRONIC Standards consists of the following items: 220 nm stray light filter standard 340 nm stray light filter standard 400 nm stray light filter standard Wavelength accuracy standard % Transmittance standard Two photometric accuracy standards with stated transmittance values near 10%T at 590 nm (working and reserve) Two photometric performance standards with stated transmittance values near 50%T at 590 nm (working and reserve) Empty filter holder control standard for photometric accuracy Carrying and storage case Certificate of calibration All standards have the same external dimensions as a standard 10 mm-pathlength square cuvette, so they will fit in standard spectrophotometer cell holders. An adapter is required to use the standards in a test-tube-holder sample compartments. Your set of SPECTRONIC Standards has been individually certified and the values are traceable to NIST. 0% Transmittance Standard The 0% transmittance (0%T) standard is completely opaque. It is used to check the 0%T reading of the SPECTRONIC spectrophotometers by blocking the light beam from the instrument's source lamp. This standard helps determine if a light leak is present and determine the correct zero setting. The 0% transmittance standard's apparent transmittance, if any, can be used to estimate the stray light in instrument. This standard is not traceable to NIST. The 0% transmittance standard cannot be used with spectrophotometers that do not provide a 0% transmittance adjustment. Using this standard in absorbance mode will produce readings that are out of range. Do not use this standard in absorption mode. SPECTRONIC Standards User Guide 3

8 Wavelength Standard The wavelength standard produces three widely spaced, symmetrical, high transmittance peaks in the regions of 400 nm, 525 nm and 780 nm. The primary peak is the certified wavelength for your wavelength standard. This value is recorded on the certificate accompanying the SPECTRONIC Standard. Only the primary peak near 525 nm is certified and NIST traceable. The peaks near 400 and 780 nm can be used for checking the repeatability of your instrument. Stray Light Standards Three stray light standards are provided with the SPECTRONIC Standards. Each standard is essentially opaque at its test wavelength and highly transmitting at longer wavelengths, as shown below. Stray light standards are not traceable to NIST. 4 SPECTRONIC Standards User Guide

9 Photometric Accuracy Standards The photometric performance standards test photometric accuracy and linearity. Transmittance of these standards is constant in the region of 590 nm, as shown below. Consequently, no bandwidth compensation is required. Empty Filter Holder The empty filter holder is identical to the holders for the other standards. The empty filter holder should be used when setting 100%T to ensure that the filter holders of the other standards do not affect the transmittance values. Recertification and repair We offer a recertification and repair service for the SPECTRONIC Standards. It is recommended that you return your standards for recertification every two years or whenever you have reason to believe that a standard no longer exhibits its certified value. Before returning your standards, please contact technical support. SPECTRONIC Standards User Guide 5

10 Handling and Care of Standards The transmittance of the photometric accuracy standards is significantly affected by the condition of the optical surfaces. Any dirt (lint, fingerprints, oil, dust, smoke film, etc.) can cause erroneous test results. The cleaning procedure for the photometric accuracy standards involves a risk of damaging the surfaces. So, prevention of soiling is far better than cleaning. Always keep the standards in their case when not in use. Always keep the cover closed. Never expose the standards to airborne contaminants like smoke, dust, oil vapor or chemical fumes. Before using the photometric accuracy standards, visually inspect them. Remove any dust or lint by gentle brushing. If it becomes necessary to clean a photometric accuracy standard with a liquid, clean only one standard at a time, and proceed as follows: 1. Read and record the standard value in percent transmittance (%T) at 590 nm before cleaning. 2. Use a nut driver to loosen the nut securing the glass filter and back the nut out almost all the way, but do not remove it. 3. Remove the filter, holding it by the edges and being careful not to scrape the glass against the filter holder especially the metal lip that projects near the bottom of the holder. Note You must use the appropriate liquid when cleaning the glass: Photometric accuracy standards - use only reagent quality isopropyl alcohol Wavelength or stray light standards - use only distilled or deionized water. 4. Clean the glass using only the recommended liquid and a soft, clean, lint-free cloth. 5. Reassemble the filter and inspect it for cleanliness and scratches. SPECTRONIC Standards User Guide 6

11 6. Read the standard value at 590 nm. The percent transmittance (%T) reading should remain the same or increase. 7. If the value has decreased, repeat the cleaning process. SPECTRONIC Standards User Guide 7

12 Instrument Setup Set up the spectrophotometer as indicated in the operator's manual to make transmittance measurements, being sure to allow the recommended warm-up time. Make all settings and readings as accurately as possible. This will minimize the borderline normal/abnormal results which give no clear indication of whether the spectrophotometer needs professional service. The SPECTRONIC Standards have the same outside dimensions as a 10 mm square cuvette. If necessary, install an appropriate adapter (see Adapter and accessory filters) in the instrument's sample compartment. Operating technique The test results you obtain using SPECTRONIC Standards will be influenced by your technique in operating the instrument as well as by the condition of the instrument. Refer to the instrument operating manual to make sure you use the instrument correctly and avoid introducing errors into the test results. Adapter and accessory filters To use the SPECTRONIC Standards with the SPECTRONIC 20 + the Standard Adapter (for ½-inch square cuvettes) must first be installed. The appropriate cuvette holder is also required. Insert standards into the cuvette holder as shown below. Orient the holder so that the window side of the standard faces to the right. Push the holder down into the sample compartment as far as possible, and close the sample compartment lid. SPECTRONIC Standards User Guide 8

13 Note To accommodate the SPECTRONIC Standards, the ½-inch cuvette adapter and the 10 mm square cuvette holder are required. It is highly recommended that the Accessory Filter Kit (for all SPECTRONIC 20 + series) be used to optimize the performance of the spectrophotometers both in routine use and during tests with SPECTRONIC Standards. SPECTRONIC Standards User Guide 9

14 Recording and Interpreting Test Results Recording test results Test results should be recorded by date in a test log one log for each instrument tested. (A sample of such a log is provided at the end of this manual.) By scanning the log entries, the user can detect gradual changes of test values over a period of time. Any conditions or events that might affect test results (for example, cleaning or lamp replacement) should be recorded in the Comments column. Interpreting test results Although every effort is made to manufacture and measure the standards as accurately as possible, the wavelength and photometric performance standards have tolerances associated with them. In interpreting test results to evaluate instrument performance, the instrument tolerance and the standard tolerance should be added to determine the acceptable total deviation from nominal. For convenience, the following table gives the maximum deviations from nominal (instrument tolerance plus standard tolerance plus allowance for indeterminate error in the test procedure). SPECTRONIC Standards User Guide 10

15 Table 1 - Maximum Acceptable Deviations from Nominal for SPECTRONIC Standards Test Standard Spectrophotometer Stray Light Wavelength Accuracy 0%T Photometric Accuracy SPECTRONIC 20 + SPECTRONIC 20D %T* ±3.5 nm 0.0** ±3.0%T GENESYS %T ±3 nm 0.0*** ±1.2%T SPECTRONIC standards can be used with the instruments listed below, however, other NIST traceable standards are also available for photometric and wavelength accuracy. For increased testing confidence, please use the Wavelength/Photometric Accuracy Standards. Contact your sales or service representative for more information. GENESYS 10 Vis 0.1%T ±2 nm 0.0*** ±1.2%T GENESYS 10UVscanning 0.1%T ±2 nm 0.1%T ±1.2%T GENESYS 6 0.1%T ±2 nm 0.1%T ±1.2%T BioMate 3 0.1%T ±2 nm 0.1%T ±1.2%T GENESYS 10S VIS <0.1%T ±2 nm 0.0*** ±1.2%T GENESYS 10S UV-VIS <0.1%T ±2 nm 0.1%T ±1.2%T GENESYS 10S Bio <0.1%T ±2 nm 0.1%T ±1.2%T BioMate 3S <0.1%T ±2 nm 0.1%T ±1.2%T *With Accessory Filter Kit highly recommended accessories **An offset of ±0.2%T is acceptable ***An offset of ±0.1%T is acceptable Note All results are pictorial. All of the energy should strike the optical surface of the cell. The size and shape of the beam should be constant for the life of the instrument.. Use of Absorbance mode The instructions in this manual assume that the spectrophotometer is used in the transmittance mode. Test results can be read in absorbance and converted mathematically to transmittance (see equations below), but the following limitations must be considered: 1. Readings near 0%T will be overrange in absorbance. Therefore, the 0%T standard cannot be used. SPECTRONIC Standards User Guide 11

16 2. Estimates of stray light will be less accurate because they normally will be overrange in absorbance. Because of these limitations, it is recommended that the transmittance mode be used for any spectrophotometer on which it is available. The accuracy of wavelength tests is the same in either the transmittance or absorbance mode. A maximum in transmittance is a minimum in absorbance, so the wavelength peaks will be seen as minima in the absorbance mode. The wavelength of a peak is the same in either mode. Equations for the conversion of absorbance readings and transmittance readings: T 1 = = 10 A 10 A (1) Where T = %T/100 A = log10t (2) Where A = absorbance T = transmittance %T = percent transmittance 12 SPECTRONIC Standards User Guide

17 0% transmittance and stray light test Test Procedures For the SPECTRONIC 20 + and SPECTRONIC 20D + instruments follow these steps: 1. Set the wavelength to 400 nm. 2. Set the filter lever to the correct wavelength range. 3. Ensure that the sample compartment is empty. 4. Use the OFF-ON/0%T knob to set the readout to 0%T. 5. Insert the empty filter holder into the holder and then place in the sample compartment. 6. Use the front right knob or 100%T/0A knob to set the readout to 100%T. 7. Replace the empty filter holder with the 0%T standard, place in the sample compartment, and record the 0%T reading. If the 0%T reading exceeds the Maximum Acceptable Deviations as listed in Table 1, correct it. Repeat steps 1 through 7 before proceeding. 8. Replace the 0%T standard with the SRE standard whose wavelength corresponds to the setting in step 1 9. Place the ( ) filter from the Accessory Filter kit in the slot on the right side of the sample compartment and observe the %T reading. 10. Subtract the reading in step 7 from the reading in step Record the result, which is the stray light at the test wavelength. This value should conform to the summed tolerances given in Table 1 SPECTRONIC Standards User Guide 13

18 12. To perform the SRE test at 340 nm, place the ( ) filter from the Accessory Filter kit in the slot on the right side of the sample compartment and repeat procedure using a wavelength of 340 nm in step 1. For other spectrophotometer models, use the following procedures: 1. Set the wavelength to 400nm. 2. Set 100%T. 3. Insert the 0%T standard and record the 0%T reading. If the 0%T reading exceeds the Maximum Acceptable Deviations as listed in Table 1, correct it and repeat steps 1 through 3 before proceeding. 4. Replace the 0%T standard with the SRE standard whose wavelength corresponds to the setting in step 1 and observe the %T reading. 5. Subtract the reading in step 3 from the reading in Record the result, which is the stray light at the test wavelength. This value should conform to the summed tolerances given in Table Repeat the procedure using the wavelengths specified for your instrument in Table SPECTRONIC Standards User Guide

19 Table 2 Wavelengths to use when testing 0% transmittance and stray light Spectrophotometer SPECTRONIC 20 + and SPECTRONIC 20D + GENESYS 20 GENESYS 10 Vis GENESYS 10 UV GENESYS 10 UVscanning GENESYS 6 BioMate 3 GENESYS 10S VIS GENESYS 10S UV-Vis GENESYS 10S Bio BioMate 3S Wavelengths 400 nm, 340 nm 400 nm, 340 nm 400 nm, 340 nm 400 nm, 340 nm, 220 nm 400 nm, 340 nm, 220 nm 400 nm, 340 nm, 220 nm 400 nm, 340 nm, 220 nm 400 nm, 340 nm 400 nm, 340 nm, 220 nm 400 nm, 340 nm, 220 nm 400 nm, 340 nm, 220 nm Wavelength accuracy test 1. If the instrument has more than one source lamp, turn on the tungsten lamp. 2. Refer to the certificate provided with your set of SPECTRONIC Standards and set the wavelength display to read approximately 10 nm lower than the certified wavelength of the Primary Wavelength Test Peak indicated on the certificate. 3. With no samples in your sample compartment, adjust 100%T, set auto zero or Measure Blank (depending on your instrument). 4. Place the wavelength standard in the sample compartment and close the lid. While watching the %T display, slowly change the wavelength toward the certified wavelength as the %T reading increases. 5. Continue to increase the wavelength until the %T value starts to decrease and record the wavelength that gives the maximum %T value. SPECTRONIC Standards User Guide 15

20 This wavelength should match the certified wavelength within summed tolerances given in Table 1. The GENESYS model spectrophotometers (excluding the GENESYS 20) wavelength accuracy tests are performed differently. The GENESYS 10 series and BioMate 3 have a Performance Validation procedure under the Test menu, where the wavelength accuracy test can be performed using the SPECTRONIC Standards. Wavelength peak selection SPECTRONIC 20 + and SPECTRONIC 20D + Ensure that the filter lever is positioned to the correct wavelength range. Note For greater accuracy on non-scanning instruments, a point-by-point scan can be used. Read and plot the filter transmittance at a series of several discrete, closely-spaced wavelength settings around the expected location of a transmission peak. Carefully reset 100%T at each wavelength with the filter removed from the light beam. The advantage of this procedure is that any slope in the instrument's spectral response curve will have less effect on where the peak is found than it would in the method of direct, continuous scanning without resetting 100%T. The disadvantage is that it is time-consuming and tedious; many settings and readings must be carefully made and plotted to obtain one result. Our recommendation is to use the method given in the General Procedure rather than try to minimize error through this more time-consuming method. Photometric accuracy/linearity test 1. Set the wavelength to 590 nm. 2. Set 0%T with the sample compartment empty. Note There are a few 20D+ models with auto 0%T capability. If you have one of these, skip this step. 3. Insert the empty filter holder into the square cuvette holder and then place in the sample compartment. Adjust the instrument to read exactly 100%T. 4. Remove the empty filter holder from the square cuvette holder and reinsert the empty holder to open the built-in occluder. 16 SPECTRONIC Standards User Guide

21 When the empty holder is inserted, the 100%T reading should remain essentially unchanged. 5. Insert the photometric performance standard whose labeled value is near 50%T and record the instrument reading. The reading should be within the summed tolerance given in Table Replace the 50%T standard with the standard whose labeled value is near 10%T and record the instrument reading. The reading should be within the summed tolerance given in Table 1. SPECTRONIC Standards User Guide 17

22 SPECTRONIC Standards Test Log DATE 0%T STRAY LIGHT 220 nm 340 nm 400 nm nm PEAK WAVELENGTH ACCURACY nm PEAK nm PEAK PHOTOMETRIC ACCURACY 10%T 50%T COMMENTS SPECTRONIC Standards User Guide 18

Installation and User Guide. FlexIR TM NIR Fiber Optic Accessory

Installation and User Guide. FlexIR TM NIR Fiber Optic Accessory Installation and User Guide FlexIR TM NIR Fiber Optic Accessory The information in this publication is provided for reference only. All information contained in this publication is believed to be correct

More information

Thermo Scientific SPECTRONIC 200

Thermo Scientific SPECTRONIC 200 molecular spectroscopy Thermo Scientific SPECTRONIC 200 Part of Thermo Fisher Scientific The New Standard for Routine Measurements Robust, Multifunction Sample Compartment Whether you measure in 10 mm

More information

Thermo Scientific SPECTRONIC 200 Education

Thermo Scientific SPECTRONIC 200 Education molecular spectroscopy Thermo Scientific SPECTRONIC 200 Education Part of Thermo Fisher Scientific Designed for the Teaching Laboratory Classroom Friendly Sample Compartment Whether you measure in 10 mm

More information

NAME SECTION PERFORMANCE TASK # 3. Part I. Qualitative Relationships

NAME SECTION PERFORMANCE TASK # 3. Part I. Qualitative Relationships NAME SECTION PARTNERS DATE PERFORMANCE TASK # 3 You must work in teams of three or four (ask instructor) and will turn in ONE report. Answer all questions. Write in complete sentences. You must hand this

More information

Thermo Scientific SPECTRONIC 200 Visible Spectrophotometer. The perfect tool. for routine measurements

Thermo Scientific SPECTRONIC 200 Visible Spectrophotometer. The perfect tool. for routine measurements Thermo Scientific SPECTRONIC 200 Visible Spectrophotometer The perfect tool for routine measurements The Standard for Routine Measurements Thermo Scientific SPECTRONIC spectrophotometers have served as

More information

Thermo Scientific SPECTRONIC 200 Visible Spectrophotometer. The perfect. teaching instrument

Thermo Scientific SPECTRONIC 200 Visible Spectrophotometer. The perfect. teaching instrument Thermo Scientific SPECTRONIC 200 Visible Spectrophotometer The perfect teaching instrument Designed for the Teaching Laboratory Thermo Scientific SPECTRONIC spectrophotometers have served as core analytical

More information

Certified Reference Materials for UV, Visible, NIR and IR Molecular Spectroscopy

Certified Reference Materials for UV, Visible, NIR and IR Molecular Spectroscopy Certified Reference Materials for UV, Visible, NIR and IR Molecular Spectroscopy RM-1N3N5DHG Set Serial No: 23135 Customer Details: Starna Cells Inc 5950 Traffic Way Atascadero CA 93422 U.S.A. The customer

More information

UV / VIS Spectrophotometer EMCLAB Instruments GmbH

UV / VIS Spectrophotometer EMCLAB Instruments GmbH A Professional Manufacturer UV / VIS Spectrophotometer EMCLAB Instruments GmbH EMCLAB Instruments GmbH Bismarckstrasse 120 47057 Duisburg Germany Phone: +49 203 3064042 Fax: +49 203 3064044 E-mail: info@emc-lab.de

More information

Agilent Cary 7000 Universal Measurement Spectrophotometer (UMS)

Agilent Cary 7000 Universal Measurement Spectrophotometer (UMS) Agilent Cary 7000 Universal Measurement Spectrophotometer (UMS) Specifications Introduction The Agilent Cary 7000 Universal Measurement Spectrophotometer (UMS) is designed for superior performance, flexibility

More information

Operator s Manual. PP016 Passive Probe

Operator s Manual. PP016 Passive Probe Operator s Manual PP016 Passive Probe 2017 Teledyne LeCroy, Inc. All rights reserved. Unauthorized duplication of Teledyne LeCroy documentation materials is strictly prohibited. Customers are permitted

More information

SP-8001 UV/Visible Spectrophotometer

SP-8001 UV/Visible Spectrophotometer SP-8001 UV/Visible Spectrophotometer Provide Accurate Results High performance UV/Visible Spectrophotometer for Research, and Laboratory work Product Features The Metertech SP-8001 UV/Visible Spectrophotometer

More information

PD-3000UV. INSTRUCTION MANUAL UV-VIS Spectrophotometer OMA IMPORTANT!

PD-3000UV. INSTRUCTION MANUAL UV-VIS Spectrophotometer OMA IMPORTANT! INSTRUCTION MANUAL UV-VIS Spectrophotometer PD-3000UV IMPORTANT! READ THIS OPERATION MANUAL! *Prior to operation, please read this manual carefully and retain it for future reference. Failure to read,

More information

2015 RIGOL TECHNOLOGIES, INC.

2015 RIGOL TECHNOLOGIES, INC. Service Guide DG000 Series Dual-channel Function/Arbitrary Waveform Generator Oct. 205 TECHNOLOGIES, INC. Guaranty and Declaration Copyright 203 TECHNOLOGIES, INC. All Rights Reserved. Trademark Information

More information

University of Wisconsin Chemistry 524 Spectroscopic Components *

University of Wisconsin Chemistry 524 Spectroscopic Components * University of Wisconsin Chemistry 524 Spectroscopic Components * In journal articles, presentations, and textbooks, chemical instruments are often represented as block diagrams. These block diagrams highlight

More information

Fiber Optic Sampling by UV/Vis and UV/Vis/NIR Spectroscopy

Fiber Optic Sampling by UV/Vis and UV/Vis/NIR Spectroscopy Fiber Optic Sampling by UV/Vis and UV/Vis/NIR Spectroscopy UV/VIS AND UV/VIS/NIR SPECTROSCOPY A P P L I C A T I O N N O T E Introduction The availability and applications of fiber optic sampling in UV/Vis

More information

30 Bending Brake. Model Assembly and Operating Instructions. Distributed exclusively by Harbor Freight Tools.

30 Bending Brake. Model Assembly and Operating Instructions. Distributed exclusively by Harbor Freight Tools. 30 Bending Brake Model 41311 Assembly and Operating Instructions Distributed exclusively by Harbor Freight Tools. 3491 Mission Oaks Blvd., Camarillo, CA 93011 Copyright 1999 by Harbor Freight Tools. All

More information

MANUAL BMS SPEC5000 Spectrophotometer

MANUAL BMS SPEC5000 Spectrophotometer MANUAL BMS SPEC5000 Spectrophotometer BMS Art. Nr. 17020 1 PRINCIPLE, APPLICATION AND CHARACTERISTICS 1.1 Principle The SPEC5000 spectrophotometer is a single beam spectrophotometer with a basic structure

More information

9 PIECE TUNGSTEN CARBIDE HOLE SAW KIT. Model 90721

9 PIECE TUNGSTEN CARBIDE HOLE SAW KIT. Model 90721 9 PIECE TUNGSTEN CARBIDE HOLE SAW KIT Model 90721 Set up And Operating Instructions Diagrams within this manual may not be drawn proportionally. Due to continuing improvements, actual product may differ

More information

Radio System Strobe Wizard Plus Freemask

Radio System Strobe Wizard Plus Freemask Radio System Strobe Wizard Plus Freemask User manual Translation of the original German user manual Doc. No.: 900.0509.00 Version: 09/2017 Contents Information about this manual and about the manufacturer...

More information

Planishing hammer stand For use with SKU Planishing hammer

Planishing hammer stand For use with SKU Planishing hammer Planishing hammer stand For use with SKU 94847 Planishing hammer Model 96300 Assembly And Operation Instructions Please Note: Planishing Hammer not included with Stand. Due to continuing improvements,

More information

Automated Double Aperture Accessory

Automated Double Aperture Accessory For the Cary 1, 3, 100, 300, 4, 5, 400, 500, 500i, 4000, 5000, 6000i, Deep UV Installation Category II Pollution Degree 2 Equipment Class I Table of Contents Introduction Theory Operation Installation

More information

UV / VIS Spectrophotometer

UV / VIS Spectrophotometer UV / VIS Spectrophotometer Single Beam Double Beam NANO EMCLAB Instruments GmbH EMCLAB Instruments GmbH Kulturstrasse 55 47055 Duisburg Germany Fon: +49 203 3064042 Fax: +49 203 41504809 E-mail: info@emc-lab.de

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 07-424 Digital Densitometer II Operators Manual March 2005 Manual No. 112111 Rev. 4 2003, 2005 Fluke Corporation, All rights reserved. Printed U.S.A. All product names are trademarks

More information

Spectrophotometer. An instrument used to make absorbance, transmittance or emission measurements is known as a spectrophotometer :

Spectrophotometer. An instrument used to make absorbance, transmittance or emission measurements is known as a spectrophotometer : Spectrophotometer An instrument used to make absorbance, transmittance or emission measurements is known as a spectrophotometer : Spectrophotometer components Excitation sources Deuterium Lamp Tungsten

More information

Model 9305 Fast Preamplifier Operating and Service Manual

Model 9305 Fast Preamplifier Operating and Service Manual Model 9305 Fast Preamplifier Operating and Service Manual This manual applies to instruments marked Rev 03" on rear panel. Printed in U.S.A. ORTEC Part No.605540 1202 Manual Revision B Advanced Measurement

More information

Miniature Spectrometer Technical specifications

Miniature Spectrometer Technical specifications Miniature Spectrometer Technical specifications Ref: MSP-ISI-TEC 001-02 Date: 2017-05-05 Contact Details Correspondence Address: Email: Phone: IS-Instruments Ltd. Pipers Business Centre 220 Vale Road Tonbridge

More information

15 Planer Stand. Model Due to continuing improvements, actual product may differ slightly from the product described herein.

15 Planer Stand. Model Due to continuing improvements, actual product may differ slightly from the product described herein. 15 Planer Stand Model 96316 Assembly And Operation Instructions Due to continuing improvements, actual product may differ slightly from the product described herein. 3491 Mission Oaks Blvd., Camarillo,

More information

SAFE USE &OPERATIONS MANUAL

SAFE USE &OPERATIONS MANUAL www.uvprocess.com Product information Updates on UV Process Supply products ITEM PART NO. VersaProbe standard package M007-092 (Standard package includes UV-A+B probe.) Individual UV-A Remote Probe M007-093

More information

Owner s Manual & Safety Instructions

Owner s Manual & Safety Instructions Owner s Manual & Safety Instructions Save This Manual Keep this manual for the safety warnings and precautions, assembly, operating, inspection, maintenance and cleaning procedures. Write the product s

More information

T92+ UV-VIS SPECTROPHOTOMETER

T92+ UV-VIS SPECTROPHOTOMETER T92+ UV-VIS SPECTROPHOTOMETER The T92+ is a high performance double beam spectrophotometer with a variable spectral bandwidth from 0.1-5nm, selected by a continuous variable slit. The instrument is fitted

More information

Signature Choral Riser Side Rail

Signature Choral Riser Side Rail Assembly/Owner s Manual Signature Choral Riser Side Rail Signature Choral 3-Step Riser with Optional Side Rail Signature Choral 4-Step Riser with Optional Side Rail CONTENTS Visit the Signature Choral

More information

3M No Polish SC/APC Angle Splice Connector Jacket for 2 x 3 mm FRP and 1.6 to 3.0 mm Cable 8802-T/APC/AS/1.6-3

3M No Polish SC/APC Angle Splice Connector Jacket for 2 x 3 mm FRP and 1.6 to 3.0 mm Cable 8802-T/APC/AS/1.6-3 3M No Polish SC/APC Angle Splice Connector Jacket for 2 x 3 mm FRP and 1.6 to 3.0 mm Cable 8802-T/APC/AS/1.6-3 Instructions October 2013 3 1.0 Table of contents 1.0 Summary...3 2.0 Cable and Fiber Preparation...4

More information

Model 9302 Amplifier-Discriminator Operating and Service Manual

Model 9302 Amplifier-Discriminator Operating and Service Manual Model 9302 Amplifier-Discriminator Operating and Service Manual Printed in U.S.A. ORTEC Part No. 733690 1202 Manual Revision C Advanced Measurement Technology, Inc. a/k/a/ ORTEC, a subsidiary of AMETEK,

More information

UV-Vis-NIR Spectrophotometer Quick Start Guide

UV-Vis-NIR Spectrophotometer Quick Start Guide UV-Vis-NIR Spectrophotometer Quick Start Guide The following instructions are provided as a Quick Start Guide for powering up, running measurements, and shutting down the Lambda 950 UV-Vis Spectrophotometer.

More information

Wenger Corporation 2010 Printed in China 02/10 Part #148H Wenger Corporation, 555 Park Drive, P.O. Box 448, Owatonna, Minnesota

Wenger Corporation 2010 Printed in China 02/10 Part #148H Wenger Corporation, 555 Park Drive, P.O. Box 448, Owatonna, Minnesota Assembly Instructions Violin/Viola Storage Rack CONTENTS Safety Precautions.................................. Warranty.......................................... Important User Information............................

More information

Content Spectrophotometers

Content Spectrophotometers Spectroph Content Spectrophotometers Selection table Spectrophotometers Page 163 PRIM Page 164 UviLine Page 166 162 Selection table Spectrophotometers PRIM Light/ PRIM Advanced UviLine 9100/ UviLine 9400

More information

Read and be familiar with this manual before operating or servicing this device. To ensure operator, technician, and patient safety, use only as

Read and be familiar with this manual before operating or servicing this device. To ensure operator, technician, and patient safety, use only as Read and be familiar with this manual before operating or servicing this device. To ensure operator, technician, and patient safety, use only as specified in this manual. Product Manufacturer: Medical

More information

Guitar Mobile Storage Rack

Guitar Mobile Storage Rack Assembly Instructions Guitar Mobile Storage Rack CONTENTS Safety Precautions.................................. 2 Warranty.......................................... 2 Operation.........................................

More information

Operating Manual. Model 721N. Visible Spectrophotometer

Operating Manual. Model 721N. Visible Spectrophotometer Operating Manual of Model 721N Visible Spectrophotometer 1 Table of Contents 1. Chief uses... 3 2. Working Conditions... 3 3. Main Specifications...3 4.Operating Principles...4 5. Optical design...4 6.

More information

ROLLER STAND WITH TWO-LEG SUPPORT ASSEMBLY AND OPERATING INSTRUCTIONS

ROLLER STAND WITH TWO-LEG SUPPORT ASSEMBLY AND OPERATING INSTRUCTIONS ROLLER STAND WITH TWO-LEG SUPPORT 95621 ASSEMBLY AND OPERATING INSTRUCTIONS 3491 Mission Oaks Blvd., Camarillo, CA 93011 Visit our Web site at http://www.harborfreight.com Copyright 2006 by Harbor Freight

More information

Visible & UV/Vis Spectrophotometers

Visible & UV/Vis Spectrophotometers Visible & UV/Vis Spectrophotometers Accurate Precise Affordable S20 Series UV/Vis Spectrophotometers The Unico S20 Series Spectrophotometers are simple, flexible, easytouse and maximize value while providing

More information

Model 863 Quad Timing Filter Amplifier Operating and Service Manual

Model 863 Quad Timing Filter Amplifier Operating and Service Manual Model 863 Quad Timing Filter Amplifier Operating and Service Manual Printed in U.S.A. ORTEC Part No. 733960 0411 Manual Revision C Advanced Measurement Technology, Inc. a/k/a/ ORTEC, a subsidiary of AMETEK,

More information

Owner s Manual & Safety Instructions

Owner s Manual & Safety Instructions Owner s Manual & Safety Instructions Save This Manual Keep this manual for the safety warnings and precautions, assembly, operating, inspection, maintenance and cleaning procedures. Write the product s

More information

9 QUICK RELEASE WOODWORKING VISE

9 QUICK RELEASE WOODWORKING VISE 9 QUICK RELEASE WOODWORKING VISE 94386 OPERATING INSTRUCTIONS Due to continuing improvements, actual product may differ slightly from the product described herein. 349 Mission Oaks Blvd., Camarillo, CA

More information

Installation and Operation Manual Document Number 000-specc

Installation and Operation Manual Document Number 000-specc SpectroClip Probe SpectroClip-R/SpectroClip-TR/Jaz-SpectroClip-TR Installation and Operation Manual Document Number 000-specc-110-02-201203 Offices: Ocean Optics, Inc. World Headquarters 830 Douglas Ave.,

More information

Beaming Optical Length Tester (BOLT)

Beaming Optical Length Tester (BOLT) O.W.L. MANUFACTURER OF QUALITY OPTICAL FIBER TEST EQUIPMENT OPTICAL WAVELENGTH LABORATORIES Beaming Optical Length Tester (BOLT) Operations Manual Version 1.0 November 8, 001 OWL Part BOLT-1 TABLE OF CONTENTS

More information

34134A AC/DC DMM Current Probe. User s Guide. Publication number April 2009

34134A AC/DC DMM Current Probe. User s Guide. Publication number April 2009 User s Guide Publication number 34134-90001 April 2009 For Safety information, Warranties, Regulatory information, and publishing information, see the pages at the back of this book. Copyright Agilent

More information

VT1586A Rack Mount Terminal Panel Installation and User s Manual

VT1586A Rack Mount Terminal Panel Installation and User s Manual VT1586A Rack Mount Terminal Panel Installation and User s Manual Manual Part Number: 82-0095-000 Rev. June 16, 2003 Printed in U.S.A. Certification VXI Technology, Inc. certifies that this product met

More information

accidents which arise due to nonobservance and the safety information herein. SPECIFICATIONS

accidents which arise due to nonobservance and the safety information herein. SPECIFICATIONS 7 TON ELECTRIC LOG SPLITTER Model: 05620 CALIFORNIA PROPOSITION 65 WARNING: You can create dust when you cut, sand, drill or grind materials such as wood, paint, metal, concrete, cement, or other masonry.

More information

Current Probes. User Manual

Current Probes. User Manual Current Probes User Manual ETS-Lindgren Inc. reserves the right to make changes to any product described herein in order to improve function, design, or for any other reason. Nothing contained herein shall

More information

Analog Voltage Detector 1-450kV Operating Instructions VDAO450TM. Patent No. 6,275,022

Analog Voltage Detector 1-450kV Operating Instructions VDAO450TM. Patent No. 6,275,022 Analog Voltage Detector 1-450kV Operating Instructions VDAO450TM Patent No. 6,275,022 CONTENTS Limitation of Warranty and Liability 2 Product Safety Information 3 Design and Function 4 Operating Procedure

More information

Spectro UV-VIS Double Beam Research Spectrophotometer UV-VIS Double Beam

Spectro UV-VIS Double Beam Research Spectrophotometer UV-VIS Double Beam Spectro UV-VIS Double Beam Research Spectrophotometer UV-VIS Double Beam Software Specifications Model UVD-3500 Spectro UV-VIS Double Beam UVD 3500 Research Spectrophotometer is a superior instrument for

More information

Model 113 Scintillation Preamplifier Operating and Service Manual

Model 113 Scintillation Preamplifier Operating and Service Manual Model 113 Scintillation Preamplifier Operating and Service Manual Printed in U.S.A. ORTEC Part No. 717560 1202 Manual Revision B Advanced Measurement Technology, Inc. a/k/a/ ORTEC, a subsidiary of AMETEK,

More information

LED Driving Technology for Long Term Flexibility Application Note

LED Driving Technology for Long Term Flexibility Application Note LED Driving Technology for Long Term Flexibility Application Note Abstract In order to guarantee constant brightness for LED illumination systems with long product cycle times, the availability of LEDs

More information

Model 533 Dual Sum and Invert Amplifier Operating and Service Manual

Model 533 Dual Sum and Invert Amplifier Operating and Service Manual Model 533 Dual Sum and Invert Amplifier Operating and Service Manual Printed in U.S.A. ORTEC Part No. 733410 1202 Manual Revision B Advanced Measurement Technology, Inc. a/k/a/ ORTEC, a subsidiary of AMETEK,

More information

USER MANUAL FOR VISIBLE SPECTROPHOTOMETER

USER MANUAL FOR VISIBLE SPECTROPHOTOMETER USER MANUAL FOR VISIBLE SPECTROPHOTOMETER 1 Table of Contents 1. MAIN USAGES...3 2. WORKING ENVIRONMENT...3 3. MAIN TECHNICAL DATA AND SPECIFICATIONS...4 4. WORKING PRINCIPLE...5 5. OPTICAL PRINCIPLE...6

More information

SHRINKER/STRETCHER SET Create radius bends and contours in sheet metal

SHRINKER/STRETCHER SET Create radius bends and contours in sheet metal Owner s Manual & Safety Instructions Save This Manual Keep this manual for the safety warnings and precautions, assembly, operating, inspection, maintenance and cleaning procedures. Write the product s

More information

ITEM Pipe Thread Kit (10 Piece Set - 7 Dies) Assembly and Operating Instructions

ITEM Pipe Thread Kit (10 Piece Set - 7 Dies) Assembly and Operating Instructions ITEM 94098 Pipe Thread Kit (10 Piece Set - 7 Dies) Assembly and Operating Instructions 3491 Mission Oaks Blvd., Camarillo, CA 93011 Copyright 2006 by Harbor Freight Tools. All rights reserved. No portion

More information

AT Advanced Wire Tracer. Users Manual

AT Advanced Wire Tracer. Users Manual AT-1000 Advanced Wire Tracer Users Manual AT-1000 Advanced Wire Tracer English Users Manual AT1000_Rev001 2008 Amprobe Test Tools. All rights reserved. Limited Warranty and Limitation of Liability Your

More information

INSTRUCTION MANUAL INF Fax: (503)

INSTRUCTION MANUAL INF Fax: (503) INSTRUCTION MANUAL INF151 1-800-547-5740 Fax: (503) 643-6322 www.ueiautomotive.com email: info@ueitest.com Introduction Congratulations on your purchase of the INF151 infrared thermometer. Like all UEi

More information

14 Piece Slide Hammer and Puller Set

14 Piece Slide Hammer and Puller Set Owner s Manual & Safety Instructions Save This This Manual Keep Keep this this manual manual for for the the safety safety warnings warnings and and precautions, assembly, assembly, operating, inspection,

More information

UVS-2800 Spectro UV-VIS Split Beam (PC) is a precise scanning

UVS-2800 Spectro UV-VIS Split Beam (PC) is a precise scanning Spectro UV-VIS Split Beam (PC) is a precise scanning Spectrophotometer with a new design of 8 microprocessor automatic 2 row cell holder that moves noiseless with a special membrane. This Split Beam Spectro

More information

Agilent G1888 Network Headspace Sampler

Agilent G1888 Network Headspace Sampler Agilent G1888 Network Headspace Sampler Safety and Regulatory Information Agilent Technologies Notices Agilent Technologies, Inc. 2004 No part of this manual may be reproduced in any form or by any means

More information

Drawer Unit Assembly. Wenger Corporation 2006 Printed in USA 12/06 Part #121B115-01

Drawer Unit Assembly. Wenger Corporation 2006 Printed in USA 12/06 Part #121B115-01 Assembly Instructions Rehearsal Resource Center Folio Box Option Model 121 Drawer Unit Assembly CONTENTS Warranty.......................................... 3 Important User Information............................

More information

USB. Part No Wavelength range. Spectral bandwidth 5 nm 4 nm Optical system

USB. Part No Wavelength range. Spectral bandwidth 5 nm 4 nm Optical system Visible range spectrophotometers V-1100 and VR-2000 V-1100 MODEL WITH MANUAL WAVELENGTH SETTINGS AND AUTOMATIC BLANK. VR-2000 MODEL WITH AUTOMATIC WAVELENGTH SETTINGS AND BLANK. V-1100 Part no. 4120025

More information

Agilent N2902A 9000 Series Oscilloscope Rack Mount Kit

Agilent N2902A 9000 Series Oscilloscope Rack Mount Kit Agilent N2902A 9000 Series Oscilloscope Rack Mount Kit Installation Guide Agilent Technologies Notices Agilent Technologies, Inc. 2009 No part of this manual may be reproduced in any form or by any means

More information

Model 416A Gate and Delay Generator Operating and Service Manual

Model 416A Gate and Delay Generator Operating and Service Manual Model 416A Gate and Delay Generator Operating and Service Manual Printed in U.S.A. ORTEC Part No. 733160 1202 Manual Revision E Advanced Measurement Technology, Inc. a/k/a/ ORTEC, a subsidiary of AMETEK,

More information

HEIGHT ADJUSTABLE WORKBENCH

HEIGHT ADJUSTABLE WORKBENCH HEIGHT ADJUSTABLE WORKBENCH Model 91 ASSEMBLY and Operating Instructions Visit our website at: http://www.harborfreight.com Read this material before using this product. Failure to do so can result in

More information

1369 & 1379 MOUNT KITS

1369 & 1379 MOUNT KITS November 15, 2014 Lit. No. 64573, Rev. 01 1369 & 1379 MOUNT KITS 1369 Ford Bronco, F 150 4X4 1992-96 1379 Ford F 250/350 4X4 1992-97 Ford F 250 2WD (over 8,500 GVWR) 1992-97 Ford F 350 2WD (over 10,000

More information

COVERT GNSS WINDOW MOUNT ANTENNA PMAN4109

COVERT GNSS WINDOW MOUNT ANTENNA PMAN4109 COVERT GNSS WINDOW MOUNT ANTENNA PMAN4109 Installation Manual Foreword This manual applies to digital mobile radios, unless otherwise specified. It includes instructions for installing the Covert GNSS

More information

Your Performance Partner

Your Performance Partner Assembly Instructions Attic Storage System Your Performance Partner CONTENTS Safety Precautions.................................. 2 Warranty.......................................... 2 Important User Information............................

More information

SlimScan BT Users Manual

SlimScan BT Users Manual SlimScan BT Users Manual ASP Microcomputers 456 North Road, Ormond, Victoria, 3204 Australia Telephone: (03) 9578-7600 FAX: (03) 9578-7727 email: solutions@asp.com.au World Wide Web: http://www.asp.com.au

More information

HP 86290B RF PLUG-IN GHz HEWLETT PACKARD

HP 86290B RF PLUG-IN GHz HEWLETT PACKARD OPERATING AND SERVICE MANUAL. HP 86290B RF PLUG-IN 2.0-18.6 GHz HEWLETT PACKARD COPYRIGHT AND DISCLAIMER NOTICE Copyright - Agilent Technologies, Inc. Reproduced with the permission of Agilent Technologies

More information

JAZ-ALP Adjustable Laser Pointer Operation Manual Document Number 013-RD

JAZ-ALP Adjustable Laser Pointer Operation Manual Document Number 013-RD JAZ-ALP Adjustable Laser Pointer Operation Manual Document Number 013-RD000-001-02-201208 Offices: Ocean Optics, Inc. World Headquarters 830 Douglas Ave., Dunedin, FL, USA 34698 Phone 727.733.2447 Fax

More information

User Manual. Flooring Removal Machine (Patent Pending) Maintenance and Operating Instructions

User Manual. Flooring Removal Machine (Patent Pending) Maintenance and Operating Instructions User Manual Flooring Removal Machine (Patent Pending) Maintenance and Operating Instructions Copyright 2012 by Carpet Concepts LLC. All Rights Reserved. For technical questions or replacement parts please

More information

HP Designjet HD Scanner and T1200 HD Multifunction Printer

HP Designjet HD Scanner and T1200 HD Multifunction Printer HP Designjet HD Scanner and T1200 HD Multifunction Printer Introductory information USB 2.0 high-speed certified Introductory Information Other sources of information The User s Guide for your scanner

More information

Educational Spectrophotometer Accessory Kit and System OS-8537 and OS-8539

Educational Spectrophotometer Accessory Kit and System OS-8537 and OS-8539 GAIN 1 10 Instruction Manual with Experiment Guide and Teachers Notes 012-06575C *012-06575* Educational Spectrophotometer Accessory Kit and System OS-8537 and OS-8539 100 CI-6604A LIGHT SENSOR POLARIZER

More information

P5100A & P5150 High Voltage Probes Performance Verification and Adjustments

P5100A & P5150 High Voltage Probes Performance Verification and Adjustments x P5100A & P5150 High Voltage Probes Performance Verification and Adjustments ZZZ Technical Reference *P077053001* 077-0530-01 xx P5100A & P5150 High Voltage Probes Performance Verification and Adjustments

More information

User s Guide. RP7000 Series Active Probe. Dec RIGOL Technologies, Inc.

User s Guide. RP7000 Series Active Probe. Dec RIGOL Technologies, Inc. User s Guide RP7000 Series Active Probe Dec. 2012 RIGOL Technologies, Inc. Guaranty and Declaration Copyright 2011 RIGOL Technologies, Inc. All Rights Reserved. Trademark Information RIGOL is a registered

More information

REF42plus HMI Cutting Tool Operating Guide

REF42plus HMI Cutting Tool Operating Guide REF42plus HMI Cutting Tool Operating Guide 2 Copyright This document and parts thereof must not be reproduced or copied without written permission from ABB, and the contents thereof must not be imparted

More information

High Temperature InfraRed Thermometer with Laser Pointer

High Temperature InfraRed Thermometer with Laser Pointer User s Manual High Temperature InfraRed Thermometer with Laser Pointer MODEL 42545 42545 Introduction Congratulations on your purchase of the Model 42545 IR Thermometer. The 42545 is capable of non-contact

More information

HAND HELD SAW W MILL

HAND HELD SAW W MILL HAND HELD SAW W MILL 92247 ASSEMBLY AND OPERATING INSTRUCTIONS 3491 Mission Oaks Blvd., Camarillo, CA 93011 Visit our Web site at http://www.harborfreight.com Copyright 2004 by Harbor Freight Tools. All

More information

Keysight 86205B RF Bridge

Keysight 86205B RF Bridge Keysight 86205B RF Bridge Operating and Service Manual Notices Keysight Technologies 2011, 2014 No part of this manual may be reproduced in any form or by any means (including electronic storage and

More information

APPLICATION NOTE. Computer Controlled Variable Attenuator for Lasers. Technology and Applications Center Newport Corporation

APPLICATION NOTE. Computer Controlled Variable Attenuator for Lasers. Technology and Applications Center Newport Corporation APPLICATION NOTE Computer Controlled Variable Attenuator for Lasers 31 Technology and Applications Center Newport Corporation Computer Controlled Variable Attenuator for Lasers This application note describes

More information

Flat Sample Holder Part Number

Flat Sample Holder Part Number Flat Sample Holder Part Number 990-00403 (Patent Pending) Operator s Manual Copyright 2016 Gamry Instruments, Inc. April 8, 2016 988-00043 Rev. 1 If You Have Problems Please visit our service and support

More information

Page 1 BIOIMAGER. Spectrophotometers. Toll Free: (855) BIO- IMAG / (855) , E: S:

Page 1 BIOIMAGER. Spectrophotometers. Toll Free: (855) BIO- IMAG / (855) , E: S: Page 1 Comparison of all models Page 2 Model BK- V1000 BK- UV1000 BK- V1200 BK- UV1200 BK- V1600 BK- UV1600 BK- V1800 BK- UV1800 BK- V1900 BK- UV1900 BK- S360 BK- S380 BK- S390 BK- D560 BK- D580 BK- D590

More information

LC200WT Security Wall Mount with Tilt for up to 32" Flat Screens

LC200WT Security Wall Mount with Tilt for up to 32 Flat Screens Page 1 of 6 The LC200WT Security Wall Mount with Tilt is designed to secure up to a 32" flat panel monitor to the wall while still allowing the monitor to tilt. The VESA mounting patterns on the front

More information

Agilent N7509A Waveform Generation Toolbox Application Program

Agilent N7509A Waveform Generation Toolbox Application Program Agilent N7509A Waveform Generation Toolbox Application Program User s Guide Second edition, April 2005 Agilent Technologies Notices Agilent Technologies, Inc. 2005 No part of this manual may be reproduced

More information

Assembly Instructions and Owner s Manual Flex TechBridge

Assembly Instructions and Owner s Manual Flex TechBridge Assembly Instructions and Owner s Manual Flex TechBridge Contents Important User Information...........................2 Safety Precautions.................................2 Required Tools.....................................3

More information

2011 / Circuit Tracer

2011 / Circuit Tracer INSTRUCTION MANUAL 2011 / 00521 Circuit Tracer Read and understand all of the instructions and safety information in this manual before operating or servicing this tool. 52044992 2008 Greenlee Textron

More information

Solid Sample Holder Accessory

Solid Sample Holder Accessory Installation category I Pollution degree 2 Equipment class III Introduction The Solid Sample Holder for the Agilent Cary Eclipse is an accessory that enables you to perform fluorescence measurements on

More information

50 Gordon Drive, Rockland, Maine sherplows.com MOUNT KIT. Installation Instructions CAUTION

50 Gordon Drive, Rockland, Maine sherplows.com MOUNT KIT. Installation Instructions CAUTION Fisher Engineering 50 Gordon Drive, Rockland, Maine 04841-2139 www.fi sherplows.com December 1, 2018 Lit. No. 77306, Rev. 00 MOUNT KIT Ram 1500 2019 - Installation Instructions Read this document before

More information

MANUAL METAL SHRINKER/STRETCHER

MANUAL METAL SHRINKER/STRETCHER MANUAL METAL SHRINKER/STRETCHER Model 95062 ASSEMBLY AND OPERATING INSTRUCTIONS Due to continuing improvements, actual product may differ slightly from the product described herein. 3491 Mission Oaks Blvd.,

More information

Z-5652 plus Series. 2D Image Hands-Free Scanner

Z-5652 plus Series. 2D Image Hands-Free Scanner Z-5652 plus Series 1 2D Image Hands-Free Scanner Revision History Changes to the original manual are listed below: Version Date Description of Version 1.0 10/02/2017 Initial release 2D Image Scan Module

More information

10 Ton Pull Back Ram

10 Ton Pull Back Ram 10 Ton Pull Back Ram Model 97207 Set up And Operating Instructions Diagrams within this manual may not be drawn proportionally. Due to continuing improvements, actual product may differ slightly from the

More information

Assembly Instructions Signature Choral Riser 4-Step Model

Assembly Instructions Signature Choral Riser 4-Step Model Assembly Instructions Signature Choral Riser 4-Step Model Contents Important User Information...........................2 General...2 Manufacturer...2 Intended Use...2 Warranty...2 Safety Precautions.................................3

More information

ATV CULTIVATOR OWNER S MANUAL

ATV CULTIVATOR OWNER S MANUAL ATV CULTIVATOR OWNER S MANUAL WARNING: Read carefully and understand all ASSEMBLY AND OPERATION INSTRUCTIONS before operating. Failure to follow the safety rules and other basic safety precautions may

More information

User Guide. Keysight N6850A Broadband Omnidirectional Antenna

User Guide. Keysight N6850A Broadband Omnidirectional Antenna User Guide Keysight N6850A Broadband Omnidirectional Antenna Notices Keysight Technologies, Inc. 2012-2015 No part of this manual may be reproduced in any form or by any means (including electronic storage

More information

Heiland electronic GmbH TD / TD1 / TD2. B&W-Densitometers. USERS MANUAL Version 5

Heiland electronic GmbH TD / TD1 / TD2. B&W-Densitometers. USERS MANUAL Version 5 Heiland electronic GmbH TD / TD1 / TD2 B&W-Densitometers USERS MANUAL Version 5 2 Table of Contents 1. GENERAL INFORMATION...4 2. SAFETY REGULATIONS...5 3. AREA OF APPLICATIONS...5 4. INSTRUMENT DESCRIPTION...6

More information

TECHNICAL MEMORANDUM

TECHNICAL MEMORANDUM TECHNICAL MEMORANDUM NOTICE: This document is version controlled and was produced as a part of the GEX Information Program which requires that all Series 100 documents be reviewed periodically to maintain

More information