1 INSTRUCTION MANUAL UV-VIS Spectrophotometer PD-3000UV IMPORTANT! READ THIS OPERATION MANUAL! *Prior to operation, please read this manual carefully and retain it for future reference. Failure to read, understand and follow the instructions in this manual may result in damage to the unit, injury to operating personnel and/or poor equipment performance. All rights are reserved. permission of APEL CO., LTD. This material may not be reproduced or copied, in whole or part, without the written OMA
2 TABLE OF CONTENTS CHAPTER 1 Introduction CHAPTER 2 Safety Instructions CHAPTER 3 Specifications CHAPTER 4 Structure Front view Control panel CHAPTER 5 Function and Operation Function Start and cautions before the start of the instrument Start and check Photometric measurement(abs.,%t) Quantitative analysis(standard curve method) Coefficient method Selection of concentration unit System setting Chapter 6 Maintenance Packing List
3 CHAPTER 1 Introduction APEL UV-VIS Spectrophotometer PD-3000UV can be used for qualitative analysis and quantitative analysis from the Ultraviolet to Near Infrared wavelength range from nm. The structure of PD-3000UV is simple but stable and accurate, and ideal for use in education, clinical laboratory, biochemistry laboratory, petro chemistry laboratory, environmental protection and various fields of quality control. CHAPTER 2 Safety Instructions 1. Operating temperature should be 5-35, and the humidity should be less than 85%. 2. Put the instrument in an area that is free of strong continuous vibration. 3. Keep the instrument away from the direct sunlight. 4. Keep the direction of air conditioner or electric fan away from the instrument so as to avoid their influence. 5. Keep the instrument away from any electric device that may generate strong magnetic or high frequency electromagnetic field. 6. Confirm the supply voltage and the frequency according to the specifications. And connect earth terminal to the ground. 7. Do not use the instrument in corrosive gases for long time, otherwise it may cause significant shortening of the life of the instrument as well as damage. 8. It is recommended that the instrument is used and stored in clean place. CHAPTER 3 Specifications Wavelength range nm Band width 2nm Wavelength accuracy ±1nm Wavelength repeatability 0.5nm Photometric accuracy ±0.3%T Drift 0.002A/h Stray light 0.1%T Data output port USB Printer port Parallel port Display 128x64 LCD Light source Deuterium lamp (D2) & Tungsten Halogen lamp Photometric Display range 0-200%T A C Dimension (L x W x H) 480x360x200mm 2
4 CHAPTER 4 Structure 1. Front view LCD display Control panel Sample compartment cover Cuvette selector arm 2. Control panel Screen Function key, Edit key Numerical keys 1)Numerical keys Used when the numerical data are input such as wavelength, concentration, date etc. 2)Function keys SET GO TO Used to set each function according to each measuring mode. Used to set the wavelength. 3
5 ZERO PRINT Used to set the value of the blank liquid to Abs or %T Used to print the measuring results. 3)Edit key Upward key Downward key START STOP CLEAR ESC Used to start or stop the test. Used to delete the numerical data etc. Used to return to the previous screen. Used to fix the numerical data etc. 4
6 CHAPTER 5 Function and Operation This chapter describes how to operate the instrument. 1. Function of the instrument(the flow of the operation) 2. Start and the cautions before the start of the instrument 3. Start and check 4. Photometric measurement(t%,abs) 5. Quantitative analysis(including standard curve) 6. Coefficient method 7. Selection of concentration unit 8. System setting 5
7 1. Function Screen open System auto-check Photometry Quantitative analysis System setting Measuring mode Absorbance Transmittance Standard curve method New curve drawing Curve drawing ON/OFF of D2 lamp ON/OFF of halogen lamp Date setting Energy Gain setting Data handling Data print Data delete Display the coordinate Data test Data print Data delete Load curve data Dark current calibration Wavelength calibration Lamp change wavelength Initialization Version No. Select a curve Wavelength set Blank 0.000A/100.0%T Display the coordinate Data test Data print Data delete Delete the standard curve Coefficient method Set parameter K Set parameter B Data test Data print Data delete Concentration unit Fig 1.1 Configuration of the instrument 6
8 2. Start and the cautions before the start of the instrument 2.1 Confirm if the environment is suitable for the instrument. 2.2 Before the power cord is connected to the wall outlet, confirm if the position of the voltage change switch suits the supply voltage. And confirm if the supply voltage is normal and the earth terminal is grounded before the use. 2.3 Before the start, make sure if nothing is in the sample compartment and the sample folder is fixed firmly. 2.4 Warm up of the instrument and the initial check-up To ensure accuracy of the measurement, always warm-up the instrument for at least 30 minutes, prior to the operation If it is the first time for the instrument to be used, inspect the stability at %T or Abs range after the warm up. Generally speaking, if the display value fluctuates just like 99.9(0.001),100.0(0.000),100.1(-0.001) at %T(Abs) range, or the least significant digit fluctuates continuously, the condition of the instrument is normal.(it shows the real value because the sensitivity is high.) If the display value fluctuates significantly, consult your supplier When the instrument is not in use for a long time, take the warm up time longer than usual, and observe the stability as new one before the use Prepare some square cuvettes before the test. Clean the inside and outside of the square cuvettes in advance. 7
9 3. Start and check 3.1 Start of the instrument After the power cord is connected, power switch is on and the instrument is started, APEL is displayed on the LCD display. 3.2 System check Few seconds after APEL is displayed on the display, self check is started. The order of the check is FILTER, LAMP CHANGE, SENSOR, D2 LAMP, HALOGEN LAMP, WAVELENGTH CALIBRATION SYSTEM PARAMETER AND DARK CURRENT. Fig. 3.2 Self check System check error When any error is detected, the error alarm sounds, and the error item is displayed at the same time. The check will resume when any key is pressed. Example:The check can not be continued and x is displayed because the dark current is too strong and goes beyond the limit. Press any key to resume the auto-check(see Fig.3.3) Fig. 3.3 System check Dark Current Error (Example) When check error is detected, consult your supplier. 8
10 3.3 System warm up The instrument will remain stable about 20 minutes after the start. And the environment of the halogen lamp will remain stable. After the system check, warm up starts for about 20 minutes. After that dark current test starts the warm up can be stopped by pressing any key. (See Fig. 3.4) Fig. 3.4 System warm up 3.4 Main menu After the warm up, main menu is displayed automatically.(see Fig. 3.5) Fig. 3.5 Main menu Photometry, Quantitative analysis, Qualitative analysis and System setting are displayed as the functions.(see Fig. 3.5) Select one item using up down keys. When is pressed the screen is changed to the selected item. 9
11 4. Photometric measurement Absorbance and/or Transmittance at fixed wavelength are measured and displayed or printed out. 4.1 Main menu of the photometry Select Photometry by up down keys in the main menu. When is pressed, the screen is changed to the main screen of Photometry.(See Fig. 4.1) Fig. 4.1 Photometry screen 4.2 Set the measuring mode When SET is pressed in the main screen of Photometry, the screen is changed to measuring mode selection screen.(see Fig.4.2) SET Fig.4.2 Measuring mode selection screen Select measuring modes (Absorbance, Transmittance, Energy) by up down keys. When item)when is pressed the item is selected.(check mark is shown at the right side of each ESC is pressed, the items are selected and the screen is returned to the main screen of the photometry. Note:When Energy is selected and is pressed in the mode selection screen, the screen is changed to Energy screen. The different gain multiplier is selected by up down keys corresponding energy is shown.(see Fig. 4.3) Fig. 4.3 Energy screen 10
12 4.3 Setting wavelength When GO TO is pressed in the main screen of photometry or measuring screen, the screen is changed to setting screen(see Fig.4.4) Fig. 4.4 Setting Wavelength When the value of the wavelength is input and screen is returned to the main screen of photometry. is pressed, the data is valid and the Note:The valid wavelength range is from 190 to 1100nm, and the data outside of this range is invalid. When the data is invalid, the screen returns to the main screen of photometry after the error alarm sounds. If the input data is wrong, the data can be deleted by pressing CLEAR A/100.0%T(Insert Blank sample to the sample holder) When ZERO is pressed in the main screen of photometry or measuring screen (with blank sample inserted), screen shows 0.000Abs. at Abs mode, or 100.0%T at %T mode Note : Make sure that the blank sample is on the optical path before 0.000Abs/100.0%T is shown. Otherwise the result may be wrong. Please be sure to use quartz square cuvette, if used wavelength is less the 340nm. 4.5 Measurement After 0.000Abs or 100.0%T is shown in the main screen of photometry or measuring screen press and press 4.5) START and move to measuring screen. Make sure the sample is on the optical path,, then the absorbance and the transmittance can be measured.(see Fig. 11
13 START Fig. 4.5 Measured results Note:Only 5 lines of the data are displayed on the screen, and the other data are shown by up and down keys. 4.6 Delete data Press in the measured results screen and select Yes and press, then the measured data is deleted.(memory capacity is 200data max.) If you want to quit deleting data, select No and press E S C CLEAR, or press, then the screen returns to the measuring screen.(see Fig. 4.6) CLEAR Fig. 4.6 Data delete screen 12
14 4.7 Data print out Press PRINT when the measured results are displayed, and select Yes and press, then the printer starts. At the same time when the print is finished, the data are deleted automatically. If you want to quit printing data, select No and press, or E S C, then the screen returns to the measuring screen.(see Fig.4.7) PRINT Fig. 4.7 Data print out screen 13
15 5. Quantitative analysis Concentration of any sample can be measured with the standard curve obtained from standard samples. If the slope and the Y-intercept of the standard straight line are known, the concentration of sample can be measured by coefficient method. 5.1 Main screen of Quantitative analysis Select Quantitative analysis by up down keys in the main menu, and press, then the screen is changed to Quantitative analysis screen.(see Fig. 5.1) Fig. 5.1 Main screen of quantitative analysis 5.2 Standard curve method Standard curve method is a quantitative analysis to know the unknown concentration of any sample with a standard curve obtained from standard samples. This section explains how to make standard curve, how to measure the concentration of samples by using the standard curve, and loading and deleting of stored standard curve Functions of standard curve method Standard curve method Draw new curve Load curve Delete curve Input sample number Select curve Select curve Input sample concentration Display message of curve Set delete curve Draw curve Record and measurement of data Delete curve Display message of the curve Record and measurement of data Wavelength setting Wavelength setting Data print out Data delete Data print out Data delete Fig. 5.2 Functions of the instrument in standard curve method 14
16 5.2.2 Standard curve method screen Select Standard curve method in the Quantitative Analysis screen by up down keys, and press, then enter into the main screen of Standard curve method.(see Fig.5.3) Fig.5.3 Standard curve method Drawing new curve Select Drawing new curve in the main screen of Standard curve method by up down keys, and press, then enter into Drawing new curve procedure. Obey the instruction shown on the screen.(see 5.4) START Fig. 5.4 Procedure of drawing new curve Procedure1: Select Drawing new curve and press number screen. Fill the number of standard samples(see Fig 5.5), press next data screen. Fill the number of data(see Fig. 5.6) and press screen., then enter into Standard sample, and move to the, then move to the next Fig. 5.5 Sample number screen Fig.5.6 Data number screen Note:The range of the number of standard curve is from 1 to 9. Outside of this range is invalid. 15
17 Procedure2 :Set the concentration of the first standard sample. Insert the blank liquid in the optical path, press ZERO first standard sample in the optical path and press sample test screen.(see Fig. 5.7 right) Press screen., set 100%T or 0.000Abs. Fill the concentration of the, then enter into the again and move to the next Fig. 5.7 Setting the concentration of the first standard sample Note:The range of the concentration is from 0 to The outside of this range is invalid. The concentration of the samples to be measured are supposed to be increasing sequentially from the first one to the last one.(the concentration of the first sample must be the lowest.) Procedure 3: Set the second sample, fill the concentration, and press, (the instrument measures the absorbance or transmittance of the standard sample automatically.) Then enter into the next screen.(see Fig. 5.8) Fig. 5.8 Setting the concentration of the second standard sample Note: After setting all the samples and filling all the concentration of the samples according to the designated number of samples and data, the standard curve is drawn automatically. Procedure 4: When all the above procedures are completed, the standard curve is drawn automatically. Press START to enter into the measuring concentration screen to measure the concentration of unknown sample by the standard curve. (See Fig.5.9) 16
18 Fig. 5.9 Drawing standard curve Note: If data error ( operation error or solution error) occurs, the standard curve can not be drawn. And the screen returns to the main screen of standard curve method automatically. Procedure 5:The concentration is measured by the standard curve. After 0.000Abs. or 100%T is set by the blank liquid in the optical path, any sample is set in the optical path. Press Press START PRINT, then the concentration is read and displayed.(see Fig.5.10), then enter into the print selection screen, select Yes, press, then test data and the standard curve are printed out. If No is selected and Fig. 5.11) is pressed, the screen returns to the previous menu.(see Fig Measuring concentration Fig Print selection Setting wavelength Press GO TO in the sample number screen of standard curve method or measuring screen, set the necessary wavelength.(see Fig.5.12) Fig.5.12 Wavelength setting screen Note: Fill the wavelength value in the space under the screen by the numerical keys. Press, then the input is completed and the screen returns to the previous one. Wavelength range of PD-3000UV is from 190 to 1100nm, and outside of the range is invalid. Press CLEAR, and delete the data and try the input again. 17
19 5.2.5 Data delete Press CLEAR Yes and press Fig. 5.13) in the standard curve method, then enter into the data delete screen. Select, all the data on the standard curve method screen are deleted. (See Fig Data delete screen Curve data load Saved standard curve can be loaded from the memory of this instrument.(newly created curve is saved automatically.) When it is necessary, select curve data load by up down keys in the main menu of standard curve method, enter drawing curve selection screen.(see Fig.5.14), then the saved curve data are displayed in the Fig Curve selection screen The newest curve is recorded at the end of the list automatically.(capacity: 200 data max) elect necessary equation, by up down keys in the curve selection screen and press Press START, then the selected curve is displayed.(see Fig.5.15), and the concentration of any sample can be measured by the selected curve. START Fig Selected curve and the measured results 18
20 5.2.7 Deleting curves Saved standard curve can be deleted. To delete a certain curve, select Deleting curves by up down keys in the standard curve method screen and press screen., then the screen is changed to Deleting curve Press after a equation is selected, select Yes in the next screen and press, then the equation is deleted. (After the deletion, the screen returns to the previous screen.) Select No in the deleting curve screen and press the screen returns to the previous one.( screen.)(see Fig.5.16) E S C and the deletion is canceled and can be also used to return to the previous Fig Deleting curves 19
21 6. Coefficient method Coefficient method is a simple application of standard curve method. Define the equation and the curve by filling each coefficients of the equation and measure the concentration of any sample by the curve. 6.1 Calculation formula Because the calculation formula of the coefficient method is defined as C=K A+B in the system, parameters K and B are to be filled. 6.2 Coefficient method screen Select Coefficient method by up down keys screen and press in the Quantitative analysis, then move to the coefficient method screen.(see Fig.6.1) Fig. 6.1 Main screen of coefficient method 6.3 Setting parameters screen Set the parameters K and B before the measurement. Press S E T in the coefficient method screen, and the screen is changed to parameter setting screen of coefficient method curve.(see Fig.6.2) Fig.6.2 Parameter setting screen of coefficient method Setting parameter K Select Setting parameter K by up down keys in the coefficient method parameter setting screen and press, then enter into the screen of setting parameter K. After the value of K is filled by numerical keys, press parameter setting screen.(see Fig.6.3), the screen returns to the Fig.6.3 Setting parameter K Note:The range of parameter K is from to
22 6.3.2 Setting parameter B Select Setting parameter B by up down keys in the coefficient method parameter setting screen and press, then enter into the screen of setting parameter B. After the value of B is filled by the numerical keys, press parameter setting screen.(see Fig.6.4), the screen returns to the Fig.6.4 Setting parameter B After filling the parameters K and B, the screen returns to the main screen of the coefficient method by pressing E S C. Note:The range of parameter B is from to Setting wavelength Press GO TO in the coefficient method screen and enter into the wavelength setting screen, then set the wavelength. After the value of the wavelength is filled by the numerical keys, press screen returns to the previous one.(see Fig.6.5), and the GO TO Fig.6.5 Wavelength setting screen Note:The range of the wavelength is from 190 to Auto zero calibration (Insert blank liquid to cuvette holder) Put the blank liquid in the optical path and press method, the display shows 0.000Abs or 100.0%T at the used wavelength. ZERO in the main screen of coefficient 21
23 6.6 Measurement, data output and printout Press START in the coefficient method screen and enter into the measurement screen. After the sample is set in the optical path, press measured(see Fig.6.6) and the concentration of the sample is If printout of the result is necessary in the coefficient method screen, please read 4.7. START START START Fig.6.6 Coefficient method screen 7.Selection of concentration unit Select Concentration unit by up down keys screen and press Fig.7.1) in the Quantitative analysis, then enter into the Concentration unit selection screen.(see Fig. 7.1 Concentration unit selection screen Select a concentration unit by up down keys in the concentration unit selection screen.(or input the number with numerical key and press.) When E S C quantitative analysis. is pressed, the setting is completed and returns to the main screen of 22
24 8. System setting This function controls the switches of halogen and D2 lamp, and arrange time, wavelength calibration, dark current, setting of lamp switching wavelength, initialization and version check.(see fig.8.1) System setting ON/OFF of D2 lamp ON/OFF of Halogen lamp Date setting Dark current cal. Wavelength cal. Setting Lamp switching wavelength Initialization Version information Fig.8.1 System setting function Select System setting by up down keys in the main menu and press, then enter into the system setting screen.(see Fig.8.2) E S C Fig.8.2 Main screen of System setting 23
25 8.1 ON/OFF of D2 lamp Select D2 lamp on in the system setting screen and press into D2 lamp ON/OFF selection screen Select ON or OFF in this screen and press ON/OFF is changed.(see Fig.8.3), then the screen enters, then the state of the D2 lamp E S C Fig.8.3 D2 lamp ON/OFF selection screen Note:To keep the life of D2 lamp longer, it is recommended that the D2 lamp is turned off after the start while the test at the UV range is unnecessary for a long time. 8.2 ON/OFF of Halogen lamp Select Halogen lamp on in the system setting screen and press enters into halogen lamp ON/OFF selection screen Select ON or OFF in this screen and press ON/OFF is changed.(see Fig.8.4), then the screen, then the state of the halogen lamp E S C Fig.8.4 Halogen lamp ON/OFF selection screen Note:To keep the life of halogen lamp longer, it is recommended that the halogen lamp is turned off after the start while the test at the VIS range is unnecessary for a long time. 24
26 8.3 Setting Date Select Setting Date in the system screen and press Setting Date screen. (See Fig.8.5), then the screen is changed to E S C Fig.8.5 Setting Date screen Move the cursor by up down keys in the Setting Date screen. The active position goes in the order of Hour, Minute, Second, Year, Month, Date, Day of the week by the key, and it goes in the opposite direction by key. Fill the time and date by numerical keys. As for day of the week, 1 means Monday, 2 means Tuesday,., 7 means Sunday and so on. Press setting screen after the data input, then the date is fixed and the screen returns to the system 8.4 Dark current calibration Dark current should be calibrated before the measurement, because it can affects the results of sample measurement when the wavelength is changed at the wavelength setting or any other reason if the time of dark current is long. Select Dark current calibration screen in the system setting screen and press the calibration starts. If E S C is pressed as instructed on the screen after the calibration, the screen returns to the system setting screen. (See Fig.8.6), then E S C Fig.8.6 Dark current calibration screen 8.5 Wavelength calibration Use this function when the accuracy of the wavelength is investigated. Select wavelength calibration in the system setting screen and press, then the calibration starts. The wavelength is calibrated automatically with the built-in filter (656.1nm). It takes about 2 minutes.(see Fig. 8.7) 25
27 E S C Fig.8.7 Wavelength calibration screen 8.6 Lamp changing wavelength Select Lamp changing in the system setting screen and press changed to setting lamp changing.(see Fig.8.8), then the screen is E S C Fig.8.8 Lamp changing wavelength screen After the value of the wavelength is filled by the numerical key in the setting lamp changing screen, press pressing E S C., and complete input, then the screen returns to the system setting screen by Note: Lamp changing wavelength means the wavelength where the lamp is changed between halogen lamp and D2 lamp. The range of the wavelength is from 325 to 375nm. 8.7 Initialization Select Initialization in the system setting screen and press changed to initialization screen.(see Fig. 8.9), then the screen is E S C Fig.8.9 Initialization screen 26
28 Select Yes in the initialization selection screen and press, the initialization starts. If No is selected and is pressed, or E S C is pressed directly, the screen returns to the previous one. Note: If Initialization is carried out, all the data in the memory (Test record, parameters, drawn standard curves etc.) are deleted. Please pay close attention when the function is selected. 27
29 Chapter 6 Maintenance Almost no daily maintenance is necessary because PD-3000UV is designed considering durability and reliability. But because there are some parts which have life span such as light source, investigate the performance by calibrating the wavelength or linearity if necessary. And, wipe off dirt frequently, if the dirt is visible. Daily maintenance and cleaning 1) Cleaning of outer case Turn off the power switch and unplug the power cord from the wall outlet. Wipe the surface with a wet soft cloth. If there are stain, use the drop of neutral detergent. 2) Cleaning of cuvette holder Turn off the power switch and unplug the power cord from the wall outlet. Wipe the surface with a wet soft cloth. (If it is necessary to remove the cuvette holder, please consult your supplier.) Replacement of halogen lamp and/or D2 lamp When halogen lamp or D2 lamp is deteriorated, replace them according to the following procedure. 1Turn off the power switch and unplug the power cord from the wall outlet. 2Remove 4 screws of the outer case, disconnect the connectors and open the case. 3Remove the black metal plate which covers the lamp. 4Remove the used lamp from the lamp holder. 5Set new lamp correctly. Caution:Leave the lamp for more than 10 minutes after the power switch is turned off before replacing the lamps, because the lamp can be too hot immediately after the use. D2 lamp should be replaced with the socket, and the halogen lamp should be replaced with the bulb alone. Do not touch the lamp with fingers directly but wrap it with tissue paper or cloth. If the oil from the fingers attaches to the lamp, the measuring accuracy can be deteriorated or the life of the lamp can become shorter. Replacement of fuses The procedure of the replacement is as follows. 1 Turn off the power switch and unplug the power cord from the wall outlet. 2 The fuse holder comes with the power unit together. Therefore insert a flat head screwdriver into the dent and remove the fuse holder. 3 Put a new fuse (2A 5) in the fuse holder and fit it to the original position. Fuse can not be deteriorated in normal use. In case the fuse burns often, there can be any problems, please consult your supplier. Wavelength calibration Calibrate the wavelength according to procedure of above
30 Packing List CONTENTS: 1. Main body 1set 2. Glass square cuvette 1box(4) 3. Quartz square cuvette 1box(2) 4. Power cord 1pc. 5. Application software 1set (CD-ROM, USB key, USB cable, Software manual) 6. Fuse(1A) 1pc. 7. Dust cover 1pc. 8. Instruction manual 1copy 29
UVS-2800 Spectro UV-VIS Split Beam (PC) is a precise scanning
Spectro UV-VIS Split Beam (PC) is a precise scanning Spectrophotometer with a new design of 8 microprocessor automatic 2 row cell holder that moves noiseless with a special membrane. This Split Beam Spectro
More informationUV / VIS Spectrophotometer EMCLAB Instruments GmbH
A Professional Manufacturer UV / VIS Spectrophotometer EMCLAB Instruments GmbH EMCLAB Instruments GmbH Bismarckstrasse 120 47057 Duisburg Germany Phone: +49 203 3064042 Fax: +49 203 3064044 E-mail: info@emc-lab.de
More informationMANUAL BMS SPEC5000 Spectrophotometer
MANUAL BMS SPEC5000 Spectrophotometer BMS Art. Nr. 17020 1 PRINCIPLE, APPLICATION AND CHARACTERISTICS 1.1 Principle The SPEC5000 spectrophotometer is a single beam spectrophotometer with a basic structure
More informationUSER MANUAL FOR VISIBLE SPECTROPHOTOMETER
USER MANUAL FOR VISIBLE SPECTROPHOTOMETER 1 Table of Contents 1. MAIN USAGES...3 2. WORKING ENVIRONMENT...3 3. MAIN TECHNICAL DATA AND SPECIFICATIONS...4 4. WORKING PRINCIPLE...5 5. OPTICAL PRINCIPLE...6
More informationSpectro UV-VIS Double Beam Research Spectrophotometer UV-VIS Double Beam
Spectro UV-VIS Double Beam Research Spectrophotometer UV-VIS Double Beam Software Specifications Model UVD-3500 Spectro UV-VIS Double Beam UVD 3500 Research Spectrophotometer is a superior instrument for
More informationOperating Manual. Model 721N. Visible Spectrophotometer
Operating Manual of Model 721N Visible Spectrophotometer 1 Table of Contents 1. Chief uses... 3 2. Working Conditions... 3 3. Main Specifications...3 4.Operating Principles...4 5. Optical design...4 6.
More informationUV / VIS Spectrophotometer
UV / VIS Spectrophotometer Single Beam Double Beam NANO EMCLAB Instruments GmbH EMCLAB Instruments GmbH Kulturstrasse 55 47055 Duisburg Germany Fon: +49 203 3064042 Fax: +49 203 41504809 E-mail: info@emc-lab.de
More informationUSB. Part No Wavelength range. Spectral bandwidth 5 nm 4 nm Optical system
Visible range spectrophotometers V-1100 and VR-2000 V-1100 MODEL WITH MANUAL WAVELENGTH SETTINGS AND AUTOMATIC BLANK. VR-2000 MODEL WITH AUTOMATIC WAVELENGTH SETTINGS AND BLANK. V-1100 Part no. 4120025
More informationT92+ UV-VIS SPECTROPHOTOMETER
T92+ UV-VIS SPECTROPHOTOMETER The T92+ is a high performance double beam spectrophotometer with a variable spectral bandwidth from 0.1-5nm, selected by a continuous variable slit. The instrument is fitted
More informationCroma Enterprise Cromtech India
UV 1601 - Split Beam & Dual (equivalent to Double beam Spectro Photometer ) Wide wavelength range, satisfying requirements of various fields. The split-beam ratio monitoring system provides accurate measurements
More informationPage 1 BIOIMAGER. Spectrophotometers. Toll Free: (855) BIO- IMAG / (855) , E: S:
Page 1 Comparison of all models Page 2 Model BK- V1000 BK- UV1000 BK- V1200 BK- UV1200 BK- V1600 BK- UV1600 BK- V1800 BK- UV1800 BK- V1900 BK- UV1900 BK- S360 BK- S380 BK- S390 BK- D560 BK- D580 BK- D590
More informationSpectrophotometer. Operational Manual
Spectrophotometer Operational Manual V1.0 Revision 1 for Alpha-1860/1900 Laxco, Inc. http://www.laxcoinc.com 1 Contents General information... 3 Safety... 3 Electrical... 3 Warning... 3 Performance...
More informationinstruments Analytical Instruments for Science
instruments Analytical Instruments for Science instruments Contents PAGE NO. Introduction 4 T60 UV-Vis Spectrophotometer 6 T70 UV-Vis Spectrophotometer 10 T80 UV-Vis Spectrophotometer 14 T90+ UV-Vis Spectrophotometer
More informationSP-8001 UV/Visible Spectrophotometer
SP-8001 UV/Visible Spectrophotometer Provide Accurate Results High performance UV/Visible Spectrophotometer for Research, and Laboratory work Product Features The Metertech SP-8001 UV/Visible Spectrophotometer
More informationModel S-520 Coin Counter / Sorter Operating Manual
Model S-520 Coin Counter / Sorter Operating Manual Table of Contents Using the Product Safely... 2 About the Warning Stickers and Points of Warning/... 2 Warning.... 3. 4 Names of the Main Parts......
More informationUSER MANUAL ENGLISH 1450 COIN COUNTER & SORTER
USER MANUAL ENGLISH 1450 COIN COUNTER & SORTER INTRODUCTION ENGLISH Thank you for purchasing the Safescan 1450 coin counter and sorter. For proper use and maintenance, we advise to read this user manual
More informationUser Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com
User Manual Digital Compound Binocular LED Microscope Model MD82ES10 MicroscopeNet.com Table of Contents i. Caution... 1 ii. Care and Maintenance... 2 1. Components Illustration... 3 2. Installation...
More informationUser Manual. Trinocular Metallurgical Microscope. MicroscopeNet.com
User Manual Trinocular Metallurgical Microscope Model M83MPTR MicroscopeNet.com Table of Contents i. Caution.. 1 ii. Care and Maintenance... 2 1. Components Illustration..... 3 2. Installation...4 3. Operation
More informationSPECTRONIC Standards User Guide
SPECTRONIC Standards User Guide The information in this publication is provided for reference only. All information contained in this publication is believed to be correct and complete. Thermo Fisher Scientific
More informationTherefore, all descriptions and illustrations in this instruction manual, including all specifications are subject to change without notice.
We are constantly endeavouring to improve our instruments and to adapt them to the requirements of modern research techniques and testing methods. This involves modification to the mechanical structure
More informationUSER MANUAL ENGLISH 2985-SX BANKNOTE COUNTER
USER MANUAL ENGLISH 2985-SX BANKNOTE COUNTER INTRODUCTION ENGLISH Thank you for purchasing Safescan 2985-SX banknote counter. For proper use and maintenance, we advise to read this user manual carefully.
More informationAutomated Double Aperture Accessory
For the Cary 1, 3, 100, 300, 4, 5, 400, 500, 500i, 4000, 5000, 6000i, Deep UV Installation Category II Pollution Degree 2 Equipment Class I Table of Contents Introduction Theory Operation Installation
More informationContents I. APPLICATIONS II. SPECIFICATIONS III. PRINCIPLES IV OPERATION V. MAINTENANCE
Contents I. APPLICATIONS...- 1 - II. SPECIFICATIONS...- 2 - III. PRINCIPLES...- 3 - IV OPERATION...- 9 - V. MAINTENANCE...- 21 - VI SUPPLEMENT...- 23 - Warning: The responsible person must be clear that
More informationTABLE OF CONTENTS. Safety notes i. Care and Maintenance. ii. 1. Components Illustration Installation of Components.. 4
TABLE OF CONTENTS Safety notes i Care and Maintenance. ii 1. Components Illustration... 1 2. Installation of Components.. 4 2.1 Installation Diagram... 4 2.2 Installation Procedures 5 3. Operation...11
More informationCONTENTS Product Accessory Software V /02 V-5100/UV /04 UV-5100B 05/06
CONTENTS Product V-5000 01/02 V-5100/UV-5100 03/04 UV-5100B 05/06 V-5600/UV-5600/V-5800/UV-5800 07/08 UV-6000 09/10 UV-6100/6100A/6100S 11/12 UV-9000/UV-9000A/9000S 13/14 B-500 15 TOC-2000/TOC-3000 16/17
More informationSarspec, Lda. - Rua Camilo Castelo Branco, 965 PQ Vila Nova de Gaia Phone:
2 3 IMPORTANT SAFETY NOTE: Before operating this device, please read carefully this User Manual and be familiar with its contents prior to using this equipment. To help avoid potential serious injury to
More informationThermo Scientific SPECTRONIC 200 Visible Spectrophotometer. The perfect tool. for routine measurements
Thermo Scientific SPECTRONIC 200 Visible Spectrophotometer The perfect tool for routine measurements The Standard for Routine Measurements Thermo Scientific SPECTRONIC spectrophotometers have served as
More informationNuclear Associates
Nuclear Associates 07-424 Digital Densitometer II Operators Manual March 2005 Manual No. 112111 Rev. 4 2003, 2005 Fluke Corporation, All rights reserved. Printed U.S.A. All product names are trademarks
More informationVWR SPECTROPHOTOMETERS
VWR SPECTROPHOTOMETERS Reliable Accurate Easy to Use Your first choice for spectrophotometry VWR SPECTROPHOTOMETERS Quality Guaranteed VWR SPECTROPHOTOMETERS Single Beam Units VWR offers a choice of 5 models
More informationThermo Scientific SPECTRONIC 200 Education
molecular spectroscopy Thermo Scientific SPECTRONIC 200 Education Part of Thermo Fisher Scientific Designed for the Teaching Laboratory Classroom Friendly Sample Compartment Whether you measure in 10 mm
More informationUser's Guide. February 2012 OPTIMA INC
User's Guide February 202 OPTIMA INC Notice! Spectrophotometer SP-3000 DB is a registered product of OPTIMA INC. No other relevant materials on our products including this publication may be reproduced,
More informationibox Spectra Small Animal Imaging System
ibox Spectra Small Animal Imaging System Installation and User Instructions UVP, LLC Ultra-Violet Products Ltd. 2066 W. 11th Street Unit 1, Trinity Hall Farm Estate Upland, CA 91786 Nuffield Road, Cambridge
More information2012 Monitored Rehab Systems E1201 Manual Kneelax. Installation and user manual
2012 Monitored Rehab Systems E1201 Manual Kneelax Installation and user manual Table of contents CHAPTER 1 Installation Manual... 3 1.1 Install Kneelax USB driver... 3 1.2 Detect COM-port... 6 1.3 Software
More informationSPECTROPHOTOMETER CM-3600A INSTRUCTION MANUAL
SPECTROPHOTOMETER CM-3600A INSTRUCTION MANUAL Before using this instrument, please read this manual. Safety Symbols The following symbols are used in this manual or CM-3600A to prevent accidents which
More informationThermo Scientific SPECTRONIC 200
molecular spectroscopy Thermo Scientific SPECTRONIC 200 Part of Thermo Fisher Scientific The New Standard for Routine Measurements Robust, Multifunction Sample Compartment Whether you measure in 10 mm
More informationTECHNICAL INSTRUCTIONS FOR DataTrace CALIBRATION PROCEDURES
TECHNICAL INSTRUCTIONS FOR DataTrace CALIBRATION PROCEDURES HUMIDITY CALIBRATION The following procedures provide instructions for performing field calibrations on DataTrace equipment. There are two types
More informationUV-2600/2700 INSTRUCTION MANUAL
206-97440A Aug. 20 Shimadzu UV-Visible Spectrophotometer UV-2600/2700 INSTRUCTION MANUAL Read the instruction manual thoroughly before you use the product. Keep this instruction manual for future reference.
More informationSPECTROPHOTOMETER CM-3700A INSTRUCTION MANUAL
SPECTROPHOTOMETER CM-3700A INSTRUCTION MANUAL Before using this instrument, please read this manual. Safety Symbols The following symbols are used in this manual or CM-3700A to prevent accidents which
More informationOperation Instructions
Operation Instructions 0 Content 1. General Information -----------------------------------------------------------------------------2 2. Appearance and parts name --------------------------------------------------------
More informationProHead. User s Guide
ProHead User s Guide 2Profoto ProHead Profoto ProHead Thank you for choosing Profoto. Thanks for showing us your confidence by investing in a ProHead unit. For more than four decades we have sought the
More informationF100 Tabletop Document Folder
F100 Tabletop Document Folder 10/2018 OPERATOR MANUAL FIRST EDITION Single Fold Letter Fold Z-fold Double Fold Specifications Paper size 11" & 14" Paper weight 60g/m 2 to 90g/m 2 Some recycled paper and
More informationThermo Scientific SPECTRONIC 200 Visible Spectrophotometer. The perfect. teaching instrument
Thermo Scientific SPECTRONIC 200 Visible Spectrophotometer The perfect teaching instrument Designed for the Teaching Laboratory Thermo Scientific SPECTRONIC spectrophotometers have served as core analytical
More informationCleaning the cutter blade
Cleaning the cutter blade < Daily maintenance > In order to use this printer in good condition, please perform the daily maintenance. Note the following when performing the maintenance. Do not lubricate
More informationAMM-1022 Digital Multimeter USER`S MANUAL
Digital Multimeter USER`S MANUAL www.tmatlantic.com CONTENTS 1. SAFETY INFORMATION.3 2. DESCRIPTION..6 3. SPECIFICATIONS.8 4. OPERATING INSTRUCTION..11 4.1 Voltage measurement...11 4.2 Current measurement
More informationUser guide ProHead Plus
User guide ProHead Plus For other languages visit: /support ProHead Plus 2 Congratulations on your new Profoto product! Thanks for showing us your confidence by investing in a ProHead unit. For more than
More informationInstalling the detector filter assembly (G )
s1 Installing the detector filter assembly (G1600-62700) The detector filter transmits light of a defined wavelength range. Figure 1 shows a typical transmission spectrum. Maximum transmission is between
More informationLab Junior Educational UV-VIS Spectrometer
www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment Lab Junior Educational UV-VIS Spectrometer K-MAC Lab Junior was developed
More informationPhotometer Maintenance Guide
Photometer Maintenance Guide Product Support Information This guide is designed to help ensure your instrument remains in a good condition and continues to deliver accurate results. It includes a series
More informationBIO SPECTROPHOTOMETER
BIO SPECTROPHOTOMETER Nano 3060UV Bio 3060-UV spectrophotometer, equipped with 6 inches LCD display, is an ideal and advanced analytical instrument for laboratory to realize wevelength scanning, kinetics
More informationElectrical Safety Tester
Electrical Safety Tester GPT-9600 Series QUICK START GUIDE GW INSTEK PART NO. 82PT-96030MA1 ISO-9001 CERTIFIED MANUFACTURER This manual contains proprietary information, which is protected by copyright.
More informationRS3040A Mini Laser Engraving Machine User Manual
RS3040A Mini Laser Engraving Machine User Manual Thank you very much for buying RS3040A laser engraving machine. In order to make sure that your laser engraving machine can work steadily for a long period
More informationContent Spectrophotometers
Spectroph Content Spectrophotometers Selection table Spectrophotometers Page 163 PRIM Page 164 UviLine Page 166 162 Selection table Spectrophotometers PRIM Light/ PRIM Advanced UviLine 9100/ UviLine 9400
More informationfi-6130 / fi-6230 Cleaning and Maintenance
fi-6130 / fi-6230 Cleaning and Maintenance fi-6130 / fi-6230 Cleaning and Maintenance fi-6130 / fi-6230 Image Scanner Operator's Guide Chapter 4 DAILY CARE This chapter describes how to clean the scanner.
More informationOtO Photonics. Light Source Datasheet. Description. Balance Light Series LS-BA. Halogen Light Series LS-HA
OtO Photonics Light Source Datasheet Description OtO Photonics provide many kinds of light sources to meet customer's need. Our light sources offer the stable and wide-band spectral output and proper for
More informationNano-100 Spectrophotometer. Brief introduction
Nano-100 Spectrophotometer Brief introduction Direct and quick measure of DNA, RNA, cell solution concentration Only need volume 0.5 to 2 µl No need cuvette or capillary tube Wavelength range 200-800 nm
More informationInstallation Instructions SKU# , Rated Voltage V~ 50Hz
Installation Instructions SKU# 240621, 240622 Rated Voltage 220-240V~ 50Hz THANK YOU FOR PURCHASING THIS QUALITY LUCCI PRODUCT, TO ENSURE CORRECT FUNCTION AND SAFETY, PLEASE READ AND SAVE ALL INSTRUCTIONS
More informationOptical Transmitter RF-T860/24. Installation & Operation Manual. Caution
Optical Transmitter RF-T860/24 Installation & Operation Manual Caution These servicing instructions are for use by qualified personnel only. To reduce the risk of electrical shock, please do not perform
More informationMulti-Channel In-Out Thermometer with Cable Free Sensor and RF Clock
Multi-Channel In-Out Thermometer with Cable Free Sensor and RF Clock MAIN FEATURES: MAIN UNIT GB MODEL: RMR182 USER'S MANUAL INTRODUCTION Congratulations on your purchase of the RMR182 Multi- Channel In-Out
More informationQuolis 5000 Light INSTALLATION AND OPERATION INSTRUCTIONS. Model AL-820 IMPORTANT
Quolis 5000 Light Model AL-80 INSTALLATION AND OPERATION INSTRUCTIONS IMPORTANT After installation is completed, check all the bolts, screws and fasteners to confirm that they are securely fastened. R
More informationOperation Manual. Chap. 1 Attention. Please read the attention carefully before operating the machine.
Chap. 1 Attention Please read the attention carefully before operating the machine. 1. No magnetic devices should be placed in the vicinity of the plotter, specifically the carriage. 2. Prevent from dropping
More informationOVA-50 Optical Variable Attenuator User s Manual
OVA-50 Optical Variable Attenuator User s Manual Shineway Technologies, Inc. All rights reserved. Safety Instructions The WARNING sign denotes a hazard. It calls attention to a procedure, practice, or
More informationPart/Reorder Number: Version 1.0
Part/Reorder Number: 870000 Version 1.0 Cross Match Technologies L SCAN 100/100R Operator s Manual Version 1.0 First Edition (August 2006) No portion of this guide may be reproduced in any form or by any
More informationPage 1 T O O L S A N D M AT E R I A L S R E Q U I R E D. 1. Screwdriver 2. Carpenter's level 3. Electrical Tape PA R T S L I S T. A.
T O O L S A N D M AT E R I A L S R E Q U I R E D 1. Screwdriver 2. Carpenter's level 3. Electrical Tape PA R T S L I S T A. Cabinet B. Wood Screw X4 C. Plastic wire connector x3 Page 1 I M P O R TA N T
More informationUser instructions Compound laboratory microscope
KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen E-mail: info@kern-sohn.com User instructions Compound laboratory microscope Tel: +49-[0]7433-9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com KERN
More informationGeneral Help. Last revised: Winter When I try to print something on the computer, it appears to work, but nothing comes out of the printer.
General Help Last revised: Winter 2015 Problem Solution When I try to print something on the computer, it appears to work, but nothing comes out of the printer. See the next item. When I try to print something
More informationUSER MANUAL PV 360. Vidifox Document Camera
Vidifox Document Camera PV 360 USER MANUAL Please read this User Manual thoroughly before you use the document camera. Keep the CD-ROM in a convenient place so you can use it quickly if you need to. Please
More informationMS6231 DIGITAL ENGINE ANALYZER OPERATOR S MANUAL CONTENTS CONTENTS
CONTENTS MS6231 DIGITAL ENGINE ANALYZER OPERATOR S MANUAL CONTENTS 1. SAFETY INFORMATION 1 1.1 PRELIMINARY 2 1.2 DURING USE 3 1.3 SYMBOLS 5 1.4 MAINTENANCE 6 2. DESCRIPTION 8 2.1 NAMES OF COMPONENTS 9
More informationUser Manual. Trinocular Infinity Compound LED Microscope. MicroscopeNet.com
User Manual Trinocular Infinity Compound LED Microscope Model M8333Z series MicroscopeNet.com Table of Contents i. Caution... 1 ii. Care and Maintenance... 2 1. Components Illustration... 3 2. Installation...
More informationRabbit Series Laser Engraving Machine HX3040. User s Manual. 1 Jinan King Rabbit Technology Development Co. Ltd.
Rabbit Laser Engraving Machine HX3040 User s Manual 1 Preface Thank you for buying our Rabbit Series HX3040 Laser Engraving Machine. You will find it has many nice features including easy operation, high
More informationVernier SpectroVis Plus Spectrophotometer (Order Code: SVIS-PL)
Vernier SpectroVis Plus Spectrophotometer (Order Code: SVIS-PL) SpectroVis Plus is a portable, visible to near-ir spectrophotometer and fluorometer. What is included with the SpectroVis Plus? One SpectroVis
More informationGLASS GLOBE MOBILE LINEAR CHANDELIER
PARTS ENCLOSED: (B) Top ceiling plate [1] (C) Bottom ceiling plate [1] (D) Glass globes [7] (E) Screws [6] (F) Nuts [5] (G) Extension rods [15] (H) Transformer [1] (I) Threaded inserts [5] (J) 1.6W LED
More informationHigh Intensity LED Stroboscope Digital Tachometer DT-361/365. Instruction manual. Be sure to read before use.
98585A High Intensity LED Stroboscope Digital Tachometer DT-361/365 Instruction manual Be sure to read before use. Before use, please carefully read these safety precautions as well as instructions, and
More informationMulti Water Quality Checker U-50 Series Instruction Manual CODE:GZ
Multi Water Quality Checker U-50 Series Instruction Manual CODE:GZ0000144342 Preface This manual describes the operation of the Multi Water Quality Checker, U-50 Series. Be sure to read this manual before
More informationMeasurement Method of High Absorbance (Low Transmittance) Samples by UH4150 INTRODUCTION
INTRODUCTION With UH4150, a detector can be selected depending on the analysis purpose. When analyzing a solid sample which doesn t contain any diffuse components, by selecting the direct light detector,
More informationOMM300. Inverted Metallurgical Microscope
OMM300 Inverted Metallurgical Microscope Instruction Manual Please read the instructions carefully before operating CONTENTS Safety 2 Parts List 2 Features 3 Assembly 5 Operation 7 Maintenance 9 Specifications
More informationSpectrophotometer CM-2500c. Instruction Manual
Spectrophotometer CM-2500c E Instruction Manual Safety Symbols The following symbols are used in this manual to prevent accidents which may occur as result of incorrect use of the instrument. Denotes a
More informationRead and be familiar with this manual before operating or servicing this device. To ensure operator, technician, and patient safety, use only as
Read and be familiar with this manual before operating or servicing this device. To ensure operator, technician, and patient safety, use only as specified in this manual. Product Manufacturer: Medical
More informationCD770 DIGITAL MULTIMETER INSTRUCTION MANUAL
CD770 DIGITAL MULTIMETER INSTRUCTION MANUAL Table of Contents 1 SAFETY PRECAUTIONS Before use, read the following safety precautions.- 1-1 Explanation of Warning Symbols 001 1-2 Warning Messages for Safe
More informationCashConcepts CCE 4000
CashConcepts CCE 4000 Coin Counter and Sorting Machine User Manual Introduction Thank you for choosing the CCE 4000. This coin counting and coin sorting machine is designed to simplify the daily cash or
More information5. Carefully remove the printer from the lower boxed foam support and place it on a solid, level base where it will be used
PROJET 1200 QUICKSTART GUIDE Before you get started you will need: Lint-free paper towels A pair of nitrile gloves Saftety glasses OPENING YOUR PROJET 1200 NOTE: Make sure you save all of your packaging
More informationUser Manual. For MAC
User Manual For MAC A3 Scanner with The TWAIN Driver (For Macintosh) English User Manual Contents Preface Chapter One Setting Up the Scanner 1 System Requirements... 1 Unpacking the Scanner... 2 The Scanner
More informationIMPORTANT SAFETY INSTRUCTIONS: -We strongly recommend that a professional electrician install all direct wire fixtures.
PRODUCT NAME: Mimi Quatrefoil Ceiling Mount ITEM NUMBER: LM199 BRA 1016 DIMENSIONS OF ITEM ASSEMBLED: 10 1/2 H x 15 1/2 DIA IMPORTANT SAFETY INSTRUCTIONS: -We strongly recommend that a professional electrician
More informationML7520 ML7530 DIOPTER ADJUSTMENT RING BINOCULAR BODY, INCLINED 30. (a) Field Iris Control Lever. (c) Filter Slots EYEPIECES, KHW10X
JAPAN DIOPTER ADJUSTMENT RING BINOCULAR BODY, INCLINED 30 (a) Field Iris Control Lever (c) Filter Slots EYEPIECES, KHW10X ANALYZER CONTROL LEVER (b) Aperture Iris Control Lever LIGHT SOURCE HOUSING VERTICAL
More informationPROGRAMMER INSTRUCTION MANUAL Please read this manual before using the machine. Please keep this manual within easy reach for quick reference.
PROGRAMMER INSTRUCTION MANUAL Please read this manual before using the machine. Please keep this manual within easy reach for quick reference. BAS-761 Thank you very much for purchasing the Brother Industrial
More informationGE Infrastructure Sensing. Druck DPI 841/842. Frequency calibrator and Frequency loop calibrator User manual - K395
GE Infrastructure Sensing Druck DPI 841/842 Frequency calibrator and Frequency loop calibrator User manual - K395 A1 B1 10 1 A 2 9 A 3 8 2 3 7 6 11 4 5 B1 12 A2 DPI 842 13 15 14 A3 19 18 17 16 27 20 21
More informationTechnical Notes. Integrating Sphere Measurement Part II: Calibration. Introduction. Calibration
Technical Notes Integrating Sphere Measurement Part II: Calibration This Technical Note is Part II in a three part series examining the proper maintenance and use of integrating sphere light measurement
More informationArched Top Lantern Pendant Assembly and Installation Instructions. Country of Destination: US/CN UK/EU/AUS Middle East
CAUTION: Arched Top Lantern Pendant Assembly and Installation Instructions Country of Destination: US/CN UK/EU/AUS Middle East BEFORE INSTALLING FIXTURE, MAKE SURE THE POWER TO THE CIRCUIT IS TURNED OFF
More informationUser instructions Metallurgical inverted microscope
KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen E-mail: info@kern-sohn.com Tel: +49-[0]7433-9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com User instructions Metallurgical inverted microscope
More informationAMANO. Electronic Calculating Time Recorder. Setup and Operations Manual
AMANO MRX-35 Electronic Calculating Time Recorder Setup and Operations Manual Proprietary Notice This document contains proprietary information and may not be reproduced in whole or in part without written
More information1. SAFETY INFORMATION.1 2. DESCRIPTION SPECIFICATIONS.6 4. OPERATING INSTRUCTION Voltage measurement Current measurement 10
CONTENTS 1. SAFETY INFORMATION.1 2. DESCRIPTION..4 3. SPECIFICATIONS.6 4. OPERATING INSTRUCTION..9 4.1 Voltage measurement...10 4.2 Current measurement 10 4.3 Resistance measurement...12 4.4 Diode test.12
More informationOperation manual. Thanks very much for purchasing this. cutter/plotter. To ensure you make best use of your machine,
Operation manual Thanks very much for purchasing this cutter/plotter. To ensure you make best use of your machine, please read this manual carefully and thorughly beforhand. IMMEDIATELY power off the machine
More informationPhysics 4C Chabot College Scott Hildreth
Physics 4C Chabot College Scott Hildreth The Inverse Square Law for Light Intensity vs. Distance Using Microwaves Experiment Goals: Experimentally test the inverse square law for light using Microwaves.
More informationEA3012 TRANSCONDUCTANCE AMPLIFIER OPERATION MANUAL
EA3012 TRANSCONDUCTANCE AMPLIFIER Ü EA3012 Transconductance Amplifier Operation Manual Version 2.00 Dec 2011 All product names are trademarks of their respective companies Guarantee and service Transmille
More informationMicroLab 500-series Getting Started
MicroLab 500-series Getting Started 2 Contents CHAPTER 1: Getting Started Connecting the Hardware....6 Installing the USB driver......6 Installing the Software.....8 Starting a new Experiment...8 CHAPTER
More informationDVM1190 DIGITAL MULTIMETER
DIGITAL MULTIMETER 1. Introduction Thank you for buying the. This digital multimeter has a large LCD, a data-hold function and a backlight. The device uses a very practical safety mechanism that keeps
More informationPEN TYPE DIGITAL MULTIMETER OPERATION MANUAL T8211D
PEN TYPE DIGITAL MULTIMETER OPERATION MANUAL T8211D T8211D 1 1. SAFETY INFORMATION BE EXTREMELY CAREFUL IN THE USE OF THIS METER. Improper use of this device can result in electric shock or destroy of
More informationLA-T SERIES. Fast and reliable minute granular flaw detection in winding wires. Winding Wire Granular Flaw Detector PARTICULAR USE SENSORS
SERIES Winding Wire Granular Flaw Detector Orders accepted till September, 2003 Production to be discontinued from April, 2004 Fast and reliable minute granular flaw detection in winding wires Slim Reliable
More informationVersaFluor Fluorometer System
VersaFluor Fluorometer System Instruction Manual Catalog Numbers 170-2402 through 170-2404 For Technical Service Call Your Local Bio-Rad Office or in the U.S. Call 1-800-4BIORAD (1-800-424-6723) Warranty
More informationUser instructions Compound laboratory microscope
KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen E-mail: info@kern-sohn.com User instructions Compound laboratory microscope Tel: +49-[0]7433-9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com KERN
More informationADVANCED USER S GUIDE
ADVANCED USER S GUIDE DCP-J125 DCP-J315W Version 0 ARL/ASA/NZ/SAF User's Guides and where do I find it? Which manual? What's in it? Where is it? Safety and Legal Quick Setup Guide Basic User's Guide Advanced
More information