The Influence of Crystal Configuration and PMT on PET Time-of-Flight Resolution

Size: px
Start display at page:

Download "The Influence of Crystal Configuration and PMT on PET Time-of-Flight Resolution"

Transcription

1 The Influence of Crystal Configuration and PMT on PET Time-of-Flight Resolution Christopher Thompson Montreal Neurological Institute and Scanwell Systems, Montreal, Canada Jason Hancock Cross Cancer Institute, Edmonton, Canada

2 Introduction: Timing alignment in PET In order to detect positron annihilations the detectors of a PET scanner must detect two coincident gamma rays This requires that gamma rays which are detected at the same time, are reported by the detectors at the same time Different cable lengths, electronic variations, and stability require that the detectors be calibrated by a timing offset.

3 Motivation for Better Timing Alignment A narrower time window improves PET images. Lower random counts. In dynamic PET studies this provides better image quality in early time frames. In whole-body PET scans this permits higher doses to be used providing better image quality with short scan times implying faster patient throughput. Substantial timing improvement would allow the use of time-of-flight information Moses W W: Time of flight in PET revisited (IEEE TNS Oct. 2003)

4 What is Timing Alignment? Timing window shown in blue

5 Why can t the resolving time be made shorter? In order to reduce the resolving time, the detectors must be very well aligned. It is a long, iterative process to perform the timing alignment. The timing response may drift over time. It is better to have noise due to increased random counts than due to poor timing alignment.

6 Positron decay A positron escapes from the parent nucleus The mass difference between the nuclei is given up as kinetic energy randomly distributed between the positron and nucleus The positron must lose its kinetic before it combines with an electron then annihilates. This must precede the detection of the gamma rays by the detectors of the PET scanner. If the energy is lost in a plastic scintillator, the time of the positron decay can be detected!!

7 Timing Using Positron Detection Central source emits positrons which deposit energy in plastic scintillator. Light flash detected by PMT recorded by timing circuit. Central positron source Gamma ray reaches any PET detector. Time difference measured. Time-difference histogram made for all (12,000!) crystals simultaneously. Time offset for each crystal is the peak of timing histogram. Timing Circuit

8 Central Time Alignment Probe Timing probe used in these studies. The connectors on the left provide V PMT bias and connect the timing anode signal from the PMT

9 Equipment and Properties Canberra 2145 Time-to-Amplitude Converter Ortec CF8000 Oct Constant Fraction Discriminator Walk: <±250 ps Fraction: 40% Internal delay jumpers Canberra 454 Quad Constant Fraction Discriminator Walk: <±100 ps (typically < 30 ps) Fraction: 20% External Delay (with 0.6 ns internal delay)

10 Experiment Setup γ-ray s time-of-flight to PET detector Scanwell PET timing alignment probe PET block detector undergoing tests MCA Σ Σ Start NIM crate Stop CFD Fast/slow Amp X4 Amp Sum Amp TAC CAMAC crate ADC sequencer ADC SCSI DEC Alpha workstation

11 Time arrival delays for crystals in HiRez Detector Arrival time ns ± ns Resolution (FWHM) 1.25 ns ± ns

12 Time arrival delays for crystals in HiRez Detector ~250 ps per frame

13 Time arrival delays for crystals in Focus Detector Arrival time 1.02 ns ± ns Resolution (FWHM) 1.21 ns ± ns

14 Time delays in Focus Detector ~250 ps per frame

15 Cause of arrival time delays in a PS-PMT The six anodes in the PS-PMT are interconnected by 1kΩ resistors and there is a small but significant (10-20 pf) capacitance between the anodes. As a result, a signal on anode 1 or 6 will have a fast, big signal at one end of the chain, and a slow, small signal at the other end. The central anodes provide slower, delayed signal to each end, so the apparent arrival time is later from central crystals when their sum signal is processed by the CFD. This effect could be eliminated by taking the timing signal from the last dynode instead of the sum of the anode signals Anodes in a PS-PMT connected with 1 kω resistors Shape of the sum signal from both ends of the anode chain when the γ-ray interacts with a crystal near each anode Measured delays for each crystal of Focus detector

16 Signals from anodes in PS-PMT The anode wires are connected with 1kΩ resistors. If they were displayed individually on an oscilloscope the signals would show a fast rise time at the anode nearest the crystal of interaction, but appear attenuated and delayed and integrated by the inter-anode capacitance at the other anodes. Response to input on Anode #3 shown for 100 nsec.

17 Cause of arrival time delays in PET detectors The four PMTs, or one quad-pmt, in a conventional PET detector do not have the same gain. To provide consistent outputs, the anode voltage is reduced on high gain PMTs, however this increases the transit time in that PMT so delays the signal at the anode. This effect is reduced by matching the PMTs gains during assembly of the block detectors, but gain adjustments are still required. t cathode anode = 2D ( n + 1) d d 2eV mn ( d + 1) D d N d = distance between dynodes = number of dynodes V = voltage across PMT e,m = charge and mass of electron The transit time from photo-cathode to anode of electrons in the Photonis Xp-1452 quad PMT used in the GE PET/CT scanner as a function of PMT anode voltage. The apparent arrival times of γ-rays in crystals in a PET detector. The PMT in the near corner has a higher gain so its anode voltage is reduced resulting in the slower electron transit time.

18 Acknowledgements NSERC (PET research grant to Chris Thompson) Sara St. James (Former student) Lissa Zyromski (IPL: Implantation of radio-isotopes) Frank Wilkinson (Alpha Spectra Inc: Scintillator) Steve Kecani (McGill Physics workshop: Machining) US Patent #7,247,844: (July-2007) An instrument and method to improve the timing alignment of a PET scanner C J Thompson and M-L Camborde

Gamma Ray Spectroscopy with NaI(Tl) and HPGe Detectors

Gamma Ray Spectroscopy with NaI(Tl) and HPGe Detectors Nuclear Physics #1 Gamma Ray Spectroscopy with NaI(Tl) and HPGe Detectors Introduction: In this experiment you will use both scintillation and semiconductor detectors to study γ- ray energy spectra. The

More information

Designing an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare

Designing an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare GE Healthcare Designing an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare There is excitement across the industry regarding the clinical potential of a hybrid

More information

COMPTON SCATTERING. Purpose. Introduction. Fundamentals of Experiment

COMPTON SCATTERING. Purpose. Introduction. Fundamentals of Experiment COMPTON SCATTERING Purpose The purpose of this experiment is to verify the energy dependence of gamma radiation upon scattering angle and to compare the differential cross section obtained from the data

More information

Instructions for gg Coincidence with 22 Na. Overview of the Experiment

Instructions for gg Coincidence with 22 Na. Overview of the Experiment Overview of the Experiment Instructions for gg Coincidence with 22 Na 22 Na is a radioactive element that decays by converting a proton into a neutron: about 90% of the time through β + decay and about

More information

Spontaneous Fission Spectrum of Neutrons from 252 Cf with Kinetic Energies Less than 1 MeV. Suraj Bastola. A senior thesis submitted to the faculty of

Spontaneous Fission Spectrum of Neutrons from 252 Cf with Kinetic Energies Less than 1 MeV. Suraj Bastola. A senior thesis submitted to the faculty of Spontaneous Fission Spectrum of Neutrons from 252 Cf with Kinetic Energies Less than 1 MeV Suraj Bastola A senior thesis submitted to the faculty of Brigham Young University in partial fulfillment of the

More information

Physics Laboratory Scattering of Photons from Electrons: Compton Scattering

Physics Laboratory Scattering of Photons from Electrons: Compton Scattering RR Oct 2001 SS Dec 2001 MJ Oct 2009 Physics 34000 Laboratory Scattering of Photons from Electrons: Compton Scattering Objective: To measure the energy of high energy photons scattered from electrons in

More information

Experiment 10. The Speed of Light c Introduction Apparatus

Experiment 10. The Speed of Light c Introduction Apparatus Experiment 10 The Speed of Light c 10.1 Introduction In this experiment you will measure the speed of light, c. This is one of the most fundamental constants in physics, and at the same time the fastest

More information

The (Speed and) Decay of Cosmic-Ray Muons

The (Speed and) Decay of Cosmic-Ray Muons The (Speed and) Decay of Cosmic-Ray Muons Jason Gross MIT - Department of Physics Jason Gross (8.13) Cosmic-Ray Muons November 4, 2011 1 / 30 Goals test relativity (time dilation) determine the mean lifetime

More information

Activities in Electronics Lab Associates are: Mrs. Arti Gupta, K.S.Golda, S.Muralithar & Dr.R.K.Bhowmik

Activities in Electronics Lab Associates are: Mrs. Arti Gupta, K.S.Golda, S.Muralithar & Dr.R.K.Bhowmik Activities in Electronics Lab. 2006 Associates are: Mrs. Arti Gupta, K.S.Golda, S.Muralithar & Dr.R.K.Bhowmik Nuclear Electronics (INGA, NAND) Pulse Shape Discriminator Electronics for NAND National Array

More information

Positron Emission Tomography

Positron Emission Tomography Positron Emission Tomography UBC Physics & Astronomy / PHYS 409 1 Introduction Positron emission tomography (PET) is a non-invasive way to produce the functional 1 image of a patient. It works by injecting

More information

Testing of the NSC Electronics Module with the GSI Clover Detector

Testing of the NSC Electronics Module with the GSI Clover Detector Testing of the NSC Electronics Module with the GSI Clover Detector Rakesh Kumar 1, P. Queiroz 2, H.-J. Wollersheim 2 (Tutor) 1 Inter University Accelerator Centre Aruna Asaf Ali Marg Post Box No 10502

More information

Time-of-flight PET with SiPM sensors on monolithic scintillation crystals Vinke, Ruud

Time-of-flight PET with SiPM sensors on monolithic scintillation crystals Vinke, Ruud University of Groningen Time-of-flight PET with SiPM sensors on monolithic scintillation crystals Vinke, Ruud IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you

More information

Index terms: PET, Silicon Photo-multiplier, Small Animal PET insert for MRI scanner. Size: 1.2x1.2 mm, Pitch: mm Thickness: 4 mm

Index terms: PET, Silicon Photo-multiplier, Small Animal PET insert for MRI scanner. Size: 1.2x1.2 mm, Pitch: mm Thickness: 4 mm Measurement of Energy and Timing Resolution of Very Highly Pixellated LYSO Crystal Blocks with Multiplexed SiPM Readout for Use in a Small Animal PET/MR Insert Christopher J. Thompson, Senior Member IEEE,

More information

Simulation of Algorithms for Pulse Timing in FPGAs

Simulation of Algorithms for Pulse Timing in FPGAs 2007 IEEE Nuclear Science Symposium Conference Record M13-369 Simulation of Algorithms for Pulse Timing in FPGAs Michael D. Haselman, Member IEEE, Scott Hauck, Senior Member IEEE, Thomas K. Lewellen, Senior

More information

Electronic Instrumentation for Radiation Detection Systems

Electronic Instrumentation for Radiation Detection Systems Electronic Instrumentation for Radiation Detection Systems January 23, 2018 Joshua W. Cates, Ph.D. and Craig S. Levin, Ph.D. Course Outline Lecture Overview Brief Review of Radiation Detectors Detector

More information

Fast first practical help -- detailed instructions will follow- preliminary Experiment F80

Fast first practical help -- detailed instructions will follow- preliminary Experiment F80 Fast first practical help -- detailed instructions will follow- preliminary Experiment F80 Measurement Methods of Nuclear and Particle Physics Introduction: This experiment is going to introduce you to

More information

ORTEC Experiment 13. Gamma-Gamma Coincidence with Angular Correlation. Equipment Required

ORTEC Experiment 13. Gamma-Gamma Coincidence with Angular Correlation. Equipment Required ORTEC Experiment 13 Equipment Required Two 905-3 2-in. x 2-in. NaI(Tl) Scintillation Detector Assemblies. Two 266 Photomultiplier Tube Bases. Two 113 Scintillation Preamplifiers. Two 556 High Voltage Power

More information

MASE: Multiplexed Analog Shaped Electronics

MASE: Multiplexed Analog Shaped Electronics MASE: Multiplexed Analog Shaped Electronics C. Metelko, A. Alexander, J. Poehlman, S. Hudan, R.T. desouza Outline 1. Needs 2. Problems with existing Technology 3. Design Specifications 4. Overview of the

More information

Performance measurements of a depth-encoding PET detector module based on positionsensitive

Performance measurements of a depth-encoding PET detector module based on positionsensitive Home Search Collections Journals About Contact us My IOPscience Performance measurements of a depth-encoding PET detector module based on positionsensitive avalanche photodiode read-out This article has

More information

K 223 Angular Correlation

K 223 Angular Correlation K 223 Angular Correlation K 223.1 Aim of the Experiment The aim of the experiment is to measure the angular correlation of a γ γ cascade. K 223.2 Required Knowledge Definition of the angular correlation

More information

Digitization of PMT signals with FADCs: comparison of simulation and measurement

Digitization of PMT signals with FADCs: comparison of simulation and measurement Digitization of PMT signals with FADCs: comparison of simulation and measurement Arno Gadola General, 10. 12.05.2010 Outline Summary of previous presentations Impact of sampling rate Verification of simulation

More information

A high-performance, low-cost, leading edge discriminator

A high-performance, low-cost, leading edge discriminator PRAMANA c Indian Academy of Sciences Vol. 65, No. 2 journal of August 2005 physics pp. 273 283 A high-performance, low-cost, leading edge discriminator S K GUPTA a, Y HAYASHI b, A JAIN a, S KARTHIKEYAN

More information

Purpose This experiment will use the coincidence method for time correlation to measure the lifetime in the decay scheme of 57

Purpose This experiment will use the coincidence method for time correlation to measure the lifetime in the decay scheme of 57 Equipment Required Two 113 Scintillation Preamplifiers Two 266 Photomultiplier Tube Bases 4001A/4002D Bin and Power Supply 414A Fast Coincidence Two 551 Timing Single-Channel Analyzers 567 Time-to-Amplitude

More information

4 Time walk correction for TOF-PET detectors based on a monolithic scintillation crystal coupled to a photosensor array

4 Time walk correction for TOF-PET detectors based on a monolithic scintillation crystal coupled to a photosensor array 4 Time walk correction for TOF-PET detectors based on a monolithic scintillation crystal coupled to a photosensor array This chapter has been published as: R. Vinke, H. Löhner, D. Schaart, H. van Dam,

More information

Week 11: Chap. 16b Pulse Shaping

Week 11: Chap. 16b Pulse Shaping Week 11: Chap. 16b Pulse Shaping Pulse Processing (passive) Pulse Shaping (active) -- Op Amps -- CR/RC network -- Bipolar pulses --- Shaping network --- Pole Zero network --- Baseline Restorer -- Delay-line

More information

ORTEC. AN34 Experiment 14 Nuclear Lifetimes and the Coincidence Method. Equipment Needed from ORTEC. Equipment Required from Other Manufacturers

ORTEC. AN34 Experiment 14 Nuclear Lifetimes and the Coincidence Method. Equipment Needed from ORTEC. Equipment Required from Other Manufacturers Equipment Needed from ORTEC Two 113 Scintillation Preamplifiers Two 266 Photomultiplier Tube Bases 4001A/4002D Bin and Power Supply 414A Fast Coincidence Two 551 Timing Single-Channel Analyzers 567 Time-to-Amplitude

More information

Advancement in development of photomultipliers dedicated to new scintillators studies.

Advancement in development of photomultipliers dedicated to new scintillators studies. Advancement in development of photomultipliers dedicated to new scintillators studies. Maciej Kapusta, Pascal Lavoutea, Florence Lherbet, Cyril Moussant, Paul Hink INTRODUCTION AND OUTLINE In the validation

More information

Physics Experiment N -17. Lifetime of Cosmic Ray Muons with On-Line Data Acquisition on a Computer

Physics Experiment N -17. Lifetime of Cosmic Ray Muons with On-Line Data Acquisition on a Computer Introduction Physics 410-510 Experiment N -17 Lifetime of Cosmic Ray Muons with On-Line Data Acquisition on a Computer The experiment is designed to teach the techniques of particle detection using scintillation

More information

ORTEC. Time-to-Amplitude Converters and Time Calibrator. Choosing the Right TAC. Timing with TACs

ORTEC. Time-to-Amplitude Converters and Time Calibrator. Choosing the Right TAC. Timing with TACs ORTEC Time-to-Amplitude Converters Choosing the Right TAC The following topics provide the information needed for selecting the right time-to-amplitude converter (TAC) for the task. The basic principles

More information

Solid-State Photomultiplier in CMOS Technology for Gamma-Ray Detection and Imaging Applications

Solid-State Photomultiplier in CMOS Technology for Gamma-Ray Detection and Imaging Applications Solid-State Photomultiplier in CMOS Technology for Gamma-Ray Detection and Imaging Applications Christopher Stapels, Member, IEEE, William G. Lawrence, James Christian, Member, IEEE, Michael R. Squillante,

More information

Timing Resolution Performance Comparison for Fast and Standard Outputs of SensL SiPM

Timing Resolution Performance Comparison for Fast and Standard Outputs of SensL SiPM Timing Resolution Performance Comparison for Fast and Standard Outputs of SensL SiPM Sergei Dolinsky, Geng Fu, and Adrian Ivan Abstract A new silicon photomultiplier (SiPM) with a unique fast output signal

More information

Silicon Photomultiplier Evaluation Kit. Quick Start Guide. Eval Kit SiPM. KETEK GmbH. Hofer Str Munich Germany.

Silicon Photomultiplier Evaluation Kit. Quick Start Guide. Eval Kit SiPM. KETEK GmbH. Hofer Str Munich Germany. KETEK GmbH Hofer Str. 3 81737 Munich Germany www.ketek.net info@ketek.net phone +49 89 673 467 70 fax +49 89 673 467 77 Silicon Photomultiplier Evaluation Kit Quick Start Guide Eval Kit Table of Contents

More information

ORTEC ORTEC. Modular Pulse- Processing Electronics. What s in this Catalog? Who Needs this Catalog?

ORTEC ORTEC. Modular Pulse- Processing Electronics.   What s in this Catalog? Who Needs this Catalog? ORTEC Modular Pulse- Processing Electronics What s in this Catalog? Tutorial Information makes this more than just a catalog. "What You Need to Know About Modular Electronic Instruments" takes you through

More information

PROGRESS in TOF PET timing resolution continues to

PROGRESS in TOF PET timing resolution continues to Combined Analog/Digital Approach to Performance Optimization for the LAPET Whole-Body TOF PET Scanner W. J. Ashmanskas, Member, IEEE, Z. S. Davidson, B. C. LeGeyt, F. M. Newcomer, Member, IEEE, J. V. Panetta,

More information

Method for digital particle spectrometry Khryachkov Vitaly

Method for digital particle spectrometry Khryachkov Vitaly Method for digital particle spectrometry Khryachkov Vitaly Institute for physics and power engineering (IPPE) Obninsk, Russia The goals of Analog Signal Processing Signal amplification Signal filtering

More information

Performance Assessment of Pixelated LaBr 3 Detector Modules for TOF PET

Performance Assessment of Pixelated LaBr 3 Detector Modules for TOF PET Performance Assessment of Pixelated LaBr 3 Detector Modules for TOF PET A. Kuhn, S. Surti, Member, IEEE, J. S. Karp, Senior Member, IEEE, G. Muehllehner, Fellow, IEEE, F.M. Newcomer, R. VanBerg Abstract--

More information

Model 935 Quad Constant-Fraction 200-MHz Discriminator Operating and Service Manual

Model 935 Quad Constant-Fraction 200-MHz Discriminator Operating and Service Manual Model 935 Quad Constant-Fraction 200-MHz Discriminator Operating and Service Manual U.S. Patent No. 4,179,644 Printed in U.S.A. ORTEC Part No. 753770 0503 Manual Revision H ii $GYDQFHG0HDVXUHPHQW7HFKQRORJ\,QF

More information

A user-friendly fully digital TDPAC-spectrometer

A user-friendly fully digital TDPAC-spectrometer Hyperfine Interact DOI 10.1007/s10751-010-0201-8 A user-friendly fully digital TDPAC-spectrometer M. Jäger K. Iwig T. Butz Springer Science+Business Media B.V. 2010 Abstract A user-friendly fully digital

More information

ORTEC. Research Applications. Pulse-Height, Charge, or Energy Spectroscopy. Detectors. Processing Electronics

ORTEC. Research Applications. Pulse-Height, Charge, or Energy Spectroscopy. Detectors. Processing Electronics ORTEC Spectroscopy systems for ORTEC instrumentation produce pulse height distributions of gamma ray or alpha energies. MAESTRO-32 (model A65-B32) is the software included with most spectroscopy systems

More information

Amptek sets the New State-of-the-Art... Again! with Cooled FET

Amptek sets the New State-of-the-Art... Again! with Cooled FET Amptek sets the New State-of-the-Art... Again! with Cooled FET RUN SILENT...RUN FAST...RUN COOL! Performance Noise: 670 ev FWHM (Si) ~76 electrons RMS Noise Slope: 11.5 ev/pf High Ciss FET Fast Rise Time:

More information

The 2017 IEEE NSS-MIC. Industrial Presentation

The 2017 IEEE NSS-MIC. Industrial Presentation Industrial Presentation 1 Introduction of new ultra high count rate Pileup Separator Processor ideal for silicon drift detector and LaBr 3 scintillation detector Tuesday, October 24 2:30:00 PM Hanover

More information

PET Detectors. William W. Moses Lawrence Berkeley National Laboratory March 26, 2002

PET Detectors. William W. Moses Lawrence Berkeley National Laboratory March 26, 2002 PET Detectors William W. Moses Lawrence Berkeley National Laboratory March 26, 2002 Step 1: Inject Patient with Radioactive Drug Drug is labeled with positron (β + ) emitting radionuclide. Drug localizes

More information

CHAPTER 8 GENERIC PERFORMANCE MEASURES

CHAPTER 8 GENERIC PERFORMANCE MEASURES GENERIC PERFORMANCE MEASURES M.E. DAUBE-WITHERSPOON Department of Radiology, University of Pennsylvania, Philadelphia, Pennsylvania, United States of America 8.1. INTRINSIC AND EXTRINSIC MEASURES 8.1.1.

More information

PoS(PhotoDet 2012)022

PoS(PhotoDet 2012)022 SensL New Fast Timing Silicon Photomultiplier Kevin O`Neill 1 SensL Technologies Limited 6800 Airport Business Park, Cork, Ireland E-mail: koneill@sensl.com Nikolai Pavlov SensL Technologies Limited 6800

More information

and N(t) ~ exp(-t/ ),

and N(t) ~ exp(-t/ ), Muon Lifetime Experiment Introduction Charged and neutral particles with energies in excess of 10 23 ev from Galactic and extra Galactic sources impinge on the earth. Here we speak of the earth as the

More information

User Guide. SIB Channel MAPMT Interface Board Hamamatsu H7546 series

User Guide. SIB Channel MAPMT Interface Board Hamamatsu H7546 series SIB164-1018 64 Channel MAPMT Interface Board Hamamatsu H7546 series Disclaimer Vertilon Corporation has made every attempt to ensure that the information in this document is accurate and complete. Vertilon

More information

CAEN. Electronic Instrumentation. CAEN Silicon Photomultiplier Kit

CAEN. Electronic Instrumentation. CAEN Silicon Photomultiplier Kit CAEN Tools for Discovery Electronic Instrumentation CAEN Silicon Photomultiplier Kit CAEN realized a modular development kit dedicated to Silicon Photomultipliers, representing the state-of-the art in

More information

Explain what is meant by a photon and state one of its main properties [2]

Explain what is meant by a photon and state one of its main properties [2] 1 (a) A patient has an X-ray scan taken in hospital. The high-energy X-ray photons interact with the atoms inside the body of the patient. Explain what is meant by a photon and state one of its main properties....

More information

Real-Time Digital Signal Processors with radiation detectors produced by TechnoAP

Real-Time Digital Signal Processors with radiation detectors produced by TechnoAP Real-Time Digital Signal Processors with radiation detectors produced by TechnoAP Lunch time Exhibitor presentation 2976-15 Mawatari, Hitachinaka-city, Ibaraki 312-0012, Japan Phone: +81-29-350-8011, FAX:

More information

Characterization of a 64 Channel PET Detector Using Photodiodes for Crystal Identification *

Characterization of a 64 Channel PET Detector Using Photodiodes for Crystal Identification * Characterization of a 64 Channel PET Detector Using Photodiodes for Crystal Identification * J. S. Huber, Member, IEEE, W.W. Moses, Senior Member, IEEE, S.E. Derenzo, Senior Member, IEEE, M.H. Ho, M.S.

More information

Cosmic Rays in MoNA. Eric Johnson 8/08/03

Cosmic Rays in MoNA. Eric Johnson 8/08/03 Cosmic Rays in MoNA Eric Johnson 8/08/03 National Superconducting Cyclotron Laboratory Department of Physics and Astronomy Michigan State University Advisors: Michael Thoennessen and Thomas Baumann Abstract:

More information

Channel-Plate Photomultipliers

Channel-Plate Photomultipliers The New Micro-Channel Channel-Plate Photomultipliers A revolution in lifetime spectroscopy? * ) F. Bečvář Charles University, Prague * ) and also New Ultra-Fast Digitizers OUTLINE The state-of-the-art

More information

Digital Signal Processing for HPGe Detectors

Digital Signal Processing for HPGe Detectors Digital Signal Processing for HPGe Detectors David Radford ORNL Physics Division July 28, 2012 HPGe Detectors Hyper-Pure Ge (HPGe) detectors are the gold standard for gamma-ray spectroscopy Unsurpassed

More information

ARTICLE IN PRESS. Nuclear Instruments and Methods in Physics Research A

ARTICLE IN PRESS. Nuclear Instruments and Methods in Physics Research A Nuclear Instruments and Methods in Physics Research A 614 (2010) 308 312 Contents lists available at ScienceDirect Nuclear Instruments and Methods in Physics Research A journal homepage: www.elsevier.com/locate/nima

More information

GAMMA-GAMMA CORRELATION Latest Revision: August 21, 2007

GAMMA-GAMMA CORRELATION Latest Revision: August 21, 2007 C1-1 GAMMA-GAMMA CORRELATION Latest Revision: August 21, 2007 QUESTION TO BE INVESTIGATED: decay event? What is the angular correlation between two gamma rays emitted by a single INTRODUCTION & THEORY:

More information

CHAPTER 7 MEDICINE IMAGING DEVICES

CHAPTER 7 MEDICINE IMAGING DEVICES CHAPTER 7 ELECTRONICS RELATED TO NUCLEAR MEDICINE IMAGING DEVICES R.J. OTT Joint Department of Physics, Royal Marsden Hospital and Institute of Cancer Research, Surrey R. STEPHENSON Rutherford Appleton

More information

Generating Jitter for Fibre Channel Compliance Testing

Generating Jitter for Fibre Channel Compliance Testing Application Note: HFAN-4.5.2 Rev 0; 12/00 Generating Jitter for Fibre Channel Compliance Testing MAXIM High-Frequency/Fiber Communications Group 4hfan452.doc 01/02/01 Generating Jitter for Fibre Channel

More information

Voltage Dividers & Electronics Scintillation detectors usually employ a Voltage Divider (VD) network to operate the PMT. This sometimes called "bleeder network" defines a potential (voltage) difference

More information

Physics 342 Laboratory. Scattering of Photons from Free Electrons: Compton Scattering

Physics 342 Laboratory. Scattering of Photons from Free Electrons: Compton Scattering RR Oct 2001 SS Dec 2001 Physics 342 Laboratory Scattering of Photons from Free Electrons: Compton Scattering Objective: To measure the energy of high energy photons scattered from electrons in a brass

More information

nanomca-sp datasheet I. FEATURES

nanomca-sp datasheet I. FEATURES datasheet nanomca-sp 80 MHz HIGH PERFORMANCE, LOW POWER DIGITAL MCA WITH BUILT IN PREAMPLIFIER Model Numbers: SP0534A/B to SP0539A/B Standard Models: SP0536B and SP0536A I. FEATURES Built-in preamplifier

More information

--- preliminary Experiment F80

--- preliminary Experiment F80 --- preliminary Experiment F80 Measurement Methods of Nuclear and Particle Physics Introduction: This experiment is going to introduce you to important counting and measuring techniques of nuclear and

More information

A Continuous Crystal Detector for TOF PET

A Continuous Crystal Detector for TOF PET 1 A Continuous Crystal Detector for TOF PET T. Szczęśniak, Member, IEEE, M. Moszyński, Fellow, IEEE, Ł. Świderski, Member, IEEE, A. Nassalski, Member, IEEE, A. Syntfeld-KaŜuch, Member, IEEE, P. Ojala,

More information

Performance characterization of a novel thin position-sensitive avalanche photodiode-based detector for high resolution PET

Performance characterization of a novel thin position-sensitive avalanche photodiode-based detector for high resolution PET 2005 IEEE Nuclear Science Symposium Conference Record M11-126 Performance characterization of a novel thin position-sensitive avalanche photodiode-based detector for high resolution PET Jin Zhang, Member,

More information

Virtual Laboratory of Nuclear Fission Virtual practicum in the framework of the project Virtual Laboratory of Nuclear Fission

Virtual Laboratory of Nuclear Fission Virtual practicum in the framework of the project Virtual Laboratory of Nuclear Fission Virtual Laboratory of Nuclear Fission Virtual practicum in the framework of the project Virtual Laboratory of Nuclear Fission Khanyisa Sowazi, University of the Western Cape JINR SAR, September 2015 INDEX

More information

ORTEC Experiment 3. Gamma-Ray Spectroscopy Using NaI(Tl) Equipment Required. Purpose. Gamma Emission

ORTEC Experiment 3. Gamma-Ray Spectroscopy Using NaI(Tl) Equipment Required. Purpose. Gamma Emission ORTEC Experiment 3 Equipment Required Electronic Instrumentation o SPA38 Integral Assembly consisting of a 38 mm x 38 mm NaI(Tl) Scintillator, Photomultiplier Tube, and PMT Base with Stand o 4001A/4002D

More information

Celesteion Time-of-Flight Technology

Celesteion Time-of-Flight Technology Celesteion Time-of-Flight Technology Bing Bai, PhD Clinical Sciences Manager, PET/CT Canon Medical Systems USA Introduction Improving the care for every patient while providing a high standard care to

More information

Highly Segmented Detector Arrays for. Studying Resonant Decay of Unstable Nuclei. Outline

Highly Segmented Detector Arrays for. Studying Resonant Decay of Unstable Nuclei. Outline Highly Segmented Detector Arrays for Studying Resonant Decay of Unstable Nuclei MASE: Multiplexed Analog Shaper Electronics C. Metelko, S. Hudan, R.T. desouza Outline 1. Resonant Decay 2. Detectors 3.

More information

PX4 Frequently Asked Questions (FAQ)

PX4 Frequently Asked Questions (FAQ) PX4 Frequently Asked Questions (FAQ) What is the PX4? The PX4 is a component in the complete signal processing chain of a nuclear instrumentation system. It replaces many different components in a traditional

More information

Ph 3324 The Scintillation Detector and Gamma Ray Spectroscopy

Ph 3324 The Scintillation Detector and Gamma Ray Spectroscopy Ph 3324 The Scintillation Detector and Gamma Ray Spectroscopy Required background reading Attached are several pages from an appendix on the web for Tipler-Llewellyn Modern Physics. Read the section on

More information

Discovery ST. An Oncology System Designed For PET/CT. Revision: B Date: 30 Jan Page 1 of 47

Discovery ST. An Oncology System Designed For PET/CT. Revision: B Date: 30 Jan Page 1 of 47 Discovery ST An Oncology System Designed For PET/CT Revision: B Date: 30 Jan 2003 Page 1 of 47 TABLE OF CONTENTS 1 Introduction...3 2 Design Requirements...4 2.1 The Design Objective...4 2.2 Design Philosophy...5

More information

nanomca 80 MHz HIGH PERFORMANCE, LOW POWER DIGITAL MCA Model Numbers: NM0530 and NM0530Z

nanomca 80 MHz HIGH PERFORMANCE, LOW POWER DIGITAL MCA Model Numbers: NM0530 and NM0530Z datasheet nanomca 80 MHz HIGH PERFORMANCE, LOW POWER DIGITAL MCA Model Numbers: NM0530 and NM0530Z I. FEATURES Finger-sized, high performance digital MCA. 16k channels utilizing smart spectrum-size technology

More information

Traditional analog QDC chain and Digital Pulse Processing [1]

Traditional analog QDC chain and Digital Pulse Processing [1] Giuliano Mini Viareggio April 22, 2010 Introduction The aim of this paper is to compare the energy resolution of two gamma ray spectroscopy setups based on two different acquisition chains; the first chain

More information

LaBr 3 :Ce, the latest crystal for nuclear medicine

LaBr 3 :Ce, the latest crystal for nuclear medicine 10th Topical Seminar on Innovative Particle and Radiation Detectors 1-5 October 2006 Siena, Italy LaBr 3 :Ce, the latest crystal for nuclear medicine Roberto Pani On behalf of SCINTIRAD Collaboration INFN

More information

Status of Primex Beam Position Monitor July 29 th, 2010

Status of Primex Beam Position Monitor July 29 th, 2010 Status of Primex Beam Position Monitor July 29 th, 2010 Anthony Tatum University of North Carolina at Wilmington The Beam Position Monitor (BPM) is used to determine the vertical and horizontal position

More information

Scintillators as an external trigger for cathode strip chambers

Scintillators as an external trigger for cathode strip chambers Scintillators as an external trigger for cathode strip chambers J. A. Muñoz Department of Physics, Princeton University, Princeton, NJ 08544 An external trigger was set up to test cathode strip chambers

More information

e t Development of Low Cost γ - Ray Energy Spectrometer

e t Development of Low Cost γ - Ray Energy Spectrometer e t International Journal on Emerging Technologies (Special Issue on NCRIET-2015) 6(2): 315-319(2015) ISSN No. (Print) : 0975-8364 ISSN No. (Online) : 2249-3255 Development of Low Cost γ - Ray Energy Spectrometer

More information

Overview 256 channel Silicon Photomultiplier large area using matrix readout system The SensL Matrix detector () is the largest area, highest channel

Overview 256 channel Silicon Photomultiplier large area using matrix readout system The SensL Matrix detector () is the largest area, highest channel 技股份有限公司 wwwrteo 公司 wwwrteo.com Page 1 Overview 256 channel Silicon Photomultiplier large area using matrix readout system The SensL Matrix detector () is the largest area, highest channel count, Silicon

More information

SiPMs for solar neutrino detector? J. Kaspar, 6/10/14

SiPMs for solar neutrino detector? J. Kaspar, 6/10/14 SiPMs for solar neutrino detector? J. Kaspar, 6/0/4 SiPM is photodiode APD Geiger Mode APD V APD full depletion take a photo-diode reverse-bias it above breakdown voltage (Geiger mode avalanche photo diode)

More information

Attenuation length in strip scintillators. Jonathan Button, William McGrew, Y.-W. Lui, D. H. Youngblood

Attenuation length in strip scintillators. Jonathan Button, William McGrew, Y.-W. Lui, D. H. Youngblood Attenuation length in strip scintillators Jonathan Button, William McGrew, Y.-W. Lui, D. H. Youngblood I. Introduction The ΔE-ΔE-E decay detector as described in [1] is composed of thin strip scintillators,

More information

Contents. Why waveform? Waveform digitizer : Domino Ring Sampler CEX Beam test autumn 04. Summary

Contents. Why waveform? Waveform digitizer : Domino Ring Sampler CEX Beam test autumn 04. Summary Contents Why waveform? Waveform digitizer : Domino Ring Sampler CEX Beam test data @PSI autumn 04 Templates and time resolution Pulse Shape Discrimination Pile-up rejection Summary 2 In the MEG experiment

More information

Scintillation Counters

Scintillation Counters PHY311/312 Detectors for Nuclear and Particle Physics Dr. C.N. Booth Scintillation Counters Unlike many other particle detectors, which exploit the ionisation produced by the passage of a charged particle,

More information

PCS-150 / PCI-200 High Speed Boxcar Modules

PCS-150 / PCI-200 High Speed Boxcar Modules Becker & Hickl GmbH Kolonnenstr. 29 10829 Berlin Tel. 030 / 787 56 32 Fax. 030 / 787 57 34 email: info@becker-hickl.de http://www.becker-hickl.de PCSAPP.DOC PCS-150 / PCI-200 High Speed Boxcar Modules

More information

Real Time Pulse Pile-up Recovery in a High Throughput Digital Pulse Processor

Real Time Pulse Pile-up Recovery in a High Throughput Digital Pulse Processor Real Time Pulse Pile-up Recovery in a High Throughput Digital Pulse Processor Paul A. B. Scoullar a, Chris C. McLean a and Rob J. Evans b a Southern Innovation, Melbourne, Australia b Department of Electrical

More information

ORTEC Experiment 1. Introduction to Electronic Signal Analysis in Nuclear Radiation Measurements. Equipment Required: Purpose. Electronic Circuits

ORTEC Experiment 1. Introduction to Electronic Signal Analysis in Nuclear Radiation Measurements. Equipment Required: Purpose. Electronic Circuits ORTEC Experiment 1 Equipment Required: 480 Pulser 113 Scintillation Preamplifier 4001A/4002D NIM Bin and Power Supply 575A Spectroscopy Amplifier 996 Timer and Counter 551 Timing Single-Channel Analyzer

More information

Nuclear Instruments and Methods in Physics Research A

Nuclear Instruments and Methods in Physics Research A Nuclear Instruments and Methods in Physics Research A 699 () Contents lists available at SciVerse ScienceDirect Nuclear Instruments and Methods in Physics Research A journal homepage: www.elsevier.com/locate/nima

More information

The Time of Flight Upgrade for CLAS at 12GeV

The Time of Flight Upgrade for CLAS at 12GeV The Time of Flight Upgrade for CLAS at 12GeV by Lewis P. Graham Bachelor of Science Benedict College, 2002 -------------------------------------------------------------------- Submitted in Partial Fulfillment

More information

event physics experiments

event physics experiments Comparison between large area PMTs at cryogenic temperature for neutrino and rare Andrea Falcone University of Pavia INFN Pavia event physics experiments Rare event physics experiment Various detectors

More information

PH2510 Nuclear Physics Laboratory Use of Scintillation Counters (NP5)

PH2510 Nuclear Physics Laboratory Use of Scintillation Counters (NP5) Physics Department Royal Holloway University of London PH2510 Nuclear Physics Laboratory Use of Scintillation Counters (NP5) 1. Introduction 1.1 Object of the Experiment The object of this experiment is

More information

The detector system of the EPOS system

The detector system of the EPOS system The detector system of the EPOS system 1. The detector arrangement 2. The lifetime system 3. Digital Doppler measurement 4. AMOC Martin-Luther-Universität RK Halle R Detector system 3 experiments: lifetime

More information

PET: New Technologies & Applications, Including Oncology

PET: New Technologies & Applications, Including Oncology PET: New Technologies & Applications, Including Oncology, PhD, FIEEE Imaging Research Laboratory Department of Radiology University of Washington, Seattle, WA Disclosures Research Contract, GE Healthcare

More information

60 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 54, NO. 1, FEBRUARY /$ IEEE

60 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 54, NO. 1, FEBRUARY /$ IEEE 60 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 54, NO. 1, FEBRUARY 2007 Prototype Parallel Readout System for Position Sensitive PMT Based Gamma Ray Imaging Systems Frezghi Habte, Member, IEEE, Peter D.

More information

ORTEC Experiment 9. Time Coincidence Techniques Applied to Absolute Activity Measurements. Equipment Required. Purpose.

ORTEC Experiment 9. Time Coincidence Techniques Applied to Absolute Activity Measurements. Equipment Required. Purpose. ORTEC Experiment 9 Equipment Required 905-3 2-inch x 2-inch (5.08-cm x 5.08-cm) NaI(Tl) Detector and PM Tube Assembly. 266 PhotoMultiplier Tube Base. Two 113 Preamplifiers. BA-015-050-1000 Partially-Depleted

More information

LIFETIME OF THE MUON

LIFETIME OF THE MUON Muon Decay 1 LIFETIME OF THE MUON Introduction Muons are unstable particles; otherwise, they are rather like electrons but with much higher masses, approximately 105 MeV. Radioactive nuclear decays do

More information

A high resolution TOF counter - a way to compete with a RICH detector?

A high resolution TOF counter - a way to compete with a RICH detector? A high resolution TOF counter - a way to compete with a RICH detector? J. Va vra, SLAC representing D.W.G.S. Leith, B. Ratcliff, and J. Schwiening Note: This work was possible because of the Focusing DIRC

More information

PET/CT Instrumentation Basics

PET/CT Instrumentation Basics / Instrumentation Basics 1. Motivations for / imaging 2. What is a / Scanner 3. Typical Protocols 4. Attenuation Correction 5. Problems and Challenges with / 6. Examples Motivations for / Imaging Desire

More information

Time-of-flight PET with SiPM sensors on monolithic scintillation crystals Vinke, Ruud

Time-of-flight PET with SiPM sensors on monolithic scintillation crystals Vinke, Ruud University of Groningen Time-of-flight PET with SiPM sensors on monolithic scintillation crystals Vinke, Ruud IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you

More information

Keyser, Ronald M., Twomey, Timothy R., and Bingham, Russell D. ORTEC, 801 South Illinois Avenue, Oak Ridge, TN 37831s

Keyser, Ronald M., Twomey, Timothy R., and Bingham, Russell D. ORTEC, 801 South Illinois Avenue, Oak Ridge, TN 37831s Improved Performance in Germanium Detector Gamma Spectrometers based on Digital Signal Processing Keyser, Ronald M., Twomey, Timothy R., and Bingham, Russell D. ORTEC, 801 South Illinois Avenue, Oak Ridge,

More information

GRETINA. Electronics. Auxiliary Detector Workshop. Sergio Zimmermann LBNL. Auxiliary Detectors Workshop. January 28, 2006

GRETINA. Electronics. Auxiliary Detector Workshop. Sergio Zimmermann LBNL. Auxiliary Detectors Workshop. January 28, 2006 GRETINA Auxiliary Detector Workshop Electronics Sergio Zimmermann LBNL 1 Outline Electronic Interface Options Digitizers Trigger/Timing System Grounding and Shielding Summary 2 Interface Options Three

More information

Investigation of a New Readout Scheme for High Resolution Scintillation Crystal Arrays Using Photodiodes

Investigation of a New Readout Scheme for High Resolution Scintillation Crystal Arrays Using Photodiodes 120s IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 44, NO. 3, JUNE 1997 Investigation of a New Readout Scheme for High Resolution Scintillation Crystal Arrays Using Photodiodes Craig S. Levin, Member, IEEE,

More information

IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 52, NO. 3, JUNE Investigation of the Block Effect on Spatial Resolution in PET Detectors

IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 52, NO. 3, JUNE Investigation of the Block Effect on Spatial Resolution in PET Detectors IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 52, NO. 3, JUNE 2005 599 Investigation of the Block Effect on Spatial Resolution in PET Detectors Nada Tomic, Student Member, IEEE, Christopher J. Thompson, Member,

More information