--- preliminary Experiment F80
|
|
- Alicia Underwood
- 5 years ago
- Views:
Transcription
1 --- preliminary Experiment F80 Measurement Methods of Nuclear and Particle Physics Introduction: This experiment is going to introduce you to important counting and measuring techniques of nuclear and particle physics. The basic detector used in this experiment is a plastic scintillator read out by a photomultiplier. This type of detector is rather simple, easy to build in very different sizes and geometries and is widely used especially to derive fast timing information if a particle hits the scintillator. Time resolutions of 100 ps can be reached if special care is taken. In this experiment scintillation counters are used to measure the energy of gamma rays and decay electrons. In a second step also the time resolution of scintillation counters is measured. The experiment is performed in the following steps: One plastic scintillation detectors is assembled (scintillator plate, light guide and Photomultiplier) using light tight tape by the students. Output pulses are inspected with an oscilloscope and average pulse height vs. high voltage of the pulses is measured Gamma ray signals in the scintillators from Co 60 and Cs 137 sources are measured with an analog to digital converter, the observed pulseheight spectra are recordes and used to calibrate the energy response of the 2 scintillators. The high energy part of the electron spectrum of a Sr-Y beta source is measured and it s endpoint determined. The time resolution of a scintillation counter is measured using two photomultipliers and two different discrimination methods The spatial resolution of particle transition from time measurements is evaluated. Please prepare yourself before you start the experiment by reading the attached documentation and the printed document available in the FP ( Radiation detection and Monitoring ), looking through the attached list of questions (which you must be able to answer). This may require to also have a look at the literature given at the end. Books can be found in the library of the FP. All measurements have to be documented in the log book. Also the spectra which you obtain should be printed out on the network printer and glued to into the logbook. Finally all results and evaluations have to be in this logbook.inbook 1
2 Practical instructions to perform the experiment: 1. wrapping of scintillators and connection to PM s Three scintillators are used: Thick scintillator: 9 mm thick Thin scintillator: 5 mm thick Scintillator for time measurements 20 cm long, 8 mm thick. Never touch the scintillator (and also not the light guide ) with bare hands use cotton tissues or gloves. Fat finger prints will lead to cracks in the surface of the scintillator. This destroys the total reflection of light. Have a look at the scintillator pieces you will probably see these effects due to mistreatment by your predecessors. Take care that the black tape will not stick to the scintillator or the light guide again this would destroy total reflection. There should be no air gap between the scintillator and the light guide and also not between the light guide and the PM cathode. A gap would lead to light losses. If the gap changes during measurement then the counter efficiency will change which wold lead to an unstable calibration.. Scintilator Lightguide Photomultiplier Socket HV divider Teflon tape Never take apart! Black tape 1st layer Tape all pieces together Figure 1 Black tape second layer signal HV 2
3 You are expected to wrap the thick scintillator detector yourself as shown in the figure above. This gives you a chance to see all pieces of the detector and also to learn something about light guiding. In addition some practical work should be fun to you. The other detectors are expected to be available fully assembled. All three pieces scintillator, light guide and PM - have to be wrapped first by the white teflon tissue tape. Each winding should overlap in area with the old one by about 1/3. What is the purpose of this tape? Next you have to wrap the pieces with the black scotch tape twice to make the counter light tight. The second layer is wound with opposite winding angle with respect to the first layer as shown on the drawing. Finally connect the three pieces with black scotch tape. Make sure that there is good mechanical contact at all surfaces to avoid air gaps which lead to a loss of light. At the end the counter should be mechanically stable and absolutely light tight, else you will not be able to measure small count rates. 2. First look at output the PM pulses vs. high voltage Set the potentiometers (voltage) of the HV power supply to zero. Put the thick counter to the black box (presently a cartoon box) and use the Co 60 source to illuminate it. Then connect the HV cable of the thick counter to one output of the HV power supply. Connect the output of the PM to the scope, switch on the HV and raise it. Starting at around -600 V you should start to see pulses. This is the moment when you should make yourself familiar with the scope and its various options which you will need: Oscilloscope operation and use The oscilloscope is an extremely valuable instrument. Never use a signal before you have seen it on the oscilloscope! You must be sure that the signal has the right shape, amplitude and rate. You better spend some time to make yourself knowlegable how to use the scope, this is not trivial the scope has a lot of knobs and options. You must be familiar with the following operation steps so try out all these operations: Selection of display channel Selection of time axis (horizontal sweep)- always look to the written screen information. Example: H: 20 ns. This means 1 cm corresponds to 20 ns. 3
4 Selection of vertical axis (voltage) for the selected channel. Screen display : example CH1 20 mv. This means 1 cm corresponds to 20 mv. Always use the full scale to view the pulses to have good resolution. Selection of trigger source (CH1 or CH2) and adjustment of trigger Threshold Switch on the persistance option which allows you to see many pulses simultaneously on the scope. The persistance time can be regulated with the turning knob at the top of the front panel. Learn how to go to single pulses again. How to display two signals simultaneously? You can independtly choose vertical and horizontal axes for the two input channels select the channel first before you move the knobs! The scope has additional useful options: e.g. determination of average pulse height over 256 pulses this is very useful for the first measurement: pulse height vs. HV. The input cables of the oscilloscope have to be terminated by 50 Ohm ( the impedance of the coaxial cable)! This is done by using a T- switch with a 50 Ohm resistor on one end and the cable at the other. This serves two purposes: a) the resistor acts as working resistor which provides the input voltage. b) termination with 50 Ohms avoids reflections of the pulses in the cable. (for experts: the basis of the PM has no working resistor see the corresponding circuit diagam, this resistor has therefore to be provided independently. In principle you could use the input resistor of the scope but this works only with AC coupling). PM Input scope 50 Ω The output pulses are expected with a typical height of ~20 mv and a pulse length of ~20 ns. Measure the average pulse height vs. HV in the range between -900 V and V and record it in form of a table in the logbook. Why does it depend so strongly on the HV? Make a graph of the result into the logbook by hand. Choose a working voltage In this range and keep it for the rest of the measurements! (block the potentiometer). For this measurement you can profit very much by using the averaging mode of the scope.note: the average pulse height depends on the threshold! Make a sketch of the pulse into the logbook with time and pulseheight in- 4
5 formation. You will have to find this kind of pulses again and again in the course of the 4 half days. Take away the source (screw it back to its absorber block ) and measure the dark pulse rate of the scintillator and their average pulse height. This rate should be low, else you have a light leak! Repeat all steps for the thin scintillator. At the end of the first half day these measurements and the graphs have to be in both logbooks! 3. calibration of the energy measured in the scintillator with Co 60 and Cs 137 sources Co 60 and Cs 137 are two commonly used γ refence sources. You find the decay schemes in the attached document Radiation Detection and Measurement by G.F. Knoll of which you can get a paper copy in the FP. Co 60 has two γ lines at and MeV Cs 137 has one γ line at 0.66 MeV This energy range is the most difficult one for spectroscopy since the dominant interaction process of γ s in this energy range is Compton scattering which leads to a continuum of electron enegies and therefore to a continuous pulse height spectrum up to the compton edge (see section theoretical background and appendix1. The energy calibration is therefore based on the evaluation of the measured pulse height spectrum and the determination of the Compton edge. You are expected to calibrate both the thick and the thin scintillator detector. The experimental steup is shown in figure 2. Put the thick scintillator and the Co60 source into the box. Switch on the crate! Connect the PM output cable to the input of the Amplifier in the NIM crate. Choose the right polarity of the input pulse! (is it positive or negative?). Choose an amplification of 100 (turning knob on top of module). Look at the output pulses on the scope using the Uni(polar) Output. Pulse length and pulse form is now very different! Find the best scope settings and make a sketch of the pulses in the logbook indicating also pulse height and pulse length. If you are satisfied with the pulses and rates then connect the output cable of the amplifier to the ADC input at the back of the PC ( the input card is labelled with ADC). This is the time to make yourself familiar with the data taking and spectroscopy program. 5
6 HV1 HV2 approx V Co and Cs sources for calibration of the energy scale measured by ADC Sr Y source for measurement of energy deposition Scintilator Sr Y source PMT2 ADC Amplifier shaper PC with spectroscopic program A=100 shaping 1micro s Peak amplitude proportional to the integral of original pulse Figure 2 Data taking program and multi channel analyser Make yourself familiar with the data taking program and its important options: The toolbar on top allows you to start the most important options. Click fields are given below in red color. Datei Datenquelle öffnen. In the window choose either detector (click on PHA) to start reading in detector data ( or choose CAM to read in a data file from disk ). Click on VKA (multi channel analyser). Look at measurement settings: you can predefine the measurement time (what is the difference between live and total?), the number of channels in the display... In the tool bar VIEW you can choose linear or log vertical scale, in the submenue View->OPTIONS you can choose different ways to show the histogram. Recommended: draw line but try it out. Start the measurement- you will see a live update of the histogram The measurement will stop after the preselected number of second or you can stop it by hand. Whenever you have to subtract background from a 6
7 spectrum you must make sure that data is recorded over exactly the same time period. If you have recorded a useful spectrum which you want to analyse later then always store it in a folder with a meaningful name on the PC. Data files are recorded in the folder F80. Create a new folder with your name and store your measured files in this folder. (Datei speichern) Please create a new folder (best with your name) on the PC: Got to folder F80, click on Datei Neu and insert your name. In furture store your files in this ew folder. You can later read in the stored histograms again and edit or manipulate them click on Datei ( Öffnen einer Datenquelle ( click on CAM Open you file and read in your spectrum again. Now you can subtract two spectra from each other if needed ( click on Analysis). You can also edit your spectrum. It is recommended to go to Options and use the possibility to smooth the histogram ( glätten) which essentially means averageing several neighboring bins. This will allow to fit lines to the histogram much easier later. (Warning: it may happen that your spectrum gets distroyed in this step, just read it in again and try with new settings). Use the option ZOOM : VIEW ZOOM and select the part of the spectrum which you want to use. Especially useful if you have to determine the end point of a spectrum. But print out always both the full spectrum and the zoomed part. You will see both the original and the zoomed spectrum (this works also during data recording). Print out the spectrum using normally a log scale : If a measurement is finished or if you have read back a stored measured histogram and if you are content with the histogram display on the screen choose the best vertical display window by the scrollbar at the side of the window then copy the histogram to the Zwischenablage (intermediate copy) by ALT COPY on the keyboard. Open the open source drawing program or Word and paste the histogram to a new window. Add a title to the histogram (this is the only safe way not to mix up printed histograms) and print it out. The printout will appear on the network printer on the same floor. 3.1 Start the Co 60 measurement with you preferred VKA settings for a time period of about 1000 seconds. Choose the best plot display, store the histogram on the PC and print it out two times 1 copy for each student. 3.2 repeat all steps for the Cs 137 source. 3.3 Repeat all steps for the thin scintillator detector using both sources. 7
8 Analysis of the printed plots: this has to happen before you go to the next measurement: Energy spectrum of beta Sr-Y source, latest end of the second half day. All results have to appear in both logbooks, the plots have to be glued into the log book. The analysis of the plots has to be done independently by each student! Have a look at the measured logarithmic pulse height spectra which shows the number of entries vs. channel number. What is the interpretation of these spectra? Have a look again at appendix 1 where the ideal energy spectra are schematically shown. Determine the Compton edge usig.the printed log plot in the following way: Roughly the spectrum near the end can be described by two straight lines, one describing the flat top and one the falling edge of the spectrum. So draw two lines to the spectrum and determine at which channel number they intersect. You can approximately identify the intersection point with the Compton edge (why?). Calculate the electron energy at the Compton edge using the known gamma energy and the Compton formula (theory section). This gives you a calibration point energy/channelnumber. Try to estimate the error on this calibration point. (we don t expect a fit to the data, though you of course can do that if you export the spectrum to your CIP pool account via network- ask the assistant how to do that). For both scintillators separately: Draw by hand the calibration points ( energy vs. channel number) into a plot in the log book. Determine a straight line through these points using also the origin as a constraint - we expect channel number zero for zero energy. Note the energy calibration constants for both detectors in a box. This is the final result of this part of the measurement. You are encouraged to add a few remarks. For ambitious experts: once you have an energy calibration from these measurements ( and if you are sufficiently clever) then you can insert it into the data taking program and measure further histograms using energy in kev directly on the horizontal axis. 4. Measurement of the beta spectrum of a Sr-Y source. The decay electron spectrum is composed of 2 beta spectra: Sr 90 with an endpoint of MeV and the spectrum of the daughter nucleus Y 90 with an endpoint of MeV. 8
9 Both are 'allowed' Beta decays e.g. the form of the electron spectrum is given by dn(e)/de ~ (E+m e c 4 )* SQRT(E 2 +2m e Ec 2 )* (Emax-E) 2 where E is the measured kinetic energy of the electron and E max the maximal kinetic energy of the spectrum. (see any textbook of Nuclear Physics). A measured spectrum is shown in figure 3. Figure 3 You are now expected to measure this spectrum and to determine the end point energy with the two scintillators using their energy calibration. Principle of detection: the decay electrons traverse the scintillators and loose energy by ionisation. The light output of the scintillator is proportional to the ionisation. If the electrons are therefore stopped in the scintillators you will be able to measure their kinetic energy. Check therefore the range of the decay electrons in the scintillator material using the range curve of figure 4. What is the maximum range for electrons near the end point? What do you expect for the pulseheight in the thick and thin scintillators? 9
10 100 Range of atomic particles in Premium plastic scintillator 10 e Range (mm) µ d α Energy (MeV) figure put the thick scintillator detector next to the Sr-Y source and the thin detector behind it and switch their HV on. Look at the shaped amplifier output pulses simultaneously on the scope by triggering on the thick scintillator. Do you see simultaneous pulses in both scintillators? What do you expect? Note down your observations! Measure the spectrum of the thick scintillator for a fixed time of ~1200 s and store it. Take off the Sr source and repeat the measurement under identical conditions. This determines the background spectrum and store it. Repeat both steps for the thin scintillator. Note:the measurement of the background spectrum has to be done under identical conditions and very near in time to exclude pulseheight trips e.g. due to temperature changes. 4.2: evaluate the end point of Sr-Y in an approximate way before you start the next measurement: 10
11 In a first step you have to subtract the background spectrum from the spectrum with source for both scintillators. This is done by loading the source spectrum and displaying it in the best way. Recommended: display also a ZOOMed spectrum in the region near to the end point. Then go to Analysis subtract. What is the origin of the background? Why does it disturb this measurement? Please print out both the original spectrum and the subtracted spectrum Try to extrapolate the subtracted energy spectrum to the endpoint. This is best done in a ZOOMed spectrum near the endpoint in a log plot. What form of the spectrum do you expect? Determine the end point energy graphically and estimate its uncertainty. Compare to the expected value. NOTE: If the spectrum of the second scintillator after subtraction shows significant energy then you have to add this energy to determine the end point energy! 4.3: repeat all steps by interchanging the thin and the thick scintillator. What do you now expect for the relative pulse heights of the two scintillators? Now you will have to add the end point energies of the two spectra to get an energy measurement of the end point. The measurement and its analysis have to be finished before you start the last part measurement of time resolution. This is before the last half day! 5. measurement of the time resolution of a scintillation detector A longer scintillator (20 cm) is available which is read out by two photomultipliers, one on each end. Discriminators are used to derive a norm pulse (discriminator output) whose leading edge gives the time stamp for the arrival of the scintillator pulse. To measure the time difference between the two norm pulses we use a Time to Amplitude Converter (TAC). This is the 3 unit wide insert in the crate. One signal is used to start the time measurement, the other is used to stop it. (Use the corresponding inputs for these signals. BUT: The stop signal has to come after the start signal under all circumstances! This is done by delaying the stop signal using the delay boxes in the crate. Use the Sr-Y source for this measurement. 11
12 You are expected to use two type of discriminators to derive the norm pulses for the time measurement. They use different ways to determine the arrival time of a pulse. 5.1 Time resolution using a leading edge discriminator The experimental setup is shown in figure 5. In a leading edge discrimination, the time pulse is generated when the input pulse exceeds a fixed threshold. Pulses of different height lead to time stamps which depend on the pulse height - high pulses cross the threshold earlier than low pulses. HV1 HV2 PMT1 Scintilator Sr Y source PMT2 min. 50 mv min 50 mv Discriminator Constant fraction or Leading edge Adjustable delay approx. 16 ns Discriminator Constant fraction or Leading edge start stop Time Amplitude Convertor (TAC) out Use 2 delays to calibrate the time measurement using the ADC ADC PC with spectroscopic program Figure Put the source to the middle of the scintillator the two PM s should see the light pulses simultaneously. Check this by looking at both output pulses simultaneously on the scope. Connect the two scintillator 12
13 signals to the inputs channels of the leading edge discriminator module. Connect the OUT channels to the two scope inputs. Trigger on channel 1 and look at both channels simultaneously using persistence display. The pulses of channel two are fluctuating in time they show a time jitter. Read off the width of the jitter- it is a measure of the time resolution. make a sketch in the logbook including time indication connect the discriminator output of one channel to the START input of the TAC module. (see figure above). Connect the discriminator of the second pulse to the input of the delay box 1. Connect the output of delay box 1 with the input of delay box two. Connect the output of delay box 2 to the stop input of the DAQ module. Set all delay buttons to OFF (no delay). Look at the TAC output pulse on the scope. Add delays by switching the buttons in the delay boxes until you see a sizeable pulse. Connect the TAC output to the ADC input at the PC and start a measurement. Choose full measurement range in the VKA menu. Add or subtract delays until the measured pulses appear in the middle of the measurement range. This is now defined as zero time difference. 5.2 Measure the time resolution of the setup. Take a spectrum for at least 600 s. The width of this spectrum measures the time resolution! Store the spectrum (uses ZOOM and linear scale). Print it out twice, one copy for each student. 5.3 Calibrate the time ( time difference vs. channel number) This is easily done by taking off or adding known delays to the stop pulse in the delay boxes. Add 4 ns delay. Do not clear the old measurement (do not click on löschen but click on start. You will see a new peak coming up at a higher channel number!. The difference in channel number of the two peak positions corresponds to a time difference of 4 ns!. Repeat the same for -4 ns and choose other delays which look reasonable to you. You must have a calibration point near the observed time resolution! Print out the spectrum (after writing the time delays into the graph!) and determine the peak positions of all peaks. Use the original setting as corresponding to time difference zero. Make a table: delay in ns vs. the channel number and plot the calibration curve into the logbook by hand. Evaluate the calibration constant from the slope of a straight line. Determine the time resolution (in ns) by measuring the Full Width Half Maximum ( FWHM)of the measured time spectrum. How is this measure related to the standard deviation of a Gaussian distribution? 13
14 5.4 Evaluation of signal velocity in a scintillation detector The light pulses generated by an electron in the scintillator propagate through the scintillator if they stay within the total reflection angle. They have of course a finite velocity. Try to estimate the travelling time over a length of 20 cm. Is this difference measureable with the present setup? Put the source near the end of the scintillator (not the light guide!). measure its position relative to the middle and record it in the logbook. Take a spectrum over 600 s. Evaluate the time difference relative to the middle position. Repeat the same step by putting the source to the other end of the scintillator. Determine the velocity of light in the scintillator and explain the result. 5.5 time resolution using a constant fraction discriminator. Repeat all steps 5.1- to 5.4 by using the constant fraction discriminator. Is it necessary to repeat the time calibration? Constant fraction discrimination: the time stamp is set when a constant fraction of the integrated pulse (const. charge fraction) has been detected. This pulse stamp is much less dependent on the pulseheight and should therefore give much better time resolution Compare the two ways of discrimination! The difference should already be visible on the scope. Summarise the results of the time measurements in a short table in the logbook and discuss the results including propagation speed. 5.7 Spatial resolution from a time of flight measurement. Very often the fast scintillation detectors are used to measure time of flight for a particle which together with a momentum measurement allows then to determine the mass of the particle. What is the arrival time difference of a pion and a kaon of identical momentum of p=1 GeV/c if their flight path has a length of 1.5 m? How well can you separate the pion from the kaon if you insert the time resolution of your constant fraction measurement? 14
15 Calculate the difference and evaluate the separation in the log book. The masses are : m π = MeV, m K = (hint: use β=v/c=p/e, where E is the total energy of the particle ). In long scintillators which are read out on both sides the position of the particle traversal can be determined by measuring the time difference of the arrival times in the two PM s. Use the measured time differences for the three source positions of your measurements with the constant fraction discriminators to determine the spatial resolution of such a measurement e.g. how well could you determine the position of the source if you measure the time difference in the two PM s. These evaluations should be finished before the end of the last half day. So you can get the final Testat on the fourth half day. Wish you fun and success during the experiment Good luck 15
16 16
Fast first practical help -- detailed instructions will follow- preliminary Experiment F80
Fast first practical help -- detailed instructions will follow- preliminary Experiment F80 Measurement Methods of Nuclear and Particle Physics Introduction: This experiment is going to introduce you to
More informationPH2510 Nuclear Physics Laboratory Use of Scintillation Counters (NP5)
Physics Department Royal Holloway University of London PH2510 Nuclear Physics Laboratory Use of Scintillation Counters (NP5) 1. Introduction 1.1 Object of the Experiment The object of this experiment is
More informationGamma Ray Spectroscopy with NaI(Tl) and HPGe Detectors
Nuclear Physics #1 Gamma Ray Spectroscopy with NaI(Tl) and HPGe Detectors Introduction: In this experiment you will use both scintillation and semiconductor detectors to study γ- ray energy spectra. The
More informationInstructions for gg Coincidence with 22 Na. Overview of the Experiment
Overview of the Experiment Instructions for gg Coincidence with 22 Na 22 Na is a radioactive element that decays by converting a proton into a neutron: about 90% of the time through β + decay and about
More informationGAMMA-GAMMA CORRELATION Latest Revision: August 21, 2007
C1-1 GAMMA-GAMMA CORRELATION Latest Revision: August 21, 2007 QUESTION TO BE INVESTIGATED: decay event? What is the angular correlation between two gamma rays emitted by a single INTRODUCTION & THEORY:
More informationORTEC Experiment 13. Gamma-Gamma Coincidence with Angular Correlation. Equipment Required
ORTEC Experiment 13 Equipment Required Two 905-3 2-in. x 2-in. NaI(Tl) Scintillation Detector Assemblies. Two 266 Photomultiplier Tube Bases. Two 113 Scintillation Preamplifiers. Two 556 High Voltage Power
More informationK 223 Angular Correlation
K 223 Angular Correlation K 223.1 Aim of the Experiment The aim of the experiment is to measure the angular correlation of a γ γ cascade. K 223.2 Required Knowledge Definition of the angular correlation
More informationExperiment 10. The Speed of Light c Introduction Apparatus
Experiment 10 The Speed of Light c 10.1 Introduction In this experiment you will measure the speed of light, c. This is one of the most fundamental constants in physics, and at the same time the fastest
More informationCOMPTON SCATTERING. Purpose. Introduction. Fundamentals of Experiment
COMPTON SCATTERING Purpose The purpose of this experiment is to verify the energy dependence of gamma radiation upon scattering angle and to compare the differential cross section obtained from the data
More informationLIFETIME OF THE MUON
Muon Decay 1 LIFETIME OF THE MUON Introduction Muons are unstable particles; otherwise, they are rather like electrons but with much higher masses, approximately 105 MeV. Radioactive nuclear decays do
More informationVirtual Laboratory of Nuclear Fission Virtual practicum in the framework of the project Virtual Laboratory of Nuclear Fission
Virtual Laboratory of Nuclear Fission Virtual practicum in the framework of the project Virtual Laboratory of Nuclear Fission Khanyisa Sowazi, University of the Western Cape JINR SAR, September 2015 INDEX
More informationPhysics Laboratory Scattering of Photons from Electrons: Compton Scattering
RR Oct 2001 SS Dec 2001 MJ Oct 2009 Physics 34000 Laboratory Scattering of Photons from Electrons: Compton Scattering Objective: To measure the energy of high energy photons scattered from electrons in
More informationElectronic Instrumentation for Radiation Detection Systems
Electronic Instrumentation for Radiation Detection Systems January 23, 2018 Joshua W. Cates, Ph.D. and Craig S. Levin, Ph.D. Course Outline Lecture Overview Brief Review of Radiation Detectors Detector
More informationSilicon Photomultiplier Evaluation Kit. Quick Start Guide. Eval Kit SiPM. KETEK GmbH. Hofer Str Munich Germany.
KETEK GmbH Hofer Str. 3 81737 Munich Germany www.ketek.net info@ketek.net phone +49 89 673 467 70 fax +49 89 673 467 77 Silicon Photomultiplier Evaluation Kit Quick Start Guide Eval Kit Table of Contents
More informationPhoton Counters SR430 5 ns multichannel scaler/averager
Photon Counters SR430 5 ns multichannel scaler/averager SR430 Multichannel Scaler/Averager 5 ns to 10 ms bin width Count rates up to 100 MHz 1k to 32k bins per record Built-in discriminator No interchannel
More informationPh 3324 The Scintillation Detector and Gamma Ray Spectroscopy
Ph 3324 The Scintillation Detector and Gamma Ray Spectroscopy Required background reading Attached are several pages from an appendix on the web for Tipler-Llewellyn Modern Physics. Read the section on
More informationEKA Laboratory Muon Lifetime Experiment Instructions. October 2006
EKA Laboratory Muon Lifetime Experiment Instructions October 2006 0 Lab setup and singles rate. When high-energy cosmic rays encounter the earth's atmosphere, they decay into a shower of elementary particles.
More informationand N(t) ~ exp(-t/ ),
Muon Lifetime Experiment Introduction Charged and neutral particles with energies in excess of 10 23 ev from Galactic and extra Galactic sources impinge on the earth. Here we speak of the earth as the
More informationAtomic and Nuclear Physics
Atomic and Nuclear Physics Nuclear physics -spectroscopy LEYBOLD Physics Leaflets Detecting radiation with a scintillation counter Objects of the experiments Studying the scintillator pulses with an oscilloscope
More informationAn ASIC dedicated to the RPCs front-end. of the dimuon arm trigger in the ALICE experiment.
An ASIC dedicated to the RPCs front-end of the dimuon arm trigger in the ALICE experiment. L. Royer, G. Bohner, J. Lecoq for the ALICE collaboration Laboratoire de Physique Corpusculaire de Clermont-Ferrand
More informationPHYSICS ADVANCED LABORATORY I COMPTON SCATTERING Spring 2002
PHYSICS 334 - ADVANCED LABORATORY I COMPTON SCATTERING Spring 00 Purposes: Demonstrate the phenomena associated with Compton scattering and the Klein-Nishina formula. Determine the mass of the electron.
More informationORTEC Experiment 1. Introduction to Electronic Signal Analysis in Nuclear Radiation Measurements. Equipment Required: Purpose. Electronic Circuits
ORTEC Experiment 1 Equipment Required: 480 Pulser 113 Scintillation Preamplifier 4001A/4002D NIM Bin and Power Supply 575A Spectroscopy Amplifier 996 Timer and Counter 551 Timing Single-Channel Analyzer
More informationPhysics 342 Laboratory. Scattering of Photons from Free Electrons: Compton Scattering
RR Oct 2001 SS Dec 2001 Physics 342 Laboratory Scattering of Photons from Free Electrons: Compton Scattering Objective: To measure the energy of high energy photons scattered from electrons in a brass
More informationORTEC. AN34 Experiment 14 Nuclear Lifetimes and the Coincidence Method. Equipment Needed from ORTEC. Equipment Required from Other Manufacturers
Equipment Needed from ORTEC Two 113 Scintillation Preamplifiers Two 266 Photomultiplier Tube Bases 4001A/4002D Bin and Power Supply 414A Fast Coincidence Two 551 Timing Single-Channel Analyzers 567 Time-to-Amplitude
More information1 Purpose of This Lab Exercise:
Physics 4796 - Experimental Physics Temple University, Spring 2010-11 C. J. Martoff, Instructor J. Tatarowicz, TA Physics 4796 Lab Writeup Hunting for Antimatter with NaI Spectroscopy 1 Purpose of This
More informationTesting of the NSC Electronics Module with the GSI Clover Detector
Testing of the NSC Electronics Module with the GSI Clover Detector Rakesh Kumar 1, P. Queiroz 2, H.-J. Wollersheim 2 (Tutor) 1 Inter University Accelerator Centre Aruna Asaf Ali Marg Post Box No 10502
More informationThe (Speed and) Decay of Cosmic-Ray Muons
The (Speed and) Decay of Cosmic-Ray Muons Jason Gross MIT - Department of Physics Jason Gross (8.13) Cosmic-Ray Muons November 4, 2011 1 / 30 Goals test relativity (time dilation) determine the mean lifetime
More informationOscilloscope Measurements
PC1143 Physics III Oscilloscope Measurements 1 Purpose Investigate the fundamental principles and practical operation of the oscilloscope using signals from a signal generator. Measure sine and other waveform
More informationCAEN Tools for Discovery
Viareggio 5 September 211 Introduction In recent years CAEN has developed a complete family of digitizers that consists of several models differing in sampling frequency, resolution, form factor and other
More informationEnergy Measurements with a Si Surface Barrier Detector and a 5.5-MeV 241 Am α Source
Energy Measurements with a Si Surface Barrier Detector and a 5.5-MeV 241 Am α Source October 18, 2017 The goals of this experiment are to become familiar with semiconductor detectors, which are widely
More informationOptical Pumping Control Unit
(Advanced) Experimental Physics V85.0112/G85.2075 Optical Pumping Control Unit Fall, 2012 10/16/2012 Introduction This document is gives an overview of the optical pumping control unit. Magnetic Fields
More informationPh 3455 The Franck-Hertz Experiment
Ph 3455 The Franck-Hertz Experiment Required background reading Tipler, Llewellyn, section 4-5 Prelab Questions 1. In this experiment, we will be using neon rather than mercury as described in the textbook.
More informationHomework Set 3.5 Sensitive optoelectronic detectors: seeing single photons
Homework Set 3.5 Sensitive optoelectronic detectors: seeing single photons Due by 12:00 noon (in class) on Tuesday, Nov. 7, 2006. This is another hybrid lab/homework; please see Section 3.4 for what you
More informationTraditional analog QDC chain and Digital Pulse Processing [1]
Giuliano Mini Viareggio April 22, 2010 Introduction The aim of this paper is to compare the energy resolution of two gamma ray spectroscopy setups based on two different acquisition chains; the first chain
More informationScintillation Counters
PHY311/312 Detectors for Nuclear and Particle Physics Dr. C.N. Booth Scintillation Counters Unlike many other particle detectors, which exploit the ionisation produced by the passage of a charged particle,
More informationORTEC Experiment 3. Gamma-Ray Spectroscopy Using NaI(Tl) Equipment Required. Purpose. Gamma Emission
ORTEC Experiment 3 Equipment Required Electronic Instrumentation o SPA38 Integral Assembly consisting of a 38 mm x 38 mm NaI(Tl) Scintillator, Photomultiplier Tube, and PMT Base with Stand o 4001A/4002D
More informationAttenuation length in strip scintillators. Jonathan Button, William McGrew, Y.-W. Lui, D. H. Youngblood
Attenuation length in strip scintillators Jonathan Button, William McGrew, Y.-W. Lui, D. H. Youngblood I. Introduction The ΔE-ΔE-E decay detector as described in [1] is composed of thin strip scintillators,
More informationPurpose This experiment will use the coincidence method for time correlation to measure the lifetime in the decay scheme of 57
Equipment Required Two 113 Scintillation Preamplifiers Two 266 Photomultiplier Tube Bases 4001A/4002D Bin and Power Supply 414A Fast Coincidence Two 551 Timing Single-Channel Analyzers 567 Time-to-Amplitude
More informationtotal j = BA, [1] = j [2] total
Name: S.N.: Experiment 2 INDUCTANCE AND LR CIRCUITS SECTION: PARTNER: DATE: Objectives Estimate the inductance of the solenoid used for this experiment from the formula for a very long, thin, tightly wound
More informationTesting the Electronics for the MicroBooNE Light Collection System
Testing the Electronics for the MicroBooNE Light Collection System Kathleen V. Tatem Nevis Labs, Columbia University & Fermi National Accelerator Laboratory August 3, 2012 Abstract This paper discusses
More informationOn the initiation of lightning in thunderclouds (Instrumentation, Supplementary information)
On the initiation of lightning in thunderclouds (Instrumentation, Supplementary information) Ashot Chilingarian 1,2, Suren Chilingaryan 1, Tigran Karapetyan 1, Lev Kozliner 1, Yeghia Khanikyants 1, Gagik
More informationFYSP1110/K1 (FYSP110/K1) USE OF AN OSCILLOSCOPE
FYSP1110/K1 (FYSP110/K1) USE OF AN OSCILLOSCOPE 1 Introduction In this exercise you will get basic knowledge about how to use an oscilloscope. You ll also measure properties of components, which you are
More informationSpontaneous Fission Spectrum of Neutrons from 252 Cf with Kinetic Energies Less than 1 MeV. Suraj Bastola. A senior thesis submitted to the faculty of
Spontaneous Fission Spectrum of Neutrons from 252 Cf with Kinetic Energies Less than 1 MeV Suraj Bastola A senior thesis submitted to the faculty of Brigham Young University in partial fulfillment of the
More informationPMT tests at UMD. Vlasios Vasileiou Version st May 2006
PMT tests at UMD Vlasios Vasileiou Version 1.0 1st May 2006 Abstract This memo describes the tests performed on three Milagro PMTs in UMD. Initially, pulse-height distributions of the PMT signals were
More informationNEEP 427 PROPORTIONAL COUNTERS. Knoll, Chapters 6 & 14 Sect. I & II
NEEP 427 PROPORTIONAL COUNTERS References: Knoll, Chapters 6 & 14 Sect. I & II a proportional counter the height of the output pulse is proportional to the number of ion pairs produced in the counter gas.
More informationMethod for digital particle spectrometry Khryachkov Vitaly
Method for digital particle spectrometry Khryachkov Vitaly Institute for physics and power engineering (IPPE) Obninsk, Russia The goals of Analog Signal Processing Signal amplification Signal filtering
More informationPhysics 131 Lab 1: ONE-DIMENSIONAL MOTION
1 Name Date Partner(s) Physics 131 Lab 1: ONE-DIMENSIONAL MOTION OBJECTIVES To familiarize yourself with motion detector hardware. To explore how simple motions are represented on a displacement-time graph.
More informationRadionuclide Imaging MII 3073 RADIONUCLIDE IMAGING SYSTEM
Radionuclide Imaging MII 3073 RADIONUCLIDE IMAGING SYSTEM Preamplifiers and amplifiers The current from PMT must be further amplified before it can be processed and counted (the number of electrons yielded
More informationECE 53A: Fundamentals of Electrical Engineering I
ECE 53A: Fundamentals of Electrical Engineering I Laboratory Assignment #1: Instrument Operation, Basic Resistor Measurements and Kirchhoff s Laws Fall 2007 General Guidelines: - Record data and observations
More informationHigh collection efficiency MCPs for photon counting detectors
High collection efficiency MCPs for photon counting detectors D. A. Orlov, * T. Ruardij, S. Duarte Pinto, R. Glazenborg and E. Kernen PHOTONIS Netherlands BV, Dwazziewegen 2, 9301 ZR Roden, The Netherlands
More informationPhysics Experiment N -17. Lifetime of Cosmic Ray Muons with On-Line Data Acquisition on a Computer
Introduction Physics 410-510 Experiment N -17 Lifetime of Cosmic Ray Muons with On-Line Data Acquisition on a Computer The experiment is designed to teach the techniques of particle detection using scintillation
More informationFilters And Waveform Shaping
Physics 3330 Experiment #3 Fall 2001 Purpose Filters And Waveform Shaping The aim of this experiment is to study the frequency filtering properties of passive (R, C, and L) circuits for sine waves, and
More informationNotes on Experiment #1
Notes on Experiment #1 Bring graph paper (cm cm is best) From this week on, be sure to print a copy of each experiment and bring it with you to lab. There will not be any experiment copies available in
More informationLab E2: B-field of a Solenoid. In the case that the B-field is uniform and perpendicular to the area, (1) reduces to
E2.1 Lab E2: B-field of a Solenoid In this lab, we will explore the magnetic field created by a solenoid. First, we must review some basic electromagnetic theory. The magnetic flux over some area A is
More informationName Date: Course number: MAKE SURE TA & TI STAMPS EVERY PAGE BEFORE YOU START EXPERIMENT 10. Electronic Circuits
Laboratory Section: Last Revised on September 21, 2016 Partners Names: Grade: EXPERIMENT 10 Electronic Circuits 1. Pre-Laboratory Work [2 pts] 1. How are you going to determine the capacitance of the unknown
More informationA digital method for separation and reconstruction of pile-up events in germanium detectors. Abstract
A digital method for separation and reconstruction of pile-up events in germanium detectors M. Nakhostin a), Zs. Podolyak, P. H. Regan, P. M. Walker Department of Physics, University of Surrey, Guildford
More information4: EXPERIMENTS WITH SOUND PULSES
4: EXPERIMENTS WITH SOUND PULSES Sound waves propagate (travel) through air at a velocity of approximately 340 m/s (1115 ft/sec). As a sound wave travels away from a small source of sound such as a vibrating
More informationScintillators as an external trigger for cathode strip chambers
Scintillators as an external trigger for cathode strip chambers J. A. Muñoz Department of Physics, Princeton University, Princeton, NJ 08544 An external trigger was set up to test cathode strip chambers
More informationSensor Calibration Lab
Sensor Calibration Lab The lab is organized with an introductory background on calibration and the LED speed sensors. This is followed by three sections describing the three calibration techniques which
More informationPHY 123/253 Shot Noise
PHY 123/253 Shot Noise HISTORY Complete Pre- Lab before starting this experiment In 1918, experimental physicist Walter Scottky working in the research lab at Siemens was investigating the origins of noise
More informationTrigger Rate Dependence and Gas Mixture of MRPC for the LEPS2 Experiment at SPring-8
Trigger Rate Dependence and Gas Mixture of MRPC for the LEPS2 Experiment at SPring-8 1 Institite of Physics, Academia Sinica 128 Sec. 2, Academia Rd., Nankang, Taipei 11529, Taiwan cyhsieh0531@gmail.com
More informationLaboratory Exercise 6 THE OSCILLOSCOPE
Introduction Laboratory Exercise 6 THE OSCILLOSCOPE The aim of this exercise is to introduce you to the oscilloscope (often just called a scope), the most versatile and ubiquitous laboratory measuring
More informationName EET 1131 Lab #2 Oscilloscope and Multisim
Name EET 1131 Lab #2 Oscilloscope and Multisim Section 1. Oscilloscope Introduction Equipment and Components Safety glasses Logic probe ETS-7000 Digital-Analog Training System Fluke 45 Digital Multimeter
More informationPositron Emission Tomography
Positron Emission Tomography UBC Physics & Astronomy / PHYS 409 1 Introduction Positron emission tomography (PET) is a non-invasive way to produce the functional 1 image of a patient. It works by injecting
More informationORTEC. Research Applications. Pulse-Height, Charge, or Energy Spectroscopy. Detectors. Processing Electronics
ORTEC Spectroscopy systems for ORTEC instrumentation produce pulse height distributions of gamma ray or alpha energies. MAESTRO-32 (model A65-B32) is the software included with most spectroscopy systems
More informationThe Influence of Crystal Configuration and PMT on PET Time-of-Flight Resolution
The Influence of Crystal Configuration and PMT on PET Time-of-Flight Resolution Christopher Thompson Montreal Neurological Institute and Scanwell Systems, Montreal, Canada Jason Hancock Cross Cancer Institute,
More informationVertical Tests of ILC Cavities and Detection of X-Rays from Field Emission
Vertical Tests of ILC Cavities and Detection of X-Rays from Field Emission Pardis Niknejadi California State Polytechnic University, Pomona, CA 91768 Elizabeth Olhsson University of Oregon, Eugene, OR
More informationExperiment 1.A. Working with Lab Equipment. ECEN 2270 Electronics Design Laboratory 1
.A Working with Lab Equipment Electronics Design Laboratory 1 1.A.0 1.A.1 3 1.A.4 Procedures Turn in your Pre Lab before doing anything else Setup the lab waveform generator to output desired test waveforms,
More informationBryn Mawr College Department of Physics Undergraduate Teaching Laboratories Alpha Particle Spectroscopy
Introduction Bryn Mawr College Department of Physics Undergraduate Teaching Laboratories Alpha Particle Spectroscopy This experiment is designed to study the detection of charged-particle radiation and
More informationLAB I. INTRODUCTION TO LAB EQUIPMENT
1. OBJECTIVE LAB I. INTRODUCTION TO LAB EQUIPMENT In this lab you will learn how to properly operate the oscilloscope Agilent MSO6032A, the Keithley Source Measure Unit (SMU) 2430, the function generator
More informationTektronix digital oscilloscope, BK Precision Function Generator, coaxial cables, breadboard, the crystal earpiece from your AM radio kit.
Experiment 0: Review I. References The 174 and 275 Lab Manuals Any standard text on error analysis (for example, Introduction to Error Analysis, J. Taylor, University Science Books, 1997) The manual for
More informationDevelopment of large readout area, high time resolution RPCs for LEPS2 at SPring-8
Development of large readout area, high time resolution RPCs for LEPS2 at SPring-8 1 Department of physics, Kyoto University Kitashirakawa-Oiwakecho, Sakyo-ku, Kyoto 606-8502, Japan E-mail: natsuki@scphys.kyoto-u.ac.jp
More informationRange of Alpha Particles in Gas (note, this is abridged from full Nuclear Decay laboratory file)
University of Illinois at Urbana-Champaign Physics 403 Laboratory Department of Physics Range of Alpha Particles in Gas (note, this is abridged from full Nuclear Decay laboratory file) 1. References 1.
More informationExperiment P49: Transistor Lab 2 Current Gain: The NPN Emitter-Follower Amplifier (Power Amplifier, Voltage Sensor)
PASCO scientific Vol. 2 Physics Lab Manual: P49-1 Experiment P49: Transistor Lab 2 Current Gain: The NPN Emitter-Follower Amplifier (Power Amplifier, Voltage Sensor) Concept Time SW Interface Macintosh
More informationSigCal32 User s Guide Version 3.0
SigCal User s Guide . . SigCal32 User s Guide Version 3.0 Copyright 1999 TDT. All rights reserved. No part of this manual may be reproduced or transmitted in any form or by any means, electronic or mechanical,
More informationLab E5: Filters and Complex Impedance
E5.1 Lab E5: Filters and Complex Impedance Note: It is strongly recommended that you complete lab E4: Capacitors and the RC Circuit before performing this experiment. Introduction Ohm s law, a well known
More informationCosmic Ray Muon Detection
Cosmic Ray Muon Detection Department of Physics and Space Sciences Florida Institute of Technology Georgia Karagiorgi Julie Slanker Advisor: Dr. M. Hohlmann Cosmic Ray Muons π + > µ + + ν µ π > µ + ν µ
More informationPhysics 4C Chabot College Scott Hildreth
Physics 4C Chabot College Scott Hildreth The Inverse Square Law for Light Intensity vs. Distance Using Microwaves Experiment Goals: Experimentally test the inverse square law for light using Microwaves.
More informationThe 34th International Physics Olympiad
The 34th International Physics Olympiad Taipei, Taiwan Experimental Competition Wednesday, August 6, 2003 Time Available : 5 hours Please Read This First: 1. Use only the pen provided. 2. Use only the
More informationExperiment P01: Understanding Motion I Distance and Time (Motion Sensor)
PASCO scientific Physics Lab Manual: P01-1 Experiment P01: Understanding Motion I Distance and Time (Motion Sensor) Concept Time SW Interface Macintosh file Windows file linear motion 30 m 500 or 700 P01
More informatione t Development of Low Cost γ - Ray Energy Spectrometer
e t International Journal on Emerging Technologies (Special Issue on NCRIET-2015) 6(2): 315-319(2015) ISSN No. (Print) : 0975-8364 ISSN No. (Online) : 2249-3255 Development of Low Cost γ - Ray Energy Spectrometer
More information3.003 Lab 3 Part A. Measurement of Speed of Light
3.003 Lab 3 Part A. Measurement of Speed of Light Objective: To measure the speed of light in free space Experimental Apparatus: Feb. 18, 2010 Due Mar. 2, 2010 Components: 1 Laser, 4 mirrors, 1 beam splitter
More informationThe Speed of Light Laboratory Experiment 8. Introduction
Exp-8-Speed of Light.doc (TJR) Physics Department, University of Windsor 64-311 Laboratory Experiment 8 The Speed of Light Introduction Galileo was right. Light did not travel instantaneously as his contemporaries
More informationUser Guide. SIB Channel MAPMT Interface Board Hamamatsu H7546 series
SIB164-1018 64 Channel MAPMT Interface Board Hamamatsu H7546 series Disclaimer Vertilon Corporation has made every attempt to ensure that the information in this document is accurate and complete. Vertilon
More informationExperiment 8: An AC Circuit
Experiment 8: An AC Circuit PART ONE: AC Voltages. Set up this circuit. Use R = 500 Ω, L = 5.0 mh and C =.01 μf. A signal generator built into the interface provides the emf to run the circuit from Output
More informationContents. Why waveform? Waveform digitizer : Domino Ring Sampler CEX Beam test autumn 04. Summary
Contents Why waveform? Waveform digitizer : Domino Ring Sampler CEX Beam test data @PSI autumn 04 Templates and time resolution Pulse Shape Discrimination Pile-up rejection Summary 2 In the MEG experiment
More informationBasic Electronics Learning by doing Prof. T.S. Natarajan Department of Physics Indian Institute of Technology, Madras
Basic Electronics Learning by doing Prof. T.S. Natarajan Department of Physics Indian Institute of Technology, Madras Lecture 26 Mathematical operations Hello everybody! In our series of lectures on basic
More informationClass #9: Experiment Diodes Part II: LEDs
Class #9: Experiment Diodes Part II: LEDs Purpose: The objective of this experiment is to become familiar with the properties and uses of LEDs, particularly as a communication device. This is a continuation
More informationFinal Report Data Acquisition Box
Final Report Data Acquisition Box By Gene Bender DeSmet Jesuit High School July 25, 2003 Contents Overview...2 A Hybrid LBNL Cosmic Ray Detector...2 The Detectors...6 Changing PMT Voltage...7 Comparator
More informationSensor Calibration Lab
Sensor Calibration Lab The lab is organized with an introductory background on calibration and the LED speed sensors. This is followed by three sections describing the three calibration techniques which
More informationExperiment 1: Fraunhofer Diffraction of Light by a Single Slit
Experiment 1: Fraunhofer Diffraction of Light by a Single Slit Purpose 1. To understand the theory of Fraunhofer diffraction of light at a single slit and at a circular aperture; 2. To learn how to measure
More informationLab 1: Pulse Propagation and Dispersion
ab 1: Pulse Propagation and Dispersion NAME NAME NAME Introduction: In this experiment you will observe reflection and transmission of incident pulses as they propagate down a coaxial transmission line
More informationAN ABSTRACT ON THE THESIS OF. David C. Vasquez for the degree of Master of Science in Radiation Health Physics presented on February 26, 2010.
AN ABSTRACT ON THE THESIS OF David C. Vasquez for the degree of Master of Science in Radiation Health Physics presented on February 26, 2010. Title: The Design, Use and Implementation of Digital Radiation
More informationPARTICLE DETECTORS (V): ELECTRONICS
Monday, April 13, 2015 1 PARTICLE DETECTORS (V): ELECTRONICS Zhenyu Ye April 13, 2015 Monday, April 13, 2015 2 References Techniques for Nuclear and Particle Physics Experiments by Leo, Chapter 15-17 Particle
More informationA Modular Readout System For A Small Liquid Argon TPC Carl Bromberg, Dan Edmunds Michigan State University
A Modular Readout System For A Small Liquid Argon TPC Carl Bromberg, Dan Edmunds Michigan State University Abstract A dual-fet preamplifier and a multi-channel waveform digitizer form the basis of a modular
More informationLLS - Introduction to Equipment
Published on Advanced Lab (http://experimentationlab.berkeley.edu) Home > LLS - Introduction to Equipment LLS - Introduction to Equipment All pages in this lab 1. Low Light Signal Measurements [1] 2. Introduction
More informationORTEC Experiment 19. Gamma-Ray Decay Scheme and Angular Correlation for 60 Co. Equipment Required. Purpose. Introduction
ORTEC Experiment 19 Equipment Required Two 905-3 NaI(Tl) 2- x 2-in. Detectors with Phototubes. Two 266 PMT Bases. Two 556 High Voltage Power Supplies. Two 113 Scintillation Preamplifiers. Two 575A Amplifiers.
More informationA high-performance, low-cost, leading edge discriminator
PRAMANA c Indian Academy of Sciences Vol. 65, No. 2 journal of August 2005 physics pp. 273 283 A high-performance, low-cost, leading edge discriminator S K GUPTA a, Y HAYASHI b, A JAIN a, S KARTHIKEYAN
More informationExponential Waveforms
ENGR 210 Lab 9 Exponential Waveforms Purpose: To measure the step response of circuits containing dynamic elements such as capacitors. Equipment Required: 1 - HP 54xxx Oscilloscope 1 - HP 33120A Function
More informationThe Oscilloscope. Vision is the art of seeing things invisible. J. Swift ( ) OBJECTIVE To learn to operate a digital oscilloscope.
The Oscilloscope Vision is the art of seeing things invisible. J. Swift (1667-1745) OBJECTIVE To learn to operate a digital oscilloscope. THEORY The oscilloscope, or scope for short, is a device for drawing
More information