PARTICLE DETECTORS (V): ELECTRONICS

Size: px
Start display at page:

Download "PARTICLE DETECTORS (V): ELECTRONICS"

Transcription

1 Monday, April 13, PARTICLE DETECTORS (V): ELECTRONICS Zhenyu Ye April 13, 2015

2 Monday, April 13, References Techniques for Nuclear and Particle Physics Experiments by Leo, Chapter Particle Data Group Review:

3 Monday, April 13, Contents Introduction Pulse Height Selection Coincidence Determination Triggers

4 Monday, April 13, Introduction NIM Electronics NIM Standard Co-axial cable and termination Discriminator Coincidence unit Logic unit Gate and delay generator Analog-to-Digital Converter Digital-to-Analog Converter Time-to-Digital Converter Single- and Multi-Channel Analyzer Multiple solutions for one application. Imaginations!

5 Monday, April 13, A Simple Counting System Preamp+Amplifier: amplify and shape the analog signal of the detector Discriminator: deliver a standard logic signal for every analog signal with an amplitude higher than the threshold; - serve a dual purpose: exclude low level electronic noise, shape the accepted signals to a form acceptable to the scaler Scaler: counts each arriving logic signal

6 Monday, April 13, Gate and Delay Generator (GDG) Gate and Delay Generators are triggerable devices which generate variable width gate pulses ranging from a few nanoseconds to a few seconds. Gate generators can be triggered by an input logic signal, or manually via a front panel button. The gate pulse then may be used to activate a certain device, e.g., a scaler. These modules are also equipped with an end marker signal which is a logic pulse issued at the end of the gate pulse.

7 Monday, April 13, Signal Height Selection In many experiments, the reaction of interest or products of the reaction are of a fixed energy or limited to a small range of energies.

8 Monday, April 13, Signal Height Selection Single-Channel Analyzer has adjustable upper (ULD) and lower (LLD) thresholds which defines an acceptable window. Pulses with amplitudes falling into this window are converted to logic signals while everything else rejected.

9 Monday, April 13, Calibration of SCA with Sources Pulse height = a + b * channel; Pulse height = a + b * channel + c * channel 2

10 Monday, April 13, Pulse Height Spectroscopy with MCA

11 Monday, April 13, Coincidence Technology

12 Monday, April 13, Coincidence Technology

13 Monday, April 13, Coincidence Technology

14 Monday, April 13, Coincidence Technology

15 Monday, April 13, Coincidence Curve Ideal coincidence curve is rectangular. Timing variations from detectors and electronics cause a jitter in the timing relation between the two signals which smear out the ideal rectangular coincidence curve. σ=resolving Time If the width of the signal is too smaller than fluctuations, true coincidences will be lost. In such cases, the widths should be enlarged.

16 Monday, April 13, Timing Measurements

17 Monday, April 13,

18 Monday, April 13, Accidental Coincidences Uncorrelated background events which happen to arrive within the resolving time of the circuit. The accidental coincidence rate ~ σn 1 N 2 The accidental coincidence rate corresponds to the baseline of the coincidence curve.

19 Monday, April 13, Combining Pulse Height Selection and Coincidence

20 Monday, April 13, Combining Pulse Height Selection and Coincidence

21 Monday, April 13, Triggers Inhibit or Busy Signal

22 Monday, April 13, Triggers Rutherford Scattering Rutherford Scattering Experiment

23 Monday, April 13, Triggers Rutherford Scattering Rutherford Scattering Experiment

24 Monday, April 13, Triggers Two Body Scattering Two-body Scattering Experiment

25 Monday, April 13, Triggers Two Body Scattering Two-body Scattering Experiment

26 Monday, April 13, Triggers Muon Lifetime

27 Monday, April 13, Triggers Muon Lifetime

Electronic Instrumentation for Radiation Detection Systems

Electronic Instrumentation for Radiation Detection Systems Electronic Instrumentation for Radiation Detection Systems January 23, 2018 Joshua W. Cates, Ph.D. and Craig S. Levin, Ph.D. Course Outline Lecture Overview Brief Review of Radiation Detectors Detector

More information

Dual 500ns ADC User Manual

Dual 500ns ADC User Manual 7072 Dual 500ns ADC User Manual copyright FAST ComTec GmbH Grünwalder Weg 28a, D-82041 Oberhaching Germany Version 2.3, May 11, 2009 Copyright Information Copyright Information Copyright 2001-2009 FAST

More information

Analog-to-Digital-Converter User Manual

Analog-to-Digital-Converter User Manual 7070 Analog-to-Digital-Converter User Manual copyright FAST ComTec GmbH Grünwalder Weg 28a, D-82041 Oberhaching Germany Version 2.0, July 7, 2005 Software Warranty FAST ComTec warrants proper operation

More information

K 223 Angular Correlation

K 223 Angular Correlation K 223 Angular Correlation K 223.1 Aim of the Experiment The aim of the experiment is to measure the angular correlation of a γ γ cascade. K 223.2 Required Knowledge Definition of the angular correlation

More information

ORTEC Experiment 13. Gamma-Gamma Coincidence with Angular Correlation. Equipment Required

ORTEC Experiment 13. Gamma-Gamma Coincidence with Angular Correlation. Equipment Required ORTEC Experiment 13 Equipment Required Two 905-3 2-in. x 2-in. NaI(Tl) Scintillation Detector Assemblies. Two 266 Photomultiplier Tube Bases. Two 113 Scintillation Preamplifiers. Two 556 High Voltage Power

More information

Coincidence Rates. QuarkNet. summer workshop June 24-28, 2013

Coincidence Rates. QuarkNet. summer workshop June 24-28, 2013 Coincidence Rates QuarkNet summer workshop June 24-28, 2013 1 Example Pulse input Threshold level (-10 mv) Discriminator output Once you have a digital logic pulse, you can analyze it using digital electronics

More information

Test Procedure for Nanometric Discriminator Cards for Wire Chambers. James Clarke

Test Procedure for Nanometric Discriminator Cards for Wire Chambers. James Clarke Test Procedure for Nanometric Discriminator Cards for Wire Chambers James Clarke November 26, 2013 To test a discriminator card, first ensure you have the proper equipment. For our setup, we had three

More information

ORTEC. Time-to-Amplitude Converters and Time Calibrator. Choosing the Right TAC. Timing with TACs

ORTEC. Time-to-Amplitude Converters and Time Calibrator. Choosing the Right TAC. Timing with TACs ORTEC Time-to-Amplitude Converters Choosing the Right TAC The following topics provide the information needed for selecting the right time-to-amplitude converter (TAC) for the task. The basic principles

More information

PH2510 Nuclear Physics Laboratory Use of Scintillation Counters (NP5)

PH2510 Nuclear Physics Laboratory Use of Scintillation Counters (NP5) Physics Department Royal Holloway University of London PH2510 Nuclear Physics Laboratory Use of Scintillation Counters (NP5) 1. Introduction 1.1 Object of the Experiment The object of this experiment is

More information

LIFETIME OF THE MUON

LIFETIME OF THE MUON Muon Decay 1 LIFETIME OF THE MUON Introduction Muons are unstable particles; otherwise, they are rather like electrons but with much higher masses, approximately 105 MeV. Radioactive nuclear decays do

More information

GAMMA-GAMMA CORRELATION Latest Revision: August 21, 2007

GAMMA-GAMMA CORRELATION Latest Revision: August 21, 2007 C1-1 GAMMA-GAMMA CORRELATION Latest Revision: August 21, 2007 QUESTION TO BE INVESTIGATED: decay event? What is the angular correlation between two gamma rays emitted by a single INTRODUCTION & THEORY:

More information

What s a Counter Plateau. An introduction for the muon Lab

What s a Counter Plateau. An introduction for the muon Lab What s a Counter Plateau An introduction for the muon Lab Counters have noise and signal If you are lucky, a histogram of the pulse heights of all the signals coming out of a photomultiplier tube connected

More information

Gamma Ray Spectroscopy with NaI(Tl) and HPGe Detectors

Gamma Ray Spectroscopy with NaI(Tl) and HPGe Detectors Nuclear Physics #1 Gamma Ray Spectroscopy with NaI(Tl) and HPGe Detectors Introduction: In this experiment you will use both scintillation and semiconductor detectors to study γ- ray energy spectra. The

More information

NIM. ADCs (Peak Sensing) Analog Pulse Processors Amplifiers (Fast) Amplifiers (Spectroscopy) Attenuators Coincidence/Logic/Trigger Units

NIM. ADCs (Peak Sensing) Analog Pulse Processors Amplifiers (Fast) Amplifiers (Spectroscopy) Attenuators Coincidence/Logic/Trigger Units The NIM-Nuclear Instrumentation Module standard is a very popular form factor widely used in experimental Particle and Nuclear Physics setups. Defined the first time by the U.S. Atomic Energy Commission

More information

NIM INDEX. Attenuators. ADCs (Peak Sensing) Discriminators. Translators Analog Pulse Processors Amplifiers (Fast) Amplifiers (Spectroscopy)

NIM INDEX. Attenuators. ADCs (Peak Sensing) Discriminators. Translators Analog Pulse Processors Amplifiers (Fast) Amplifiers (Spectroscopy) NIM The NIM-Nuclear Instrumentation Module standard is a very popular form factor widely used in experimental Particle and Nuclear Physics setups. Defined the first time by the U.S. Atomic Energy Commission

More information

An ASIC dedicated to the RPCs front-end. of the dimuon arm trigger in the ALICE experiment.

An ASIC dedicated to the RPCs front-end. of the dimuon arm trigger in the ALICE experiment. An ASIC dedicated to the RPCs front-end of the dimuon arm trigger in the ALICE experiment. L. Royer, G. Bohner, J. Lecoq for the ALICE collaboration Laboratoire de Physique Corpusculaire de Clermont-Ferrand

More information

Purpose This experiment will use the coincidence method for time correlation to measure the lifetime in the decay scheme of 57

Purpose This experiment will use the coincidence method for time correlation to measure the lifetime in the decay scheme of 57 Equipment Required Two 113 Scintillation Preamplifiers Two 266 Photomultiplier Tube Bases 4001A/4002D Bin and Power Supply 414A Fast Coincidence Two 551 Timing Single-Channel Analyzers 567 Time-to-Amplitude

More information

ORTEC. AN34 Experiment 14 Nuclear Lifetimes and the Coincidence Method. Equipment Needed from ORTEC. Equipment Required from Other Manufacturers

ORTEC. AN34 Experiment 14 Nuclear Lifetimes and the Coincidence Method. Equipment Needed from ORTEC. Equipment Required from Other Manufacturers Equipment Needed from ORTEC Two 113 Scintillation Preamplifiers Two 266 Photomultiplier Tube Bases 4001A/4002D Bin and Power Supply 414A Fast Coincidence Two 551 Timing Single-Channel Analyzers 567 Time-to-Amplitude

More information

Quad Analog-to-Digital Converter Technical Documentation

Quad Analog-to-Digital Converter Technical Documentation 7074 Quad AnalogtoDigital Converter Technical Documentation copyright FAST ComTec GmbH Grünwalder Weg 28a, D82041 Oberhaching Germany Version 2.2, February 25, 2005 Table of Contents Table of Contents

More information

LeCroy Research Systems Model 365AL, Model 465, and Model 622 Logic Units

LeCroy Research Systems Model 365AL, Model 465, and Model 622 Logic Units Page 1 of 5 365AL DUAL 4-FOLD MAJORITY LOGIC UNIT 465 TRIPLE 4-FOLD LOGIC UNIT (Note - the 465 is no longer available) 622 QUAD 2 LOGIC UNIT NIM Packaging High Speed Multiple Input Multiple Output Selectable

More information

COMPTON SCATTERING. Purpose. Introduction. Fundamentals of Experiment

COMPTON SCATTERING. Purpose. Introduction. Fundamentals of Experiment COMPTON SCATTERING Purpose The purpose of this experiment is to verify the energy dependence of gamma radiation upon scattering angle and to compare the differential cross section obtained from the data

More information

The (Speed and) Decay of Cosmic-Ray Muons

The (Speed and) Decay of Cosmic-Ray Muons The (Speed and) Decay of Cosmic-Ray Muons Jason Gross MIT - Department of Physics Jason Gross (8.13) Cosmic-Ray Muons November 4, 2011 1 / 30 Goals test relativity (time dilation) determine the mean lifetime

More information

A high-performance, low-cost, leading edge discriminator

A high-performance, low-cost, leading edge discriminator PRAMANA c Indian Academy of Sciences Vol. 65, No. 2 journal of August 2005 physics pp. 273 283 A high-performance, low-cost, leading edge discriminator S K GUPTA a, Y HAYASHI b, A JAIN a, S KARTHIKEYAN

More information

Status of Primex Beam Position Monitor July 29 th, 2010

Status of Primex Beam Position Monitor July 29 th, 2010 Status of Primex Beam Position Monitor July 29 th, 2010 Anthony Tatum University of North Carolina at Wilmington The Beam Position Monitor (BPM) is used to determine the vertical and horizontal position

More information

Model 935 Quad Constant-Fraction 200-MHz Discriminator Operating and Service Manual

Model 935 Quad Constant-Fraction 200-MHz Discriminator Operating and Service Manual Model 935 Quad Constant-Fraction 200-MHz Discriminator Operating and Service Manual U.S. Patent No. 4,179,644 Printed in U.S.A. ORTEC Part No. 753770 0503 Manual Revision H ii $GYDQFHG0HDVXUHPHQW7HFKQRORJ\,QF

More information

Chapter 6 Pulse Processing

Chapter 6 Pulse Processing Med Phys 4RA3, 4RB3/6R3 Radioisotopes and Radiation Methodology 6-6.. Introduction Chapter 6 Pulse Processing Most radiation detectors require pulse (or signal) processing electronics so that energy or

More information

Fast first practical help -- detailed instructions will follow- preliminary Experiment F80

Fast first practical help -- detailed instructions will follow- preliminary Experiment F80 Fast first practical help -- detailed instructions will follow- preliminary Experiment F80 Measurement Methods of Nuclear and Particle Physics Introduction: This experiment is going to introduce you to

More information

A CMOS INTEGRATED CIRCUIT FOR PULSE-SHAPE DISCRIMINATION*

A CMOS INTEGRATED CIRCUIT FOR PULSE-SHAPE DISCRIMINATION* A CMOS INTEGRATED CIRCUIT FOR PULSE-SHAPE DISCRIMINATION* S. S. Frank, M. N. Ericson, M. L. Simpson, R. A. Todd, and D. P. Hutchinson Oak Ridge National Laboratory, Oak Ridge, TN 3783 1 Abstract and Summary

More information

CAEN Tools for Discovery

CAEN Tools for Discovery Viareggio 5 September 211 Introduction In recent years CAEN has developed a complete family of digitizers that consists of several models differing in sampling frequency, resolution, form factor and other

More information

and N(t) ~ exp(-t/ ),

and N(t) ~ exp(-t/ ), Muon Lifetime Experiment Introduction Charged and neutral particles with energies in excess of 10 23 ev from Galactic and extra Galactic sources impinge on the earth. Here we speak of the earth as the

More information

Highly Segmented Detector Arrays for. Studying Resonant Decay of Unstable Nuclei. Outline

Highly Segmented Detector Arrays for. Studying Resonant Decay of Unstable Nuclei. Outline Highly Segmented Detector Arrays for Studying Resonant Decay of Unstable Nuclei MASE: Multiplexed Analog Shaper Electronics C. Metelko, S. Hudan, R.T. desouza Outline 1. Resonant Decay 2. Detectors 3.

More information

Experiment 10. The Speed of Light c Introduction Apparatus

Experiment 10. The Speed of Light c Introduction Apparatus Experiment 10 The Speed of Light c 10.1 Introduction In this experiment you will measure the speed of light, c. This is one of the most fundamental constants in physics, and at the same time the fastest

More information

PX4 Frequently Asked Questions (FAQ)

PX4 Frequently Asked Questions (FAQ) PX4 Frequently Asked Questions (FAQ) What is the PX4? The PX4 is a component in the complete signal processing chain of a nuclear instrumentation system. It replaces many different components in a traditional

More information

INDEX. Firmware for DPP (Digital Pulse Processing) DPP-PSD Digital Pulse Processing for Pulse Shape Discrimination

INDEX. Firmware for DPP (Digital Pulse Processing) DPP-PSD Digital Pulse Processing for Pulse Shape Discrimination Firmware for DPP (Digital Pulse Processing) Thanks to the powerful FPGAs available nowadays, it is possible to implement Digital Pulse Processing (DPP) algorithms directly on the acquisition boards and

More information

Digital trigger system for the RED-100 detector based on the unit in VME standard

Digital trigger system for the RED-100 detector based on the unit in VME standard Journal of Physics: Conference Series PAPER OPEN ACCESS Digital trigger system for the RED-100 detector based on the unit in VME standard To cite this article: D Yu Akimov et al 2016 J. Phys.: Conf. Ser.

More information

nanomca 80 MHz HIGH PERFORMANCE, LOW POWER DIGITAL MCA Model Numbers: NM0530 and NM0530Z

nanomca 80 MHz HIGH PERFORMANCE, LOW POWER DIGITAL MCA Model Numbers: NM0530 and NM0530Z datasheet nanomca 80 MHz HIGH PERFORMANCE, LOW POWER DIGITAL MCA Model Numbers: NM0530 and NM0530Z I. FEATURES Finger-sized, high performance digital MCA. 16k channels utilizing smart spectrum-size technology

More information

nanodpp datasheet I. FEATURES

nanodpp datasheet I. FEATURES datasheet nanodpp I. FEATURES Ultra small size high-performance Digital Pulse Processor (DPP). 16k channels utilizing smart spectrum-size technology -- all spectra are recorded and stored as 16k spectra

More information

Moderne Teilchendetektoren - Theorie und Praxis 2. Dr. Bernhard Ketzer Technische Universität München SS 2013

Moderne Teilchendetektoren - Theorie und Praxis 2. Dr. Bernhard Ketzer Technische Universität München SS 2013 Moderne Teilchendetektoren - Theorie und Praxis 2 Dr. Bernhard Ketzer Technische Universität München SS 2013 7 Signal Processing and Acquisition 7.1 Signals 7.2 Amplifier 7.3 Electronic Noise 7.4 Analog-to-Digital

More information

Physics Experiment N -17. Lifetime of Cosmic Ray Muons with On-Line Data Acquisition on a Computer

Physics Experiment N -17. Lifetime of Cosmic Ray Muons with On-Line Data Acquisition on a Computer Introduction Physics 410-510 Experiment N -17 Lifetime of Cosmic Ray Muons with On-Line Data Acquisition on a Computer The experiment is designed to teach the techniques of particle detection using scintillation

More information

ORTEC Experiment 9. Time Coincidence Techniques Applied to Absolute Activity Measurements. Equipment Required. Purpose.

ORTEC Experiment 9. Time Coincidence Techniques Applied to Absolute Activity Measurements. Equipment Required. Purpose. ORTEC Experiment 9 Equipment Required 905-3 2-inch x 2-inch (5.08-cm x 5.08-cm) NaI(Tl) Detector and PM Tube Assembly. 266 PhotoMultiplier Tube Base. Two 113 Preamplifiers. BA-015-050-1000 Partially-Depleted

More information

nanomca datasheet I. FEATURES

nanomca datasheet I. FEATURES datasheet nanomca I. FEATURES Finger-sized, high performance digital MCA. 16k channels utilizing smart spectrum-size technology -- all spectra are recorded and stored as 16k spectra with instant, distortion-free

More information

DSA-LX. Digital Signal Analyzer. Radiation Safety. Amplified.

DSA-LX. Digital Signal Analyzer. Radiation Safety. Amplified. Radiation Safety. Amplified. DSA-LX Digital Signal Analyzer Nuclear Healthcare Homeland Security & Defense Labs and Education Industrial and Manufacturing KEY FEATURES Integrated desktop MCA based on Digital

More information

Model 310H Fast 800V Pulse Generator

Model 310H Fast 800V Pulse Generator KEY FEATURES Temperature Stability +/-5ppm 100 V to 800 V into 50 Ω

More information

Traditional analog QDC chain and Digital Pulse Processing [1]

Traditional analog QDC chain and Digital Pulse Processing [1] Giuliano Mini Viareggio April 22, 2010 Introduction The aim of this paper is to compare the energy resolution of two gamma ray spectroscopy setups based on two different acquisition chains; the first chain

More information

ORTEC. Research Applications. Pulse-Height, Charge, or Energy Spectroscopy. Detectors. Processing Electronics

ORTEC. Research Applications. Pulse-Height, Charge, or Energy Spectroscopy. Detectors. Processing Electronics ORTEC Spectroscopy systems for ORTEC instrumentation produce pulse height distributions of gamma ray or alpha energies. MAESTRO-32 (model A65-B32) is the software included with most spectroscopy systems

More information

Positron Emission Tomography

Positron Emission Tomography Positron Emission Tomography UBC Physics & Astronomy / PHYS 409 1 Introduction Positron emission tomography (PET) is a non-invasive way to produce the functional 1 image of a patient. It works by injecting

More information

The Time of Flight Upgrade for CLAS at 12GeV

The Time of Flight Upgrade for CLAS at 12GeV The Time of Flight Upgrade for CLAS at 12GeV by Lewis P. Graham Bachelor of Science Benedict College, 2002 -------------------------------------------------------------------- Submitted in Partial Fulfillment

More information

Photon Counters SR430 5 ns multichannel scaler/averager

Photon Counters SR430 5 ns multichannel scaler/averager Photon Counters SR430 5 ns multichannel scaler/averager SR430 Multichannel Scaler/Averager 5 ns to 10 ms bin width Count rates up to 100 MHz 1k to 32k bins per record Built-in discriminator No interchannel

More information

AN ABSTRACT ON THE THESIS OF. David C. Vasquez for the degree of Master of Science in Radiation Health Physics presented on February 26, 2010.

AN ABSTRACT ON THE THESIS OF. David C. Vasquez for the degree of Master of Science in Radiation Health Physics presented on February 26, 2010. AN ABSTRACT ON THE THESIS OF David C. Vasquez for the degree of Master of Science in Radiation Health Physics presented on February 26, 2010. Title: The Design, Use and Implementation of Digital Radiation

More information

Model 542 Linear Gate and Stretcher Operating and Service Manual

Model 542 Linear Gate and Stretcher Operating and Service Manual Model 542 Linear Gate and Stretcher Operating and Service Manual NOTE: A substitution for the dual diode package (MSD6100) may have been made in this unit. If so, two 1N4153 diodes were used to replace

More information

The behavior of the FastADC in time domain

The behavior of the FastADC in time domain August 29, 2000 The behavior of the FastADC in time domain F. Tonisch 1. General remarks The 8-channel FastADC was developed for use with the readout electronic of the Waveguide Beam Position Monitors

More information

Contents. Why waveform? Waveform digitizer : Domino Ring Sampler CEX Beam test autumn 04. Summary

Contents. Why waveform? Waveform digitizer : Domino Ring Sampler CEX Beam test autumn 04. Summary Contents Why waveform? Waveform digitizer : Domino Ring Sampler CEX Beam test data @PSI autumn 04 Templates and time resolution Pulse Shape Discrimination Pile-up rejection Summary 2 In the MEG experiment

More information

Scintillators as an external trigger for cathode strip chambers

Scintillators as an external trigger for cathode strip chambers Scintillators as an external trigger for cathode strip chambers J. A. Muñoz Department of Physics, Princeton University, Princeton, NJ 08544 An external trigger was set up to test cathode strip chambers

More information

Digital coincidence acquisition applied to portable β liquid scintillation counting device

Digital coincidence acquisition applied to portable β liquid scintillation counting device Nuclear Science and Techniques 24 (2013) 030401 Digital coincidence acquisition applied to portable β liquid scintillation counting device REN Zhongguo 1,2 HU Bitao 1 ZHAO Zhiping 2 LI Dongcang 1,* 1 School

More information

Testing of the NSC Electronics Module with the GSI Clover Detector

Testing of the NSC Electronics Module with the GSI Clover Detector Testing of the NSC Electronics Module with the GSI Clover Detector Rakesh Kumar 1, P. Queiroz 2, H.-J. Wollersheim 2 (Tutor) 1 Inter University Accelerator Centre Aruna Asaf Ali Marg Post Box No 10502

More information

Testing the Electronics for the MicroBooNE Light Collection System

Testing the Electronics for the MicroBooNE Light Collection System Testing the Electronics for the MicroBooNE Light Collection System Kathleen V. Tatem Nevis Labs, Columbia University & Fermi National Accelerator Laboratory August 3, 2012 Abstract This paper discusses

More information

Amptek Silicon Drift Diode (SDD) at High Count Rates

Amptek Silicon Drift Diode (SDD) at High Count Rates Amptek Silicon Drift Diode (SDD) at High Count Rates A silicon drift diode (SDD) is functionally similar to a SiPIN photodiode but its unique electrode structure reduces the electronic noise at short peaking

More information

NEEP 427 PROPORTIONAL COUNTERS. Knoll, Chapters 6 & 14 Sect. I & II

NEEP 427 PROPORTIONAL COUNTERS. Knoll, Chapters 6 & 14 Sect. I & II NEEP 427 PROPORTIONAL COUNTERS References: Knoll, Chapters 6 & 14 Sect. I & II a proportional counter the height of the output pulse is proportional to the number of ion pairs produced in the counter gas.

More information

K. Desch, P. Fischer, N. Wermes. Physikalisches Institut, Universitat Bonn, Germany. Abstract

K. Desch, P. Fischer, N. Wermes. Physikalisches Institut, Universitat Bonn, Germany. Abstract ATLAS Internal Note INDET-NO-xxx 28.02.1996 A Proposal to Overcome Time Walk Limitations in Pixel Electronics by Reference Pulse Injection K. Desch, P. Fischer, N. Wermes Physikalisches Institut, Universitat

More information

Model 566 Time-to-Amplitude Converter (TAC) Operating and Service Manual

Model 566 Time-to-Amplitude Converter (TAC) Operating and Service Manual Model 566 Time-to-Amplitude Converter (TAC) Operating and Service Manual Printed in U.S.A. ORTEC Part No. 678950 0411 Manual Revision F Advanced Measurement Technology, Inc. a/k/a/ ORTEC, a subsidiary

More information

Week 11: Chap. 16b Pulse Shaping

Week 11: Chap. 16b Pulse Shaping Week 11: Chap. 16b Pulse Shaping Pulse Processing (passive) Pulse Shaping (active) -- Op Amps -- CR/RC network -- Bipolar pulses --- Shaping network --- Pole Zero network --- Baseline Restorer -- Delay-line

More information

Development of Personal Dosimeter Using Electronic Dose Conversion Method

Development of Personal Dosimeter Using Electronic Dose Conversion Method Proceedings of the Korean Nuclear Spring Meeting Gyeong ju, Korea, May 2003 Development of Personal Dosimeter Using Electronic Dose Conversion Method Wanno Lee, Bong Jae Lee, and Chang Woo Lee Korea Atomic

More information

MASE: Multiplexed Analog Shaped Electronics

MASE: Multiplexed Analog Shaped Electronics MASE: Multiplexed Analog Shaped Electronics C. Metelko, A. Alexander, J. Poehlman, S. Hudan, R.T. desouza Outline 1. Needs 2. Problems with existing Technology 3. Design Specifications 4. Overview of the

More information

ORTEC Experiment 19. Gamma-Ray Decay Scheme and Angular Correlation for 60 Co. Equipment Required. Purpose. Introduction

ORTEC Experiment 19. Gamma-Ray Decay Scheme and Angular Correlation for 60 Co. Equipment Required. Purpose. Introduction ORTEC Experiment 19 Equipment Required Two 905-3 NaI(Tl) 2- x 2-in. Detectors with Phototubes. Two 266 PMT Bases. Two 556 High Voltage Power Supplies. Two 113 Scintillation Preamplifiers. Two 575A Amplifiers.

More information

Two Slit Interference, One Photon at a Time

Two Slit Interference, One Photon at a Time 1 of 5 20/02/2007 11:11 AM home about us unique support users prices newsletters contact us products overview diode laser spectroscopy earth's field nmr earth's field nmr gradient/field coil system faraday

More information

nanomca-sp datasheet I. FEATURES

nanomca-sp datasheet I. FEATURES datasheet nanomca-sp 80 MHz HIGH PERFORMANCE, LOW POWER DIGITAL MCA WITH BUILT IN PREAMPLIFIER Model Numbers: SP0534A/B to SP0539A/B Standard Models: SP0536B and SP0536A I. FEATURES Built-in preamplifier

More information

Generation of Gaussian Pulses using FPGA for Simulating Nuclear Counting System

Generation of Gaussian Pulses using FPGA for Simulating Nuclear Counting System Generation of Gaussian Pulses using FPGA for Simulating Nuclear Counting System Mohaimina Begum Md. Abdullah Al Mamun Md. Atiar Rahman Sabiha Sattar Abstract- Nuclear radiation counting system is used

More information

Hall C Winter Collaboration Meeting. Hall C Analyzer Update

Hall C Winter Collaboration Meeting. Hall C Analyzer Update Hall C Winter Collaboration Meeting Hall C Analyzer Update Jefferson Lab On Behalf of the Software Working Group Hall A/C Software Analysis Workshop Hall A/C software analysis workshop took place June

More information

Summary. Introduction

Summary. Introduction Performance of an Enhanced Throughput Feature in a High-Count Rate System Ronald M Keyser, Senior Member, and Rex C Trammell, Senior Member ORTEC 801 South Illinois Avenue Oak Ridge, TN 37831-0895 Summary

More information

PMT tests at UMD. Vlasios Vasileiou Version st May 2006

PMT tests at UMD. Vlasios Vasileiou Version st May 2006 PMT tests at UMD Vlasios Vasileiou Version 1.0 1st May 2006 Abstract This memo describes the tests performed on three Milagro PMTs in UMD. Initially, pulse-height distributions of the PMT signals were

More information

A NOVEL FPGA-BASED DIGITAL APPROACH TO NEUTRON/ -RAY PULSE ACQUISITION AND DISCRIMINATION IN SCINTILLATORS

A NOVEL FPGA-BASED DIGITAL APPROACH TO NEUTRON/ -RAY PULSE ACQUISITION AND DISCRIMINATION IN SCINTILLATORS 10th ICALEPCS Int. Conf. on Accelerator & Large Expt. Physics Control Systems. Geneva, 10-14 Oct 2005, PO2.041-4 (2005) A NOVEL FPGA-BASED DIGITAL APPROACH TO NEUTRON/ -RAY PULSE ACQUISITION AND DISCRIMINATION

More information

Diamond sensors as beam conditions monitors in CMS and LHC

Diamond sensors as beam conditions monitors in CMS and LHC Diamond sensors as beam conditions monitors in CMS and LHC Maria Hempel DESY Zeuthen & BTU Cottbus on behalf of the BRM-CMS and CMS-DESY groups GSI Darmstadt, 11th - 13th December 2011 Outline 1. Description

More information

Integrating Analogue to Digital Converter (ADC)

Integrating Analogue to Digital Converter (ADC) Integrating Analogue to Digital Converter (ADC) Integrate signal during application of gate - another time variant filter convert charge to digital number = convolution of pulse shape with gate so w(t)

More information

PACS codes: Qx, Nc, Kv, v Keywords: Digital data acquisition, segmented HPGe detectors, clock and trigger distribution

PACS codes: Qx, Nc, Kv, v Keywords: Digital data acquisition, segmented HPGe detectors, clock and trigger distribution Clock and Trigger Synchronization between Several Chassis of Digital Data Acquisition Modules W. Hennig, H. Tan, M. Walby, P. Grudberg, A. Fallu-Labruyere, W.K. Warburton, XIA LLC, 31057 Genstar Road,

More information

ORTEC Experiment 1. Introduction to Electronic Signal Analysis in Nuclear Radiation Measurements. Equipment Required: Purpose. Electronic Circuits

ORTEC Experiment 1. Introduction to Electronic Signal Analysis in Nuclear Radiation Measurements. Equipment Required: Purpose. Electronic Circuits ORTEC Experiment 1 Equipment Required: 480 Pulser 113 Scintillation Preamplifier 4001A/4002D NIM Bin and Power Supply 575A Spectroscopy Amplifier 996 Timer and Counter 551 Timing Single-Channel Analyzer

More information

--- preliminary Experiment F80

--- preliminary Experiment F80 --- preliminary Experiment F80 Measurement Methods of Nuclear and Particle Physics Introduction: This experiment is going to introduce you to important counting and measuring techniques of nuclear and

More information

ORTEC ORTEC. Modular Pulse- Processing Electronics. What s in this Catalog? Who Needs this Catalog?

ORTEC ORTEC. Modular Pulse- Processing Electronics.   What s in this Catalog? Who Needs this Catalog? ORTEC Modular Pulse- Processing Electronics What s in this Catalog? Tutorial Information makes this more than just a catalog. "What You Need to Know About Modular Electronic Instruments" takes you through

More information

Mass Spectrometry and the Modern Digitizer

Mass Spectrometry and the Modern Digitizer Mass Spectrometry and the Modern Digitizer The scientific field of Mass Spectrometry (MS) has been under constant research and development for over a hundred years, ever since scientists discovered that

More information

The Speed of Light Laboratory Experiment 8. Introduction

The Speed of Light Laboratory Experiment 8. Introduction Exp-8-Speed of Light.doc (TJR) Physics Department, University of Windsor 64-311 Laboratory Experiment 8 The Speed of Light Introduction Galileo was right. Light did not travel instantaneously as his contemporaries

More information

Detecting and Suppressing Background Signal

Detecting and Suppressing Background Signal Detecting and Suppressing Background Signal Valerie Gray St. Norbert College Advisors: Dr. Michael Wiescher Freimann Professor Nuclear Physics University of Notre Dame Dr. Ed Stech Associate Professional

More information

Instructions for gg Coincidence with 22 Na. Overview of the Experiment

Instructions for gg Coincidence with 22 Na. Overview of the Experiment Overview of the Experiment Instructions for gg Coincidence with 22 Na 22 Na is a radioactive element that decays by converting a proton into a neutron: about 90% of the time through β + decay and about

More information

On the initiation of lightning in thunderclouds (Instrumentation, Supplementary information)

On the initiation of lightning in thunderclouds (Instrumentation, Supplementary information) On the initiation of lightning in thunderclouds (Instrumentation, Supplementary information) Ashot Chilingarian 1,2, Suren Chilingaryan 1, Tigran Karapetyan 1, Lev Kozliner 1, Yeghia Khanikyants 1, Gagik

More information

Silicon Photomultiplier Evaluation Kit. Quick Start Guide. Eval Kit SiPM. KETEK GmbH. Hofer Str Munich Germany.

Silicon Photomultiplier Evaluation Kit. Quick Start Guide. Eval Kit SiPM. KETEK GmbH. Hofer Str Munich Germany. KETEK GmbH Hofer Str. 3 81737 Munich Germany www.ketek.net info@ketek.net phone +49 89 673 467 70 fax +49 89 673 467 77 Silicon Photomultiplier Evaluation Kit Quick Start Guide Eval Kit Table of Contents

More information

Exercise 4 - THE OSCILLOSCOPE

Exercise 4 - THE OSCILLOSCOPE Exercise 4 - THE OSCILLOSCOPE INTRODUCTION You have been exposed to analogue oscilloscopes in the first year lab. As you are probably aware, the complexity of the instruments, along with their importance

More information

Nuclear Experiment Phys 318/317 Room 208 Instructor Richard Lindgren Room 302 Ext

Nuclear Experiment Phys 318/317 Room 208 Instructor Richard Lindgren Room 302 Ext Revised Aug 28, 2008 Nuclear Experiment Phys 318/317 Room 208 Instructor Richard Lindgren Room 302 Ext 2-2691 ral5q@virginia.edu 1 NOTE: Some of the figures referred to in this document can be found in

More information

Digital Signal Processing for HPGe Detectors

Digital Signal Processing for HPGe Detectors Digital Signal Processing for HPGe Detectors David Radford ORNL Physics Division July 28, 2012 HPGe Detectors Hyper-Pure Ge (HPGe) detectors are the gold standard for gamma-ray spectroscopy Unsurpassed

More information

Beam Condition Monitors and a Luminometer Based on Diamond Sensors

Beam Condition Monitors and a Luminometer Based on Diamond Sensors Beam Condition Monitors and a Luminometer Based on Diamond Sensors Wolfgang Lange, DESY Zeuthen and CMS BRIL group Beam Condition Monitors and a Luminometer Based on Diamond Sensors INSTR14 in Novosibirsk,

More information

The software and hardware for the ground testing of ALFA- ELECTRON space spectrometer

The software and hardware for the ground testing of ALFA- ELECTRON space spectrometer Journal of Physics: Conference Series PAPER OPEN ACCESS The software and hardware for the ground testing of ALFA- ELECTRON space spectrometer To cite this article: A G Batischev et al 2016 J. Phys.: Conf.

More information

A user-friendly fully digital TDPAC-spectrometer

A user-friendly fully digital TDPAC-spectrometer Hyperfine Interact DOI 10.1007/s10751-010-0201-8 A user-friendly fully digital TDPAC-spectrometer M. Jäger K. Iwig T. Butz Springer Science+Business Media B.V. 2010 Abstract A user-friendly fully digital

More information

Data Compression and Analysis Methods for High- Throughput Radiation Detector Systems

Data Compression and Analysis Methods for High- Throughput Radiation Detector Systems 1 Data Compression and Analysis Methods for High- Throughput Radiation Detector Systems John Mattingly Associate Professor, Nuclear Engineering North Carolina State University 2 Introduction The capabilities

More information

Investigation of a Cs137 and Ba133 runs. Michael Dugger and Robert Lee

Investigation of a Cs137 and Ba133 runs. Michael Dugger and Robert Lee Investigation of a Cs137 and Ba133 runs Michael Dugger and Robert Lee 1 Cs137 Using run 149 One million triggers Doing a quick analysis with fits: Not using Kei s noise corrections at the moment 2 ADC

More information

Noise Characteristics Of The KPiX ASIC Readout Chip

Noise Characteristics Of The KPiX ASIC Readout Chip Noise Characteristics Of The KPiX ASIC Readout Chip Cabrillo College Stanford Linear Accelerator Center What Is The ILC The International Linear Collider is an e- e+ collider Will operate at 500GeV with

More information

Model 2145 Time to Amplitude Converter

Model 2145 Time to Amplitude Converter 9231617B Model 2145 Time to Amplitude Converter User s Manual Copyright 2005, Canberra Industries, Inc. All rights reserved. The material in this document, including all information, pictures, graphics

More information

Model 9302 Amplifier-Discriminator Operating and Service Manual

Model 9302 Amplifier-Discriminator Operating and Service Manual Model 9302 Amplifier-Discriminator Operating and Service Manual Printed in U.S.A. ORTEC Part No. 733690 1202 Manual Revision C Advanced Measurement Technology, Inc. a/k/a/ ORTEC, a subsidiary of AMETEK,

More information

Implementation of A Nanosecond Time-resolved APD Detector System for NRS Experiment in HEPS-TF

Implementation of A Nanosecond Time-resolved APD Detector System for NRS Experiment in HEPS-TF Implementation of A Nanosecond Time-resolved APD Detector System for NRS Experiment in HEPS-TF LI Zhen-jie a ; MA Yi-chao c ; LI Qiu-ju a ; LIU Peng a ; CHANG Jin-fan b ; ZHOU Yang-fan a * a Beijing Synchrotron

More information

PX5 User Manual and Operating Instructions

PX5 User Manual and Operating Instructions PX5 User Manual and Operating Instructions Amptek, Inc. 14 DeAngelo Dr. Bedford, MA 01730 PH: +1 781-275-2242 FAX: +1 781-275-3470 sales@amptek.com www.amptek.com Other PX5 related documents: PX5 Quick

More information

S4 OPERATION DESCRIPTION IN PAMELA INSTRUMENT

S4 OPERATION DESCRIPTION IN PAMELA INSTRUMENT For internal use only S4 OPERATION DESCRIPTION IN PAMELA INSTRUMENT 1 1. DESTINATION The bottom detector S4 of PAMELA instrument is intended for: choice of useful events by energy deposition in S4 detector

More information

nanomca-ii-sp datasheet

nanomca-ii-sp datasheet datasheet nanomca-ii-sp 125 MHz ULTRA-HIGH PERFORMANCE DIGITAL MCA WITH BUILT IN PREAMPLIFIER Model Numbers: SP8004 to SP8009 Standard Models: SP8006B and SP8006A I. FEATURES Finger-sized, ultra-high performance

More information

Photon Count. for Brainies.

Photon Count. for Brainies. Page 1/12 Photon Count ounting for Brainies. 0. Preamble This document gives a general overview on InGaAs/InP, APD-based photon counting at telecom wavelengths. In common language, telecom wavelengths

More information

LED Displacement Sensor

LED Displacement Sensor LED Displacement Sensor Low-cost LED Displacement Sensor Offers 10-micron Resolution of 25 mm (0.98 in) with measurement range of ±4 mm (0.16 in). Easy-to-use, built-in amplifier. Fast 5-millisecond response.

More information