Energy resolution and transport properties of CdTe-Timepix-Assemblies

Size: px
Start display at page:

Download "Energy resolution and transport properties of CdTe-Timepix-Assemblies"

Transcription

1 Journal of Instrumentation OPEN ACCESS Energy resolution and transport properties of CdTe-Timepix-Assemblies To cite this article: D Greiffenberg et al View the article online for updates and enhancements. Related content - Study of a GaAs:Cr-based Timepix detector using synchrotron facility P. Smolyanskiy, D. Kozhevnikov, O. Bakina et al. - X-ray absorption and charge transport in a pixellated CdTe detector with single photon processing readout E Fröjdh, B Norlin, G Thungström et al. - Imaging properties of small-pixel spectroscopic x-ray detectors based on cadmium telluride sensors Thomas Koenig, Julia Schulze, Marcus Zuber et al. Recent citations - Energy calibration of photon counting detectors using a single monochromatic source C. Feng et al - Laboratory based study of dynamical processes by 4D X-ray CT with subsecond temporal resolution D. Vavík et al - Characterization of a pixelated CdTe Timepix detector operated in ToT mode T. Billoud et al This content was downloaded from IP address on 25/01/2018 at 18:56

2 PUBLISHED BY IOP PUBLISHING FOR SISSA 12 th INTERNATIONAL WORKSHOP ON RADIATION IMAGING DETECTORS, JULY 11 th 15 th 2010, ROBINSON COLLEGE, CAMBRIDGE U.K. RECEIVED: November 15, 2010 ACCEPTED: December 10, 2010 PUBLISHED: January 11, 2011 Energy resolution and transport properties of CdTe-Timepix-Assemblies D. Greiffenberg, a,1 A. Fauler, a A. Zwerger b and M. Fiederle a a Freiburg Materials Research Institute (FMF), Albert-Ludwigs-Universität Freiburg, Stefan-Meier-Strasse 21, D Freiburg, Germany b Physikalisches Institut, Albert-Ludwigs-Universität Freiburg, Hermann-Herder-Strasse 3, D Freiburg, Germany Dominic.Greiffenberg@fmf.uni-freiburg.de ABSTRACT: CdTe is a promising material for the detection of γ- and X-rays as 1 mm thick CdTe sensors offer an absorption efficiency that is higher than 50 % for photon energies up to 120 kev. Therefore 1mm thick CdTe from Acrorad has been flip-chipped as the sensor material on a Timepix readout-electronics ASIC at the Freiburg Materials Research Center (FMF). The transport properties of the CdTe material are investigated by determining the µτ-product of the charge carriers by illuminating the non-collecting electrode with alpha particles. The method commonly used to determine the µτ-product for non-pixelated devices was adapted for the Medipix pixelated readout-electronics. The position of the alpha particles on the pixel matrix provides spatial information, which was used to create a mapping of the µτ-product. The energy resolutions and the positions of the noise edge of two CdTe-Timepix assemblies, one with a pixel pitch of 110 x 110 µm 2 and another one with a pixel pitch of 55 x 55 µm 2 were investigated by performing a threshold scan of a 241 Am source. A comparison of the energy spectra obtained with the two assemblies with different pixel pitches confirms the strong dependence of charge sharing on the energy spectra, as expected. KEYWORDS: Charge transport and multiplication in solid media; X-ray detectors 1 Corresponding author. c 2011 IOP Publishing Ltd and SISSA doi: / /6/01/c01058

3 Contents 1 Introduction 1 2 Experimental setup 2 3 Results Determination of the charge carrier transport properties Energy spectra 5 4 Conclusions 7 1 Introduction CdTe is a promising material for the detection of γ- and X-rays with photon energies up to 120 kev. 1mm thick CdTe offers an absorption efficiency higher than 50 % within this energy regime due to its high atomic number (Z=48, 52). The sensor material was provided by Acrorad [1]. Further processing and flip-chipping with the pixelated readout electronics Timepix was done at the Freiburg Materials Research Center (FMF) [2, 3]. The assembly had a pixel pitch of 110 x 110 µm 2 and had intentionally ohmic contacts. The transport properties of the charge carriers in the CdTe sensor, namely the mobility-lifetime product (µτ-product), have a major influence on the charge collection efficiency (CCE) of the detectors [4]. The µτ-product was investigated by using alpha particles from a 241 Am source. The alpha particles are absorbed in the first 30 microns of the sensor and the charge carriers have to drift almost through the complete sensor to the pixel contacts. This method is commonly used for nonpixelated systems, where the induced charges are generating pulse amplitudes, which are measured for various applied bias voltages. As the slope of the Hecht relationship is relevant for the value of the µτ product, a proportional conversion of the number of charge carriers to pulse height is a necessary requirement. Finally by fitting the slope using a Hecht relationship, the µτ-product can be extracted [5]. The Hecht-relationship is shown in equation (1.1). Q = Q 0 [ µe τ e U bias d 2 ( 1 e x 0 d µeτeu bias ) + µ hτ h U bias d 2 ( 1 e (d x 0 )d µ h τ h U bias )] (1.1) Here Q refers to the pulse height, which is correlated with the induced charge. µ is the mobility and τ the lifetime of the charge carriers. The index e refers to electrons, while h refers to holes. U bias is the bias voltage on the sensor and d is the sensors thickness. After absorption of the alpha particles, the generated charge cloud will expand while drifting towards the collecting electrode. As the CdTe-Timepix assembly is a pixelated readout system, the charge carriers will influence charge in a circular shaped cluster expanding over several pixels. 1

4 Therefore the method described before had to be adapted, which was achieved by using the Timeover-Threshold (ToT) readout mode of the Timepix chip [6, 7]. This readout mode is measuring the time, which the signal of the charge sensitive amplifier (CSA) is above the energy threshold. The ToT is measured by counting the number of period lengths given by an internal frequency, being further referred to as ToT counts. In the case of the Timepix pixel electronics the pulse length is directly correlated to the pulse height, thus providing a measure for the pulse height in each pixel [7]. The overall charge created by an alpha particle will create pulse heights in several pixels which when summed up corresponds to the pulse height as would be induced in a non-pixelated readout system. Thus, a reconstruction of the overall pulse height is possible by summing the pulse heights in each pixel of the cluster. Moreover, the method allows spatially resolved µτ-mappings of the sensor, which then can be correlated with other properties of the assembly (e.g. quantum efficiency). The energy resolution and the energy level of the noise edge of the assembly was determined by performing an energy calibration. For this purpose, a scan of the energy threshold was performed, while the assembly was illuminated with a 241 Am source, where the alpha particles were shielded. More details of the method are given elsewhere [8, 9]. In order to compare the results with an CdTe assembly with a pixel pitch of 55 x 55 µm 2, the same procedure of obtaining an energy resolution was performed with such an assembly. 2 Experimental setup A Timepix chip from the Medipix2 collaboration was used as the readout electronics [10]. The Timepix chip consists of 256 x 256 = pixels with a pixel size of 55 x 55 µm 2 covering an active area of 1.98 cm 2. It is designed as hybrid readout electronics and is capable of collecting either electrons or holes. Thus different sensor materials like Si, GaAs and CdTe can be used. The Timepix chip also supports different readout modes like the Time-over-Threshold (ToT) mode, which measures the time, while the signal is over the energy threshold in units of the clock period of an external signal. The readout modes have been described in several publications [6, 7]. In the Timepix readout electronics, the charge sensitive amplifiers (CSA) are discharged by a current source which is controlled by the DAC IKrum. The shaped pulse decreases linearly, creating a triangular shaped pulse [6, 7]. The rise time of the CSA is constant over a wide range of induced charge. One consequence of this behavior is a direct correlation between the pulse length and the pulse height [7]. Therefore, the pulse length measured using the ToT mode is a good measure for the pulse height. For the determination of the µτ-product, a 1 mm thick CdTe sensor was flip-chipped to Timepix readout electronics. By connecting every 2 nd pixel, a pixel pitch was 110 x 110 µm 2 was achieved [3]. The system consisting of sensor material and readout electronics will be referred to as an assembly. The contacts were intentionally ohmic, thus the sensor acted as a photoresistor, allowing positive or negative voltage to be applied and the µτ-product to be measured for either electrons or holes. The clock frequency which determines the resolution of the ToT mode was set to 80 MHz, and the energy threshold was set a value of (15.3 ± 0.2) kev. Before these measurements, a threshold equalization with respect to the noise edge was performed to ensure an equal response of the pixels. 2

5 Alpha particles emitted by a 241 Am source with a mean energy of 5.48 MeV were used [11]. The 241 Am source was placed around 4 mm above the sensor surface. In order to reduce the interaction of the alpha particles with the ambient air, the measurements have been performed in a vacuum chamber with a residual pressure of 0.5 mbar. The interaction of the alpha particles with the sensor material happens close below the surface. Therefore, this method of measuring the µτproduct is adaptable for electrons as well as holes, as the contribution from the opposite charge is negligible. Thus, equation 1 can be reduced to one sort of charge carrier, while furthermore being adjusted with the fitting parameter U 0. Q(U) = Q 0 µ τ ( ( )) d 2 (U U d 2 µ τ (U U 0) 1 e 0 ) (2.1) The fitting parameter U 0 takes into account the effect of the energy threshold. Descriptively, it is the minimum voltage that needs to be applied in order to induce enough charge in a single pixel and the pulse height reaches above the energy threshold. For the measurement of the energy spectra the 241 Am source was used, but alpha particles were shielded. An energy calibration of the energy threshold (THL) was performed using the single photon counting mode, which registers the number of photons whose pulses were above the energy threshold. The energy calibration was done by scanning the THL, and therefore by measuring the integral spectrum. Finally, the 241 Am spectrum can be extracted by differentiation. More details on the procedure are described elsewhere [8, 9]. Two Timepix assemblies were used for the measurements. One assembly was the assembly described above with a pixel pitch 110 µm and quasi-ohmic contacts, the other assembly was a M-π-n diode structure described in publication [12] and has a pixel pitch of 55 µm In both cases, electrons were collected and the sensor voltage was set to 500 V. Prior to the measurements, a threshold equalization was performed with respect to the noise edge. Moreover, a moving average was applied to the spectra. In the case of the 110 x 110 µm 2 assembly, the spectrum was obtained by using the complete pixel matrix, while for the 55 x55 µm 2 assembly the analysis area was restricted to 15 x 15 pixels. 3 Results 3.1 Determination of the charge carrier transport properties For the measurement of the µτ-products, it was essential to obtain the position as well as the ToT counts of single clusters created by alpha particles, thus the acquisition time was set to 2 ms, to avoid overlapping of individual clusters. In a first approach, the mean number of pixels in a cluster was measured versus the applied bias voltage (figure 1). As a necessary first step, an algorithm was used to check each cluster, to see if the complete cluster was registered by the pixel matrix, or cut, e.g. by hitting the sensor edge or by hitting dead pixels. Figure 1 shows the mean cluster size for a voltage range between 50 to 500 V in case of electron collection. The maximum cluster size is around 35 pixels for the smallest bias voltage of 50 V. The cluster size decreases with increasing bias voltage due to the shorter drift time and therefore the shorter time for the charge cloud to diffuse. When fitting the data with a function describing 3

6 Figure 1. Number of pixels in a cluster versus the applied bias voltage. The saturation value given by fitting with a function describing an exponential decay results a minimal cluster size of (14.19±0.14) pixels, which corresponds to a diameter of (468±5)µm, presuming circular shaped clusters. an exponential decrease, a minimal cluster size of (14.19 ± 0.14) pixels is obtained. Presuming circular shaped clusters, that corresponds to a mean diameter of the clusters of (468 ± 5) µm. At the beginning of the measurements the mean µτ-product was determined. In order to investigate the functionality of the procedure, as a first step the spatial information of the position of the clusters as neglected. In case of electron collection, the bias voltage was varied between 10 and 70 V, while in case of hole collection it was varied between 25 and 200 V. The resulting values of the reconstructed pulse heights versus bias voltage for electron collection can be seen on the left side in figure 2. By using equation 2, a mean µτ-product for electrons of µ e τ e = (1.9 ± 0.6) cm 2 /V was measured. This value is within the range of values reported by other authors for Acrorad material [1, 13 16]. The saturation value of the reconstructed pulse height in units of ToT counts resulting from the fit is Q 0 = (3753 ± 112). By using the energy calibration for the ToT counts, the energy corresponding to Q 0 is (4.41 ± 0.13) MeV. Thus, a charge collection efficiency for the electron collection is estimated to be (80.4 ± 2.4) %. The slope of the reconstructed pulse heights versus bias voltage, in case of hole collection, can be seen on the right side of figure 2. By fitting the Hecht relationship, a mean µτ-product of µ h τ h = (0.75 ± 0.25) cm 2 /V is obtained. Again, this value is within the range reported by different authors [1, 13 16]. The saturation value was Q 0 = (1591 ± 204), which corresponds to an energy of (1.86 ± 0.24) MeV and results a charge collection efficiency of (34.0 ± 4.4) %. In order to exploit the additional spatial information provided by the position of the clusters, the 128 x 128 pixel matrix was divided into 16 x 16 sub-squares, so each square contained 8 x 8 pixels. An algorithm checked if the complete cluster lay within the square and then reconstructed the overall pulse height by summing up the ToT counts for each single pixel of the cluster. A spatial mapping of the µτ-product was obtained with a spatial resolution of 880 x 880 µm 2. These 4

7 Figure 2. (left) Measured pulse heights versus bias voltage in case of electron collection fitted with equation 2 describing the Hecht relationship. The values resulting from the fit are: µ e τ e = (1.9 ± 0.6) cm 2 /V, Q 0 = (3753 ± 112), U 0 = (7.4 ± 1.7) V. (right) Measured pulse heights versus bias voltage in case of hole collection fitted with equation 2 describing the Hecht relationship. The values resulting from the fit are: µ h τ h = (0,75 ± 0,25) cm 2 /V,Q 0 = (1591 ± 204), U 0 = (13,5 ± 7,3) V. measurements have only been performed for electron collection as this is the typical operation mode due to the better transport properties of the electrons. The resulting mapping of the values for the µτ-product for electrons is shown in figure 3. The mapping reveals slightly higher values for the transport properties in the lower right part of the sensor compared to the mean value at around ( ) cm 2 /V. The transport properties in the upper half of the sensor are slightly reduced to values of around ( ) cm 2 /V. As the charge collection efficiency is connected with the µτ-product, a flood image was taken in order to compare the count rate of the pixels with the transport properties in the respective areas. The flood images were created by using an X-ray source with an acceleration voltage of 20 kv to ensure an absorption close below the sensor surface. The bias voltage was set to 500 V and prior to the measurements a threshold equalization with respect to the noise edge was performed. The flood image is shown in figure 3 on the left side. There is a correlation visible between the counting behavior of the pixels and the transport properties of the charge carriers. 3.2 Energy spectra The energy spectra, obtained using a threshold scan of a 241 Am source, are shown in figure 4. On the left side of figure 4, the energy spectrum obtained with the assembly with quasi-ohmic contacts and a pixel pitch of 110 x 110 µm 2 is shown. The spectrum shown includes all pixels from the pixel matrix. The highest energetic peak with a photon energy of 59.5 kev at a mean THL value of (509.4 ± 0.2) can clearly be resolved. The count rate floor for THL values below can be attributed to events caused by charge sharing of 59.5 kev photons, which are only partially collected in a single pixel. There several peaks are visible for THL values below 350, which can be assigned to the respective photon energies emitted by the source or by fluorescence effects of the CdTe. For example, the self-absorption of the Te-fluorescence (27.4 kev) by the Cd K-edge located at

8 Figure 3. (left) Flood image with an X-ray tube. The acceleration voltage was 20 kv and the bias voltage was 500 V. Prior to the measurements a threshold equalization was performed with respect to the noise edge. (right) µτ-mapping for electrons with a bilinear filter applied. The two dots with higher values for the µτproduct in the µτ-mapping are arising from the fit, which is indicated by relatively high errors for these two values. Figure 4. (left) Spectrum of a 241 Am source obtained by performing a threshold scan with an assembly with quasi-ohmic contacts with a pixel pitch of 110 x 110 µm 2. Several peaks can be resolved. (right) Spectrum of a 241 Am source obtained by performing a threshold scan with an assembly with a M-π-n diode structure with a pixel pitch of 55 x 55 µm 2. No peaks can be resolved, only shoulders are visible. kev can be seen. After being absorbed by the Cd K-edge, photons with an energy of 23.1 kev originating from the Cd-fluorescence will contribute to the spectrum. Using the assignment of the THL values to its respective photon energies, an energy calibration was made, this exhibits a linear dependence (figure 5). The energy resolution of the 59.5 kev peak was determined exploiting the energy calibration, finally resulting a FWHM of the 59.5 kev peak of 5.6 %. Moreover, the position of the noise edge with respect to energy could be measured to be at (5.42 ± 0.14) kev. The THL position of the noise edge was extracted from the information provided by the threshold equalization. When performing the determination of the energetic position of the noise edge with a standard silicon assembly, a value of (4.0 ± 0.3) kev was measured. On the right side of figure 4, the energy spectrum of the 241 Am source is shown, using the a M-π-n diode structure on a Timepix readout electronics with a pixel pitch of 55 x 55 µm 2. Single 6

9 Figure 5. Energy calibration for the µm 2 pixel pitch assembly wih quasi-ohmic contacts. peaks could not be resolved, as it was possible for the assembly with the 110 x 110 µm 2 pixel pitch. Two shoulders are visible, which can most probably be attributed to the 59.5 kev photons and the 23.1 kev Cd-fluorescence. The reason that the peaks cannot be resolved, is assumed to be due to charge sharing, as it was shown by Nillson et al. [17]. 4 Conclusions The transport properties of the charge carriers of a CdTe sensor from Acrorad on a Timepix readout electronics with a pixel pitch of 110 x 110 µm 2 have been investigated. The method is an adaption to pixelated readout systems, of the procedure to determine the µτ-product using alphaparticles. The results obtained are comparable to the results presented by other authors. Moreover, the spatial information was exploited to create a mapping of the µτ-product and a correlation between the counting behavior of the assembly and the distribution of the µτ-products in the corresponding areas. The energy resolution and the energetic position of the noise edge were determined by using a threshold scan of a 241 Am source. The FWHM of the 59.5 kev peak was 5.6 %, the position of the noise edge was at (5.42 ± 0.14) kev. Measurements with a silicon assembly resulted a noise edge energy of (4.0 ± 0.3) kev. Thus, with respect to the noise edge, the CdTe assemblies are comparable to the silicon assemblies. An energy spectrum taken with an assembly with a pixel pitch of 55 x 55 µm 2 revealed, that it s not possible to resolve single peaks, which is most probably due to charge sharing. 7

10 Acknowledgments The work was funded by the,,bundesministerium für Bildung und Forschung (Förderkennzeichen: 05 KS7VFA). The work was carried out within the Medipix collaboration. ( The authors thank Prof. P. H. Butler from the University of Canterbury for proofreading this article. References [1] Acrorad, [2] M. Fiederle, H. Braml and A. Fauler, Development of flip-chip bonding technology for (Cd,Zn)Te, IEEE Trans. Nucl. Sci., 51 (2004) [3] A. Zwerger and M. Fiederle, Medipix2: processing and measurements of GaAs pixel detectors, Nucl. Instrum. Meth. A 576 (2007) 23. [4] Y. Du et al., Temporal response of CZT detectors under intense irradiation, IEEE Trans. Nucl. Sci. 50 (2003) [5] K. Hecht, Zum Mechanismus des lichtelektrischen Primärstromes in isolierenden Kristallen, Z. Physik A 77 (1932) 235. [6] X. Llopart et al., Timepix, a 65k programmable pixel readout chip for arrival time, energy and/or photon counting measurements, Nucl. Instrum. Meth. A 581 (2007) 485. [7] Jan Jakùbek, Precise energy calibration of pixel detector working in time-over-threshold mode, article in press in Nucl. Instrum. Meth. A (corrected proof). [8] M. Fiederle et al., Energy calibration measurements of MediPix2, Nucl. Instrum. Meth. A 591 (2008) 75. [9] Ch. Broennimann et al., The PILATUS 1M detector, J. Synchrotron Rad. 13 (2006) 120. [10] Medipix, [11] WWW Table of Radioactive Isotopes, Radiation Search, [12] D. Greiffenberg et al., Preliminary characterisation of CdTe M-π-n diode structures, article in press in Nucl. Instrum. Meth. A (corrected proof). [13] G. Sato et al., Characterization of CdTe/CdZnTe detectors, IEEE Trans. Nucl. Sci. 49 (2002) [14] P.J. Sellin et al., Drift mobility and mobility-lifetime products in CdTe:Cl grown by the travelling heater method, IEEE Trans. Nucl. Sci. 52 (2005) [15] H. Shiraki et al., Improvement of the productivity in the growth of CdTe single crystal by THM for the new PET system, IEEE Nucl. Sci. Symp. Conf. Rec. (2007) [16] T. Takahasi et al., High resolution Schottky CdTe diode detector, IEEE Trans. Nucl. Sci. 49 (2002) [17] H.-E. Nilsson et al., Simulation of photon and charge transport in X-ray imaging semiconductor sensors, Nucl. Instrum. Meth. A 487 (2002)

NOT FOR DISTRIBUTION JINST_128P_1010 v2

NOT FOR DISTRIBUTION JINST_128P_1010 v2 Pixel sensitivity variations in a CdTe-Medipix2 detector using poly-energetic x-rays R Aamir a, S P Lansley a, b,*, R Zainon a, M Fiederle c, A. Fauler c, D. Greiffenberg c, P H Butler a, d d, e, f, A

More information

Marten Bosma 1, Alex Fauler 2, Michael Fiederle 2 en Jan Visser Nikhef, Amsterdam, The Netherlands 2. FMF, Freiburg, Germany

Marten Bosma 1, Alex Fauler 2, Michael Fiederle 2 en Jan Visser Nikhef, Amsterdam, The Netherlands 2. FMF, Freiburg, Germany Marten Bosma 1, Alex Fauler 2, Michael Fiederle 2 en Jan Visser 1 1. Nikhef, Amsterdam, The Netherlands 2. FMF, Freiburg, Germany Digital Screen film Digital radiography advantages: Larger dynamic range

More information

Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC

Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC R. Bellazzini a,b, G. Spandre a*, A. Brez a, M. Minuti a, M. Pinchera a and P. Mozzo b a INFN Pisa

More information

The Influence of Edge Effects on the Detection Properties of Detector Grade Cadmium Telluride

The Influence of Edge Effects on the Detection Properties of Detector Grade Cadmium Telluride The Influence of Edge Effects on the Detection Properties of Detector Grade Cadmium Telluride M.J. Bosma a, M.G. van Beuzekom a, S. Vähänen b, J.Visser a a. National Institute for Subatomic Physics, Nikhef,

More information

Count rate linearity and spectral response of the Medipix3RX chip coupled to a 300μm silicon sensor under high flux conditions

Count rate linearity and spectral response of the Medipix3RX chip coupled to a 300μm silicon sensor under high flux conditions Journal of Instrumentation OPEN ACCESS Count rate linearity and spectral response of the Medipix3RX chip coupled to a 300μm silicon sensor under high flux conditions To cite this article: E Frojdh et al

More information

MEASUREMENT OF TIMEPIX DETECTOR PERFORMANCE VICTOR GUTIERREZ DIEZ UNIVERSIDAD COMPLUTENSE DE MADRID

MEASUREMENT OF TIMEPIX DETECTOR PERFORMANCE VICTOR GUTIERREZ DIEZ UNIVERSIDAD COMPLUTENSE DE MADRID MEASUREMENT OF TIMEPIX DETECTOR PERFORMANCE VICTOR GUTIERREZ DIEZ UNIVERSIDAD COMPLUTENSE DE MADRID ABSTRACT Recent advances in semiconductor technology allow construction of highly efficient and low noise

More information

Single Photon X-Ray Imaging with Si- and CdTe-Sensors

Single Photon X-Ray Imaging with Si- and CdTe-Sensors Single Photon X-Ray Imaging with Si- and CdTe-Sensors P. Fischer a, M. Kouda b, S. Krimmel a, H. Krüger a, M. Lindner a, M. Löcker a,*, G. Sato b, T. Takahashi b, S.Watanabe b, N. Wermes a a Physikalisches

More information

The Medipix3 Prototype, a Pixel Readout Chip Working in Single Photon Counting Mode with Improved Spectrometric Performance

The Medipix3 Prototype, a Pixel Readout Chip Working in Single Photon Counting Mode with Improved Spectrometric Performance 26 IEEE Nuclear Science Symposium Conference Record NM1-6 The Medipix3 Prototype, a Pixel Readout Chip Working in Single Photon Counting Mode with Improved Spectrometric Performance R. Ballabriga, M. Campbell,

More information

Electron-Bombarded CMOS

Electron-Bombarded CMOS New Megapixel Single Photon Position Sensitive HPD: Electron-Bombarded CMOS University of Lyon / CNRS-IN2P3 in collaboration with J. Baudot, E. Chabanat, P. Depasse, W. Dulinski, N. Estre, M. Winter N56:

More information

236 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 59, NO. 1, FEBRUARY 2012

236 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 59, NO. 1, FEBRUARY 2012 236 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 59, NO. 1, FEBRUARY 2012 Characterization of the H3D ASIC Readout System and 6.0 cm 3-D Position Sensitive CdZnTe Detectors Feng Zhang, Cedric Herman, Zhong

More information

Properties of Irradiated CdTe Detectors O. Korchak M. Carna M. Havranek M. Marcisovsky L. Tomasek V. Vrba

Properties of Irradiated CdTe Detectors O. Korchak M. Carna M. Havranek M. Marcisovsky L. Tomasek V. Vrba E-mail: korchak@fzu.cz M. Carna E-mail: carna@fzu.cz M. Havranek E-mail: havram@fzu.cz M. Marcisovsky E-mail: marcisov@fzu.cz L. Tomasek E-mail: tamasekl@fzu.cz V. Vrba E-mail: vrba@fzu.cz Institute of

More information

High collection efficiency MCPs for photon counting detectors

High collection efficiency MCPs for photon counting detectors High collection efficiency MCPs for photon counting detectors D. A. Orlov, * T. Ruardij, S. Duarte Pinto, R. Glazenborg and E. Kernen PHOTONIS Netherlands BV, Dwazziewegen 2, 9301 ZR Roden, The Netherlands

More information

Charge Loss Between Contacts Of CdZnTe Pixel Detectors

Charge Loss Between Contacts Of CdZnTe Pixel Detectors Charge Loss Between Contacts Of CdZnTe Pixel Detectors A. E. Bolotnikov 1, W. R. Cook, F. A. Harrison, A.-S. Wong, S. M. Schindler, A. C. Eichelberger Space Radiation Laboratory, California Institute of

More information

Simulation and test of 3D silicon radiation detectors

Simulation and test of 3D silicon radiation detectors Simulation and test of 3D silicon radiation detectors C.Fleta 1, D. Pennicard 1, R. Bates 1, C. Parkes 1, G. Pellegrini 2, M. Lozano 2, V. Wright 3, M. Boscardin 4, G.-F. Dalla Betta 4, C. Piemonte 4,

More information

Avalanche statistics and single electron counting with a Timepix-InGrid detector

Avalanche statistics and single electron counting with a Timepix-InGrid detector Avalanche statistics and single electron counting with a Timepix-InGrid detector Michael Lupberger EUDET Annual Meeting 29.09-01.10.2010 DESY, Hamburg, Germany Outline Hardware Timepix Chip + InGrid Experimental

More information

The pixel readout of Micro Patterned Gaseous Detectors

The pixel readout of Micro Patterned Gaseous Detectors The pixel readout of Micro Patterned Gaseous Detectors M. Chefdeville NIKHEF, Kruislaan 409, Amsterdam 1098 SJ, The Netherlands chefdevi@nikhef.nl Abstract. The use of pixel readout chips as highly segmented

More information

Title detector with operating temperature.

Title detector with operating temperature. Title Radiation measurements by a detector with operating temperature cryogen Kanno, Ikuo; Yoshihara, Fumiki; Nou Author(s) Osamu; Murase, Yasuhiro; Nakamura, Masaki Citation REVIEW OF SCIENTIFIC INSTRUMENTS

More information

Study of gain fluctuations with InGrid and TimePix

Study of gain fluctuations with InGrid and TimePix Study of gain fluctuations with InGrid and TimePix Michael Lupberger 5th RD51 Collaboration Meeting 24-27 May 2010 Freiburg, Germany Summary Hardware Timepix Chip + InGrid Experimental setup and calibration

More information

Medipix Project: Characterization and Edge Analysis

Medipix Project: Characterization and Edge Analysis Medipix Project: Characterization and Edge Analysis Matthijs Damen Student ID: 5887453 University of Amsterdam and Nikhef, National Institute for Subatomic Physics July 5, 2011 Abstract This thesis is

More information

Large-Area CdTe Photon-Counting Pixel Detectors

Large-Area CdTe Photon-Counting Pixel Detectors Large-Area CdTe Photon-Counting Pixel Detectors Tilman Donath, Application Scientist 22.6.2015, DIR2015, Ghent DECTRIS Ltd. 5400 Baden Switzerland www.dectris.com Agenda 1. Introduction Hybrid Photon Counting

More information

Soft X-ray sensitivity of a photon-counting hybrid pixel detector with a Silicon sensor matrix.

Soft X-ray sensitivity of a photon-counting hybrid pixel detector with a Silicon sensor matrix. Soft X-ray sensitivity of a photon-counting hybrid pixel detector with a Silicon sensor matrix. A. Fornaini 1, D. Calvet 1,2, J.L. Visschers 1 1 National Institute for Nuclear Physics and High-Energy Physics

More information

Author(s) Osamu; Nakamura, Tatsuya; Katagiri,

Author(s) Osamu; Nakamura, Tatsuya; Katagiri, TitleCryogenic InSb detector for radiati Author(s) Kanno, Ikuo; Yoshihara, Fumiki; Nou Osamu; Nakamura, Tatsuya; Katagiri, Citation REVIEW OF SCIENTIFIC INSTRUMENTS (2 2533-2536 Issue Date 2002-07 URL

More information

2015 JINST 10 C Interconnect and bonding techniques for pixelated X-ray and gamma-ray detectors

2015 JINST 10 C Interconnect and bonding techniques for pixelated X-ray and gamma-ray detectors PUBLISHED BY IOP PUBLISHING FOR SISSA MEDIALAB 10 th INTERNATIONAL CONFERENCE ON POSITION SENSITIVE DETECTORS 7 12 SEPTEMBER 2014, UNIVERSITY OF SURREY, GUILDFORD, SURREY, U.K. RECEIVED: October 7, 2014

More information

Strip Detectors. Principal: Silicon strip detector. Ingrid--MariaGregor,SemiconductorsasParticleDetectors. metallization (Al) p +--strips

Strip Detectors. Principal: Silicon strip detector. Ingrid--MariaGregor,SemiconductorsasParticleDetectors. metallization (Al) p +--strips Strip Detectors First detector devices using the lithographic capabilities of microelectronics First Silicon detectors -- > strip detectors Can be found in all high energy physics experiments of the last

More information

Real Time Pulse Pile-up Recovery in a High Throughput Digital Pulse Processor

Real Time Pulse Pile-up Recovery in a High Throughput Digital Pulse Processor Real Time Pulse Pile-up Recovery in a High Throughput Digital Pulse Processor Paul A. B. Scoullar a, Chris C. McLean a and Rob J. Evans b a Southern Innovation, Melbourne, Australia b Department of Electrical

More information

Implementation of A Nanosecond Time-resolved APD Detector System for NRS Experiment in HEPS-TF

Implementation of A Nanosecond Time-resolved APD Detector System for NRS Experiment in HEPS-TF Implementation of A Nanosecond Time-resolved APD Detector System for NRS Experiment in HEPS-TF LI Zhen-jie a ; MA Yi-chao c ; LI Qiu-ju a ; LIU Peng a ; CHANG Jin-fan b ; ZHOU Yang-fan a * a Beijing Synchrotron

More information

CADMIUM Telluride (CdTe) and Cadmium Zinc Telluride

CADMIUM Telluride (CdTe) and Cadmium Zinc Telluride Evaluation of 5 mm-thick CdTe Detectors from the Company Acrorad Alfred Garson III 1, Ira V. Jung 1, Jeremy Perkins 1, and Henric Krawczynski 1 arxiv:astro-ph/511577v1 18 Nov 25 Abstract Using 2 2.5 cm

More information

X-ray Detectors: What are the Needs?

X-ray Detectors: What are the Needs? X-ray Detectors: What are the Needs? Sol M. Gruner Physics Dept. & Cornell High Energy Synchrotron Source (CHESS) Ithaca, NY 14853 smg26@cornell.edu 1 simplified view of the Evolution of Imaging Synchrotron

More information

Fabrication and Test of Pixelated CZT Detectors with Different Pixel Pitches and Thicknesses

Fabrication and Test of Pixelated CZT Detectors with Different Pixel Pitches and Thicknesses Fabrication and Test of Pixelated CZT Detectors with Different Pixel Pitches and Thicknesses Qiang Li, Alfred III Garson, Paul Dowkontt, Jerrad Martin, Matthias Beilicke, Ira Jung, Michael Groza, Arnold

More information

Production of HPDs for the LHCb RICH Detectors

Production of HPDs for the LHCb RICH Detectors Production of HPDs for the LHCb RICH Detectors LHCb RICH Detectors Hybrid Photon Detector Production Photo Detector Test Facilities Test Results Conclusions IEEE Nuclear Science Symposium Wyndham, 24 th

More information

Citation X-Ray Spectrometry (2011), 40(4): 2. Right final form at

Citation X-Ray Spectrometry (2011), 40(4): 2.   Right final form at TitleSi PIN X-ray photon counter Author(s) Nakaye, Yasukazu; Kawai, Jun Citation X-Ray Spectrometry (2011), 40(4): 2 Issue Date 2011-03-24 URL http://hdl.handle.net/2433/197743 This is the peer reviewed

More information

Pixel hybrid photon detectors

Pixel hybrid photon detectors Pixel hybrid photon detectors for the LHCb-RICH system Ken Wyllie On behalf of the LHCb-RICH group CERN, Geneva, Switzerland 1 Outline of the talk Introduction The LHCb detector The RICH 2 counter Overall

More information

Study of irradiated 3D detectors. University of Glasgow, Scotland. University of Glasgow, Scotland

Study of irradiated 3D detectors. University of Glasgow, Scotland. University of Glasgow, Scotland Department of Physics & Astronomy Experimental Particle Physics Group Kelvin Building, University of Glasgow Glasgow, G12 8QQ, Scotland Telephone: ++44 (0)141 339 8855 Fax: +44 (0)141 330 5881 GLAS-PPE/2002-20

More information

arxiv: v1 [physics.ins-det] 9 Dec 2017

arxiv: v1 [physics.ins-det] 9 Dec 2017 Properties of GaAs:Cr-based Timepix detectors P. Smolyanskiy a,b,1, G. Chelkov a, S. Kotov a, U. Kruchonak a, D. Kozhevnikov a, S. Pospisil c, I. Stekl c, A. Zhemchugov a arxiv:1712.3369v1 [physics.ins-det]

More information

GAMPIX: a Gamma Camera for Homeland Security

GAMPIX: a Gamma Camera for Homeland Security GAMPIX: a Gamma Camera for Homeland Security H. Lemaire a, K. Amgarou b, F. Carrel a, N. Menaa b, V. Schoepff a a CEA, LIST, Gif-sur-Yvette, F-91191, France b AREVA CANBERRA, 1 rue des hérons, Saint-Quentin-en-Yvelines,

More information

Time-of-flight PET with SiPM sensors on monolithic scintillation crystals Vinke, Ruud

Time-of-flight PET with SiPM sensors on monolithic scintillation crystals Vinke, Ruud University of Groningen Time-of-flight PET with SiPM sensors on monolithic scintillation crystals Vinke, Ruud IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you

More information

CALCULATION OF THE MODULATION TRANSFER FUNCTION OF HYBRID SEMICONDUCTOR PIXEL DETECTORS FOR X-RAY IMAGING

CALCULATION OF THE MODULATION TRANSFER FUNCTION OF HYBRID SEMICONDUCTOR PIXEL DETECTORS FOR X-RAY IMAGING CALCULATION OF THE MODULATION TRANSFER FUNCTION OF HYBRID SEMICONDUCTOR PIXEL DETECTORS FOR X-RAY IMAGING L. del Risco Norrlid, C. Rönnqvist, K. Fransson, R. Brenner, L. Gustafsson, F.Edling, S. Kullander

More information

ARTICLE IN PRESS. Nuclear Instruments and Methods in Physics Research A

ARTICLE IN PRESS. Nuclear Instruments and Methods in Physics Research A Nuclear Instruments and Methods in Physics Research A 614 (2010) 308 312 Contents lists available at ScienceDirect Nuclear Instruments and Methods in Physics Research A journal homepage: www.elsevier.com/locate/nima

More information

Lecture 18: Photodetectors

Lecture 18: Photodetectors Lecture 18: Photodetectors Contents 1 Introduction 1 2 Photodetector principle 2 3 Photoconductor 4 4 Photodiodes 6 4.1 Heterojunction photodiode.................... 8 4.2 Metal-semiconductor photodiode................

More information

Today s Outline - January 25, C. Segre (IIT) PHYS Spring 2018 January 25, / 26

Today s Outline - January 25, C. Segre (IIT) PHYS Spring 2018 January 25, / 26 Today s Outline - January 25, 2018 C. Segre (IIT) PHYS 570 - Spring 2018 January 25, 2018 1 / 26 Today s Outline - January 25, 2018 HW #2 C. Segre (IIT) PHYS 570 - Spring 2018 January 25, 2018 1 / 26 Today

More information

THE USE OF CdTe DETECTORS FOR DENTAL X-RAY SPECTROMETRY

THE USE OF CdTe DETECTORS FOR DENTAL X-RAY SPECTROMETRY 2007 International Nuclear Atlantic Conference - INAC 2007 Santos, SP, Brazil, September 30 to October 5, 2007 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-02-1 THE USE OF CdTe DETECTORS

More information

Charge Sharing Effect on 600 µm Pitch Pixelated CZT Detector for Imaging Applications *

Charge Sharing Effect on 600 µm Pitch Pixelated CZT Detector for Imaging Applications * Charge Sharing Effect on 600 µm Pitch Pixelated CZT Detector for Imaging Applications * Yin Yong-Zhi( 尹永智 ), Liu Qi( 刘奇 ), Xu Da-Peng( 徐大鹏 ), Chen Xi-Meng( 陈熙萌 ) School of Nuclear Science and Technology,

More information

http://clicdp.cern.ch Hybrid Pixel Detectors with Active-Edge Sensors for the CLIC Vertex Detector Simon Spannagel on behalf of the CLICdp Collaboration Experimental Conditions at CLIC CLIC beam structure

More information

2 Pixel readout of Micro-Pattern Gas Detectors. The InGrid Concept

2 Pixel readout of Micro-Pattern Gas Detectors. The InGrid Concept 53 Studies of sensitive area for a single InGrid detector A. Chaus a,b, M.Titov b, O.Bezshyyko c, O.Fedorchuk c a Kyiv Institute for Nuclear Research b CEA, Saclay c Taras Shevchenko National University

More information

arxiv: v2 [physics.ins-det] 17 Oct 2015

arxiv: v2 [physics.ins-det] 17 Oct 2015 arxiv:55.9v2 [physics.ins-det] 7 Oct 25 Performance of VUV-sensitive MPPC for Liquid Argon Scintillation Light T.Igarashi, S.Naka, M.Tanaka, T.Washimi, K.Yorita Waseda University, Tokyo, Japan E-mail:

More information

P ILC A. Calcaterra (Resp.), L. Daniello (Tecn.), R. de Sangro, G. Finocchiaro, P. Patteri, M. Piccolo, M. Rama

P ILC A. Calcaterra (Resp.), L. Daniello (Tecn.), R. de Sangro, G. Finocchiaro, P. Patteri, M. Piccolo, M. Rama P ILC A. Calcaterra (Resp.), L. Daniello (Tecn.), R. de Sangro, G. Finocchiaro, P. Patteri, M. Piccolo, M. Rama Introduction and motivation for this study Silicon photomultipliers ), often called SiPM

More information

The on-line detectors of the beam delivery system for the Centro Nazionale di Adroterapia Oncologica(CNAO)

The on-line detectors of the beam delivery system for the Centro Nazionale di Adroterapia Oncologica(CNAO) The on-line detectors of the beam delivery system for the Centro Nazionale di Adroterapia Oncologica(CNAO) A. Ansarinejad1,2, A. Attili1, F. Bourhaleb2,R. Cirio1,2,M. Donetti1,3, M. A. Garella1, S. Giordanengo1,

More information

Large area position-sensitive CVD diamond detectors for X-ray beam monitoring with extreme position resolution

Large area position-sensitive CVD diamond detectors for X-ray beam monitoring with extreme position resolution Large area position-sensitive CVD diamond detectors for X-ray beam monitoring with extreme position resolution M. Pomorski, P. Bergonzo, Ch. Mer, M. Rebisz-Pomorska D. Tromson, N. Tranchant Diamond Sensors

More information

The LHCb Vertex Locator (VELO) Pixel Detector Upgrade

The LHCb Vertex Locator (VELO) Pixel Detector Upgrade Home Search Collections Journals About Contact us My IOPscience The LHCb Vertex Locator (VELO) Pixel Detector Upgrade This content has been downloaded from IOPscience. Please scroll down to see the full

More information

the need for an intensifier

the need for an intensifier * The LLLCCD : Low Light Imaging without the need for an intensifier Paul Jerram, Peter Pool, Ray Bell, David Burt, Steve Bowring, Simon Spencer, Mike Hazelwood, Ian Moody, Neil Catlett, Philip Heyes Marconi

More information

Position-Sensitive Coincidence Detection of Nuclear Reaction Products with Two Timepix Detectors and Synchronized Readout

Position-Sensitive Coincidence Detection of Nuclear Reaction Products with Two Timepix Detectors and Synchronized Readout Position-Sensitive Coincidence Detection of Nuclear Reaction Products with Two Timepix Detectors and Synchronized Readout 1, Vaclav Kraus, Stanislav Pospisil Institute of Experimental and Applied Physics

More information

Update to the Status of the Bonn R&D Activities for a Pixel Based TPC

Update to the Status of the Bonn R&D Activities for a Pixel Based TPC EUDET Update to the Status of the Bonn R&D Activities for a Pixel Based TPC Hubert Blank, Christoph Brezina, Klaus Desch, Jochen Kaminski, Martin Killenberg, Thorsten Krautscheid, Walter Ockenfels, Simone

More information

Development of Integration-Type Silicon-On-Insulator Monolithic Pixel. Detectors by Using a Float Zone Silicon

Development of Integration-Type Silicon-On-Insulator Monolithic Pixel. Detectors by Using a Float Zone Silicon Development of Integration-Type Silicon-On-Insulator Monolithic Pixel Detectors by Using a Float Zone Silicon S. Mitsui a*, Y. Arai b, T. Miyoshi b, A. Takeda c a Venture Business Laboratory, Organization

More information

A Readout ASIC for CZT Detectors

A Readout ASIC for CZT Detectors A Readout ASIC for CZT Detectors L.L.Jones a, P.Seller a, I.Lazarus b, P.Coleman-Smith b a STFC Rutherford Appleton Laboratory, Didcot, OX11 0QX, UK b STFC Daresbury Laboratory, Warrington WA4 4AD, UK

More information

ITk silicon strips detector test beam at DESY

ITk silicon strips detector test beam at DESY ITk silicon strips detector test beam at DESY Lucrezia Stella Bruni Nikhef Nikhef ATLAS outing 29/05/2015 L. S. Bruni - Nikhef 1 / 11 Qualification task I Participation at the ITk silicon strip test beams

More information

Lecture 2. Part 2 (Semiconductor detectors =sensors + electronics) Segmented detectors with pn-junction. Strip/pixel detectors

Lecture 2. Part 2 (Semiconductor detectors =sensors + electronics) Segmented detectors with pn-junction. Strip/pixel detectors Lecture 2 Part 1 (Electronics) Signal formation Readout electronics Noise Part 2 (Semiconductor detectors =sensors + electronics) Segmented detectors with pn-junction Strip/pixel detectors Drift detectors

More information

CDTE and CdZnTe detector arrays have been recently

CDTE and CdZnTe detector arrays have been recently 20 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 44, NO. 1, FEBRUARY 1997 CMOS Low-Noise Switched Charge Sensitive Preamplifier for CdTe and CdZnTe X-Ray Detectors Claudio G. Jakobson and Yael Nemirovsky

More information

Full characterization tests of Micromegas with elongated pillars

Full characterization tests of Micromegas with elongated pillars University of Würzburg Full characterization tests of Micromegas with elongated pillars B. Alvarez1 Gonzalez, L. Barak1, J. Bortfeldt1, F. Dubinin3, G. Glonti1, F. Kuger1,2, P. Iengo1, E. Oliveri1, J.

More information

Pixel sensors with different pitch layouts for ATLAS Phase-II upgrade

Pixel sensors with different pitch layouts for ATLAS Phase-II upgrade Pixel sensors with different pitch layouts for ATLAS Phase-II upgrade Different pitch layouts are considered for the pixel detector being designed for the ATLAS upgraded tracking system which will be operating

More information

arxiv: v2 [physics.ins-det] 14 Jul 2015

arxiv: v2 [physics.ins-det] 14 Jul 2015 April 11, 2018 Compensation of radiation damages for SOI pixel detector via tunneling arxiv:1507.02797v2 [physics.ins-det] 14 Jul 2015 Miho Yamada 1, Yasuo Arai and Ikuo Kurachi Institute of Particle and

More information

Studies on MCM D interconnections

Studies on MCM D interconnections Studies on MCM D interconnections Speaker: Peter Gerlach Department of Physics Bergische Universität Wuppertal D-42097 Wuppertal, GERMANY Authors: K.H.Becks, T.Flick, P.Gerlach, C.Grah, P.Mättig Department

More information

Multianode Photo Multiplier Tubes as Photo Detectors for Ring Imaging Cherenkov Detectors

Multianode Photo Multiplier Tubes as Photo Detectors for Ring Imaging Cherenkov Detectors Multianode Photo Multiplier Tubes as Photo Detectors for Ring Imaging Cherenkov Detectors F. Muheim a edin]department of Physics and Astronomy, University of Edinburgh Mayfield Road, Edinburgh EH9 3JZ,

More information

Soft X-Ray Silicon Photodiodes with 100% Quantum Efficiency

Soft X-Ray Silicon Photodiodes with 100% Quantum Efficiency PFC/JA-94-4 Soft X-Ray Silicon Photodiodes with 1% Quantum Efficiency K. W. Wenzel, C. K. Li, D. A. Pappas, Raj Kordel MIT Plasma Fusion Center Cambridge, Massachusetts 2139 USA March 1994 t Permanent

More information

High energy X-ray emission driven by high voltage circuit system

High energy X-ray emission driven by high voltage circuit system Journal of Physics: Conference Series OPEN ACCESS High energy X-ray emission driven by high voltage circuit system To cite this article: M Di Paolo Emilio and L Palladino 2014 J. Phys.: Conf. Ser. 508

More information

Design and Test of a 65nm CMOS Front-End with Zero Dead Time for Next Generation Pixel Detectors

Design and Test of a 65nm CMOS Front-End with Zero Dead Time for Next Generation Pixel Detectors Design and Test of a 65nm CMOS Front-End with Zero Dead Time for Next Generation Pixel Detectors L. Gaioni a,c, D. Braga d, D. Christian d, G. Deptuch d, F. Fahim d,b. Nodari e, L. Ratti b,c, V. Re a,c,

More information

Chapter 2 Solid-State Detectors for Small-Animal Imaging

Chapter 2 Solid-State Detectors for Small-Animal Imaging Chapter 2 Solid-State Detectors for Small-Animal Imaging Paolo Russo and Alberto Del Guerra 1 Introduction Semiconductor detector technology, initially developed for high energy physics applications, has

More information

Use of a pulsed laser to study properties of CdZnTe pixel detectors

Use of a pulsed laser to study properties of CdZnTe pixel detectors Use of a pulsed laser to study properties of CdZnTe pixel detectors A. E. Bolotnikov, S. E. Boggs, W. R. Cook, F. A. Harrison, S. M. Schindler Space Radiation Laboratory, California Institute of Technology

More information

Improvement of the CdTe Diode Detectors using a Guard-ring Electrode

Improvement of the CdTe Diode Detectors using a Guard-ring Electrode Improvement of the CdTe Diode Detectors using a Guard-ring Electrode Kazuhiro Nakazawa, Kousuke Oonuki, Takaaki Tanaka, Yoshihito Kobayashi, Ken ichi Tamura, Takefumi Mitani, Goro Sato, Shin Watanabe,

More information

Pulse Shape Analysis for a New Pixel Readout Chip

Pulse Shape Analysis for a New Pixel Readout Chip Abstract Pulse Shape Analysis for a New Pixel Readout Chip James Kingston University of California, Berkeley Supervisors: Daniel Pitzl and Paul Schuetze September 7, 2017 1 Table of Contents 1 Introduction...

More information

Ultra fast single photon counting chip

Ultra fast single photon counting chip Ultra fast single photon counting chip P. Grybos, P. Kmon, P. Maj, R. Szczygiel Faculty of Electrical Engineering, Automatics, Computer Science and Biomedical Engineering AGH University of Science and

More information

AIDA-2020 Advanced European Infrastructures for Detectors at Accelerators

AIDA-2020 Advanced European Infrastructures for Detectors at Accelerators Grant Agreement No: 654168 AIDA-2020 Advanced European Infrastructures for Detectors at Accelerators Horizon 2020 Research Infrastructures project AIDA -2020 MILESTONE REPORT SMALL-SIZE PROTOTYPE OF THE

More information

Highly Miniaturised Radiation Monitor (HMRM) Status Report. Yulia Bogdanova, Nicola Guerrini, Ben Marsh, Simon Woodward, Rain Irshad

Highly Miniaturised Radiation Monitor (HMRM) Status Report. Yulia Bogdanova, Nicola Guerrini, Ben Marsh, Simon Woodward, Rain Irshad Highly Miniaturised Radiation Monitor (HMRM) Status Report Yulia Bogdanova, Nicola Guerrini, Ben Marsh, Simon Woodward, Rain Irshad HMRM programme aim Aim of phase A/B: Develop a chip sized prototype radiation

More information

The PERDaix Detector. Thomas Kirn I. Physikalisches Institut B. July 5 th 2011, 6 th International Conference on New Developments In Photodetection

The PERDaix Detector. Thomas Kirn I. Physikalisches Institut B. July 5 th 2011, 6 th International Conference on New Developments In Photodetection Proton Electron Radiation Detector Aix la Chapelle The PERDaix Detector Thomas Kirn I. Physikalisches Institut B July 5 th 2011, 6 th International Conference on New Developments In Photodetection Motivation

More information

MEASUREMENT AND MODELING OF BLOCKING CONTACTS FOR CADMIUM TELLURIDE GAMMA RAY DETECTORS

MEASUREMENT AND MODELING OF BLOCKING CONTACTS FOR CADMIUM TELLURIDE GAMMA RAY DETECTORS MEASUREMENT AND MODELING OF BLOCKING CONTACTS FOR CADMIUM TELLURIDE GAMMA RAY DETECTORS A Thesis presented to the Electrical Engineering Faculty of California Polytechnic State University, San Luis Obispo

More information

Interpixel crosstalk in a 3D-integrated active pixel sensor for x-ray detection

Interpixel crosstalk in a 3D-integrated active pixel sensor for x-ray detection Interpixel crosstalk in a 3D-integrated active pixel sensor for x-ray detection The MIT Faculty has made this article openly available. Please share how this access benefits you. Your story matters. Citation

More information

Improvement of Energy Resolutions for Planar TlBr Detectors Using the Digital Pulse Processing Method

Improvement of Energy Resolutions for Planar TlBr Detectors Using the Digital Pulse Processing Method CYRIC Annual Report 2009 III. 5. Improvement of Energy Resolutions for Planar TlBr Detectors Using the Digital Pulse Processing Method Tada T. 1, Tanaka T. 2, Kim S.-Y. 1, Wu Y. 1, Hitomi K. 1, Yamazaki

More information

Physics Laboratory Scattering of Photons from Electrons: Compton Scattering

Physics Laboratory Scattering of Photons from Electrons: Compton Scattering RR Oct 2001 SS Dec 2001 MJ Oct 2009 Physics 34000 Laboratory Scattering of Photons from Electrons: Compton Scattering Objective: To measure the energy of high energy photons scattered from electrons in

More information

Multi-Element Si Sensor with Readout ASIC for EXAFS Spectroscopy 1

Multi-Element Si Sensor with Readout ASIC for EXAFS Spectroscopy 1 Multi-Element Si Sensor with Readout ASIC for EXAFS Spectroscopy 1 Gianluigi De Geronimo a, Paul O Connor a, Rolf H. Beuttenmuller b, Zheng Li b, Antony J. Kuczewski c, D. Peter Siddons c a Microelectronics

More information

Instruction manual and data sheet ipca h

Instruction manual and data sheet ipca h 1/15 instruction manual ipca-21-05-1000-800-h Instruction manual and data sheet ipca-21-05-1000-800-h Broad area interdigital photoconductive THz antenna with microlens array and hyperhemispherical silicon

More information

Receiver Design for Passive Millimeter Wave (PMMW) Imaging

Receiver Design for Passive Millimeter Wave (PMMW) Imaging Introduction Receiver Design for Passive Millimeter Wave (PMMW) Imaging Millimeter Wave Systems, LLC Passive Millimeter Wave (PMMW) sensors are used for remote sensing and security applications. They rely

More information

Role of guard rings in improving the performance of silicon detectors

Role of guard rings in improving the performance of silicon detectors PRAMANA c Indian Academy of Sciences Vol. 65, No. 2 journal of August 2005 physics pp. 259 272 Role of guard rings in improving the performance of silicon detectors VIJAY MISHRA, V D SRIVASTAVA and S K

More information

50 MHz Voltage-to-Frequency Converter

50 MHz Voltage-to-Frequency Converter Journal of Physics: Conference Series OPEN ACCESS 50 MHz Voltage-to-Frequency Converter To cite this article: T Madden and J Baldwin 2014 J. Phys.: Conf. Ser. 493 012008 View the article online for updates

More information

Moderne Teilchendetektoren - Theorie und Praxis 2. Dr. Bernhard Ketzer Technische Universität München SS 2013

Moderne Teilchendetektoren - Theorie und Praxis 2. Dr. Bernhard Ketzer Technische Universität München SS 2013 Moderne Teilchendetektoren - Theorie und Praxis 2 Dr. Bernhard Ketzer Technische Universität München SS 2013 7 Signal Processing and Acquisition 7.1 Signals 7.2 Amplifier 7.3 Electronic Noise 7.4 Analog-to-Digital

More information

Partial Replication of Storms/Scanlan Glow Discharge Radiation

Partial Replication of Storms/Scanlan Glow Discharge Radiation Partial Replication of Storms/Scanlan Glow Discharge Radiation Rick Cantwell and Matt McConnell Coolescence, LLC March 2008 Introduction The Storms/Scanlan paper 1 presented at the 8 th international workshop

More information

Applications of a Pixellated Detection System to Digital Mammography

Applications of a Pixellated Detection System to Digital Mammography Applications of a Pixellated Detection System to Digital Mammography Valeria Rosso Dipartimento di Fisica, Universita di Pisa and Sezione INFN Pisa, Italy + valeria.rosso@pi.infn.it Outline The detection

More information

Simulation of High Resistivity (CMOS) Pixels

Simulation of High Resistivity (CMOS) Pixels Simulation of High Resistivity (CMOS) Pixels Stefan Lauxtermann, Kadri Vural Sensor Creations Inc. AIDA-2020 CMOS Simulation Workshop May 13 th 2016 OUTLINE 1. Definition of High Resistivity Pixel Also

More information

Fundamentals of CMOS Image Sensors

Fundamentals of CMOS Image Sensors CHAPTER 2 Fundamentals of CMOS Image Sensors Mixed-Signal IC Design for Image Sensor 2-1 Outline Photoelectric Effect Photodetectors CMOS Image Sensor(CIS) Array Architecture CIS Peripherals Design Considerations

More information

AIDA-2020 Advanced European Infrastructures for Detectors at Accelerators. Deliverable Report. CERN pixel beam telescope for the PS

AIDA-2020 Advanced European Infrastructures for Detectors at Accelerators. Deliverable Report. CERN pixel beam telescope for the PS AIDA-2020-D15.1 AIDA-2020 Advanced European Infrastructures for Detectors at Accelerators Deliverable Report CERN pixel beam telescope for the PS Dreyling-Eschweiler, J (DESY) et al 25 March 2017 The AIDA-2020

More information

DALLA LUCE VISIBILE AI RAGGI X: NUOVI RIVELATORI DI IMMAGINI PER RAGGI X A DISCRIMINAZIONE IN ENERGIA ED APPLICAZIONI

DALLA LUCE VISIBILE AI RAGGI X: NUOVI RIVELATORI DI IMMAGINI PER RAGGI X A DISCRIMINAZIONE IN ENERGIA ED APPLICAZIONI DALLA LUCE VISIBILE AI RAGGI X: NUOVI RIVELATORI DI IMMAGINI PER RAGGI X A DISCRIMINAZIONE IN ENERGIA ED APPLICAZIONI D. Pacella ENEA - Frascati LIMS, Frascati 14-15 ottobre 2015 Come per la fotografia:

More information

Performance characterization of a novel thin position-sensitive avalanche photodiode-based detector for high resolution PET

Performance characterization of a novel thin position-sensitive avalanche photodiode-based detector for high resolution PET 2005 IEEE Nuclear Science Symposium Conference Record M11-126 Performance characterization of a novel thin position-sensitive avalanche photodiode-based detector for high resolution PET Jin Zhang, Member,

More information

SPECTROMETRIC CHARACTERISTIC IMPROVEMENT OF CdTe DETECTORS*

SPECTROMETRIC CHARACTERISTIC IMPROVEMENT OF CdTe DETECTORS* SPECTROMETRIC CHARACTERISTIC IMPROVEMENT OF CdTe DETECTORS* Abstract V. I. Ivanov, V. Garbusin, P. G. Dorogov, A. E. Loutchanski, V. V. Kondrashov Baltic Scientific Instruments, RITEC Ltd., P. O. Box 25,

More information

portable scanner based on monolithic semi-insulating

portable scanner based on monolithic semi-insulating The 6 th International Workshop on Radiation Imaging Detectors, 25-29 July 2004, Glasgow Digital X-ray X portable scanner based on monolithic semi-insulating insulating GaAs detectors: General description

More information

Time Resolution Studies with Timepix3 Assemblies with Thin Silicon Pixel Sensors

Time Resolution Studies with Timepix3 Assemblies with Thin Silicon Pixel Sensors CLICdp-Pub-19-1 15 January 19 Time Resolution Studies with Timepix3 Assemblies with Thin Silicon Pixel Sensors N. Alipour Tehrani ú, D. Dannheim ú, A. Fiergolski ú, D. Hynds ú1), W. Klempt ú, X. Llopart

More information

SUPPLEMENTARY INFORMATION

SUPPLEMENTARY INFORMATION SUPPLEMENTARY INFORMATION Supplementary Information Real-space imaging of transient carrier dynamics by nanoscale pump-probe microscopy Yasuhiko Terada, Shoji Yoshida, Osamu Takeuchi, and Hidemi Shigekawa*

More information

arxiv: v2 [physics.ins-det] 14 Jan 2009

arxiv: v2 [physics.ins-det] 14 Jan 2009 Study of Solid State Photon Detectors Read Out of Scintillator Tiles arxiv:.v2 [physics.ins-det] 4 Jan 2 A. Calcaterra, R. de Sangro [], G. Finocchiaro, E. Kuznetsova 2, P. Patteri and M. Piccolo - INFN,

More information

SILICON DRIFT DETECTORS (SDDs) [1] with integrated. Preliminary Results on Compton Electrons in Silicon Drift Detector

SILICON DRIFT DETECTORS (SDDs) [1] with integrated. Preliminary Results on Compton Electrons in Silicon Drift Detector Preliminary Results on Compton Electrons in Silicon Drift Detector T. Çonka-Nurdan, K. Nurdan, K. Laihem, A. H. Walenta, C. Fiorini, B. Freisleben, N. Hörnel, N. A. Pavel, and L. Strüder Abstract Silicon

More information

Results of FE65-P2 Pixel Readout Test Chip for High Luminosity LHC Upgrades

Results of FE65-P2 Pixel Readout Test Chip for High Luminosity LHC Upgrades for High Luminosity LHC Upgrades R. Carney, K. Dunne, *, D. Gnani, T. Heim, V. Wallangen Lawrence Berkeley National Lab., Berkeley, USA e-mail: mgarcia-sciveres@lbl.gov A. Mekkaoui Fermilab, Batavia, USA

More information

Gamma Ray Spectroscopy with NaI(Tl) and HPGe Detectors

Gamma Ray Spectroscopy with NaI(Tl) and HPGe Detectors Nuclear Physics #1 Gamma Ray Spectroscopy with NaI(Tl) and HPGe Detectors Introduction: In this experiment you will use both scintillation and semiconductor detectors to study γ- ray energy spectra. The

More information

Recent Development and Study of Silicon Solid State Photomultiplier (MRS Avalanche Photodetector)

Recent Development and Study of Silicon Solid State Photomultiplier (MRS Avalanche Photodetector) Recent Development and Study of Silicon Solid State Photomultiplier (MRS Avalanche Photodetector) Valeri Saveliev University of Obninsk, Russia Vienna Conference on Instrumentation Vienna, 20 February

More information

Gas scintillation Glass GEM detector for high-resolution X-ray imaging and CT

Gas scintillation Glass GEM detector for high-resolution X-ray imaging and CT Gas scintillation Glass GEM detector for high-resolution X-ray imaging and CT Takeshi Fujiwara 1, Yuki Mitsuya 2, Hiroyuki Takahashi 2, and Hiroyuki Toyokawa 2 1 National Institute of Advanced Industrial

More information