1 N-SIM guide NIKON IMAGING KING S COLLEGE LONDON

2 Starting-up / Shut-down The NSIM hardware is calibrated after system warm-up occurs. It is recommended that you turn-on the system for at least 30 minutes prior to use for optimal performance. A digital thermometer attached to the microscope stand can be used to gauge when the system is ready the temperature of the microscope will rise by approximately 2 o C when all the system components and PC are switched on. To switch on the system: 1. Switch on the laser bed (see figures 2-4) and select either multi or single to choose the appropriate optical fibre depending on whether you are doing 2D/3D (multi) or TIRF- SIM (single). 2. Switched on the XY stage, transmitted light and 561nm laser (by turning the key) as shown in figure Switch on the remaining lasers (by turning the keys) and Intensilight unit as shown in figure Switch on the PC; doing so switches on the SIM illuminator and EMCCD camera. 5. The microscope stand (Ti) should already by powered on as should be the piezo z drive (see figure 5). Figure 1 To shut-down the system: Power 1. Close NIS Elements. 2. Shut-down the PC. 3. Turn-off the lasers and lamps. Figure 2 Figure 3

3 561nm laser XY-stage Transmitted light 405, 488, 640nm lasers Fluorescence lamp Figure 3 Figure 4 Figure 5 Figure 6

4 A note about user rights This NSIM system is driven through the use of optical configurations to enable new users to get up and running quickly and to make day-to-day use of the system straight-forward. In order to prevent modifications to the base set of configurations each new user will be given common privileges in NIS-Elements. These privileges are restricted so that shared optical configurations cannot be modified. However, common users can still make new configurations and mark them as private so that other users cannot change the configuration. To enable laser power and camera settings to be saved into the configuration (required when using ND Acquisition) it is also necessary to make a change to the General settings for each user. This change is made as follows: 1. Click the cog icon or go to Edit Options. 2. In General change Optical Configuration to save all camera settings to optical configuration automatically. Each time you change the laser power and camera settings these values will automatically get saved to the optical configuration. Note these settings are not applied globally they are saved only in your instance of the optical configuration.

5 Viewing your sample through the eye-pieces Since the N-SIM system is essentially a wide-field microscope we choose to drive the system using the Optical configuration panel ( OC panel ). The OC panel is used to hold shortcuts to microscope configurations for performing a specific task. The configurations are grouped on the panel according to mode of operation they control. To view your sample through the eye-pieces use the Eyes group of configurations. These configurations: 1. Sets the Light Path to E100 (eyes) 2. Sets turret 2 to the empty position. 3. Sets turret1 to the appropriate filter set. 4. Use the Intensilight remote control (see figure 6) to open the lamp shutter. 5. Also not that the laser interlock is engaged when the light path is set to E100 which prevents control of the lasers on the N-SIM pad. Switching to an optical configuration which uses the camera (R100) releases the interlock. Figure 7 Figure 8

6 Image acquisition SIM acquisitions are possible with both the 60x water and 100x oil immersion objectives. The choice between the two is driven by how thick or how deep into your sample you need to image. A depth of approximately 20µm and 7µm is possible with the 60x and 100x objectives respectively. If you plan to use the TIRF- SIM mode you will need the 100x objective. The optimal resolution is achieved with thin samples (low out-of-focus light) and the 100x objective. The microscope stand (Ti-E) includes two filter turrets (see the Ti Pad); turret 2 is used when performing SIM acquisitions while an empty position (no filter cube) is selected in turret 1. The filter blocks in turret 2 are specially designed and factory aligned filter sets to minimised chromatic misalignment for multi-channel images. The example shown here is the Ti Pad configuration for a 488nm SIM acquisition. SIM acquisitions are performed using the N-SIM pad. This is a plugin for NIS- Elements that drives both acquisitions and image reconstruction. Optical configurations have been created for each channel of each type of acquisition. 3D-SIM, single channel, single z-plane 1. Select the appropriate optical configuration (3D-SIM 405, 3D-SIM 488, 3D-SIM 561, 3D-SIM 640) depending on the channel you want to acquire. 2. Make sure the appropriate grating is selected in the N-SIM pad (100 EX V-R for the 100x objective lens). 3. Make sure to select the correct fibre input Multi for multi-mode fibre. 4. Make sure the correct fibre is selected on the N-SIM illuminator. 5. Set a laser power (for the appropriate laser wavelength) and camera exposure time (on the DU-897 Settings ) to obtain an appropriate signal level. This will be ~4000 ADU with the camera in 14-bit mode and ~12,000 ADU with the camera in 16-bit mode (recommended see DU-897 Settings ). It is important for final image quality that you do not exceed these values. 6. Make sure to select 3D-SIM from the drop-down menu below the Capture button. 7. Press the Capture button to acquire a single z-plane. 3D-SIM ND Acquisition ND acquisition is used to acquire multi-dimensional data, for example a z-stack in multiple channels. Select the appropriate tab(s) to add dimensions to the dataset. Here we will describe how to acquire a z-stack in two channels. Figure 9 Figure 10 Figure 11

7 1. Check the λ (lambda) tab to add optical configurations to the acquisition. Simply click the checkbox on the first line and then select the optical configuration from the dropdown menu. Note that acquisition is sequential. 2. To add another channel, check the box on the next line and select a configuration from the dropdown menu. Here we have selected 3D-SIM 488 and 3D-SIM If you require more than one z-slice, check the Z tab and choose one of three acquisition modes: a. Top and bottom b. Symmetric (selected in the figure) c. Asymmetric 4. Choose a step size of at least 120nm (or multiples thereof) ignoring the suggested step size of 200nm and set the range or number of steps. 5. Make sure the appropriate grating is selected in the N-SIM pad (100 EX V-R for the 100x objective lens). 6. Make sure to select the correct fibre input Multi for multi-mode fibre. 7. Make sure the correct fibre is selected on the N-SIM illuminator. 8. Set a laser power (for the appropriate laser wavelength) and camera exposure time (on the DU-897 Settings ) to obtain an appropriate signal level in each channel to be recorded. This will be ~4000 ADU with the camera in 14-bit mode and ~12,000 ADU with the camera in 16-bit mode (recommended see DU-897 Settings ). It is important for final image quality that you do not exceed these values. 9. Make sure to select 3D-SIM from the drop-down menu below the Capture button. 10. Press Run now. Figure 12 Figure 13

8 3D-SIM image reconstruction During 3D-SIM acquisition a total of fifteen images are captured, 5 images with varying grating phase at 3 different grating rotations (separated by 120 o ). This of course limits the rate at which frames can be recorded which is an important consideration for live cell imaging. For 3D-SIM 1 frame per second is possible while for 2D- and TIRF-SIM a slightly faster rate of around 1.2 frames per second is possible since a total of 9 images are recorded in these modes. Figure 14 Image reconstruction works best when the recommended number of grey levels are collected and when good grating contrast is achieved. The final can however be improved by adjusting the image reconstruction parameters. These are accessed on the N-SIM pad; choose between Preview, Slice, Stack and Batch reconstruction. See the Advanced section of this document for a description of the reconstruction parameters. To reconstruct: 1. Tune the reconstruction parameters by clicking the Param button below the Reconstruct Slice button on the N-SIM pad. 2. Adjust the Illumination modulation contrast. Use larger values for situations where the grating contrast is poor: for example for thick samples with lots of out-of-focus light. Start with a value of 0.5 when the grating lines are clearly visible (see the example opposite). 3. Adjust the High resolution noise suppression to remove any high frequency reconstruction artefacts. This can occur for samples with lots of out-of-focus light. Start with a value of 0.25 for samples with good grating contrast. Use of large values of this parameter will compromise the resolution of fine structure in the image. Figure 15

9 4. Adjust the Out of focus blur suppression parameter. This is effectively optical sectioning and can help to remove any outof-focus light from the image. Start with a value of 0.3; larger values suppress out-of-focus light to a greater extent. Note that this parameter is unavailable for 2D-SIM, TIRF-SIM and Stack reconstruction. 5. If reconstructing ND acquisition data you can tune the parameters for each channel by ticking Use different settings for each Channel and select each channel from the dropdown list. Tick Use best K-Vectors found in: and choose All Frames from the drop-down list if you have multiple z-slices. 6. Press Reconstruct to perform image reconstruction. 7. Press Apply to store these parameters enabling them to be used with the Preview Slice and Reconstruct Slice buttons on the N-SIM pad. Figure 16 After image reconstruction the final image will typically have a large baseline offset. To remove this offset you will need to adjust the black level on the LUT for the image. Click the right mouse button and select LUTs from the Visualisation controls in the resultant context menu. Figure 17

10 Figure 18

11 Capturing a widefield image It is possible to reconstruct widefield images from the 15 phase images captured during a SIM acquisition but this requires splitting the raw data using a macro. The easier method is to record a new image with the microscope in its widefield mode. In this mode only the zero order of the illumination grating is transmitted through the microscope resulting in epi-like excitation of the sample. A single z plane can be captured use the SIM Full Illumination configurations on the OC panel (click the button corresponding to your channel of choice either Full 405, Full 488, Full 561 or Full 640 ). This will configure the N-SIM pad as follows: 1. On the N-SIM pad the Widefield option is selected from the dropdown menu under the Live button. 2. The Widefield option under the capture button is selected. You should then set the exposure time ( DU-897 settings ) and laser power to achieve the same number of ADUs as captured in the SIM data (either 4,000 or 12,000 for 14- or 16-bit modes respectively). Pressing the Capture button on the N-SIM pad captures and image at the current z- plane and with the current laser power and camera settings. Figure 19 Multi-channel widefield images can be captured using ND Acquisition. Simply select the λ (lambda) tab, click the checkbox on the first row and select the SIM Full Illumination configuration corresponding to your channel of choice. Continue to add channels by clicking the checkboxes on subsequent rows. Synchronizing SIM and widefield images The Synchronizer is a tool that allows you to compare two images by locking the pan and zoom. The widefield image is as captured and has a format of 512x512 pixels whereas the SIM image has extra pixels introduced during the image reconstruction and is 1024x1024. So, before using the synchronizer you will need to resize the widefield image to match the SIM image. On the NIS Elements toolbar go to Image Size resize to set the widefield image to 1024x1024. Click the right mouse button and select Synchronizer from the Visualisation controls. Add both images to the Synchronizer to lock the pan and zoom. Figure 20 Figure 21

12 Notes about image reconstruction The intention of this section of the notes is to give a brief (non-mathematical) description of the process of SIM image reconstruction and the parameters used. A more in-depth discussion of SIM can be found in the Structured Illumination Methods chapter of the Handbook of confocal biological microscopy by James Pawley. The two most important concepts relating to SIM are spatial frequency and the modulation transfer function (MTF). Images with small details have high spatial frequency and conversely large details (and out-of-focus light) have large spatial frequencies. The spatial frequency content of an image can be examined by performing a Fourier transform (FFT) into frequency space. Low frequency components lie at the centre of the FFT image (see examples in figure 22). Figure22 In a microscope the resolution and performance can be characterised by the MTF which is a measurement of how specimen contrast is transferred through the optical system to the image plane. The microscope objective lens is the most important component contributing to the MTF and forms a pass band for the microscope with a cut-off at a spatial frequency defined by the numerical aperture (NA) of the lens. The example shows that when an image of a periodic structure is transmitted through the microscope the high frequency details get blurred and there is a resultant loss of contrast. The purpose of SIM is to restore the high frequency detail that gets lost when transmitted through a microscope objective lens. Figure23 This is achieved by illuminating the sample with patterned or structured light and relying on the Moire effect to transmit high frequency information through the objective lens. The Moire effect is a simple pseudo-interference effect where two overlapping, slightly mismatched, patterns produces a low frequency fringes. The key concept is that the low frequency fringes are unique to the overlap between the two patterns (see figure 24). In SIM the spatial frequency of the illumination pattern is arranged to be just on the cut-off frequency of the objective lens while the Moire fringes produced by the overlap of the sample and the illumination pattern has a low enough frequency to be transmitted by the lens. By adjusting the position of the illumination pattern in real space (3 rotations and 5 lateral shifts or phases for each rotation for 3D Figure 24 SIM) it is possible to extend the cut-off limit of the objective lens by almost a factor of 2. Knowledge of the rotation and the phase shift allows the pass band captured in each image to be moved to its true position in frequency space (as illustrated in figure 25). A reverse Fourier transform then produces an image with the high frequency information restored.

13 Figure 25 Image reconstruction parameters The following is adapted from the N-SIM manual. Illumination modulation contrast This parameter allows adjustment of the relative amplitudes of the MTF for the different phase images captured during a SIM acquisition. Matching the separate image components captured at each grating rotation and phase in frequency space is dependent of the contrast of the illumination pattern. The purpose of this parameter is to improve image reconstruction when the contrast of the illumination pattern is low (i.e. when stripe visibility is low). This can happen in highly scattering samples or where there is a large amount of out-of-focus light. Conventional High frequency noise suppression During image reconstruction image filtering and apodization takes place. The first is to remove noise from the image and the second is remove artefacts that can be introduced during the filtering process. The High frequency noise suppression parameter can be tuned in order to remove these high frequency artefacts but at the expense of high frequency information in the image. Large value MTF Super resolution 2D SIM 3D SIM Spatial frequency Small value MTF Spatial frequency Small value MTF Spatial frequency Figure 26 Large value MTF Spatial frequency Figure 27

14 Out of focus blur suppression Out-of-focus light can be removed during image reconstruction by tuning this parameter. Put simply larger values produce more optical sectioning. Small value Z Large value Z X,Y X,Y Spread of intensity distribution of luminescent point in real space Figure 28 Recommended starting parameters Figure 29

Nikon SIM-E & A1-R System

Nikon SIM-E & A1-R System Nikon SIM-E & A1-R System USER GUIDE LSU Health Sciences Center Shreveport Research Core Facility June 01 2017 Chaowei Shang 1 Table of Content 1. Start Up the System... Page 3 Hardware and microscope

More information

LSM 510 Meta Training Notes

LSM 510 Meta Training Notes LSM 510 Meta Training Notes Turning on the system Turn on X-Cite power supply. This supplies light for epifluorescence for viewing your samples through the microscope. Turn on the remote control switch.

More information

LSM 510 Training Notes

LSM 510 Training Notes LSM 510 Training Notes Turning on the system Turn on the arc lamp, found on the bench top left of the microscope. This supplies light for epifluorescence for viewing your samples through the microscope.

More information

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling Zeiss LSM 510 Confocor III Training Notes Center for Cell Analysis & Modeling Confocor 3 Start Up Go to System Module Turn on Main Switch, System/ PC, and Components Switches Do you need the arc lamp?

More information

Nikon Instruments Europe

Nikon Instruments Europe Nikon Instruments Europe Recommendations for N-SIM sample preparation and image reconstruction Dear customer, We hope you find the following guidelines useful in order to get the best performance out of

More information

Diskovery Spinning Disk Guide

Diskovery Spinning Disk Guide Diskovery Spinning Disk Guide qbi.microscopy@uq.edu.au Getting started The microscope and its peripherals (Fig. 1a) should always be turned on, but if they are not, turn them on in the following way: 1.

More information

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017.

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017. Nikon AZ100 Laser Scanning Macro Confocal Microscope Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin May 2017 Contents 1 Introduction 2 2 Hardware - Startup 2 3 Software/Operation 4 3.1 Multidimensional

More information

Microscopy from Carl Zeiss

Microscopy from Carl Zeiss Microscopy from Carl Zeiss Contents Page Contents... 1 Introduction... 1 Starting the System... 2 Introduction to ZEN Efficient Navigation... 5 Setting up the microscope... 10 Configuring the beam path

More information

LSM 780 Confocal Microscope Standard Operation Protocol

LSM 780 Confocal Microscope Standard Operation Protocol LSM 780 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Sign on log sheet according to Actual start time 2. Check Compressed Air supply for the air table 3. Switch

More information

Training Guide for Carl Zeiss LSM 510 META Confocal Microscope

Training Guide for Carl Zeiss LSM 510 META Confocal Microscope Training Guide for Carl Zeiss LSM 510 META Confocal Microscope AIM 4.2 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn ON Components and System/PC switches

More information

Title: Nikon A1R Confocal User Manual

Title: Nikon A1R Confocal User Manual Title: Nikon A1R Confocal User Manual Date of first issue: 23/10/2015 Date of review: Version: Admin For assistance or to report an issue Office: CG.07 or CG.05 Email: Igmm-imaginghelpdesk@igmm.ed.ac.uk

More information

Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope

Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope ZEN 2009 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn Chameleon TiS laser key from Standby

More information

Leica SPEII confocal microscope. Short Manual

Leica SPEII confocal microscope. Short Manual Leica SPEII confocal microscope Short Manual Switching ON sequence: 1. Turn on the Workstation under the bench (top, far right). 2. Turn on the Supply Unit - Laser box (big green switch first and then

More information

Nikon Eclipse Ti A1-A Confocal Operating Manual. Start-up. Microscope

Nikon Eclipse Ti A1-A Confocal Operating Manual. Start-up. Microscope Nikon Eclipse Ti A1-A Confocal Operating Manual Start-up 1. Turn on Excite Fluorescent light power supply- metal halide. a. Cool down as for mercury bulb b. Wheel closed liquid light guide 2. Turn on power

More information

Quick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s

Quick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s LSM 5 MP, LSM 510 and LSM 510 META M i c r o s c o p y f r o m C a r l Z e i s s Quick Guide Laser Scanning Microscopes LSM Software ZEN 2007 August 2007 We make it visible. Contents Page Contents... 1

More information

LSM 710 Confocal Microscope Standard Operation Protocol

LSM 710 Confocal Microscope Standard Operation Protocol LSM 710 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Switch on Main power switch 2. Switch on System / PC power button 3. Switch on Components power button 4.

More information

EPIFLUORESCENCE &/OR BRIGHTFIELD MICROSCOPY

EPIFLUORESCENCE &/OR BRIGHTFIELD MICROSCOPY EPIFLUORESCENCE &/OR BRIGHTFIELD MICROSCOPY TURN ON THE FOLLOWING EQUIPMENT The fluorescent light (if needed) The power strip for the microscope and accessories The CoolSNAP HQ camera on the right (Turn

More information

Zeiss 780 Training Notes

Zeiss 780 Training Notes Zeiss 780 Training Notes Turn on Main Switch, System PC and Components Switches 780 Start up sequence Do you need the argon laser (458, 488, 514 nm lines)? Yes Turn on the laser s main power switch and

More information

3. are adherent cells (ie. cells in suspension are too far away from the coverslip)

3. are adherent cells (ie. cells in suspension are too far away from the coverslip) Before you begin, make sure your sample... 1. is seeded on #1.5 coverglass (thickness = 0.17) 2. is an aqueous solution (ie. fixed samples mounted on a slide will not work - not enough difference in refractive

More information

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope AIM 4.2 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Verify that main power switches on the

More information

Nikon A1R. Multi-Photon & Laser Scanning Confocal Microscope. Kyle Marchuk Adam Fries Jordan Briscoe Kaitlin Corbin. April 2017.

Nikon A1R. Multi-Photon & Laser Scanning Confocal Microscope. Kyle Marchuk Adam Fries Jordan Briscoe Kaitlin Corbin. April 2017. Nikon A1R Multi-Photon & Laser Scanning Confocal Microscope Kyle Marchuk Adam Fries Jordan Briscoe Kaitlin Corbin April 2017 Contents 1 Introduction 2 2 Start-Up 2 3 Imaging 4 3.1 Sample Alignment...........................................

More information

Training Guide for Carl Zeiss LSM 880 with AiryScan FAST

Training Guide for Carl Zeiss LSM 880 with AiryScan FAST Training Guide for Carl Zeiss LSM 880 with AiryScan FAST ZEN 2.3 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2018) Power ON Routine 1 2 Turn ON Main Switch from the remote control

More information

Zeiss 880 Training Notes Zen 2.3

Zeiss 880 Training Notes Zen 2.3 Zeiss 880 Training Notes Zen 2.3 1 Turn on the HXP 120V Lamp 2 Turn on Main Power Switch Turn on the Systems PC Switch Turn on the Components Switch. 3 4 5 Turn on the PC and log into your account. Start

More information

Practical work no. 3: Confocal Live Cell Microscopy

Practical work no. 3: Confocal Live Cell Microscopy Practical work no. 3: Confocal Live Cell Microscopy Course Instructor: Mikko Liljeström (MIU) 1 Background Confocal microscopy: The main idea behind confocality is that it suppresses the signal outside

More information

Leica TCS SP8 Quick Start Guide

Leica TCS SP8 Quick Start Guide Leica TCS SP8 Quick Start Guide Leica TCS SP8 System Overview Start-Up Procedure 1. Turn on the CTR Control Box, EL6000 fluorescent light source for the microscope stand. 2. Turn on the Scanner Power

More information

Nikon A1Rsi Confocal Start-Up Sequence

Nikon A1Rsi Confocal Start-Up Sequence 1. Turn the key on the Nikon LUN-V Laser Launch. Nikon A1Rsi Confocal Start-Up Sequence 2. Press the button the left side of the A1Rsi Controller unit. 3. Turn on the power strip underneath the microscope.

More information

NIS-Elements C (For CONFOCAL MICROSCOPE A1) Instructions (Ver. 4.40)

NIS-Elements C (For CONFOCAL MICROSCOPE A1) Instructions (Ver. 4.40) M487E 15.4.NF.17 (1/4) *M487EN17* NIS-Elements C (For CONFOCAL MICROSCOPE A1) Instructions (Ver. 4.40) Preface Thank you for purchasing the Nikon products. This instruction manual has been prepared for

More information

Training Guide for Leica SP8 Confocal/Multiphoton Microscope

Training Guide for Leica SP8 Confocal/Multiphoton Microscope Training Guide for Leica SP8 Confocal/Multiphoton Microscope LAS AF v3.3 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn ON power switch for epifluorescence

More information

Leica SP8 TCS Users Manual

Leica SP8 TCS Users Manual Version : 07/08/0 Leica SP8 TCS Users Manual Start up:. Turn the PC Microscope, Scanner Power, Laser Power, and the Laser Emission key to on (bottom right of desk).. Turn on the fluorescent lamp (top left

More information

Things to check before start-up.

Things to check before start-up. Byeong Cha Page 1 11/24/2009 Manual for Leica SP2 Confocal Microscope Enter you name, the date, the time, and the account number in the user log book. Things to check before start-up. Make sure that your

More information

Swept-Field User Guide

Swept-Field User Guide Swept-Field User Guide Note: for more details see the Prairie user manual at http://www.prairietechnologies.com/resources/software/prairieview.html Please report any problems to Julie Last (jalast@wisc.edu)

More information

ZEISS LSM510META confocal manual

ZEISS LSM510META confocal manual ZEISS LSM510META confocal manual Switching on the system 1) Switch on the Remote Control button located on the table to the right of the microscope. This is the main switch for the whole system including

More information

LSM 800 Confocal Microscope Standard Operation Protocol

LSM 800 Confocal Microscope Standard Operation Protocol LSM 800 Confocal Microscope Standard Operation Protocol Turning on the system 1. Switch on the Main switch (labeled 1 and 2 ) mounted on the wall. 2. Turn the Laser Key (labeled 3 ) 90 clockwise for power

More information

START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7

START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7 Leica DMI AF6000LX Table of contents START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7 ACQUIRE MODULE 6 SETTING THE LIGHTPATH 6

More information

Operating Instructions for Zeiss LSM 510

Operating Instructions for Zeiss LSM 510 Operating Instructions for Zeiss LSM 510 Location: GNL 6.312q (BSL3) Questions? Contact: Maxim Ivannikov, maivanni@utmb.edu 1 Attend A Complementary Training Before Using The Microscope All future users

More information

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING Operations Guide Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING Nikon Eclipse 90i Operations Guide STARTUP Startup Powering Up Fluorescence

More information

Zeiss LSM 780 Protocol

Zeiss LSM 780 Protocol Zeiss LSM 780 Protocol 1) System Startup F Please note the sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.

More information

Leica SP8 TCS Users Manual

Leica SP8 TCS Users Manual Leica SP8 TCS Users Manual Follow the procedure for start up and log on as posted in the lab. Please log on with your account only and do not share your password with anyone. We track and confirm usage

More information

Boulevard du Temple Daguerrotype (Paris,1838) a busy street? Nyquist sampling for movement

Boulevard du Temple Daguerrotype (Paris,1838) a busy street? Nyquist sampling for movement Boulevard du Temple Daguerrotype (Paris,1838) a busy street? Nyquist sampling for movement CONFOCAL MICROSCOPY BioVis Uppsala, 2017 Jeremy Adler Matyas Molnar Dirk Pacholsky Widefield & Confocal Microscopy

More information

Leica TCS SP8 Quick Start Guide

Leica TCS SP8 Quick Start Guide Leica TCS SP8 Quick Start Guide Leica TCS SP8 System Overview Start-Up Procedure 1. Turn on the CTR Control Box, Fluorescent Light for the microscope stand. 2. Turn on the Scanner Power (1) on the front

More information

Nikon C1si Spectral Laser Scanning Confocal Microscope. User Guide

Nikon C1si Spectral Laser Scanning Confocal Microscope. User Guide Nikon C1si Spectral Laser Scanning Confocal Microscope User Guide Contents: C1Si Turn-On/ShutDown Procedures... 2 Overview... 4 Setup for epi-illumination to view through the eyepieces:... 5 Setup for

More information

Zeiss LSM 880 Protocol

Zeiss LSM 880 Protocol Zeiss LSM 880 Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.

More information

ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL

ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL START THE SYSTEM... 2 START ZEN SOFTWARE... 3 SET THE TEMPERATURE AND THE CO2 CONTROLLERS... OBSERVATION AT OCULARS... 5 STATIF PRESENTATION... 6 ACQUIRE ONE

More information

User manual for Olympus SD-OSR spinning disk confocal microscope

User manual for Olympus SD-OSR spinning disk confocal microscope User manual for Olympus SD-OSR spinning disk confocal microscope Ved Prakash, PhD. Research imaging specialist Imaging & histology core University of Texas, Dallas ved.prakash@utdallas.edu Once you open

More information

3D light microscopy techniques

3D light microscopy techniques 3D light microscopy techniques The image of a point is a 3D feature In-focus image Out-of-focus image The image of a point is not a point Point Spread Function (PSF) 1D imaging 2D imaging 3D imaging Resolution

More information

QUICKSTART GUIDE: WIDEFIELD WF3 Zeiss Cell Observer Live Cell Imaging System (SAF, ROOM 409) Imperial College London

QUICKSTART GUIDE: WIDEFIELD WF3 Zeiss Cell Observer Live Cell Imaging System (SAF, ROOM 409) Imperial College London Imperial College London Facility for Imaging by Light Microscopy QUICKSTART GUIDE: WIDEFIELD WF3 Zeiss Cell Observer Live Cell Imaging System (SAF, ROOM 409) Observing Life As It Happens Startup procedure...

More information

QUICKSTART GUIDE: WIDEFIELD HWF1 Zeiss Cell Observer Live Cell Imaging System (HAMMERSMITH, L BLOCK, ROOM 314) Imperial College London

QUICKSTART GUIDE: WIDEFIELD HWF1 Zeiss Cell Observer Live Cell Imaging System (HAMMERSMITH, L BLOCK, ROOM 314) Imperial College London Imperial College London Facility for Imaging by Light Microscopy QUICKSTART GUIDE: WIDEFIELD HWF1 Zeiss Cell Observer Live Cell Imaging System (HAMMERSMITH, L BLOCK, ROOM 314) Observing Life As It Happens

More information

Contents. Introduction

Contents. Introduction Contents Page Contents... 1 Introduction... 1 Starting the System... 2 Introduction to ZEN Efficient Navigation... 5 Setting up the microscope... 10 Configuring the beam path and lasers... 12 Scanning

More information

Leica DMI 4000 tutorial

Leica DMI 4000 tutorial Leica DMI 4000 tutorial Before using the Leica DMI 4000, You will need to put down your name on the reservation system = 1 Welcome to the Leica DMI4000 Microscope tutorial How to start up the system (p.3)

More information

Using Autofocus in NIS-Elements

Using Autofocus in NIS-Elements Using Autofocus in NIS-Elements Overview This technical note provides an overview of the available autofocus routines in NIS-Elements, and describes the necessary steps for using the autofocus functions.

More information

OPERATING INSTRUCTIONS

OPERATING INSTRUCTIONS Zeiss LSM 510 M eta Confocal M icroscope OPERATING INSTRUCTIONS Starting the System: 1. Turn the black knob on the laser box one-quarter turn from Off to On. You will hear the laser cooling mechanisms

More information

Super Resolution Microscope N-SIM E. Super Resolution Microscope

Super Resolution Microscope N-SIM E. Super Resolution Microscope Super Resolution Microscope N-SIM E Super Resolution Microscope Explore Nano world with Nikon N-SIM E is a streamlined, affordable superresolution system that provides double the resolution of conventional

More information

FLUORESCENCE MICROSCOPY. Matyas Molnar and Dirk Pacholsky

FLUORESCENCE MICROSCOPY. Matyas Molnar and Dirk Pacholsky FLUORESCENCE MICROSCOPY Matyas Molnar and Dirk Pacholsky 1 The human eye perceives app. 400-700 nm; best at around 500 nm (green) Has a general resolution down to150-300 μm (human hair: 40-250 μm) We need

More information

Title: Leica SP5 Confocal User Manual

Title: Leica SP5 Confocal User Manual Title: Leica SP5 Confocal User Manual Date of first issue: 23/10/2015 Date of review: Version: Admin For assistance or to report an issue Office: CG07 or 05 Email: Igmm-imaginghelpdesk@igmm.ed.ac.uk Website:

More information

Thor. Nikon Inverted Structured Illumination (N-SIM) & A1 Laser Scanning Confocal Microscope User Guide. v. 1.0 (12/2015)

Thor. Nikon Inverted Structured Illumination (N-SIM) & A1 Laser Scanning Confocal Microscope User Guide. v. 1.0 (12/2015) Thor Nikon Inverted Structured Illumination (N-SIM) & A1 Laser Scanning Confocal Microscope User Guide v. 1.0 (12/2015) Objectives 10x/0.45NA (WD 4 mm) Plan Apo 20x/0.75NA (WD 1 mm) Plan Apo 40x/0.95NA

More information

Zeiss Deconvolution Microscope: A Quick Guide

Zeiss Deconvolution Microscope: A Quick Guide Zeiss Deconvolution Microscope: A Quick Guide Start-up Uncover microscope. Do not put dust cover on the floor. Plug in both cameras. The default camera is the AxioCam HRm (monochrome camera) for fluorescence

More information

Zeiss Axio Imager.A1 manual

Zeiss Axio Imager.A1 manual Zeiss Axio Imager.A1 manual Power-up protocol 1. Mercury lamp 2. Power strip on shelf 3. Computer The Mercury lamp should always be first-on and last-off. This prevents any electrical surges caused by

More information

TRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO

TRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO TRAINING MANUAL Multiphoton Microscopy LSM 510 META-NLO September 2010 Multiphoton Microscopy Training Manual Multiphoton microscopy is only available on the LSM 510 META-NLO system. This system is equipped

More information

Confocal and 2-photon Imaging. October 15, 2010

Confocal and 2-photon Imaging. October 15, 2010 Confocal and 2-photon Imaging October 15, 2010 Review Optical Elements Adapted from Sluder & Nordberg 2007 Review Optical Elements Collector Lens Adapted from Sluder & Nordberg 2007 Review Optical Elements

More information

Nasmyth Ultraview Vox User Protocol

Nasmyth Ultraview Vox User Protocol Nasmyth Ultraview Vox User Protocol Switch on all wall sockets labelled Nasmyth, switch camera on (power supply located on table behind monitor), switch on laser switch in laser rack, switch computer on

More information

The following units are required for an ApoTome imaging workstation:

The following units are required for an ApoTome imaging workstation: V VKN fã~öé=^åèìáëáíáçå=jççìäéë= ^éçqçãé= déåéê~ä= The ApoTome software module controls the ApoTome hardware (control box and slider) and coordinated image acquisition using a digital camera, such as the

More information

Overview. About other software. Administrator password. 58. UltraVIEW VoX Getting Started Guide

Overview. About other software. Administrator password. 58. UltraVIEW VoX Getting Started Guide Operation 58. UltraVIEW VoX Getting Started Guide Overview This chapter outlines the basic methods used to operate the UltraVIEW VoX system. About other software Volocity places great demands on the computer

More information

Nikon Eclipse Ti2-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol

Nikon Eclipse Ti2-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol Nikon Eclipse Ti-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol Please sign on the log sheet before switching on system. Turn on system Turn on A only if confocal mode or laser

More information

TRAINING MANUAL. Olympus FV1000

TRAINING MANUAL. Olympus FV1000 TRAINING MANUAL Olympus FV1000 September 2014 TABLE OF CONTENTS A. Start-Up Procedure... 1 B. Visual Observation under the Microscope... 1 C. Image Acquisition... 4 A brief Overview of the Settings...

More information

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center For any questions or concerns, please contact: Linda Nieman lnieman@mgh.harvard.edu Office: (617) 643-9684 Cell: (512) 565-8076 Chenyue

More information

Quick Start Guide. Leica SP5 X

Quick Start Guide. Leica SP5 X Quick Start Guide Leica SP5 X Please note: Some of the information in this guide was taken from Leica Microsystems Leica TCS SP5 LAS AF Guide for New Users. This work is licensed under the Creative Commons

More information

LEICA TCS SP5 AOBS TANDEM USER MANUAL

LEICA TCS SP5 AOBS TANDEM USER MANUAL LEICA TCS SP5 AOBS TANDEM USER MANUAL STARTING THE SYSTEM...2 THE LAS AF SOFTWARE...3 THE «ACQUIRE» MENU...5 CHOOSE AND CREATE A SETTING...6 THE CONTROL PANEL...8 THE DMI6000B MICROSCOPE...10 ACQUIRE ONE

More information

REMEMBER: You have 5GB of disk space on this microscope. Check before you start if you have room for your experiment. If not delete your old data.

REMEMBER: You have 5GB of disk space on this microscope. Check before you start if you have room for your experiment. If not delete your old data. 1 Use of the Zeiss LSM 510 Inverted Firstly please be aware that this microscope should be treated with respect and care at all times. Rules of use: This Microscope can only be used by Masters by Research

More information

Bi/BE 227 Winter Assignment #3. Adding the third dimension: 3D Confocal Imaging

Bi/BE 227 Winter Assignment #3. Adding the third dimension: 3D Confocal Imaging Bi/BE 227 Winter 2016 Assignment #3 Adding the third dimension: 3D Confocal Imaging Schedule: Jan 20: Assignment Jan 20-Feb 8: Work on assignment Feb 10: Student PowerPoint presentations. Goals for this

More information

Cell Biology and Bioimaging Core

Cell Biology and Bioimaging Core Cell Biology and Bioimaging Core Leica TCS SP5 Operating Instructions Starting up the instrument 1. First, log in the log book located on the confocal desk. Include your name, your lab s PI, an account

More information

ZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide

ZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide ZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide Matyas Molnar, Biovis 2016 Starting the microscpe 1. Check the microscope if everything looks clean and normal. If not, report it in the logbook.

More information

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009 Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009 Introduction of Fluoresence Confocal Microscopy The first confocal microscope was invented by Princeton

More information

Guide to Confocal 5. Starting session

Guide to Confocal 5. Starting session Guide to Confocal 5 Remember that when booking and before starting session you can check for any problems at https://www.bris.ac.uk/biochemistry/uobonly/cif/index.html Starting session Switch on microscope

More information

Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each

Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each lens with a 1 Airy unit pinhole setting. Many water lenses

More information

More fancy SPIM, Even fancier SPIM

More fancy SPIM, Even fancier SPIM More fancy SPIM, Even fancier SPIM Last class Light sheet microscopy Fancy SPIM (ispim, dspim, etc ) This class Multi camera SPIM SIM SPIM Bessels d x,y = λ em 2 NA d z = 2 NA λ ex + n(1 cosθ λ em 1 IsoView

More information

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer Page 1 of 11 Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer The Aramis Raman system is a software selectable multi-wavelength Raman system with mapping capabilities with a 400mm monochromator and

More information

Internal Medicine Imaging Core Emory University Department of Medicine

Internal Medicine Imaging Core Emory University Department of Medicine Internal Medicine Imaging Core Emory University Department of Medicine 1 OPERATION OF THE ZEISS LSM 510 META YOU MUST SIGN UP TO USE THE MICROSCOPE OR COMPUTER EVERY TIME NO EXCEPTIONS Before attempting

More information

Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope

Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope ZEN 2012 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 If you require fluorescence imaging,

More information

3D light microscopy techniques

3D light microscopy techniques 3D light microscopy techniques The image of a point is a 3D feature In-focus image Out-of-focus image The image of a point is not a point Point Spread Function (PSF) 1D imaging 1 1 2! NA = 0.5! NA 2D imaging

More information

1 Set up the confocal light path for imaging a green dye (Alexa488-EGFP). For example, the

1 Set up the confocal light path for imaging a green dye (Alexa488-EGFP). For example, the 1 Set up the confocal light path for imaging a green dye (Alexa488-EGFP). For example, the light path as shown here using the 488 nm LASER (Laser Unit 1) reflecting off of the 405/488 nm Dichroic mirror

More information

High resolution extended depth of field microscopy using wavefront coding

High resolution extended depth of field microscopy using wavefront coding High resolution extended depth of field microscopy using wavefront coding Matthew R. Arnison *, Peter Török #, Colin J. R. Sheppard *, W. T. Cathey +, Edward R. Dowski, Jr. +, Carol J. Cogswell *+ * Physical

More information

Zeiss AxioImager.Z2 Brightfield Protocol

Zeiss AxioImager.Z2 Brightfield Protocol Zeiss AxioImager.Z2 Brightfield Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge

More information

Olympus Fluoview 1000S Spectral Confocal Microscope Introduction to the NRI-MCDB Microscopy Facility Spectral Confocal Microscope

Olympus Fluoview 1000S Spectral Confocal Microscope Introduction to the NRI-MCDB Microscopy Facility Spectral Confocal Microscope Olympus Fluoview 1000S Spectral Confocal Microscope Introduction to the NRI-MCDB Microscopy Facility Spectral Confocal Microscope Improved Optics More Lasers 405 diode 440 diode 488 Argon 515 Argon 559

More information

Point Spread Function Estimation Tool, Alpha Version. A Plugin for ImageJ

Point Spread Function Estimation Tool, Alpha Version. A Plugin for ImageJ Tutorial Point Spread Function Estimation Tool, Alpha Version A Plugin for ImageJ Benedikt Baumgartner Jo Helmuth jo.helmuth@inf.ethz.ch MOSAIC Lab, ETH Zurich www.mosaic.ethz.ch This tutorial explains

More information

1 Co Localization and Working flow with the lsm700

1 Co Localization and Working flow with the lsm700 1 Co Localization and Working flow with the lsm700 Samples -1 slide = mousse intestine, Dapi / Ki 67 with Cy3/ BrDU with alexa 488. -1 slide = mousse intestine, Dapi / Ki 67 with Cy3/ no BrDU (but with

More information

Motorized Axio Observer Start-up instructions

Motorized Axio Observer Start-up instructions Start-up instructions 1. If using fluorescence turn on Fluorescent light source. TL light Source (Hal 100) 2. Turn on microscope using switch on lower left side of the microscope. 3. If imaging, turn on

More information

Wide-Field TCSPC FLIM with bh SPC-150 N TCSPC System and Photek FGN Detector

Wide-Field TCSPC FLIM with bh SPC-150 N TCSPC System and Photek FGN Detector Wide-Field TCSPC FLIM with bh SPC-150 N TCSPC System and Photek FGN 392-1000 Detector Abstract: We present a wide-field TCSPC FLIM system consisting of a position-sensitive MCP PMT of the delay-line type,

More information

b. Turn the power switch and key to on position for blue laser.

b. Turn the power switch and key to on position for blue laser. OLYMPUS FLUOVIEW 300 CONFOCAL MICOSCOPE OPERATION PROCEDURE 1. Turn ON microscope in this order: 1) Turn on mercury lamp (Note: once the mercury lamp is turned off, DO NOT turn it back on for at least

More information

STEM Spectrum Imaging Tutorial

STEM Spectrum Imaging Tutorial STEM Spectrum Imaging Tutorial Gatan, Inc. 5933 Coronado Lane, Pleasanton, CA 94588 Tel: (925) 463-0200 Fax: (925) 463-0204 April 2001 Contents 1 Introduction 1.1 What is Spectrum Imaging? 2 Hardware 3

More information

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5.

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5. Smith College August 2010 Operating Checklist for using the Laser Scanning Confocal Microscope Leica TCS SP5. CONTENT, page no. Startup, 1 Initial set-up, 1 Software, 2 Microscope Specimen observation

More information

SPINNING DISK CSU-X1 USER MANUAL

SPINNING DISK CSU-X1 USER MANUAL SPINNING DISK CSU-X1 USER MANUAL Starting the temperature controller... 2 Starting the CO2 controller... 3 Start the spinning disk... 4 Sample observation with the oculars... 5 Spatial sampling, Pixel

More information

KEYENCE VKX LASER-SCANNING CONFOCAL MICROSCOPE Standard Operating Procedures (updated Oct 2017)

KEYENCE VKX LASER-SCANNING CONFOCAL MICROSCOPE Standard Operating Procedures (updated Oct 2017) KEYENCE VKX LASER-SCANNING CONFOCAL MICROSCOPE Standard Operating Procedures (updated Oct 2017) 1 Introduction You must be trained to operate the Laser-scanning confocal microscope (LSCM) independently.

More information

Microscopy Live Animal Imaging

Microscopy Live Animal Imaging Microscopy Live Animal Imaging A collaborative environment that provides the knowledge, instruments, and expertise needed to visualize life at scales ranging from single molecules to entire animals. Project

More information

Widefield-NikonEclipseTE200-ORCA Nikon Eclipse TE200 Inverted Microscope with Hamamatsu 1394 Orca-ER Cooled CCD Camera and Micromanager Software

Widefield-NikonEclipseTE200-ORCA Nikon Eclipse TE200 Inverted Microscope with Hamamatsu 1394 Orca-ER Cooled CCD Camera and Micromanager Software Widefield-NikonEclipseTE200-ORCA Nikon Eclipse TE200 Inverted Microscope with Hamamatsu 1394 Orca-ER Cooled CCD Camera and Micromanager Software September 2007 Check website for most current User Guide

More information

Instructions for the Experiment

Instructions for the Experiment Instructions for the Experiment Excitonic States in Atomically Thin Semiconductors 1. Introduction Alongside with electrical measurements, optical measurements are an indispensable tool for the study of

More information

Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev. Microscopy course, Michmoret Dec 2005

Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev. Microscopy course, Michmoret Dec 2005 Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev Why use confocal microscopy? Principles of the laser scanning confocal microscope. Image resolution. Manipulating the

More information

Enhancement of the lateral resolution and the image quality in a line-scanning tomographic optical microscope

Enhancement of the lateral resolution and the image quality in a line-scanning tomographic optical microscope Summary of the PhD thesis Enhancement of the lateral resolution and the image quality in a line-scanning tomographic optical microscope Author: Dudás, László Supervisors: Prof. Dr. Szabó, Gábor and Dr.

More information

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center For any questions or concerns, please contact: Linda Nieman lnieman@mgh.harvard.edu Office: (617) 643-9684 Cell: (512) 565-8076 Chenyue

More information

Bio 407. Applied microscopy. Introduction into light microscopy. José María Mateos. Center for Microscopy and Image Analysis

Bio 407. Applied microscopy. Introduction into light microscopy. José María Mateos. Center for Microscopy and Image Analysis Center for Microscopy and Image Analysis Bio 407 Applied Introduction into light José María Mateos Fundamentals of light Compound microscope Microscope composed of an objective and an additional lens (eyepiece,

More information

Adding scale bars to images

Adding scale bars to images Adding scale bars to images ImageJ/FIJI IVision Nikon Nis- Elements Imaris 8 Imagej/FIJI 1. If you are viewing images acquired in Micro- Manager and you selected the lens before acquiring the images then

More information