TRAINING MANUAL. Olympus FV1000

Size: px
Start display at page:

Download "TRAINING MANUAL. Olympus FV1000"

Transcription

1 TRAINING MANUAL Olympus FV1000 September 2014

2 TABLE OF CONTENTS A. Start-Up Procedure... 1 B. Visual Observation under the Microscope... 1 C. Image Acquisition... 4 A brief Overview of the Settings... 4 Image Acquisition for Single dye... 6 Image Acquisition for Double and Triple labelled samples... 8 Image Acquisition for samples labelled with more than 3 fluorophores... 9 Adding the DIC or Bright Field Channel Z-stacks D. Reload the Image Parameters E. Improving your Image F. Other Important Functions...17 G. Shut Down Procedure H. Appendix I : Tile Images OlympusTraining Manual

3 A. Start-Up Procedure 1. Turn on the mercury burner (#1) 3 2. Turn the Microscope Box on (#2) 3. Turn the Scan Head box on (#3) 4. Turn the Stage box on - back right (#4) Turn the Diode laser power switch on (it has to be on even if you are not using this laser) 6. Turn the Argon laser power switch and key on Turn the HeNe1 laser key on (this is the 543 laser) 8. Turn the computer on and login 9. Double click in the software icon and login again B. Visual Observation under the Microscope Focus Knob - there are 2 buttons right beside this knob: ESC = puts the objective all the way down Focus = toggles between fine and coarse mode for the focus focus up and down these buttons control the intensity of the transmitted light OlympusTraining Manual 1

4 B. Visual Observation under the Microscope continued... You can select an objective lens and a fluorescent filter cube by using the hand switch. You can also use the microscope controller box in the software to choose the objective. PS: if you can't find the microscope controller window, go to DEVICE in the main tab of the Olympus Fluoview software. Open the drop down menu and select "Microscope controller" Here are the specifications of the objectives in this system: Please note that the only oil objective in this system is the 60X. All the others are dry. To load your sample: 1- choose the objective you want to use 2- open the doors of the chamber 3- move the microscope's arm back (be careful not to go too far otherwise the condensor will hit the chamber) 4- lower the objective before loading your sample to avoid damaging the objective. Use the functions focus up and down in the microscope to do that 5- load your sample 6- move the stage around using the stage controller at your left side 7- move the microscope s arm back To focus your sample: 1-If you want to use transmitted light to focus your sample, select the transmitted light button in the image acquisition control window OlympusTraining Manual 2

5 B. Visual Observation under the Microscope continued... To focus your sample continued: 2- if you want to acquire a DIC image of your sample, make sure you match the DIC prism with the objective used. Position #3 = 20X Position #4 = 40X Position #5 = 63X Just rotate the prisms to put the right one in place You don't need to worry about this if you are only observing your sample 3- clockwise rotation of the focus knob will move the objective down and counter-clockwise will move it up. Right bellow, to the side, of the focus knob, there is a button to toggle between fine and coarse mode for the focus. 4- To control the intensity of the transmitted light, change the voltage settings in the software 5- if you want to focus your sample using fluorescence light, select the EPI lamp button in the image acquisition control window 6- now, select the filter cube you want to use (according to your fluorophore) in the hand switch controller (left side of the microscope) or in the microscope controller window 7- open the fluorescence light shutter to let the light comes through the objective (to open, move the slider to the left) 8- you can control the intensity of the light by opening or closing this shutter 9- use the hand switch controller to switch between different filters 10- to turn the light off, close the shutter or click the EPI lamp button again OlympusTraining Manual 3

6 C. Image Acquisition A brief overview of the settings: 1- Acquisition Setting, Data manager and Explorer Windows Scan mode (unidirectional and bidirectional) Number of pixels Scan speed (the speed the laser dwells in each pixel). As lower as better the image, however higher the bleaching. As a starting point you can use 4.0us/Pixel In the Area you can control the zoom and rotation of your Zoom in function Laser output adjustment The time scan tab controls the parameters for time lapse In the Microscope tab you can select the objective you want to use, set the parameters for Z stack and change the focus (bigger arrow heads are for coarse focus adjustment and fine arrow heads are for fine focus change) The data manager window contains all the image information. You can right click on the image to save it. Use the explorer window to find and organize your files OlympusTraining Manual 4

7 C. Image Acquisition continued... A brief overview of the settings continued: 2- Image Acquisition Control Window Transmitted light observation (for visual observation) Fluorescence observation (for visual observation) Scan buttons select depth for Z-stacks and time for time series Dye List Optical Path Diagram Scanner setting (for photobleaching) Adjustment for each channel (do not change the gain) Pinhole adjustment Image information (for pixel size and resolution Sequential mode (for up to 3 channels). For more than that use the virtual channel under the dye list 3-2D or 3D View window Kalman filter = for line and frame average Light intensity adjustment for halogen bulb OlympusTraining Manual In this window, you can merge the channels and reconstruct the Z-stack. You can also find information about the distribution of the fluorescence intensity of the pixels in your image (histogram) and change the colour of specific channels. If you can t visualize the channels separately, click on the arrow head (left corner of the window). A new window will open. Click on Ch to separate the channels and Z to visualize all the stacks in the gallery mode. 5

8 C. Image Acquisition continued... Image Acquisition for Single Dye: 1- Click on the Dyelist button at the Image Acquisition control window. 2- Choose your dye from the available list. If you already have a dye in the Selected Dyes area, make sure you click ALL CLEAR before adding new ones (unless you will use the same dye) 3- Double click on the dye to add to the Selected Dye tab 4- After adding all the desired dyes, click on APPLY 5- Close the Dyelist window To check and/or modify the light path configuration of your dye, click on the Light and Path button Here you will find all the options for setting the light path configurations. You can change the filters, mirrors and channels if you want. Please note that there's no spectral detectors in this system (no channel S). For a fast, low resolution scan, you can use the Focus x2 or Focus x4 scanning mode. If you want to use this mode to adjust your detector settings (HV and offset), you MUST select AutoHV in the Acquisition Setting windows. With this function selected, as the scan speed becomes slower (for the final image), the saturation level will be maintained. If you forget to select this function, when you press XY for the final image you will see that your image is saturated. As the scan speed is much slower in the final image, the laser spends more time in each pixel, acquiring more signal. If you don't have the auto HV to compensate for that your image will be saturated. Another option would be adjusting the settings of your image while scanning in the XY REPEAT mode. This mode uses the parameters that you set for the final image and it will reflect exactly how your final image will look like. OlympusTraining Manual 6

9 C. Image Acquisition continued... Image Acquisition for Single Dye continued: Adjust the sensitivity of the detector by changing the HV and offset values. The HV will change the brightness of your image and the offset will reduce the noise (background). Press keybord Ctrl+H key to visualize the dynamic range of the detector. The maximum intensity of your image is 4096 (12bit) image. Intensities above that is shown in red and intensities bellow 0 is shown in blue. If you are planning to quantify this image you want NO RED and SOME BLUE If your signal is too bright, you can also adjust the laser power in the Image acquisition control or in the acquistion setting windows DO NOT CHANGE THE GAIN. Leave it as 1. To stop the scanning, click STOP Before scan your final image (XY), make sure to define: a- the right scan time (you can start with 4us/pixel) b- the right aspect ratio (recommended: 1024X1024) c- the right zoom (see instructions on improving your image topic) OlympusTraining Manual 7

10 C. Image Acquisition continued... Image acquisition for double and triple labelled samples: If you have fluororphores that overlap, you MUST use sequential scan mode. 1-Go to the Dyelist panel and add all the fluorophores you want to image (even if they overlap) 2- In the Image Acquisition Control window, select sequential. You can choose to switch between tracks using line or frame mode. If you choose "line", all the lasers will illuminate your sample at the same time and the system will switch between groups line by line. You will visualize all channels almost at the same time. If you choose "frame", the lasers will illuminate your sample separately (each laser at a time) and the system will switch between groups after a full channel has been scanned. 3- To save time during acquisition, you can group the fluorophores that don't overlap to be scanned at the same time. For example, you can drag Alexa Fluor 633 from group 3 to group 1 to be scanned together with Alexa Fluor 488. Then you leave the overlaping fluorophore in a separate group OlympusTraining Manual 8

11 C. Image Acquisition continued... Image acquisition for double and triple labelled samples continued... 4-To adjust the HV and offset for each channel: a- if you have each fluorophore in a separate group, you can start scanning the image and adjust each channel b- if you have combined fluorophores groups, you MUST select the group before doing the adjustments. Otherwise you will keep changing the parameters and not seeing the effect until you select that group 6- After checking and adjusting all your parameters, click XY for the final sequential image Image acquisition for samples labelled with more than 3 fluorophores: As we only have 3 photodetectors in the system, you will have to create a virtual channel in order to use the same channel for 2 fluorophores. 1- Open the DyeList window and select all the fluorophores you want to use. 2- Now, click on Virtual channels scan on the DyeList window. The Virtual channel controller window is automatically open. 3- Each phase represents a sequential track. Drag the fluorophores to different phases, grouping the ones that don t overlap. 4- To visualize the fluorophores in each phase, you must select the phase on the Virtual Channel Controller. Once you select the phase, you can then adjust the parameter for the dyes in that phase. 5-After doing all the adjustments, click START to scan all the phases. ** If you are acquiring a Z-stack using virtual channels, make sure to check the parameter for the Z-stacks in all phases. They must be the same. OlympusTraining Manual 9

12 C. Image Acquisition continued... Image acquisition for samples labelled with more than 3 fluorophores continued... Phase 1" Phase 2" Adding the DIC or Bright Field Channel Please refer to page 3 for DIC prism. For regular bright field image, make sure to change the position of the prism slider to BF (position 1). 1- On the TD1 channel, select the laser that you want to use for this channel. It can be any of the laser lines. 2- Then, activate the TD1 and adjust the HV and offset parameters 3-Click on Fast2X or XY repeat to scan a preview image and adjust the parameters OlympusTraining Manual 10

13 C. Image Acquisition continued... Z-stacks Use the Microscope tab in the Acquisition Settings window to set the StepSize values, bottom and top of your image. 1- Hit clear start and end 2- Start scanning 3- Change the focus all the way down (clock wise in the focus knob). You can use the arrow head buttons to shift the focal point bigger moves 1.0um smaller move 0.1um 4- Once you find the lower limit (bottom) of your image, click SET 0 and END SET. This will be your 0um position 5- Now start changing the focus in the opposite direction (counter clockwise to go up or use arrowhead buttons). You can visualize how far you are going into the sample by looking at the depth 6- Once you find the upper limit of your image (top), click START SET 7- Press the Stop button to stop scanning 8- Enter the StepSize. Optimized values can be referred to by using the Op button. 6 to shift the focal point OlympusTraining Manual 11

14 C. Image Acquisition continued... Z-stacks continued Adjust: a- scan speed b- aspect ratio c- zoom d- HV and offset for each channel 10- Select Depth on Image Acquisition control window 11- Click on XYZ button to scan the series 12- Click on SeriesDone. A 2D ViewLiveImage is displayed on the window bar for the series that have been acquired. To reconstruct the Z-series 1- Click on button to select the type of projection To save the display image, right-click on the image, select Save Display and save the image with a new name. OlympusTraining Manual 12

15 D. Reload the image parameters 1- Go to File and open an exiting image 2- Click on to open the 2D control Panel 3- Click on to reload the parameters E. Improving your image Proper Pixel Size: The best aspect ratio and zoom depend on the resolution you want to achieve. Open the Image Information window and check the optical resolution of the system for the conditions you have. The optical resolution value for X,Y and Z will change when you change the objective and the wavelength. The Pixel size is the size of each pixel in your image using the parameters that you set. According to the Nyquist sampling Theorem, in order to image all the details of an object the pixel size should be between 2 to 2.5 times smaller than the object. For example, if your optical resolution is 200nm, your pixel size should be between 75 to 100nm. If the pixels in your image are bigger than 100nm, you may lose details. If it is smaller than 75nm you will be over sampling without gaining additional information. You can change the pixel size by changing the zoom or the aspect ratio. Roughly you want your pixel size to be half of your optical resolution value. OlympusTraining Manual 13

16 E. Improving your image continued... Proper Pixel Size Nyquist Criterion requires a sampling frequency of at least 2x the highest spatial frequency to accurately preserve the spatial resolution in the resulting digital image Too few pixels - pixel size is too large for maximal resolution Too many pixels - too many pixels (oversampling) does not add spatial resolution Optimal # of pixels 2-2.5x the resolution limit Example: R=0.61 /NA = (0.61x488nm)/1.4 =~200nm R/2 = 100nm R/2.5 = 75nm Optimal pixel size is 75nm-100nm Pixel too big 300nm Pixel = resolution limit 200nm Pixel < resolution limit Optical Image 200nm 300nm 200nm 100nm 100nm 200nm Digital Image OlympusTraining Manual 14

17 To open this window, click on on the Image Acquisition window Resolution of the system based on objective and wavelength used Resolution using your parameters (should be about half of the optical resolution) 512 x nm pixel size Pixel too big 988 x nm pixel size Pixel < resolution limit OlympusTraining Manual 15

18 E. Improving your image continued... Averaging The signal-to-noise ratio can be substantially improved by slowing the scan speed and averaging over several scans Average Averaging removes noise detected by the PMT. In general, the low the Gain is set to, the less sensitive the detector is and the less noise detected by the PMT. Therefore, a lower Gain setting requires less averaging. Line Average - each line the # of times indicated and then the average signal of each scan for each pixel is calculated Frame Average - the entire frame is scanned the # of times indicated and then the average signal of each scan for each pixel is calculated select your line average by clicking on "Kalman" under filter mode (Image acquisition window). Image acquisition is repeated to the specific number of times to provide an averaged image. This function helps reducing the noise and roughness of the image but it increases the scan time. OlympusTraining Manual 16

19 F. Other Important Functions Saving the image: Make sure to save it as OIB format (see File format bellow) To save your image you have 3 options: 1-Go to File - SAVE AS - select your folder in GONAD and save your image 2-close the image. When you do that, you will see a pop up window asking what to do with the image. Click on "Save and Close". Select your folder in GONAD and save your image 3-Right click on the image icon on the Data Manager window and select SAVE AS. select your folder in GONAD and save your image File Format: OIF = creates a folder which contains a 16-bit TIFF image and an "accessory file" with the image information OIB = creates a single file with the original image and information Zooming: If you want to image an specific region of your image, click on Zoom scan (Aquisition Setting window) and draw an ROI on your image. Check your pixel size to make sure it is still about half of your optical resolution. If its too small, you have 2 options: 1- decrease the zoom factor 2- decrease the aspect ratio: click on the Zoom scan again to reset the zoom - change the pixel format of your image (for example from 1024X1024 to 512X512) - scan the image again - select Zoom scan again and draw your ROI Changing the colour of the channels: 1-click on the LUT settings tab in the Live view window 2-select the channel you want to change the colour in the channel drop down bar 3-select the new colour OlympusTraining Manual 17

20 G. Shut Down Procedure 1- Save all your images 2- Before shutting everything down, check the online calendar to see if there is anybody coming after you. If the next user is coming within 1 hour after you, close the software, log off and leave the other components on. If the next user is coming with more than 1 hour after you, proceed with the shut down steps listed bellow. 3- Turn the keys in both lasers off. Wait until it the lasers cool down (about 5min) 4- Remove your sample, clean the objective and press ESC to bring the objective all the way down 5-Close the software and shut computer off 6-Close the shutter of the fluorescence lamp (all the way to the right) 7- Turn components 1-4 off 8- Close the doors of the microscope chamber 9- After 5 min cool down, turn both switches in the laser boxes off (one in the black middle box and the other in the lower gray box) 10- Sign the user sheet OlympusTraining Manual 18

21 H. APPENDIX I Tile Images: For Tile Images, you must first adjust the image parameters for your sample following the steps of this manual (light path configuration, laser power, HV, offset, zoom, pixel size and scan speed). Go to the main tab of the software - DEVICES - and open MULTI-AREA TIME LAPSE A warning window will pop-up. Please hit CANCEL. Two new windows will open: The Multi-Area Time Lapse Controller Window Go to OPTIONS - PREFERENCES Click on the Mosaic icon to open the Mosaic Outline window If the Registered Point List Window doesn t open automatically when you click on Multi-Area Time Lapse, then click on this icon Under INDEX, type how much overlap between the tiles you want (you can use any value from 90 to 100%). I have tested 95% and it worked well. You must do this step before registering the positions (unless you are not changing this value). OlympusTraining Manual 19

22 H. APPENDIX I Tile Images continued... To mark the positions, use the transmitted light or EPI lamp. Under the microscope, move your slide to the edge where your sample starts. Mark this position on the Mosaic Outline window. Move to the next edge and mark the position. You should do this until you form a square (4 positions) or a different geometric shape that includes all the edges of your sample. As you mark the positions, you can see geometric shape forming at the Multi-Area Time Lapse Controller Window. If all the positions are fine, click on ADD GROUP CLICK ON YES The positions you selected are now registered at the Registered Point List Window If you want to delete a position, highlight the position and click on the garbage bin icon. If you want to delete all the positions, click on NEW and don t save the old positions. OlympusTraining Manual 20

23 H. APPENDIX I Tile Images continued... Go to OPTIONS - FILE AND FOLDER Choose where you want to save your files. Suggestion: since the tile files are quite large, I suggest creating a folder on Desktop to save your images and then transferring to gonad after finishing. Please don t forget to remove your files from this computer. Define the name of the subfolder Before start the tile, go to Focus2X or XY REPEAT to make sure your sample is in focus and all the parameters are fine. Now, click on READY on the Registered Point List Window and Image Stitching If your file is too large, you may have to do this step in the another computer and even using another software (ImageJ). Go to DEVICES - MULTI-AREA TIME LAPSE VIEWER Open the MATL_Mosaic document Click on Actions - Stitch OlympusTraining Manual 21

Leica TCS SP8 Quick Start Guide

Leica TCS SP8 Quick Start Guide Leica TCS SP8 Quick Start Guide Leica TCS SP8 System Overview Start-Up Procedure 1. Turn on the CTR Control Box, EL6000 fluorescent light source for the microscope stand. 2. Turn on the Scanner Power

More information

Leica SP8 TCS Users Manual

Leica SP8 TCS Users Manual Version : 07/08/0 Leica SP8 TCS Users Manual Start up:. Turn the PC Microscope, Scanner Power, Laser Power, and the Laser Emission key to on (bottom right of desk).. Turn on the fluorescent lamp (top left

More information

b. Turn the power switch and key to on position for blue laser.

b. Turn the power switch and key to on position for blue laser. OLYMPUS FLUOVIEW 300 CONFOCAL MICOSCOPE OPERATION PROCEDURE 1. Turn ON microscope in this order: 1) Turn on mercury lamp (Note: once the mercury lamp is turned off, DO NOT turn it back on for at least

More information

Leica TCS SP8 Quick Start Guide

Leica TCS SP8 Quick Start Guide Leica TCS SP8 Quick Start Guide Leica TCS SP8 System Overview Start-Up Procedure 1. Turn on the CTR Control Box, Fluorescent Light for the microscope stand. 2. Turn on the Scanner Power (1) on the front

More information

Things to check before start-up.

Things to check before start-up. Byeong Cha Page 1 11/24/2009 Manual for Leica SP2 Confocal Microscope Enter you name, the date, the time, and the account number in the user log book. Things to check before start-up. Make sure that your

More information

Quick Start Guide. Leica SP5 X

Quick Start Guide. Leica SP5 X Quick Start Guide Leica SP5 X Please note: Some of the information in this guide was taken from Leica Microsystems Leica TCS SP5 LAS AF Guide for New Users. This work is licensed under the Creative Commons

More information

ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL

ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL START THE SYSTEM... 2 START ZEN SOFTWARE... 3 SET THE TEMPERATURE AND THE CO2 CONTROLLERS... OBSERVATION AT OCULARS... 5 STATIF PRESENTATION... 6 ACQUIRE ONE

More information

Zeiss 880 Training Notes Zen 2.3

Zeiss 880 Training Notes Zen 2.3 Zeiss 880 Training Notes Zen 2.3 1 Turn on the HXP 120V Lamp 2 Turn on Main Power Switch Turn on the Systems PC Switch Turn on the Components Switch. 3 4 5 Turn on the PC and log into your account. Start

More information

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5.

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5. Smith College August 2010 Operating Checklist for using the Laser Scanning Confocal Microscope Leica TCS SP5. CONTENT, page no. Startup, 1 Initial set-up, 1 Software, 2 Microscope Specimen observation

More information

LSM 780 Confocal Microscope Standard Operation Protocol

LSM 780 Confocal Microscope Standard Operation Protocol LSM 780 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Sign on log sheet according to Actual start time 2. Check Compressed Air supply for the air table 3. Switch

More information

Zeiss LSM 880 Protocol

Zeiss LSM 880 Protocol Zeiss LSM 880 Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.

More information

LSM 510 Training Notes

LSM 510 Training Notes LSM 510 Training Notes Turning on the system Turn on the arc lamp, found on the bench top left of the microscope. This supplies light for epifluorescence for viewing your samples through the microscope.

More information

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling Zeiss LSM 510 Confocor III Training Notes Center for Cell Analysis & Modeling Confocor 3 Start Up Go to System Module Turn on Main Switch, System/ PC, and Components Switches Do you need the arc lamp?

More information

LSM 710 Confocal Microscope Standard Operation Protocol

LSM 710 Confocal Microscope Standard Operation Protocol LSM 710 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Switch on Main power switch 2. Switch on System / PC power button 3. Switch on Components power button 4.

More information

Nikon SIM-E & A1-R System

Nikon SIM-E & A1-R System Nikon SIM-E & A1-R System USER GUIDE LSU Health Sciences Center Shreveport Research Core Facility June 01 2017 Chaowei Shang 1 Table of Content 1. Start Up the System... Page 3 Hardware and microscope

More information

LEICA TCS SP5 AOBS TANDEM USER MANUAL

LEICA TCS SP5 AOBS TANDEM USER MANUAL LEICA TCS SP5 AOBS TANDEM USER MANUAL STARTING THE SYSTEM...2 THE LAS AF SOFTWARE...3 THE «ACQUIRE» MENU...5 CHOOSE AND CREATE A SETTING...6 THE CONTROL PANEL...8 THE DMI6000B MICROSCOPE...10 ACQUIRE ONE

More information

Leica SP8 TCS Users Manual

Leica SP8 TCS Users Manual Leica SP8 TCS Users Manual Follow the procedure for start up and log on as posted in the lab. Please log on with your account only and do not share your password with anyone. We track and confirm usage

More information

Cell Biology and Bioimaging Core

Cell Biology and Bioimaging Core Cell Biology and Bioimaging Core Leica TCS SP5 Operating Instructions Starting up the instrument 1. First, log in the log book located on the confocal desk. Include your name, your lab s PI, an account

More information

Supplemental Method Information Zeiss LSM710

Supplemental Method Information Zeiss LSM710 Supplemental Method Information Zeiss LSM710 1 Under the Light Path window set up the confocal for imaging a green dye (Alexa488-EGFP). For example, set up the light path as shown here using the 488 nm

More information

LSM 510 Meta Training Notes

LSM 510 Meta Training Notes LSM 510 Meta Training Notes Turning on the system Turn on X-Cite power supply. This supplies light for epifluorescence for viewing your samples through the microscope. Turn on the remote control switch.

More information

Title: Leica SP5 Confocal User Manual

Title: Leica SP5 Confocal User Manual Title: Leica SP5 Confocal User Manual Date of first issue: 23/10/2015 Date of review: Version: Admin For assistance or to report an issue Office: CG07 or 05 Email: Igmm-imaginghelpdesk@igmm.ed.ac.uk Website:

More information

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center For any questions or concerns, please contact: Linda Nieman lnieman@mgh.harvard.edu Office: (617) 643-9684 Cell: (512) 565-8076 Chenyue

More information

1 Set up the confocal light path for imaging a green dye (Alexa488-EGFP). For example, the

1 Set up the confocal light path for imaging a green dye (Alexa488-EGFP). For example, the 1 Set up the confocal light path for imaging a green dye (Alexa488-EGFP). For example, the light path as shown here using the 488 nm LASER (Laser Unit 1) reflecting off of the 405/488 nm Dichroic mirror

More information

Nikon Eclipse Ti A1-A Confocal Operating Manual. Start-up. Microscope

Nikon Eclipse Ti A1-A Confocal Operating Manual. Start-up. Microscope Nikon Eclipse Ti A1-A Confocal Operating Manual Start-up 1. Turn on Excite Fluorescent light power supply- metal halide. a. Cool down as for mercury bulb b. Wheel closed liquid light guide 2. Turn on power

More information

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017.

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017. Nikon AZ100 Laser Scanning Macro Confocal Microscope Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin May 2017 Contents 1 Introduction 2 2 Hardware - Startup 2 3 Software/Operation 4 3.1 Multidimensional

More information

Leica Sp5 II Confocal User Guide

Leica Sp5 II Confocal User Guide Leica Sp5 II Confocal User Guide Turning on the Confocal System (instructions are posted in the room) 1. Turn on Laser Power Button 2. Turn Key to On position 3. Turn on Scanner Power Button 4. Turn on

More information

ZEISS LSM510META confocal manual

ZEISS LSM510META confocal manual ZEISS LSM510META confocal manual Switching on the system 1) Switch on the Remote Control button located on the table to the right of the microscope. This is the main switch for the whole system including

More information

Title: Nikon A1R Confocal User Manual

Title: Nikon A1R Confocal User Manual Title: Nikon A1R Confocal User Manual Date of first issue: 23/10/2015 Date of review: Version: Admin For assistance or to report an issue Office: CG.07 or CG.05 Email: Igmm-imaginghelpdesk@igmm.ed.ac.uk

More information

Olympus Fluoview 1000S Spectral Confocal Microscope Introduction to the NRI-MCDB Microscopy Facility Spectral Confocal Microscope

Olympus Fluoview 1000S Spectral Confocal Microscope Introduction to the NRI-MCDB Microscopy Facility Spectral Confocal Microscope Olympus Fluoview 1000S Spectral Confocal Microscope Introduction to the NRI-MCDB Microscopy Facility Spectral Confocal Microscope Improved Optics More Lasers 405 diode 440 diode 488 Argon 515 Argon 559

More information

Zeiss LSM 780 Protocol

Zeiss LSM 780 Protocol Zeiss LSM 780 Protocol 1) System Startup F Please note the sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.

More information

Leica SPEII confocal microscope. Short Manual

Leica SPEII confocal microscope. Short Manual Leica SPEII confocal microscope Short Manual Switching ON sequence: 1. Turn on the Workstation under the bench (top, far right). 2. Turn on the Supply Unit - Laser box (big green switch first and then

More information

Zeiss 780 Training Notes

Zeiss 780 Training Notes Zeiss 780 Training Notes Turn on Main Switch, System PC and Components Switches 780 Start up sequence Do you need the argon laser (458, 488, 514 nm lines)? Yes Turn on the laser s main power switch and

More information

Guide to Confocal 5. Starting session

Guide to Confocal 5. Starting session Guide to Confocal 5 Remember that when booking and before starting session you can check for any problems at https://www.bris.ac.uk/biochemistry/uobonly/cif/index.html Starting session Switch on microscope

More information

MIF ZEISS LSM510 CONFOCAL USER PROTOCOL

MIF ZEISS LSM510 CONFOCAL USER PROTOCOL MIF ZEISS LSM510 CONFOCAL USER PROTOCOL START-UP Turn on the Mercury Bulb Power Supply (if needed). Power-on the Control Box. Turn on the computer. Open the LSM 510 software. Choose Scan New Images and

More information

Microscopy from Carl Zeiss

Microscopy from Carl Zeiss Microscopy from Carl Zeiss Contents Page Contents... 1 Introduction... 1 Starting the System... 2 Introduction to ZEN Efficient Navigation... 5 Setting up the microscope... 10 Configuring the beam path

More information

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center For any questions or concerns, please contact: Linda Nieman lnieman@mgh.harvard.edu Office: (617) 643-9684 Cell: (512) 565-8076 Chenyue

More information

Training Guide for Leica SP8 Confocal/Multiphoton Microscope

Training Guide for Leica SP8 Confocal/Multiphoton Microscope Training Guide for Leica SP8 Confocal/Multiphoton Microscope LAS AF v3.3 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn ON power switch for epifluorescence

More information

Operating Instructions for Zeiss LSM 510

Operating Instructions for Zeiss LSM 510 Operating Instructions for Zeiss LSM 510 Location: GNL 6.312q (BSL3) Questions? Contact: Maxim Ivannikov, maivanni@utmb.edu 1 Attend A Complementary Training Before Using The Microscope All future users

More information

OPERATING INSTRUCTIONS

OPERATING INSTRUCTIONS Zeiss LSM 510 M eta Confocal M icroscope OPERATING INSTRUCTIONS Starting the System: 1. Turn the black knob on the laser box one-quarter turn from Off to On. You will hear the laser cooling mechanisms

More information

Quick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s

Quick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s LSM 5 MP, LSM 510 and LSM 510 META M i c r o s c o p y f r o m C a r l Z e i s s Quick Guide Laser Scanning Microscopes LSM Software ZEN 2007 August 2007 We make it visible. Contents Page Contents... 1

More information

LSM 800 Confocal Microscope Standard Operation Protocol

LSM 800 Confocal Microscope Standard Operation Protocol LSM 800 Confocal Microscope Standard Operation Protocol Turning on the system 1. Switch on the Main switch (labeled 1 and 2 ) mounted on the wall. 2. Turn the Laser Key (labeled 3 ) 90 clockwise for power

More information

Nikon A1R. Multi-Photon & Laser Scanning Confocal Microscope. Kyle Marchuk Adam Fries Jordan Briscoe Kaitlin Corbin. April 2017.

Nikon A1R. Multi-Photon & Laser Scanning Confocal Microscope. Kyle Marchuk Adam Fries Jordan Briscoe Kaitlin Corbin. April 2017. Nikon A1R Multi-Photon & Laser Scanning Confocal Microscope Kyle Marchuk Adam Fries Jordan Briscoe Kaitlin Corbin April 2017 Contents 1 Introduction 2 2 Start-Up 2 3 Imaging 4 3.1 Sample Alignment...........................................

More information

ZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide

ZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide ZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide Matyas Molnar, Biovis 2016 Starting the microscpe 1. Check the microscope if everything looks clean and normal. If not, report it in the logbook.

More information

Nikon C1si Spectral Laser Scanning Confocal Microscope. User Guide

Nikon C1si Spectral Laser Scanning Confocal Microscope. User Guide Nikon C1si Spectral Laser Scanning Confocal Microscope User Guide Contents: C1Si Turn-On/ShutDown Procedures... 2 Overview... 4 Setup for epi-illumination to view through the eyepieces:... 5 Setup for

More information

Training Guide for Carl Zeiss LSM 510 META Confocal Microscope

Training Guide for Carl Zeiss LSM 510 META Confocal Microscope Training Guide for Carl Zeiss LSM 510 META Confocal Microscope AIM 4.2 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn ON Components and System/PC switches

More information

Quick Start FLUOVIEW FV1000

Quick Start FLUOVIEW FV1000 Quick Start FLUOVIEW FV1000 CONFOCAL LASER SCANNING BIOLOGICAL MICROSCOPE FV10-ASW Ver1.3 Petition Thank you for your purchase of Olympus microscope at this time. Prior to using this microscope, read this

More information

Nikon A1Rsi Confocal Start-Up Sequence

Nikon A1Rsi Confocal Start-Up Sequence 1. Turn the key on the Nikon LUN-V Laser Launch. Nikon A1Rsi Confocal Start-Up Sequence 2. Press the button the left side of the A1Rsi Controller unit. 3. Turn on the power strip underneath the microscope.

More information

Simplified Instructions: Olympus Widefield Microscope S1230

Simplified Instructions: Olympus Widefield Microscope S1230 Contents General Microscope Operation Simple Image Capture Multi-Wavelength Capture Z-Series Timelapse Combining Capture Modes Synopsis of Other Functions Pages 2-23 24-40 41-47 48-56 57-59 60-68 69-83

More information

START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7

START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7 Leica DMI AF6000LX Table of contents START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7 ACQUIRE MODULE 6 SETTING THE LIGHTPATH 6

More information

Leica SP8 Resonant Confocal. Quick-Start Guide

Leica SP8 Resonant Confocal. Quick-Start Guide Leica SP8 Resonant Confocal Quick-Start Guide Contents Start-up Preparing for Imaging Part 1 On the scope Part 2 Software interface Part 3 Heat & CO2 incubation Part 4 Other hardware options Shut-down

More information

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope AIM 4.2 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Verify that main power switches on the

More information

Training Guide for Carl Zeiss LSM 880 with AiryScan FAST

Training Guide for Carl Zeiss LSM 880 with AiryScan FAST Training Guide for Carl Zeiss LSM 880 with AiryScan FAST ZEN 2.3 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2018) Power ON Routine 1 2 Turn ON Main Switch from the remote control

More information

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009 Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009 Introduction of Fluoresence Confocal Microscopy The first confocal microscope was invented by Princeton

More information

Everest System / Slidebook Operating Procedures

Everest System / Slidebook Operating Procedures Everest System / Slidebook Operating Procedures NOTICE: This guide is meant to supplement training, not replace it. All users must be trained first hand by a core employee. Training of others in your lab

More information

Microscope Confocal Sp2 Upright.

Microscope Confocal Sp2 Upright. Microscope Confocal Sp2 Upright. Welcome to the Leica Sp2 Confocal Upright tutorial. Before using the Sp2 Invert, You will need to put down your name on the reservation system = http://svintranet.epfl.ch/index.php?optio

More information

Zeiss LSM880 Operating Instructions. UTMB Optical Microscopy Core Jan. 16, 2018

Zeiss LSM880 Operating Instructions. UTMB Optical Microscopy Core Jan. 16, 2018 Zeiss LSM880 Operating Instructions UTMB Optical Microscopy Core Jan. 16, 2018 1 1. Power up the microscope Sing the LOGBOOK Steps below will provide power to the computer and all of the microscope components.

More information

Leica TCS SL Confocal Training. Neuroscience Imaging Core Staff. Core Director. Facility Manager

Leica TCS SL Confocal Training. Neuroscience Imaging Core Staff. Core Director. Facility Manager Leica TCS SL Confocal Training Neuroscience Imaging Facility The Ohio State University Rightmire Hall 614-292-1367 Staff Core Director Anthony Brown, Ph. D. 060 Rightmire Hall 614-292-1205 brown.2302@osu.edu

More information

CONFOCAL MICROSCOPE (Zeiss LSM 510 META v4.2)

CONFOCAL MICROSCOPE (Zeiss LSM 510 META v4.2) Wellcome Trust Centre for Human Genetics Molecular Cytogenetics and Microscopy Core CONFOCAL MICROSCOPE (Zeiss LSM 510 META v4.2) 1) STARTING THE SYSTEM Abridged INSTRUCTIONS Switch on the mercury bulb

More information

Zeiss Axio Imager.A1 manual

Zeiss Axio Imager.A1 manual Zeiss Axio Imager.A1 manual Power-up protocol 1. Mercury lamp 2. Power strip on shelf 3. Computer The Mercury lamp should always be first-on and last-off. This prevents any electrical surges caused by

More information

MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL

MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL START-UP On the Switchbox, turn both black switches to the ON position. Wait for the microscope to boot up completely (watch the screen on the side of the microscope).

More information

Using the Nikon TE2000 Inverted Microscope

Using the Nikon TE2000 Inverted Microscope Wellcome Trust Centre for Human Genetics Molecular Cytogenetics and Microscopy Core Using the Nikon TE2000 Inverted Microscope Fluorescence image acquisition using Scanalytic s IPLab software and the B&W

More information

Contents. Introduction

Contents. Introduction Contents Page Contents... 1 Introduction... 1 Starting the System... 2 Introduction to ZEN Efficient Navigation... 5 Setting up the microscope... 10 Configuring the beam path and lasers... 12 Scanning

More information

Practical work no. 3: Confocal Live Cell Microscopy

Practical work no. 3: Confocal Live Cell Microscopy Practical work no. 3: Confocal Live Cell Microscopy Course Instructor: Mikko Liljeström (MIU) 1 Background Confocal microscopy: The main idea behind confocality is that it suppresses the signal outside

More information

Confocal imaging on the Leica TCS SP8. 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software:

Confocal imaging on the Leica TCS SP8. 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software: Confocal imaging on the Leica TCS SP8 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software: 4) Do not touch the microscope while the software is initializing. Choose your options: Turn

More information

RENISHAW INVIA RAMAN SPECTROMETER

RENISHAW INVIA RAMAN SPECTROMETER STANDARD OPERATING PROCEDURE: RENISHAW INVIA RAMAN SPECTROMETER Purpose of this Instrument: The Renishaw invia Raman Spectrometer is an instrument used to analyze the Raman scattered light from samples

More information

Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope

Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope ZEN 2009 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn Chameleon TiS laser key from Standby

More information

Zeiss Axiovert 135 Fluorescence Microscope Quick Guide / Operations Manual (v. 1.0 February 09)

Zeiss Axiovert 135 Fluorescence Microscope Quick Guide / Operations Manual (v. 1.0 February 09) University of Chicago Integrated Light Microscopy Core Dr. Vytas Bindokas, Director http://digital.bsd.uchicago.edu By: Christine Labno, Assistant Director Room: AB-129 Phone: 4-9040 Zeiss Axiovert 135

More information

Microscope Confocal LSM510 META

Microscope Confocal LSM510 META Microscope Confocal LSM510 META Welcome to the Zeiss LSM 510 Meta Confocal tutorial. Before using the LSM 510 META, Log off any other computer that is open with your personal login. You will need to put

More information

Diskovery Spinning Disk Guide

Diskovery Spinning Disk Guide Diskovery Spinning Disk Guide qbi.microscopy@uq.edu.au Getting started The microscope and its peripherals (Fig. 1a) should always be turned on, but if they are not, turn them on in the following way: 1.

More information

SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014

SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014 CIAN LSM1 or LSM2 short instructions, version 1.4, September 2014 page 1 of 6 SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014 Before starting To work with LSM1

More information

Overview. About other software. Administrator password. 58. UltraVIEW VoX Getting Started Guide

Overview. About other software. Administrator password. 58. UltraVIEW VoX Getting Started Guide Operation 58. UltraVIEW VoX Getting Started Guide Overview This chapter outlines the basic methods used to operate the UltraVIEW VoX system. About other software Volocity places great demands on the computer

More information

3. are adherent cells (ie. cells in suspension are too far away from the coverslip)

3. are adherent cells (ie. cells in suspension are too far away from the coverslip) Before you begin, make sure your sample... 1. is seeded on #1.5 coverglass (thickness = 0.17) 2. is an aqueous solution (ie. fixed samples mounted on a slide will not work - not enough difference in refractive

More information

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING Operations Guide Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING Nikon Eclipse 90i Operations Guide STARTUP Startup Powering Up Fluorescence

More information

OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop

OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop 1. Log in usage using the SMIF web site 2. Turn power

More information

Olympus Confocal Microscope User Guide Last updated

Olympus Confocal Microscope User Guide Last updated 1 Olympus Confocal Microscope User Guide Last updated 11-22-2017 Contents: Laser Safety Training Requirement... 1 Mercury Lamp Precautions... 1 Biosafety requirements and rules for work in the MIC... 2

More information

REMEMBER: You have 5GB of disk space on this microscope. Check before you start if you have room for your experiment. If not delete your old data.

REMEMBER: You have 5GB of disk space on this microscope. Check before you start if you have room for your experiment. If not delete your old data. 1 Use of the Zeiss LSM 510 Inverted Firstly please be aware that this microscope should be treated with respect and care at all times. Rules of use: This Microscope can only be used by Masters by Research

More information

ScanArray Overview. Principle of Operation. Instrument Components

ScanArray Overview. Principle of Operation. Instrument Components ScanArray Overview The GSI Lumonics ScanArrayÒ Microarray Analysis System is a scanning laser confocal fluorescence microscope that is used to determine the fluorescence intensity of a two-dimensional

More information

Swept-Field User Guide

Swept-Field User Guide Swept-Field User Guide Note: for more details see the Prairie user manual at http://www.prairietechnologies.com/resources/software/prairieview.html Please report any problems to Julie Last (jalast@wisc.edu)

More information

Olympus Time-lapse Microscope Basic operation

Olympus Time-lapse Microscope Basic operation Olympus Time-lapse Microscope Basic operation To start up the microscope 1. Switch on the Olympus UCB. (label as ) 1 Power Switch 2 2. Switch on the MT10. (label as ) Power Switch Page 1 of 18 3 3. Switch

More information

Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope

Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope ZEN 2012 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 If you require fluorescence imaging,

More information

Zeiss LSM 510 Multiphoton Confocal Microscope

Zeiss LSM 510 Multiphoton Confocal Microscope Zeiss LSM 510 Multiphoton Confocal Microscope Quick Start User Guide LSU Health Sciences Research Core Facility Table of Contents 1 Safety... Page 3 2 Turn On the System... Page 4 3 Start Up the ZEN Software.

More information

Zeiss LSM 510 Multiphoton Confocal Microscope

Zeiss LSM 510 Multiphoton Confocal Microscope Zeiss LSM 510 Multiphoton Confocal Microscope User Guide LSU Health Sciences Center-Shreveport Research Core Facility Table of Contents 1 Safety... Page 3 2 Turn On the System... Page 4 3 Start Up the

More information

BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions

BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions Starting up: Schematic of Olympus BX-61. 1. Turn on Olympus microscope power box (left of microscope) with toggle switch on

More information

1 Co Localization and Working flow with the lsm700

1 Co Localization and Working flow with the lsm700 1 Co Localization and Working flow with the lsm700 Samples -1 slide = mousse intestine, Dapi / Ki 67 with Cy3/ BrDU with alexa 488. -1 slide = mousse intestine, Dapi / Ki 67 with Cy3/ no BrDU (but with

More information

Instructions for the Leica SP5 II laser scanning confocal microscope

Instructions for the Leica SP5 II laser scanning confocal microscope Instructions for the Leica SP5 II laser scanning confocal microscope Content: Check-in and Start up Set up acquistion parameters Optimize acquistion parameters Acquire a z-stack Sequential scan Check out

More information

Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each

Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each lens with a 1 Airy unit pinhole setting. Many water lenses

More information

Widefield 1. Switching on

Widefield 1. Switching on Widefield 1 Switching on 1. Ignite DG5 lamp - must be switched on first (if previous user has switched off, wait 30 min before igniting) 2. Wait 5s and then turn on the main DG5 controller switch. 3. DG5

More information

Nasmyth Ultraview Vox User Protocol

Nasmyth Ultraview Vox User Protocol Nasmyth Ultraview Vox User Protocol Switch on all wall sockets labelled Nasmyth, switch camera on (power supply located on table behind monitor), switch on laser switch in laser rack, switch computer on

More information

Widefield-NikonEclipseTE200-ORCA Nikon Eclipse TE200 Inverted Microscope with Hamamatsu 1394 Orca-ER Cooled CCD Camera and Micromanager Software

Widefield-NikonEclipseTE200-ORCA Nikon Eclipse TE200 Inverted Microscope with Hamamatsu 1394 Orca-ER Cooled CCD Camera and Micromanager Software Widefield-NikonEclipseTE200-ORCA Nikon Eclipse TE200 Inverted Microscope with Hamamatsu 1394 Orca-ER Cooled CCD Camera and Micromanager Software September 2007 Check website for most current User Guide

More information

Brief manual how to start and close the Leica sp2 Confocal. (TCS SP2 AOBS system mounted on a DM IRE2)

Brief manual how to start and close the Leica sp2 Confocal. (TCS SP2 AOBS system mounted on a DM IRE2) Brief manual how to start and close the Leica sp2 Confocal (TCS SP2 AOBS system mounted on a DM IRE2) A. Switching on hardware B. Acquiring and saving images C. Switching off the microscope D. Good working

More information

3 Choose the Channels button and set the Channel Settings. Set the Pinhole to 1 Airy unit.

3 Choose the Channels button and set the Channel Settings. Set the Pinhole to 1 Airy unit. 1 Set up the confocal light path for imaging a green dye (e.g. Alexa488-EGFP). For example, under the Configuration Control window the light path could be set up as shown here using the 488 nm LASER (found

More information

Locating Molecules Using GSD Technology Project Folders: Organization of Experiment Files...1

Locating Molecules Using GSD Technology Project Folders: Organization of Experiment Files...1 .....................................1 1 Project Folders: Organization of Experiment Files.................................1 2 Steps........................................................................2

More information

TRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO

TRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO TRAINING MANUAL Multiphoton Microscopy LSM 510 META-NLO September 2010 Multiphoton Microscopy Training Manual Multiphoton microscopy is only available on the LSM 510 META-NLO system. This system is equipped

More information

Internal Medicine Imaging Core Emory University Department of Medicine

Internal Medicine Imaging Core Emory University Department of Medicine Internal Medicine Imaging Core Emory University Department of Medicine 1 OPERATION OF THE ZEISS LSM 510 META YOU MUST SIGN UP TO USE THE MICROSCOPE OR COMPUTER EVERY TIME NO EXCEPTIONS Before attempting

More information

Bi/BE 227 Winter Assignment #3. Adding the third dimension: 3D Confocal Imaging

Bi/BE 227 Winter Assignment #3. Adding the third dimension: 3D Confocal Imaging Bi/BE 227 Winter 2016 Assignment #3 Adding the third dimension: 3D Confocal Imaging Schedule: Jan 20: Assignment Jan 20-Feb 8: Work on assignment Feb 10: Student PowerPoint presentations. Goals for this

More information

User Guide to the IBIF Leica TCS SP8 MP Confocal Microscope

User Guide to the IBIF Leica TCS SP8 MP Confocal Microscope User Guide to the IBIF Leica TCS SP8 MP Confocal Microscope This version: 7.24.14. Introduction The IBIF confocal microscope is made available on a fee-for-use-hour basis to all users who have been trained.

More information

NIS-Elements C (For CONFOCAL MICROSCOPE A1) Instructions (Ver. 4.40)

NIS-Elements C (For CONFOCAL MICROSCOPE A1) Instructions (Ver. 4.40) M487E 15.4.NF.17 (1/4) *M487EN17* NIS-Elements C (For CONFOCAL MICROSCOPE A1) Instructions (Ver. 4.40) Preface Thank you for purchasing the Nikon products. This instruction manual has been prepared for

More information

Olympus xcellence Software - basic user guide

Olympus xcellence Software - basic user guide Olympus xcellence Software - basic user guide This is a basic overview of setting up time lapse experiments using Olympus's xcellence software on BIU's IX81 inverted phase contrast system - the software

More information

Operation Of The Leica SP8 Multiphoton Confocal System Using Single Or Multiple Fluorochromes

Operation Of The Leica SP8 Multiphoton Confocal System Using Single Or Multiple Fluorochromes University of South Carolina Scholar Commons Theses and Dissertations 2018 Operation Of The Leica SP8 Multiphoton Confocal System Using Single Or Multiple Fluorochromes Amy E. Rowley University of South

More information

Leica Confocal - 2. Instructions for Leica SP2 Confocal Equipped with Visible Laser Lines

Leica Confocal - 2. Instructions for Leica SP2 Confocal Equipped with Visible Laser Lines Leica Confocal - 2 Instructions for Leica SP2 Confocal Equipped with Visible Laser Lines Imaging Technology Group Revised as of 08/2011 by Dianwen Zhang (Email: zhangdw@illinois.edu) Revision History Authors

More information