Motorized Axio Observer Start-up instructions
|
|
- Gilbert Parker
- 5 years ago
- Views:
Transcription
1 Start-up instructions 1. If using fluorescence turn on Fluorescent light source. TL light Source (Hal 100) 2. Turn on microscope using switch on lower left side of the microscope. 3. If imaging, turn on computer and start Axiovision. TL Aperture ) Diaphragm TL Field Diaphragm Condenser FL light source (HBO 100) FL Filter Turret Trouble Shooting: P: The microscope is on but there is no transmitted light. S: Hold a piece of white paper over the Hal 100 light source. If there is no light, the bulb may be burnt out. If there is light, check to make sure the field diaphragm is are open. Next be sure that H is the selected contrast method. Lastly be sure the light is being sent to the eyes. P: The microscope is on but there is not Fluorescent light. S: Be sure the Fl light source is on, the top light on the power supply should be lit, if not you may need to turn the power supply off and on again. If the top light is lit and there is still no light be sure the RL shutter, RL aperture diaphragm and RL field diaphragm are all open. If there is still no light, check to make sure the correct filter spot is selected in the Fl Filter Turret. P: Everything is on but the image in my live window is all black (or white). S: Make sure light is being sent to the camera. Click the measure button in Axio Vision to take an exposure reading and click the linear button to properly display the image on the screen.
2 How to Set up Koehler Illumination Step 1: The Field Diaphragm 1. Using a low power objective (10x or 20x) focus your specimen 2. Correct for color temperature by increasing the light intensity and adding neutral density filters. 3. Move the condenser to it s bottom most position. 4. Open the aperture diaphragm all the way. 5. Close down the field diaphragm until it s edges are visible in the field of view. 6. Focus the image of the condenser by raising or lowering the condenser carrier. 7. Center the image of the field diaphragm with the centering screws located on the sides of the condenser. Step 2: The Aperture Diaphragm 1. Remove one eyepiece to view the back focal plane of the microscope. 2. Close the aperture Diaphragm so it illuminates approx 80% of the field of view. 3. Return the eyepiece and observe your specimen 4. Slightly adjust the position of the aperture diaphragm until desired contrast/resolution is achieved. Note: The above process should be followed every time a different objective is moved into place. Only after this procedure has become second habit, will you know when shortcuts are o.k. Benefits of Koehler Illumination: * Evenly illuminated image * Minimized reflection or glare or effect of dust in the system * Minimum heating of specimen * Field diaphragm becomes a focus reference for easy focusing of specimen * Microscope is all set for employing contrasting techniques (Brightfield, Oblique, Phase, DIC, etc.)
3 Field Diaphragm Neutral density control Condenser focus knob Contrast Technique Control Aperture Diaphragm Light intensity control Specimen focus knob
4 Setting up Phase Contrast Phase contrast is ideal for examining thin, unstained samples, e.g. culture cells. It allows the human eye to see the differences in refractive index and thickness of different cell components. It does this by using matching optical modulators, a phase stop in the condenser, and a phase ring in the objective. NOTE: not all objective contain phase rings objectives with green writing on them are phase objectives and have the phase ring inside. The size of the phase ring will be written on the objective as Ph1 Ph2 or Ph3 Step 1: Matching the Objective to the Condenser In order to work the phase ring in the objective must match the phase stop in the condenser. The matching is simple. If using a Ph1 objective you must select Ph1 in the condenser. If using Ph2 objective, select Ph2 in the condenser, and if using a Ph3 objective select Ph3 in the condenser. Step 2: Aligning the Phase Rings 1 - Observe the back focal plane, either by using a phase telescope or by removing the eyepiece. - Ideal contrast is achieved when the two rings are superimposed (B). If the rings are not superimposed (A) follow the next two steps to fix the problem. - Using 1.5 mm ball drives adjust the alignment of the phase stop by inserting them into spaces 1 & 1 and turning until the images are super imposed. - Replace the eyepiece and observe the specimen. Contrast should be drastically improved. - 1 This step only needs to be done if the contrast is not optimal or if the microscope is being set up for the first time.
5 Setting up Differential Interference Contrast DIC short for Differential Interference Contrast is a contrasting technique that allows the user to view thin transparent specimens at high resolution as well as a high level of contrast. It does this using a combination polarizer s and wollaston prisms in the beam path of the microscope. Step 1: Match DIC sliders to objective and condenser position. Each objective has a matching DIC slider (pictured at right). Each DIC slider is optimized for a specific DIC Setting which will be written on the DIC slider as either roman numeral I, II, or III. Step 2: Move the fluorescence filter turret to a position that contains an analyzer. (Usually marked analyzer or DIC ) Step 3: Adjust for Koehler illumination. Step 4: Adjust the final image by turning the small knurled silver knob on the DIC slider above the objective. Key: 1. Nuetral Density 2. Condenser 3. DIC Slider 4. Analyzer Slider (option1) 5. Analyzer Slider (option 2) Important notes: DIC can only be done through glass surfaces. It will NOT work with plastic dishes! If your final image does not look correct check to make sure the analyzer is crossed with the polarizer of the DIC prism in the condenser. You can check this by pulling out the DIC slider, and with no sample present and the light on the image should black.
6 Setting up Fluorecence Fluorescence: Light generated by a high performance lamp reaches the exciter filter, the filtered light is reflected by a beam splitter (dichroic) on focused on the specimen via the objective. The specimen absorbs the light and then emits it s fluorescent signal which is transmitted through the dichroic and then passes through an emission filter before reaching our eye. Step 1: Switch on the Fl. light source and allow to warm up for approx 15 min. Step 2: On the reflector turret select the desired filter cube. (Depending on dye used) Step 3: Close the transmitted light shutter and open the reflected light shutter. Step 4: Focus the specimen. Step 5: Adjust the reflected light field diaphragm to be just barely larger than the objective field of view. Fl Lamp Alignment Occasionally it will be necessary to replace and align the FL light source. In the past this was tricky, now it is really very simple! Step 1: Remove the lamp housing from the microscope and remove the cover. This is done using the red handled 3mm tool with each microscope. Step 2: Remove the old bulb by pushing the clamping lever (7) and pulling straight up on the cooling body (1) Then while holding bulb, push the clamping lever on the cooling body (4) to release bulb. Step 3: Insert the new bulb securing it into place with the clamping levers. Step 4: Properly dispose of old bulb adhering to building safety codes. Step 5: Replace lamp housing cover and secure lamp housing to microscope. Step 6: Align bulb by pulling out the alignment window on the side of the microscope and using the 3mm red handled tool in the two alignment slots on the side of the lamp housing. Once you see the mirror image and the real image side by side and focused, use the control knob for the collector (5) to create a completely even illumination across the field. See figure at right. 1. Cooling Body 2. Fixation screws 3. Lamp mount 4. Clamping lever 5. Control knob for collector 6. Bulb 7. Clamping lever
7 Navigating the Touch Screen: (aka: the TFT) Your microscope can be controlled in three ways. 1. by manually using the buttons on the stand 2. by using the computer and Axiovision software 3. by using the touch screen to the side of the microscope. The choice is completely up to you. This short instruction page will guide you through the most basic functions of the TFT. 1. The Home Page: This screen displays a snapshot of all the current microscope settings. From this screen you have several control options: - You can also control the shutters for Transmitted Light (Bright field, DIC, or Phase) or Reflected Light (Fluorescence). - By touching the Load Position button you can lower the stage to put your sample on and then return the stage to it s prior position by hitting the work position button. - By touching the Microscope button a new screen will appear. 2. The Microscope Page: From this screen there are three main functions. - Individual component control: by touching the tabs at the top of this screen you can select the individual components of the microscope, such as objective, and reflector cube. Example: by touching 63x the microscope will move the 63x objective into position over the sample. - Contrast Manager: the buttons on the lower portion of this screen allow you to quickly switch between different contrast methods. By clicking DIC the microscope will move all the necessary components for viewing your sample in DIC, you can then touch the FL button and the microscope will move into position for viewing your sample in Fluorescence. - Shutter Control: As on the main page you also have control of the transmitted light shutter and the reflected light shutter..
INSTRUCTIONS FOR COURSE WORK 4 (AxioVert) Instructor: Anne Vaahtokari (MIU) 1. Purpose of the work
INSTRUCTIONS FOR COURSE WORK 4 (AxioVert) Instructor: Anne Vaahtokari (MIU) 1. Purpose of the work In this work, you will get familiar with an inverted epifluorescence microscope. Also, you will learn
More informationDIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands
DIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands Differential Interference Contrast (DIC) imaging is a technique used to increase contrast in brightfield images. In confocal systems,
More informationCharacterization Microscope Nikon LV150
Characterization Microscope Nikon LV150 Figure 1: Microscope Nikon LV150 Introduction This upright optical microscope is designed for investigating up to 150 mm (6 inch) semiconductor wafers but can also
More informationDIC Imaging using Laser Scanning Microscopes (LSM) on Inverted Stands
DIC Imaging using Laser Scanning Microscopes (LSM) on Inverted Stands Differential Interference Contrast (DIC) imaging is a technique used to increase contrast in brightfield images. In confocal systems,
More informationNikon E800 Microscope. Operating Instructions
Nikon E800 Microscope Operating Instructions B Watson 12/2005 Table of contents: 1. The Nikon E800 Microscope 2. Turning the system ON and OFF 3. Selecting the light path 4. Operating in transmitted light
More informationNikon E800 Operating Instructions.
Nikon E800 Operating Instructions. You can request electronic copies of this manual by contacting lshats@jhsph.edu Copies are also available on the JHU MMI Department web site. Please send your comments
More informationZeiss Axio Imager.A1 manual
Zeiss Axio Imager.A1 manual Power-up protocol 1. Mercury lamp 2. Power strip on shelf 3. Computer The Mercury lamp should always be first-on and last-off. This prevents any electrical surges caused by
More informationImaging Introduction. September 24, 2010
Imaging Introduction September 24, 2010 What is a microscope? Merriam-Webster: an optical instrument consisting of a lens or combination of lenses for making enlarged images of minute objects; especially:
More informationEPIFLUORESCENCE &/OR BRIGHTFIELD MICROSCOPY
EPIFLUORESCENCE &/OR BRIGHTFIELD MICROSCOPY TURN ON THE FOLLOWING EQUIPMENT The fluorescent light (if needed) The power strip for the microscope and accessories The CoolSNAP HQ camera on the right (Turn
More informationUsing the Nikon TE2000 Inverted Microscope
Wellcome Trust Centre for Human Genetics Molecular Cytogenetics and Microscopy Core Using the Nikon TE2000 Inverted Microscope Fluorescence image acquisition using Scanalytic s IPLab software and the B&W
More informationZeiss Axioskop II. The AIF's "routine" light microscope. (Installed 8/24/04)AxioCam installed July 11th 2005
Zeiss Axioskop II The AIF's "routine" light microscope. (Installed 8/24/04)AxioCam installed July 11th 2005 Featuring: Phase Contrast Darkfield DIC/Nomarski Brightfield Fluorescent filters for Dapi, FITC,Rhodamine
More informationNikon Ti-E Microscope Manual. Rightmire Hall Ohio State University. Director: Tony Brown Rightmire
Nikon Ti-E Microscope Manual Rightmire Hall Ohio State University Director: Tony Brown Rightmire 060 292-1205 brown.2302@osu.edu Facility Manager: Paula Monsma Rightmire 062 293-0939 292-1367 monsma.1@osu.edu
More informationTherefore, all descriptions and illustrations in this instruction manual, including all specifications are subject to change without notice.
We are constantly endeavouring to improve our instruments and to adapt them to the requirements of modern research techniques and testing methods. This involves modification to the mechanical structure
More informationZeiss LSM 880 Protocol
Zeiss LSM 880 Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.
More informationANSWER KEY Lab 2 (IGB): Bright Field and Fluorescence Optical Microscopy and Sectioning
Phys598BP Spring 2016 University of Illinois at Urbana-Champaign ANSWER KEY Lab 2 (IGB): Bright Field and Fluorescence Optical Microscopy and Sectioning Location: IGB Core Microscopy Facility Microscope:
More information3. are adherent cells (ie. cells in suspension are too far away from the coverslip)
Before you begin, make sure your sample... 1. is seeded on #1.5 coverglass (thickness = 0.17) 2. is an aqueous solution (ie. fixed samples mounted on a slide will not work - not enough difference in refractive
More informationNikon E800 Operating Instructions.
Nikon E800 Operating Instructions. You can request electronic copies of this manual by contacting imaging@fhcrc.org. Copies are also available on the Scientific Imaging web site. Please send your comments
More informationOlympus IX71 Microscope and DP71 Camera Instructions
Olympus IX71 Microscope and DP71 Camera Instructions Microscopy in Medicine (MiM) Core Emory University Department of Medicine 1 Olympus IX71 Image Capture Procedure 2 3 1. STARTING-UP PROCEDURE: Remove
More informationBasics of Light Microscopy and Metallography
ENGR45: Introduction to Materials Spring 2012 Laboratory 8 Basics of Light Microscopy and Metallography In this exercise you will: gain familiarity with the proper use of a research-grade light microscope
More informationWidefield-NikonEclipseTE200-ORCA Nikon Eclipse TE200 Inverted Microscope with Hamamatsu 1394 Orca-ER Cooled CCD Camera and Micromanager Software
Widefield-NikonEclipseTE200-ORCA Nikon Eclipse TE200 Inverted Microscope with Hamamatsu 1394 Orca-ER Cooled CCD Camera and Micromanager Software September 2007 Check website for most current User Guide
More informationCAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE
University of Virginia ITC Academic Computing Health Sciences CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE Introduction The Olympus BH-2 microscope in ACHS s microscope lab has objectives from
More informationML7520 ML7530 DIOPTER ADJUSTMENT RING BINOCULAR BODY, INCLINED 30. (a) Field Iris Control Lever. (c) Filter Slots EYEPIECES, KHW10X
JAPAN DIOPTER ADJUSTMENT RING BINOCULAR BODY, INCLINED 30 (a) Field Iris Control Lever (c) Filter Slots EYEPIECES, KHW10X ANALYZER CONTROL LEVER (b) Aperture Iris Control Lever LIGHT SOURCE HOUSING VERTICAL
More informationZeiss Axiovert 135 Fluorescence Microscope Quick Guide / Operations Manual (v. 1.0 February 09)
University of Chicago Integrated Light Microscopy Core Dr. Vytas Bindokas, Director http://digital.bsd.uchicago.edu By: Christine Labno, Assistant Director Room: AB-129 Phone: 4-9040 Zeiss Axiovert 135
More informationAxioVision 4.5 Brightfield Image Capture Procedure
AxioVision 4.5 Brightfield Image Capture Procedure 1. STARTING-UP PROCEDURE: Remove blue dust cover and place on shelf under microscope. Turn on the halogen lamp by pushing the switch at the back right
More informationLeica SP8 TCS Users Manual
Version : 07/08/0 Leica SP8 TCS Users Manual Start up:. Turn the PC Microscope, Scanner Power, Laser Power, and the Laser Emission key to on (bottom right of desk).. Turn on the fluorescent lamp (top left
More informationReichert Univar Manual
Reichert Univar Manual Translated from the 11/1975 German language edition, with slight modifications. William R. Porter San Marcos CA USA 2017 v 1.3 Notes This is a very slightly-modified, new (2017)
More informationSTART-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7
Leica DMI AF6000LX Table of contents START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7 ACQUIRE MODULE 6 SETTING THE LIGHTPATH 6
More informationBiology 29 Cell Structure and Function Spring, 2009 Springer LABORATORY 1: THE LIGHT MICROSCOPE
Biology 29 Cell Structure and Function Spring, 2009 Springer LABORATORY 1: THE LIGHT MICROSCOPE Prior to lab: 1) Read these instructions (p 1-6) 2) Go through the online tutorial, the microscopy pre-lab
More informationUser instructions Biological inverted microscope
KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen E-mail: info@kern-sohn.com User instructions Biological inverted microscope Tel: +49-[0]7433-9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com KERN
More informationIndian Institute of technology Madras Presents NPTEL NATIONAL PROGRAMME ON TECHNOLOGY ENHANCED LEARNING
Indian Institute of technology Madras Presents NPTEL NATIONAL PROGRAMME ON TECHNOLOGY ENHANCED LEARNING Lecture - 5 Materials Characterization Fundamentals of Optical microscopy Dr. S. Sankaran Associate
More informationLSM 510 META in Chang Gung University
Content LSM 510 META in Chang ung University LSM 510 META 路 理 The features and applications of LSM 510 META 01-09 Introduction of the hardware 10-12 Fluorescence observation in conventional microscope
More informationBASICS IN BIOIMAGING AND OPTICS PLATFORM EPFL SV PTBIOP LIGHT MICROSCOPY
BASICS IN LIGHT MICROSCOPY OVERVIEW 1. Motivation 2. Basic in optics 3. How microscope works 4. Illumination and resolution 5. Microscope optics 6. Contrasting methods -2- MOTIVATION Why do we need microscopy?
More informationSHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014
CIAN LSM1 or LSM2 short instructions, version 1.4, September 2014 page 1 of 6 SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014 Before starting To work with LSM1
More informationPractical Light Microscopy
Biomedical & X-ray Physics Kjell Carlsson Important: Study the preparatory exercises carefully before the lab session starts! Practical Light Microscopy Laboratory instructions for course SK2500/01, Physics
More informationInstruction Manual T Binocular Acromat Research Scope T Trinocular Acromat Research Scope
Research Scope Instruction Manual T-29031 Binocular Acromat Research Scope T-29041 Trinocular Acromat Research Scope T-29032 Binocular Semi-Plan Research Scope T-29042 Trinocular Semi-Plan Research Scope
More informationOperating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5.
Smith College August 2010 Operating Checklist for using the Laser Scanning Confocal Microscope Leica TCS SP5. CONTENT, page no. Startup, 1 Initial set-up, 1 Software, 2 Microscope Specimen observation
More informationInstructions for Making On-Line Reservations for Microscopes in NB11-204
Instructions for Making On-Line Reservations for Microscopes in NB11-204 1. Log into Mail using Mail.swmed.edu 2. Log in using your university id and password. 3. Click the Calendar Tab at the top right
More informationZeiss AxioImager.Z2 Brightfield Protocol
Zeiss AxioImager.Z2 Brightfield Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge
More informationWidefield 1. Switching on
Widefield 1 Switching on 1. Ignite DG5 lamp - must be switched on first (if previous user has switched off, wait 30 min before igniting) 2. Wait 5s and then turn on the main DG5 controller switch. 3. DG5
More informationCX-PCD INSTRUCTIONS PHASE CONTRAST TURRET CONDENSER
INSTRUCTIONS CX-PCD PHASE CONTRAST TURRET CONDENSER This instruction manual is for the Olympus Phase Contrast Turret Condenser Model CX-PCD. To obtain optimum performance and to familiarize yourself fully
More informationSimplified Instructions: Zeiss Brightfield Microscope S1000
Contents General Microscope Set-Up Adjust Illumination Focus Condenser Open Software Image Capture Settings Shading & Color Corrections Image Capture & Viewing Scaling and Measurements Synopsis of Other
More informationBX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions
BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions Starting up: Schematic of Olympus BX-61. 1. Turn on Olympus microscope power box (left of microscope) with toggle switch on
More informationSTEINDORFF METALLURGICAL MICROSCOPE. NYMCS-620 Instruction Manual
METALLURGICAL MICROSCOPE NYMCS-620 Instruction Manual It is recommended strongly that you study this manual thoroughly before using the microscope. Retain this manual in an easily accessible place near
More informationMicroscopy: Fundamental Principles and Practical Approaches
Microscopy: Fundamental Principles and Practical Approaches Simon Atkinson Online Resource: http://micro.magnet.fsu.edu/primer/index.html Book: Murphy, D.B. Fundamentals of Light Microscopy and Electronic
More informationZoom Stereo Microscope NYMCS-360 Instruction Manual
Zoom Stereo Microscope NYMCS-60 Instruction Manual This manual is written for stereo microscope NYMCS-60. To ensure the safety, obtain optimum performance and to familiarize yourself fully with the use
More informationLight Microscopy. Upon completion of this lecture, the student should be able to:
Light Light microscopy is based on the interaction of light and tissue components and can be used to study tissue features. Upon completion of this lecture, the student should be able to: 1- Explain the
More informationTABLE OF CONTENTS. Safety notes i. Care and Maintenance. ii. 1. Components Illustration Installation of Components.. 4
TABLE OF CONTENTS Safety notes i Care and Maintenance. ii 1. Components Illustration... 1 2. Installation of Components.. 4 2.1 Installation Diagram... 4 2.2 Installation Procedures 5 3. Operation...11
More information2/4/15. Brightfield Microscopy! It s all about Magnification..! or is it?!
Brightfield Microscopy It s all about Magnification.. or is it? 1 What actually does go into chosing a microscope Choice depends on what you need the microscope to do. Do you want to magnify stained specimens?
More informationEyepieces KHW10X. Diopter Adjustment Ring. Binocular Body Inclined 30. Binocular Clamp Screw. Analyzer control Lever. Reflected Light Illuminator
JAPAN Eyepieces KHW10X Diopter Adjustment Ring Binocular Body Inclined 30 Binocular Clamp Screw Analyzer control Lever Reflected Light Illuminator Ball-Bearing Objective Nosepiece Objectives Large Scan
More informationGuide to Confocal 5. Starting session
Guide to Confocal 5 Remember that when booking and before starting session you can check for any problems at https://www.bris.ac.uk/biochemistry/uobonly/cif/index.html Starting session Switch on microscope
More informationZeiss AxioObserver with ApoTome
Zeiss AxioObserver with ApoTome Quick Start User Guide LSU Health Sciences Center-Shreveport Research Core Facility (RCF) Microscopy Table of Contents 1 Start up the system.. Page 3 2 Touch screen controller
More informationmicroscopy A great online resource Molecular Expressions, a Microscope Primer Partha Roy
Fundamentals of optical microscopy A great online resource Molecular Expressions, a Microscope Primer http://micro.magnet.fsu.edu/primer/index.html Partha Roy 1 Why microscopy Topics Functions of a microscope
More informationMicroscope ECLIPSE 80i Instructions <Reference>
M318E 03.12.CF.1(2/2) Microscope ECLIPSE 80i Instructions Introduction Thank you for purchasing this Nikon product. This instruction manual, which describes basic microscope operations, is
More informationThings to check before start-up.
Byeong Cha Page 1 11/24/2009 Manual for Leica SP2 Confocal Microscope Enter you name, the date, the time, and the account number in the user log book. Things to check before start-up. Make sure that your
More informationQUICKSTART GUIDE: WIDEFIELD HWF1 Zeiss Cell Observer Live Cell Imaging System (HAMMERSMITH, L BLOCK, ROOM 314) Imperial College London
Imperial College London Facility for Imaging by Light Microscopy QUICKSTART GUIDE: WIDEFIELD HWF1 Zeiss Cell Observer Live Cell Imaging System (HAMMERSMITH, L BLOCK, ROOM 314) Observing Life As It Happens
More informationBiological Microscope Manual
Version No.: V1.2 Series Biological Microscope Manual This manual expatiates the using method, troubleshooting and maintenance about MT-50 series biological microscope. Please study this manual thoroughly
More informationLeica TCS SP8 Quick Start Guide
Leica TCS SP8 Quick Start Guide Leica TCS SP8 System Overview Start-Up Procedure 1. Turn on the CTR Control Box, EL6000 fluorescent light source for the microscope stand. 2. Turn on the Scanner Power
More informationOlympus Fluoview 1000S Spectral Confocal Microscope Introduction to the NRI-MCDB Microscopy Facility Spectral Confocal Microscope
Olympus Fluoview 1000S Spectral Confocal Microscope Introduction to the NRI-MCDB Microscopy Facility Spectral Confocal Microscope Improved Optics More Lasers 405 diode 440 diode 488 Argon 515 Argon 559
More informationSWIFT SERIES M2252DGL MICROSCOPE
SWIFT SERIES M2252DGL MICROSCOPE The M2252DGL Series is ideal for elementary to high school classrooms. Built to withstand student use, this series has locked-on eyepieces, objectives, illuminator housing
More informationUser Manual. Trinocular Metallurgical Microscope. MicroscopeNet.com
User Manual Trinocular Metallurgical Microscope Model M83MPTR MicroscopeNet.com Table of Contents i. Caution.. 1 ii. Care and Maintenance... 2 1. Components Illustration..... 3 2. Installation...4 3. Operation
More informationChapter 1 Parts. Figure 1.1. Parts of a Compound Light Microscope
Chapter 1 Parts Chapter 1 Parts Figure 1.1 illustrates the parts of an upright compound microscope and indicates the terminology that I use in these notes. Figure 1.1. Parts of a Compound Light Microscope
More informationMicroscopy. The dichroic mirror is an important component of the fluorescent scope: it reflects blue light while transmitting green light.
Microscopy I. Before coming to lab Read this handout and the background. II. Learning Objectives In this lab, you'll investigate the physics of microscopes. The main idea is to understand the limitations
More informationSTEINDORFF. Inverted Fluorescence Microscope NYMCS-702. Instruction Manual
Inverted Fluorescence Microscope NYMCS-702 Instruction Manual This manual is written for Inverted Fluorescence Microscope NYMCS-702. For safety and for keeping the best performance, making you familiar
More informationZeiss Deconvolution Microscope: A Quick Guide
Zeiss Deconvolution Microscope: A Quick Guide Start-up Uncover microscope. Do not put dust cover on the floor. Plug in both cameras. The default camera is the AxioCam HRm (monochrome camera) for fluorescence
More informationNikon Eclipse Ti A1-A Confocal Operating Manual. Start-up. Microscope
Nikon Eclipse Ti A1-A Confocal Operating Manual Start-up 1. Turn on Excite Fluorescent light power supply- metal halide. a. Cool down as for mercury bulb b. Wheel closed liquid light guide 2. Turn on power
More informationBrightfield Microscopy and Image Acquisition on Spotcam1. by Ryan Taylor/Nancy Kleene Last modified 10/02/05 by Birgit Ehmer
Brightfield Microscopy and Image Acquisition on Spotcam1 by Ryan Taylor/Nancy Kleene Last modified 10/02/05 by Birgit Ehmer Log onto the computer. Enter your username and password to log onto the server.
More informationLeica SP8 TCS Users Manual
Leica SP8 TCS Users Manual Follow the procedure for start up and log on as posted in the lab. Please log on with your account only and do not share your password with anyone. We track and confirm usage
More informationMicroscopy. Matti Hotokka Department of Physical Chemistry Åbo Akademi University
Microscopy Matti Hotokka Department of Physical Chemistry Åbo Akademi University What s coming Anatomy of a microscope Modes of illumination Practicalities Special applications Basic microscope Ocular
More informationUser Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com
User Manual Digital Compound Binocular LED Microscope Model MD82ES10 MicroscopeNet.com Table of Contents i. Caution... 1 ii. Care and Maintenance... 2 1. Components Illustration... 3 2. Installation...
More informationUser instructions Compound laboratory microscope
KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen E-mail: info@kern-sohn.com User instructions Compound laboratory microscope Tel: +49-[0]7433-9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com KERN
More informationMICROSCOPE LAB. Resolving Power How well specimen detail is preserved during the magnifying process.
AP BIOLOGY Cells ACTIVITY #2 MICROSCOPE LAB OBJECTIVES 1. Demonstrate proper care and use of a compound microscope. 2. Identify the parts of the microscope and describe the function of each part. 3. Compare
More informationVery short introduction to light microscopy and digital imaging
Very short introduction to light microscopy and digital imaging Hernan G. Garcia August 1, 2005 1 Light Microscopy Basics In this section we will briefly describe the basic principles of operation and
More informationMicroscopy from Carl Zeiss
Microscopy from Carl Zeiss Contents Page Contents... 1 Introduction... 1 Starting the System... 2 Introduction to ZEN Efficient Navigation... 5 Setting up the microscope... 10 Configuring the beam path
More informationTherefore, all descriptions and illustrations in this instruction manual, including all specifications are subject to change without notice.
We are constantly endeavouring to improve our instruments and to adapt them to the requirements of modern research techniques and testing methods. This involves modification to the mechanical structure
More informationUser instructions Compound laboratory microscope
KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen E-mail: info@kern-sohn.com User instructions Compound laboratory microscope Tel: +49-[0]7433-9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com KERN
More informationObserving Microorganisms through a Microscope
2016/2/19 PowerPoint Lecture Presentations prepared by Bradley W. Christian, McLennan Community College CHAPTER 3 Observing Microorganisms through a Microscope 1 Figure 3.2 Microscopes and Magnification.
More informationSystems Biology. Optical Train, Köhler Illumination
McGill University Life Sciences Complex Imaging Facility Systems Biology Microscopy Workshop Tuesday December 7 th, 2010 Simple Lenses, Transmitted Light Optical Train, Köhler Illumination What Does a
More informationMicroscopy Training & Overview
Microscopy Training & Overview Product Marketing October 2011 Stephan Briggs - PLE OVERVIEW AND PRESENTATION FLOW Glossary and Important Terms Introduction Timeline Innovation and Advancement Primary Components
More informationLEO 912 TEM Short Manual. Prepared/copyrighted by RH Berg Danforth Plant Science Center
LEO 912 TEM Short Manual Prepared/copyrighted by RH Berg Danforth Plant Science Center Specimen holder [1] Never touch the holder (outside of the O-ring, double-headed arrow) because finger oils will contaminate
More informationWhere Image Quality Begins
Where Image Quality Begins Filters are a Necessity Not an Accessory Inexpensive Insurance Policy for the System The most cost effective way to improve repeatability and stability in any machine vision
More informationwww.cairn-research.co.uk Free Phone: 08453301267 (UK only) Tel: +44 (0) 1795590140 Fax: +44 (0) 1795594510 Important Information Please Read Before Installing and Operating Your Optosplit II For maximum
More informationTopic 1 - What is Light? 1. Radiation is the type of energy transfer which does not require... A matter B heat C waves D light
Grade 8 Unit 1 Test Student Class Topic 1 - What is Light? 1. Radiation is the type of energy transfer which does not require... A matter B heat C waves D light 2. Light-producing technologies, such as
More informationResolution. Diffraction from apertures limits resolution. Rayleigh criterion θ Rayleigh = 1.22 λ/d 1 peak at 2 nd minimum. θ f D
Microscopy Outline 1. Resolution and Simple Optical Microscope 2. Contrast enhancement: Dark field, Fluorescence (Chelsea & Peter), Phase Contrast, DIC 3. Newer Methods: Scanning Tunneling microscopy (STM),
More informationOlympus Time-lapse Microscope Basic operation
Olympus Time-lapse Microscope Basic operation To start up the microscope 1. Switch on the Olympus UCB. (label as ) 1 Power Switch 2 2. Switch on the MT10. (label as ) Power Switch Page 1 of 18 3 3. Switch
More informationMicroscopy Techniques that make it easy to see things this small.
Microscopy Techniques that make it easy to see things this small. What is a Microscope? An instrument for viewing objects that are too small to be seen easily by the naked eye. Dutch spectacle-makers Hans
More informationWITec Alpha 300R Quick Operation Summary October 2018
WITec Alpha 300R Quick Operation Summary October 2018 This document is frequently updated if you feel information should be added, please indicate that to the facility manager (currently Philip Carubia,
More informationWeek IX: INTERFEROMETER EXPERIMENTS
Week IX: INTERFEROMETER EXPERIMENTS Notes on Adjusting the Michelson Interference Caution: Do not touch the mirrors or beam splitters they are front surface and difficult to clean without damaging them.
More informationOperation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009
Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009 Introduction of Fluoresence Confocal Microscopy The first confocal microscope was invented by Princeton
More informationEverest System / Slidebook Operating Procedures
Everest System / Slidebook Operating Procedures NOTICE: This guide is meant to supplement training, not replace it. All users must be trained first hand by a core employee. Training of others in your lab
More informationWeek IV: FIRST EXPERIMENTS WITH THE ADVANCED OPTICS SET
Week IV: FIRST EXPERIMENTS WITH THE ADVANCED OPTICS SET The Advanced Optics set consists of (A) Incandescent Lamp (B) Laser (C) Optical Bench (with magnetic surface and metric scale) (D) Component Carriers
More informationFigure 3.4 Approximate size of various types of cells. ~10 um. Red Blood Cells = mm 1500 um. Width of penny Pearson Education, Inc.
Figure 3.4 Approximate size of various types of cells. ~10 um Red Blood Cells 1.5mm 1500 um Width of penny = 1500 Figure 4.3 The limits of resolution (and some representative objects within those ranges)
More informationLaboratory Introduction
Laboratory Introduction There are two basic categories of microscopes: light microscopes and electron microscopes. Light, or optical, microscopes require light waves to provide the illumination while electron
More informationUser instructions Transmitted light phase contrast microscope
KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen E-Mail: info@kern-sohn.com Tel: +49-[0]7433-9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com User instructions Transmitted light phase contrast
More informationIntroduction. Laboratory Equipment & Supplies. Model 1333PHi Shown (Phase Contrast) (2) Eyepieces (Eyecups installed) Diopter Adjustment Mechanism
Introduction With the invention of the microscope in the early 17th century, it was made possible to view objects which were too small for the human eye to see. As the microscope evolved, the structure
More informationManual for BMS E1 eplan series, compound microscope
Manual for BMS E1 eplan series, compound microscope The compound microscope allows it to study, at cell level, structures of textures of botanical and zoological nature. (e.g. slides of roots, leaves and
More informationFLUORESCENCE MICROSCOPY. Matyas Molnar and Dirk Pacholsky
FLUORESCENCE MICROSCOPY Matyas Molnar and Dirk Pacholsky 1 The human eye perceives app. 400-700 nm; best at around 500 nm (green) Has a general resolution down to150-300 μm (human hair: 40-250 μm) We need
More informationOMM300. Inverted Metallurgical Microscope
OMM300 Inverted Metallurgical Microscope Instruction Manual Please read the instructions carefully before operating CONTENTS Safety 2 Parts List 2 Features 3 Assembly 5 Operation 7 Maintenance 9 Specifications
More informationMICROSCOPY FOR THE DEVELOPMENTAL BIOLOGY STUDENT...
MICROSCOPY FOR THE DEVELOPMENTAL BIOLOGY STUDENT... You will be using two configurations of microscope during the course of the semester to observe specimens and record your results: compound microscopes
More information2 How to operate the microscope/obtain an image
Morgagni Operating Instructions 50079 010912 2-1 2 ow to operate the microscope/obtain an image 2.1 Starting the microscope 2.1.1 Starting the microscope with several manually-operated steps 1. Turn on
More informationPractical work no. 3: Confocal Live Cell Microscopy
Practical work no. 3: Confocal Live Cell Microscopy Course Instructor: Mikko Liljeström (MIU) 1 Background Confocal microscopy: The main idea behind confocality is that it suppresses the signal outside
More information