Mammography: Physics of Imaging
|
|
- Nigel Nichols
- 5 years ago
- Views:
Transcription
1 Mammography: Physics of Imaging Robert G. Gould, Sc.D. Professor and Vice Chair Department of Radiology and Biomedical Imaging University of California San Francisco, California Mammographic Imaging: Uniqueness Low subject contrast High spatial resolution requirements Complicated structures Priors critical Dedicated imaging equipment Use compression Page 1
2 Factors Affecting Subject Contrast X-ray energy: contrast decreases with increasing energy Scattered radiation Function of tissue volume irradiated For small objects (e.g. Ca specks), image blurring Tissue Characteristics Glandular tissue Fat Calcium Atomic number ~ Density, gm/cm 3 ~ Page 2
3 Mammography: Subject Contrast t calcium speck µ Ca X-rays Compression paddle µ tissue KeV Tissue µ, cm -1 Ca µ, cm -1 Δµ C, %* Grid I 2 I 1 C s ~ Log I 1 -Log I 2 ~ Δµt *For 100 µm Ca particle w/o scatter! C s (%) = Log I 1 /I 2 = Δµt x 100! Mammographic vs. Standard Radiographic Equipment Lower x-ray energy < 35 KVp vs > 40 KVp Smaller focal spot mm vs mm Shorter source - receptor distance cm vs 100 cm Apply compression Page 3
4 Mammography X-ray Sources Anode materials Molybdenum (Z=42) Rhodium (Z=45) X-ray production efficiency ~ 40% less than in tungsten Nominal focal spot sizes Large 0.3 mm Small 0.1 mm Typical exposure times ~ >1 sec Element Atomic Number Melting Point, C o Relative X-ray Emission W Mo Rh Characteristic Radiation Produced at specific energies Equal to differences in binding energies of orbital electron shells For standard x-ray source, typically < 15% of output Energetic free electron (Kinetic energy > K shell binding energy) Characteristic x-ray (hν = K shell BE - M shell BE) e - e - In mammography, can exceed 25% of output Page 4
5 Characteristic Radiation Use to increase contrast No Bremsstrahlung low energy penalty Increase output relative to Bremsstrahlung Filter with same material Materials relatively translucent to their characteristic emissions Energy, KeV Molybdenum Rhodium " Molybdenum X-Ray Spectrum Filter! 0.3 mm polycarbonate" Number of photons" 20000" 15000" 10000" 0.3 mm polycarbonate mm Mo" K edge Mo" 19.9 KeV" µ (cm -1 ) " 100 0" 0" 5" 10" 15" 20" 25" 30" 10 Photon energy, KeV" Page 5
6 Mammography X-Ray Beam Relative Intensity" Entrance spectrum" Relative Intensity" Exit" spectrum" Increasing breast thickness/ density" 10" 20" 30" Photon Energy, KeV" 10" 20" 30" Photon Energy, KeV" As breast thickness/density increases, characteristic radiation has lesser effect on image! Mammographic Tubes Benefit of characteristic radiation in image formation greatest for thin or fatty breasts Bremsstrahlung component above the K-edge becomes increasingly dominate in exit beam with increasing breast thickness Page 6
7 Compression Reduces anatomical motion Reduces dose Reduces exposure latitude Spreads out detail Less superimposition X-rays! X-rays! Compression! paddle! Decreases distortion Detector! Scatter in Mammography Exit surface to detector ~ 2-3 cm Little air gap effect Scatter/primary ratio ~ 0.5 If contrast is 5% w/o scatter, contrast reduced to 3.4% with 0.5 S/P 33% contrast reduction Use ~5:1 grid Increases dose by 2-3x (Bucky factor) Improves contrast by ~ 40% Contrast of 100 µm Ca speck, % No Sca5er to Primary Ra:o KeV Sca5er Page 7
8 Hologic HTC Grid High-Transmission Cellular Focused, 2 dimensional grid Air interspacing ~ 4-10% improvement in contrast compared to 5:1 linear grid After J. Gray" Mammography Dose Levels Screening exam 2 views per breast Peak (skin) dose 8-10 mgy/view By law in the USA, glandular dose < 3 mgy/view Based on the ACR Phantom = average compressed breast of 50% adipose and 50% glandular tissue 4.2 cm thick (3.5 cm Lucite cm Paraffin) Function of KVp, anode material, filtration (HVL) Page 8
9 Digital Imaging in Mammography Improved diagnostic capability? True in dense breasts Use of digital processing Spatially adaptive Feature enhancing CAD Include in PACS w/o digitizing film Better image management Dose reduction? Digital Mammography-Outcomes ACRIN study Comparison made to screen/film systems Apparently no light shed on differences between FFDM systems Greater sensitivity in woman < 50 years old and in woman with dense breasts Equal to screen/film systems otherwise Page 9
10 Digital Mammographic Systems CR! Carestream DirectView (computed radiography)! Fuji FCRm (computed radiography)! Indirect Conversion DR! General Electric Senographe 2000D, DS, Essential, Senographe Care! Direct Conversion DR! Hologic Loard Selenia, Selenia S, Selenia Dimensions, Selenia Encore! Siemens Mammomat Novation DR, Mammomat Novation S, Mammomat Inspiration, Mammomat Inspiration Pure! Scanning systems! Fischer Senoscan! Sectra MicroDose Mammography L30! Fuji CR Use conventional mammo unit (eg GE DMR) BaFBr phosphor Coverage (plate size): 18 x 24 cm or 24 x 30 cm 50 µm pixels 3328 x4096 (24 MB) image size 14 bit dynamic range Page 10
11 GE Senographe Indirect DR using CsI phosphor Mo and Rh anodes, Mo and Rh filters Coverage: 19.2 x 23 cm or 24 x 30.7 cm 100 µm pixels 1914 x 2294 (9 MB) or 2394 x 3062 (14 MB) image size 14 bit dynamic range Hologic Selenia Direct DR using amorphous selenium Mo anode, Mo and Rh filters Coverage: 24 x 29 cm 70 µm pixels 3328 x 4096 (24 MB) image size 14 bit dynamic range Page 11
12 Photon counting, multi-slit scanning system Scan time: 3-15 sec Direct detector of crystalline silicon W target/al filter Coverage: 24 x 26 cm 50 µm pixels 4800 x 5200 (37 MB) image size 12 bit dynamic range Sectra MicroDose Image Interpretation All systems come with a dual, high resolution display system 5 megapixels Obfuscation on using display for images from other vendors Critical features: Zoom and pan Ability to annotate and save annotations to PACS Easy comparison with priors CAD software Page 12
13 Digital Mammo: Image Quality Compared to Screen/film Lower limiting spatial resolution < 10 lp/mm vs > 12 lp/mm Higher DQE More pronounced at lower spatial frequencies (< 2 lp/mm) Density more uniform across the image Adaptive filtering Skin line to chest wall Screen-film vs Digital Page 13
14 Digital Mammography-Dose Considerations Geometric efficiency (fill factor) < 100% 50-80% depending on pixel size Screen-film systems = 100% Quantum noise Influence on processing algorithms? Influence on CAD performance? Dose savings remain potential Developments in Mammography Computer-aided diagnosis (CAD) Widespread for screening mammography Improved performance compared to singleset of eyes Relative improvement depends on experience of mammographer Little or no improvement over double reading by experienced mammographers Higher false positive rate Tomosynthesis Dedicated breast CT systems Page 14
15 Digital Tomosynthesis exposures! Acquire a series of low dose projection images at different angles Electronically align and weight images to isolate a plane through the breast Scroll through breast 15-25! Digital Tomosynthesis Active area of development by multiple vendors Hologic Selenia Dimensions system has FDA approval Can total dose be kept same as for screening mammogram? Page 15
16 Selenia Dimensions 3D FDA approved Allows both 3D and 2D images to be obtained with same compression ~140 µm pixels (3D) 70 µm pixels (2D) Scans 15 in 3.7 sec Obtains 15 projections by pulsing the x-ray tube Continuous tube movement W target x-ray tube 200 ma capable Al filter (3D) Rh or Ag filters (2D) Selenia Dimensions 3D Dose 2D: 12 mgy 3D Tomo: 14.5 mgy Combo: 26.5 mgy Each projection: 1 mgy Low dose detector performance critical (DQE) Grid removed for tomo acquisition Page 16
17 Selenia Dimensions 3D: Reconstruction 1 mm thick tomographic slices Number of slices = compressed breast thickness/1mm ~ 20 to 80 slices per tomographic image Pixels binned into 2x2 size (140 µm) 2-5 sec reconstruction time Breast CT Experimental Challenge to minimize dose Challenge to achieve adequate spatial resolution Challenge to image near the chest wall Impressive preliminary results Page 17
18 Conclusions Digital mammography is currently the gold standard for breast imaging Not a dose saving technique Remains an active area of development Tomographic acquisitions Page 18
Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School
Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Outline Physics aspects of breast tomosynthesis Quality control of breast tomosynthesis
More informationAcceptance Testing of a Digital Breast Tomosynthesis Unit
Acceptance Testing of a Digital Breast Tomosynthesis Unit 2012 AAPM Spring Clinical Meeting Jessica Clements, M.S., DABR Objectives Review of technology and clinical advantages Acceptance Testing Procedures
More informationQuality Control of Full Field Digital Mammography Units
Quality Control of Full Field Digital Mammography Units Melissa C. Martin, M.S., FACMP, FACR, FAAPM Melissa@TherapyPhysics.com 310-612-8127 ACMP Annual Meeting Virginia Beach, VA May 2, 2009 History of
More informationDigital Breast Tomosynthesis
Digital Breast Tomosynthesis OLIVE PEART MS, RT(R) (M) HTTP://WWW.OPEART.COM 2D Mammography Not 100% effective Limited by tissue superimposition Overlapping tissue can mask tumors False negative Overlapping
More informationMammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over
Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over a lifetime Breast cancer screening programs rely on
More informationImaging Technique Optimization of Tungsten Anode FFDM System
Imaging Technique Optimization of Tungsten Anode FFDM System Biao Chen a*, Andrew P. Smith b, Zhenxue Jing a, Elena Ingal a a Hologic, Inc. 600 Technology Drive, DE 1970 b Hologic, Inc. 35 Crosby Drive,
More informationPractical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities
Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities Melissa Martin, M.S., FAAPM, FACR, FACMP AAPM Annual Meeting - Philadelphia July 19, 2010 MO-B-204C-1 Educational Objectives
More informationIntroduction. Digital Mammography QA: Comparing the Manufacturers Recommendations. What is QC and why is it important? Review & compare QC tests
Slide 1 Digital Mammography QA: Comparing the Manufacturers Recommendations Eric A. Berns, Ph.D. Slide 2 Introduction What is QC and why is it important? Review & compare QC tests Key take home points
More informationRadiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893
Digital Radiography D. J. Hall, Ph.D. x20893 djhall@ucsd.edu Background Common Digital Modalities Digital Chest Radiograph - 4096 x 4096 x 12 bit CT - 512 x 512 x 12 bit SPECT - 128 x 128 x 8 bit MRI -
More informationGE Healthcare. Senographe 2000D Full-field digital mammography system
GE Healthcare Senographe 2000D Full-field digital mammography system Digital has arrived. The Senographe 2000D Full-Field Digital Mammography (FFDM) system gives you a unique competitive advantage. That
More informationFeatures and Weaknesses of Phantoms for CR/DR System Testing
Physics testing of image detectors Parameters to test Features and Weaknesses of Phantoms for CR/DR System Testing Spatial resolution Contrast resolution Uniformity/geometric distortion Dose response/signal
More informationPredicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities
Journal of Physics: Conference Series PAPER OPEN ACCESS Predicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities Recent citations - Resolution Properties of a
More informationCollimation Assessment Using GAFCHROMIC XR-M2
Collimation Assessment Using GAFCHROMIC XR-M2 I. Introduction A method of collimation assessment for GE Senographe full-field digital mammography (FFDM) systems is described that uses a self-developing
More informationMammography Solution. AMULET Innovality. The new leader in the AMULET series. Tomosynthesis, 3D mammography and biopsy are all available.
Mammography Solution AMULET Innovality The new leader in the AMULET series. Tomosynthesis, 3D mammography and biopsy are all available. FUJIFILM supports the Pink Ribbon Campaign for early detection of
More informationX-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto
X-ray Imaging PHYS Lecture Carlos Vinhais Departamento de Física Instituto Superior de Engenharia do Porto cav@isep.ipp.pt Overview Projection Radiography Anode Angle Focal Spot Magnification Blurring
More informationOverview. Professor Roentgen was a Physicist!!! The Physics of Radiation Oncology X-ray Imaging
The Physics of Radiation Oncology X-ray Imaging Charles E. Willis, Ph.D. DABR Associate Professor Department of Imaging Physics The University of Texas M.D. Anderson Cancer Center Houston, Texas Overview
More informationCOMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR
This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura New Technology https://www.youtube.com/watch?v=ptkzznazb 7U COMPUTED
More informationSurveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation
Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation AAPM Annual Clinical Meeting Indianapolis, IN August 5, 2013 Learning Objectives Become familiar with the recommendations and
More informationCurrent technology in digital image production (CR/DR and other modalities) Jaroonroj Wongnil 25 Mar 2016
Current technology in digital image production (CR/DR and other modalities) Jaroonroj Wongnil 25 Mar 2016 Current technology in digital image production (CR/DR and other modalities) 2/ Overview Digital
More informationX-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage
X-rays Ouch! 1 X-rays X-rays are produced when electrons are accelerated and collide with a target Bremsstrahlung x-rays Characteristic x-rays X-rays are sometimes characterized by the generating voltage
More information10/3/2012. Study Harder
This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura Study Harder CR detection is inefficient, inferior to film screen
More information10/26/2015. Study Harder
This presentation is a professional collaboration of development time prepared by: Rex Christensen Terri Jurkiewicz and Diane Kawamura Study Harder CR detection is inefficient, inferior to film screen
More informationPerformance and care. all in one
Performance and care all in one INNOVATION IS WHAT DRIVES US THINKING ABOUT THE FUTURE Preventive diagnostics remains an essential weapon in defeating breast cancer. Metaltronica s forward-thinking design
More informationTECHNICAL DATA. GIOTTO IMAGE SDL/W is pre-arranged for Full Field Digital Biopsy examination with the patient in prone position.
Ver. 01/06/07 TECHNICAL DATA GIOTTO IMAGE SDL/W LOW DOSE, FULL FIELD DIGITAL MAMMOGRAPHY UNIT USING AMORPHOUS SELENIUM (a-se) TECHNOLOGY DETECTOR (pre-arranged for stereotactic biopsy with the same digital
More informationIntroduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image
Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology
More informationSTEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS
STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS JAMES A. TOMLINSON, M.S. Diagnostic Radiological Physicist American Board of Radiology Certified Medical Physics Consultants, Inc. Bio 28 yrs experience 100%
More informationX-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah
X-RAY IMAGING EE 472 F2017 Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook Stewart C. Bushong, Radiologic Science for Technologists: Physics, Biology, and Protection, 10 th ed., Mosby,
More informationBeam-Restricting Devices
Beam-Restricting Devices Three factors contribute to an increase in scatter radiation: Increased kvp Increased Field Size Increased Patient or Body Part Size. X-ray Interactions a some interact with the
More informationACPSEM Position Paper RECOMMENDATIONS FOR A DIGITAL MAMMOGRAPHY QUALITY ASSURANCE PROGRAM V4.0
Heggie et al ACPSEM Position Paper: Digital Mammography V4.0 ACPSEM Position Paper RECOMMENDATIONS FOR A DIGITAL MAMMOGRAPHY QUALITY ASSURANCE PROGRAM V4.0 JCP Heggie 1, P Barnes 2, L Cartwright 3, J Diffey
More informationInvestigation of the line-pair pattern method for evaluating mammographic focal spot performance
Investigation of the line-pair pattern method for evaluating mammographic focal spot performance Mitchell M. Goodsitt, a) Heang-Ping Chan, and Bob Liu Department of Radiology, University of Michigan, Ann
More informationX-ray Tube and Generator Basic principles and construction
X-ray Tube and Generator Basic principles and construction Dr Slavik Tabakov - Production of X-rays OBJECTIVES - X-ray tube construction - Anode - types, efficiency - X-ray tube working characteristics
More informationUnit thickness. Unit area. σ = NΔX = ΔI / I 0
Unit thickness I 0 ΔI I σ = ΔI I 0 NΔX = ΔI / I 0 NΔX Unit area Δx Average probability of reaction with atom for the incident photons at unit area with the thickness of Delta-X Atom number at unit area
More information3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development
Teaching Radiographic Technique in a Digital Imaging Paradigm Objectives 1. Discuss the historical development of digital imaging. Dawn Couch Moore, M.M.Sc., RT(R) Assistant Professor and Director Emory
More informationFFDM in the Field: Physicist's Role in the QC of Mammography Laser Printers May Carl R. Keener, Ph.D., DABMP, DABR
FFDM in the Field: Physicist's Role in the QC of Mammography Laser Printers May 2010 Carl R. Keener, Ph.D., DABMP, DABR keener@marpinc.com MARP Medical & Radiation Physics, Inc. Physicist's Role in the
More informationOptimization of Digital Mammography Resolution Using Magnification Technique in Computed Radiography 1
Optimization of Digital Mammography Resolution Using Magnification Technique in Computed Radiography 1 Gham Hur, M.D., Yoon Joon Hwang, M.D., Soon Joo Cha, M.D., Su Young Kim, M.D., Yong Hoon Kim, M.D.
More informationDo you have any other questions? Please call us at (Toll Free) or , or
INSTRUCTIONS Read the appropriate course/ textbook. This is an open book test. A score of 75% or higher is needed to receive CE credit. You will have a maximum of three attempts to pass this course. Please
More informationKODAK DIRECTVIEW CR Mammography Feature User s Guide
KODAK DIRECTVIEW CR Mammography Feature User s Guide 17 September 2010 9G3741 Version 1.0 Carestream Health, Inc. 150 Verona Street Rochester, NY 14608 CARESTREAM, DIRECTVIEW, and DRYVIEW are trademarks
More informationI. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA
FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE
More information- KiloVoltage. Technique 101: Getting Back to Basics
Why do I need to know technique? Technique 101: Getting Back to Basics Presented by: Thomas G. Sandridge, M.S., M.Ed., R.T.(R) Program Director Northwestern Memorial Hospital School of Radiography Chicago,
More informationNew spectral benefi ts, proven low dose
New spectral benefi ts, proven low dose Philips MicroDose mammography SI, technical data sheet Philips MicroDose SI with single-shot spectral imaging is a fullfi eld digital mammography solution that delivers
More informationProtocol for the Quality Control of the Physical and Technical Aspects of Digital Breast Tomosynthesis Systems
Protocol for the Quality Control of the Physical and Technical Aspects of Digital Breast Tomosynthesis Systems Draft version 0.10 February 2013 European Reference Organisation for Quality Assured Breast
More informationImage Quality. HTC Grid High Transmission Cellular Grid provides higher contrast images
B R E A S T I M A G I N G S O L U T I O N S Setting the benchmark for mammography M-IV Series Innovations in breast imaging The Lorad M-IV Series exemplifies Hologic s commitment to developing advanced
More informationRADIOGRAPHIC EXPOSURE
RADIOGRAPHIC EXPOSURE Receptor Exposure Receptor Exposure the that interacts with the receptor. Computed Radiography ( ) requires a. Direct Digital Radiography (DR) requires a. Exposure Indicators Exposure
More informationQuality Control for Stereotactic Breast Biopsy. Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc
Quality Control for Stereotactic Breast Biopsy Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc. 716-924-0350 Methods of Imaging Guided Breast Biopsy Ultrasound guided, hand-held needle
More informationImage Quality. HTC Grid High Transmission Cellular Grid provides higher contrast images
B R E A S T I M A G I N G S O L U T I O N S Setting the benchmark for mammography M-IV Series Innovations in breast imaging The Lorad M-IV Series exemplifies Hologic's commitment to developing advanced
More informationA Comprehensive Review of Image Production
A Comprehensive Review of Image Production Presented by: John Fleming, M.Ed., RT(R)(MR)(CT) St. Petersburg College Office: (727) 341-3758 E-mail: flemingj@spcollege.edu Lesson Objectives: ARRT Content
More informationPublished text: Institute of Cancer Research Repository Please direct all s to:
This is an author produced version of an article that appears in: MEDICAL PHYSICS The internet address for this paper is: https://publications.icr.ac.uk/1316/ Copyright information: http://www.aip.org/pubservs/web_posting_guidelines.html
More informationVisibility of Detail
Visibility of Detail Radiographic Quality Quality radiographic images represents the, and information is for diagnosis. The of the anatomic structures and the accuracy of their ( ) determine the overall
More information4/19/2016. Quality Control Activities for the RadiologicTechnologist. Objectives. 3D Tomosynthesis QC differences
Quality Control Activities for the RadiologicTechnologist Quality Control Tests 2D QC Tomosynthesis QC DICOM Printer Quality Control Weekly Detector Flat Field Calibration Weekl Artifact Evaluation Weekly
More informationDRAFT Technical evaluation of Philips Microdose SI digital mammography system
DRAFT Technical evaluation of Philips Microdose SI digital mammography system NHSBSP Equipment Report 1310 August 2013 About the NHS Cancer Screening Programmes The national office of the NHS Cancer Screening
More informationMILADY. Product Data. Page 1 of 8
Page 1 of 8 The MILADY Mammographic Unit offers the best quality-to-price ratio to our customers worldwide. The unit advanced technology together with the application of industrial production standards,
More informationProtocol for the Quality Control of the Physical and Technical Aspects of Digital Breast Tomosynthesis Systems
Protocol for the Quality Control of the Physical and Technical Aspects of Digital Breast Tomosynthesis Systems Draft version 0.15 January 2014 European Reference Organisation for Quality Assured Breast
More informationInvestigation of Effective DQE (edqe) parameters for a flat panel detector
Investigation of Effective DQE (edqe) parameters for a flat panel detector Poster No.: C-1892 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Exhibit D. Bor 1, S. Cubukcu 1, A. Yalcin 1, O.
More informationExposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?
Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them? Definition and Validation of Exposure Indices Ingrid Reiser, PhD DABR Department of Radiology University of Chicago
More information2 nd generation TOMOSYNTHESIS
2 nd generation TOMOSYNTHESIS 2 nd generation DBT true innovation in breast imaging synthesis graphy Combo mode Stereotactic Biopsy Works in progress: Advanced Technology, simplicity and ergonomics Raffaello
More informationFFDM -FCRm QC Requirements- What You REALLY Need to Know
FFDM -FCRm QC Requirements- What You REALLY Need to Know Melissa C. Martin, M.S., FACR, FAAPM, FACMP AAPM Annual Meeting July 28, 2008 FDA Approval for Mammography Fuji FCRm System approved for Mammography
More informationGE Healthcare. Essential for life. Senographe Essential Full-Field Digital Mammography system
GE Healthcare Essential for life Senographe Essential Full-Field Digital Mammography system Excellence in FFDM is a process. An ongoing quest, fueled by our continuing breakthroughs in breast cancer detection
More informationDigital Imaging Considerations Computed Radiography
Digital Imaging Considerations Digital Radiography Computed Radiography o Cassette based Direct or Indirect Digital Radiography o Cassetteless Computed Radiography 1 CR Image Acquisition Most like conventional
More informationRAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II
RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II APPROVED 12/O2/2011 EFFECTIVE SPRING 2013-14 Prefix & Number RAD 150 Course Title: Radiologic Exposure Technique II & Lab Purpose of this submission: New Change/Updated
More informationDistributed source x-ray tube technology for tomosynthesis imaging
Distributed source x-ray tube technology for tomosynthesis imaging Authors: F. Sprenger a*, X. Calderon-Colon b, Y. Cheng a, K. Englestad a, J. Lu b, J. Maltz c, A. Paidi c, X. Qian b, D. Spronk a, S.
More informationBreast Imaging Basics: Module 10 Digital Mammography
Module 10 Transcript For educational and institutional use. This test bank is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions.
More informationTailoring automatic exposure control toward constant detectability in digital mammography
Tailoring automatic exposure control toward constant detectability in digital mammography Elena Salvagnini a) Department of Imaging and Pathology, Medical Physics and Quality Assessment, KUL, Herestraat
More informationTHE ART OF THE IMAGE: IDENTIFICATION AND REMEDIATION OF IMAGE ARTIFACTS IN MAMMOGRAPHY
THE ART OF THE IMAGE: IDENTIFICATION AND REMEDIATION OF IMAGE ARTIFACTS IN MAMMOGRAPHY William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org INTRODUCTION
More informationPRACTICE GUIDELINE FOR DETERMINANTS OF IMAGE QUALITY IN DIGITAL MAMMOGRAPHY
The American College of Radiology, with more than 30,000 members, is the principal organization of radiologists, radiation oncologists, and clinical medical physicists in the United States. The College
More informationMaximizing clinical outcomes
Maximizing clinical outcomes Digital Tomosynthesis Dual Energy Subtraction Automated Long Length Imaging Improved image quality at a low dose Xray Xray Patented ISS capture technology promotes high sensitivity
More informationMAMMOGRAPHY - HIGH LEVEL TROUBLESHOOTING
MAMMOGRAPHY - HIGH LEVEL TROUBLESHOOTING Maynard High New York Medical College SS2001-M.High 1 Objectives: Review MQSA and ACR annual QC tests as opportunities for troubleshooting before a significant
More informationAbstract. ZEIGLER, GARY BOYCE, II. Direct Detection of Microcalcification Pairs in Simulated
Abstract ZEIGLER, GARY BOYCE, II. Direct Detection of Microcalcification Pairs in Simulated Digital Mammograms. (Under the Direction of Professor Kuruvilla Verghese.) Using the MCMIS (Monte Carlo for Mammography
More informationSYLLABUS. TITLE: Equipment Operation I. DEPARTMENT: Radiologic Technology
CODE: RADT 156 INSTITUTE: Health Science TITLE: Equipment Operation I DEPARTMENT: Radiologic Technology COURSE DESCRIPTION: This course covers the principles of equipment operation and maintenance of radiographic
More informationIntroduction of Computed Radiography in Two Mammography Services: Image Quality and Dose Analysis
Introduction of Computed Radiography in Two Mammography Services: Image Quality and Dose Analysis Rosangela Requi Jakubiak* a, Humberto Remigio Gamba a, Maria Manuela Ramos a, Gislene Gabrielle Faversani
More informationHISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD
CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging Shawn D. Teague, MD DISCLOSURES 3DR- advisory committee CT PHYSICS WITH AN EMPHASIS ON APPLICATION IN THORACIC AND CARDIAC IMAGING
More informationAcquisition, Processing and Display
Acquisition, Processing and Display Terri L. Fauber, R.T. (R)(M) Department of Radiation Sciences School of Allied Health Professions Virginia Commonwealth University Topics Image Characteristics Image
More informationDigital Mammography Quality Control for the Mammographic Physicist
Ontario Breast Screening Program Digital Mammography Quality Control for the Mammographic Physicist Authors: G.E. Mawdsley, A.K. Bloomquist, M.J. Yaffe March 2014 Revision 3.2 Mammographic Physics Consulting
More informationCR Basics and FAQ. Overview. Historical Perspective
Page: 1 of 6 CR Basics and FAQ Overview Computed Radiography is a term used to describe a system that electronically records a radiographic image. Computed Radiographic systems use unique image receptors
More informationK-edge subtraction X-ray imaging with a pixellated spectroscopic detector
K-edge subtraction X-ray imaging with a pixellated spectroscopic detector Silvia Pani Department of Physics, University of Surrey Summary Hyperspectral imaging K-edge subtraction X-ray imaging for mammography
More informationDose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp
Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division
More informationThe importance of radiation quality for optimisation in radiology
Available online at http://www.biij.org/2007/2/e38 doi: 10.2349/biij.3.2.e38 biij Biomedical Imaging and Intervention Journal COMMENTARY The importance of radiation quality for optimisation in radiology
More informationPhase Imaging Using Focused Polycapillary Optics
Phase Imaging Using Focused Polycapillary Optics Sajid Bashir, Sajjad Tahir, Jonathan C. Petruccelli, C.A. MacDonald Dept. of Physics, University at Albany, Albany, New York Abstract Contrast in conventional
More informationX-RAYS - NO UNAUTHORISED ENTRY
Licencing of premises Premises Refer Guidelines A radiation warning sign and warning notice, X-RAYS - NO UNAUTHORISED ENTRY must be displayed at all entrances leading to the rooms where x-ray units are
More informationDIGITAL IMAGE PROCESSING IN X-RAY IMAGING
DIGITAL IMAGE PROCESSING IN X-RAY IMAGING Shalini Kumari 1, Bachan Prasad 2,Aliya Nasim 3 Department of Electronics And Communication Engineering R.V.S College of Engineering & Technology, Jamshedpur,
More informationDosepix Detector as kvp-meter in Radiology and Mammography: First steps
Dosepix Detector as kvp-meter in Radiology and Mammography: First steps F.Bisello, I.Ritter, F.Tennert, A.Zang MediPix Collaboration Meeting, 19th February 2014, CERN Protect, Enhance, and Save Lives -
More informationThe Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System
The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System Manus
More informationChromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC
Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC R. Bellazzini a,b, G. Spandre a*, A. Brez a, M. Minuti a, M. Pinchera a and P. Mozzo b a INFN Pisa
More informationMultiple Choice Identify the letter of the choice that best completes the statement or answers the question.
RA110 test 3 Multiple Choice Identify the letter of the choice that best completes the statement or answers the question. 1. An object 35 cm in width is radiographed at 100 cm SID and at a 50 cm SOD. What
More informationLECTURE 1 The Radiographic Image
LECTURE 1 The Radiographic Image Prepared by:- KAMARUL AMIN ABDULLAH @ ABU BAKAR UiTM Faculty of Health Sciences Medical Imaging Department 11/23/2011 KAMARUL AMIN (C) 1 Lesson Objectives At the end of
More informationADVANCED MEDICAL SYSTEMS PTE LTD Singapore Malaysia India Australia
Innovative design is combined with cutting-edge technology to yield a definitive diagnosis and never before seen ergonomics GIOTTO CLASS is the result of 25 years of experience in the research and development
More informationComparison of computed radiography and filmõscreen combination using a contrast-detail phantom
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 1, WINTER 2003 Comparison of computed radiography and filmõscreen combination using a contrast-detail phantom Z. F. Lu,* E. L. Nickoloff, J.
More information1. Carlton, Richard R., and Arlene M. Adler. Principles of Radiographic Imaging: An Art and a Science, 5th edition (2013).
CODE: RADT 151 INSTITUTE: Health Science TITLE: Radiographic Exposure DEPARTMENT: Radiologic Technology COURSE DESCRIPTION: This course covers the principles of radiographic exposure selection and manipulation
More informationDigital radiography (DR) post processing techniques for pediatric radiology
Digital radiography (DR) post processing techniques for pediatric radiology St Jude Children s Research Hospital Samuel Brady, MS PhD DABR samuel.brady@stjude.org Purpose Review common issues and solutions
More informationMammograph FFDM PRODUCT DATA
PRODUCT DATA Product data Rev.1 (May 2010) Page Description and Configurations 3 Technical Features 7 Mammography Unit 8 Acquisition Unit 12 Diagnostic Unit (Optional) 12 Size and Dimensions 13 Classifications
More informationDisclosures. Outline 7/31/2017. Current Implementation Status of IEC Standard : Exposure Index (EI) for Digital Radiography
Current Implementation Status of IEC Standard 62494-1: Exposure Index (EI) for Digital Radiography July 31, 2017 Ryan Fisher, PhD, DABR Katie Hulme, MS, DABR None Disclosures Outline Review of IEC Standard
More informationDigital radiography: Practical advantages of Digital Radiography. Practical Advantages in image quality
Digital radiography: Digital radiography is set to become the most common form of processing radiographic images in the next 10 years. This is due to a number of practical and image quality issues. Practical
More informationAmorphous Selenium Direct Radiography for Industrial Imaging
DGZfP Proceedings BB 67-CD Paper 22 Computerized Tomography for Industrial Applications and Image Processing in Radiology March 15-17, 1999, Berlin, Germany Amorphous Selenium Direct Radiography for Industrial
More informationSetting up digital imaging department!
Outline Setting up digital imaging department! From screen/film to digital radiography PACS/Tele radiology Setting up digital department Digital Imaging Napapong Pongnapang, Ph.D. Department of Radiological
More informationDigital mammography: DQE versus optimized image quality in clinical environment an on site study
Digital mammography: DQE versus optimized image quality in clinical environment an on site study Nadia Oberhofer* a, Alessandro Fracchetti a, Margareth Springeth a, Ehrenfried Moroder a a Health Service
More informationPD233: Design of Biomedical Devices and Systems
PD233: Design of Biomedical Devices and Systems (Lecture-8 Medical Imaging Systems) (Imaging Systems Basics, X-ray and CT) Dr. Manish Arora CPDM, IISc Course Website: http://cpdm.iisc.ac.in/utsaah/courses/
More informationExposure System Selection
Principles of Imaging Science II (RAD120) Exposure Systems Exposure System Selection Radiographic exposure is a very complex process Best technique systems manipulate one variable while holding others
More informationSeminar 8. Radiology S8 1
Seminar 8 Radiology Medical imaging. X-ray image formation. Energizing and controlling the X-ray tube. Image detectors. The acquisition of analog and digital images. Digital image processing. Selected
More informationACR AAPM SIIM Practice Guideline for Determinants of Image Quality in Digital Mammography
J Digit Imaging (2013) 26:10 25 DOI 10.1007/s10278-012-9521-3 EDITORIAL ACR AAPM SIIM Practice Guideline for Determinants of Image Quality in Digital Mammography Kalpana M. Kanal & Elizabeth Krupinski
More informationAN ABSTRACT OF THE THESIS OF. W. Scott Helms for the degree of Master of Science in Radiation Health Physics
AN ABSTRACT OF THE THESIS OF W. Scott Helms for the degree of Master of Science in Radiation Health Physics presented on November 24, 2014 Title: A Quantitative Comparison of Cardiovascular Imaging Systems
More informationof sufficient quality and quantity
of sufficient quality and quantity The patient s body attenuates the beam as it passes though the body More energy is deposited in organs located near the entry of the beam than near the exit of the beam
More information